You are on page 1of 23

chapter

14
The Gamma Camera:
Performance
Characteristics

The performance of a gamma camera system blurring caused by the collimator, is called the
is defined by the sharpness and detail of the intrinsic spatial resolution of the camera.
images it produces, the efficiency with which Intrinsic resolution is limited primarily by
it detects incident radiation, its ability to two factors. The first is multiple scattering of
measure the energy of the incident rays (to -ray photons within the detector. If a photon
minimize scatter), and the counting rate it can undergoes Compton scattering within the
handle without significant dead time losses. A detector crystal and the residual scattered
gamma camera is not capable of producing photon also is detected, but at some distance
perfect images of the radionuclide distribu- away, the two events are recorded as a single
tion. Certain inherent imperfections arise event occurring at a location along the line
from the performance characteristics of the joining the two interaction sites. This is not
detector, its associated electronic circuitry, a serious cause of degraded resolution for
and the collimator. Image artifacts also can be photon energies 300keV in which multiple
caused by malfunctions of various camera scatter Compton interactions in NaI(Tl) are
components. In this chapter, we describe the almost negligible. Even at 662keV, Anger
major factors that determine gamma camera calculated that for a detector thickness of
performance and examine the limitations that 6.4mm, less than 10% of photons are mis-
can lead to artifacts in gamma camera images placed by more than 2.5mm as a result of
and their correction. Standard tests of gamma multiple scattering events.1
camera performance also are summarized. The second, and primary, cause of limited
intrinsic resolution is statistical fluctuation
in the distribution of light photons among
A. BASIC PERFORMANCE photomultiplier (PM) tubes from one scintil-
lation event to the next. The problem is
CHARACTERISTICS
exactly analogous to the statistical fluctua-
tions observed in radioactive decay, discussed
1. Intrinsic Spatial Resolution in Chapter 9. If a certain PM tube records, on
Spatial resolution is a measure of the sharp- average, N light photons from scintillation
ness and detail of a gamma camera image. events occurring at a certain location in the
Sharp edges or small, pointed objects produce detector crystal, the actual number recorded
blurred rather than sharply defined images. from one event to the next varies with a stan-
Part of the blurring arises from collimator dard deviation given by N . Thus if a very
characteristics discussed in Sections C and D narrow beam of rays is directed at a point
and part arises in the sodium iodide [NaI(Tl)] on the detector, the position of each event as
detector and positioning electronics. The limit determined by the positioning circuitry or
of spatial resolution achievable by the detec- computer algorithm is not precisely the same.
tor and the electronics, ignoring additional Rather, they are distributed over a certain
209
210 Physics in Nuclear Medicine

area, the size of which depends on the mag- their distribution are therefore proportional
nitude of these statistical fluctuations. to 1/ N . This causes noticeably greater blur-
A detailed method for measuring and char- ring at lower -ray energies. An example of
acterizing intrinsic spatial resolution is dis- the change of intrinsic spatial resolution as
cussed in Section E.1. Typically, a lead mask a function of -ray energy is shown in Figure
containing a number of narrow (~1mm) slits 14-1.
is placed on the face of the gamma camera Intrinsic resolution also depends on detec-
(without the collimator) and the camera is tor crystal thickness. Thicker detectors result
irradiated using a 99mTc (140-keV) point in greater spreading of scintillation light
source. The resulting image is a series of lines before it reaches the PM tubes. Furthermore,
corresponding to the locations of the slits there is a greater likelihood of detecting
(e.g., see Fig. 14-10A). The resolution is cal- multiple Compton-scattered events in thicker
culated as the full width at half maximum detectors, particularly with higher-energy
(FWHM) of a profile drawn perpendicular to radionuclides. These are the primary reasons
the image of the lines at various locations in why gamma cameras use relatively thin
the field of view. The intrinsic spatial resolu- detectors in comparison with NaI(Tl) systems
tion of modern large field-of-view gamma that are used for counting applications.
cameras measured with 99mTc in this manner Figure 14-2 shows an example of the intrin-
is in the range of 2.9- to 4.5-mm FWHM. sic spatial resolution versus crystal thickness
Because the resolution is considerably worse for 140-keV rays.
than the width of the slits, the contribution of Intrinsic resolution improves with increased
the slits themselves to the measured resolu- efficiency of collection of scintillation photons.
tion is very small (10% for measured resolu- Modern cameras are substantially improved
tion 2.5mm). over earlier versions in this regard because of
Intrinsic resolution becomes worse with the use of more efficient PM tubes and of
decreasing -ray energy because lower-energy better techniques for optical coupling between
rays produce fewer light photons per scin the detector crystal and the PM tubes. The
tillation event, and smaller numbers of light use of greater numbers of smaller PM tubes
photons result in larger relative statistical (5-cm-diameter tubes have become the stan-
fluctuations in their distribution (Chapter 9, dard, and some gamma cameras have as
Section B.1). As a rule of thumb, intrinsic many as 110 PM tubes per head) and improved
resolution is proportional to 1/ E, in which electronics also have contributed to this
E is the -ray energy. This follows because improvement. Accurate corrections for non-
the number of scintillation light photons linearity (see Section B.1) and nonuniformity
produced, N, is roughly proportional to E (see Section B.2) have also resulted directly
and the relative statistical fluctuations in in improvements in intrinsic resolution, as

4.5
Intrinsic resolution (mm FWHM)

4
FIGURE 14-1 Intrinsic spatial resolution
of a gamma camera as a function of -ray
energy for a 6.3-mm-thick NaI(Tl) crystal.
(Compiled with data from Sano RM,
3.5 Tinkel JB, LaVallee CA, Freedman GS:
Consequences of crystal thickness reduc-
tion on gamma camera resolution and sen-
sitivity. J Nucl Med 19:712-713, 1978; and
Muehllehner G: Effect of crystal thickness
3 on scintillation camera performance. J
Nucl Med 20:992-993, 1979.)

2.5
0 50 100 150 200 250 300 350 400
-ray energy (keV)
14 The Gamma Camera: Performance Characteristics 211

5.5
FIGURE 14-2 Intrinsic spatial resolution
of a gamma camera at 140keV as a function
of crystal thickness. (Compiled with data 5

Intrinsic resolution (mm, FWHM)


from Sano RM, Tinkel JB, LaVallee CA,
Freedman GS: Consequences of crystal
thickness reduction on gamma camera reso- 4.5
lution and sensitivity. J Nucl Med 19:712-
713, 1978; Muehllehner G: Effect of crystal
thickness on scintillation camera perfor-
mance. J Nucl Med 20:992-993, 1979; Royal 4
HD, Brown PH, Claunch BC: Effects of
reduction in crystal thickness on Anger
camera performance. J Nucl Med 20:977- 3.5
980, 1979; Chapman D, Newcomer K,
Berman D, etal: Half-inch versus quarter-
inch Anger camera technology: Resolution
3
and sensitivity differences at low photo-
peak energies. J Nucl Med 20:610-611,
1979; and unpublished data from Dr. Joel
Karp, University of Pennsylvania, Philadel- 2.5
phia, PA.) 0 5 10 15 20 25 30
Crystal thickness (mm)

discussed in the following sections. The best 100% efficient for energies up to approxi-
reported intrinsic resolution for a large field- mately 100keV for all crystal thicknesses,
of-view gamma camera is just below 3mm but then shows a rather marked decrease in
FWHM at 140keV ( 99mTc). Significant efficiency at higher energies, depending on
improvements beyond approximately 2mm crystal thickness. At 140keV (-ray energy of
99m
FWHM will be difficult to achieve, owing to Tc), the difference in efficiency between
the ultimate limitation of the light photon 6.4-mm and 12.7-mm-thick crystals is approx-
yield of NaI(Tl). In most practical situations, imately 20% and the photopeak detection
however, the intrinsic spatial resolution efficiency is in the 70% to 90% range. At
makes a negligible contribution to the overall approximately 500keV, the standard gamma
system resolution of the gamma camera, camera (detectors 0.64-0.95-cm-thick) is less
which is largely determined by the resolution than 20% efficient at converting incident
of the collimator (see Sections C and D). rays into photopeak pulses.
At high energies, the performance of gamma
2. Detection Efficiency cameras with 0.64- to 1.27-cm-thick crystals
The gamma camera employs a sodium iodide is limited by decreasing detection efficiency
crystal that is relatively thin in comparison (as well as increasing collimator septal
with most other sodium iodide detectors used penetrationsee Section C.2). Deteriorating
in nuclear medicine: 6.4 to 12.7mm versus 2 intrinsic spatial resolution becomes the limit-
to 5cm for probe counting systems, scanners, ing factor at lower energies. Because of these
and so on. The trade-off in gamma cameras is tradeoffs, the optimal -ray energy range is
between detection efficiency (which improves approximately 100 to 200keV for most gamma
with thicker crystals) and intrinsic spatial cameras. Some gamma cameras are now fitted
resolution (which improves with thinner with thicker crystals (12.7-25.4mm), enabling
crystalssee Fig. 14-2). The gamma camera is them to achieve improved efficiency for
designed to provide acceptable detection effi- imaging positron-emitting radionuclides at
ciency while maintaining high intrinsic spatial 511keV (Chapter 18, Section B.4). This comes
resolution in the energy range of 100-200keV. at the expense of some loss of intrinsic spatial
As a result, the detection efficiency of the resolution (see Fig. 14-2) when these systems
gamma camera detector is somewhat less than are used in the 100-200-keV energy range.
would be desirable at higher -ray energies.
Figure 14-3 shows photopeak detection 3. Energy Resolution
efficiency versus -ray energy for the gamma It is not unusual in a typical patient study for
camera detector for a range of NaI(Tl) crystal there to be more Compton-scattered than
thicknesses. The gamma camera is nearly unscattered rays striking the detector (see
212 Physics in Nuclear Medicine

NaI(Tl) thickness (cm)


100
Photopeak detection efficiency (%)

5.08

2.54

10 1.27

0.64

1
0 100 200 300 400 500 600 700
-ray energy (keV)
FIGURE 14-3 Photopeak detection efficiency versus -ray energy for NaI(Tl) detectors of different thicknesses.
(Adapted from Anger HO: Radioisotope cameras. In Hine GJ [ed]: Instrumentation in Nuclear Medicine, Vol 1. New
York, 1967, Academic Press, p 506.)

Fig. 11-9). Because the Compton-scattered 500


photons have lower energy, it is possible to 400
Counts

discriminate against them using pulse-height 300 PHA


window
analysis. The energy resolution of the detec- 200
tor determines the efficiency with which this 100
can be accomplished. Good energy resolution
is perhaps the most important performance 0 50 100 150
feature of the camera system for this purpose. Energy (keV)
Energy resolution, like intrinsic spatial FIGURE 14-4 Energy spectrum from a gamma camera
resolution, depends largely on statistical fluc- measured using a point source of 99mTc in air. The energy
tuations in the number of light photons col- resolution at 140keV in this example is 10.5%. A typical
15% energy window (approximately 130 to 150keV) is
lected from a scintillation event (Chapter 10, shown superimposed on the spectrum. PHA, pulse-height
Section B.7). Thus good light collection analyzer. (Data courtesy Dr. Magnus Dahlbom, UCLA
efficiency is a prerequisite for good energy School of Medicine, Los Angeles, CA.)
resolution. As well, because the number
of light photons released in a scintillation
event increases almost linearly with -ray
energy, E, (Fig. 10-11), energy resolution 130keV should reject 140-keV rays that
improves approximately in proportion to have been scattered through angles greater
1/ E (Fig. 10-13). than approximately 45 degrees. However,
The energy resolution for gamma cameras because the spectrum for scattered rays is
is typically in the 9% to 11% range for 99mTc. blurred in the same way as the spectrum for
Figure 14-4 shows a typical gamma camera unscattered ones, the rejection efficiency for
spectrum for 99mTc with the pulse-height ana- this scattering angle is only approximately
lyzer (PHA) window set to 130 to 150keV. 50%; half of the events produce pulses above
This corresponds to approximately a 15% the threshold, and half below it. This percent-
energy window, which is a common setting for age would apply for 45-degree scattered
clinical studies. As illustrated by the figure, 140-keV rays and a 130-keV lower energy
most of the events in the photopeak are level, regardless of the energy resolution of
accepted within this window. According to the detector. Gamma rays scattered through
Equation 6-11, a low-energy threshold of greater angles are rejected more efficiently,
14 The Gamma Camera: Performance Characteristics 213

and those scattered through smaller angles cardiac studies. Dead time corrections can be
are rejected less efficiently. applied; however, these corrections generally
Two advantages are obtained with improved become increasingly inaccurate as counting
energy resolution. First, the photopeak losses increase.
becomes narrower, resulting in more efficient Because pulse pile-up can occur between
detection of unscattered photons within the any two events in the pulse-height spectrum,
chosen energy window. This increases the system counting losses are determined by
number of valid events recorded and improves total-spectrum counting rates. Most gamma
the statistical quality of the image. Second, cameras behave as paralyzable systems. The
rays scattered through large angles are apparent dead time for a selected energy
rejected more efficiently, because their energy window depends on the window fraction, that
spread within the pulse-height spectrum is is, the fraction of the total spectrum counting
also smaller. Thus image contrast is improved. rate occurring within that window. The smaller
It also is true that rays scattered through the window fraction, the larger the apparent
smaller angles are detected somewhat more dead time. Thus the apparent dead time is
efficiently, because of the narrowing of their longer when a photopeak window is used than
distribution as well. However, the increased when a full-spectrum window is used. The
efficiency for recording photopeak events apparent dead time also is longer when
more than offsets this effect, in terms of scattered radiation is present, because this
contrast-to-noise ratio (Chapter 15, Section also adds to the counting rate outside the
D.2). Alternatively, one can take advantage of photopeak window (Fig. 14-5). Therefore,
the improved energy resolution to use a nar- when specifying gamma camera dead time, it
rower PHA window, trading back some of the is important to note the conditions of mea
increased efficiency for recording photopeak surement. Dead time values as short as 1 to 2
events for improved rejection of small-angle sec can be obtained in the absence of scatter-
scatter. Either way, improved energy resolu- ing material with a full-spectrum window;
tion results in better image quality. however, under clinically realistic conditions
(99mTc source in scattering material, 15% pho-
4. Performance at High topeak window), system dead times of 4 to 8
Counting Rates sec are more typical. For a dead time of 5 sec,
At high counting rates, there is increased like- counting losses are approximately 20% for a
lihood of recording two events at the same counting rate of 4 104 counts per second (cps).
time. The most troublesome effect is known as Dead time losses are not serious in most
pulse pile-up (Chapter 8, Section B.3). Pulse static imaging studies, but they can be impor-
pile-up has two undesirable effects on gamma tant in certain high-counting-rate applica-
camera performance: counting losses and tions (e.g., first-pass cardiac studies) in which
image distortion. counting rates as high as 105 cps may occur.
Counting losses cause inaccurate counting Pile-up rejection circuitry (see Chapter 8,
rates to be recorded at higher counting rates. Section B.3) is used to achieve higher usable
The inaccuracies are described by conven- counting rates in such situations. Another
tional dead time models (Chapter 11, Section approach for shortening camera dead time is
C) and may be significant in some high-count- by the use of analog buffers, or derandomiz-
rate quantitative studies, such as first-pass ers. These are electronic circuits that hold a
Observed counting rate

Source Source
Source behind in air No losses in air
Counts

10 cm water

Source behind 10 cm water

Energy True counting rate


FIGURE 14-5 Effect of scattered radiation on counting losses. Scattered radiation decreases the window fraction
recorded with a photopeak window (left), thus causing an apparent increase in dead time counting losses (right).
214 Physics in Nuclear Medicine

voltage level or pulse from one circuit compo- (see Chapter 13, Section B.2), two events
nent (e.g., an amplifier) until the next circuit detected simultaneously at different locations
in the pulse-processing sequence (e.g., the in the detector are recorded as a single event
PHA) is ready to receive it. with energy equal to the sum of the two
Similarly, in digital gamma cameras, data events, at a location somewhere between
can be buffered in memory until the computer them (Fig. 14-6). If both are valid photopeak
is ready to process them. Both these approaches events, their total energy exceeds the value
result in a decrease in the apparent dead that would be accepted by the PHA window
time of the camera by effectively changing the and both events are rejected, resulting in
arrival times of the pulses. This, however, counting losses. On the other hand, it is pos-
means that the simple dead time models and sible for two Compton-scattered rays to
corrections presented in Chapter 11, Section have a total energy that falls within the
C can no longer be used, and more complex selected energy window, so that two invalid
modeling of system dead time must be carried events are accepted as a single valid event.
out to produce accurate correction at high The visible result at very high counting rates
counting rates. is to add a diffuse background to the image,
It also is possible to physically shorten the as illustrated in Figure 14-7. Note as well the
dead time of a camera by shortening the image in the upper right-hand corner of this
charge integration time from the PM tubes figure, showing how contrast can be restored
and using electronic circuitry that returns the by shielding high-activity areas outside the
signal to baseline after the chosen integration imaging area of interest (e.g., with a thin
time.2 Clearly, this also decreases the amount sheet of lead).
of signal used for determining event location. Early pile-up rejection methods were based
For example, with a charge integration time on measuring the length of a pulse. If the
of 0.4 sec, only 81% of the scintillation light pulse did not return close to baseline level
is collected, compared with 98% for a 1-sec within the time expected given the decay time
integration time. This causes a degradation of of NaI(Tl), it was assumed that pile-up of
intrinsic spatial resolution and energy resolu- two pulses had occurred and the event was
tion. Some gamma cameras have a variable rejected, resulting in the loss of both rays.
integration time, in which the charge integra- This improved image quality but resulted in
tion is automatically shortened as the count- an effective increase in system dead time,
ing rate increases. because many events were rejected at high
Other means for shortening dead time are counting rates.
to bypass altogether the pile-up rejection cir- Many gamma cameras now incorporate cir-
cuits and nonuniformity correction circuitry cuits that continuously monitor the decay of
(see discussion on pile-up correction later in a pulse and use a method based on pulse-tail
this section and on nonuniformity and its cor- extrapolation for pile-up correction. Consider
rection in Section B.3). The signal processing two -ray interactions that occur close together
that occurs in these circuits slows down the in time and create overlapping pulses. When
rate at which the camera can handle indi- the second ray arrives, the decay of the
vidual events, and bypassing them can pulse created by the first ray immediately
shorten system dead time from typical values
of 4 to 8 sec down to 1 to 3 sec. Some
cameras provide an optional high count rate
mode of operation in which some or all of
these corrections are turned off by software Source
control. This mode is intended specifically for Pile-up
applications requiring high counting rates, Source
such as first-pass cardiac studies. Normal
mode, in which all corrections are employed,
is used for routine imaging to obtain the
desired high-quality images. Obviously speci-
fications for gamma camera dead time should
indicate whether any circuits were bypassed
to achieve the reported value.
FIGURE 14-6 Images of two 99mTc point sources of rela-
The second undesirable effect of pulse tively high activities (~370 MBq each). Events appearing
pile-up is image distortion. Using standard in the band between the two point-source locations are
pulse-positioning logic for gamma cameras mispositioned events caused by pulse pile-up.
14 The Gamma Camera: Performance Characteristics 215

105

Observed counting rate (cps)


FIGURE 14-7 Demonstration of pile-up
effects on images of a brain phantom. 0.4 min
Times required to record 1.2 106 counts 0.4 min
104
are indicated. At very high counting 0.6 min
rates there is a noticeable loss of image
contrast, which can be restored by 0.9 min
shielding useless high-activity areas
from the detector (top right-hand image).
1.8 min

3.2 min
103
1.0 10 100
Activity in phantom (mCi)

deviates from the expected exponential decay of NaI(Tl). At the same time, this extrapo-
and the gamma camera signal is switched to lated tail is also sent to the second amplifier
a second amplifier circuit. Estimator circuitry circuit and subtracted from the second pulse.
in the first amplifier circuit completes the This removes the contribution of the pulse
signal from the first ray by extrapolating the generated by the first ray from that of the
remainder of the tail of the pulse with an second ray. This process is summarized
exponential function based on the decay time in Figure 14-8. The pulse-tail extrapolation

Without pulse-tail extrapolation:

P2 P1 + P2
P1

Amplifier

With pulse-tail extrapolation:

FIGURE 14-8 Illustration of pile-up correc- P2 P1 Tail


P1
tion using pulse-tail extrapolation techniques. P1
See text for details. (Adapted from Lewellen
TK, Pollard KR, Bice AN, Zhu JB: A new clini-
cal scintillation camera with pulse tail extra
polation electronics. IEEE Trans Nucl Sci
37:702-706, 1990.) Amplifier 1

Estimator
P2

Subtractor

Amplifier 2

P2 + P1 Tail
216 Physics in Nuclear Medicine

technique results in both events being displacement distance of a radiation source


retained and allows them to contribute to the across the face of the detector. For example,
image, providing they also meet the PHA when a source is moved from the edge of one
requirements. This method is very effective, of the PM tubes toward its center, the light
unless the two pulses occur nearly simultane- collection efficiency of that PM tube increases
ously (within a few tens of nanoseconds of more rapidly than the distance the source is
each other), in which case the extrapolation moved. This causes the image of a line source
is of limited accuracy. crossing in front of a PM tube to be bowed
With modern digital gamma cameras, it toward its center. The result is a characteristic
also is possible to use the spatial distribution pincushion distortion in areas of a gamma
of PM tube signals to further reduce pile-up. camera image lying directly in front of the PM
For pile-up events occurring at different loca- tubes, and barrel distortion between them.
tions in the detector crystal, two distinct clus- Differences in sensitivity among the PM tubes,
ters of PM tubes will produce signals. If the nonuniformities in optical light guides, as well
light distributions produced by the two events as PM tube or electronic malfunctions, also
on the PM tubes do not overlap, or only can cause nonlinearities.
slightly overlap, the events can be clearly Figure 14-10A, shows an image of a
separated and retained. straight-line test pattern recorded on a

B. DETECTOR LIMITATIONS:
NONUNIFORMITY AND
NONLINEARITY

1. Image Nonlinearity
A basic problem arising in the detector and
electronics is image nonlinearity. Straight-line
objects appear as curved-line images. An
inward bowing of line images is called pin-
cushion distortion; an outward bowing is Pincushion Barrel
called barrel distortion (Fig. 14-9). Non distortion distortion
linearities result when the X- and Y-position FIGURE 14-9 Appearance of straight-line objects with
signals do not change linearly with pincushion and barrel distortions.

A B
FIGURE 14-10 A, Illustration of nonlinearities in images of a straight-line test pattern obtained with a gamma camera.
Image demonstrates subtle waviness in the lines. B, Flood-field image obtained by exposing the same camera to a
uniform radiation field. This is the image obtained in the absence of any corrections for nonuniformity. Notice that the
photomultiplier tube pattern can be seen. The gray levels in this image are confined to a narrow display window to
improve visualization of the artifacts. (Images courtesy Dr. Magnus Dahlbom, UCLA School of Medicine, Los Angeles,
CA.)
14 The Gamma Camera: Performance Characteristics 217

modern gamma camera to demonstrate the in areas of barrel distortion events are pushed
general appearance of nonlinearities. On outward from the center, causing an apparent
close inspection, some waviness of the lines is cold spot. Because of the characteristic pin-
apparent. On properly functioning cameras, cushion distortions occurring in front of PM
including the one illustrated, the nonlineari- tubes, it is common to see a pattern of hot
ties themselves (including the pincushion spots at the locations of the PM tubes on an
distortions in front of PM tubes) are barely otherwise uniform gamma camera image.
perceptible and rarely interfere directly with Other causes of nonlinearities (e.g., PM tube
image interpretation; however, they can have failure, crystal cracking, and collimator
significant effects on image nonuniformities, defects) also can result in nonuniformities.
as discussed in the following section. Another characteristic nonuniformity is a
bright ring around the edge of the image. This
2. Image Nonuniformity artifact, called edge packing, results from a
A more noticeable problem is image non somewhat greater light collection efficiency
uniformity. Exposing the detector crystal to a for events near the edge versus central regions
uniform flux of radiation produces a flood- of the detector crystal. This is the result of
field image with small but noticeable nonuni- internal reflections of scintillation light from
formities in intensity, even with a properly the sides of the detector crystal back into the
functioning camera. These variations may be PM tubes near the edge. Also, for events
equivalent to counting rate variations of 10% occurring toward the center of the crystal,
or more. A flood-field image from a gamma there are always PM tubes on either side of
camera demonstrating image nonuniformity the event location, whereas at the edges of the
is shown in Figure 14-10B. Intrinsic flood- crystal there are PM tubes only to one side.
field images are acquired with the collimator Thus events at the very edges are not distrib-
removed, using a point source placed far uted uniformly across the edge, but are
enough from the surface of the gamma camera pulled toward the center, compounding the
to give uniform irradiation of the surface (dis- edge-packing artifact. The portion of the
tance equal to 4-5 times the camera diame- image demonstrating this artifact usually is
ter). Extrinsic flood-field images are acquired masked on the image display and therefore is
with the collimator in place using a disk or not a part of the useful field of view (UFOV ).
thin flood phantom that covers the area of the Typically, 5cm or more of the detector width
detector. 99mTc or 57Co are the two most is eliminated by the mask. When specifying
commonly used radionuclides for flood-field gamma camera detector dimensions, it is
measurements. important to distinguish between the physi-
There are two primary causes of gamma cal dimensions of the crystal and the dimen-
camera nonuniformities. The first is non- sions of the useful imaging area.
uniform detection efficiency arising from (1) Both nonuniformity and edge-packing
small differences in the pulse-height spec- artifacts are related to the pattern of the
trum for different PM tubes and (2) position- distribution of scintillation light falling on
dependent collection efficiency of scintillation the PM tubes. For this reason, they also have
light, particularly for events located over the an energy-dependent component. When the
gaps and dead areas between the PM tubes gamma camera is used to image higher-
compared with events located directly over energy radionuclides, interactions, on average,
the center of a PM tube. The differences in occur deeper in the crystal, closer to the PM
PM tube response can be minimized by careful tubes. This produces a more narrow light
selection and tuning of all of the PM tubes of spread distribution on the PM tubes and gen-
a gamma camera; however, position-dependent erally results in a worsening of detector
effects on the pulse-height spectrum remain. nonuniformity.
If a fixed pulse-height window is used for all
output pulses, the result is an apparent dif- 3. Nonuniformity Correction
ference in detection efficiency owing to differ- Techniques
ences in the window fraction for different All modern gamma cameras incorporate tech-
areas of the crystal (see Fig. 13-6, top). niques that attempt to correct the causes of
The second cause of nonuniformities is nonuniformity described in the preceding
image nonlinearities described in Section section. All of these techniques begin with
B.1. In areas of pincushion distortion events spatially varying energy corrections, normally
are crowded toward the center of the distor- derived from an intrinsic flood-field image.
tion, causing an apparent hot spot, whereas The flood-field image is divided into a matrix
218 Physics in Nuclear Medicine

of small, square elements, typically 128 128 counts recorded in each pixel are thrown out,
elements (or pixels). Using the PHA, the depending on the relative value for that pixel
channel number (pulse amplitude) of the photo in the energy-corrected flood-field image.
peak in the pulse-height spectrum is deter- For example, if the value in the normalized
mined for each element. This information is intensity-correction matrix is 110, then 1 of
stored in a 128 128 look-up table and used every 11 counts is subtracted from the patient
to set regionally varying PHA windows for image at that location. This process is some-
subsequent studies on patients. For example, times called count skimming.
if a 20% window is chosen for a patient study, Most modern digital gamma cameras
and the center of the photopeak is found in replace the second step described in the pre-
PHA channel 100 in a particular pixel in ceding paragraph with a correction for image
the flood-field image, then events at that nonlinearity, which more directly attacks the
location having Z-signal amplitudes between major underlying cause of image nonunifor-
PHA channels 90 and 110 are accepted in mity. For nonlinearity corrections, another
patient imaging studies. If the center of the flood-field image is obtained, this time with a
photopeak is in channel 110 at another loca- sheet of lead having a uniformly spaced array
tion, events for which the Z-signal falls within of small holes (~1-mm diameter, ~4-mm sepa-
the range of 99 to 121 are accepted at that ration) placed directly on the gamma camera
location. The position-dependent PHA window face (no collimator). The locations of the images
corrects for variations in the pulse-height of these holes are compared with their known
spectrum across the face of the camera detec- locations in the lead sheet to derive a matrix
tor. It also provides a partial correction for of offsets, x and y, for each (X,Y ) location on
image nonuniformity. the detectors, which is stored as another
The second step in the nonuniformity cor- look-up table. When an event is detected, its X
rection is to account for the remaining regional and Y coordinates are computed using conven-
variations in image intensity, largely caused tional positional circuitry or algorithms. These
by detector nonlinearity. In one older method, values then are corrected using the positional
the correction is based directly on variations offsets for that location stored in the look-up
in intensity of the energy-corrected flood-field table. The offsets and the corresponding
image. The number of counts recorded within look-up table usually are measured and gener-
each pixel in that image is stored in a matrix ated at the factory prior to shipment.
and compared to the smallest number Figure 14-11 shows the same data as Figure
recorded in the pixel array. This is used to 14-10, after the corrections for nonuniformity
derive a matrix of normalized intensity values, and nonlinearity described in the preceding
which range from 100 for the coldest pixel paragraphs have been applied. Figure 14-12,
to higher values for other pixels. In subse- showing intensity profiles across the flood-
quent patient studies, a certain fraction of the field images in Figures 14-10 and 14-11,

A B
FIGURE 14-11 Straight-line test pattern (A) and uniform flood-field (B) images after nonuniformity corrections are
applied. Compare with Figure 14-10. (Images courtesy Dr. Magnus Dahlbom, UCLA School of Medicine, Los Angeles,
CA.)
14 The Gamma Camera: Performance Characteristics 219

1.2 1.2

Normalized pixel intensity 1 1

Normalized pixel intensity


0.8 0.8

0.6 0.6

0.4 0.4

0.2 0.2

0 0
0 50 100 150 200 250 0 50 100 150 200 250
Pixel Pixel
No correction Corrected
FIGURE 14-12 Profiles through the uniform flood-field images in Figures 14-10 and 14-11 showing relative uniformity
of flood-field image with and without nonuniformity correction. The standard deviation is improved from 3.4% to 1.9%
after correction.

clearly illustrates the improvements. The the tube output has changed by more than 1%
examples in these figures are for 99mTc. Note from the original reference value.
that different correction matrices must be In newer gamma cameras, the large
obtained for each radionuclide used, because number of PM tubes makes the manual
the effects corrected for generally vary with method impractical. Many digital gamma
-ray energy, for example, because of different cameras therefore contain tuning circuitry
average depths of interaction in the NaI(Tl) that allows the output of each individual PM
crystal. tube to be automatically adjusted to a set of
Improvements in camera uniformity also reference outputs. One automated approach
have contributed to improvements in intrinsic involves the use of light-emitting diodes
resolution. Earlier cameras used thicker light (LEDs) that are coupled to the neck of each
guides and large-diameter PM tubes, in part PM tube. These LEDs are pulsed to produce
to achieve satisfactory uniformity, at the a light signal on the photocathode of the PM
expense of somewhat degraded spatial resolu- tube that does not vary with time. The PM
tion. Because of effective uniformity correc- tube signals are then monitored and the pre-
tions, newer gamma cameras can use thinner amplifier adjusted electronically if the PM
light guides (or eliminate the light guide tube signal has drifted.
entirely) and smaller PM tubes, both of which A second approach uses two narrow energy
contribute to more accurate event localization windows, placed just above the photopeak
and improved intrinsic spatial resolution. position to minimize the influence of scatter
(Fig. 14-13). The count ratio between the two
4. Gamma Camera Tuning energy windows during flood-field irradiation
The nonuniformity corrections described pre- by the radionuclide of interest is measured for
viously require that the gamma camera each PM tube. This ratio remains constant,
remain very stable over time. However, the unless the PM tube signal drifts over time. If
gain of PM tubes invariably changes as the the count ratio changes, the PM tube pream-
tubes age. The high-voltage supply and ampli- plifier is adjusted electronically to restore the
fier gain can also drift over time. A method to ratio to its original value.
tune the PM tubes to ensure consistent per- Some of these tuning methods also can be
formance over time is therefore necessary. adapted so that they are continuous, in the
On many older systems, the tuning is done sense that the camera is tuned dynamically
manually. One method involves irradiation of every few seconds during a patient study.
the gamma camera detector through a lead This can be used to adjust the energy
mask with holes centered over each of the PM windows in real time, compensating for any
tubes. The output of each PM tube is exam- drift that occurs during the course of a study.
ined and the preamplifier gain is adjusted if The major cause of drift on such short
220 Physics in Nuclear Medicine

500 Window 1
400
Window 2

Counts
300
200
100

0 50 100 150
Energy (keV)
FIGURE 14-13 The ratio of counts detected in two narrow energy windows can be used to determine whether a pho-
tomultiplier tube is drifting. The windows are defined on the high side of the photopeak to avoid any contribution from
scatter. The ratio is virtually independent of source distribution and the amount of scattering material present.

timescales usually is related to count-rate the collimator be designed carefully. Poor


effects. At high counting rates, a small base- design can result only in poorer overall per-
line shift can occur in the signal amplitudes formance. Design considerations for parallel-
owing to overlap of pulses, and continuous hole collimators are discussed in this section.
adjustment of the energy window minimizes Design characteristics for converging and
such effects, keeping the energy window cen- diverging collimators are similar to those of
tered over the photopeak irrespective of the the parallel-hole type. Design characteristics
counting rate. Continuous tuning also is of pinhole collimators are not discussed in
important in single photon emission com- detail here but are described in references 4
puted tomographic imaging (Chapter 17), in and 5. The analysis to be presented for
which rotation of the gamma camera through parallel-hole collimators is similar to that
the earths magnetic field can result in presented by Anger in reference 1, which may
changes of PM tube gain. A detailed discus- be consulted for a more detailed discussion.
sion of automatic tuning methods can be
found in reference 3. 2. Septal Thickness
A primary consideration in collimator design
is to ensure that septal penetration by rays
C. DESIGN AND PERFORMANCE crossing from one collimator hole into another
CHARACTERISTICS OF is negligibly small. This is essential if an
PARALLEL-HOLE COLLIMATORS accurate -ray image is to be projected by the
collimator onto the camera detector. No thick-
ness of septal material is sufficient to stop all
1. Basic Limitations in rays, so the usual criteria is to accept some
Collimator Performance reasonably small level of septal penetration
The collimator is a weak link for the perfor- (e.g., ~5%).
mance of a gamma camera system, as indeed The required septal thickness can be deter-
it is in any nuclear medicine imaging system mined by analysis of Figure 14-14. The short-
employing the principles of absorptive colli- est path length for rays to travel from one
mation. Collimator efficiency, defined as the hole to the next is w. Septal thickness t is
fraction of rays striking the collimator that related to w, and to the length l and diameter
actually pass through it to project the -ray d of the collimator holes, by
image onto the detector, is typically only a few
percent or less. Collimator resolution, which t 2dw/(l w) (14-1)
refers to the sharpness or detail of the -ray
image projected onto the detector, also is If septal penetration is to be less than 5%, the
rather poor, generally worse than the intrin- transmission factor for the thickness w must
sic resolution of the camera detector and be
electronics.
Because it is a limiting factor in camera
system performance, it is important that ew 0.05 (14-2)
14 The Gamma Camera: Performance Characteristics 221

upper limit of approximately 150keV and


Detector
medium-energy collimators of approximately
400keV. High-energy collimators are used for
imaging positron-emitting radionuclides at
511keV.
w Collimator
EXAMPLE 14-1
l septa Calculate the septal thickness required for
low-energy (150keV) and medium-energy
(400keV ) lead collimators having hole diam-
eters of 0.25cm and lengths of 2.5cm.
d t
Answer
FIGURE 14-14 Minimum path length w for a ray The linear attenuation coefficient of lead at
passing through the collimator septa from one hole to the 150keV is l = 1.91cm2/g 11.34g/cm3 =
next depends on length l and diameter d of the collimator 21.66cm1 and at 400keV is l = 0.22cm2/g
holes and on septal thicknesses t.
11.34g/cm3 = 2.49cm1 (Appendix D). There-
fore from Equation 14-5 for the low-energy
collimator
where is the linear attenuation coefficient 6 0.25 / 21.66
of the septal material. Because e3 is approxi- t
mately 0.05, this implies 2.5 ( 3 / 21.66 )
0.029 cm
w 3 (14-3)
and for the medium-energy collimator
w 3/ (14-4) 6 0.25 / 2.49
t
and thus 2.5 ( 3 / 2.49 )
6d / 0.465 cm
t (14-5)
l (3 / ) As shown by this example, thicknesses
It is desirable that septal thickness t be as needed for low-energy collimators are only a
small as possible so that the collimator septa few tenths of a millimeter, which is in the
obstruct the smallest possible area of detector range of lead foil thicknesses and approaches
surface and collimator efficiency is maxi- the limits of lead thicknesses that can be used
mized. This objective is realized by using a without loss of necessary mechanical strength.
material with a large value of for the colli- Indeed, low-energy collimators generally are
mator septa. Materials of high atomic number quite fragile, and their septa can be damaged
Z and high density are preferred. Lead (Z = easily by mechanical abuse (such as dropping
82, = 11.3g/cm3) is the material of choice for or stacking on sharp objects). Medium-energy
reasons of cost and availability; however, collimators require substantially greater
other materials, including tantalum (Z = 73, septal thicknesses, typically a few millimeters
= 16.6g/cm3), tungsten (Z = 74, = 19.3g/ of lead. Alternatively, medium-energy collima-
cm3), gold (Z = 79, = 19.3g/cm3) and even tors can be made thicker (larger l in Equation
depleted uranium (Z = 92, = 18.9g/cm3) 14-5).
have been employed in experimental Low-energy -ray emitters (e.g., 99mTc,
applications. 140keV) can be imaged using medium-energy
As discussed in Chapter 6, Section D.1, collimators. This is done, however, with an
attenuation coefficients of heavy elements unnecessary sacrifice of collimator efficiency
depend strongly on -ray energy in the nuclear because the collimator septa are unnecessarily
medicine energy range. Thus the required thick. (See Table 14-1 for comparative efficien-
septal thickness also depends strongly on cies of low- and medium-energy collimators.)
the -ray energy for which the collimator is Low-energy collimators are used whenever
designed to be used. Commercially available possible to obtain maximum collimator effi-
collimators are categorized according to the ciency. When choosing a collimator, however,
maximum -ray energy for which their septal one must consider not only the energy of the
thickness is considered to be adequate. rays to be imaged but also the energies of any
Low-energy collimators generally have an other rays emitted by the radionuclide of
222 Physics in Nuclear Medicine

interest or by other radionuclides that may be Collimator resolution


present as well (e.g., residual activity from
another study or radionuclide impurities).
Higher-energy rays may be recorded by Projected
radiation
Compton downscatter into a lower-energy profile FWHM
analyzer window. If the collimator septa are
too thin, the collimator may be virtually trans-
parent to higher-energy rays, causing a rela-
tively intense foggy background image to
be superimposed on the desired image, with l
a resulting loss of image contrast. Whether
a low-energy collimator can be used when Multihole
collimator d
higher-energy rays are present depends on
the energy and intensity of those emissions
and requires experimental evaluation in spe- b
cific cases.
3. Geometry of Collimator Holes
Point or
Collimator performance also is affected by the line source
geometry of the collimator holes, specifically,
FIGURE 14-15 Radiation profile (point- or line-spread
their shape, length, and diameter. The pre- function) for a parallel-hole collimator. The full width at
ferred hole shape, to maximize the exposed half maximum (FWHM) of the profile is used to charac-
area of detector surface for a given septal terize collimator resolution.
thickness, is round or hexagonal, with the
holes arranged in a close-packed hexagonal
array, or square holes in a square array. Tri-
angular holes also have been used. to septal penetration. For 2.5-cm thick low-
Collimator hole length and diameter affect energy collimators (150keV), the difference
strongly both collimator resolution and colli- between effective and actual length is approx-
mator efficiency. Collimator resolution Rcoll is imately 0.1cm, whereas for 2.5-cm thick
defined as the FWHM of the radiation profile medium-energy collimators (400 keV) it is
from a point or line source of radiation pro- approximately 0.8cm.
jected by the collimator onto the detector (Fig.
EXAMPLE 14-2
14-15). This profile is also called the point-
spread function (PSF) or line-spread function Calculate the resolution (FWHM) of the low-
(LSF). Collimator resolution Rcoll is given by* energy collimator described in Example 14-1,
at source depths b = 0 and b = 10cm, assum-
Rcoll d(leff + b) /leff (14-6) ing it has a septal thickness of 0.03cm.

where b is the distance from the radiation Answer


source to the collimator and d is the diameter The effective length of the collimator is
and leff = l 21 the effective length of the
collimator holes. Here is the linear attenu- leff = 2.5 cm (2 / 21.66) cm 2.4 cm
ation coefficient of the collimator material.
The effective length of the collimator holes is Thus for b = 0
somewhat less than their actual length owing
Rcoll 0.25 (2.4 + 0) / 2.4 cm

0.25 cm
*Some versions of Equation 14-6 include additional cor- and at b = 10 cm
rection terms involving the thickness of the detector
crystal, reflecting the fact that the image actually is Rcoll 0.25 (2.4 + 10) / 2.4 cm
formed at some depth within the detector crystal. Because
photons of different energies penetrate to different 1.3 cm
average depths within the crystal, the correction actually
is photon-energy dependent, a point not noted in some This example illustrates the strong depen-
texts. The correction is small and for simplicity is omitted
from Equation 14-6, as well as from Equations 14-10 and dence of collimator resolution on the distance
14-13 for the converging and diverging collimators pre- of the source from the collimator.
sented later in this chapter.
14 The Gamma Camera: Performance Characteristics 223

Collimator efficiency g, defined as the frac- camera systems also provide a selection of col-
tion of rays passing through the collimator limators with different combinations of resolu-
per ray emitted by the source is given by tion and efficiency. Those with good resolution
but poor efficiency generally are described as
g K 2 (d/leff )2 [ d2 / (d + t)2 ] (14-7) high-resolution collimators, whereas those
with the opposite characteristics are described
where t is septal thickness and K is a constant as high-sensitivity collimators. Those with
that depends on hole shape (~0.24 for round characteristics intermediate to the extremes
holes in a hexagonal array, ~0.26 for hexago- are referred to as general purpose, all
nal holes in a hexagonal array, ~0.28 for purpose, or by other similar names.
square holes in a square array1). Equation Equation 14-6 indicates that collimator
14-7 applies to a source in air and assumes resolution becomes poorer as source-to-
no attenuation of radiation by intervening collimator distance b increases. Thus struc-
body tissues. tures closest to the collimator are imaged
Several aspects of Equations 14-6 and 14-7 with sharpest detail. Figure 14-16 shows
should be noted. First, resolution improves as graphically the relationship between colli-
the ratio of hole diameter to effective length mator resolution and source-to-collimator
(d/leff) is made smaller. Long, narrow holes distance for three different collimators pro-
provide images with the best resolution; vided by one commercial manufacturer. Typi-
however, collimator efficiency decreases cally, collimator resolution deteriorates by a
approximately as the square of the ratio of factor of 2 at a distance of 4-6cm from the
hole diameter to length (d/leff)2. Thus an collimator.
approximate relationship between collimator On the other hand, according to Equation
efficiency, g, and spatial resolution, Rcoll, is 14-7, collimator efficiency for a source in air
is independent of source-to-collimator dis-
g ( Rcoll )2 (14-8) tance b. This rather surprising result is
obtained provided the counting rate for the
Therefore for a given septal thickness, colli- entire detector area is measured. The reason
mator resolution is improved only at the for this is illustrated by Figure 14-17. As the
expense of decreased collimator efficiency, source is moved farther away from the colli-
and vice versa. mator, the efficiency with which radiation is
transmitted through any one collimator hole
EXAMPLE 14-3 decreases in proportion to 1/b2 (inverse-square
Calculate the efficiency g of the collimator law), but the number of holes through which
described in Examples 14-1 and 14-2, assum- radiation can pass to reach the detector
ing it has hexagonal holes in a hexagonal increases in proportion to b2. The two effects
array. cancel each other, with the result that total
counting rateand thus collimator efficiency
Answer does not change with source-to-collimator dis-
For hexagonal holes in a hexagonal array, tance. Another illustration of this effect is
K = 0.26. Thus, shown in Figure 14-18. As source-to-collima-
tor distance increases, the maximum height
g (0.26)2 (0.25 / 2.4)2 [(0.25)2 / (0.25 + 0.03)2 ] of the PSF or LSF decreases, but the width
(0.0676) (0.0109) (0.797) increases (and resolution becomes poorer), so
that the total area under the curve (total
5.85 104
detector counting rate) does not change.
(photons transmitted/photons emitted))
Invariance of collimator efficiency with
This example illustrates the relatively small source-to-collimator distance applies to point
fraction of emitted rays that are transmitted sources, line sources, and uniform sheet
by a typical gamma camera collimator. sources in air with parallel-hole collimators;
however, it applies only to uniform sheet
Equation 14-7 also demonstrates the effect sources with converging, diverging, or pinhole
of septal thickness on efficiency. Medium- collimators (Section D). When the source is
energy collimators have lower efficiencies embedded at different depths in the patient,
than low-energy collimators because of their attenuation effects also must be considered.
greater septal thicknesses. Septal penetration and scatter of photons
In addition to providing low- and medium- from the walls of the collimator holes also are
energy collimators, manufacturers of gamma not considered in the earlier analysis.
224 Physics in Nuclear Medicine

2.0

1.8

1.6
Collimator resolution FWHM (cm)

1.4
ity
1.2 itiv
ens
hs
1.0 Hig e
pos
al pur
ner
0.8 Ge

0.6 tion
solu
igh re
H
0.4

0.2

0.0
0 2 4 6 8 10 12 14 16
Source-to-collimator distance (cm)
FIGURE 14-16 Collimator resolution versus source-to-collimator distance for three different collimators. (Adapted
from Hine GJ, Paras D, Warr CP: Recent advances in gamma-camera imaging. Proc SPIE 152:123, 1978.)

Single-hole
efficiency  1/b2

FIGURE 14-17 Explanation for con-


Detector stant counting rate (collimator efficiency)
area Radiation versus source-to-collimator distance for a
exposed source point source in air and a parallel-hole
 b2 collimator. Efficiency for a single hole
decreases as 1/b2, but number of holes
b passing radiation (area of detector
exposed) increases as b2.

Collimator
Detector

Counting rate

FIGURE 14-18 Point-spread functions versus dis-


tance for a parallel-hole collimator. Area under
Source-to-collimator curve is proportional to collimator efficiency and
distance does not change with distance.

Distance
14 The Gamma Camera: Performance Characteristics 225

TABLE 14-1
PERFORMANCE CHARACTERISTICS OF SOME TYPICAL COMMERCIALLY MANUFACTURED
PARALLEL-HOLE COLLIMATORS

Resolution Rcoll
Recommended Max. (FWHM at
Collimator Type Energy (keV) Efficiency, g 10cm)
Low-energy, high-resolution 150 1.84 104 7.4mm
Low-energy, general-purpose 150 2.68 10 4
9.1mm
Low-energy, high-sensitivity 150 5.74 10 4
13.2mm
Medium-energy, high-sensitivity 400 1.72 104 13.4mm
Adapted from Hine GJ, Erickson JJ: Advances in scintigraphic instruments. In Hine GJ, Sorenson JA (eds):
Instrumentation in Nuclear Medicine, Vol 2. New York, 1974, Academic Press.
FWHM, full width at half maximum.

Table 14-1 summarizes the physical con- differences between system resolutions for
struction and typical performance character- cameras having substantially different intrin-
istics of a number of collimators. Collimator sic resolutions (e.g., 4mm vs. 8mm), but the
resolution is the FWHM for a source at 10cm difference in system resolutions for cameras
from the face of the collimator. Collimator effi- having small differences in intrinsic resolu-
ciency g refers to the relative number of rays tions (e.g., 4mm vs. 5mm) is minor and not
transmitted by the collimator and reaching clinically significant. Small differences in
the detector per ray emitted by the source. intrinsic resolution may be apparent on bar-
Note that the approximate relationship pattern images or on images of very superfi-
between collimator efficiency and resolution cial structures in the patient, but they usually
given by Equation 14-8 is verified by these are not apparent on images of deeper-lying
data. Note also the relatively small values for structures.
collimator efficiency. System resolution also is degraded by scat-
tered radiation. This is discussed in Chapter
4. System Resolution 15, Section C. The method for combining com-
The sharpness of images recorded with a ponent resolutions to determine system reso-
gamma camera is limited by several factors, lution also is discussed in Appendix G.
including intrinsic resolution, collimator reso-
lution, scattered radiation, and septal pene-
tration. In terms of the FWHM of a PSF or D. PERFORMANCE CHARACTERISTICS
LSF, the most important factors are the OF CONVERGING, DIVERGING, AND
intrinsic resolution Rint of the detector and PINHOLE COLLIMATORS
electronics, and the collimator resolution Rcoll.
The combined effect of these two factors is to Figure 14-20 illustrates the important design
produce a system resolution Rsys that is some- parameters for converging, diverging, and
what worse than either one alone. System pinhole collimators. Equations for collimator
resolution Rsys (FWHM) is given by resolution, Rcoll, and efficiency, g, for these col-
limators are as follows:
Rsys = 2
Rint + Rcoll
2 (14-9)
Converging Collimator:
Because collimator resolution depends on
source-to-collimator distance, system resolu- Rcoll [ d(leff
+ b) / leff
][1 / cos ]
tion also depends on this parameter. Figure (14-10)
[1 (leff
/ 2) /( f + leff
)]
14-19 shows system resolution versus source-
to-collimator distance for a typical parallel-
hole collimator and different values of intrinsic g K 2 (d / leff
)2 [ d2 /(d + t)2 ][ f 2 /( f b)2 ]
resolution. At a distance of 5-10cm (typical (14-11)
depth of organs inside the body), system
resolution is much poorer than intrinsic where
resolution and is determined primarily by
collimator resolution. There are significant (l 2 1 ) / cos leff / cos (14-12)
leff
226 Physics in Nuclear Medicine

1.8

m
1.6 8m
t

R in m
4m
1.4 t

R in
0
System resolution, FWHM (cm)

t

1.2 R in

FIGURE 14-19 System resolution versus


1.0 source-to-collimator distance for a typical
parallel-hole collimator and different
values of intrinsic resolution. At most
0.8 typical organ depths, system resolution
is determined primarily by collimator
resolution.
0.6

0.4 Typical
organ
depths
0.2

0.0
0 2 4 6 8 10 12 14 16
Source-to-collimator distance (cm)

f
d d detector


l l l
d

b b
b
f 



Converging Diverging Pinhole


FIGURE 14-20 Parameters for collimator resolution Rcoll, and efficiency, g, for Equations 14-10 to 14-18.

Diverging Collimator: Pinhole Collimator:

Rcoll [ d(leff
+ b) / leff
][1 / cos ][1 + (leff
/ 2 f )] Rcoll deff ,R (l + b) /l (14-15)
(14-13)
g K 2 (d / leff
)2 [ d2 /(d + t)2 ][( f + l) /( f + l + b)]
(14-14) g deff ,g 2 cos3 / (16b2 ) (14-16)
14 The Gamma Camera: Performance Characteristics 227

where image magnification factor is 2, the equiva-


ln (2) lent FWHM in the imaged plane is 1cm.
deff ,R = d+ tan These equations may be compared with
2 (14-17) Equations 14-6 and 14-7 for the parallel-hole
and collimator. They are similar except for the
presence of additional terms involving colli-
deff ,g = d[ d + (2 /) tan( / 2)] + [(2 / 2 ) tan 2 ( / 2)] mator focal lengths f and, for off-axis sources,
(14-18) the angle between the source, the focal point
(or pinhole), and the central axis of the colli-
leff is the effective collimator length, mator. The equations illustrate that for con-
accounting for septal penetration at different verging and diverging collimators, resolution
off-axis locations (see also Equation 14-6). For is best at the center ( = 0, cos = 1).
the pinhole collimator, deff,R and deff,g are the The performance characteristics of differ-
effective pinhole diameters, for resolution ent types of collimators are compared in
and sensitivity, respectively. deff takes into Figure 14-21, which shows system resolution
account the penetration of gamma rays and efficiency versus distance, including
through the edges of the pinhole aperture, but effects of camera intrinsic resolution as well
still assumes parallel rays, normally incident as collimator magnification. Equations 14-10,
on the detector surface.4,5 These expressions 14-13, and 14-15 show that resolution always
for deff also assume that the pinhole aperture is best with the source as close as possible to
has a knife-edge geometry, as illustrated in the collimator. Changes in collimator effi-
Figure 14-20. ciency with distance depend on whether the
The equations for collimator resolution Rcoll radiation source is a point source or a uniform
refer to the equivalent FWHM of the PSF or sheet source.
LSF, corrected for magnification or minifica- For a point source (Fig. 14-21, right), col-
tion of the image by the collimator described limator efficiency increases with increasing
by Equations 13-3, 13-5, and 13-6. Thus, if source-to-collimator distance for the con
the collimator projects a profile with a 2-cm verging collimator. Maximum efficiency is
FWHM measured on the detector and the obtained at the collimator convergence point

20 250

Diverging
16 200 Converging
Relative geometric efficiency
System resolution (mm)

Parallel-hole
12 150

Parallel-hole
8 100

Pinhole

4 Converging
50
Diverging
Pinhole

0 0
0 5 10 15 20 0 5 10 15 20
A B
Source-to-collimator distance (cm)
FIGURE 14-21 Performance characteristics ( A, system resolution; B, point-source geometric efficiency in air) versus
source-to-collimator distance for four different types of gamma camera collimators. (Adapted with permission from
Society of Nuclear Medicine from Moyer RA: A low-energy multihole converging collimator compared with a pinhole
collimator. J Nucl Med 15:59-64, 1974.)
228 Physics in Nuclear Medicine

(~35cm), where rays are transmitted distance; they also have a quite limited field
through all of the collimator holes, and then of view because of magnification effects at
decreases beyond that point. Point-source typical imaging distances (Equation 13-3).
collimator efficiency decreases with distance Generally they are used for imaging smaller
for the diverging and pinhole collimators, organs, such as the thyroid and heart, which
more severely for the latter. For an extended, can be positioned close to the collimator. They
large-area sheet source, sufficiently large to also are useful with high-energy -ray emit-
cover the entire field of view of the collimator, ters because they can be designed to reduce
efficiency does not change with source-to- septal penetration problems.
collimator distance for all of these collima- Differences between the resolution and
tors. Again, for sources embedded within a field-of-view obtained at different source-to-
patient, attenuation effects also must be collimator distances with parallel-hole, con-
accounted for. verging, diverging, and pinhole collimators are
Figure 14-21 illustrates that the converg- further illustrated by Figure 14-22. The distor-
ing collimator offers the best combination of tions caused by changing magnification with
resolution and efficiency at typical imaging depth for different structures inside the body
distances (5 to 10cm); however, the field-of- sometimes make images obtained with the
view is also somewhat limited at these dis- converging, diverging, and pinhole collimators
tances (Equation 13-6 and Example 13-2), difficult to interpret (see Fig. 13-9).
and for this reason converging collimators are
most useful with cameras having relatively
large-area detectors. Diverging collimators E. MEASUREMENTS OF GAMMA
offer a larger imaging area (Example 13-1) CAMERA PERFORMANCE
but at the cost of both resolution and effi-
ciency. Pinhole collimators offer very good It is important to define standardized experi-
resolution and reasonable efficiency at close mental protocols for measuring gamma
distances but lose efficiency very rapidly with camera performance that produce consistent

At
collimator
face

10 cm
distance

20 cm
distance

Diverging Parallel-hole Converging


FIGURE 14-22 Bar-pattern images demonstrating changing field size and resolution obtained versus distance for
three collimator types.
14 The Gamma Camera: Performance Characteristics 229

results, are easily reproducible, and that do (<10,000 cps for a small field-of-view gamma
not require specialized equipment. Such pro- camera) to avoid pile-up-related mis
tocols can then be used in comparing one positioning. Two sets of images are taken,
gamma camera with another, in performing with a 90-degree rotation of the test pattern
acceptance testing on a newly installed between acquisitions so that both X and Y
gamma camera, and as part of a quality resolution are measured. Data are acquired
assurance program to ensure that the camera until the peak channel has at least 1000
is always performing to its specifications. counts. Images are acquired in a matrix with
The exact regulations for gamma camera pixel sizes less than 1 10 of the expected reso-
quality assurance and the guidelines for lution (typically <0.35mm). Profiles through
acceptance testing vary with locality. For the images of the line sources are taken at
example, in the United States, the Joint different locations across the gamma camera
Commission requires that instruments be face and fitted to a Gaussian function
tested prior to initial use and that the per (Chapter 9, Section B.3). The FWHM (Fig.
formance of a gamma camera be tested at 14-15) and full width at tenth maximum
least once a year. Because of the rapidly (FWTM) of the profiles are measured in both
changing regulatory environment, and differ- X and Y directions. The reported measure-
ences between states and countries, a detailed ments usually are average measurements
review of the requirements of regulatory across the UFOV, and the average across the
agencies is beyond the scope of this book. In central field-of-view (CFOV ) that has linear
this section, we therefore briefly summarize dimensions scaled by 75% with respect to the
the more common measurements that are dimensions of the UFOV. Typical values of
performed to assess gamma camera perfor- intrinsic spatial resolution are 2.5 to 3.5mm.
mance (whether they be for acceptance
testing or for quality assurance). The proto- 2. System Resolution
cols presented here are largely based on the This measurement is made with the collima-
recommendations of the National Electrical tor in place and should be repeated for each
Manufacturers Association.6 A typical quality collimator of interest. The source consists of
assurance program might involve daily mea- two 1-mm-diameter line sources, placed 5cm
sures of flood-field uniformity, weekly checks apart at a distance of 10cm from the front
of spatial resolution and spatial linearity, face of the collimator. The measurement also
and semiannual checks of other performance can be performed with the addition of a scat-
parameters. It is important that all measure- tering medium by placing 10cm of plastic
ments be taken under the same conditions between the sources and the collimator, and
(pulse-height window width, correction algo- 5cm of the same material behind the sources.
rithms, and correction circuitry on or off ) as Images are acquired (typically several million
are used for routine clinical studies. More events, at a rate of <30,000 cps to avoid pile-
detailed information on performance mea- up) and profiles taken through the image of
surements, quality assurance, and accep- the line sources are fitted to Gaussian func-
tance testing can be found in references tions to determine FWHM and FWTM as
7 to 9. described for intrinsic resolution. The results
vary widely depending on the exact type of
1. Intrinsic Resolution collimator used but are typically in the range
Intrinsic resolution is determined without a of 8 to 14mm for 99mTc.
collimator using a linearity test pattern, such
as the one shown in Figure 14-10 (left), placed 3. Spatial Linearity
directly on the surface of the NaI(Tl) crystal This measurement uses the same slit pattern
housing. The width of the strips in the pattern (Fig. 14-10A) and conditions as for the intrin-
is approximately 1mm, which is significantly sic resolution measurement. Once again,
smaller than the resolution expected in the measurements are taken with two orienta-
measurement. A point source (usually 99mTc tions of the test pattern, rotated by 90 degrees,
or 57Co) is placed at a distance equal to five to provide linearity measurements in both X
times the UFOV from the gamma camera and Y directions. Two measurements can be
face. The UFOV corresponds to the field of made from the resulting images. The differ-
view of the gamma camera after masking off ential spatial linearity is the deviation of the
the portion of the camera face affected by measured distance di between two slits from
edge-packing effects. Data are acquired with the actual distance D between them calcu-
the system count rate below 30,000 cps lated for each row i in the image. The means
230 Physics in Nuclear Medicine

and the standard deviations are reported for across all rows and columns of the image. It
the X and Y directions across the UFOV and is defined as
the CFOV and are defined as:
(high low )
Differential Uniformity (%) = 100
n (high + low )
(d i D)
(14-19) (14-22)
i =1
Mean =
n where high refers to the maximum count
n difference for any five consecutive pixels (row
(d i D)2 (14-20) or column) in the image and low refers to
Standard Deviation = i =1 the minimum count different for any five con-
n 1 secutive pixels. This usually is reported for
the UFOV.
In addition, the absolute spatial linearity is For convenience, uniformity measurements
defined as the maximum deviation of the often are made with the collimator in place
location of the slits from their true location. (extrinsic uniformity). A thin flood-field source
Once again this is assessed for the UFOV of 99mTc or a disk source of 57Co that covers the
and the CFOV. It is not easy to detect small active area of the gamma camera is placed on
nonlinearities using these techniques, and top of the collimator to provide uniform irra-
tests of uniformity (discussed in the next diation. This protocol is more practical for
section) usually are better at revealing the routine quality assurance because the mea-
effects of small nonlinearities. surement can be done without removing the
collimator. Extrinsic uniformity measure-
4. Uniformity ments also have the advantage that they
Intrinsic uniformity is determined from flood- reveal any defects or problems caused by the
field images acquired without a collimator. A collimator itself.
99m
Tc source is placed at a distance of approxi-
mately 5 the UFOV from the front face of 5. Counting Rate Performance
the gamma camera. The source activity is As described in Section A.4, most gamma
such that the counting rate on the gamma cameras behave as paralyzable counting
camera is less than approximately 30,000 cps. systems with the observed count rate described
Flood-field images are acquired so that there as a function of the true count rate by Equa-
are a minimum of 4000 counts in each pixel tion 11-18. The basis for measurement of the
of the image and then smoothed with a 9-point dead time, , is the two-source method
(3 3) smoothing filter with the following described in Chapter 11, Section C.4. Two
99m
weightings: Tc sources are placed approximately 1.5m
away from the camera face. The total activity
1 2 1 should be sufficient to cause approximately a
2 4 2 20% loss in the observed counting rate rela-
tive to the true counting rate. Counting rates
1 2 1 then are measured with both sources present,
and then with each individual source present.
Integral uniformity is based on the Care must be taken that all measurements
maximum and minimum pixel counts in the are performed with exactly the same source
image and is defined as geometry, that pile-up rejection electronics or
any other high counting rate correction cir-
Integral Uniformity (%) cuitry is turned on, and that source decay is
negligible (<1%) during the course of the mea-
max. pixel count min. pixel count
= 100 surement. The dead time can then be calcu-
max. pixel count + min. pixel count lated from Equation 11-25. The observed
(14-21) count rate at which a 20% counting rate loss
occurs, R20%, is also often quoted, and this can
This is calculated for the UFOV and CFOV. be computed from Equation 11-18 using the
Integral uniformity values are typically 2% fact that Ro = 0.8Rt as
to 4%.
Differential uniformity is based on the 0 .8
R20% = ln(0.8) (14-23)
change in counts of five consecutive pixels
14 The Gamma Camera: Performance Characteristics 231

6. Energy Resolution Sensitivity (cps/Bq)


Energy resolution is measured with a flood counts in ROI background counts in ROI
illumination of the gamma camera face, =
time(sec) source activity (Bq)
without a collimator, using a 99mTc source sus-
(14-24)
pended 5 UFOV above the camera face. The
resulting pulse-height spectrum is analyzed
to determine the FWHM of the 99mTc photo- A general-purpose collimator typically has a
peak. It usually is reported in keV or con- sensitivity on the order of 2 to 3 104 cps/Bq
verted to a percent energy resolution based or 0.02% to 0.03%.
on the energy of the photopeak (see Equation
10-3). Typical values are in the range of 8%
to 11% for 99mTc. REFERENCES
1. Anger HO: Radioisotope cameras. In Hine GJ, editor:
7. System Sensitivity Instrumentation in Nuclear Medicine, Vol 1. New York,
1967, Academic Press, pp 485-552.
System sensitivity needs to be measured sep- 2. Muehllehner G, Karp JS: A positron camera using
arately for each collimator. In general, the position-sensitive detectors: PENN-PET. J Nucl Med
sensitivity of low-energy collimators is mea- 27:90-98, 1986.
sured with 99mTc (E = 140keV ), that of 3. Graham LS: Automatic tuning of scintillation cameras:
A review. J Nucl Med Tech 14:105-110, 1986.
medium-energy collimators is measured with 4. Accorsi R, Metzler SD: Analytic determination of the
111
In (E = 172, 247keV), and 131I (E = resolution-equivalent effective diameter of a pinhole
364keV ) is used for high-energy collimators. collimator. IEEE Trans Med Imag 23:750-763, 2004.
A solution of the radionuclide (known total 5. Smith MF, Jaszczak RJ: The effect of gamma ray pen-
activity) is placed in a 10-cm diameter dish to etration on angle-dependent sensitivity for pinhole
collimation in nuclear medicine. Med Phys 24:1701-
a depth of 2 to 3mm. The shallow depth mini- 1709, 1997.
mizes self-absorption by the source. The 6. Performance measurements of scintillation cameras.
source is placed 10cm from the front face of Standards Publication No. NU-1-2007. Washington,
the collimator and an image is acquired. The DC, 2007, National Electrical Manufacturers
Association.
sensitivity is calculated by drawing a circular 7. Simmons GH, editor: The Scintillation Camera, New
region of interest around the image of the York, 1988, Society of Nuclear Medicine.
dish and integrating all the counts in that 8. Graham LS: Scintillation camera imaging perfor-
region. A second image is recorded for an mance and quality control. In Henkin RE, Boles MA,
equal imaging time with the source removed Karesh SM, et al, editors: Nuclear Medicine, St. Louis,
1996, Mosby, pp 125-146.
to provide a measure of the background. The 9. Murphy PH: Acceptance testing and quality control of
same region of interest is applied to this gamma cameras, including SPECT. J Nucl Med
image. The sensitivity is given by 28:1221-1227, 1987.

You might also like