You are on page 1of 3

26 tccc NoRTH

AMER|CA
voLutvlE
22 NUtulBER
I JANUARY
2004 ww. chro matog ra phyon I i ne.com

Broad Peaks

eala that are too broad can mean The literaturethat comeswith a column
that analysts are nor using their rypicallyreportsplatenumbersof approxi-
Iiquid chromarography (LC) mately80,000/mfor 5-pm d, mediaand
columns very efficiently. Narrow peaks 100,000/mfor 3-pm do media.Don'rlook
can translate into faster runs, because less for this kind of performancewith your
time is necessaryto obtain baselinesepara- method,though.The manufacturers t€st
tion. This montht "LC Tioubleshooting" columns under very carefully controlled
takes a look at how to determine if peaks
are broader than they should be and dis-
cussessome of the most common system- &W&*p*#{iips:$s*s*$.#*ww
msx
related causesofpeak broadening. For this
Do your peaks
iliscussion, I'll focus on reversed-phaseiso- $"#p*r*& *&m#ws*mm
needto go on a diet? cratic separations. dp*'mm#m$ss:p
es:*:arw$wws
.*w
What ls Fat? tr#tre#gffi #ss?d#$-,ss?*
{3f
As we all know from our interactions with
others or examinations of ourselves in a
mp;rrxs*sx"
mirror, classifring a person as being over-
weight involves opinion more than quan- conditions with test compounds such as
tification in most cases.\(hat consdtutes toluene and methyl b..rro"t., which pro,
an LC peak that is too broad also involves a vide ideal chromatographic behavior. A
certain amount of opinion. However, some method for analysisof a pharmaceutical
simple quantitative measuresof peak per- compound in plasma, a pesticide in hog fat,
formance generally serve befter than a sim- or a synthetic intermediate will not behave
ple visual examination of the chromato- in such an ideal manner. As a guide for rea-
gram to determine if a peak is too broad. sonable chromatographic performance with
The peak widrh is a poor measurement a real compound, I use the following esti-
of a chromatographic peak, becausethe mate:
peak width increasesproportionally with
the retention time (rq) for isocratic separa- N * 3000Ltdo t3l
tions. The first step in quantifying the
broadnessof a peak should be to measure where Z is the column length in cenrime-
the plate number (.A/) for the peak of inter- ters and dois the packing parricle diameter
est. The plate number can be measuredin in micrometers. If I observeplate numbers
one or two ways: within approximately 20o/oof this esrimate,
I dont worry much about column perfor-
N: 5.54 (tyl*o)2 tll mance.A l5-cm long, 5-pm doClS col-
umn probably won't generatemore than
approximately 9000 plates, even though its
manufacturer might report .Ay'valuesof
N: 16(txl*)2 t2l approximately 12,000.

where wg.5 and w are the peak width ls Tailing a Problem?


measured at half the peak height and at Another m€asurement that I make before
the baseline between tangents drawn to drawing any conclusions about excessive
the sides of the peak, respectively. I prefer peak broadening is the amounr of peak tail-
lohn W Dolan the half-height method becauseit is easier ing. The two most common measuresof
LC Troubleshooting Editor especiallyiftwo adjacent peaks are not peak tailing are the USP (U.S. Pharma-
baselineresolved.Most dara sysrems copeia) tailing factor (76) and the peak
should be able to determine l/usinq either asymmeuy factor (A,). Z6is calculated as
method.
28 rcccNoRrH
AMERiCA
voLUI\tE
22NUN1BER
1 JANUARY
2004
the ratio of the peak width to twice the
front half-width of the peak as measuredat
where V2 is the square of the peak volume plumbing,
r,r,""rU.r,ffili"fi;
(in milliliter$ and the subscripts indicare extracolumn contributions from the tubing,
5o/oof the peak height. l, is the back half- the contributions from the column. the fittings, or detector.Ifyou are suspicious,
width divided by the fiont half-width of injector, the tubing, the fittings, and the make sure to minimize excesstube lengths
the peak measuredat l0o/o of its height. detector. The total peak volume is the and keep the tubing diameter no larger
Generally, the tailing factor is used in the square root of the sum of the volumetric than 0.007-in. i.d.
pharmaceutical industry and the asymme- peak variancesof the various components
try factor for other applications - ir of equation 4. The contributions from the
doesnt really marter from a technical stand- injector, tubing, fittings, and detector (and #exsf ,i*s g&# ffi#d#dk
point which method you use, as long as sometimes a term is added for the time
you consistently use the same method. constant or data rate) are called extracol- $dcxf*#s??# drrxps **ti*&
Once again, the column manufacturer's umn effects, becausethey are factors out- s*w;xl$s,s'r*r *q*$*'i{}rr
test compounds and real samplesare differ, side the column. 'i*v'twx;,.s# #d&
ent. The test compounds exhibit little, if Consider two scenarios:First, the volume rrd?s,#$f#r"
any tailing, whereas it is rare to have a per- contribution of the column is large com-
fectly symmetric peak with a real sample.
?"s',{:&jfg;,*?*s
es*$*srp?#,'*s
pared with the extracolumn contributions.
As long as the tailing facror or the asymme- In this case,the percentagecontribution to ##,flrdlf'# f4| $#ffidr$,f#f-
try factor is no more than approximately the overall peak volume by the extracolumn
1.5, it generally is not worth trying to
#,$-
#p#.#${'}/{},'${.s,flffi
effects is small and usually can be ignored.
improve. Larger values of peak tailing could Second, the extracolumn effects are much
mean that unwanted secondary interactions larger and are significant when compared Contrast this column with the 50 mm X
are occurring. Techniques to reduce peak with the contributions by the column. So, 2.1 mm, 3-p"m d, column that generates
tailing have been discussedin previous "LC narrower peaks are more strongly affected peals lessthan one-tenth the volume of the
toubleshooting" columns (for example, see by extracolumn efFectsthan broader peaks. larger column. A tube end that is poorly
reference l). In an isocratic chromatogram, the earlier- seatedin a fitting or the accidental use of a
eluted peala are narrower than later-eluted piece of 0.01O-in.i.d. tubing can dramati-
Extracolumn Effects ones (Figure 1). k follows thar exrracolumn cally increasethe peak width.
Let's assumethat peak tailing is acceptable. effects negatively infuence peaks with If you arent convinced, make some trial
You obtained a plate number for a peak smaller retention dmes more than they do calculadons using equation 4 and add 10
that was lessthan approximately 80% of those with larger retention times. If you pL of extracolumn volume to the column-
the value estimared from equation 3. If atl observed more broadening (smaller values generatedpeak volumes shown in Table I.
the peals in the chromatogram are well of ly') earlier in the chromatogram, exrra- The small peak volumes generatedby 2.1-
separated,it still might not be worth column effects could be responsible. and 1.0-mm i.d. columns are good exam-
addressingthe problem. If resolution has ples of why it is difiicult to obtain column
suffered or if you would like to improve Column Size plate numbers that are close to those
resolution so that you can speed the separa- An additional paramerer that plays a very reported by the manufacturer with test
tion by obtaining the same resolution in imponant role in the infuence of extracol- compounds. It is easyto seehow chro-
lesstime, then it is time to look more umn effects is the column size.Just as the matographers can get comfortable using
closely at the chromatogram. Measure the peak volume drops with smaller rerenrion 150 mm x 4.5 mm columns for routine
plate number for peaks eluted at the begin- times, so do smaller volume columns gener- work and get into trouble by switching to
ning, middle, and end of the chromato- ate smaller peak volumes. Refer to Thble I 50 mm X 2.1 mm columns, unlessthey
gram. If the plate number improves with to seethis influence. Ive calculated peak take care to minimize the extracolumn con-
retention time, extracolumn effects are a volumes based upon equation 3 for peaks tributions from the tubing, fittings, and
likely problem source. Exrracolumn effects with retention factors (2) of 1 and 5 for detector.
are refected in the peak volume: three column configurations. The 150 mm
x 4.6 mm,5-vm d, column that most of What about the Injector?
V?",: VLt + Vl^i+ V?"a+ rt1 us use for routine LC-W work generates The sample injection processcan con-
t' /f2i t ' L r / "z d e r taj
relatively large peak volumes, and, unless tribute to peak broadening in two different
you are quite sloppy with the system ways: First, if too large a sample volume is

150mmX 4.6mm,5-pmdo 1.5 9000 12s 380


50mmX 2.1mm,5-pmdo 0.1 3000 t5 45
50mm X 2.1mm,3-pmdo 0.1 5000 10 35
* C o l u m nd e a dv o l u m e .
Time
tPlate numbersobtainedusingestimateof equation3.
f Peakvolumefor peakswith k = 1.
Figure 1: lsocraticchromatogramshowing $Peakvolumefor peakswith k : 5.
broadeningof laterpeaks.
30 rcccNoRTH
AMERICA
voLUt\4E
22NU|\.4BER
i JAiluARy
2004 natographyonline.com

injected, it can causeincreasedpeak the injection along the top of the column approximately 3-s wide. To obtain 10-20
widths. Second, if the injection solvent is and causesband broadening and some- data points acrossthis peak, the data rate
stronger than the mobile phase, it can times tailing, distortion, or splitting of all would have to be 4 Hz or more.
wash some of the sample molecules down the peaks in the chromatogram (see the
the column until rhe solvent is diluted by discussion in reference2 for an example). Conclusions
the mobile phase. Again, the injection volume and injection Ive considered only a few ofthe sources
For the first injection-related contribu- solvent strength musr be balanced. For ofband broadening in this "LC Tiouble-
tion to band spreading, think of an example, a 5-p,L injection of sample in a shooting" column, but these variables are
extreme case in which the sample volume strong-solvent mobile phase is unlikely to ones that can infuence peak broadening in
is the same as the column volume and in causemuch problem with a 150 mm X any separation. Control these variables and
which the mobile phase is used as the 4.6 mm column, but a 20-pL injection you can feel fairly confident that you are
injection solvent. The first sample mole- might. The best advice here is to deter- not doing anything stupid that causes
cules injected would migrate a significant mine the effect empirically. \(hen the unwanted peak broadening.
distance down the column before the last injection solvent is stronger than the Another easily controlled source of band
of the sample molecules were injected. mobile phase, double and halve the iniec- broadening is column temperature. If the
This situation would generatean extremely tion volume and seeif it has any pracrical column temperature is constant, both axi-
broad band. The opposite would be an effect on the separation and then you can ally and radially, you shouldnt have
infinitely small injection size in which all decide if modification of the method is unwanted temperature-related peak distor-
sample molecules reach the column simul- necessary. tion. Use a column oven and make sure
taneously.Realistic injection sizesare On the other hand, if the injection sol- the solvent at the inlet to the column is
somewhere between these rwo exffemes. vent is weaker than the mobile phase,you within +5'C of the column temperarure,
How much sample can be injected with- usually can inject much more than the and you should be safe. (Reference3 has a
out deleterious peak broadening? lf a 5o/o l5% guideline. This larger injection vol- casestudy of extracolumn and temperature
loss in resolution is acceptable, approxi- ume is possible becausethe migration of effects.)
mately l5o/o of the peak volume of the first peaks in the injection solvent is slower Ifyou still observeexcessivelybroad
peak ofinterest can be injected, ifthe than in the mobile phase,and it has the peaks after these sourcesare eliminated,
mobile phase is used as the injection sol- effect of compressing the peak at the top of look to chemical interactions such as exces-
vent. Thble I shows that this percenrage the column during injection. This tech- sive interactions with surface silanol
would allow approximately 20 pL for the nique of on-column concentration can be groups, slow diffirsion (especiallya prob-
150 mm x 4.6 mm column, but only a handy tool to enable the injection ofa lem with high molecular weight samples),
2 p,L for the 50 mm X 2.1 mm column. problematic sample. or some other sample-relatedsource.
Many of the LC-tandem mass spectrome- I remember a method in which the
chromatographer needed to inject 50 pL References
of sample that was extracted into metha- (1) J.\7. Dolan,LCGC2T(7),612-616 (2003).
ffix*"s.m
&mexdsptr###fr*?W nol, but the mobile phasewas 50o/ometha- (2) C. HawkinsandJ.'$7.Dolan, LCGC 2l(t2),
1134-rr38Q00r.
tr#ess##&y *&xs ff##& nol. By diludng the sample fourfold with
(3) R.M. Minikis andJ.ltr(r.
Dolan,LCGC2t(rI),
water, it was possible to inject the same 1050-1054 Q003\.
w€p$*s$??#$xrwryws€wd-, sample mass in 200 pL and avoid the hor-

&ax**'y6xscsf rible peak broadening encounrered when


50 pL of 100%omethanol was used as the
JohnW. Dolan
"LC Troubleshoot-
g&rmwmfmgnr.w,rxxs
&* erw injection solvent. Again, an empirical test ing" editor lohn
should help guide you toward a suitable W. Dolan is vice-
€$d*$3f
fwxrepwffigd- injection volume in a weak solvent. presidentof BAS|
r#$'s$a.fl#dpwm&w- Northwest Labora-
tory of McMinnville,
And Finally, the Data System
Oregon;a training
I mentioned that sometimes equation 4 consultantfor Rheo-
try (MS-MS) methods in my laboratory is written to include a band-spreading dyne LLC,the LC
that use 50 mm X 2.1 mm columns also term causedby the data sysrem. If the data ResourcesTraining
call for 5-10 pL injections. Extra band rate is too slow, an insufficient number of Group, of Walnut Creek,California;and
spreading caused by this peak volume is a member of LCGC'seditorial advisoryboard.
data points would be gathered, and peak
Direct correspondence about this column
expected, but rypical chromatograms have broadening could appear.The general rule to "LC Troubleshooting,"LCGC,Woodbridge
only two well-resolved peafts; the addi- is that 10-20 data points should be gath- CorporatePlaza,485 Route 1 South, Building
tional selectiviry of the mass spectrometer ered acrossthe peak. If the 150 mm x F, FirstFloor, lselin,Nl 08830,e-mail
enables users to obtain acceptable quantita- 4.5 mm column of Thble I was operated at John.DoIan@BioanaIyti cal.com.
For an ongoing discussionof LCtrouble-
tive results. I ml/min, the first peak would be approx-
shooting with John Dolan and other chro-
The second injection-related contribu- imately 0.125-min or approximately 7-s matograp he rs,visit th e Chromatogr aphy
tion to band spreading is the injection sol- wide. A data rate of 2 Hz or more should Forum discussiongroup at http:llwww.
'$7hen
vent. a srrong injection solvent be sufficient for this case.The 50 mm x chromforum.com.
washessome of the sample molecules 2.1 mm, 3-p,m d, column would use a
down the column and dilutes the sample fow rate of 0.2 ml/min for the same lin-
in the mobile phase, this processsmears ear velocity and would generatea peak

You might also like