Operation And Calibration Of Friability Test Apparatus

1.0 OBJECTIVE
To lay down the procedure for operation and calibration of friability test apparatus
2.0 RESPONSIBILITY
Quality Control Chemist / QA Chemist
3.0 ACCOUNTABILITY
Quality Control Manager / QA Manager.
4.0 PROCEDURE
4.1 GENERAL CLEANING

Clean the apparatus free of dust with dry cloth from outside every day.
Clean the rotating disc and unloading pan from inside with dry cloth at the start and end of every
operation. Wet cloth may be used occasionally if cleaning is not proper with dry cloth but this should
follow drying of the inner surface by first wiping with dry cloth and exposure to atmosphere air.

4.2 OPERATING INSTRUCTIONS

Ensure that apparatus is properly connected with power supply.


Weigh tablets to be tested and record the weight {W I}.
Switch on the main switch of power supply.
Press MODE key and select count mode.
Press SET button and set the RPM to 25 with the help of numerical keys (0 to 9) & press ENTER
Key located on the front side of the instrument.
Load the weighed tablets in the drum, close the lid and secure it by tightening he screw and press ENTER
and then press START to run test.
After the completion of set rotations, the apparatus will automatically unload the tablets into the tray.
Observe the tablets.
Dedust the tablets, weigh and record the weight {W 2}.
Calculate the percentage Friability by the following formula;

Initial Wt. of Tablets (W1) Wt. of tablets After 100 rotation (W2) X 10

Initial Wt. of Tablets (W1)


Dstroy the tested tablets as per SOP

4.3 CALIBRATION PROCEDURE

Switch on the power supply.


Set the RPM to 25 and start the machine simultaneously with the stop watch. Count the actual rotations
and not the time required for the same.
Similarly set the RPM to 100 and note the time required and actual rotations.
Apparatus is in proper working condition if,
Time required for 25 rotations is 1 min 05 sec.
Time required for 100 rotations is 4 min 20 sec.
Record the observation in the calibration record as per Annexure 1.
Affix a Calibration Status label on the instrument.
In case of any discrepancy, report the observations to QC manager / QA Manager and notify the defect to
Engg. Department. Affix an UNDER MAINTENANCE label on the instrument.

FREQUENCY
Once in a month and after each maintenance job.
ANNEXURE 1

Revision No.: 00
REF. SOP NO.: Page No. : 1 of 1
Effective from:

Name of Instrument : Friability Test Apparatus


Instrument Ser. No. :
Make & Model : Electrolab.
Calibration date :
Calibration due on :

Sr. SET ROTATIONS ROTATION OBSERVED MA TIME


No. ROTATION DISPLAYED NUALLY ACCEPTANCE
CRITERIA

1 25 1 min 05 sec

2 100 4 min 20 sec

REMARKS : Satisfactory / Not satisfactory.

CALIBRATED BY: CHECKED BY :

DATE: DATE :

Prepared By Checked By Approved By Authorised by