EASYLAB
Controller TCU3
for Fume Cupboards and Air Management Systems

The art of handling air


Contents · Description

Description __________________________________ 2 Construction · Dimensions ______________________15


EASYLAB system ____________________________ 3 Technical data________________________________16
Expansion modules __________________________ 4 Fume cupboard controller
Areas of application · control strategies ____________ 5 Specification text ____________________________17
Fume cupboard control ________________________ 6 Order code ________________________________18
Room control Room controller
Room management function RMF ______________12 Specification text ____________________________19
Controller inputs and outputs · Air terminal units ____14 Order code ________________________________20

TVLK with TCU3 TCU3 with all expansion modules

The EASYLAB controller TCU3 is the core of the TROX Air terminal units
EASYLAB system. With EASYLAB, good functionality and high – TVLK, variable volume flow controller for laboratories
levels of safety can be easily combined with optimum energy – TVRK, plastic variable volume flow controller
efficiency. The system is prepared for future expansions or
– TVR, circular variable volume flow controller for supply or
changes in use.
extract air
The TCU3 controller was developed for fume cupboard control, – TVJ · TVT, rectangular variable volume flow controller for
for the supply or extract air control of laboratories, and for supply or extract air
various tasks in hospitals and clean rooms.
– TVZ · TVA, variable volume flow terminal units for supply or
Additional functions such as the integration into a LonWorks extract air
network can be achieved with various expansion modules.
Characteristics, dimensions, and technical data on the air
The TCU3 automatic controller is combined with an air terminal terminal units can be found in the corresponding technical
unit to provide a flow rate control package. Every controller is leaflets.
factory tested and set to the required flow rates.
Further, current information on design can be found on our
Website and in our “Air Management Systems” design manual.

2
EASYLAB system

The EASYLAB system makes it possible to control Advantages of the EASYLAB system
safety-related flow rates in a rapid, reliable, and demand-based • Modular, expandable system
manner. Among other things, this includes fume cupboard • Innovative operating mode and room control strategies to
control and the supply air, extract air, and differential pressure meet individual project requirements
control in laboratories, clean rooms, and rooms in hospitals with
• Configurable monitoring functions and fault signalling
similar requirements.
(e.g. consolidated alarms)
In addition to the controller, expansion modules, sensors, and • Simple plug-and-play installation and plug-in communication
control panels, EASYLAB mainly comprises extensive software cable
for solving control-issues. The TCU3 controller forms the basis
for all closed-loop and open-loop control tasks and for this • Easy installation due to user-guided EasyConnect
purpose is combined with the corresponding air terminal units, configuration software
actuators, and pressure transducers. • Easily visible alarm indicator lights on both sides of the outer
casing
The modular structure offers numerous function expansion
options to adapt to project-specific requirements. Complicated • Central configuration of all parameters for higher order room
room air management is possible with numerous configurable functions (room management function)
closed-loop control strategies. The optimum EM-LON
expansion, for example, allows integration into a LonWorks
network.

Areas of application

TVLK, volume flow controller for Laboratory


aggressive media
TVLK
TVZ · TVA

TVRK

EASYLAB
TVJ · TVT

TVR

Clean room Closed-loop control in potentially Hospital


explosive atmospheres

3
Expansion modules

Function expansion
Expansion module Order code
The TCU3 controller can be adapted to the particular
requirements with various expansion modules.
EM-TRF
With the exception of the battery, all expansion modules are Supply voltage 230 V AC
T
connected to the TCU3 within the controller casing. The battery
is attached to the TVLK casing. The expansion modules can be
EM-TRF-USV
delivered already installed at the factory or retrofitted. U
Supply voltage 230 V AC and UPS

EM-LON
L
LonWorks interface

EM-AUTOZERO
Z
Automatic zero balance

EM-LIGHT

Lighting connecting socket

¹ Only in connection with T or U

Expansion modules

EM-TRF
Power pack for connection of the controller to a 230 V AC
supply voltage.
For more information, see Technical Leaflet No. 5.3/4/EN/..

EM-TRF
EM-TRF-USV
Power pack with uninterruptible power supply for connection
of the controller to a 230 V AC supply voltage. An electronic
system with an emergency power supply battery keeps the
controller running in case of a supply voltage failure.
For more information, see Technical Leaflet No. 5.3/4/EN/..
Battery pack
EM-TRF-USV

EM-AUTOZERO
Automatic zero balance with a solenoid valve for the
long-term stabilisation of the flow rate measurement. The
zero point voltage of the pressure transducer is measured
EM-AUTOZERO cyclically and the slow zero point drift is thus reliably
prevented. A manual zero balance during maintenance
is no longer necessary.

EM-LON
LonWorks communication interface FT10 for integrating the
controller into the centralised BMS.
For more information, see Technical Leaflet No. 5.3/5/EN/..
EM-LON

EM-LIGHT
The lighting control system enables the control of the fume
cupboard interior lighting on the control panel. 230 V AC
is controlled with a maximum of 500 W. EM-LIGHT enables
the simple plug-in connection of the lighting using a WAGO
WINSTA Midi Type 770 plug. EM-LIGHT is possible only in
EM-LIGHT connection with EM-TRF or EM-TRF-USV.

4
Areas of application · control strategies

Fume cupboard control Extract air control in extract air master system
FH function RE/LAB function
Control of the flow rates required due to the testing of the Control of the room extract air and limitation of the total extract
fume cupboards according to EN 14175. The monitoring of air flow rate of the room to a defined value. The flow rates of all
the aerodynamic function is integrated. Ideally, this closed-loop extract air controllers are taken into consideration. If a difference
control is used with an adaptively variable flow rate in order to in comparison with the required total extract air flow rate exists,
achieve energy savings with optimum safety. this difference will be controlled with the room extract.
Supply air control in extract air master system Differential pressure control in extract air master
RS/LAB function system
Control of the demand-based supply air flow rate related to RC/LAB function
the current overall extract air flow rate of the room. In the Link the room air balance control with the differential pressure
process, a configurable difference to achieve the underpressure control to achieve a rapidly balanced and stable pressure. The
required in DIN 1946, Part 7, is taken into consideration. difference between supply and extract air is variable in the case
of differential pressure control.

Fume cupboard control Closed-loop supply air control

Supply air controller with


room management function (RMF)
BMS
RMF

Communication cable

500 mm

Fume Fume
cupboard 1 cupboard n

Closed-loop extract air control Differential pressure control

Extract air controller with Supply air controller with


room management function (RMF) room management function (RMF)

RN RMF

Communication cable Communication cable

Room pressure controller

BE-LCD-01

Fume Fume Fume Fume


cupboard 1 cupboard 2 cupboard 3 cupboard 4

5
Fume cupboard control

TCU3 as fume cupboard controller FH-DS


For fume cupboard control, the TCU3 electronic controller is Variable flow rate control based on the measured opening
usually used in combination with a plastic volume flow controller height of the sash, with linear characteristic. Sash position
type TVLK · TVRK. sensor included in delivery.
– Control checked according to EN 14175, Part 6, by an FH-DV
independent testing institute and suitable for all fume Safety-optimised flow rate control based on the measured
cupboards tested according to EN 14175. opening height of the sash and theoretical calculation of the
– Fume cupboard control with integrated monitoring of the incoming flow velocity. Sash position sensor included in delivery.
aerodynamic function with optical and acoustic alarms FH-2P
according to EN 14175 Flow rate control with two setpoint values, controlled using a
switch contact provided by others.
Equipment functions and control strategies
FH-3P
FH-VS
Flow rate control with three setpoint values, controlled using
Variable flow rate control based on the measured incoming two switch contacts provided by others.
flow velocity. High thermal loads are detected and removed
FH-F
by an increased flow rate. Additional monitoring option for the
incoming flow velocity. A velocity transducer is included in Constant value flow rate control
delivery.

TVLK volume flow controller FH-VS function

TCU3

VS-TRD
TVLK velocity transducer

Control panels for fume cupboard FH-DS or FH-DV function


controllers TCU3

BE-LCD-01 BE-SEG-01 DS-TRD-01


sash position sensor

6
Fume cupboard control

Monitoring functions and alarms Operating modes


– Monitoring of the flow rate with configurable optical and – Standard operation with evaluation of the connected sensor
acoustic alarm system
– Optional monitoring of the incoming flow velocity with – Special operating modes: Increased operation, reduced
configurable optical and acoustic alarm (only for FH-VS operation, shut-off, and open position
equipment function)
– Monitoring of the sash position and signalling when the
Selection of the operating mode
maximum set opening position is exceeded (EN 14175). The The selection of the operating mode can take place from
acoustic signalling of this event can be manually suppressed three control levels with configurable prioritisation and isolation
or it can be time limited. options and affect all controllers in a room or individual
– Optical and acoustic signalling of the monitoring functions on controllers, e.g. for 24-hour operation.
the EASYLAB BE-SEG-01 or BE-LCD-01 control panels. Two – From the centralised BMS with LonWorks variables or
control panels, e.g. for fume cupboards with double sided switch contacts:
openings, can be connected. room operating mode or fume cupboard operating mode
– Alarm signalling can be individually configured, e.g. for alarm – On the room control panel or switch contacts:
suppression during operation or for the alarm light to flash or room operating mode or fume cupboard operating mode
be permanently on. – From the control panel of the fume cupboard:
fume cupboard operating mode

Control diagrams
FH-VS FH-3P
V
V 1
Flow rate
Incoming flow velocity V₃
Vmax

V₂ Flow rate
0.5 v in m/s

V₁
Vmin

0
Closed Open S₁ S₂ Switch

FH-DS FH-2P
V
V 1
Flow rate
Vmax
Incoming flow velocity

V₂ Flow rate

0.5 v in m/s
V₁
Vmin
0
Closed Open S₁ Switch

FH-DV FH-F

V 1 V
Flow rate
Incoming flow velocity
Vmax

0.5 v in m/s Flow rate

Vmin

0
Closed Open Closed Open

7
Fume cupboard control

Special functions Interface to the centralised BMS


– Consideration of diversity factors in connection with the Analogue signal (0 - 10 V)
EASYLAB room controllers – Actual flow rate of the controller
– Support of fume cupboards with supportive flow technology – Controller damper blade position
– Support of safe operation with extract air fume scrubbers – Total extract air flow rate, total supply air flow rate, or setpoint
– Connection of a motion sensor to signal “Sash Opened” value of the total supply air flow rate
or to reduce the incoming flow velocity when the room is Digital switch contact
unoccupied – Alarm state of the controller (change-over contact)
– Smoke extract function for opening or closing the control LonWorks interface with EM-LON expansion module
damper blade The current operating values and fault messages of the fume
– Control of a sash moving mechanism cupboard controller are transferred via the network:
– Control of the fume cupboard interior lighting on the control actual flow rate, damper blade position, operating mode,
panel incoming flow velocity, sash opening (DS and DV), and
– Adjustable controller behaviour in case of a failure of the switch stage (2P and 3P)
supply voltage using the expansion module EM-TRF-USV Further information on this topic, especially on the interface
definitions, can be found in Technical Leaflet No. 5.3/4/EN...

Diversity control BE-SEG-01 control panel


Large laboratory buildings often do not have the capacity for
the simultaneous maximum operation of all controllers. The
diversity control permits the safe operation of these specific
plants and an economic operation of all plant. Status display
1. The total extract air flow rate is monitored. Configurable
optical and acoustic alarm on the room control panel.
2. The total extract air flow rate is monitored and, when
exceeded, limited by a reduction of the flow rate of Alarm suppression
individual fume cupboards.
Sash monitoring (EN 14175)

Fume cupboard extract air

1 Selection of the operating


mode
open 1

open 4
open 3
open 2
oard

oard
oard
cupb e

cupb e
cupb e
oard
cupb e
Fum

Fum
Fum
Fum

Sash open

Sash close
Total room extract air

2 Lighting

Local control
open 4
oard
cupb e
Fum

Display (incoming flow velocity,


status messages)
open 3
oard
cupb e
Fum
open 2
oard
cupb e
Fum
open 1
oard
cupb e
Fum

1 Flow rate reduction at fume cupboards 3 and 4 due to diversity control


2 A flow rate reduction to the set maximum value of the total extract air
is achieved.

8
Fume cupboard control

Control inputs and outputs


Function Required inputs
4 analogue inputs
Up to 4 analogue inputs with configurable characteristic for the FH-2P 1
integration of variable flow rates. FH-3P 2
– 3 inputs for FH-VS, FH-DS, and FH-DV equipment function
– 4 inputs for FH-2P, FH-3P, and FH-F equipment function Monitoring of the maximum (variable)
1
opening of the sash
6 digital switching inputs
The table opposite shows the possible equipment functions and Extract air fume scrubber 1
number of necessary inputs.
Supportive flow technology 1
3 analogue outputs
Motion detector 1
First actual flow rate, second controller damper blade position
and third total extract flow rate, or total supply air flow rate, or Smoke extract function 1
setpoint value of the total supply air flow rate Selection of operating mode using switch
6 digital switching outputs contacts (only for fume cupboards For each operating mode:
Alarm signalling, control of extract air fume scrubbers, without the use of the room operating 1
supportive flow technology, sash moving mechanism, and mode)
fume cupboard interior lighting
Temporary integration of constant flow
Up to 6
rates with switch contacts

BMS

Analogue
LonWorks Switch contacts
0 - 10 V DC

Communication cable

TROX
adapter module
(TAM)

Fume cupboard 1 Fume cupboard 2 Fume cupboard 3 Fume cupboard 4

BE-LCD-01 control panel

9
Room control

TCU3 for room control Differential pressure control in extract air master system
PC/LAB
For the control of variable supply or extract air flow rates, the
electronic controller TCU3 is used in combination with an air Link the room air balance control with the differential pressure
terminal unit types TVR · TVRK · TVJ · TVT · TVZ · TVA. control to achieve a rapidly balanced and stable pressure. The
difference between supply and extract air is variable in the case
Equipment functions and control strategies of differential pressure control.
Extract air control in extract air master system RE/LAB Balancing
Control of the room extract air and limitation of the total extract
air flow rate of the room to a defined value. The flow rates of all For balancing, all supply and extract flow rates of a room
extract air controllers are taken into consideration. If a difference are recorded and summed. The following are taken into
in comparison with the required total extract air flow rate exists, consideration:
this difference will be controlled with the room extract. – Maximum of 24 EASYLAB controllers
Supply air control in extract air master system RS/LAB – Analogue inputs
Control of the demand-based supply air flow rate related to the – Digital inputs
current overall extract air flow rate of the room. In the process, a – Parameters
configurable difference to achieve the underpressure required in
DIN 1946, Part 7, is taken into consideration.

RS equipment function

Room management function (RMF)

Communication line X

X
X

Fume Fume
cupboard 1 cupboard 2

X = Constant supply air/extract air difference


for air transfer flow

RE equipment function

RN Room management function (RMF)

Communication cable Total extract air X


Total supply air

Room extract air Extract air of fume cupboards

Fume Fume Fume Fume


cupboard 1 cupboard 2 cupboard 3 cupboard 4

X = Flow rate difference to achieve the underpressure

10
Room control

Functions Room management function


– Stable room balance due to rapid control response – Connection of room control panels
– Flow rate monitoring with optical alarm on the controller – Central room interface
casing, as well as individually configurable alarm signalling – Central configuration of room parameters
– Override of the room operating mode by an individual
operating mode Operating modes
– Extract air balance optimisation through the reduction of the – Standard operation
room extract air if a sufficient flow rate is provided through the – Special operating modes: Increased operation, reduced
fume cupboards (introduction at the end of 2010) operation, shut-off, and open position
– Automatic flow rate distribution of the supply or extract air
of a room in case of several EASYLAB controllers. Flow rate
Selection of the operating mode
distribution with same percentage or configurable to any – From the centralised BMS with LonWorks variables or
value. switch contacts
– Adjustable controller behaviour in case of a failure of the – From a maximum of 2 room control panels
supply voltage using the expansion module EM-TRF-USV

Pressure control BE-LCD-01 room control panel

Display of the setpoint and actual values of a differential


pressure control

Room control

∆p V

Total extract air x


Total supply y
air

0 Pa

−10 Pa

Door OPEN Door CLOSED

x = Difference, regulated by the pressure controller


y = Limited volume flow change from the pressure cascade

11
Room control
Room management function RMF

Functional description Central configuration of the room parameters and


functions
The room management function (RMF) is a software option
to achieve functions across rooms. Within an EASYLAB system – Selection of the minimum total extract air of the room
with a maximum of 24 controllers, the room management – Selection of the supply air/extract air difference
function is activated on one controller – supply air, extract air, or – Selection of the differential pressure control parameters
EASYLAB adapter module. This controller is configured for the – Consideration of constant, non-switched flow rates
connection of the project-specific, cross-room functions. Monitoring of the room functions
Central interface point for room functions – Exceeding of the specified total extract air (monitoring of
– Central interface point for room functions diversity) or limitation to the specified total extract air
– Operating mode selection for all controllers in the room (diversity control)
– Flow rate change for temperature control Individually configurable consolidated alarms
– Flow rate change for external pressure control Integration of room control panels
– Connection of the room pressure transducer for internal – Connection of 2 BE-LCD-01 room control panels
differential pressure control – Condition display of the room monitoring functions
– Display of current room parameters, e.g. the actual and
setpoint value of the total extract air flow rate

Room management function RMF

Minimum total ∆p actual


extract air pressure value

Room operating Setpoint


mode pressure value

RMF
Change signal
Central
temperature
selections
control

Signalling Centralised BMS


interface/room

Room control panel Room alarms

12
Room control
Room management function RMF

LonWorks interface for the room with EM-LON expansion Interface to the centralised BMS
module Analogue signal (0 - 10 V)
– Selection of the room operating mode – Actual flow rate of the controller
– Prioritisation of the operating mode selections – Controller damper blade position
(centralised BMS/local)
– Total extract air flow rate, total supply air flow rate, or setpoint
– Connection of a volume flow change signal for temperature value of the total supply air flow rate
or differential pressure control
Digital switch contact
– Switchover between two setpoint values of a differential
– Alarm state of the controller (change-over contact)
pressure control
LonWorks interface with EM-LON expansion module
– Feedback signals from the EASYLAB system (room)
The current operating values and fault messages of the room
– Actual room operation mode
controller are transferred via the network:
– Total extract air flow rate actual value actual flow rate, setpoint flow rate, and operating mode.
– Total supply air flow rate actual value
Further information on this topic, especially on the interface
– Evaluated damper positions
definitions, can be found in Technical Leaflet No. 5.3/4/EN...
– Consolidated alarm
– Setpoint and actual value of the internal room pressure
control
– Room pressure alarm

Room control

BMS

Room management function


(RMF)

LON

Communication cable

Fume Fume Fume Fume


cupboard 1 cupboard 2 cupboard 3 cupboard 4 – 22

13
Room control
Controller inputs and outputs · Air terminal units

Without activated room management function 3 analogue outputs


4 analogue inputs First actual flow rate, second controller damper blade position
and third total extract flow rate, or total supply air flow rate, or
For the integration of a maximum of 4 variable flow rates
setpoint value of the total supply air flow rate
6 digital switching inputs
6 digital switching inputs
For the integration of temporary constant flow rates with switch
Possible functions:
contacts
– Selection of room operating mode:
3 analogue outputs Standard Mode, reduced operation, increased operation,
First actual flow rate, second controller damper blade position shut-off, open position
and third total extract flow rate, or total supply air flow rate, or – Room operating mode prioritisation:
setpoint value of the total supply air flow rate centralised BMS or local selection
With activated room management function: – Switchover between two setpoint pressure values in case of
internal differential pressure control
4 analogue inputs
– Flow rate change for temperature control – Consideration of door contact for internal differential pressure
control
– Flow rate change for external pressure control
– Integration of temporarily constant flow rates at unoccupied
– Actual room pressure value for internal differential pressure inputs
control
2 interfaces
– Integration of variable flow rates at unoccupied inputs
For the connection of a maximum of two room control panels

Air terminal units with EASYLAB controller TCU3 for room control

TVR TVZ · TVA

TVJ · TVT TVRK

14
Construction · Dimensions

Characteristics Construction features


– Electronic volume flow controller for fume cupboards, room – TCU3 controller installed and wired on the air terminal unit at
supply or extract air can be used in conjunction with air the factory
terminal unit types TVLK · TVRK · TVR · TVJ · TVT · TVZ · TVA – Electronics with failsafe saving of all configuration data in one
– Also for room or duct pressure control EEPROM
– Rapid actual value measurement – All important connections with plug and socket connections
– Fast and stable control algorithm; correction time depending on the outside of the casing
on duct pressure ≤ 2 s – Communication system with plug-in cables to connect all
– Permanent function monitoring of the system and the controllers in a room
connected sensors – Integrated terminal resistor for terminating the communication
– Intelligent communication system with automatic detection of cable, activation configurable and visible from the outside
the various equipment functions (indicator light)
– Prepared for wireless integration of a notebook for – Connections for functional expansions
configuration using the Bluetooth module (BlueCON) – Sockets inside the casing for optional expansion modules
– The controller casing can be opened without tools
– Fixing of the controller casing to the air terminal unit without
tools
– Plastic ABS casing, blue (RAL 5002)

TCU3
122

220¹
27¹
267

282

¹ Only in combination with TVRK

15
Technical data

Connections Status displays


Communication interfaces to other controllers Condition and fault display (with easily visible indicator
– Two RJ45 sockets for plug-in communication cables lights) on the outside of the casing
type S-FTP – Alarm display on both sides
– Screw terminals for the connection of two communication – Controller function display
cables type S-FTP cut from a wire roll – Communication diagnosis
Inputs and outputs – Activated terminal resistor
– 5 analogue inputs, 0 - 10 V, with configurable characteristic Additional condition display for diagnosis on the PCB
– 6 floating digital inputs inside the casing
– 4 analogue outputs, 0 - 10 V, with configurable characteristic – Display of the switching states of the digital inputs and outputs
– 6 digital outputs as relay change-over contacts 250 V, 12 A – Display of supply voltage
Additional connections
– 2 interfaces for the connection of EASYLAB control panels
– Service interface for configuration and diagnosis on the
controller and control panel
– EM-LON expansion module: LonWorks interface
– EM-LIGHT expansion module: lighting connecting socket

Connections and status displays 1 Alarm indicator light


2 Sash contact according to EN 14175
3 Connections for control panels or service
4 Connection for the actuator
5 Connection for the sensor system
6 Controller communication
7 Indicator lights for:
communication,
activated terminal resistor,
controller function (heartbeat)
1 8 Connection for fume cupboard interior lighting
2 3 4 5 6 7 1 8

Technical data

Supply voltage 24 V AC/DC ±15%, optional 230 V AC, 50/60 Hz

Power rating
Fume cupboard controller with control panel 35 VA
Room controller 29 VA
Room controller with room control panel 33 VA
with all expansion modules max. 40 VA

Micro fuse 5 × 20 mm 2A slow-acting, 250 V


Diaphragm pressure transducer with room air induction to protect the
Measurement of the flow rates
measuring point
Actuator Rapid response with high resolution, run time for 90° 3 s
Flow rate correction time ≤ 2 s, depending on duct pressure
Time for return of controller readiness after
500 ms
supply voltage failure

Temperature range
Operation 0°C to +50°C
Storage −10°C to +70°C

Humidity
Operation <90%, non-condensing
Storage <90%, non-condensing

Area of application within enclosed rooms


Protection level IP 20
IEC protection class III safety extra-low voltage (SELV)

16
Fume cupboard controller
Specification text

Specification text for the fume cupboard Alarm monitoring and signalling which is individually configurable,
controller* e.g. for alarm suppression in certain operating modes or compilation
of a consolidated alarm.
Circular variable volume flow controller made of flame-resistant
polypropylene (PPs) for variable volume flow systems and fume Operating mode selection of main control point, room control
cupboards. For the flow rate control of extract air containing panel, or fume cupboard control panel with flexible isolation and
contaminated or aggressive materials since all components coming prioritisation options. Supported special operating modes: increased
into contact with the air stream are made of plastic. operation, reduced operation, shut-off, and open position
... (The complete text can be found in Technical Leaflet No. Additional special functions
5.3/1/EN/..) – Support of fume cupboards with supportive flow technology
Control components: – Safe control of an extract air fume scrubber
Electronic controller for demand-based, variable flow rate control – Connection of motion sensors
of fume cupboards including the integrated monitoring of the – Smoke extract function
aerodynamic function according to EN 14175 with optical and – Control of the fume cupboard interior lighting
acoustic signalling. – Control of a sash moving mechanism
Supported equipment functions: – Consideration of diversity factors in connection with
– FH-VS TROX EASYLAB room controllers
Flow rate control on the basis of the measured incoming flow 5 analogue inputs for connecting the sensor system and integrating
velocity under consideration of thermal loads. Monitoring of up to 4 variable flow rates.
the flow rate and incoming flow velocity. 6 floating digital inputs for the integration of constant flow rates
– FH-DS and/or the control of special functions.
Flow rate control based on the measured opening height of the 3 analogue outputs for signalling actual flow rates and the controller
sash, with linear characteristic. Monitoring and signalling of the damper blade position as well as the total flow rate of the room
maximum set sash opening according to EN 14175. (supply or extract air).
– FH-DV 1 digital output for flow rate alarm with configurable alarm
Safety-optimised flow rate control based on the measured conditions.
opening height of the sash. Monitoring and signalling of the
5 digital outputs for supporting the various special functions.
maximum set sash opening according to EN 14175.
– FH-2P Rapid response actuator (run time 90° < 3 s)
Flow rate control with two setpoint values and monitoring of Supply voltage 24 V AC
the flow rate. Expansion modules:
– FH-3P – EM-TRF
Flow rate control with three setpoint values and monitoring of Supply voltage 230 V AC
the flow rate. – EM-TRF
– FH-F Supply voltage 230 V AC with uninterruptible power supply.
Constant value flow rate control and monitoring of the flow rate. Range of functions in UPS operation: alarm in case of power
Control electronics on microprocessor basis with configuration supply failure, monitoring and alarm for the controller function,
settings in EEPROM memory and thus secured in case of power continuation of normal control operation or, alternatively: open
failure. 5 analogue inputs, 6 floating digital inputs, 3 analogue position, shut-off, or lock damper blade position.
outputs, and 6 digital outputs as relay change-over contacts. Static – EM-AUTOZERO
diaphragm pressure transducer with room air induction to protect the Automatic zero point correction
measuring point. Maintaining of the flow rates through a permanent – EM-LON
setpoint/actual value comparison in a closed loop with a limitation to
minimum and maximum flow rates. LonWorks interface
– EM-LIGHT
Special characteristics:
Connection socket for simple connection of the fume cupboard
– Integrated system and sensor monitoring
interior lighting
– Intelligent communication system (plug-and-play)
* Text for a volume flow controller type TVLK with EASYLAB
– Easy configuration, commissioning, and diagnosis with
user-guided EasyConnect service software
All important communication and periphery connections on the
outside of the casing and thus easy to connect. Indicator lights on
the outside of the casing for alarm display (on both sides), controller
function (heartbeat), and communication.
Communication system with plug-in communication cables,
automatic connection setup without manual network configuration,
easy to expand with additional controllers (usually without
integration issues).
Integration of variable or constant flow rates in the room balance
using analogue signals, switch contacts, or constant values.
Signalling of actual flow rate values, damper blade positions, faults,
and status messages to centralised BMS with analogue or switch
outputs.
Connection of one or two (for fume cupboards with double sided
openings) adaptable EASYLAB control panels BE-SEG-01 or
BE-LCD-01 with optical and acoustic signalling. Acoustic alarm can
be deactivated and limited in terms of time.

17
Fume cupboard controller
Order code

Order code

TVLK / 250 / GK / ELAB / FH-VS / U L Z S / 500 – 900 m³/h

Air terminal unit¹ Module expansions Operating values


Type Option 1 FH-VS –
Nominal size Supply voltage FH-DS min – max
24 V AC, no entry required FH-D min – max
ccessories FH-2P –
T EM-TRF
Control components 230 V AC FH-3P – –
ELAB EASYLAB controller with rapid U EM-TRF-USV FH-F Constant value
response actuator 230 V AC with uninterrupted
Equipment function for power supply
fume cupboard control Option 2
FH-VS With incoming air flow sensor, LonWorks interface
incoming air flow velocity None, no entry required
control strategy L EM-LON
FH-DS With sash position sensor,
Option 3
linear control strategy
Automatic zero balance
FH-DV With sash position sensor,
safety-optimised control None, no entry required
strategy Z EM-AUTOZERO
FH-2P With switching stages via Solenoid valve for automatic
switch contacts (switches by zero balance
others), Two-point control Option 4
FH-3P With switching stages via Wired socket for lighting control
switch contacts (switches by None, no entry required
others), Three-point control S EM-LIGHT
FH-F Control of a fixed flow rate Wired socket for lighting control
¹ Entries to relate to the air terminal unit

Order example for the EASYLAB fume cupboard


controller
Volume flow controller type TVLK, plastic controller PP,
nominal size 250 mm, with flanges and matching flanges
EASYLAB control component with rapid response actuator,
fume cupboard control, and free issue velocity transducer
provided.
Expansions: supply voltage 230 V AC, automatic zero
balance, connecting socket for fume cupboard interior
lighting.
Operating values: Vmin = 300 m³/h and Vmax = 1,200 m³/h

Make: TROX
Type: TVLK-FL / 250-0 / GK / ELAB / FH-VS / TZS / 300 – 1200 m³/h

18
Room controller
Specification text

Specification text for room controller* Additional functions with activated room management function:
Variable volume flow controller in circular construction for variable – Connection of one or two adaptable EASYLAB room control
volume flow systems, for supply or extract air, in 7 nominal sizes. panels BE-LCD-01
Consisting of the casing with control damper, differential pressure – Central operating mode selection for all controllers in a room
sensor, and control components. Control damper closed blade – Central transfer point for individual or consolidated faults
leakage complies with EN 1751, Class 4 (nominal sizes 100 and – Monitoring of the room functions for an exceedence of the
125, class 3). Position of the control damper blade visible from the minimum total extract air flow rate and exceedence of the total
outside on the shaft extension. extract air flow rate (diversity)
... (The complete text can be found in Technical Leaflet No. 5/3/EN/..) – Limitation of the total extract air flow rate (diversity control)
Control components: – Operating mode selection of main control point, room control
Electronic controller for demand-based, variable flow rate control of panel, or fume cupboard control panel with flexible lisolation
supply or extract air in laboratories and for various tasks in hospitals and prioritisation options. Supported special operating modes:
and clean rooms. increased operation, reduced operation, shut-off, and open
position
Supported equipment functions:
– Central configuration of the room parameters
– RE / LAB
– Flow rate change for temperature control
Control of the room extract air and limitation of the total extract air
flow rate of the room to a defined value. The flow rates of all – Flow rate change for external pressure control
extract air controllers are taken into consideration. If a difference in – Connection of the pressure transducer for differential pressure
comparison with the required total extract air flow rate exists, this control
difference will be controlled with the room extract. – Individually configurable consolidated alarms
– RS / LAB 5 analogue inputs for connecting the sensor system and integrating
Control of the demand-based supply air flow rate related to the up to 4 variable flow rates or, in case the room management function
current overall extract air flow rate of the room. In the process, a is activated, for connection of the temperature flow rate change, the
configurable difference to achieve the underpressure required in pressure flow rate change, or the pressure transducer.
DIN 1946, Part 7, is taken into consideration. 6 volt-free digital inputs for the integration of constant flow rates or, in
– PC / LAB case the room management function is activated, for room operating
Linking of the closed loops for room air balance and pressure mode selection and prioritisation, switchover between two setpoint
for rapid response and stable differential control. The difference pressure values, or the connection of a door contact.
between supply and extract air is variable in the case of differential 3 analogue outputs for the signalling first actual flow rate, second
pressure control. controller damper blade position and third total extract air flow rate,
Control electronics on microprocessor basis with configuration or total supply air flow rate, or setpoint value of the total supply air
settings in EEPROM memory and thus secured in case of power flow rate
failure. 5 analogue inputs, 6 floating digital inputs, 3 analogue 1 digital output for flow rate alarm with configurable alarm conditions.
outputs, and 6 digital outputs as relay change-over contacts. Static
5 digital outputs for supporting the activated room management
diaphragm pressure transducer with room air induction to protect the
function: configurable consolidated alarm, pressure alarm (in case of
measuring point. Maintaining of the flow rates through a permanent
differential pressure control), or lighting control.
setpoint/actual value comparison in a closed loop with a limitation to
minimum and maximum flow rates. Rapid response actuator (run time 90° < 3 s)
Special characteristics: Supply voltage 24 V AC
– Integrated system and sensor monitoring Expansion modules:
– Intelligent communication system (plug-and-play) – EM-TRF
– Easy configuration, commissioning, and diagnosis with Supply voltage 230 V AC
user-guided EasyConnect service software – EM-TRF-USV
All important communication and periphery connections on the Supply voltage 230 V AC with uninterruptible power supply.
outside of the casing and thus easy to connect. Indicator lights on Range of functions in UPS operation: alarm in case of power
the outside of the casing for alarm display (on both sides), controller supply failure, monitoring and alarm for the controller function,
function (heartbeat), and communication. continuation of balancing function and normal control operation
Communication system with plug-in communication cables, or, alternatively: open position, shut-off, or lock damper blade
automatic connection setup without manual network configuration, position.
easy to expand with additional controllers (usually without integration – EM-AUTOZERO
issues). Automatic zero point correction
Integration of variable or constant flow rates in the room balance – EM-LON
using analogue signals, switch contacts, or constant values. LonWorks interface for the integration of the room into a higher
Signalling of actual flow rate values, damper blade positions, faults, level network.
and status messages to centralised BMS with analogue or switch
outputs. * Text for a volume flow controller type TVR with EASYLAB

Alarm monitoring and signalling which is individually configurable,


e.g. for alarm suppression in certain operating modes or compilation
of a consolidated alarm.
Balancing of up to 24 controllers on the communication cable (fume
cupboards, extract air, supply air). Automatic distribution of the
extract or supply air flow rates across several EASYLAB controllers
for one room or alterative individually configurable distribution.

19
Room controller
Order code

Order code

TVR / 250 / FL / ELAB / RS / U L Z / LAB /

10

Air terminal unit¹ Module expansions Application


Type Option 1 LAB Laboratory
Nominal size Supply voltage LAB-RMF Laboratory with room
24 V AC, no entry required management function
ccessories
T EM-TRF 10 Operating values
Control components 230 V AC LAB no paramters
ELAB EASYLAB controller with rapid U EM-TRF-USV LAB-RMF 7 parameters
response actuator 230 V AC with uninterrupted Flow rate in m³/h, pressure in Pa
quipment function power supply – Total room extract air
for room control – standard operation
Option 2 – Total room extract air
RS Supply air control LonWorks interface – reduced operation
(Room Supply) None, no entry required – Total room extract air
RE Extract air control L EM-LON – increased operation
(Room Extract) – Constant room supply air
Option 3
PC Pressure control – Constant room extract air
Automatic zero balance
– Supply-extract air difference
None, no entry required – Room pressure (only for rooms with a
Z EM-AUTOZERO pressure controller)
Solenoid valve for automatic
zero balance

¹ Entries to relate to the air terminal unit


Subject to change · All rights reserved © TROX GmbH (9/2011)

Order example for the EASYLAB room controller


Circular volume flow controller type TVRD with acoustic
cladding and flange on both ends, nominal size 160.
EASYLAB control component with rapid response actuator,
supply air control for laboratories (extract air system)
Expansions: automatic zero balance.
Supply voltage 24 V AC.

Make: TROX
Type: TVRD-FL /160/ ELAB / RS / Z / LAB