Professional Documents
Culture Documents
Model No.
Contents
1 Installation preparation
1.1 Preparing working-area
1.2 Working plateform and Transportation
1.3 Safety
1.4 Installation instruction
2 Brief introduction
2.1 Safety introduction
2.2 How to use this manual?
2.3 Handling method
to inextricable problem during operation
2.4 Warnings & Cautions
3 Summary
3.1 Brief introduction
3.2 Basic description of the
instrument:component descripition
3.3 controller
4 Instrumentation
4.1 calibration of the instrument(before
operation)
4.2 How to use the instrument?
4.3 Testing steps
4.4 Testing way
5 Routine maintenance
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
5.1 Warnings&Cautions
5.2 Important brief summary
5.3 Instrument calibration
6 troubleshooting
7. Oxygen index influence factor and
discussion
A Appendix A Technicl parameters brief
summary
B Appendix B Packing List
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
1. Installation preparation
1.1 Prepare working-area
Please ensure
the following preparation work ready before the installation
personnel of instrument arrive
Preparing working-area
Transport the instrument to the installation
site
Place the working plateform of the
instrument
Cautions
The instrument is working under high voltage and must be
processed by professional personel,otherwise will result in
instrument damage. Power supply requirement of the instrument.
stable voltage and frequency
. good grounding
1.2 Working plateform and Transportation
Please arrange and transport
working plateform as below requirement:
Ensure the working plateform is strong enough to stand whole
machine.
Cautions
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
Open the
instrument,and place on the draught cupboard or working
plateform.Get the metal net out of incendiary cylinder and put glass
bead,and then put metal net,sample holder and then put glass
incendiary cylinder.
2
Connect
the output end of trimix behind the control cabinet with plastic
intensifier tube and host machine bottom and clamp tightly. Attention:
nitrogen and oxygen input match well the front nitrogen and oxygen
adjusting position, trimix output is in the middle.
3
Prepare oxygen
and nitrogen cylinder, install the reducing valve.Connect the
nitrogen and oxygen interface on the control cabinet behind board
and their cylinder reducing valve output with plastic intensifier
tube and clamp tightly. Cylinder and reducing valve should be
prepared by the user.
4Igniter is a firing gun which
is especially designed for ignition
5 Install the burning barrels on the
engine base.
2 Brief introduction
The chapter contains the following contents
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
of Nations.
Ensure all
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
Name:Yunuo (Suzhou) Instrument Co.,Ltd.
26
Address:No 26,Muxuxi Rd,Jinqiao Development Zones,Mudu
Town,Suzhou
0512-82256088 Tel:0512-82256088
0512-68560932 Fax:0512-68560932
2.4 Warnings & Cautions
WARNINGS
CAUTIONS
In this manual,
WARNINGS:warn the operating personnel to avoid possible personal
injury; CAUTIONS :remind the operating personnel to prevent
possible instrument damage.The universal WARNINGS & CAUTIONS as
below can apply to all equipments in Yunuo.
User should never ignore the
safety performance of the instrument.
Laypeople may damage the equipment or cause
personal injury when he disassemble mechanical component of the
instrument or make reinstallation.
Make sure
all power supply are switch off before make maintenance or repair
Must ensure the proper grounding of the instrument any time and
ensure the proper connecting of the main power supply polarity
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
50 3
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
instrument
.
flowmeter
piezometer
3.3
The controller includes the following part:
1. display area
2. operational key
3.3.1 display area
LCD
3.3.2 operational key
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
4 Instrumentation
4.1 calibration of the instrument(before
operation)
()
0.25-0.4Mpa
0.10.01Mpa
10L0.5L/min,
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
99.2% 99.3%
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
0.05-0.15MPa
30
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
Turn off the instrument power supply; break the host power supply
switch and air source valve,keep the instrument and site clean and
clear.
4.3 Testing steps:
1 15 50mm
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
0.10.01Mpa 10L/min
30S
Open the nitrogen pressure maintaining valve and oxygen pressure
maintaining valve, and the pressure value on the instrument is 0.1
0.01Mpa and adjust the flow together to make nitrogen and oxygen
mixed flow to be 10L0.5L/min(the max diameter of spherical float)
and the value displayed on the digital display window is just the
value of oxygen density.If increase the oxygen density,then increase
oxygen flow and decrease nitrogen flow;otherwise conversely.
4( 12mm20mm)
.
Use the ignitor(flame length is about 12mm-20mm) to ignite the
top of test sample and withdraw after the top of test sample is burning
completely.
5Start to timing,observe
burning and record burning phenomenon.
6
Change the test sample after burning and repeat the previous steps
and adjust oxygen density based on burning phenomenon.
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
Ignite the ignitor and make the flame bottom visible part can
touch test sampe top and cover the whole top surface.Dont make the
flame to touch edge and side surface of the test sample.
2 180
Timing is over when the test sample meets or exceeds the marking
line in 180seconds;adjust the oxygen density,wipe the burning barrel
and surface dirt on the ignitor.Make the next experiment after the
temperature of burning barrel return back to normal or change another
burning barrel with normal temperature.
3 30s
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
Ignite the ignitor ,down and move the ignitor to make the flame
visible part can touch the top surface of test sampe and put on
vertical side surface about 6mm.Start timing.The longest action time
of flame is 30s and start to timing for 180s after 30s.Move the ignitor
left and right and observe the test sample every 5s til the vertical
side surface burning or visible burning part reach marking line,then
move the ignitor,start to timing and observe the the burning length
of the test sample.
2 180
Timing is over when the test sample meets or exceeds the marking
line within 180s.Adjust oxygen density,wipe the burning barrel and
surface dirt on the ignitor to make the temperature of burning barrel
return back to normal or change another burning barrel with normal
temperature.
3
Switch off the power supply ,air source and clean the residue
after testing.
5 Routine maintenance
5.1 Warnings& Cautions
Please read the
following contents carefully before any operation:
Warnings
5.2 Important brief summary
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
Cautions
1 1MPa
Wipe the burning barrel and surface dirt of the ignitor termly
and clean the burning dripping on the metal net of burning barrel
to keep the smooth air.
5
Please contact us if any problem.Dont discharge the instrument
by yourself,expecially for the component.
6
Friendly reminder:Keep the oxygen cylinder and nitrogen
cylinder far away from fire source and be safety,as they are high
pressure vessel.
5.3 Calibration instrument
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
error range.
Please inpect the instrument calibration value every year to
ensure the same value as experiment result.
6 Troubleshooting
If any problem of the instrument or misoperation,and still cant
solve as per our listed dealing method,please contact
us:breakdown,possible reasons,inspect steps,solution.
7 Oxygen index influence factor and
discussion
Due to too many
factors,now analyse main factors only:
1 gas-flow rate
,
The faster of the gas-flow rate,the more oxygen flow in the unit
time and it will be helpful for test sample burning;on the other
hand,overlarge gas-flow rate will take away mass heat and influence
burning.
2 Oxygen density and purity
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
index.It has little influence under room temperature but has various
influence on the polymer in high temperature.
6
REV A
Instrument Name Digital Readout Oxygen Index Tester YN61037
Model No.
11 testing type
self-supporting material and non
self-supporting material
12 power supply220VAC50Hz
13 electricity DC 1A
REV A