TESTBEAM SLOW CONTROL OCTOBER

2014

Eugenia Price
University of Bristol

1
2

•  Open remote desktop connection to PCLHCB66

•  On the Desktop open the folder ‘Telecope Motion Control rev 2.0’

•  Open the vi ‘Telescope Motion Control.vi’

•  Select Data Acquisition and Write Data To File

•  Turn on T1, T2 and RH1

•  Run the VI

Eugenia Price
3

Monitoring
•  The Temperature and Humidity
can be monitored via the charts
to the Right of the VI

•  The Values are also displayed to


the left of the temperature chart
4

•  Turn on PS1 (Telescope),


PS2(Telescope), PS3 (DUT)

•  Select whether the PS3 DUT


bias is +ve or –ve (green +ve red
–ve), it is not possible to set the
voltage with a different sign to
the chosen bias

•  Set ‘default value’ of power


supplies to 1V (this is an initial
value to be set and cannot
exceed ±10V, if the DUT Bias is
–ve set this to -1V)

Eugenia Price
5

•  Login and check the High


voltage Limits

•  This section is password


protected

•  The hardware limit should


be set to one of the
accepted values for the
Keithleys

•  NOTE: To change the


hardware limit ramp down,
turn off output, set value and
turn back on (the OVP is set
when first initialized)

•  The software limits can be


set to any value

•  This section also protects


homing of the stages
Eugenia Price
6

•  Press ‘Turn on Selected Power Supplies’

•  To ramp up the voltage enter the desired


value into ‘set voltage’ for the
corresponding power supply and press
‘set PSX’

•  When ramping up ensure that the set


value is reached. If it is not then pre ‘set
PSX’ again

•  Use the reset buttons to reset the power


supplies this ramps down the voltage to
0, resets the instrument and then resets
to the ‘default voltage’ value (to be safe
you can set to zero before pressing
‘reset’)

Eugenia Price
7

•  To see the actual Current and


Voltage measurement see the
‘Voltages’ Section, the large
circular indicator shows if the
power supply is on or off

•  This also displays the stage


positions

•  The Current can also be


monitored using the centre graph
of the right of the vi

Eugenia Price
8

•  To ramp down the power


supplies set the voltage as
previously to 0.

•  Again ensure the voltage


reaches zero and press set
again if it is not reached

•  Once all Power supplies are


ramped down select ‘’Stop High
Voltage Output”

•  To turn the high voltage back on,


as previously, check limits, select
the power supplies and press
‘Turn on Selected Power
Supplies’

Eugenia Price
9

•  To Move the Telescope enter the


desired value, select steps/ mm
and press move

•  To Home login to the protected


tab

•  The current position can be seen


by looking at the web cam image

Eugenia Price
10

•  Set the position and angles of


the Centre Stage and Outer
Stage using the windows to the
left of the VI

•  To Home these stages again


login to the protected tab

•  To Terminate Movement press


‘Stop Movements’

•  When you are finished press


‘Stop Program’ to terminate the
entire vi

Eugenia Price
11

Power supply on after vi terminated


•  If the Power supplies do not display ‘output off’ then run
the following vi’s to ramp them down and turn the output
off

•  Go to Desktop/Eugenia/Telescope_Slow_Control_2014

•  Run the FAULT_RUN_PS1...vi for the given power supply