Instrument Name Digital Readout Oxygen Index Tester YN61037

Model No.

Contents

1 Installation preparation
1.1 Preparing working-area
1.2 Working plateform and Transportation
1.3 Safety
1.4 Installation instruction
2 Brief introduction
2.1 Safety introduction
2.2 How to use this manual?
2.3 Handling method
to inextricable problem during operation
2.4 Warnings & Cautions
3 Summary
3.1 Brief introduction
3.2 Basic description of the
instrument:component descripition
3.3 controller
4 Instrumentation
4.1 calibration of the instrument(before
operation)
4.2 How to use the instrument?
4.3 Testing steps
4.4 Testing way
5 Routine maintenance

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

5.1 Warnings&Cautions
5.2 Important brief summary
5.3 Instrument calibration
6 troubleshooting
7. Oxygen index influence factor and
discussion
A Appendix A Technicl parameters brief
summary
B Appendix B Packing List

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

1. Installation preparation
1.1 Prepare working-area
Please ensure
the following preparation work ready before the installation
personnel of instrument arrive
Preparing working-area
Transport the instrument to the installation
site
Place the working plateform of the
instrument
Cautions


The instrument is working under high voltage and must be
processed by professional personel,otherwise will result in
instrument damage. Power supply requirement of the instrument.
stable voltage and frequency
. good grounding
1.2 Working plateform and Transportation
Please arrange and transport
working plateform as below requirement:

Ensure the working plateform is strong enough to stand whole
machine.
Cautions

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

Please dont push or pull the operating panel during


transportation,can only take the hard metal appearance as point of
force application for transportation.
1.3 Safety
Keep safety awareness in mind always
Operate the inside
component by professional personnel.
220V Working voltage of the instrument is 220V.
Warning
Please operate when power
supply interrupt in order to avoid hurt.
Cautions
Please connect
ground wire in the earthing makr,otherwise may result in instrument
damage.
1.4 Installation instruction

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

Open the
instrument,and place on the draught cupboard or working
plateform.Get the metal net out of incendiary cylinder and put glass
bead,and then put metal net,sample holder and then put glass
incendiary cylinder.
2

Connect
the output end of trimix behind the control cabinet with plastic
intensifier tube and host machine bottom and clamp tightly. Attention:
nitrogen and oxygen input match well the front nitrogen and oxygen
adjusting position, trimix output is in the middle.
3

Prepare oxygen
and nitrogen cylinder, install the reducing valve.Connect the
nitrogen and oxygen interface on the control cabinet behind board
and their cylinder reducing valve output with plastic intensifier
tube and clamp tightly. Cylinder and reducing valve should be
prepared by the user.
4Igniter is a firing gun which
is especially designed for ignition
5 Install the burning barrels on the
engine base.
2 Brief introduction
The chapter contains the following contents

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

. How to use this manual?



Handling method to inextricable problem during operation
Safety instruction
Warnings & Cautions
Please must read and underestand the
contents of this Chapter
2.1 Safety instruction
Incorrect
operation of this manual will result in possible personal injury.

In order to avoid personal injury, please must read the following


safety precautions firstly before operating any instrument and
follow strictly:
Keep safety awareness in mind always;
All contents
must be read when you start to operate the instrument or have
maintenance
Must be familiar with all
security features and warning words;
Must operate the

instrument or have maintenance strictly according to safety rules

of Nations.

Ensure all

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

operating personnel and maintenance personnel can use and


maintenance this instrument safely
Must inspect the
safety device every day and ensure its normal working
2.2 How to use this manual?
The purpose of this
manual is to help you to use and maintenance correctly and safely.
This manual contains the following
contents:
Installation preparation
Brief instruction
Summary
Instrumentation
Routine maintenance
troubleshooting
Oxygen index influence factor and
discussion
Appendix
2.3 Handling method
to inextricable problem during operation

When you meet inextricable
problem during operation or maintenance or feel confused how to
operate its next step, please stop operation and contact our company
immediately
Contact info:

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.


Name:Yunuo (Suzhou) Instrument Co.,Ltd.
26
Address:No 26,Muxuxi Rd,Jinqiao Development Zones,Mudu
Town,Suzhou
0512-82256088 Tel:0512-82256088
0512-68560932 Fax:0512-68560932
2.4 Warnings & Cautions
WARNINGS
CAUTIONS
In this manual,
WARNINGS:warn the operating personnel to avoid possible personal
injury; CAUTIONS :remind the operating personnel to prevent
possible instrument damage.The universal WARNINGS & CAUTIONS as
below can apply to all equipments in Yunuo.
User should never ignore the
safety performance of the instrument.

Laypeople may damage the equipment or cause
personal injury when he disassemble mechanical component of the
instrument or make reinstallation.
Make sure
all power supply are switch off before make maintenance or repair

Must ensure the proper grounding of the instrument any time and
ensure the proper connecting of the main power supply polarity

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

Keep in mind that the


power voltage is enough to cause personal injury.

If the tested material may release pungent smell or toxic gas,


please must connect exhaust equipment and drain the gas outside
safely.

More WARNINGS & CAUTIONS are included in this manual. These


additional safety precautions cant replace any other safety
standard which all user must follow.
3. Summary
YN61037
This chapter contains summary of YN61037 Digital readout oxygen
index Tester

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

3.1 Brief introduction


GB2406-2009

50 3

OI Tester is compliant with GB2406-2009 and Used for testing the


oxygen concentration (volume percent) needed when polymer is burning.
With the following features:
(1) Its simply structured, and easy to operate,
(2) It can be used as a means of identifying polymers
combustibility, also a relevant research tool. Thereby, we can get
a better understanding of the polymer burning.
(3) Suitable for testing plastic, rubber, bubble and other
materials combustibility.
(4) Its result is accurate. So, its an essential testing
instrument when researching and producing antiflaming materials.
3.2 Basic description of the instrument:
component description
Please learn
well the following component and their function before operating the

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

instrument
.

flowmeter

piezometer

3.3
The controller includes the following part:
1. display area
2. operational key
3.3.1 display area
LCD
3.3.2 operational key

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

power supply switch keyPress this


key to have electricity
flow adjusting valvePress this key
to adjust flow
reducing valvePress this key to adjust
pressure
potentionmeter adjusting knob

correct oxygen density value (behind the machine)

4 Instrumentation
4.1 calibration of the instrument(before
operation)
()
0.25-0.4Mpa
0.10.01Mpa
10L0.5L/min,

Full calibration: plug in and turn on the main valve of the

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

cylinder which has known oxygen density value (there is gasing


calibration value on the cylinder) and adjust the reducing valve,
pressure is 0.25-0.4Mpa;adjust the stabilivolt clockwise on the
lower right corner of the instrument panel, the pressure
display0.10.01Mpa;adjust theflow counter-clockwise on the
right side of the piezometer, flowmeter display 10L0.5L/min.The
display value no on the instrument is already compliant with known
oxygen density value, otherwise should adjust full and
counter-clockwise close stabilivolt
4.2 how to use?
This chapter
describe the basic using way. Please operate as follow steps:
1 GB2406-2009 41 /
14.1L/min
X
GB/T5454-1997 As per
GB2406-2009 requirement, mixed gas flow velocity in the burning
barrels is 41cm/s,and thus get the mixed gas flow is 14.1L/min
Can also calculate the required flow based on special standard.
Formula: burning barrel sectional area X flow velocity(pay attention
the the unit conversion).Some standard like: GB/T5454-1997 listed
the mixed gas flow directly, and adjust the flow based on it.
2

Making test sample


based on sample standard, use self-supporting material test sample

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

holder or non self-supporting material test sample holder,install


on the top of adjustable sliding sleeve.Can discharge the adjustable
sliding sleeve and put the test sample holder on the sliding sleeve
directly when the height is beyond of standard requirement.
3

99.2% 99.3%

Full scale adjusting..The most oxygen which user use is industry


oxygen,and its purity is different due to different producing area.So
please mark oxygen density when change the oxygen cylinder every
time.See the following steps:close the nitrogen cylinder,open the
oxygen cylinder and clean for several minitues.Make the control
cabinet is full of nitrogen,adjust thefull scale button behind
of the control cabinet,change the previous oxygen density to be oxygn
cylinder density.Generally,the reference value is 99.2% or about
99.3%.And also inquiry oxygen producer directly.
4
18%
25%
If still cant learn about the oxygen density based on documents
or experience,you can burn the test sample in the air.If the testing
sample is burning quickly,and the oxygen density is 18%;if the
testing sample extinguish immediately,please choose the oxygen

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

density 25% or more based on ignition difficulty and extingush


situation after no fire.
5
0.2-0.3MPa

0.05-0.15MPa

30

Close the flow adjusting valve and flo valve on the


flowmeter,open the oxygen cylinder valve and nitrogen cylinder
valve,adjust the reducing valve and make the pressure output about
0.2-0.3Mpa,then adjust pressure maintaining valve to keep oxygen and
nitrogen pressure on the control panel to be 0.05-0.15MPathen adjust
the flow valve,press the power supply switch and turn on it. (before
adjusting,please pull the hand knob cover and push in,the hand knob
will not rotate if not pull out.)Pay attention to the oxygen density
on the LDC screen and the mixed gas flow meets the required
flow.Please remember to do it carefully and slowly during
adjusting.The display will not have change immediately when move the
adjusting valve knob,because the air should be mixed completely in
the air cylinder.Waiting for 30 seconds after adjusting value is
stable and then wash the burning barrel to make the inside air to
be similar as adjustd mixed gas.

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

Ignite the test sample as per standard requirement after


washing( see the ignitor using as above-mentioned),move the fire ways
at once after test sample burning.
7
Record the testing process as per standard and fill the testing
report after test.
8

Turn off the instrument power supply; break the host power supply
switch and air source valve,keep the instrument and site clean and
clear.
4.3 Testing steps:
1 15 50mm

Get the standard testing sample at least 15pcs, number each of


them,mark any one side about 50mm,and insert another side to the
testing fixture in the burning barrel,hang the test sample on the
testing sample fixture and cover incendiary cylinder.
2
18%
21% 25%
Estimate the oxygen density based on experience or test sample
burning in the air.If the testing sample is burning quickly,and the

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

oxygen density is 18%;if the testing sample burning slowly or


intermittent, the oxygen density 21% ; and at least 25% when extingush
after no fire.
3 0.10.01Mpa
10L0.5L/min(
)()

0.10.01Mpa 10L/min
30S
Open the nitrogen pressure maintaining valve and oxygen pressure
maintaining valve, and the pressure value on the instrument is 0.1
0.01Mpa and adjust the flow together to make nitrogen and oxygen
mixed flow to be 10L0.5L/min(the max diameter of spherical float)
and the value displayed on the digital display window is just the
value of oxygen density.If increase the oxygen density,then increase
oxygen flow and decrease nitrogen flow;otherwise conversely.
4( 12mm20mm)
.
Use the ignitor(flame length is about 12mm-20mm) to ignite the
top of test sample and withdraw after the top of test sample is burning
completely.
5Start to timing,observe
burning and record burning phenomenon.
6
Change the test sample after burning and repeat the previous steps
and adjust oxygen density based on burning phenomenon.

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

4.4 Testing method


A Top ignition method
1

Ignite the ignitor and make the flame bottom visible part can
touch test sampe top and cover the whole top surface.Dont make the
flame to touch edge and side surface of the test sample.
2 180

Timing is over when the test sample meets or exceeds the marking
line in 180seconds;adjust the oxygen density,wipe the burning barrel
and surface dirt on the ignitor.Make the next experiment after the
temperature of burning barrel return back to normal or change another
burning barrel with normal temperature.
3 30s

Move the ignitor if the material are not on ignition within 30


seconds,and adjust oxygen density and start experiment again.
BDiffused ignition method
1
6mm
30s 30s 180s 5s

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

Ignite the ignitor ,down and move the ignitor to make the flame
visible part can touch the top surface of test sampe and put on
vertical side surface about 6mm.Start timing.The longest action time
of flame is 30s and start to timing for 180s after 30s.Move the ignitor
left and right and observe the test sample every 5s til the vertical
side surface burning or visible burning part reach marking line,then
move the ignitor,start to timing and observe the the burning length
of the test sample.
2 180

Timing is over when the test sample meets or exceeds the marking
line within 180s.Adjust oxygen density,wipe the burning barrel and
surface dirt on the ignitor to make the temperature of burning barrel
return back to normal or change another burning barrel with normal
temperature.
3
Switch off the power supply ,air source and clean the residue
after testing.
5 Routine maintenance
5.1 Warnings& Cautions
Please read the
following contents carefully before any operation:
Warnings
5.2 Important brief summary

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

Cautions
1 1MPa

The inside pressure of oxygen cylinder and nitrogen cylinder


should not be less than 1Mpa,Please make replacement on time if the
pressure is lower than 1Mpa.
2Ignite the ignitor with lighter.
3
Adjust the flow evenly,dont be unstable.
4

Wipe the burning barrel and surface dirt of the ignitor termly
and clean the burning dripping on the metal net of burning barrel
to keep the smooth air.
5
Please contact us if any problem.Dont discharge the instrument
by yourself,expecially for the component.
6
Friendly reminder:Keep the oxygen cylinder and nitrogen
cylinder far away from fire source and be safety,as they are high
pressure vessel.
5.3 Calibration instrument

Usually,calibrate the instrument termly.Calibrate the


instrument once the calibration value is beyond of permissible

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

error range.

Please inpect the instrument calibration value every year to
ensure the same value as experiment result.
6 Troubleshooting


If any problem of the instrument or misoperation,and still cant
solve as per our listed dealing method,please contact
us:breakdown,possible reasons,inspect steps,solution.
7 Oxygen index influence factor and
discussion
Due to too many
factors,now analyse main factors only:
1 gas-flow rate

,

The faster of the gas-flow rate,the more oxygen flow in the unit
time and it will be helpful for test sample burning;on the other
hand,overlarge gas-flow rate will take away mass heat and influence
burning.
2 Oxygen density and purity

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

Obviously,the higher oxygen density will be useful for test


sample burning;otherwise,its bad for test sample
burning.Therefore,oxygen density control and gas system joint should
be calibrated termly.
100

Usually,we regard oxygen and nitrogen purity to be 100%,but there


is a little errorfor industrial gas.The lower purity,the bigger
error;by the way,the decline of pressure in the cylinder has
influence on the oxygen density.
3

Test sample size&making test sample burning speed is related


with test sample shape,size and making method.
4

Pressure control Some polymer oxygen index is sensitive for the


decompression,it may resulted by continously removing the propduct
and speed up degradation of the polymer.
5

Temperature Temperature has big influence on the oxygen

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

index.It has little influence under room temperature but has various
influence on the polymer in high temperature.
6

Plasticizer and padding The testing value of oxygen index is


always the one which most volatile component,thats oxygen index
of plasticizer.

A Appendix A Technicl parameters brief summary


Technicl parameters as below:
1 inner diameter of burning barrel 100mm
2 height of burning barrel 450mm
3 flowmeter precision 2.5 level
4 piezometer precison 2.5 level
5 piezometer precision resolution ratio
0.01MPa
6 resolution ratio of oxygen density
indicator 0.1%
7 air sourceGB3863 GB3864
GB3863 require oxygen cylinder to have pressure
regulating valve,GB3864 require nitrogen cylinder to have pressure
regulating valve.
8 test environment temperature 10-35
humidity 45-75%
9 pressure input0.2-0.3MPa
10 working pressure0.05-0.15MPa

REV A

Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.

11 testing type
self-supporting material and non
self-supporting material
12 power supply220VAC50Hz
13 electricity DC 1A

REV A