Artculo original
ABSTRACT
Keywords: Introduction and objectives: Blood pressure measurement methods and conditions are determinants of
Ambulatory blood pressure monitoring hypertension diagnosis. A recent British guideline recommends systematic 24-h ambulatory blood
Primary care pressure monitoring. However, these devices are not available at all health centers and they can only be
White coat hypertension
used by 1 patient per day. The aim of this study was to test a new blood pressure recording method to see
if it gave the same diagnostic results as 24-h blood pressure monitoring.
Methods: One-hour blood pressure monitoring under routine clinical practice conditions was compared
with standard method of day time recording by analyzing the coefcient of correlation and Bland-
Altman plots. The Kappa index was used to calculate degree of agreement. Method sensitivity and
specicity were also analyzed.
Results: Of the 102 participants, 89 (87.3%) obtained the same diagnosis regardless of method, with high
between-method agreement (k= 0.81; 95% condence interval, 0.71-0.91). We observed robust
correlations between diastolic (r = 0.85) and systolic blood pressure (r = 0.76) readings. Sensitivity and
specicity for the new method for diagnosing white coat hypertension were 85.2% (95% condence
interval 67.5%-94.1%) and 92% (95% condence interval, 83.6%-96.3%), respectively.
Conclusions: One-hour blood pressure monitoring is a valid and reliable method for diagnosing
hypertension and for classifying hypertension subpopulations, especially in white coat hypertension and
refractory hypertension. This also leads to a more productive use of monitoring instruments.
* Autor para correspondencia: Centro de Salud Les Corts (CAPSE), Meja Lequerica s/n, 08028 Barcelona, Espana.
Correo electronico: asiso@clinic.ub.es (A. Siso-Almirall).
0300-8932/$ see front matter 2013 Sociedad Espanola de Cardiologa. Publicado por Elsevier Espana, S.L.U. Todos los derechos reservados.
http://dx.doi.org/10.1016/j.recesp.2013.12.019
Documento descargado de http://www.revespcardiol.org el 21/02/2017. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
La hipertension arterial (HTA) es, junto con el tabaquismo y la Los participantes procedan de una poblacion de pacientes
hipercolesterolemia, uno de los tres factores de riesgo cardiovascu- atendidos en un centro de salud urbano. Los criterios de inclusion
lar mas frecuentes en Espana: su prevalencia en adultos > 18 anos fueron: edad > 18 anos e indicacion de MAPA en condiciones de
es de aproximadamente un 38% de los varones y un 28% de las practica clnica real ante tres situaciones clnicas (diagnostico
mujeres, proporcion que va en aumento en relacion con el de HTA de novo, sospecha clnica de HBB y seguimiento de la HTA
envejecimiento de la poblacion1,2. Desde el punto de vista macro- por sospecha clnica de HTA enmascarada o HR). Se asumieron
economico, los costes sanitarios directos asociados a la HTA como criterios de exclusion los mismos que consideran las guas
representaron un 2,6-3,9% del gasto sanitario total espanol3. Las del NICE7 y europeas8,9 (es decir, la presencia de brilacion
consecuencias de esta situacion se observan en la consultas de los auricular u otros trastornos del ritmo cardiaco que pudieran
centros de salud de atencion primaria, donde la HTA es uno de interferir en las lecturas) y la falta de colaboracion como criterio de
los motivos de consulta mas frecuentes3. exclusion adicional. Siguiendo las indicaciones de la Gua Practica
El diagnostico de la HTA se puede realizar mediante diferentes de Hipertension Arterial para Atencion Primaria25, se considero
metodos, que dieren en la frecuencia de toma de los valores de valido un resultado de MPA1 con un mnimo de lecturas correctas
presion arterial (PA). En Espana, PAPPS4 recomienda la triple igual al 70%. Al nalizar ambas pruebas, se excluyo a los
medicion de la PA en diferentes semanas en el centro de salud, tal y participantes cuyos porcentajes de lecturas correctas con MPA1
como se hace en la mayor parte del mundo. Sin embargo, este y MAPA fueron < 70%25,26.
metodo tiene limitaciones importantes: no reeja las verdaderas El periodo de reclutamiento fue de junio de 2011 a junio de
variaciones de la PA a lo largo del da, no permite descartar la 2012.
hipertension de bata blanca (HBB) subyacente y tiene escasa
capacidad predictiva del riesgo cardiovascular5,6. Procedimiento
Las recientes guas del NICE7 britanicas y la gua de practica
clnica en HTA de la Sociedad Europea de Hipertension Arterial y la Ambos metodos de medicion y registro (MPA1 y MAPA) se
Sociedad Europea de Cardiologa8,9 recomiendan la monitorizacion realizaron con el mismo dispositivo (SpacelabsW 90207)27. Dicho
ambulatoria de la presion arterial (MAPA) durante 24 h con un dispositivo incluye un manguito estandar que se adapta al brazo e
dispositivo electronico o, en su defecto, la automedicion de la insua aire a intervalos programables y una unidad portatil que
presion arterial domiciliaria (AMPA) como metodos validos para registra y almacena los datos durante esos intervalos.
el diagnostico y el seguimiento de la HTA. Mediante la AMPA, es el La MAPA se programo para obtener registros de PA cada 20 min
propio paciente quien realiza los registros de PA, lo que evita los durante el da y cada 30 min durante la noche28. Se recomendaba a
factores asociados a cambios bruscos de la PA. Sin embargo, la los participantes que realizaran sus actividades diarias habituales,
validez de este metodo esta afectada por la propia variabilidad de evitaran las que requirieran esfuerzo fsico intenso y se mantu-
la PA, la abilidad de la tecnica de medicion y la gran variedad vieran en reposo en el momento del registro. Para la comparacion
de estandares de calibracion de los dispositivos utilizados1014. En con MPA1, se tuvo en cuenta los valores registrados de PA con
cuanto a la MAPA, presenta entre sus ventajas el analisis de la MAPA en periodo diurno (de 8.00 a 20.00). Entre 24 y 48 h despues
variabilidad de los periodos diurno y nocturno de la PA1518 y del registro con MAPA, los participantes realizaron MPA1. En ese
aporta informacion sobre el patron de descenso nocturno y el intervalo se comprobo que no se produjeran cambios en las
aumento de la PA matinal relacionados con la enfermedad condiciones basales del participante. La medicion de la PA
cerebrovascular1921. Los resultados de la MAPA permiten mediante MPA1 se realizo a intervalos programados de 6 min
clasicar a los pacientes en diferentes subgrupos: normotension, durante 1 h; se desestimaron dos tomas adicionales al procedi-
HBB, hipertension enmascarada e hipertension refractaria (HR). miento, una inicial y otra al nal. Todas las MPA1 se aplicaron en
Entre los pacientes hipertensos de grado 1, la prevalencia de HBB horario matinal, entre las 9.00 y las 13.00. El paciente reciba la
oscila en torno al 15%18, y el efecto de bata blanca es una de las informacion acerca del estudio, se le informaba del procedimiento
principales indicaciones de la MAPA. A pesar de las ventajas de la en la consulta del medico de familia y se proceda al registro en una
MAPA, su implementacion sistematica en el primer nivel sala tranquila dentro del propio centro de salud.
asistencial queda restringida a solo un paciente por da, ademas
del elevado coste de los dispositivos validados.
Los estudios que han analizado la manera de obtener Variables de estudio
mediciones similares al registro de la MAPA reejan resultados
dispares2224. Por este motivo, el objetivo de este estudio ha sido La variable principal del estudio fue la clasicacion de los
validar un nuevo metodo de monitorizacion de la presion arterial pacientes en las subpoblaciones de HTA: normotension (PA
de 1 h (MPA1) de duracion utilizando el mismo dispositivo de consulta < 140/90 mmHg y MAPA o MPA1 < 135/85 mmHg);
Documento descargado de http://www.revespcardiol.org el 21/02/2017. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
HBB (PA consulta 140/90 mmHg y MAPA o MPA1 < 135/85 mmHg); utilizo la prueba de normalidad de Shapiro-Wilk para comprobar la
hipertension enmascarada (PA consulta < 140/90 mmHg y MAPA distribucion normal de las variables.
o MPA1 135/85 mmHg), y HR (PA consulta 140/90 mmHg y Se calculo el coeciente de correlacion de Pearson entre MPA1 y
MAPA o MPA1 135/85 mmHg). En el caso de no disponer del registro MAPA. Se aplico el metodo de Bland-Altman29 con la represen-
de PA en consulta, se utilizo la media de las ultimas tres tomas de tacion graca de las correlaciones de los intervalos (las diferencias
registros de AMPA mas recientes aportados por el paciente: esto entre las mediciones contra la media) para conrmar la
sucedio en 19 casos. independencia entre las diferencias obtenidas con cada metodo
Los datos de la variables sociodemogracas procedentes de la y la magnitud.
historia clnica fueron: edad y sexo, peso, talla e ndice de masa Se calcularon las prevalencias de HBB, hipertension enmascarada
corporal, antecedentes de factores de riesgo cardiovascular, habito y HR. Se estimo la proporcion de participantes bien clasicados con
tabaquico, consumo de alcohol, sedentarismo, diabetes mellitus, MPA1 respecto a MAPA y se calculo el ndice kappa30 para medir el
dislipemia y obesidad. grado de concordancia entre los dos metodos para clasicar a los
participantes en subpoblaciones de hipertensos. Se calcularon la
sensibilidad, la especicidad y los valores predictivos positivo y
Tamano muestral
negativo para el diagnostico de HBB, HR e hipertension enmasca-
rada. Todos los resultados de concordancia y correlacion se basaron
Asumiendo una proporcion de pacientes bien clasicados con
en los promedios de las lecturas de presion arterial sistolica (PAS) y
MPA1 respecto a MAPA en funcion de las subpoblaciones de HTA
presion arterial diastolica (PAD) de cada paciente.
alrededor del 90%22, con una precision del 5%, un nivel de conanza
Se llevo a cabo un analisis de sensibilidad para comparar las
del 90% y perdida del 20%, se estimo que sera necesario incluir en
caractersticas basales de los pacientes incluidos y excluidos
el estudio una muestra mnima de 98 participantes.
(< 70% de lecturas correctas con MPA1 y MAPA) en el analisis nal.
Se consideraron estadsticamente signicativos los valores de
Aspectos eticos p < 0,05. Se calcularon los intervalos de conanza del 95% (IC95%).
El analisis estadstico se realizo mediante el programa estadstico
Este estudio ha sido aceptado por el Comite Etico de R version 2.15.1 para Windows31.
Investigacion Clnica del Hospital Clnico, con numero de registro
2011/6743. Los pacientes que aceptaron participar fueron debi-
damente informados y manifestaron su aprobacion a traves del RESULTADOS
consentimiento informado. Los procedimientos utilizados siguie-
ron todas las clausulas de la Declaracion de Helsinki para la Caractersticas de los participantes
realizacion de estudios biomedicos y el respeto de los derechos
humanos. Se solicitaron en total 158 MAPA, de las que 123 (77,8%) se
incluyeron en el estudio. En la gura 1 se detalla el diagrama
de ujo con los motivos de exclusion. De los 102 pacientes
Analisis estadstico nalmente analizados, 45 (44,1%) eran mujeres, con medias de
edad de 60,2 13,6 anos e ndice de masa corporal de 27,5 3,9. Las
Las variables categoricas se presentan mediante frecuencia caractersticas basales de los pacientes en el momento de su inclusion
absoluta (%). Se proporciona la media desviacion estandar o la se muestran en la tabla 1. Las indicaciones de MAPA de los
mediana [intervalo intercuartlico] para las variables continuas. Se participantes fueron las siguientes: 49 (48,0%), diagnostico de HTA;
Excluidos (35)
Rechazaron participar (20)
Problemas tcnicos con el dispositivo (9)
Indicacin por el mdico de familia responsable (3)
Iniciar el tratamiento farmacolgico (3)
Excluidos (21)
< 70% de lecturas correctas con MPA1 (21) y
con MAPA diurna (4). Cuatro participantes
tuvieron < 70% de lecturas correctas
con los 2 mtodos.
Figura 1. Diagrama de ujo de la inclusion de participantes. MAPA: monitorizacion ambulatoria de la presion arterial; MPA1: monitorizacion de la presion arterial
de 1 h.
Documento descargado de http://www.revespcardiol.org el 21/02/2017. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
PAD PAS
120
180
100
140
80
60 100
I 12 24 36 48 60 F I 12 24 36 48 60 F
Figura 2. Distribucion de las lecturas de la presion arterial diastolica y la presion arterial sistolica con monitorizacion de la presion arterial de 1 h. Las lneas verdes
discontinuas representan los puntos de corte de normalidad para la prueba (presion arterial diastolica, 85 mmHg; presion arterial sistolica, 135 mmHg). La lectura
inicial (I) corresponde al momento de colocar el dispositivo y la lectura nal (F), al momento de retirarlo (test pre y post). PAD: presion arterial diastolica; PAS:
presion arterial sistolica. Esta gura se muestra a todo color solo en la version electronica del artculo.
Documento descargado de http://www.revespcardiol.org el 21/02/2017. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
PAD PAS
100 150
140
90
MAPA
MAPA
130
80
120
70
110
60 100
MPA1 MPA1
Figura 3. Comparacion de la presion arterial diastolica y la presion arterial sistolica entre monitorizacion de la presion arterial de 1 h y monitorizacion ambulatoria
de la presion arterial. Los puntos representan la media de presion arterial diastolica (o presion arterial sistolica) de los sujetos obtenida mediante los registros de
monitorizacion de la presion arterial de 1 h (eje horizontal) y los registros de monitorizacion ambulatoria de la presion arterial (eje vertical). Las lneas rojas indican
las correlaciones entre estos valores y r indica el valor del coeciente de correlacion. MAPA: monitorizacion ambulatoria de la presion arterial; MPA1:
monitorizacion de la presion arterial de 1 h; PAD: presion arterial diastolica; PAS: presion arterial sistolica. Esta gura se muestra a todo color solo en la version
electronica del artculo.
(IC95%, 76,2-94,4) y especicidad del 92,3% (IC95%, 81,8%-97,0%) el primer nivel asistencial. En este sentido, en la busqueda de
de la MPA1 para diagnosticar HR. La prevalencia observada de nuevos metodos de medicion de la PA que permitan simplicar la
hipertension enmascarada fue la mas baja (10,8%), con sensibilidad MAPA, un reciente estudio desarrollado en Pases Bajos22 ha
y especicidad de la MPA1 del 72,7% (IC95%, 43,4-90,3) y el 100,0% propuesto el registro de la PA durante 30 min a intervalos de 5 min
(IC95%, 96,0-100,0) respectivamente (tabla 3). en 84 sujetos comparado con la MAPA de 24 h. Dicho metodo
permitio clasicar de la misma manera al 87% de los sujetos,
DISCUSION resultados practicamente identicos a los observados en nuestro
estudio. Parece evidente que la medicion de la PA obtenida en
El metodo de MPA1 permite clasicar de la misma manera el periodos de registro simplicados es superponible a los resultados
87,3% de los casos respecto a la MAPA de 24 h. Ademas, la MPA1 obtenidos en la fase diurna de la MAPA. Sin embargo, hay cierta
obtiene correlaciones elevadas tanto en PAS (r = 0,76) como en PAD variabilidad y, por lo tanto, escaso consenso sobre cual debera ser
(r = 0,85) frente a la MAPA, y elevadas sensibilidad y especicidad el periodo mnimo de registro de PA que permita una concordancia
para los diagnosticos de HBB y HR. suciente respecto a un registro diurno de MAPA. Wong y Yeo32
Este nuevo metodo de registro, por su sencillez y reproduci- estudiaron a sujetos con alteraciones del sueno y descenso
bilidad, permitira optimizar el uso de los dispositivos de MAPA en nocturno de PA, pero asumiendo periodos de registro mas largos,
PAD PAS
15 20 Media + 2 DE = 19,4
Media + 2 DE = 13,1
10
10
Diferencia (MPA1 MAPA)
Media = 1,5
Media = 0,3 0
0
5
10
10
Media 2 DE = 12,6
20 Media 2 DE = 22,5
15
20 30
Figura 4. Bland-Altman plots; el eje horizontal representa la media de las determinaciones con los 2 metodos y el eje vertical, su diferencia. Las lneas verdes
discontinuas indican la media de las diferencias y las lneas rojas discontinuas, sus intervalos de conanza. DE: desviacion estandar; MAPA: monitorizacion
ambulatoria de la presion arterial; MPA1: monitorizacion de la presion arterial de 1 h; PAD: presion arterial diastolica; PAS: presion arterial sistolica. Esta gura se
muestra a todo color solo en la version electronica del artculo.
Documento descargado de http://www.revespcardiol.org el 21/02/2017. Copia para uso personal, se prohbe la transmisin de este documento por cualquier medio o formato.
CONCLUSIONES 12. Viera AJ, Hinderliter AL, Kshirsagar AV, Fine J, Dominik R. Reproducibility of
masked hypertension in adults with untreated borderline ofce blood pressure:
comparison of ambulatory and home monitoring. Am J Hypertens. 2010;23:
La MPA1 obtiene elevadas sensibilidad y especicidad para el 11907.
diagnostico de HBB y de HR. Este nuevo metodo se ha demostrado 13. Hanninen MR, Niiranen TJ, Puukka PJ, Jula AM. Comparison of home and
ambulatory blood pressure measurement in the diagnosis of masked hyper-
tan util como la MAPA de 24 h en el diagnostico y el seguimiento de tension. J Hypertens. 2010;28:70914.
diferentes subpoblaciones de HTA. Es previsible que su aplicabi- 14. Gaborieau V, Delarche N, Gosse P. Ambulatory blood pressure monitoring
lidad en la practica clnica real pueda convertirlo en un metodo versus self-measurement of blood pressure at home: correlation with target
organ damage. J Hypertens. 2008;26:191927.
diagnostico de referencia en cualquier nivel asistencial. 15. Martnez-Moya L, Garca-Noain A, Lobo-Escolar A, Gonzalvo-Liarte C, Martnez-
Berganza A, Ca-Gomez P. Inuence of siesta in the estimation of blood pressure
AGRADECIMIENTOS control in patients with hypertension. Hypertension. 2007;50:e14.
16. Bilo G, Giglio A, Styczkiewicz K, Caldara G, Maronati A, Kawecka-Jaszcz K, et al. A
new method for assessing 24-h blood pressure variability after excluding the
Agradecemos la colaboracion de los participantes (pacientes del contribution of nocturnal blood pressure fall. J Hypertens. 2007;25:205866.
Centro de Atencion Primaria Les Corts) y de todo el Equipo de 17. De la Sierra A, Redon J, Banegas JR, Segura J, Parati G, Gorostidi M, et al.
Prevalence and factors associated with circadian blood pressure patterns in
Atencion Primaria de Les Corts (Barcelona), pero especialmente al hypertensive patients. Hypertension. 2009;53:46672.
equipo de enfermera que ha hecho posible la realizacion de las 18. Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, et al. Recommen-
monitorizaciones. dations for blood pressure measurement in humans and experimental animals:
part 1: blood pressure measurement in humans: a statement for professionals
from the Subcommittee of Professional and Public Education of the American
Heart Association Council on High Blood Pressure Research. Circulation.
FINANCIACION 2005;111:697716.
19. Coll-de-Tuero G, Saez M, Roca-Saumell C, Rodriguez-Poncelas A, Franco P, Dalfo
Este trabajo se ha nanciado por medio de la Beca Fin de A, et al. Evolution of target organ damage by different values of self-blood
pressure measurement in untreated hypertensive patients. Am J Hypertens.
Residencia 2011 otorgada por el Consorci dAtencio Primaria de Salut 2012;25:125663.
de lEixample. 20. Andrade SS, Serro-Azul JB, Nussbacher A, Giorgi D, Pierri H, Gebara O, et al.
Daytime systolic blood pressure load and previous stroke predict cardiovascu-
lar events in treated octogenarians with hypertension. J Am Geriatr Soc.
CONFLICTO DE INTERESES 2010;58:22324.
21. Pickering T, Schwartz J, Verdecchia P, Imai Y, Kario K, Eguchi K, et al. Prediction
of strokes versus cardiac events by ambulatory monitoring of blood pressure:
Ninguno. results from an international database. Blood Press Monit. 2007;12:3979.
22. Van der Wel MC, Buunk IE, Van Weel C, Thien TA, Bakx JC. A novel approach to
ofce blood pressure measurement: 30-minute ofce blood pressure vs day-
BIBLIOGRAFIA time ambulatory blood pressure. Ann Fam Med. 2011;9:12835.
23. Culleton BF, McKay DW, Campbell NR. Performance of the automated BpTRU
measurement device in the assessment of white-coat hypertension and white-
1. Banegas JR, Graciani A, De la Cruz-Troca JJ, Leon-Munoz LM, Guallar-Castillon P, coat effect. Blood Press Monit. 2006;11:3742.
Coca A, et al. Achievement of cardiometabolic goals in aware hypertensive 24. Garca-Donaire JA, Dalfo Baque A, Sanclemente Anso C, Urdiales Castillo D,
patients in Spain: a nationwide population-based study. Hypertension. Martnez Deben F, Ortega Lopez N, et al. Medida de la presion arterial en
2012;60:898905. consulta y automatizada (BPTruW) para evaluar el efecto de bata blanca. Med
2. Baena Dez JM, Del Val Garca JL, Tomas Pelegrina J, Martnez Martnez JL, Martn Clin (Barc). 2012;138:597601.
Penacoba R, Gonzalez Tejon I, et al. Epidemiologa de las enfermedades car- 25. Societat Catalana de Medicina Familiar i Comunitaria. Gua de Practica de
diovasculares y factores de riesgo en atencion primaria. Rev Esp Cardiol. Hipertension Arterial para Atencion Primaria. 4.a ed. Barcelona: EdiDe; 2011.
2005;58:36773. 26. OBrien E, Asmar R, Beilin L, Imai Y, Mancia G, Mengden T, et al. Practice
3. Suarez C, Gabriel R. Epidemiologa de la hipertension arterial en Espana. guidelines of the European Society of Hypertension for clinic, ambulatory and
Cardiovascular Risk Factors. 2000;9:23946. self blood pressure measurement. J Hypertens. 2005;23:697701.
4. Maiques Galan A, Brotons Cuixart C, Villar Alvarez F, Navarro Perez J, Lobos- 27. OBrien E, Mee F, Atkins N, OMalley K. Accuracy of the SpaceLabs 90207
Bejarano JM, Ortega Sanchez-Pinilla R, et al; Grupo de Prevencion Cardiovas- determined by the British Hypertension Society protocol. J Hypertens.
cular del PAPPS. Recomendaciones preventivas cardiovasculares. Aten Primaria. 1991;9:5734.
2012;44 Supl 1:315. 28. OBrien E, Petrie J, Littler W, De Swiet M, Padeld PL, Altman DG, et al. An outline
5. Omboni S, Guarda A. Impact of home blood pressure telemonitoring and blood of the revised British Hypertension Society protocol for the evaluation of blood
pressure control: a meta-analysis of randomized controlled studies. Am J pressure measuring devices. J Hypertens. 1993;11:6779.
Hypertens. 2011;24:98998. 29. Bland JM, Altman DG. Statistical methods for assessing agreement between two
6. Banegas JR, Segura J, Sobrino J, Rodrguez-Artalejo F, De la Sierra A, De la Cruz JJ, methods of clinical measurement. Lancet. 1986;1:30710.
et al. Effectiveness of blood pressure control outside the medical setting. 30. Fleiss JL. Statistical methods for rates and proportions. 2.a ed. Nueva York:
Hypertension. 2007;49:628. Wiley; 1981.
7. Krause T, Lovibond K, Cauleld M, McCormack T, Williams B; Guideline Devel- 31. R Development Core Team. R: a language and environment for statistical
opment Group. Management of hypertension: summary of NICE guidance. BMJ. computing. Viena: R Foundation for Statistical Computing; 2012.
2011;343:d4891. 32. Wong RC, Yeo TC. Ofce-hour ambulatory blood pressure monitoring is
8. Mancia G, Fagard R, Narkiewicz K, Redon J, Zanchetti A, Bohm M, et al. 2013 ESH/ sufcient for blood pressure diagnosis. J Hum Hypertens. 2006;20:4403.
ESC Guidelines for the management of arterial hypertension: The Task Force for 33. Sheps SG, Bailey KR, Zachariah PK. Short-term (six hour), ambulatory blood
the management of arterial hypertension of the European Society of Hyperten- pressure monitoring. J Hum Hypertens. 1994;8:8738.
sion (ESH) and of the European Society of Cardiology (ESC). J Hypertens. 34. Stergiou GS, Nasothimiou EG, Omboni S, De la Sierra A, Ruilope LM, Parati G.
2013;31:1281357. Diagnostic accuracy of home blood pressure monitoring: the MEDIT-HABP
9. Mancia G, Fagard R, Narkiewicz K, Redon J, Zanchetti A, Bohm M, et al. Gua de Study (Mediterranean Home vs. Ambulatory BP). J Hypertens. 2011;29 Suppl
practica clnica de la ESH/ESC 2013 para el manejo de la hipertension arterial. A:e83.
Grupo de Trabajo para el manejo de la hipertension arterial de la Sociedad 35. Myers MG, Valdivieso M, Kiss A. Use of automated ofce blood pressure
Europea de Hipertension (ESH) y la Sociedad Europea de Cardiologa (ESC) Rev measurement to reduce the white coat response. J Hypertens. 2009;27:2806.
Esp Cardiol. 2013;66:880.e164. 36. De la Sierra A, Segura J, Banegas JR, Gorostidi M, De la Cruz JJ, Armario P, et al.
10. Hodgkinson J, Mant J, Martin U, Guo B, Hobbs FD, Deeks JJ, et al. Relative Clinical features of 8295 patients with resistant hypertension classied on
effectiveness of clinic and home blood pressure monitoring compared with the basis of ambulatory blood pressure monitoring. Hypertension. 2011;57:
ambulatory blood pressure monitoring in diagnosis of hypertension: system- 898902.
atic review. BMJ. 2011;342:d3621. 37. Gutierrez-Misis A, Sanchez-Santos MT, Banegas JR, Zunzunegui MV, Sanchez-
11. Stergiou GS, Bliziotis IA. Home blood pressure monitoring in the diagnosis and Martnez M, Castell MV, et al. Asociacion entre presion arterial y mortalidad en
treatment of hypertension: a systematic review. Am J Hypertens. 2011;24: una cohorte de individuos de edad igual o superior a 65 anos de Espana: un
12334. modelo dinamico. Rev Esp Cardiol. 2013;66:46471.