Scope of Accreditation
For
Alliance Calibration
11402 Reading Road
Cincinnati, OH 45241
Sidney Taylor
513-769-1200
Calibration
Amount of Substance pH/Conductivity
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-)
4 pH 0.022 pH
pH Meters 7 pH 0.032 pH
10 pH 0.040 pH
Compared to aqueous
1 S 0.35 S/cm
solutions
10 S 0.50 S/cm
Conductivity Meters
100 S 4.1 S/cm
1 000 S 26 S/cm
Electrical Capacitance
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-) 3
Up to 400 pF 14 pF
Capacitance Source Fluke 5500A
(0.4 to 1.1) nF (0.11 + 0.007 4x) nF
@ 1 kHz
Electrical Current
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-) 3
Up to 3A (0.005 3 + 0.001 2X) mA
Agilent 34401A
(3 to 10) A 0.52 % of reading
DC Current Measure Fluke 87
(10 to 200) A (1 + 0.021X) A
Fluke 336
(200 to 500) A (0.51 + 0.023X) A
Up to 3.3 mA (0.000 086 + 0.000 14X) mA
(3.3 to 33) mA (0.000 32 + 0.000 12X) mA
(33 to 330) mA (0.002 9 + 0.000 12X) mA
Fluke 5500A
DC Current Source (0.33 to 2.2) A (0.000 051 + 0.000 35X) A
w/ Coil
(2.2 to 11) A (0.000 40 + 0.000 70X) A
(11 to 200) A (0.11 + 0.002 2X) A
(200 to 500) A (0.11 + 0.002 2X) A
AC Current Measure 40 Hz to 1 kHz
10 mA to 1 A (0.000 69 + 0.001X) A
(1 to 3) A 0.31 % of reading Agilent 34401A
(3 to 10) A 0.52 % of reading Fluke 336
(10 to 200) A (1 + 0.021X) A
(200 to 500) A (0.64 + 0.023X) A
Electrical Power
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-) 3
Electrical Resistance
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-) 3
Up to 100
(0.004 7 + 0.000 088X)
(100 to 1 000)
(0.000 3 + 0.000 13X)
(1 to 10) k
(0.000 015 + 0.000 12X) k
(10 to 100) k Agilent 34401A
Resistance Measure 0.012% of reading
(0.1 to 1 M
(0.000 001 9 + 0.000 1X) M
(1 to 10) M
(0.03 + 0.009 1X) M
(10 to 100) M
(0 to 5 000) m (0.004 4 + 0.000 17X) m Agilent 34401A
Up to 11 (0.007 0 + 0.000 17X)
(11 to 33) (0.012 + 0.000 14X)
(33 to 110) (0.012 + 0.000 1X)
(110 to 330) (0.012 + 0.000 1X)
(0.33 to 1.1) k (0.000 069 + 0.000 1X) k
(1.1 to 3.3) k (0.000 076 + 0.000 11X) k
(3.3 to 11) k (0.000 7 + 0.000 1X) k
(11 to 33) k (0.000 62 + 0.000 12X) k
Resistance Source Fluke 5500A
(33 to 110) k (0.008 4 + 0.000 12X) k
(110 to 330) k (0.006 9 + 0.000 14X) k
(0.33 to 1.1) M (0.000 064 + 0.000 17X) M
(1.1 to 3.3) M (0.000 079 + 0.000 17X) M
(3.3 to 11) M (0.000 64 + 0.000 69X) M
(11 to 33) M (0.000 64 + 0.001 2X) M
(33 to 110) M (0.006 4 + 0.005 8X) M
(110 to 330) M (0.01 + 0.005 9X) M
500
5 m Fluke 5500A
Resistance Source 50 m (0.001 3 + 0.005 8X) m Agilent 34401A
500 m Shunts
5
RTD Resistance Simulation
(-200 to 800) C (0.058) C Fluke 5500A
Pt 385, 100
Electrical Voltage
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-) 3
Up to 1 V (0.000 003 8 + 0.000 070X) V
(1 to 10) V (0.000 010 + 0.000 046X) V Agilent 34401A
(10 to 100) V 0.006 1 % of reading
DC Voltage Measure (100 to 1 000) V 0.006 6 % of reading
Ross Engineering HV
(1 to 80) kV 0.12 % of reading
Probe
(0 to 330) mV (5.5 + 0.055X) V
(0.33 to 3.3) V (17 + 43.1X) V
DC Voltage Source (3.3 to 33) V (180 + 42.8X) V Fluke 5500A
(33 to 330) V (90 + 55.2X) V
(0.3 to 1) kV (14 + 43.9X) V
Indicator
0.001 in resolution (580 + 0.06L) in
0.000 5 in resolution (290 + 0.12L) in Indicator Tester
0.000 1 in resolution (0 to 4) in (58 + 0.59L) in Gage Blocks
0.000 05 in resolution (29 + 1.1L) in
0.000 01 in resolution (5.2 + 3.6L) in
Universal Measuring
(0 to 24) in (2.6 + 3.9L) in Gage Blocks
Machine
Optical Comparators
Magnification 5x to 100x (610 + 1.2L) in
Glass Scale
Linearity (0 to 6) in (140 + 0.67L) in Length Standards
Spheres
Squareness (0 to 6) in 93 in
Microscopes
(0 to 1) in (54 + 45L) in Gage Blocks
Stage Travel
Mass Flow
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-) 3
492 ft/min
Gas Flow Speed 984 ft/min
1.9 ft/min + 3.8 % of reading TSI 9535 VelociCalc
(Anemometers) 1 969 ft/min
2 953 ft/min
(50 to 500) CCM 1.2 CCM + 0.5 % of reading
Gas Flow Meter Up to 50 SLM 0.12 SLM + 0.5 % of reading Alicat Flow Standard
(50 to 250) SLM 1.2 SLM + 0.8 % of reading
Liquid Flow
(0.3 to 50) gpm 0.008gpm + 1.4 % of reading Scale
(gravimetric)
Mass Force
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-) 3
Bench Micrometer
(4 to 40) ozf 2.1 ozf Force Gage
Measuring Force
1 grf to 45 kgf (5.7 + 1.1F) gf Class F Masses
Force Gages (0.5 to 50) lbf (0.003 4 + 0.000 3W) lbf
Class 7 Masses
(50 to 250) lbf (0.009 7 + 0.000 2W) lbf
(10 to 500) lbf 0.83 lbf + 0.1% of reading
(501 to 1 000) lbf 1.1 lbf + 0.051% of reading
Load Cells Load Cells
(1 001 to 5 000) Lbf 5.9 lbf + 0.032% of reading
(5 001 to 20 000) lbf 58 lbf + 0.003% of reading
Mass Hardness
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-)
HRBw
Low 4.6 HRBw
Indirect Verification of Middle 2.2 HRBw
Rockwell Hardness and High 2.5 HRBw Indirect Verification
Rockwell Superficial HRC per ASTM E18
Hardness Testers Low 2 HRC
Middle 1.9 HRC
High 1.1 HRC
Mass Masses
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment Measurement (+/-) 3
1g 0.13 mg
2g 0.22 mg
5g 0.19 mg
10g 0.22 mg
Masses Class 1 Masses
20g 0.25 mg
50g 0.49 mg
100g 0.77 mg
200g 1.3 mg
Mass Volume
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-)
Pipettes (10 to 1 000) L 0.28 L + 0.06 % of reading Gravimetric
Thermodynamics Humidity
Calibration Expanded Uncertainty of
Range Remarks
Parameter/Equipment 1 Measurement (+/-) 3
Vasaila MI70 /
Relative Humidity (0 to 80) % RH (1.2 + 0.005 4T) % RH
HMP77
Timers and Stopwatches (0.1 to 10) min 0.034 sec + 0.005 9% of reading HP 53132 Counter
Calibration and Measurement Capability (CMC) is expressed in terms of the measurement parameter, measurement range,
expanded uncertainty of measurement and remarks. The expanded uncertainty of measurement is expressed as the standard
uncertainty of the measurement multiplied by a coverage factor of 2 (k=2), corresponding to a confidence level of
approximately 95%.
Notes:
1) Laboratory offers calibration services at the laboratory's own facilities and at the client or other agreed upon facilities.
2) Uncertainty shown is per wire for thread wire sets.
3) L = length in inches, D = diameters in inches, T = temperature / RH applied, X = flow / frequency / volts / ohms /
amps / capacitance applied, M = mass applied, F = force in kg, P = pressure applied, V = volume, W = weight in lb
4) 1 mil = 0.001 in