1. Description
The signet 8150 Flow Totalizer is a battery-powered instrument capable of providing uninterrupted ow and total volume information for
4 years and beyond. The unique features of the 8150 include:
Easy setup and display selection with 4-button keypad
User selectable security access to prevent unwanted programming changes.
Displays ow rate from 0.001 to 9999 engineering units with an auto-ranging decimal point.
Three totalizers, one permanent and two that are independently resettable.
Displays elapsed time between operating periods.
Integral mount and panel mount options
3.6 V Lithium batteries last 4 years nominal in most applications.
Non-volatile memory stores all programming and totalizer values even when batteries are removed.
Dimensions
Optional
Rear 92 mm
96 mm 82 mm
Cover 96 mm (3.8 in.)
41 mm 96 mm 42 mm 60 mm 106 mm
56 mm
(1.7 in.) (2.3 in.)
97 mm
SIDE VIEW 102 mm (4.0 in.)
2. Specications
General
Compatibility: Signet 515, 525, 2517 Flow Sensors, Electrical:
3000 Irrigation Sensor Battery: Two 3.6V Lithium thionyl chloride, AA-size
Input Frequency Range: 1 to 400 Hz Sensor power output: +3.6 VDC @ 20A
Battery life: 4 years nominal @ 50C (122F)
Accuracy: Low Battery indication: Battery symbol on LCD display
Frequency: 0.5% of reading
Environmental:
Enclosure: Operating Temperature: -10C to 65C (14F to 149F)
Rating: NEMA 4X/IP65 (front panel) Storage Temperature: -40 to 100C (-40 to 212F)
Dimensions: 1/4 DIN (96 x 96 x 50 mm) (3.8 x 3.8 x 2.0 in.) Relative Humidity: 0 to 95% Non-condensing
Case material: PBT resin
Keypad material: Sealed 4-key silicon rubber Shipping Weight: 0.5 kg (1.1 lbs.)
Flow
Totalizer
Flow (3-8150-1)
Totalizer
(3-8150-1)
Terminal Cover
(4-3111)
Integral
adapter Kit
(3-8051)
Integral Mount
3-8150-PO
Liquid-tight Connector
3-8150-P1 (3-9000.392-1)
3-8150-TO
3-8150-VO Integral
Flow Sensor Integral
(3-8510-XX) Adapter kit
(3-8051)
Integral
Flow Sensor
Signet (3-8510)
Fittings
Signet
Fittings
1. Insert the wires from the sensor through the yellow conduit
base and locking ring.
3-8150
Flow Totalizer
2. Insert the locking ring into the conduit base, aligning it so
that the square tab is close to the conduit ports.
The parts identied in Bold type are required for this installation. The parts identied in Bold type are required for this installation.
Other parts are shown for reference only. Other parts are shown for reference only.
Totalizer
Panel Mount
(3-8150-1P)
Signet Total Flow
ENTER
Standard Mount
Paddlewheel
Signet (P51530)
Fittings
Universal
Mounting Kit
(3-8050) Signet
Fittings
The 8150 Panel-mount Totalizer is a standard 1/4 DIN package. Use a 92x92 mm punch tool to make the panel cutout.
Minimum spacing of 25 mm (1 in.) between panel units is recommended.
panel
gasket
mounting
terminals bracket
latch 7
sr Gn
SenIEL
(SH
D)
d
sr IN
SenD)
t- 6 (RE
Pwr Outpu sr V+
System
-
4 SenAC K)
Loop (BL
2 Pwr
5
92 x 9 1
System
Loop
+
3
Outpu
t+
(3.62 2 mm
x 3.62
in.)
Panel Mount
Installation Detail
Panel cutout quick-clips
Instructions 10 mm
1. Remove 10mm (3/8 in.) of insulation from sensor cable conductors.
2. Press down on orange lever to open terminal.
3. Insert wire into terminal until it hits bottom.
4. Release the lever to secure wire. Terminal no. Function
1 Signal Ground
2 Sensor signal
3 Open Collector Signal
4 DC Power to sensor
5 Ground
6 Ext. Reset
splice together
outside of the terminal
E
IT
WH
BLACK/SHIE
LD
+ +
+ + 1
RED
1 2
RED
2 #1 #2 3
#1 #2 3 4
4
BLAC
5 K
5 6
6
- - - - Connect shield
Connect shield to Ground wire
to Ground wire
Totalizer 1 Totalizer 1
Remote reset Remote reset
ENTER ENTER
ENTER ENTER
ENTER
5 sec. ENTER
ENTER
Totalizers are shipped from the factory with these standard settings:
Example: Change the security code from the factory standard 0-0-0-0 to custom setting 1-0-0-1
1.Hold the ENTER key for 2 seconds, then enter the current 8. Press the UP and DOWN keys together to exit the
SECURITY CODE. Calibrate menu, store the new settings and return to normal
If working with a new unit, press the ENTER key again. The operation.
display shows the rst menu item (Auto CAL) NOTE:
2. Press the UP key three times to scroll to Sec Code. Record and store your security code in a safe place!
3. Press RIGHT ARROW to edit the code.
The leading digit on the display will
ash.
4. Press the UP key one time to
scroll the ashing digit to 1. 1
Signet Total Flow Signet Total Flow
to 1.
7. Press the ENTER Signet Total Flow Signet Total Flow Signet Total Flow
Signet Total Flow
key to complete ENTER ENTER ENTER
6 7 8
Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow
ENTER ENTER
ENTER ENTER ENTER
ENTER
During normal operation the 8150 displays the ow rate and one selected totalizer value. Any one of the three totalizers can be set
as the standard display: The other two totalizers can be viewed by pressing the keypad. The display will automatically return to the
standard selection after ve minutes.
ENTER ENTER
Total 1 (tot1) can be reset from the keypad or from the external RESET
(see wiring, section 4) without the security code.
Total 1 is identied by a ashing display every six seconds.
Application: Use Total 1 to measure water usage for a recurring
period, as for a daily discharge volume.
Total 2 (tot2) can be reset only by entering the security code in the Signet Total Flow Signet Total Flow
ENTER
ENTER
Signet Total Flow
ENTER
"PERMANENT" label.)
3. Press RIGHT ARROW key. (PERMANENT 2s
Enter Security code
label begins to ash
4. Press UP key one time. Display changes to
ashing tot 1". 2 3 4 5
5. Press ENTER key to complete the edit
process.
6. Press UP and DOWN keys together to store Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow
ENTER
1 2
Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow
2s
8s 5m
1 2 3
Signet Total Flow Signet Total Flow
Signet Total Flow
ENTER ENTER
ENTER
2s
Enter Security code
4 5
Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow
8s
To prevent this loss, enter the security code, then enter any menu item and induce a "STORING" message.
Whenever the 8150 stores a setting, it also stores all current totalizer data:
Example: The Totalizer ow rate shows 60 GPM, while an external reference indicates a true ow rate of 70 GPM.
To adjust the 8150, change the Flow rate from 60 GPM to 70 GPM using AutoCAL.
1. Hold the ENTER key for 2 seconds, then enter the SECURITY CODE.
Auto CAL is the rst item in the menu.
2. Press the RIGHT key to select the Auto CAL function.
The display will show Auto CAL and the current ow rate will be ashing.
3. Press the RIGHT key again to change the ow rate. The display shows "Set Flo" and the rst digit of the ow rate will begin
ashing.
5. Press the RIGHT key to advance the ashing element to the "6".
5. Press the UP key one time to change the "6" to "7".
6. Press the ENTER key to complete the automatic calibration process.
The display shows a new K-factor with the rst digit ashing. This K-factor is based on the change in ow rate.
7. Press the ENTER key again to accept the new value.
NOTE: If the display shows "ERR SetFlo" the procedure was unsuccessful because the calculated K-factor is less than 0.001 or
greater than 999999. Verify the ow rate and start the Auto CAL procedure from step 1.
8. Press UP and DOWN keys together to store the new value in the memory.
The display will show "storing" for a few seconds, then return to normal operation.
1 2
Signet Total Flow
Signet Total Flow
Signet Total Flow
ENTER
ENTER
ENTER
2s
Enter Security code
3 4 5 6
Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow
ENTER
ENTER
ENTER
ENTER
ENTER
1 2
Signet Total Flow Signet Total Flow Signet Total Flow
*
ENTER ENTER ENTER
2s
Enter Security code
**
3 4 5 6 7
Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow
ENTER
8 9
Signet Total Flow Signet Total Flow Signet Total Flow
Signet Total Flow
**
ENTER ENTER
ENTER
ENTER
For example, if the ow RATE registers in Litres per minute, the totalizer may be set to 1 (factory standard), so it counts in 1-Litre
increments, or it may be set to 1000, so it counts in 1Kilolitre (1m3) increments.
By converting the Flow K-factor, the totalizer can also be set to count in other engineering units.
See sec. 17: "Flow and Total K-factor selection" for additional information about Total K-factor adjustments.
Example: Change the totalizer from 1 kiloLitre increments to count in 10-kiloLitre increments.
1. Hold the ENTER key for 2 seconds, then enter the SECURITY CODE.
The display shows the rst item in
the menu, AutoCAL.
2. Press the DOWN key two times to
scroll to the Total K-factor.
3. Press the RIGHT key to select the
1 Signet Total Flow Signet Total Flow
ENTER
Example:
The Flow K-factor is set at 480.19 pulses per Gallon.
The totalizer registers 10 Gallons when the actual volume is known to be 11 Gallons.
The error is 1 Gallon divided by 10 Gallons , or -10%. (The totalizer is counting 10% low, and the ow rate is reading 10% slow.)
Reduce the Flow K-factor by 10%: 480.19 - 10% = 432.17.
Change the Flow K-factor to 432 pulses per Gallon.
The result: The totalizer must count 10% fewer pulses from the ow sensor to register one Gallon, so both the totalizer and the ow
rate will increase by 10%.
1 2 **
Signet Total Flow
Signet Total Flow
Signet Total Flow
ENTER
ENTER
2s
ENTER
3 4 * 5 *
**
6 Signet Total Flow Signet Total Flow Signet Total Flow
ENTER ENTER
ENTER
Finished Editing?
* Exit Without Changing?
Example: Change the maximum decimal display from from hundredths to tenths.
hundredths to tenths: 5. Press the ENTER key to return to the menu.
1. Hold the ENTER key for 2 seconds, then enter the SECURITY The decimal will stop ashing.
CODE. 6. Press UP and DOWN keys together to exit the menu and
The display shows the rst item in the menu, AutoCAL. return to normal operation.
2. Press the DOWN key four times to scroll to the Decimal setting. The display shows "Storing" for a few seconds, then returns
The display shows four dashes and the current decimal setting. to the normal operating display.
3. Press the RIGHT key to select the decimal for editing.
The decimal point will begin to ash.
4. Press the DOWN key one time to move the ashing decimal
ENTER ENTER
2s
Enter Security code
2 3 4 5 *
Signet Total Flow Signet Total Flow Signet Total Flow
Signet Total Flow
ENTER ENTER
ENTER
ENTER
6
**
Signet Total Flow
Signet Total Flow Signet Total Flow
ENTER
ENTER ENTER
Finished Editing?
* Exit Without Changing?
1
Signet Total Flow Signet Total Flow
ENTER ENTER
2 3 4 * 5
Signet Total Flow Signet Total Flow Signet Total Flow Signet Total Flow
**6
Signet Total Flow Signet Total Flow Signet Total Flow
ENTER ENTER
ENTER
No SPEED, no SENSITIVITY
With SPEED averaging set to 0 (zero) and with SENSITIVITY set
to zero, the ow rate may be very unstable. This line represents
the actual output of the ow sensor as it responds to unstable
ow conditions in the pipe.
SPEED only
With SPEED set to 60 seconds and SENSITIVITY still set to zero
the ow rate is stabilized, but a sharp change in ow rate is not
represented for 60 seconds or longer. (dotted green line).
NOTE: The SENSITIVITY function is ineffective if the SPEED function is set to zero (seconds).
1
Signet Total Flow Signet Total Flow
ENTER ENTER
2 3 4 5
ENTER ENTER
ENTER
ENTER
ENTER ENTER
ENTER
In AutoCal, the calculated K-factor is Press RIGHT key to start Auto CAL procedure again.
outside the range of the8150. Make sure that the ow rate entered is accurate.
(less than 0.001 or
greater than 99999)
Both batteries are too depleted to Replace battery #1, then replace battery #2.
safely store settings.
If you want the and you want the TO- Set the Flow and set the Total
FLOW RATE to read in: TALIZER to count in: K-factor to: K-factor to:
17. Flow and Total K-factor selection
Liters Liters K(Liter) 1
Pages 17-19 provide K-factors for the Liters Kiloliters K(Liter) 1000
Signet 515, 525, and 2517 ow sensors. Liters cubic meters K(Liter) 1000
Use this table to convert the K-factor to
cubic meters cubic meters K(Liter) x 1000 1
other units of measure, and to set the Total
K-factor. cubic meters MegaLiters K(Liter) x 1000 1000
Kiloliters Kiloliters K(Liter) x 1000 1
NOTE: KiloLiters MegaLiters K(Liter) x 1000 1000
The maximum K-factor is 999999.
MegaLiters MegaLiters K(Liter) x 1 000 000 1
The minimum K-factor is 0.001.
U.S. gallons U.S. gallons K(gal) 1
U.S. gallons U.S. gallons x 1000 K(gal) 1000
U.S. gallons cubic feet K(gal) 7.4805
U.S. gallons acre inches K(gal) 27154
U.S. gallons Acre feet K(gal) 325848
U.S. gallons Kiloliters K(gal) 264.2
Acre-Inches Acre-Inches K(gal) x 27154 1
The following calibration data is reprinted from the instruction manual for the Signet 515 Flow sensor for your convenience.
SCH 80 CPVC TEES FOR SCH 80 CPVC 3/4 CR4T007 202.61 53.530
PIPE 1 CR4T010 127.14 33.590
1/2 CPV8T005 480.19 126.87 1-1/4 CR4T012 61.910 16.357
3/4 CPV8T007 257.72 68.090 1-1/2 CR4T015 40.410 10.676
1 CPV8T010 174.67 46.148 2 CR4T020 22.300 5.8917
1-1/4 CPV8T012 83.390 22.032 GALVANIZED IRON TEES ON SCH 40
1-1/2 CPV8T015 58.580 15.477 PIPE
SCH 40 WROUGHT STEEL PIPE SCH 5S STAINLESS STEEL PIPE STD WROUGHT STEEL PIPE SCH 40S STAINLESS STEEL
PER ANSI B36.10 PER ANSI B36.19 PER ANSI B36.10 PIPE PER ANSI B36.19
K-FACTOR K-FACTOR K-FACTOR K-FACTOR K-FACTOR K-FACTOR K-FACTOR K-FACTOR
PIPE PULSES/ PULSES/ PIPE PULSES/ PULSES/ PIPE PULSES/ PULSES/ PIPE PULSES/ PULSES/
SIZE U.S. GAL LITER SIZE U.S. GAL LITER SIZE U.S. GAL LITER SIZE U.S. GAL LITER
1 1/2 122.000 32.232 1 1/2 104.200 27.5297 1 1/2 122.000 32.2325 1 1/2 122.000 32.2325
2 78.690 20.790 2 67.160 17.7437 2 78.690 20.7900 2 78.690 20.7900
2 1/2 55.630 14.697 2 1/2 46.060 12.1691 2 1/2 55.630 14.6975 2 1/2 55.630 14.6975
3 35.530 9.3871 3 29.790 7.8705 3 35.530 9.3871 3 35.530 9.3871
3 1/2 26.070 6.8877 3 1/2 22.060 5.8283 3 1/2 26.070 6.8877 3 1/2 26.070 6.8877
4 19.840 5.2417 4 16.890 4.4624 4 19.840 5.2417 4 19.840 5.2417
5 12.090 3.1942 5 10.6500 2.8137 5 12.090 3.1942 5 12.090 3.1942
6 8.0410 2.1244 6 7.1160 1.8801 6 8.0410 2.1244 6 8.0410 2.1244
8 4.3500 1.1493 8 3.8700 1.0225 8 4.3500 1.1493 8 4.3500 1.1493
10 2.6080 0.6890 10 2.3570 0.6227 10 2.6080 0.6890 10 2.6080 0.6890
12 1.7610 0.4653 12 1.6060 0.4243 12 1.7400 0.4597 12 1.7400 0.4597
14 1.4250 0.3765 14 1.2980 0.3429 14 1.3950 0.3686 14 * *
16 1.0590 0.2798 16 0.9620 0.2542 16 1.0220 0.2700 16 * *
18 0.8180 0.2161 18 0.7400 0.1955 18 0.7800 0.2061 18 * *
20 0.6460 0.1707 20 0.5900 0.1559 20 0.6150 0.1625 20 * *
22 * * 22 0.4790 0.1266 22 0.4970 0.1313 22 * *
24 0.4350 0.1149 24 0.3990 0.1054 24 0.4110 0.1086 24 * *
SCH 80 WROUGHT STEEL PIPE SCH 10S STAINLESS STEEL XS WROUGHT STEEL SCH 80S STAINLESS STEEL
PER ANSI B36.10 PIPE PER ANSI B36.19 PIPE PER ANSI B36.10 PIPE PER ANSI B36.19
K-FACTOR K-FACTOR K-FACTOR K-FACTOR K-FACTOR K-FACTOR K-FACTOR K-FACTOR
PIPE PULSES/ PULSES/ PIPE PULSES/ PULSES/ PIPE PULSES/ PULSES/ PIPE PULSES/ PULSES/
SIZE U.S. GAL LITER SIZE U.S. GAL LITER SIZE U.S. GAL LITER SIZE U.S. GAL LITER
1 1/2 136.100 35.9577 1 1/2 113.600 30.0132 1 1/2 136.100 35.9577 1 1/2 136.100 35.9577
2 88.590 23.4055 2 72.560 19.1704
2 88.590 23.4055 2 88.590 23.4055
2 1/2 62.810 16.5945 2 1/2 48.750 12.8798
2 1/2 62.810 16.5945 2 1/2 62.810 16.5945
3 39.990 10.5654 3 31.250 8.2563
3 39.990 10.5654 3 39.990 10.5654
3 1/2 29.220 7.7199 3 1/2 23.010 6.0793
3 1/2 29.220 7.7199 3 1/2 29.220 7.7199
4 22.160 5.8547 4 17.540 4.6341
4 22.160 5.8547 4 22.160 5.8547
5 13.420 3.5456 5 10.8700 2.8719
5 13.420 3.5456 5 13.420 3.5456
6 9.0160 2.3820 6 7.2410 1.9131
6 9.0160 2.3820 6 9.0160 2.3820
8 4.8190 1.2732 8 3.9520 1.0441
10 2.3880 0.6309
8 4.8190 1.2732 8 4.8190 1.2732
10 2.8970 0.7654
12 1.9620 0.5184 12 1.6200 0.4280 10 2.7730 0.7326 10 2.7730 0.7326
Flow Totalizer
Signet Total Flow Panel Mount
ENTER
3-8150-1P
Terminal Cover
4-3111
Flow
Totalizer
3-8050 (3-8150-1)
Universal mounting kit
Integral
Integral System adapter Kit
3-8150-PO (3-8051)
3-8150-P1
3-8150-TO
3-8150-VO
Integral
Flow Sensor
(3-8510-XX)
George Fischer Signet, Inc. 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. Tel. (626) 571-2770 Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfsignet.com Or call (in the U.S.): (800) 854-4090
3-8150.090-1 Rev. D 8/05 English George Fischer Signet, Inc. 2001 Printed in U.S.A. on recycled paper