You are on page 1of 3

VENTILATOR

Brand Maquet
Type ServoiUniversalBasicEdition
Origin Sweden
TechnicalDataofServoiUniversalBasicEdition
1 Applicableto :Adultchildinfant(allpatientcategorysoftware)
Patientweight :0.5250Kg

2 General
Modularandupgradeablesystem :Interchangeablemodulewithothersystem
Dimension :Userinterface:355x53x295mm
:PatientUnit:300x205x415mm
Weight(withoutstand) :20kg(withouttrolley)
Methodoftriggering :FlowandPressure(canbeupgradedtoneuraltrigger)
Max.OperatingPressure :115cmH2O
BiasFlow :Infant:0.5l/min
:Adult:2l/min
InletGasPressure :200650kPa/2.06.5bar/2994PSI
UnavailableGas/lossofGaspressure :Theflowfromanunavailablegas(air/O2)isautomaticallycompensatedfor
sothatpatientgetsthepresetvolume
PowerSupply :100120VAC10%5060Hzor22024010%5060Hz
External :1215VDC,10A
Batterybackup :2Batterymodule,upto6batterycanbeincluded
BatteryCapacity :Rechargeable,12V,3.5Aheach
BatteryBackupTime :atleast3H(whenUsing6Batteries)
MaxPowerConsumption :140W
Operatingtemperature :+10to+40C
Relativehumidity :1595%noncondensing
Atmosphericpressure :660to1060hPa
Oxygensensorlifetime :>5years
3 GeneralFeatures TheSuctionSupportfunctionpausestheventilatorfromcyclingduringa
trachealsuctionprocedureincludingconfigurablepre/postoxygen
concentration.
UserinterfacecanbeoperatedviatouchscreenorbymeansoftheMain
RotaryDial.DirectaccessknobsforvitalsettingssuchasPEEP,O2
concentration.
Left&RightOrientation
Possibilitytosaveeventlog,trends,dataofscreenpictures(bitmap)
WillautomaticallypromptaPreusecheckwhereallimportantparameters
arecalibratedorcheckedanddocumentedinthememory.Patientcircuitcan
betestedseparatelyfromthePreusecheck.
Insupportedmodestheterminationcriteriaforthebreathissetbytheuser
(InspiratoryCycleoff),inordertoovercomecausesofpatientventilatory
asynchrony
Cleaningprocedureinvolvesonlyonesingleinterchangeablecomponent.
Theflowsensoriscompatibleforallcategoriespatient(donotneedtobe
replacedwhenchangingpatientcategory)
Ventilatorwillreturnautomaticallytothesupportedmode,afteranapnea
periodthathascausedtheventilatortostartBackupventilation
IncludingNebulizer
4 UserInterface Canberotatedandtiltedformaximumflexibility.Canberemovedand
clampedforflexibilityinplacemente.g.bed,wallmount,IVpole,ceiling
supplyunitanda2.9mcableincludingcablemanagement.
Display12,1inch,Color,TouchScreen,highvisibilityfromanyposition
Automaticbrightnessadjusted
5 VentilationModes
PressureControlledVentilation :PC
VolumeControlledVentilation :VC
PressureSupportedVentilation/ContinousPositive :PS/CPAP
AirwayPressureVentilation
SynchronizedIntermittentMandatoryVentilationBased :SIMV(PC)+PS
onPressureControlledVentilationwithPressureSupport

SynchronizedIntermittentMandatoryVentilationBased :SIMV(VC)+PS
onVolumeControlledVentilationwithPressureSupport

NonInvasivepressurecontrolledventilation :NIVPressureControl
NonInvasivepressureSupportventilation :NIVPressureSupport

6 VentilationSetting
Tidalvolume :2to4000mL
InspiratoryMinuteVolume :0.360l/min
Apnea,timetoalarm :545s
PC/PSabovePEEP(cmH2O) :0(120PEEP)
PC/PSabovePEEPinNIV(cmH2O) :0(32PEEP)
PEEP :050cmH2O
CMVFrequency :4150bpm
SIMVFrequency :160bpm
Breathcycletime :0.515s
Phigh :(PEEP+1)50
Thigh :0.210s
TPEEP :0.210s
PSabovePhigh :0(120Phigh)
O2Concentration :21100%
I:ERatio :1:104:1
TInsp. :0.15s
Tpause :030%ofbreathcycletime
Tpause :01.5s
Flowtriggersensitivitylevel :0100%
Pressuretriggersensitivity :200cmH2O
Insp.Risetime :00.4s
Insp.Cycleoff :170%ofpeakflow
Nebulizertime :530min
OxygenBreath :100%for1minute
StartBreath :initiationof1breath
Alarmsilence/reset :2minutessilenceandresetoflatchedalarms
ComplianceCompensation :On/Off
Nebulizer :On/Off

7 Alarms
:Airwaypressuremax
:Exp.Minutevolumemax/min
:Apnea
:RespiratoryFrequency
:Exp.Pressuremax/min
:HighContinousPressure
:O2concentration
:Gassupplymax/min
:Battery
:Leakage
:NopatienteffortinNIV
:TechnicalAlarm

8 Waveformandlooppresentation
Realtimewaveform :PressureWaveform
:FlowWaveform
:VolumeWaveform
:Realtimecombination:PVLoopandflow,Pressure,Volumewaveform
Loops :Volume/Pressure
:Flow/Volume

9 Logfunction
EventLog :Alarm
:VentilatorSetting
:ApneaPeriods
:Immediatefunction
ServiceLog :TechnicalAlarm
:TestResult
:PreventiveMaintenance
:ServiceHistory
:Configurationlog.

11 Communication/Interface
:SerialPort

12 Accessories:
Testlung 1pcs
TestTube 1pcs
Expiratorycassete 2pcs
AGAO2Hose,3m 1pc
AGAAirHose,3m 1pc
Mobilecart 1set
SupportArm 1set
Nebulizer 1set
Bacterialfilter50Pcs 1box
HumidifierHolder 1pc
F&PHumidifierMR850,English 1set
Temperature&FlowProbe 1pc
ReuseablekitMR850 1set
HumidifierChamberAdult 1pc
HumidifierChamberInfant 1pc
SetPatientTubeSilic.Adult 1set
SetPatientTubeSilic.Infant 1set
Angledadapter2215mm/15mm 1pack