You are on page 1of 6

Measuring Cardiac Output: Intermittent Bolus Thermodilution Method

Anna Gawlinski

Crit Care Nurse. 2004;24: 74-78


© 2004 American Association of Critical-Care Nurses
Published online http://ccn.aacnjournals.org

Personal use only. For copyright permission information:


http://ccn.aacnjournals.org/cgi/external_ref?link_type=PERMISSIONDIRECT

Subscription information
http://ccn.aacnjournals.org/subscriptions

Information for authors


http://ccn.aacnjournals.org/misc/ifora.shtml

Submit a manuscript
http://www.editorialmanager.com/ccn

Email alerts
http://ccn.aacnjournals.org/subscriptions/etoc.shtml

Critical Care Nurse is the official peer-reviewed clinical journal of the American
Association of Critical-Care Nurses, published bi-monthly by The InnoVision Group,
101 Columbia, Aliso Viejo, CA 92656. Telephone: 949-362-2000. Fax:
949-362-2049. Copyright 2004 by AACN. All rights reserved.

Downloaded from ccn.aacnjournals.org by on August 16, 2010


ProtocolsforPractice

by a computer and converted into a


Measuring Cardiac measurement of cardiac output. Car-
diac output is inversely proportional

Output: Intermittent Bolus to the area under the curve.


The normal cardiac output curve
has a rapid smooth upstroke and a
Thermodilution Method gradual downstroke (Figure 1). A
small area under the curve indicates
Anna Gawlinski, RN, DNSc, CS-ACNP a high cardiac output. The faster
blood flows through the heart, the
earlier the peak and sharper the drop,
because the catheter senses temper-

Q What are the physiological


principles underlying the intermittent
proximal port of a pulmonary artery
catheter located in the right atrium.
ature change over a short period. A
low cardiac output results in a larger
area under the curve. When blood
flows slowly (low cardiac output), the
bolus thermodilution (TDCO) method In the atrium, the injectate mixes with area under the curve (temperature
of determining cardiac output? the blood and passes through the tri- change over time) is greater because
cuspid valve into the right ventricle. the catheter senses changes in tem-
The TDCO method is the one clini- A thermistor within the catheter perature over a longer period. The
cians use most often to measure cardiac senses the change in blood tempera- curves vary according to the patient’s
output. It is based on the principles of ture as the blood passes the catheter clinical condition and according to
dilution. A known quantity of an indi- tip located in the pulmonary artery.1 deviations in technique (Figure 2).
cator (ie, a contrast agent) is injected A curve that shows the change in Concerns about contamination
into the bloodstream. Blood flow and temperature over time is calculated of prefilled syringes of injectate and
blood volume are calculated by meas-
uring the concentration of the indica-
tor downstream at a distal arterial site
at selected times. The TDCO method Peak
uses a cold solution to create a thermal 80%
Decreasing temperature

deficit as a variant of the indicator-dilu-


tion method. A bolus of sterile solution Curve onset 30%
(ie, the injectate) that is colder than
the patient s blood is injected into the
Baseline Downslope Extrapolated area

Author Time
Anna Gawlinski is a clinical nurse spe-
cialist in the cardiac care unit and car-
diac observation unit at UCLA Medical Figure 1 An ideal thermodilution curve. Administration of the injectate is charac-
terized by a rapid upslope to a peak, a gradual downslope, and an exponential
Center and is an assistant professor of decay of the thermal signals. The cardiac output computer begins integration of
nursing at the UCLA School of Nursing, the area under the thermodilution curve at the instant of injection and terminates
Los Angeles, Calif. integration when the exponential decay reaches a value of about 30%. The com-
To purchase reprints, contact The InnoVision Group, puter then extrapolates the exponential decay to baseline. In this way, any artifact
101 Columbia, Aliso Viejo, CA 92656. Phone, (800) introduced by recirculation of indicator is minimized.
809-2273 or (949) 362-2050 (ext 532); fax, (949)
362-2049; e-mail, reprints@aacn.org.

74 CRITICALCARENURSE Vol 24, No. 5, OCTOBER 2004


Downloaded from ccn.aacnjournals.org by on August 16, 2010
warming of injectate due to handling
led to the design of closed delivery sys- Normal curve
tems (Figure 3). The closed system is Injection Smooth upstroke to peak,

Temperature
designated for use with a pulmonary then gradual downslope
to baseline
artery catheter and a cardiac output Computer looks for a
computer or with iced and room-tem- smooth curve
Time
perature injectate. The closed system
also incorporates a flow-through tem- Variation in normal curve Computer extrapolates
perature probe that measures the tem- Curve downslope to calculate

Temperature
Injection area under curve
extrapolated
perature of the injectate near the site Respiratory variation
of injection. The obvious advantages Increase baseline from
recirculation of injectate
are that closed systems do not require
Time Respiratory variation
preparation of individual syringes,
eliminate inefficiencies, and reduce A
Normal high cardiac output
multiple entries into a sterile system.2
Small area under the
Temperature

Injection
curve is typical of a
Q: How accurate is the TDCO high cardiac output
method? (small change in injectate
temperature over time)
Time
The accuracy of the method is
Normal low cardiac output
related to how closely the observed
signal (ie, measurement of cardiac
Temperature

Injection
Large area under the curve
output) matches an accepted standard seen in patients with low
value. Forrester et al3 found a corre- cardiac output (greater
lation coefficient of 0.993 between a change in temperature
over time)
mechanical pump with a known flow Time
and TDCO measurements. In other Uneven injection technique
Temperature

studies4-6 with flowmeters, correlation


coefficients were 0.97 to 1.0. Other Injection
researchers have used the direct Fick Uneven upstroke on curve
method as the gold standard for meas-
Temperature

urement of cardiac output. The Fick Time


Injection
principle states that the amount of a Severe artifact on both
substance taken up by the body per upstroke and downstroke
of curve
unit of time is equal to the arterial
Time
level of the substance minus the B
Prolonged injection time
venous level of the substance multi-
Temperature

Injection
plied by the flow.6,7 Cardiac output
Injectate delivered in over
can be calculated by the Fick method 4 seconds
by dividing the amount of oxygen 10
consumed by the body by the arterial- seconds
Time
venous oxygen difference.8 Numerous
Figure 2 A, Variations in the normal cardiac output curve seen in certain
studies4,9-14 have addressed the corre- clinical conditions. B, Abnormal cardiac output curves that will produce an
lation between the direct Fick method erroneous cardiac output value.
and the TDCO method. In all but 2 Reprinted with permission from Love M, Lough ME, Bloomquist J. Cardiovascular laboratory
assessment and diagnostic procedures. In: Thelan LA, Davie JK, Urden LD. Textbook of Critical Care
studies,15,16 the correlations between Nursing: Diagnosis and Management. St Louis, Mo: CV Mosby;1990:246.

values obtained with the 2 methods

CRITICALCARENURSE Vol 24, No. 5, OCTOBER 2004 75


Downloaded from ccn.aacnjournals.org by on August 16, 2010
ProtocolsforPractice

Q: Is the continuous cardiac out-


put (CCO) method as accurate is the
IV solution
TDCO method?
container
Clinical studies have also com-
pared the accuracy of the CCO and
IV spike
the TDCO methods. These studies
Snap Clamp are synthesized and critiqued in the
“Annotated Bibliography” section of
To continuous the research-based protocol on meas-
flush solution
urement of cardiac output.19 Although
Thermodilution the studies report good correlations
Injectate
catheter
temperature probe (0.84 to 0.94) between measurements
obtained with the 2 methods, com-
parisons between CCO and TDCO
Syringe barrel
with thermal shield
values have intrinsic methodologi-
cal issues. The TDCO method itself
is not a gold standard for measure-
ment of cardiac output. Under ideal
Right atrial Syringe plunger with circumstances, TDCO measurements
injection hub contamination sheath have a 10% error rate related to oper-
check
3-Way valve ator error, temperature transduction,
stopcock Flow-through housing
Reusable and instrument inaccuracies. Studies
injectate that compare values obtained with
probe
assembly the CCO method with values obtained
with an inaccurate gold standard such
as the TDCO method may lead to
erroneous estimation of the accuracy
Cardiac output cable of the CCO device. Further studies
Figure 3 Schematic illustration of the closed injectate delivery system for use are necessary to compare CCO values
with iced injectate for determination of cardiac output. with more precise measurement of
Reprinted with permission from Abbott Critical Care, Mountain View, Calif. cardiac output (such as those obtained
by using the Fick or the indicator-
dilution method) in a variety of criti-
were 0.91 to 0.98.6 The TDCO Several technical considerations must cally ill patients. Such confirmatory
method is an acceptable substitute be understood in order to minimize studies are necessary to verify the
for the Fick method or the indicator- the potential for error. These include accuracy of CCO technology.
dilution method. the position of the pulmonary artery Other methodological problems
Many of the derived hemody- catheter, volume and temperature of limit the generalizability of these
namic indexes and the clinical thera- the injectate, the phase in the respi- studies to other critically ill popula-
pies for critically ill patients depend ratory cycle for administration of the tions. The majority of the studies
on accurate measurement of cardiac injectate, the patient’s body position, were done during periods of hemo-
output. Critical care nurses routinely effects of concomitant intravenous dynamic stability in patients who had
use the TDCO method to measure infusions, and the effect of positive had coronary artery bypass graft sur-
cardiac output in critically ill patients end-expiratory pressure. Even under gery. Many of these studies used
and are responsible for the accuracy ideal circumstances, TDCO measure- small sample sizes (N = 12 to 35) and
and analysis of the obtained values. ments have a 10% error.17,18 multiple observations. Therefore,

76 CRITICALCARENURSE Vol 24, No. 5, OCTOBER 2004


Downloaded from ccn.aacnjournals.org by on August 16, 2010
measurements within each study Q: What is the procedure for tate may decrease the temperature
were not completely independent obtaining a TDCO measurement? difference (signal) and may yield erro-
of one another. For example, if a neous values for cardiac output. In
subject in a study had some charac- For accurate TDCO curves, the most normothermic patients, 5 mL
teristic that led to errors in measure- signal-to-noise ratio must be ade- of iced injectate can be used if fluid
ment, multiple measurements in quate for the cardiac output moni- restriction is warranted. Before
that subject would lead to overrepre- tor to sense a change in temperature using a smaller volume, clinicians
sentation and possibly negative bias over time. The signal is the tempera- must verify that the values obtained
in the results. Using larger sample ture difference between the injectate with the smaller volume are compa-
sizes and limiting the number of and the patient’s blood; the noise is rable to values obtained with a
data points per patient would con- the cycling variation in blood tem- larger volume.
trol for this possible variation. perature. The difference between • To ensure the validity and reli-
Power analysis was not used to the temperature of the injectate and ability of the measurement, check
determine appropriate sample size the temperature of the patient’s the following:
(see the “Annotated Bibliography” blood should be 10˚C.2 Theoretically, – position of the pulmonary
in the protocol19 for more details). 10 mL of iced injectate produces a artery catheter,
Rigorous studies that control for the greater signal-to-noise ratio than does – computation constant,
issues just described are needed in 10 mL of room-temperature injectate – catheter size,
critically ill patients to establish the or smaller volumes (eg, 5 mL) of – temperature of the
accuracy of newer technologies of either iced or room-temperature injectate,
noninvasive measurements of car- injectate. Use of room-temperature – volume of injectate, 10 mL
diac output. injectate or smaller volumes of injec- (or 5 mL if iced), and

Downloaded from ccn.aacnjournals.org by on August 16, 2010


ProtocolsforPractice

– patient’s body position. as patients who have low ejection 9. Kadota LT. Theory and application of ther-
modilution cardiac output measurement: a
• Use appropriate TDCO tech- fraction, low cardiac output, high review. Heart Lung. 1985;14;605-616.
10. Wyse SD, Pfitzner J, Rees A, Lincoln JCR,
nique: cardiac output, or hypothermia. Branthwaite MA. Measurement of cardiac
– Inject 10 mL within 4 • Replicate studies with small output by thermal dilution in infants and
children. Thorax. 1975;30:262-265.
seconds; 5 mL, within 2 seconds. volumes of injectate in subgroups of 11. Freed MD, Keane JF. Cardiac output meas-
ured by thermodilution in infants and chil-
– Administer the injectate critically ill patients, such as patients dren. J Pediatr. 1978;92:39-42.
when the patient is at end- who have low ejection fraction, low 12. Branthwaite MA, Bradley RD. Measurement
of cardiac output by thermodilution in
expiration. cardiac output, high cardiac output, man. J Appl Physiol. 1968;24:434-438.
or hypothermia. 13. Vandermoten P, Bernard R, de Hamptinne
– Obtain 3 measurements. J, Gillet JM, Lenaers A. Cardiac output mon-
– Assess the cardiac output • Replicate studies comparing itoring during the acute phase of myocar-
dial infarction: accuracy and precision of
curve. the accuracy of closed system of the thermodilution method. Cardiology.
– Average values that are injectate delivery with the accuracy 1977;62:291-295.
14. Zisserman D, Mantle JA, Smith LR, Rogers
within 10% of the median value. of using prefilled syringes in sub- WJ, Russell RO Jr, Rackley CE. Clinical com-
parison of thermal dilution cardiac output
– Starting at baseline, a groups of critically ill patients. to the Fick and angiographic methods
normal cardiac output curve has • Replicate studies in patients with [abstract]. Clin Res. 1979;27:736A.
15. Hodges M, Downs JB, Mitchell LA. Ther-
a smooth rapid upstroke and a cardiac conditions in which the TDCO modilution and Fick cardiac index determi-
nations following cardiac surgery. Crit Care
gradual downstroke. method is considered less accurate but Med. 1975;3:182-184.
for which cardiac output and other 16. Stawicki JJ, Holford FD, Michelson EL,
Josephson ME. Multiple cardiac output meas-
Q: What further research on meas- hemodynamic measurements are urements in man. Chest. 1979;76:193-197.
17. Moore FA, Haenel JB, Moore EE. Alterna-
urement of cardiac output is needed? used, such as valvular disease (tricus- tives to Swan-Ganz cardiac output monitor-
pid regurgitation), dysrhythmias, and ing. Surg Clin North Am. 1991;71:699-721.
18. Burchell SA, Yu M, Takiguchi SA, Ohta RM,
Although many research studies dilated heart chambers with increased Myers SA. Evaluation of a continuous car-
diac output and mixed venous oxygen satu-
have been done on measurement of ventricular dimensions. ration catheter in critically ill surgical
cardiac output and on newer meth- • Replicate studies of the effects patients. Crit Care Med. 1997;25:388-391.
19. Gawlinski A. Cardiac Output Monitoring.
ods, additional investigation and of the patient’s body position on Aliso Viejo, Calif: American Association of
TDCO measurements in critically ill Critical Care Nurses: 1997.
replication are needed. Further
research is needed to do the following: patients. Note
• Describe chronobiological fluc- This article was first published in
tuations in cardiac output. References
1. Kallasian KG, Raffin, TA. The technique of Critical Care Nurse April 2000.
• Determine the validity and thermodilution cardiac output measurement.
J Crit Illness. 1996;11:249-256.
reliability of noninvasive methods of 2. Gardner PE, Bridges ET. Hemodynamic This article is based on the protocol
measuring cardiac output, such as monitoring. In: Woods SL, Sivarajan-
Froelicher ES, Halpenny CJ, Motzer SU, eds. Cardiac Output Monitoring by
Doppler flow imaging and echocar- Cardiac Nursing. 3rd ed. Philadelphia, Pa:
Anna Gawlinski. It was taken from
JB Lippincott; 1995:424-458.
diography and thoracic electric bioim- 3. Forrester JS, Ganz W, Diamond G, McHugh the Hemodynamic Monitoring series
pedence, in various populations of T, Chonette DW, Swan HJC. Thermodilution
cardiac output determination with a single (Product #170709) of AACN’s
critically ill patients. flow-directed catheter. Am Heart J. 1972; Protocols for Practice. Protocols can
83:306-311.
• Determine the validity and reli- 4. Hoel BL. Some aspects of the clinical use of be obtained from AACN, 101
ability of values obtained by using thermodilution in measuring cardiac output.
Scand J Clin Lab Invest. 1978;38:383-388. Columbia, Aliso Viejo, CA 92656-
the CCO method in various popula- 5. Sanmarco ME, Philips CM, Marquez LC, 1491, (800) 899-AACN, (949) 362-
Hall D, Davila JA. Measurement of cardiac
tions of critically ill patients. output by thermodilution. Am J Cardiol. 2000. Product #170704: $11, AACN
• Determine the effect on clinical 1971;29:54-58.
members; $14, nonmembers.
6. Sommers MS, Woods SL, Courtade MA.
outcomes and cost of using CCO Issues in methods and measurement of
versus TDCO methods in critically thermodilution cardiac output. Nurs Res.
1993;42:228-233.
ill patients. 7. Ganong WF. Review of Medical Physiology.
16th ed. East Norwalk, Conn: Appleton &
• Replicate studies with iced and Lange; 1993.
room-temperature injectate in sub- 8. Berne RM, Levy MN. Cardiovascular Physi-
ology. 6th ed. St Louis, Mo: Mosby-Year
groups of critically ill patients, such Book; 1992.

78 CRITICALCARENURSE Vol 24, No. 5, OCTOBER 2004


Downloaded from ccn.aacnjournals.org by on August 16, 2010

You might also like