You are on page 1of 70

Application Note

Onsite PD measurement on MV and HV cables

Author
Udo Ranninger | Udo.Ranninger@omicronenergy.com
Daniel Gebhardt | Daniel.Gebhardt@omicronenergy.com

Date
Sep 30, 2016

Related OMICRON Product


MPD 600, MPD 540, MCT 085, MCT 100, MCT 120, MCC 205, MCC 210, CAL 542, CPL 542, MCU 502,
UCS1

Application Area
MV and HV cable

Keywords
High Voltage Test, Cable, Fault Localization, Coupling Methods

Version
V1.2

Document ID
ANP_16004_ENU

Abstract
This application note gives an overview of PD measurement on medium- and high-voltage cables and their
accessories using different coupling methods. Basic theoretical information and fault localization techniques
are also explained.
General information

OMICRON electronics GmbH including all international branch offices is henceforth referred to as
OMICRON.
The product information, specifications, and technical data embodied in this application note represent the
technical status at the time of writing and are subject to change without prior notice.
We have done our best to ensure that the information given in this application note is useful, accurate and
entirely reliable. However, OMICRON does not assume responsibility for any inaccuracies which may be
present.
OMICRON translates this application note from the source language English into a number of other
languages. Any translation of this document is done for local requirements, and in the event of a dispute
between the English and a non-English version, the English version of this note shall govern.
All rights including translation are reserved. Reproduction of any kind, for example, photocopying,
microfilming, optical character recognition and/or storage in electronic data processing systems, requires the
explicit consent of OMICRON. Reprinting, wholly or partly, is not permitted.
OMICRON 2016. All rights reserved. This application note is a publication of OMICRON.

OMICRON 2016 Page 2 of 70


Table of content

1 Safety instructions ................................................................................................................................5


1.1 Operator qualifications ....................................................................................................................5
1.2 Safety standards and rules .............................................................................................................5
1.2.1 Safety standards ............................................................................................................................... 5
1.2.2 Safety rules ....................................................................................................................................... 5
1.3 Orderly measures ...........................................................................................................................6
1.4 Disclaimer .......................................................................................................................................6
2 General Information about MV/HV cables ...........................................................................................7
2.1 Overview .........................................................................................................................................7
2.1.1 Design code (VDE) ........................................................................................................................... 8
2.1.2 Main components of a crosslinked polyethylene cable (XLPE) ......................................................... 8
2.2 High frequency propagation along a cable .................................................................................. 12
2.2.1 Measurement frequency selection .................................................................................................. 14
2.3 Essential information prior to a PD cable measurement ............................................................. 17
3 Cable accessories and how to decouple the PD signals ............................................................... 18
3.1 End terminations .......................................................................................................................... 18
3.1.1 Coupling methods used at end terminations ................................................................................... 20
3.2 Cross Bonding Link Box ........................................................................................................... 26
3.2.1 Sheath Voltage Limiter - SVL .......................................................................................................... 27
3.2.2 Coupling methods used at link boxes ............................................................................................. 28
3.3 Joint ............................................................................................................................................. 29
3.3.1 Coupling methods used at joints ..................................................................................................... 30
4 Test setup, calibration and software settings ................................................................................. 31
4.1 Coupling capacitor ....................................................................................................................... 31
4.1.1 Connection, calibration and software settings ................................................................................. 32
4.1.2 What type of quadripole is needed?................................................................................................ 35
4.2 HFCT ........................................................................................................................................... 36
4.2.1 Connection and software settings ................................................................................................... 37
4.2.2 Saturation of HFCTs (MCT 100) ..................................................................................................... 39
4.3 UCS1 (UHF) ................................................................................................................................ 40
5 Phase synchronized measurement without coupling capacitor ................................................... 44
6 Typical PRPD patterns on cables ..................................................................................................... 46
6.1 Corona ......................................................................................................................................... 46
6.2 Surface Discharge ....................................................................................................................... 48
6.3 Void .............................................................................................................................................. 49

OMICRON 2016 Page 3 of 70


6.4 Developing fault in XLPE ............................................................................................................. 50
6.5 Floating potential ......................................................................................................................... 51
7 Fault localization on cables ............................................................................................................... 52
7.1 TDR Time Domain Reflectometry............................................................................................. 52
7.2 STDR Statistical Time Domain Reflectometry .......................................................................... 54
7.3 STDR Dual End measurement ................................................................................................. 58
8 Measurement example ....................................................................................................................... 59
8.1 Test Object and general information ........................................................................................... 59
8.2 What does the setup look like? .................................................................................................... 59
8.3 What can be expected? ............................................................................................................... 60
8.4 FFT background noise .............................................................................................................. 60
8.5 PD calibration and cable length determination ............................................................................ 61
8.6 Voltage calibration ....................................................................................................................... 63
8.7 Voltage sweep / PD measurement .............................................................................................. 63
8.8 Fault localization .......................................................................................................................... 64
9 List of literature................................................................................................................................... 69

OMICRON 2016 Page 4 of 70


1 Safety instructions

1.1 Operator qualifications

Working on high-voltage assets can be extremely dangerous. Consequently, only personnel qualified, skilled
and authorized in electrical engineering and trained by OMICRON are allowed to handle the OMICRON
measurement equipment. Before starting to work, clearly establish the responsibilities. Personnel receiving
training, instructions, directions, or education on the measurement equipment must be under constant
supervision of an experienced operator while working with the equipment. The operator is responsible for the
safety requirements during the whole procedure.

1.2 Safety standards and rules

1.2.1 Safety standards


Handling OMICRON measurement equipment must comply with the internal safety instructions and
additional safety-relevant documents.
In addition, observe the following safety standards, if applicable:
EN 50191 (VDE 0104) "Erection and Operation of Electrical Test Equipment"
EN 50110-1 (VDE 0105 Part 100) "Operation of Electrical Installations"
IEEE 510 "IEEE Recommended Practices for Safety in High-Voltage and High-Power Testing"
Moreover, observe all applicable regulations for accident prevention in the country and at the site of
operation.
Before installing OMICRON measurement equipment and its system components, read the safety
instructions in this user manual carefully.
Do not handle OMICRON measurement equipment without understanding the safety information in this
manual. If you do not understand some safety instructions, please contact OMICRON before proceeding.
Maintenance and repair of OMICRON measurement equipment and its accessories are only permitted by
qualified experts at OMICRON service centers.

1.2.2 Safety rules


Always observe the five safety rules:
Disconnect completely.
Secure against re-connection.
Verify that the installation is dead.
Carry out grounding and short-circuiting.
Provide protection against adjacent live parts.

OMICRON 2016 Page 5 of 70


1.3 Orderly measures

The user manuals of OMICRON measurement equipment or alternatively the e-book has always to be
available on the site where the equipment is operated.
The users of OMICRON measurement equipment must read this manual before installing the equipment and
they must observe the safety, installation, and operation instructions therein.
The OMICRON measurement equipment and its accessories may be used only as described in this user
manual. Any other use is not in accordance with the regulations. The manufacturer and the distributor are
not liable for damage resulting from improper usage. The user alone assumes all responsibility and risk.
Opening OMICRON measurement equipment or its accessories invalidates all warranty claims.

1.4 Disclaimer

If the equipment is used in a manner not specified by the manufacturer, the protection provided by the
equipment may be impaired.

OMICRON 2016 Page 6 of 70


2 General Information about MV/HV cables

2.1 Overview

Due to different application areas and historical reasons, several types of MV and HV cables, each with their
advantages and disadvantages are in use out in the field. The most common types are impregnated paper
three core single sheath (MV); belted (MV); external gas pressure; internal gas pressure and thermoplastic-
(XLPE, EPR) cables, where the core conductor and sheath exist in different designs and materials, including
single or multi core, cooper, aluminium, lead, etc.

Cable type Pro Contra Where used

Good experience for a long time Moisture sensitive


Immune against single faults due to Complex assembling process
Impregnated paper dielectric and self-healing
Not suitable for big height difference 1-400 kV
(tan ~ 0,3-0,9%) Good thermal properties (up to 80C)
High weight
Good electric properties (r ~ 3,5)

High elastic strength


XLPE / VPE
Good thermal properties (up to 90C) Complex production process
(tan ~ 0,05%) 1-500 kV
Good electric properties (r ~ 2,3) Flammable
Simple assembling process

EPR
Very flexible 1-110 kV
(tan ~ 0,5%) Low tensile strength
Large permissible temperature range Special cases
(r ~ 3)

Table 1: Overview - insulation material

OMICRON 2016 Page 7 of 70


2.1.1 Design code (VDE)
The design code of a cable is used to provide a basic information like core conductor and insulation material,
cable design or voltage level within one term. An example of a VDE design code is given in Figure 1.

Figure 1: Design code example of a typical MV cable

2.1.2 Main components of a crosslinked polyethylene cable (XLPE)


The insulation of an extruded plastic cable does not consist of liquids which can fill up cavities or barriers
that can stop the growing of PD channels, as is the case in paper cables. In case of PD activity, it is only a
matter of time until the insulation fails. The main difference of XLPE compared to PE is the cross connection
of the C2H4 chains, whereby higher operation temperatures can be achieved.

Figure 2: Simplified construction of a XLPE cable

OMICRON 2016 Page 8 of 70


Core conductor
Design and size of the core conductor mainly depend on the nominal current (losses due to increase of AC
resistance) and the used dielectric. Aluminium (light, flexible, cheaper) and copper (good specific resistance,
smaller, more robust, more expensive) are typical core conductor materials.

Figure 3: Overview of core conductor types and some examples

Type Pro Contra Where used

Homogeneous surface Inflexible


Single core spherical Up to 800mm
Cheap Huge current suppression
conductor aluminum core
Good filling factor (skin-effect)

Multi core spherical Flexible


Current suppression Up to 630mm
conductor Simple setup

Profile conductor Stabile Partly flexible


PE and oil cable
waveguide Homogeneous surface High current suppression

Segment conductor Very flexible Complex Big cross section


waveguide Low current suppression Bad filling factor Waveguide for oil cable

Flexible
Multi core oval conductor Current suppression External gas pressure
Simple setup

Table 2: Characteristic of different core conductors

OMICRON 2016 Page 9 of 70


Semi conductive layer
Semi conductive layers adjacent to the core conductor and the sheath are used to smoothen the field
distribution and to prevent areas with elevated electric fields.

Figure 4: Manage electrical field using semi conductive layers

Sheath
The metallic sheath of an MV or HV cable is used as the return path for the capacitive load and fault current.
It is a protection against accidental contact, electromagnetic shielding and it builds a basic mechanical
protection. Beside the electrical sheath, an additional armor is often used to protect the cable from external
mechanical forces (Except when using the steel tube technique).
A layer sheath is a combination of wires and conductive tape.
Depending on the cable type, additional layers build the total cable construction. As PE for instance is not
resistant to water, the cable sheath should be totally water tight. Often there is an additional fleece
preventing longitudinal water distribution.

Type Pro Contra Where used

Light
Cooper wire sheath
Small cable diameter Low mechanical protection PE - cable
Aluminium layer sheath
Longitudinal watertight

PE - cable
No corrugation essential
Heavy Single wire of external
Lead sheath Small cable diameter
Pressure protection (oil cable) gas pressure cable
Longitudinal watertight
Low pressure oil cable

Light PE - cable
Big cable diameter due to
Good mechanical protection corrugation Single core gas
Aluminium wave sheath
pressure cable
High short current Complex longitudinal water tight
conductivity Low pressure oil cable

High mechanical protection fix gas pressure cable


Steel tube
good reduction factor huge building effort high pressure oil cable

Table 3: Characteristic of different sheath types

OMICRON 2016 Page 10 of 70


Common faults causing PD in a XLPE cable
There are three main failure mechanisms that exist in XLPE cables: Cavities, water/electrical trees and
protrusions.
In case of a cavity, the lower permittivity r within the gas-filled cavity leads to a higher electric field strength
in this specific area within the insulation. Once the local electric field strength exceeds the dielectric strength
of the insulation, the consequence will be overstressing of the insulation leading to partial discharge.
Water can enter into the insulation system once the increased water protection is damaged. Due to the
water vapour pressure, water can diffuse into the XLPE. A water tree will start growing towards the core
conductor as the water molecules are polarized. It will grow in the direction of the electrical field. The field
increases due to the higher electrical conductance and the decreased insulation distance. A water tree
cannot be detected via a PD measurement, however it can lead to an electrical tree once the electric field
strength is too high.
A defect in the inner or outer semi conductive layers, like a tip, leads to an inhomogeneous electrical field.
The elevated field strength, once exceeding the dielectric strength of the insulation, will cause partial
discharge and the evolution of an electrical tree.

Figure 5: Common faults causing PD activity within a cable, electric field lines are shown in blue

Figure 6: Electrical tree growing from the main conductor towards the counter electrode. For the photo the XLPE has been heated to
become clear (around 115C).

OMICRON 2016 Page 11 of 70


2.2 High frequency propagation along a cable

PD signals in general are very broad-band signals. The unavoidable low-pass characteristic of a cable
influences the signal. Higher frequencies will get damped while the signal propagates from the fault to the
sensor. The amount of damping as well as the frequencies that are damped depend mainly on the distance
between fault and sensor.
A high frequency PD impulse travels in both directions between the core conductor and the sheath. The
recharging current, which represents the PD impulse can therefore be measured either at the core conductor
or the sheath of the cable. An inductive PD coupler (HFCT) should not cover both conductors adjacent to the
insulation material, as the signal would be eliminated due to invers polarity.

Figure 7: Simplified cable circuit diagram, low pass behaviour depends on core conductor, semi conductive layers, insulation material
and cable length

Figure 8: With increasing distance from fault location, the PD signal loses its broad-band characteristic due to the low-pass
characteristics of the cable under test. (Colour of the frequency curves correspond to the impulses in Figure 7).

Using a higher frequency range often allows a less disturbed measurement, while it limits the range from a
sensor in which faults are detectable. Vice versa, lower frequencies are often more disturbed but allow for a
'deeper' look inside the cable.

OMICRON 2016 Page 12 of 70


The damping effect of a calibrator impulse along an HV cable in dependence of the measuring frequency is
shown in the example below.

Figure 9: Damping of a calibration impulse along a HV cable in dependence of the measuring frequency - each joint represents 500m.

Figure 10: Synchronous multichannel PD measurement - damping along a HV XLPE cable

The MPD allows selectable center frequency as well as the bandwidth for the PD measurement. For a first
measurement, a bandwidth of 300 kHz with a center frequency below 1 MHz (e.g. 400 kHz) is a good
starting point. Depending on the disturbances and the cable length between consecutive sensors, these
settings have to be adapted.

OMICRON 2016 Page 13 of 70


2.2.1 Measurement frequency selection
When selecting the measuring frequency and bandwidth, several points have to be considered. Different
filter settings might be used to measure and to locate partial discharge.

"Visibility" of PD
The low-pass characteristic of the cable damps higher signal frequencies. Higher measurement frequency
reduces the "visibility". It is not possible to measure PD far away from the measuring point at high
frequencies. Therefore the center frequency should be selected as low as possible to see far inside the
cable. On the other hand, using a high center frequency allows for a very sensitive measurement that
focuses on the point of measurement, which is often the termination. Care needs to be taken when not
following the above mentioned points, as resonance effects in the cable or on the termination might lead to
changes in sensitivity.

Disturbances Noise
Disturbances and noise are not equally distributed over the frequency range. There might be certain
frequencies which are noisier than others. Very often, noise is dominant in the low frequency range. It might
be possible to avoid disturbances and to get a better signal-to-noise ratio by carefully choosing the
measuring frequency and bandwidth.

Figure 11: Disturbances are dominate in the low frequency range up to 1,5MHz the measurement filter should be set next to the noise
hump but as low as possible

OMICRON 2016 Page 14 of 70


TDR
The TDR view is not effected by the selected measuring filter. The signal represents the whole input signal
including, background noise, disturbances and real PD events. If the signal is too noisy, it is possible to
activate an averaging of the signal. As noise is a random, statistically evenly distributed signal, averaging it
will lead to cancellation. The PD signal on the other hand will be more or less the same for all triggered
events and as such will be highlighted. Averaging of repetitive signals is the way of choice in enhancing the
signal-to-noise ratio.
In special cases a 4.7 MHz or even 2.3 MHz low-pass filter might be activated within the MPD software to
smooth the TDR signal. This filter is activated as one of the first elements in the whole signal chain. Be
aware that it influences the whole input signal noise as well as PD and therefore needs to be used with

extreme care. PD measurements with upper cut-off frequencies 2 = above the low-pass filter
2
frequency will not show the expected behaviour.

Figure 12: TDR example joints can be seen as impedance changes where HF signals are partly reflected.

Filter bandwidth (sTDR)


As the sTDR uses a statistical approach, some of the limitations mentioned earlier do not apply. Good
results usually can be achieved, enabling the high bandwidth filter option and selecting a bandwidth of
1 MHz. This results in a small dead time, where no pulses can be detected. The localization precision is
increased as a direct consequence.
The correct ratio between the center frequency and the filter bandwidth is essential to achieve a successful
PD localization. The center frequency fc has to be at least 1.3 times higher than the used bandwidth f: It can
also be chosen to gain the best signal-to-noise ratio.
The resulting dead time of the selected filter is shown in the Q tab of the MPD & MI Software. These values
are only representing the dead time of the used filter. Therefore the low-pass behaviour of the test setup and
the asset itself also have to be taken into account.

f Minimum fc Filter dead time

300 kHz 390 kHz 3.3 s

1 MHz 1.3 MHz 1 s

3 MHz 3.9 MHz 333 ns

Table 4: Values only valid for "high bandwidth filters".

OMICRON 2016 Page 15 of 70


Double pulse resolution
The selected filter bandwidth is directly related to the double pulse resolution of the system. The bandwidth
of the filter is inverse proportional to the double pulse resolution. The PD impulse and its reflection cannot be
separated if the time in between is less than the total filter dead time. As a result, the system is 'blind' to PD
happening close to the cable end.

Figure 13: Influence of wide band filter compared to narrow band filter.

PD occurring with a very high repetition rate, like corona, might blind the system so that other PD events
cannot be measured, respectively the sTDR will not work.
It is beneficial to use a wide filter bandwidth as positive superposition, negative superposition or even a total
cancellation of the PD signal caused by reflections and its resonances can be avoided. Resonance might
be seen in the FFT view.

Figure 14: Several resonances can be seen.

One would think that the highest filter bandwidth leads to the best result, however there is one drawback. A
low-pass damped PD impulse is low in amplitude but relatively long in time. The peak detector will not
recognize the signal as a PD impulse in case the rise time is longer than the filter dead time. In such a case
the statistical localization might not work, thus the bandwidth should be reduced.
The statistical fault localization can be done using a recorded measurement, however it is recommended to
perform a quick localization onsite, allowing the shift of the measuring frequency if required.

OMICRON 2016 Page 16 of 70


2.3 Essential information prior to a PD cable measurement

Cable type
Cable rating
Cable length
Single phase or 3 phase
Type of end termination
Joints How many joints and their exact position
Reason/expectation of measurement
Test setup
Coupling method
Surge arrester
Sheath grounding
Online or offline test (power frequency or VLF)
Test procedure
Relevant standard
Test reports of previous measurements
Weather (Indoor/Outdoor)

OMICRON 2016 Page 17 of 70


3 Cable accessories and how to decouple the PD signals

3.1 End terminations

End terminations are used to manage the electrical field at the end of MV or HV cables, similar as bushings
are used at transformers. Different solutions exist due to the variety of insulation materials. The most
common techniques are geometric (HV/MV cable) and refractive/resistive field control (MV cable).

Figure 15: Cable end without field control Tangential concentration of the electric field at the surface of the insulation material. The

inception voltage can be estimated by knowing the dielectric strength, insulation capacitance and surface capacitance. =

Figure 16: Geometric field control Deflector attached to the semi conductive layer increases the insulation capacitance which is
directly related to the PD inception voltage. In addition, the electric field is displaced into the solid insulation (higher r). There is no
tangential concentration of the electric field at the surface of insulation anymore. Most common solution for HV cable termination/joints.

Figure 17: Resistive and non-linear field control A semi conductive tube or tube with microvaristors (siliciumcarbit, zinkoxid) is
attached to the outer semicon layer of the cable. The higher the electric field, the more conductive the tube gets, thus the field is
automatically controlled especially in the areas of high field strength.

OMICRON 2016 Page 18 of 70


Figure 18: Refractive field control The electric field lines are refracted at the attached tube due to the way higher relative permittivity of
the used material. The changed field distribution is leading to a lower electric field at the surface/border of the semicon layer.

Figure 19: MV end termination using refractive field control technique.

Figure 20: Capacitive field control - Defined voltage drop across the grading layers reduces the local electrical field.

OMICRON 2016 Page 19 of 70


3.1.1 Coupling methods used at end terminations

Coupling capacitor / measuring tap


A coupling capacitor has to be used to perform a PD measurement according the IEC 60270 standard. The
setup provides information about voltage and PD activity. Connect the HV terminal of the coupling capacitor
to the HV side of the end termination. A toroid and conductive tube, which increases the diameter of the
main HV connection and lowers the surface field strength, should be used to ensure a PD-free setup.
Conductive tape has to be used, in case the tube is broken. If not, floating potential can occur. The
measuring impedance CPL 542 can also be connected directly to the end termination measuring tap, if
present and accessible.

DANGER
Death or severe injury caused by high voltage or current
Switch off the high voltage. Always obey the five safety rules and follow the
detailed safety instructions.
When mounting Measurement Equipment and its signal cables, please observe the
high-voltage safety distances between the Measurement Equipment (incl. cable)
and HV, as well as the phases and phase to ground. Always adhere to
international standards and local safety regulations.
Ensure that the rating of the Measurement Equipment fits the requirements of the
grid or test procedure.
All capacity (coupling capacitor or DUT) can store and recover dangerous charge
(and high voltage). Therefor ground all capacities carefully between the HV tests.

Figure 21: Basic connection diagram of a coupling capacitor at a cable end termination.

OMICRON 2016 Page 20 of 70


Figure 22: High-voltage filter including a coupling capacitor (left side) connected to a HV cable end termination.

HFCT
The HFCT has to be installed at the grounding wire between the cable sheath and the end termination.
If more than one grounding wire is used, it should be considered to extend the length of one of these wires
to feed both grounding wires through the HFCT. Otherwise, the HFCT will measure only parts of the high-
frequency signals. The percentage of the signals measured is defined by the high-frequency impedance of
the wires. If the impedance of both wires would be similar, a maximum of 50% of the signal will be detected.
The test or grid frequency which is required for a phase-synchronized measurement can be transferred by
the sensor (HFCT) in case the current is high enough. An air gap should be added in case the current is
higher than 15 A (for MCT 120) to avoid saturation effects (7,5 A for MCT 100). A Rogowski Coil around the
MV or HV cable (connected to the V input of the MPD) might be used if the current through the HFCT is too
low.
The grounding of the cable sheath heavily influences the signal quality, thus there is a difference if the
sheath is connected directly to ground or via SVL.

DANGER
Death or severe injury caused by high voltage or current
Switch off the high voltage. Always obey the five safety rules and follow the
detailed safety instructions. Keep safety distance to high voltage.
Do not mount the HFCT to any grounding cable or electrical conductor where the
occurrence of high currents is possible during mounting. Due to electromagnetic
force, the split-core elements could be contracted, resulting in possible injuries
(trapped fingers) or damage to equipment.
Never open grounding connections where the occurrence of any currents is
possible.

OMICRON 2016 Page 21 of 70


Figure 23: Basic connection diagram of an HFCT at a cable end termination.

Figure 24: An HFCT installed at the sheath grounding wire of a HV cable termination.

OMICRON 2016 Page 22 of 70


Figure 25: HFCTs installed at the connection of cable sheath to a GIS substation.

OMICRON 2016 Page 23 of 70


Figure 26: HFCTs around the cable screen/sheath of an MV cable termination.

Figure 27: HFCT installed at the sheath of a HV cable close to the end termination.

OMICRON 2016 Page 24 of 70


UCS1
The UCS1 is a UHF PD sensor designed to measure and monitor cable end terminations. The directional
sensor detects PD signals as transient voltage drops across the insulation of HV cable terminations and can
be used with or without parallel installed grounding connections. It is not designed to be used as a
mechanical support. In order to achieve the best response characteristic, the UCS1 should be installed as
close as possible to the termination, using short and low inductive braids.

DANGER
Death or severe injury caused by high voltage or current
Switch off the high voltage. Always obey the five safety rules and follow the
detailed safety instructions.
When mounting Measurement Equipment and its signal cables observe the high-
voltage safety distances. Always adhere to international standards and local safety
regulations.
Ensure that the rating of the Measurement Equipment fits the requirements of the
grid or test procedure.

The UHF approach ensures a very sensitive (local) PD measurement in environments with high disturbance
levels.

Figure 28: Basic connection diagram of a UCS1 sensor at different cable end termination designes.

OMICRON 2016 Page 25 of 70


3.2 Cross Bonding Link Box

A current flow in the core conductor of an MV/HV cable causes a magnetic field which induces a voltage into
the sheath of the cable. The longer the cable, the higher and more dangerous the voltage-to-ground on the
sheath can be. The sheath has to be bonded to ground potential in order to limit the voltage to an acceptable
level. If the sheath is grounded at both ends or at multiple points, then the induced voltage causes a
circulating current in the sheath. This current will cause additional heating and results in a reduction of the
cable carrying capacity (losses). To minimize such losses, different bonding techniques like single point (only
for short cables), both ends (additional cable for circulation current) or cross bonding can be used.
Losses in long 3-phase HV cable systems can be reduced best by using the cross bonding technique. The
length of the cable is divided into three approximately equal sections, where the screen is cross bonded in
each joint (link box) and connected to ground at the end of the circuit. The vector sum of the induced voltage
is practically zero in case of a symmetrical operation (same current in all phases, 120 phase shift). [2]

Figure 29: Principle cross bonding setup, sheath voltage distribution above the connection diagram.

OMICRON 2016 Page 26 of 70


3.2.1 Sheath Voltage Limiter - SVL
Sheath voltage limiters (also called surge arresters) are installed for instance in link boxes to limit the
induced voltage in the screen, which can occur due to fault currents or asymmetric load. Lightning, switching
impulses (transient overvoltage) or earth faults are typical error causes.

Figure 30: Sheath voltage limiter (bottom) inside a bonding box.

Build up from a stack of varistors (e.g. a metal-oxide varistor MOV) causes the voltage above a certain
response voltage to be reduced to a pre-set voltage level. Because the construction is made of multiple
single varistors, difficulties during PD measurement might occur. During normal operation, each inner
varistor pill is on floating potential and might produce partial discharges.

Figure 31: Principle drawing of a sheath voltage limiter. The arrester pills are shown in red.

OMICRON 2016 Page 27 of 70


3.2.2 Coupling methods used at link boxes

HFCT
HFCTs can be used to decouple the high-frequency PD signals at the cross-bonding connections, as these
impulses travel between the main conductor and the cable screen. The signal propagation can be very
complex, therefore it is beneficial to have a connection straight through, especially during tests with an
external voltage source. The HFCTs have to be connected around the cross-bonding connections.

DANGER
Death or severe injury caused by high voltage or current
Switch off the high voltage. Always obey the five safety rules and follow the
detailed safety instructions. Keep safety distance to high voltage.
Note: The cable sheath may lead dangerous high voltage.
Do not mount the HFCT to any grounding cable or electrical conductor where the
occurrence of high currents is possible during mounting. Due to electromagnetic
force, the split-core elements could be contracted, resulting in possible injuries
(trapped fingers) or damage to equipment.
Never open grounding connections where the occurrence of any currents is
possible.

Figure 32: HFCTs installed in a cross-bonding box, on every phase of a three-phase cable system.

OMICRON 2016 Page 28 of 70


Figure 33: HFCT connected at a straight through cross bonding box.

3.3 Joint

The acceptance level of a newly-manufactured MV and HV cable is less than 10 pC and therefore often not
an issue within the first years of service. End terminations and joints have to be built onsite, hence the risk of
failures at those components is much higher. Inadequate field control or voids (e.g. cavities, dirt, etc.) due to
the onsite fabrication can lead to an inhomogeneous electric field that can cause PD activity.
Different cable types and voltage levels require different joint techniques, therefore several solutions do
exist.

Figure 34: Component joint, two component material (silicon rubber) shrinks and applied pressure on the single parts.

OMICRON 2016 Page 29 of 70


3.3.1 Coupling methods used at joints

HFCT
HFCTs can be connected around the bypassed sheath at the cable joint to decouple PD signals. The
physical principle is the same as when measuring at cross-bonding connections.

Figure 35: HFCT installed at a cable joint with a sheath bypass.

OMICRON 2016 Page 30 of 70


4 Test setup, calibration and software settings

4.1 Coupling capacitor

Figure 36: PD measurement using a coupling capacitor at the end termination.

The IEC 60270 states that a coupling capacitor has to be used when performing a PD measurement. Such
an approach provides information about the voltage, which is needed for a phase-synchronized PD
measurement. OMICRON offers standard coupling capacitors from 12 kV up to 100 kV with and without
external measuring impedance (quadripole / CPL 542).
The more energy that is saved in the capacitor, the higher the recharging current is in case of a PD event,
and the sensitivity is higher. Therefore a capacitor value should be as high as possible.

DANGER
Death or severe injury caused by high voltage or current
Switch off the high voltage. Always obey the five safety rules and follow the
detailed safety instructions.
When mounting Measurement Equipment and its signal cables, please observe the
high-voltage safety distances between Measurement Equipment (incl. cable) and
HV, as well as the phases and phase to ground. Always adhere to international
standards and local safety regulations.
Ensure that the rating of the Measurement Equipment fits the requirements of the
grid or test procedure.
All capacity (coupling capacitor or DUT) can store and recover dangerous charge
(and high voltage). Therefor ground all capacities between the HV tests carefully.

OMICRON 2016 Page 31 of 70


4.1.1 Connection, calibration and software settings

Figure 37: Typical cable PD test setup according to IEC60270 measurement on ground potential.

When using a coupling capacitor (CC) without an integrated measuring impedance, the low side of the
coupling capacitor has to be connected to the input (red socket) of the CPL542 (basic test setup with
measurement on ground potential). Connect the PD output of the CPL to the PD input of the MPD and do
the same for V. The black socket or grounding screw has to be connected via a short, low inductive cable to
ground potential. Please note that the measuring impedance and acquisition unit can be placed on various
positions, such as on HV potential or within the test object path due to the fiber optic approach. Therefore
different setups have their benefits.

PD calibration using a coupling capacitor


The PD calibrator CAL542 has to be connected as close as possible to the device under test. Inject a
reasonable charge value in the expected range and perform the calibration. To do so, enter the target value
in the Q tab, check if fQIEC is 300 Hz (output frequency of the calibrator) and press compute. The divider
factor (Charge) will change when the measuring frequency is changed. Therefore the calibration has to be
performed again in case of changing the center frequency or bandwidth.

Figure 38: PD calibration using a coupling capacitor.

OMICRON 2016 Page 32 of 70


Figure 39: Software settings Q tab Filter settings, charge calibration.

The same setup but usually a higher charge value might be used to calibrate the propagation speed or cable
length. To do so, use the Measure- / Time- (TDR) or Localization- / Loc.- (sTDR) tab. Detailed information
can be found in the TDR and sTDR chapters.

Figure 40: Software settings PD tab External quadripole, preamplifier auto gain checked, trigger settings.

OMICRON 2016 Page 33 of 70


A correct trigger is essential for analyzing the signal spectrum and using TDR. To trigger to the calibration
impulse or the highest PD impulse, the Trigger Mode has to be set first to edge or peak. Then change the
trigger level until a stabile moving impulse is shown in the Time view. If the signal is frozen, the trigger level
has to be changed. The maximum time resolution is limited to 32 s. A pre-trigger delay of 16 s is selected
as default. In case of long cables the pre trigger might be shifted, in order to measure the reflection after
16 s. There is no difference in terms of how to use the trigger function if the measurement is made using a
coupling capacitor or HFCT.

The influence of using short or long banana cables connected to the calibrator is shown below. It is highly
recommend to use short and low inductive connections to avoid oscillations and resonances. The reflection
might not be visible in case of an oscillating signal. Respectively a localization might not be possible.

Figure 41: CAL542 impulse using a 0.1 m cable.

Figure 42: CAL542 impulse using a 10 m cable.

OMICRON 2016 Page 34 of 70


Voltage calibration using a coupling capacitor
To perform the voltage calibration, apply about 10% of the nominal voltage; choose MPD 600 as the voltage
trigger source; select for display; enter the target voltage and press compute. The feature allowing each
unit to trigger itself might be used in case the PD and V information is gained at several coupling capacitors.

Figure 43: Software settings V tab Trigger Source MPD 600, Select for display, V calibration.

4.1.2 What type of quadripole is needed?


The CPL is a coupling device (CD) (or also called quadripole) and should be used since the MPD input
voltage (100 mV > V-Input@MPD < 60 V) and current (PD-Input@MPD < 15 mA) is limited. The capable
current of the each CPL type depends on the used inductance and capacitance. Due to the higher
capacitance, CPLs with higher current ratings are often required in case of a measurement in the test object
path. The coupling capacitor, CPL (4 F, 30 F, 120 F, 270 F) and MPD (1 F) form a capacitive voltage
divider, thus the divider factor (test voltage see V-tab) can be approximately calculated.

Figure 44: Equivalent circuit diagram

It is recommended to double check the software calculated divider factor. As an example, a 2 nF coupling
capacitor in combination with a 30 F CPL leads to a divider factor (test voltage) in the range of 1:15500

OMICRON 2016 Page 35 of 70


4.2 HFCT

Figure 45: Omicrons split core MCT 100 and MCT 120 HFCT inductive sensors.

PD causes electromagnetic signals. Inductive sensors pick up the magnetic part of the electric signal by the
same principle as a "real" transformer. HFCTs are often used if a ground/earth connection is available.
Therefore the HFCT is placed around these connections and transfers the high-frequency impulse to a
secondary winding. The main benefit of using HFCTs is the possibility to measure PD pulses not at high
voltage potential but at grounding connections without opening them.

DANGER
Death or severe injury caused by high voltage or current
Switch off the high voltage. Always obey the five safety rules and follow the
detailed safety instructions. Keep safety distance to high voltage.
Do not mount the HFCT to any grounding cable or electrical conductor where the
occurrence of high currents is possible during mounting. Due to electromagnetic
force, the split-core elements could be contracted, resulting in possible injuries
(trapped fingers) or damage to equipment.
Never open grounding connections where the occurrence of any currents is
possible.

OMICRON 2016 Page 36 of 70


4.2.1 Connection and software settings
The HFCT has to be connected directly to the PD input port of the MPD without using an external
quadripole. Use a BNC cable which is as short as possible.

Figure 46: HFCT connected to the PD input of a MPD

The MPD contains an internal quadripole. This has to be enabled in order to separate and detect the 50 Hz
components of the coupled signal directly from the PD input path. Select the PD tab in the OMICRON
software. Set all units where a HFCT is connected to internal quadripole and do not use the V port. The
HFCT couples inductive with the current to be measured. In order to gather stable PD patterns, it will be
sufficient when the current has the same frequency as the voltage. However there will be a phase shift to the
normally used voltage signal. This shift can be compensated in the Q tab of the software (). See the
chapter phase synchronization for more information.

Performance check HFCTs


PD measurements require charge calibration. Due to the non-conventional PD measurements done with
HFCTs, it is not possible to calibrate the measuring circuit in accordance with IEC 60270. The calibration
factor of the HFCT depends mainly on a frequency-dependent material constant and the numbers of
secondary turns on the ferrite. Thus the sensor can be checked independently from the test circuit. The
calibration factor will change when the measuring frequency is changed, therefore the sensitivity check has
to be performed again in case of changing the center frequency or bandwidth.
For the sensitivity check, a PD pulse calibrator with short-circuited output port can be put through the inner
ferrite window. Select internal quadripole, inject a defined charge level, set this value as target value and
press compute in the Q tab of the software.

Figure 47: HFCT performance check.

OMICRON 2016 Page 37 of 70


Figure 48: Use "internal quadripole" in case HFCT is used to gain 50/60 Hz frequency information.

All other software settings remain as shown in the chapter coupling capacitor.

fcenter [MHz] MCT 120 (0 mm gap) MCT 120 (0.4mm gap)

0.1 4.3 - 9.2

0.2 4.0 7.2 8.0 12.7

0.5 3.7 5.8 4.5 9.5

1.0 3.6 5.2 3.7 6.1

2.0 3.6 5.0 3.7 5.3

3 14 3.6 4.9 3.7 5.1

15 3.5 4.9 3.7 5.1

16 20 3.5 5.3 3.7 5.5

Table 5: Typical divider factor (charge) values of MCT 120 with and without air gap.

The amount of primary turns has to be considered by entering the corrected value in the calibration text field
of the Q tab. For example, a calibrator value of 100 pC injected over 5 turns to the HFCT means 5x 100 pC
= 500 pC.

OMICRON 2016 Page 38 of 70


4.2.2 Saturation of HFCTs (MCT 100)
The magnetic material will be saturated with high current through the sheath leading through the HFCT. This
can easily be observed in the shape of the measured test voltage, which in fact is the induced current.
With heavy saturation, the measured PD level will be negatively affected as well. A heavy saturation of the
HFCT not only affects the transmission characteristics of the test frequency, but also of higher frequencies.
Connecting a charge calibrator under load conditions leads to the examples shown below. Depending on the
magnitude of saturation, the PD signals get heavily under estimated.

Figure 49: Calibration impulse (10nC) measurement with a heavily saturated HFCT load conditions with 200A through the HFCT.

Beside the saturated voltage waveform (green), the actual current through the HFCT was a sinusoidal
current a line can be seen in the histogram. This line was generated by constant calibrator pulses of 10nC.
Due to the core saturation, the true PD value could not be measured!
The effect of saturation can be reduced by introducing an air gap to the HFCT. Shown below is the same
test setup as above but with an air gap.

Figure 50: Calibration impulse (10nC) measurement with a heavily saturated HFCT after introducing a 0.2 mm air gap load conditions
with 200A through the HFCT.

OMICRON 2016 Page 39 of 70


Figure 51: Calibration impulse (10nC) measurement with a heavily saturated HFCT after introducing a 1.8 mm air gap load conditions
with 200A through the HFCT.

The air gap will not only avoid saturation at higher currents, but it also influences the calibration factor and
the dynamic range of the HFCT. The standard airgap foil provided by OMICRON is 0.1 mm.

4.3 UCS1 (UHF)

The UCS1 has a signal side on top and a ground side next to the TNC connector. The ground side must
always have a solid, low inductive grounding, whereas the signal side has to be connected to the cable
sheath. The coaxial cable towards the UHF converter must have a certain distance from the voltage lead
(signal side) due to safety reasons. The distance shouldnt be less than 15 cm. The UHF converter attached
to the MPD 600 has to be connected to the TNC connector of the sensor. Besides the safety distance, the
cable should be as short as possible to ensure a sensitive measurement.
Due to safety reasons, the senor should not be connected while the cable is energized.

OMICRON 2016 Page 40 of 70


Figure 52: Schematic of UCS1 connection.

DANGER
Death or severe injury caused by high voltage or current
Switch off the high voltage. Always obey the five safety rules and follow the
detailed safety instructions.
When mounting the Measurement Equipment and its signal cables, please observe
the high-voltage safety distances. Always adhere to international standards and
local safety regulations.
Ensure that the rating of the Measurement Equipment fits the requirements of the
grid or test procedure.

In the PD tab, select the UHF preamplifier, deactivated Auto gain and set the gain regulator to position 6-8.
In case the signal is very low, the gain level might be changed. The UCS1 is not designed to transfer the
50/60 Hz signal content, thus an external voltage trigger (phase synchronization) or additional channel has
to be used. Prior to starting the UHF sweep in the UHF tab, the start and stop frequencies should be
selected. Typical measuring frequencies are 300 MHz up to 600 MHz. The upper limit of the sensor is
1000 MHz. It is recommended to perform several sweeps, as partial discharge might not be active at the
time of a single sweep. After a few sweeps, the sweep feature has to be deselected and the measuring
frequency (V_UHF tab) has to be tuned to a proper frequency range to ensure the highest signal-to-noise
ratio. The lower curve in the UHF view represents the noise and continues the signals, whereas the upper
curve contains the noise, continues signals as well as sporadic PD impulses (maximum value).
Using the UHF 620, several narrowband filters, a 70 MHz medium band filter and a 1.9 GHz wideband filter
can be selected. This allows a precisely tuneable measuring frequency. Please note that in case of using a
narrowband filter, the center frequency might have to be shifted during the measurement.

OMICRON 2016 Page 41 of 70


Figure 53: UCS1 Software settings PD tab UHF Preamp. Selected, Auto gain deselected, gain level 6-8.

Figure 54: UCS Software settings UHF tab Set start and stop frequency, hold Min/Max selected, activated and deactivate f-sweep.

OMICRON 2016 Page 42 of 70


Figure 55: UCS1 Software settings V-UHF tab and UHF view Choose proper fcenter and f.

OMICRON 2016 Page 43 of 70


5 Phase synchronized measurement without coupling capacitor
It is important to measure partial discharges phase-synchronized as this makes it possible to interpret the
partial discharge events with respect to their phase position. There are multiple ways to do a phase-
synchronized measurement without using a conventional coupling capacitor measuring impedance setup. In
case several MPDs are connected to one MCU, select the MPD which measures the voltage as a trigger
source in the V tab.

HFCT
In case the current through the HFCT is high enough, the pattern can be triggered using an internal
quadripole, as mentioned in the HFCT chapter.

Rogowski Coil (current clamp)


If the current through the sheath, which leads through the HFCT, is not high enough, an additional external
current clamp is necessary. This current clamp, as shown in the photo below, can be connected to the
voltage port of the MPD 600 to receive a synchronization signal even with very low sheath current. In this
case, the coupling setting of the unit with a connected current clamp has to be set to "external quadripole". In
an unlikely event the 50/60 Hz current of the special current clamp leads to a backwards saturation of the
HFCT, it is recommended to connect the HFCT via a measuring impedance to the MPD.

Figure 56: Example setup of an HFCT and a special current clamp connected to an MPD.

Even when using an external current clamp, the PD pattern will stay phase-shifted in relation to the test
voltage. This phase shift will most probably be dependent on the load conditions and will not be constant
over the time! A phase shift in the PD pattern can be compensated by entering the angle in the Q tab.
The only way to get the real test voltage in on-line measurements is to either have an MPD connected to a
voltage measuring device (e.g. capacitive divider, secondary side of a VT) or to know or estimate the phase
difference between voltage and current. In general, the additional current clamp can be connected around
the sheath or the whole MV/HV cable. A rough estimation in no-load situations is a phase shift of 80 to 90.

OMICRON 2016 Page 44 of 70


Light sensor
In case it is not possible to use any of the options above, the MPD offers the possibility to synchronize itself
to the voltage supply of an artificial light source. The MCU and the CAL 542 are equipped with a light sensor
on top. A light source can be used to get a phase-synchronized measurement. This light source has to be
powered by a voltage which is phase-synchronized to the voltage on the cable/asset to be measured. This
way is not as accurate as any kind of electrical synchronization. Nevertheless, it enables phase-
synchronized measurements. Please note that plugging in the light to the power mains close to the test
object (same building) is usually sufficient. However the light source must be a light bulb, rather than an LED
or other light source that might work with DC. Furthermore, a lot of dimmable lights do not work either, as
they are driven by pulse-width modulation.
Trigger Source Line Trigger 50/60 Hz has to be selected, to enable the synchronization on a light signal.

Figure 57: MPD software settings using a light source as trigger Select Line Trigger 50/60 Hz as Trigger source.

Voltage divider
In general, all kinds of sinusoidal voltage signals below 60 V can be connected to the voltage input port of
the MPD 600. This signal might be gained form a LV side of a VT or 230 V to e.g. 50 V transformer.

Internal Trigger
If none of the above-mentioned solutions is feasible, the internal trigger of the MPD 600 might be used.
Select the trigger source Internal Trigger and change the Int. trigger freq. until a stabile PRPD can be
obtained.

OMICRON 2016 Page 45 of 70


6 Typical PRPD patterns on cables
The following examples show four typical PD patterns. PD patterns measured in the field will most likely not
be so clear to identify, as shown in this example.
If the synchronization signal is derived from an HFCT or a current clamp, the displayed 'voltage' is actually
the measured current. If the current has the same frequency as the voltage, it is sufficient for the generation
of stable PRPD pattern. In the following descriptions 'voltage' always refers to the actual voltage.

6.1 Corona

Figure 58: Corona visible in both half waves Source is most likely on high voltage potential.

How to identify/distinguish:
o Inception (not applicable during online measurement)
Starts in the negative half wave of the test voltage if corona is on high voltage potential or in
the positive if it is on ground potential.
o No change in amplitude on change in voltage level (not applicable during online
measurement).
o If discharge patterns appear in both half waves, they will have very different amplitudes
while being 180 apart.
o Reflections: Full reflections can be seen at short cable lengths.
o Partial reflections can be created on joints, which are installed in the cable under test.

OMICRON 2016 Page 46 of 70


Corona discharges, even if clearly identified, should be avoided as they are of high amplitude and
are lowering the sensitivity of the measuring system.

Corona

First reflection

Second reflection

Figure 59: Corona on ground potential Measured at a cable with reflections visible.

Due to the change of the characteristic impedance on joints and on the end of a cable, a PD impulse
propagating in the cable will be partially transmitted and partially reflected. The ratio of transmission to
reflection depends on the size of the change of characteristic impedance.
Due to the damping of the signal, the reflection is always much lower in amplitude than the original signal.
Thus if the cable or the distance between joints gets too long, the reflection is not visible anymore.

OMICRON 2016 Page 47 of 70


6.2 Surface Discharge

Figure 60: PRPD of surface discharge

How to identify/distinguish
o Two clouds are visible (almost symmetric in size and shape), which have phase difference
of 180 to each other.
o The clouds are displayed in both half-waves.
o Sometimes the amplitude of one cloud is higher than of the other one. But the density acts
the other way around.
o Dominant in the first (0 to 90) and third quadrant (180 to 270). This applies only, if phase
correction was carried out

OMICRON 2016 Page 48 of 70


6.3 Void

Figure 61: PRPD of a void discharge in solid material.

How to identify/distinguish
o The discharge pattern will always appear around the zero crossings of the test voltage.
o It will grow to both sides on increasing test voltage (not applicable during online
measurement).
o The dumpling shape pattern will appear on both half waves with comparable amplitude
(attention logarithmic scaling).

OMICRON 2016 Page 49 of 70


6.4 Developing fault in XLPE

Figure 62: Developing void fault in XLPE cable Approximately one minute measuring time.

Figure 63: Developing void fault in XLPE cable Approximately two minutes measuring time.

Figure 64: Developing void fault in XLPE cable Approximately three minutes measuring time.

OMICRON 2016 Page 50 of 70


6.5 Floating potential

Figure 65: PRPD of floating potential.

In the PRPD view, a potential fault within a surge arrester with more exact floating potential will be located
close to the zero crossing of the test voltage.
Compared to contact discharge, the repetition rate is usually lower and the PD pulses do not disappear with
time.

OMICRON 2016 Page 51 of 70


7 Fault localization on cables
Signal propagation along a MV/HV cable system can be very complex, thus analyzing the test results is one
of the main challenges. Time Domain Reflectometry (TDR) is a very well-known tool, however it is limited
especially in noisy situations with multiple reflections. The statistical TDR (sTDR) which is later described, is
more robust regarding noise and visualise the multiple reflection points.

Udo Ranninger
7.1 TDR Time Domain Reflectometry

Udo Ranninger
Figure 66: TDR schematic A single PD impulse causes traveling waves on the cable in both directions (green and blue).

Udo Ranninger

Figure 67: TDR schematic The impulse traveling to the near end (green) gets measured directly at the time t 1.

Figure 68: TDR schematic The impulse traveling to the far end (blue) gets reflected and is measured at the near end at the time t 2.
The time difference t is directly proportional to the distance of the fault to the far end.

A single PD pulse on an expanded test object (cable) causes traveling waves in both directions. One pulse
will directly be measured (green in the example) at the near end. The other pulse reaching the far cable end
will be reflected. It will also be measured at the near end after the time t. The time delay of these two
pulses is directly proportional to the position of the PD fault. If the propagation speed is known, the fault

distance from the far end can be calculated easily by: =
2

As TDR is a well-known method the OMICRON software, as most solutions, expects a propagation speed of
to be entered to return correct results.
2
OMICRON 2016 Page 52 of 70
Software Settings TDR
Prior of using the localization tool, the PD impulse has to be triggered. Cursors can be activated in the
Measuring tab allowing a precise time measurement. Enter the propagation speed and set the cursors to the
beginning or peak of the original impulse and its reflection. As mentioned earlier, the shown distance
represents twice the signal path and therefore has to be divided by two. Discharges along a cable lead to
results seen from the far end. This approach can be used to calibrate the cable length propagation speed
respectively or to locate discharges.

Figure 69: Determination of total cable length (400m) using TDR. In addition a joint can be seen at 2.9 s with respect to the near end
of the cable (point of measurement).

OMICRON 2016 Page 53 of 70


7.2 STDR Statistical Time Domain Reflectometry

Besides the well-known TDR, there exists a second way for fault localization on cables. Statistical TDR often
allows for good localization results in on-line/field measurements as it is able to scope better with a higher
noise level.
As sTDR is a partly automated statistical approach, the real propagation speed v is used to calculate the
distance from the source to the far end. For the best results, use a high bandwidth filter as mentioned in the
chapter measurement frequency selection.
The sTDR functionality is shown in the following two examples:
Example 1
Given is a cable consisting of three parts with two joints in between. The propagation speed on the cable is
taken to be 200 m/s. In a real cable the propagation speed is lower, depending on the type of cable.
There is an HFCT sensor around the grounding of the sheath at the near end. The far end is left open. A PD
source in the first section of a cable emits a signal that is propagating along the cable in both directions.
Near end

PD

Far end
source
Joint Joint

400m 600m 200m

Measuring point

Figure 70: PD signal propagating from the source in both directions along the cable.

While one signal (green) can be measured directly on the near end, the other signal (blue) has to travel
along the whole cable and through both joints before getting reflected at the far end.

PD
source
Joint Joint

2 s 3 s 1 s

Measuring point Joints: Far end:


80% transmission 100% reflection
20% reflection

Figure 71: The signal propagating into the cable (blue) is partly reflected and partly transmitted on every joint along its propagation path.

On the way along the cable, the signal will get partly transmitted and partly reflected on every change in
characteristic impedance. The most common change is on joints in the cable. In this example, 80% of the
signal is transmitted and 20% is reflected.

OMICRON 2016 Page 54 of 70


Amplitude

10 pC

4 s 10 s 12 s Time
0 s

Figure 72: Scope view of the main signals and the noise for the given example.

On the near end, the original signal, as well as all reflected signals will be measured after a certain time, as
long as the damping along the propagation path is not too high. The figure above shows a schematic scope
view of the signals being measured at the near end. For the reason of simplification only the first reflection is
shown.

Amplitude Amplitude

10 pC
100%

0 4 10 12 Time 0 5 10 Time
[s] [s]

Figure 73: Scope view and schematic of signal correlation for sTDR.

At this point the sTDR kicks in. The user selects a certain area in the PRPD view. In this example, the user
selects only the PD source in the cable (green in the scope view). Other signal sources like possible corona
or noise are neglected.
Every single event in the selected area is correlated to all other signals measured. The results of these
correlations time difference between signals and relative amplitude with respect to the starting signal are
now plotted into a histogram view.
As noise events are not phase stable and happen with a statistical manner, they result in a homogeneous
distribution of correlations. The signals from the PD source (green in the scope view) in the cable and the
corresponding reflections are recurring events, which are happening always after the same time distance
and with the same amplitude. So they will pile up in the histogram view of the sTDR, generating clusters.
As stated at the beginning, the software will calculate the real distance for a selected cluster to the far end of
the reflection path.
12s 200 ms
= = = 1200m
2 2

OMICRON 2016 Page 55 of 70


A second example shows the main steps for a PD source somewhere midway along the cable.
Example 2:
Given is the same cable as above. The only difference is that the PD source is located somewhere between
the two joints.

PD
source
Joint Joint

2 s 1 s 1 s

Measuring point Joints: Far end:


80% transmission 100% reflection
20% reflection

Figure 74: The signal propagates in both directions (green and blue). It is partly reflected and partly transmitted on every join along its
propagation path.

As in the example above, the selected signals are correlated to every signal measured and a histogram view
of the amplitude and time difference correlations is generated.

Amplitude Amplitude

10 pC
100%

0 4 8 12 Time 0 5 10 Time
[s] [s]

Figure 75: Scope view and schematic of signal correlation for the sTDR of the second example.

The highest reflection can be seen at = 4 s, therefore the distance of the PD source to the far end will be

4s 200 ms
= = = 400m
2 2

OMICRON 2016 Page 56 of 70


Software settings STDR
Select the Localization tab after choosing a proper STDR filter. The MPD channel you want to refer to has to
be selected as a trigger source. Enter the propagation speed, select Start detection and draw a rectangle
around the discharges shown in the PRPD view. The selection will be correlated in time and amplitude with
discharges occurring within the correlation depth. The results are shown in the Localization tab, where the
cluster with the highest possibility is automatically selected. The corresponding fault location given from the
far end of the reflection path is shown on the right side.

Figure 76: Determination of total cable length (400 m) using STDR.

OMICRON 2016 Page 57 of 70


7.3 STDR Dual End measurement

By using the statistical fault localization mode, it is possible to perform a dual-end measurement. For fault
localization, the dual-end measurement is more accurate than the measurement with one channel. The
original impulse is measured as always. This advantage can be used for measurements on short cables or
for measurements on joints and terminations.
Please note that the scope function (TDR, Time view) of multiple channels is not synchronized, therefore a
dual-end measurement is not possible using TDR.
The main software and filter settings have to be chosen, such as using SDTR with a single channel.
After selecting the discharge pattern in the PRPD view, the second MPD unit has to be selected. Now the
marked discharges of MPD unit 1 (Measuring point 1) are correlated with the signals of MPD unit 2
(Measuring point 2). The shown distance is calculated from the middle of both units and has to be divided by
2. If the distance is positive, the position of the source is towards to the trigger unit. If the distance is
negative, the second MPD channel is closer.
An example of how to use the MPD software in terms of dual-end measurements is shown in the next
chapter Measurement example.

PD source

Measuring Measuring
point 1 Center position point 2
of both sensors
Figure 77: Dual End measuremt

OMICRON 2016 Page 58 of 70


8 Measurement example

8.1 Test Object and general information

Cable type: MV Cable - NA2XS(F)2Y 1x50 RM/16 12/20kV


Cable rating: 20 kV
Cable length: 1200 m
Single phase or 3 phase: Single
Type of end termination: Cold shrink
How many joints: 5
SVL: No
Sheath: Grounded on both sides
Reason of measurement: Demonstration
Online or Offline test (power frequency or VLF): Offline 50 Hz
Test procedure:
Relevant Standard: IEC 60502-2
Weather (Indoor / Outdoor): Indoor

8.2 What does the setup look like?

The test object is a 1200 meter long MV cable with five joints and shrink-end terminations. To demonstrate
different types of coupling methods, respectively fault localization techniques, a coupling capacitor and two
HFCTs are used.

Figure 78: Test setup

OMICRON 2016 Page 59 of 70


8.3 What can be expected?

As there are five joints, multiple reflections can be expected. The graphic below shows the complexity of the
signal propagation within the cable. Please note that only a few scenarios are shown. It is shown that
different reflections lead to the same reflection time. This will form a single cluster in the STDR view,
resulting in a complex interpretation. Measuring at a joint for instance can lead to at least three clusters,
representing the distance of the unit to the cable end, the distance to the beginning of the cable and the fault
position.

Figure 79: A few selected examples of possible reflection paths.

8.4 FFT background noise

The background noise is dominant in the lower frequency range. Therefore a starting center frequency of
3 MHz and a bandwidth of 1 MHz (high bandwidth filter) was selected.

Figure 80: FFT of background noise.


OMICRON 2016 Page 60 of 70
8.5 PD calibration and cable length determination

The PD calibration should be performed for multiple filter settings (fcentre, f) to ensure a calibrated
measurement, as a frequency shift could be essential after applying the test voltage. A calibration signal of
1 nC was injected between the main conductor and the screen of the Test Object. The point of injection was
at the coupling capacitor side.

Figure 81: PD calibration - Coupling capacitor.

The divider factor (charge see Q-tab) of the coupling capacitor unit (at 3 MHz) was 3.501. A factor of 5 was
selected for the HFCT units.
Divider factor for a coupling capacitor:
2 MHz = 3.48
3 MHz = 3.50
4 MHz = 2.82
5 MHz = 4.13
6 MHz = 5.67
7 MHz = 5.67
As expected, multiple reflections within the cable, respectively between the joints, can be seen. The length
of the cable is 1200 m, thus the resulting propagation speed is 163 m/s.

OMICRON 2016 Page 61 of 70


Figure 82: Time View: TDR reflections of the CAL impulse Results have to be seen from the near end (point of measurement) and
divided by 2.

Figure 83: Loc. View: STDR - Reflections of the CAL impulse.

OMICRON 2016 Page 62 of 70


8.6 Voltage calibration

The voltage calibration was done at 2 kV.

Figure 84: Voltage calibration

8.7 Voltage sweep / PD measurement

The voltages slowly increased to 11 kV. This level also represents the PD inception voltage.

Figure 85: Trend of applied voltage and PD activity.

OMICRON 2016 Page 63 of 70


Figure 86: Discharges obtained at the coupling capacitor unit.

8.8 Fault localization

Due to the high noise level in the lower frequency range and the relatively low discharge level, the TDR
approach could not be used to locate the fault position.

Figure 87: Triggered Coupling Capacitor (PD) input signal.

OMICRON 2016 Page 64 of 70


Localization using STDR and the coupling capacitor unit
Beside the fault position, various clusters are generated due to reflections between joints.

Figure 88: STDR fault localization - Result has to be seen from the far end of the reflection path (cable end).

OMICRON 2016 Page 65 of 70


Localization using STDR and the unit connected to the HFCT 1

Figure 89: STDR fault localization - Result has to be seen from the far end of the reflection path (cable end).

As the result is 400 m from the cable end, the fault is most likely in the joint marked red.

Figure 90: Schematic of fault position

OMICRON 2016 Page 66 of 70


Dual End Localization
The HFCT2 is selected and also used as Trigger Source in the Localization tab. The marked cluster will be
correlated with another unit once the unit is selected (see Figure 95).

Figure 91: Dual End settings for the trigger unit (HFCT2)

OMICRON 2016 Page 67 of 70


Figure 92: Dual End correlation after selecting HFCT1

The distance shown of 562.9 m has to be divided by 2. The signal source/fault has to be closer to the unit
HFCT2 as the result is positive. The resulting fault position is 520 m + 281 m = 801 m seen from the near
end of the cable.

Figure 93: Schematic of fault location using dual end STDR

OMICRON 2016 Page 68 of 70


9 List of literature

Kabelhandbuch, 8. Auflage; Rolf Rdiger Cichowski (Hrsg.)

Kabelanlagen fr Hoch- und Hchstspannung, Egon F. Peschke und Rainer v. Olshausen

Special Bonding of High Voltage Power Cables, Cigre, Working Group B1.18 2

OMICRON 2016 Page 69 of 70


Support
When you are working with our products we want to provide you with the greatest possible
benefits. If you need any support, we are here to assist you.

24/7 Technical Support Get Support


www.omicronenergy.com/support
At our technical support hotline, you can reach well-educated technicians for all of your
questions. Around the clock competent and free of charge.
Make use of our 24/7 international technical support hotline:
Europe / Middle East / Africa +43 59495 4444
Americas +1 713 830-4660 +1 800-OMICRON
Asia-Pacific +852 3767 5500
Additionally, on our website you can find our Service Center or Sales Partner closest
to you.

Customer Area Stay Informed


www.omicronenergy.com/customer
The Customer Area on our website is an international knowledge exchange platform.
Download the latest software updates for all our products and share your own experiences
in our user forum.
Browse through the knowledge library and find application notes, conference papers,
articles about daily working experiences, user manuals and much more.

OMICRON Academy Learn More


www.omicronenergy.com/academy
Learn more about your product in one of the training courses offered by the OMICRON
Academy.

For more Information, additional literature,


and detailed contact information of our
worldwide office please visit our website.

OMICRON
www.omicronenergy.com
Subject to change without notice.

You might also like