MODEL 224/224C

DIFFERENTIAL PRESSURE UNIT (DPU)


(PRIME M224C for Industrial Service and
Barton® brand M224 (non-C) for
Nuclear Service)

Installation Manual
(Replaces/Supersedes ID#10085)
Version 06D76a
ID#10087
4/2006
[This manual is for DPU only - see separate instrument manual.]
CONTENTS
Prepages ............................................................................... Page 2
Section 1 - Introduction
1-1. General.................................................................... Page 3
1-2. Product Description .................................................. Page 3
1-3. Specifications ........................................................... Page 3
Table 1-1. DPU Material/Range Specifications ............ Page 4
1-4. Theory of Operation ................................................. Page 5
Section 2 - Installation
2-1. General.................................................................... Page 7
2-2. DPU Mounting ......................................................... Page 7
2-3. DPU Piping.............................................................. Page 7
2-4. DPU Piping/Startup Examples................................... Page 9
Section 3 - Maintenance and Calibration
3-1. Required Tools ......................................................... Page 19
3-2. Test/Calibration Equipment ....................................... Page 19
3-3. Preventative Maintenance .......................................... Page 19
3-4. Calibration ............................................................... Page 19
3-5. DPU Cleaning/Inspection.......................................... Page 19
A. Pressure Check .................................................... Page 20
B. Cleaning and Inspection Procedure......................... Page 20
Table 3-1. Bolt Torque Ratings .............................. Page 21/22
3-6. Servicing .................................................................. Page 23
3-7. Attaching Drive Arm to Torque Tube .......................... Page 24
3-8. Drive Arm Tightness Test .......................................... Page 24
Section 4 - Troubleshooting
Table 4-1. Troubleshooting ............................................... Page 25
Section 5 - Dimensions ........................................................... Page 26/27
Section 6 - Parts Drawing/List................................................. Page 28

Measurement Products
(Formerly Barton Instrument Systems)
SAFETY
Before installing this instrument, become familiar with the installation instructions
in Section 2 and in the separate instrument manual.
DANGER notes indicate the presence of a hazard which will cause severe
personal injury, death, or substantial property damage if warning is ignored.
WARNING notes indicate the presence of a hazard which can cause severe
personal injury, death, or substantial property damage if warning is ignored.
CAUTION notes indicate the presence of a hazard which will or can cause
moderate personal injury or property damage if warning is ignored.
DANGER, WARNING, and/or CAUTION notes that appear on the following pages
of this manual should be reviewed before proceeding: 3, 7, 8, 19, 20, and 23.
(Important! Before installing or operating this instrument, review all safety
notices contained in the separate actuated instrument manual.)
PRODUCT WARRANTY STATEMENT
Product warranty for this instrument is as stated on the back cover of this manual.
NOTICE: DPU warranty will be voided if the following limitations are exceeded:
Temperature - Do not subject the DPU to temperatures above 200°F or below -60°F
(+40°F minimum for water filled units).
Pressure - Do not subject the DPU to operating pressures in excess of the working
pressure rating stamped on the unit or attached dataplate.
Corrosion - Do not subject the DPU to incompatible process media.
Sealed Components - Do not loosen or remove the torque tube gland nut, the drive
arm hole plug, or the torque tube housing from the centerplate. To do so will cause loss of
bellows fill fluid and render the unit inoperable.
Vibration/Shock - Do not subject the DPU to severe mechanical or hydraulic shock.
Should any after-delivery problems arise, please contact PRIME Measurement Products
Customer Service, during normal business hours, at (626) 961-2547.
"C" VERSION DESIGN CHANGE
The 224C is a redesigned version of the 224 DPU. The "C" version is identical in func-
tion, performance, installation, and operation to the previous version - redesign was for
improved manufacturing only. This design change does not affect the instrument being
actuated. Barton-brand Model 224 (Non-C) DPU configurations are for
Nuclear service only.
RECORD OF CHANGES
VERSION DATE DESCRIPTION (initial issue 95E1 [5/95])
99D8 8/99 Updated specs., Table 4-1, Table 5-1, and Section 6 drawings
01D18d 4/01 Revised to Booklet Format
03G50c 7/03 Rev. Co. Name/Logo; Rev. Sections 1, 2, & 3 for PED compliance
03H50b 8/03 Rev. Spec. Table 1-1; Combined 224 and 224C (replaces 10080); Add
(ID#10087) 224 DPU information to Section 6
04C50b 8/03 Produced separate manual under Barton brand for Nuclear/Government
(ID#10085) Service (No Tech. Changes from ID#10085)
06C75m 4/06 Eliminated separate Barton-brand manual (ID#10085); Rev. Safety/ "C"
(ID#10087) Ver. statements; Rev. Tables 1-1, 3-1, and 6-1; Added Piping/Startup
Examples (para. 2-4); Rev. housing bolt caution to warning (page 8); Rev.
Figure 6-1; non-technical changes throughout

06D76a 4/06 Removed Government Service designation; No Tech. Changes.

2
SECTION 1 - INTRODUCTION

1-1. General
The Model 224/224C Differential Pressure Unit (DPU) is a mechanical device
that measures differential pressure relative to a gas or liquid flowing through a
process system, or to the level of a liquid contained in a process vessel.
For process flow measurements, the DPU is connected across a primary device
(e.g., venturi, orifice plate, or flow tube) located in the process system.
For liquid level measurements, the DPU may be connected in a variety of ways
to measure the difference in pressure caused by variations in the level of the
liquid in the process vessel.
1-2. Product Description
The Model 224/224C DPU is a dual bellows assembly enclosed within pressure
housings. The dual bellows assembly consists of two opposing internally con-
nected liquid-filled bellows, a center-block, range springs, overrange valves, and
a torque tube assembly.
The pressure housings are connected by pipe or tubing to the primary device
located in the system piping. Variations in differential pressure within the pres-
sure housings cause the bellows to expand or contract in a linear direction
towards the side having the lowest pressure.
The linear movement of the bellows is converted into angular rotation when
transmitted to the torque tube shaft by the drive arm and this mechanical motion
actuates the mechanism of the process monitoring instrument.
The process monitoring instrument that is connected to the torque tube assem-
bly may be an indicator, a switch, a transmitter, a recorder, or other process
control device.
1-3. Specifications
The Model 224/224C DPU is available in various pressure ratings to measure
to specific ranges between 0-30 inches of water column and 0-1000 psi, with
safe working pressure (SWP) ratings from 500 to 10,000 psi (224 Non-C lim-
ited to 500 to 3,000 psi SWP). See Table 1-1, page 4.
The 224C Bellows Unit Assembly (BUA) is produced in three bellows sizes:
1-5/8 inch, 5/8 inch, and 3/4 inch diameter.
The 1-5/8 inch diameter BUA accommodates differential pressures to 55 psi.
The range springs of this BUA are contained within the bellows and do not
come in contact with the measured liquid.
The 3/4 inch BUA accommodates differential pressures up to 400 psi and the
5/8 inch BUA up to 1,000 psi. The range springs of these BUAs are grouped
around the outside of the bellows and must be of a material that is compatible
with the liquid being measured.

CAUTION
Ambient temperatures below +40°F should never be allowed for water
filled (D-fill) units.

3
4

224/224C DPU Specifications


Table 1-1. 224/224C DPU Material/Range Specifications (4/06a)
BODY AVAILABLE DIFFERENTIAL PRESSURE RANGES PRESSURE CONNECTIONS

Stainless Steel or
Beryllium Copper Bellows (224C Only) Inconel Bellows
SWP psi (bar) Housing Material Inconel Bellows Top Bottom
1-5/8" (41mm) O.D. 1-5/8" (41mm) O.D. 3/4" (19mm) O.D. 3/4" (19mm) O.D. 5/8" (16mm) O.D.

500 Forged Brass 0-30" w.c. (0-75 mbar) to 0-61 psi (0-4.2 bar) to 0-400 psi (0-27.6 bar) to
(ASTM-B124#2) 0-60 psi (0-4.1 bar) 0-400 psi (0-27.6 bar) 0-500 psi (0-34 bar) 1/4" NPT 1/4" NPT
(34)

500 Cold Rolled Steel (C1018) 0-30" w.c. (0-75 mbar) to 0-61 psi (0-4.2 bar) to 0-400 psi (0-27.6 bar) to
1/4" NPT 1/4" NPT
(34) Stainless Steel (316) 0-60 psi (0-4.1 bar) 0-400 psi (0-27.6 bar) 0-500 psi (0-34 bar)
224 (Non-C)*

1,000 Copper Nickel (70-30)


(MIL-C-15726) MS-16142-4 MS-16142-4
(69)

1,500 Cold Rolled Steel (C1018)


224C

1/4" NPT 1/4" NPT


(103) Stainless Steel (316)
0-60" w.c. (0-149 mbar) to
Cold Rolled Steel (C1018) 0-60 psi (0-4.1 bar)
3,000 0-60" w.c. (0-149 mbar) to 0-61 psi (0-4.2 bar) to 0-400 psi (0-27.6 bar) to
Stainless Steel (316) 1/2" NPT 1/4" NPT
(207) 0-60 psi (0-4.1 bar) 0-400 psi (0-27.6 bar) 0-1000 psi (0-69 bar)
Monel 0-61 psi (0-4.2 bar) to
6,000 Cold Rolled Steel (C1018) 0-400 psi (0-27.6 bar)
1/2" NPT 1/4" NPT
(414) Stainless Steel (316)

Alloy Steel (4140) 0-100" w.c. (0-248 mbar) to 0-100" w.c. (0-248 mbar) to 1/4" NPT 1/4" NPT
10,000
0-60 psi (0-4.1 bar) 0-60 psi (0-4.1 bar) 9/16 AMINCO 9/16 AMINCO
(689)
(1/4" ODT) (1/4" ODT)

Net Volume L.P. Head 1.66" (27.2 cc) 1.66" (27.2 cc) 2.51" (41.1 cc) 2.51" (41.1 cc) 2.61" (42.8 cc)
cu.in. (cu. cm) H.P. Head 1.55" (25.4 cc) 1.55" (25.4 cc) 2.42" (39.7 cc) 2.42" (39.7 cc) 2.50" (40.9 cc)

Displacement cu.in. (cu. cm) for full-scale Travel 0.14" (2.3 cc) 0.14" (2.3 cc) 0.03" (0.49 cc) 0.03" (0.49 cc) 0.025" (0.41 cc)

Performance: Torque Tube Rotation = 8° ±10%; Torque Tube Material = Beryllium Copper (BeCu); Temperature Limits = -40°F/°C to +180°F (+82°C); Maximum Non-linearity = per Range;
Repeatability = 0.2% of full scale differential pressure (see DP Indicators, Switches, & Transmitters bulletin #21920 for additional information).

Notes: Zero center or split ranges available on special order (e.g., 0-60" w.c. (0-149 mbar) range may be ordered 30-0-30" w.c. (75-0-75 mbar) or (15-0-45" w.c. (37-0-112 mbar). Absolute pressure ranges available from 30" w.c. (75
mbar) to 600 psi (41.4 mbar). Other sizes and types of connections (welding stubs, MA, A.N.D., etc.) available upon request. Outline dimension drawings available upon request. Metric conversions ( ) are approximate. M224C with
NACE (MR-01-75) compliant materials available upon request. *Model 224 (Non-C)IRUVSHFL¿F1XFOHDUVHUYLFHDSSOLFDWLRQVRQO\
1-4. Theory of Operation
The M224/224C DPU measures the differential pressure (DP) in a process
system relative to process functions and produces a mechanical output that actu-
ates process monitoring instruments and process control devices.

LP Connection
HP Connection Torque Tube
Center Plate
High Pressure
Housing Low Pressure
Housing
High Pressure
Bellows
Low Pressure
Range Spring Bellows

Range Spring

Temperature
Compensator
Figure 1-1. DPU Components
The high-pressure (HP) housing is connected by pipe or tubing to the high-
pressure side of the primary device. The low-pressure (LP) housing is connected
to the low-pressure side of the primary device.
Any pressure changes within the housings causes the bellows to move in the
direction of least pressure. As the bellows move laterally, the connecting drive
arm follows the motion of the bellows, and twists the torque tube.
The torque tube shaft, which is freely supported within the torque tube at its
outer end, but attached to the torque tube and drive arm at the inner end, rotates
through the same angle as the drive arm. The rotation of the torque tube shaft
provides the mechanical motion required to actuate process instruments, such as
recorders, indicators, transmitters, controllers, and switches.
If the bellows are subjected to a pressure greater than the range of the DPU,
they move their normal travel range and a small amount of overtravel until
the overrange valve seals against the valve seat. As the valve closes, the fill
liquid is trapped in the bellows. Since the fill fluid is essentially noncompressible,
the bellows are fully supported and will not be damaged regardless of the
overpressure (up to the full rated pressure of the instrument) applied to them.
Also, since dual overrange valves are used, full protection against overrange is
provided in either direction.
An additional convolution in the high-pressure side of the bellows provides
for expansion or contraction of the fill fluid relative to ambient temperature
changes. This extra convolution acts as an accumulator permitting the fill fluid
to change volume without materially affecting the pressure within the bellows or
changing the physical relationship of the two bellows.

5
SECTION 2 - INSTALLATION

2-1. General
The instrument should be inspected at time of unpacking to detect any damage
that may have occurred during shipment. Note: The DPU was checked for
accuracy at the factory — do not change any of the settings during examina-
tion or accuracy will be affected.
For applications requiring special cleaning/precautions, a polyethylene
bag is used to protect the instrument from contamination. This bag
should be removed only under conditions of extreme cleanliness.
2-2. DPU Mounting
A. Flush or Panel
1. Cut an opening in the panel and drill four holes (17/64"), per outline
dimension drawings in Section 5.
2. Pass the instrument through the cutout in the panel (from panel front).
3. Attach the indicator to the panel with four screws.
C. Pipe
Refer to Section 5. The unit must be mounted approximately level for
proper operation.
2-3. DPU Piping

WARNING
HIGH PRESSURE HAZARD. TO PREVENT PERSONAL
INJURY OR DAMAGE TO EQUIPMENT, DIRECT ALL PIPING
AWAY FROM THE OPERATOR WHILE CONNECTING THE
DPU TO THE SYSTEM PIPING.
WARNING
EXPLOSION HAZARD. NO ORGANIC COMPOUNDS,
OIL, GREASE, DIRT, OR SCALE OF ANY KIND CAN BE
TOLERATED IN AN OXYGEN INSTALLATION.
WARNING
FOR INSTALLATIONS WHERE THE PRESSURE COULD
EXCEED THE RATED MAXIMUM SAFE WORKING
PRESSURE OF THE DPU, THE PIPING SYSTEM MUST
INCLUDE PROTECTIVE MEASURES TO PREVENT OVER
PRESSURE, IN ACCORDANCE WITH APPLICABLE LOCAL
AND NATIONAL PIPING CODES.
WARNING
THE MODEL 224/224C DPU IS NOT DESIGNED FOR
HIGH STATIC OR DYNAMIC LOADS AT IMPULSE LINE
CONNECTIONS. THE IMPULSE PIPING SYSTEM MUST BE
DESIGNED WITH ADEQUATE SUPPORTS TO MINIMIZE
THE LOADS AT THE DPU.

7
2-3. DPU Piping (Cont.)
WARNING
DO NOT REUSE HOUSING BOLTS. SEE HOUSING BOLT
WARNING ON PAGE 20.
CAUTION
DPU pressure rating has been determined using standard methods
without specific considerations for corrosion (internal or external) or
fatigue. The system designer should appropriately derate any DPU
where these considerations are significant issues. Do not subject DPU
to unnecessary shock/overrange pressure during operation.
CAUTION
DPU bellows protects torque tube from exposure to process fluid.
Cyclic conditions can cause undetected bellows failure resulting in
exposure of the torque tube to the process fluid.
NOTICE
Do not share filling or vapor return lines with DPU piping lines.
Standard Piping Practices (observe when installing piping):
1. Shorten distance between primary device and DPU as much as possible.
Distances exceeding 100 feet are not recommended. For distances up to
50 feet, use 1/4-inch pipe or tubing. For runs 50 to 100 feet, use 1/2-
inch pipe or tubing. The recommended limitation does not apply if an air
purge or blow-back system is used.
2. Slope all piping at least 1-inch/linear foot to avoid liquid/gas entrapment.
3. Provide two feet of un-insulated piping between the DPU and the primary
device for each 100 degrees in excess of +180°F (+162°C), if process
media exceeding +200°F (+93.3°C) must be measured.
4 Assure that the temperature of the DPU never exceeds 180°F (+162°C).
When steam tracing is necessary, the steam pressure should not exceed five
pounds per square inch and insulation should not be used. If pressure must
exceed five pounds per square inch, limit the length of tubing around the
DPU to two turns and do not insulate.
5. When severe pulsation is present, install a suitable pulsation dampening
device upstream of the DPU; otherwise, accuracy will be affected.
6. Mount DPU on a solid support to minimize vibration. Tighten all points,
using a suitable compound; leaks in piping can cause measurement errors.
7. Rotate the housing as necessary to place the connection in the proper posi-
tion. The DPU has connections in the pressure housings to accommodate
various pipe sizes (refer to Section 5).
8. Install a valve manifold connecting the DPU and the source of differential
pressure to facilitate operation and checking of the DPU.
9. Install all shutoff and bypass valves so they are easily accessible from front
of instrument. Locate block valves at the source of differential pressure.
10. For gas service, it is recommended that zero check be performed with both
block valves closed. If gas flow is pulsating, a standing wave effect may be
in the process line displacing the indicator (appearing to be a zero error).

8
2-4. DPU Piping/Startup Examples
"Typical" piping diagrams and startup procedures are presented on pages 9-17.
For specific application/installation/piping information, contact PRIME Factory.
Review all WARNINGS, CAUTIONS, AND NOTICES (pages 7 and 8)
before installation or startup.
• Flow Applications (pages 9-13) (HP = High Pressure)
Ensure DPU HP housing is connected to the primary device upstream tap.
NOTICE: To prevent overheating DPU during blow-down, operator should
monitor temperature by placing his hands on pipe between DPU and the
manifold pipe containing the vent valves.
• Liquid Level Applications (pages 13-16)
The process media may be used as a reference leg seal fluid when it is of
a type that will condense in the reference leg under all conditions. If the
process or process media characteristics are such that the above conditions
cannot be met, a special reference leg seal fluid will be required.
If special seal fluid media characteristics are such that the above conditions
cannot be met, a special reference leg seal fluid will be required. The spe-
cial seal fluid must not be volatile and must not be miscible with the process
media. The difference in densities (special seal fluid vs process media) will
require compensation in calculating differential pressure range of the DPU.
• Liquid Specific Gravity Applications (page 17)
This method of piping is used for determining specific gravity changes in a
process media using differential pressure instruments.
A. Gas Flow, DPU Above Run
Recommended for use whenever possible, as the DPU is self-draining.
However, NOT recommended when hydrates are present.
1. Open bypass valve(s) and close vent valve.
2. Open both shutoff valves
and one block valve to
pressurize DPU, then close
block valve.
3. Close one bypass valve and
check system for leaks.
If output travels upscale,
check for low-pressure
piping leaks. If output
travels downscale, check for
high-pressure piping leaks.
Figure 2-1. Gas Flow,
4. Repair piping if necessary and repeat steps DPU Above Run
1 through 3 until output remains stationary
at zero.
5. Close both shutoff valves and open bypass valve(s).
6. Open both block valves and slowly open both shutoff valves.
7. Close bypass valves and if two bypass valves are used, open vent valve.
8. Close bypass valves, and if two bypass valves are used, open vent
valve.

9
2-4. DPU Piping/Startup Examples (Continued)
Review all WARNINGS, CAUTIONS, AND NOTICES (pages 7 and 8)
before installation or startup.
B. Gas Flow, DPU Below Run
Used only when necessary to mount the DPU below run. Drip pots are
required when wet gas is present.
1. Open bypass valve(s) and
close vent valve.
2. Open both shutoff valves
and one block valve to
pressurize DPU, then close
block valve.
3. Close one bypass valve
and check system for
leaks. If output travels
Figure 2-2. Gas Flow,
upscale, check for low-
DPU Below Run
pressure piping leaks. If
output travels downscale, check for high-pressure piping leaks.
4. Repair piping if necessary and repeat steps 1 through 3 until output
remains stationary at zero.
5. Close both shutoff valves and open bypass valve(s).
6. Open both block valves and slowly open both shutoff valves.
7. Close bypass valves, and if two bypass valves are used, open vent
valve.
8. If drip pots are used, open drip valves and blow out accumulated liquid.
C. Gas Flow, Hydrates Present
The following is used if hydrates or heavy solids are present, piping and
shutoff not less than 1/2-inch in diameter. Bypass the manifold above to
isolate the meter from con-
necting piping. Drip pots
prevent plugging.
1. Open bypass valve(s) and
close vent valve.
2. Open both shutoff valves
and one block valve to
pressurize the DPU, then
close block valve.
3. Close one bypass valve
and check system for
leaks. If output travels
upscale, check for low-pressure piping leaks. Figure 2-3. Gas Flow,
If output travels downscale, check for high- Hydrates Present
pressure piping leaks.
4. Repair piping if necessary and repeat steps 1 through 3 until output
remains stationary at zero.
5. Close bypass valves. If two bypass valves are used, open vent valve.
6. Drain drip pots of hydrates at regular intervals.
10
2-4. DPU Piping/Startup Examples (Continued)
Review all WARNINGS, CAUTIONS, AND NOTICES (pages 7 and 8)
before installation or startup.
D. Steam Flow, DPU Below Run
For this application, condensing reservoirs and piping to orifice taps must
be level. Assure that the reservoir and steam lines are at the same level.
Two-inch pipe crosses may be used as seal pots.
1. Close vent valves if used, open
bypass and shutoff valves.
2. Remove condensing reservoir side
and fill plugs.
3. Pour water into both reservoirs
until piping and DPU housings
are filled. Piping and housing
chambers shall be free of bubbles.
The pointer will rest at zero (or
output will be 4 mA or 10 mA as
applicable) when the instrument
and piping are completely filled.
4. Install side and fill plugs in reservoirs.
5. Close shutoff valves and open block valves.
6. Slowly open both shutoff valves simultaneously and check for leaks.
7. Close bypass valve.
NOTICE: Ensure that plugs are used and never valves on the DPU.
E. Liquid Flow, DPU Above Run
The following steps should be used if sediments may be present. Inspect
piping periodically. Not recommended for hot or gassy liquids.
1. Close both shutoff valves and
open both block valves.
2. Open bypass valve. Crack
vent valves or loosen plugs
from top ports of DPU body
housings.
3. Crack and close shutoff valves
alternately until liquid is free
of bubbles and spills out of
both upper DPU body ports.
4. Close vent valves or tighten
plugs. Close block valves and
Figure 2-5. Steam
open shutoff valves.
Flow, DPU Above Run
5. The pointer will rest at zero (or output will
be 4 mA or 10 mA as applicable). If not and no leaks are detected,
the housing and/or piping are not completely filled with liquid. Repeat
steps 1 through 4 until output remains stationary at the lowest value.
6. Slowly open both block valves and close bypass valve.

11
2-4. DPU Piping/Startup Examples (Continued)
Review all WARNINGS, CAUTIONS, AND NOTICES (pages 7 and 8)
before installation or startup.
F. Liquid Flow, DPU Below Run
The following steps are recommended for hot or gassy liquids. Periodic
inspections of piping are recommended.
1. Close both shutoff valves
and open both block
valves.
2. Open bypass valve. Crack
vent valves or loosen
plugs from top ports of
DPU pressure housings.
3. Crack and close shutoff
valves alternately until the
liquid is free of bubbles
and spills out of both
upper DPU body ports. Figure 2-6. Liq.Flow,
DPU Below Run
4. Close vent valves or tighten plugs. Close block
valves and open shutoff valves.
5. The pointer will rest at zero (or output will be 4 mA or 10 mA as appli-
cable). If not and no leaks are detected, the housing and/or piping are
not completely filled with liquid. Repeat steps 1 through 4 until output
remains stationary at the lowest value.
6. Slowly open both block valves and close bypass valve.
7. For service with hot or gassy liquids, fill both sides of the manifold
through the fill tee, with the liquid to be measured cooled to +200°F
(+93.3°C) or less, expel gas bubbles from DPU and piping.
8. Open vent valve in the bypass valve. Tighten the fill plug when bubble
free liquid flows.
G. Corrosive Liquid Flow
1. Close shutoff valves and open
block valves.
2. Open bypass valve and close
DPU drain plugs.
3. Remove fill and side plugs
from seal pots.
4. Fill seal pots, piping and DPU
housing with immiscible seal
fluid by pouring into upper fill
ports. DPU housing tubing,
and scale pots must be filled to seal pot side Figure 2-7. Corrosive
ports with bubble-free liquid. The pointer will Liquid Flow
indicate zero (or 4 mA or 10 mA as appli-
cable) when high and low pressure chambers are filled with liquid.
5. Install side plugs.
(continued on next page...)

12
2-4. DPU Piping/Startup Examples (Continued)
Review all WARNINGS, CAUTIONS, AND NOTICES (pages 7 and 8)
before installation or startup.
G. Corrosive Liquid Flow (Continued)
6. Slowly open each shutoff valve alternately, until bubble-free line liquid
spills from both upper fill plugs.
7. Replace fill plugs.
8. Open and close block valves. Check for piping leaks (repair as needed).
9. Close shutoff valves and open block valves.
10. Slowly open both shutoff valves and close bypass valve.
H. Cool Non-Condensing Liquid, DPU Level with Tank Bottom
High-pressure side of DPU is connected to bottom outlet of tank. The low-
pressure side is connected to top or gas outlet of tank.
NOTICE: Do not share filling
or vapor return lines with DPU Figure 2-8. Cool
piping. Non-Condensing Liq.
The following steps are suitable
for piping layout for water, oil, or
other media that will not con-
dense in low-pressure piping.
1 . Close both shutoff valves,
open lower block valve and
crack vent valve.
2. Slowly open lower shutoff
valve. When bubble-free
liquid spills from vent, close
vent valve.
3. Open upper block valve and
slowly open shutoff valve.
4. Crack drain valve to remove any condensation and close drain valve.
I. Cool Non-Condensing Liquid, DPU Below Tank
High-pressure side of DPU is connected to the bottom outlet of tank. The
low-pressure side is connected to top or gas outlet of tank. The following
steps are suitable for piping Figure 2-9. Cool
layout for water, oil, or other Non-Condensing Liq.
media that will not condense in (DPU Below Tank)
low-pressure piping.
1 . Close both shutoff valves,
open lower block valve and
crack vent valve.
2. Slowly open lower shutoff
valve. When bubble-free
liquid spills from vent,
close vent valve.
(continued on next page...)

13
2-4. DPU Piping/Startup Examples (Continued)
Review all WARNINGS, CAUTIONS, AND NOTICES (pages 7 and 8)
before installation or startup.
I. Cool Non-Condensing Liquid, DPU Below Tank (Continued)
3. Open upper block valve and slowly open shutoff valve.
4. Crack drain valve to remove any condensation and close drain valve.
NOTICE: Do not share filling or vapor return lines with DPU piping.
J. DPU Below Tank with Reference Leg
Cool Liquids:
Use of reference leg cancels out the "dead leg" (piping from the tank
bottom to center line of meter body). Seal fluid in reference leg must not
volatilize. Process media can be used as a reference leg seal fluid if it is
such that it will condense in reference leg under all conditions. Otherwise,
special, immiscible seal fluid must be used. Difference in densities between
process media and seal fluid must be considered when computing differen-
tial pressure range of DPU.
1. Partially fill reference leg by opening bottom block valve, both shutoff
valves, and bypass valve.
2. Crack drain plugs on the
DPU housing and vent the
valve. Close when clear,
bubble-free liquid flows.
3. Close bypass and shutoff
valve on reference leg.
4. Remove plug from the top
port in 2-inch pipe cross
connection, and fill the
reference leg manually.
5. Open top shutoff valve
and crack vent valve until
bubbles are expelled.
Leave the reference leg
full.
6. Replace plug in pipe cross
and close vent valve.
Figure 2-10. DPU
7. Slowly open upper block valve.
Below Tank w/Ref. Leg
Hot Liquids (Volatile):
1. Open both shutoff valves, bypass valve and vent valve.
2. Remove plug from the top port in 2-inch pipe cross. Fill both high and
low DPU housings with liquid until it runs out of vent valve. Use liquid
to be measured, cooled below +200°F (+93.3°C) or other suitable seal
liquid. Expel bubbles in DPU housing.
(continued on next page...)

14
2-4. DPU Piping/Startup Examples (Continued)
Review all WARNINGS, CAUTIONS, AND NOTICES (pages 7 and 8)
before installation or startup.
J. DPU Below Tank with Reference Leg (Continued)
Hot Liquids (Volatile): (Continued)
3. Close bypass and shutoff valve on reference leg.
4. Remove plug from top port in 2-inch pipe cross connection, and fill
reference leg manually.
5. Open top shutoff valve and crack vent valve until bubbles are expelled.
Leave reference leg full.
6. Replace plug in pipe cross and close vent valve.
7. Slowly open upper block valve.
NOTICE: Do not share filling or vapor return lines with DPU piping.
If bypass valve is opened at any time, the reference leg must be filled again.
If no bypass is installed, disregard steps 1, 2, and 3. Open the lower block
valve in step 6 and shut off valve in step 7.
K. Liquid CO2
The instrument may be located above or below the vessel. The recom-
mended vapor generator is a 12-inch length of 1 to 1-1/2-inch diameter
pipe. Avoid traps or pockets between the vapor generator and the tank.
Install the inverted "U" shaped gas trap inside the vessel.
1. Close shutoff valves.
2. Open drain valve and DPU housing drain
plugs to remove all liquid from the system,
and close drains.
3. Open bottom block valve slowly
to allow liquid to enter gas gen-
erator. Assure that vent valve is
closed.
4. Open top block valve and
shutoff valves.
NOTICE: Do not insu-
late the piping below
the lower block valve. Do not share
filling and vapor return lines with
the DPU piping lines.
Figure 2-11. Liq. CO2
L. Cryogenic Liquids
Instrument may be located above or below the vessel. The recommended
vapor generator is a spiral of 3/8-inch tubing. Install the inverted "U"
shaped gas trap inside vessel. Meters for oxygen service are specially
cleaned and packaged in polyethylene bags, and must be kept extremely
clean.
1. Close shutoff valves.
2. Open drain valve and DPU housing drain plugs to remove all liquid
from system, and close drains.
(continued on next page...)

15
2-4. DPU Piping/Startup Examples (Continued)
Review all WARNINGS, CAUTIONS, AND NOTICES (pages 7 and 8)
before installation or startup.
L. Cryogenic Liquids (Continued)
3. Open bottom block
valve slowly to
allow liquid to enter
gas generator.
Assure that vent
valves are closed.
4. Open top block
valve and open
shutoff valves.

Figure 2-12. Cryo Liq.


NOTICE: Do not share filling or vapor return
lines with DPU piping.

M. Bubbler System
Recommended when piping connections must be made at the top of the
tank or whenever solids or sludge are present.
1 . Set bubbler input gas regulator pressure slightly higher than process
vessel pressure.
2. Open block valves and
shutoff valves, and close
bypass valve.
3. Check sight-flow into bub-
bler tube.
4. Adjust bubbler system to
minimum flow by adjust-
ing input gas regulator.

NOTICE: Do not share filling or vapor return lines with DPU piping.

16
2-4. DPU Piping/Startup Examples (Continued)
Review all WARNINGS, CAUTIONS, AND NOTICES (pages 7 and 8)
before installation or startup.
N. Liquid Specific Gravity
This method is for determining specific gravity changes in a process media
using a differential pressure instrument.
1 . Set bubbler input gas regulator pressure slightly higher than process
vessel pressure.
2. Open block valves and
shutoff valves, and close
bypass valve.
3. Check sight-flow into bub-
bler tube.
4. Adjust bubbler system
to minimum flow by
adjusting digit flow gages
throttling valves.

Figure 2-14. Liq. Specifc Gravity

17
SECTION 3 - MAINTENANCE

3-1. Required Tools


Tool Purpose
Screwdriver Bracket Screws
50 lb. Torque wrench Pressure housing bolts
3-2. Test/Calibration Equipment

Figure 3-1. Test/Calibration Setup


3-3. Preventative Maintenance
• DPU Piping — Periodically inspect the integrity of the DPU piping. Tighten
all pipe joints as necessary.
• DPU Inspection and Cleaning — See DPU Cleaning and Inspection Proce-
dure that follows.
3-4. Calibration
For calibration procedures, refer to separate actuated instrument manual.

3-5. DPU Cleaning and Inspection


WARNING
(High Pressure Gas Installations
with pressures greater than 200 psig)
HIGH PRESSURE GAS HAZARD ON DISASSEMBLY OF DPU. TO
PREVENT POSSIBLE SEVERE PERSONAL INJURY, DEATH, OR
SUBSTANTIAL PROPERTY DAMAGE DUE TO THE RELEASE OF
INTERNAL PRESSURE, PERFORM THE PRESSURE CHECK
PROCEDURE THAT FOLLOWS (Step A) BEFORE REMOVING
THE DPU HOUSING BOLTS.

WARNING
THE DPU MAY BE UNDER PRESSURE. ENSURE THAT
THE PIPING SYSTEM IS COMPLETELY DEPRESSURIZED
BEFORE REMOVING THE METER FOR MAINTENANCE OR
INSPECTION.
(Continued on next page...)
19
3-5. DPU Cleaning and Inspection (Continued)
(See WARNINGS on page 19 before proceeding.)

WARNING
(Housing Bolts)
DO NOT REUSE HOUSING BOLTS. IF BOLTS ARE
DISTURBED, REPLACE WITH NEW BOLTS, PER TABLE 3-1.
REUSE OF HOUSING BOLTS, ESPECIALLY IN CRITICAL
APPLICATIONS LIKE HYDROGEN SULFIDE AND SALT WATER
EXPOSURES, CAN RESULT IN SEVERE INJURY, DEATH
OR SUBSTANTIAL PROPERTY DAMAGE DUE TO BOLT
FAILURE.
NOTICE
If accumulation of solids or semi-solids is extensive, remove the
housings carefully to prevent damage to the bellows.

A. Pressure Check
WARNING
FAILURE TO PERFORM THIS PROCEDURE CAN RESULT IN
SEVERE INJURY, DEATH OR SUBSTANTIAL PROPERTY
DAMAGE DUE TO THE RELEASE OF INTERNAL PRESSURE.
This procedure should be performed prior to removing the DPU housing
bolts, especially if the DPU has been installed in gas applications with
working pressures greater than 200 psig.
1. Back off all housing bolts 4 turns.
2. Check for internal pressure by attempting to move the housing in and
out along the bolts.
a. If the housing moves freely — no pressure is present — servicing
or repair may continue.
b. If the housing does not move freely — the bellows may
be pressurized and is POTENTIALLY HAZARDOUS if further
disassembled.
Tighten the bolts and return the unit to the factory or
authorized service center for repair.
Tag the unit and specify “Gas in Bellows”.
B. Cleaning/Inspection Procedure
Instruments used in services where solids or semi-solids may accumulate
inside the pressure housings require periodic inspection and cleaning, as
follows.
1. Remove the DPU from service and remove the pressure housings.
2. Carefully remove the pressure housings from the bellows unit assembly.

NOTICE
If the accumulation of material inside the housing is extensive, rapid
removal of the housing may damage the bellows convolutions.

20
3-5. DPU Cleaning and Inspection (Continued)
(See WARNINGS on pages 19 and 20 before proceeding.)
B. Cleaning/Inspection Procedure (Continued)
3. Remove the accumulation from between the bellows convolutions and
from the inside of the housings. Use a solvent if possible. Do not use
a sharp instrument to clean between convolutions.
4. Assure that there are no broken range springs.
5. Replace the housings and O-rings.
6. Review Housing Bolt WARNING (page 20) before proceeding.
See Table 3-1 (below and page 22) for pressure housing bolt size,
material, part number, and torque values.

Table 3-1. Bolt Torque Ratings


HOUSING BOLT TORQUE ROTATION
LUBE
LB-FT (Degrees)
SWP MATL. MATL. SIZE PART NO. (Note 1)
(Note 2) (Note 3)

ALUM
400 STEEL 1/4-28 x 2 0220-0035J NO 8-10 135
(Note 6)

ALUM
400 18-8 1/4-28 x 2 0220-1016J NO 8-10 135
(Note 6)

500 BRASS STEEL 1/4-28 x 2 0220-0035J NO 8-10 135

18-8
500 BRASS 1/4-28 x 2 0220-1010J YES 10-12 450
(Note 5)

500 BRASS K-MONEL 1/4-28 x 2 0220-1018J YES 8-10 180

500 STEEL STEEL 1/4-28 x 2.5 0220-0029J NO 12-14 180

500 SST 17-4 PH 1/4-28 x 2.5 0220-1017J YES 12-14 180

A286
500 SST (Salt Spray & 1/4-28 x 2.5 0220-1112J YES 16-18 —
NACE)

STEEL
500 SST 1/4-28 X 1/2 S408-0067Z NO 12 —
(NACE)

500 Cu-Ni K-MONEL 1/4-28 x 2.5 0220-0045J YES 9-11 180

1,000 Cu-Ni K-MONEL 1/4-28 x 2.5 0220-0045J YES 9-11 180

1,500 STEEL STEEL 1/4-28 x 2.5 0220-0029J NO 12-14 180

1,500 STEEL K-MONEL 1/4-28 x 2.5 0220-0045J YES 12-14 180

1,500 SST STEEL 1/4-28 x 2.5 0220-0029J NO 12-14 180

1,500 SST 17-4 PH 1/4-28 x 2.5 0220-1017J YES 12-14 180

A286
1,500 SST (Salt Spray & 1/4-28 x 2.5 0220-1112J YES 16-18 —
NACE)

(table continued on next page . . .)

Notes: (1) /XEULFDQW/LJKWO\DSSO\0RO\NRWH*QSDVWH 8 RQWKH¿UVWWZRWKUHDGVRQO\XQOHVVRWKHU-


ZLVHVSHFL¿HGGRQRWOXEHEHDULQJVXUIDFH6SHFLDODSSOLFDWLRQVVXFKDV1XFOHDU&U\RJHQLFDQGR[\JHQVHUYLFHV
PD\UHTXLUHVSHFLDOOXEULFDQWV&KHFN%20IRUVSHFL¿FVVXFKDVWKUHDGKHDGDQGJDVNHWVXUIDFHV (2) Torque
on bolts is accomplished in 3 or 4 steps. Tighten uniformly. (3) PRIME recommends using a torque wrench
whenever installing housing bolts. Rotation of bolt head is measured after bolt is "snug" (approx. 1/16 turn
past head contact) with approximately 2 LB-FT torque. Do not exceed this rotation. To tighten bolts without torque
wrench, use rotation values. (4) Lube threads per Note 1 and under the head of the MP35N bolts. (5) Denotes for
commercial applications only. (6) For reference only - 400 psi ALUM housing is obsolete.

21
3-5. DPU Cleaning and Inspection (Continued)
(See WARNINGS on pages 19 and 20 before proceeding.)
B. Cleaning/Inspection Procedure (Continued)
6. (Continued)
Review Housing Bolt WARNING (page 20) before proceeding.
Table 3-1. Bolt Torque Ratings (Continued)
HOUSING BOLT TORQUE ROTATION
LUBE
LB-FT (Degrees)
SWP MATL. MATL. SIZE PART NO. (Note 1)
(Note 2) (Note 3)

3,000 STEEL STEEL 3/8-24 x 2.5 0220-0022J NO 30 90

3,000 SST 17-4 PH 3/8-24 x 1.5 0220-1028J YES 40 90

3,000 SST 17-4 PH 3/8-24 x 3.0 0220-1015J YES 35 90

A286
3,000 SST (Salt Spray & 3/8-24 x 2.5 0220-1111J YES 34-36 —
NACE)

STEEL
3,000 SST 3/8-24 x 2-1/2 S408-0065Z NO 35 —
(NACE)

3,000 SST 17-4 PH 3/8-24 x 2.5 0220-1014J YES 35 90

3,000 MONEL 17-4 PH 3/8-24 x 2.5 0220-1014J YES 45 180

A286
3,000 MONEL (Salt Spray & 3/8-24 x2.5 0220-1111J YES 19-21 —
NACE)

6,000 STEEL STEEL 3/8-24 x 2.5 0220-0022J NO 35 135

6,000 SST 17-4 PH 3/8-24 x 2.5 0220-1014J YES 40 135

MP35N
6,000 SST 3/8-24 x 2.5 0220-1109J (Note 4) 50-52 —
(Salt Spray)

10,000 STEEL STEEL 3/8-24 x 2.5 0220-0022J YES 40 180

10,000 SST 17-4 PH 3/8-24 x 2.5 0220-1014J YES 45 180

10,000 K-MONEL 17-4 PH 3/8-24 x 2.5 0220-1014J YES 45 180

10,000 K-MONEL STEEL 1/2-13 x 3.5 S408-0113C NO 40 180

Notes: (1) /XEULFDQW/LJKWO\DSSO\0RO\NRWH*QSDVWH 8 RQWKH¿UVWWZRWKUHDGVRQO\XQOHVVRWKHU-


ZLVHVSHFL¿HGGRQRWOXEHEHDULQJVXUIDFH6SHFLDODSSOLFDWLRQVVXFKDV1XFOHDU&U\RJHQLFDQGR[\JHQVHUYLFHV
PD\UHTXLUHVSHFLDOOXEULFDQWV&KHFN%20IRUVSHFL¿FVVXFKDVWKUHDGKHDGDQGJDVNHWVXUIDFHV (2) Torque
on bolts is accomplished in 3 or 4 steps. Tighten uniformly. (3) PRIME recommends using a torque wrench
whenever installing housing bolts. Rotation of bolt head is measured after bolt is "snug" (approx. 1/16 turn
past head contact) with approximately 2 LB-FT torque. Do not exceed this rotation. To tighten bolts without torque
wrench, use rotation values. (4) Lube threads per Note 1 and under the head of the MP35N bolts. (5) Denotes for
commercial applications only. (6) For reference only - 400 psi ALUM housing is obsolete.

22
3-6. Servicing
Range Change
To change the range of the DPU, the bellows unit assembly (BUA) must be
replaced with a unit of the desired range.
BUA Replacement
WARNING
REVIEW ALL WARNINGS, CAUTIONS, AND NOTICES UNDER DPU
CLEANING AND INSPECTION (PAGES 19 AND 20) BEFORE PER-
FORMING THIS PROCEDURE.
NOTICE
Do not loosen drive-arm hold plug (located on the top of the BUA center plate)
when removing the mounting bracket. If the plug is loosened, the fill fluid in the
bellows will be lost and the unit will be rendered inoperable.

1. Disconnect the DPU piping, remove the instrument from service, and
remove the mounting bracket from the DPU.
2. Loosen the actuated instrument drive arm and separate it from the torque
tube shaft. Do not disconnect the actuated instrument linkage.
NOTICE
See the actuated instrument's manual for particular components that
must be removed to gain access to the DPU mounting fasteners.

3. Perform Pressure Check Procedure (A) under DPU Cleaning and Inspec-
tion on page 20. If internal pressure was found, do not continue
this procedure.
4. Remove pressure housing bolts and pressure housings.

NOTICE
It is recommended that new O-Rings be used whenever pressure hous-
ings are replaced.

5. Install the pressure housings onto the new BUA and attach with
appropriate housing bolts (see Table 3-1 for torque requirements).
6. Re-attach DPU (See para. 3-7, page 24 for procedure) to the actuated
instrument.
NOTICE
(Drive Arm/Torque Tube Tightness Test)
For nuclear seismic/high impact qualified units, a tightness test must
be performed whenever a DPU is attached to an instrument. See para.
3-8., page 24 for specific instructions.

7. Install the assembled instrument into service (in reverse order of removal)
and calibrate, per separate actuated instrument manual.

23
3-7. Attaching Drive Arm to Torque Tube
NOTICE
See separate actuated instrument manual for instrument specific details.
Drive Arm Assembly Procedure
LINK SINTERED
CLAMP BLO CK
.03 (approx)
DRIVE
ARM
M224/M224C

LINK SLO TTED


CLAMP BLO CK
DRIVE CLAMP
3/32" ALLEN SHAFT SCREW
ARM
W RENCH 1/8" W RENCH 0163-0044C

TO TIGHTEN

SHAFT CLAMP
SCREW TO TIGHTEN

Figure 3-2. Attaching Drive Arm to Torque Tube


1. Slip drive arm over torque tube shaft; clear end of torque-tube housing by
approximately 0.030-inches before securing to prevent interference.
2. To tighten drive arm assembly onto torque-tube shaft:
a. Support block/shaft and tighten clamp screw until snug to shaft.
b. Still supporting block/shaft, tighten clamp screw an additional:
• Sintered: 1/3 to 1/2 turn (This screw can normally turn one full
revolution before breaking.)
• Slotted: 1/4 to 1/3 turn (The slot in the slotted clamp block
should still be open.)
3-8. Drive Arm Tightness Test Procedure
(Required for Nuclear Seismic/High Impact Shock Qualified Units)
This procedure tests drive arm to torque-tube attachment tightness by apply-
ing torque developed by DPU onto a fixed drive arm. Care should be taken
to apply pressure slowly as the torque is being applied to the connection
through the torque-tube drive shaft and not the torque-tube itself.
With pointer at normal 0% torque-tube rotation position (max. minimum scale
position or 0% on a normal 0 to 100% scale unit), adjust drive arm stop bracket
(or use alternate means) to prevent pointer from moving (stop bracket interferes
with drive arm movement). Note: On reverse acting and split range units, the
DPU must be pressurized to move pointer to max. minimum scale position and
on suppressed units it will be necessary to apply pressure to establish a reference
point to check for “zero” shift.
Pressurize DPU as required to full calibrated scale differential pressure (100% of
the full scale range). This achieves 8-degree of torque-tube drive shaft equivalent
torque onto the connection.
Observe shift in unit “zero” following DPU depressurization (as required) and
drive arm stop bracket readjusting (to allow free movement of the drive arm and
pointer). A downscale (counter-clockwise) shift in “zero” of greater than 1/2% is
indicative of drive arm slippage necessitating further clamp block tightening.

24
SECTION 4 - TROUBLESHOOTING

Table 4-1. Troubleshooting

Note: See actuated instrument manual for additional troubleshooting informa-


tion.

25
SECTION 5 - DIMENSIONS

Figure 5-1. 224/224C Dimensional Drawing (Part 1 of 2)


(based on 224C-12303, Rev. 00)
26
Figure 5-1. 224/224C Dimensional Drawing (Continued) (Part 2 of 2)
27

(based on 224C-12303, Rev. 00)


SECTION 6 - PARTS DRAWING/LIST
Figure 6-1. 224/224C DPU Parts Drawing
28
224/224C DPU Parts List (See Notes and NOTICE, page 33)
Table 6-1. 224/224C DPU Parts List
Item Description Part Number Qty.
1 Bellows Unit Assembly, 3/4" or 5/8" Specify -
2 Bellows Unit Assembly, 1-5/8" Specify -
3 Spacer, Bracket 0224-0302C 1
4 O-Ring, Housing Gasket 2
Buna-N 0001-0028R
Viton 0001-0039R
Viton (NACE) 0001-1164R
Viton (10,000 psi units only) 0001-0062R
EPT 0001-1054R
5 Bracket, Mounting (Not for use with Brass Housings) 0273-0001C 2
6 Nut, Hex, Steel, 5/16-18 CP 0500-0033J 4
7 Washer, Split Lock, 5/16" Steel 0003-0006K 4
8 2" Pipe Mounting Kit (w/U-bolts, Nuts, & Washer) 0440-0001J 2
9 Not Used
10 Screw, Hex Hd., 1/4-28 x 1/2" 0116-1011J 8
11 Screw, Rd. Hd., Steel CP, 10-32 x 3/8" 0111-0057J 4
12 Bracket, Flush (500 psi Forged Brass only) 0224-0235B 1
13 Screw, Fil. Hd., SST, 10-32 x 1/2" 0114-0031J 4
14 Washer, Split Lock, #10, SST 0003-0032K 4
15 Bracket, Universal Mounting (Wall or Yoke Mtg.) 0273-0003C 1
16 Nut, Hex, Steel CP, 1/4-28 0500-0011J 4
17 Washer, Lock, Ext. Tooth, SST, 1/4" 0003-0068K 12
18 Screw, Fl. Hd., Steel CP, 1/4-28 x 1/2" 0112-0018J 4
19 Housing, Pressure (See Note 5) 2
224C DPU
Brass, Forged, 500 psi, 1/4 NPT Ports 224C-1016C
Brass, Forged 500 psi, 1/8 NPT Ports 224C-1018C
Steel, 500/1500 psi, 1/4 NPT Ports 224C-1113C
Steel, 500/1500 psi, 1/4 X 1/2 NPT Ports 224C-1123C
Steel, 500/1500 psi, 1/2 NPT Ports 224C-1130C
Steel, 3000/6000 psi, 1/2 X 1/4 NPT Ports 224C-1108C
Steel, 3000/6000 psi, 1/4 NPT Ports 224C-1120C
Steel, 3000/6000 psi, 1/2 NPT Ports 224C-1128C

29
224/224C DPU Parts List (See Notes and NOTICE, page 33) (Cont.)
Table 6-1. 224/224C DPU Parts List (Continued)

Item Description Part Number Qty.


19 Housing, Pressure (See Note 5) (Continued) 2
224C DPU (Continued)
Steel, 10,000 psi, 1/4 NPT Ports 224C-1049C
Steel, 10,000 psi, 9/16 AMINCO Ports 224C-1189C
Cu-Ni, 1000 psi, 1/4 NPT Ports 224C-1021C
Cu-Ni, 1000 psi, 1/2 NPT Ports 224C-1022C
Cu-Ni, 1000 psi, 1/8 NPT Ports 224C-1292C
SST, 500/1500 psi, 1/2 NPT Ports 224C-1032C
SST, 500/1500 psi, 1/4 NPT Ports 224C-1033C
SST, 500/1500 psi, 3/8 NPT Ports 224C-1036C
SST, 500/1500 psi, 1/8 NPT Ports 224C-1238C
SST, 500/1500 psi, 1/4 NPT Ports (NACE) 0224-1621C
SST, 500/1500 psi, 1/4 X 1/2 NPT Ports (NACE) 0224-1626C
SST, 500/1500 psi, 1/2 NPT Ports (NACE) 0224-1697C
SST, 3000/6000 psi, 1/4 NPT Ports 224C-1040C
SST, 3000/6000 psi, 1/2 NPT Ports 224C-1041C
SST, 3000/6000 psi, 1/4 X 1/2 NPT Ports 224C-1042C
SST, 3000/6000 psi, 9/16 AMINCO Ports 224C-1187C
SST, 3000/6000 psi, Seal Welded, 3/4 0224-1641C
SST, 3000/6000 psi, 1/4 X 1/2 NPT Ports 0224-1637C
(NACE)
Monel, 3000 psi, 1/4 X 1/2 NPT Ports 224C-1028C
Monel, 3000 psi, 1/2 NPT Ports 224C-1029C
Monel, 3000 psi, 1/4 NPT Ports 224C-1121C
Monel, 3000 psi, 1/4 X 1/2 NPT Ports (NACE) 0224-1685C
K-Monel, 10,000 psi, 9/16 AMINCO Ports 224C-1215C
224 (Non-C Version) DPU (See Note 3)
Brass, Forged, 500 psi, Absolute (No Ports) 0224-1602C
Brass, Forged, 500 psi, 1/4 NPT Ports 0224-1532C
Brass, Forged, 500 psi, 1/8 NPT Ports 0224-1606C
Steel, 500/1500 psi, 1/4 NPT Ports 0224-1576C

30
224/224C DPU Parts List (See Notes and NOTICE, page 33) (Cont.)
Table 6-1. 224/224C DPU Parts List (Continued)

Item Description Part Number Qty.


19 Housing, Pressure (See Note 5) (Continued) 2
224 (Non-C Version) DPU (See Note 3) (Continued)
Steel, 500/1500 psi, 1/2 NPT Ports 0224-1577C
Steel, 500/1500 psi, Absolute (No Ports) 0224-1664C
Steel, 500/1500 psi, 3/8 NPT Ports 0224-1667C
Steel, 500/1500 psi, 1/4 X 1/2 NPT Ports 0224-1613C
Steel, 3000/6000 psi, Absolute (No Ports) 0224-1715C
Steel, 3000/6000 psi, 1/4 NPT Ports 0224-1555C
Steel, 3000/6000 psi, 1/2 NPT Ports 0224-1578C
Steel, 3000/6000 psi, 1/4 X 1/2 NPT Ports 0224-1607C
Steel, 10,000 psi, 1/4 NPT Ports 0224-1662C
Steel, 10,000 psi, 9/16 AMINCO Ports 0224-1663C
SST, 500/1500 psi, 1/4 NPT Ports 0224-1543C
SST, 500/1500 psi, 1/2 NPT Ports 0224-1579C
SST, 500/1500 psi, 1/4 X 1/2 NPT Ports 0224-1610C
SST, 500/1500 psi, Absolute (No Ports) 0224-1618C
SST, 500/1500 psi, 1/4 NPT Ports, Nipple Mod. 0224-1619C
SST, 500/1500 psi, 1/8 NPT Ports 0224-1675C
SST, 500/1500 psi, Seal Welded, 1/2 NPT Ports 0224-1679C
SST, 500/1500 psi, 3/8 NPT Ports 0224-1721C
SST, 500/1500 psi, 1/2 NPT Ports (NACE) 0224-1697C
SST, 500/1500 psi, 1/4 NPT Ports (NACE) 0224-1621C
SST, 500/1500 psi, Seal Welded, 1/4 NPT Ports 0224-1622C
SST, 500/1500 psi, 1/4 X 1/2 NPT Ports (NACE) 0224-1626C
SST, 3000/6000 psi, 1/4 NPT Ports 0224-1556C
SST, 3000/6000 psi, 1/2 NPT Ports 0224-1580C
SST, 3000/6000 psi, 1/4 X 1/2 NPT Ports 0224-1608C
SST, 3000/6000 psi, Seal Welded, 3/4 0224-1641C
SST, 3000/6000 psi, 9/16 AMINCO Ports 0224-1645C
SST, 3000/6000 psi, 1/4 X 1/2 Ports, Nipple 0224-1647C
Mod.

31
224/224C DPU Parts List (See Notes and NOTICE, page 33) (Cont.)
Table 6-1. 224/224C DPU Parts List (Continued)
Item Description Part Number Qty.
19 Housing, Pressure (See Note 5) (Continued) 2
224 (Non-C Version) DPU (See Note 3) (Continued)
SST, 3000/6000 psi, 1/4 X 1/2 NPT Ports 0224-1637C
(NACE)
Cu-Ni, 1000 psi, 1/4 NPT Ports 0224-1581C
Cu-Ni, 1000 psi, 1/4 X 1/2 NPT Ports 0224-1616C
Cu-Ni, 1000 psi, Absolute (No Ports) 0224-1651C
Cu-Ni, 1000 psi, 1/2 NPT Ports 0224-1655C
Monel, 3000 psi, 1/4 X 1/2 NPT Ports (NACE) 0224-1685C
Monel, 3000 psi, 1/4 X 1/2 NPT Ports 0224-1687C
Monel, 3000 psi, 1/2 NPT Ports 0224-1696C
20 Spacer, DPU Mounting 0224-1547C 1
21 Bolt, Housing Refer to Table 4
(Refer to Housing Bolt WARINING on page 20.) 3-1, pg. 21/22
for Part No.
22 Plug, Pipe/Port 2
Pipe Plug, 1/4" NPT Brass 0224-0100C
Pipe Plug, 1/8" NPT, Brass 0224-0058C
Pipe Plug, 1/8" NPT, SST 0224-1676C
Pipe Plug, 1/4" NPT, Steel CP 0199-0191C
Pipe Plug, 1/4" NPT, Monel 0199-0235C
Pipe Plug, 1/4" NPT, SST 0199-0214C
Pipe Plug, 1/2" NPT, Steel CP 0199-0192C
Pipe Plug, 1/2" NPT, SST 0199-0215C
Pipe Plug, 1/2" NPT, Monel 0199-0234C
Pipe Plug, 3/8" NPT, SST 0605-0010L
23 Screw, 1/4-28 x 5/8", Steel, CP 0111-0033J 4
24 Washer, Lock, Ext. Tooth, SST, 1/4" 0003-0068K 4
25 Bracket, Mounting (Not for use with Brass Housings) 0224-0055C 2

32
224/224C DPU Parts List (See Notes and NOTICE, below) (Cont.)
Table 6-1. 224/224C DPU Parts List (Continued)
Item Description Part Number Qty.
26 Nut, Hex, 1/4-28, Steel, CP 0500-0011J 4
27 Washer, Ext. Tooth, 1/4" SST 0003-0068K 4
28 Screw, Fl. Hd., 1/4-28 x 1/2", Steel, CP 0112-0018J 4
Notes:
(1) Typically two Pipe Plugs required, size depends upon ports to be plugged.
(2) Stainless steel housing screws are recommended when the instrument is installed in
a corrosive environment.
(3) Non-C Versions for specific nuclear service applications only.
(4) When ordering, please specify model number of instrument. Minimum parts order is
$100.00. For actuated instrument parts, refer to separate instrument manual.
(5) Housing pressure ratings are for housings only - not DPU or instrument.

NOTICE
(Use Identical Replacement Parts)
Always replace parts with identical parts (same material, rating(s), certs,
size, etc.), unless otherwise directed by PRIME Factory.

33
(Blank Page)

34
(Blank Page)

35
Product Warranty
A. Warranty
PRIME Measurement Products, LLC warrants that at the time of shipment the products manu-
factured by PRIME Measurement Products, LLC. and sold hereunder will be free from defects In
material and workmanship, and will conform to the specifications furnished by or approved by
PRIME Measurement Products, LLC.
B. Warranty Adjustment
(1) If any defect within this warranty appears, Buyer shall notify PRIME Measurement Prod-
ucts, LLC immediately.
(2) PRIME Measurement Products, LLC agrees to repair or furnish a replacement for, but
not install, any product which within one (1) year from the date of shipment by PRIME
Measurement Products, LLC shall, upon test and examination by PRIME Measurement
Products, LLC, prove defective within the above warranty.
(3) No product will be accepted for return or replacement without the written authorization
of PRIME Measurement Products, LLC. Upon such authorization, and in accordance with
instructions by PRIME Measurement Products, LLC, the product will be returned shipping
charges prepaid by Buyer. Replacements made under this warranty will be shipped prepaid.
C. Exclusions from Warranty
(1) THE FOREGOING WARRANTY IS IN LIEU OF AND EXCLUDES ALL OTHER EXPRESSED
OR IMPLIED WARRANTIES OF MERCHANTABILITY, OR FITNESS FOR A PARTICULAR
PURPOSE, OR OTHERWISE.
(2) Components manufactured by any supplier other than PRIME Measurement Products, LLC shall
bear only the warranty made by the manufacturer of that product, and PRIME Measurement
Products, LLC assumes no responsibility for the performance or reliability of the unit as a whole.
(3) "In no event shall PRIME Measurement Products, LLC be liable for indirect, incidental, or
consequential damages nor shall the liability of PRIME Measurement Products, LLC arising in
connection with any products sold hereunder (whether such liability arises from a claim based
on contract, warranty, tort, or otherwise) exceed the actual amount paid by Buyer to PRIME
Measurement Products, LLC for the products delivered hereunder."
(4) The warranty does not extend to any product manufactured by PRIME Measurement Products,
LLC which has been subjected to misuse, neglect, accident, improper installation or to use in
violation of instructions furnished by PRIME Measurement Products, LLC.
(5) The warranty does not extend to or apply to any unit which has been repaired or altered at
any place other than at the PRIME Measurement Products, LLC factory or service locations by
persons not expressly approved by PRIME Measurement Products, LLC.

Disclaimer
PRIME Measurement Products, LLC makes no warranty of any kind with regard to this manual, materials attached to or provided with, includ-
ing, but not limited to, implied warranties of merchantability and fitness for a particular purpose. PRIME Measurement Products, LLC shall
not be liable for errors contained in this manual, including materials attached to, or provided with, or for incidental or consequential damages
in connection with the furnishing, performance, or use of this information. The information contained in this manual (along with attached
or provided materials) is subject to change without notice. The contents of this manual and all attached/provided materials may be revised
without prior notice.

Product Brand
Barton® brand products are manufactured and sold by PRIME Measurement Products, LLC (formerly Barton Instrument Systems) in Nuclear
and Government Markets. Former Barton® brand Industrial Products are manufactured and sold by PRIME Measurement Products, LLC as
PRIME products. Barton® is a trademark of NuFlo Technologies, Inc..

PRIME Measurement Products, LLC


(Formerly Barton Instrument Systems)
900 S. Turnbull Canyon Rd.
City of Industry, CA 91745 USA
(626) 961-2547
On the Web at www.prime-measurement.com
©Copyright 2003-2006, PRIME Measurement Products, LLC. All rights reserved.
MODELS 288A/288C/290D
DIFFERENTIAL PRESSURE INDICATING-SWITCHES
(PRIME M288C for Industrial Service and
Barton® brand M288A (non-C) for
Nuclear Service)
288A uses 224 DPU
288C/290D use 224C DPU
Model 288C

Installation Manual
Version 06G08a
[This manual is for indicating-switch only - see separate ID#10300
224/224C DPU manual.] 7/2006
CONTENTS
Section 1 - Introduction (Specifications 1-4, page 5) .................... Page 3
Section 2 - Installation .............................................................. Page 7
Section 3 - Maintenance:
3-1. DPU Installation/Maintenance/Repair ............................. Page 13
3-2. Tools............................................................................. Page 13
3-3. Bezel/Lens (or Cover) Installation and Removal ................ Page 13
3-4. Calibration Check .......................................................... Page 14
3-5. Pointer Installation and Removal...................................... Page 15
3-6. Indicator Calibration ....................................................... Page 15
3-7. Complete Calibration ..................................................... Page 16
3-8. Changing Switch Setpoint............................................... Page 18
3-9. Preventative Maintenance ............................................... Page 20
3-10. Locking Drive Arm to Torque Tube................................ Page 20
3-11. Drive Arm Tightness Test ............................................. Page 21
3-12. Troubleshooting ........................................................... Page 22
Section 4 - Parts Drawing/List:
Figure 4-1. 288A/288C Parts Drawing .................................. Page 25
Table 4-1. 288/290 General Parts List ................................... Page 26
Table 4-2. External Wire Lead Part Numbers ........................... Page 30
Figure 4-2. 288C SST Specific Parts Drawing ......................... Page 31
Table 4-3. 288C SST Specific Parts List ................................. Page 31
Figure 4-3. 290D Specific Parts Drawing ................................ Page 32
Figure 4-4. 290D Detail Drawing ........................................... Page 33
Table 4-4. 290D Specific Parts List ........................................ Page 33
Section 5 - Install/Dimensional Drawings:
Figure 5-1. 288A/288C OD Drawing .................................... Page 34/35
Figure 5-2. 290D OD Drawing .............................................. Page 36/37

Measurement Products
(Formerly Barton Instrument Systems)
SAFETY
Before installing this instrument, become familiar with the installation instructions
in Section 2 and in the separate 224/224C DPU manual.
DANGER notes indicate the presence of a hazard which will cause severe
personal injury, death, or substantial property damage if warning is ignored.
WARNING notes indicate the presence of a hazard which can cause severe
personal injury, death, or substantial property damage if warning is ignored.
CAUTION notes indicate the presence of a hazard which will or can cause
moderate personal injury or property damage if warning is ignored.
DANGER, WARNING, and/or CAUTION notes that appear on the following pages
of this manual should be reviewed before proceeding: 13. (Important! Before
installing or operating this instrument, review all safety notices contained in
the separate 224/224C DPU manual, per models listed on front page.)

PRODUCT WARRANTY STATEMENT


The product warranty applicable to this PRIME Measurement Products, LLC is as stated
on the back cover of this manual.
Should any after-delivery problems arise, please contact PRIME Measurement Products
Customer Service, during normal business hours, at (626) 961-2547.

DPU "C" VERSION DESIGN CHANGE


The 288C and 290D utilize a 224C DPU, which is a redesigned version of the 224 DPU.
The "C" version is identical in function, performance, installation, and operation to the
previous version - redesign was for improved manufacturing only. This design change
does not affect the instrument being actuated. Barton-brand instruments, using the
Model 224 (Non-C) DPU, are for Nuclear service only.

RECORD OF CHANGES
CHANGE DATE DESCRIPTION
NO.
03E08n 5/03 Booklet; Combined 288/290 & 289/291 manuals (replaces
ID#10300 & 10310); Assigned unique ID# (10305); Revised
Wiring diagrams, Fig. 2-1; Reorganized; Removed explosion-
proof "A" Case version information; Add Drive Arm to Torque
Tube Locking procedure/Tightness Test; Add Bezel/Lens
install/removal procedure; Updated ODD drawings; Updated
Parts Drawings; Updated Parts Lists
03H10f 8/03 Combined 288A/288C/290D Versions; Removed 289 and 291
information (Replaces 288/290 portions of (10305) Manual;
Rev. Co. Name/Logo to PRIME Measurement Products;
04A05a 1/04 Rev. para. 1-3: ref.: Std. models can have one or two alarm sw.
04C05a 6/04 Corrected version number to 04C05a. No other changes.

06G08a 7/06 Rev. pg. 2 statements; Updated to new wire color coding
scheme {Add Table 2-1 - Wire Color Coding, Rev. wiring
diagrams throughout, and Rev. Section 4 throughout (parts
drawings/lists)]; Updated Section 5; Ref. Barton brand for
Nuclear Service (cover) ; Non-technical changes throughout

2
SECTION 1 - INTRODUCTION

1-1. General
The weatherproof Models 288A, 288C, and the explosion-proof Model 290D
are DP Indicating Switches. The 288A and 288C have a NEMA-4 watertight
die-cast aluminum case (finished with a weather-resistant black epoxy resin
paint). The cover lens is secured in the bezel with an elastomer ring to reduce
the possibility of accidental breakage. This ring also acts as a seal between the
bezel and the case to ensure a moisture, fume and dust-free atmosphere for the
indicator and switch mechanism.
Model 290D has an explosion-proof case that is certified for Class I, Division 1,
Groups B, C & D service. The large cover lens allows maximum readability of
the indicating pointer.
Switches and all adjustments are readily accessible when the cover is removed.
The built-in switches energize either single or dual alarm circuits when the mea-
sured differential pressures exceed predetermined limits. These limits may be
either maximum, minimum, or both.
1-2. Main Components
Stop Screw Stop Screw
Actuating
Low Alarm Cam High Alarm
Switch Switch
Switch
Lock Switch
Lock
Plunger
Screw Plunger
Screw
Differential
Drive Arm
Low Alarm High Alarm
Relay Relay
Switch
Link

Hex Hub
Terminal Zero
Strip Adjust

Low Alarm Movement


Switch Assembly
Adjust
High Alarm
Switch
Pointer Range Adjust
Adjust
Figure 1-1. Switch Components

A. Indicating Switch
B. Differential Pressure Unit (DPU)
• 288A is actuated by a Model 224 DPU
• 288C and 290D are actuated by a Model 224C DPU
For detailed information on the DPU, see separate 224/224C DPU
manual.
3
1-3. Indicating Switch (refer to Figures 1-2/2-1 to 2-6 and Table 2-1)
Rotation of DPU torque tube shaft is coupled through connecting linkage within
the switch case to move the pointer across the scale plate. An actuating cam,
directly connected to torque tube shaft, rotates with the motion of the shaft.
Two cam follower roller/actuator arm assemblies, one for each switch, respond
to torque tube rotation by opening and closing the switches as they ride on and
off the cam. The levels of DP at which the switches actuate are adjustable with
high and low alarm switch adjustments on the scale plate.
Standard models can have one or two alarm switches. Each switch can be con-
nected to operate normally-opened or normally-closed. The direct-set switch
contacts are adjustable over a scale range of 5-95% nominal.
BOTH SWITCHES RELAXED Mid-point DP
50

CAM 0 100

SWITCH “A” SWITCH “B” CLOSED OPEN


SWITCH
(LEFT) (RIGHT) CONTACTS CONTACTS
SWITCH LOW HIGH A (LEFT) YELLOW & BLUE RED & YELLOW
PLUNGER
SCREW B (RIGHT) VIOLET & ORANGE VIOLET & BLACK
ORANGE
YELLOW

VIOLET

BLACK

Note: Table above references Current Configurations


BLUE
RED

color coding (refer to Table 2-1).

TERMINAL
BLOCK IN
CASE

SWITCH “A” ACTUATED/SWITCH “B” RELAXED Zero DP


50

0 100

SWITCH “A” SWITCH “B” CLOSED OPEN


(LEFT) (RIGHT) SWITCH
CONTACTS CONTACTS
LOW HIGH
A (LEFT) RED & YELLOW YELLOW & BLUE

B (RIGHT) VIOLET & ORANGE VIOLET & BLACK

Note: Table above references Current Configurations


color coding (refer to Table 2-1).

SWITCH “A” RELAXED/SWITCH “B” ACTUATED Maximum DP


50

0 100

SWITCH “A” SWITCH “B” CLOSED OPEN


SWITCH
(LEFT) (RIGHT) CONTACTS CONTACTS
LOW HIGH A (LEFT) YELLOW & BLUE RED & YELLOW

B (RIGHT) VIOLET & BLACK VIOLET & ORANGE

Note: Table above references Current Configurations


color coding (refer to Table 2-1).

Figure 1-2. Switch Actuation Example


(current configuration color codes - see Table 2-1)
Notes: (1) Cam rotates counter-clockwise with increased pressure; (2) Switches shown w/low at 25%
DP and high at 75% DP; (3) To change setpoint: loosen lock screw, move switch plate, tighten screw,
and test setpoint; (4) Internal wires use No. 22 AWG and external wires use No. 18 AWG
4
1-4. Specifications
General:
Actuating Unit (DPU) ............... per specific instrument model (refer to page 1
of this manual)
Dial Size ................................. 6-inches (150 mm)

Accuracy of Indication ............. ±1%* of full scale DP


Accuracy at
Point of Switch Actuation......... ±1.5%* of full scale differential pressure
*NOTICE: Accuracy varies with DP range, bellows size, etc.. For general
accuracy information, refer to Ind/Switch/Xmtr Product Bulletin (ID#21920).
For specific model/configuration accuracy specifications, consult Factory.
Temperature Limits (Ambient)... -40°F/°C to +180°F (+80°C)

Switch:
Accuracy of:
Switch Repeatability .......... ±0.25% of full scale
Switch Deadband .............. ±5% (SPDT); ±6% (DPDT)

Switch Type............................ Mechanical, Snap-Acting; all switches are SPDT


(DPDT switches are stacked SPDT switches
with a common actuator)
Contact Type .......................... Single Pole, Double Throw (SPDT)
Double Pole, Double Throw (DPDT)
Adjustment ............................. All switches are adjustable between 5% and
95% of factory calibrated scale
Activation ............................... Switches can be calibrated to actuate either
increasing or decreasing scale
Switch Contact Rating:
2.5 Amps @ 250 VAC, 50/60 Hz ‚
5.0 Amps @ 250 VAC, 50/60 Hz
5.0 Amps @ 125 VAC, 50/60 Hz ‚
3.0 Amps @ 30 VDC (Resistive) ‚ƒ
0.4 Amps @ 125 VDC (Resistive)
0.2 Amps @ 125 VDC (Inductive) ‚„
0.1 Amp @ 30 VDC (Inductive) ƒ„
See Notes on page 6.

(continued on next page...)

5
1-4. Specifications (continued)
Relay:
Contact Types......................... Single Pole, Double Throw (SPDT)
Double Pole, Double Throw (DPDT)
Relay Contact Ratings:
25 Amps @ 150/250 VAC, 50/60 Hz, SPDT
25 Amps @ 300 VAC, 50/60 Hz, DPDT
5.0 Amps @ 600 VAC, 50/60 Hz
13 Amps @ 28 VDC ƒ
20 Amps @ 15 VDC ƒ
0.75 HP @ 200/240 VAC, 50/60 Hz
1.0 HP @ 200/240 VAC, 50/60 Hz, SPDT
1.5 HP @ 200/240 VAC, 50/60 Hz, DPDT
10 Amps (Resistive) @ 120 VAC ‚
5 Amps (Inductive) @ 120 VAC ‚„
10 Amps (Resistive) @ 26.5 VDC ‚ƒ
5 Amps (Inductive) @ 26.5 VDC ‚ƒ„
0.5 HP @ 120 VAC ‚
1.0 HP @ 120 VAC ‚

Relay Coil Voltages:


6 VAC @ 5VA Max.
12 VAC @ 5VA Max.
24 VAC @ 5VA Max.
120 VAC @ 5VA Max.
230 VAC @ 5VA Max.
6 VDC @ 2W Max.
12 VDC @ 2W Max.
24 VDC @ 2W Max.
110 VDC @ 2W Max.

Switch/Relay Wiring:
Size........................................ 22 AWG for internal Switch/Relay Coil Wires;
18 AWG for all external wires and relay con-
tacts
Insulation................................ Standard (CSA) - PVC;
Radiation Resistant - Tefzel

Notes:
‚CSA Approved
ƒ CE Approved (Voltage limited to less than 50 VDC or 35 VAC for CE
applications)
„Arc suppression recommended for inductive loadings.

6
SECTION 2 - INSTALLATION

2-1. General
The instrument should be inspected at time of unpacking to detect any dam-
age that may have occurred during shipment. Note: The unit was checked for
accuracy at the factory — do not change any of the settings during examina-
tion or accuracy will be affected.
For applications requiring special cleaning/precautions, a polyethylene
bag is used to protect the instrument from contamination. This bag
should be removed only under conditions of extreme cleanliness.

2-2. Mounting/Piping/DPU Installation


Refer to the appropriate DPU Manual for the Model being installed.
2-3. Electrical Connection (Switches/Relays)
Units are supplied with either single or dual alarm switches and/or relays (de-
pending on customer order). The direct-set switch contacts are adjustable over
the entire scale range.
Table 2-1 shows switch and relay wiring color coding for legacy and current
configurations. Figures 2-1 through 2-6 show switch and relay wiring.
The high switch and low switch set point adjustment procedures are covered in
paragraph 3-7, page 16.
For physical location of switches, see page 3.
2-4. Switch Use
Switch contact life is influenced by various application conditions such as tem-
perature, humidity, airborne contamination, vibration, amount of plunger travel,
cycling rate, and rate of plunger travel (and others), as well as by the electrical
(circuit) characteristics.
NOTICE: Arc suppression for inductive loads will prolong the life of the switch
contacts.
Note: Due to their size, subminiature switches have small mechanical clear-
ances; therefore, no rating above 250 VAC has been established.
2-5. Startup
For startup procedures, Warnings, and information, refer to the separate DPU
Manual that is appropriate for the indicator-switch model (see front page of this
manual).
NOTICE
To ensure the unit calibration is within factory-set calibration tolerances, per-
form the Calibration Check procedure in paragraph 3-4, page 13.

7
TABLE 2-1. SWITCH/RELAY WIRE COLOR CODING (4/06c)
CURRENT CONFIGURATIONS (Note 2)
8

LEGACY CONFIGURATIONS 288A, 288C, 289A, 290D, 291B, 321, 322, 1512
LEGACY CONFIGURATIONS
288A, 288C, 289A (290D/291B after Aug. 2005)
290B, 290D, 291B
(prior to Jan. 2005 - No CSA label or (288A/288C/289A w/CSA Label after Jan. 2005)
(prior to Aug. 2005) (288A/288C/289A w/Tefzel Insulation after Apr. 2006)
prior to Apr. 2006 w/Tefzel insulation)
(All other Models after Jun. 2006 [est.])

NO C NC NO C NC NO C NC

SPDT SWITCHES
Low Red Yellow Blue Red Yellow Blue Red Yellow Blue
High Black Green White Black Violet White Black Violet Orange
DPDT SWITCHES
Low #1 Red Yellow Blue Red Yellow Blue Red Yellow Blue
Low #2 White/Red White/Yellow White/Blue White/Red White/Yellow White/Blue White/Red White/Yellow White/Blue
High #1 Black Green White Black Violet White Black Violet Orange
High #2 White/Black White/Green White/Violet White/Black/Gray White/Black/Violet White/Black/Blue White/Black White/Violet White/Orange
4-INDEPENDENTLY ADJUSTABLE SWITCHES
Low #1 Red Yellow Blue Red Yellow Blue Red Yellow Blue
Low #2 White/Red White/Yellow White/Blue White/Red White/Yellow White/Blue White/Red White/Yellow White/Blue
High #1 Black Green White Black Violet White Black Violet Orange
High #2 White/Black White/Green White/Violet White/Black/Gray White/Black/Violet White/Black/Blue White/Black White/Violet White/Orange

SWITCHES FOR RELAYS


Low Red Yellow Blue (Note 1) Red Yellow Blue (Note 1) White/Brown Brown White (Note 1)

High Black Green White (Note 1) Black Violet White (Note 1) White/Gray Gray White (Note 1)

RELAYS
Low #1 Gray Blue Brown Gray Blue Brown Red Yellow Blue

Low #2 White/Gray White/Blue White/Brown White/Gray White/Blue White/Brown White/Red White/Yellow White/Blue

High #1 Violet White Orange White/Red White Orange Black Violet Orange

High #2 White/Violet White/Black White/Orange White/Violet White/Black White/Orange White/Black White/Violet White/Orange

Coil Wiring: Legacy Ver.: Low = Red and High = Black Current Ver.: Low = White/Brown and High = White/Gray

Note 1: Wire is NOT connected. Note 2: The Current Configurations wiring information does not apply to Model 288C for military service.
9
Closed in (shelf) condition. C= Common.
position above the pointer scale position. NO = Normally Open in (shelf) condition. NC = Normally
low switch set to trip at a position below the pointer scale position, and the high switch set to trip at a
NOTICE: Figures 2-1 through 2-6 show: switch & relay contacts in the relaxed (shelf) condition, the
WHITE/RED
WHITE/YELLOW
NO C NC

WHITE/BLUE
RED RED
1

YELLOW YELLOW
2

GREEN
NO C NC

BLUE BLUE
3

CASE GROUND
LOW DPDT SWITCH

WHITE/RED
4

WHITE/YELLOW
5

RED RED
1

WHITE/BLUE
6

YELLOW YELLOW
2
NO C NC

BLUE BLUE
3
LOW SPDT SWITCH

GREEN
CASE GROUND
WHITE/RED RED RED
1

WHITE/YELLOW YELLOW YELLOW


2

NO C NC

WHITE/BLUE
NO C NC

BLUE BLUE
3
4

RED RED
5

1
YELLOW YELLOW
6

2
NO C NC
7

BLUE BLUE

3
ORANGE ORANGE
8

WHITE/RED

4
WHITE/YELLOW VIOLET VIOLET

5
NC C NO

WHITE/BLUE BLACK BLACK

6
9 10

WHITE/ORANGE

7
WHITE/VIOLET
LOW & HIGH SPDT SWITCHES

8
GREEN
WHITE/BLACK

9
CASE GROUND
ORANGE ORANGE
VIOLET VIOLET

NC C NO
BLACK BLACK

10 11 12
WHITE/ORANGE LOW & HIGH DPDT SWITCHES
GREEN WHITE/VIOLET
CASE GROUND

NC C NO
WHITE/BLACK
ORANGE ORANGE
8

VIOLET VIOLET
9
NC C NO

BLACK BLACK
10

WHITE/ORANGE

5
WHITE/VIOLET

6
GREEN
WHITE/BLACK

7
CASE GROUND
ORANGE ORANGE

8
VIOLET VIOLET
HIGH SPDT SWITCH

9
NC C NO
BLACK BLACK

10
Figure 2-1. Low/High SPDT Switch Diagrams (current configuration color codes - see Table 2-1)

Figure 2-2. Low/High DPDT Switch Diagrams (current configuration color codes - see Table 2-1)
WHITE/ORANGE
GREEN WHITE/VIOLET
CASE GROUND

NC C NO
WHITE/BLACK

HIGH DPDT SWITCH


10

3-SWITCHES
SPDT LOW & DPDT HIGH 4-SWITCHES
3-SWITCHES
DPDT LOW & SPDT HIGH INDEPENDENTLY ADJUSTABLE

NO C NC NC C NO NC C NO

NO C NC NO C NC NC C NO NO C NC NO C NC NC C NO NC C NO

WHITE/ORANGE
WHITE/ORANGE

WHITE/BLUE
WHITE/YELLOW
WHITE/BLUE

WHITE/VIOLET
WHITE/YELLOW
YELLOW

WHITE/VIOLET
ORANGE
BLUE

YELLOW

WHITE/BLACK
ORANGE
YELLOW
RED

VIOLET
BLACK

WHITE/RED
WHITE/BLACK

VIOLET
ORANGE

BLACK
BLUE

VIOLET
BLACK

BLUE
RED
WHITE/RED

RED
1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10
1 2 3 4 5 6 7 8 9 10 11 12
WHITE/ORANGE

WHITE/YELLOW
CASE GROUND
WHITE/VIOLET

CASE GROUND
WHITE/BLACK

WHITE/ORANGE

CASE GROUND
WHITE/BLUE

WHITE/YELLOW

WHITE/VIOLET
WHITE/RED

WHITE/BLACK
WHITE/BLUE
ORANGE
ORANGE

YELLOW
YELLOW

WHITE/RED
VIOLET
GREEN
VIOLET

BLACK

GREEN
BLACK

ORANGE
BLUE

GREEN
BLUE

YELLOW
RED

VIOLET
RED

BLACK
BLUE
RED
Figure 2-3. 3-Switches Figure 2-4. 4-Switches, Independently Adjustable
(current configuration color codes - see Table 2-1) (current configuration color codes - see Table 2-1)

NOTICE: Figures 2-1 through 2-6 show: switch & relay contacts in the relaxed (shelf) condition, the low switch set to trip at a position below the pointer scale position, and the
high switch set to trip at a position above the pointer scale position. NO = Normally Open in (shelf) condition. NC = Normally Closed in (shelf) condition. C= Common.
LOW & HIGH
SWITCHES WITH
SPDT RELAYS
LOW SWITCH
HIGH SWITCH
WITH SPDT RELAY
WITH SPDT RELAY
NO C NC NO C NC NC C NO NC C NO

WHITE/GRAY
WHITE/GRAY
WHITE/BROWN
WHITE/BROWN

WHITE
WHITE

GRAY
GRAY
BROWN
BROWN

WHITE
WHITE

WIRE NUT (TYPICAL)

NC BLUE NC BLUE ORANGE NC ORANGE NC


C YELLOW C YELLOW VIOLET C VIOLET C

NO RED NO RED BLACK NO BLACK NO

WHITE/BROWN WHITE/BROWN WHITE/GRAY WHITE/GRAY

WHITE/BROWN WHITE/BROWN WHITE/GRAY WHITE/GRAY

1 2 3 4 5 1 2 3 4 5 6 7 8 9 10 6 7 8 9 10
WHITE/BROWN

CASE GROUND

WHITE/BROWN

CASE GROUND

CASE GROUND
WHITE/GRAY
WHITE/GRAY

ORANGE
YELLOW

ORANGE
YELLOW
BROWN

GREEN

VIOLET
BROWN

GREEN

GREEN
VIOLET

BLACK
BLACK

GRAY
BLUE

GRAY
BLUE
RED

RED

Figure 2-5. Low/High Switch(es) w/SPDT Relay(s) Diagrams (current configuration color codes - see Table 2-1)

NOTICE: Figures 2-1 through 2-6 show: switch & relay contacts in the relaxed (shelf) condition, the low switch set to trip at a position below the pointer scale position, and the
11

high switch set to trip at a position above the pointer scale position. NO = Normally Open in (shelf) condition. NC = Normally Closed in (shelf) condition. C= Common.
12

LOW & HIGH SWITCHES


WITH DPDT RELAYS

LOW SWITCH HIGH SWITCH


WITH DPDT RELAY WHITE/RED NO C NC NC C NO
WHITE/BLACK
WITH DPDT RELAY
WHITE/YELLOW WIRE NUT WHITE/VIOLET

WHITE/BROWN
WHITE/RED (TYPICAL) NC C NO
WHITE/BLACK
NO C NC

WHITE/GRAY
WHITE/BLUE WHITE/ORANGE
WHITE/YELLOW WHITE/VIOLET
WHITE/BROWN

BROWN

WHITE/GRAY
WHITE/BLUE WHITE/ORANGE

WHITE
WHITE
NC NC

GRAY
C C
BROWN

WHITE
WHITE

NC NC

GRAY
NO NO
C C

NO NC BLUE ORANGE NC NO
C YELLOW VIOLET C
NC BLUE ORANGE NC
NO RED BLACK NO
C YELLOW VIOLET C

NO RED WHITE/BROWN WHITE/GRAY BLACK NO

WHITE/BROWN WHITE/GRAY
WHITE/BROWN WHITE/GRAY

WHITE/BROWN WHITE/GRAY

1 2 3 4 5 1 2 3 4 5 6 7 8 9 10 6 7 8 9 10

WHITE/BROWN
WHITE/BROWN

CASE GROUND
CASE GROUND

CASE GROUND
WHITE/GRAY

WHITE/GRAY
ORANGE
YELLOW
YELLOW

ORANGE
BROWN
BROWN

GREEN
GREEN

GREEN
VIOLET

VIOLET
BLACK

BLACK
BLUE
BLUE

GRAY

GRAY
RED
RED

Figure 2-6. Low/High Switch(es) w/DPDT Relay(s) Diagrams (current configuration color codes - see Table 2-1)

NOTICE: Figures 2-1 through 2-6 show: switch & relay contacts in the relaxed (shelf) condition, the low switch set to trip at a position below the pointer scale position, and the
high switch set to trip at a position above the pointer scale position. NO = Normally Open in (shelf) condition. NC = Normally Closed in (shelf) condition. C= Common.
SECTION 3 - MAINTENANCE AND CALIBRATION

3-1. DPU Installation/Maintenance/Repair


DPU inspection, cleaning, service, repair, range change, and BUA replacement
procedures (along with applicable WARNINGS, CAUTIONS, and NOTICES)
are presented in the separate 224/224C DPU manual.
3-2. Tools
Table 3-1. Tools
Equipment Purpose
Pointer Puller Pointer Removal
Small Screwdriver Calibration Adjustment
Medium Screwdriver Bezel Removal and Replacement
1/4" and 1/8" Open-end Wrenches Zero (1/4") and Range (1/8") Adjustments
1/8 Hex Allen Wrench Switch Setpoint Adjustment

3-3. Bezel/Lens (or Cover) Installation and Removal


WARNING
(EXPLOSIONPROOF UNITS)
PRIOR TO LOOSENING ANY ATTACHING HARDWARE, HOUSING
BOLTS, OR COVER ASSEMBLY, ASSURE SURROUNDING AREA
IS, AND REMAINS, WELL VENTILATED.
PRIOR TO LOOSENING OR REMOVAL CASE/COVER ASSEMBLY,
ALL ELECTRICAL POWER SUPPLIES MUST BE TURNED OFF.
THE COVER MUST BE REMOVED TO CALIBRATE THE
INSTRUMENT.
BEFORE ANY MAINTENANCE/CALIBRATION, REVIEW THE
APPROPRIATE (SEPARATE) DPU MANUAL.

For non-SST/non-explosionproof units, the bezel gasket (p/n 0277-


0026C) must be installed as shown below (also ref. Fig. 4-1):

Bezel
To remove the bezel and cover
(0277-0029C) lens on non-SST/non-explosion-
Cover (Lens)
(0181-0038C) proof units:
Gasket 1. Loosen three screws on the
(0277-0026C) front of bezel.
Case 2. Tilt out bottom of bezel and
Figure 3-1. Bezel/Lens slide bezel upward.

The two snubbers (p/n 0266-0028C) on the scaleplate should not be


compressed against the lens cover and the pointer should not touch the lens.
(Continued on next page...)

13
3-3. Bezel/Lens (or Cover) Installation and Removal (Continued)
NOTICE
Ensure correct bezel gasket orientation before placing instrument
back in service. Incorrect bezel gasket orientation will cause the
instrument indicator to jam, resulting in inaccurate readings.
For 288C SST Version units, refer to Fig. 4-2, page 31.
For Explosionproof units, cover is unscrewed to gain access to the inter-
nal components. When re-installing cover, tighten cover securely and inspect
lens for cracks or other defects that may affect the explosionproof rating.
Refer to Fig. 4-3, page 32.
3-4. Calibration Check
To ensure the unit calibration is within factory-set calibration tolerances, per-
form the following procedure. NOTICE: Review all procedures, WARN-
INGS/NOTICES in the separate 224/224C DPU manual BEFORE
performing this procedure.
1. Mount instrument in an approx. level position and connect to a standard
pressure source (see separate 224/224C manual).
2. If the zero indication is incorrect,
remove bezel/lens assembly (cover
on explosionproof units) and re-adjust
zero, per the following: Pointer "slipped" HUB
rotated on hub
With a 1/4" open-end wrench (in-
cluded in calibration kit 0288-1032B),
hold the hexagon pointer hub fixed
and rotate the pointer with fingers
until the pointer accurately indicates
Hub held in place
ZERO on the scale. See Figure 3-2. with 1/4" wrench
Replace bezel/lens or cover.
3. To test for reverse travel, connect
pressure source to LP housing and
Figure 3-2. "Slipping" Pointer
vent HP housing. Apply pressures
approximately 150% of the DP range.
The pointer should move approximately 5% to 10% below zero.
4. To test for overtravel, connect pressure source to HP housing and vent LP
housing. Apply pressures approximately 150% of DP range. The pointer
should move approximately 5% to 10% above full scale.
5. Apply 0, 50, and 100% of full scale pressure. If indication is within speci-
fied limits, no adjustments are necessary. If indication is not within specified
limits, perform a complete calibration, per paragraph 3-7, page 16.
6. Make sure instrument zero indication is correct; otherwise, repeat step 2.
7. Verify switch set points (refer to paragraph 3-8, page 18).

14
3-5. Pointer Installation and Removal
During adjustment and calibration of the unit, it may be necessary to remove
and reinstall the pointer, per the following procedures:
A. Pointer Installation
1. Position pointer on movement shaft with pointer set at zero scale.
It may be necessary to enlarge the hub hole, using a tapered broach
(included in the toolkit (p/n 0288-1032B).
2. Lightly tap pointer hub with a hand-set or other flat-end tool. Use
perpendicular blows to avoid bending shaft.
3. Check indicating switch for calibration over its entire range (refer
to Switch Calibration in this section). If indicating switch is correctly
calibrated, secure pointer to movement shaft by tapping hub with a
hand-set or other flat-end tool.
B. Pointer Removal
Pointer is removed with Pointer Puller (included in the calibration toolkit
0288-1032B), see Figure 3-3.
1. Slide pointer puller along pointer until pin protruding from tip of screw
in pointer puller is directly over movement shaft and arms of pointer
puller are directly under pointer. POINTER PULLER
2. Gently turn knurled head of
screw clockwise, pushing pin
against movement shaft and
HUB
lifting pointer with arms. Finger
pressure should be sufficient to POINTER
pull the pointer free. If more
pressure is required, an Allen
wrench (inserted into head of
HUB
the screw) can be used. How-
MOVEMENT SHAFT
ever, care should be exercised:
too much pressure can cause the Figure 3-3. Pointer Puller
pin to break. (p/n 0163-0005B)

3-6. Indicator Calibration (DP=Differential Pressure)


1. Securely mount instrument
in an approx. level position
and connect DPU into the
test setup, as described in
the appropriate (separate)
DPU manual.
2. Align linkage between drive
arm and movement, per
Figure 3-4.
3. Check pointer for zero
indication. If necessary, set
pointer to zero, by slipping
pointer on hub, per para-
graph 3-4, step 2.

Figure 3-4. Range/Linearity Adj. (50% DP)


15
3-6. Indicator Calibration (Continued) (DP=Differential Pressure)
4. Apply 100% pressure. If pointer exceeds full-scale, lengthen movement
range arm.
5. Release pressure. Set pointer to zero, by slipping pointer on hub, per para-
graph 3-4, step 2.
6. Repeat steps 4 and 5, as necessary, to obtain correct zero/full-scale.
7. Apply 50% DP. If pointer does not indicate 50% scale, a linearity adjust-
ment is necessary. Loosen drive arm screw and move arm to shift pointer
in direction of error a distance of about 10 times linearity error. Tighten
drive arm screw.
8. Release pressure and reset pointer at zero. Check the span. If gear in
movement reaches limit of travel as a result of linearity adjustment (step 7),
slip range arm along gear approximately 5 degrees from normal 37.5 de-
gree angle to approximately 43 degrees (see step 2). Range arm is slipped
by applying pressure to range arm with thumb, while holding gear firmly in
place. Retest pointer response at 50%, 0%, and 100% of full-scale differen-
tial pressure, and adjust linkage until readings are acceptable.
9. Apply 0%, 25%, 50%, 75%, 100%, 75%, 50%, 25%, and 0% of full-scale
differential pressure consecutively to instrument without overshoot.
Lightly tap indicator to overcome friction. Pointer should accurately indicate
each applied pressure.
10. Test instrument repeatability by applying 0%, 50%, 0%, 50% of full-scale
differential pressure. Indicator should accurately indicate each applied pres-
sure.
11. Set stops to prevent pointer from striking snubbers on scale. Test setting by
moving pointer from zero position to 50% position manually and then let-
ting pointer return freely. If necessary, tap pointer hub to tighten it to shaft.
3-7. Complete Calibration (See Warning on page 13)
Before performing complete calibration procedure, verify that instrument is out
of calibration, by performing calibration check procedure (para. 3-4, page 14).
If instrument is out of calibration, before performing complete calibration proce-
dure, remove bezel/lens and inspect switch mechanism to verify the following:
• The roller rotates without wobble or binding.
• The cam does not touch the roller side shields.
• The actuator arm moves
freely on its pivot. CAM
Switch Roller
• All switch mounting screws
are tight.
• Linkages are straight and do Actuator Arm
not bind at the pivots. Plunger
Correct any problems that are Screw
encountered. Stop CAM
Screw
Figure 3-5. Cam/Follower Position

16
3-7. Complete Calibration (Continued) (See Warning on page 13)
A. Calibration Setup
1. Connect lamp or buzzer to switch output leads. Connect a test voltage
to switch input terminals on terminal strip. (A low voltage is recom-
mended for safety.) If relay is installed in instrument, coil voltage must
be applied to switch.
2. Unlock switch plate and move
plate until roller is
positioned at top
of the cam.
3. Advance plunger
screw until switch
actuates, then
advance plunger
screw an addition-
al 60° (one flat).
4. Exercise switch roller across
top of cam to verify steady
operation. Advance stop screw
to touch switch, then back out
Figure 3-6.
screw 1.5 turns (9 flats).
Linkage Arrangement
B. Calibration Procedure (see Figure 3-6)
To calibrate switch linkage (required when unit is rebuilt):
1. Loosen three linkage screws and turn crank to 12 o’clock position.
2. Use a 1/8-inch Allen wrench to hold index shaft and slip index pointer
to 0 on switch index. Tighten screw on crank to mid-slot position.
3. Turn switch index pointer to “1” (index numbers refer to numbers on
outer edge of scale plate).
4. Apply 10% differential pressure and adjust switch plate until switch
actuates. Lock the two linkage screws.
5. Rotate index pointer to “9.” Apply 90%
differential pressure and adjust crank radius until switch actuates.
6. Recheck 10% and 90% set points. Adjust crank radius and index
pointer until both set points are 2% accurate (nominal).
7. If switch is to be field-set at low differential pressure values (1% or 2%
of pressure range), check crank to prevent a top-dead-center position.
Otherwise, minimum set point position will be restricted and set point
may become reversed.
8. Adjust switch to actuate at desired pressure by applying test pressures
in a decreasing direction, in discrete steps. Allow unit and pressure
system to stabilize. Then change pressure a small amount. The magni-
tude of the pressure change is determined by desired accuracy of test.
Tighten lock screw before testing switch performance.

(Continued on next page...)


17
B. Calibration Procedure (Cont.) (see Fig. 3-6) (See Warning, page 13)
To calibrate switch linkage (required when unit is rebuilt):
8. (Cont.)
The high switch is usually set to actuate at increasing pressure — when
calibrating high switch, apply test pressure in an increasing direction.
This amount of loading will prevent cam-runout of a similar condition.
Excessive plunger loading (more than 3 flats) may cause roller to drag
on cam. Cam friction will be apparent by excessive hysteresis, erratic
pointer readings and inconsistent switch operation.
9. Check switch deadband, (actuate to reset) by applying differential pres-
sures in a decreasing then increasing (opposite for high switch direc-
tion). Observe pressures. To reduce deadband, advance plunger screw
(two flats maximum).
10. Adjust high switch to actuate at desired pressure. The procedure is the
same as for the low switch.
3-8. Changing Switch Setpoint
(Tools: Screwdriver, 1/8 Hex Allen wrench) (See Warning on page 13)
Set Point — The measured pressure at which the snap-switch actuates and
thereby changes the states of the N.O and N.C. contacts. For example, the set
point of the low switch is 24 psid with decreasing pressure.
Deadband — The difference in measured pressures between switch-actuation
and switch-reset. Deadband is usually expressed as percent of full scale (% of
F.S.). Deadband is not adjustable. For example, the switch above was found to
reset at 26.4 psid with increasing pressure. The deadband was 2.4 psi, or 4% of
the full scale (0 to 60 psid).
A. In-Service Instruments (calibration pressures
cannot be applied)
1. Remove bezel/cover. Do not
remove pointer or scale.
2. Insert hex wrench in switch
adjust post, (item A, Fig. 3-7).
3. Loosen switch lock-screw,
(item B), 1/2 to 1 turn.
4. With hex wrench, move index
pointer (item C) to new set
point as indicated on switch
index (item D).
5. If possible, check set point by varying pro-
cess pressures and observing pointer readings
when switch actuates. (Open manifold bypass
valve slowly and watch for “pointer-jump” at
set point or by electrical signal.) Improve set-
ting if necessary, repeat test several times to
verify stability.
Note: Switch index has 10 div., marked 0, 5, and 10.
These match markings on outer edge of scale. Figure 3-7.
Setpoint Adjustment
18
3-8. Changing Switch Setpoint (Cont.) (See Warning on page 13)
A. In-Service (Continued)
Example: Scale has range of 0-60 psid. Set point is 24 psid, with decreas-
ing pressure, (24/60 x 10 = 4). Move index pointer (item C) to division 4.
Start from bottom of switch index (0 on low, 10 on high). Tighten switch
lock (Item B) snug plus 1/4 turn. Do not overtighten.This will place set
point within ±2% of full scale.
B. Out-of-Service (Disconnected from process lines or mounted on bench)
1. Drain and vent housings.
2. Attach calibration pressure source (air or N2) to HP housing.
3. Apply pressures and observe pointer readings for accuracy. Use a pres-
sure standard (Heise gage or equiv.) for reference. Change pressures
slowly in discrete steps. A “bleed-pressure” method may cause errors.
4. Change switch set point as described in Part A.
5. Check set point by changing measured pressure to actuate switch.
For Example: To verify low-switch set point to 24 psid, apply approxi-
mately 30 psi. Then reduce pressure to approximately 25 psid, hold
for a few seconds, then continue in 1/4 psid steps until switch actu-
ates. If set point is incorrect, continue instructions in Part A.
6. To measure switch deadband, reduce pressure to zero, then increase
pressure until switch resets.
7. To verify repeatability of set point, repeat step 5 several times. For
improved accuracy, use smaller increments of pressure. Allow extra
time for slow response gages and for test systems that have long runs
of small bore tubing.
8. High alarm switches right side of the scale) are adjusted in a similar
manner. Apply increasing pressure to establish the switch set point,
decreasing pressure to measure deadband.
C. Notes
1. Always check set point after tightening switch lock screws.
2. Either switch may be set at any point of the scale, except allowance
must be made for deadband to enable switch to reset itself. For ex-
ample, high switch (right side of scale) may be set at 100% of F.S., but
should not be set near zero. Also, low switch may be set at zero, but
not near full scale. (Observe deadband values for specific models).
3. If switch performance is unsatisfactory (set point does not repeat, dead-
band is excessive, pointer exhibits hysteresis, contacts are unstable.
etc.) remove scale and inspect switch and mechanism. Scale is split for
removal without pulling pointer.
4. Switch Variations: (a) SPDT: This is the standard model (low, high, or
both); (b) DPDT: Two switches are stacked and actuated by a single
lever (low, high, or both); (c) Three (or four) SPDT: Switches have
independent switch points - No.1 and 3 are usually low switches set for
decreasing pressures. Nos. 2 and 4 for increasing pressures;
(d) Hermetically Sealed Switch (Micro 11HM41): this switch has
wide deadband, case is modified for electrical clearances, amperes are
reduced for 60 Hz service.
19
3-9. Preventative Maintenance
• Indicating Switch — Periodically inspect alarm switch mechanism to verify
that all mounting screws are seated properly. Inspect linkage for wear.
Inspect integrity of electrical circuits. Tighten as necessary.
• DPU — Warning/Caution notices, inspection/cleaning procedures, and
maintenance procedures are presented in the appropriate (separate) DPU
manual. Do Not perform any maintenance/repair on the instru-
ment or DPU without first reviewing all procedures and Warn-
ing/Caution notices in the separate DPU manual.

3-10. Locking Drive Arm to Torque Tube


Refer to the appropriate (separate) DPU manual. For explosionproof units,
refer to Section 4, 290D Detail Drawing, page 33 (See Warning on page 13).

Figure 3-8. Locking Drive Arm to Torque Tube


1. Slip drive arm over torque tube shaft; clear end of torque-tube housing by
approximately 0.030-inches before securing to prevent interference.
2. To tighten drive arm assembly onto torque-tube shaft:
a. While supporting block/shaft, tighten clamp screw until snug to shaft.
b. Still supporting block/shaft, tighten clamp screw an additional:
• Sintered: 1/3 to 1/2 turn (This screw can normally turn one full
revolution before breaking.)
• Slotted: 1/4 to 1/3 turn (The slot in the slotted clamp block should
still be open.)
NOTICE: For Seismic and High Shock Qualified Units, perform Drive
Arm Tightness Test, per paragraph 3-11.

20
3-11. Drive Arm Tightness Test (See Warning on page 13)
This procedure tests the drive arm to torque tube attachment for tightness, by
applying torque developed by the DPU onto a fixed drive arm. Care should be
taken to apply pressure slowly, as torque is being applied to the connection
through the torque tube drive shaft.
With pointer at normal 0% torque tube rotation position (max. minimum scale
position or 0% on a normal 0 to 100% scale unit), adjust drive arm stop bracket
(or use alternate means) to prevent pointer from moving (stop bracket interferes
with drive arm movement). Note: On reverse acting/split range units, it will be
necessary to pressurize DPU to move pointer to max. minimum scale position,
and on suppressed units, it will be necessary to apply pressure to establish a
reference point to check for "zero" shift.
Pressurize DPU to full calibrated scale DP (100% of full scale range) to achieve
8-degrees of torque tube drive shaft equivalent torque onto the connection.
Observe shift in the unit "zero" following DPU depressurization (as required) and
drive arm stop bracket readjusting (to allow free movement of drive arm and
pointer). A downscale (counter-clockwise) shift in "zero" of greater than 1/2% is
indicative of drive arm slippage necessitating further clamp block tightening.

21
3-12. Troubleshooting (See Warning on page 13)
Refer to Table 3-1 and the separate 224/224C DPU manual.
Table 3-1. Troubleshooting
Problem Possible Probable Cause Corrective Action
Source
Low or Primary Orifice installed backwards or oversized Replace Orifice
No Element
Flow Blocked Upstream from run Clean out run or open
Indication or DPU
valve
(Refer
to DPU Loss of liquid in Reference Leg (liquid Refill Reference Leg
Manual) level)
Density changes in process media or Refill Reference Leg
Reference Leg with same density liquid
as process media
Primary Pressure tap holes plugged and/or piping Clean out piping
Element plugged
to DPU
Bypass Valve Open or Leaking Close bypass valve(s)
Piping
and/or repair leaks
(Refer
to DPU Liquids or Gases trapped in piping Vent piping
Manual)
Block or shutoff valves closed Open block or shutoff
valves
Piping leaks on HP side Repair leaks
Bellows Housing filled with solids restricting bel- Clean out housing
Unit lows movement
(Refer
Gas (liquid service) or liquid (gas service) Vent housing
to DPU
trapped in housing
Manual)
HP housng gasket leak Replace gasket
DPU tampered with Return BUA for repair
Indicator, Loose linkage or movement Tighten or replace
Alarm
Out of Calibration Calibrate
Switch
Mecha- Pointer loose Tighten pointer
nism
Dirty or corroded mechanism Clean or replace
Wiring interfering with movement Re-route wiring
Dirty mechanism Clean mechanism
Continued on next page...

22
3-12. Troubleshooting (Continued) (See Warning on page 13)
Refer to Table 3-1 and the separate 224/224C DPU manual.
Table 3-1. Troubleshooting (Continued)

Problem Possible Probable Cause Corrective Action


Source
High Primary Orifice partially restricted or too small Clean out or replace
Indication Source
Primary Piping leaks on LP side Repair leaks
Element
to DPU
piping
Bellows Gas (liquid service) or liquid (gas service) Vent housing
Unit trapped in housing
(Refer
LP housng gasket leak Replace gasket
to DPU
Manual) Range Spring broken or DPU tampered Return BUA for repair
with
Indicator, Loose linkage or movement Tighten or replace
Alarm
Out of Calibration Calibrate
Switch
Mecha-
nism
Erratic Primary Flow pulsating Install dampening
Indication Element device upstream of
DPU run
Primary Liquid (gas service) or gas bubble (liquid Remove
Element service) trapped in piping
to DPU
Vapor generator incorrectly installed Re-pipe
piping
Reference Leg gassy or Liquid vaporizing See piping instructions
in DPU Manual
Bellows Obstructed Bellows Travel Clean Bellows
Unit
Gas trapped in DPU HP or LP housing Remove (see Startup
(Refer
procedure)
to DPU
Manual) Linkage dragging or dirty Adjust or Clean
Pointer Dragging on Scaleplate Ajust Pointer Position
Inac- Power Blown Fuse Replace Fuse
curate or Supply
Broken or loose wire Repair
No
Electrical Alarm Switch not properly Adjusted Adjust Switch
Alarm Switch
Dirty or Burned Replace switch con-
tacts
Continued on next page...

23
3-12. Troubleshooting (Continued) (See Warning on page 13)
Refer to Table 3-1 and the separate 224/224C DPU manual.
Table 3-1. Troubleshooting (Continued)

Problem Possible Probable Cause Corrective Action


Source
Switch Process Transients or surges cause switches to Add time delay gages
Drifts (set- Changes actuate prematurely or add time circuit
point not
Setpoint and/or Deadband are too wide Specify DP Range as
repeat-
in pressure valves low as practical; set-
able)
point repeatability and
deadband are percent-
age of full range
Electrical overloads affect the spring prop- Examine circuits for
erties of the leaf actuator in the switch Voltage, amperes
DC inductive loads cause arcing and burn- Consider arc-suppres-
ing of contacts sion devices or relays
Accumulation of fluids in piping generate Vent gas or drain liq-
artificial signal uids from signal lines
Calibra- Failure to check setpoint after locking Verify setpoint repeat-
tion Tech- ability after locking
niques switch plate
Rapid pressure change or venting system During calibration,
make pressure changes
in slow, discrete steps
Pressure application in reverse Test low-alarm with
decreasing pressure
and high-alarm with
increasing pressure
Reference Gage Inaccuracy Suitable pressure stan-
dard such as manom-
eter, dead-weight tester,
or Heise-type gage
may be required
Damage to Switch Contacts Adjust plunger Screws
carefully to avoid
damage to internal
parts of switch

24
SEE

SECTION 4 - PARTS DRAWING/LIST


DETAIL B
23 (Low)
23
24 40
KIT
25 27 26
26 (High) 3 Standard
288A/288C
28 (contains 4 ea. Case Shown
39 of items)
29 29 2
29 38
Figure 4-1. 288A/288C Parts Drawing

32 24 1
29 30 27 33 4 2X
(Ref. Tables 4-1/4-2 for part numbers)

32 30 25
34
33 28
35 31 52
34
31 5
35 36
6 COVER (LENS)
(High) 29 BEZEL (0181-0038C)
29 (0277-0029C)
37 2X 7 GASKET
(0277-0026C)
37 29 DETAIL A - Standard 288A/288C
(Item 24 Shown)
11
36
(Low) 21 4 43
4X 19 8
9 6X 41
40
10 38
42
7 2X
44
12
18
45
16 DETAIL B
(Item 25 - Low SPDT SW Ver. Shown)
22 3X 17 2X 13
SEE DETAIL A 20
Standard 288A/288C 14 2X
Bezel/Lens Shown
15

For 288C SST specific parts, ref. Fig. 4-2/Table 4-3.


50 For 290D specific parts, ref. Fig. 4-3/Table 4-4
25
Table 4-1. 288/290 General Parts List
(For 288C SST specific parts, see Fig. 4-2 and Table 4-3)
(For 290D specific parts, see Fig. 4-3/4-4 and Table 4-4)

PART NO. PER


ITEM DESCRIPTION
(•)=model specific UNIT
1 Differential Pressure Unit See separate 1
288A uses 224 DPU 224/224C
288C/290D use 224C DPU DPU Manual
2 Case Assembly: 1
288A/288C (Standard) 0288-0038B (•)
288C Stainless Steel Version See Fig. 4-2
290D See Fig. 4-3
3 Panel Mount Kit (Std.288A/288C Only -not available 0040-1007T (•) 1
for SST Ver. or 290D) (4 ea. of items listed below)
Set Screw, Slotted, 1/4-20x1, STL 0340-0003J (•)
Nut, 1/4-20, STL. 0500-0010J (•)
4 Screw, Flat, Skt Cap, 10-32 x 1/2, SST (Riser/DPU) 0240-0019J 8
5 Movement Riser 0277-0035C 2
6 Movement Assembly 0288-0035B 1
7 Screw, Bd Hd, 6-32 x 3/16, SST 0117-0013J 4
8 Stop Bracket 0288-0028C 1
9 Screw, Bd Hd, 4-40 x 3/16, SST 0117-0012J 1
10 Link Assembly 0288-0036B 1
11 Drive Arm Assembly 0288-0048B 1
12 Bracket, Terminal 1
10-Position (Standard) 0288-0029C
10-Position (For 14 AWG Terminal Blocks) 0288-1161C
12-Position S469-0060Z
13 Strip, Insulator 1
10-Position (Standard) 0038-1345T
10-Position (For 14 AWG Terminal Blocks) 0038-1379T
12-Position 0038-1351T
14 Screw, Rd Hd, 6-32 x 3/8, BRS, NP 0111-0015J 2
15 Terminal Block: 1
10-Position (Standard) 0038-0033T
10-Position, 14 AWG 0288-1160C
12-Position S469-0061Z
16 Scaleplate: 1
288A/288C/290D, White 0288-1014C (•)
288A/288C/290D, Black 0288-0031C (•)
288C Stainless Steel Version See Fig. 4-2
17 Snubber 0226-0028C 2
18 Pointer 1
White 0288-0030B
Black 0288-0031B

See notes, page 29.


26
Table 4-1. 288/290 General Parts List (Continued)
PART NO. PER
ITEM DESCRIPTION
(•)=model specific UNIT
19 Screw, Sl Fil., Hd., 4-40 X 3/16, SST 0114-0023J 4
20 Screw, Bezel 3
Standard 288A/288C 0181-1001C (•)
288C Stainless Steel Version See Fig. 4-2
21 Stud, Drive-Lok, Retaining, Bezel 0004-0005K (•) 1
(Standard 288A/288C Only)
22 Bezel Assembly (Bezel, Cover (lens) and Gasket) 1
288A/288C Standard, Bezel Assembly 0277-0018B (•)
Bezel (part of 0277-0018B) 0277-0029C (•)
Cover (Lens) (part of 0277-0018B) 0181-0038C (•)
Gasket (part of 0277-0018B) 0277-0026C (•)
288C Stainless Steel Version See Fig. 4-2
290D See Fig. 4-3
23 Switch and Plate Assembly: A/R
PVC Insulation (CSA)
Low, SPDT 290B-1100B
High, SPDT 290B-1006B
Low, DPDT S401-0053Z
High DPDT S401-0056Z
Low Relay (not available on 288C SST Version) 290B-1101B
High Relay (not available on 288C SST Version) 290B-1008B
Low #1, Independently Adjustable* S469-0058Z
High #1, Independently Adjustable* S469-0059Z
Low #2, Independently Adjustable* S469-0020Z
High #2, Independently Adjustable* S469-0043Z
Tefzel Insulation (Radiation Resistant)
Low, SPDT S666-0135Z
High, SPDT S666-0137Z
Low, DPDT S401-0110Z
High DPDT S401-0113Z
Low Relay (not available on 288C SST Version) S666-0189Z
High Relay (not available on 288C SST Version) S666-0196Z
Low #1, Independently Adjustable* S666-0135Z
High #1, Independently Adjustable* S666-0137Z
Low #2, Independently Adjustable* S666-0136Z
High #2, Independently Adjustable* S666-0138Z
Teflon Insulation (High Temp./Hermetic SW)
Low SPDT 0288-1132B
High SPDT 0288-1163B

*Note: Independently Adjustable Switches - SW#1 LOW (upper left);


SW#2 HIGH (upper right); SW#3 LOW (lower right); SW#4 HIGH
(lower left)
24 Washer, 1/2 OD x 0.14 ID, SST (1 per switch) 0317-0019C A/R
25 Washer, Spring (1 per switch) 0257-0019C A/R
26 Retaining Ring (1 per switch) 0087-0015T A/R

See notes, page 29.


27
Table 4-1. 288/290 General Parts List (Continued)

PART NO. PER


ITEM DESCRIPTION
(•)=model specific UNIT
27 Screw, Lock (1 per switch) A/R
Switch #1 and #2 0317-0012C
Switch #3 and #4 S469-0045Z
28 Link Plate (2 req. when both low and high sw used) 0288-0016C A/R
29 Washer, Flat, #3, SST (3 per Link Plate) 0003-0045K A/R
30 Grip Ring (1 per Link Plate) 0087-0011T A/R
31 Drive Plate Assembly A/R
Low Alarm 0288-0026B
High Alarm 0288-0027B
32 Screw, 3-48 x 3/16, SST, Binding Hd. (2 per Link A/R
0117-0007J
Plate)
33 Strap 0288-0017C 1
34 Washer, Split Lock, #8, SST 0003-0036K 1
35 Screw, 8-32 x 1/4, SST, Binding Hd. 0117-0016J 1
36 Switch Adjustment Index Assembly 1
Low Switch 0288-0028B
High Switch 0288-0029B
37 Screw, Bd. Hd., 3-48 x3/16, SST 0117-0007J 1
38 Switch Plate (1 per switch) A/R
Low Switch 0288-0022B
High Switch 0288-0023B
Low Switch, #1 (3 or 4 SW Config.) (Not Shown) S469-0056Z
High Switch #2 (3 or 4 SW Config.) (Not Shown) S469-0057Z
Low Switch, #3 (3 or 4 SW Config.) (Not Shown) S469-0033Z
High Switch, #4 (3 or 4 SW Config.) (Not Shown) S469-0035Z
39 Spacer, Switch Plate (used w/single High SW or 3 0258-0007C 1
Ind. Adj. SWs when #3 position is not used)
40 Switch and Wire Assemblies: A/R
Low #1, PVC Insulation (CSA) 290B-1102B
Low #1, Tefzel Insulation (Radiation Resistant) S666-0141Z
Low #2, PVC Insulation (CSA) S401-0016Z
Low #2, Tefzel Insulation (Radiation Resistant) S666-0142Z
High #1, PVC Insulation (CSA) 290B-1004B
High #1, Tefzel Insulation (Radiation Resistant) S666-0143Z
High #2, PVC Insulation (CSA) S401-0017Z
High #2, Tefzel Insulation (Radiation Resistant) S666-0144Z
Low Relay, PVC Insulation (CSA) 0288-0004B
Low Relay, Tefzel Insulation (Radiation Resistant) S666-0188Z
High Relay, PVC Insulation (CSA) 290B-1059B
High Relay, Tefzel Insulation (Radiation Resistant) S666-0197Z
Low Hermetic SW, Teflon Insulation (High Temp.) 0288-1131B
High Hermetic SW, Teflon Insulation (High Temp.) 0288-1162B
41 Alarm Actuator (Arm) (1 per switch) A/R
Low Switch 0251-0014B
High Switch 0251-0015B
42 Pivot Plate (1 per switch) 0251-0050C A/R

See notes, page 29.


28
Table 4-1. 288/290 General Parts List (Continued)

PART NO. PER


ITEM DESCRIPTION
(•)=model specific UNIT
43 Dielectric Shield (1 per switch) A/R
SPDT Switches 0290-1020C
DPDT Switches 0290-1021C
44 Washer, #2, Lock, Int. Tooth (2 per switch) 0003-0048K A/R
45 Nut, Hex, #2 (2 per switch) 0500-0030J A/R
46 Low Relay Assy., PVC Insulation (‚ = Non-CE Use Only) (Not Shown) 1
(Relays not available for 288C Stainless Steel version):
Note: Consult Factory for relays with Tefzel insulated wires.
Coil Voltage Wired SPDT Wired DPDT
6 VAC 290B-1018B 290B-1038B
12 VAC 290B-1019B 290B-1039B
24 VAC 290B-1020B 290B-1040B
‚115 VAC 290B-1021B 290B-1041B
‚230 VAC 290B-1022B 290B-1042B
6 VDC 290B-1023B 290B-1043B
12 VDC 290B-1024B 290B-1044B
24 VDC 290B-1025B 290B-1045B
‚110 VDC 290B-1026B 290B-1046B
47 High Relay Assy., PVC Insulation (‚ = Non-CE Use Only) (Not Shown) 1
(Relays not available for 288C Stainless Steel version):
Note: Consult Factory for relays with Tefzel insulated wires.
Coil Voltage Wired SPDT Wired DPDT
6 VAC 290B-1028B 290B-1048B
12 VAC 290B-1029B 290B-1049B
24 VAC 290B-1030B 290B-1050B
‚115 VAC 290B-1031B 290B-1051B
‚230 VAC 290B-1032B 290B-1052B
6 VDC 290B-1033B 290B-1053B
12 VDC 290B-1034B 290B-1054B
24 VDC 290B-1035B 290B-1055B
‚110 VDC 290B-1036B 290B-1056B
48 Screw, Relay Mount, 4-40 x 3/8, SST, Fl. Hd. 0112-1022J 2
(Qty. is per relay) (Not Shown)
49 Wire Nut (Relays Only) (2 per Relay) (Not Shown) 0109-0017T A/R
50 External Wire Leads See Table 4-2, A/R
on page 30
51 ID. Tag (Not Shown) 0199-0125C 1
52 Disc, Sealer, 1/2" (w/3M 468 Adhesive) 0192-1035T 1
53 Mounting Bracket (Refer to separate 224/224C DPU Manual.) A/R
Notes: A/R indicates as required. When ordering parts, specify serial number of instru-
ment. For 288C SST specific parts, see Fig. 4-2 and Table 4-3. For 290D
specific parts, see Fig. 4-3/4-4 and Table 4-4.
Scaleplate identification:
If the plate shows an SCR number, this will identify it. Otherwise, provide the follow-
ing information: Square root or linear graduations, Scale (e.g., 0-100, 25-0-100, etc.),
Number of graduations (linear scales only), and Data (e.g., PSID, inches water column,
etc.) If the plate has multiple scales or colors (other than black and white), consult factory
for pricing/description.

29
30

Table 4-2. External Wire Lead Part Numbers (Table 4-1, Item 50)
Length
Low #1 High #1 Low #2 High #2 Low Relay High Relay Ground
(feet)
PVC Insulation (CSA)
2.5 0288-0021B S655-0056B S401-0018Z S401-0093Z 290B-1083B 290B-1084B IT10-1063B
4.5 S655-0060B S655-0059B 291B-1014B
10 S401-0085Z S401-0086Z S401-0087Z S401-0088Z 291B-1015B
15 0288-1153B 0288-1154B S401-0172B S401-0173B 291B-1012B
20 S401-0122Z S401-0123Z S401-0124Z S401-0125Z 291B-1016B
40 0288-1156B 0288-1157B 291B-1017B
Tefzel Insulation (Radiation Resistant)
2.5 S666-0213Z S666-0140Z S666-0166Z S666-0167Z S594-0137Z S594-0138Z S666-0270B
10 S666-0190Z S666-0191Z S666-0241Z S666-0242Z S666-0271B
15 S666-0194Z S666-0195Z S666-0268B S666-0269B S666-0272B
20 S666-0201Z S666-0202Z S666-0203Z S666-0204Z S666-0273B
Teflon Insulation (High Temp./Hermetic Sw)
2.5 S822-0003Z S822-0004Z S822-0055B
Item#1 - Case Assembly)

(Item# 52, Figure 4-1)

Ref. Table 4-3 for part numbers.


w/3M 468 Adhesive
(not included with

Disc, Sealer, 1/2"

p/n: 0192-1035T
8

(See Fig. 4-1/Table 4-1)


Indicator/Switch Parts
For Common
NOTICE
1

DETAIL A
4X
7
(Bezel, Gasket, & Case)

6
DETAIL A

5
4
3
2

Figure 4-2. 288C (SST Version) Specific Parts Drawing

Table 4-3. 288C SST Version Specific Parts List (Ref. Fig. 4-2)
PER
ITEM DESCRIPTION PART NO.
UNIT
1 Case Assembly, SST S961-0052B 1
2 Screw, Bezel S961-0041C 9
3 Bezel S961-0037C 1
4 Gasket, Lens S961-0039C 1
5 Lens S961-0038C 1
6 Scaleplate 1
Black S961-0058C
White S961-0057C
7 Riser, Scaleplate 0279-0003C 4
8 Mounting Plate Assembly S961-0059B 1
— For Common Parts, refer to 288 (Fig. 4-1/Table 4-1)
Note: When ordering parts, please specify the serial number of the instrument.
31
32

3
Figure 4-3. 290D Specific Parts Drawing

11 9 4X

6
5 7
Screw 1
(Soc. Hd. Cap, 8-32 x 3/4, SST)
10
(0220-1089J) NOTICE 4
(included in item 11)
For Common
Indicator/Switch/Relay Parts
(See Fig. 4-1/Table 4-1)

O-Ring
(NBR 269-70 SHORE)
(0001-1157R)
(included in item 11)
Ref. Table 4-4 for part numbers.
Figure 4-4. 290D Detail Drawing
(Attaching DPU to Instrument)

Table 4-4. 290D Specific Parts List (Ref. Fig. 4-3.)


PER
ITEM DESCRIPTION PART NO.
UNIT
1 Case Assembly, Explosionproof IT10-1091B 1
2 Case (Body), Explosionproof (part of item 1) 0290-1018C 1
3 Screw, Hex Hd., 5/16-18 (part of item 1) 0116-1036J ?
4 Ground Wire, Green (part of item 1) IT10-1063B 1
5 Screw, Sl. Hex. Hd. (part of item 1) 0117-1012J 1
6 Int. Tooth Lock Washer (part of item 1) 0003-0066K 1
7 Ext. Tooth Lock Washer (part of item 1) 0003-0050K 1
8 Plate, Mounting, Assy. 0290-0002B 1
9 Riser, Scale Plate 0279-0003C 4
10 Screw, Mounting Plate, 8/32 x 3/16, Bin. Hd. 0117-0014J 3
11 Cover Assembly, Explosionproof IT10-1090B 1
12 Adapter, Explosionproof (Fig. 4-4, Note [2]) S655-0048C 1
— For Common Parts, refer to 288 (Fig. 4-1/Table 4-1)
Note: When ordering parts, please specify the serial number of the instrument.

33
SECTION 5 - INSTALL/DIMENSIONAL DRAWINGS
(Metric conversions are approximate.)

Figure 5-1. 288A/288C OD Drawing (Part 1 of 2)


34
(Metric conversions are approximate.)

Figure 5-1. 288A/288C OD Drawing (Part 2 of 2)

Table 5-1. 288A/288C Dimensions ‡


PRESSURE DIM. A* DIM. B* DIM. C* DIM. D* PRESSURE CONNECTION
HOUSING
RATING INCHES INCHES INCHES INCHES
MATERIAL TOP BOTTOM
PSI (MPa) (mm) (mm) (mm) (mm)

1/2 NPT** 1/4 NPT**


500 to 1500 4 2-15/16 2-41/64 2-5/8
ALL 9/16-18 UNF*** 9/16-18 UNF***
(3.4 to 10.3) (101.6) (74.6) (67.1) (66.7)
1/4 NPT 1/4 NPT
3000 to 1/2 NPT 1/2 NPT
4-3/8 3-1/4 2-49/64 2-3/4
10,000 ALL 7/16 MS 7/16 MS
(111.1) (82.6) (70.2) (69.9)
(20.7 to 68.9) 1/8 NPT 1/8 NPT
*Note: Dim. A ±0.03; Dim. B ±0.01; Dim. C ±0.01; Dim. D ±0.03
**Note: Can be reversed when ordered or rotated 180° in the field.
***Note: Suitable for use with Aminco fittings (American Inst. Co., Silver Springs, MD) or equiv..
All standard pipe fittings furnished by customer.
(Metric conversions are approximate.)


NOTICE: 288A version configurations limited to Nuclear applications; also
refer to separate 224/224C DPU Manual. For dimensions of Stainless
Steel version (SST) of 288C, contact Factory.

35
(Metric conversions are approximate.)

Figure 5-2. 290D OD Drawing (Part 1 of 2)


36
Figure 5-2. 290D OD Drawing (Part 2 of 2)

Table 5-2. 290D Dimensions


PRESSURE DIM. A* DIM. B* DIM. C* DIM. D* PRESSURE CONNECTION
HOUSING
RATING INCHES INCHES INCHES INCHES
MATERIAL TOP BOTTOM
PSI (MPa) (mm) (mm) (mm) (mm)

1/2 NPT** 1/4 NPT**


500 to 1500 4 2-15/16 2 1
ALL 9/16-18 UNF*** 9/16-18 UNF***
(3.4 to 10.3) (101.6) (74.6) (50.8) (25.4)
1/4 NPT 1/4 NPT
3000 to 1/2 NPT 1/2 NPT
4-3/8 3-1/4 2-1/4 1-1/8
10,000 ALL 7/16 MS 7/16 MS
(111.1) (82.6) (57.2) (28.6)
(20.7 to 68.9) 1/8 NPT 1/8 NPT
*Note: Dim. A ±0.03; Dim. B ±0.01; Dim. C ±0.01
**Note: Can be reversed when ordered or rotated 180° in the field.
***Note: Suitable for use with Aminco fittings (American Inst. Co., Silver Springs, MD) or equiv..
All standard pipe fittings furnished by customer.
(Metric conversions are approximate.)

(Metric conversions are approximate.)

37
(Blank Page)

38
(Blank Page)

39
Product Warranty
A. Warranty
PRIME Measurement Products, LLC warrants that at the time of shipment the products manu-
factured by PRIME Measurement Products, LLC and sold hereunder will be free from defects In
material and workmanship, and will conform to the specifications furnished by or approved by
PRIME Measurement Products, LLC.
B. Warranty Adjustment
(1) If any defect within this warranty appears, Buyer shall notify PRIME Measurement Prod-
ucts, LLC immediately.
(2) PRIME Measurement Products, LLC agrees to repair or furnish a replacement for, but
not install, any product which within one (1) year from the date of shipment by PRIME
Measurement Products, LLC shall, upon test and examination by PRIME Measurement
Products, LLC, prove defective within the above warranty.
(3) No product will be accepted for return or replacement without the written authorization
of PRIME Measurement Products, LLC. Upon such authorization, and in accordance with
instructions by PRIME Measurement Products, LLC, the product will be returned shipping
charges prepaid by Buyer. Replacements made under this warranty will be shipped prepaid.
C. Exclusions from Warranty
(1) THE FOREGOING WARRANTY IS IN LIEU OF AND EXCLUDES ALL OTHER EXPRESSED
OR IMPLIED WARRANTIES OF MERCHANTABILITY, OR FITNESS FOR A PARTICULAR
PURPOSE, OR OTHERWISE.
(2) Components manufactured by any supplier other than PRIME Measurement Products, LLC shall
bear only the warranty made by the manufacturer of that product, and PRIME Measurement
Products, LLC assumes no responsibility for the performance or reliability of the unit as a whole.
(3) "In no event shall PRIME Measurement Products, LLC be liable for indirect, incidental, or
consequential damages nor shall the liability of PRIME Measurement Products, LLC arising in
connection with any products sold hereunder (whether such liability arises from a claim based
on contract, warranty, tort, or otherwise) exceed the actual amount paid by Buyer to PRIME
Measurement Products, LLC for the products delivered hereunder."
(4) The warranty does not extend to any product manufactured by PRIME Measurement Products,
LLC which has been subjected to misuse, neglect, accident, improper installation or to use in
violation of instructions furnished by PRIME Measurement Products, LLC.
(5) The warranty does not extend to or apply to any unit which has been repaired or altered at any
place other than at PRIME Measurement Products, LLC factory or service locations by persons
not expressly approved by PRIME Measurement Products, LLC.

Disclaimer
PRIME Measurement Products, LLC makes no warranty of any kind with regard to this manual, materials attached to or provided with, includ-
ing, but not limited to, implied warranties of merchantability and fitness for a particular purpose. PRIME Measurement Products, LLC shall
not be liable for errors contained in this manual, including materials attached to, or provided with, or for incidental or consequential damages
in connection with the furnishing, performance, or use of this information. The information contained in this manual (along with attached
or provided materials) is subject to change without notice. The contents of this manual and all attached/provided materials may be revised
without prior notice.

Product Brand
Barton® brand products are manufactured and sold by PRIME Measurement Products, LLC (formerly Barton Instrument Systems) in Nuclear
and Government Markets. Former Barton® brand Industrial Products are manufactured and sold by PRIME Measurement Products, LLC as
PRIME products. Barton® is a trademark of NuFlo Technologies, Inc..

PRIME Measurement Products, LLC


(Formerly Barton Instrument Systems)
900 S. Turnbull Canyon Rd.
City of Industry, CA 91745 USA
(626) 961-2547
On the Web at www.prime-measurement.com