Scanning tunneling microscope

A scanning tunneling microscope (STM) is an instrument for imaging surfaces at the atomic level. Its
development in 1981 earned its inventors, Gerd Binnig and Heinrich Rohrer (at IBM Zürich), the Nobel
Prize in Physics in 1986.[1][2] For a STM, good resolution is considered to be 0.1 nm lateral resolution and
0.01 nm (10 pm) depth resolution.[3] With this resolution, individual atoms within materials are routinely
imaged and manipulated. The STM can be used not only in ultra-high vacuum but also in air, water, and
various other liquid or gas ambients, and at temperatures ranging from near zero kelvin to over
1000 °C.[4][5]
Image of reconstruction
STM is based on the concept of quantum tunneling. When a conducting tip is brought very near to the
on a clean gold (100)
surface to be examined, a bias (voltage difference) applied between the two can allow electrons to tunnel surface
through the vacuum between them. The resulting tunneling current is a function of tip position, applied
voltage, and the local density of states (LDOS) of the sample.[4] Information is acquired by monitoring the
current as the tip's position scans across the surface, and is usually displayed in image form. STM can be a
challenging technique, as it requires extremely clean and stable surfaces, sharp tips, excellent vibration control,
[6]
and sophisticated electronics, but nonetheless many hobbyists have built their own.

The silicon
Contents atoms on the
surface of a
1 Procedure crystal of silicon
2 Instrumentation carbide (SiC).
3 Other STM related studies Image obtained
using an STM.
4 Principle of operation
5 Early invention
6 See also
7 References
8 Further reading
9 External links
An STM image of a single-walled
carbon nanotube

Procedure
First, a voltage bias is applied and the tip is brought close to the sample by coarse
sample-to-tip control, which is turned off when the tip and sample are sufficiently
close. At close range, fine control of the tip in all three dimensions when near the
sample is typically piezoelectric, maintaining tip-sample separation W typically in the
4-7 Å (0.4-0.7 nm) range, which is the equilibrium position between attractive
(3<W<10Å) and repulsive (W<3Å) interactions.[4] In this situation, the voltage bias
will cause electrons to tunnel between the tip and sample, creating a current that can Play media
be measured. Once tunneling is established, the tip's bias and position with respect to Scanning tunneling microscope operating
the sample can be varied (with the details of this variation depending on the principle

experiment) and data are obtained from the resulting changes in current.

If the tip is moved across the sample in the x-y plane, the changes in surface height and density of states causes changes in current. These
changes are mapped in images. This change in current with respect to position can be measured itself, or the height, z, of the tip
corresponding to a constant current can be measured.[4] These two modes are called constant height mode and constant current mode,
respectively. In constant current mode, feedback electronics adjust the height by a voltage to the piezoelectric height control mechanism.[7]
This leads to a height variation and thus the image comes from the tip topography
across the sample and gives a constant charge density surface; this means contrast on
the image is due to variations in charge density.[8] In constant height mode, the voltage
and height are both held constant while the current changes to keep the voltage from
changing; this leads to an image made of current changes over the surface, which can
be related to charge density.[8] The benefit to using a constant height mode is that it is
faster, as the piezoelectric movements require more time to register the height change
in constant current mode than the current change in constant height mode.[8] All
images produced by STM are grayscale, with color optionally added in post-
processing in order to visually emphasize important features.
A close-up of a simple scanning tunneling
In addition to scanning across the sample, information on the electronic structure at a microscope head using aplatinum–iridium
given location in the sample can be obtained by sweeping voltage and measuring tip.

current at a specific location.[3] This type of measurement is called scanning tunneling


spectroscopy (STS) and typically results in a plot of the local density of states as a
function of energy within the sample. The advantage of STM over other measurements of the density of states lies in its ability to make
extremely local measurements: for example, the density of states at an impurity site can be compared to the density of states far from
impurities.[9]

Framerates of at least 25 Hz enable so called video-rate STM.[10][11] Framerates up to 80 Hz are possible with fully working feedback that
adjusts the height of the tip.[12] Due to the line-by-line scanning motion, a proper comparison on the speed requires not only the framerate,
but also the number of pixels in an image: with a framerate of 10 Hz and 100x100 pixels the tip moves with a line frequency of 1 kHz,
whereas it moves with only with 500 Hz, when measuring with a faster framerate of 50 Hz but only 10x10 pixels. Video-rate STM can be
used to scan surface diffusion.[13]

Instrumentation
The components of an STM include scanning tip, piezoelectric controlled height and
.[7]
x,y scanner, coarse sample-to-tip control, vibration isolation system, and computer

The resolution of an image is limited by the radius of curvature of the scanning tip of
the STM. Additionally, image artifacts can occur if the tip has two tips at the end
rather than a single atom; this leads to “double-tip imaging,” a situation in which both
tips contribute to the tunneling.[3] Therefore, it has been essential to develop processes
for consistently obtaining sharp, usable tips. Recently, carbon nanotubes have been
used in this instance.[14]

The tip is often made of tungsten or platinum-iridium, though gold is also used.[3]
Tungsten tips are usually made by electrochemical etching, and platinum-iridium tips Schematic view of an STM
by mechanical shearing.[3]

Due to the extreme sensitivity of tunnel current to height, proper vibration insulation or an extremely rigid STM body is imperative for
obtaining usable results. In the first STM by Binnig and Rohrer, magnetic levitation was used to keep the STM free from vibrations; now
mechanical spring or gas spring systems are often used.[4] Additionally, mechanisms for reducingeddy currents are sometimes implemented.

Maintaining the tip position with respect to the sample, scanning the sample and acquiring the data is computer controlled.[7] The computer
may also be used for enhancing the image with the help ofimage processing[15][16] as well as performing quantitative measurements.[17][18]

Other STM related studies


Many other microscopy techniques have been developed based upon STM. These include photon scanning microscopy (PSTM), which uses
an optical tip to tunnel photons;[3] scanning tunneling potentiometry (STP), which measures electric potential across a surface;[3] spin
polarized scanning tunneling microscopy (SPSTM), which uses a ferromagnetic tip to tunnel spin-polarized electrons into a magnetic
sample,[19] and atomic force microscopy(AFM), in which the force caused by interaction between the tip and sample is measured.
Other STM methods involve manipulating the tip in order to change the topography of
the sample. This is attractive for several reasons. Firstly the STM has an atomically
precise positioning system which allows very accurate atomic scale manipulation.
Furthermore, after the surface is modified by the tip, it is a simple matter to then
image with the same tip, without changing the instrument. IBM researchers developed
a way to manipulate xenon atoms adsorbed on a nickel surface.[3] This technique has
been used to create electron "corrals" with a small number of adsorbed atoms, which
allows the STM to be used to observe electron Friedel oscillations on the surface of
the material. Aside from modifying the actual sample surface, one can also use the
STM to tunnel electrons into a layer of electron beamphotoresist on a sample, in order Nanomanipulation via STM of aself-
to do lithography. This has the advantage of offering more control of the exposure assembled organic semiconductor
monolayer (here: PTCDA molecules) on
than traditional electron beam lithography. Another practical application of STM is
graphite, in which the logo of theCenter
atomic deposition of metals (gold, silver, tungsten, etc.) with any desired (pre-
for NanoScience (CeNS), LMU has been
programmed) pattern, which can be used as contacts to nanodevices or as nanodevices written.
themselves.

Variable temperature STM was used to investigate temperature dependency of molecular rotations on
single crystalline surfaces.[20] Rotating molecules appear blurred compared to non-rotating ones.

Recently groups have found they can use the STM tip to rotate individual bonds within single molecules.
The electrical resistance of the molecule depends on the orientation of the bond, so the molecule
effectively becomes a molecular switch.

Image of a graphite
Principle of operation surface at an atomic
level obtained by an
Tunneling is a functioning concept that arises from quantum mechanics. Classically, an object hitting an STM.
impenetrable barrier will not pass through. In contrast, objects with a very small mass, such as the
electron, have wavelike characteristics which permit such an event, referred to as
tunneling.

Electrons behave as beams of energy, and in the presence of a potential U(z), assuming
1-dimensional case, the energy levels ψn(z) of the electrons are given by solutions to
Schrödinger’s equation,

where ħ is the reduced Planck’s constant, z is the position, and m is the mass of an
electron.[4] If an electron of energy E is incident upon an energy barrier of height U(z), The first STM produced commercially,
1986.
the electron wave function is a traveling wave solution,

where

if E > U(z), which is true for a wave function inside the tip or inside the sample.[4] Inside a barrier, E < U(z) so the wave functions which
satisfy this are decaying waves,

where
quantifies the decay of the wave inside the barrier
, with the barrier in the +z direction for .[4]

Knowing the wave function allows one to calculate the probability density for that electron to be
Condensed matter
found at some location. In the case of tunneling, the tip and sample wave functions overlap such
experiments
that when under a bias, there is some finite probability to find the electron in the barrier region and
even on the other side of the barrier.[4] Let us assume the bias is V and the barrier width is W. This
probability, P, that an electron at z=0 (left edge of barrier) can be found at z=W (right edge of
barrier) is proportional to the wave function squared,

.[4]
ARPES
If the bias is small, we can let U − E ≈ φM in the expression for κ, where φM, the work function,
ACAR
gives the minimum energy needed to bring an electron from an occupied level, the highest of Neutron scattering
which is at the Fermi level (for metals at T=0 kelvins), to vacuum level. When a small bias V is X-ray spectroscopy
applied to the system, only electronic states very near the Fermi level, within eV (a product of
Quantum oscillations
Scanning tunneling microscopy
electron charge and voltage, not to be confused here with electronvolt unit), are excited.[4] These
excited electrons can tunnel across the barrier. In other words, tunneling occurs mainly with
electrons of energies near the Fermi level.

However, tunneling does require that there be an empty level of the same ener
gy as the
electron for the electron to tunnel into on the other side of the barrier. It is because of
this restriction that the tunneling current can be related to the density of available or
filled states in the sample. The current due to an applied voltage V (assume tunneling
occurs sample to tip) depends on two factors: 1) the number of electrons between Ef
and eV in the sample, and 2) the number among them which have corresponding free
states to tunnel into on the other side of the barrier at the tip.[4] The higher the density
of available states the greater the tunneling current. When V is positive, electrons in
the tip tunnel into empty states in the sample; for a negative bias, electrons tunnel out
of occupied states in the sample into the tip.[4]

Mathematically, this tunneling current is given by

One can sum the probability over energies between Ef − eV and Ef to get the number
of states available in this energy range per unit volume, thereby finding the local
density of states (LDOS) near the Fermi level.[4] The LDOS near some energy E in an A large scanning tunneling microscope, in
the labs of the London Centre for
interval ε is given by
Nanotechnology

and the tunnel current at a small bias V is proportional to the LDOS near the Fermi level, which gives important information about the
sample.[4] It is desirable to use LDOS to express the current because this value does not change as the volume changes, while probability
density does.[4] Thus the tunneling current is given by
where ρs(0,Ef) is the LDOS near the Fermi level of the sample at the sample surface.[4] This current can also be expressed in terms of the
LDOS near the Fermi level of the sample at the tip surface,

The exponential term in the above equations means that small variations in W greatly influence the tunnel current. If the separation is
[8]
decreased by 1 Å, the current increases by an order of magnitude, and vice versa.

This approach fails to account for the rate at which electrons can pass the barrier. This rate should affect the tunnel current, so it can be
treated using the Fermi's golden rule with the appropriate tunneling matrix element. John Bardeen solved this problem in his study of the
metal-insulator-metal junction.[21] He found that if he solved Schrödinger’s equation for each side of the junction separately to obtain the
wave functions ψ and χ for each electrode, he could obtain the tunnel matrix, M, from the overlap of these two wave functions.[4] This can be
applied to STM by making the electrodes the tip and sample, assigning ψ and χ as sample and tip wave functions, respectively, and
[4]
evaluating M at some surface S between the metal electrodes, where z=0 at the sample surface and z=W at the tip surface.

Now, Fermi’s Golden Rule gives the rate for electron transfer across the barrier, and is written

gy.[4] The tunnel matrix element, given by


where δ(Eψ–Eχ) restricts tunneling to occur only between electron levels with the same ener

gy.[4]
is a description of the lower energy associated with the interaction of wave functions at the overlap, also called the resonance ener

Summing over all the states gives the tunneling current as

where f is the Fermi function, ρs and ρT are the density of states in the sample and tip, respectively.[4] The Fermi distribution function
describes the filling of electron levels at a given temperature .T

Early invention
An earlier, similar invention, the Topografiner of R. Young, J. Ward, and F. Scire from the NIST,[22] relied on field emission. However,
Young is credited by the Nobel Committee as the person who realized that it should be possible to achieve better resolution by using the
tunnel effect.[23]

See also
Microscopy Electrochemical scanning tunneling microscope
Scanning electron microscope Atomic force microscope
Scanning probe microscopy Electron microscope
Scanning tunneling spectroscopy Spin polarized scanning tunneling microscopy

References
1. Binnig, G.; Rohrer, H. (1986). "Scanning tunneling microscopy". IBM Journal of Research and Development. 30 (4): 355–
69.
2. Press release for the 1986 Nobel Prize in physics(http://nobelprize.org/nobel_prizes/physics/laureates/1986/press.html)
3. C. Bai (2000). Scanning tunneling microscopy and its applications(https://books.google.com/?id=3Q08jRmmtrkC&pg=P
A34
5). New York: Springer Verlag. ISBN 3-540-65715-0.
4. C. Julian Chen (1993).Introduction to Scanning Tunneling Microscopy (http://www.columbia.edu/~jcc2161/documents/STM
_2ed.pdf) (PDF). Oxford University Press.ISBN 0-19-507150-6.
5. SPECS. "STM 150 Aarhus - High Stability Temperature Control" (http://www.specs.de/cms/upload/PDFs/ApplNotes/STM/A
Note_HSTC-STM.pdf)(PDF). specs.de. Retrieved 23 February 2017.
6. "STM References - Annotated Links for Scanning T
unneling Microscope Amateurs"(http://www.e-basteln.de/index_r.htm).
Retrieved July 13, 2012.
7. K. Oura; V. G. Lifshits; A. A. Saranin; A. V. Zotov & M. Katayama (2003).Surface science: an introduction(https://books.go
ogle.com/?id=TTPMbOGqF-YC&pg=PP1). Berlin: Springer-Verlag. ISBN 3-540-00545-5.
8. D. A. Bonnell & B. D. Huey (2001). "Basic principles of scanning probe microscopy". In D. A. Bonnell.
Scanning probe
microscopy and spectroscopy: Theory, techniques, and applications(2 ed.). New York: Wiley-VCH. ISBN 0-471-24824-X.
9. Pan, S. H.; Hudson, EW; Lang, KM; Eisaki, H; Uchida, S; Davis, JC (2000). "Imaging the ef
fects of individual zinc impurity
atoms on superconductivity in Bi2Sr2CaCu2O8+delta".Nature. 403 (6771): 746–750. arXiv:cond-mat/9909365 (https://arxi
v.org/abs/cond-mat/9909365) . Bibcode:2000Natur.403..746P (http://adsabs.harvard.edu/abs/2000Natur .403..746P).
doi:10.1038/35001534 (https://doi.org/10.1038%2F35001534). PMID 10693798 (https://www.ncbi.nlm.nih.gov/pubmed/106
93798).
10. G. Schitter; M. J. Rost (2008)."Scanning probe microscopy at video-rate"(https://web.archive.org/web/20090909035928/htt
p://www.materialstoday.com:80/view/2194/scanning-probe-microscopy-at-videorate/) . Materials Today. UK: Elsevier. 11
(special issue): 40–48.doi:10.1016/S1369-7021(09)70006-9(https://doi.org/10.1016%2FS1369-7021%2809%2970006-9) .
ISSN 1369-7021 (https://www.worldcat.org/issn/1369-7021). Archived from the original (http://www.materialstoday.com/vie
w/2194/scanning-probe-microscopy-at-videorate/)(PDF) on 2009-09-09.
11. R. V. Lapshin; O. V. Obyedkov (1993). "Fast-acting piezoactuator and digital feedback loop for scanning tunneling
microscopes" (http://www.lapshin.fast-page.org/publications.htm#fast1993) (PDF). Review of Scientific Instruments. 64 (10):
2883–2887. Bibcode:1993RScI...64.2883L (http://adsabs.harvard.edu/abs/1993RScI...64.2883L) . doi:10.1063/1.1144377 (h
ttps://doi.org/10.1063%2F1.1144377).
12. M. J. Rost; et al. (August 2016)."Scanning probe microscopes go video rate and beyond"(http://dx.doi.org/10.1063/1.1915
288). Review of Scientific Instruments. US: American Institute of Physics.76: 053710. Bibcode:2005RScI...76e3710R (htt
p://adsabs.harvard.edu/abs/2005RScI...76e3710R) . doi:10.1063/1.1915288 (https://doi.org/10.1063%2F1.1915288).
ISSN 1369-7021 (https://www.worldcat.org/issn/1369-7021). Check date values in: |year= / |date= mismatch(help)
13. B. S. Swartzentruber (1996). "Direct measurement of surface dif
fusion using atom-tracking scanning tunneling microscopy".
Physical Review Letters. 76 (3): 459–462. Bibcode:1996PhRvL..76..459S (http://adsabs.harvard.edu/abs/1996PhRvL..76..4
59S). doi:10.1103/PhysRevLett.76.459(https://doi.org/10.1103%2FPhysRevLett.76.459) . PMID 10061462 (https://www.ncb
i.nlm.nih.gov/pubmed/10061462).
14. "STM carbon nanotube tips fabrication for critical dimension measurements".
Sensors and Actuators A: Physical. 123–124:
655–659. 2005. doi:10.1016/j.sna.2005.02.036(https://doi.org/10.1016%2Fj.sna.2005.02.036)
.
15. R. V. Lapshin (1995). "Analytical model for the approximation of hysteresis loop and its application to the scanning tunneling
microscope" (http://www.lapshin.fast-page.org/publications.htm#analytical1995)(PDF). Review of Scientific Instruments. 66
(9): 4718–4730. Bibcode:1995RScI...66.4718L (http://adsabs.harvard.edu/abs/1995RScI...66.4718L) .
doi:10.1063/1.1145314 (https://doi.org/10.1063%2F1.1145314). (Russian translation (http://www.lapshin.fast-page.org/publi
cations.htm#analytical1995)is available).
16. R. V. Lapshin (2007). "Automatic drift elimination in probe microscope images based on techniques of counter-scanning
and topography feature recognition"(http://www.lapshin.fast-page.org/publications.htm#automatic2007)(PDF).
Measurement Science and Technology. 18 (3): 907–927. Bibcode:2007MeScT..18..907L (http://adsabs.harvard.edu/abs/20
07MeScT..18..907L). doi:10.1088/0957-0233/18/3/046(https://doi.org/10.1088%2F0957-0233%2F18%2F3%2F046) .
17. R. V. Lapshin (2004). "Feature-oriented scanning methodology for probe microscopy and nanotechnology"(http://www.laps
hin.fast-page.org/publications.htm#feature2004)(PDF). Nanotechnology. 15 (9): 1135–1151.
Bibcode:2004Nanot..15.1135L (http://adsabs.harvard.edu/abs/2004Nanot..15.1135L) . doi:10.1088/0957-4484/15/9/006(htt
ps://doi.org/10.1088%2F0957-4484%2F15%2F9%2F006) .
18. R. V. Lapshin (2011). "Feature-oriented scanning probe microscopy". In H. S. Nalwa.Encyclopedia of Nanoscience and
Nanotechnology (http://www.lapshin.fast-page.org/publications.htm#fospm2011) (PDF). 14. USA: American Scientific
Publishers. pp. 105–115.ISBN 1-58883-163-9.
19. R. Wiesendanger; I. V. Shvets; D. Bürgler; G.Tarrach; H.-J. Güntherodt & J.M.D. Coey (1992). "Recent advances in spin-
polarized scanning tunneling microscopy".Ultramicroscopy. 42–44: 338–344. doi:10.1016/0304-3991(92)90289-V(https://d
oi.org/10.1016%2F0304-3991%2892%2990289-V) .
20. T. Waldmann; J. Klein; H.E. Hoster; R.J. Behm (2012). "Stabilization of Large Adsorbates by Rotational Entropy: Aime-
T
Resolved Variable-Temperature STM Study".ChemPhysChem. 14: 162–169. doi:10.1002/cphc.201200531(https://doi.org/
10.1002%2Fcphc.201200531). PMID 23047526 (https://www.ncbi.nlm.nih.gov/pubmed/23047526).
21. J. Bardeen (1961). "Tunneling from a many particle point of view". Phys. Rev. Lett. 6 (2): 57–59.
Bibcode:1961PhRvL...6...57B (http://adsabs.harvard.edu/abs/1961PhRvL...6...57B) . doi:10.1103/PhysRevLett.6.57(https://
doi.org/10.1103%2FPhysRevLett.6.57).
22. R. Young; J. Ward; F. Scire (1972). "The Topografiner: An Instrument for Measuring Surface Microtopography"(https://web.
archive.org/web/20030508182417/http://www .nanoworld.org:80/museum/young2.pdf)(PDF). Rev. Sci. Instrum. 43: 999.
Bibcode:1972RScI...43..999Y (http://adsabs.harvard.edu/abs/1972RScI...43..999Y) . doi:10.1063/1.1685846 (https://doi.org/
10.1063%2F1.1685846). Archived from the original (http://www.nanoworld.org/museum/young2.pdf) (PDF) on 2003-05-08.
23. "The Topografiner: An Instrument for Measuring Surface Microtopography"(http://nvl.nist.gov/pub/nistpubs/sp958-lide/214-2
18.pdf) (PDF). NIST.

Further reading
Tersoff, J.: Hamann, D. R.: Theory of the scanning tunneling microscope,Physical Review B 31, 1985, p. 805 - 813.
Bardeen, J.: Tunnelling from a many-particlepoint of view, Physical Review Letters 6 (2), 1961, p. 57-59.
Chen, C. J.: Origin of Atomic Resolution on Metal Surfaces in Scanning unneling
T Microscopy, Physical Review Letters 65
(4), 1990, p. 448-451
G. Binnig, H. Rohrer, Ch. Gerber, and E. Weibel, Phys. Rev. Lett. 50, 120 - 123 (1983)
G. Binnig, H. Rohrer, Ch. Gerber, and E. Weibel, Phys. Rev. Lett. 49, 57 - 61 (1982)
G. Binnig, H. Rohrer, Ch. Gerber, and E. Weibel, Appl. Phys. Lett., Vol. 40, Issue 2, pp. 178-180 (1982)
R. V. Lapshin (2004). "Feature-oriented scanning methodology for probe microscopy and nanotechnology" (PDF).
Nanotechnology. UK: IOP. 15 (9): 1135–1151. Bibcode:2004Nanot..15.1135L. doi:10.1088/0957-4484/15/9/006.
ISSN 0957-4484.
R. V. Lapshin (2011). "Feature-oriented scanning probe microscopy". In H. S. Nalwa.Encyclopedia of Nanoscience and
Nanotechnology (PDF). 14. USA: American Scientific Publishers. pp. 105–115.ISBN 1-58883-163-9.
Fujita D., Sagisaka K. "Topical review: Active nanocharacterization of nanofunctional materials by scanning tunneling
microscopy". Sci. Technol. Adv. Mater. 9: 013003. Bibcode:2008STAdM...9a3003F. doi:10.1088/1468-6996/9/1/013003.
Roland Wiesendanger (1994).Scanning probe microscopy and spectroscopy: methods and applications . Cambridge
University Press. ISBN 0-521-42847-5.
Theory of STM and Related Scanning Probe Methods.Springer Series in Surface Sciences, Band 3. Springer , Berlin 1998
Stabilization of Large Adsorbates by Rotational Entropy: A iTme-Resolved Variable-Temperature STM Study,
ChemPhysChem, doi: 10.1002/cphc.201200531

External links
NobelPrize.org website about STM, including an interactive STM simulator.
Build a simple STM with a cost of materials less than $100.00 excluding oscilloscope
animations and explanations on various types of microscopes including electron microscopes
(Université Paris Sud)

Retrieved from "https://en.wikipedia.org/w/index.php?title=Scanning_tunneling_microscope&oldid=810331165


"

This page was last edited on 14 November 2017, at 16:38.

Text is available under theCreative Commons Attribution-ShareAlike License ; additional terms may apply. By using this site, you
agree to the Terms of Use and Privacy Policy. Wikipedia® is a registered trademark of theWikimedia Foundation, Inc., a non-
profit organization.

You might also like