PRODUCT DATA ............................................................................................................................................... 2
MIRROR DRF: FULL DIGITAL IMAGE ACQUISTION SYSTEM ......................................................................... 8
TOMORIX – TOMOSYNTHESIS (OPTIONAL) .................................................................................................. 13
FINANCIAL OFFER .......................................................................................................................................... 14
GENERAL CONDITIONS OF SALE .................................................................................................................. 17
1
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
PRODUCT DATA
GENERAL
MECHANICAL CHARACTERISTICS
2
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
PATIENT TABLETOP
250 x 70 cm.
Tabletop dimensions Width 27 mm
Useful are: 232x50,5 cm
Standard tabletop (white): carbon fiber covered with laminated
Filtration 0,7 mm Al @ 100 kV
Max patient weight 225 Kg (without limitation)
Tabletop Optional tabletop (black): carbon fiber
Filtration 0,4 mm Al @ 100 kV
Max patient weight 225 Kg (without limitation)
Tabletop side profiles of 6 mm, ready to attach some accessories.
Tabletop lateral movement ±15 cm
Tube column longitudinal excursion: 188 cm
Spot film device (center) excursion: 168,5 cm
The movement of both column and flat panel allows for patient
total scanning: 209 cm at adjustable speed up to 15 cm/sec,
Longitudinal excursion controlled through joystick.
Important:
The movement of the longitudinal tabletop is not necessary as the
complete exposure of the patient is guaranteed by the field of
movement of the column and the flat panel.
Tube angulations for oblique
± 40°with I.I. parallax correction
projections
Tabletop tilting range -90° to +90° continuously
X-ray tube assembly rotation range -180° to +180° motorized, control from console
SID From 105 to 180 cm continuously or with presets
ACCESSORIES
Removable footrest with surface 400x600 mm;
Standard Shoulder rest
Pair of ergonomic handlers
Paper roll bearer
LDC glass bearer
Compression band
Optional OB-GYN legs bearer
Lateral cassette holder for oblique projection
Hand safety stripes
Additional pedal (in examination room) for RAD/Fluoro
3
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
DYNAMIC CHARACTERISTICS
Tabletop rise time from lower to max height (horizontal position) 15 sec
Tabletop combined rotation from 0° to +90° 25 sec
Tabletop combined rotation from 0° to -90° 26 sec
Tabletop combined rotation from -90° to +90° 37 sec
Tabletop lateral displacement From 1cm/sec to 2,5 cm/sec
Lined translation (tube+receptor) along the horizontal axis 12 sec
Rotation only from 0° to +90° 16 sec
Focus to film extension from 105 to 180 cm 18 sec
TOMOGRAPHY
Planigraph with homothetic linear movement and electronic fulcrum
Type
calculation
Stratum max height 400 mm (physical limit 450 mm)
Increase / decrease Manual, 1 mm step
Automatic (auto step function) with step mm program and selectable
according to starting tomo angle
Speed 10° - 21° / sec. Adjustable
Preset: 8° - 15° - 20° - 30° - 40°
Tilting angles According to customer preference within max 80° with respect to the
chosen anatomical area and FFD
Up to 5 speeds can be chosen that represent a percentage of the max
speed: 3525 cm/sec (21°/sec)
For each tilting angles the speed can be decreased in 5 steps of approx. 10%
Tomo timings
per step.
A tomography at 40° and FFD 105cm at the maximum speed will take
approx. 2,2 sec.
Direction Bi-directional in each position of table and FPD/column group
Sequence program with outward and inward emission up to the limits set
Sequence tomography
by the operator or to the reaching of preset limit; stratum, area, etc.
Tomography is achieved at different receptor position according to the
Receptor movement range
angle, to the FFD and the selected stratum.
4
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
5
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
COLLIMATOR
Model R225 ACS inclusive of retractable tape meter
Functioning Manual with push buttons and knob
Automatic, microprocessor controlled and CAN-BUS interface
Field Square & rectangular
Inherent filtration 2 mm Al eq.
Square field covering at 1 mt FFD 430x430 mm
Field light indicator > 160 lux
Light indicator accuracy < 1% FFD
Single: direct line projection
Laser pointer
Double: cross projection
OPTIONS
Laser pointer Double (cross) line projection
Disk support with automatic filter exchange device; it can be manual
of automatically controlled by CAN-BUS
Additional filtering (optional) 1mm Al + 0.1mm Cu
1mm Al + 0.2mm Cu
2mm Al
RO314 Integrated camera for patient view
RO320 Collimator motorized rotation
REMOTE CONSOLE
19” Control console Medical Grade Panel PC Touch Screen with all
commands.
Description N. 4 keyboards duplicating the most frequent commands: tube
angulation movements; FFD; tabletop tilting and elevation
movements; longitudinal and transversal displacement.
Dimensions: 19”
Resolution: 1600 x 1200 pixel
Brightness: 350 cd/m2
Touch screen characteristics
Number of colors: 32 bit
View angle: 170° H – 170° V
Aspect ratio: 4:3
Dedicated LAN connected to control CPU
Available connections Standard LAN for networking
N. 4 RS 232 ports
ADDITIONAL CONTROLS
N.2 additional membrane keyboards are located on the image receptor font and a X-ray tube front
6
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
ENDEAVOR RF
HIGH FREQUENCY GENERATOR WITH TOUCH SCREEN CONSOLE
Power 65 kW 80 kW
Tube protection Max load, anode thermal load, Max load, anode thermal load,
anode rotation, anode thermal anode rotation, anode thermal
safety, filament overcharge, max safety, filament overcharge, max
voltage protection voltage protection
RADIOGRAPHY
7
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
FLUOROSCOPY
Frequency 50/60 Hz
Weight 75 Kg
Origin Italy
The Mirror DRF is a full field direct to digital imaging system with flat panel detector. Choice is made
between two state of the art flat panels, namely Pixium RF 4343 and Pixium RF 4343 FL.
The Mirror DRF makes for fast examination and therefore high patient workflow, as well as offering cost
savings and excellent image quality at lower doses.
It covers a huge range of examinations in General Radiology studies like:
Muskoloskeletal Tmography
Swallowing Limphography
Chest Myelography
Genitourinary ERCP procedures
Gastrointestinal Long leg % spine stitching;
Interventional (Angiography)
8
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
DETECTOR SPECIFICATIONS
MODEL PIXIUM RF 4343 PIXIUM RF 4343 - FL
Technology Amorphous silicon - CSi scintillator
Detector Area 43x43 cm
Pixel size 148x148 µm
Matrix size 2880x2881 pixels 2880x2880 pixel
Pixel data 16 bits
X-ray exposure dose 0.2 – 3 µGy
X-ray saturation dose 85 µGy
Maximum linear dose 50 µGy
X-ray generator voltage 40-150 kV
DQE @ 0 pl/mm 65 %
MTF @ 1 pl/mm 63 %
MTF @ 2 pl/mm 32%
Number of models 15 (4 RAD, 2 DSA, 9 Fluoroscopy) 17 (5 RAD, 12 Fluoroscopy)
Overall size 518x508x81 mm 500x490x45 mm
Weight 25 kg 14 kg
Cooling Ambient air flow, no special cooling devices needed
OPERATIONAL MODE
Continuous Fluoroscopy
43x43 (960x960x16bits) 18 fps 16 fps
30x30 (1024x1024x16bits) 15 fps 12 fps
20x20 (673x673x16 bits) 30 fps 20 fps
15x15 (1024x1024x16 bits - 6 fps
Pulsed Fluoroscopy
43x43 (960x960x16bits) 15 fps 12 fps
30x30 (1024x1024x16bits) 15 fps 12 fps
20x20 (673x673x16 bits) 15 fps 12 fps
15x15 (1024x1024x16 bits - 6 fps
Radiography
43x43 cm (2880x2881x16 bits) 3 fps 2 fps
43x43 cm (1440x1441x16 bits) 8 fps 6 fps
9
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
Unlimited APR programs. More than 30 preset parameters for each anatomical study.
Anatomical presentation of the images for the best intuitive use;
Choice of the lowest X dose for each selected study;
Reduce examination time;
Better comfort for the patient and the operator.
Dose Saving
- Virtual collimator;
- Virtual scanning.
Digital Tomography
Combining the flat detector image quality with the remote control table ergonomics, the digital
Tomography becomes again very affective.
Images processing
10
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
Networking
STITCHING (OPTIONAL)
Images Stitching - Integrated procedures for leg & spine images stitching.
The stitching function, (usually used for spine and legs scan) is needed for the automatic
reconstruction of an X-ray image starting from a series of images acquired at fixed
frequency during the scanning of the patient.
The image is reconstructed, keeping all original pixels, and can be viewed on the
monitor, processed, printed or sent to the network.
As for standard acquisition, stitching is done giving the x-ray command from the
generator control panel – the system automatically generates the required exposures
(2,3,4) each time irradiating a different part of the patient.
After the exposure, the system automatically processes the acquired images and then
recomposes them creating a single image shown on the monitor after approx 30 seconds.
11
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
Windows XP or Win 7 / 32 bit, Win 7 /6 4 bit OS, intuitive icon, 3F mouse, keyboard
User
Multi language information: Italian, English, French, German, Spanish, other languages upon Ѵ
interface
request
Archive Patient data archive: work-list, studies to do, studies completed, studies documented (Print,
Ѵ
display store, etc)
Frame area : 1280x1024 pixels
Image area: 1024x1204 pixels, overwrite tools (patient data, image data, dose, symbols and
Operative graphics)
Ѵ
panel Icons area: pre-acquisition data selection, post-processing functions, images destination for
reporting, system status, exposure dose rate evaluation.
Thumbnails of the main 6 images/run acquired.
Control Desk: 19” LCD medical display, DICOM LUT, monochrome image, colours for icons
Ѵ
Display and graphic (live image display)
In room: one/two 19” LCD high brightness (1.500 cd/m2), medical display, DICOM LUT, native
Ѵ
monochrome (live and reference images)
12
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
slide
Main Controller
Control PC for the complete Rx diagnostic with interface & controls of: remote table,
collimator, generator, ione chamber, grid, DAM, PID, PU4343
PC Cabinet Ѵ
PID
PC for images processing. PC architecture: Windows XP OS, Pentium 4/3 GHz, 8 GB RAM, 250
GB HD (minimum configuration)
The tomosynthesis is a digital technique that allows the reconstruction of volumetric images starting from a finite
number of bidimensional projections taken at different tube angulations. This is given by the separation of the
acquisition process from the visualization one. Among the advantages with respect to CT scan technique, it is well
accepted by those patients affected by severe claustrophobia. Tomosynthesis features several applications like:
- Knee pain
- Spontaneous osteonecrosis
- Bone on bone
- Worn cartilage
TOMOSNTHESIS SPECIFICATIONS
Frame rate 8 fps
Single exposure duration 4msec
Number of exposures per exam Up to 60
Projection angle 40°
Scanning speeds 7.5 sec
Stratum selection Can be set by the operator from 0 to 400 mm
13
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
FINANCIAL OFFER
Remote controlled tilting table with ± 90° tilting and elevating table top set for
direct to digital Flat panel detector 43x43cm
Minimum height from ground 43 cm
Tabletop in laminated with abs. <0.7 mm Al eq.
SID: from 105 to 180 cm
Total Amount inclusive of wooden boxing (fumigation and vacuum not incl) € 137.800,00
14
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
15
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
16
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
Packaging
Included (fumigation not included)
Payment
Advanced or L/C at the order. In case of L/C and administrative fee of +€ 300,00 must be added
Warranty
As per General Warranty Conditions (ANNEX 1).
Training
Training for technicians who will install CAT Medical Systems devices is strongly recommended.
Warranty conditions may not be applicable to devices not installed by trained technicians. Training session
is free of charge if performed at CAT Medical Systems premises except for travel, meals and
accommodation, which are at customer charge.
CAT Medical Systems will release a Training Certificate.
Delivery Time
For any legal purpose, CAT Medical Systems recognizes its official commitment on delivery time expressed in
its Order Acknowledgement, which is determined in reason of volume ordered and type of device.
Other conditions
All other conditions are stated in our ANNEX 2) General Supplying Conditions.
ANNEX 1)
17
CAT MEDICAL SYSTEMS SPA – VIA E.AMALDI, 14
MONTEROTONDO, (ROMA) ITALY VAT NO. 13/01/2015
IT-02318260649
c. All components and parts of the product up to the complete product substitution;
To benefit of warranty right, the customer must take charge of sending back the defective component by
mean of express courier within 2 working days from receipt of the spare parts required. After evaluation at
our premises, if any modification is found or any damage not depending on the manufacturing process is
pointed, CAT Medical Systems will issue a report and will invoice the full cost of the spare part delivered
and its transport. Without receiving any notice of the defective part shipment within 30 days from
shipment of the required spare part, CAT Medical Systems will invoice the full cost of the spare part
delivered and its transport;
X-Ray inserts are subject to 12 months warranty Pro-Rata-Temporis, all conditions for X-Ray insert are the
same as for other parts or components;
___________________________
18