Pulse Sequences
Application Brochure
MAGNETOM Amira & Sempra
This brochure informs the user about
the pulse sequences of syngo MR for the
MR tomographs listed on the front page.
It addresses medical personnel working
in the area of MR tomography.
Contents
Basics Overview 2
Sequence variants 4
Gradient echo,
sequence description (I) 38
Gradient echo,
sequence description (II) 40
Gradient echo,
sequence description (III) 42
Gradient echo,
sequence description (IV) 44
Gradient echo,
sequence description (V) 46
Gradient echo,
parameter ranges (I) 48
Gradient echo,
parameter ranges (II) 50
Contents
Gradient echo,
parameter ranges (III) 52
Gradient echo,
parameter ranges (IV) 54
TrueFISP,
sequence description (I) 58
TrueFISP,
sequence description (II) 60
TrueFISP,
sequence description (III) 62
TrueFISP,
sequence description (IV) 64
TrueFISP,
sequence description (V) 66
TrueFISP,
sequence description (VI) 68
TrueFISP,
sequence description (VII) 70
EPI sequences,
sequence description 72
EPI sequences,
parameter ranges 74
Parameter ranges 80
Contrast-enhanced angiography,
parameter ranges (I) 110
Contrast-enhanced angiography,
parameter ranges (II) 112
Rephased-dephased angiography,
sequence description 114
Phase-contrast angiography,
sequence description 120
2
3
Sequence variants
qSWctDCFsR
(2) (4) (6) (8)
4
On-screen display
The sequence variants are shown in
different ways on-screen.
In the image text: The complete
sequence variant is displayed
(e.g., fl3d1r_tf)
In the parameter card: Only the
abbreviation of the sequence variant
is shown (sequence code)
Abbreviation Definition
ci CISS
(Constructive Interference in Steady State)
csi Chemical shift imaging
de DESS (Dual Echo Steady State)
epfid EPI-FID
epir EPI spin echo with inversion recovery
epse EPI spin echo
fi FISP (Fast Imaging with Steady State),
gradient echo sequence without
RF spoiling
fid Free induction decay
fl FLASH (Fast Low Angle Shot), gradient
echo sequence with active RF spoiling
h HASTE (Half Fourier Single Shot TurboSE)
hir HASTE with inversion recovery
ir Spin echo with inversion recovery
me MEDIC (Multi Echo Data Image
Combination)
ps PSIF (time-reversed FISP)
resolve RESOLVE (REadout Segmentation Of Long
Variable Echo trains),
multi-shot, diffusion-weighted EPI
Abbreviation Definition
se Spin echo
spc SPACE (Sampling Perfection with
Application-optimized Contrast using
different flip angle Evolutions)
svs Single voxel spectroscopy
swi Susceptibility-weighted imaging
tfl TurboFLASH
tgir Turbo gradient spin echo with
inversion recovery
tgse Turbo gradient spin echo
tir Turbo spin echo with inversion recovery
tirB Turbo spin echo with inversion recovery
and BLADE trajectory
tirBR Turbo spin echo with inversion recovery,
BLADE trajectory and “restore pulse”
tirR Turbo spin echo with inversion recovery
and “restore pulse”
tse Turbo Spin Echo
tseB Turbo spin echo with BLADE trajectory
tseBR Turbo spin echo with BLADE trajectory
and “restore pulse”
tseR Turbo spin echo with “restore pulse”
Abbreviation Definition
W Basic syngo WARP mode:
For high bandwidth optimization
V VAT (View Angle Tilting):
To reduce in-plane susceptibility distortions
S SEMAC (Slice Encoding for Metal Artifact
Correction): To reduce through-plane
susceptibility distortions
a
For a detailed description, please refer to “Operator
Manual – Ortho”.
Dimension (D)
Abbreviation Definition
1d 1-dimensional data acquisition
2d 2-dimensional data acquisition
3d 3-dimensional data acquisition
Contrast (C)
Abbreviation Definition
1, 2, ..., 16, ... Number of reconstructed images with
different contrasts
8
Flow compensation (F)
Abbreviation Definition
r Flow compensation in the readout and
slice selection direction
rr Flow compensation in the readout
direction only
rs Flow compensation in the
slice direction only
rd Interleaved acquisition with alternating
rephased and dephased measurements.
No flow compensation; is displayed only
if another character follows
pc Flow-sensitive gradients for
phase-contrast angiography
9
Sequence variant components (III)
Abbreviation Definition
2, 3, ..., 15, ... Number of segments in the k-space
for segmented sequences:
EPI factorTurbo factorsegments
2, 3, 4, 5, ... Medic sequences:
Number of combined echoes
If the number of segments 1
and the number of combined echoes 1
(is shown only, if another character
follows)
10
Excitation pulse (R)
Abbreviation Definition
ns Non-selective excitation
t10, t20, ..., t100 TONE pulse; number indicates the value of
the TONE ramp parameter (in percent)
11
Sequence nomenclature (I)
Abbreviation Definition
ciss Constructive Interference Steady State
dess Dual Echo Steady State
ep Echo Planar Imaging
fl FLASH (fast low angle shot)
gre Gradient echo sequence
haste Half Fourier Single Shot Turbo Spin Echo
medic Multi-Echo Data Image Combination
PETRA Pointwise Encoding Time reduction with
Radial Acquisition
psif Time inverted FISP sequence (FISP
fast imaging with steady state precession)
se Spin echo sequence
tfl TurboFLASH
tgse Turbo gradient spin echo
trufi TrueFISP (FISP fast imaging with steady
state precession)
tse Turbo spin echo
12
Abbreviation Definition
_bold Sequence for BOLD imaging
(BOLD Blood Oxygen Level Dependency)
_cb For determining the contrast bolus
_ce For contrast-enhanced angiography
_ct For angiography with continuous table
move
_diff Sequence for diffusion contrast
_fid Free induction decay, e.g., ep_fid:
Gradient echo variant of EPI sequence
_field_mapping For generating a field map for BOLD
post-processing
_fq Flow quantification
_histo Spectroscopy sequence for fat and iron
quantification in the liver
_IRT Sequence for interactive real-time imaging
_map Sequence for myocardial mapping
_mc Multi-contrast sequence
_nav Sequence with navigator
_pace Sequence with prospective motion
correction
_pasl Sequence for perfusion imaging
(PASL = Pulsed Arterial Spin Labeling)
_pc Phase-contrast angiography
_peri_tof Time-of-Flight sequence, optimized for
peripheral angiography
13
Sequence nomenclature (II)
Abbreviation Definition
_r With flow compensation
_rd Rephased-dephased angiography sequence
_se Spin echo sequence, e.g., ep_se:
Gradient echo variant of EPI sequence
_seg Segmented sequence, also refer to ..seg..
_sheco Shared echo sequence
(double contrast TSE)
_slaser Semi-LASER variant
_st Spectroscopy sequence for STEAM tech-
nique (STEAM Stimulated Echo Acquisi-
tion Method)
_therm Temperature-sensitive sequences
_tof Time-of-Flight (inflow angiography)
_trufi TrueFISP (FISP fast imaging with steady
state precession), refer to trufi as well
_vibe Volume interpolated breathhold
examination
_vfl Variable flip angle
14
Abbreviation Definition
..b.. Bandwidth per pixel; the value follows in
Hertz, e.g., se_15b130
..perf.. To measure First Pass perfusion
(cardiac imaging)
..r.. With flow compensation, also refer to _r
..seg.. Segmented sequence, also refer to _seg
..ti.. To determine TI (cardiac imaging)
..2d 2D imaging
..3d 3D imaging
.._15.. Echo time, e.g., 15 ms
15
Basic sequences: Spin echo with fixed
temporal sequence, sequence description
a
only available at systems with a field strength of 1.5 T
16
se_15b130 se_17rb130a
TE variable 3
Contrasts 1 1
Magn. Prep. IR IR
Recon. mode Magnitude/Realb Magnitude/Realb
Fat suppression Fat saturation/ Fat saturation
water excitation
Saturation Regular/Parallel Regular/Parallel
regions
MTC 3 3
Phase Partial 3 3
Fourier
Averaging Short term/ Short term/
mode long term long term
Multi-slice mode Interleaved/ Interleaved/
sequential sequential
Flow comp. No Yes
Dimension 2D 2D
RF pulse type Normal Normal
Gradient mode Normal Normal
b
may be selected only if Magn.Preparation IR has been selected
17
Basic sequences: Spin echo with optimized
temporal sequence, sequence description
18
se_mc Multi-contrast sequence with up to
32 contrasts. The sequence design is
derived from the TurboSE technique.
Since stimulated echoes contribute
to the signal with TurboSE, the second
echo provides a higher signal contribution
than the first echo.
Use
For acquiring T2 relaxation curves.
19
Basic sequences: Spin echo with optimized
temporal sequence, parameter ranges
se se_mc
TE variable 3 3
Contrasts 12 132
Bandwidth 3 3
variable
Magn. Prep. IR IR
a
Recon. mode Magnitude/Real Magnitude/Reala
Fat suppression Fat saturation/ Fat saturation/SPAIR
water excitation/
SPAIR
Saturation Regular/Parallel Regular/Parallel
regions
MTC 3 3
Phase Partial 3 3
Fourier
Asymm. Echo Off/allowed
Averaging mode Short term/ Short term/
long term long term
Multi-slice mode Interleaved/ Interleaved/
sequential sequential
Flow No No
compensation
Dimension 2D 2D
20
se se_mc
RF pulse type Fast/Normal/ Normal/Low SAR
Low SAR
a
may be selected only if Magn.Preparation IR has been selected
21
Basic sequences: Turbo spin echo,
sequence description (I)
a
only available at systems with a field strength of 3 T
23
Basic sequences: Turbo spin echo,
sequence description (II)
a
Please adhere to all safety instructions regarding
implants (-> Operator Manual - MR System)
24
If VATb is selected in addition, image distor-
tions can be further reduced. However, VAT
may introduce some blurring in the image.
To reduce VAT-related blurring, increase the
readout bandwidth and reduce the slice
thickness. The VAT effect can be scaled
from 0% (no VAT) to 100% (full VAT).
SEMAC is a technique to correct for
through-plane distortions, i.e., distortions
of the excited slice plane. The number of
SEMAC encoding steps depends on the
actual amount of distortions, for most
applications a range between 9–15 is
appropriate. Since the scan time is signifi-
cantly increased, techniques like parallel
imaging or partial Fourier should be
applied.
Do not use conventional spectral fat sup-
pression (including SPAIR) in the presence
of strong off-resonance effects, use inver-
sion recovery based fat suppression (STIR)
instead.
b
VAT: View Angle Tilting technique, eliminates the sig-
nal shift in the frequency-encoding direction caused
by off-resonance effects
25
Basic sequences: Turbo spin echo,
sequence description (III)
27
Basic sequences: Turbo spin echo,
sequence description (IV)
a
only available at systems with a field strength of 1.5 T
28
spc/tse_vfl 3D TurboSE sequence optimized for
3D data acquisition, T1/T2/PD weighting,
and dark fluid contrast. Significant SAR
reduction through the variable flip angle
technique.
Use
SPACE applications.
Display of the entire head with high spatial
resolution within an acceptable measure-
ment time.
PD, PD FatSat and T1 weighting with high
spatial resolution for orthopedic applica-
tions.
Basis for non-CE MR angiography protocols
with NATIVE SPACE.a
a
with NATIVE licencse
30
Recommendations: If SAR is exceeded,
either the flip angle of the refocusing pulse
can be reduced and/or RF pulse type “Low
SAR” can be selected. A smaller flip angle
(e.g., approx. 150) does not have a notice-
able effect on image contrast. As an alter-
native, the hyperecho techniquea may be
used for SAR reduction.
a
only available at systems with a field strength of 3 T
31
Basic sequences: Turbo spin echo,
parameter ranges (I)
tse tse_dixon
Turbo factor 1129 1129
TE variable 3 3
Contrasts 13
Bandwidth 3 3
variable
Magn. Prep. IR IR
Recon. mode Magnitude/Realb Magnitude
Fat suppression Fat saturation/
SPAIR
Saturation Regular/Parallel Regular/Parallel
regions
MTC 3 3
c
Phase Partial 3
Fourier
Averaging mode Short term/ Short term/
long term long term
Multi-slice mode Interleaved/ Interleaved/
sequential sequential
Flow No/Readout/Slice No/Readout/Slice
compensation
Dimension 2D/3D 2D
Slice resolutiona 3
Slice Partial 3
Fouriera
32
tse tse_dixon
RF pulse type Fast/Normal/ Fast/Normal/
Low SAR Low SAR
Gradient mode Fast/Normal/ Fast/Normal/
Whisper Whisper
Dark blood 3 3
iPAT 3 3
Respiratory Breathholdd/Trigger Trigger
control
Hyperecho 3 3
BLADE Trajectory 3
Restore 3 3
Acoustic noise 3
reduction
WARP 3
e
VAT 0100%
SEMACe 632
a
only if Dimension 3D is selected
b
may be selected only if Mag. Preparation IR has been selected
c
parameter cannot be freely set, depends on selected TE
d
Multi-breathhold possible, depending on protocol setup
e
only if WARP is selected
33
Basic sequences: Turbo spin echo,
parameter ranges (II)
tse_sheco spc/tse_vfl
Turbo factor 57 10599
TE variable 3 3
Contrasts 2 1
Bandwidth 3 3
variable
Magn. Prep. IR IR, DIR
Recon. mode Magnitude/Realb Magnitude/Realb
Fat suppression Fat saturation Fat saturation/
SPAIR/
water excitation
Saturation Regular/Parallel Regular/Parallel
regions
MTC 3 3
Phase Partial 3 3c
Fourier
Averaging mode Short term/
long term
Multi-slice mode Interleaved/
sequential
Flow No/Readout/Slice No/Readout
compensation
Dimension 2D 3D
Elliptical 3
scanninga
Slice resolutiona 3
Slice Partial 3
Fouriera
34
tse_sheco spc/tse_vfl
Excitation Slab-selective/
pulsesa non-selective
RF pulse type Fast/Normal/ Fast/Normal/
Low SAR Low SAR
Gradient mode Fast/Normal/ Fast/Normal/
Whisper Whisper
Dark blood 3 3
iPAT 3
2
iPAT 3
Respiratory Breathhold/Trigger
control
Restore 3
d
NATIVE Off/TD scout/
3D mode
Flow sensitivityd Default/Weak/
Medium/Strong
Blood Body region/Free
suppression
a
only if Dimension 3D is selected
b
may be selected only if Mag. Preparation IR has been selected
c
parameter cannot be freely set, depends on selected TE
d
with NATIVE license; “NATIVE” available only in case of ECG or pulse trigger,
“Flow sensitivity” available only if NATIVE is not off
35
Basic sequences: Turbo spin echo,
parameter ranges (III)
haste
Turbo factor 1512d
TE variable 3
Contrasts 14
Bandwidth 3
variable
Magn. Prep. IR
Recon. mode Magnitude/Realb
Fat suppression Fat saturation/
SPAIR
Saturation Regular/Parallel
regions
MTC 3
Phase Partial 3c
Fourier
Averaging mode Long term
Multi-slice mode Single-shot
Flow No/Readout/Slice
compensation
Dimension 2D/3D
Slice resolution 3a
Slice Partial 3a
Fourier
36
haste
RF pulse type Fast/Normal/
Optimized
Gradient mode Fast/Normal/
Whisper
Dark blood 3
iPAT 3
Respiratory Breathholde/Trigger
control
Restore 3
a
only if Dimension 3D is selected
b
may be selected only if Mag. Preparation IR has been selected
c
parameter cannot be freely set, depends on selected TE
d
parameter cannot be freely set, depends on base and phase resolution
e
Multi-breathhold possible, depending on protocol setup
37
Basic sequences: Gradient echo,
sequence description (I)
38
Use
Abdominal imaging In-phase
Opposed-phase in one measurement.
3D FLASH sequence for T1-weighted
ortho imaging and neuro imaging.
Fast 3D sequence for contrast agent
enhancement studies (e.g., dynamic
breast diagnostics).
Segmented CINE FLASH sequence, can
be combined with line or grid tagging.
Triggered segmented TurboFLASH
sequence (activated IR pulse) for “late
enhancement” imaging of the myocar-
dium.
Susceptibility-weighted imaging (SWI).
Evaluation of metastases and lymph nodes
in the thorax, abdomen, and pelvis.
40
medic MEDIC sequence (Multi Echo Data Image
Combination), gradient echo sequence
with flow compensation. When combined
echoes 1, the sequence behaves like a
FLASH sequence with flow compensation.
Use
T2*-weighted imaging with a high signal-
to-noise ratio for the spine (especially
transverse cervical and thoracic spine) and
ortho (knee and shoulder) imaging.
T2*-weighted 3D imaging of the cervical
spine with a small flip angle and adjusted
TR.
41
Basic sequences: Gradient echo,
sequence description (III)
Use
3D tfl (= MPRAGE) for fast T1-weighted
volume measurements of the head.
3D tfl with double inversion contrast
(= MP2RAGE) for homogenous
T1-weighted head imaging with high
CNR and optionally T1 mapping. Acti-
vated by setting two IR contrasts.
2D tfl with dark blood preparation for
low resolution imaging of the cardiac
anatomy.
2D tfl combined with saturation recovery
preparation for contrast agent enhance-
ment studies.
2D tfl to determine the transit time of a
test bolus as preparation for
CE angiography.
42
Please note: The 3D tfl MP2RAGE protocols
are optimized for consistent T1 mapping in
the range of biological T1 values in the
human tissue. Mapping of T1 values
exceeding the range between 400 ms and
1800 ms becomes increasingly imprecise.
T1 maps are presented along with a color
bar that displays the color mapping
between 0 and 2000 ms according to the
current window levels. Computed T1 maps,
however, may contain values between
0 and 4095 ms, thus exceeding the range
of the color bar.
43
Basic sequences: Gradient echo,
sequence description (IV)
44
Parameter ranges
Slice thickness 3 mm
Row and 2 mm
column spacing
iPAT acceleration 3
factor
In-plane 96 96
matrix size
Number of 64
slices
Orientations Sagittal: A>>P or P>>A PE-direction
Coronal: R>>L or L>>R PE-direction
Outputs
The overall quality rating is displayed in the
image text of the original series.
Image quality
High in the case of a high-quality scan
Not assessed in the case of non-performed quality
assessment (i.e., inconsistent protocol
parameters, error detected during the
processing)
To be confirmed if quality criteria are not fulfilled (e.g.,
motion, aliasing artifacts and other image
degradations)
45
Basic sequences: Gradient echo,
sequence description (V)
46
Parameters under syngo
Parameter Parameter card Comment
Segments Sequence/Part 2 If a fat saturation pulse is
selected, the number of
repetitions acquired with
each fat saturation can
be set with the parame-
ter Segments.
TI Contrast/Common If an inversion pulse is
selected, TI determines
the duration between
the inversion pulse and
the beginning of the
acquired repetitions.
Turbo Sequence/Part 2 If an inversion pulse is
Factor selected, Turbo Factor
determines the number
of repetitions between
two inversion pulses.
TR Contrast/Common If an inversion pulse is
selected, the repetition
time of the single repeti-
tions is set to its mini-
mum and the repetition
time corresponds to the
time between two inver-
sion pulses.
Radial Resolution/Common The number of radial
Views views can be changed on
the resolution card.
47
Basic sequences: Gradient echo,
parameter ranges (I)
gre
Segments 1127
TE variable 3
Contrasts 112
Bandwidth variable 3
Magn. Prep. IR, SRa
Reconstruction mode Magnitude/Phase
Fat suppression Fat saturation/
water excitation
Saturation regions Regular/Parallel/Tracking
MTC 3
Phase Partial Fourier 3
Asymm. Echo 3
Averaging mode Short term/long term
Multi-slice mode Interleaved/sequential
Flow compensation Yes/No/Readout/Slice/
Slice&Readout
Dimension 2D/3D
Elliptical scanningb 3
b
Slice resolution 3
48
gre
b
Slice Partial Fourier 3
Excitation pulsesb Slab-selective/non-selective
RF pulse type Fast/Normal/Low SAR
RF spoiling On/off
Gradient mode Fast/Normal/Whisper
Dark blood 3a
Grid/line tagging 3
iPAT 3
iPAT2 3
Respiratory control Breathholdc
Readout mode Monopolar/Bipolar
SWI 3
MapIt 3
Acoustic noise reduction 3
a
if segments 1 (segmented TurboFLASH)
b
only if Dimension 3D is selected
c
Multi-breathhold possible, depending on protocol setup
d
when switched on, DIXON reconstruction can be selected
49
Basic sequences: Gradient echo,
parameter ranges (II)
fl3d_vibe
Segments
TE variable 3a
Contrasts 112
Bandwidth variable 3
Magn. Prep.
Reconstruction mode Magnitude
Fat suppression Fat saturation/
water excitation/
SPAIR
Dixon 3
d
LiverLab 3
Saturation regions Regular/Parallel
MTC
Phase Partial Fourier 3
Asymm. Echo Off/Weak/Strong
Averaging mode Short term
Multi-slice mode Interleaved/sequential
Flow compensation No
Dimension 3D
Elliptical scanningb 3
b
Slice resolution 3
50
fl3d_vibe
b
Slice Partial Fourier 3
Excitation pulses Slab-selective
RF pulse type Fast/Normal
RF spoiling On
Gradient mode Fast/Normal/Whisper
Dark blood
3D centric reordering 3
iPAT 3
iPAT2 3
c
CAIPIRINHA 3
Respiratory control Breathhold
Readout mode Monopolar/Bipolar
Liver registration 3
MapIt 3
a
sequence allows a freely-selectable TE, an automatic minimal TE,
or an automatic TE set to the “in-phase” or “opposed-phase” time
b
only if Dimension 3D is selected
c
CAIPIRINHA is a generalization of iPAT2 allowing undersampling on a sheared
grid. (For a detailed description, please refer to “Supplement - Parameters
and Image text”.)
d
depends on the DIXON and DIXON evaluation options
51
Basic sequences: Gradient echo,
parameter ranges (III)
medic tfl
Segments
Combined 112
echoes
TE variable 3 Min.
Contrasts 1 1
IR contrasts 1 2d
Bandwidth 3 3
variable
Magn. Prep. IR, SR
Recon. mode Magnitude Magnitude/Realb
Fat suppression Fat saturation/ Water excitation
water excitation
Saturation Regular/Parallel
regions
MTC 3
Phase Partial 3 3
Fourier
Asymm. Echo 3
Averaging mode Short term/ Long term
long term
Multi-slice mode Interleaved/ Single-shot
sequential
Flow Yes No
compensation
Dimension 2D/3D 2D/3D
52
medic tfl
Elliptical 3 3c
scanninga
Slice resolutiona 3 3
Slice Partial 3 3
Fouriera
Excitation Slab-selective Slab-selective/
pulsesa non-selective
RF pulse type Fast/Normal Fast/Normal
RF spoiling On/off On/off
Gradient mode Fast/Normal/ Fast/Normal/
Whisper Whisper
Dark blood 3
3D centric 3
reordering
Grid/line tagging
iPAT 3 3
Respiratory Breathholde/
control Trigger
MapIt 3
a
only if Dimension 3D is selected
b
may be selected only if Mag. Preparation IR has been selected
c
freely selectable if 3D centric reordering is activated
d
may be selected only if Mag. Preparation IR or SR has been selected
e
Multi-breathhold possible, depending on protocol setup
53
Basic sequences: Gradient echo,
parameter ranges (IV)
PETRA
Segments 2100
TE variable 3
Bandwidth variable 3
Magn. Prep. IR
Fat suppression Fat saturation
Phase Partial Fourier 3
Dimension 3D
Acoustic noise reduction 3
54
55
Basic sequences:
Steady-state gradient echo,
sequence description
56
psif
TE variable Min.
Contrasts 1
Bandwidth variable 3
Recon. mode Magnitude
Fat suppression Water excitation
Phase Partial Fourier 3
Averaging mode Short term/long term
Multi-slice mode Sequential
Flow compensation Readout/slice
Dimension 2D/3D
Elliptical scanninga 3
a
Slice resolution 3
Slice Partial Fouriera 3
a
Excitation pulses Slab-selective
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
iPAT 3
Diffusion weighting Yes
a
only if Dimension 3D is selected
57
Basic sequences: TrueFISP,
sequence description (I)
Recommendations: Minimize TR by
selecting a high readout bandwidth to
avoid interference streaks in the image.
Use
T2 weighting for neuro applications, used,
for example, with uncooperative patients.
T2 weighting for breathhold studies in the
abdomen without motion artifacts.
58
TE variable Min.
Contrasts 1
Bandwidth variable 3
Magn. Prep. IR, SR
Reconstruction mode Magnitude
Fat suppression Fat saturation/
water excitation
Phase Partial Fourier 3
Asymm. Echo 3
Averaging mode Short term/long term
Multi-slice mode Sequential
Flow compensation Noa
Dimension 2D/3D
b
Elliptical scanning 3
Slice resolutionb 3
b
Slice Partial Fourier 3
Excitation pulsesb Slab-selective
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
iPAT 3
2b
iPAT 3
Respiratory control Breathholdc
a
TrueFISP is essentially flow-compensated from one RF pulse to the next
b
only if Dimension 3D is selected
c
Multi-breathhold possible, depending on protocol setup
59
Basic sequences: TrueFISP,
sequence description (II)
CV (without ECG)
Use
For displaying dynamic processes.
To display cardiac motion without a breath-
hold phase.
To examine cardiac volume in a breathhold
phase; multi-phase measurement of multi-
ple sequential slices is possible.
a
TE is automatically set to the minimum and cannot be changed
b
TrueFISP is essentially flow-compensated from one RF pulse to the next
c
only if Dimension 3D is selected
61
Basic sequences: TrueFISP,
sequence description (III)
CV (without ECG)
Use
To measure First Pass perfusion.
For sequential multi-slice applications.
62
Segments Variableb
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation IR, SR, selective, non-selective
Reconstruction mode Magnitude
Fat suppression Possible
Saturation regions
MTC
Phase Partial Fourier 3
Asymm. Echo 3
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation Noc
Dimension 2D/3D
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging
iPAT 3
a
TE is automatically set to the minimum and cannot be changed
b
max. number of segments depends on the number of lines
c
TrueFISP is essentially flow-compensated from one RF pulse to the next
63
Basic sequences: TrueFISP,
sequence description (IV)
CV
Use
To display cardiac motion. Provides good
contrast of blood and myocardium.
64
Segments 331
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression
Saturation regions
MTC
Phase Partial Fourier 3
Asymm. Echo
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation No
Dimension 2D (only untriggered)
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging
iPAT 3
Respiratory control Breathholdb
a
TE is automatically set to the minimum and cannot be changed
b
Multi-breathhold possible, depending on protocol setup
65
Basic sequences: TrueFISP,
sequence description (V)
CV
Use
For 2D or 3D applications.
To characterize tissue and angiography.
66
Segments Variableb
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation IR, SR, selective, non-selective
Reconstruction mode Magnitude
Fat suppression 3
Saturation regions Regular/Parallel
MTC
Phase Partial Fourier 3
Asymm. Echo 3
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation Noc
Dimension 2D/3D
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
Dark blood 3
Tagging
iPAT 3
Respiratory control Breathholdd
a
TE is automatically set to the minimum and cannot be changed
b
max. number of segments depends on the number of lines
c
TrueFISP is essentially flow-compensated from one RF pulse to the next
d
Multi-breathhold possible, depending on protocol setup
67
Basic sequences: TrueFISP,
sequence description (VI)
CV
Use
Identifying a suitable TI to modify image
contrast.
68
Segments Variablea
TE variable
Contrasts Variable
Bandwidth variable 3
Magn. Preparation TI Scout
Reconstruction mode Magnitude
Fat suppression
Saturation regions
MTC
Phase Partial Fourier
Asymm. Echo 3
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation
Dimension 2D
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging
iPAT
a
max. number of segments depends on the number of lines
69
Basic sequences: TrueFISP,
sequence description (VII)
70
Segments Variable
TE variable 3
Contrasts 1
Bandwidth variable 3
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression
Saturation regions
MTC
Phase Partial Fourier 3
Asymm. Echo 3
Averaging mode Short term/long term
Multi-slice mode Sequential/
Single measurement
Flow compensation
Dimension 2D
RF pulse type Fast/Normal/Low SAR
RF spoiling
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging
iPAT 3
71
Basic sequences: EPI sequences,
sequence description
72
ep_seg_se Segmented spin echo EPI sequence. Faster
than spin echo sequences, less off reso-
nance effects than a single-shot spin echo
EPI sequence.
Use
Fast T2-weighted 2D/3D imaging in the
head.
73
Basic sequences: EPI sequences,
parameter ranges
ep2d_se
EPI Factor 1256a
TE variable 3
Contrasts 1
Bandwidth variable Min ... max
Magn. Preparation IR
Recon. mode Magnitude
Fat suppression Fat saturation/
water excitation/SPAIR
Saturation regions Regular/Parallel
MTC 3
Phase Partial Fourier 3
Averaging mode Long term
Multi-slice mode Interleaved
Flow compensation No
Dimension 2D
RF pulse type Normal/Low SAR
RF spoiling Off
Gradient mode Fast/Normal
iPAT 3
Respiratory control
74
ep_seg_fid ep_seg_se
1127 1127
3 3
1 1
Min ... max Min ... max
IR
Magnitude Magnitude
Fat saturation/ Fat saturation/
water excitation water excitation
Regular/Parallel Regular/Parallel
3 3
3 3
Long term Long term
Interleaved/ Interleaved/
sequential sequential
No No
2D/3D 2D/3D
Normal Normal
On/Off Off
Fast/Normal Fast/Normal
b
Breathhold Breathholdb
a
parameter cannot be freely set, depends on base and phase resolution
b
Multi-breathhold possible, depending on protocol setup
75
CISS & DESS: Sequence description
76
dess ciss
TE variable 3 Min.
Contrasts 1 1
Bandwidth variable 3 3
Recon. mode Magnitude Magnitude
Fat suppression Water excitation Water excitation
Phase Partial Fourier 3 3
Averaging mode Short term/ Short term
long term
Multi-slice mode Sequential Sequential
Flow compensation Readout No
Dimension 3D 2D/3D
a
Elliptical scanning 3 3
a
Slice resolution 3 3
Slice Partial Fouriera 3 3
a
Excitation pulses Slab-selective/ Slab-selective
non-selective
RF pulse type Fast/Normal Fast/Normal
RF spoiling Off Off
Gradient mode Fast/Normal/ Fast/Normal/
Whisper Whisper
iPAT 3
a
only if Dimension 3D is selected
77
TGSE: Sequence description
78
Please note: Increasing the EPI factor
increases the sequence’s susceptibility sen-
sitivity; distortion artifacts may increase.
79
TGSE: Parameter ranges
tgse
Turbo factor 165
EPI factor 121
TE variable 3b
Contrasts 1
Bandwidth variable 3
Magn. Preparation IR
Reconstruction mode Magnitude/Reala
Fat suppression Fat saturation
Saturation regions Regular/Parallel
MTC 3
Phase Partial Fourier
Averaging mode Short term/long term
Multi-slice mode Interleaved/sequential
Flow compensation No
Dimension 2D
RF pulse type Fast/Normal/Low SAR
Gradient mode Fast/Normal/Whisper
Dark blood 3
Respiratory control Breathholdc
a
may be selected only if Magn.Preparation IR has been selected
b
applies only, when turbo factor 1
c
Multi-breathhold possible, depending on protocol setup
80
81
AutoAlign: Sequence description
82
83
Cardiac imaging:
Sequence description (I)
CV
Use
To display cardiac motion of a complete
cardiac cycle. Provides good contrast of the
blood and myocardium.
84
Segments Variableb
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression
Saturation regions
MTC
Phase Partial Fourier
Asymm. Echo
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation No
Dimension 2D
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging
iPAT 3
Respiratory control Breathholdc
a
TE is automatically set to the minimum and cannot be changed
b
max. number of segments depends on the number of lines
c
Multi-breathhold possible, depending on protocol setup
85
Cardiac imaging:
Sequence description (II)
CV
Use
To display cardiac motion. Provides good
contrast of blood and myocardium.
For short breathhold studies or multiple
slice measurements within one breathhold
phase.
86
Segments Variableb
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression
Saturation regions
MTC
Phase Partial Fourier
Asymm. Echo
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation Noc
Dimension 2D
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging
iPAT 3
Respiratory control Breathholdd
a
TE is automatically set to the minimum and cannot be changed
b
max. number of segments depends on the number of lines
c
TrueFISP is essentially flow-compensated from one RF pulse to the next
d
Multi-breathhold possible, depending on protocol setup
87
Cardiac imaging:
Sequence description (III)
CV
Use
To display cardiac motion.
88
Segments Variableb
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression
Saturation regions
MTC
Phase Partial Fourier
Asymm. Echo
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation No
Dimension 2D
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging
iPAT
Respiratory control Breathholdc
a
TE is automatically set to the minimum and cannot be changed
b
max. number of segments depends on the number of lines
c
Multi-breathhold possible, depending on protocol setup
89
Cardiac imaging:
Sequence description (IV)
CV
Use
To display cardiac motion. Especially well
suited for the display of cardiac valve
diseases.
90
Segments
TE variable 3
Contrasts 1
Bandwidth variable 3
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression
Saturation regions Regular/Parallel
MTC
Phase Partial Fourier
Asymm. Echo 3
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation Yes/No/Slice&Readout
Dimension 2D
RF pulse type Fast/Normal
RF spoiling On
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging 3
iPAT 3
Respiratory control Breathholda
a
Multi-breathhold possible, depending on protocol setup
91
Cardiac imaging:
Sequence description (V)
CV
Use
To display cardiac motion.
92
Segments Variablea
TE variable 3
Contrasts 1
Bandwidth variable 3
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression Fat saturation
Saturation regions Regular/Parallel
MTC 3
Phase Partial Fourier 3
Asymm. Echo 3
Averaging mode Short term/long term
Multi-slice mode Sequential
Flow compensation Yes/No/Slice&Readout
Dimension 2D
RF pulse type Fast/Normal
RF spoiling On/off
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging 3
iPAT
Respiratory control Breathholdb
a
max. number of segments depends on the number of lines
b
Multi-breathhold possible, depending on protocol setup
93
Cardiac imaging:
Sequence description (VI)
CV
Use
To display cardiac motion without a breath-
hold phase. To examine cardiac volume in a
breathhold phase; multi-phase measure-
ment of multiple sequential slices is possi-
ble.
94
Segments
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression
Saturation regions
MTC
Phase Partial Fourier
Asymm. Echo
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation Nob
Dimension 2D
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging
iPAT 3
a
TE is automatically set to the minimum and cannot be changed
b
TrueFISP is essentially flow-compensated from one RF pulse to the next
95
Cardiac imaging:
Sequence description (VII)
CV
Use
To display cardiac motion in real time.
96
Segments
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression
Saturation regions
MTC
Phase Partial Fourier
Asymm. Echo
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation No
Dimension 2D
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging
iPAT
Respiratory control Breathholdb
a
TE is automatically set to the minimum and cannot be changed
b
Multi-breathhold possible, depending on protocol setup
97
Cardiac imaging:
Sequence description (VIII)
98
Segments Variableb
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation IR, SR, selective, non-selective
Recon. mode Magnitude
Fat suppression 3
Saturation regions Regular/Parallel
d
Inversion regions up to 4
(with different TI)
Phase Partial Fourier 3
Asymm. Echo 3
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation Noc
Dimension 2D/3D
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/
Whisper
Dark blood 3
iPAT 3
a
TE is automatically set to the minimum and cannot be changed
b
max. number of segments depends on the number of lines
c
TrueFISP is essentially flow-compensated from one RF pulse to the next
d
with NATIVE license
99
Cardiac imaging:
Sequence description (IX)
CV
Use
Display of individual slices within a breath-
hold phase.
100
Segments 123
TE variable 3
Contrasts 1
Bandwidth variable 3
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression Fat saturation
Saturation regions Regular/Parallel
MTC 3
Phase Partial Fourier 3
Asymm. Echo 3
Averaging mode Short term/long term
Multi-slice mode Sequential
Flow compensation Yes
Dimension 2D
RF pulse type Fast/Normal
RF spoiling On/off
Gradient mode Fast/Normal/Whisper
Dark blood
Tagging 3
iPAT
Respiratory control Breathholda
a
Multi-breathhold possible, depending on protocol setup
101
Cardiac imaging:
Sequence description (X)
CV, CV epi
Use
To measure First Pass perfusion.
For sequential multi-slice applications.
102
Segments Fixedb
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation IR, SR, selective, non-selective
Reconstruction mode Magnitude
Fat suppression Yes (Gre epi)
Saturation regions
MTC
Phase Partial Fourier
Asymm. Echo 3
Averaging mode Long term
Multi-slice mode Single-shot
Flow compensation Noc (Gre)
Dimension 2D/3D
RF pulse type Fast/Normal
RF spoiling On (Gre and Gre epi)
Gradient mode Fast/Normal/Whisper
PAT Integrated/tPAT
a
TE is automatically set to the minimum and cannot be changed
b
max. number of segments depends on the number of lines
c
TrueFISP is essentially flow-compensated from one RF pulse to the next
103
Cardiac imaging:
Sequence description (XI)
CV
Use
To display late enhancement.
104
Segments Variable and fixedb
TE variable a
Contrasts 1
Bandwidth variable 3
Magn. Preparation IR, selective, non-selective
Reconstruction mode Magnitude/Real
Fat suppression
Saturation regions
MTC
Phase Partial Fourier
Asymm. Echo 3
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation Noc (Gre)
Dimension 2D
RF pulse type Fast/Normal
RF spoiling On (Gre)
Gradient mode Fast/Normal/Whisper
PAT Integrated
a
TE is automatically set to the minimum and cannot be changed
b
max. number of segments depends on the number of lines
(variable = segmented, fixed = single-shot)
c
TrueFISP is essentially flow-compensated from one RF pulse to the next
105
Myocardial mapping with BEAT_map:
Sequence description
Use
Generating parametric maps of the heart.
For a detailed description, please refer to
“Operator Manual – Cardio”.
Parameter ranges
Segments Variablea
TE variable b
Contrasts 132
Bandwidth variable 3
Magn. Preparation Non-selective, IR,
T2 preparation
Reconstruction mode Magnitude/Phase
Fat suppression 3
MTC
106
Phase Partial Fourier 3
Sequence type GRE/TRUFI
Asymm. Echo 3
Averaging mode
Multi-slice mode Sequential
Flow compensation Yesc
Dimension 2D
RF pulse type Fast/Normal
RF spoiling Off
Gradient mode Fast/Normal/Whisper
Dark blood 3
Trajectory Cartesian
iPAT 3
Respiratory control Breathholdd
Optimization Min TE, Min TE/TR
a
max. number of segments depends on the number of lines.
T1 and T2 mapping are single-shot techniques.
b
TE is automatically set to the minimum and cannot be changed
c
TrueFISP is essentially flow-compensated from one RF pulse to the next.
GRE can be combined with flow compensation.
d
Multi-breathhold possible, depending on protocol setup
107
Angiography:
Contrast-enhanced angiography,
sequence description
Use
Optimized for CE angiography.
Angiography of the carotid, thoracic, and
pulmonary arteries
Basis for Care Bolus and Test Bolus proto-
cols for determining arrival of contrast
agent.
Table stepping angiography for display
of peripheral vessels.
Cardiac imaging to examine
“Myocardial Viability” with triggered
segmented 3D FLASH sequence with
activated IR pulse.
108
Recommendations: For Care examina-
tions, select “3D centr. reordering” on the
Angio parameter card.
109
Angiography:
Contrast-enhanced angiography,
parameter ranges (I)
tfl_cb
Segments
TE variable Min.
Contrasts 1
Bandwidth variable 3
Magn. Prep. IR/SR
Recon. mode Magnitude/Realc
Fat suppression Water excitation
Saturation regions
MTC
Phase Partial Fourier 3
Asymm. Echo Off/allowed
Averaging mode Long term
Multi-slice mode Single-shot
Flow compensation No
110
tfl_cb
Dimension 2D
Elliptical scanningd
d
Slice resolution
d
Slice Partial Fourier
Excitation pulse Slice-selective
RF pulse type Fast/Normal
RF spoiling On
Gradient mode Fast/Normal
Dark blood 3
CARE Bolus
iPAT
2
iPAT
a
if segments 1 (segmented TurboFLASH)
b
Partial Fourier is selected. However, the setting cannot be changed
c
may be selected only if Magn. Preparation IR or SR has been selected
d
only if Dimension 3D is selected
111
Angiography:
Contrast-enhanced angiography,
parameter ranges (II)
fl3d_ce
Segments 1127
TE variable 3a
Contrasts 1
Bandwidth variable 3
Magn. Prep. IRb
Recon. mode Magnitude
Fat suppression Fat saturation/water excitation
Saturation regions Regular/Parallel
MTC
Phase Partial Fourier 3
Asymm. Echo Allowed
Averaging mode Short term
Multi-slice mode Interleaved/
sequential
Flow compensation No
Dimension 2D/3D
Elliptical scanningc 3d
c
Slice resolution 3
c
Slice Partial Fourier 3
112
fl3d_ce
Excitation pulse Slab-selective/non-selective
RF pulse type Fast/Normal/Low SAR
RF spoiling On
Phase enc. rewinder 3
Gradient mode Fast/Normal/Whisper
Dark blood
3D centric reordering 3
e
Time to center 3
iPAT 3
2
iPAT 3
Respiratory control
Readout mode
Liver registration
MapIt
a
sequence allows a freely-selectable TE, an automatic minimal TE,
or an automatic minimal TE and TR
b
if segments 1 (segmented TurboFLASH)
c
only if Dimension 3D is selected
d
not available if iPAT is activated
e
freely selectable if 3D centric reordering is activated
113
Angiography:
Rephased-dephased angiography,
sequence description
114
TE variable 3
Contrasts 1
Bandwidth variable 3
Reconstruction mode Magnitude
Fat suppression Fat saturation
Saturation regions Regular/Parallel
MTC 3
Phase Partial Fourier 3
Averaging mode Short term/long term
Multi-slice mode Sequential
Flow compensation Yes
Dimension 3-D
Elliptical scanning 3
Slice resolution 3
Slice Partial Fourier 3
Excitation pulse Slab-selective
RF pulse type Normal
RF spoiling On
Gradient mode Fast/Normal/Whisper
115
Angiography: Inflow angiography (ToF),
sequence description
116
fl_peri_tof 2D FLASH sequence.
Use
Peripheral angiography.
ECG-triggered measurement of vessels in
the legs at a high heart rate or in the kid-
neys, arms, and carotid arteries.
117
Angiography: Inflow angiography (ToF),
parameter ranges
fl_tof
TE variable 3
Contrasts 1
Bandwidth variable 3
Recon. mode Magnitude
Saturation regions Regular/Parallel/Tracking
MTC 3
Phase Partial Fourier 3
Asymm. Echo Allowed
Averaging mode Short term/long term
Multi-slice mode Sequential
Flow compensation Yes
Dimension 2D/3D
Elliptical scanninga 3
a
Slice resolution 3
Slice Partial Fouriera 3
Excitation pulse Slab-selective
RF pulse type Normalb
RF spoiling On
Gradient mode Fast/Normal/Whisper
iPAT 3
a
only if Dimension 3D is selected
b
if Dimension 3D is selected TONE pulses are used, which can be optimized
for different flow velocities via the TONE ramp parameter
118
fl_peri_tof
Segments 123
TE variable 3
Contrasts 1
Bandwidth variable 3
Reconstruction mode Magnitude
Fat suppression
Saturation regions Regular/Parallel/Tracking
MTC 3
Phase Partial Fourier 3
Asymm. Echo Off/allowed
Averaging mode Short term/long term
Multi-slice mode Sequential
Flow compensation Yes
Dimension 2D
Excitation pulse Slab-selective
RF pulse type Fast/Normal
RF spoiling On
Gradient mode Fast/Normal/Whisper
119
Angiography: Phase-contrast angiography,
sequence description
120
Segments 125
TE variable 3
Contrasts 1
Bandwidth variable 3
Reconstruction mode Magnitude
Saturation regions Regular/Parallel/Tracking
MTC 3
Phase Partial Fourier
Asymm. Echo 3
Averaging mode Off/Weak/Strong
Multi-slice mode Sequential
Flow compensation Yes
Dimension 2D/3D
Elliptical scanninga 3
a
Slice resolution 3
Slice Partial Fouriera
Excitation pulse Slab-selective
RF pulse type Fast/Normal
RF spoiling On
Gradient mode Fast/Normal/Whisper
Venc cm/s 1999
a
only if Dimension 3D is selected
121
Angiography: Fast dynamic imaging,
sequence description
122
TE variable Min.
Contrasts 1
Bandwidth variable 3
Reconstruction mode Magnitude
MTC
Phase Partial Fourier 3
Asymm. Echo Off/Weak/Strong
Averaging mode Short term
Multi-slice mode Interleaved/sequential
Flow compensation No
Dimension 3D
Elliptical scanninga 3
Slice resolution 3
Slice Partial Fourier 3
Excitation pulses Slab-selective/
non-selective
RF pulse type Fast/Normal/Low SAR
RF spoiling On
Gradient mode Fast/Normal/Whisper
iPAT 3
iPAT2 3
Central region A 4–100%
Sampling density B 0–50%
Temporal interpolation 3
Time stamp in image 3
a
not available if iPAT is activated
123
Flow quantification and phase-contrast
angiography: Sequence description (I)
124
Use flow quantification
To display and quantify blood flow.
Through-plane measurements for flow
quantification and in-plane measure-
ments for flow display.
Triggered, retrospective gating possible,
to display and quantify blood flow during
complete heart cycle coverage. Particu-
larly suited for patients with arrhythmias.
Enables flow measurement in breathhold
technique using phase sharing.
125
Flow quantification and phase-contrast
angiography: Sequence description (II)
126
Segments Variable
TE variable 3
Contrasts 1
Bandwidth variable 3
Reconstruction mode Magnitude
Asymm. Echo 3
Averaging mode Short term/long term
Multi-slice mode Sequential
Flow compensation Yes
Dimension 2D/3D
a
Elliptical scanning 3
Slice resolutiona 3
Excitation pulse Slab-selective
RF pulse type Fast/Normal
RF spoiling On
Gradient mode Fast/Normal/Whisper
Venc cm/s 1999
a
only if Dimension 3D is selected
127
Diffusion imaging: Sequence description
128
resolve Multi-shot, diffusion-weighted, readout-
segmented EPI sequence.
During diffusion preparation, multi-shot
DWI is sensitive to brain motion caused by
CSF pulsation. This leads to non-linear
phase errors that vary from shot to shot.
The effect of these phase errors can be
minimized by using a readout-segmented
EPI readout in conjunction with
2D navigator phase correction and naviga-
tor-based reacquisition.
Use
For improved image quality in neuro imag-
ing compared to standard single-shot EPI
sequence.
129
Diffusion imaging: Parameter ranges (I)
ep2d_diff
a
EPI Factor 1256
TE variable 3
Contrasts 1
Bandwidth variable Min ... max
Magn. Preparation IR
Recon. mode Magnitude
Fat suppression Fat saturation/
water excitation/SPAIR
Saturation regions Regular/Parallel
MTC 3
Phase Partial Fourier 3
Averaging mode Long term
Multi-slice mode Interleaved
Flow compensation No
Dimension 2D
RF pulse type Normal/Low SAR
RF spoiling Off
130
ep2d_diff
Gradient mode Fast/Normal
b-value 010000
Online DW and ADC image, FA
reconstruction image, TRACE, colored FA
image
iPAT 3
Respiratory control Breathholdd/Trigger
Readout segments
Reacquisition modec
Echo spacing 3
variable
Optimization Min TE
a
parameter cannot be freely set, depends on base and phase resolution
b
parameter cannot be freely set, depends on other parameters
c
automatically set to “off” when physiological triggering is active
d
Multi-breathhold possible, depending on protocol setup
131
Diffusion imaging: Parameter ranges (II)
qDWI resolve
a
EPI Factor 1−256 1−256
TE variable -b -b
Contrasts 2 2
b
Bandwidth variable - -b
Magn. Preparation IR IR
Recon. mode Magnitude Magnitude
Fat suppression Fat saturation/ Fat saturation/
water excitation/ water excitation/
SPAIR SPAIR
Saturation regions Regular/Parallel Regular/Parallel
MTC 3 3
Phase partial Fourier3 -
Readout partial - 3
Fourier
Averaging mode Long term Long term
Multi-slice mode Interleaved Interleaved
Flow compensation No No
Dimension 2D 2D
RF pulse type Normal/Low SAR Normal/Low SAR
RF spoiling Off Off
Excitation pulses - -
132
qDWI resolve
Gradient mode Fast/Normal/ Fast/Normal/
Whisper Whisper
b-value 0- 10000 0- 10000
Online DW and ADC DW and ADC
reconstruction image, FA image, image, FA image,
TRACE, colored FA TRACE, colored FA
image image
iPAT 3 3
Respiratory control - -
Readout segments 3 3
Reacquisition modecOn/off On/off
Echo spacing 3 3
variable
Optimization Min TE/Min TR Min TE/Min TR
a
parameter cannot be freely set, depends on base and phase resolution
b
parameter cannot be freely set, depends on other parameters
c
automatically set to “off” when physiological triggering is active
133
Perfusion imaging: Sequence description
134
tgse_pasl Segmented turbo gradient spin echo
sequence for high resolution imaging or
extremely short measurement times.
At an EPI factor of 1 and a Turbo factor 1,
the sequence behaves like a tse sequence.
At a Turbo factor of 1 and an EPI factor 1,
the sequence behaves like a segmented
EPI sequence.
Use
For perfusion imaging in the head. As a
result, you obtain perfusion-weighted
images.
135
Perfusion imaging: Parameter ranges
ep2d_fid ep2d_pasl
a
EPI Factor 1256 1128
TE variable 3 3
Contrasts 1 1
Bandwidth variable Min ... max Min ... max
Magn. Preparation PASL
Recon. mode Magnitude Magnitude
Fat suppression Fat saturation Fat saturation
(strong/weak)
Saturation regions Regular/Parallel PICORE Q2Tc
MTC 3
Phase Partial Fourier 3 3
Averaging mode Long term Long term
Multi-slice mode Interleaved Interleaved
Flow compensation No No
Dimension 2D 2D
RF pulse type Normal Normal
RF spoiling Off Off
Gradient mode Fast/Normal Fast
Online TTP, GBP, PBP, Perfusion-weighted
reconstruction relCBV, relCBF, images, relCBF
relMTT
Flow limit 0100 cm/s
Bolus duration Variabled
iPAT 3 3
Quality check On/Off/
On - extended
136
tgse_pasl
Turbo factor 165
EPI factor 1127
TE variable b
Bandwidth variable 3
Magn. Preparation PASL
Reconstruction mode Magnitude
Fat suppression Fat saturation
Saturation regions FAIR QIIc
MTC
Phase Partial Fourier
Averaging mode Long term
Multi-slice mode
Flow compensation No
Dimension 3D
RF pulse type Normal
Gradient mode Fast/Normal
Online reconstruction Perfusion-weighted images
Bolus duratione Variable
f
Inversion array size Min. 1
a
parameter cannot be freely set, depends on base and phase resolution
b
TE is automatically set to the minimum and cannot be changed
c
parameter “Perfusion mode”
d
parameter depends on TR
e
parameter depends on TI and inversion array size
f
maximum depends on TR and TI
137
BOLD imaging: Sequence description
138
gre_field_ 2D gradient echo sequence that generates
mapping images at two different echo times. The
first TE is variable, the second TE is fixed
relative to the first TE. The different echo
times correspond to the in-phase condi-
tions of fat and water protons.
Use
When phase images are selected, you
obtain a single phase differential image
that can be used as a “Fieldmap” during
BOLD post-processing. The information
is required when overlaying anatomical
images generated with other techniques,
e.g., SE or MPRAGE.
Prerequisites for using the BOLD post-
processing card:
The phase range of PI to PI corre-
sponds to image pixel values of 0 to
4095
Image geometry and slice orientation
have to be identical to the EPI images
The sequence has to be applied prior to
the corresponding EPI measurement
139
BOLD imaging: Parameter ranges
ep2d_bold
a
EPI Factor 1128
TE variable 3
Contrasts 1
Bandwidth variable Min...max
Magn. Preparation
Reconstruction mode Magnitude
Fat suppression Fat saturation/
water excitation
Saturation regions Regular/Parallel
MTC 3
Phase Partial Fourier 3
Asymm. Echo
Averaging mode Long term
Multi-slice mode Interleaved
Flow compensation No
Dimension 2D
RF pulse type Normal
RF spoiling Off
Gradient mode Fast/Normal
iPAT 3
a
parameter cannot be freely set, depends on base and phase resolution
140
ep2d_pace gre_field_mapping
1128
3 3
1 2
Min...max 3
Magnitude Magnitude/Phase
Fat saturation/ Fat saturation
water excitation
Regular/Parallel Regular/Parallel/Tracking
3 3
3 3
3
Long term Short term/long term
Interleaved Interleaved/sequential
No Yes
2D 2D
Normal Fast/Normal
Off On/off
Fast/Normal Fast/Normal/Whisper
3
141
Interactive real-time imaging:
Sequence description
143
Interactive real-time imaging:
Parameter ranges
a
TE is automatically set to the minimum and cannot be changed
144
Sequence type Gre
TE variable 3
Contrasts 1
Bandwidth variable 3
Magn. Preparation 3
Recon. mode Magnitude
Fat suppression Fat saturation/
water excitation
Saturation regions Regular
MTC
Phase Partial Fourier 3
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation Yes/no
Dimension 2D
RF pulse type Fast/Normal/Low SAR
RF spoiling On/Off
Gradient mode Fast/Normal/Whisper
Dark blood 3
Trajectory Radial
Slices 1
145
Spectroscopy: Sequence description (I)
Available sequences
svs_se
svs_st
svs_st_histo (HISTO)
csi_se
csi_st
csi_slasera
fid
a
only available at systems with a field strength of 3 T
STEAM (_st)
Localization with three orthogonal slices
90–90–90
Measured signal: stimulated echo
Short TEs possible
Only half of the spin echo signal
Semi-LASER (_slaser)
Localization with three orthogonal slices
90–180–180–180–180
180 pulses are adiabatic full passage
pulses and used in pairs
Measured signal: full spin echo
147
Spectroscopy: Parameter ranges (I)
svs_se svs_st
TE variable 3 3
Mixing time (TM) 3
variable
Bandwidth variable 3 3
Spectral fat 3
suppression
Saturation regions Regular
Averaging
mode
Dimension
Phase cycling 3 3
Triggering ECG/respiratory
k-space
weighting
Excitation of VOI 3 3
Multiple 3 3
measurements
AutoAlign 3 3
Fully excited VOI
Water reference 3 3
scan
148
csi_se csi_st
TE variable 3 3
Mixing time (TM) 3
variable
Bandwidth variable 3 3
Spectral fat 3
suppression
Saturation regions Regular
Averaging mode Short term/ Short term/
long term long term
Dimension 2D/3D 2D/3D
Phase cycling
Triggering
k-space weighting Full/elliptical/ Full/elliptical/
weighted weighted
Excitation of VOI 3 3
Multiple 3 3
measurements
AutoAlign 3 3
Fully excited VOI 3
Water reference
scan
149
Spectroscopy: Parameter ranges (II)
csi_slasera fid
TE variable 3 b
Mixing time (TM)
variable
Bandwidth variable 3 3
Spectral fat
suppression
Saturation regions Regular
Averaging mode Short term/
long term
Dimension 2D/3D
Phase cycling 3
Triggering ECG/respiratory
k-space weighting Full/elliptical/
weighted
Excitation of VOI 3
Multiple 3 3
measurements
AutoAlign 3
Fully excited VOI
Water reference
scan
150
a
only available at systems with a field strength of 3 T
b
TE is automatically set to the minimum and cannot be changed
151
Spectroscopy: Sequence description (II)
Advantages
Complete data acquired in one breath-
hold
Accurate fat evaluation due to integrated
T2 correction
Estimation of water R2 relaxivity, which
correlates with the amount of iron in the
liver
152
TE variable 3
Mixing time (TM) variable 3
Bandwidth variable
Spectral fat suppression
Saturation regions
Averaging mode
Dimension
Phase cycling 3
Triggering
k-space weighting
Excitation of VOI 3
Multiple measurements 5 (fixed)
AutoAlign 3
Fully excited VOI
Water reference scan
153
Thermometry: Sequence description
154
ep_seg_therm
EPI Factor 1
TE variable
Contrasts 1
Bandwidth variable
Magn. Preparation
Reconstruction mode Magnitude/Phase
Fat suppression
Phase Partial Fourier
Asymm. Echo Off
Averaging mode Short term
Multi-slice mode Sequential
Flow compensation
Dimension 2D
Excitation pulse Slice-selective
RF pulse type Normal
RF spoiling On
Gradient mode Fast
iPAT
Trajectory Cartesian
Slices 1
155
Susceptibility-weighted imaging (SWI)
156
Method
For data acquisition, a high-resolution
3D gradient echo sequence with echo
times of 40 ms at a field strength of 1.5 T
or 20 ms at 3 T is used. Using the long echo
times, the MR signals in venous blood and
normal brain tissue are able to create a
phase difference. This phase difference is
used for contrasting where veins and hem-
orrhages are shown as dark.
Image reconstruction
Two image series are computed In the
Magnitude reconstruction mode:
MIP reconstructions of the SWI images
across eight slices each (Sliding Window
mode)
SWI images without MIP reconstruction
157
BLADE: Application/sequence description
158
Sequence When using the BLADE technique, each
description echo train of the sequence generates a
low-resolution image. The phase-encoding
direction of this image is turned from one
excitation to the next. Subsequently, the
individual, low-resolution images are com-
bined into a high-resolution image.
Sampling of
the k-space
with BLADE:
Each echo train
fills an
individual
stripe, the
so-called blade
159
BLADE: Sequence parameters
160
In the Sequence/Part 2 parameter card
Turbo factor: number of lines per blade
(if GRAPPA is not used)
Motion correction On:
Corrects patient motion within the image
plane (in-plane) which happens between
the acquisition of two blades
161
BLADE: Application notes
162
6: The true FoV of the sequence is not dis-
played. It corresponds to the blue circle in
the square FoV (visible with graphic slice
positioning). In the figure, the true
expanded FoV (with phase-oversampling)
is shown as a green circle. Due to the dif-
ferent phase-encoding directions of the
163
TSE DIXON: Application
a
Please adhere to all safety instructions regarding
implants (-> Operator Manual - MR System)
164
Image examples:
top left: reliable FatSat
top right: with MR Conditional implants
bottom left: fat image
bottom right: water image
165
SPACE: Application
166
Body imaging Pelvis: in place of several 2D acquisitions
PACE MRCP during free breathing
167
SPACE: Sequence description/protocols
168
Optimized protocols
The following applications in the
Siemens protocol tree include optimized
SPACE protocols:
Head: T1, T2, dark fluid, Double Inversion
Recovery, and TrueIR contrast
Spine: T2 contrast
Body: T2 contrast, MRCP
Ortho: PD and T1 contrast (with and
without fat suppression)
169