You are on page 1of 1

INFANT INCUBATOR CABINET 800 A ZD

STANDARD COMPOSITION :
Servo Control Mode : Air temperature, Skin temperature, Display humidify
Instant Parameter Displaying : Set temperature, air temperature, skin temperature,
internal humidify, heat power rate
Audible & Visual Alarms : Power failure, sensor failure, higher & lower temp. air & skin temp
override, fan failure, system failure
> 37C Preset Mode : power failure reset memory, alarm silence, keypad lock, self-test function
Hood with dual slope : 6 hand-ports, left & right cabinets rail
RS-232 output connector (use for network system)

TECHNICAL SPECIFICATION
Power requirement : 400VA
Temperature control mode : Air and skin servo control
Air temp control range : 20C ~ 37C
Temperature override mode to : 20C ~ 39C
Skin temperature control range : 30C ~ 37C
Temperature override mode to : 30C ~ 38C
Skin temperature sensor accuracy : 0.3C
Temperature fluctuation : 0.5C
Mattress temperature uniformity : 0.8C
Temperature rise time : 30 min
Noise level within hood environment : 55 dB

ALARMS
Power failure alarm
Fan failure alarm
Sensor failure alarm
System failure alarm
Alarm silence function
High air temperature alarm : +1.5C
Low air temperature alarm : -2.5C
High skin temperature alarm : +1.0C
Low skin temperature alarm : -1.0C
Over skin temperature alarm : 38.0C ; 40.0C
Over skin temperature alarm : 38.0C ; 39.0C

OPERATIONAL CONDITION
Environment temperature : +20C ~ +30C
Environment relative humidity : 5% ~ 99%
Environment air velocity of flow : <0.3 m/s

OTHERS
RS-232 connector
Rail
Infant bed tilt adjusting : 10 step less
Mattress tray max. load : 10 kg
Transfusion pole max load : 2 kg
Cabinet max. load : 6 kg
Gross weight : 120 kg