This manual contains proprietary information, protected by international copyright laws. No part of this document
may be reproduced without the written agreement of the publisher, FLINTAB.
FLINTAB reserves the right to revise this manual and alter its content without notification at any time.
ANNEX A containing installation schematic drawings and connector pin outs comprises of 12 pages.
OPTIONS
a. RS485 serial port adapter for bi-directional communication with a host computer.
b. 2nd analog load cell input board (PCB765).
c. Analogue output 0/4 - 20mA or 0-10V, 16 bits resolution. (PCB 761) remotely powered.
2.1 Mounting
The mounting location must be such that the instrument is not subject to excessive vibrations, heat or
humidity. Avoid direct sunlight on the front of the instrument. The unit has to be installed at the right
height to allow an easy reading of the display and keyboard operation.
2.2 Wiring
Use load cell cable 6 x 0,5 mm2 shielded for the sensor. Use 3 x 0,34 mm2 shielded for RS232C
connection and 2x0,34 mm2 twisted pair and shielded for RS485 connection. .
Table top
All connections to the instrument are made through the rear panel connectors. Strain reliefs are
supplied with the connectors. The shield should be connected to the metal frame of the connector.
Stainless steel
- Remove the rear panel and lift it carefully.
-.Insert cables via the cable glands. Strip and connect cables according to connector diagram in
Annex A
-.Connect the shields of the cables between the plastic part and metal case of cable glands or on
screws supporting
the pcb’s.
- Re-install the rear panel.
2.3 Power
The instrument is powered from external power supply (9-15VDC/500mA) or battery.
As the instrument is computer controlled it requires clean power for reliable operation.
Power supplied should come from a source that is isolated from other process equipment.
A main adapter 9VDC/500mA is recommended for operation.
LOAD CELL : 5VDC, fixed or alternating polarity (Setup selectable) for 10 load cells of 350
OHM each.
EXCITATION
CAUTION : Do not run signal cables together with power cables. Connect the shielding
where indicated on the drawing only. Never use a megger to check wiring.
Never use plastic insulating tape on load cell connections.
Be sure the AC power socket outlet is properly protected. For optimum EMC
performance,
keep the length of shield inside the enclosure as short as possible.
WEIGHT DISPLAY
Six digit LED for the display of measured weight (gross or net), and error messages.
Digit height approx. 20mm LED.
OVER-RANGE is indicated by vvvvvv.
UNDER-RANGE is indicated by uuuuuu.
It is illuminated when the scale weight is stable and within 1/4 division of actual zero.
It is illuminated when the scale has been tared and the display is in NET mode.
It is illuminated when the weight from scale #2 is displayed (if 2nd analogue load cell input
option -PCB765- is installed).
When both scale 1 and scale 2 are illuminated the sum of the weight is displayed.
ANALOG INPUT
LOADCELL EXC/TION +5V switched polarity or + 5 VDC with sense.
CONNECTION 6 wire technique. Max 10 load cells 350 Ohm each.
FLI 5204 WEIGHT INDICATOR - TECHNICAL MANUAL 7 of 28
SIGNAL RANGE -0.25 to 2.0 mV/V (Gain=10), -0.25 to 4.0 mV/V (Gain=20).
SENSITIVITY min 0,4 µV / digit (VSI) approved scales.
min 0,1 µV / digit non-approved scales.
INPUT AMPLIFIER Input noise 0.3µVp-p. Input bias current 10nA typical.
A/D CONVERTER Sigma delta 550.000 internal counts max.
.Conversion speed 3 , 7, 14, 28, 57,70 Hz (selectable).
LINEARITY Within 0,002% of full scale.
SPAN TEMP - COEFF < 2 ppm / oC.
ZERO TEMP - COEFF < 2 ppm / oC.
LONG TERM STABILITY 0.005 % of full scale per year.
2nd ANALOG INPUT Two inputs for the connection of analog load cells.
(OPTIONAL) Technical specifications of 2nd analog input are the same to the
.primary .analog input.
.Each input has its own configuration and calibration parameters.
DIGITAL INPUT (x1) 9-24 VDC , positive common optoisolated to 2.5 KV.
.Input resistance 3.3 KΩ.
.ON delay 2 msec max. OFF delay 2msec max.
DIGITAL OUTPUT (x2) 24 VDC ± 10 % transistor (SOURCE) darlington, positive common.
.Max current 100 mA - leakage current 100 µA.
.Max off - state voltage 30 VDC.
ON delay 2 msec max - OFF delay 2 msec max.
.All outputs optoisolated to 2.5 KV.
ANALOGUE OUTPUT
CURRENT or VOLTAGE Hardware selectable via jumper JP1 of printed circuit board 761.
CURRENT OUTPUT 0 - 20 mA or 4 - 20 mA. .Max load resistance 1KΩ (line + termination).
VOLTAGE OUTPUT 0.02 - 10 V. Min load resistance 1KΩ.
RESOLUTION Internal 16 bit. External 16 bit or in accordance with regulation.
LINEARITY Better than 0,01% of FSR.
THERMAL STABILITY 50 PPM/0C typical.
SHORT CIRCUIT Yes, 25mA indefinite duration.
PROTECTION
4 SETUP
In the setup dialogue the following two numeric scroll edit procedures are used.
The Audit Trail Counter is incremented every time a weight parameter or Calibration is changed
regardless if the change was saved in EEPROM or not.
The A.T.CNT (Audit Trail Counter) is provided for Weights & Measures Authorities to check if any
calibration attempt has been made since the last inspection.
* Clear A/D converter error (8.4): If enabled A/D converter errors (due to excessive noise) will be
cleared automatically, when the cause is no long present. If not
the unit will require operator intervention to acknowledge the
error.
* AC/DC Excitation (8.8) : AC: Switching polarity excitation, at a rate defined by the
conversion rate. Switching excitation results in more stable zero.
DC: DC excitation selected for the load cells.
DIALOGUE E-CAL (Electronic calibration with mV/V form Load cell data sheets)
Display shows ZERO. Press < or ~ to scroll through ZERO, SPAN or press ) to exit to E-CAL
Press ( to enter the displayed dialogue.
Enter the overall mV/V of the dead-load (Load cell zero balance + scale dead-load).
Display shows weight. Press ) to exit to ZERO then <. Display shows SPAN. Press (
Key in the overall mV/V value (calculated from the load cell data sheet), corresponding to a calibration
Weight equal to the max capacity of the scale and press (
The entered value is used to compute the span factor after which the weight on the scale is displayed.
Press ) to exit to SPAN.
EXAMPLE:
Assume a scale of Max= 30/60kg , e=0.010/0.020kg , with 4 loadcells of rated capacity
50kg -2mV/V each and the dead load is 1.940kg.
Loadcell data:
L/C1 : 1.9793 mV/V output at 50kg and zero balance : 0.0257 mV/V.
L/C2 : 1.9392 mV/V output at 50kg and zero balance : 0.0276 mV/V.
L/C3 : 1.9577 mV/V output at 50kg and zero balance : 0.0553 mV/V.
L/C4 : 1.9640 mV/V output at 50kg and zero balance : -0.0022 mV/V.
LC RATED OUTPUT average = (1.9793 + 1.9392 + 1.9577 +1.9640) / 4 = 1.9600 mV/V.
LC OUTPUT AT SCALE MAX = 1.9600*60 / 4*50 = 0.5880 mV/V.
L/C ZERO BALANCE average = [0.0257 + 0.0276 + 0.0553 + (- 0.0022)] / 4 = 0.0266 mV/V.
SCALE DEDLOAD : 1.9600 mV/V . 1.940 kg = 0.0190 mV/V.
(4x50kg)
OVERALL DEADLOAD = L/C ZERO BALANCE + SCALE DEADLOAD = 0.0456 mV/V.
Enter the value of 0.0456 for the dead load calibration and 0.5880 for span calibration.
5.1 PRINTER OUTPUT (SETUP 2 2.t=01, 03, 04, 05, 06, 07, 08, 10, 20)
The data to be printed, printer type and paper dimensions is defined in SETUP 2
ERROR
POSSIBLE CAUSE ACTION TO BE TAKEN
DISPLAY
Err 01 SYSTEM ROM : Faulty EPROM. Contact manufacturer.
Err 02 DATA RAM : Faulty CMOS RAM. Contact manufacturer.
Err 04 CALIBRATION DATA : Faulty EEPROM. Contact manufacturer.
Err 05 SCALE or A/D CONVERTER. Check scale, cable, connectors,
contact manufacturer.
Err 06 LOW INPUT VOLTAGE . Check power supplied to the
instrument.
Err 15 System has been initialised due to power Zero scale.
failure or soft reset.
Err 16 System DATE – TIME is wrong. Enter new date – time.
Err 20 PRINTER IS NOT ON-LINE. Check printer cables. Press ( to retry
Either not connected or out of paper or Or ) to abort.
failed.
Err 26 No paper for EPSON TM295 printer. Supply with paper.
Err 30 Protocol not ready. Computer not Check computer cables. Press ( to
connected or communication link failed. retry or ) to abort.
Err 33 Protocol not Acknowledge. No correct Check computer cables. Press ( to
response has been received from host retry or ) to abort.
computer.
Err 50 Small sample count. Retry.
Err 51 Small sample weight. Retry.
Err 55 Alibi memory full. Acknowledge the error. The unique
identification number will reset to
0000.
Err 56 No Alibi memory in NET mode (only Printout aborted.
gross).
Err 57 Alibi memory corrupted. The Alibi memory can not be cleared
but next records will be corrected.
Err 67 Corrupted totaliser. Clear totalising memory.
Err 69 TOTALISER OVERFLOW. Print / clear totalisers. Overflow is not
Because totalisers have not been cleared critical.
for But remember that 1 total capacity
long time. must be added to the indication, each
time an overflow occurs.
8 FUNCTION MENU
The function menu enables the selection of a number of software programs indirectly.
Press key > and hold pressed for ~ 3 sec.
Display shows Fn 00 . Key in the function code desired (using multi digit entry procedure).
The corresponding program will be activated. To exit the function selection, press ( while Fn 00 is
displayed. To exit from any function program, press ). The function menu will not operate if function
operation is disabled (SETUP 1.8=0) or while the sum of scale1 and scale 2 is displayed.
The following functions are available.
Fn 01 EDIT SETPOINTS
Display shows SetP 1 briefly, then the existing setpoint value. Key in the desired value.
Display shows SetP 2 briefly, then the existing setpoint value. Key in the desired value.
The setpoints are saved in EEPROM.
Fn 02 DISPLAY REMAINING BATTERY CAPACITY
The battery charge will be displayed as a percentage of the remaining capacity.
Fn 05 SET DATE – TIME – SERIAL Nr. (Option and enabled only if SETUP 1.6 = 1)
-Display shows: DDMMYY or MMDDYY (depends on setup 1.4)
DD=Day, MM=Month, YY=Year. Use MULTI DIGIT EDIT to edit the date.
-Display shows: HHmmSS. HH=Hour, mm=minutes, SS=Seconds. Edit the time.
-Display shows: nnnnn, the print serial number. Edit the print number.
The program ends and returns to weight display.
Fn 06 HIGH RESOLUTION
The weight display accuracy will be increased 10 times.
The display flashes. Press ) to exit and return to normal resolution.
Fn 20 - 30 PRINT TYPE SELECT
Equivalent to 2.t in SETUP 2 (i.e. Fn23 corresponds to 2.t=03)
The functions are activated only if SETUP 2.8=1
Fn 20 DISABLE PRINTER OUTPUT
Fn 21 PRINT FORMAT 1
Fn 22 CONTINUOUS WEIGHT OUTPUT
Fn 23 PRINT FORMAT 3
Fn 24 PRINT FORMAT 4
Fn 25 PRINT FORMAT 5
Fn 26 PRINT FORMAT 6
Fn 27 PRINT FORMAT 7
Fn 28 PRINT FORMAT 8
Fn 29 PRINT FORMAT 9
Fn 30 PRINT FORMAT 10
Fn 40 PIN CHANGE
Used to change the current Personal Identification Number that enables access to the
calibration procedure. (The unit is factory set so that the PIN is not active i.e. 000000).
The display shows Pin 0 briefly then 000000. Key in the old pin and press (.
The display shows Pin 1 if the pin was correct else the unit resets.
Key in the new Pin and press (. The display shows Pin 2. Key in the same Pin for
validation and press (. The new pin is stored and the display shows PASS briefly.
If the two entries are not the same, FAIL is displayed briefly and the program returns
with the old pin remaining in memory.
Fn 41 DOWNLOAD CUSTOM PRINT FORMAT . 1 (GROSS MODE)
FORMAT 1 ticket is printed if setup 2t=20 and when the indicator operates in gross
mode. The ticket may be downloaded from a PC to the LD5204/6 via serial port 1
(RS232C). A max of 511 bytes may be stored. If more characters are received they will
be lost. The ticket format may be created or edited with ASCII characters, comprising of
fixed texts and print fields. When a print operation takes place LD 5204/6 will insert the
print fields.
Use a serial communication program (e.g. Hyper terminal, Procomm) and set serial
port, baud rate/data bits. Download procedure:
Select Fn 41. The display will show ACCESS if PIN is activated. Key in the PIN and
press ( .
The display shows For 1 .The default ticket is uploaded to the PC and the LD5204/6
is ready to receive format 1. The ticket may be used for editing of a new one may be
downloaded.
Download the ticket to the indicator and press ) to acknowledge the end of
transmission.
Display shows Prn 1 briefly and the received ASCII block is transmitted to the PC for
verification. The received ticket is stored in non volatile memory.
Fn 42 DOWNLOAD CUSTOM PRINT FORMAT . 2 (NET MODE)
Identical procedure with Fn 41. The display shows For 2 .
FORMAT 2 ticket is printed if setup 2t=20 and when the indicator operates in net mode.
Fn 43 COPY LAST PRINTOUT
Used to reprint the ticket just printed, in case the paper is damaged.
Fn 48 CALIBRATION NR CHECK Refer to chapter 4.1
Fn 49 SETUP & CALIBRATION Refer to chapter 4
Fn 50 PIECE COUNT
The display will show the number of pieces on the scale calculated using the average
piece weight already in memory. (identical operation with key &).
Fn 51 SAMPLE & DISPLAY PIECES
-Display shows Pc XX. Enter the sample size.
-Display shows the number of pieces (piece count symbol illuminates).
-Press < to display the average piece weight. (Press again to return to pieces display)
or press > to tare the scale or press ( to print weight and pieces.
Press & to return to weight mode.
Fn 52 EDIT AVERAGE PIECE WEIGHT & DISPLAY PIECES
-Display shows XXXXX. Edit the average piece weight in * 100 accuracy.
-Display shows the number of pieces (piece count symbol illuminates
-Press < to display the average piece weight (press again to return to pieces display)
or press > to tare the scale or press (to print weight and pieces.
Press & to return to weight mode.
Fn 53 SAMPLE & DISPLAY AVERAGE PIECE WEIGHT
Display shows Pc XX. Enter the sample size.
Display shows the average piece weight in * 100 accuracy. Press ) to exit.
Fn 55 VIEW ALIBI MEMORY LOCATION / PRINT 10 NEXT SERIAL NUMBERS
The display shows n 1 2 3 4 where 1234 is the serial number of the last record.
Key in the 4 digit serial number desired and/or press (. The display flashes the weight
of this record. Press ( to print the serial number displayed plus the next nine locations.
Fn 56 PRINT ALL ALIBI MEMORY
Automatically prints a list of the contents of the Alibi memory. The printer must be
capable of printing on 80 columns paper compressed mode.
Empty locations are printed as - - - - - - .
Corrupted locations are printed as * * * * * * .
Alibi memory functions and storage are executed only if enabled from SETUP (SETUP 2.t=13).
10 SERVICE
- There no serviceable parts. The unit must be serviced by trained personnel only. The user may
check load cell connection and power supply.
- LOAD CELLS
Load cells are reliable and very rarely present errors. Check input and output resistance, and
resistance between any terminal and shield. Check load cell connection and cable.
- POWER SUPPLY
Check 9-15VDC power supply&resettable fuse F4 on PCB 801.
- SETPOINT OUTPUTS
Check 24VDC power supply&resettable fuse F3 on PCB 801.
ADJUSTME
NT
TOTALISING RESERVED RESERVED DATE RESERVED REAL TIME RESERVED RESERVED AUTO
MEMORY FORMAT CLOCK & RAM POWER
TOTALISER OFF NO OF
00-99 MIN SAMPLE
SETUP1
OPERATION 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.t 1.E
PRINT PAGE PAPER LEFT PAGE LINE PRINTER RESERVED PRINT RESERVED RESERVED WAIT
TYPE LENGTH REVERCE MARGIN HEADER TERMINATION MODEL BELOW UNLOAD
MIN CAP
SET2 \
COM 2.t 2.L 2.r 2.A 2.F 2.E 2.1 2.2 2.3 2.4 2.5 2.6
INSTRUMENT TIMEOUT HAND OPERATOR HOST REMOTE RESERVED RESERVED DDEBUG BAUD RATE DATA BIT
COMMUNICATION CONTROL SHAKE DISABLE ENQUIRY KBD SERIAL SERIAL
COMMAND CHANNEL 2 CHANNEL 2
TYPE
SET3\
2COM 3.t 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.b 3.d
TILT SWITCH
SETUP 4 4.t