ENGINEER'S STAMP CONTRACTOR'S STAMP

BLACK & VEATCH


This data has been reviewed and assigned the status

NEN - No Exceptions Noted


Submittal Closed

EN - Exceptions Noted
Resubmit and proceed with manufacture

RFC - Returned For Correction


Resubmit - do not proceed with manufacture

RC - Record Copy 
Submittal Closed

RWOR - Returned Without Review


Submittal Closed

NA - Not Acceptable
Resubmit - do not proceed with manufacture

By B. Schulz Mar 25, 2015

D ISSUED FOR APPROVAL H.A.WOO T.K.KANG 12 Mar, 2015


C ISSUED FOR APPROVAL H.A.WOO Y.R.JEON 13 Nov, 2014
B ISSUED FOR APPROVAL H.A.WOO Y.R.JEON 06 Aug, 2014
A ISSUED FOR APPROVAL H.A.WOO Y.R.JEON 30 May, 2014
REV. DESCRIPTION PREPARED CHECKED DATE

MINERA ESCONDIDA Ltda

ESCONDIDA WATER SUPPLY PROJECT

CONTRACTOR

Project No TITLE
LOCATION
DATASHEETS OF ANALYZER INSTRUMENTS
W13018
Escondida

SCALE DOCUMENT No

REV. D
W13018-YT01-CQ-800214

SUBCONTRACTOR

SHEET 1 OF 220
ESCONDIDA
WATER SUPPLY PROJECT

DATASHEETS OF ANALYZER INSTRUMENTS

TABLE OF CONTENTS

1. BILL OF MATERIALS
1.1 BOM LIST

2. DATA SHEET
2.1 ANALYZER LIST & DATA SHEET

3. CATALOGUE
3.1 ANALYZER

REV. D W13018-YT01-CQ-800214
ESCONDIDA
WATER SUPPLY PROJECT

DATASHEETS OF ANALYZER INSTRUMENTS

1. BILL OF MATERIALS

1.1 BOM LIST

REV. D W13018-YT01-CQ-800214
Rev No. BY DATE REVISION
A H.A. WOO 30 May. 2014 Issued for Approval
BILL OF MATERIALS B H.A. WOO 06 Aug. 2014 Issued for Approval
C H.A. WOO 13 Nov. 2014 Issued for Approval
D H.A. WOO 12 Mar. 2015 Issued for Approval

ESCONDIDA WATER SUPPLY PROJECT


INDOOR /
NO. TAG NO ITEMS TYPE Q'TY AREA Group NO.
OUTDOOR
1 514161 AIT 01821 EXISTING SWRO SWPS SUP SDI SDI 1
2 514161 AIT 01822 EXISTING SWRO SWPS SUP PH pH 1
3 514161 AIT 01823 EXISTING SWRO SWPS SUP TURB TURB 1
INTAKE OUTDOOR 514161-AP-00001
4 514161 AIT 01824 EXISTING SWRO SWPS SUP COND COND 1
5 514161 AIT 01825 EXISTING SWRO SWPS SUP CL2 CL2 1
6 514161 AIT 01826 EXISTING SWRO SWPS SUP HYD HYD 1
7 514161 AIT 01906 PLANT 1+2 SWPS SUP SDI SDI 1
8 514161 AIT 01901 PLANT 1+2 SWPS SUP pH pH 1
9 514161 AIT 01902 PLANT 1+2 SWPS SUP TURB TURB 1
10 514161 AIT 01903 PLANT 1+2 SWPS SUP COND COND 1
11 514161 AIT 01904 PLANT 1+2 SWPS SUP CL2 CL2 1
12 514161 AIT 01905 PLANT 1+2 SWPS SUP HYD HYD 1
INTAKE OUTDOOR 514161-AP-00002
13 514161 AIT 01916 PLANT 3 SWPS SUP SDI SDI 1
14 514161 AIT 01911 PLANT 3 SWPS SUP pH pH 1
15 514161 AIT 01912 PLANT 3 SWPS SUP TURB TURB 1
16 514161 AIT 01913 PLANT 3 SWPS SUP COND COND 1
17 514161 AIT 01914 PLANT 3 SWPS SUP CL2 CL2 1
18 514161 AIT 01915 PLANT 3 SWPS SUP HYD HYD 1
19 514162 AIT 02001 P1 1ST STG PF 1-12 SDI SDI 1
20 514162 AIT 02002 P1 1ST STG PF 1-12 CL2 CL2 1
21 514162 AIT 02003 P1 1ST STG PF 1-12 TURB TURB 1
22 514162 AIT 02005 P1 1ST STG PF 1-12 INLET pH pH 1
1ST DMF OUTDOOR 514162-AP-00003
23 514162 AIT 02014 P2 1ST STG PF 13-24 SDI SDI 1
24 514162 AIT 02012 P2 1ST STG PF 13-24 CL2 CL2 1
25 514162 AIT 02013 P2 1ST STG PF 13-24 TURB TURB 1
26 514162 AIT 02015 P2 1ST STG PF 13-24 INLET pH pH 1
27 514162 AIT 02021 P3 1ST STG PF 25-36 SDI SDI 1
28 514162 AIT 02022 P3 1ST STG PF 25-36 CL2 CL2 1
1ST DMF OUTDOOR 514162-AP-00004
29 514162 AIT 02023 P3 1ST STG PF 25-36 TURB TURB 1
30 514162 AIT 02025 P3 1ST STG PF 25-36 INLET pH pH 1
31 514162 AIT 02100 P1 1ST STG PF 1 CELL A DIS TURB TURB 1 1ST DMF OUTDOOR 514162-AP-00005
32 514162 AIT 02560 P2 1ST STG PF 24 CELL B DIS TURB TURB 1 1ST DMF OUTDOOR 514162-AP-00006
33 514162 AIT 02120 1st PRESSURE FILTERED WATER (1&2) TURB 1 1ST DMF OUTDOOR 514162-AP-00007
34 514162 AIT 02580 P3 1ST STG PF 25 CELL A DIS TURB TURB 1 1ST DMF OUTDOOR 514162-AP-00008
35 514162 AIT 02800 P3 1ST STG PF 36 CELL B DIS TURB TURB 1 1ST DMF OUTDOOR 514162-AP-00009
36 514162 AIT 02600 1st PRESSURE FILTERED WATER (1&2) TURB 1 1ST DMF OUTDOOR 514162-AP-00010
37 514162 AIT 03047 P1 2ND STG PF 1 CELL A DIS TURB TURB 1 2ND DMF OUTDOOR 514162-AP-00011
38 514162 AIT 03672 P2 2ND STG PF 16 CELL B DIS TURB TURB 1 2ND DMF OUTDOOR 514162-AP-00012
39 514162 AIT 03072 1st PRESSURE FILTERED WATER (1&2) TURB 1 2ND DMF OUTDOOR 514162-AP-00003
40 514162 AIT 03697 P3 2ND STG PF 17 CELL A DIS TURB TURB 1 2ND DMF OUTDOOR 514162-AP-00014
41 514162 AIT 03997 P3 2ND STG PF 24 CELL B DIS TURB TURB 1 2ND DMF OUTDOOR 514162-AP-00015
42 514162 AIT 03722 1st PRESSURE FILTERED WATER (1&2) TURB 1 2ND DMF OUTDOOR 514162-AP-00016
514162-AP-00017
43 514162 AIT 03711 P1+2 WWT CL2 CL2 1 WWT OUTDOOR
w/pump
514162-AP-00018
44 514162 AIT 03811 P3+4 WWT CL2 CL2 1 WWT OUTDOOR
w/pump
45 514163 AIT 04002 P1+2 RO SYST HDR CL2 CL2 1
46 514163 AIT 04003 P1+2 RO SYST HDR TURB TURB 1
47 514163 AIT 04004 P1+2 RO SYST HDR COND COND 1
48 514163 AIT 04005 P1+2 RO SYST HDR SDI SDI 1
49 514163 AIT 04006 P1+2 RO SYST HDR PH pH 1
50 514163 AIT 04001 P1+2 RO SYST HDR ORP ORP 1
C/F INDOOR 514163-AP-00019
51 514163 AIT 04022 P3+4 RO SYST HDR CL2 CL2 1
52 514163 AIT 04023 P3+4 RO SYST HDR TURB TURB 1
53 514163 AIT 04024 P3+4 RO SYST HDR COND COND 1
54 514163 AIT 04025 P3+4 RO SYST HDR SDI SDI 1
55 514163 AIT 04026 P3+4 RO SYST HDR PH pH 1
56 514163 AIT 04021 P3+4 RO SYST HDR ORP ORP 1
Rev No. BY DATE REVISION
A H.A. WOO 30 May. 2014 Issued for Approval
BILL OF MATERIALS B H.A. WOO 06 Aug. 2014 Issued for Approval
C H.A. WOO 13 Nov. 2014 Issued for Approval
D H.A. WOO 12 Mar. 2015 Issued for Approval

ESCONDIDA WATER SUPPLY PROJECT


INDOOR /
NO. TAG NO ITEMS TYPE Q'TY AREA Group NO.
OUTDOOR
57 514163 AIT 04041 RO TRAIN 1 INL COND COND 1 RO MEM INDOOR 514163-AP-00020
58 514163 AIT 04042 ERS BP 1 DIS COND COND 1 RO MEM INDOOR 514163-AP-00021
59 514163 AIT 04670 RO TRAIN 1 DIS COND COND 1 RO MEM INDOOR 514163-AP-00022
60 514163 AIT 04111 RO TRAIN 2 INL COND COND 1 RO MEM INDOOR 514163-AP-00023
61 514163 AIT 04112 ERS BP 2 DIS COND COND 1 RO MEM INDOOR 514163-AP-00024
62 514163 AIT 04685 RO TRAIN 2 DIS COND COND 1 RO MEM INDOOR 514163-AP-00025
63 514163 AIT 04181 RO TRAIN 3 INL COND COND 1 RO MEM INDOOR 514163-AP-00026
64 514163 AIT 04182 ERS BP 3 DIS COND COND 1 RO MEM INDOOR 514163-AP-00027
65 514163 AIT 04700 RO TRAIN 3 DIS COND COND 1 RO MEM INDOOR 514163-AP-00028
66 514163 AIT 04251 RO TRAIN 4 INL COND COND 1 RO MEM INDOOR 514163-AP-00029
67 514163 AIT 04252 ERS BP 4 DIS COND COND 1 RO MEM INDOOR 514163-AP-00030
68 514163 AIT 04715 RO TRAIN 4 DIS COND COND 1 RO MEM INDOOR 514163-AP-00031
69 514163 AIT 04321 RO TRAIN 5 INL COND COND 1 RO MEM INDOOR 514163-AP-00032
70 514163 AIT 04322 ERS BP 5 DIS COND COND 1 RO MEM INDOOR 514163-AP-00033
71 514163 AIT 04730 RO TRAIN 5 DIS COND COND 1 RO MEM INDOOR 514163-AP-00034
72 514163 AIT 04391 RO TRAIN 6 INL COND COND 1 RO MEM INDOOR 514163-AP-00035
73 514163 AIT 04392 ERS BP 6 DIS COND COND 1 RO MEM INDOOR 514163-AP-00036
74 514163 AIT 04745 RO TRAIN 6 DIS COND COND 1 RO MEM INDOOR 514163-AP-00037
75 514163 AIT 04461 RO TRAIN 7 INL COND COND 1 RO MEM INDOOR 514163-AP-00038
76 514163 AIT 04462 ERS BP 7 DIS COND COND 1 RO MEM INDOOR 514163-AP-00039
77 514163 AIT 04760 RO TRAIN 7 DIS COND COND 1 RO MEM INDOOR 514163-AP-00040
78 514163 AIT 04531 RO TRAIN 8 INL COND COND 1 RO MEM INDOOR 514163-AP-00041
79 514163 AIT 04532 ERS BP 8 DIS COND COND 1 RO MEM INDOOR 514163-AP-00042
80 514163 AIT 04775 RO TRAIN 8 DIS COND COND 1 RO MEM INDOOR 514163-AP-00043
81 514163 AIT 04601 RO TRAIN 9 INL COND COND 1 RO MEM INDOOR 514163-AP-00044
82 514163 AIT 04602 ERS BP 9 DIS COND COND 1 RO MEM INDOOR 514163-AP-00045
83 514163 AIT 04790 RO TRAIN 9 DIS COND COND 1 RO MEM INDOOR 514163-AP-00046
84 514165 AIT 04810 P1+2 RO CIP PH pH 1 CIP OUTDOOR 514163-AP-00047
insertion type 2" pipe
85 514165 AIT 04985 P1+2 NEUTRALN TNK PH pH 1 NEUT OUTDOOR
mounting
86 514165 AIT 04870 P3+4 RO CIP PH pH 1 CIP OUTDOOR 514165-AP-00048
insertion type 2" pipe
87 514165 AIT 04990 P3+4 NEUTRALN TNK PH pH 1 NEUT OUTDOOR
mounting
88 514162 AIT 03005 P1+2 2ND STG PF 1-16 SDI SDI 1
89 514162 AIT 03001 P1+2 2ND STG PF 1-16 CL2 CL2 1
2ND DMF OUTDOOR
90 514162 AIT 03002 P1+2 2ND STG PF 1-16 TURB TURB 1
91 514162 AIT 03004 P1+2 2ND STG PF 1-16 PH pH 1
92 514164 AIT 05004 RO PROD TNK 1 PERMEATE PH pH 1 514162-AP-00049
93 514164 AIT 05005 RO PROD TNK 1 INL COND COND 1
94 514164 AIT 05003 RO PROD TNK 1 INL PH pH 1 POTABLE OUTDOOR
95 514164 AIT 05006 RO PROD TNK 1 INL CL2 CL2 1
96 514164 AIT 05007 RO PROD TNK 1 INL PHOSPHATE PHOSPHATE 1
97 514162 AIT 03009 P3 2ND STG PF 17-24 SDI SDI 1
98 514162 AIT 03006 P3 2ND STG PF 17-24 CL2 CL2 1
2ND DMF OUTDOOR
99 514162 AIT 03007 P3 2ND STG PF 17-24 TURB TURB 1
100 514162 AIT 03008 P3 2ND STG PF 17-24 PH pH 1
101 514164 AIT 05014 RO PROD TNK 2 PERMEATE PH pH 1 514162-AP-00050
102 514164 AIT 05015 RO PROD TNK 2 INL COND COND 1
103 514164 AIT 05013 RO PROD TNK 2 INL PH pH 1 PORTABLE OUTDOOR
104 514164 AIT 05016 RO PROD TNK 2 INL CL2 CL2 1
105 514164 AIT 05017 RO PROD TNK 2 INL PHOSPHATE PHOSPHATE 1

NOTE.
ESCONDIDA
WATER SUPPLY PROJECT

DATASHEETS OF ANALYZER INSTRUMENTS

2. DATA SHEET

2.1 ANALYZER LIST & DATA SHEET

REV. D W13018-YT01-CQ-800214
Doc. No. : W13018-YT01-CQ-800214
INSTRUMENT DATA SHEET
FOR Rev. No. : D
ANALYZER
2 OF 105

ANALYZER TAG LIST


NO. TAG NO INSTRUMENT TYPE SHEET NO. REMARK
1 514161 AIT 01822 PH ANALYZER 4 FLOW THROUGH
2 514161 AIT 01901 PH ANALYZER 5 FLOW THROUGH
3 514161 AIT 01911 PH ANALYZER 6 FLOW THROUGH
4 514162 AIT 02005 PH ANALYZER 7 FLOW THROUGH
5 514162 AIT 02015 PH ANALYZER 8 FLOW THROUGH
6 514162 AIT 02025 PH ANALYZER 9 FLOW THROUGH
7 514162 AIT 03004 PH ANALYZER 10 FLOW THROUGH
8 514162 AIT 03008 PH ANALYZER 11 FLOW THROUGH
9 514163 AIT 04006 PH ANALYZER 12 FLOW THROUGH
10 514163 AIT 04026 PH ANALYZER 13 FLOW THROUGH
11 514165 AIT 04810 PH ANALYZER 14 FLOW THROUGH
12 514165 AIT 04870 PH ANALYZER 15 FLOW THROUGH
13 514165 AIT 04985 PH ANALYZER 16 RETRACTABLE
14 514165 AIT 04990 PH ANALYZER 17 RETRACTABLE
15 514164 AIT 05004 PH ANALYZER 18 FLOW THROUGH
16 514164 AIT 05003 PH ANALYZER 19 FLOW THROUGH
17 514164 AIT 05014 PH ANALYZER 20 FLOW THROUGH
18 514164 AIT 05013 PH ANALYZER 21 FLOW THROUGH
19 514163 AIT 04001 ORP ANALYZER 22 FLOW THROUGH
20 514163 AIT 04021 ORP ANALYZER 23 FLOW THROUGH
21 514161 AIT 01824 CONDUCTIVITY ANALYZER 24 FLOW THROUGH
22 514161 AIT 01903 CONDUCTIVITY ANALYZER 25 FLOW THROUGH
23 514161 AIT 01913 CONDUCTIVITY ANALYZER 26 FLOW THROUGH
24 514163 AIT 04004 CONDUCTIVITY ANALYZER 27 FLOW THROUGH
25 514163 AIT 04024 CONDUCTIVITY ANALYZER 28 FLOW THROUGH
26 514163 AIT 04041 CONDUCTIVITY ANALYZER 29 FLOW THROUGH
27 514163 AIT 04042 CONDUCTIVITY ANALYZER 30 FLOW THROUGH
28 514163 AIT 04670 CONDUCTIVITY ANALYZER 31 FLOW THROUGH
29 514163 AIT 04111 CONDUCTIVITY ANALYZER 32 FLOW THROUGH
30 514163 AIT 04112 CONDUCTIVITY ANALYZER 33 FLOW THROUGH
31 514163 AIT 04685 CONDUCTIVITY ANALYZER 34 FLOW THROUGH
32 514163 AIT 04181 CONDUCTIVITY ANALYZER 35 FLOW THROUGH
33 514163 AIT 04182 CONDUCTIVITY ANALYZER 36 FLOW THROUGH
34 514163 AIT 04700 CONDUCTIVITY ANALYZER 37 FLOW THROUGH
35 514163 AIT 04251 CONDUCTIVITY ANALYZER 38 FLOW THROUGH
36 514163 AIT 04252 CONDUCTIVITY ANALYZER 39 FLOW THROUGH
37 514163 AIT 04715 CONDUCTIVITY ANALYZER 40 FLOW THROUGH
38 514163 AIT 04321 CONDUCTIVITY ANALYZER 41 FLOW THROUGH
39 514163 AIT 04322 CONDUCTIVITY ANALYZER 42 FLOW THROUGH
40 514163 AIT 04730 CONDUCTIVITY ANALYZER 43 FLOW THROUGH
41 514163 AIT 04391 CONDUCTIVITY ANALYZER 44 FLOW THROUGH
42 514163 AIT 04392 CONDUCTIVITY ANALYZER 45 FLOW THROUGH
43 514163 AIT 04745 CONDUCTIVITY ANALYZER 46 FLOW THROUGH
44 514163 AIT 04461 CONDUCTIVITY ANALYZER 47 FLOW THROUGH
45 514163 AIT 04462 CONDUCTIVITY ANALYZER 48 FLOW THROUGH
46 514163 AIT 04760 CONDUCTIVITY ANALYZER 49 FLOW THROUGH
47 514163 AIT 04531 CONDUCTIVITY ANALYZER 50 FLOW THROUGH
48 514163 AIT 04532 CONDUCTIVITY ANALYZER 51 FLOW THROUGH
49 514163 AIT 04775 CONDUCTIVITY ANALYZER 52 FLOW THROUGH
50 514163 AIT 04601 CONDUCTIVITY ANALYZER 53 FLOW THROUGH
51 514163 AIT 04602 CONDUCTIVITY ANALYZER 54 FLOW THROUGH
52 514163 AIT 04790 CONDUCTIVITY ANALYZER 55 FLOW THROUGH
53 514164 AIT 05005 CONDUCTIVITY ANALYZER 56 FLOW THROUGH
54 514164 AIT 05015 CONDUCTIVITY ANALYZER 57 FLOW THROUGH
Doc. No. : W13018-YT01-CQ-800214
INSTRUMENT DATA SHEET
FOR Rev. No. : D
ANALYZER
3 OF 105

ANALYZER TAG LIST


NO. TAG NO INSTRUMENT TYPE SHEET NO. REMARK
53 514161 AIT 01825 CHLORINE ANALYZER 58 FREE CHLORINE
54 514161 AIT 01904 CHLORINE ANALYZER 59 FREE CHLORINE
55 514161 AIT 01914 CHLORINE ANALYZER 60 FREE CHLORINE
56 514162 AIT 02002 CHLORINE ANALYZER 61 FREE CHLORINE
57 514162 AIT 02012 CHLORINE ANALYZER 62 FREE CHLORINE
58 514162 AIT 02022 CHLORINE ANALYZER 63 FREE CHLORINE
59 514162 AIT 03001 CHLORINE ANALYZER 64 FREE CHLORINE
60 514162 AIT 03006 CHLORINE ANALYZER 65 FREE CHLORINE
61 514162 AIT 03711 CHLORINE ANALYZER 66 TOTAL CHLORINE
62 514162 AIT 03811 CHLORINE ANALYZER 67 TOTAL CHLORINE
63 514163 AIT 04002 CHLORINE ANALYZER 68 FREE CHLORINE
64 514163 AIT 04022 CHLORINE ANALYZER 69 FREE CHLORINE
65 514164 AIT 05006 CHLORINE ANALYZER 70 FREE CHLORINE
66 514164 AIT 05016 CHLORINE ANALYZER 71 FREE CHLORINE
67 514161 AIT 01823 TURBIDITY ANALYZER 72 SCATTERED LIGHT
68 514161 AIT 01902 TURBIDITY ANALYZER 73 SCATTERED LIGHT
69 514161 AIT 01912 TURBIDITY ANALYZER 74 SCATTERED LIGHT
70 514162 AIT 02003 TURBIDITY ANALYZER 75 SCATTERED LIGHT
71 514162 AIT 02013 TURBIDITY ANALYZER 76 SCATTERED LIGHT
72 514162 AIT 02023 TURBIDITY ANALYZER 77 SCATTERED LIGHT
73 514162 AIT 02100 TURBIDITY ANALYZER 78 SCATTERED LIGHT
74 514162 AIT 02560 TURBIDITY ANALYZER 79 SCATTERED LIGHT
75 514162 AIT 02120 TURBIDITY ANALYZER 80 SCATTERED LIGHT
76 514162 AIT 02580 TURBIDITY ANALYZER 81 SCATTERED LIGHT
77 514162 AIT 02800 TURBIDITY ANALYZER 82 SCATTERED LIGHT
78 514162 AIT 02600 TURBIDITY ANALYZER 83 SCATTERED LIGHT
79 514162 AIT 03047 TURBIDITY ANALYZER 84 SCATTERED LIGHT
80 514162 AIT 03672 TURBIDITY ANALYZER 85 SCATTERED LIGHT
81 514162 AIT 03072 TURBIDITY ANALYZER 86 SCATTERED LIGHT
82 514162 AIT 03697 TURBIDITY ANALYZER 87 SCATTERED LIGHT
83 514162 AIT 03997 TURBIDITY ANALYZER 88 SCATTERED LIGHT
84 514162 AIT 03722 TURBIDITY ANALYZER 89 SCATTERED LIGHT
85 514163 AIT 04003 TURBIDITY ANALYZER 90 SCATTERED LIGHT
86 514163 AIT 04023 TURBIDITY ANALYZER 91 SCATTERED LIGHT
87 514162 AIT 03002 TURBIDITY ANALYZER 92 SCATTERED LIGHT
88 514162 AIT 03007 TURBIDITY ANALYZER 93 SCATTERED LIGHT
89 514161 AIT 01826 HYDROCARBON ANALYZER 94 UV FLUORESCENCE
90 514161 AIT 01905 HYDROCARBON ANALYZER 95 UV FLUORESCENCE
91 514161 AIT 01915 HYDROCARBON ANALYZER 96 UV FLUORESCENCE
94 514161 AIT 01821 SDI ANALYZER 97 ONE CHANNEL
95 514161 AIT 01906
SDI ANALYZER 98 TWO CHANNEL
96 514161 AIT 01916
97 514162 AIT 02001
SDI ANALYZER 99 TWO CHANNEL
98 514162 AIT 02014
99 514162 AIT 02021 SDI ANALYZER 100 ONE CHANNEL
100 514162 AIT 03005 SDI ANALYZER 101 ONE CHANNEL
101 514162 AIT 03009 SDI ANALYZER 102 ONE CHANNEL
102 514163 AIT 04005
SDI ANALYZER 103 TWO CHANNEL
103 514163 AIT 04025
104 514164 AIT 05007 PHOSPHATE 104 COLORIMETRIC
105 514164 AIT 05017 PHOSPHATE 105 COLORIMETRIC
SHEET 4 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01822

2 Quantity 1 SET
GENERAL 3 Service EXISTING SWRO SWPS SUP PH

4 P&ID Number 001-2

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 11.5 14.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 7.7

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 5 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01901

2 Quantity 1 SET
GENERAL 3 Service PLANT 1+2 SWPS SUP PH

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 7.7

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 6 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01911

2 Quantity 1 SET
GENERAL 3 Service PLANT 3 SWPS SUP PH

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 7.7

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 7 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02005

2 Quantity 1 SET
GENERAL 3 Service P1 1ST STG PF 1-12 INLET PH

4 P&ID Number 2

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 7.0

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 8 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02015

2 Quantity 1 SET
GENERAL 3 Service P2 1ST STG PF 13-24 INLET PH

4 P&ID Number 003

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 7.0

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 9 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02025

2 Quantity 1 SET
GENERAL 3 Service P3 1ST STG PF 25-36 INLET PH

4 P&ID Number 004

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 7.0

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 10 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03004

2 Quantity 1 SET
GENERAL 3 Service P1+2 2ND STG PF 1-16 PH

4 P&ID Number 007

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16 20

9 pH Min / Nor / Max pH 7

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 11 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03008

2 Quantity 1 SET
GENERAL 3 Service P3 2ND STG PF 17-24 PH

4 P&ID Number 009

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16 20

9 pH Min / Nor / Max pH 7

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 12 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04006

2 Quantity 1 SET
GENERAL 3 Service P1+2 RO SYST HDR PH

4 P&ID Number 018

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16 20

9 pH Min / Nor / Max pH 7

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 13 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04026

2 Quantity 1 SET
GENERAL 3 Service P3+4 RO SYST HDR PH

4 P&ID Number 018

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16 20

9 pH Min / Nor / Max pH 7

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 14 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514165 - AIT - 04810

2 Quantity 1 SET
GENERAL 3 Service P1+2 RO CIP PH

4 P&ID Number 025

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 4.0 6.2


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 2~12

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 15 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514165 - AIT - 04870

2 Quantity 1 SET
GENERAL 3 Service P3+4 RO CIP PH

4 P&ID Number 026

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 4.0 6.2


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 2~12

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 16 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514165 - AIT - 04985

2 Quantity 1 SET
GENERAL 3 Service P1+2 NEUTRALN TNK PH

4 P&ID Number 027

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.6 0.8


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 2~12

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A201 / 20m

20 Chamber Type Retractable

21 Chamber Material (Model) 316SS (CPA450-0K110)

22 Process Connection 1-1/4" ANSI 150# RF FLANGE

23
24
25
26
27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 17 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514165 - AIT - 04990

2 Quantity 1 SET
GENERAL 3 Service P3+4 NEUTRALN TNK PH

4 P&ID Number 028

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.6 0.8


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 2~12

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A201 / 20m

20 Chamber Type Retractable

21 Chamber Material (Model) 316SS (CPA450-0K110)

22 Process Connection 1-1/4" ANSi 150# RF FLANGE

23
24
25
26
27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 18 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514164 - AIT - 05004

2 Quantity 1 SET
GENERAL 3 Service RO PROD TNK 1 PERMEATE PH

4 P&ID Number 029

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 2.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 7.6

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 19 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514164 - AIT - 05003

2 Quantity 1 SET
GENERAL 3 Service RO PROD TNK 1 INL PH

4 P&ID Number 029

5
6 Fluid PRODUCT WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 2.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 7.6

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 20 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514164 - AIT - 05014

2 Quantity 1 SET
GENERAL 3 Service RO PROD TNK 2 PERMEATE PH

4 P&ID Number 030

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 2.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 7.6

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 21 OF 105
pH ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514164 - AIT - 05013

2 Quantity 1 SET
GENERAL 3 Service RO PROD TNK 2 INL PH

4 P&ID Number 030

5
6 Fluid PRODUCT WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 2.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 pH Min / Nor / Max pH 7.6

10 Manufacturer Endress + Hauser

11 Model No. CPS11D-7BA21

12 Cell Measuring Range 0-14pH

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Connection Size & Type PG13.5

19 Special Cable ( Model / Length) CYK10-A881 / 1m

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 to 14 pH

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.02 pH

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 22 OF 105
ORP ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 02 Apr. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04001

2 Quantity 1 SET
GENERAL 3 Service P1+2 RO SYST HDR ORP

4 P&ID Number 018

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16 20

9 ORP Min / Nor / Max mV

10 Manufacturer Endress + Hauser

11 Model No. CPS12D-7NA21

12 Cell Measuring Range -999 to 999 mV

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Special Cable ( Model / Length) CYK10-A881 / 1m

19 Connection Size & Type PG13.5

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range -999 to 999 mV

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 1mV

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 23 OF 105
ORP ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 02 Apr. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04021

2 Quantity 1 SET
GENERAL 3 Service P3+4 RO SYST HDR ORP

4 P&ID Number 018

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16 20

9 ORP Min / Nor / Max mV

10 Manufacturer Endress + Hauser

11 Model No. CPS12D-7NA21

12 Cell Measuring Range -999 to 999 mV

13 Cell Constant N/A

14 Electrode Material GLASS ELECTRODE

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 16 barG


SENSOR
17 Element Temperature Rating 135 degC
18 Special Cable ( Model / Length) CYK10-A881 / 1m

19 Special Cable PG13.5

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-MAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range -999 to 999 mV

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 1mV

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 24 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01824

2 Quantity 1 SET
GENERAL 3 Service EXISITING SWRO SWPS SUP COND

4 P&ID Number 001-2

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 11.5 14.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 51900

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 60,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 ~ 60,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 25 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01903

2 Quantity 1 SET
GENERAL 3 Service PLANT 1+2 SWPS SUP COND

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 51900

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 60,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 ~ 60,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 26 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01913

2 Quantity 1 SET
GENERAL 3 Service PLANT 3 SWPS SUP COND

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 51900

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 60,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 ~ 60,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 27 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04004

2 Quantity 1 SET
GENERAL 3 Service P1+2 RO SYST HDR COND

4 P&ID Number 018

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 60,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 60,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 28 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04024

2 Quantity 1 SET
GENERAL 3 Service P3+4 RO SYST HDR COND

4 P&ID Number 018

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 60,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 60,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 29 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04041

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 1 INL COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 61,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 61,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 30 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04042

2 Quantity 1 SET
GENERAL 3 Service ERS BP 1 DIS COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 62,000 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 62,000 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 31 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04670

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 1 DIS COND

4 P&ID Number 022

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 1.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 550

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,400 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,400 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 32 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04111

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 2 INL COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 61,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 61,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 33 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04112

2 Quantity 1 SET
GENERAL 3 Service ERS BP 2 DIS COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 62,000 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 62,000 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 34 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04685

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 2 DIS COND

4 P&ID Number 022

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 1.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 550

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,400 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,400 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 35 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04181

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 3 INL COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 61,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 61,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 36 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04182

2 Quantity 1 SET
GENERAL 3 Service ERS BP 3 DIS COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 62,000 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 62,000 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 37 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04700

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 3 DIS COND

4 P&ID Number 022

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 1.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 550

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,400 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,400 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 38 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04251

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 4 INL COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 61,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 61,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 39 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04252

2 Quantity 1 SET
GENERAL 3 Service ERS BP 4 DIS COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 62,000 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 62,000 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 40 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04715

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 4 DIS COND

4 P&ID Number 022

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 1.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 550

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,400 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,400 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 41 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04321

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 5 INL COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 61,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 61,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 42 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04322

2 Quantity 1 SET
GENERAL 3 Service ERS BP 5 DIS COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 62,000 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 62,000 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 43 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04730

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 5 DIS COND

4 P&ID Number 022

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 1.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 550

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,400 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,400 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 44 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04391

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 6 INL COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 61,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 61,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 45 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04392

2 Quantity 1 SET
GENERAL 3 Service ERS BP 6 DIS COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 62,000 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 62,000 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 46 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04745

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 6 DIS COND

4 P&ID Number 022

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 1.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 550

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,400 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,400 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 47 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04461

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 7 INL COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 61,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 61,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 48 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04462

2 Quantity 1 SET
GENERAL 3 Service ERS BP 7 DIS COND

4 P&ID Number 022

5
6 Fluid SEA WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 62,000 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 62,000 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 49 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04760

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 7 DIS COND

4 P&ID Number 022

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 1.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 550

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,400 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,400 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 50 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04531

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 8 INL COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 61,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 61,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 51 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04532

2 Quantity 1 SET
GENERAL 3 Service ERS BP 8 DIS COND

4 P&ID Number 022

5
6 Fluid SEA WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 61,200 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 61,200 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 52 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04775

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 8 DIS COND

4 P&ID Number 022

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 1.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 550

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,400 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,400 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 53 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04601

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 9 INL COND

4 P&ID Number 022

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 61,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 61,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 54 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04602

2 Quantity 1 SET
GENERAL 3 Service ERS BP 9 DIS COND

4 P&ID Number 022

5
6 Fluid SEA WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 6.5 7.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 54300

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 62,000 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 62,000 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 55 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04790

2 Quantity 1 SET
GENERAL 3 Service RO TRAIN 9 DIS COND

4 P&ID Number 022

5
6 Fluid PERMEATE

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 1.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 550

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,400 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1/2" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,400 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 56 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514164 - AIT - 05005

2 Quantity 1 SET
GENERAL 3 Service RO PROD TNK 1 INL COND

4 P&ID Number 029

5
6 Fluid PRODUCT WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 2.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 600

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 57 OF 105
CONDUCTIVITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514164 - AIT - 05015

2 Quantity 1 SET
GENERAL 3 Service RO PROD TNK 2 INL COND

4 P&ID Number 030

5
6 Fluid PRODUCT WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 2.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Conductivity Min / Nor / Max uS / cm 600

10 Manufacturer Endress + Hauser

11 Model No. CLS50D-AA1B71

12 Cell Measuring Range 0 - 1,500 uS/cm

13 Cell Constant 1.98/cm

14 Electrode Material PEEK

15 Temp. Compensation YES (AUTO)

16 Element Pressure Rating 20 barG


SENSOR
17 Element Temperature Rating 125 degC
18 Connection Size & Type G3/4

19 Special Cable 1m (Fixed Cable)

20 Cell Chamber Type FLOW CHAMBER

21 Cell Chamber Material Polypropylene (PP)

22 Cell Chamber Connection 1/2" NPT(F)

23 Sample Flow Rate 0.1 - 10 l/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer Endress + Hauser

28 Model No. CM42-LAB300EAS00

29 Output Signal FOUNDATION Fieldbus

30 Power Supply 9 to 32 V DC

31 Display LCD DISPLAY

32 Enclosure Class IP66


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection M20 x 1.5

35 Set Scale Range 0 - 1,500 uS/cm

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy 0.5 % from measured value ±1.7 μS/cm

38 Ambient Temperature Rating -30 to +70°C

39 Relative Humidity Rating 10 to 95% RH non-condensing

Note.
SHEET 58 OF 105
FREE CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01825

2 Quantity 1 SET
GENERAL 3 Service EXISTING SWRO SWPS SUP CL2

4 P&ID Number 001-2

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 11.5 14.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440001)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440001)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 59 OF 105
FREE CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01904

2 Quantity 1 SET
GENERAL 3 Service PLANT 1+2 SWPS SUP CL2

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440001)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440001)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 60 OF 105
FREE CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01914

2 Quantity 1 SET
GENERAL 3 Service PLANT 3 SWPS SUP CL2

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440001)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440001)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 61 OF 105
FREE CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02002

2 Quantity 1 SET
GENERAL 3 Service P1 1ST STG PF 1-12 CL2

4 P&ID Number 002

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440001)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440001)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 62 OF 105
FREE CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02012

2 Quantity 1 SET
GENERAL 3 Service P2 1ST STG PF 13-24 CL2

4 P&ID Number 003

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440001)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440001)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 63 OF 105
FREE CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02022

2 Quantity 1 SET
GENERAL 3 Service P3 1ST STG PF 25-36 CL2

4 P&ID Number 004

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440001)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440001)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 64 OF 105
FREE CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03001

2 Quantity 1 SET
GENERAL 3 Service P1+2 2ND STG PF 1-16 CL2

4 P&ID Number 007

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440001)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440001)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 65 OF 105
FREE CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03006

2 Quantity 1 SET
GENERAL 3 Service P3 2ND STG PF 17-24 CL2

4 P&ID Number 009

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440001)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440001)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 66 OF 105
TOTAL CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03711

2 Quantity 1 SET
GENERAL 3 Service P1+2 WWT CL2

4 P&ID Number 016

5
6 Fluid WASTE WATER

PROCESS 7 Pressure Min / Nor / Max bar.G ATM


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440002)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440002)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 67 OF 105
TOTAL CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03811

2 Quantity 1 SET
GENERAL 3 Service P3+4 WWT CL2

4 P&ID Number 017

5
6 Fluid WASTE WATER

PROCESS 7 Pressure Min / Nor / Max bar.G ATM


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440002)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440002)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 68 OF 105
FREE CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04002

2 Quantity 1 SET
GENERAL 3 Service P1+2 RO SYST HDR CL2

4 P&ID Number 018

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440001)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440001)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 69 OF 105
FREE CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04022

2 Quantity 1 SET
GENERAL 3 Service P3+4 RO SYST HDR CL2

4 P&ID Number 019

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440001)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440001)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 70 OF 105
TOTAL CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514164 - AIT - 05006

2 Quantity 1 SET
GENERAL 3 Service RO PROD TNK 1 INL CL2

4 P&ID Number 029

5
6 Fluid PRODUCT WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 2.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440002)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440002)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 71 OF 105
TOTAL CHLORINE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514164 - AIT - 05016

2 Quantity 1 SET
GENERAL 3 Service RO PROD TNK 2 INL CL2

4 P&ID Number 030

5
6 Fluid PRODUCT WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 2.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Chlorine Min / Nor / Max mg/L

10 Manufacturer HACH

11 Model No. CL17 (5440002)

12 Measuring Principle DPD (COLORIMETRIC)

13 Measuring Type SAMPLING

14 Measuring Rangeability ±5% or ±0.035 mg/L as CL2, whichever is greater

15 Measuring Period every 2' 30"

DETECTOR 16 Cycle Time 2.5 minutes


(ANALYZER) 17 Light Source Class 1 LED (light emitting diode)

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating 5 barG


21 Temperature Rating 40 degC
22 Analyzer Inlet Connection 1/4" OD polyethylene tube

23 Analyzer Drain Connection 1/2" ID flexible hose

24 Sample Flow Rate 200 - 500 ml/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. CL17 (5440002)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP62


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" conduit

36 Set Scale Range 0~5 mg/L


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±5% or ±0.035 mg/L as CL2, whichever is greater

39 Ambient Temperature Rating 5 - 40 ℃

40 Relative Humidity Rating 90% at 40°C

Note.
SHEET 72 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01823

2 Quantity 1 SET
GENERAL 3 Service EXISTING SWRO SWPS SUP TURB

4 P&ID Number 001-2

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 11.5 14.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 20

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable 1

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0 ~ 40 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 73 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01902

2 Quantity 1 SET
GENERAL 3 Service PLANT 1+2 SWPS SUP TURB

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 20

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0 ~ 40 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 74 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01912

2 Quantity 1 SET
GENERAL 3 Service PLANT 3 SWPS SUP TURB

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 20

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0 ~ 40 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 75 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02003

2 Quantity 1 SET
GENERAL 3 Service P1 1ST STG PF 1-12 TURB

4 P&ID Number 002

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 3.0

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~7 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 76 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02013

2 Quantity 1 SET
GENERAL 3 Service P2 1ST STG PF 13-24 TURB

4 P&ID Number 003

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 3.0

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~7 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 77 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02023

2 Quantity 1 SET
GENERAL 3 Service P3 1ST STG PF 25-36 TURB

4 P&ID Number 004

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 3.0

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~7 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 78 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02100

2 Quantity 1 SET
GENERAL 3 Service P1 1ST STG PF 1 CELL A DIS TURB

4 P&ID Number 005

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 3.0

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~7 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 79 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02560

2 Quantity 1 SET
GENERAL 3 Service P2 1ST STG PF 24 CELL B DIS TURB

4 P&ID Number 005

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 3.0

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~7 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 80 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02120

2 Quantity 1 SET
GENERAL 3 Service 1st PRESSURE FILTERED WATER (1&2)

4 P&ID Number 005

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 3.0

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~7 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 81 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02580

2 Quantity 1 SET
GENERAL 3 Service P3 1ST STG PF 25 CELL A DIS TURB

4 P&ID Number 006

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 3.0

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~7 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 82 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02800

2 Quantity 1 SET
GENERAL 3 Service P3 1ST STG PF 36 CELL B DIS TURB

4 P&ID Number 006

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 3.0

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~7 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 83 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02600

2 Quantity 1 SET
GENERAL 3 Service 1st PRESSURE FILTERED WATER (1&2)

4 P&ID Number 006

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 3.0

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~7 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 84 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03047

2 Quantity 1 SET
GENERAL 3 Service P1 2ND STG PF 1 CELL A DIS TURB

4 P&ID Number 010

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 0.5

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~2 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 85 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03672

2 Quantity 1 SET
GENERAL 3 Service P2 2ND STG PF 16 CELL B DIS TURB

4 P&ID Number 010

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 0.5

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~2 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 86 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03072

2 Quantity 1 SET
GENERAL 3 Service 1st PRESSURE FILTERED WATER (1&2)

4 P&ID Number 010

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 0.5

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~2 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 87 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03697

2 Quantity 1 SET
GENERAL 3 Service P3 2ND STG PF 17 CELL A DIS TURB

4 P&ID Number 011

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 0.5

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~2 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 88 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03997

2 Quantity 1 SET
GENERAL 3 Service P3 2ND STG PF 24 CELL B DIS TURB

4 P&ID Number 011

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 0.5

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~2 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 89 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03722

2 Quantity 1 SET
GENERAL 3 Service 1st PRESSURE FILTERED WATER (1&2)

4 P&ID Number 011

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 0.5

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~2 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 90 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04003

2 Quantity 1 SET
GENERAL 3 Service P1+2 RO SYST HDR TURB

4 P&ID Number 018

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 0.5

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~2 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 91 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514163 - AIT - 04023

2 Quantity 1 SET
GENERAL 3 Service P3+4 RO SYST HDR TURB

4 P&ID Number 018

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 0.5

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~2 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 92 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03002

2 Quantity 1 SET
GENERAL 3 Service P1+2 2ND STG PF 1-16 TURB

4 P&ID Number 007

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 0.5

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~2 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 93 OF 105
TURBIDITY ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03007

2 Quantity 1 SET
GENERAL 3 Service P3 2ND STG PF 17-24 TURB

4 P&ID Number 009

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Turbidity Min / Nor / Max NTU 0.5

10 Manufacturer HACH

11 Model No. ULTRATURB (LPV415.99.82002)

12 Measuring Principle 90° scattered light

13 Measuring Type SAMPLING

14 Measuring Rangeability ±0.003 FNU ±0.5% of reading

15 Measuring Period CONTINUOUS

DETECTOR 16 Automatic Cleaning WIPER


(ANALYZER) 17 Light Source LED

18 Temp. Compensation 1m

19 Special Cable N/A

20 Pressure Rating 6 barG


21 Temperature Rating 50 degC
22 Analyzer Inlet Connection 1/4" NPT F

23 Analyzer Drain Connection 1/2" NPT F

24 Sample Flow Rate 200 to 750 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer HACH

29 Model No. SC200 (LPV404.99.00504)

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

33 Enclosure Class IP66


CONVERTER
34 Mounting Type WALL MOUNTING
(ANALYZER)
35 Conduit Connection 1/2" NPT

36 Set Scale Range 0~2 NTU


37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±0.003 FNU ±0.5% of reading

39 Ambient Temperature Rating -20 to 60°C

40 Relative Humidity Rating 0 to 95% RH non-condensing

Note.
SHEET 94 OF 105
Hydrocarbon ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01826

2 Quantity 1 SET
GENERAL 3 Service EXISTING SWRO SEPS SUP HYD

4 P&ID Number 001-2

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 11.5 14.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Hydrocarbon Min / Nor / Max ppm

10 Manufacturer Turner Designs Hydrocarbon Instrument

11 Model No. TD-4100XD

12 Measuring Range 0 to 10 ppm

13 Measuring Principle UV Fluorescence

14 Mesuring Period CONTINUOUS

15 Temp. Compensation N/A


DETECTOR
16 Element Pressure Rating 1.4 barG
(ANALYZER)
17 Element Temperature Rating 88 degC
18 Connection Size & Type 1/2" NPT(F)

19 Special Cable N/A

20 Paper filter exchange period N/A

21 Accuracy Stability : 10 % or better over 6 months

22 Sample Flow Rate 7.5 to 11.5 L/min

23 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 24 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

25 Sample Flow DRAIN

26 Manufacturer Turner Designs Hydrocarbon Instrument

27 Model No. TD-4100XD

28 Output Signal 4 - 20mA

29 Power Supply 120VAC, 50 Hz

30 Display LCD DISPLAY

31 Enclosure Class IP66


CONVERTER
32 Mounting Type WALL MOUNTING
(ANALYZER)
33 Conduit Connection 1/2" NPT

34 Set Scale Range 0 to 10 ppm

35 Range Adjustment ADJUSTABLE AT SITE

36 Typical Accuracy Stability : 10 % or better over 6 months

37 Ambient Temperature Rating 0 to 49 °C

38 Relative Humidity Rating 10 – 90% RH

Note.
SHEET 95 OF 105
Hydrocarbon ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01905

2 Quantity 1 SET
GENERAL 3 Service PLANT 1+2 SWPS SUP HYD

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Hydrocarbon Min / Nor / Max ppm

10 Manufacturer Turner Designs Hydrocarbon Instrument

11 Model No. TD-4100XD

12 Measuring Range 0 to 10 ppm

13 Measuring Principle UV Fluorescence

14 Mesuring Period CONTINUOUS

15 Temp. Compensation N/A


DETECTOR
16 Element Pressure Rating 1.4 barG
(ANALYZER)
17 Element Temperature Rating 88 degC
18 Connection Size & Type 1/2" NPT(F)

19 Special Cable N/A

20 Paper filter exchange period N/A

21 Accuracy Stability : 10 % or better over 6 months

22 Sample Flow Rate 7.5 to 11.5 L/min

23 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 24 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

25 Sample Flow DRAIN

26 Manufacturer Turner Designs Hydrocarbon Instrument

27 Model No. TD-4100XD

28 Output Signal 4 - 20mA

29 Power Supply 120VAC, 50 Hz

30 Display LCD DISPLAY

31 Enclosure Class IP66


CONVERTER
32 Mounting Type WALL MOUNTING
(ANALYZER)
33 Conduit Connection 1/2" NPT

34 Set Scale Range 0 to 10 ppm

35 Range Adjustment ADJUSTABLE AT SITE

36 Typical Accuracy Stability : 10 % or better over 6 months

37 Ambient Temperature Rating 0 to 49 °C

38 Relative Humidity Rating 10 – 90% RH

Note.
SHEET 96 OF 105
Hydrocarbon ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01915

2 Quantity 1 SET
GENERAL 3 Service PLANT 3 SWPS SUP HYD

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Hydrocarbon Min / Nor / Max ppm

10 Manufacturer Turner Designs Hydrocarbon Instrument

11 Model No. TD-4100XD

12 Measuring Range 0 to 10 ppm

13 Measuring Principle UV Fluorescence

14 Mesuring Period CONTINUOUS

15 Temp. Compensation N/A


DETECTOR
16 Element Pressure Rating 1.4 barG
(ANALYZER)
17 Element Temperature Rating 88 degC
18 Connection Size & Type 1/2" NPT(F)

19 Special Cable N/A

20 Paper filter exchange period N/A

21 Accuracy Stability : 10 % or better over 6 months

22 Sample Flow Rate 7.5 to 11.5 L/min

23 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 24 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

25 Sample Flow DRAIN

26 Manufacturer Turner Designs Hydrocarbon Instrument

27 Model No. TD-4100XD

28 Output Signal 4 - 20mA

29 Power Supply 120VAC, 50 Hz

30 Display LCD DISPLAY

31 Enclosure Class IP66


CONVERTER
32 Mounting Type WALL MOUNTING
(ANALYZER)
33 Conduit Connection 1/2" NPT

34 Set Scale Range 0 to 10 ppm

35 Range Adjustment ADJUSTABLE AT SITE

36 Typical Accuracy Stability : 10 % or better over 6 months

37 Ambient Temperature Rating 0 to 49 °C

38 Relative Humidity Rating 10 – 90% RH

Note.
SHEET 97 OF 105
SDI ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514161 - AIT - 01821

2 Quantity 1 SET
GENERAL 3 Service EXISTING SWRO SWPS SUP SDI

4 P&ID Number 001-2

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 11.5 14.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 SDI Min / Nor / Max

10 Manufacturer MABAT

11 Model No. SDI-2200-1

12 Cell Measuring Range 0.00-99.99

13 Measuring Principle ASTM D4189-95 Test Method

14 Measuring Period 5, 10 and 15 minute intervals, as defined in the ASTM D4189-07

15 Temp. Compensation N/A


DETECTOR
16 Element Pressure Rating 6 barG
(ANALYZER)
17 Element Temperature Rating 40 degC
18 Connection Size & Type 3/8" O.D Tube

19 Special Cable N/A

21 Paper filter exchange period 90 measurements per roll

22 Accuracy Less than 0.1 SDI Units

23 Sample Flow Rate 1.5 L/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer MABAT

28 Model No. SDI-2200-1

29 Output Signal 4-20 mA

30 Power Supply 120VAC, 50 Hz

31 Display LCD DISPLAY

32 Enclosure Class IP53


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection 1/2" NPT

35 Set Scale Range 0.00-99.99

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy Less than 0.1 SDI Units

38 Ambient Temperature Rating 10 to 40 °C

39 Relative Humidity Rating 10 – 90% RH

Note.
SHEET 98 OF 105
SDI ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

514161 - AIT - 01906


1 Tag No
514161 - AIT - 01916

2 Quantity 1 SET, Two-channel


GENERAL PLANT 1+2 SWPS SUP SDI (514161-AIT-1906)
3 Service
PLANT 3 SWPS SUP SDI (514161-AIT-1916)

4 P&ID Number 001-1

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 10.9 11.5


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 SDI Min / Nor / Max

10 Manufacturer MABAT

11 Model No. SDI-2200-2

12 Cell Measuring Range 0.00-99.99

13 Measuring Principle ASTM D4189-95 Test Method

14 Measuring Period 5, 10 and 15 minute intervals, as defined in the ASTM D4189-07

15 Temp. Compensation N/A


DETECTOR
16 Element Pressure Rating 6 barG
(ANALYZER)
17 Element Temperature Rating 40 degC
18 Connection Size & Type 3/8" O.D Tube

19 Special Cable N/A

21 Paper filter exchange period 90 measurements per roll

22 Accuracy Less than 0.1 SDI Units

23 Sample Flow Rate 1.5 L/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer MABAT

28 Model No. SDI-2200-2

29 Output Signal 4-20 mA

30 Power Supply 120VAC, 50 Hz

31 Display LCD DISPLAY

32 Enclosure Class IP53


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection 1/2" NPT

35 Set Scale Range 0.00-99.99

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy Less than 0.1 SDI Units

38 Ambient Temperature Rating 10 to 40 °C

39 Relative Humidity Rating 10 – 90% RH

Note.
SHEET 99 OF 105
SDI ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

514162 - AIT - 02001


1 Tag No
514162 - AIT - 02014

2 Quantity 1 SET, two-channel


GENERAL P1 1ST STG PF 1-12 SDI (514162-AIT-2001)
3 Service
P2 1ST STG PF 13-24 SDI (514162-AIT-2014)

4 P&ID Number 002 / 003

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 SDI Min / Nor / Max

10 Manufacturer MABAT

11 Model No. SDI-2200-2

12 Cell Measuring Range 0.00-99.99

13 Measuring Principle ASTM D4189-95 Test Method

14 Measuring Period 5, 10 and 15 minute intervals, as defined in the ASTM D4189-07

15 Temp. Compensation N/A


DETECTOR
16 Element Pressure Rating 6 barG
(ANALYZER)
17 Element Temperature Rating 40 degC
18 Connection Size & Type 3/8" O.D Tube

19 Special Cable N/A

21 Paper filter exchange period 90 measurements per roll

22 Accuracy Less than 0.1 SDI Units

23 Sample Flow Rate 1.5 L/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer MABAT

28 Model No. SDI-2200-2

29 Output Signal 4-20 mA

30 Power Supply 120VAC, 50 Hz

31 Display LCD DISPLAY

32 Enclosure Class IP53


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection 1/2" NPT

35 Set Scale Range 0.00-99.99

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy Less than 0.1 SDI Units

38 Ambient Temperature Rating 10 to 40 °C

39 Relative Humidity Rating 10 – 90% RH

Note.
SHEET 100 OF 105
SDI ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 02021

2 Quantity 1 SET
GENERAL 3 Service P3 1ST STG PF 25-36 SDI

4 P&ID Number 004

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 7.4 8.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 SDI Min / Nor / Max

10 Manufacturer MABAT

11 Model No. SDI-2200-1

12 Cell Measuring Range 0.00-99.99

13 Measuring Principle ASTM D4189-95 Test Method

14 Measuring Period 5, 10 and 15 minute intervals, as defined in the ASTM D4189-07

15 Temp. Compensation N/A


DETECTOR
16 Element Pressure Rating 6 barG
(ANALYZER)
17 Element Temperature Rating 40 degC
18 Connection Size & Type 3/8" O.D Tube

19 Special Cable N/A

21 Paper filter exchange period 90 measurements per roll

22 Accuracy Less than 0.1 SDI Units

23 Sample Flow Rate 1.5 L/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer MABAT

28 Model No. SDI-2200-1

29 Output Signal 4-20 mA

30 Power Supply 120VAC, 50 Hz

31 Display LCD DISPLAY

32 Enclosure Class IP53


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection 1/2" NPT

35 Set Scale Range 0.00-99.99

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy Less than 0.1 SDI Units

38 Ambient Temperature Rating 10 to 40 °C

39 Relative Humidity Rating 10 – 90% RH

Note.
SHEET 101 OF 105
SDI ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03005

2 Quantity 1 SET
GENERAL 3 Service P1+2 2ND STG PF 1-16 SDI

4 P&ID Number 007

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 SDI Min / Nor / Max

10 Manufacturer MABAT

11 Model No. SDI-2200-1

12 Cell Measuring Range 0.00-99.99

13 Measuring Principle ASTM D4189-95 Test Method

14 Measuring Period 5, 10 and 15 minute intervals, as defined in the ASTM D4189-07

15 Temp. Compensation N/A


DETECTOR
16 Element Pressure Rating 6 barG
(ANALYZER)
17 Element Temperature Rating 40 degC
18 Connection Size & Type 3/8" O.D Tube

19 Special Cable N/A

21 Paper filter exchange period 90 measurements per roll

22 Accuracy Less than 0.1 SDI Units

23 Sample Flow Rate 1.5 L/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer MABAT

28 Model No. SDI-2200-1

29 Output Signal 4-20 mA

30 Power Supply 120VAC, 50 Hz

31 Display LCD DISPLAY

32 Enclosure Class IP53


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection 1/2" NPT

35 Set Scale Range 0.00-99.99

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy Less than 0.1 SDI Units

38 Ambient Temperature Rating 10 to 40 °C

39 Relative Humidity Rating 10 – 90% RH

Note.
SHEET 102 OF 105
SDI ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514162 - AIT - 03009

2 Quantity 1 SET
GENERAL 3 Service P3 2ND STG PF 17-24 SDI

4 P&ID Number 009

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 5.3 6.9


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 SDI Min / Nor / Max

10 Manufacturer MABAT

11 Model No. SDI-2200-1

12 Cell Measuring Range 0.00-99.99

13 Measuring Principle ASTM D4189-95 Test Method

14 Measuring Period 5, 10 and 15 minute intervals, as defined in the ASTM D4189-07

15 Temp. Compensation N/A


DETECTOR
16 Element Pressure Rating 6 barG
(ANALYZER)
17 Element Temperature Rating 40 degC
18 Connection Size & Type 3/8" O.D Tube

19 Special Cable N/A

21 Paper filter exchange period 90 measurements per roll

22 Accuracy Less than 0.1 SDI Units

23 Sample Flow Rate 1.5 L/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer MABAT

28 Model No. SDI-2200-1

29 Output Signal 4-20 mA

30 Power Supply 120VAC, 50 Hz

31 Display LCD DISPLAY

32 Enclosure Class IP53


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection 1/2" NPT

35 Set Scale Range 0.00-99.99

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy Less than 0.1 SDI Units

38 Ambient Temperature Rating 10 to 40 °C

39 Relative Humidity Rating 10 – 90% RH

Note.
SHEET 103 OF 105
SDI ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

514163 - AIT - 04005


1 Tag No
514163 - AIT - 04025

2 Quantity 1 SET
GENERAL P1+2 RO SYST HDR SDI (514163-AIT-4005)
3 Service
P3+4 RO SYST HDR SDI (514163-AIT-4015)

4 P&ID Number 018

5
6 Fluid SEAWATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 3.7 6.7


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 SDI Min / Nor / Max

10 Manufacturer MABAT

11 Model No. SDI-2200-2

12 Cell Measuring Range 0.00-99.99

13 Measuring Principle ASTM D4189-95 Test Method

14 Measuring Period 5, 10 and 15 minute intervals, as defined in the ASTM D4189-07

15 Temp. Compensation N/A


DETECTOR
16 Element Pressure Rating 6 barG
(ANALYZER)
17 Element Temperature Rating 40 degC
18 Connection Size & Type 3/8" O.D Tube

19 Special Cable N/A

21 Paper filter exchange period 90 measurements per roll

22 Accuracy Less than 0.1 SDI Units

23 Sample Flow Rate 1.5 L/min

24 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING 25 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

26 Sample Flow DRAIN

27 Manufacturer MABAT

28 Model No. SDI-2200-2

29 Output Signal 4-20 mA

30 Power Supply 120VAC, 50 Hz

31 Display LCD DISPLAY

32 Enclosure Class IP53


CONVERTER
33 Mounting Type WALL MOUNTING
(ANALYZER)
34 Conduit Connection 1/2" NPT

35 Set Scale Range 0.00-99.99

36 Range Adjustment ADJUSTABLE AT SITE

37 Typical Accuracy Less than 0.1 SDI Units

38 Ambient Temperature Rating 10 to 40 °C

39 Relative Humidity Rating 10 – 90% RH

Note.
SHEET 104 OF 105
PHOSPHATE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514164 - AIT - 05007

2 Quantity 1 SET
GENERAL 3 Service RO PROD TNK 1 INL PHOSPHATE

4 P&ID Number 029

5
6 Fluid PRODUCT WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 2.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Phosphate Min / Nor / Max mg/L

10 Manufacturer Endress + Hauser

11 Model No. CA71PH-A12B2A1

12 Measuring Principle COLORIMETRIC

13 Measuring Type SAMPLING

14 Measuring Rangeability ±2%

15 Measuring Period Continuous

DETECTOR 16 Cycle Time Typically, 9.5 minutes at 25 °C


(ANALYZER) 17 Light Source LED

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating ATM barG


21 Temperature Rating 0 - 40 ℃ degC
22 Analyzer Inlet Connection 1/4" OD TUBE

23 Analyzer Drain Connection 1/2" OD TUBE

24 Sample Flow Rate 55 - 300 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer Endress + Hauser

29 Model No. CA71PH-A12B2A1

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

IP43
33 Enclosure Class
CONVERTER
( WITH IP65 WEATHER PROOF ENCLOSURE ) D
(ANALYZER) 34 Mounting Type WALL MOUNTING

35 Conduit Connection 1/2" conduit

36 Set Scale Range 2 mg/L PO4-P (6.132 mg/L PO4)

37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±2%

39 Ambient Temperature Rating 0 - 40 ℃

40 Relative Humidity Rating BELOW THE CONDENSATION LIMIT

Note.
SHEET 105 OF 105
PHOSPHATE ANALYZER
SPEC.NO. REV.

NO BY DATE REVISION D
A H.A. WOO 30 MAY. 2014 Issued for Approval CONTRACT DATE
B H.A. WOO 06 Aug. 2014 Issued for Approval Yokogawa Electric Korea 12 Mar. 2015

C H.A. WOO 13 Nov. 2014 Issued for Approval REQ. P.O.


D H.A. WOO 12 Mar. 2015 Issued for Approval

BY CHK'D APPR.
H.A. WOO T.K.KANG S.H. PARK

1 Tag No 514164 - AIT - 05017

2 Quantity 1 SET
GENERAL 3 Service RO PROD TNK 2 INL PHOSPHATE

4 P&ID Number 029

5
6 Fluid PRODUCT WATER

PROCESS 7 Pressure Min / Nor / Max bar.G 0.0 0.5 2.0


DATA 8 Temperature Min / Nor / Max deg C 11.5 16.0 20.0

9 Phosphate Min / Nor / Max mg/L

10 Manufacturer Endress + Hauser

11 Model No. CA71PH-A12B2A1

12 Measuring Principle COLORIMETRIC

13 Measuring Type SAMPLING

14 Measuring Rangeability ±2%

15 Measuring Period Continuous

DETECTOR 16 Cycle Time Typically, 9.5 minutes at 25 °C


(ANALYZER) 17 Light Source LED

18 Temp. Compensation N/A

19 Special Cable N/A

20 Pressure Rating ATM barG


21 Temperature Rating 0 - 40 ℃ degC
22 Analyzer Inlet Connection 1/4" OD TUBE

23 Analyzer Drain Connection 1/2" OD TUBE

24 Sample Flow Rate 55 - 300 mL/min

25 Sample Connections ( Inlet / Outlet ) 1/2" ANSI 150# FF / 1" ANSI 150# FF
SAMPLING
26 Sample Line Materials & Sizes TEFLON / 1/2" & 1/4"

27 Sample Flow DRAIN

28 Manufacturer Endress + Hauser

29 Model No. CA71PH-A12B2A1

30 Output Signal 4 - 20mA

31 Power Supply 120VAC, 50 Hz

32 Display LCD DISPLAY

IP43
33 Enclosure Class
CONVERTER
( WITH IP65 WEATHER PROOF ENCLOSURE ) D
(ANALYZER) 34 Mounting Type WALL MOUNTING

35 Conduit Connection 1/2" conduit

36 Set Scale Range 2 mg/L PO4-P (6.132 mg/L PO4)

37 Range Adjustment ADJUSTABLE AT SITE

38 Typical Accuracy ±2%

39 Ambient Temperature Rating 0 - 40 ℃

40 Relative Humidity Rating BELOW THE CONDENSATION LIMIT

Note.
ESCONDIDA
WATER SUPPLY PROJECT

DATASHEETS OF ANALYZER INSTRUMENTS

3. CATALOGUE

3.1 Analyzer

REV. D W13018-YT01-CQ-800214
TI00381C/07/EN/14.13 Products Solutions Services
71224431

Technical Information
Liquiline M CM42
Memosens: pH/ORP, oxygen, conductivity
Analog sensors: pH/ORP, conductivity, concentration,
resistance

Two-wire transmitter for Ex and non-Ex areas

Application Your benefits


Liquiline M CM42 is a modular two-wire transmitter for all • Cost-saving:
areas of process engineering. – Simple commissioning with Quick Setup and Navigator
Depending on the version ordered, Liquiline has one or two – Memosens: Plug & play laboratory-calibrated sensors
analog current outputs. In addition, it can be connected to – Predictive maintenance system detects when a sensor
fieldbuses in accordance with the FOUNDATION Fieldbus, needs to be cleaned, calibrated or replaced.
PROFIBUS PA and HART protocol. – Less storage thanks to modular design
The transmitter is suitable for pollution degree 3. – Effective asset management thanks to Fieldcare and W@M
• Safe:
The highly robust plastic version and the hygienic stainless steel
– Memosens: Active display of cable interruption
version are tailored to the following applications:
– User-guided commissioning, graphic display and plain text
• Chemical processes
guidance for maximum operating safety
• Pharmaceuticals industry
– Approvals: ATEX, FM, CSA, NEPSI, TIIS
• Foodstuff technology
– User administration: Code-protected commissioning
• Applications in hazardous areas
– pH glass with Memosens: SIL2 measuring point with TÜV
approval
Liquiline M CM42

Table of contents

Function and system design . . . . . . . . . . . . . . . . . . . . . .3 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29


Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Ambient temperature limits . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Equipment architecture . . . . . . . . . . . . . . . . . . . . . . . . . .4
Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
DAT memory modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Relative humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Suitable sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Pollution degree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

Dependability. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . 30
Reliability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Maintainability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Measured variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Operation concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Measuring ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Display characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Binary input (Memosens): pH/ORP, oxygen, conductivity . . . . 9
Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Analog input: pH / ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Analog input: Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Product page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Product configurator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Product structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Output signal range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Ex specification current output . . . . . . . . . . . . . . . . . . . . . . . . . 12 Certificates and approvals . . . . . . . . . . . . . . . . . . . . . . 34
Ex specification PROFIBUS and FOUNDATION Fieldbus . . . . 12 4-character . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Current output, passive . . . . . . . . . . . . . . . . . . . . . . . . 12
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Signal characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Mounting kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Weather protection cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Active barrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Fieldbus accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Measuring cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Grounding the housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Supply and signal voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Software update and upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Sensor connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Sensor connection: Memosens sensors . . . . . . . . . . . . . . . . . . 18
Sensor connection: analog pH/ORP sensors . . . . . . . . . . . . . . 19
Sensor connection: analog conductivity sensors . . . . . . . . . . . 23

Performance characteristics . . . . . . . . . . . . . . . . . . . . 25
Reference temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Current output response time . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Maximum measured error Memosens . . . . . . . . . . . . . . . . . . . 25
Tolerance of current outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Temperature compensation conductivity . . . . . . . . . . . . . . . . . 25
Temperature adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Wall securing plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Weather protection cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Installation options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Installation in Ex area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

2 Endress+Hauser
Liquiline M CM42

Function and system design


Measuring system A complete measuring system consists of:
• Liquiline M CM42 transmitter with mounting plate (e.g. for wall mounting)
• sensor and suitable sensor cable
• assembly suitable for the sensor (optional)
• post retainer (optional)
• weather protection cover (optional)

a0002012

Examples of a measuring system

pH / ORP (analog) Conductivity, inductive Conductivity, conductive Memosens (digital)


• CM42-P... measurement (analog) measurement (analog) • CM42-K/L/M/N/O...
• Measuring cable CPK9 • CM42-I... • CM42-C... • Measuring cable CYK10
• Cleanfit CPA471 • Dipfit CLA111 assembly • Measuring cable CPK9 • (Unifit CPA442
assembly • Indumax CLS50 sensor • Condumax CLS16 sensor assembly)
• Orbisint CPS11 sensor • Sensor 1)

1)
e.g. CPS11D (pH glass), CPS471D (pH ISFET), CPS341D (enamel pH), CPS16D (pH/ORP combined), CLS15D/
CLS16D/CLS21D (LFc), CLS50D/CLS54D (LFi), COS22D/COS51D (DO)
You can combine your measuring point with a range of assemblies and sensors. For more information,
see the "Accessories" section or the specified documentation.

NOTICE
Effect of climatic conditions: rain, snow, direct sun
Impaired operation to complete transmitter failure
‣ When installing outside, always use the weather protection cover (see accessories).

Endress+Hauser 3
Liquiline M CM42

Equipment architecture
Software You can choose from the following software packages:
• Standard:
Standard application for the most common measuring points
• Advanced:
Many additional functions that increase safety and quality

Package Features

pH /ORP (glass/ISFET) Conductivity Oxygen

Standard Analog sensors Analog sensors Memosens sensors


– Offset and two-point – Sample calibration – Slope calibration
calibration – Temperature calibration: one- – in air (100% rF)
– Sample calibration point – in water (100% air-
– Calibration with standard – Temperature compensation: saturated)
buffers linear, NaCl, ultrapure water – in air (with input of current
– Manual buffer specification (NaCl, HCl) absolute air pressure and
– Temperature – Simulation of current output relative humidity)
compensation – Self-diagnosis – Zero point calibration
– Temperature adjustment – Concentration measurement – Sample calibration
– Isotherm intersection – Clock – Temperature adjustment
– Simulation of current – Medium compensation
Memosens sensors
output – Calibration stability settings
as for analog with the addition
– Self-diagnosis – Simulation of current output
of:
– Calibration stability – Self-diagnosis
– Sensor information
settings – Clock
– Sensor check
– Clock – Sensor information
Memosens sensors – Sensor check
as for analog with the
addition of:
– Sensor information
– Sensor check

Advanced "Standard" software package with the addition of:


Analog sensors Analog sensors Memosens sensors
– Medium compensation – Calibration with separate – Polarization voltage setting
– Calibration timer installation factor (inductive – Calibration timer
– Sensor condition check measurement only) – Sensor statistics
(SCC) – Polarization detection – Operated hours counter
– Process check system (PCS) (conductive measurement – Sterilization cnt.
only)
Memosens sensors
– Temperature compensation
as for analog with the
addition of via user table
– Two-point temperature
– Operated hours counter
adjustment: offset and slope
– Sterilization cnt.
– USP alarm and prealarm
Memosens sensors
as for analog with the addition
of
– Operated hours counter
– Sterilization cnt.

All devices, regardless of measurement parameter


– Logbooks
– Data logbook
– Free assignment of measured values to current outputs (optional)
– Switching on and off of diagnostic function
– Extended user administration
– Current output tables

4 Endress+Hauser
Liquiline M CM42

DAT memory modules There are 3 different types of DAT module


• SystemDAT
– For changing the sensor type, software updates (more recent software version)
and changing the language package
– Part of the scope of delivery of the version ordered and available as an accessory
– Optional accessory
• FunctionDAT
– For software upgrades (additional functionality)
– Extended functional range (2nd current output)
– Optional accessory
• CopyDAT
– Memory for own configuration settings
– Optional accessory
A FunctionDAT is never available for a SIL device since the device already has all the possible
functions and its functions can therefore not be extended.
A SystemDAT is also not available for the SIL device since it would not be possible to ensure the
"functional safety" otherwise.

Suitable sensors pH/ORP


• Analog and Memosens glass electrodes
• Analog and Memosens ISFET sensors
• Analog and Memosens ORP sensors
• Memosens pH/ORP combined sensors
• Analog and Memosens enamel pH electrodes
• Analog single electrodes (glass or antimony)

Conductivity
• Analog and Memosens, conductive sensors
– Two-electrode sensors
– Four-electrode sensors
• Analog and Memosens, inductive sensors

Oxygen
Amperometric sensors:
• with Memosens technology
• 12 and 40 mm design

Dependability
Reliability Memosens
MEMO SENS

Memosens makes your measuring point safer:


• No-contact, digital signal transmission allows optimum galvanic isolation
• No contact corrosion
• Completely watertight
• Sensors can be calibrated in a lab, which increases the availability of measured values
• Predictive maintenance thanks to recording of sensor data, e.g.:
– Total hours of operation
– Hours of operation with very high or very low measured values
– Operating hours at high temperatures
– Number of steam sterilizations
– Sensor status

Quick Setup
To the first measured value within 1 minute
Once you have set up the few parameters in the Quick Setup menu, the measuring point is ready to
measure. The first measured value is reliably displayed.

Endress+Hauser 5
Liquiline M CM42

Process Check System (PCS): Life check


This function is used to check the measuring signal for stagnation. An alarm is triggered if the
measuring signal does not change over a certain period (several measured values). The reason for such
behavior can be contamination, clogging or similar.

yT

a0013107 a0013106

Normal measuring signal, no alarm Stagnating signal, alarm is triggered


y Measuring signal tD Specified time interval
yT Minimum signal fluctuation tA Time at which the alarm is triggered

Sensor Condition Check (SCC, pH only)


This function monitors the electrode status and the degree of electrode aging. The status is displayed
with the messages "SCC electrode condition bad" or "SCC electrode condition sufficient". The status of
the electrode is updated after every calibration.

Sensor Check System (SCS, pH only)


The Sensor Check System monitors the pH glass resistance or reference resistance (for analog sensors
only), thus indicating possible incorrect measurements that may occur as a result of damage to or
clogging of the pH electrode.
In addition, the SCS detects glass breakage in the case of classic glass electrodes as well as leaks in the
case of ISFET sensors.

Polarization monitoring (conductivity only, conductively measured)


Polarization effects in the boundary layer between the sensor and the measuring solution limit the
measuring range of conductive conductivity sensors.
The transmitter can detect and report polarization effects using an intelligent signal evaluation
process.

United States Pharmacopeia (USP) and European Pharmacopeia (EP) (conductivity only)
The requirements for ultrapure water in the pharmaceutical industry are determined primarily by the
American USP and European EP.
The transmitter complies with USP/EP requirements for conductivity measuring systems:
• Exact temperature measurement at the place of conductivity measurement
• Simultaneous display of uncompensated conductivity values and temperature is possible
• Display resolution 0.01 μS/cm
• Exact factory calibration of the transmitter with traceable precision resistance (optional)
• Exact adjustment of sensors in the factory as per ASTM D 1125–91 or ASTM D 5391–99 (optional)
• Temperature-dependent measured value monitoring as per USP and EP.
The limit functions for pharmaceutical water are implemented in the "Advanced" software package in
accordance with USP and EP: For conductivity measurements, the limit functions for pharmaceutical
water are implemented in the software in accordance with USP and EP:
• Water for Injection (WFI) as per USP <645> and EP
• Highly purified water (HPW) as per EP
• Purified water (PW) as per EP
The uncompensated conductivity value and the temperature are measured for the USP/EP limit
functions. The measured values are compared with the tables set down in the standards. An alarm is
triggered if the limit value is exceeded. In addition, a prealarm can be set that flags undesirable
operating statuses before they occur.

6 Endress+Hauser
Liquiline M CM42

Application-optimized calibration models (oxygen only)


In separate functions, the transmitter allows process-adjusted zero point sensor calibration or sensor
calibration via the slope.
There are different calibration models for this, ranging from simple slope calibration in air that is
saturated with water vapor to slope calibration with input of the absolute air pressure and relative
humidity at the measuring location.
The latter model allows in-process calibration during operation as well as during sterilization and
cleaning.
The number of calibrations and sterilizations are tracked separately for the sensor and membrane cap.
The relevant counter can be reset if the membrane cap is replaced.

Maintainability Modular design

a0010477 a0010476

Liquiline inside (version with sensor module, without wiring) CPU and sensor module

Sensor monitor
You will find the sensor monitor in the DIAG menu. Important sensor data incl. warning and alarm
limits are displayed either graphically or numerically.

a0010475 a0015155

Sensor monitor, graphical display (example) Sensor monitor, numerical display (example)

Endress+Hauser 7
Liquiline M CM42

Safety User administration


The device has a user administration function in order to avoid unscheduled changes to the measuring
point.
You need to log on as an Expert to switch on the user administration function. The first time you log
onto the device, you will therefore be asked to enter a password (the "Admin" user name is already
entered).
The user administration function makes two different modes available in the Advanced version:
Roles
• There are 3 fixed user roles (Expert, Maintenance, Operator).
• "Experts" always have all the rights. "Operator" is the role with the least amount of rights.
• Each role has its own individual password. This password can be changed.
• No other user roles can be created.
User accounts
• You can create and manage a maximum of 15 user accounts.
• You need to be logged on as an "Expert" to be able to manage accounts.
• In each user account you specify the user name and the password and assign the new user one of 3
user roles (Operator, Maintenance and Expert).
• Multiple user accounts with the "Expert" role are possible.

SIL
A TÜV-approved SIL2 version of the Liquiline M CM42 is also available (CM42-M* only).
Safety functions
• Safe output of the digitized value at the current output
• Monitoring of the measured value for leaving a defined interval
• Safe calibration and adjustment
Additional information and Safety Manuals can be found at:
www.endress.com/SIL

Reordering validated software


You can order new devices with older, validated software so you do not have to constantly validate new
software versions of new devices. This is possible as long as allowed by the hardware version.

8 Endress+Hauser
Liquiline M CM42

Input
Measured variables --> Documentation of the connected sensor

Measuring ranges --> Documentation of the connected sensor

Binary input (Memosens): Cable specification


pH/ORP, oxygen,
conductivity CYK10 with Memosens Max. cable length 100 m (330 ft)
Fixed cable with Memosens Max. cable length 55 m (180 ft)
MEMO SENS

Ex specification

Intrinsically safe sensor circuit with protection type: Ex ia IIC1), TIIS Ex ib IIC
Intrinsically safe sensor circuit with protection type: Ex ic IIC2)
Intrinsically safe sensor circuit with protection type: Ex ia Ga IIC3)
Max. output voltage Uo: 5.04 V
Max. output current Io: 80 mA
Max. output Po: 112 mW
For connecting the measuring cable CYK104)
For connecting fixed cable of sensor CLS50D5)

1) CM42-*G*********, CM42-*X*********, CM42-*Z*********


2) CM42-*V*********
3) CM42?*I*********
4) CM42-K**********, CM42-M**********, CM42-N**********
5) CM42-L**********

Endress+Hauser 9
Liquiline M CM42

Analog input: pH / ORP Cable specification

Without SCS Max. cable length 50 m (160 ft)


With SCS Max. cable length 20 m (65 ft)

Applicable temperature sensors


• Pt100
• Pt1000
• NTC 30K

Ex specification

Intrinsically safe sensor circuit with protection type: Ex ia IIC1)


Intrinsically safe sensor circuit with protection type: Ex ic IIC2)
Intrinsically safe sensor circuit with protection type: Ex ia Ga IIC3)
Glass ISFET
Max. output voltage Uo 10.08 V 10.08 V
Max. output current Io 4.1 mA 50.7 mA
Max. output Po 10.2 mW 128 mW
Max. external inductance Lo 1 mH 1 mH
Max. external capacitance Co 250 nF 250 nF
Connection class as per NE1164) SensISCO1X -

1) CM42-*G*********, CM42-*X*********, CM42-*Z*********


2) CM42-*V*********
3) CM42-*I**********
4) CM42-*G*********

When pH/ORP glass electrodes are connected to terminals 317, 318, 320, 111, 112 and 113, the
device corresponds to connection class 1 as per NAMUR Recommendation NE116 (SensISCO).
Terminals 315 and 316 may not be connected for this categorization.
The device is labeled SensISCO1X.

Input impedance
> 1*1012 Ω (under nominal operating conditions)

Input leakage current


< 1*10-13 A (under nominal operating conditions)

10 Endress+Hauser
Liquiline M CM42

Analog input: Conductivity Cable specification

Conductivity/specific resistance, conductively


measured1)
Two-electrode sensor Max. cable length 100 m (330 ft)
10 μS.k to 20 mS.k / 0.1 MΩ/k to 50 Ω/k Max. cable length 50 m (160 ft)
5 μS.k to 20 mS.k / 0.2 MΩ/k to 50 Ω/k Max. cable length 15 m (50 ft)
0.1 μS.k to 20 mS.k / 20 MΩ/k to 50 Ω/k
Conductivity, measured conductively
Four-electrode sensor
10 μS.k to 1.5 S.k Max. cable length 100 m (330 ft)
0.1 μS.k to 20 mS.k Max. cable length 15 m (50 ft)
Conductivity, measured inductively2) Max. cable length 55 m (180 ft)

1) with cable CYK71, CPK9 or fixed cable


2) with cable CLK5, CLK6 or fixed cable

Applicable temperature sensors


• Pt100
• Pt1000

Ex specification, conductive sensors

Intrinsically safe sensor circuit with protection type: Ex ia IIC1)


Intrinsically safe sensor circuit with protection type: Ex ic IIC2)
Intrinsically safe sensor circuit with protection type: Ex ia Ga IIC3)
Max. output voltage Uo 10.08 V
Max. output current Io 23 mA
Max. output Po 57 mW
Max. external inductance Lo 300 μH
Max. external capacitance Co 50 nF

1) CM42-*G*********, CM42-*X*********, CM42-*Z*********


2) CM42-*V*********
3) CM42?*I*********

Ex specification, inductive sensors

Intrinsically safe sensor circuit with protection type: Ex ia IIC1)


Intrinsically safe sensor circuit with protection type: Ex ic IIC2)
Intrinsically safe sensor circuit with protection type: Ex ia Ga IIC3)
Max. output voltage Uo 10.08 V
Max. output current Io 64 mA
Max. output Po 128 mW
For connecting the inductive sensors CLS50, CLS54

1) CM42-*G*********, CM42-*X*********, CM42-*Z*********


2) CM42-*V*********
3) CM42?*I*********

Endress+Hauser 11
Liquiline M CM42

Output
Output signal 1x 4 to 20 mA, passive, potentially isolated against sensor circuit1)2)
2x 4 to 20 mA, passive, potentially isolated against sensor circuit and against each other 1)2)3)
PROFIBUS PA, potentially isolated against sensor circuit1)2)4)
FOUNDATION Fieldbus, potentially isolated against sensor circuit1)2)5)

Signal on alarm 3.6 to 22.0 mA (3.6 mA fixed value when using HART communication)
digitally via fieldbus6)

Load Max. load with supply voltage of 24 V: 500 Ω


Max. load with supply voltage of 30 V: 750 Ω

Output signal range 3.6


to 22.0 mA

Ex specification current
Intrinsically safe supply and signal circuits, passive
output
Max. input voltage Ui 30 V
Max. input current Ii 100 mA
Max. input Pi 750 mW
Max. internal inductance Li 29 μH (output 1)
24 μH (output 2)
Max. internal capacitance Ci 1.2 nF (output 1)
0.2 nF (output 2)

Ex specification PROFIBUS
Suitable for use as a field device in a FISCO system as per EN/IEC 60079-27
and FOUNDATION Fieldbus
Max. input voltage Ui 17.5 V
Max. input current Ii 380 mA
Max. input Pi 5.32 W
Max. internal inductance Li <10 μH
Max. internal capacitance Ci < 5 nF

Current output, passive


Span 3.6
to 22.0 mA

Signal characteristics Linear

Cable specification Cable type: shielded wire, Ø 2.5 mm (14 AWG)

1) Potential isolation for Memosens in sensor plug


2) For inductive sensors with Memosens protocol CLS50D and CLS54D not potentially isolated against sensor circuit!
3) Current output 1 and current output 2 (optional)
4) For version with PROFIBUS PA
5) For version with FOUNDATION Fieldbus
6) For version with PROFIBUS PA or FOUNDATION Fieldbus

12 Endress+Hauser
Liquiline M CM42

Power supply
Grounding the housing Plastic housing

! WARNING
Electric voltage at ungrounded cable mounting rail
No shock protection
‣ Connect the cable mounting rail to the foundation ground using a separate ≥2.5 mm2 (14 AWG)
functional ground.

a0003617 1 Cable mounting rail


Grounding the housing 2 ≥2.5 mm2 (14 AWG) functional ground

Stainless steel housing

! WARNING
Electric voltage at ungrounded housing
No shock protection
‣ Connect the external ground connection of the housing to the foundation ground using a separate
wire (GN/YE) (≥2.5 mm2, corresponds to 14 AWG).

a0003616 1 Outer ground connection


Grounding the housing 2 ≥2.5 mm2 (14 AWG) wire (GN/YE)

Endress+Hauser 13
Liquiline M CM42

Supply and signal voltage


! WARNING
In sensor CLS50D, the internal sensor circuits are connected to the shielded wire of the supply
cable and are therefore also connected to ground via the shielded connection of the CM42
→ ä 18.
The overheating of electronic components constitutes a fire and explosion hazard and may result in
serious injury and death.
‣ Never connect the supply circuit to the ground.
‣ Use an active barrier (e.g. RN221) to ensure safe galvanic isolation during operation in explosive
atmospheres (see Accessories).

NOTICE
The internal sensor circuits of CLS50D and CLS54D are also connected to ground via the shielded
wire of the supply cable and via the shielded connection of the CM42 → ä 18.
Incorrect connection of the supply circuit results in the overheating of components which in turn
results in the transmitter being destroyed.
‣ Never connect the supply circuit to the ground.
For sensor CLS50, the following applies:
The internal sensor circuits in the sensor are connected to the shielded wire of the supply cable.
This results in the sensor circuits also being connected to the ground. The transmitter CM42
features safe galvanic isolation between the supply and sensor circuit.

4 to 20 mA
‣ Connect the transmitter with a shielded two-wire cable.
 How the shield is connected depends on the interference influence expected. To suppress
electrical fields, it suffices to ground the shield at one end. If you also want to suppress
interference from a magnetic alternating field, you must ground the shield at both ends.
In the case of a SIL device, you must ground both current outputs at both ends.
The second current output can be ordered as an option
(see "Ordering information").

133
4 ... 20 mA

+
Out 1

134 –

233
4 ... 20 mA

+
Out 2

234 –

a0005037 a0005038

View in device (CPU module) Wiring diagram

14 Endress+Hauser
Liquiline M CM42

4 to 20 mA / HART
For safe communication via the HART protocol and for compliance with NAMUR NE 21, you must use
a two-wire cable that is grounded at both ends.
‣ Connect the transmitter to a two-wire cable that is grounded at both ends.

a0002365 a0003100

View in device (CPU module) Wiring diagram

PROFIBUS PA and FOUNDATION Fieldbus


Use a fieldbus cable grounded at both ends (device and PCS).
There are various ways of connecting the unit:
1. Two-wire cable grounded at both ends, "hard grounding"
(generally to be given priority over "capacitive connection to ground")
2. If there is a risk of large potential equalization currents.
Shielded two-wire cabling, "capacitive connection to ground" (shield grounded on device side via
capacitor, "C-module" accessory required)
Do not use in the hazardous area!
3. Use the fieldbus connection socket (accessories)

"Hard grounding"
1. Connect the cable shield to the cable mounting rail.
2. Connect the cable cores as per the assignment (→ å ).

a0004060 a0001640

View in device (CPU module) Wiring diagram

Endress+Hauser 15
Liquiline M CM42

"Capacitive ground connection"


1. Pull back the braided shield, put the stranded extension wire of the C-module (item 1) onto the
exposed shield and tighten the clip:

2. Connect the extension wire to the cable mounting rail.


3. Connect the cable cores as per the assignment (→ å ).

a0004071 a0004074

View in device (CPU module) Wiring diagram

"Fieldbus connection socket"


1. Screw the fieldbus connection socket into the housing bushing.
2. Trim the connection cores of the socket to approx. 15 cm.
3. Connect the cable cores as per the assignment. In doing so, you must place the cable shield (GN/
YE) on the cable mounting rail (→ å ).

a0002378 a0004063

View in device (CPU module) Wiring diagram

16 Endress+Hauser
Liquiline M CM42

Cable specification
max. cable cross-section: 2.5 mm2 (i14 AWG), GND 4 mm2 (i12 AWG)

Supply voltage

18
17
16.5
15.5

V
14

12.5 A
11.5 B

10
3.6 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
mA
a0008804

Minimum supply voltage at the transmitter depending on the output current


A with HART communication
B without HART communication

Profibus PA / FOUNDATION Fieldbus: 9 to 32 V DC (non-Ex)


9 to 17.5 V DC (Ex)
Bus current consumption: 22 mA

Sensor connection Explanation of abbreviations in the following diagrams:

Abbreviation Meaning

pH pH signal

Ref Signal from reference electrode


Src Source

Drn Drain

PM Potential matching
U+
Power supply of the digital sensor
U–

Com A
Communication signals of the digital sensor
Com B

ϑ Signal of the temperature sensor

d.n.c. Do not connect!

NOTICE
No shielding against electrical and magnetic interference
Interference can give erroneous measurement results
‣ You must connect shielded connections or terminals with functional earth ( ) ) (there is no
protective earth in the case of the plastic housing ( )).
‣ Keep magnetic interference fields away from the sensor since inductive conductivity sensors use
magnetic fields.

Endress+Hauser 17
Liquiline M CM42

Sensor connection: pH/ORP incl. combined sensors, oxygen, conductively measured conductivity
Memosens sensors

a0001087 a0001078

View in device (sensor module) Wiring diagram

Inductively measured conductivity

a0014525 a0020732

View in device (sensor module) Wiring diagram CLS50D, CLS54D

18 Endress+Hauser
Liquiline M CM42

Sensor connection: Glass electrodes, with PAL (symmetrical)


analog pH/ORP sensors
CM42 CPK9 Sensor

GN
112
J
WH
113

YE
111

315

CPK1
316

BN
320 PM

BK
318 Ref

317 pH

319

a0001088 a0001079

View in device (sensor module) Wiring diagram

Glass electrodes, without PAL (asymmetrical)

CM42 CPK9 / CPK1 Sensor

GN
112
J
WH
113

YE
111

315

316 CPK1

BK
320

318 Ref

317 pH

319
d.n.c.

BN

a0001089 a0001080

View in device (sensor module) Wiring diagram

Endress+Hauser 19
Liquiline M CM42

ISFET sensors, with PAL (symmetrical)

CM42 CPK12 Sensor

GN
112
J
WH
113

YE
111

315 Src

Ref
RD
316 Drn

BN
320 PM

318 BK

317

319

a0001090 a0001076

View in device (sensor module) Wiring diagram

ISFET sensors, without PAL (asymmetrical)

CM42 CPK12 Sensor

GN
112
J
WH
113

YE
111

315 Src

Ref
RD
316 Drn

BK
320

318 d.n.c.

317

319

BN

a0001084 a0001077

View in device (sensor module) Wiring diagram

20 Endress+Hauser
Liquiline M CM42

Pfaudler electrodes
With PML (symmetrical) With PML (symmetrical)
Pfaudler electrode, absolute Pfaudler electrode, relative
Type 03/type 04 Type 18/type 40
CM42 Pfaudler
18/40
BN 3
112

GN
113 4

BK
111 6

315

316

BU PM
320 5

318

WH pH
317 1

BN Ref
319 2
OG
8

VT Shield
9
a0010467 a0010468

Wiring diagram Wiring diagram

With PML (symmetrical) Without PML (asymmetrical)


pH Reiner Pfaudler electrode, absolute
Type 03/type 04

a0010469 a0010470

Wiring diagram Wiring diagram

Endress+Hauser 21
Liquiline M CM42

Individual electrodes (e.g. CPS64 glass or antimony), without PAL (asymmetrical)

a0010465 a0010466

View in device (sensor module) Wiring diagram

Glass electrode and ORP sensor for rH measurement


For rH measurement, connect a pH
CM42 CPK9 Sensor
combination electrode (e.g. CPS11 with
sensor cable CPK9) and an ORP sensor
GN
112 (e.g.
 CPS12 with sensor cable CPK1).
WH
113

YE
111

315
CPK1

316

320 ORP

BK
318 Ref

317 pH

d.n.c.
BK
319

BN

a0010558

Wiring diagram

22 Endress+Hauser
Liquiline M CM42

Sensor connection: Conductive sensors, two-electrode sensors


analog conductivity sensors
CM42 CPK9 / CYK71 Sensor

GN
112
J
WH
113

YE
111

219

BK
220

d.n.c.
222

221

BN

a0001086 a0001083

View in device (sensor module) Wiring diagram

Conductive sensors, four-electrode sensors

a0002384 a0002371

View in device (sensor module) Wiring diagram

Endress+Hauser 23
Liquiline M CM42

Inductive sensors

a0001085 a0001082 a0020733

View in device (sensor Wiring diagram CLS52, CLS54 Wiring diagram CLS50
module)

24 Endress+Hauser
Liquiline M CM42

Performance characteristics
Reference temperature Standard 25 °C (77 °F)
configurable from –5 to 100 °C (23 to 212 °F)7)

Current output response time t90 = max. 500 ms for an increase from 4 to 20 mA

Maximum measured error Thanks to digital data transmission, the measured value supplied by the sensor at the sensor input is
Memosens passed on accurately. Measurement accuracy depends solely on the connected sensor and the quality
of its adjustment.

Tolerance of current outputs 25 μA, additionally

Repeatability --> Documentation of the connected sensor

Temperature compensation
Types of compensation Range
conductivity
None
Linear α = 0.00 to 20.00 %.K–1
NaCl to IEC 746-3 0 to 100 °C (32 to 212 °F)
Natural waters to IEC 7888 0 to 35 °C (32 to 95 °F)
Ultrapure water NaCl 0 to 100 °C (32 to 212 °F)
Ultrapure water HCl (also for NH3) 0 to 60 °C (32 to 140 °F)
4 user-definable tables 1)

1) with "Advanced" software package

Temperature adjustment
Temperature offset –5 to +5 °C (23 to 41 °F)
Temperature increase 0.9 ... 1.1 1)

1) with "Advanced" software package

Installation
Wall securing plate
3)
.4
(0
11

220 (8.66) 112 (4.41)


x

2 (0.08)
2

85 (3.35) 85 (3.35) 86 (3.39)

122°
95 (3.74)
80 (3.15)
70 (2.76)

2 x R8 2 x R8 2 x R3

R3
8)

70 (2.76)
.2
(0

4)

141 (5.55) 2 x R70


7

.2
Ø

(0
x

161 (6.34)
4

Ø
x

196 (7.72) mm (inch)


4

a0010363

Mounting plate

7) with "Advanced" software package

Endress+Hauser 25
Liquiline M CM42

NOTICE
Weather protection cover Effect of climatic conditions (rain, snow, direct sun etc.)
Impaired operation to complete transmitter failure
‣ When installing outside, always use the weather protection cover (accessory).

Weather protection cover for plastic housing

a0001671

Weather protection cover

Weather protection cover for stainless steel housing

a0001676

Weather protection cover CYY101


A Wall mounting
B Pipe or circular post mounting
1 Circular post fixture (accessory)

To mount the weather protection cover on pipes or round posts, you additionally require a round
post mount, --> "Accessories". (or "Installation options")

26 Endress+Hauser
Liquiline M CM42

Installation options
1 2

4 3
a0002166 a0003092 a0005036

Wall mounting Post mounting Panel mounting


– Weather protection cover (optional)
1 Liquiline
2. 3 Mounting plate (1x accessory)
4 Pipe or post (circular/square)

Wall mounting Pipe mounting Panel mounting

Plastic housing

without weather Mounting plate Standard Mounting kit: 51518263 Installation 51518173
protection cover kit:

with weather Protection 51517382 Mounting kit: 51518263


protection cover cover Protection 51517382
cover

Stainless steel
housing

without weather Mounting plate Standard Mounting kit: 51518286 Installation 51518284
protection cover kit:

with weather Protection CYY101-A Protection CYY101-A


protection cover cover cover
Circular post 50062121
fixture:

Installation in Ex area CM42-*G

a0001312–en

Installation in Ex area
1 Sensor in Ex version 5 Active barrier, e.g. Preline RN221
2 Intrinsically safe sensor circuit Ex ia 6 Signal line HART/PROFIBUS/FF
3 Transmitter 7 Fieldcare via PROFIBUS/FF
4 Supply and signal circuit Ex ib (4 to 20 mA) 8 HART handheld terminal

Endress+Hauser 27
Liquiline M CM42

CM42-*V

a0014574–en

Installation in Ex area
1 Sensor in Ex version 5 Active barrier, e.g. Preline RN221
2 Intrinsically safe sensor circuit Ex ic 6 Signal line HART/PROFIBUS/FF
3 Transmitter 7 Fieldcare via PROFIBUS/FF
4 Supply and signal circuit Ex nA (4 to 20 mA)8 HART handheld terminal

CM42-*P/S

a0014572–en

Installation in Ex area
1 Sensor in Ex version 5 Active barrier, e.g. Preline RN221
2 Intrinsically safe sensor circuit 6 Signal line HART/PROFIBUS/FF
3 Transmitter 7 Fieldcare via PROFIBUS/FF
4 Supply and signal circuit (4 to 20 mA) 8 HART handheld terminal

28 Endress+Hauser
Liquiline M CM42

CM42-*X/Z

a0014573–en

Installation in Ex area
1 Sensor in Ex version 5 Active barrier, e.g. Preline RN221
2 Intrinsically safe sensor circuit Ex ia 6 Signal line HART/PROFIBUS/FF
3 Transmitter 7 Fieldcare via PROFIBUS/FF
4 Supply and signal circuit Ex nA (4 to 20 mA)8 HART handheld terminal

Environment
Ambient temperature range Non-Ex version
–30 to 70 °C (–20 to 160 °F)
Ex version: ATEX II (1)2G
–20 to 50 °C (T6)
–20 to 55 °C (T4)
Ex version: ATEX II 3G
-10 to 50 °C (T6)
Ex version: FM
–20 to 50 °C (0 to 120 °F) (T6)
Ex version: CSA
–20 to 50 °C (0 to 120 °F) (T6)
–20 to 55 °C (0 to 130 °F) (T4)

Ambient temperature limits –30 to 80 °C (–20 to 175 °F)

Storage temperature –40 to 80 °C (–40 to 175 °F)


–25 to 85 °C (–13 to 185 °F)

Electromagnetic Interference emission and interference immunity to EN 61326-1: 2004, Category B (residential
compatibility environments)

Degree of protection IP66 / IP67 / NEMA 4X

Relative humidity 10 ... 95%,


not condensing

Pollution degree The product is suitable for pollution degree 3.

Endress+Hauser 29
Liquiline M CM42

Mechanical construction
Dimensions Plastic housing

177 (6.97)
164 (6.46)
148 (5.83)
110 (4.33)
220 (8.66) 130 (5.12)

2 (0.08)
138 (5.43)
144 (5.67)
144 (5.67)

95 (3.74)
(0.87)

mm (inch)
22

a0011834

Dimensions

Stainless steel housing


165 (6.50)
150 (5.91)
134 (5.28)
110 (4.33)

26
(1.02)
132.5 (5.22)
134.5 (5.30)
2 (0.08)

136.5 (5.37)
136.5 (5.37)
220 (8.66)
174 (6.85)
174 (6.85)

95 (3.74)
(0.39)

mm (inch)
10

a0011833

Dimensions

30 Endress+Hauser
Liquiline M CM42

Weight Plastic housing


1.5 kg (3.3 lbs)

Stainless steel housing


2.1 kg (4.6 lbs)

Material
Plastic housing
Housing PC?FR (polycarbonate, flame-retardant)
Housing seals Foamed silicone, EPDM
Stainless steel housing
Housing Stainless steel 1.4301 (AISI 304)
Housing seals EPDM
Plastic and stainless steel housing
Module housing PC (polycarbonate)
Soft keys TPE
Cable mounting rail Stainless steel 1.4301 (AISI 304)
Display panel PC-FR (polycarbonate, flame-retardant)

Operability
Operation concept The unique operating concept sets new standards:
• Fewer user errors thanks to very easy operation
• Rapid configuration using the Navigator.
• Intuitive configuration and diagnosis due to plain text display

a0001984 a0010473

Navigator Plain text menu

Display characteristics LCD display: FSTN technology (FSTN = Foil Super Twisted Nematic)
Size: 94 x 76 mm (3.7 x 3.0")
Resolution: 240 x 160 dots

Endress+Hauser 31
Liquiline M CM42

Operating elements

a0010480

Operation overview
1 Alarm LED
2 Display, current display: pH measuring mode
3 Navigator
4-7 Softkeys
8 Display of softkey function (menu-dependent)

Ordering information
Product page You can create a valid and complete order code on the Internet using the Configurator tool.
Enter the following address in the browser to launch the product page:
www.products.endress.com/cm42

Product configurator 1. On the right-hand side of the product page, you will find the following selection options:

2. Click "Configure this product".


3. The Configurator opens in a new window. You can now configure your device, and you will receive
the valid and complete order code for it.
4. Now export the order code as a PDF or Excel file. To do so, click the specific button at the top of
the page.

32 Endress+Hauser
Liquiline M CM42

Product structure Product structures always reflect the status at time of going to press. You can create an up-to-date
and complete order code on the internet using the Configurator.

Sensor input
A No module
C Conductivity, conductive measurement, analog sensor
I Conductivity, inductive measurement, analog sensor
K Digital sensor: conductivity, conductive measurement
L Digital sensor: conductivity, inductive measurement
M Digital sensor: pH/ORP measurement, glass sensor
N Digital sensor: pH measurement, ISFET sensor
O Digital sensor: amperometric oxygen measurement
P pH (glass/ISFET) or ORP, analog sensor

Approval
A Non-hazardous area, CSA GP
G ATEX II (1)2G Ex ib [ia Ga] IIC T6 Gb / II 3 D tD A22 IP67 T85°C
P FM IS NI Cl. I, Div. 1&2, Groups A-D
S CSA IS NI Cl. I, II, III, Div. 1&2, Groups A-G
T TIIS Ex ib IIC T4
V ATEX/NEPSI II 3G Ex nA[ic] IIC T6 / II 3 D tD A22 IP67 T85°C
X ATEX II (2)3G Ex nA[ia] IIC T6 / II 3 D tD A22 IP67 T85°C
Z NEPSI Ex nA[ia] EC T6

Certificate
A Not selected
B Test certificate to EN 10204, 3.1
C Test certificate to EN 10204, 3.1, factory calibration certificate

Output
0 1 current output 4 to 20 mA, HART®
1 2 current outputs 4 to 20 mA, 1x HART®
2 PROFIBUS PA
3 FOUNDATION Fieldbus
4 2 current outputs 4 to 20 mA, SIL

Housing
0 Plastic housing
1 Housing made of stainless steel 1.4301 (AISI 316L )

Cable entry
0 M20 x 1.5
1 NPT ½"
2 G½

Software
EA Standard version
EB Advanced version

Device language
C Cz/En (Docs, Czech)
D De/En (Docs, German)
E En/De (Docs, English)
F Fr/En (Docs, French)
J Ja/En (Docs, Japanese)
K Kr/En (Korean documentation)
L Pl/En (Docs, Polish)
N Nl/En (Docs, Dutch)
P Pt/En (Portuguese documentation)
R Ru/En (Russian documentation)
S Es/En (Docs, Spanish)
T It/En (Docs, Italian)
W Sv/En (Swedish documentation)
Z Zh/En (Chinese documentation)

Documentation
0 Installation and configuration

Additional fittings
0 Basic version
1 SystemDAT CY42-S1

CM42– Complete order code

Endress+Hauser 33
Liquiline M CM42

Optional
Measuring point identification (Tag), stainless steel 1.4404
Add a 1 to the end of the order code.

Scope of delivery Depending on the version, the scope of delivery comprises::


• 1 transmitter in the version ordered
• 1 mounting plate incl. 4 flat head screws
• 1 sheet of adhesive labels (nameplate,
terminal connection diagrams)
• 1 test certificate to EN 10204-3.1
• 1 factory calibration certificate
• Operating Instructions Part 1 and 2, BA00381C "Commissioning" and BA00382C "Operation" in the
language ordered
• 1 CD-ROM with additional documentation
• 1 manufacturer's certificate
• Safety Manual
• Maintenance Documentation SIL 2

Certificates and approvals


4-character Declaration of conformity
The product meets the requirements of the harmonized European standards.
It therefore complies with the statutory requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.

Ex approval Depending on the version ordered:


• ATEX II (1)2G Ex ib [ia Ga] IIC T6 Gb / II 3 D tD A22 IP67 T85°C
• ATEX II 3G Ex nA[ic] IIC T6 / II 3 D tD A22 IP67 T85°C (applied for)
• ATEX II (2)3G Ex nA[ia] IIC T6 / II 3 D tD A22 IP67 T85°C (applied for)
• NEPSI Ex nA[ia] IIC T6
• NEPSI Ex nA[ic] IIC T6
• CSA IS NI Cl.I, II, III, Div. 1&2, Grps. A-G
• FM IS NI Cl.I, Div. 1&2, Grps. A-D
• TIIS Ex ib IIC T4

34 Endress+Hauser
Liquiline M CM42

Accessories
The following is a list of the most important accessories available at the time this documentation
was issued. For accessories not listed here, please contact your local service or sales center.

Mounting kits Post retainer for plastic housing


• 1 Mounting plate
• 2 Threaded rods
M5x75 mm A2
• 2 Hexagonal nuts M5 A2, DIN 934
• 2 spring washers A2 DIN127, form B5 (M5)
• 2 Washers A 5.3, DIN125 A2
• Order No. 51518263
Post retainer for stainless steel housing
• 1 Mounting plate
• 2 Threaded rods
M5x75 mm A2
• 2 Hexagonal nuts M5 A2, DIN 934
• 2 spring washers A2 DIN127, form B5 (M5)
• 2 Washers A 5.3, DIN125 A2
• Order No. 51518286
Panel installation kit for plastic housing for panel cutout 138x138 mm (5.43x5.43 inch)
• 1 Panel installation seal
• 2 Tensioning screws M6x150 mm
• 4 Hexagonal nuts M6, DIN934 A2
• 4 spring washers, A2 DIN127, form B6
• 4 Washers A6.4, DIN125 A2
• Order No. 51518173
Panel installation kit for stainless steel housing for panel cutout 138x138 mm (5.43x5.43 inch)
• 1 Panel installation seal
• 2 Tensioning screws M6x150 mm
• 4 Hexagonal nuts M6, DIN934 A2
• 4 spring washers, A2 DIN127, form B6
• 4 Washers A6.4, DIN125 A2
• Order No. 51518284

Weather protection cover Weather protection cover for plastic housing


• Order No. 51517382
Weather protection cover for stainless steel housing
• Order No. CYY101-A

Active barrier Active barrier RN221N


• With power supply for safe isolation of 4 to 20 mA standard signal circuits
• Technical Information TI073R/09/EN

Endress+Hauser 35
Liquiline M CM42

Fieldbus accessories HART handheld terminal DXR375


• For communicating with a HART-compatible device via a 4 to 20 mA line
• Order No. DXR375
Commubox FXA291
• Interface module between CDI and USB PC interface
• Technical Information TI405C/07
• Order No. 51516983
Fieldbus connection socket
• FOUNDATION Fieldbus M20 7/8" connection
• Order No. 51517974
M12 connector
• Four-pole metal connector for mounting on transmitter
• For connecting to connection box or cable jack. Cable length 150 mm (5.91").
• Order No. 51502184
C-module accessories bag
• 1 capacitor for connecting the cable shielding to ground potential
• Kit documentation SD108C/07/A3
• Order No. 71003097

Measuring cable Memosens data cable CYK10


• For digital sensors with Memosens technology
pH, ORP, oxygen (amperometric), chlorine, conductivity (conductive)
• Order as per product structure (--> Online Configurator, www.products.endress.com/cyk10)
• Technical Information TI00118C/07/EN
Memosens data cable CYK11
• Extension cable for digital sensors with Memosens protocol
• Order as per product structure (--> Online Configurator, www.products.endress.com/cyk11)
Measuring cable CPK9
• For sensors with ESA plug-in head, for high-temperature and high-pressure applications, IP 68
• Order as per order structure
• Technical Information TI00501C/07/EN
Special measuring cable CPK12
• For ISFET sensors and pH/ORP glass electrodes with TOP68 plug-in head
• Order as per order structure
• Technical Information TI00118C/07/EN
Measuring cable CYK71
• non-terminated cable for connecting sensors and for extending sensor cables
• sold by the meter, order numbers:
– non-Ex version, black: 50085333
– Ex version, blue: 50085673
Measuring cable CLK6
• Extension cable for extending CLS50/52/54 and transmitter via VBM junction box, sold by the
meter
• Order No. 71183688

36 Endress+Hauser
Liquiline M CM42

Sensors pH/ORP glass sensors


Orbisint CPS11/CPS11D
• pH electrode for process engineering
• Optional SIL version for connecting to SIL transmitter
• With dirt-repellent PTFE diaphragm
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps11 or
www.products.endress.com/cps11d)
• Technical Information TI00028C/07/EN
Orbisint CPS12/CPS12D
• ORP electrode for process engineering
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps12 oder
www.products.endress.com/cps12d)
• With dirt-repellent PTFE diaphragm
• Technical Information TI00367C/07/EN
Ceraliquid CPS41/CPS41D
• pH electrode with ceramic diaphragm and KCl liquid electrolyte;
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps41 or
www.products.endress.com/cps41d)
• Technical Information TI00079C/07/EN
Ceraliquid CPS42/CPS42D
• Redox electrode with ceramic diaphragm and KCl liquid electrolyte;
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps42 or
www.products.endress.com/cps42d)
• Technical Information TI00373C/07/EN
Ceragel CPS71/CPS71D
• pH electrode with double-chamber reference system and integrated bridge electrolyte;
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps71 or
www.products.endress.com/cps71d)
• Technical Information TI00245C/07/EN
Ceragel CPS72/CPS72D
• ORP electrode with double-chamber reference system and integrated bridge electrolyte;
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps72 or
www.products.endress.com/cps72d)
• Technical Information TI00374C/07/EN
Orbipore CPS91/CPS91D
• pH electrode with open aperture diaphragm for media with high contamination potential;
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps91 or
www.products.endress.com/cps91d)
• Technical Information TI00375C/07/EN
Orbipore CPS92/CPS92D
• ORP electrode with open aperture diaphragm for media with high contamination potential;
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps92 or
www.products.endress.com/cps92d)
• Technical Information TI00435C/07/EN

pH/ORP combined sensors


Memosens CPS16D
• pH ORP combined sensor for process engineering, with dirt-repellent PTFE diaphragm
• With Memosens technology
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps16d)
• Technical Information TI00503C/07/EN
Memosens CPS76D
• pH ORP combined sensor for process engineering, hygiene and sterile applications
• With Memosens technology
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps76d)
• Technical Information TI00506C/07/EN

Endress+Hauser 37
Liquiline M CM42

Memosens CPS96D
• pH and ORP combined sensor for chemical processes
• With poison-resistant reference with ion trap
• With Memosens technology
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps96d)
• Technical Information TI00507C/07/EN

pH?ISFET sensors
Tophit CPS471/CPS471D
• Sterilizable and autoclavable ISFET sensor for food and pharmaceuticals, process engineering,
• Water treatment and biotechnology;
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps471 or
www.products.endress.com/cps471d)
• Technical Information TI00283C/07/EN
Tophit CPS441/CPS441D
• Sterilizable ISFET sensor for media with low conductivity, with
• liquid KCl electrolyte;
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps441 or
www.products.endress.com/cps441d)
• Technical Information TI00352C/07/EN
Tophit CPS491/CPS491D
• ISFET sensor with open aperture diaphragm for media with high contamination potential;
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps491 or
www.products.endress.com/cps491d)
• Technical Information TI00377C/07/EN

Enamel pH electrode
Ceramax CPS341D
• pH electrode with pH-sensitive enamel
• For the toughest requirements in terms of measurement accuracy, pressure, temperature, sterility
and operating life
• Order as per product structure (--> Online Configurator, www.products.endress.com/cps341d)
• Technical Information TI00468C/07/EN

Inductive conductivity sensors


Indumax P CLS50/CLS50D
• Highly-resistant inductive conductivity sensor for standard, Ex and high-temperature applications
• With or without Memosens technology
• Order as per product structure (--> Online Configurator, www.products.endress.com/cls50d or .../
cls50)
• Technical Information TI00182C/07/EN
Indumax H CLS52
• Inductive conductivity sensor with fast-acting temperature sensor for applications in the food
industry
• Order as product structure (--> Online Configurator, www.products.endress.com/cls52)
• Technical Information TI00167C/07/EN
Indumax H CLS54D
• Inductive conductivity sensor with certified, hygienic design for foodstuffs, beverages,
pharmaceuticals and biotechnology
• Order as per product structure (--> Online Configurator, www.products.endress.com/cls54d)
• Technical Information TI00508C/07/EN
Indumax H CLS54
• Inductive conductivity sensor with certified, hygienic design for foodstuffs, beverages,
pharmaceuticals and biotechnology
• Order as product structure (--> Online Configurator, www.products.endress.com/cls54)
• Technical Information TI00400C/07/EN

38 Endress+Hauser
Liquiline M CM42

Conductive conductivity sensors


Condumax W CLS12
• For process temperatures up to 160 °C (320 °F) and process pressures up to 40 bar (580 psi)
• Order as product structure (--> Online Configurator, www.products.endress.com/cls12)
• Technical Information TI00082C/07/EN
Condumax W CLS13
• For process temperatures up to 250 °C (480 °F) and process pressures up to 40 bar (580 psi)
• Order as product structure (--> Online Configurator, www.products.endress.com/cls13)
• Technical Information TI00083C/07/EN
Condumax CLS15/CLS15D
• Conductive conductivity sensor for pure and ultrapure water and Ex applications
• With Memosens protocol (CLS15D) or analog (CLS15)
• Order as per product structure (--> Online Configurator, www.products.endress.com/cls15d)
• Technical Information TI00109C/07/EN
Condumax H CLS16/CLS16D
• Hygienic, conductive conductivity sensor for pure, ultrapure water and Ex applications
• With EHEDG and 3A approval
• Optionally available with Memosens (CLS16D)
• Order as per product structure (--> Online Configurator, www.products.endress.com/cls16d)
• Technical Information TI00227C/07/EN
Condumax W CLS19
• Inexpensive, conductive conductivity sensor for pure and ultrapure water applications;
• Order as per product structure (--> Online Configurator, www.products.endress.com/cls19)
• Technical Information TI00110C/07/EN
Condumax W CLS21/CLS21D
• Two-electrode sensor in version with plug-in head and fixed cable
• Optionally available with Memosens (CLS21D)
• Order as per product structure (--> Online Configurator, www.products.endress.com/cls21d)
• Technical Information TI00085C/07/EN

Amperometric oxygen sensors


Oxymax COS22D
• Sterilizable sensor for dissolved oxygen
• Order as per product structure (--> Online Configurator, www.products.endress.com/cos22d)
• Technical Information TI00446C/07/EN
Oxymax COS51D
• Amperometric sensor for dissolved oxygen, with Memosens technology
• Order as per product structure (--> Online Configurator, www.products.endress.com/cos51d)
• Technical Information TI00413C/07/EN

Software update and upgrade CY42 DAT module


• Function upgrade, update and memory module
• Ordering as per order structure

Version
S1 SystemDAT, for software update,
changeover to other parameters, language catalog extension
F1 FunctionDAT for extending the function to 2 current outputs
F2 FunctionDAT for extending the function to advanced software
C1 CopyDAT, for saving the configuration and transferring it to other devices

CY42- Complete order code

Endress+Hauser 39
www.addresses.endress.com
Technical Information

Orbisint CPS11 and CPS11D


pH electrodes, analog or with digital Memosens technology
For standard applications in process and environment technology, with
dirt-repellent PTFE diaphragm, optional built-in temperature sensor

Application
• Long-term monitoring and limit monitoring of processes with
stable process conditions
– Pulp and paper industry
– Plastics chemistry
– Chemical processes
– Power plants (e.g. flue gas washers)
– Incineration plants
– Food industry
– Breweries
• Water treatment
– Drinking water
– Cooling water
– Well water
With ATEX, FM and CSA approval for application in hazardous
areas

Your benefits
• Robust electrode requiring low maintenance thanks to large
PTFE ring diaphragm
• Application under pressures of up to 16 bar (232 psi)
• Process glass for highly alkaline applications available
• Built-in Pt 100 or Pt 1000 temperature sensor for effective
temperature compensation (optional)
• Certified biocompatibility
• Sterilizable
• Long service life thanks to double junction system of metal
lead and thus long electrode poison diffusion path
• Poison-resistant reference with ion trap (optional, CPS11D
only)
Further benefits offered by Memosens technology
• Maximum process safety through contactless inductive signal
transmission
• Data safety through digital data transmission
• Easy handling thanks to storage of sensor-specific data in the
sensor
• Predictive maintenance possible thanks to registration of
sensor load data in the sensor

TI028C/07/en/02.08
71068748
Orbisint CPS11 / CPS11D

Function and system design


Measuring principle pH measurement
The pH value is used as a unit of measurement for the acidity or alkalinity of a liquid medium. The membrane
glass of the electrode supplies an electrochemical potential which is dependent upon the pH value of the
medium. This potential is generated by the selective penetration of H+ ions through the outer layer of the
membrane. An electrochemical boundary layer with an electric potential forms at this point. An integrated
Ag/AgCl reference system serves as reference electrode.
The transmitter converts the measured voltage into the corresponding pH value using the Nernst equation.

General properties • Low maintenance


The sterilizable dirt-repellent PTFE ring diaphragm of the electrode prevents blocking and assures long-time
stability and accuracy.
• Long service life
The double junction system of the metal lead offers better protection from electrode poisons and guarantees
a considerably longer service life.
• Durability
Depending on the ordered version, the electrode is pressure proof up to 16 bar (232 psi) and can be applied
at temperatures of up to 135 °C (275 °F.)

Important properties of Maximum process safety


CPS11D The inductive and non-contacting measured value transmission of Memosens guarantees maximum process
safety and offers the following benefits:
• All problems caused by moisture are eliminated.
– The plug-in connection is free from corrosion.
– Measured value distortion from moisture is not possible.
– The plug-in system can even be connected under water.
• The transmitter is galvanically decoupled from the medium. The result: No more need to ask about
"symmetrically high-impedance" or "unsymmetrical" (for pH/ORP measurement) or an impedance
converter.
• EMC safety is guaranteed by screening measures for the digital measured value transmission.
• Application in explosion-hazardous areas is unproblematic; the integrated electronics are intrinsically safe.
Data safety through digital data transfer
The Memosens technology digitalizes the measured values in the sensor and transfers them to the transmitter
contactlessly and free from interference potential. The result:
• An automatic error message is generated if the sensor fails or the connection between sensor and transmitter
is interrupted.
• The availability of the measuring point is dramatically increased by immediate error detection.
Easy handling
Sensors with Memosens technology have integrated electronics that allow for saving calibration data and
further information such as total hours of operation and operating hours under extreme measuring conditions.
When the sensor is mounted, the calibration data are automatically transferred to the transmitter and used to
calculate the current measured value. Storing the calibration data in the sensor allows for calibration away from
the measuring point. The result:
• Sensors can be calibrated unter optimum external conditions in the measuring lab. Wind and weather do
neither affect the calibration quality nor the operator.
• The measuring point availability is dramatically increased by the quick and easy replacement of precalibrated
sensors.
• The transmitter does not need to be installed close to the measuring point but can be placed in the control
room.
• Maintenance intervals can be defined based on all stored sensor load and calibration data and predictive
maintenance is possible.
• The sensor history can be documented on external data carriers and evaluation programs at any time. Thus,
the current application of the sensors can be made to depend on their previous history.
Communication with the transmitter
Always connect digital sensors to a transmitter with Memosens technology. Data transmission to a transmitter
for analog sensors is not possible.

2 Endress+Hauser
Orbisint CPS11 / CPS11D

Data storage of CPS11D Digital sensors are able to store the following system data in the sensor.
• Manufacturing data
– Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Calibrated slope at 25 °C (77 °F)
– Calibrated zero point at 25 °C (77 °F)
– Temperature offset
– Number of calibrations
– Serial number of the transmitter used for the last calibration
• Application data
– Temperature application range
– pH application range
– Date of first commissioning
– Maximum temperature value
– Operating hours at temperatures above 80 °C (176 °F) and 100 °C (212 °F)
– Operating hours at very low and very high pH values (Nernst voltage below -300 mV, above +300 mV)
– Number of sterilizations
– Glass membrane impedance
These system data can be displayed with Mycom S and Liquiline M transmitters.

Measuring system A complete measuring system comprises:


• CPS11 pH electrode or CPS11D digital sensor
• Transmitter, e.g. Liquiline M CM42 (with Memosens technology for CPS11D)
• Special measuring cable, e.g. CPK9 or Memosens data cable CYK10
• Immersion, flow or retractable assembly, e.g. Cleanfit P CPA472

1 4

2
3

a0003706

Measuring system for pH measurement


1 Cleanfit P CPA472 retractable assembly
2 CPS11/CPS11D pH electrode
3 Special measuring cable CPK9 for electrodes with TOP68 plug-in head / CYK10 for digital sensors
4 Liquiline M CM42 transmitter

Endress+Hauser 3
Orbisint CPS11 / CPS11D

Input
Measured variables pH value
Temperature

Measuring ranges Electrode version AA, AS (for water / wastewater):


pH: 1 to 12 pH
Temperature: -15 to 80 °C (5 to 176 °F)
Electrode version BA (for process applications, sterilizable)
pH: 0 to 14 pH
Temperature: 0 to 135 °C (32 to 275 °F)
Electrode version FA (for hydrofluoric acid):
pH: 0 to 10 pH
Temperature: 0 to 70 °C (32 to 158 °F)
Electrode version BT with ion trap (for pulp, paper, chemical applications)
pH: 0 to 14
Temperature: 0 to 135 °C (32 to 275 °F)

" Caution!
Please note the process operating conditions.

Installation
Installation instructions Do not install the electrode upside down. The inclination angle must be at least 15° from the horizontal. A
smaller inclination angle is not permitted as such an inclination results in air cushion forming in the glass
sphere. This might impair full wetting of the pH membrane with inner electrolyte.

" Caution!
• Make sure that the assembly’s threaded connection for the electrode is clean and well running before
installing the electrode.
• Hand tighten the electrode (3 Nm)! (Given value only applies to installation in Endress+Hauser assemblies.)
• Make sure to follow the installation instructions in the operating instructions of the used assembly.

A
15° 15°

a0003133

Electrode installation; inclination angle min. 15° from the horizontal


A Permitted inclination angle
B Non-permitted inclination angle

4 Endress+Hauser
Orbisint CPS11 / CPS11D

Environment
Ambient temperature

" Caution!
Danger of frost damage
Do not use the electrode at temperatures below –15 °C (5 °F).

Storage temperature 0 to 50 °C (32 to 122 °F)

Ingress protection IP 67: GSA plug-in head (with closed plug-in connection)
IP 68: TOP68 plug-in head, autoclavable up to 135 °C (275 °F), 1 m (3.28 ft) water column,
50 °C (122 °F), 168 h
IP 68: Memosens plug-in head, 10 m (32.81 ft) water column, 25 °C (77 °F), 45 days, 1M KCl

Process
Process temperature Version AA, AS: -15 to 80 °C (5 to 176 °F)
Version BA, BT: 0 to 135 °C (32 to 275 °F)
Version FA: 0 to 70 °F (32 to 158 °F)

Process pressure 0 to 6 bar (0 to 87 psi) / 16 bar (232 psi) (versions CPS11-xBAxESA, CPS11D-7BAxx, CPS11D-7BTxx)

Pressure temperature load [psi] p [bar]


curve
18
240 A
232 16
200 14
12
160
10
120 8
87
B
80 6
4
40
2
0 0 T [°C]
-15 0 20 40 60 80 100 120 135
T [°F]
275
5 50 100 150 200 250 300
a0009229

Pressure temperature load curve


A Versions CPS11-xBAxESA, CPS11D-7BAxx, CPS11D-7BTxx
B CPS11 and CPS11D (except CPS11-xBAxESA, CPS11D-7BAxx, CPS11D-7BTxx)

Conductivity min. 50 μS/cm

pH range Version AA, AS: 1 to 12 pH


Version BA, BT: 0 to 14 pH
Version FA: 0 to 10 pH

" Caution!
Danger of electrode damage
Do not operate the electrodes in applications outside the given specifications!

Endress+Hauser 5
Orbisint CPS11 / CPS11D

Mechanical construction
Design, dimensions
mm (inch)

mm (inch)
1
1
2
2

(4.72, 8.86, 14.17, 16.73)


3

(4.72, 8.86, 14.17, 16.73)

120, 225, 360, 425


120, 225, 360, 425
4
4

5
5

6
6

7
7

12 / 12 / 8
0.47 0.47
a0009222 a0009223

CPS11 with GSA plug-in head CPS11 with TOP68, built-in temperature sensor
1 GSA plug-in head, Pg 13.5 1 TOP68 plug-in head, Pg 13.5
2 Viton O-ring with thrust collar 2 Viton O-ring with thrust collar
3 Ag/AgCl metal lead - reference 3 Ag/AgCl metal lead - reference
4 "Advanced Gel" electrolyte 4 "Advanced Gel" electrolyte
5 Ag/AgCl metal lead - pH 5 Ag/AgCl metal lead - pH
6 PTFE diaphragm 6 PTFE diaphragm
7 pH glass membrane 7 pH glass membrane
8 Pt 100 temperature sensor

Design, dimensions CPS11D


mm (inch) mm (inch)
50 (1.97)

50 (1.97)

1
1
2
2

3
3
(4.72, 8.86, 14.17)

(4.72, 8.86, 14.17)

4 4
120, 225, 360

120, 225, 360

5 5

6
6

7
7
8
12 8 12
(0.47) (0.47) 9
a0009225 a0009224

CPS11D with Memosens, built-in temperature sensor CPS11D-7BTxx


1 Memosens plug-in head, Pg 13.5 1 Memosens plug-in head, Pg 13.5
2 Viton O-ring with thrust collar 2 Viton O-ring with thrust collar
3 Ag/AgCl metal lead - reference 3 Ag/AgCl metal lead - reference
4 "Advanced Gel" electrolyte 4 "Advanced Gel" electrolyte
5 Ag/AgCl metal lead - pH 5 Ion trap
6 PTFE diaphragm 6 Ag/AgCl metal lead - pH
7 pH glass membrane 7 PTFE diaphragm
8 Temperature sensor 8 pH glass membrane
9 Temperature sensor

6 Endress+Hauser
Orbisint CPS11 / CPS11D

Weight approx. 0.1 kg (0.22 lbs)

Material Electrode shaft process glass


pH membrane glasses types A, B, F
Metal lead Ag/AgCl
Diaphragm ring-shaped Teflon® diaphragm, sterilizable

Process connection Pg 13.5

Temperature sensor CPS11: Pt 100, Pt 1000


CPS11D: NTC

Plug-in heads CPS11:


ESA plug-in head Pg 13.5, TOP68 for electrodes with or without temperature sensor,
16 bar (232 psi) triple safety overpressure, Ex
GSA plug-in head Pg 13.5 for electrodes without temperature sensor
CPS11D: Memosens plug-in head for digital, contactless data transmission

Reference system CPS11: Ag/AgCl metal lead with Advanced Gel 3 M KCl, AgCl free
CPS11D:
Versions AA, AS, BA, FA Ag/AgCl metal lead with Advanced Gel 3 M KCl, AgCl free
Version BT Ag/AgCl metal lead with ion trap and Advanced Gel 3 M KCl

Certificates and approvals


Ex approval CPS11 (TOP68)
• ATEX II 1G EEX ia IIC T3/T4/T6
• FM Class I Div. 2, in combination with the Liquiline M CM42 and Mycom S CPM153 transmitters

CPS11D
• ATEX II 1G EEX ia IIC T3/T4/T6
• FM / CSA Class I Div. 2, in combination with the Liquiline M CM42 and Mycom S CPM153 transmitters

! Note!
Ex versions of digital sensors with Memosens technology are indicated by an orange-red ring in the plug-in
head.

Biocompatibility Biocompatibility validated according to:


• ISO 10993-5:1993
• USP, current revision

TÜV certificate TOP68 plug-in head


Pressure resistance 16 bar (232 psi), min. triple overpressure safety

Memosens plug-in head


Pressure resistance 16 bar (232 psi), min. triple overpressure safety

Electromagnetic compatibility Interference emission and interference immunity complies with EN 61326: 1997 / A1: 1998
of CPS11D

Endress+Hauser 7
Orbisint CPS11 / CPS11D

Ordering information
Product structure CPS11 Electrode type
1 without temperature sensor
2 with built-in Pt 100 (not available with GSA plug-in head)
3 with built-in Pt 1000 (not available with GSA plug-in head)

Application range
AA pH = 1 to 12, T = -15 to 80 °C (5 to 176 °F), 6 bar (87 psi)
AS pH = 1 to 12, T = -15 to 80°C (5 to 176 °F), 6 bar (87 psi), salt ring
BA pH = 0 to 14, T = 0 to 135 °C (32 to 275 °F), sterilizable, 16 bar (232 psi) in combination with ESA plug-in head
FA pH = 0 to 10, T = 0 to 70 °C (32 to 158 °F), HF resistant up to 1 g/l, 6 bar (87 psi)

Shaft length
2 120 mm (4.72")
4 225 mm (8.86")
5 360 mm (14.17")
6 425 mm (16.73")

Plug-in head
ESA Plug-in head Pg 13.5, TOP68, 16 bar (232 psi) in combination with BA application range, Ex
GSA Plug-in head Pg 13.5, DIN coax, non-Ex

CPS11- complete order code

Product structure CPS11D Version


7 max. 135 °C (275 °F), built-in temperature sensor

Application range
AA pH = 1 to 12, T = -15 to 80 °C (5 to 176 °F), 6 bar (87 psi)
AS pH = 1 to 12, T = -15 to 80 °C (5 to 176 °F), 6 bar (87 psi), salt ring
BA pH = 0 to 14, T = 0 to 135 °C (32 to 275 °F), 16 bar (232 psi), sterilizable
BT pH = 0 to 14, T = 0 to 135 °C (32 to 275 °F), 16 bar (232 psi), ion trap
FA pH = 0 to 10, T = 0 to 70 °C (32 to 158 °F), 6 bar (87 psi), HF resistant up to 1 g/l

Shaft length
2 120 mm (4.72")
4 225 mm (8.86")
5 360 mm (14.17")
6 425 mm (16.73")

Approval
1 Non-hazardous area
G ATEX II 1G EEx ia IIC T3/T4/T6

CPS11D- complete order code

Accessories
! Note!
In the following sections, you find the accessories available at the time of issue of this documentation.
For information on accessories that are not listed here, please contact your responsible service.

Transmitters Liquiline M CM42


• Modular two-wire transmitter, stainless steel or plastic, field or panel instrument,
• various Ex approvals (ATEX, FM, CSA, Nepsi, TIIS),
• HART, PROFIBUS or FOUNDATION Fieldbus available
• Ordering acc. to product structure, see Technical Information (TI381C/07/en)
Liquisys M CPM223/253
• Transmitter for pH and ORP, field or panel-mounted housing,
• HART or PROFIBUS available
• Ordering acc. to product structure, see Technical Information (TI194C/07/en)
Mycom S CPM153
• Transmitter for pH and ORP, one or two channel version, Ex or non-Ex,
• HART or PROFIBUS available
• Ordering acc. to product structure, see Technical Information (TI233C/07/en)

8 Endress+Hauser
Orbisint CPS11 / CPS11D

Assemblies (Selection) • Cleanfit W CPA450


Manually operated, retractable assembly for installation of 120 mm (4.72") pH/ORP electrodes in tanks and
pipes,
Ordering acc. to product structure, see Technical Information (TI183C/07/en)
• Cleanfit P CPA471
Compact retractable stainless steel assembly for installation in tanks and pipes, manual or pneumatic
operation
Ordering acc. to product structure, see Technical Information (TI217C/07/en)
• Cleanfit P CPA472
Compact retractable plastic assembly for installation in tanks and pipes, manual or pneumatic operation,
Ordering acc. to product structure, see Technical Information (TI223C/07/en)
• Cleanfit P CPA472D
Robust retractable assembly for pH, ORP and other industry sensors, manual or pneumatic operation,
heavy-duty version;
Ordering acc. to product structure, see Technical Information (TI403C/07/en)

a0003135 a0003137 a0009269

Cleanfit W CPA450 Cleanfit P CPA471 or 472 Cleanfit P CPA472D

Endress+Hauser 9
Orbisint CPS11 / CPS11D

• Cleanfit P CPA473
Retractable stainless steel process assembly, with ball valve for a particularly safe and reliable separation of
the medium from the environment,
Ordering acc. to product structure, see Technical Information (TI344C/07/en)
• Cleanfit P CPA474
Retractable plastic process assembly, with ball valve for a particularly safe and reliable separation of the
medium from the environment,
Ordering acc. to product structure, see Technical Information (TI345C/07/en)
• Cleanfit H CPA475
Retractable assembly for installation in tanks and pipes under sterile conditions,
Ordering acc. to product structure, see Technical Information (TI240C/07/en)
• Unifit H CPA442
Process assembly for the food industry, biotechnology and pharmaceutical industry,
Ordering acc. to product structure, see Technical Information (TI306C/07/en)

a0003138 a0003136 a0003139


Cleanfit P CPA473 or 474 Cleanfit H CPA475 Unifit H CPA442
• Dipfit W CPA111
Plastic immersion and installation assembly for open and closed tanks,
Ordering acc. to product structure, see Technical Information (TI112C/07/en)
• Dipfit P CPA140
Immersion assembly for pH/ORP electrodes for demanding processes,
Ordering acc. to product structure, see Technical Information (TI178C/07/en)
• Flowfit P CPA240
Flow assembly for pH/ORP electrodes, for demanding processes,
Ordering acc. to product structure, see Technical Information (TI179C/07/en)

a0003140 a0003141 a0003142


Dipfit W CPA111 Dipfit P CPA140 Flowfit P CPA240

10 Endress+Hauser
Orbisint CPS11 / CPS11D

• Flowfit W CPA250
Flow assembly for pH/ORP electrodes,
Ordering acc. to product structure, see Technical Information (TI041C/07/en)
• Probfit H CPA465
Retractable assembly for installation in tanks and pipes under sterile conditions,
Ordering acc. to product structure, see Technical Information (TI146C/07/en)
• Ecofit CPA640
Process connection adapter and cable set for 120 mm (4.72") pH/ORP electrodes,
Ordering acc. to product structure, see Technical Information (TI264C/07/en)

a0003143 a0003144 a0003145

Flowfit W CPA250 Probfit H CPA465 Ecofit CPA640

Buffer solutions High-quality buffer solutions of Endress+Hauser


The secondary buffer solutions have been referenced to primary reference material of the PTB (German Federal
Physico-technical Institute) and to standard reference material of NIST (National Institute of Standards and
Technology) according to DIN 19266 by a DKD (German Calibration Service) accredited laboratory.

pH value
A pH 2.00 (accuracy ± 0.02 pH)
C pH 4.00 (accuracy ± 0.02 pH)
E pH 7.00 (accuracy ± 0.02 pH)
G pH 9.00 (accuracy ± 0.02 pH)
I pH 9.20 (accuracy ± 0.02 pH)
K pH 10.00 (accuracy ± 0.05 pH)
M pH 12.00 (accuracy ± 0.05 pH)

Quantity
01 20 x 18 ml (0.68 fl.oz) only buffer solutions pH 4.00 and 7.00
02 250 ml (8.45 fl.oz)
10 1000 ml (0.26 US gal)
50 5000 ml (1.32 US gal) canister for Topcal S

Certificates
A Buffer analysis certificate

Version
1 Standard

CPY20- complete order code

Endress+Hauser 11
Measuring cables CPK9 special measuring cable
• For sensors with TOP68 plug-in head, for high-temperature and high-pressure applications, IP 68
• Ordering acc. to product structure, see Technical Information (TI118C/07/en)
CPK1 special measuring cable
• For pH/ORP electrodes with GSA plug-in head
• Ordering acc. to product structure, see Technical Information (TI118C/07/en)
CPK12 special measuring cable
• For pH/ORP glass electrodes and ISFET sensors with TOP68 plug-in head
• Ordering acc. to product structure, see Technical Information (TI118C/07/en)
CYK10 Memosens data cable
• For digital sensors with Memosens technology
• Ordering according to product structure, see below

Certificates
A Standard, non Ex
G ATEX II 1G EEx ia IIC T6/T4

Cable length
03 Cable length: 3 m (9.8 ft)
05 Cable length: 5 m (16 ft)
10 Cable length: 10 m (33 ft)
15 Cable length: 15 m (49 ft)
20 Cable length: 20 m (66 ft)
25 Cable length: 25 m (82 ft)
88 ... m length
89 ... ft length

Ready-made
1 Wire terminals

CYK10- complete order code

! Note!
Ex versions of CYK10 are indicated by an orange-red coupling end.

Instruments International

Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland

Tel.+41 61 715 81 00
Fax+41 61 715 25 00
www.endress.com
info@ii.endress.com

TI028C/07/en/02.08
71068748
Printed in Germany / FM+SGML 6.0 / DT
Technical Information

Orbisint CPS12/CPS12D/CPS13
ORP electrodes, analog and digital with Memosens technology and
reference electrode
Each with dirt-repellent PTFE diaphragm for standard applications in
process and environmental technology

Application
• Long-term monitoring and limit monitoring of processes with
stable process conditions
– Paper industry
– Plastics chemistry
– Power plants (e.g. flue gas washers)
– Incineration plants
– Food industry
– Breweries
• Water treatment
– Drinking water
– Cooling water
– Well water

0 With ATEX, FM1 and CSA1 approval


for application in hazardous areas

Your benefits
• Robust electrode requiring low maintenance thanks to large
PTFE ring diaphragm
• Application under pressures of up to 16 bar (232 psi)
• Long service life due to double junction system of metal lead
and thus long electrode poison diffusion path
• Various measuring elements for application in oxidising and
reducing media
• Three lengths: 120, 225 and 360 mm (4.72, 8.86,
14.17 inches)
Further benefits offered by Memosens technology
• Maximum process safety through contactless inductive signal
transmission
• Data safety through digital data transmission
• Easy handling due to storage of sensor-specific data
• Predictive maintenance possible thanks to registration of
sensor load data

1
approval for digital sensors pending

TI367C/07/en/05.05
51513586
Orbisint CPS12 / CPS12D / CPS13

Function and system design


Measuring principle Redox measurement
The redox potential is a unit of measurement for the state of equilibria between oxidising and reducing
components of a medium. Redox potential is measured similarly to the pH value. A platinum or gold electrode
is used instead of pH-sensitive membrane glass. Analog to the pH measurement, an integrated Ag/AgCl
reference system is used as a reference electrode.

General properties • Low maintenance


The dirt-repellent, sterilisable PTFE ring diaphragm of the CPS12 prevents blocking and assures long-time
stability and accuracy.
• Long service life
The double junction system of the metal lead offers better protection from electrode poisons and guarantees
a considerably longer service life.
• Durability
Depending on the ordered version, the electrode is pressure proof up to 16 bar / 232 psi and can be applied
at temperatures of up to 135 °C / 275 °F.

Important properties CPS12D Maximum process safety


The inductive and non-contacting measured value transmission of Memosens guarantees maximum process
safety and offers the following benefits:
• All problems caused by moisture are eliminated.
– The plug-in connection is free from corrosion.
– Measured value distortion from moisture is not possible.
– The plug-in system can even be connected under water.
• The transmitter is galvanically decoupled from the medium. The result: No more need to ask about
"symmetrically high-impedance" or "unsymmetrical" or an impedance converter.
• EMC safety is guaranteed by screening measures for the digital measured value transmission.
Data safety through digital data transfer
The Memosens technology digitalizes the measured value in the sensor and transfers it to the transmitter via a
contactless connection. The result:
• An automatic error message is generated if the sensor fails or the connection between sensor and transmitter
is interrupted.
• The availability of the measuring point is dramatically increased by immediate error detection.
• The digital signals are suitable for application in hazardous areas; the integrated electronics are intrinsically
safe.
Easy handling
Sensors with Memosens technology have integrated electronics that allow for saving calibration data and
further information such as total hours of operation and operating hours at very high temperatures. When the
sensor is mounted, the calibration data are automatically transferred to the transmitter and used to calculate
the current redox potential. Storing the calibration data in the sensor allows for calibration and adjustment
away from the measuring point. The result:
• The sensors can be calibrated unter optimum external conditions in the measuring lab. Wind and weather
do neither affect the calibration quality nor the operator.
• The measuring point availability is dramatically increased by the quick and easy replacement of precalibrated
sensors.
• The transmitter does not need to be installed close to the measuring point but can be placed in the control
room.
• Maintenance intervals can be defined based on all stored sensor load and calibration data and predictive
maintenance is possible.
• The sensor history can be documented on external data carriers and evaluation programs at any time. Thus,
the current application of the sensors can be made to depend on their previous history.
Communication with the transmitter
Always connect the digital sensor to a digital transmitter with Memosens technology. Data transmission to an
analog transmitter is not possible.

2 Endress+Hauser
Orbisint CPS12 / CPS12D / CPS13

Data storage of CPS12D Digital sensors are able to store the following system data in the sensor.
• Manufacturing data
– Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Calibrated offset (operating mode "mV")
– % slope (operating mode "%")
– Number of calibrations
– Serial number of the transmitter used for the last calibration
• Application data
– Temperature application range
– Redox application range
– Date of first commissioning
– Operating hours
These system data can be displayed with the Mycom S or the Liquiline M CM42 transmitter.

Measuring system A complete measuring system comprises:


• CPS12 or CPS12D ORP electrode
• transmitter, e.g. Liquisys M CPM223/253 (with Memosens technology for CPS12D)
• special measuring cable, e.g. CPK9 or Memosens data cable CYK10 for CPS12D
• immersion, flow or retractable assembly, e.g. Cleanfit P CPA472

a0003690

Measuring system for measurement of redox potential


1 Retractable Cleanfit P CPA472 assembly
2 CPS12 / CPS12D ORP electrode
3 CPK9 special measuring cable (for electrodes with TOP68 plug-in head) / CYK10 for digital sensors
4 Liquisys M CPM253 transmitter

Endress+Hauser 3
Orbisint CPS12 / CPS12D / CPS13

Input
Measured variables Redox potential

Measuring range -1500 ... 1500 mV

" Caution!
Please note the process operating conditions.

Installation
Installation instructions Do not install the electrode upside down. The inclination angle must be at least 15° from the horizontal. A
smaller inclination angle is not permitted as such an inclination results in air cushion forming. This might impair
the contact of reference and metal lead.

" Caution!
• Make sure that the assembly’s threaded connection for the electrode is clean and well running before
installing the electrode.
• Hand tighten the electrode (3 Nm)! (Given value only applies to installation in Endress+Hauser assemblies.)
• Make sure to follow the installation instructions in the operating instructions of the used assembly.

A
15° 15°

a0003686

Electrode installation; inclination angle min. 15° from the horizontal


A Permitted inclination angle
B Non-permitted inclination angle

Environment
Ambient temperature

" Caution!
Danger of frost damage
Do not use the electrode at temperatures below -15 °C / 5 °F.

Storage temperature 0 ... 50 °C / 32 ... 122 °F

Ingress protection IP 67: GSA plug-in head (with closed plug-in connection)
IP 68: TOP 68 plug-in head (1 m / 3.28 ft water column, 50 °C / 122 °F, 168 h)
IP 68: Memosens plug-in head (10 m / 32.8 ft water column, 25 °C /77 °F, 45 days, 1 M KCl)

4 Endress+Hauser
Orbisint CPS12 / CPS12D / CPS13

Process
Process temperature -15 ... 135 °C / 5 ... 275 °F

Process pressure 0 ... 16 bar / 0 ... 232 psi

Application CPS12, • Gold electrode


CPS12D: for oxidising media, e.g. cyanide oxidation, nitrite oxidation, ozone measurement, hydrogen
superoxide measurement
• Platinum electrode
for reducing media, e.g. chromate reduction, chlorine dosing in swimming pools
CPS13: Single reference electrode, used in combination with the single pH electrode CPS64

" Caution!
Danger of damage to the electrode
Never use the electrode in applications outside the given specifications!

Mechanical construction
Design, dimensions CPS12
mm / inch
mm / inch

52.3 / 2.06
30.2 / 1.19

1 1
2
2

3 3
120, 225, 360 / 4.72, 8.86, 14.17

120, 225, 360 / 4.72, 8.86, 14.17


4
4

5
5
6
6
7
12 / 0.47 7
12 / 0.47
a0003689-en a0003688-en

CPS12 with GSA plug-in head CPS12 with TOP68 plug-in head
1 GSA plug-in head, PG 13.5 1 TOP68 plug-in head, Pg 13.5
2 Viton O-ring with thrust collar 2 Viton O-ring with thrust collar
3 Inner metal lead 3 Inner metal lead
4 Ag/AgCl lead for the reference 4 Ag/AgCl lead for the reference
5 "Advanced Gel" electrolyte 5 "Advanced Gel" electrolyte
6 PTFE diaphragm 6 PTFE diaphragm
7 Gold pin or platinum ring 7 Gold pin or platinum ring

Endress+Hauser 5
Orbisint CPS12 / CPS12D / CPS13

Design, dimensions CPS12D


mm / inch

50 / 1.97
1

4.72, 8.86, 14.17


120, 225, 360 /

8
12 / 0.47
a0003696-en

CPS12D with Memosens plug-in head


1 Memosens plug-in head
2 Viton O-ring with thrust collar
3 Inner metal lead
4 Ag/AgCl lead for the reference
5 "Advanced Gel" electrolyte
6 NTC 30K temperature sensor
7 PTFE diaphragm
8 Gold pin or platinum ring

Design, dimensions CPS13


mm / inch
mm / inch
52.3 / 2.06
30.2 / 1.19

1
1
2
2

3
(80 / 3.15 for CPA320)

4
3
(80 / 3.15 for CPA320)
120 / 4.72

120 / 4.72

5
5
12 / 0.47
12 / 0.47
a0004256-en a0004255-en

CPS13 with GSA plug-in head CPS13 with TOP68 plug-in head
1 GSA plug-in head, Pg 13.5 1 TOP68 plug-in head, Pg 13.5
2 Viton O-ring with thrust collar 2 Viton O-ring with thrust collar
3 Ag/AgCl reference lead 3 Ag/AgCl reference lead
4 "Advanced Gel" electrolyte 4 "Advanced Gel" electrolyte
5 PTFE diaphragm 5 PTFE diaphragm

6 Endress+Hauser
Orbisint CPS12 / CPS12D / CPS13

Weight 0.1 kg / 0.2 lb.

Material Electrode shaft glass, suitable for processes


Redox measuring element platinum ring or gold pin
Diaphragm ring-shaped Teflon® diaphragm, sterilisable

Process connection Pg 13.5

Plug-in heads CPS12, CPS13:


ESA: ESA plug-in head Pg 13.5, TOP68, 16 bar / 232 psi, Ex
GSA: GSA plug-in head Pg 13.5
CPS12D: Memosens plug-in head for digital, contactless data transmission,
16 bar / 232 psi, Ex or non-Ex

Reference system Ag/AgCl lead with Advanced Gel 3 M KCl, AgCl free

Certificates and approvals


Ex approval CPS12 (ESA) / • ATEX II 1G EEX ia IIC T3/T4/T6
CPS13 (ESA) • FM Class I Div. 2, in combination with the Liquiline M CM42 and Mycom S CPM153 transmitters

Ex approval CPS12D • ATEX II 1G EEX ia IIC T3/T4/T6


• FMa / CSAa Class I Div. 2, in combination with the Liquiline M CM42 and Mycom S CPM153 transmitters

! Note!
Ex versions of digital sensors with Memosens technology are indicated by an orange-red ring in the plug-in
head.

Biocompatibility Biocompatibility validated according to:


• ISO 10993-5:1993
• USP, current revision

TÜV certificate TOP68 and Pressure resistance 16 bar, min. triple overpressure safety
Memosens plug-in head

Electromagnetic compatibility Interference emission and interference immunity complies with EN 61326: 1997 / A1: 1998
of CPS12D

a) approval pending

Endress+Hauser 7
Orbisint CPS12 / CPS12D / CPS13

Ordering information
Product structure CPS12 Electrode type
0 Standard type

Measuring surface
NA Gold pin
PA Platinum ring

Shaft length
2 120 mm / 4.72"
4 225 mm / 8.86"
5 360 mm / 14.17"

Type of head
ESA Plug-in head Pg 13.5, TOP68, 16 bar, Ex
GSA Plug-in head Pg 13.5

CPS12- complete order code

Product structure CPS12D Version


7 Basic version, max. 135 °C / 275 °F

Measuring surface
NA Gold pin
PA Platinum ring

Shaft length
2 120 mm / 4.72"
4 225 mm / 8.86"
5 360 mm / 14.17"

Approval
1 Non-hazardous areas
G ATEX II 1G EEx ia IIC T3/T4/T6

CPS12D- complete order code

Product structure CPS13 Electrode type


0 Standard type

Electrolyte
TA Advanced Gel filling (120 mm / 4.72" only)
TD Advanced Gel filling / double reference (80 mm / 3.15" only)

Shaft length
1 80 mm / 3.15"
2 120 mm / 4.72"

Type of head
ESA Plug-in head Pg 13.5, TOP68, 16 bar / 232 psi, Ex
GSA Plug-in head Pg 13.5

CPS13- complete order code

8 Endress+Hauser
Orbisint CPS12 / CPS12D / CPS13

Accessories
! Note!
In the following sections, you find the accessories available at the time of issue of this documentation.
For information on accessories that are not listed here, please contact your responsible service.

Assemblies • Cleanfit W CPA450


Manually operated, retractable assembly for installation of 120 mm / 4.72" pH/ORP electrodes in tanks and
pipes,
Ordering acc. to product structure, see Technical Information (TI183C/07/en)
• Cleanfit P CPA471
Compact retractable stainless steel assembly for installation in tanks and pipes, manual or pneumatic
operation
Ordering acc. to product structure, see Technical Information (TI217C/07/en)
• Cleanfit P CPA472
Compact retractable plastic assembly for installation in tanks and pipes, manual or pneumatic operation,
Ordering acc. to product structure, see Technical Information (TI223C/07/en)
• Cleanfit P CPA473
Retractable stainless steel process assembly, with ball valve for a particularly safe and reliable separation of
the medium from the environment,
Ordering acc. to product structure, see Technical Information (TI344C/07/en)
• Cleanfit P CPA474
Retractable plastic process assembly, with ball valve for a particularly safe and reliable separation of the
medium from the environment,
Ordering acc. to product structure, see Technical Information (TI345C/07/en)

a0003135 a0003137 a0003138

Cleanfit W CPA450 Cleanfit P CPA471 or 472 Cleanfit P CPA473 or 474

Endress+Hauser 9
Orbisint CPS12 / CPS12D / CPS13

• Cleanfit H CPA475
Retractable assembly for installation in tanks and pipes under sterile conditions,
Ordering acc. to product structure, see Technical Information (TI240C/07/en)
• Unifit H CPA442
Process assembly for the food industry, biotechnology and pharmaceutical industry, with EHEDG and 3A
certificates,
Ordering acc. to product structure, see Technical Information (TI306C/07/en)
• Dipfit W CPA111
Plastic immersion and installation assembly for open and closed tanks,
Ordering acc. to product structure, see Technical Information (TI112C/07/en)

a0003139 a0003140
a0003136

Cleanfit H CPA475 Unifit H CPA442 Dipfit W CPA111

• Dipfit P CPA140
Immersion assembly for pH/ORP electrodes for demanding processes,
Ordering acc. to product structure, see Technical Information (TI178C/07/en)
• Flowfit P CPA240
Flow assembly for pH/ORP electrodes, for demanding processes,
Ordering acc. to product structure, see Technical Information (TI179C/07/en)
• Flowfit W CPA250
Flow assembly for pH/ORP electrodes,
Ordering acc. to product structure, see Technical Information (TI041C/07/en)

a0003141 a0003142 a0003143

Dipfit P CPA140 Flowfit P CPA240 Flowfit W CPA250

10 Endress+Hauser
Orbisint CPS12 / CPS12D / CPS13

• Probfit H CPA465
Retractable assembly for installation in tanks and pipes under sterile conditions,
Ordering acc. to product structure, see Technical Information (TI146C/07/en)
• Ecofit CPA640
Process connection adapter and cable set for 120 mm pH/ORP electrodes,
Ordering acc. to product structure, see Technical Information (TI264C/07/en)

a0003144 a0003145

Probfit H CPA465 Ecofit CPA640

Buffer solutions Technical buffer solutions for ORP electrodes


• +220 mV, pH 7.0, 100 ml (0.026 US gal.); order no. CPY3-0
• +468 mV, pH 0.1, 100 ml (0.026 US gal.); order no. CPY3-1

Transmitters • Liquisys M CPM223/253


Transmitter for pH and redox, field or panel-mounted housing,
Hart® or PROFIBUS available
Ordering acc. to product structure, see Technical Information (TI194C/07/en)
• Mycom S CPM153
Transmitter for pH and redox, one or two channel version, Ex or Non-Ex,
Hart® or PROFIBUS available
Ordering acc. to product structure, see Technical Information (TI233C/07/en)
• Liquiline M CM42
Modular two-wire transmitter for Ex and non-Ex areas
Hart®, PROFIBUS or FOUNDATION Fieldbus available
Ordering acc. to product structure, see Technical Information (TI381C/07/en)

Endress+Hauser 11
Measuring cables • CPK9 special measuring cable
For sensors with TOP68 plug-in head, for high-temperature and high-pressure applications, IP 68
Ordering acc. to product structure, see Technical Information (TI118C/07/en)
• CPK1 special measuring cable
For pH/redox electrodes with GSA plug-in head
Ordering acc. to product structure, see Technical Information (TI118C/07/en)
• CYK10 Memosens data cable
For digital sensors with Memosens technology
Ordering according to product structure, see below

Certificates
A Standard, non Ex
G ATEX II 1G EEx ia IIC T6/T4

Cable length
03 Cable length: 3 m / 9.84 ft
05 Cable length: 5 m / 16.41 ft
10 Cable length: 10 m / 32.81 ft
15 Cable length: 15 m / 49.22 ft
20 Cable length: 20 m / 65.62 ft
25 Cable length: 25 m / 82.03 ft
88 ... m length
89 ... ft length

Ready-made
1 Wire terminals

CYK10- complete order code

! Note!
Ex versions of CYK10 are indicated by an orange-red coupling end.

International Head Quarters

Endress+Hauser
GmbH+Co. KG
Instruments International
Colmarer Str. 6
79576 Weil am Rhein
Deutschland

Tel. +49 76 21 9 75 02
Fax +49 76 21 9 75 34 5
www.endress.com
info@ii.endress.com

TI367C/07/de/05.05
51513586
Printed in Germany / FM+SGML 6.0 / DT
Technical Information

Indumax CLS50/CLS50D
Inductive conductivity sensor for standard, hazardous and high-
temperature applications, analog or digital with Memosens protocol

Application
Indumax CLS50/CLS50D conductivity sensors are especially
suitable for application in the chemical industry and process
engineering. The six-decade measuring range and the high
chemical resistance of the materials in contact with medium
(PFA or PEEK) permit to use this sensor in a number of various
applications, e.g.:
• Concentration measurement of acids and bases
• Quality monitoring of chemical products in tanks and pipes
• Phase separation of product/product mixtures
Digital CLS50D sensors are used with Liquiline CM44x or
Liquiline CM42. Analog CLS50 sensors are used with the
transmitters Liquiline CM42, Liquisys CLM223/253 or
Mycom CLM153.

Your benefits
• High durability
– High chemical resistance thanks to PFA coating
– PEEK version for high temperatures up to 180 ˚C (356 ˚F)
• Low risk of soiling
– Dirt-repellent PFA surface
– Large sensor opening
• Easy installation
– Can be installed in T-pieces ≥ DN 80 with the outgoing
diameter reduced to ≥ DN 50
– Total cable length up to 55 m (180 ft)
• Wide measuring range from 2 μS/cm to 2000 mS/cm
• Integrated, coated Pt 100 temperature sensor, error class A
• Ex approval EEx ia IIC T6/T4

TI00182C/07/EN/15.12
71186403
Indumax CLS50/CLS50D

Function and system design


Measuring principle Inductive conductivity measurement
A generator (1) generates an alternating magnetic field in the primary coil (2) which induces a current in the
medium (3). The strength of the induced current depends on the conductivity and thus the ion concentration
of the medium. The current flow in the medium generates another magnetic field in the secondary coil (4).
The resulting current induced in the coil is measured by the receiver (5) and processed to determine the
conductivity.

Inductive conductivity measurement


1 5
1 Generator
2 Primary coil
3 Current flow in the medium
4 Secondary coil
5 Receiver

Benefits of inductive conductivity measurement


• No electrodes, therefore no polarization
2 4 • Accurate measurement in media or solutions with
a high soiling degree and a tendency to deposition
• Complete galvanic separation of measurement and
medium

3
a0004894

Important properties • Wide measuring range


The sensor’s measuring range comprises six decades, from 2 μS/cm to 2000 mS/cm.
• Durability
The materials in contact with medium (PEEK, PFA) feature a very high chemical resistance. In addition, the
PEEK version is suitable for application at high temperatures (CLS50 up to 180 ˚C (356 ˚F) and CLS50D up
to 125 ˚C (260 ˚F)).
• Low risk of soiling
Thanks to its large opening, the sensor is not susceptible to soiling. The PFA version requires even less
cleaning thanks to its dirt-repellent surface.

Advantages of digital sensors Digital sensors are able to store the following system data in the sensor:
with Memosens protocol • Manufacturing data
(CLS50D only) – Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Cell constant
– Change in cell constant
– Number of calibrations
– Serial number of the transmitter used for the last calibration
• Application data
– Temperature application range
– Conductivity application range
– Date of first commissioning
– Maximum temperature value
– Operating hours at high temperatures

2 Endress+Hauser
Indumax CLS50/CLS50D

Measuring system CLS50D


A complete measuring system comprises:
• a CLS50D conductivity sensor with fixed cable
• a transmitter, e.g. Liquiline CM44x
Optional:
• Weather protection cover for the field installation of the transmitter
• Assembly for sensor installation in tanks or in pipework, e.g. CLA111

1 2

4 3
a0012990

Example of a measuring system


1 Transmitter Liquiline CM44x
2 Weather protection cover
3 Pipe socket for flange DN50 PN/6
4 Sensor CLS50D, flange version DN50 PN/6, fixed cable with M12 plug

Endress+Hauser 3
Indumax CLS50/CLS50D

CLS50
A complete measuring system comprises:
• a CLS50 conductivity sensor with fixed cable
• a transmitter, e.g. Liquiline CM42
Optional:
• Weather protection cover for the field installation of the transmitter
• Assembly for sensor installation in tanks or in pipework, e.g. CLA111

1 2

5 3

4
a0007045

Example of a measuring system


1 Indumax CLS50
2 Liquiline CM42 transmitter
3 Weather protection cover
4 Pipework
5 Pipe socket

Input
Measured variable • Conductivity
• Temperature

Cell constant k = 1.98 cm–1

Measuring frequency 2 kHz

Measuring range Conductivity: 2 μS/cm to 2000 mS/cm (uncompensated)


Temperature: –20 to +180 ˚C (-4 to +350 ˚F)

Temperature measurement CLS50D


Pt 1000 (class A acc. to IEC 60751)
CLS50
Pt 100 (class A acc. to IEC 60751)

4 Endress+Hauser
Indumax CLS50/CLS50D

Wiring
Cable specification The sensor is supplied with a fixed cable. The connection to the transmitter can be extended using the CYK81
(CLS50D) or the CLK5 (CLS50) special measuring cable.

RD
WH BU
BN U+ RD
WH U– BU
RD
GN COM A GN

Pt 100
YE COM B WH
YE
BN n.c.
a0013000 a0012998

CYK81 for cable extension for CLS50D CLK5 for cable extension for CLS50
– max. total cable length: 100 m (330 ft) – max. total cable length: 55 m (180 ft)

Please note that the residual coupling increases when the cable is extended.

Performance characteristics
Maximum measured error • -20 to 100 ˚C (-4 to 212 ˚F):
±(5 μS/cm + 0.5 % of the measured value)
• > 100 ˚C (212 ˚F):
±(10 μS/cm + 0.5 % of the measured value)

Temperature response time • PEEK versions:


approx. 7 min
• PFA versions:
approx. 11 min

Installation conditions
Installation position Install the sensor in such a way that the sensor opening is oriented in the flow direction of the medium. The
sensor head must be completely immersed in the medium.

a0007035

Installation position of the sensor


1 Flow direction of medium
2 Minimum water level in the pipe
a Sensor distance from the pipe wall

Endress+Hauser 5
Indumax CLS50/CLS50D

Installation factor In narrow installation conditions, the conductivity measurement is affected by the pipe walls. This effect is
compensated by the so-called installation factor. The transmitter corrects the cell constant by multiplication
with the installation factor.
The value of the installation factor depends on the diameter and the conductivity of the pipe as well as the
sensor’s distance from the wall.
If the distance from the wall is sufficient (a > 15 mm (0.59"), from DN 80), it is not necessary to consider the
installation factor (f = 1.00).
If the distance from the wall is smaller, the installation factor increases in case of electrically insulating pipes
(f > 1) and decreases in case of electrically conductive pipes (f < 1).
The installation factor can be measured using calibration solutions or it can be approximately determined from
the following diagram.

1.40
2

1.20

1.00

1
0.80
0 5 10 15 20 25 a [mm]

0.20 0.39 0.59 0.79 0.98 a [inch]


a0005441

Relationship between installation factor f and distance from wall a


1 Electrically conductive pipe wall
2 Insulating pipe wall

Air set CLS50


To compensate residual coupling in the cable and between the two sensor coils, you must perform a zero
calibration in air ("air set") before installing the sensor.
For further information, refer to the Operating Instructions of your transmitter.
CLS50D
The digital sensor is already adjusted at the factory, an on-side compensation is not necessary.

6 Endress+Hauser
Indumax CLS50/CLS50D

Installation of sensors with The sensor is suitable for installation in T-pieces ≥ DN 80 with the outgoing diameter reduced to ≥ DN 50.
flange ! WARNING
Leakages
Danger of injuries by leaking medium
► Tighten the nut with a torque of 20 Nm.
► To avoid leakages, regularly check the tightness of the nut.

Flange, not in contact with medium

a0007061

Fixed flange, not in contact with medium (order option "process connection": 5, 6, 7)
1 Flange (stainless steel) 4 O-ring
2 Nut 5 Sensor
3 Sealing disk (GYLON)

Flange, in contact with medium

a0007062

Fixed flange, in contact with medium (order option "process connection": 3, 4)


1 Flange (stainless steel) 3 O-ring
2 Nut 4 Sensor

Endress+Hauser 7
Indumax CLS50/CLS50D

Lap-joint flange, not in contact with medium

a0007063

Lap-joint flange, not in contact with medium (order option "process connection": A, B, C)
1 Lap-joint flange (PP-GF) 4 O-ring
2 Nut (stainless steel) 5 Sensor
3 Flange (PVDF)

Installation of sensor assembly

a0007064

Installation of sensor with assembly


1 CLA111 with suspension bracket 3 CLA140 with flange connection
2 CLA111 with flange connection 4 CYA611

Environment
Ambient temperature range CLS50D
–10 to +60 ˚C (+10 to +140 ˚F)
CLS50
–10 to +70 ˚C (+10 to +160 ˚F)

Storage temperature –20 to +80 ˚C (-4 to +180 ˚F)

Ingress protection IP 68 / NEMA 6 (installed with original sealing)

8 Endress+Hauser
Indumax CLS50/CLS50D

Process
Process temperature CLS50D

Without flange DN50, ANSI 2" JIS PVDF lap-joint flange

Sensor CLS50D-*1/2 CLS50D-*3/4/5/6/8 CLS50D-*7 CLS50D-*A/B/C

PEEK –20 to 125 °C –20 to 125 °C –20 to 125 °C –20 to 125 °C


(-4 to 260 °F) (-4 to 260 °F) (-4 to 260 °F) (-4 to 260 °F)
PFA –20 to 110 °C –20 to 125 °C –20 to 125 °C –20 to 125 °C
(-4 to 230 °F) (-4 to 260 °F) (-4 to 260 °F) (-4 to 260 °F)

CLS50

Without flange DN50, ANSI 2" JIS PVDF lap-joint flange

Sensor CLS50-*1/2 CLS50-*3/4/5/6/8 CLS50-*7 CLS50-*A/B/C

PEEK –20 to 180 °C –20 to 180 °C –20 to 180 °C –20 to 125 °C


(-4 to 360 °F) (-4 to 360 °F) (-4 to 360 °F) (-4 to 260 °F)

PFA –20 to 125 °C –20 to 125 °C –20 to 125 °C –20 to 125 °C


(-4 to 260 °F) (-4 to 260 °F) (-4 to 260 °F) (-4 to 260 °F)

Process pressure max. 20 bar (290 psi), acc. to the sensor version, see pressure-temperature diagram

Pressure-temperature
diagram

a0018311

Pressure-temperature-diagram of CLS50D
1 PEEK sensor, without flange
2 PFA sensor, without flange (blue line)
3 PEEK or PFA sensor, with DN50/ANSI 2" flange (red line)
4 PEEK or PFA sensor, with JIS flange
5 PEEK or PFA sensor, with lap-joint flange PVDF (green line)

Endress+Hauser 9
Indumax CLS50/CLS50D

a0018312

Pressure-temperature-diagram of CLS50
1 PEEK sensor, without flange
2 PFA sensor, without flange or with DN50/ANSI 2" flange (blue line)
3 PEEK sensor, with DN50/ANSI 2" flange (red line)
4 PFA sensor, with JIS flange (black line)
5 PEEK or PFA sensor, with lap-joint flange PVDF (green line)
6 PEEK sensor, with JIS flange (grey line)

Mechanical construction
Sensor dimensions

a0007070
a0007069

Dimensions of sensor version with G ¾ thread Dimensions of sensor version with NPT 1" thread
* Dimensions of PEEK version

10 Endress+Hauser
Indumax CLS50/CLS50D

Flange dimensions 1 2

b
d2 d2

a a
Øk Øk
D D
a0007071

Flange dimensions
1 Lap-joint flange (PP-GF)
2 Fixed flange (stainless steel)

Lap-joint flange PP-GF DN50 PN10 ANSI 2" 150 lbs JIS 10K 50A

D 165 165 152

Øk 125 121 120

d2 4 x 18 8 x 19 4 x 19

b 18 18 18

a 78 78 78

Screws M16 M16 M16

Fixed flange SS 316 L DN50 PN16 ANSI 2" 300 lbs JIS 10K 50A

D 165 165.1 155

Øk 125 127 120

d2 4 x 18 8 x 19 4 x 19

b 18 22.2 16

a 27 27 27

Screws M16 M16 M16

Weight approx. 650 g (1.43 lbs)

Material Sensor: PEEK, PFA (depending on ordered version)


Sensor seal: VITON, CHEMRAZ (depending on ordered version)
Process connections:
G ¾: CLS50-**A: stainless steel 1.4571 (AISI 316Ti)
CLS50-**B/C: PEEK GF30
CLS50D-**D: stainless steel 1.4571 (AISI 316Ti)
CLS50D-**B/C: PEEK GF30
NPT 1": PEEK
Fixed flange: stainless steel 1.4404 (AISI 316 L)
Sealing disk: GYLON (ceramic-filled PTFE)
Lap-joint flange: PP-GF
Flange combined with lap-joint flange: PVDF

Endress+Hauser 11
Indumax CLS50/CLS50D

Chemical durability Medium Concentration PEEK PFA CHEMRAZ VITON

Sodium hydroxide 0 to 50 % 20 to 100 ˚C not suitable 0 to 150 ˚C not suitable


solution NaOH (68 to 212 ˚F) (32 to 302 ˚F)
0 to 10 % 20 to 100 ˚C 20 to 80 ˚C 0 to 150 ˚C 0 to 120 ˚C
Nitric acid (68 to 212 ˚F) (68 to 176 ˚F) (32 to 302 ˚F) (32 to 248 ˚F)
HNO3 0 to 40 % 20 ˚C (68 ˚F) 20 to 60 ˚C 0 to 150 ˚C 0 to 120 ˚C
(68 to 140 ˚F) (32 to 302 ˚F) (32 to 248 ˚F)

Phosphoric acid 0 to 80 % 20 to 100 ˚C 20 to 60 ˚C 0 to 150 ˚C 0 to 120 ˚C


H3PO4 (68 to 212 ˚F) (68 to 140 ˚F) (32 to 302 ˚F) (32 to 248 ˚F)
0 to 2.5 % 20 to 80 ˚C 20 to 100 ˚C 0 to 150 ˚C 0 to 120 ˚C
Sulphuric acid (68 to 176 ˚F) (68 to 212 ˚F) (32 to 302 ˚F) (32 to 248 ˚F)
H2SO4 0 to 30 % 20 ˚C (68 ˚F) 20 to 100 ˚C 0 to 150 ˚C 0 to 120 ˚C
(68 to 212 ˚F) (32 to 302 ˚F) (32 to 248 ˚F)

0 to 5 % 20 to 100 ˚C 20 to 80 ˚C 0 to 150 ˚C 0 to 120 ˚C


Hydrochloric acid (68 to 212 ˚F) (68 to 176 ˚F) (32 to 302 ˚F) (32 to 248 ˚F)
HCl 0 to 10 % 20 to 100 ˚C 20 to 80 ˚C 0 to 150 ˚C 0 to 120 ˚C
(68 to 212 ˚F) (68 to 176 ˚F) (32 to 302 ˚F) (32 to 248 ˚F)

No responsibility is taken for the correctness of this information.

Ordering information
Product page You can create a complete and valid order code by using the configurator on the internet product page.
Enter the following address to access the product page:
www.products.endress.com/cls50
www.products.endress.com/cls50d

Online configurator 1. You can choose from the following options on the product page located on the right:

2. Click "Configure this product".


3. The configurator opens in a separate window. You can now configure your device and receive the
complete order code that applies for the device.
4. Afterwards, export the order code as a PDF or Excel file. To do so, click the appropriate button at the top
of the page.

12 Endress+Hauser
Indumax CLS50/CLS50D

Product structure The following product structure represents the status of printing. You can create a complete and valid
order code on the Internet using the configurator tool.

CLS50D
Approval
AA Non-hazardous area
BA ATEX II 1G Ex ia IIC T3/T4/T6
BV ATEX II 3G Ex nL IIC T4/T6 + NEPSI Ex nL IIC T4/T6
C2 CSA IS NI Cl . I, II, III, Div. 1&2, Group A-G
FB FM IS NI Cl . I, II, III, Div. 1&2, Group A-G
NA NEPSI Ex ia IIC T4/T6

Process connection
A Lap joint flange DN 50 PN 10, PP-GF
B Lap joint flange ANSI 2" 150 lbs, PP-GF
C Lap joint flange JIS 10 K 50 A, PP-GF
1 Thread G ¾
2 Thread NPT 1", PEEK
3 Flange DN 50 PN 16, stainless steel 1.4404 (AISI 316 L)
4 Flange ANSI 2" 300 lbs, stainless steel 1.4404 (AISI 316 L)
5 Flange DN 50 PN 16, stainless steel 1.4404 (AISI 316 L), PTFE sealing disk
6 Flange ANSI 2" 300 lbs, stainless steel 1.4404 (AISI 316 L), PTFE sealing disk
7 Flange JIS 10 K 50 A, stainless steel 1.4404 (AISI 316 L), PTFE sealing disk
8 Flange DN 50 PN 16, stainless steel 1.4404 (AISI 316 L), welded PFA sealing plate sensor

Material: sensor , seal, adapter


B PEEK, VITON, PEEK
C PEEK , CHEMZRAZ, PEEK
D PFA, CHEMRAZ, stainless steel 1.4571 (AISI 316 Ti)

Cable length
1 3 m (9.84 ft)
2 7 m (23 ft)
3 15 m (49 ft)
7 ... m fixed cable, max. 50 m
8 ... ft fixed cable, max. 164 ft

Cable connection
1 Fixed cable, terminated cable cores
2 Fixed cable, M12 plug

CLS50D- order code

Calibration (single choice only)


F1 Cell constant

Dienstleistung (multiple choice possible)


HC LABS free (LABS = paint-wetting impairment substances)

TAG (multiple choice possible)


1 TAG, see additional specification
2 Feldbus TAG

Endress+Hauser 13
Indumax CLS50/CLS50D

CLS50
Approval
A Non-hazardous area
G ATEX II 1G Ex ia IIC T4/T6 Ga
H NEPSI Ex ia IIC T4/T6
L Non-hazardous area, PWIS free
O FM IS NI Cl . I, II, III, Div. 1&2, Group A-G
S CSA IS NI Cl . I, II, III, Div. 1&2, Group A-G
T TIIS
V ATEX II 3G Ex nL IIC T4/T6 + NEPSI Ex nL IIC T4/T6

Process connection
A Lap joint flange DN 50 PN 10, PVDF
B Lap joint flange ANSI 2" 150 lbs, PVDF
C Lap joint flange JIS 10 K 50 A, PVDF
1 Thread G ¾
2 Thread NPT 1", PEEK
3 Flange DN 50 PN 16, stainless steel 1.4404 (AISI 316 L)
4 Flange ANSI 2" 300 lbs, stainless steel 1.4404 (AISI 316 L)
5 Flange DN 50 PN 16, stainless steel 1.4404 (AISI 316 L), PTFE sealing disk
6 Flange ANSI 2" 300 lbs, stainless steel 1.4404 (AISI 316 L), PTFE sealing disk
7 Flange JIS 10 K 50 A, stainless steel 1.4404 (AISI 316 L), PTFE sealing disk
8 Flange DN 50 PN 16, stainless steel 1.4404 (AISI 316 L), welded PFA sealing plate sensor

Material: sensor, seal, adapter


A PFA, CHEMRAZ, stainless steel 1.4571 (AISI 316 Ti)
B PEEK, VITON, PEEK
C PEEK , CHEMZRAZ, PEEK

Cable length and temperature range


1 5 m (16 ft) fixed cable, max. 125 ˚C (257 ˚F)
2 10 m (32 ft) fixed cable, max. 125 ˚C (257 ˚F)
3 20 m (65 ft) fixed cable, max. 125 ˚C (257 ˚F)
4 fixed cable of specific length, max. 55 m (180 ft), max. 125 ˚C (257 ˚F)
5 5 m (16 ft) fixed cable, max. 180 ˚C (356 ˚F) (PEEK only, version for non-hazardous areas only)
6 10 m (32 ft) fixed cable, max. 180 ˚C (356 ˚F) (PEEK only, version for non-hazardous areas only)

CLS50- complete order code

Certificates and approvals


Ex approvals CLS50D
• ATEX II 1G Ex ia IIC T4/T6 Ga
CLS50
• ATEX II 1G Ex ia IIC T4/T6 Ga
• ATEX II 3G Ex nL IIC T4/T6 + NEPSI Ex nL IIC T4/T6
• FM/CSA IS NI Cl. I, II, III, Div. 1 & 2; Group A - G in connection with a Liquiline CM42 or Mycom
CLM153 transmitter
• NEPSI Ex ia IIC T4/T6

14 Endress+Hauser
Indumax CLS50/CLS50D

Accessories
Measuring cables For CLS50D
CYK11 Memosens data cable
• Extension cable for digital sensors with Memosens protocol
• Ordering as per product structure (--> Online configurator, www.products.endress.com/cyk11)

For CLS50
Extension cable CLK6
• For inductive conductivity sensors, for extension via the VBM junction box, sold by the meter
• Order no.: 71183688
Junction box VBM
• For cable extension
• 10 terminals
• Cable entries: 2 x Pg 13.5 or 2 x NPT ½"
• Material: aluminum
• Ingress protection: IP 65 (i NEMA 4X)
• Order numbers:
– cable entries Pg 13.5: 50003987
– cable entries NPT ½": 51500177

Assemblies Dipfit CLA111


• Immersion assembly for open and closed tanks with flange DN 100
• Ordering acc. to product structure, --> online configurator: www.products.endress.com/cla111
• Technical Information TI00135C/07/EN
Dipfit CLA140
• For the inductive sensors CLS50/CLS50D
• Immersion assembly with flange connection for highly demanding processes
• Ordering acc. to product structure, --> online configurator: www.products.endress.com/cla140
• Technical Information TI00196C/07/EN
Immersion assembly Dipfit CYA611
• For sensor immersion in basins, open channels and tanks, PVC
• Ordering acc. to product structure, --> online configurator: www.products.endress.com/cya611
• Technical Information TI00166C/07/EN

Calibration solutions Precision solutions, traceable to SRM (standard reference material) by NIST, for qualified calibration of
conductivity measurement systems according to ISO 9000, with temperature table
• CLY11-B
149.6 μS/cm (reference temperature 25 ˚C / 77 ˚F), 500 ml / 16.9 fl.oz
Order no. 50081903
• CLY11-C
1.406 mS/cm (reference temperature 25 ˚C / 77 ˚F), 500 ml / 16.9 fl.oz
Order no. 50081904
• CLY11-D
12.64 mS/cm (reference temperature 25 ˚C/ 77 ˚F), 500 ml / 16.9 fl.oz
Order no. 50081905
• CLY11-E
107.0 mS/cm (reference temperature 25 ˚C / 77 ˚F), 500 ml / 16.9 fl.oz
Order no. 50081906

Endress+Hauser 15
TI00182C/07/EN/15.12
71186403
FM9
Technical Information

Stamolys CA71PH
Phosphate analyzer
Compact photometric analysis system for the phosphate measurement in
sewage treatment plants and cooling water cycles

Application
• Monitoring and optimising the cleaning capacity of sewage
treatment plants
• Controlling of precipitant addition
• Monitoring of activated sludge basins
• Monitoring of sewage treatment plant outlet
• Monitoring of cooling water cycles

Your benefits
• Stainless steel or glass-fibre reinforced carbon housing
available
• Two channel version available
• Measured value storage using integrated data logger
• Automatic calibration and self-cleaning
• Free selectable measuring, cleaning and calibration intervals

TI356C/07/en/05.09
71095964
Stamolys CA71PH

Function and system design


Measuring principle After sample conditioning, the analyzer sample pump conveys a part of the filtrate to a mixing vessel. The
reagent pump adds reagent at a specific ratio. As a result of the reaction, the sample turns a characteristic
colour. The photometer determines the sample’s absorption of an emitted light at a specific wavelength (s. Fig.,
Pos. 2). The wavelength is parameter specific. The absorbance is proportional to the concentration of the
specified parameter in the sample (Pos. 3). Additionally, the absorption of a reference light is determined to
receive a genuine measuring result. The reference signal is subtracted from the measuring signal to prevent any
effects due to turbidity, contamination and ageing of the LEDs.
The temperature in the photometer is controlled thermostatically so that the reaction is reproducible and takes
place within a short period of time.

1 Reference LED
2 Emitter LED
3 Sample
4 Reference detector
5 Measuring detector

a0001612

Photometric principle

Phosphorus and Phosphates Phosphorus is mainly present in natural bodies of water and in wastewater as phosphate.
The phosphate gets into water through:
• fertiliser washed out of the earth
• biological and industrial wastes and wastewater
• additives in water treatment systems (anticorrosion agents)
Phosphate is essential to life in specific proportions for animals and plants. Too much of it can, however, lead
to eutrophication1).

Orthophosphate and total Phosphate is generally subdivided into:


phosphate • orthophosphate
• condensed phosphate:
– metaphosphate
– pyrophosphate
– polyphosphate
• organically bound phosphate
Only orthophosphate can be determined without disintegration of a sample since it can be detected directly
photometrically. One speaks in this case of the determination of "reactive" phosphorus.
The measurement results can be indicated in various ways:
• PO4, phosphate
• PO4-P, phosphate-phosphorus
• P2O5, phosphorus pentoxide
E+H offers two methods for determining reactive phosphorus (depending on the measuring range):
• The Molybdenum Blue Method (2 reagents, specifications A+E)
• The Molybdate Vanadate Method (1 reagent, specifications B+D)

1) eutrophication = an increase in the amount of nutrients in bodies of water which can lead to undesirable usuriousness of certain plants.

2 Endress+Hauser
Stamolys CA71PH

Photometric orthophosphate Molybdenum blue method in deviation to DIN EN 1189


determination (versions PH-A and PH-E)
In acid solutions molybdate ions and antimony ions form in conjunction with phosphate an antimonyl
phosphoro molybdato complex. The complex is reduced to phosphoro molybdato blue by ascorbic acid.
The absorption is determined at a wavelength of 880 nm (PH-A) resp. of 660 nm (PH-E). The absorption
intensity is proportional to the orthophosphate concentration in the sample.
The reference wavelength is 565 nm.
Molybdate vanadate method
(versions PH-B and PH-D)
Vanadate ions and molybdate ions form in conjunction with phosphate a yellow vanadato molybdato
phosphorous acid.
The absorption is determined at a wavelength of 430 nm. The absorption intensity is proportional to the
orthophosphate concentration in the sample.
The reference wavelength is 565 nm.

Interferences No interferences up to the given concentration:

Concentration [mg/l (ppm)] Interference

10,000 SO42–
1,000 Cl–

500 Na+, K+, Ca2+

50 CO32–, NO3–, Fe2+, Fe3+, Zn2+, Cu2+, Ni2+, Cr3+, Co2+, Hg2+
25 Sn2+

10 Pb2+

5 Ag+
0.5 Cr6+, can be eliminated by higher ascorbic acid dosage

Turbidity: sample has to be filtered before analysis

Sample conditioning Micro/ultrafiltration (Stamoclean CAT430, optional)


A membrane filter element is suspended directly into the wastewater basin or channel. A hose pump is located
in a pump box on the basin rim.The pump creates a vacuum between the membrane and the carrier plate of
the filter element. This vacuum makes the filtrate pass through the filter membrane. Suspended materials,
particles, algae and bacteria are collected on the surface of the membrane.
Due to alternating pumping and pause, intervals of more than one month are achieved between cleaning
cycles. Parallel connection of two or four filter elements increases the sampling quantity up to approx. 1 l/h
(0.26 gal/hr).
The hose pump pressure transports the sample to a collecting vessel near the analyzer over a distance of 20 m
(66 ft). For distances up to 100 m (330 ft) the sample is transported to the collecting vessel by means of
compressed air. The analyzers suck the needed sample volume from the collecting vessel.

Membrane filtration (Stamoclean CAT411, optional)


A sample flow of 0.8 to 1.8 m3/h (3.5 to 8 gal/min) is continuously conducted through the micro filter via a
pressure pipe. A part of the sample passes the filter membrane and is then conveyed to the measuring device
as filtrate.
Sampling is based on the cross flow filtration principle. The PTFE filter membrane separates particles with sizes
> 0.45 μm from the filtrate. These particles are collected in front of the membrane and are washed away with
the sample flow.
The medium is conducted in a meander-like channel through the filter element. This results in a constantly
high flow rate. The high flow rate generates the self cleaning effect. Therefore, mechanical drives for the
generation of a flow at the filter surface are not necessary.

Backwash filter (Stamoclean CAT221, optional)


A sample flow of 1 to 2.5 m3/h (4.4 to 11 gal/min) is permanently conveyed through the backwash filter by
means of a sampling pump or compressed air or rinse water. The filtrate passes through the wedge wire sieve
and is then transported to the measuring device.
Clogging is minimized by the flow at the wedge wire sieve. Automatic backwashing results in a filter operating
time of several weeks.

Endress+Hauser 3
Stamolys CA71PH

The automatic backwashing and a small compressor or compressed air resp. rinse water supply guarantee
low-maintenance and low-energy operation.

Customer specific solution


Before analysis, the sample has to be conditioned and to be transported to an external or to the delivered
collecting vessel.

Measuring system A complete measuring system comprises:


• An analyzer
• A sample conditioning system (optionally):
– Micro filtration / ultra filtration Stamoclean CAT430 or Stamoclean CAT411
– Backwash filter Stamoclean CAT221
– Customer specific solution
• Collecting vessel (see product structure)

Micro / ultra filtration


1 Control box
1 2 3 2 Pump
3 Control unit
4 Collecting unit (optional)
5 Collecting vessel
6 Overflow
7 Analyzer
8 Aeration basin
4 5 6 7 9 Membrane filter

9 8

a0001616

Measuring system with Stamoclean CAT430

1 Stamoclean CAT411
2 Inlet
3 Sample pump or hydraulic main
4 Filtrate line
5 Collecting vessel
6 Overflow
7 Analyzer
8 Analyzer sample line
9 Outlet
a0001615

Measuring system with Stamoclean CAT411

Backwash filter
1 Stamoclean CAT221
2 Compressor or compressed air
3 Sample pump or hydraulic main
1 4 Sample outlet
5 Collecting vessel
6 Overflow
2 7 Analyzer

7 6 5 4 3

a0001617

Measuring system with Stamoclean CAT221

4 Endress+Hauser
Stamolys CA71PH

Standard applications Sewage treatment plant outlet monitoring


Sampling from hydraulic main and analyzer in measuring station:
• Backwash filter Stamoclean CAT221 (order no. CAT221-Axxx)
• Compressor for CAT221 (order no. 51511143)
• Analyzer with collecting vessel, Stamolys CA71PH-A1xB2A1
Sampling from open channel
Local filtration and analyzer in measuring station (up to 20 m (66 ft) distance):
• Ultra filtration Stamoclean CAT430, plate filter with hose heating for max. 20 m (66 ft) distance to the
analyzer (order no. CAT 430-A1F0A3A)
• Filter element holder with horizontal slide (order no. 51511374)
• Analyzer with collecting vessel, Stamolys CA71PH-A1xB2A1
Precipitant addition monitoring in the biology
Local filtration and analyser in measuring station (up to 100 m (328 ft) distance):
• Ultra filtration Stamoclean CAT430, plate filter with hose heating over 18 m (59 ft) , remaining distance
freeze free installed, sample transportation by compressed air up to 100 m (order no. CAT 430-A4F0A3A)
• Filter element holder with vertical slide (order no. 51511354)
• Analyzer with collecting vessel, Stamolys CA71PH-A1xB2A1 or CA71PH-B1xB2A1
Precipitant addition monitoring in the biology
Local filtration and analyser in measuring station (up to 20 m (66 ft) distance):
• Ultra filtration Stamoclean CAT430, plate filter with hose heating over max. 20 m (66 ft) distance to the
analyzer (order no. CAT430-A1F0A3A)
• Filter element holder with vertical slide (order no. 51511354)
• Analyzer with collecting vessel, Stamolys CA71PH-A1xB2A1 or CA71PH-B1xB2A1

Input
Measured variable PO4-P [mg/l]

Measuring range PH-A


0.05 to 2.5 mg/l (0.05 to 2.5 ppm)
PH-B
0.5 to 20 mg/l (0.5 to 20 ppm)
PH-D
0.5 to 50 mg/l (0.5 to 50 ppm)
PH-E
0.05 to 10 mg/l (0.05 to 10 ppm)

Wavelength PH-A
880 nm
PH-B and PH-D
430 nm
PH-E
660 nm

Reference wavelength 565 nm

Output
Output signal 0/4 to 20 mA

Signal on alarm Contacts: 2 limit contacts (per channel), 1 system alarm contact
optional: end of measurement (with two channel version display of channel no. available)

Endress+Hauser 5
Stamolys CA71PH

Load max. 500 Ω

Data interface RS 232 C

Data logger 1024 data pairs per channel with date, time and measured value
100 data pairs with date, time and measured value for calibration factor determination (diagnostic tool)

Load capacity 230 V / 115 V AC max. 2 A, 30 V DC max. 1 A

Power supply
Electrical connection Caution!
The following figure (→ å 1) shows the connection department sticker as an example. Terminal assignment
and cable core colors can be different to the originals.
For connecting your analyzer only use the terminal assignment of the connection department sticker in the
device (→ å 2)!

Photometer +17V Photometer -17V Photometer 0V fx Kanal 1 Kanal 2


Analog Out Channel 1
Schirm MP5 MP6 Channel 2
BK 65 0/4-20mA COM
F4 M 0.2 A M 0.2 A Shield 28 22 16

ALARM 2
F5

ALARM 2

OPTION
Schirm Schirm
GN 64 0V BK 58 Shield 40 Shield NC
27 21 15
Schirm Schirm keine Probe
80 V4 Shield BN 63 -17 V YE 57 I2 + NO
Screen 75 70 No sample 39
26 20 14
n.b.
COM
SCHRITTMOTOR

OR 79 74 V3 69 n.c. WH 62 +17 V 56 Reserve 38 I2 -


25 19 13

STÖRUNG
STEP MOTOR

ALARM 1
ALARM 1

ERROR
n.b. Mess 2 Schirm n.b.
BK 78 73 V2 68 YE 61 fx 55 37 31 NC
n.c. Meas 2 Shield n.c. 24 18 12
n.b. n.b.
LEAK S.
LECKF.

RD 77 72 V 1 BU 67 60
n.c. WH 54 +24V 36 I1 + 30 NO
n.c. 23 17 11
YE 76 71 0 V BN 66 GY 59 Status GN 53 0V 35 I1 - 29 0V RELAIS / Relay:
max. Last / load: 2A bei / at 115/230V AC, 1A bei / at 30V DC

Klemmen führen auch bei ausgeschaltetem Gerät Spannung !!!

Terminals voltage to ground even when the unit is switched off !!!

F2 NETZ
AC Motor MAINS
BU 7 L2 T 0.1A
AC MOTOR

YE 6 N 3 L

RD L1 Elektronik F1 N
5 2
Electronics
GN/YE PE T 0.5A
4 1 PE

a0001369

Fig. 1: Example of the connection sticker

2 3

Photometer +17V Photometer -17V Photometer 0V fx Kanal 1 Kanal 2 Klemmen führen auch bei ausgeschaltetem Gerät Spannung !!!
Analog Out Channel 1
Schirm MP5 MP6 Channel 2
BK 65 0/4-20mA COM
F4 M 0.2 A M 0.2 A Shield 28 22 16 Terminals voltage to ground even when the unit is switched off !!!
ALARM 2

F5
ALARM 2

OPTION

Schirm Schirm
GN 64 0V BK 58 Shield 40 Shield NC
27 21 15
Schirm Schirm keine Probe
80 V4 Shield BN 63 YE 57 NO
Screen 75 -17 V
70 No sample 39 I2 +
26 20 14 F2 NETZ
n.b. AC Motor MAINS
BU L2
SCHRITTMOTOR

OR 79 74 V3 69 WH 62 +17 V 56 Reserve 38 I2 - COM 7 T 0.1A


n.c.
STEP MOTOR

25 19
STÖRUNG

13
ALARM 1
ALARM 1

ERROR

n.b. Mess 2 n.b.


AC MOTOR

BK 78 73 V2 68 YE 61 fx 55 37 Schirm
31 NC YE 6 N 3 L
n.c. Meas 2 Shield n.c. 24 18 12
n.b. n.b. Elektronik
LEAK S.

F1
LECKF.

RD 77 72 V 1 BU 67 60 WH 54 +24V 36 I1 + 30 NO RD 5 L1 2 N
n.c. n.c. 23 17 11 Electronics
GN/YE PE T 0.5A
YE 76 71 0 V BN 66 GY 59 Status GN 53 0V 35 I1 - 29 0V RELAIS / Relay: 4 1 PE
max. Last / load: 2A bei / at 115/230V AC, 1A bei / at 30V DC

1
a0001372

Fig. 2: Analyzer from top (open version resp. swung out)


1 Connection department sticker
2 Printed circuit board with terminal strip
3 Backside of the analyzer

Supply voltage 115 V AC / 230 V AC ±10%, 50/60 Hz

6 Endress+Hauser
Stamolys CA71PH

Power consumption approx. 50 VA

Current consumption approx. 0.2 A at 230 V


approx. 0.5 A at 115 V

Fuses 1 x time-lag 0.5 A for electronics


2 x medium time-lag 0.2 A for photometer
1 x time-lag 0.5 A for motors

Performance characteristics
Time between two tmes = reaction time + rinse time + waiting time + rinse again time + filling time + sampling time +
measurements reagent refusal time (min. waiting time = 0 min)

Maximum measured error ±2 % of measuring range end

Measuring interval tmes to 120 min

Reaction time 6 minutes

Sample requirement 15 ml (0.51 fl.oz.) per measurement

Reagent requirement PH-A: 2 x 0.1 ml (0.0034 fl.oz.)


PH-B: 1 x 0.1 ml (0.0034 fl.oz.)
PH-D: 1 x 0.20 ml (0.0068 fl.oz.)
PH-E: 2 x 0.20 ml (0.0068 fl.oz.)
0.43 l (0.11 US.gal) (PH-A/B) resp.
0.86 l (0.23 US.gal) (PH-D/E) per reagent per month with 10 minute measuring interval

Calibration interval 0 to 720 h

Rinse interval 0 to 720 h

Rinse time selectable from 20 to 300 s (standard = 60 s)

Rinse again time 30 s

Filling time 25 s

Maintenance interval 6 months (typical)

Servicing requirement 15 minutes per week (typical)

Environment
Ambient temperature 5 to 40 °C (40 to 100 °F), avoid strong fluctuations

Humidity below the condensation limit, installation in usual, clean rooms


outdoor installation only possible with protective devices (customer supplied)

Ingress protection IP 43

Endress+Hauser 7
Stamolys CA71PH

Process
Sample temperature 5 to 40 °C (40 to 100 °F)

Sample flow rate min. 5 ml (0.0013 US.gal.) per min

Consistence of the sample low solid content (< 50 ppm)

Sample inlet Unpressurized

Mechanical construction
Design, dimensions Stainless steel housing

3(0.12) Detail A 30(1.18)


38(1.5)

15(0.59)
163(6.42)
110(4.33)
Ø 27(1.06)

15(0.59) 9(0.35)

250 (9.84)
330 (13.0)
160 (6.30)

15(0.59) 140 (5.51)


100(3.94)
60(2.36)
120(4.72)

A
Ø 26 (1.02)
243 (9.57)

200 (7.87)
Ø 21(0.83)
165 (6.50)

125(4.92)

47(1.85)

580 (22.8)
648 (25.5)
555 (21.8)
524 (20.6)

78(3.07)
119(4.69)

Ø 21(0.83)

326 (12.8)

Ø 26(1.02)

360 (14.2) 207 (8.15)

436 (17.2) mm(inch)


a0001361

Stainless steel version

8 Endress+Hauser
Stamolys CA71PH

GFR housing

163 (6.42)

250 (9.84)

120 (4.72)
Ø 26(1.02)
Ø 16(0.63)
160 (6.30)
140 (5.51)
100(3.94)
330 (13.0)
60 (2.36)

Ø 26(1.02)
(4.72)
7-11/

120
0.28 - 0.43

243 (9.57)
(M 6)

Ø 21(0.83)
165 (6.50)

125 (4.92)
200 (7.87)

47(1.85)
555 (21.8)

(0.83)

648 (25.5)
Ø 21
78(3.07)
119/4.69
Ø 26 (1.02)
326 (12.8)

Ø 16 (0.63)

360 (14.2)
436 (17.2) Ø 27 (1.06)
465 (18.3) 120(4.72)
500 (19.7)
mm(inch)
a0001354-en

GFR version

Open version

381 (15.0)

351 (13.8)

19 (0.75) 197 (7.76)

192 (7.56)
Ø6 (0.24)

RS 232
540 (21.3)

562 (22.1)

175 (6.89)

325 (12.8)

9 (0.35) 8 (0.31) mm (inch)


a0001356

Open version (without housing)

Endress+Hauser 9
Stamolys CA71PH

Note!
With the open version, you need an additional platform for the reagents. Mount this platform max. 35 cm
(13.8 inch) below the pumps. The reagent bottels have the following dimensions: 90 x 90 x 215 mm (3.54 x
5.54 x 8.46 inch). The number of bottles varies from 2 to 5 depending on the analyzer version.
For these versions, the outlet pipe must be installed right of the analyzer. See the supplement to the Operating
Instructions.
The outlet pipe must be mounted to a wall so that the sample outlet hoses from the photometer have a gradient
of 5 to 10 %. If neccessary, extend the hoses.

Collecting vessel

145 (5.71)
4 x Ø 6.6 (0.26)
27 (1.06) 60 (2.36)

70(2.76) 42(1.65)

48(1.89)

79 (3.11)

120 (4.72)
130 (5.12)

140 (5.51)
17 (0.67)
10 (0.39)
3 x Ø 28/20 (1.1/0.79)
115 (4.53)
Ø 34/25 (1.34/0.98)
4 x Ø 5.5/10.4 (0.22/0.41)

Ø 56 (2.2)
*
Ø 50H7 (1.97H0.28)

60(2.36)

80(3.15)
100 (3.9)
*

M20x1.5
Ø 24 (0.94) *
mm (inch) 70 (2.76)
a0001363 a0001346

Collecting vessel at analyzer (optional) Collecting vessel dimensions


1 Ventilation * variable, freely adjustable dimensions
2 Sample inlet from sampling 6 Sampling for analyzer
3 Collecting vessel 7 Sample overflow
4 Electrical connections 8 Analyzer outlet
5 Analyzer sample inlet

Weight GFR housing approx. 28 kg (62 lbs)


Stainless steel housing approx. 33 kg (73 lbs)
Without housing approx. 25 kg (55 lbs)

Materials Housing: Stainless steel 1.4301 (AISI 304) or


glass-fibre reinforced carbon(GFR)
Front windows: Polycarbonate®
Endless hose: C-Flex®, Norprene®
Pump hose: Tygon®, Viton®
Valves: Tygon®, silicone
Connecting the sample line One channel version
Collecting vessel (at analyzer, with or without level measurement)
Connection hose ID 3.2 mm (0.13 inch)

Customer collecting vessel


Connection hose ID 1.6 mm (0.06 inch)
Max. distance from collecting vessel to analyzer 1 m (3.3 ft)
Max. height difference from collecting vessel to analyzer 0.5 m (1.6 ft)

Two channel version


• Depending on the ordered version, one or two collecting vessels (with or without level measurement) are
included in the scope of delivery.
• Level measurement is only possible for one channel.

10 Endress+Hauser
Stamolys CA71PH

• Only one collecting vessel can be mounted at the housing. The second is to be placed nearby the analyzer.

Sample outlet Connection Hose ID 6.4 mm (0.25 inch)


– Max. length of closed loop: 1 m (3.28 ft)
– Open outlet downgrade installed
– No combination of several devices to a closed-loop system
Min. volume per measurement 20 ml (0.68 fl.oz.)

Human Interface
Display and 1 2
operating elements

M CE E K RS 232

4 3
a0001679

Display and operating elements


1 LED (measured value)
2 LC display (measured value and status)
3 Serial interface RS 232
4 Operating keys and control LEDs

Endress+Hauser 11
Stamolys CA71PH

Ordering information
Product structure Measuring range
A Measuring range 0.05 to 2.5 mg/l PO4-P (blue)
B Measuring range 0.5 to 20 mg/l PO4-P (yellow)
D Measuring range 0.5 to 50 mg/l PO4-P (yellow)
E Measuring range 0.05 to 10 mg/l PO4-P (blue)
Y Special version acc. to customer’s specification

Sample transfer
1 Sample transfer from one measuring point (one-channel version)
2 Sample transfer from two measuring points (two-channel version)

Power supply
0 230 V AC / 50 Hz
1 115 V AC / 60 Hz
2 115 V AC / 50 Hz
3 230 V AC / 60 Hz

Collecting vessel for up to 3 analysers


A Without collecting vessel
B With collecting vessel without level measurement
C With collecting vessel with level measurement (one-channel version only)
D With two collecting vessels without level measurement (two-channel version)

Housing version
1 Without housing
2 With GFK housing
3 With stainless steel 1.4301 (AISI 304) housing

Communication
A 0/4 to 20 mA, RS 232

Additional equipment
1 Quality certificate
2 Quality certificate + set of inactive reagents PH-A+E
3 Quality certificate + 3 sets of inactive reagents PH-A+E
4 Quality certificate + set of inactive reagents PH-B+D
5 Quality certificate + 3 sets of inactive reagents PH-B+D

CA71PH - complete order code

Scope of delivery The scope of delivery comprises:


• an analyzer with mains plug
• a cleaning injector
• a tin of silicone spray
• a Norprene hose, length 2.5 m (8.2 ft), ID 1.6 mm (0.06 inch)
• a C-flex hose, length 2.5 m (8.2 ft), ID 6.4 mm (0.25 inch)
• a C-flex hose, length 2.5 m (8.2 ft), ID 3.2 mm (0.13 inch)
• two hose fittings of each size:
– 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch)
– 1.6 mm x 3.2 mm (0.06 inch x 0.13 inch)
– 6.4 mm x 3.2 mm (0.25 inch x 0.13 inch)
• two T-hose fittings of each size:
– 1.6 mm x 1.6 mm x 1.6 mm (0.06 inch x 0.06 inch x 0.06 inch)
– 3.2 mm x 3.2 mm x 3.2 mm (0.13 inch x 0.13 inch x 0.13 inch)
• an interference suppressor for the current output
• 4 edge covers
• a quality certificate
• Operating Instructions (English).
Note!
Please, order reagents separately with analyzer version CA71XX-XXXXXX1.
With all other versions, inactive reagents are included in the scope of delivery. You have to mix the reagents
before using them. Please, read the instructions attached to the reagents.

12 Endress+Hauser
Stamolys CA71PH

Certificates and approvals


4 approval Declaration of conformity
The product meets the requirements of the harmonized European standards. It thus complies with the legal
requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.

Test reports Quality certificate


Depending on the order code, you receive a quality certificate.
With the certificate the manufacturer confirms compliance with all technical regulations and the successful
individual testing of your product.

Accessories
Reagents and • Reagent set active, 1 l of each reagent PH1+PH2 (blue); order no. CAY240-V10AAE
standard solutions • Reagent set inactive, 1 l of each reagent PH1+PH2 (blue); order no. CAY240-V10AAH
• Reagent active PH1, 1 l (yellow); order no. CAY243-V10AAE
• Cleaning agent 1 l; order no. CAY241-V10AAE
• Standard solution 1.0 mg/l PO4 - P; order no. CAY242-V10C01AAE
• Standard solution 1.5 mg/l PO4 - P; order no. CAY242-V10C03AAE
• Standard solution 2.0 mg/l PO4 - P; order no. CAY242-V10C02AAE
• Standard solution 5 mg/l PO4 - P; order no. CAY242-V10C05AAE
• Standard solution 10 mg/l PO4 - P; order no. CAY242-V10C10AAE
• Standard solution 15 mg/l PO4 - P; order no. CAY242-V10C15AAE
• Standard solution 20 mg/l PO4 - P; order no. CAY242-V10C20AAE
• Standard solution 25 mg/l PO4 - P; order no. CAY242-V10C25AAE
• Standard solution 30 mg/l PO4 - P; order no. CAY242-V10C30AAE
• Standard solution 40 mg/l PO4 - P; order no. CAY242-V10C40AAE
• Standard solution 50 mg/l PO4 - P; order no. CAY242-V10C50AAE

Cleaner for hoses • Cleaning agent, alkaline, 100 ml (3.4 fl.oz.); order no. CAY746-V01AAE
• Cleaning agent, acidic, 100 ml (3.4 fl.oz.); order no. CAY747-V01AAE

Collecting vessel – for sampling from pressurised systems


– results in an unpressurised continuous sample stream
• Collecting vessel without level measurement; order no. 51512088
• Collecting vessel with level measurement (conductive); order no. 51512089

Maintenance kit • Maintenance kit CAV 740:


– 1 set pump hoses yellow/blue
– 1 set pump hoses black/black
– 1 set hose connectors per hose set
order no. CAV 740-1A

Additional accessories • Interference suppressor for control, power and signal lines
order no. 51512800
• Silicon spray
order no. 51504155
• Valve set, 2 pieces, for two-channel version
order no. 51512234
• Upgrade kit for upgrading from one-channel to two-channel version
order no. 51512640

Endress+Hauser 13
Stamolys CA71PH

Documentation
• Technical Information Stamoclean CAT430, TI 338C/07/en
• Technical Information Stamoclean CAT411, TI 349C/07/en
• Technical Information Stamoclean CAT221, TI 384C/07/en

14 Endress+Hauser
Stamolys CA71PH

Endress+Hauser 15
Instruments International

Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland

Tel.+41 61 715 81 00
Fax+41 61 715 25 00
www.endress.com
info@ii.endress.com

TI356C/07/en/05.09
71095964
FM+SGML 6.0
DATA SHEET

ULTRATURB plus sc

Turbidity
Sensor

Features and Benefits


Wide Measurement Range – 0 to 1000 FNU (NTU)
The Hach ULTRATURB plus sc Sensor measures turbidity
from 0 to 1000 FNU (NTU) for a wide variety of low to
medium range turbidity applications. Use it to effectively
monitor conventional filtered water processes, effluent water,
distribution systems, and elevated tanks where small-sized
particulates or air entrapment may be an issue.

Self-cleaning Sample Chamber Option


The ULTRATURB sensor is available with an automatically
The Hach ULTRATURB plus sc Sensor is a precisely measures DW
self-cleaning sample chamber that guarantees stable
turbidity of ultra-clear to medium turbid media for optimum
measured values. The silicon wiper blades inside the sample
filtration control in municipal and industrial water processing
chamber are held in place with a magnetic coupling wheel
to further keep the sample chamber intact.
plants. Data display and processing of measured data WW
is performed by Hach sc Digital Controllers.

Seawater Version PW
The new ULTRATURB plus Seawater sc is resistant to
salt concentrations up to 65 g/L.
IW
Ratio Methodology Detects Sample Chamber
Fouling Dry Verification Modules
The ULTRATURB sensor uses ratio methodology with
The Calibration/Verification
detection at 90° and 180° to minimize reflections in the
Module (CVM) is an optical
measuring cell and on the windows. In this way, fouling
standard for rapid and
of the sample chamber is detected earlier than with a
straightforward checking
single detection system.
of the ULTRATURB
sensor. It facilitates
Compatible with Hach Multi-Sensor, Multi- problem-free routine
Parameter Digital Controller checks and confirmation
The ULTRATURB sensor can be used with Hach’s sc Digital of measurements without
Controllers. The sc200 controller accepts up to two sensors. the need to mix liquid
The sc1000 accepts up to eight sensors. Multiple controllers calibration standards. CVM
can be networked to accommodate many more sensors and modules are available in
parameters, reducing the cost per measuring point. Just different turbidity ranges from
plug in any Hach “plug and play” digital sensor and it’s 0.6 to 25 FNU (NTU).
ready to use without software configuration. “Plug and play”
connectivity means there’s no complicated wiring or set up.
Network the ULTRATURB plus sensor with any of Hach’s
digital sensors for measuring dissolved oxygen, ORP, Complies with ISO 7027
conductivity, and many other parameters. The Hach ULTRATURB plus sc Sensor meets the ISO 7027
turbidity requirements for on-line process turbidity.

DW = drinking water WW = wastewater municipal PW = pure water / power


IW = industrial water E = environmental C = collections FB = food and beverage
2

Specifications*
Measurement Technique Verification Materials
90° scattered light in accordance with StablCal or dry standard CVM module Window: quartz
DIN EN ISO 7027 infrared light 860 nm
Ambient Temperature Measurement chamber: Noryl, GFN2
Range Sensor: 36 to 104°F (2 to 40°C) Wiper blade: silicon
0.0001 to 1000 FNU, freely programmable
(0.0001 to 250 EBC = 2500 ppm SiO2) Display unit: 14 to 104°F (-10 to 40°C) Wiper shaft: stainless steel 1.4571
(Note: 1 FNU is equivalent to 1 NTU) Housing: ASA
Sample Temperature
Resolution 122°F (50°C) maximum
Enclosure Rating
0.0001 to 0.9999 / 1.000 to 9.999 / IP 65
Sample Flow
10.00 to 99.99 / 100 to 1000 FNU (NTU)
0.2 to 1 L/min
Dimensions
Precision
Sample Pressure 9.9 x 9.4 x 4.3 in. (250 x 240 x 110 mm)
±0.008 FNU ±1% of reading
(0 to 10 FNU) 87 PSI at 68°F (6 bar at 20°C)
Weight
Sample Connection Approx. 3.3 lbs. (1.5 kg)
Repeatability
±0.003 FNU ±0.5% of reading 13 mm I.D. hose or fixed connection
with with G+F system parts Display Unit
(0 to 2 FNU) Hach sc200 Digital Controller or Hach
Automatic Cleaning sc1000 Universal Controller
Variation Coefficient
1% in accordance with DIN 38402 Wiper

Response Time
1 to 60 seconds (programmable)

*Specifications subject to change without notice.

Engineering Specifications
1. The sensor shall continuously measure turbidity in water 8. The transmitter enclosure shall be rated at IP 65.
using detectors at 90 and 180 degrees.
9. The sensor shall be model ULTRATURB plus sc Sensor
2. The measurement technology shall use infrared pulse manufactured by Hach Company.
scattered light process according DIN EN ISO 7027.
Optional:
3. The measuring range shall be from 0.0001 to 1000 NTU.
1. The sensor shall be equipped with a self-cleaning sample
4. The precision shall be ±0.5% or ±0.008 NTU of the chamber that uses a silicon wiper that is held in place
measured value. magnetically.
5. The response time shall be approximately 1 to 60 seconds. 2. The sensor shall include a calibration/verification module
with a factory-certified calibration value for periodic
6. The flow rate of sample shall be 0.2 to 1 L/minute. verification of calibration.
7. The verification of calibration for the sensor shall be by
StablCal or dry standard CVM module.
3

Dimensions
The sensor should be installed in an accessible location. It can be mounted on a flat, vertical surface (such as a panel, stand,
etc.). It should allow for access for any checking or maintenance. Sample flow should meet the specifications above.
4

At Hach, it’s about learning


Ordering Information from our customers and
providing the right answers.
ULTRATURB plus sc Sensor It’s more than ensuring the
LPV415.52.11002 Sensor with 0.35 m cable quality of water—it’s about
ensuring the quality of life.
LPV415.52.21002 Sensor with 5 m cable
When it comes to the things
LPV415.52.31002 Sensor with 10 m cable
that touch our lives...
LPV415.52.41002 Sensor with 15 m cable
LPV415.52.51002 Sensor with 20 m cable Keep it pure.
LPV415.52.61002 Sensor with 30 m cable
Make it simple.
LPV415.52.71002 Sensor with 50 m cable
Be right.
ULTRATURB plus sc Sensor with Auto-clean
LPV415.52.10002 Sensor with 0.35 m cable and Auto-clean system
LPV415.52.20002 Sensor with 5 m cable and Auto-clean system
For current price information,
LPV415.52.30002 Sensor with 10 m cable and Auto-clean system
technical support, and ordering
LPV415.52.40002 Sensor with 15 m cable and Auto-clean system assistance, contact the Hach
LPV415.52.50002 Sensor with 20 m cable and Auto-clean system office or distributor serving
LPV415.52.60002 Sensor with 30 m cable and Auto-clean system your area.
LPV415.52.70002 Sensor with 50 m cable and Auto-clean system
LPV415.99.22002 Sensor Seawater sc with 5 m cable
LPV415.99.32002 Sensor Seawater sc with 10 m cable In the United States, contact:
LPV415.99.42002 Sensor Seawater sc with 15 m cable
LPV415.99.52002 Sensor Seawater with 20 m cable HACH COMPANY World Headquarters
LPV415.99.62002 Sensor Seawater sc with 30 m cable P.O. Box 389
LPV415.99.72002 Sensor Seawater sc with 50 m cable Loveland, Colorado 80539-0389
LPV415.99.82002 Sensor Seawater sc with 1 m cable U.S.A.
Telephone: 800-227-4224
sc200 / sc1000 Controller Options Fax: 970-669-2932
E-mail: orders@hach.com
sc200 for Hach Digital Sensors www.hach.com
LXV404.99.00552 sc200 controller, 2 channel, digital
LXV404.99.00502 sc200 controller, 1 channel, digital
LXV404.99.00542 sc200 controller, 2 channel, digital & mA input U.S. exporters and customers in Canada,
sc1000 for Hach Digital Sensors Latin America, sub-Saharan Africa, Asia,
LXV400.99.1R572 sc1000 Probe Module; 4 sensors - analog and Australia/New Zealand, contact:
LXV400.99.1B572 sc1000 Probe Module; 4 sensors - RS485 HACH COMPANY World Headquarters
LXV400.99.1F572 sc1000 Probe Module; 4 sensors - PROFIBUS DP P.O. Box 389
LXV400.99.1R572 sc1000 Probe Module; 6 sensors - analog Loveland, Colorado 80539-0389
U.S.A.
Dry/Secondary Verification Modules Telephone: 970-669-3050
LZV414.00.00000 CVM Module; 0.6 NTU Fax: 970-461-3939
LZV414.00.10000 CVM Module; 1.5 NTU E-mail: intl@hach.com
www.hach.com
LZV414.00.20000 CVM Module; 6 NTU
LZV414.00.30000 CVM Module; 15 NTU
LZV414.00.40000 CVM Module; 25 NTU In Europe, the Middle East, and
Mediterranean Africa, contact:
Spare Parts
HACH LANGE GmbH
LZV275 Wiper blades; set of 4 Willstätterstraße 11
DOC023.52.03231 User Manual D-40549 Düsseldorf
GERMANY
Calibration Tools Tel: +49 (0) 211 5288-0
246142 Formazin; 4000 NTU, 500 mL Fax: +49 (0) 211 5288-143
LZV451 Calibration/Cleaning Kit E-mail: info@hach-lange.de
LZV325 Filters with Apparatus; for zero calibration (0.2 micron) www.hach-lange.com

Lit. No. 2620 Rev 1


J10 Printed in U.S.A.
©Hach Company, 2010. All rights reserved.
In the interest of improving and updating its equipment, Hach Company reserves the right to alter specifications to equipment at any time.
DATA SHEET

CL17 Chlorine Analyzer

Chlorine
Overview
With over 60 years of industry leadership, Hach provides
the best products and application knowledge for chlorine
monitoring. Our portfolio includes the market- leading CL17
chlorine analyzer and the CLF10 sc or CLT10 sc reagentless
chlorine analyzers, as well as Hach laboratory colorimeters,
spectrophotometers and chemistries. Let Hach guide you to
the best disinfection solution.

Features and Benefits


Accurate Results
The Hach CL17 Chlorine Analyzer uses colorimetric DPD The Hach CL17 Chlorine Analyzer uses fast, reliable,
chemistry to monitor water continuously for free or total and economical DPD chemistry for up to 30 days of
residual chlorine which is the same method (Standard unattended operation. No electrodes and minimal use DW
Method 4500-Cl G) as used for grab samples. This of reagents means low operating costs.
analysis method is not affected by changes in sample pH,
temperature, chlorine concentration (within the measurement WW
range), pressure or flow, thus offering more accuracy than Method of Analysis
other methods in the market today. Free Residual Chlorine—The CL17 Chlorine Analyzer
uses an aqueous buffered colorimetric indicator— PW
Simple, Predictable Maintenance N,N-diethylphenylenediamine (DPD)—to determine levels
of chlorine. DPD turns a magenta color in response to the
Monthly routine maintenance for the CL17 can be performed
amount of free residual chlorine (as hypochlorous acid or IW
in 15 minutes and includes changing reagents and cleaning
hypochlorite ion). The reaction takes place at a buffered
the colorimetric cell. No special tools are required. Under
pH of 6.3 to 6.6.
typical use, the CL17 will operate unattended for 30 days.
Total Residual Chlorine—To measures total residual
Re-Calibration Not Necessary chlorine (free residual chlorine plus mono-, di- and
Calibration of the CL17 with a chlorine standard or against a trichloramines) an additional reagent is used. By adding
reference analysis is possible; however, it is neither potassium iodide to the sample, chloramines in the sample
necessary nor recommended due to its factory- established oxidize iodide to iodine, which then oxidizes the DPD
embedded calibration curve. indicator to the magenta color at a buffered pH of 5.1.

Manual or Automatic Feed Pump Control Applications


Both an alarm for manual chlorine feed control and a 4-20 Drinking Water—The CL17 Chlorine Analyzer can be used
mA output for automatic control of chlorine feed pumps is in finished water where residual chlorine levels must be
available in the CL17 Analyzer. When set for manual control, maintained during distribution. It can also be used to
the instrument notifies the operator of out-of-limit chlorine monitor raw water to facilitate preoxidation, disinfection,
levels for manual intervention in the disinfection process as and control of taste and odor problems.
needed. For automatic control of chlorine feed pumps, the Wastewater—Large tubing and fittings and a Self-Cleaning
CL17 can use one of two methods: By-Pass Y-Strainer Kit for the CL17 ensures continuous
1. On-Off Control turns the feed pumps on when chlorine operation without clogging.
levels fall too low for disinfection and off when levels rise Industrial—The CL17 can be used to monitor disinfection
above a pre-set limit. and prevent biological build-up in applications that involve
2. Proportional Control adjusts the amount of chlorine in chemical or industrial processes (feed water), heating and
proportion to the strength of the CL17’s output signal. cooling water, or food and beverage applications. It is also
useful in systems that use reverse osmosis, to protect
expensive cellulose acetate membranes.

DW = drinking water WW = wastewater municipal PW = pure water / power


IW = industrial water E = environmental C = collections FB = food and beverage
2

Specifications*
Range Interferences Certification
0 to 5 mg/L free or total residual chlorine Other oxidizing agents such as bromide, CE approved
chlorine dioxide, permanganate and ETL listed to UL 1262
Accuracy ozone will cause a positive interference. ETL certified to CSA 22.2 No. 142
±5% or ±0.035 mg/L as CL2, whichever Hexavalent chromium will cause a
is greater positive interference: Enclosure
1 mg/L Cr6+ = approximately 0.02 mg/L ABS plastic, two clear polycarbonate
Precision as Cl2. windows,
±5% or 0.005 mg/L as CL2, whichever is Hardness must not exceed 1,000 mg/L IP62-rated with the gasketed door
greater as CaCO3. latched

Minimum Detection Limit Recorder Outputs Mounting


0.035 mg/L One 4-20 mA with an output span Wall mount
programmable over any portion of the
Cycle Time 0 to 5 mg/L range, 130 V isolation from Display
2.5 minutes earth ground, 500 ohm maximum LCD, 3-digit measurement readout
and six-character alphanumeric scrolling
One isolated recorder output, 4–20 mA
Inlet Pressure to Instrument text line
(can be adjusted to 0–20 mA),
1 to 5 psig (1.5 psig is optimum) recommended load impedance 3.6 to Light Source
500 ohms. Optional AquaTrend®
Inlet Pressure to Sample Conditioning Class 1 LED (light emitting diode) with
Network interface
1.5 to 75 psig a peak wavelength of 520 nm;
Alarm Relay Outputs 50,000 hours estimated minimum life
Air Purge (optional)
Two alarms selectable for sample Power
0.1 cfm instrument quality air at 20 psig concentration alarm, analyzer system
maximum warning, or analyzer system shut-down 100 to 115/230 Vac, 50/60 Hz
alarm. Each is equipped with an SPDT (switch selectable), 95 VA maximum,
Sample Flow relay with contacts rated for 5A resistive 2.5 Amp fuse
200 to 500 mL per minute minimum load at 230 Vac.
required Dimensions
Sample Inlet Connection 34.3 x 41.9 x 19.1 cm
Sample Temperature 1/4-inch OD polyethylene tube, quick- (13.5 x 16.5 x 7.5 in.)
5 to 40°C (41 to 104°F) disconnect fitting
Shipping Weight
Operating Temperature Drain Connection 11.3 kg (25 lbs.)
5 to 40°C (41 to 104°F) 1/2-inch ID flexible hose, hose barb
Operating Humidity Air Purge (optional)
90% at 40°C (90% at 104°F) maximum 1/4-inch OD tube, quick-disconnect
fitting, 0.1 cfm instrument quality air at
20 psig maximum

*Specifications subject to change without notice.

Sample Inlet Sample Valve Principle of Operation


The CL17 Chlorine Analyzer has three operating components:
(Under Pressure)

Light Source To Waste


• A linear peristaltic pump to precisely control the volume of
incoming samples and reagents.
• A colorimeter with seal-free, solid-state mixing system that
includes a self-cleaning stir bar.
• One-month supply of reagents (indicator and buffer)
Sample Cell Photo Cell A zero reference point is established with the first sample in
Peristaltic
Pump
Magnetic the cycle by measuring blank absorbance. (This compensates
Stirrer
Buffer Indicator for the sample’s color intensity and turbidity before the
chlorine measurement is made.) Then, indicator and buffer
reagents are added to the sample while a magnetic stirrer
mixes the solution and the sample changes color. A compact
colorimeter then measures the light transmitted through the
sample. The measured color intensity is compared to a
reference standard. Finally, the sample cell is flushed with new
sample so that the cycle can repeat itself every 2.5 minutes.
3

Engineering Dimensions
Specifications The CL17 is designed to be wall-mounted with four 1/4-inch screws. Adequate
clearance must be left at the sides and bottom of the case for plumbing and
1. The chlorine analyzer shall employ a electrical connections. The sample inlet connection is 1/4-inch quick-disconnect
DPD colorimetric method of fitting and the drain connection is 1/2-inch I.D. flexible hose. Electrical connections
measurement using DPD indicator are inside the instrument case. Holes for three 1/2-inch conduit fittings are provided.
and a buffer solution.
2. The analyzer shall be capable of SIDE VIEW FRONT VIEW
measuring free or total residual 4.02" Air Purge Quick Connect
(102.17 mm) .25 O.D. Polyethylene Tubing
chlorine by changing the indicator 1.56" (Cat. No. 47438-00) Four 1/4" Mounting Bolts
(39.68 mm)
and buffer solutions.

(67.55 mm)
(103.88 mm)
3. A measurement shall be taken every

2.66"
Relay/Alarm

5.61" (142.48 mm)

4.09"
Contact
2.5 minutes and results displayed by Connections
a three digit LCD readout in the Chlorine

range of 0 to 5 mg/L. mg/L


µg/L

4. The analyzer shall be designed for Network and


Alarm

4-20 mA

17.88" (454.05 mm) Mounting Pattern


30-days unattended operation and Connections
use only 473 mL of each reagent per
month. Power
Connections
5. The analyzer shall operate with an
LED light source with a peak
wavelength of 510 nm.
6. The instrument shall measure a
sample blank before each sample
measurement to provide automatic
zero reference to compensate for
sample color and turbidity and
changes in light intensity due to
voltage fluctuations or light
source aging.
1.06" (26.99 mm)
7. The instrument shall provide a
4.38"
minimum detection limit of (111.12 mm)
7.50" (190.50 mm)
Mounting Pattern
0.035 mg/L or better, precision
better than ± 5% or 0.005 mg/L
as Cl2, and accuracy better than BOTTOM VIEW
± 5% or 0.035 mg/L as Cl2.
8. The analyzer shall be
microprocessor-controlled and
provide a 4-20 mA recorder output 19.95"
(506.78 mm)
as well as 2 alarms.
15.21" (386.30 mm)
9. Each alarm shall be user-selectable
for sample concentration alarms R 14.89
(high or low), analyzer system (378.18 mm)

warnings, or analyzer system shut-


down alarms. Overflow Drain
Recommended
10. The sample concentration alarms .50 I.D. Tubing
shall be fully adjustable through the
entire range.
11. The system warning shall activate for 7.06"
(179.39 mm)
minor variations in analyzer
performance.
12. A system alarm shall activate for 0.06" (1.59 mm)
Clearance
major variations in analyzer For Vents

performance and it shall shut down


the analyzer until corrective action is .50 I.D. Drain Tube Sample Sample Inlet Quick Connect
.25 O.D. Polyethylene Tubing
taken. (Cat. No. 47438-00)

13. The microprocessor shall provide


self-diagnostic functions accessible
16. Recorder span minimum and 18. The enclosure shall have two clear
through an alphanumeric, menu-
maximum values shall be operator polycarbonate windows for viewing
driven keyboard.
programmable at the menu-driven the measurement readout and
14. Two SPDT normally open/normally keypad over the entire operating reagent levels.
closed dry contact relays rated at 5A range. 19. Power requirements shall be
resistive load at 230 Vac shall be
17. The chlorine analyzer shall be 100-115/230 Vac, 50/60 Hz,
provided.
housed in an IP-62 rated, ABS switch selectable, 95 VA maximum.
15. Recorder outputs shall be a plastic enclosure designed for wall 20. The instrument shall be the Model
4-20 mA. mounting. CL17 Chlorine Analyzer,
manufactured by Hach Company.
4

At Hach, it’s about learning


Ordering Information from our customers and
providing the right answers.
Hach CL17 Chlorine Analyzers are shipped with a one-month supply of reagents, It’s more than ensuring the
maintenance kit, installation kit, and manual. (The power cord is ordered separately.) quality of water—it’s about
ensuring the quality of life.
5440001 Model CL17 Free Residual Chlorine Analyzer
When it comes to the things
5440002 Model CL17 Total Residual Chlorine Analyzer that touch our lives...
5440003 Model CL17 Free Residual Chlorine Analyzer Keep it pure.
with AquaTrend® Network Capability
Make it simple.
5440004 Model CL17 Total Residual Chlorine Analyzer
with AquaTrend® Network Capability Be right.

Accessories
5448800 Power Cord, 125V, 10A, 1.83 m (6 ft.)
For current price information,
5448900 Power Cord, 230V, 10A, 1.83 m (6 ft.), continental European plug technical support, and ordering
assistance, contact the Hach
5444300 Maintenance Kit, 1 year, includes tubing, caps, funnel, and fittings
office or distributor serving
5444301 Maintenance Kit with preassembled tubing, 1 year, includes tubing, caps, your area.
funnel, and fittings
4643600 Flow Meter with 1/4-inch OD tubing In the United States, contact:
4427800 Serial I/O Kit HACH COMPANY World Headquarters
5449000 CL17 CAL/Verification Kit P.O. Box 389
Loveland, Colorado 80539-0389
Reagents U.S.A.
Telephone: 800-227-4224
Reagent sets include all three of the required reagents [DPD indicator powder (added
to indicator solution), indicator and buffer solutions] is sufficient for a 30-day Fax: 970-669-2932
operating period. E-mail: orders@hach.com
www.hach.com
2556900 Reagent Set, CL17 free chlorine
2557000 Reagent Set, CL17 total chlorine U.S. exporters and customers in Canada,
2297255 DPD Indicator Powder (free and total) Latin America, sub-Saharan Africa, Asia,
and Australia/New Zealand, contact:
2314011 Free Chlorine Indicator Solution, 473 mL
HACH COMPANY World Headquarters
2314111 Free Chlorine Buffer Solution, 473 mL
P.O. Box 389
2263411 Total Chlorine Indicator Solution, 473 mL Loveland, Colorado 80539-0389
U.S.A.
2263511 Total Chlorine Buffer Solution, 473 mL
Telephone: 970-669-3050
2835900 Calibration Refill Kit Fax: 970-461-3939
E-mail: intl@hach.com
www.hach.com

In Europe, the Middle East, and


Mediterranean Africa, contact:
HACH LANGE GmbH
Willstätterstraße 11
D-40549 Düsseldorf
GERMANY
Tel: +49 (0) 211 5288-0
Fax: +49 (0) 211 5288-143
E-mail: info@hach-lange.de
www.hach-lange.com

Lit. No. 1626 Rev 2


B11 Printed in U.S.A.
©Hach Company, 2011. All rights reserved.
In the interest of improving and updating its equipment, Hach Company reserves the right to alter specifications to equipment at any time.
TD-4100 XD H Y D R O C A R B O N S I N WAT E R
Continuous On-Line Monitor

Continuous on-line monitoring with the TD-4100 XD provides the most responsive feedback loop
for measuring hydrocarbons in water. Continuous monitoring is reliable, effective, and
recognized for its ability to improve process management for treating, discharging and
detecting hydrocarbons in water. Compared to laboratory grab sample analysis, on-line
monitor-ing provides cost effective, continuous, remote, operator unattended measurement of
hydrocarbons in water.

N O N - C O N TA C T, N O N - H I G H S E N S I T I V I T YA N D
FOULING FLOW CELL SELECTIVITY
SENSITIVE
The TD-4100 XD does not have a glass flow
cell. Hydrocarbons are detected in a stream BTEX, gasoline, diesel, jet fuel, crude oil, aromatic
of water falling through an open chamber; solvents and refined petroleum products are detected by
the water does not contact, dirty or foul the the TD-4100 XD from low ppb (mg/L) to high ppm
optical windows. A proprietary Air Curtain (mg/L). For example, the TD-4100 XD can detect 1 ppb
system keeps optical windows fog-free in hot of diesel fuel in water free of interfering compounds.
water applications.

SELECTIVE
LOW MAINTENANCE
The TD-4100 XD continuously measures fluorescent
The instrument is stable within 10% over 6 hydrocarbons in water. Fluorescence occurs when a
months. Other than sample line molecule absorbs light energy and emits light energy at
maintenance, routine maintenance involves longer wavelengths.
changing a lamp twice a year. System
checks are easily performed with the
EFFECTIVE MONITORING IN
CheckPOINT® solid standard.
D I R T Y WAT E R

Fluorescence technology makes the TD -4100 XD


D I R E C T, C O N T I N U O U S
resistant to interferences from turbid or dirty water that
MONITORING
impact on-line UV, IR absorption, or light scatter
The TD-4100 XD monitors a flowing water instruments. Most substances absorb light, but very few
stream continuously. No chemicals, no pre- fluoresce; if a substance does not fluoresce at the
treatment, no mechanical manipulation or specific wavelengths for the monitored hydrocarbon, it
mixing of the sample is required to monitor will not interfere.
hydrocarbons in water.

O P E R AT O R F R I E N D LY
A C C U R AT E
The TD-4100 XD is designed for easy operation. Simple
The TD-4100 XD directly measures on-board software controls alarms, 4-20 mA output,
fluorescing hydrocarbons in water with diagnostics and calibration. “Turner Designs
accuracies that consistently correlate to Hydrocarbon Instruments is the recognized expert for oil
regulated laboratory methods in most cases. in water monitoring technology.”
TD-4100 XD

T Y P I C A L S P E C I F I C A T I O N S
Envelope Dimensions: 55.5 cm W x 63.5 cm D x 241 cm H [22" W x 25" D x 95" H]

Weight: 82 Kg [180 lbs] plus accessories

Power Requirements: 100-240 VAC, 50/60 Hz ± 10%, 200 W, 1 ph


or 21-56 VDC, 200 W startup, 50-60 W operation (optional)

Inlet Plumbing Requirements: ½" MNPT (standard) or ½" tube

Outlet Plumbing Requirements: 1-1/2" MNPT

Inlet Sample Flowrate: 7.5 -11.5 L/min [2-3 US gallons/min], optional sample pump

Inlet Sample Pressure: 34 -136 kPag [5-20 psig]

Outlet Sample Pressure: Atmospheric (standard) or optional sample return pump

Sample Temperature: 0 -88oC [32-190oF] standard, higher temperatures optional

Ambient Temperature: 0 - 49oC [0-120oF] standard, optional cold and high


temperature protection

Operational Principle: Fluorescence

Detection Range: 1 ppb - 1000 ppm depending upon target hydrocarbon and
water quality

Stability: 10 % or better over 6 months

Response Time: < 10 seconds continuous real-time response

Calibration: Multiple - point or un- calibrated

Alarms: Baseline, early warning, high alarm, system-function, local


display and audible tone

Alarm Outputs: Two user-settable, independently-protected, solid -state AC relay


standard or optional dry contact relays

Analog Output: 4-20 mA or 0 -20 mA, isolated, powered (standard), other


TURNER DESIGNS
Hydrocarbon Instruments
protocol options available

Diagnostics: System failure reports to relay and local display 2023 N. Gateway Blvd., Suite 101
Fresno, California 93727 USA
Security: Two level password protected, lockable cabinet
TEL : +1 559 253 1414
Electronics Cabinet: 316 stainless steel, NEMA 4X, IP 66 FAX: +1 559 253 1090
Air Purge Options: ATEX Zone1, ATEX Zone 2, Class1 Division 1, Class1 sales@ oilinwatermonitors.com
Division 2, or non-hazardous environment purge www.oilinwatermonitors.com
Technical data sheet

SDI-2000
Silt Density Index

ƒ Fully complies with the ASTM D4189-07


ƒ CE certified
ƒ Up to 4 sampling points
ƒ Up to 85 cycles (to complete one paper filter roll) without
operator interference
ƒ Several modes for operation and calculation per channel
ƒ Three built-in operational modes:
1. Manual (one-cycle only)
2. Auto (programmable time interval)
3. Remote (pre-started selected channel from remote station)

ƒ Two built-in calculation formula can program per channel:


1. Standard mode test for low SDI water
(less than 5.0 for 15minutes)
2. Auto Time mode test for HIGH SDI water
The SDI 2200 (Silt Density Index) is a full automatic, (greater than 5.0 for 15 minutes)
multi channels on line analyzers and monitoring systems. * Measure time 1-15 minutes
The SDI 2200 is a unique analyzer and designed for
continuous, unattended operation in desalination and Principal of operation
other water treatment plants.
Standard Test Mode

Silt Density Index (SDI) is measured by comparing the flow rate of a known
volume of water, where the water volume is kept under constant pressure.

The flow rate is measured twice; first time is when the water flows for
reference through a standard unused filter, second time is after the filter is
exposed for 15 min. or less.

The water sample flow is always kept under constant pressure.


The drop in flow rate is a direct measure of the silt build-up on the filter.

The pressure is kept constant by precisely regulating air pressure applied on


Standards
the water surface without the need of any mechanical device.
MABAT work to ISO 14001 and ISO 9001 standards,
and is a member of SEMI. Auto Time Test Mode

All products are manufactured according to SEMI In cases where the silt density index (SDI) reading is unstable and readings
guidelines and are CE marked. are higher than 5 (i.e. Plugging Factor is higher than 75%), during the 15
minutes check time, the analyzer can be programmed to work in Auto Time
Member of
Test Mode based on the following test method:

ƒ The plugging factor calculations will proceed and will be recorded at


intervals of one-minute elapsed
ƒ Calculations will stop once plugging factor 75% has been reached
ƒ New SDI readings will be displayed based on the last calculated factor
The exact check time has no valuable meaning (as in Standard Test Mode)
In general, the higher the SDI reading the more silt is in the water.
This document contains confidential information.
Copies or disclosure to third parties without the prior When plugging factor is calculated to be lower than 75%, the analyzer will
written agreement of MABAT is expressly prohibited. switch and calculate the test results based on the Standard Method.
Technical data sheet

SDI-2000
Silt Density Index

System types

The SDI format definition: SDI 220 N-A-E-T

N number of channels (1-4)


MABAT can offer a turn-key solution if required and A number of analog outputs (1-4)
have their own experienced team or approved local E Ethernet communication
contractors to ensure project installation and T temperature measurement
acceptance dead-lines are achieved.

MABAT have a range of products to suit most Options


applications and can also provide custom built ƒ Optional analog output (programmable range) per channel for
systems for non-standard requirements. Site visits remote control and monitoring purposes
can be arranged to ensure the end user’s exact
ƒ Booster pump - When supply is in low pressure, one booster
needs are understood and catered for.
pump support all channels
Please contact our offices to obtain advice and ƒ Temperature measurement and compensation (if necessary)
quotations.

For applications outside any system’s specification or Dimensions


capabilities please refer to our local representative or Height Width Depth
the MABAT Headquarters at Kiryat Gat. Cabinet 1,126 mm 525 mm 310 mm
Wall mounted
MABAT have more than 25 years experience
supplying systems to the Semiconductor and Solar
Cell manufacturing industries. Utilities needed

Input pressure 35 – 80 psi

Flow rate 11/2 l/min.


Air pressure 80 – 120 psi

Electrical 110 – 230 VAC at 4 Amp. Or 24 VDC

Filter paper type

Standard Millipore 0.45 micron roll paper

Length of roll 6 meters, up to 85 measurements per roll

Contact details Control System Communication


Kiryat-Gat, ISRAEL
Tel. + 972.8.660.1320
ƒ CPU ƒ RS488
US toll free + 1.866.908.6297 ƒ Color touch screen ƒ RS232
UK +44 (0) 1825 733503
info@mabat-systems.com
www.mabat-systems.com
Reliability

Uptime >99.9%
This document contains confidential information.
Repeatability less than 0.1 SDI units
Copies or disclosure to third parties without the prior
written agreement of MABAT is expressly prohibited.