Professional Documents
Culture Documents
C RYSTAL
engineering corporation
PN: 1762 Rev E, 2/2009
Page 1 of 8
Operating Instructions
To ensure safe and accurate operation, please be familiar with the following operations and
functions.
WARNING:
Severe injury or damage can occur through improper use of pressure instruments!
Do not exceed recommended pressure limits of tubing and ttings. Be certain all pressure
connections are secured. Never disconnect pressure instrumentation without rst relieving
system pressure.
CAUTION:
Never insert any object (other than a 18 NPT tting) into the inlet ports. The sensor
diaphragm is very thin and can be damaged or destroyed by solid or sharp objects. Clean
the sensor with appropriate solvents, only.
The internal pressure sensor measures the difference between a sealed high vacuum
reference pressure, and the pressure (or vacuum) applied to the pressure port. The pressure
being measured can be either liquid or gaseous, providing it is compatible with the materials
listed in the specication section.
PTFE tape should be used with any ttings installed into either of the 18 female NPT ports.
Two ports are provided so that additional tees should not be required. Plug any unused
ports.
MultiCals have been calibrated for use with meters that have 10M input impedance, as do
most handheld meters, and some benchtop multimeters. If your meter has higher or lower
input impedance, add 0.1% to the specication of the MultiCal, or refer to the Calibration
section of this manual for other methods of eliminating the error introduced by an impedance
mismatch.
Pressure Measurement
Follow this procedure to correctly use the MultiCal pressure module.
1. Plug the MultiCal directly into the voltage input terminals of the multimeter or with the
patch cord set. The patch cord set consists of a 48 cord with double banana plugs
on each end and a double banana jack splice. Plug the splice into the MultiCal and
the patch cord into the splice. Plug the other end of the patch cord into the multimeter.
Polarity is marked on the MultiCal and on the patch cord and splice.
3. Check the battery condition: With the MultiCal still in the OFF position, the
multimeter must indicate a minimum reading of 100 mV. Readings less than 100 mV
indicate the battery must be replaced. To be sure that your pressure measurements are
accurate, always check the battery condition rst, and replace the battery if necessary.
4. Turn on the MultiCal by sliding the switch to the range you intend to use.
Readings may change when the MultiCal is shifted from a vertical to a horizontal
orientation. This is due to the oil lling that transmits the pressure signal from the
stainless steel diaphragm to the silicon sensor. The magnitude of the shift is typically
less than 0.15 mmHgA.
To check at zero pressure, use a vacuum pump and a thermocouple type vacuum
gauge. Set the Multical to the mmHgA scale.
Zero the MultiCal when the vacuum achieved is 10 millitorr or less. With a small
screwdriver adjust the small trimming potentiometer on the end of the MultiCal until your
meter reads precisely zero.
Another way to adjust zero is to compare the barometric reading with a barometric
reference standard. If any difference is noticed between the MultiCal and the reference,
the zero trimming potentiometer can be adjusted so the readings match.
Battery Replacement
1. Set the power switch to the OFF position.
2. Disconnect the MultiCal from the multimeter and any pressure connections.
3. Turn the MultiCal so the power switch is facing down. Remove the single screw located
between the banana plugs.
4. Grasp the one case half in each hand. Pull the two halves apart, beginning at the end
with the banana plugs.
6. Reassemble the MultiCal. To reassemble, mate the two case halves at the end opposite
the banana plugs, then snap the two halves together.
C RYSTAL
engineering corporation
PN: 1762 Rev E, 2/2009
Page 3 of 8
MultiCal Calibration
A calibration cycle of 1 year is recommended to maintain the MultiCal within specications.
Procedure
1. Connect the MultiCal to the pressure calibration system and to the multimeter. Be sure
that the pressure calibration system and the connection to the MultiCal is leak free.
2. Allow the MultiCal to stabilize at room temperature, away from drafts, for at least 30
minutes before proceeding with calibration. Turn on the multimeter and allow it to warm
up per the multimeter operating instruction. Set the multimeter to the appropriate DC volt
or millivolt scale.
3. Verify the condition of the battery and replace the battery if necessary. Follow the Battery
Replacement procedure to disassemble the MultiCal, but do not remove the battery or
reassemble the MultiCal.
4. Set the switch on the MultiCal to the mmHgA position (switch in position closest to the
banana plugs). Allow the MultiCal to warm up for two minutes.
5. Apply full vacuum to the MultiCal. When the vacuum gauge indicates 10 milliTorr adjust
the Fine Zero potentiometer (R1) for a reading of 0.01mV 0.01mV. If the zero adjustment
range of the Fine Zero potentiometer is inadequate, the coarse zero must be readjusted.
When resetting the adjustment range of the Fine Zero potentiometer (R1) should rst
be set to the center of its rotational range. To center the Fine Zero, rotate the Fine
Zero potentiometer fully clockwise until a click is heard. Then rotate the FINE ZERO
potentiometer counter-clockwise 7 turns. Adjust the COARSE ZERO pot until the output
is as close to zero as possible. Trim in the nal reading to 0 mV 0.01 mV with the FINE
ZERO potentiometer.
R1 FINE ZERO
R12 mmHgA
R15
-
PSIA
Specifications
The following specications apply at 23C, 5C for 1 year after calibration, and when
used with measuring devices having 10M input impedance, or shunted so that their input
impedance is 10M. Also, the module must be "zeroed" prior to taking measurements to
achieve the pressure specications.
C RYSTAL
engineering corporation
PN: 1762 Rev E, 2/2009
Page 5 of 8
Pressure
Range Accuracy
0 to 760mmHgA (0.08% of Reading + 0.2 mmHgA)
0 to 30 PSIA (0.08% of Reading + 0.01 PSIA)
Maximum Working Pressure: 100 PSI
Burst Pressure: 500 PSI
Wetted Materials: 316 Stainless Steel, PTFE impregnated
hard anodized aluminum, and buna-n rubber.
Unless otherwise specied at time of purchase: 1 PSI = 51.7149 mmHg (Mercury at 0C)
Operating Environment
Temperature Derating
Add to Basic Accuracy Specication. C = ambient temperature.
General Specications
EC Declaration of Conformity
I/We
Crystal Engineering Corporation
of
declare that
Pressure Module
MultiCal Series (All Versions)
I hereby declare that the equipment named above has been designed to comply with the
relevant sections of the above referenced specifications. The unit complies with all essential
requirements of the Directives
Director of Engineering
(TITLE OF AUTHORIZED PERSON)
4 February 2009
(SIGNATURE OF THE AUTHORIZED PERSON) (DATE OF ISSUE)
C RYSTAL
engineering corporation
PN: 1762 Rev E, 2/2009
Page 7 of 8
Warranty
Crystal Engineering Corporation warrants MultiCals to be free from defects in material and
workmanship under normal use and service for one (1) year from date of purchase to the
original purchaser. It does not apply to batteries or when the product has been misused,
altered or damaged by accident or abnormal conditions of operation.
Crystal Engineering will, at our option, repair or replace the defective device free of charge
and the device will be returned, transportation prepaid. However, if we determine the failure
was caused by misuse, alteration, accident or abnormal condition of operation, you will be
billed for the repair.
CRYSTAL ENGINEERING CORPORATION MAKES NO WARRANTY OTHER THAN
THE LIMITED WARRANTY STATED ABOVE. ALL WARRANTIES, INCLUDING IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE,
ARE LIMITED TO A PERIOD OF ONE (1) YEAR FROM THE DATE OF PURCHASE.
CRYSTAL ENGINEERING SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL OR
CONSEQUENTIAL DAMAGES, WHETHER IN CONTRACT, TORT OR OTHERWISE.
Note (USA only): Some states do not allow limitations of implied warranties or the exclusion
of incidental or consequential damages, so the above limitations or exclusions may not apply
to you. This warranty gives you specic legal rights and you may have other rights which
vary from state to state.
If calling, have ready the model number, serial number, date of purchase and reason for
return. You will receive instructions for returning the device to Crystal Engineering.