You are on page 1of 6

Analyst, November 1997, Vol.

122 (1283–1288) 1283

Analytical Bias: the Neglected Component of


Measurement Uncertainty†

Jean S. Kane
Robert J. Kane Associates Inc., Brightwood, VA 22715, USA

All analysts are accustomed to reporting measurement to illustrate what is required. Two examples from the literature
results accompanied by either the standard deviation of describing NIST certifications3,4 provide similar illustration.
individual results from the mean or the standard error of Most routine laboratories will not be able to develop uncertainty
the mean. These statements indicate repeatability of statements for their data in this manner; they will instead be
measurement under unchanged conditions, or drawn largely on data from their routine quality control and
reproducibility, where the time period over which quality assurance programs and the more in-depth knowledge of
measurements are taken is the source of changed measurement uncertainty developed by metrology laborato-
conditions. Repeatability generally produces a smaller ries.
deviation between replicates than does reproducibility. Uncertainty in the broadest terms means the doubt about the
However, most of the variability of measurements made validity of any measurement results. While doubt about validity
by different laboratories, or using different methods in a arises from error, error and uncertainty are not the same. The
single laboratory, is not accounted for by either doubt may exist about the exactness of the results, i.e., as a
repeatability or reproducibility attributable to time result of repeatability or reproducibility attributable to time
period. This is evident in all interlaboratory data sets; period factors. Alternatively, it may exist about the validity of a
these frequently contain between-laboratory and result attributable to inappropriate sample collection or han-
between-method discrepancies that are very large in dling, to heterogeneity leading to subsampling error or to the
comparison with the uncertainty of measurements presence of analytical bias.
typically reported by individual laboratories. Recently In order to estimate uncertainty reliably for a measurement, it
issued ISO guidelines and related documents address is essential that the measurement process be well defined and
these discrepancies as a legitimate component of that the process be under statistical control when the measure-
measurement uncertainty, and recommend expanding the ment is made. In defining the measurement process, the
concept to include the deviation of a measurement from laboratory will determine the degree of accuracy necessary for
the true value of the measurand, so long as this deviation ‘fitness of purpose’ and will identify those factors in the process
is small relative to fitness for purpose requirements. which have greatest influence on the measurement result.
When the bias of measurement renders the result Influences may be either random, arising from unpredictable
unsuitable for purpose, however, that bias is a significant variations, or systematic, arising from an influence quantity
effect and must be removed by use of a correction factor. whose effect is either constant or varies in a predictable way. If
To correct for significant bias, or to include smaller bias an influence quantity produces a systematic effect that is
as an uncertainty component, laboratories must evaluate significant from a fitness-for-purpose perspective, correction is
and quantify the bias in their measurements to the fullest essential, and the uncertainty associated with the correction
extent possible. This paper presents uncertainty must be included in the overall uncertainty of measure-
statements developed in accordance with the ISO ment.1,2,5
guidelines for several reference sample measurements. If, instead, the systematic effect is small relative to acceptable
Some are very complex, drawn from published work of uncertainty for the purpose, the bias itself is the source of
metrology laboratories. Others, drawn from the author’s uncertainty, and its magnitude may be included as an un-
data for the Japanese Sedimentary rock reference certainty component, as illustrated in example 4 of Appendix A
materials, are more suitable for routine laboratory use. in the Eurachem document.2 A NIST document6 illustrating
tests for bias in comparison with reference values is instructive
Keywords: Analytical bias; measurement uncertainty;
in this regard. According to this reference,6 not all deviations of
reference materials
a measurement result represent bias: only those which are
significant in comparison with the uncertainty of the reference
value. However, NIST includes any deviations of reference
The Guide to the Expression of Uncertainty in Measurement1 material (RM) results from reference values in its uncertainty
established general rules for evaluating and stating measure- estimates,3 when an existing NIST Standard Reference Material
ment uncertainty across a broad spectrum of measurements. (SRM) is analyzed concurrently with a new SRM undergoing
This guide has particular relevance to laboratories whose data certification analysis. This uncertainty is termed ‘recovery of
will be used to develop reference values for geological standard’ ( = measured/certified) in ref. 3 and in the text and
reference materials. These laboratories will need to establish Tables 2 and 5 in this paper.
uncertainties for all contributed data in keeping with the Many potential soruces of uncertainty are listed in the
principles of this important new ISO guide. Eurachem document.2 The list is intended to be illustrative,
A detailed study of measurement uncertainty requires rather than definitive. These have been broadly categorized as
extensive effort, and is typically undertaken only in national sampling, random variation and analytical bias, and are shown
metrology laboratories. The ISO guide1 provides guidance on in Table 1. It should be noted that spurious error, also included
the process, and the Eurachem Working Group on Uncertainty in Table 1, is an extreme case of random error; uncertainty
in Chemical Measurement2 has published a number of examples estimates do not allow for the possibility of spurious error.
Sampling as a source of uncertainty is the subject of the other
† Presented at Geoanalysis 97: 3rd International Conference on the Analysis of papers in this issue.7–9 The focus of all that follows will be on
Geological and Environmental Materials, Vail, CO, USA, June 1–5, 1997. analytical bias alone among these potential soruces of un-
1284 Analyst, November 1997, Vol. 122

certainty. The goal is to illustrate the use of control data and the acceptable, and outlier rejection before deriving recommended
like in estimating overall uncertainty. This process is included values was greater than 50% in some cases. This is seen in Figs.
as appropriate in the ISO guide and related documents,1,2,5 but 1 and 2, presenting the G-1 and W-1 data from Fairbairn et al.’s
illustrative examples are very limited in comparison with those report.10 Youden11 attributed most discrepancies of this type to
showing the more exhaustive metrological approaches. systematic error or analytical bias inherent in methods being
used.
It is these systematic errors leading to analytical bias, and the
Analytical Bias
doubt about the validity of the result due to them, that will be
All analysts are very familiar with instances of disagreement addressed in all that follows. Some of these errors and resulting
between results that are nominally for the same measurand. uncertainties can be evaluated from the statistical distribution of
These discrepancies have been the topic of much discussion a series of measurement results, and can therefore be charac-
among geoanalysts since Fairbairn et al.’s study10 first made it terized by a standard deviation, in just the same way that
clear how common they are. Data ranges for all major rock- random variations are. Other components are evaluated from
forming constituents other than SiO2 were completely un- assumed probability distributions based on experience or other
information. These two categories for evaluating uncertainties
are referred to as Type A and Type B in the ISO guide.1
Table 1 Potential sources of uncertainty in analytical measurements

Sampling uncertainty—
Metrology Laboratory Approach
Sample was inappropriate to the measured purpose The isotope dilution mass spectrometric method is a definitive
Sample was suitable, but has degraded in handling and/or storage method, commonly used by metrology laboratories for certifica-
between collection and measurement (see ref. 7) tion analyses. When properly applied and fully under control,
Field sampling uncertainty was too large to be ‘fit for purpose’ (see ref.
8)
results will be obtained that are essentially free from analytical
Measurand is inhomogeneously distributed in the sample bias. The uncertainty in result for the homogeneous sample will
be determined entirely by repeatability of measurement, with
Note: Sampling error due to heterogeneity of sample is more properly
described as sub-sampling error, leading to between-unit sub-sampling
RSDs of about 0.2%. This small uncertainty of measurement is
variance. The uncertainty due to sub-sampling error may be quantified
using analysis of variance where the unit of issue is the classification
variable (Type A evaluation) or by modeling mineralogy and using the
theoretical Poisson distribution for the count of mineralized grains in the
subsample (Type B evaluation) (see ref. 9)
Random error—
Measurement repeatability for sample measured repetitively
Measurement repeatability for replicate subsamples of a single sample
Note: Random error can never be assigned a definitive value. The
standard deviation of replicate measurements from the arithmetic mean is
a measure of uncertainty due to some random effects, but is not the
random error2
Note: The standard deviation of the measured values from the arithmetic
mean can be appropriately estimated from 15 or fewer measurements,
unless very high precision is required, in which case a larger number of
replicates is needed. This estimation of uncertainty is Type A
Spurious error—
Transcription error in recording analytical mass and/or signal
Misreading of analog signal output
Undetected instrument malfunction during measurement Fig. 1 Interlaboratory ranges relative to the recommended values for
Sample loss due to poor technique, spillage, etc., that did not result in major oxides in G-1. Data from Fairbairn et al.10 Open bar, full range; solid
discarding sample bar, 1s range for recommended value.
Note: Results affected by spurious error are invalid. They should be
discarded and there should be no attempt to incorporate the spurious error
into statistical analysis of error sources1,2,5
Analytical bias—
Incomplete knowledge of factors having critical impact on measurement
accuracy
No correction or incomplete correction for matrix effects (physical
interferences)
No correction or incomplete correction for spectral interferences
Incomplete separation and/or preconcentration of measurand from
matrix
Calibration error due to uncertainty in reference values used for
calibration
Approximations and assumptions incorporated in measurement proce-
dure
Uncertainty in masses, volumetric apparatus, etc.
Uncertainties in values of parameters used in data reduction (e.g., atomic
mass and isotopic abundances of the elements for mass spectrome-
try)
Instrument resolution or discrimination threshold
Inadequate control of environmental conditions needed for measurement Fig. 2 Interlaboratory ranges relative to the recommended values for
accuracy major oxides in W-1. Data from Fairbairn et al.10 Open bar, full range; solid
bar, 1s range for recommended value.
Analyst, November 1997, Vol. 122 1285

obtained at great effort and cost, however, and should not be the (sample vial to digestion beaker, beaker to column, beaker after
analytical goal except when necessary for making valid evaporation to ignition crucible) to be very important. They also
decisions based on the measurement results. considered ignition losses due to evaporation and/or splattering.
Similarly, gravimetric analysis when carried out under full Ability to achieve the certified result for concurrent analysis of
control can produce results that have random uncertainties previously certified materials (‘recovery of standard’) was
similar to the overall uncertainty of the isotope dilution evaluated. Additionally, incompleteness of the chromato-
measurement. Freedom from systematic error in the gravimetric graphic separation or failure to collect 100% of the sodium
analysis, however, is less assured than with isotope dilution fraction introduced uncertainty. Finally, the chemical blanks
measurement. Two examples taken from NIST certifications are and any weighing or volumetric errors were evaluated.
illustrative.3,4 The NaCl SRM 919a is the starting material for Only measurement replication (random error) and ‘recovery
the Na spectrometric solution SRM 3152 and the Cl anion of standard’ were evaluated by statistical means (Type A). All
chromatography solution SRM 3182. The solution SRM 3181 is others were evaluated in other ways (Type B). The individual
the sulfate anion chromatography solution standard. uncertainties were first combined within each type following
For the certification of both, NIST employed gravimetry. the law of propagation of errors, and then the uncertainties
Error evaluation examined incomplete precipitation, con- evaluated by Type A and Type B means were combined and
tamination of the precipitate and losses in sample handling at expanded to provide the final uncertainty for the certification. A
every step in the process, among other things. coverage factor of 2 for the expansion gives a level of
The method used for SRM 919a certification3 is a modifica- confidence of approximately 95%, and a factor of 3 gives
tion and improvement of the definitive method first developed approximately 99% confidence. The recommended coverage
as a National Committee for Clinical Laboratory Standards factor1 for general use is 2. The detailed evaluation taken
method. The sample was first decomposed and then separated directly from Moody and Vetter3 is given in Table 2.
on an AG 50W-X8 ion-exchange column. The sodium fraction In the case of the gravimetric analysis of the sulfate solution
was collected and converted into the sulfate using H2SO4; SRM 3181,4 classical gravimetry was coupled with in-
evaporation produced the salt, which was then ignited and strumental techniques; the latter allowed corrections to be made
weighed to complete the gravimetric determination. The ion- for any Ba that remained unprecipitated, for any BaCl2 or
exchange separation of sodium from any Li, K, Mg and Ca also K2SO4 occluded in the BaSO4 precipitate and for any sulfate
in the sample avoided an error due to the co-existence of other volatilized during ignition. Uncertainties in the contribution of
sulfates in the ignited residue. the blank and from measurement replication were determined
To summarize briefly the error sources examined, Moody using a Type A evaluation. Mechanical losses and all in-
and Vetter3 considered the sample handling and transfer steps strumentally determined corrections were estimated using Type

Table 2 Uncertainty of measurement of sodium for the certification of SRM 919a3

Source of uncertainty Relative standard uncertainty (%)


Type A evaluation used—
Measurement replication 0.007
‘Recovery of standard’ 0.019
Combined Type A ABBBBBB BB2 = 0.020
(0.007)2 + (0.019)
Type B evaluation used—
Sampling handling 0.036
Sample volume 0.002
Separation—loading/retention 0.005
Separation—loss/contamination 0.001
Weighing and ignition 0.015
Combined Type B ABBBBBBBBBBBBBBBBBBBB BB
(0.036)2 + (0.002)2 + (0.005)2 + (0.001)2 + (0.015) 2 = 0.040

Combined uncertainty ABBBBBB BB2 = 0.0443


(0.020)2 + (0.040)
Coverage factor 2
Expanded uncertainty (relative 0.0089

Table 3 Measurement uncertainty for the certification of sulfate SRM 31814

Source of uncertainty Relative standard uncertainty (%)


Type A evaluation used—
Measurement replication 0.018
Blank 0.009
Combined Type A ABBBBBB BB2 = 0.020
(0.018)2 + (0.009)
Type B evaluation used—
Mechanical loss 0.006
BaCl2 correction 0.060
K2SO4 correction 0.003
Filtrate sulfate correction 0.033
Volatile sulfate correction 0.035
Combined Type B ABBBBBBBBBBBBBBBBBBBBB
(0.006)2 + (0.060)2 + (0.003)2 + (0.033)2 + (0.035)2 = 0.077
Combined uncertainty ABBBBBB BB2 = 0.080
(0.020)2 + (0.077)
Coverage factor 2
Expanded uncertainty (relative) 0.160
1286 Analyst, November 1997, Vol. 122

B methods. The results taken directly from Vetter et al.4 are 22 for BHVO-1, n = 24 for BIR-1, n = 24 for G-2) that were
given in Table 3. made at widely spaced times over several years between 1984
It can be seen from these examples that carefully applied and 1990. The changed conditions resulting from the time
gravimetric procedures can produce total uncertainties, includ- intervals included such things as replacement torches and
ing analytical bias, as small as those for the isotope dilution mirrors in the spectrometer, different decomposition solutions
mass spectrometric method. The potential for biased results in for both the control sample RMs and the calibration RMs and
gravimetric methods is also seen clearly in these examples. Note variations in instrument set-up, including manual wavelength
that separation step errors in the case of the Na certification and peaking and torch height adjustments. However, the same
of precipitation errors in the case of Ba account for a analytical procedure and instrument were used in all measure-
considerable portion of the total uncertainty reported. Laborato- ments and the same analyst performed the measurements, over
ries applying either of these methods under less strignent the entire time period involved. The detailed within-run
control will produce results with greater associated un- repeatability data are no longer available, but within-run RSDs
certainties. were typically only a fraction (less than one fifth) of the
reproducibility data shown in Table 4.
Table 4 also provides bias estimations or ‘recovery of
The Routine Laboratory Approach
standard’ data, both as the difference between the mean of the
The average laboratory cannot devote the time and effort to replicate control sample measurements and the reference value
method development and to evaluating sources of uncertainty and as the maximum deviation encountered during the time
that a national metrology laboratory such as NIST does. Yet the period involved. There is no case where the bias estimated from
average laboratory needs to be able to state uncertainties for its the mean deviation is statistically significant6 in comparison
measurements in a way that allows users of the data to judge with either the certified value uncertainties (see below) or the
their validity in addition to their repeatability. This is the case measurement reproducibility. Additionally, it is acceptably
regardless of the use to which the data are put; it is particularly small from a fitness for purpose perspective to remain
the case when the data will contribute to a reference sample uncorrected. Acceptable errors for the method are a function of
characterization. The magnitude of the uncertainty is dictated concentration13 and for SiO2 are 1% relative to the reference
by requirements for the end use of the data, by ‘fitness for value.
purpose’ considerations. For certification purposes, the un- It is presumed that sub-sampling error is not a factor, based
certainty should be 3–10 times smaller than that of routine on prior homogeneity studies conducted in establishing refer-
analyses.12 ence values for the USGS rock reference materials. Data
It has been stated previously1,2,5 that many aspects of a affected by spurious error are absent from the control chart data
laboratory’s quality control/quality assurance program will summarized in Table 4. Weighing and volume errors are
provide data that can be used in the estimation of uncertainty. estimated from balance and glassware calibrations, in the same
Information from calibration certificates for balances and manner as reported by NIST3,4 and summarized in Tables 2 and
laboratory glassware provides Type B estimations of un- 3.
certainties due to weighing and volume errors. Information on In using the control chart data to establish uncertainties in
reference sample certificates of analysis provides ‘recovery of results for the Japanese Geological Survey sedimentary rocks,
standard’ uncertainty estimates and/or calibration error esti- the principal factors to consider are how well known the USGS
mates, depending on how the laboratory used the reference rock references values are and how the difference between the
sample. Similarly, information from control charts, collabor- matrix of the sedimentary versus the igneous rocks might affect
ative studies and proficiency tests can be used to estimate the ‘recovery of standard’ value for the analytical samples in
laboratory uncertainty. Evaluations could be either Type A or comparison with the controls.
Type B, depending on whether sufficient data existed to The uncertainties in reference values for the USGS rock
produce a statistical standard deviation or whether judgment standards are comparable to those for NIST SRM certified
instead was used. These approaches to uncertainty evaluation values established using two or more reference independent
will often provide a single combined uncertainty of several methods of analysis, but slightly larger than those based on
related error sources that the metrology laboratory might definitive method certification.15–17 For example, uncertainties
evaluate individually. for reference values of SiO2 as stated on the certificates of
For example, the laboratory measuring Ba or S gravimet- analysis18,19 are 0.20% relative for the NIST obsidian and basalt
rically as the BaSO4 precipitate might analyze a number of SRMs 278 and 688. For the USGS rock RMs, the uncertainties
reference samples, and determine their precipitation losses and are estimated as twice the stanard error of compilation means,
contaminants (analytical bias) collectively with a Type A and range from 0.1% relative for G-1 and W-1 to 0.30% for
evaluation from the average ‘recovery of standard’ for the DTS-1 and PCC-1.15 The uncertainty in the reference values
several RMs, rather than following the NIST process. This is a
valid approach, and will result in an uncertainty statement that
suits the accuracy requirements of the laboratory measure- Table 4 ICP-AES control data14
ments.
The approach will be illustrated with an evaluation of
Deviation (relative) from reference
uncertainty for major oxide data that the author contributed for
value (%)15
six of the Japanese Geological Survey sedimentary RMs, three Reproducibility
sediments, two slates and one chert, having analyzed the over multi-year of control data
samples in replicate (three splits, digested in duplicate with each Oxide RM period (as RSD) mean of extreme value
digestion solution analyzed on three different days) by ICP- SiO2 AGV-1 1.58 20.12 +1.7
AES.13 The decomposition was by lithium metaborate–tetra- BHVO-1 1.68 +0.60 21.64
borate fusion, and a two-point calibration using similarly Al2O3 BIR-1 1.9 +0.20 +2.6
prepared USGS rock RMs was employed. Two additional G-2 2.0 +0.5 +3.2
USGS rock RMs were included in each experimental run for Fe2O3 AGV-1 1.8 20.20 +3.3
BIR-1 1.7 +1.9 22.4
control purposes.
Na2O G-2 2.9 21.7 22.7
The control data14 given in Table 4 provide reproducibility BHVO-1 4.0 +0.4 +2.2
data based on replicate measurements (n = 26 for AGV-1, n =
Analyst, November 1997, Vol. 122 1287

shown in Table 5 is the pooled uncertainty for the two RMs used simultaneously with the Japanese sedimentary RMs to provide
as controls for SiO2 analysis, a Type A evaluation. the estimates of ‘recovery of standard’, the reference values of
While long experience with the method suggests that there is the newly characterized reference samples20 are used for that
no correlation between magnitude of bias and matrix type,13,14 purpose (see Table 6).
this has not specifically been evaluated. To be certain that the These data reflect a larger relative deviation of mean
uncertainty is not underestimated, a rectangular distribution is measured results from reference values than the USGS control
assumed, and the uncertainty is stated without specifying the sample data had. This may result because uncertainties in the
confidence level. In this case, standard uncertainty for the bias reference values themselves are larger than for the USGS rocks,
estimate is the worst case control value deviation from the being based on a 8–10-fold smaller database. Alternatively, it
reference value divided by AB 3.2,3 could result from extrapolation of the calibration curve for some
It is recognized that the calibration range for the measure- individual analyses, as noted above. Were we to examine data
ments does not encompass extreme values, e.g., the 98.8% SiO2 from the proficiency test program, even larger deviations of
of the Japanese sedimentary chert sample JCh-1. A bias larger individual laboratory data from the central value would be
than observed for the two control samples is likely to result from noted, as would greater uncertainty in the central value.
extrapolation of the calibration curve beyond its linear range.13 However, all of these approaches to estimating measurements
This is another reason for assuming a rectangular distribution uncertainties follow the principles of the process outlined in the
for ‘recovery of standard’ values, rather than a normal ISO Guide,1 and are recommended to all geoanalytical
distribution, for which uncertainty would be based on the laboratories, particularly those participating in collaborative
average deviation of all control results from the reference programs to establish values for reference materials.
value.
Applying the above to the control data, total uncertainties
were calculated as shown in Table 5. The expanded uncertainty Conclusion
shown can be expected to estimate reliably not only the total
uncertainties in results for the two control samples measured Most laboratories appropriately evaluate and report the repeata-
using the ICP-AES method, but more generally for most rock bility of their laboratory measurements. However, repeatability
samples measured using the same analytical method. They is only one component of the total measurement uncertainty, or
included not only the random component of uncertainty that is doubt about the validity of measurement. Approaches to
customarily reported in all of the analytical literature, but also evaluating measurement uncertainty more completely have
the analytical bias component that is often neglected. been presented.
A laboratory’s participation in collaborative measurement These approaches to evaluating measurement uncertainty,
programs can similarly provide a means of uncertainty although less rigorous than the approach taken by metrology
estimation. In this case, rather than using control samples run laboratories, will help geoanalytical laboratories to identify any
of their standard operating procedures (SOPs) that do not
produce ‘fit for purpose’ data. This will lead to improved SOPs,
Table 5 Estimation of uncertainty for SiO2 in control samples AGV-1 and and therefore to overall improvement of the reliability of
BHVO-114 laboratory data. Improved data in turn will permit laboratories
participating in proficiency tests to see continuous improve-
ments in their respective z-scores. Alternatively, it will permit
Type Source of uncertainty Relative standard uncertainty (%) laboratories producing ‘fit-for purpose’ data to refine their
A Repeatability* 0.34 uncertainty estimates to include any acceptably small bias that
Uncertainty in reference 0.17 contributes to uncertainty. Were this to be done, comparability
value15 or harmony of data from collaborative programs would improve
B ‘Recovery of standard’ 0.17/AB 3 = 0.98 and the definition of reference values and their uncertainties for
B Weighing errors 0.1 geochemical RMs would improve accordingly.
B Volume errors 0.02
Combined Type A ABBB
(0.34)BBBBB
2 + (0.017)2 = 0.0038

Combined Type B ABBBBBBBBBB


(098)2 + (0.1)2 + (0.02)2 = 0.985
References
Combined uncertainty ABBBBBB BB
(0.38)2 + (0.985) 2 = 1.06

Coverage factor 2 1 ISO, Guide to the Expression of Uncertainty in Measurement, ISO,


Expanded uncertainty (rela- 2.12 Geneva, 1993.
tive) 2 Eurachem Working Group, Quantifying Uncertainty in Analytical
Measurements, Laboratory of the Government Chemist, Teddington,
* Maximum repeatability value = reproducibility value (Table 4) 3 UK, 1995.
0.2. 3 Moody, J. R., and Vetter, T. W., J. Res. Natl. Inst. Stand. Technol.,
1996, 101, 155.
4 Vetter, T. W., Pratt, K. W., Turk, G. C., Beck, C. M., and Butler, T.
A., Analyst, 1995, 120, 2025.
Table 6 ‘Recovery of standard’ uncertainty estimate for SiO2 in rocks based 5 Taylor, B. N., and Kuyatt, C. E., Guidelines for Evaluating and
on participation in collaborative certification of Japanese sedimentary rock Expressing the Uncertainty of NIST Measurement Results, NIST
standards Technical Note 1297, US Government Printing Office, Washington,
DC, 1993.
Relative deviation 6 Becker, D., Christiansen, R., Currie, L., Diamondstone, B., Eber-
Contributed data Reference value of contributed data hardt, K., Gills, T., Hertz, H., Klouda, G., Moody, J., Parris, R.,
Standard (mass (mass fraction as %) from reference Schaffer, R., Steel, E., Taylor, J., Watters, R., and Zeisler, R., NIST
Standard fraction as %) (mean ± s) value (%) Spec. Publ., 1992, No. 829.
JSd-1 66.17 66.55 ± 0.354 20.57 7 Horowitz, A. J., Analyst, 1997, 122, 1193.
JSd-2 60.82 60.78± 0.411 +0.066 8 Ramsey, M. J., Analyst, 1997, 122, 1255.
JSd-3 76.03 76.00± 0.606 +0.039 9 Kane, J. S., Analyst, 1997, 122, 1289.
JSI-1 59.20 59.47± 0.263 20.46 10 Fairbairn, H. W., Schlecht, W. G., Stevens, R. E., Dennen, W. H.,
JSI-2 59.8 59.45± 0.223 +0.58 Ahrens, L. H., and Cheyes, F., US Geol. Surv. Bull., 1951, No.
JCh-1 98.15 97.81± 0.484 +0.35 980.
11 Youden, W. J., Anal. Chem., 1960, 32, 23A.
1288 Analyst, November 1997, Vol. 122

12 Uriano, G. A., and Gravatt, C. A., CRC Crit. Rev. Anal. Chem., 1977, 18 Uriano, G. A., NBS Certificate for Analysis for SRM 278, National
6, 361. Bureau of Standards, Washington, DC, 1981.
13 Kane, J. S., and Dorrzapf, A. F., Jr., in GeoExpo/86: Exploration of 19 Uriano, G. A., NBS Certificate for Analysis for SRM 278, National
the North American Cordillera, ed. Elliott, I. L., and Smee, B. M., Bureau of Standards, Washington, DC, 1981.
Association of Exploration Geochemists, Vancouver, BC, Canada, 20 Imai, N., Sakuramachi, H., Terashima, S., Itoh, S., and Ando, A.,
1987, pp. 184–188. Geostand. Newsl., 1996, 20, 165.
14 Kane, J. S., unpublished control sample data.
15 Gladney, E. S., Burns, C. E., and Roelandts, I., Geostand. Newsl.,
1983, 7, 1988, 12, 253. Paper 7/04789D
16 Kane, J. S., J. Geochem. Explor., 1992, 44, 37. Received July 7, 1997
17 Kane, J. S., Spectrochim. Acta, Part B, 1991, 46, 1623. Accepted September 22, 1997

You might also like