Professional Documents
Culture Documents
RESULTS
Figure 2. Freeze cycle: Gas to be processed is contained in Results are shown in Table 1. The mean freezing
bag (A). This is pumped via an inlet port (B) through the chamber temperature was 139.2 (0.1), at an ambi-
freezing chamber at 139.2C (C) where the xenon selec- ent temperature of 19.5 (1.7)C mean (sd). It was
tively solidifies. The remaining gases (efflux gas) emerge noted that xenon always started to thaw at a chamber
from the distal end of the chamber, pass through the xenon temperature of between 107C and 108C, in-
analyzer (D) and are directed by the outlet valve (E) to
atmosphere via hose (F). The xenon analyzer displays the keeping with the quoted boiling point of 107C
xenon content of the efflux gas which should be low. Thaw 3C (32). Purity and yield are shown graphically in
cycle: The gas pump stops, inlet port (B) automatically (Fig. 4).
closes, the chamber temperature increases to 100C where- The volume of gas processed was measured by
upon the xenon in the freezing chamber (C) thaws, becomes summation of the gas volume in the xenon and waste
gaseous and expands out via analyzer (D) to the outlet valve
(E), which now directs the xenon via hose (G) to a collection gas collection bags after each run (Table 1, columns 3
bag. The valves are all oxygen compatible (Model E3A. ACL and 5). There was a small difference between the
srl. Caponago, Italy). The microprocessor system (H) con- calculated total xenon content of the gas being pro-
trols the valves, gas pump, and the temperature of the cessed and that of the same gas once processed. This
chamber during each freeze/thaw cycle. The refrigeration could result, for example, from under-measurement of
unit forms the base of the machine (I).
the xenon collection bag or over-measurement of the
waste gas bag volumes. The opposite measurement
error could cause a slight apparent xenon gain.
The machine functioned automatically under com-
Yield calculations could therefore be performed in two
puter control. Once the chamber had cooled to
different ways. Although there is very little practical
139.2C, the inlet valve opened and a gas pump
difference between the results of either calculation, for
propelled the gas to be processed through the freezing
the purpose of Figure 4, we have erred on the side of
chamber. The xenon froze, forming xenon ice,
caution and presented the lowest, most conservative
whereas the unfrozen efflux waste gas from the cham-
yield values.
ber outlet was directed to a gas collection bag. After 5
The xenon concentration in the efflux gas was
L of gas had been processed, the pump stopped, the
rather independent of the concentration in the input
inlet valve closed and the chamber warmed to
gas (Fig. 5). The freezing chamber temperature influ-
100C, allowing frozen xenon to thaw to gas. The
enced the efficiency of recovery, as evidenced by the
chamber outlet valve changed, to direct the expanding
percentage of xenon present in the efflux gas, which
xenon gas to a xenon collection bag. After each run,
increased as temperature increased (Fig. 6).
the volume of gas in the efflux waste and xenon
storage bags (V1) and (V2) was measured by emptying
each bag in aliquots with a calibrated 200-mL syringe. DISCUSSION
The purity of recovered xenon was measured using This study demonstrated that the device developed
the xenon analyzer. This analyzer has been extensively for cryogenic recovery of xenon from waste anesthetic
evaluated and found to be very accurate relative to a gases was able to extract xenon reliably, and that the
1314 Selective Xenon Recovery System ANESTHESIA & ANALGESIA
Figure 3. Control algorithm for the machine.
Table 1. Performance of Machine with a Series of Test Gas Mixtures of Varying Xenon Content
Efflux waste Xenon recovery Xenon recovery
Input gas Input gas xenon Efflux waste gas gas xenon bag volume bag xenon
volume (mL) concentration (%) volume (mL) concentration (%) (mL) concentration (%)
5100 48.9 2950 7.5 2150 96.1
5650 39.0 3650 7.0 2000 98.2
4610 27.9 3400 6.8 1210 94.1
4730 20.4 3950 6.8 780 90.4
4560 10.6 4300 6.8 260 72.5
5940 48.2 3350 7.2 2590 95.8
6160 40.0 3900 7.1 2260 96.1
5560 29.8 4150 7.0 1410 94.8
5890 20.8 4900 6.8 990 92.2
5600 10.7 5300 6.8 300 73.4
6600 49.2 3620 7.3 2980 97.7
6360 39.7 4040 7.4 2320 96.5
6290 30.0 4700 7.1 1590 95.3
5545 20.1 4690 7.1 855 92.4
5610 10.5 5330 7.1 280 76.7
All gas processed at 300 mL/min. Ambient temperature 19.5 (1.7)C. Cooling chamber temperature 139.2 (0.1)C Mean (SD).
Vol. 105, No. 5, November 2007 2007 International Anesthesia Research Society 1315
gas as it passes through the freezing chamber can be
estimated, from the heat capacities of oxygen and
xenon, to be approximately 52 J/min. Improving heat
transfer properties where the chamber and its outer
cooling sleeve are in contact would permit both
faster gas flows and more rapid freeze/thaw cycling
times.
For a 70% xenon, 30% O2 inhaled gas mixture
(approximately 1 MAC), collection of useful gas for
recovery could include: gas flushed from a closed
breathing system to counteract N2 build up, xenon-
rich gas remaining in the circle and that eliminated
from the body in the first few exhaled breaths at the
end of an anesthetic. This would likely produce sev-
eral liters of gas with 20% xenon content, from
which our machine should be able to recover 70% of
the xenon.
For xenon, very low-flow circuits are not as efficient
as one might assume and it is clear that xenon
Figure 4. Purity and yield of recovered xenon relative to the recovery in this situation would greatly reduce overall
xenon content of the input gas (flow, 300 mL/min; tempera- xenon use. For example, Burov et al. (20,35) described
ture, 139.2C).
a very-low flow protocol in which anesthesia is main-
tained with FGFs of 300 mL/min xenon and 300
purity of the extracted xenon exceeded 90%. The mL/min oxygen. If, during maintenance, we assume
percentage unfrozen xenon in the efflux gas is mainly typical uptake values of 60 mL/min xenon and 250
determined by the temperature of the freezing cham- mL/min oxygen (metabolic), this would still generate
ber and independent of the input gas xenon fraction. an overspill or wastage rate of 240 mL/min (14.4 L/h)
Since the experiments were performed at atmospheric xenon (i.e., 80% of the xenon FGF) and 3 L/h oxygen.
pressure, the expected percentage of unfrozen xenon A closed-circuit breathing system would be a better
can be derived from the vapor pressure. At 139.2C, choice of breathing system as it would minimize the
the machine operated exactly as predicted from pub- volume of gas to be processed (36).
lished physical chemistry data; however, at tempera- From these closed-circuit breathing systems, the
tures close to its freezing point, the efficiency of xenon waste gas collected would mainly be produced by
recovery appeared to be much better than predicted denitrogenation circle flushes with fresh gas, re-
(Fig. 6) (32). This requires some consideration. sidual xenon purged from the circle, and the initial
The published vapor data in the 139 to 112 exhaled gas collected during emergence from anesthe-
temperature range are for xenon gas in equilibrium sia. This gas would again be rich in xenon. By extrapo-
(by sublimation) with frozen solid xenon. However, lation of data from Luttropp et al. and Ferrari et al. (in
before multiple layers of solid xenon can form, gas which a denitrogenation flush was included) a 2-h
will first be adsorbed onto the surfaces of the copper closed circle anesthetic would consume 10.9 L ($109)
rings in the freezing chamber. Xenon adsorbs much and 18.6 L ($186) of xenon, respectively (23,24). Fur-
more readily to copper than to its own solidified thermore, by returning all waste gas collected in this
molecules (i.e., xenon ice); therefore, the intense way to the xenon manufacturer, Ferrari et al. (24)
cooling we used may not be required if the copper managed to recover 60% of the total xenon used. The
surface area can be increased (34). cost of industrially reprocessed xenon has been
Also, the xenon content decreases as the gas moves estimated at $3 per liter, and thus if 60% could be
along the chamber. Further xenon can only reach the recycled, this would reduce the above estimates for a
distal chamber by diffusion from the proximal section 2-h anesthetic from $109 $186 to $63$108. It is likely,
(diffusion limited transport). therefore, that even with closed-circuit breathing sys-
It is clear that we observed the mixed effects of tems, a recovery device could still usefully reduce
several processes. Potential improvements to the de- overall xenon consumption per anesthetic, a 50%
sign would include the use of a honeycomb structure reduction perhaps being a realistic target. By combin-
within the freezing chamber to increase the copper ing xenon recovery with closed circuit use, it is
surface area (for which xenon has a high affinity) and conceivable that a 2-h administration for anesthesia or
a further reduce operating temperature. neuroprotection could be achieved using less than 10
The gas processing rate could also be improved. L xenon overall.
Currently, every 5 L of gas processed takes 53 min, There are a number of issues that would have to be
mostly due to the pauses in processing during cooling addressed before a cryogenic machine of this type could
and thawing. The heat transfer (removal) rate from the be used in a clinical setting. Waste gas from a circle
1316 Selective Xenon Recovery System ANESTHESIA & ANALGESIA
Figure 5. Xenon content of efflux gas from
freezing chamber is not materially affected by
xenon content of the input gas being processed
(flow, 300 mL/min; temperature, 139.2C).
system might be very humid, containing carbon dioxide pathways could be sterilized between patients, the
and possibly volatile anesthetic vapors. These could be sterility of the gas itself would be less of an issue than
removed by removing gas for processing from a point if recovered xenon were intended for other patients.
distal to the circle soda lime canister to ensure carbon Although the removable chamber can be sterilized
dioxide removal, by using a dehumidification chamber more readily than, for example, membrane or zeolite-
containing silica gel beads for example or by using a based devices, sterility concerns over the gas itself
volatile vapor adsorbent, such as activated charcoal, as mean that, realistically, recycling would need to be
seen in the Physioflex closed circuit machine (37). Alter- restricted to the same patient episode only. It is more
natively, a precooling chamber at approximately likely that such a device might form part of a theater
50C could selectively freeze water vapor, volatiles, complex central xenon recovery facility, from which
and any residual carbon dioxide. the relatively pure xenon could be returned at inter-
From a legal perspective, the recovery process vals to a gas manufacturer for final purification,
could be considered the manufacture of a medical gas sterilization, and recertification as a medical gas.
and the operator to be a medical gas manufacturer, In conclusion, we have demonstrated that an auto-
responsible for its medical purity (typically 99.9% or mated xenon recovery machine running on 240 V AC,
above) and freedom from microbial contamination. If 13 A standard electrical current with autoclavable
recovered xenon were only to be reused during the components is technically feasible. It functioned, as
same patient anesthetic then, so long as the gas intended, without the use of refrigerants (e.g., liquid
Vol. 105, No. 5, November 2007 2007 International Anesthesia Research Society 1317
nitrogen) that require replenishment. Within-patient 13. Homi HM, Yokoo N, Ma D, Warner DS, Franks NP, Maze M,
Grocott HP. The neuroprotective effect of xenon administration
xenon recycling and reuse might prove feasible, but during transient middle cerebral artery occlusion in mice.
concerns over purity, gas sterility, and legal issues Anesthesiology 2003;99:876 81
preclude interpatient xenon reuse. A machine such as 14. Ma D, Yang H, Lynch J, Franks NP, Maze M, Grocott HP. Xenon
this might be best suited for use in a central xenon attenuates cardiopulmonary bypass-induced neurologic and
neurocognitive dysfunction in the rat. Anesthesiology 2003;
recovery facility. Even with the most efficient breath- 98:690 8
ing systems, this recovery method could still usefully 15. Dingley J, Tooley J, Porter H, Thoresen M. Xenon provides
reduce overall xenon consumption. Such unusual at- short-term neuroprotection in neonatal rats when administered
after hypoxia-ischemia. Stroke 2006;37:501 6
tention to detail may be necessary if xenon finds a 16. Arrowsmith JE, Grocott HP, Reves JG, Newman MF. Central
clinical role as a neuroprotective drug, since demand nervous system complications of cardiac surgery. Br J Anaesth
could easily outstrip a fixed supply. Regardless of the 2000;84:378 93
purchase price, medical xenon will always be a scarce 17. Grocott HP. Cerebral injury during cardiac surgery: under-
standing the need for a neuroprotective strategy. J Anasth
commodity. Ten liters of xenon per 2 h delivery period Intensivbehandl 2002;2:45 6
may be a realistic overall target using closed circuit 18. Altschuler EL. Xenon as neuroprotectant in acute stroke? Med
breathing systems in combination with a xenon recov- Hypotheses 2001;56:227 8
19. Goto T, Suwa K, Uezono S, Ichinose F, Uchiyama M, Morita S.
ery device of this type. The blood-gas partition coefficient of xenon may be lower than
generally accepted. Br J Anaesth 1998;80:255 6
ACKNOWLEDGMENTS 20. Burov NE, Makeev GN, Potanov VN, Kornienko L. Alternative
The authors thank Dr. Alexei Moozyckine, Swansea, UK, means for reducing the cost of xenon anesthesia. Anesteziol
for technical assistance; Mr. John Minister, NBS Cryore- Reanimatol 1997:71 4
21. Lynch C III, Baum J, Tenbrinck R. Xenon anesthesia. Anesthe-
search, Tollesbury, UK, for advice and construction of part of siology 2000;92:865 8
the machine; Dr. Per Blom and Wolfgang Shmehl of Linde 22. Nalos M, Wachter U, Pittner A, Georgieff M, Radermacher P,
Gas therapeutics, Lidingo, Sweden for donation of xenon gas. Froeba G. Arterial and mixed venous xenon blood concentra-
tions in pigs during wash-in of inhalational anaesthesia. Br J
REFERENCES Anaesth 2001;87:497 8
23. Luttropp HH, Thomasson R, Dahm S, Persson J, Werner O.
1. Boomsma F, Rupreht J, Man in t Veld AJ, de Jong FH, Dzoljic
M, Lachmann B. Haemodynamic and neurohumoral effects of Clinical experience with minimal flow xenon anesthesia. Acta
xenon anaesthesia. A comparison with nitrous oxide. Anaesthe- Anaesthesiol Scand 1994;38:1215
sia 1990;45:273 8 24. Ferrari A, Erdmann W, Del Tacca M, Formichi B, Volta CA,
2. Dingley J, King R, Hughes L, Terblanche C, Mahon S, Hepp M, Ferrari E, Bissolotti G, Giunta F. Xenon anesthesia: clinical
Youhana A, Watkins A. Exploration of xenon as a potential results and recycling of gas. Appl Cardiopulm Pathophysiol
cardiostable sedative: a comparison with propofol after cardiac 1998;7:1535
surgery. Anaesthesia 2001;56:829 35 25. Luttropp HH, Rydgren G, Thomasson R, Werner O. A
3. Goto T, Hanne P, Ishiguro Y, Ichinose F, Niimi Y, Morita S. minimal-flow system for xenon anesthesia. Anesthesiology
Cardiovascular effects of xenon and nitrous oxide in patients 1991;75:896 902
during fentanyl-midazolam anaesthesia. Anaesthesia 2004; 26. Lachmann B, Armbruster S, Schairer W, Landstra M, Trouwborst
59:1178 83 A, Van Daal GJ, Kusuma A, Erdmann W. Safety and efficacy of
4. Luttropp HH, Romner B, Perhag L, Eskilsson J, Fredriksen S, xenon in routine use as an inhalational anaesthetic. Lancet
Werner O. Left ventricular performance and cerebral haemo- 1990;335:14135
dynamics during xenon anaesthesia. A transoesophageal 27. Tenbrinck R, Hahn MR, Gultuna I, Hofland J, van Dijk G,
echocardiography and transcranial Doppler sonography Baumert J, Hecker K, Erdmann W. The first clinical experiences
study. Anaesthesia 1993;48:10459 with xenon. Int Anesthesiol Clin 2001;39:29 42
5. Marx T, Froeba G, Wagner D, Baeder S, Goertz A, Georgieff M. 28. Bedi A, Murray JM, Dingley J, Stevenson MA, Fee JP. Use of
Effects on haemodynamics and catecholamine release of xenon xenon as a sedative for patients receiving critical care. Crit Care
anaesthesia compared with total i.v. anaesthesia in the pig. Br J Med 2003;31:2470 7
Anaesth 1997;78:326 7 29. Dingley J, Findlay GP, Foex BA, Mecklenburgh J, Esmail M,
6. Preckel B, Mullenheim J, Moloschavij A, Thamer V, Schlack W. Little RA. A closed xenon anesthesia delivery system. Anesthe-
Xenon administration during early reperfusion reduces infarct siology 2001;94:173 6
size after regional ischemia in the rabbit heart in vivo. Anesth 30. Bader S, Brand T. Reclaiming volatile and gaseous anesthetics.
Analg 2000;91:132732 Anasthesiol Intensivmed Notfallmed Schmerzther 1997;32:46 8
7. Goto T, Saito H, Nakata Y, Uezono S, Ichinose F, Morita S. 31. Greenspan M. Propagation of sound in five monoatomic gases.
Emergence times from xenon anaesthesia are independent of J Acoust Soc Am 1956;28:644 8
the duration of anaesthesia. Br J Anaesth 1997;79:5959
32. Lide DR. Handbook of chemistry and physics. 73rd ed. Boca
8. Rossaint R, Reyle-Hahn M, Schulte Am Esch J, Scholz J, Scherpereel P,
Raton, FL: CRC Press, 1992
Vallet B, Giunta F, Del Turco M, Erdmann W, Tenbrinck R,
33. King R, Bretland M, Wilkes A, Dingley J. Xenon measurement in
Hammerle AF, Nagele P. Multicenter randomized comparison of
the efficacy and safety of xenon and isoflurane in patients under- breathing systems: a comparison of ultrasonic and thermal
going elective surgery. Anesthesiology 2003;98:6 13 conductivity methods. Anaesthesia 2005;60:1226 30
9. Petzelt C, Blom P, Schmehl W, Muller J, Kox WJ. Prevention of 34. Da Silva JL, Stampfl C, Scheffler M. Adsorption of Xe atoms on
neurotransmitter over-release by xenon. J Anasth Intensivbe- metal surfaces: new insights from first principles calculations.
handl 2002;2:S117 Phys Rev Lett 2003;90:066104
10. Ma D, Wilhelm S, Maze M, Franks NP. Neuroprotective and 35. Burov NE, Kornienko L, Makeev GN, Potapov VN. Clinical and
neurotoxic properties of the inert gas, xenon. Br J Anaesth experimental study of xenon anesthesia. Anesteziol Reanimatol
2002;89:739 46 1999:56 60
11. Petzelt C, Blom P, Schmehl W, Muller J, Kox WJ. Prevention of 36. Baum JA. New and alternative delivery concepts and tech-
neurotoxicity in hypoxic cortical neurons by the noble gas niques. Best Pract Res Clin Anaesthesiol 2005;19:41528
xenon. Life Sci 2003;72:1909 18 37. Nathan N, Sperandio M, Erdmann W, Westerkamp B, Van Dijk
12. Petzelt C, Blom P, Schmehl W, Mueller J, Kox W. Xenon G, Scherpereel P, Feiss P. PhysioFlex: a target-controlled self-
prevents cellular damage in differentiated PC-12 cells exposed regulating closed-circuit inhalation anesthesia regulator. Ann Fr
to hypoxia. BMC Neurosci 2004;5:55 Anesth Reanim 1997;16:534 40