Professional Documents
Culture Documents
Version 2.0
4.30-0
Note
7
This operation manual is valid for the program version beginning in
,
1 October 1997 for mobile instruments with the MS DOS operation sys-
tern,
.,p'
.<:
,@
8
.ij
.#
x.&'
0 .,.....,.,.,.
.$
.. <p
..-..- :.A$<
The text and illustrations have been ;ompiled with utmost care. How-
ever, it is impossible to absolutely exclude errors.
With the proper sample preparation, the analyses using spark excita-
tion can be favorably compared with the results from large laboratory
spectrometers.
Note:
All safety precautions must be followed when working with
the system. Please read the safety instructions in this
manual.
1.1 The SPECTROPORT
1.2 The SPECTROTEST
1.3 The SPECTROTEST JR
1.4 General
lntroduction
Profiling
Measuring Process
Recalibration
Grade Recognition
Storage of Measurement Data
Type Recalibration
The Password
Individual Settings
Support Functions for Approximate Analysis
4.1 Introduction
4.2 Sorting without ldentification of the Reference
Sample (without Approximate Analysis Program)
4.3 Sorting with ldentification of the Reference Sample
(with Approximate Analysis Program)
4.4 Creating a New Sorting Program
4.5 Support Functions for Sorting
gSPECTRO
Analytical ~rntrurnentll
After reaching the control site, operation can follow immediately. Sin-
gle sided heating, due to heaters, ovens, or the sun should be avoided.
When working with instruments with the UV 150 minioptic (option) for
the determination of phosphorus and sulfur in iron base, the mini should
alwavs be transported in the case in which it was delivered. When trans-
porting the spectrometer in a van, etc., the minioptic must be discon-
nected and placed in its case.
Please make sure that the argon inlet on the instrument is always closed
during transportation.
0 Remove the minioptic from its case and slide it onto the mounting
plate on the pistol until it hits the stop. Connect the supply cable
securely with the connection on the front of the instrument.
8 Fix the optic with the screw on the rear side of the mounting plate.
Turn the screw until it presses the metal sheet against the brass
block, then unscrew the screw by a half of a rotation.
Q Screw the special argon adapter for the analysis of phosphorus and
sulfur onto the pistol.
O Seal the spark opening on the adapter with tape, for example (the
spacer should not be used, because the diameter of the opening
can be enlarged).
The adapter must be closed in this way so that the main argon stream
flows through the optic, thus supplying it with argon.
The pistol of the mobile instruments can be flushed with argon. The
argon and argon container must be supplied by the customer.
The analytical programs were discussed when the instrument was pur-
chased. The instrument has been calibrated by SPECTRO according
to your needs as set by you and your sales representative.
Chassis Construction
SPECTROPORT Front Panel
SPECTROPORT Rear Side
Chassis Construction
SPECTROTEST Front Panel
SPECTROTEST Left Side
SPECTROTEST Right Side
Chassis Construction
SPECTROTEST JR Front Panel
SPECTROTEST JR Rear Side
Chassis Construction
The SPECTROPORT is a modularly constructed analytical system. All
components are easily and quickly accessible. This makes the instru-
ment extremely easy to maintain and service.
The outer panels are fastened with Philips head screws and can be
opened with the removal of a few screws. The optical system, with space
for a maximum of 20 channels, the argon valves, the ventilator and an
electronics rack with the evaluation unit and the computer are located
in the upper section of the instrument.
The entire power supply, the source and the high voltage supply are
located in the lower section of the instrument.
Light is transported from the control pistol to the optic by a fiber optic.
Note:
The SPECTROPORT should never be placed on its side
during transport. The PM tubes may be shaken out of
their holders!
SPECTROPURT Front Panel
10. Argon Inlet (only with option spark source) - The argon tube
delivered with the instrument is connected here in order to sup-
ply the instrument with argon.
11. Argon Outlet (only with option spark source) - Argon and con-
densate from the pistol is lead out of the instrument through the
argon outlet.
12. Flow Meter (only with option spark source) - The flow meter
shows the flow amount in liter/h. With normal use, the analytical
argon flow should be ca. 200 I/h.
13. Air Filter - The spectrometer is ventilated through this filter. Pull
off the cover to change the filter.
14. Instrument Fuse - A 6.3 A fuse is used to protect the instru-
ment.
15. Power Cable - This cable is used to connect the instrument to
the mains supply: 220/230 V l o % , 50/60 Hz or 110 V * l o % ,
50/60 Hz. The mains supply must be protected with a 16 A fuse.
The general assembly and individual components of the SPECTRO-
TEST are described in this section. The functions of the various parts
are also explained and an introduction to operation of the mobile spec-
trometer is given.
Chassis Construction
The SPECTROTEST is a modularly constructed analytical system. All
components are easily and quickly accessible. This makes the instru-
ment extremely easy to maintain and service.
The outer panels are fastened with Philips head screws and can be
opened with the removal of a few screws.
The evaluation unit and the computer are located in an electronics rack
behind the rear outer panel.
--
Chassis Construction
The SPECTROTEST JR is a modularly constructed analytical system.
All components are easily and quickly accessible. This makes the in-
strument extremely easy to maintain and service.
The outer panels are fastened with Philips head screws and can be
opened with the removal of a few screws.
The evaluation unit and the computer are located in an electronics rack
behind the rear outer panel. The source is located above the evalua-
tion unit.
Argon Inlet (only with option spark source) - The argon tube
delivered with the instrument is connected here in order to con-
nect the instrument to the pressure reducer.
Printer Connection - A dot-matrix printer can be connected to
the instrument here.
Robotic Connection (option) - A Spectrotest Robotic can be
connected to the Robotic connection.
Flow Meter (only with option spark source) - The flow meter
shows the flow amount in liter/h. With normal use, the analytical
argon flow should be ca. 200 I/h.
Argon Outlet (only with option spark source) - Argon and con-
densate from the pistol is lead out of the instrument to the water
bottle through the argon outlet.
Water Bottle (only with option spark source) - The water bottle
must be at least 3/4 full so that a given back pressure can build
up in the argon system.
Argon Bottle (only with option spark source) - A 10 1 (50 cubic
foot) argon bottle must be supplied by the user. It can be fas-
tened to the instrument here.
Power Socket - The plug is grounded through the mains supply
and can be used to plug in a small grinder, for example.
Printer Socket (option) - This 220 V- plug is intended only for
use with the dot-matrix printer optionally delivered with the in-
strument.
Source Switch - Turn the source on and off with this switch, in
order to service the pistol, for example.
Power Cable - This cable is used to connect the instrument to
the mains supply: 220/230 V 2 lo%, 50/60 Hz or 110 V lo%,
50/60 Hz. The mains supply must be protected with a 16 A slow-
blow fuse.
The Control Pistol
Concord - The removable concord protects the fiber optic from
mechanical damage.
Fiber Optic Holder - The fiber optic is fastened to the holder by
a rotatable rubberpiece. A lens is assembled in front of the fiber
optic and adjusted onto the spark plasma.
Spark Adapter (only with option spark source) - The spark
adapter must be screwed onto the control pistol. It has a con-
nection for an argon tube and a quartz exit window which trans-
mits light to the fiber optic.
Reset Button - Various software functions (calculation of the
average, closing of error messages, etc.) can be performed with
the Reset button.
E. Start Button - The measurement procedure is started with the
start button.
E Good/ Bad Display - This is a good (green)/bad (red) LED dis-
play on the rear side of the pistol for mix-up (sorting) control.
G. Arc Adapter - The arc adapter is plugged onto the control pis-
tol.
H. Argon Tubing (only with option spark source) - The spark
adapter is supplied with argon through the argon tube.
1. Electrode Allen Screw - The electrode can be removed or fas-
tened with the Allen screw by using an Allen wrench (SW2).
J. Argon Outlet (only with option spark source) - Argon, with con-
* There is no water bottle for the densate, is lead from the control pistol through the argon outlet
SPECTROPORT to the water bottle:
Control Pistol
The control pistol is connected to the spectrometer by a 4 m (ca. 12')
long supply hose. The hose can be optionally extended to 10 m (ca. 33').
The light weight and the simple but stable construction ensure optimal
handling for the component of the spectrometer that is subject to the
most wear. The control pistol is suitable for many different applications
through the use of various adapters.
Arc Source
The arc source (arc excitation) is activated when a mix-up (sorting)
control program or an approximate analysis program is selected. The
direct current arc can be selected in three different strengths (maxi-
mum 4 A) depending on the application. The arc is ignited by a control-
led single spark. The ignition voltage is 12 kV; the burn voltage 35 V.
Computer
Utilizing the given software, the computer in the mobile spectrometer
controls all processes, evaluates data and monitors the electronics.
A 3.5" hard drive and an easily accessible 1.44 MB, 3.5" floppy drive
have been installed for data storage.
Printer (Option)
For the documentation of data, a dot-matrix printer for 8 %" continuous
Please see the printer manufactur- paper has been installed in a metal chassis in the SPECTROTESTI
er,s manual for information about SPECTROTEST JR. A printer can be optionally connected to the paral-
operation and maintenance. lel interface on the SPECTROPORT.
Air Filter
The spectrometer electronics are ventilated with cooling air through
this filter. Depending on the environmental conditions (dust, oil fumes,
etc.) it may be necessary to occasionally clean or exchange the filters
on-site. This can be done by opening the left side panel.
CSPEmRo
AnaQtkd ~nsuurnents
Cursor Control
Help and Options
The mobile spectrometers are software controlled analytical instru-
ments. To use the instrument correctly, it is necessary to understand
the software.
Due to space restrictions, the function key field is not available on the
SPECTROPORT.
Cursor Control
The measured values can be stored and processed with various op-
tions in the software. All of the possible options can be listed on the
screen by pressing the <F9> key. The various functions can be se-
lected and started with the <Cursor> and <Enter> keys. Press the
<F1> key instead of the <Enter> key to obtain further help about an
option. A help text is diaplayed on the screen.
The letters behind the options are the "shortcut" keys that can be used
to start a function. Lower or upper case letters can be used.
Working with the option list (<F9>) is very useful in the learning period,
to get used to the software.
The lower row on the screen displays any available options and how
you can exit a dialog.
The following keys can be used throughout most of the software:
Following is a list of the functions that can be used in all dialogs for
stored data:
Select Program
I Approx. Analysis I
Sorting
Burn 11918
See chapter 3 for more information The desired working mode can be selected with the <A.-analysis> or
lhe <Sorting> function keys on the function key field or with the <+JI> and
mode and chapter 4 for the mix-up
<Enter> keys on the keyboard.
(sorting) control mode.
e SPECTRO
Analytlel lnatrumentlr
- - -
Entering a Grade 13
Changing/Copying a Grade 15
Special Functions under <F9> in Dialog <B> 15
3.5.1 Grade Identification 16
3.5.2 Grade Verification 17
3.9.1 <X> Output Mode
3.9.2 <M> Save and Delete Programs
3.9.3 <D> Program Data
3.9.4 <I> General System Settings
3.9.5 <T> Source Parameters
3.9.6 cP> Reports
For most elements, the approximate analysis with spark excitation un-
der argon (option) can be compared with laboratory analysis if the sam-
ples are prepared accordingly. Control of safety related parts, sorting
of LA-metals and exact identification of unknown control pieces are the
most common applications.
The source modes are listed as "Arc" or "Spark" in the row marked
See also the following page. "Source" in the list of Approximate Analysis programs .
If the instrument has not been used for a long period of time, it may be
necessary to flush it for 2-3 min. (<F> key).
Below is an example for an instrument that has been calibrated for iron
based materials (iron alloys):
Select the desired program with the c++> keys (the currently selected
program is highlighted). Load the selected program with the <Enter> key.
Application Window
C i Si Rn P S Cr b Mi A1
I
Total counter
Information field;
System messages
I
Date
I Counter for
single measurements
Entry field for
Information field for
quality identification
Overview of functions
sample identification
Help function
SPECTRU
Analytical Instruments
m 0 Press the <0>key to begin the profiling procedure. The red <Pro-
file> function key on the SPECTROTESTISPECTROTEST JR can be
pressed as an alternative. The following screen is displayed:
Q Wait ca. three minutes for the mercury lamp to stabilize, then begin
with the profiling procedure.
* For the SPECTROPORT: Turn O Rotate the profile dial* clockwise in steps of 25 scale divisions and
the profile screw on the front compare the current intensity value with the previous value.
panel 1/4 of a rotation each time,
using an Allen wrench..
ri
SPECTRO
Analytical Irstruments
The profiling procedure is then completed. Press the < E s o key on the
foil keyboard or the <Reset> button on the pistol to return to the Ap-
proximate Analysis mode.
Please note that single measurements can only be deleted when they
are displayed on the screen. To delete a single measurement, simply
press the number key on the foil keyboard which is shown before the
measurement values. For example, press the <4> key to delete the
first single measurement when four measurements have been con-
ducted.
After calculating the average value, all single measurements are de-
leted from the memory.
A tilde (-) shows that the calibration range has been ex-
ceeded. The value displayed is extrapolated
and not actually measured against calibration
samples.
The measurement values can also be stored for use (e.g., quality as-
surance) at a later time.
The system prompts the operator to measure the first recalibration sam-
ple. The elements that are highlighted and presented in a box are
recalibrated with this sample.
Measure the respective sample several times. The values shown are
intensity ratios and may have a spread of ca. +15% with arc and +3%
with spark excitation. The measurements should be deleted if the de-
viation is greater than this.
A minimum of three individual measurements should be used to calcu-
late the average.
Note:
In the recalibration mode, only the measurements shown
on the screen are used by the computer to calculate the
average.
A low and a high sample are assigned to most of the elements for the
recalibration.
After the average calculation for the last recalibration sample, the
recalibration factors appear on the screen.
The factors give information about the quality of the recalibration. All
factors should be near 1.00. If a factor is smaller than 0.50 or larger
then 2.00 the message "Recalibration Error" appears.
Note:
A bad recalibration may be caused by a defect in the in-
strument, or switching of the recalibration samples, poor
maintenance of the pistol, etc.
SPECTRO
Analytical lrntruments
The instrument also works with values smaller than 0.50 or larger than
2.00. However, the accuracy of the calibration should be checked with
a control sample.
f 8.97
Recalibratlon data
Printout
After checking to be sure that all of these requirements have been met,
switch to the list of Approximate Analysis programs using the <Select
pgm> function key or the <L> key in the list of Approximate Analysis
programs. Select and load the program that is to be the global pro-
gram. Proceed as follows:
0 Open the "Program" dialog with the <D> key,
Q select "Progam Information",
Q select "Global program",
0 enter "Y",
Q close the "Program Information" menu with the < E s o key,
O press the < E s o key again to close the "Program" dialog.
The new global program must then be stored. This can be done in the
"Storage Dialog" opened with the <M> key, entry of the password and
confirmation of the password with the <Enter>key.
Note:
Only one global program can be defined for each matrix
and source type!
Note:
If a program is selected and the <Enter> key pressed by
mistake, a window in which values can be assigned to
the elements opens. These values are used by the op-
matic Program Finder" sec- tional "APF" (Automatic Program Finder) software. Close
tion in chapter 5. the dialog with the <Esc> key.
Entering a Grade
Load the analytical program for which the databank is to be created.
Open the dialog for entry of grades with the cBr key and entry of the
password:
(Ti)
(Ti)
See the illustration on the following Begin entry of a new quality with the <Ins> key. A new dialog opens:
Enter the grade number in the respective text box.
I -
.........
Grade No.
..-.....
-....
: Anl....
- -. . . . . . . . . . . . . . . . . .
~, ~
.-..
.
.. -
...
Grade
-
name : (Ti )
I
Note:
This entry must be made (it serves as the search crite-
ria) !
An entry can be made in the "Grade name" text box, e.g., the short
name or company identification.
Press the <Enter> key. A list of the elements calibrated in this approxi-
mate analysis program appears.
Grade identification is used to search for the grade or grades into which
the analysis on the screen fits.
The system searches the grades databank for the fitting grade after
each measurement or average calculation. The cQ> key can be used
to continue the search for additional grades into which the analysis
fits. Manual search using the <Q> key can also be used in the grade
verification mode and Approximate Analysis when grade identification
is not activated.
When an average is calculated, the control values for the grade are
shown on the screen in addition to the calculated average:
The message "* not found *" is displayed if no grade can be identified.
This may be because the analysis lies outside of all of the stored toler-
ances.
3.5.2 Grade Verification
A list with the possible grades can be called up to the screen with the
4 - 9 ) function key and the grade can be selected.
. .....................................................................................................................................................
I Si tln P S Cr no Hi A1
If the tested material does not fit into the selected grade, then, after
calculation of the average, the values which are above or below the
allowed limits are displayed together with the message "No matchlout
of spec..press any keyn.
A key on the foil keyboard or the <Reset> key on the pistol must be
pressed in order to continue. This also ensures that the message can-
not be overseen.
C Si tln P S Cr no A 1
Depending on the test application, it can be important to print the ana-
lytical data at a later time.
The contents of a data record can be called-up onto the screen so that
the information can be viewed. Select the respective analysis with the
<Cursor> keys and confirm with the <Enter> key.
Note:
Optional conversion programs, that converts the stored
measurement data into a format that can be used i n other
programs (e.g., MS Word, MS Excel or Lotus), can be
purchased from SPECTRO.
Please contact your SPECTRO rep- A special software, DIA-4, can be also be purchased
resentatilye for additional informa- through SPECTRO. This is program for statistical con-
tion. trol of the measurement data.
Depending on the analytical application, it may be useful to activate
type recalibration (standardization). Before this mode can be activated,
analyses must be entered in a list in the same way that grades are
entered in the dialog opened with <B>.
The list for type recalibration is opened for the selected Approximate
Analysis program with the <Zs key and entry of a password:
Press the <Insert> key to enter an analysis for a new type recalibration
grade. The entry screen opens:
...................................................................................................................................................
Type Sanple : 1
Grade Wane : RN 16/5
cF1O> Store
Enter the analysis for the reference material and save it with the <F10>
key.
The type recalibration factors can be viewed or printed after confirm-
ing with the <Enter> key:
( Recalibration Factors I
I Recal ibration Data I
I Printout I
Ni2 A1 Co Cu CuZ Nb Pb Ti V
Take Recalibration
I
The type recalibration is stored on the hard drive and the type
recalibration factors are applied to analyses when the Yes menupoint
is chosen.
Once the type recalibration has been activated, the <F6> key can
be used to return to the "normal" analytical mode.
a The cF7> key can be used to return again to the type recalibration
mode.
As with the <R> function for a recalibration, the type recalibration fac-
tors can be displayed by activating the cF4s key.
CHAPTER 3 4- PAGE
24
5 P E ~ n
...... ~ , Tr>-y
3x..&??~ ~J@~
, , *>. ,\':.:... :.i
?
-
Entry of a password is necessary in order to open some dialogs or to
store changes on the hard drive.
The ensures that only authorized personnel are able to make any
changes to the system.
During installation of the system, the person(s) resposible for the sys-
tem is shown the method for changing the password. In this way it is
possible to regularly change the password, ensuring protection of the
system.
Note:
If the password is changed, it should be noted, but not
near the system.
The dialog described below is used to set the parameters for the out-
put options. For example, automatic printing of the average value, turn-
ing on grade identification, etc.The output types are displayed on the
left side of the monitor and the selected settings on the right:
C Si tln Cr no Ni A1 Co Cu
I Output mode
Sample I.D.
Grade I.D. is Off
Uithout sample I.D.
I
Automatic Storage of hnalysis No storage
Printer node No printout
Printout of Recalibration data No printout
Calculation for Reference Calculation is On
Analysis to e x t . PC disabled
Show the grades single form
Automatic Program Finder Not Active
SCREEN OUTPUT
Small characters: analysis in standard size characters, a
maximum of four measurements at the
same time on the screen.
Small with average: like above, but always with the current av-
erage.
Large font one line: analysis in large characters, one measure-
ment can be displayed.
Large font multiple lines: depends on the number of elements being
analyzed.
Large font with average: like above, but always with the current av-
erage.
GRADE MODE
Grade I.D. is Off: only manual grade identification with the
<Q> key.
Grade identification: automatic grade identification after each
measurement and each average value.
Grade verification: the system checks after each measurement
See also chapter 3.5.2: "Grade Veri- and each average value whether or not the
fication ". analysis fits into the given grade.
SPECTRO
Analytlel lrntrumentm
SAMPLE I.D.
Without sample I.D.: no entry for sample I. D. is available.
With sample I.D.: after each average calculation, the system
requests a sample I.D. or name, in order to
simplify assignment of the analysis for print-
ing or storage. Entry of sample I.D. can be
manually activate with the <A> key or
skipped with the < E s o key.
PRINTER MODE
Printout disabled: automatic printing is turned off.
Average: each average value is automatically printed.
Print each burn single measurements and averages are
and average: automatically printed.
ANALYSIS TO EXT. PC
Disabled: Data transfer is not activated.
Averages: Averages are transferred using the
SPECTRO 4.3 format.
To store, copy or delete programs, open the following dialog with the
<M> key:
I
Saue a new progran
Updating progan -FE-BlUO-
Delete a program
Note:
All of the other dialog options contain settings for the pro-
gram information which, when incorrectly changed, can
lead to poor measurements and functions which do not
work correctly. Changes should only be made by person-
nel trained by SPECTRO.
I
Elerent selection
Remlibrat. sanplcs
Recalibration values i V U Fe
Channels
Reference monitoring
Interelement corrections
Print progran data
Matrix correction enabled
REFERENCE MONITORING
Tolerances for the intensity for the reference element (e.g., Fe in Fe-
base) can be entered in this dialog. This can help in the recognition of
poor measurements.
If the given values are exceeded, the following message appears on
the screen:
This dialog should not normally be selected, because all of the ana-
lytical programs can only function properly when the correct values
have been set.
I Reports I
C
I h i n t a report
Stare a report
h i n t stored reports
Create text for a l l test records
1"
I
Fe
PRINTOUT LAST MEASUREMENT/PRINTOUT
LAST AVERAGE
The last single measurementlaverage is printed.
PRINT A REPORT
The report is printed. If "Text for all test records" has been entered,
then the system prompts for them before printing begins and customer-
specific information, for example, can be entered.
STORE A REPORT
If the report is to be printed at a later time, then it is possible to store
it here. This is only possible if "Text for all test records" has been
See below. previouslv entered and stored.
Your company name, for example, can be entered as a header for the
report in the dialog that opens. It is also possible to enter, for exam-
ple, the customer name, operator's name, charge, supplier, etc. as
criteria for the report.
A list of the functions available for Approximate Analysis can be opened
with the <F9> key.
Auerage E
Enable/disable flush *ode F
Suitch to sorting mode J
Recalibration K
Load a program L
Display state of neasuring N
Prof ile 0
Register: Stored measurements G
Print/Store records P
<E> AVERAGE
With the cE> key, the average of any number of measurements can
be calculated.
<K> RECALIBRATION
See also chapter 3.4. E!l The recalibration (standardization) is used to adjust the system.
To load a different Approximate Analysis program, press the <L> key.
A list of analytical programs is displayed. If changes have been made
to the currently loaded program, then the following question appears
on the screen:
The changes are stored on the hard drive when "Update the program"
is selected. The system switches directly to the list of programs if "No
updaten is chosen.
This function is generally used only for maintenance and service op-
erations.
<O> PROFILING
See chapter 3.2. Profiling is used to adjust the optical pathway.
<P> PRINTISTORE
RECORDS
<F7> ANALYSIS w r r ~ o u
TYPE
~ RECALIBRATION
The type recalibration is deactivated.
4.2.1 Calibration
4.2.2 Testing
4.3.1 Calibration
4.3.2 Testing
Sorting task
Reference element
Tolerances for sorting
Switch source mode
Source conditions
Tolerances for calibration
No of measurements for calibration
Switch drift correction
No. of repeats for material mix
Attach an analytical program
Store I delete a program
Print program data
Counter for electrode cleaning
New electrode before the calibration
<K> New calibration
<L> Select program
<Z> Display intensities
<O> Profile
<P> Storelprint Test Record
<R> Check calibration
<Y> Change to APPROX. ANALYSIS
<W> Counter for electrode replacement
<D> Program dialog
cF> Enableldisable flush (analytical flow)
<Alt> + <Q> Exit to DOS
Mix-up control (sorting) is a comparison measurement for the quick
control of large amounts of the same material.
If the measured value lies outside of the given tolerance values, then
an error message is displayed on the screen. The elements that are
being monitored and that lie outside of the tolerance values are dis-
played. The test pistol is blocked until the error message has been
confirmed. A new measurement can then be conducted.
A report about the tests can be printed when finished with the mix-up
control.
Select Program
Approx. Analysis
Sorting
Burn 11918
The programs available for Sorting (mix-up control) are listed after the
"Sorting" option has been selected with the <Cursor> keys and con-
firmed with the <Enter> key or with the <Sorting> function key.
The desired program can be selected with the <Cursor> keys and
For more information about creat- loaded with the <Enter>
ing a new Sorting program, see
chapter 4.4. A new program can be created by pressing the <Ins> key in this list.
SPECTRO
Analytknl Instruments
Sorting programs are often created only for arc excitation. This is be-
cause sorting (mix-up control) is a measuring method for which the
time required often plays a deciding roll; the quicker the procedure can
be performed the better. An arc measurement requires only about three
seconds, whereas a spark measurement can last about eleven sec-
onds. The time for sample preparation for spark excitation must also
be taken into consideration.
In the example, the entry "-I-"can be seen under the "Pgm" heading for
one program, but a program name (e.g., FE-01MO) is entered for two
of the programs. This is because there are two modes for working with
the Sorting program. One mode is to work without identification of the
reference sample and the other mode is to work with identification of
the reference sample. The difference in the two modes consists of as-
signment of an analytical program. When a program is assigned, it is
also possible to obtain an approximate analysis.
The two modes for working with the Sorting mode are described in the
following sections. There is also a description of the procedure for cre-
ating a new Sorting program.
K SPECTRO
Analytical Instruments
Note:
See also chapter 3.2. Before beginning with Sorting (mix-up control), the opti-
cal system of the spectrometer should be profiled.
4.2.1 Calibration
Measurement 1
seethe illustration on thefollowing A message appears on the screen when the calibration has been suc-
Page. cessfully completed.
>
P-
CHAPTER 4 -IPAGE
= 4 SPECTR W
Fe1 : 56340
Start sorting
4.2.2 Testing
If the test piece conforms to the tolerances set for the reference sam-
ple, then the "Pass" counter and the "Total" counter are both increased
by one.
Pass : 1
Continue sorting
If the test piece lies outside of the set tolerances, then an error mes-
sage appears on the screen. All of the elements which fail are dis-
played.
1 ~ ~ P EInstruments
CTRO
Analytical
1 SORTING [MIX-UP CONTROL]
The error message must be confirmed (reset) before sorting can con-
tinue.
The elements that lie outside of the tolerance ranges are displayed in
the error message along with the amount by which they deviate. For
example, in the example above, "V = 1399". This means that the inten-
sity for vanadium is 1399% higher than the expected intensity stored
for the reference material for vanadium.
As for during the calibration, the system prompts for a repeat measure-
ment on the same piece:
Total : 2
Pass : 2
Repeat
If three fail measurements are obtained in sequence, then the test piece
is probably a material mix-up.
This is displayed on the screen:
..........-..........-.-.......--..--.-.......-.-............... .-...........................................................................
I
H
Total : 4 Fe1 : 47151
Pass : 3 Fail : 1
Continue sorting
Remove the test piece which has failed and test the next piece.
Note:
Before beginning with Sorting (mix-up control), the opti-
See also chapter 3.2. cal system of the spectrometer should be profiled.
Note:
Because this working mode refers back t o an Approxi-
mate Analysis program, the respective analytical program
must be recalibrated before Sorting can begin!
4.3.1 Calibration
Measurement 1
A message appears on the screen when the calibration has been suc-
cessfully completed.
The calibration can be directly accepted with the <Start> button on the
test pistol and sorting can be started.
The current program step can be stopped with the <Esc> button. The
system asks if the grade identification is correct:
\I
Identification 0.k. ?
I
Select either "Yes" or "No" to switch into the sorting mode or to start a
new calibration, respectively.
The Sorting window opens when the calibration has been completed
and accepted. Testing of the pieces can then begin.
Fel : 45862
Calibration finished
Start sorting
/
{~o~*n~;a$~;;~,>.;y.:.'f;.:.';.'<
.~3~&:~::!:.~:::,;.:,'::]:,:;::~$~:$~~~~ ?:{if,
:'i:';.'{ j:..;:;.;'.::.;<..:;;<:;::{B~~~y~f~~
9&.:1
4.3.2 Testina
Pass : 1
Continue sorting
If the test piece lies outside of the set tolerances, then an error mes-
sage appears on the screen. All of the elements which fail are dis-
played.
The error message must be confirmed (reset) before sorting can con-
tinue.
SPECTRO
Anslytltlel Instruments
As for during the calibration, the system prompts for a repeat measure-
ment on the same piece:
Pass : 1
i
Repeat 1
IB2K?7w!37 llBt29 1
cF1> Help cF9> Opt~ons
If three fail measurements are obtained in sequence, then the test piece
is probably a material mix-up. This is displayed on the screen:
See the illustration on the following The approximate analysis for the piece that does not fit within the toler-
Page. ances is displayed when the error message is closed.
>
.. press any key .. NOmatchlout of spec cF5> Printout
Pass : 1 Fail : 1
Continue sorting
Remove the test piece which has failed and test the next piece.
SPECTRO
Analytical Instruments
After pressing the <Sorting> function key or using the <Cursor> and
<Enter> keys to select the Sorting mode or (from a Sorting program)
pressing the <L> key, the list of Sorting programs is displayed.
Note:
This dialog can only be closed after entries have been
made for the first four dialog points and the program has
been stored.
Enter the information starting from the top and proceeding to the bot-
tom of the dialog.
SORTING TASK
Enter a description of the program here.
REFERENCE ELEMENT
In order to minimize external influences, the test channels are meas-
ured in relation to the reference channel. The ratios generated in this
way are used for sorting and the approximate analysis; i.e., intensity
ratios are used during sorting instead of concentrations.
A reference channel must be assigned in order for this to work.
1
General settings for sorting node
Sorting task
Reference element
Tolerances for sorting Element
Suitch source node
Source conditions
Tolerances for calibration
No. of measurings for calibration
Suitch drift correction
No.
Attach
of repeats
an analytical
for material
programmix Fe2
Fe1
Note:
Sorting with an assigned Approximate Analysis program
is an exception to this rule, as the reference element is
set b y the analytical program. I n this case, no entry
should be made i n this dialog point.
I no
At
s t F=Z
Pr
Co
CI
si
no
Cr
Fs
NIZ
iee
15
e
B
t3
e . i .
ss
ZB
B
8
z
-/-
8
off
!
The tolerances are stored and the window closed with the c E s o
key.
Note:
When the Sorting program is calibrated, the reference
sample used to set the tolerances must be used to cali-
brate the program ! This is the only way to ensure proper
operation of the Sorting program.
S w r r c ~SOURCE M O D E
As discussed in the introduction, Sorting is generally performed with
arc excitation. This is also the standard setting for the software.
If spark excitation is, for some reason, required, then this can be set
here.
Note:
This dialog point is unnecessary if a Sorting program has
been assigned. The settings for the Sorting program are
used.
S OURCE COND~TIONS
The source must be activated, before measurements can be conducted.
A measurement time for a single measurement must be entered.
Note:
This dialog point is unnecessary if a Sorting program has
been assigned. The settings for the Sorting program are
used.
The source parameters for several alloys are given in the table below:
13 13 13 3 I
Prespark
10 7 7 7 7
i n seconds
SPECTRO
Analytlcsl Instruments
The standard setting for this point is three. If more than three measure-
ments are desired for calculation of the average for the calibration, this
value can be set up to a maximum of ten measurements.
The number of repeat measurements can be set from zero to four. When
this function is set to zero, then for each poor measurement, the op-
erator must decide whether to accept the test piece anyway (with only
one measurement, it could be a poor measurement) or to discard the
piece (it may be a material mix-up).
Note:
The "Fail" counter i s always increased by one when a
material mix-up i s displayed.
ATTACHAN ANALYTICAL PR OG R A M
See also chapter 3.8.2. When the password has been correctly entered and confirmed, a win-
dow in which various options for storage and deleting of programs are
listed opens.
Select progran L
Display intensities 7.
Prof i le 0
StordPrint Test Record P
Check calibration R
Change to hPPROX. hNALYSIS Y
Counter for electrode replacement W
Program dialog D
Enable/disable flush(analytical flow) F
Exit to DOS <ALT>+<Q>
cO> PROFILE
See chapter 3.2. See the "Profiling" section in the "Approximate Analysis" chapter.
cP> STOR~PRINT
TEST RECORD
This function is used for the documentation an printing of reports. An
additional window with various options opens when the <P> key is
pressed.
Exit without printout Return to Sorting with this function.
Print a report Directly print the report for the current Sort-
ing procedure.
Store a report Store the test report for the current Sorting
procedure.
Print stored reports Select and print stored test report(s). A list
of stored reports opens.
Text for all test records Define the text for all test reports. This is
similar to the text for Approximate Analysis
test reports.
SPECTRO
Analytical Instruments
Drive and file names The drive and file name under which the test
for storage report is to be stored can be entered here.
<R> CHECKC A M B R A ~ O N
This function can be used to check the calibration with the reference
sample during sorting, without conducting the complete calibration pro-
cedure. The test is started when <R> is pressed.
Check calibration
When the reference sample has been successfully measured, the sys-
tem returns to the Sorting mode.
<F> ENABL~DISABLE
FLUSH (ANALYTICAL FLOW)
If spark excitation is used for the Sorting program, this key can be used
to flush the system with argon before beginning to calibrate.
Several settings must be made in the system software before APF can
be used.
See chapter 3.4.1. Follow the steps described in the "Global Recalibration" chapter. When
the list of program assignments is opened with the <H> key, minimum1
maximum tolerance values must be entered for various elements (similar
to those for grade verification).
Load the first assigned program and press the <Enter> key. The fol-
lowing window opens:
t ion
1 <FSC> ~ x i t ~ F I O >Store
Enter the tolerances that are to be used in the analytical program. Load
the next program and enter the tolerances. Repeat this for each ana-
lytical program.
I Output mode I
Scree z e : small
Grade Automatic Program Finder dentif icat ion
--
Print 1
Print Semi-Automatic
Full-flutomatic
D isp 1 Each measuring
Iuerages
t disabled
enabled
Data d
Automatic Program Finder disabled
There are four different options that can be selected under the "Auto-
matic program finder" dialog point.
disabled: APF is deactivated
Semi-Automatic: APF switches to the respective analytical pro-
gram after the first measurement. All additional
measurements and averages are conducted in
the program selected by the system.
Full-Automatic: APF switches to the respective analytical pro-
gram after the first measurement. All additional
measurements are conducted in the program
selected by the system. After the average has
been calculated, the system returns to the glo-
bal program. The average is also carried into
the global program. Display of the average re-
mains on the screen until a new measurement
is conducted.
Each measurement: APF switches to the respective analytical pro-
gram after the first measurement. As soon as
the measurement value is displayed, the sys-
tem returns to the global program. The meas-
urement value is also carried into the global
program. This procedure is repeated for each
measurement. The same procedure is automati-
cally started after an average is calculated.
Analytical Instruments
MINIOPTIC
UV 150 M M
@ Activate the analytical argon flow by pressing the<F> key. The mode
for the argon flow is displayed in the status line on the screen.
0 After the argon system in the instrument has been flushed for
ca. 1 min., connect the minioptic to the argon supply. This proce-
dure is necessary to make sure that any contamination in the sys-
tem cannot influence or be deposited in the minioptic.
See chapier 3.4. @ Recalibrate (standardize) the analytical program in the normal way.
0 Open the <N> dialog and switch to the display of raw intensities.
@ Use the Allen key (2 mm) delivered with the instrument to rotate the
Allen screw on the minioptic by a quarter rotation, clockwise.
@ Then conduct steps 8 and Q turning the Allen screw counter clock-
wise until the maximum value for phosphorus (the profile position
for the minioptic) has been found.
See also chapter 3.2. This procedure can be compared with the profile procedure using the
mercury vapor lamp.
SPECTRO
Analytical Instruments
Note:
The adapter should always be closed when no measure-
ments are being conducted with the instrument. This
ensures that the constant argon flow can properly main-
tain the argon supply for the minioptic!
Backing-Up Data ii
Loading (Restoring) Data iii
Maintenance Note iv
Air Filter iv
Keyboard and Monitor Glass iv
Printer iv
Argon System (Option) v
Burn Spot v
Test Pistol vi
During Testing (Arc Excitation) vi
During Testing (Spark Excitation) vii
General vii
The Test Pistol viii
SPECTROPORT X
SPECTROTEST xiii
SPECTROTEST JR xvi
Backing-Up Data
Please have formatted 3.5" HD (1.44 MB) diskettes available for the
data back-up. Place a a diskette of the type mentioned above into the
floppy drive in the instrument and open the dialog with copy functions
with the <C>key.
ly functions
If more than one diskette is required for the back-up, the system auto-
matically prompts the operator for additional diskettes. The diskettes
should be numbered in the order in which they are used. This is neces-
sary if it is necessary to restore software from the diskettes. They must
be restored in the order in which they were backed-up.
The analytical programs for the Approximate Analysis mode are saved
on diskette when this option is chosen.
SAVE SORTING FILES
The sorting programs for the Sorting (mix-up control) mode are saved
when this option is chosen.
All of the lists of grades (qualities) are backed-up with this option.
All of the files in which analytical data is stored are backed-up when
this option is chosen.
All of the reports stored for the Approximate Analysis mode are saved
when this option is chosen.
This must also be started from the Approximate Analysis mode. Open
the dialog with copy functions using the <C> key. Select the "Load data
from floppy disk" option after entering the password in order to restore
See above. data. Functions corresponding to those for storage of data are avail-
able in the window that opens.
The respective data can then be restored from the diskettes onto the
instrument computer.
SPECTRO
Anelytlral Instnunents
Maintenance Note
-
SPECTRO Service in USA
Tel.: 508-342-4471
Fax: 508-342-8695
Air Filter
Air to cool the instrument is filtered through this air filter. This must be
periodically changed (depending on the air contamination and use of
the instrument) to ensure proper cooling of the instrument.
Note:
Disconnect the instrument from the mains power.
Note:
Do not use any sharp objects for cleaning.
Printer
The printer should be periodically cleaned with a small vacuum cleaner
(when the paper is changed). The print head should be cleaned with a
damp, soft cloth. All other maintenance (papertribbon change, etc.)
should be taken from the manufacturer's manual.
Argon System (Option)
The SPECTROTESTISPECTROTEST JR and the SPECTROPORT can
be operated with an argon bottle (mobile operation) or with a stationary
argon supply system (stationary operation).
The argon flow rates are set to the correct values during manufacture.
If the values deviate from those listed above when tested, then they
can be adjusted to the correct values with the respective knobs.
The "spark adapter" must be closed when the settings for the argon
flow rates are adjusted.
ARGON FLUSH
If the instrument has not been used with spark excitation for a long
period of time or if the argon bottle has been changed, the analytical
flow can be manually activated in a spark program with the <F> key to
quickly flush the system with argon.
Activate the analytical flow with the <F> key and close the adapter and
the argon outlet. Quickly open the outlet or the adapter to remove the
condensate with the built-up pressure.
WATER BOT~LE
(NOT SPECTROPORT)
The water bottle on the rear side of the instrument functions as valve.
In order for the argon to exit through the bottle, the argon must flow
with enough pressure to overcome the water in the bottle.
Burn Spot
A spark discharge generates a plasma which contains information about
the contents of elements in the sample.
SPECTRO
Analytical Instruments
The spark discharge is divided into the prespark and the measurement
(integration).
When this homogenization phase has finished and all of the precipi-
tates, even in deeper layers, have been removed, then the sample is in
the stationary spark condition. Concentrated discharges occur. The
sample suface is remelted point for point and is characterized by a
metallic, shiny surface with a homogeneous crater. A black conden-
sate forms along the edge of the burn spot.
The actual measurement begins (integration).
The optimal burn spot depends on numerous factors and may appear
differently from system to system. For this reason, it is best to have the
installation engineer (or a service engineer) explain and demonstrate
good versus bad burn spots for a particular system.
Test Pistol
The metal condensate on the tip of the electrode is from the samples
sparked. It can be removed with the spark plug brush (wooden handle)
delivered with the instrument. The adapter can be removed and then
the electrode cleaned for difficult to remove contamination.
The lense or window in front of the fiber optic should be cleaned with a
cotton swab.
General
Note:
For safety reasons, the source must be switched off dur-
ing maintenance and when changing the adapter.
There is a spring below the electrode for better adjustment in the test
pistol. The electrode gap should be set with the adjuster to 1.5 mm for
arc excitation and 2.0 mm for spark excitation (option).
SILVER ELECTRODE
The silver electrode is only intended for use with arc excitation.
Before sharpening the tip ca. 500 measurements per side are pos-
sible.
It is not possible to analyze the element silver with this electrode.
The silver electrode is only intended for use with arc excitation.
Before sharpening the tip IW ca. 1000 measurements per side are pos-
sible.
It is not possible to analyze the element copper with this electrode.
TUNGSTEN
The tungsten electrode is only intended for use with spark excitation.
The tip of the electrode is worn very little in an argon atmosphere.
A special tool is needed to sharpen the tip of this type of electrode.
For the spark adapter (option), it is also necessary to remove and clean
the quartz window.
The window or lense in front of the fiber optic holder are held in place
by a metal cap and must be replaced when extremely dirty.
Note:
If the SPECTROTESTISPECTROTEST J R or SPECTRO-
PORT is equipped for the analysis of carbon, then only
diffuse quartz windows may be installed.
Careful sample preparation is required for accurate spectrometer analy-
ses. The type of sample preparation depends on the sample material.
Catalogues and/or informational Copper and aluminum samples should be prepared on high speed lathes
rnaterial is available upon request. Or milling machines.
A belt grinder or a hand held disk grinder with a rough grinding disk is
normally sufficient for sample preparation in the factory.
Magnetic samples strongly divert the light arc. This is why such sam-
ples cannot always be sparked. This is also true for samples on which
the surface does not conduct electricity because it has been treated or
has developed a patina. This can generally be solved by lightly grind-
ing the surface of the material.
Portable spectrometer for sorting control and the sorting, identification
and analysis of metals. It can be equipped with a maximum of 20 ana-
lytical channels.
ARC SOURCE
DC arc, maximal current intensity of 3.2 A
choice of 3 current intensities
Ignition of DC arc by controlled single spark
Gate trigger voltage: 12 kV
Arc voltage: 35 V
SOWARE
Operation can be driven by menu or by function keys
User-friendly through context sensitive help
Operator instructions on screen - if necessary
SORTING CONTROL
Up to 100 programs for the sorting mode
Tolerances relative to calibration values
Symmetrical and asymmetrical tolerances
Failure messages with display of elements and relative deviation
Test report can be printed and stored
Editable text for test reports
Automatic calibration control using given tolerances and
numbers of measurements
Programmable operation control (number of burns, etc.)
Programmable counter for electrode cleaning
ELECTRICAL REQUIREMENTS
230 V/11 OV +1OO/o/-15%, 50160 HZ
800 W during integration
150 W in standby mode
16 A slow-blow fuse
OPTIONS
Printer
Special adapters for different applications
Minioptic for the determination of P and S
Automatic Program Finder (APF)
Subject to change
Stabilized against fluctuations in temperature by the geometric
construction and materials used
Paschen-Runge mounting
Rowland circle diameter (focal length): 750 rnm
Concave holographic grating: 2400 grooves/mm
Grating material: Zerodur
Effective wavelength range: 185 - 530 nrn
Reciprocal dispersion: 0.55 nmlmm (first order)
Hg line as optical monitor
SO~ARE
Operation can be driven by menu or by function keys
User-friendly through context sensitive help
Operator instructions on screen - if necessary
GRADE IDENTIFICATION
Storage of grade types for each analytical program
Display of grade numbertgrade type
Grade identification (system searches for corresponding grade)
or grade verification (given grade is checked)
Up to 64 analytical programs for the approximate analysis mode
Output of intensities, intensity ratios, standardized and
corrected intensity ratios
Output of analyses in weight percent
Standardization (one or two point)
Additive and multiplicative interelement correction
matrix correction
Automatic line switches
Calculation of average, standard deviation and
variation coefficient for any number of burns
Marking of results beyond calibration range
Entry of sample identification information
Transmission of analytical results to an external PC,
additional printers and/or monitors
Automatic hardware diagnostics
Storage of reports and analytical results
ELECTRICAL REQUIREMENTS
230 VI11 OV +lo%/-1 5%, 50160 H Z
800 W during integration
150 W in standby mode
16 A slow-blow fuse
OPTIONS
Special adapters for different applications
Minioptic for the determination of P and S
Automatic Program Finder (APF)
Subject to change
Stabilized against fluctuations in temperature by the
geometric construction and materials used
Paschen-Runge mounting
Rowland circle diameter (focal length): 500 mm
Concave holographic grating: 2924 grooves/mm
Grating material: Zerodur
Effective wavelength range: 185 - 520 nm
Reciprocal dispersion: 0.68 nm/mm (first order)
Hg line as optical monitor
A RC SOURCE
DC arc, maximal current intensity of 3.2 A
choice of 3 current intensities
Ignition of DC arc by controlled single spark
Gate trigger voltage: 12 kV
Arc voltage: 35 V
SO~ARE
Operation can be driven by menu or by function keys
User-friendly through context sensitive help
Operator instructions on screen - if necessary
GRADE IDENTIFICATION
Storage of grade types for each analytical program
Display of grade numberlgrade type
Grade identification (system searches for corresponding grade) or
grade verification (given grade is checked)
Up to 64 analytical programs for the approximate analysis mode
Output of intensities, intensity ratios, standardized and
corrected intensity ratios
Output of analyses in weight percent
Standardization (one or two point)
Additive and multiplicative interelement correction
matrix correction
Automatic line switches
Calculation of average, standard deviation and variation
coefficient for any number of burns
Marking of results beyond calibration range
Entry of sample identification information
Transmission of analytical results to an external PC,
additional printers and/or monitors
Automatic hardware diagnostics
Storage of reports and analytical results
ELECTRICAL REQUIREMENTS
230 V/11OV +1OW-15%, 50160 Hz
800 W during integration
150 W in standby mode
16 A slow-blow fuse
OPTIONS
Special adapters for different applications
Minioptic for the determination of P and S
Automatic Program Finder (APF)
Subject to change
AUFTRAGSBESCHREIBUNG / JOBDESCRIPTION
FUR / FOR
AUFTRAG NR. / ORDER NO.
Verteiler/Distribution: HGW,HEK,NDL,SUG,EDH
LABTEC,HAD,HGD,FRM,KUH, ?
02.04.1998
Analytical Instruments GmbH
.
BoschstraOe10 D-47533 Kleve
Optik 1 : 29241400
I l l I I 1 I I I I I I I I I I
INr. / E l . IWellenl I T lSbr I FE-BASE ( I I I 11121314151
8 8 ! S , , . . , . ,
a. ) No guarantee !
9678/98
ez.:
02.04.1998 SPECTRO
epr.:
ea.:
I Analytrd Instruments
- 1
1 Analytical Instruments GmbH
.
BoschstraOe10 D-47533 Kleve
.
gs (02821) 892-0 Tx 811 734 sptro d
Fax(02821)23144
HARD WAR& /SOFTWARE- STATUS
Taktfrequenz: MHz
Drucker:
Ausfiihrung
Fabnkat, Typ: SIN:
O A4-Fomat CI A3-Fonnat Cl Endlospapier C1 Einzelblatt
I
Ia EPC 3 a EPC 386 EI EPC 486 I
Taktfrequenz: J
- MHz BIOS: %3) - TJ- Besonderheiten:
V. Wcitcre Mn~nnlimcn/Sonstigcs: