TRASFORMADOR DE POTENCIAL "TP"

TIPO DE CONFIGURACION: INGRESANDO VALORES


TENSION DE PRUEBA : 40V

VALOR
VALORES EXPERIMENTALES (LABORATORIO) TEORICO

RATIO CORRIENTE % DIF


RELACION Vp/Vs
(proporcion) (mA) (error)
220/240 0.9168 34 0.01 0.9167
220/120 1.8485 31 0.83 1.8333
220/6 36.986 31 0.87 36.6667
220/3 73.998 31 0.91 73.3333
CUADRO 01

TRASFORMADOR DE POTENCIAL "TP"


TIPO DE CONFIGURACION: MODO AUTOMATICO
TENSION DE PRUEBA : 40V
VALOR
VALORES EXPERIMENTALES (LABORATORIO) TEORICO
RATIO CORRIENTE % DIF
RELACION Vp/Vs
(proporcion) (mA) (error)
220/240 0.9168 34 .. 0.9167
220/120 1.8486 32 .. 1.8333
220/6 36.984 31 .. 36.6667
220/3 73.992 31 .. 73.3333
CUADRO 02

VALOR
VALORES EXPERIMENTALES (LABORATORIO) TEORICO
RATIO CORRIENTE % DIF
RELACION Vp/Vs
(proporcion) (mA) (error)
5/5 1.0003 1