You are on page 1of 5

DAISY SHANE L. ATAYAN Molecular orbital–Describes the distribution Ion-exchange chromatography.

tion Ion-exchange chromatography. Anions such components with a constant total concentra-
MS 1 of an e- within a molecule. as —SO_3 or cations such as —N(CH3)_3 tion. A Scatchard plot is used to measure an
CHEMISTRY 220 – CHEAT SHEET 1 Monochromatic light - very narrow range of are covalently attached to the stationary solid equilibrium constant. Youshould be able to
MARCH 1, 2018 wavelengths (“one color”). phase, usually a resin. apply Excel Solver to fit nonlinear
Molecular exclusion chromatography. Also experimental measurementssuch as
CHAPTER 18 Monochromator - device (usually a grating absorbance or fluorescence to the equilibrium
or prism) that disperses light into its called size exclusion, gel filtration, or gel
Spectrophotometry–measure chemical permeation chromatography, this technique expression to find an equilibrium constant. The
component wavelengths and selects a narrow method of continuous variation(Job’s plot)
concentrations using light(in dilute solution) band of wavelengths to pass through the exit separates molecules by size, with the larger
solutes passing through most quickly. allows us to determine the stoichiometry of a
Colorimetry –absorption of visible light slit. complex. In flow injection analysis, sample
Wave-Particle Duality of Light: Phosphorescence- Emission of light during a Affinity chromatography. This most selective injected into a flowing carrierstream is mixed
ℎ𝑐 transition bet states of different spin kind of chromatography employs specific with a color-forming reagent and passed into
𝑐 = 𝜈𝜆 𝐸 = ℎ𝜈 = interactions between one kind of solute a flow-through detector. Analyte disperses
𝜆 multiplicity (e.g., triplet S singlet). It’sslower
than fluorescence, with emission occurring molecule and a second molecule that is and reacts withreagent without reaching
Molarabsorptivityamt light absorbed at𝜆
~104 to 102 s after absorption of a photon. covalently attached (immobilized) to the equilibrium. Precision depends on carryingout
Chromophore responsible for light absorption stationary phase. the process reproducibly. In sequential
Photon - “particle” of light with energy hv, injection, sampleand reagent are separately
𝝀𝒎𝒂𝒙 Color absorbed Color where h is Planck’s constant & v is the f. A chromatogram is a graph showing the
observed s detector response as a function of elution aspirated into the carrier by acomputer-
380–420Violet Green-yellow Quenching- Process in which emission from an time. controlled syringe pump. After being
420–440Violet-blue Yellow excited molecule isdecreased by energy stationary in thereaction coil, analyte,
transfer to another molecule called a The retention time, tr, for each component is product, and reagent are pushed throughthe
440–470Blue Orange the time that elapses between injection of the
470–500Blue-green Red quencher. detector.Immunoassays use antibodies to
mixture onto the column and the arrival of detect the analyte of interest. Inan enzyme-
500–520Green Purple Raman scattering - Scattering of light in that component at the detector.
520–550Yellow-green Violet linked immunosorbent assay, signal is
which the wavelength of scattered light is amplified by coupling analyte to an enzyme
550–580 Yellow Violet-blue changed from that of incident light by an Retention volume, Vr, is the volume of mobile
580–620 Orange Blue phase required to elute a particular solute that catalyzes many cycles of a
energy correspondingto vibrational energy of reactionproducing a colored or fl uorescent
620–680Red Blue-green the molecule responsible for scattering.In from the column.
680–780Red Green product. Time-resolved fl
Stokes Raman scattering, the molecule gains Adjusted retention time: t’R = tR - tM uorescencemeasurements provide sensitivity
vibrational energy andscattered light has less 𝑡𝑟−𝑡𝑚
Chemiluminescence - emission from a Retention factor: 𝑘 = by separating analyte fl uorescencein time
chemical reaction energy than incident light. In anti-Stokes 𝑡𝑚
and wavelength from background
Ramanscattering, an excited molecule loses time solute spends in stationary phase
fluorescence.Luminescence intensity is
Basic components of a spectrophotometer: vibrational energy and scatteredlight has
𝑘=
time solute spends in mobile phase
include a radiation source, a monochromator, proportional to the concentration of
more energy than incident light. the emitting species if the concentration is low
a sample cell, and a detector. Resolution:
Reagent blank - A solution prepared from all 2Δ𝑡𝑟 enough.
Minimize errors:if the absorbance falls in the of the reagents except analyte. The blank
range A < 0.3–2 𝑅=
measures the response of the analytical w𝑏 + w𝑎 CHAPTER 20
In spectrophotometric titration:absorbance is method to impurities in the reagents or any Number of plates on column:
L 16𝑡𝑟 2 Attenuated total reflectance - An analytical
monitored as titrant is added.In many other effects caused by any component other technique based on passage oflight through a
𝑁= =
reactions, there is an abrupt change in slope than the analyte. The reagent blank, unlike H 𝑤2 waveguide or optical fi ber by total internal
when the equivalence point is reached. When the method blank, is not subjected to all refl ection. Theabsorption of the cladding is
a molecule absorbs light, it is promoted to an sample preparation steps before analysis. CHAPTER 19
sensitive to the presence of analyte. Some
excited state from which it returns to the Refractive index - The speed of light in any Flow injection analysis - sample is injected ofthe evanescent wave is absorbed in the
ground state by nonradiative processes or by medium is c/n, where c isthe speed of light in into flowing liquid carrier containing a cladding during each reflection inthe presence
fluorescence (singlet S singlet emission) or vacuum and n is the refractive index of the reagent that reacts with the of analyte. The more analyte, the more signal
phosphorescence (triplet S singlet emission). medium.The refractive index also measures analyte.Additional reagents might be added is lost.
Emission intensity - proportional to concat the angle at which a light ray is bentwhen it farther downstream. As sample flowsfrom
Bandwidth- Usually, the range of
low conc; at high conc, emission decreases passes from one medium into another. injector to detector, the sample zone
wavelengths or frequencies of anabsorption
because of selfabsorption by the analyte. broadens and reacts with reagentto form a
LASER - Light Amplification thru Stimulated or emission band, typically measured at a
product to which the detector responds.
Excitation spectrum - graph of emission Emission of Radiation height equal to halfof the peak height. Also,
intensity vs excitation wavelength; observed Immunoassay - An analytical measurement the width of radiation emerging from the exit
Stray light wavelengths outside the
at lower energy than the absorption spectrum using antibodies. slitof a monochromator.
bandwidth expected from the monochromator
and tends to be the mirror image of the Method of continuous variation - Procedure Beam chopping - technique using a rotating
absorption spectrum. Electrons emitted from the photosensitive
for finding the stoichiometryof a complex by beam chopper to modulatethe signal in a
surface strike a second surface, called a
***A molecule that is not fluorescent can be preparing a series of solutions with different spectrophotometer at a frequency at which
dynode, which is positive with respect to the
analyzed by attaching a fluorophore to it or by metal-toligandratios. The ratio at which the noise is reduced. In atomic absorption,
photosensitive emitter.
its ability to enhance or diminish the extreme response (such as spectrophotometric periodic blocking of the beam allows a
fluorescence of a sensor compound. Light A thermocouple is a junction between two absorbance) occurs corresponds to the distinction tobe made between light from the
emitted by a chemical reaction— different electrical conductors. stoichiometry of the complex. Also called source and light from the flame.
chemiluminescence—is also used for A ferroelectric material, such as deuterated Job’s method.
Blackbody radiation - Radiation emitted by a
quantitative analysis. triglycine sulfate, has a permanent electric Quantum yield - In photochemistry, the blackbody. The energy andspectral
Absorbance, A = log (P0/P), P0radiant power polarization because of alignment of the fraction of absorbed photons thatproduce a distribution of the emission depend only on the
also called optical density molecules in the crystal. particular result. For example, if a molecule temperature of theblackbody.
Snell’s (refraction) Law: n1sinϴ1 = n2sin ϴ2 can isomerize froma cis to a trans isomer
Absorption spectrum - graph of absorbance Charge coupled device - An extremely
when light is absorbed, the quantum yield for
vs𝜆, frequency, or wavenumber Interferenceis any effect that changes the sensitive detector in which lightcreates
isomerization is the number of molecules that
signal while analyte concentration remains electrons and holes in a semiconductor
Beer’s law- Relates the absorbance, A, of a isomerize divided by thenumber that absorb
unchanged. Interference can be corrected by material. The electronsare attracted to
sample to its concentration, c,pathlength, b, photons. Quantum yield is in the range 0 to 1.
removing the source of interference or by regions near positive electrodes, where the
and molar absorptivity, ε: A =εbc. More Quenching - Process in which emission from electrons are“stored” until they are ready to
correctly calledBeer-Lambert-Bouguer law. preparing standards that exhibit the same
interference. an excited molecule is decreased by energy be counted. The number of electronsin each
Derivatization – Chemical alteration to attach transfer to another molecule called a pixel (picture element) is proportional to the
a group to a molecule sothat the molecule can Chemical interference is caused by any quencher. number of photonsstriking the pixel.
be detected. Thiscan alter volatility or component of the sample that decreases the
extent of atomization of analyte. Scatchard plot - A graph used to fi nd the Diffraction - The bending of light rays by a
solubility to allow easier separation. equilibrium constant for a reactionsuch as X 1 grating. This process occurswhen
Electromagnetic spectrum –whole range of Ionization interference can be a problem in P Δ PX. It is a graph of [PX]/[X] versus [PX] or electromagnetic radiation passes through or is
electromagnetic rad: visible, radio, X-rays, IR the analysis of alkali metals at relatively low anyfunctions proportional to these quantities. reflected from slitswith a spacing comparable
temperature and in the analyses of other The magnitude of the slope ofthe graph is the to the wavelength. Interference of waves
Electronic transition –an electron goes from elements at higher temperature. equilibrium constant. fromadjacent slits produces a spectrum of
one energy level to another.
MALDI: matrix-assisted laser radiation, with each wavelength
Sequential injection - Analytical technique
Excitation spectrum - graph of luminescence desorption/ionization emerging at a different angle.
related to flow injection. Sample and
(measured at fixed 𝜆) versus excitation reagents are taken into a holding coil through Dispersion - A measure of the ability of a
frequency or wavelength. Closely corresponds Distribution coefficient:
𝑡𝑜𝑡𝑎𝑙 𝑐𝑜𝑛𝑐𝑒𝑛𝑡𝑟𝑎𝑡𝑖𝑜𝑛 𝑖𝑛 𝑝ℎ𝑎𝑠𝑒 2 a multiportvalve. After a suitable reaction monochromator to separatewavelengths
to an absorption spectrum bc luminescence is differing by Dl through the angle Df. The
𝑡𝑜𝑡𝑎𝑙 𝑐𝑜𝑛𝑐𝑒𝑛𝑡𝑟𝑎𝑡𝑖𝑜𝑛 𝑖𝑛 𝑝ℎ𝑎𝑠𝑒 1 time, flow is reversed and the zones of
generally proportional to the absorbance. greater the dispersion,the greater the angle
Chromatography operates on the same reagent,product, and sample are pushed
Excited state - state of an atom or a molecule principle as extraction, but one phase is held through a detector to measure theamount of separating two closely spaced wavelengths.
having more than minimum possible energy. in place while the other moves past it. product. Flow is not continuous, so sequential Fora prism, dispersion refers to the rate of
Fluorescence -emits a photon10-8 to 10-4 s injection consumesless reagents than does flow change of refractive index withwavelength,
The mobile phase (the solvent moving through injection. dn/dl.
after absorbing a photon. Results from a the column) in chromatography is either a
transition between states of thesame spin liquid or a gas. The stationary phase (the one The absorbance of a mixture is the sum of Ferroelectric material - A solid with a
multiplicity (e.g., singlet S singlet). that stays in place inside the column) is most absorbances of the individualcomponents. At permanent electric polarization(dipole) in the
Frequency - The number of cycles per unit commonly a viscous liquid chemically bonded a minimum, you should be able to fi nd the absence of an external electric field. The
time for a repetitive event. to the inside of a capillary tube or onto the concentrationsof two species in a mixture by polarization resultsfrom alignment of
surface of solid particles packed in the writing and solving two simultaneousequations molecules within the solid.
Hertz - SI unit of f, s-1, or reciprocal seconds for absorbance at two wavelengths. This
column. Fourier analysis - Process of decomposing a
Irradiance- Power per unit area (W/m2) EM procedure ismost accurate if the two function into an infi nite
Fluid entering the column is called eluent. absorption spectra have regions where they
rad beam akaradiant power or intensity. series of sine and cosine terms. Because each
Adsorption chromatography. A solid donot overlap very much. With a term represents a certain frequency
Luminescence- emission of light by molecule spreadsheet, you should be able to usematrix
stationary phase and a liquid or gaseous or wavelength, Fourier analysis decomposes a
Masking - Adding a masking agent to sample mobile phase are used. operations to solve n simultaneous Beer’s law function into itscomponent frequencies or
to prevent other components from interfering equations for ncomponents in a solution, with wavelengths.
Partition chromatography. A liquid measurements at n wavelengths. Youshould be
in a chemical analysis.
stationary phase is bonded to a solid surface, able to use Excel Solver to decompose a Grating - Either a refl ective or a transmitting
Molar absorptivity - tells how much light is which is typically the inside of the silica (SiO2) surface etched with closelyspaced lines; used
spectrum into a sumof spectra of the
absorbed at particular𝜆 by a particular chromatography column in gas components by minimizing the function to disperse light into its component
substance aka extinction coefficient. chromatography. .Isosbestic (crossing) points are observed when wavelengths.
a solution containsvariable proportions of two
Greenhouse gas - A component of Earth’s Thermocouple - An electrical junction across Chemical interference - In atomic measured. Liquids may be atomized by a
atmosphere that absorbs infraredradiation which a temperaturedependentvoltage exists. spectroscopy, any chemical reactionthat plasma, a furnace, or a flame. Flame
from the ground and reradiates some of it Thermocouples are calibrated for decreases the efficiency of atomization. temperatures are usually in the range2 300–
back to the ground, therebykeeping Earth measurementof temperature and usually Detection limit -The smallest quantity of 3 400 K. The choice of fuel and oxidant
warmer than it would be in the absence of the consist of two dissimilar metals in contact with analyte that is “significantlydifferent” from a determines the temperature of the flame and
greenhouse gas. each other. blank. The detection limit is often taken as the affects the extent of spectral, chemical, or
Laser- Source of intense, coherent Waveguide - A thin layer or hollow structure concentrationof analyte that gives a signal ionization interference that will be
monochromatic radiation. Light is producedby in which electromagneticradiation is totally equal to three times the standard deviationof encountered. Temperature instabilityaffects
stimulated emission of radiation from a reflected. signal from a blank. Also called lower limit of atomization in atomic absorption and has an
medium in which anexcited state has been detection even largereffect on atomic emission, because
. Tungsten and silicon carbide lamps the excited-state population isexponentially
pumped to a high population. Coherence behaveapproximately as blackbodies, which Doppler effect -A molecule moving toward a
means thatall light exiting the laser has the sensitive to temperature. An electrically
are objects that absorb alllight striking them source of radiation experiencesa higher heated graphitefurnace requires less sample
same phase. and reradiate a characteristic spectrum of frequency than one moving away from the than a flame and has a lower detectionlimit.
Monochromator - A device (usually a grating radiation.Emission of radiant energy from the source. In an inductively coupled plasma, a radio-
or prism) that disperseslight into its component surface of the blackbodyis proportional to the Graphite furnace - A graphite tube that can frequency inductioncoil heats Ar1 ions to 6
wavelengths and selects a narrow band of fourth power of temperature and shifts to be heated electrically to about2 500 K to 000–10 000 K. At this high
wavelengthsto pass through the exit slit. shorter wavelengths as temperature increases. decompose and atomize a sample for atomic temperature,emission is observed from
Optical fiber - Fiber that carries light by total Lasers provide highintensity,coherent, spectroscopy electronically excited atoms and ions.There is
internal reflection becausethe transparent monochromatic radiation by stimulated little chemical interference in an inductively
emissionfrom a medium in which an excited Heisenberg uncertainty principle - Certain coupled plasma,the temperature is very
core has a higher refractive index than the pairs of physical quantitiescannot be known
surroundingcladding. state has been pumped to ahigher population stable, and little self-absorption is observed.
than that of a lower state. Sample cells must simultaneously with arbitrary accuracy. If dE is Plasma emission spectroscopy does not
Optode - A sensor based on an optical fi ber. betransparent to the radiation of interest. A the uncertaintyin the energy difference require a light source andis capable of
Also called optrode reference sample compensatesfor reflection between two atomic states and dt is the measuring |70 elements simultaneously with a
Photoconductive detector - A detector whose and scattering by the cell and solvent. A lifetimeof the excited state, their product chargeinjection device detector. Background
conductivity changes whenlight is absorbed by gratingmonochromator disperses light into its cannot be known more accuratelythan dEdt $ correction for a given emissionpeak is based
the detector material. component wavelengths.The finer a grating is h/(4p), where h is Planck’s constant. A similar on subtracting the intensity of neighboring
ruled, the higher the resolution and the relation holdsbetween the position and the pixels in thedetector. The lowest detection
Photodiode array - An array of greaterthe dispersion of wavelengths over momentum of a particle. If position is limits are obtained by directing theplasma
semiconductor diodes used to detect light.The angles. Narrow slits improveresolution but knownvery accurately, then the uncertainty in into a mass spectrometer that separates and
array is normally used to detect light that has increase noise, because less light reaches the momentum is large, and vice versa. measures ions from the plasma. In flame and
been spread into itscomponent wavelengths. detector.A bandwidth that is one-fi fth of the furnace atomic absorption spectroscopy,a
One small band of wavelengths falls on each Inductively coupled plasma - A high-
width of the spectral peak isa good temperature plasma that derives itsenergy hollow-cathode lamp made of the analyte
detector. compromise between maximizing signal-to- element providesspectral lines sharper than
from an oscillating radio-frequency fi eld. It is
Photomultiplier tube - One in which the noise ratio andminimizing peak distortion. used to atomize asample for atomic emission those of the atomic vapor. The
cathode emits electrons whenstruck by light. Stray light introduces absorbance errorsthat spectroscopy. inherentlinewidth of atomic lines is limited by
The electrons then strike a series of dynodes are most serious when the transmittance of a the Heisenberg uncertaintyprinciple. Lines in a
(plates thatare positive with respect to the sample is verylow. Filters pass wide bands of Ionization interference - In atomic fl ame, furnace, or plasma are broadened by
cathode), and more electrons are wavelength and reject other bands.A spectroscopy, loss of signal intensityas a result a
releasedeach time a dynode is struck. As a photomultiplier tube is a sensitive detector of of ionization of analyte atoms. factor of 10–100 by the Doppler effect and
result, more than 106 electrons mayreach the visible and ultravioletradiation; photons cause Ionization suppressor - An element used in by atomic collisions. Correctionfor background
anode for every photon striking the cathode. electrons to be ejected from a metallic atomic spectroscopy todecrease the extent of emission from the flame is possible by
cathode. The signal is amplified at each ionization of the analyte. electricallypulsing the lamp on and off or
Phototube - A vacuum tube with a successive dynode on whichthe photoelectrons
photoemissive cathode. The electric current Isobaric interference - In mass spectrometry, mechanically chopping thebeam. Light
impinge. Photodiode arrays and charge scattering and spectral background can be
flowing between the cathode and the anode coupleddevices are solid-state detectors in overlap of two peaks withnearly the same
is proportional to theintensity of light striking mass. For example, 41K1 and 40ArH1 differ subtracted bymeasuring absorption with a
which photons create electrons andholes in deuterium lamp or by Zeeman background
the cathode. semiconductor materials. Coupled to a by 0.01atomic mass unit and appear as a
single peak unless the spectrometerresolution correction, in which the atomic energy levels
Photovoltaic detector - A photodetector with polychromator, thesedevices can record all are alternatelyshifted in and out of resonance
a junction across which thevoltage changes wavelengths of a spectrum simultaneously, is great enough to separate them.
with the lamp frequency by a magnetic field.
when light is absorbed by the detector withresolution limited by the number and Laser-induced breakdownspectroscopy - Chemical interference can be reduced by
material spacing of detector elements. An image Semiquantitative measurementof elements in a addition ofreleasing agents, which prevent the
Polychromator - A device that spreads light intensified charge coupled device converts surface by vaporizing a small patch with a analyte from reacting with interferingspecies.
into its component wavelengthsand directs incoming photons to electrons, amplifies the short laserpulse and measuring atomic Ionization interference in flames is suppressed
each small band of wavelengths to a different electrons with a microchannel plate, converts emission from the plasma above the surface by addingeasily ionized elements such as
regionwhere it is detected by a photodiode electrons back to photons with a fluorescent matrix - The medium containing analyte (that Cs.In inductively coupled plasma–mass
array. screen, and then detectsthe photons. Common is, everything in the sampleother than spectrometry, isobaricinterference occurs
infrared detectors include thermocouples, analyte). For many analyses, it is important between species with the same mass and
Refraction - Bending of light when it passes ferroelectric materials, and photoconductive that standards beprepared in the same charge.Interference can be eliminated if mass
between media with differentrefractive and photovoltaic devices.Photoacoustic matrix as the unknown. spectral resolution is suffi -
indexes. spectroscopy detects infrared absorption by ciently great, but this is not often the case. A
Matrix modifier - Substance added to sample
Resolution - How close two bands in a convertingsome absorbed energy to sound for atomic spectroscopy tomake the matrix dynamic reaction cellreduces isobaric
spectrum or a chromatogram canbe to each waves that are detected. Cavity more volatile or the analyte less volatile so interference by using an exothermic reaction
other and still be seen as two peaks. In ringdownspectroscopy measures very low that the matrixevaporates before analyte of agas such as NH3, N2O, or CO to remove
chromatography, it is defined as the absorption by gases or liquidsby circulating a does. interfering molecular ionssuch as ArO1 or to
difference in retention times of adjacent laser beam through the sample thousands of transform analyte into a molecular ion that
peaks divided bytheir width. In mass times tomultiply the pathlength. The decay Nebulization - In atomic spectroscopy, this canbe measured without interference.For
spectrometry, resolution is the smallest time of the circulating laser poweris a device breaks the liquid sample into a mist of qualitative and semiquantitative analysis,
difference inm/z values that can be detected measure of absorption by the sample. fine droplets. solids and liquidscan be sampled by laser
as separate peaks. Report the m/z value Piezoelectric crystal - A crystal that deforms ablation. Ablated material is swept throughan
atwhich resolution is measured. CHAPTER 21 when an electric field isapplied. inductively coupled plasma to detect multiple
Ablation - Vaporization of a small volume of elements by atomicemission or mass
Root-mean-square (rms) noise - Standard Plasma - A gas that is hot enough to contain spectrometry. When emission is measured, the
deviation of the noise in a region material by a laser pulse. free ions and electrons, aswell as neutral technique is called laser-induced breakdown
where the signal is flat: Aerosol - A suspension of very small liquid or molecules. spectroscopy. Matrixmatchedstandards are
solid particles in air or gas.Examples include Pressure broadening - In spectroscopy, line necessary for semiquantitative analysis.X-ray
where Ai is the measured signal for the ith fog and smoke. broadening due to collisionsbetween fl uorescence is the emission of X-rays by a
data point, A is the mean signal,and n is the molecules. material afterit has absorbed X-rays.
number of data points. Atomic absorption - A technique in which the
absorption oflight by free gaseous atoms in a Releasing agent - In atomic spectroscopy, a Handheld spectrometers enable
Signal averaging - Improvement of a signal flame or furnace is used to measure substance that prevents chemicalinterference. qualitativeand semiquantitative measurements
by averaging successivescans. The signal theconcentration of atoms of elemental composition in the field. Each
increases in proportion to the number of scans Self-absorption - In a luminescence element emits characteristic, narrow X-ray
accumulated.The noise increases in proportion Atomic emission - A technique in which the measurement, a high concentration ofanalyte lines, whoseenergies identify the element and
to the square root of the number ofscans. emission oflight by thermally excited atoms in molecules can absorb energy from excited whose integrated intensities areproportional
Therefore, the signal-to-noise ratio improves a plasma, flame, or furnace is used tomeasure analyte. Also calledinner filter effect. If the to the concentration of the element. Many
in proportion to thesquare root of the number the concentration of atoms. absorbed energy is dissipated as heat elementscan be measured at levels below
of scans collected. Atomic fluorescence - A technique in which instead of light, fluorescence does not 0.01 wt% with |10% accuracywhen suitable
electronic transitionsof atoms in a flame, increase in proportion to analyte care is exercised. X-rays from the
Snell’s law - concentration. Analyte concentration can be so
furnace, or plasma are excited by light, and spectrometersource ionize electrons from the
Stray light -Unintentional light that reaches the fluorescence is observed at a right angle high that fluorescence decreases with inner (K and L) shells of atoms.When outer
thedetector but has not passed through the to the incident beam. increasing concentration. In flame emission electrons fall into the vacancies, characteristic
sample or is not part of the narrow set of atomic spectroscopy, there is alower X-rayscalled Ka, Kb, La, and Lb are emitted.
wavelengths expected from the Atomization - Process in which a compound is concentration of excited-state atoms in the
decomposed into its atomsat high You must verify that bothexpected lines for
monochromator. cool, outer part of the flame than in the hot, each element are observed in the spectrum
temperature. inner flame. The cool atoms can absorb
Surface plasmon resonance - A sensitive beforeyou conclude that the element is
means to measure the bindingof molecules to Background correction - In atomic emission from the hot ones and thereby present.
a thin gold layer on the underside of a prism. spectroscopy, a means of distinguishingsignal decrease the observed signal.
Lightdirected through the prism is reflected due to analyte from signal due to absorption, Spectral interference - In atomic spectroscopy,
from the gold surface. There is onenarrow emission, or scattering by the flame, furnace, any physical process thataffects light intensity
range of angles at which reflection is nearly 0 plasma, or sample matrix. at the analytical wavelength. Created by
because the goldabsorbs the light to set up Boltzmann distribution - Relative population substancesthat absorb, scatter, or emit light of
oscillations (called plasmons) of the electron of two states at thermalequilibrium: the analytical wavelength.
cloud in the metal. When a thin layer of X-ray fluorescence - Emission of X-rays from
material (such as protein or DNA)binds to the where Ni is the population of the state, gi is a material following theabsorption of X-rays
side of the gold away from the prism, the the degeneracy of the state, Eiis the energy by the material.
electrical propertiesof the gold are changed of the state, k is Boltzmann’s constant, and T is
and the reflectivity changes. temperaturein kelvins. Degeneracy refers to In atomic spectroscopy, absorption, emission,
the number of states with the same energy. or fl uorescence from gaseous atoms is
CHAPTER 22 Ions are created or desorbed in the ion collision-induced dissociation. Desorption insensitive to almost everything other than the gas (or, very slowly, in a solid). Diffusion coefficient
source of a mass spectrometer. Neutral molecules are electrospray ionization (DESI) - A solvent is intended analyte. Also called mass spectrometry– - Defined by Fick’s first law of diffusion: J = -
converted into ions by electron ionization (which electrosprayed onto a surface to dissolve analyte mass spectrometry (MS–MS) or tandem mass D(dc/dx), where J is the rate at which molecules
produces a molecular ion, M1?, and many fragments) from the surface into aerosol microdroplets, which can spectrometry. Three-dimensional quadrupole ion diffuse across a plane of unit area and dc/dx is the
or by chemical ionization (which tends to create MH1 be analyzed with a mas spectrometer. Direct trap – An instrument that separates gaseous ions by concentration gradient in the direction of diffusion
and few fragments). A magnetic sector mass analysis in real time (DART) - A DART source trapping them in stable three dimensional trajectories Distribution coefficient - For a solute partitioned
spectrometer separates gaseous ions by accelerating produces excited He or N2, which is directed at the inside a metal chamber to which a radio-frequency between two phases, the distribution coeffi cient is the
them in an electric field and defl ecting ions of surface of an object to be sampled in ambient electric field is applied. Application of an oscillating total concentration of all forms of solute in phase 2
different mass-to-charge (m/z) ratio through different atmosphere. The excited species react with ambient electric field between the ends of the chamber divided by the total concentration in phase 1. Eluate -
arcs in a magnetic field. Ions are detected by an moisture to create protonated water clusters that destabilizes the trajectories of ions with a particular What comes out of a chromatography column. Also
electron multiplier, which works like a photomultiplier react with analyte M to make MH1. The MH1 is mass-to-charge ratio, expelling them from the cavity called effluent. Eluent - Solvent applied to the
tube. A conversion dynode may be necessary prior to measured by mass spectrometry. Double-focusing and into a detector. Time-of-flight mass beginning of a chromatography column. Elution -
the electron multiplier to ensure that all ions produce mass spectrometer - A spectrometer that uses electric spectrometer - Ions of different mass accelerated Process of passing a liquid or a gas through a
a similar electrical response at the detector. The mass and magnetic sectors in series to obtain high through the same electric fi eld have different chromatography column. Extraction - Process in which
spectrum is a graph of detector response versus m/z resolution. Electron ionization - used to create ions velocities. The lighter ions move faster than the a solute is transferred from one phase to another.
value. A double-focusing mass spectrometer attains from gaseous molecules in the inlet of a mass heavier ions. The time-of-flight spectrometer finds the Analyte is sometimes removed from a sample by
high resolution by employing an electric sector with spectrometer by bombardment of the gas with high mass-to-charge ratio by measuring the time that each extraction into a solvent that dissolves the analyte.
the magnetic sector to select ions with a narrow range energy electrons. Electrospray ionization - method group of ions requires to travel a fixed distance to Gel permeation chromatography - See molecular
of kinetic energy. Other mass separators include the for interfacing liquid chromatography to mass the detector. Transmission quadrupole mass exclusion chromatography. Laminar flow - Motion
transmission quadrupole mass spectrometer, the time- spectrometry. A high potential applied to the liquid at spectrometer – A mass spectrometer that separates with a parabolic velocity profi le of fl uid through a
of-flight mass spectrometer, the three-dimensional the column exit creates charged droplets in a fi ne ions by passing them between four metallic cylinders tube. Motion is fastest at the center and zero at the
quadrupole ion trap, the linear ion trap, and the aerosol. Gaseous ions are derived from ions that to which are applied direct current and oscillating walls. Linear velocity - In chromatography, the
orbitrap. The time-of-fl ight instrument is capable of were already in the mobile phase on the column. It is electric fields. Resonant ions with the right mass-to- distance per unit time travelled by the mobile phase.
high acquisition rates and has a nearly unlimited common to observe protonated bases (BH1), ionized charge ratio pass through the chamber to the detector Also called linear fl ow rate, in contrast to volume
upper mass range. Resolving power is defi ned as acids (A2), and complexes formed between analyte, while nonresonant ions are deflected into the flow rate Longitudinal diffusion - Diffusion of solute
m/Dm or m/m1/2, where m is the mass being M (which could be neutral or charged), and stable cylinders and are lost. CHAPTER 23 A solute can be molecules parallel to the direction of travel through a
measured, Dm is the difference in mass between two ions such as NH14, Na1, HCO22, or CH3CO22 that extracted from one phase into another in which it is chromatography column. Mass transfer - Net
peaks that are separated with a 10% valley were already in solution. Extracted ion more soluble. The ratio of solute concentrations in movement of a molecule from one location to another
between them, and m1/2 is the width of a peak at chromatogram - Chromatogram made by collecting each phase at equilibrium is called the partition by mechanisms such as diffusion, convection, and
half-height. Time-of-flight and orbitrap spectrometers consecutive full-range mass spectra, but selecting just coefficient. If more than one form of the solute exists, deliberate mixing. In chromatography, mass transfer
can provide high-resolution spectra with part per one value of m/z for display. Most time is spent we use a distribution coefficient instead of a partition refers to movement of solute between the mobile and
million mass accuracy. In a mass spectrum, the monitoring values of m/z that will not be displayed. A coefficient. We derived equations relating the stationary phases. Molecular exclusion
molecular ion is found from the highest m/z value of selected ion chromatogram provides greater signal- fraction of solute extracted to the partition or chromatography - technique in which the stationary
any “signifi cant” peak that cannot be attributed to to-noise ratio than an extracted ion chromatogram distribution coefficient, volumes, and pH. Many small phase has a porous structure into which small
isotopes or background signals. For a given because all of the time is taken to monitor just one or extractions are more effective than a few large molecules can enter but large molecules cannot.
composition, you should be able to predict the a few values of m/z in the selected ion extractions. A metal chelator, soluble only in organic Molecules are separated by size, with larger
relative intensities of the isotopic peaks at M11, M12, chromatogram. Ion mobility spectrometer - An solvents, can extract metal ions from aqueous molecules moving faster than smaller ones. Also called
and so on. Among common elements, Cl and Br have instrument that measures the drift time of gaseous ions solutions, with selectivity achieved by adjusting pH. size exclusion, gel filtration, gel permeation, or size
particularly diagnostic isotope patterns. From a migrating in an electric fi eld against a flow of gas. Crown ethers and salts containing a hydrophobic ion exclusion chromatography. Open tubular column - In
molecular composition, the rings 1 double bonds The “spectrum” of detector current versus drift time is act as phase transfer agents to bring a hydrophilic chromatography, a hollow capillary column whose
equation helps us propose structures. An organic really an electropherogram of a gas. Linear ion from an aqueous phase into an organic phase. In walls are coated with stationary phase Partition
compound with an odd number of nitrogen atoms will quadrupole ion trap - An instrument that separates adsorption and partition chromatography, a chromatography - A technique in which separation is
have an odd mass. Fragment ions arising from bond gaseous ions by trapping them in stable trajectories continuous equilibration of solute between mobile and achieved by equilibration of solute between two
cleavage and rearrangements provide clues to inside a linear quadrupole by use of radio-frequency stationary phases occurs. Eluent goes into a column phases Partition coefficient K- The equilibrium
molecular structure. Gas emerging from a capillary fields. Ions can be expelled from the trap in order of and eluate comes out. Columns may be packed with constant for the reaction in which a solute is
gas chromatography column can go directly into the increasing m/z for mass spectrometry. Magnetic stationary phase, or may be open tubular with partitioned between two phases: solute (in phase 1) Δ
ion source of a well-pumped mass spectrometer to sector mass spectrometer - A device that separates stationary phase bonded to the inner wall. In ion- solute (in phase 2). Relative retention - In
provide qualitative and quantitative information gaseous ions that have the same kinetic energy by exchange chromatography, the solute is attracted to chromatography, the ratio of adjusted retention times
about the components of a mixture. For capillary passing them through a magnetic field perpendicular the stationary phase by coulombic forces. In molecular for two components. If component 1 has an adjusted
liquid chromatography with a flow rate of |500 to their velocity. Trajectories of ions with a certain exclusion chromatography, the fraction of stationary retention time of T’r1 and component 2 has an
nL/min, liquid can be admitted directly to a well mass-to-charge ratio are bent exactly enough to phase pore volume available to solute decreases as adjusted retention time of alpha = T’r2/T’r1, the
pumped electron ionization source. For liquid reach the detector. Other ions are deflected too much the size of the solute molecules increases. Affinity relative retention is a 5 t¿r2/t¿r1. Also called
chromatography with wider columns and higher fl ow or too little. Mass spectrometry - A technique in which chromatography relies on specific, noncovalent separation factor. Resolution - How close two bands
rates, electrospray ionization employs high voltage at gaseous molecules are ionized, accelerated by an interactions between the stationary phase and one in a spectrum or a chromatogram can be to each
the exit of the column, combined with coaxial N2 gas electric field, and then separated according to their solute in a complex mixture. The relative retention of other and still be seen as two peaks. In
fl ow, to create a fi ne aerosol of charged droplets mass-to-charge ratio. Mass-to-charge ratio, m/z - two components is the quotient of their adjusted chromatography, it is defined as the difference in
containing ions that were already present in the liquid The mass of an ion in daltons divided by the charge retention times. The retention factor for a single retention times of adjacent peaks divided by their
phase. Analyte is often associated with other ions to of the ion measured in multiples of the elementary component is the adjusted retention time divided by width. In mass spectrometry, resolution is the smallest
give species such as [MNa]1 or [M(CH3CO2)]2. charge. For 23Na1, for example, m/z ¯ 23/1 5 23. the elution time for solvent. Retention factor gives the difference in m/z values that can be detected as
Control of pH helps ensure that selected analytes are Matrix-assisted laser desorption/ionization (MALDI) ratio of time spent by solute in the stationary phase separate peaks. Report the m/z value at which
in anionic or cationic form. Another interface between - A gentle technique for introducing predominantly to time spent in the mobile phase. When a separation resolution is measured. Retention factor - In
liquid chromatography and mass spectrometry is singly charged, intact macromolecular ions into the is scaled up from a small load to a large load, linear chromatography, the adjusted retention time for a
atmospheric pressure chemical ionization, which gas phase. An intimate solid mixture of analyte plus a velocity is held constant and the crosssectional area peak divided by the time for the mobile phase to
utilizes a corona discharge to create a variety of large excess of a small, ultraviolet-absorbing of the column should be increased in proportion to the travel through the column. Retention factor is also
gaseous ions from aerosol droplets. Both atmospheric molecule is irradiated by a pulse from an ultraviolet loading. Plate height (H 5 s2/x) is related to the equal to the ratio of the time spent by the solute in
pressure chemical ionization and electrospray tend to laser. The small molecule (the matrix) absorbs the breadth of a band emerging from the column. The the stationary phase to the time spent in the mobile
produce unfragmented ions. Collisionally activated radiation, becomes ionized, evaporates, and expands smaller the plate height the sharper the band. The phase. Also called capacity factor, capacity ratio, and
dissociation to produce fragment ions is controlled by in a supersonic jet that carries analyte into the gas number of plates for a Gaussian peak is N 5 partition ratio. Retention time - The time from
the cone voltage at the mass spectrometer inlet. phase. Matrix ions apparently transfer charge to the 5.55t2r/w21/2. Plate height is approximately the injection for an individual solute to reach the detector
Electrospray of proteins typically creates an array of analyte. Molecular ion - In mass spectrometry, an ion length of column required for one equilibration of of a chromatography column. Retention volume - The
highly charged ions such as MHn1n. Matrix-assisted that has not lost or gained any atoms during solute between mobile and stationary phases. volume of solvent needed to elute a solute from a
laser desorption/ionization is a gentle way to ionization. Nitrogen rule - A compound with an odd Resolution of neighboring peaks is the difference in chromatography column. Silanization - Treatment of
produce predominantly singly charged, intact protein number of nitrogen atoms—in addition to C, H, retention time divided by the average width a chromatographic solid support or glass column with
ions for mass spectrometry. A reconstructed total ion halogens, O, S, Si, and P—will have an odd nominal (measured at the baseline, w54s). Resolution is hydrophobic silicon compounds that bind to the most
chromatogram shows the signal from all ions above a mass. A compound with an even number of nitrogen proportional to 1N. Doubling the length of a column reactive Si—OH groups. It reduces irreversible
chosen m/z emerging from chromatography as a atoms (0, 2, 4, …) will have an even nominal mass. increases resolution by 12. Some retention (k < 0.5 to adsorption and tailing of polar solutes. Stationary
function of time. An extracted ion chromatogram Nominal mass - Integer mass of the species with the 20) is essential to achieve resolution, but greater phase – phase that does not move through the
shows the signal for one ion taken from the complete most abundant isotope of each of the constituent retention increases separation time without substantial column. Van Deemter equation - Describes the
mass spectrum. Selected ion monitoring of one or a atoms. For C, H, and Br, the most abundant isotopes improvements in resolution. Relative retention (a) dependence of chromatographic plate height, H, on
few values of m/z improves signal-to-noise because are 12C, 1H, and 79Br. Therefore, the nominal mass strongly influences chromatographic resolution. The linear velocity, ux: H = A +B/ux +Cux. The constant A
all of the time is spent measuring just that one or a of C2H5Br is (2 3 12) 1 (5 3 1) 1 (1 3 79) 5 108. standard deviation of a diffusing band of solute is depends on band-broadening processes such as
few ions. In selected reaction monitoring, a precursor Orbitrap mass spectrometer - A device that traps s512Dt, where D is the diffusion coefficient and t is multiple fl ow paths that are independent of flow
ion isolated by one mass fi lter passes into a collision ions in stable orbits around a central electrode. Ions time. The van Deemter equation describes band rate. B depends on the rate of diffusion of solute in
cell in which it breaks into products. One (or more) oscillate from one end of the trap to the other, broadening on the chromatographic column: H < A 1 the mobile phase. C depends on the rate of mass
product ion is then selected by a second mass fi lter inducing image currents in the outer electrodes. B/ux 1 Cux, where H is plate height, ux is linear transfer (that is, the equilibration time) between the
for passage to the detector. This process is extremely Fourier analysis of the image currents gives m/z for velocity, and A, B, and C are constants. The first term stationary and mobile phases. CHAPTER 24 In gas
selective for just one analyte and vastly increases the the oscillating ions. Precursor ion - In tandem mass represents irregular flow paths, the second, chromatography, a volatile liquid or gaseous solute is
signal-to-noise ratio for this analyte. Electrontransfer spectrometry (selected reaction monitoring), the ion longitudinal diffusion, and the third, the finite rate of carried by a gaseous mobile phase over a stationary
dissociation is employed in protein sequencing to selected by the fi rst mass separator for mass transfer of solute between mobile and phase on the inside of an open tubular column or on a
break amide bonds in a polypeptide without fragmentation in the collision cell. Product ion - In stationary phases. The optimum flow rate, which solid support. Long, narrow, fused-silica open tubular
breaking other bonds. Several methods can ionize tandem mass spectrometry (selected reaction minimizes plate height, is faster for gas columns have low capacity but give excellent
molecules from the surface of an object in ambient monitoring), the fragment ion from the collision cell chromatography than for liquid chromatography. The separation. They can be wall-coated or porous layer.
atmosphere. Direct analysis in real time (DART) and a selected by the fi nal mass separator for passage number of plates and the optimal flow rate increase Packed columns provide high capacity but poor
low-temperature plasma use electronically excited through to the detector. Reconstructed total ion as the stationary phase particle size or open tubular resolution. Solutes must be partially in the vapour
helium or nitrogen to ionize analytes. Desorption chromatogram - In chromatography, a graph of the column diameter decreases. In gas chromatography, phase to be eluted from a gas chromatography
electrospray ionization (DESI) directs electrosprayed sum of intensities of all ions detected at all masses open tubular columns can provide higher resolution or column. Solute partial pressure and retention are
solvent onto a surface to dislodge ions. An ion (above a selected cutoff) versus time. Resolving shorter analysis times than packed columns. Bands governed by thermodynamics, meaning that column
mobility spectrometer separates gas-phase ions by power - In mass spectrometry, resolving power can spread during injection, during detection, and in temperature controls retention. Each liquid stationary
their different mobilities in an electric field at be defined as m/Dm, where Dm is the separation of connecting tubing, as well as during passage through phase most strongly retains solutes in its own polarity
atmospheric pressure. Field asymmetric waveform two peaks when the overlap at the base is 10% of the separation column. The observed variance of the class (“like dissolves like”). Solid stationary phases,
spectrometry separates ions based on differences of the peak height and m is the smaller of the two m/z band is the sum of the variances for all mechanisms of including porous carbon, alumina, and molecular
mobility in a high electric field at which mobility is not values. Alternatively, resolving power can be taken as spreading. Fronting and tailing can be corrected by sieves, are used to separate permanent gases. The
a linear function of electric field. Ion mobility can add m/m1/2, where m1/2 is the width of a peak at half injecting less sample and by masking strong retention index measures elution times in relation to
another dimension to mass spectrometry by the maximum height. In this case, the dip between two adsorption sites on the stationary phase. Adjusted those of linear alkanes. Temperature or pressure
separating isobaric ions according to differences in barely resolved peaks is 8% below the peak heights. retention time = Tr’ = tr - tm, where tr is the retention programming reduces elution times of strongly
mobility. Atmospheric pressure chemical ionization Rings 1 double bonds formula - The number of rings time of a solute and tm is the time needed for mobile retained components. Without compromising
- A method to interface liquid chromatography to 1 double bonds in a molecule with the formula phase to travel the length of the column. Adsorption separation effi ciency, the linear velocity may be
mass spectrometry. Liquid is nebulized into a fine CcHhNnOx is c 2 h/2 1 n/2 1 1, where c includes all chromatography - solute equilibrates between the increased when H2 or He, instead of N2, is used as
aerosol by a coaxial gas flow and the application of Group 14 atoms (C, Si, Ge, Sn, Pb, which all make mobile phase and adsorption sites on the stationary carrier gas. Split injection provides high-resolution
heat. Electrons from a high-voltage corona discharge four bonds), h includes H 1 halogens (which make one phase. Affinity chromatography - a particular solute separations of relatively concentrated samples.
create cations and anions from analyte exiting the bond), and n is the number of Group 15 atoms (N, P, is retained by a column by virtue of a specifi c Splitless injection of very dilute samples requires
chromatography column. The most common species As, Sb, Bi, which make three bonds). Group 16 atoms interaction with a molecule covalently bound to the solvent trapping or cold trapping to concentrate
observed with this interface is MH1, the protonated (which make two bonds) do not affect the result. stationary phase. Asymmetry factor – In C, the solutes at the start of the column to give sharp bands.
analyte, with little fragmentation. Base peak - Most Selected ion chromatogram - A graph of detector parameter describing the shape of a peak. B is the On-column injection is best for quantitative analysis
intense peak in a mass spectrum. Chemical ionization response versus time when a mass spectrometer distance from the peak apex to the back of the peak and for thermally unstable solutes. Quantitative
- gentle method of producing ions for a mass monitors just one or a few species of selected mass- measured at 10% of peak height. A is the distance analysis is usually done with internal standards in gas
spectrometer without extensive fragmentation of the to-charge ratio, m/z, emerging from a from the peak apex to the front of the peak at 10% chromatography. Coelution of an unknown peak with
analyte molecule, M. A reagent gas such as CH4 is chromatograph. Also called mass chromatogram. of peak height. B/A 5 1 is a symmetrical peak. B/A . a spike of a known compound on several columns with
bombarded with electrons to make CH15, which Selected ion monitoring - Use of a mass 1 is tailing and B/A , 1 is fronting Baseline resolution different retention mechanisms is useful for qualitative
transfers H1 to M, giving MH1. Collisionally spectrometer to monitor species with just one or a few – In C, when two adjacent peaks are suffi ciently identifiation of the unknown peak. Mass spectral
activated dissociation - Fragmentation of an ion in a mass-to-charge ratios, m/z. Selected reaction resolved that the signal between the peaks returns to detection provides qualitative information to help
mass spectrometer by high-energy collisions with gas monitoring - A technique in which the precursor ion the baseline (resolution . 1.5). Allows determination of identify unknowns. The mass spectrometer is more
molecules. In atmospheric pressure chemical ionization selected by one mass separator passes through a accurate peak areas. Chromatogram - graph sensitive and less subject to interference when
or electrospray interfaces, collisionally activated collision cell in which the precursor breaks into several showing chromatography detector response as a selected ion monitoring or selected reaction
dissociation at the inlet to the mass filter can be fragment ions (product ions). A second mass separator function of elution time or volume. Diffusion - Net monitoring is employed. Thermal conductivity
promoted by varying the cone voltage. In tandem then selects one (or a few) of these ions for detection. transport of a solute from a region of high detection has universal response but is not very
mass spectrometry, dissociation occurs in a collision Selected reaction monitoring improves concentration to a region of low concentration caused sensitive. Flame ionization detection is sensitive
cell between the two mass separators. Also called chromatographic signal-to-noise ratio because it is by the random movement of molecules in a liquid or
enough for most columns and responds to most trace analysis and quantitative analysis. The entire tmF, where F is the volume flow rate. Derivatization - electroosmotic mobility (which is the same for all
organic compounds. Electron capture, nitrogen- sample in a low-boiling solvent is directed to the Chemical alteration to attach a group to a molecule species). Zone dispersion (broadening) arises mainly
phosphorus, flame photometry, photoionization, column, where the sample is concentrated by solvent so that the molecule can be detected conveniently. from longitudinal diffusion and the finite length of the
chemiluminescence, and atomic emission detectors are trapping (condensing the solvent below its boiling Alternatively, treatment can alter volatility or injected sample. Stacking of solute ions in the
specific for certain classes of compounds or individual point) or coldntrapping (condensing solutes far below solubility to allow easier separation. Dwell volume - capillary occurs when the sample has a low
elements. You need to define the goal of an analysis their boiling range). The column is then warmed to Volume in chromatography between the point of conductivity. Electroosmoti flow is reduced at low pH
before developing a chromatographic method. One initiate separation. Stir-bar sorptive extraction - mixing solvents and the start of the column. because surface SiOO2 groups are protonated.
key to successful chromatography is a clean sample. Sample preparation method similar to solid phase Electrochemical detector - LC detector that measures SiOO2 groups can be masked by polyamine cations,
Solid-phase microextraction, stir-bar sorptive microextraction, except the sorptive layer is coated current when an electroactive solute emerges from the and the wall charge can be reversed by a cationic
extraction, purge and trap, and thermal desorption on the outside ofba stirring bar. Coating volume is column and passes over a working electrode held at surfactant that forms a bilayer along the wall.
can isolate volatile components from complex greater than the fi ber volume in solidphase a fixed potential with respect to a reference Covalent coatings reduce electroosmosis and wall
matrices. After sample preparation the remaining microextraction, so it provides ~102 times greater electrode. Also called amperometric detector. Eluent adsorption. Hydrodynamic sample injection uses
decisions for method development are to select a sensitivity for traces of analyte. Analyte is removed strength - A measure of the ability of a solvent to pressure; electrokinetic injection uses an electric field.
detector, a column, and the injection method, in that from the coating by thermal desorption for elute solutes from a normal-phase chromatography Ultraviolet absorbance is commonly used for
order. Carrier gas - Mobile phase gas in gas chromatography. Temperature programming - column. It is a measure of the solvent adsorption detection. Micellar electrophoretic chromatography
chromatography Cold trapping - Splitless gas Raising the temperature of a gas energy on bare silica. Evaporative light-scattering uses micelles as a pseudostationary phase to
chromatography injection technique in which solute is chromatographyncolumn during a separation to detector - LC detector that makes a fine mist of eluate separate neutral molecules and ions. Capillary gel
condensed far below its boiling point in a narrow reduce the retention time of late eluting components. and evaporates solvent from the mist in a heated electrophoresis separates macromolecules by sieving.
band at the start of the column. Electron capture Thermal conductivity - Rate at which a substance zone. The remaining particles of liquid or solid solute In contrast with molecular exclusion chromatography,
detector - Gas chromatography detector that is transports heat (energy per unit time per unit area) fl ow past a laser beam and are detected by their small molecules move fastest in gel electrophoresis.
particularly sensitive to compounds with halogen through a temperature gradient (degrees per unit ability to scatter the light. Gradient elution - Microfluidic devices (“lab-on-a-chip”) use
atoms, nitro groups, and other groups with high distance). Energy flow [J/(s x m2)] = –k(dT/dx), Chromatography in which the composition of electroosmotic or hydrodynamic flow in
electron affi nity. Makeup gas (N2 or 5% CH4 in Ar) where k is the thermal conductivity [W/(m x K)] and thenmobile phase is progressively changed to lithographically fabricated channels to conduct
is ionized by b-rays from 63Ni to liberate electrons dT/dx is the temperature gradient (K/m). CHAPTER increase the eluent strength of the solvent. High- chemical reactions and chemical analysis. Affinity
that produce a small, steady current. High-electron- 25 In high-performance liquid chromatography performance liquid chromatography - using very chromatography - A technique in which a particular
affi nity analytes capture some of the electrons and (HPLC), solvent is pumped at high pressure through a small stationary phase particles and high pressure to solute is retained by a column by virtue of a specific
reduce the detector current. Flame ionization column containing stationary phase particles with force solvent through the column. Hydrophilic interaction with a molecule covalently bound to the
detector - A GC detector in which solute is burned in diameters of 1.5–5 mm. The smaller the particle size, interaction chromatography - Chromatographic stationary phase. Anion exchanger - An ion
an H2-air flame to produce CHO1 ions. The current the more effi cient the column, but the greater the separation of polar solutes with a hydrophilic exchanger with positively charged groups covalently
carried through the flame by these ions is resistance to fl ow. Microporous silica particles with a stationary phase using mixed organicaqueous eluent. attached to the support. It can reversibly bind anions.
proportional to the concentration of susceptible covalently bonded liquid phase such as octadecyl Eluent strength increases with decreasing organic Capillary electrophoresis - Separation of a mixture
species in the eluate. Gas chromatography - A form groups (—C18H37) are most common. Polarity index solvent. Commonly called HILIC. Isocratic elution - into its components by a strong electric fi eld imposed
of chromatography in which the mobile phase is a measures the ability of a solvent to dissolve polar Chromatography using a single solvent (or single between the two ends of a narrow capillary tube fi
gas. General elution problem - In chromatography, molecules. Mobile phase strength measures the ability solvent mixture) for the mobile phase. Microporous lled with electrolyte solution. Unlike chromatography,
the inability of a single isothermal or isocratic of a given solvent to elute solutes from the column. In particles - Chromatographic stationary phase there is no stationary phase in electrophoreses.
condition to provide adequate separation in normal-phase chromatography and hydrophilic consisting of porous particles 1.5–10 mm in diameter, Solutes are separated by differences in mobility.
reasonable time for solutes with a wide range of interaction chromatography, the stationary phase is with high efficiency and high capacity for solute. Capillary gel electrophoresis - A form of capillary
retention. Guard column - In high-performance liquid polar and a less polar solvent is used. Mobile phase Mobile phase strength - Refers to the ability of a electrophoresis in which the tube is fi lled with a
chromatography, a short column packed with the strength increases as the polarity of the mobile phase to elute a particular compound from a polymer gel that serves as a sieve for
same material as the main column and placed solventincreases. Reversed-phase chromatography liquid chromatography column. Also called solvent macromolecules. The largest molecules have the
between the injector and the main column. The guard employs a nonpolar stationaryphase and polar strength. Normal-phase chromatography - A slowest migration through the gel. Also called
column removesnimpurities that might irreversibly bind solvent. Mobile phase strength increases as the chromatographic separation utilizing a polar capillary sieving electrophoresis. Capillary zone
to the main column and degrade it. Also called polarity of the solvent decreases. Most separations of stationary phase and a less polar mobile phase electrophoresis - A form of capillary electrophoresis
precolumn. In gas chromatography, the guard column organic compounds can be done on reversed-phase Polarity index - In partition liquid chromatography, in which ionic solutes are separated because of
is a length of empty, silanized capillary ahead of the columns. Polar compounds that are not retained on refers to the ability of a solvent to dissolve polar differences in their electrophoretic mobility. Cation
chromatography column. Nonvolatile residues reversed-phase columns can be separated by compounds. Polarity arises from dipole, hydrogen- exchanger - An ion exchanger with negatively
accumulate in the guard column. Headspace - Gas hydrophilic interaction chromatography. Polar bond-donating, and hydrogen-bond-accepting charged groups covalently attached to the support. It
phase above a solid or liquid sample within a sealed compounds that only dissolve in organic solvents can interactions. Refractive index detector - Liquid can reversibly bind cations. Cross-linking - Covalent
vial. The concentration of a volatile component in the be separated by normal-phase chromatography. chromatography detector that measures the change in linkage between different strands of a polymer.
headspace is proportional to its concentration in the Normal-phase chromatography or porous graphitic refractive index of eluate as solutes emerge from the Donnan exclusion - Ions of the same charge as those
solid or liquid. Molecular sieve - crystalline solid carbon is good forseparating isomers. Chiral phases column. Reversed-phase chromatography - Liquid fixed on an ion exchange resin are repelled from the
particle with pores the size of small molecules. are used for optical isomers. If a solution of organic chromatography in which the stationary phase is less resin. Thus, anions do not readily penetrate a cation-
Zeolites (sodium aluminosilicates) are a common type. solvent and water is used in reversed phase polar than the mobile phase. Supercritical fluid - fluid exchange resin, and cations are repelled from an
On-column injection - Technique used in gas chromatography, mobile phase strength increases as whose temperature is above its critical temperature anion exchange resin. Electrokinetic injection - In
chromatography to place a thermally unstable the percentage of organic solvent increases. If the and whose pressure is above its critical pressure. It capillary electrophoresis, the use of an electric field
sample directly on the column without excessive solvent has a fixed composition, the process is called has properties of both a liquid and a gas. to inject sample into the capillary. Because different
heating in an injection port. Solute is condensed at the isocratic elution. In gradient elution, mobile phase Superficially porous particle - A stationary phase species have different mobilities, the injected sample
start of the column by low temperature, and then the strength is increased during chromatography by particle for liquid chromatography containing a thin, does not have the same compositionnas the original
temperature is raised to initiate chromatography. increasing the percentage of strong (less polar) porous outer layer and a dense, nonporous core. sample. Electroosmosis - Bulk flow of fluid in a
Open tubular column - a hollow capillary column solvent. A short guard column containing the same Mass transfer is faster in the superfi cally porous capillary tube induced by an electric field. Mobile
whose walls are coated with stationary phase Packed stationary phase as the analytical column is placed particle than in a fully porous particle of the same ions in the diffuse part of the double layer at the wall
column - column filled with stationary phase particles. before the analytical column to protect it from diameter. Also called fused-core particle. Ultra-high- of the capillary serve as the “pump.” Also called
Phase ratio - the relative volume of mobile phase to contamination with particulate matter or irreversibly performance liquid chromatography – Liquid electroendosmosis Electrophoresis - Migration of ions
stationary phase in the column. In older literature, the adsorbed solutes. A high-quality pump provides chromatography with stationary phase particles in the in solution in an electric fi eld. Cations move toward
phase ratio may be given as Phi which is the smooth solvent flow. The injection valve allows rapid, 1.5–2 microm range. Higher pressure is required to the cathode and anions move toward the anode. Ions
reciprocal of Beta m. Purge and trap - A method for precise sample introduction. The column is best housed force liquid through a UHPLC column than through can be separated from one another by their differing
removing volatile analytes from liquids or solids, in an oven to maintain a reproducible temperature. columns with larger particle size. Ultraviolet detector rates of migration in a strong electric field.
concentrating the analytes, and introducing them into Column efficiency increases with smaller particles and – LC detector that measures ultraviolet absorbance of Equivalent - redox reaction, amount of reagent that
a gas chromatograph. A carrier gas bubbled through at elevated temperature because the rate of mass solutes emerging from the column CHAPTER 25 Ion- can donate or accept one mole of electrons. For an
a liquid or solid extracts volatile analytes, which are transfer between phases is increased. Mass exchange chromatography employs resins and gels acid-base reaction, the amount of reagent that can
then trapped in a tube containing adsorbent. After spectrometric detection provides quantitative and with covalently bound charged groups that attract donate or accept one mole of protons. For ion
analyte has been collected, the adsorbent tube is qualitative information for each substance eluted solute counterions (and that exclude ions having the exchange, the amount of reagent that will exchange
heated and purged in the reverse direction to desorb from the column. Ultraviolet detection is most common, same charge as the resin). Ion-exchange binding with one mole of monovalent cation or anion. Gel -
the analytes, which are collected by cold trapping in and it can provide qualitative information if a increases with increased ion charge and Chromatographic stationary phase particles, such as
a gas chromatography column. Retention gap - In photodiode array is used to record a full spectrum of polarizability, and it depends inversely on the radius Sephadex ormpolyacrylamide, which are soft and
gas chromatography, a 3- to 10-m length of empty, each analyte. Refractive index detection has universal of the hydrated ion. Ion binding results in release of pliable. Gradient elution - Chromatography in which
silanized capillary ahead of the chromatography response but is not very sensitive. Evaporative light an equivalent ionic charge from the resin. Polystyrene the composition of the mobile phase is progressively
column. The retention gap improves the peak shape scattering and the charged aerosol detectors respond resins are useful for small ions. Greater cross-linking changed to increase the eluent strength of the solvent.
of solutes that elute close to solvent when large to the mass of each nonvolatile solute. Electrochemical of the resin increases the capacity, selectivity, and Hydrodynamic injection - In capillary
volumes of solvent are injected or when the solvent and fl uorescence detectors are extremely sensitive, time needed for equilibration. Ion-exchange gels electrophoresis, the use of a pressure difference
has a very different polarity from that of the but selective. In supercritical fluid chromatography, based on cellulose and dextran have large pore sizes between the two ends of the capillary to inject
stationary phase. Sample preparation - Transforming nonvolatile solutes are separated by a process whose and low charge densities, suitable for the separation sample into the capillary. Injection is achieved by
a sample into a state that is suitable for analysis. This efficiency, speed, and detectors more closely of macromolecules. Certain inorganic solids have ion- applying pressure on one end, by applying suction on
process can include concentrating a dilute analyte resemble those of gas chromatography than those of exchange properties and are useful at extremes of one end, or by siphoning. Hydrophobic interaction
and removing or masking interfering species. Selected liquid chromatography. Steps in method development: temperature or radiation. Ion exchangers operate by chromatography - Chromatographic separation
ion monitoring - Use of a mass spectrometer to (1) determine the goal of the analysis, (2) select a the principle of mass action, with a gradient of based on the interaction of a hydrophobic solute with
monitor species with just one or a few mass-to-charge method of sample preparation, (3) choose a detector, increasing ionic strength most commonly used to effect a hydrophobic stationary phase. Indirect detection –
ratios, m/z. Selected reaction monitoring - A and (4) use a systematic procedure to select the a separation. Zwitterionic hydrophilic interaction C detection based on the absence of signal from a
technique in which the precursor ion selected by one mobile phase for isocratic or gradient elution. chromatography can separate anions and cations on background species. For example, in ion
mass separator passes through a collision cell in which Aqueous acetonitrile, methanol, and tetrahydrofuran the same column. In suppressed-ion chromatography, chromatography, a light-absorbing ionic species can
the precursor breaks into several fragment ions are customary solvents for reversed-phase a separator column separates ions of interest, and a be added to the eluent. Nonabsorbing analyte
(product ions). A second mass separator then selects separations. Retention is optimized by varying the suppressor membrane converts eluent into a nonionic replaces an equivalent amount of light-absorbing
one (or a few) of these ions for detection. Selected amount of organic solvent in the mobile phase. form so that analytes can be detected by their eluent when analyte emerges from the column,
reaction monitoring improves chromatographic signal- Relative retention is adjusted by fine-tuning the electrical conductivity. Eluent and suppressor can be thereby decreasing the absorbance of eluate.
to-noise ratio because it is insensitive to almost organic solvent concentration, altering the generated continuously by electrolysis. Alternatively, Interstitial volume - The volume of mobile phase
everything other than the intended analyte. Also temperature, and by choice of a new organic solvent nonsuppressed ion chromatography uses an ion- outside of the stationary phase in a molecular
called mass spectrometry–mass spectrometry (MS– or column. Column length and particle size can be exchange column and low-concentration eluent. Ion- exclusion chromatography column. The total volume of
MS) or tandem mass spectrometry. Septum - A disk, changed to improve efficiency or speed of analysis. pair chromatography employs an ionic surfactant in mobile phase is the interstitial volume plus the volume
usually made of silicone rubber, covering the injection Criteria for a successful separation are 0.5 # k # 20, the eluent to make a reversed-phase column function of solvent inside the stationary phase particles.
port of a gas chromatograph. The sample is injected resolution $ 2.0, operating pressure # 15 MPa (for as an ion-exchange column. Molecular exclusion Formerly called void volume. Ion chromatography -
by syringe through the septum. Solid-phase conventional equipment), and asymmetry factor in the chromatography is used for separations based on HPLC ion exchange separation of ions. See also
microextraction - Extraction of compounds from range 0.9–1.5. Retention of weak acids and bases is size and for molecular mass determinations of suppressed-ion chromatography and single-column
liquids or gases into a coated fiber dispensed from a governed by their ionization state, which is controlled macromolecules. Molecular exclusion is based on the ion chromatography. Ion-exchange chromatography
syringe needle. After extraction, the fiber is by the mobile phase pH. In selecting a mobile phase inability of large molecules to enter pores in the - A technique in which solute ions are retained by
withdrawn into the needle and the needle is inserted buffer, consider the desired pH, buffering capacity, stationary phase. Small molecules enter these pores oppositely charged sites in the stationary phase Ion-
through the septum of a chromatograph. The fiber is solubility, and compatibility with detection. A wide and therefore exhibit longer elution times than large pair chromatography - Separation of ions on
extended inside the injection port and adsorbed gradient is a good first choice to determine whether molecules. In affinity chromatography, the stationary reversed-phase highperformance liquid
solutes are desorbed by heating (for gas to use isocratic or gradient elution. If it is not possible phase retains one particular solute in a complex chromatography column by adding to the eluent a
chromatography) or solvent (for liquid to separate all components with 0.5 # k # 20, mixture. After all other components have been eluted, hydrophobic counterion that pairs with analyte ion
chromatography). Solvent trapping - Splitless gas gradient elution is needed. Retention factors the desired species is liberated by a change in and is attracted to stationary phase. Micellar
chromatography injection technique in which solvent is measured at a few solvent compositions can be fi t to conditions. In hydrophobic interaction electrokinetic chromatography - A form of capillary
condensed below its boiling point at the start of the the linear solvent-strength model to enable computer chromatography, high concentrations of ammonium electrophoresis in which a micelle-forming surfactant is
column. Solutes dissolve in a narrow band in the optimization of a separation. Bonded stationary sulfate induce proteins to adhere to a hydrophobic present. Migration times of solutes depend on the
condensed solvent Spiking - Addition of a known phase - In high-performance liquid chromatography, stationary phase. A gradient of decreasing salt fraction of time spent in the micelles. Micelle - An
compound (usually at a known concentration) to an a stationary liquid phase covalently attached to the concentration is applied to increase the solubility of aggregate of molecules with ionic headgroups and
unknown. In isotope dilution mass spectrometry, the solid support Charged aerosol detector - Sensitive, proteins in water and to elute them from the column. long, nonpolar tails. The inside of the micelle
spike is the added, unusual isotope. Spike is a noun nearly universal detector for liquid chromatography In capillary zone electrophoresis, ions are separated resembles hydrocarbon solvent, whereas the outside
and a verb. In co-chromatography, spiking is the in which solvent is evaporated from eluate to leave by differences in mobility in a strong electric field interacts strongly with aqueous solution. Mobility -
simultaneous chromatography of a known compound an aerosol of fi ne particles of nonvolatile solute. applied between the ends of a silica capillary tube. The terminal velocity that an ion reaches in a field of
with an unknown. If a known and an unknown have Aerosol particles are charged by adsorption of N1 2 The greater the charge and the smaller the 1 V/m. Velocity = mobility * field. See also apparent
the same retention time on several columns, they are ions and fl ow to a collector that measures total hydrodynamic radius, the greater the electrophoretic mobility and electrophoretic mobility. Molecular
probably identical. Also called fortifi cation. Split charge reaching the detector versus time. Dead mobility. Normally, the capillary wall is negative, and exclusion chromatography - the stationary phase
injection - used in capillary gas chromatography to volume - Volume of mobile phase within a solution is transported from anode to cathode by has a porous structure into which small molecules can
inject a small fraction of sample onto the column; the chromatography column, equal to the entire space electroosmosis of cations in the electric double layer. enter but large molecules cannot. Molecules are
rest of the sample is blown out to waste. Splitless accessible to solvent and small solutes. Includes pores Solute cations arrive first, followed by neutral species, separated by size, with larger molecules moving
injection - used in capillary gas chromatography for in the particles and volume between particles. If time followed by solute anions (if electroosmosis is stronger faster than smaller ones. Also called size exclusion,
required for solvent or unretained solute to travel than electrophoresis, which typically is true). Apparent gel filtration, gel permeation, or size exclusion
through the column is tm, the dead volume is Vm = mobility is the sum of electrophoretic mobility and chromatography.

You might also like