This FAQ Booklet is kind of service guide for our engineer to answer the daily questions from our distributors and
end-user.
Questions from customers cover a wide range of topics, for easy discussion, we focus on Customer service.
Note: Below figure shows the whole procedure how to install the IV set correctly.
1).IV set installation incorrect, make sure the IV set install correct and the indicator is on green, otherwise it will
leakage., turn on the infusion-tube drop speed regulator to check whether there is free flow before staring
operation . It can start work if no leakage.
1 / 13
2). Try to test with other kinds of pipe to confirm it. It will lead to leakage if the pipe material is too thinness.
3).Replace the pump body if there is still leakage after above checking procedure.
TO MAX
AFR: According to related legal rules, the alarming voice of infusion pumps should be enough loud in order to
remind the user.
AFR: According to related legal rules, the alarming voice of infusion pumps should not be forbidden by manual.
The alarming triggered by the pumps will give warning to nurses or doctors.
Q3. Why the power connector does not have ground connector?
AFR: According to the security rules and related laws, It’s not necessary design power ground connector for
syringe and infusion pumps, Our design is totally meets the related rules.
AFR: Actually, the infusion pump can support all kinds of brand tube.
AFR: 1). The purpose of this function is convenience for users’ operation.
Let’s suppose, In the hospital, There are may be more than one kind of tube under operation(for example, they
have 3 kinds of brand tube , Brand 1,Brand 2,Brand 3 )
As mentioned above, the accuracy value will be different due to different brand tube, In this case, it necessary to
do accuracy, the user can save any one brand as any one kind of three type after calibration. for example.
Brand 1===>Type A(after adjustment)
Brand 2===>Type B(after adjustment)
Brand 3===>Type C(after adjustment)
Q8. Why can’t adjust the drop speed? For example, The max drop speed is 6 dropml/h.
AFR: 1).This issue caused by the configuration. Press the keys and simultaneously to enter the
setup menu. Press the key or to choose the sub-menu of ‘Drop/ML’. Then press to change
the numeric. 20 drop/ml as default from the manufactory, then exit the menu, the max drop speed can achieve to
133drop/min. press and keys simultaneously to transform the RATE and DROP mode.
3 / 13
NOTE: The configuration of Drop/ml depends on the tube. Normally, there are two kinds of tube in the markets. 20drop/ml and
AFR: 1) Open rear panel to check if cable for the screen is loosen. If yes, fix it.
2) If the cable and its connection is OK, the screen is defective. Please disassemble the pump and main board to
replace the screen.
AFR: 1) Open the rear panel to check if the motor works normally. (If it does not work, please replace the main
board).
2) If motor works normally, then the pump itself is blocked. Please replace it test.
AFR: 1) Press both ‘ ’ and ‘ ’ button simultaneously to check the real pressure occlusion value (the last
line value). The value without IV set is around from 45-65 and around 90-110 after installing IV set, if is normal.
Then press “Stop” and ‘Select’ buttons simultaneously to check the pressure occlusion valve value. Normally it
should be Middle level, Please change to High level.
4 / 13
Real Pressure occlusion value
2) If the currently occlusion value is too high (for example around 1000), Then adjust the pressure regulator with
screw driver, the pressure regulator located in the bottom.. There is one small hole in the bottom of the pump and
one screen inside, Please circumgyrate the screen check the currently pressure occlusion value until around 50
5 / 13
3).The pressure sensor damaged if the sensitivity value can’t be adjusted. Please replace it. Please refer to the
service manual for disassembly.
AFR: 1) .IV SET is not well installed. (Indicator does not light up).
2).IV set is well installed but indicator still does not light up. Press both ‘ ’ and ‘ ’ keys simultaneously
and check currently bubble value (the second line value). When the IV set not well installed and the air inside that
the air bubble value should be Yes, when the IV set well installed that the air bubble value should be No.(As the
below picture shows)
3) Through the above steps, install the IV set and press both ‘ ’ and ‘ ’ keys simultaneously and the
interface shows Handle: Open Bubble: No .It mentioned that the middle of IV set not put in the end, please
reinstall it.
2).The cable between disconnection between battery and the main-board, please check the connection. 3).The
main board has defective.
Note:
1. When connect the cable between battery and main-board, please do not make connection reverse, otherwise
it will burn the main-board.
6 / 13
Cable between battery and main-board
AFR: The pump with temperature is normally, because the motor inside under working will generate heat energy.
The temperature will be higher especially when the infusion rate is high speed and long time operation.
AFR: 1) Check every key on keyboard in standby status. If any key is not working, replace the key board.
2) If it can be started but the all the keys are not working, check the cable connected between keyboard and main
bard. If cable is ok, replace the main board.
AFR: 1). It can’t be turned on using AC power. Check the AC power supply source and cable.
2).It can’t be turned on using internal battery (using AC power supply can work). Possible causes: battery failure.
No battery inside or the battery can’t recharge.
3).It can’t start neither with AC power supply nor battery. Solution: check the cable connection between main
board and power supply board. If the cable connection is ok, replace the power supply board.
AFR: 1).Please press the ‘STOP’ button, then press the ‘POWER’ button last 3-5 second.
3) If the above still can’t be turned off , please replace the power supply board
Q18. The accuracy is correct at the first time after doing calibration, but why the accuracy is still
inaccuracy after several hours operation?
AFR: 1).First make sure the calibration procedure is correct(Refer to the accuracy calibration guide), the indicator
located in the front panel should be on green, otherwise, please re-install the IV set again.
7 / 13
2). Have you changed the tube position after each 4hours operation according to the requirement? If not, the
leakage will happen due to the tube distortion after long time operation.
3).If it still happen like this even change the tube position, probably it caused by the quality of the tube. Please
replace another kind of tube to test.
Q19. The pump can’t make occlusion alarming even the pressure is too high.
AFR: 1) The pressure sensor or pressure board damaged. Please do across testing to confirm it.
2). Press the keys and simultaneously to enter the setup menu. Press the key or
to choose the sub-menu of ‘Lock time’. Then press to choose ‘OFF’ to freeze this function.
Note: If active the ‘Lock time’ function, the key board will frozen after minutes.(Max to six minutes)
As shown below, if the ‘Lock time’ setup as 1min, the key board will be frozen after 1min, there is one lock logo in
the right corner.
Q21. The pump can’t make alarming when there is bubble in the tube.
AFR: 1) Please check this function active or not. Press ‘STOP ’ and ‘MENU ’ keys simultaneously. It
indicates the bubble detection function was canceled if the ‘Air level’ setup as ‘Close’.As shown below. Please
adjust it to level 2 or level 3.
2). Please replace the ultrasonic sensor or pressure board to test it if the error still happened. Please check the
service manual for details.
8 / 13
Ultrasonic sensor
Q22. The pump can’t make occlusion alarming even the pressure is too high.
AFR: 1) The pressure sensor or pressure board damaged. Please do across testing to confirm it.
9 / 13
Alarming information related infusion pumps
10 / 13
11 / 13
12 / 13
13 / 13