You are on page 1of 8

See

discussions, stats, and author profiles for this publication at: http://www.researchgate.net/publication/12072668

Evaluation of a sensor for low interface pressure


applications

ARTICLE in MEDICAL ENGINEERING & PHYSICS DECEMBER 2000


Impact Factor: 1.83 DOI: 10.1016/S1350-4533(00)00080-1 Source: PubMed

CITATIONS READS

57 17

3 AUTHORS, INCLUDING:

Martin Ferguson-Pell
University of Alberta
63 PUBLICATIONS 969 CITATIONS

SEE PROFILE

All in-text references underlined in blue are linked to publications on ResearchGate, Available from: Martin Ferguson-Pell
letting you access and read them immediately. Retrieved on: 21 December 2015
Medical Engineering & Physics 22 (2000) 657663
www.elsevier.com/locate/medengphy

Technical note
Evaluation of a sensor for low interface pressure applications
a,*
Martin Ferguson-Pell , Satsue Hagisawa b, Duncan Bain a

a
Centre for Disability Research and Innovation, Institute of Orthopaedics and Musculo-Skeletal Sciences, Royal Free and University College
Medical School, Brockley Hill, Stanmore, Middlesex HA7 4LP, UK
b
School of Nursing, Oita Medical University, Hasama, Oita 879-5593, Japan

Received 16 August 2000; received in revised form 21 November 2000; accepted 7 December 2000

Abstract

An ultra-thin, small sensor has recently been developed, FlexiForce (Tekscan, Boston, MA, USA), which may be effective for
the measurement of low interface pressure between the skin, support surfaces and pressure garments.
To evaluate the suitability of the sensor for these applications, drift, repeatability, linearity, hysteresis and curvature effects were
tested under laboratory conditions.
The drift was 1.72.5%/logarithmic time, the repeatability was 2.36.6% and the linearity was 1.99.9% in the range of forces
of 1050 g applied. The hysteresis was 5.4% on average. The output offset of the sensor increased with decreasing radius of
curvature for radii less than 32 mm compared with a flat surface when no pressure was applied. The sensitivity to pressure decreased
with curvature for radii less than 32 mm.
It was found that the sensor had acceptable drift, repeatability, linearity and hysteresis. However, a significant curvature effect
was observed indicating that the sensor is suitable for direct measurement on surfaces with the radii greater than 32 mm under
static conditions. 2001 IPEM. Published by Elsevier Science Ltd. All rights reserved.

Keywords: Low interface pressure; Pressure garments; Pressure sensor calibration; Curvature effect

1. Introduction pressure, and conform to the curvature of the skin sur-


face.
External pressures applied to the surface of the skin The accuracy of interface pressure measurements
are commonly employed to reduce oedema in the relies upon sensor characteristics and measurement pro-
residual limb following amputation [1]. Similarly, press- cedures which may give rise to a variety of errors [3].
ure garments are used to prevent and reduce hyper- Specifications for the ideal sensor to measure the inter-
trophic scars following burn injury [2]. In both these face stresses associated with the skin of the human body
examples it is probable that application of a specific were proposed by Ferguson-Pell [4] and Grant [5]. Sen-
range of pressures can expedite and optimise the out- sors should be:
come. Furthermore, prescription of garment size based
on objective measurements of interface pressure is more (a) small and thin, e.g. 1 mm in thickness, 10 mm
likely to provide a consistent outcome for different in diameter, and highly flexible;
patients. (b) have a continuous output;
In order to measure the interface pressure between (c) be able to measure shear as well as normal
support surfaces, pressure garments and the skin, the forces;
sensor should be small, thin, flexible and sensitive to low (d) be free from error of measurement on curved
surfaces and from the effects of temperature
and moisture;
(e) low cost.
* Corresponding author. Tel.: +44-20-8909-5471; fax: +44-20-
8385-7151. With regard to specification (d), previous studies have
E-mail address: m.ferguson-pell@ucl.ac.uk (M. Ferguson-Pell). emphasized that curvature effects must be taken into

1350-4533/01/$ - see front matter 2001 IPEM. Published by Elsevier Science Ltd. All rights reserved.
PII: S 1 3 5 0 - 4 5 3 3 ( 0 0 ) 0 0 0 8 0 - 1
658 M. Ferguson-Pell et al. / Medical Engineering & Physics 22 (2000) 657663

account when using the transducers [47]. Bain [8] processes it, and sends it to the computer through a serial
examined this effect using a Talley pneumatic transducer port. Software is supplied to control the sampling rate,
loaded by a hemispherical dome at different radii (19 gain and displays data dynamically. The system is shown
87 mm) and reported that curvature had a large influence in Fig. 1.
on the output of the transducer. A similar study by Buis
and Convery [9] reported that spherical curvatures intro- 3.2. Test apparatus
duced significant reductions in the outputs using force
sensing resistors (#9810, Tekscan, Boston, MA, USA). Fig. 2 represents a loading rig to test for drift, repeat-
In recent years new types of sensor have been ability, linearity and hysteresis. A low friction tilt table
developed for the measurement of interface pressure. was used to apply axial load to the sensor. The sensor
Grant [5], Wytch et al. [6] and Olson [10] summarize was placed in a sandwich between a plastic flat surface
the characteristics of the sensors. However, until now no and a disc made from polyurethane (PU), 9 mm in diam-
sensor has met all the above specifications. eter and 3 mm in thickness, attached to another plastic
An ultra-thin, small (0.1 mm in thickness, 9.5 mm in flat surface. The weight of the table was counterbalanced
diameter) and flexible sensor, FlexiForce, has recently using fixed weights, about an edge fulcrum. This con-
been developed by Tekscan. The specifications of this strained the applied force to a single degree of freedom,
device indicate that it may be suitable for the measure- in an arc closely approximating the vertical. Deviation
ment of low interface pressure between the skin, support from the vertical due to the arc effect during the deflec-
surfaces and pressure garments. tion of the assembly comprising sensor and disc was cal-
culated as being less than 0.05. Deviation from a
straight line projection was calculated as being less than
2. Purpose 1 m. The simple edge fulcrum design provided near-
zero hysteresis due to friction, while effectively guiding
The aim of this study was to evaluate the FlexiForce the applied weights.
sensor for the measurement of interface pressure, parti- Conditioning of the sensor was performed by apply-
cularly in more challenging applications such as beneath ing a weight of 60 g, four times prior to testing, as rec-
bandages, compression stockings and pressure garments, ommended by the manufacturers. Data samples were
where pressure is relatively low. obtained at 8 Hz for 10 s.
Readings were taken after achieving a stable level,
mostly 1 min after loading. Room temperature was kept
3. Method at 2425C, which is close to the surface temperature of
skin of the lower limb.
3.1. The FlexiForce sensor

The FlexiForce sensor employs an ink, the electrical 3.2.1. Drift test
resistance of which varies with applied force. The active Drift is the change in sensor output when a constant
area of the sensor is produced lithographically with con- force is applied over a period of time. Its was tested by
ductive strips connecting them to electronic conditioning applying weights of 50 and 30 g to the sensor for 2 h.
circuitry. The resistance change of the sensors varies Readings were taken every 5 min for the first 20 min
inversely with the applied force. then every 10 min up to 2 h.
The sensor is constructed from two layers of a poly-
ester film. On each layer, a conductive material (silver) 3.2.2. Repeatability test
is applied, followed by a layer of pressure-sensitive ink. Repeatability is the ability of the sensor to respond in
The two layers are laminated together to form the sensor. the same way to the same applied force for successive
In this study ultra-sensitive sensors operating in the loadings. It was tested by applying weights of 50, 30
range 00.45 kg were employed. Sensors operating over and 10 g to the sensor every 5 min and readings were
higher force ranges (11, 68, 55 kg) are available rou- taken. This procedure was repeated 20 times.
tinely but were not tested in this study. Since the inter-
face pressure reported for stump bandages and pressure 3.2.3. Linearity test
garments are relatively low, 2030 mmHg [2,11], a Linearity is a measure of the proportionality of the
pressure of 50 mmHg was used as a maximum pressure sensors response to the applied load over the range of
for the evaluation. The active area of the sensor is loading. A weight of 50 g was applied to the sensor and
7.122105 m2, therefore, a force of 50 g corresponds to readings were taken 1 min after the loading. This pro-
a pressure of 51.6 mmHg. In this study ultra-sensitive cedure was repeated 10 times.
sensors operating in the range 00.45 kg were employed. Similarly, tests using weights of 40, 30, 20 and 10 g
A small sensor-handle gathers data from the sensor, were carried out.
M. Ferguson-Pell et al. / Medical Engineering & Physics 22 (2000) 657663 659

Fig. 1. FlexiForce sensor system.

sensor by inflation. The apparatus used in this test is


provided diagrammatically in Fig. 3.
Readings were taken 1 min after loading with press-
ures of 0, 10, 20, 30, 40 and 50 mmHg. These procedures
were repeated five times for each pipe.

4. Data analysis

The data were saved in ASCII data format and pro-


cessed using Microsoft EXCEL version 7.0 to calcu-
late the mean and standard deviation (SD) and to rep-
resent the data graphically.
Drift was calculated by dividing the gradient of linear
Fig. 2. Sensor calibration rig used for drift, repeatability, linearity
and hysteresis tests.
regression of log-time vs. output by the starting value of
output. For repeatability and linearity, the coefficient of
variation was calculated by dividing the SD by the mean.
For the hysteresis test, the biggest difference in the read-
3.2.4. Hysteresis test ings between increasing and decreasing output values at
Hysteresis is the difference in the sensor output the same applied force was taken and divided by the
response during increased loading and decreased loading maximum output reading in the range of loading. For
at the same force. Its was examined by increasing and
decreasing the applied force consecutively, starting from
0 g up to 50 g, with 10-g increments and then decreasing
back to 0 g. The test was repeated 10 times.

3.2.5. Curvature test


The curvature effect indicates whether the reading
from the sensor is affected by a curved surface. The sur-
face may have cylindrical or spherical curvature,
depending on the intended application. ABS cylindrical
pipes with different radii of curvature (8.0, 10.4, 13.6,
16.1, 32.5 and 51.7 mm) were used for the test. The
sensor was carefully taped inside of the pipe using 3M
Micropore without generating extra tension on the
sensor and a condom connected to a pressure gauge and
pump was placed in the pipe to apply the pressure to the Fig. 3. Sensor calibration device used for curvature test.
660 M. Ferguson-Pell et al. / Medical Engineering & Physics 22 (2000) 657663

the curvature test, the offset, which is the output when ation at each force were 1.9%, 3.8%, 4.0%, 6.0% and
the pressure is 0 mmHg, and the sensitivity to pressure, 9.9%, respectively.
which is represented as the gradient of the linear
regression of applied pressure vs. output, were examined 5.4. Hysteresis
to evaluate how these are influenced by a radius of cur-
vature compared with a flat surface. The meanSD of the maximum difference in output
for increasing and decreasing loads was 5.42.5%. A
typical example of the data is shown in Fig. 5.
5. Results
5.5. Curvature
5.1. Drift
When the pressures were applied to the curved sur-
Weights of 50 and 30 g were applied to the sensor face, the outputs of the sensor changed not only the off-
for 2 h. The output reading started to increase immedi- set but also the sensitivity, as indicated in Fig. 6.
ately upon application of the load and at 5 min had The offset on a flat surface was 0 g, however, the
reached a level that was 97.3% and 95.3% of the stable offset outputs with different radii increased with decreas-
values for 50 and 30 g. At 10 min the readings for 50 ing radius. The offset vs. curvature for radii in the range
and 30 g reached 98.5% and 97.6% of the stable 8.0100.0 mm are presented graphically in Fig. 7.
values, respectively. The sensitivity to pressure on a flat surface was 1.19
The drift for 50 and 30 g was 1.7% and g/mmHg. Changes in sensitivity with curvature for radii
2.5%/logarithmic time, respectively. in the range 8.051.7 mm are illustrated in Fig. 8.

5.2. Repeatability
6. Discussion
When weights of 50, 30 and 10 g were applied, the
mean and SD of the readings for 20 measurements were According to Tekscans literature, drift is less than
45.651.03, 26.740.86 and 7.280.48 g, respectively. 3%/logarithmic time, repeatability is within 5%, linearity
The coefficient of variation for 50, 30 and 10 g was is within 5%, hysteresis is less than 4.5% of full scale
2.3%, 3.2% and 6.6%, respectively. when 50% of the full force range is applied. There is no
information on curvature effects.
5.3. Linearity When comparing our results with the literature, drift
was 1.72.5%/logarithmic time; repeatability was 2.3
Fig. 4 indicates the results of the linearity test. The 6.6%; linearity was 1.99.9%, however, when the read-
mean and SD of readings for 50, 40, 30, 20 and 10 g ings were taken 10 min following loading, the coefficient
were 47.190.89, 37.511.45, 27.531.16, 17.711.07 of variation became much smaller (2.34.3%); hysteresis
and 7.370.73 g, respectively. The coefficients of vari- was 5.4% on average.

Fig. 4. The test results for linearity. Bars indicate SD.


M. Ferguson-Pell et al. / Medical Engineering & Physics 22 (2000) 657663 661

Fig. 5. Typical data for the hysteresis test. Bars indicate SD.

Fig. 6. The sensor response for surfaces with different radii of curvature. Bars indicate SD.

The drift test in this study indicated that there was an degradation in accuracy is permissible and the compat-
increase in sensor output with time, achieving approxi- ible calibration methods are used.
mately 96% and 98% of the stable level at 5 and 10 min The repeatability, linearity and hysteresis are accept-
following loading, respectively. Although FlexiForce able. The hysteresis of the FlexiForce sensor has been
provides continuous real-time output, the values substantially improved compared with earlier Tekscan
presented immediately after loading are smaller than the sensors in which hysteresis was 21.7% [12]. In the lin-
stable values from the results of the drift and linearity earity test, the coefficient of variation was higher at
tests described above. This suggests that the sensor is lower forces (9.9% at 10 g) compared with higher forces
best suited for measurements under static conditions. (1.9% at 50 g). This may be due to the limited sensitivity
Dynamic measurements are possible in situations where of the sensor at such low forces, suggesting that the out-
662 M. Ferguson-Pell et al. / Medical Engineering & Physics 22 (2000) 657663

Fig. 7. The effect of radius curvature on the offset output. Bars indicate SD.

Fig. 8. The effect of radius of curvature on sensitivity to pressure applied.

put obtained from a force of less than 10 g may not substantially. Combining these curvature effects (offset
be accurate. and sensitivity) it is apparent that the smallest radius of
The curvature test generated significant inaccuracies curvature recommended is approximately 32 mm and
in the output of data at small radii of curvature. When that the error is dominated by the offset effect. However,
the offset reading was plotted against radius of curva- if the offset at 0 mmHg is subtracted from all readings,
ture, an inverse relationship was found (Fig. 7). For cur- then the device can be used for radii down to 16 mm
vatures with a radius of less than 32.5 mm the offset where the sensitivity effect becomes significant.
output increased with decreasing radius. From these Therefore, it was concluded that drift, repeatability,
results the acceptable range in radius of curvature in linearity and hysteresis of the sensor used in this study
terms of offset would be for radii greater than approxi- are acceptable and the sensor is suitable on flat and
mately 32 mm. The output sensitivity to pressure was curved surfaces with the radii greater than approximately
also calculated for radii of curvature in the range of flat 32 mm under the static conditions.
to 8.0 mm and was found to be unaffected by curvatures Further study is needed to evaluate the performance of
down to approximately 32 mm. For radii of curvature the sensor on spherical surfaces and also how the sensor
less than approximately 32 mm the sensitivity decreased performs when its total area is not completely loaded.
M. Ferguson-Pell et al. / Medical Engineering & Physics 22 (2000) 657663 663

Acknowledgements [5] Grant LJ. Interface pressure measurement between a patient and
a support surface. CARE Br J Rehab Tissue Viability
1985;1(1):79.
We would like to thank Mr Peter Lennon for provid- [6] Wytch R, Neil G, Kalisse C. Skin-orthosis interface pressure
ing technical support for this study. This study was transducers. CARE Sci Practice 1989;7(4):1004.
funded in part by the Engineering and Physical Sciences [7] Veraart JCJM, Pronk G, Neumann HAM. Pressure difference of
Research Council, JREI Programme Grant No. elastic compression stockings at the ankle region. Dermatol Surg
1997;23:9359.
GR/L86708/01, the Royal National Orthopaedic Hospi- [8] Bain DS. Development of a phantom for the assessment of patient
tal NHS Trust and ASPIRE. support systems for the prevention of pressure sores. PhD thesis,
University of Surrey, 1997.
[9] Buis AWP, Convery P. Calibration problems encountered while
monitoring stump/socket interface pressures with force sensing
References resistors: techniques adopted to minimise inaccuracies. Prosthet
Orthot Int 1997;21:17982.
[1] Lambert A, Johnson J. Stump shrinkers: a survey of their use. [10] Olson J. Conventional pressure sensors. In: Webster JG, editor.
Phys Ther 1995;81(4):2346. Prevention of pressure sores. Bristol: Adams Hilger, 1991:121
[2] Mann R, Yeong EK, Moore ML, Engrav LH. A new tool to meas- 41.
ure pressure under burn garments. J Brun Care Rehabil [11] Isherwood PA, Robertson JC, Rossi A. Pressure measurements
1997;18(2):1603. beneath below-knee amputation stump bandages: elastic bandag-
[3] Beebe D. Accuracy of pressure and shear measurement. In: ing, the Puddifoot dressing and a pneumatic bandaging technique
Webster JG, editor. Prevention of Pressure Sores. Bristol: Adams compared. Br J Surg 1975;62:9826.
Hilger, 1991:15574. [12] Ferguson-Pell M, Cardi M. Prototype development and compara-
[4] Ferguson-Pell MW. Design criteria for the measurement of press- tive evaluation of wheelchair pressure mapping system. Assist
ure at body/support interfaces. Eng Med 1980;9(4):20914. Technol 1993;5:7891.

You might also like