18 Cell Holder

For the Cary 50

Part Number: 00 100727 00

Installation category I
Pollution degree 2
Safety class 3 (EN 61010-1)

Table of Contents
Introduction
Safety information
Specifications
Connections
Hose connections
Weight
Dimensions
Getting started
Unpacking
Installation
Using the Temperature Probe
Maintenance
Introduction
The 18 Cell Holder consists of two rows of nine staggered cell holders, enabling you
to perform multicell measurements for up to 18 cuvettes.

A circulating water bath, such as the PCB-150 (part number 79 100296 00), may be
used to maintain the temperature of the cells. You may also mount the optional
Temperature Probe accessory (part number 00 100408 00) on top of the 18 Cell
Holder. The two temperature probes can then be inserted in selected cuvettes to allow
direct monitoring of the temperature inside the cuvette. When not in use these probes
can be 'parked' in the channels alongside the front cells.

The 18 Cell Holder requires either the Accessory Cable kit (to connect it to the Cary
50) or, if you are using the 18 Cell Holder in a Dissolution system, the Dissolution
Accessory Cable kit (to connect it to the Cary 50 and the peristalic pump).

Before using this accessory, you must read the Safety Practices and Hazards section in
your Cary operation manual.

If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.

Safety information
Before using this accessory, you must read the Safety Practices and Hazards section in
your Cary operating instructions.

If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Specifications
For environmental and power supply requirements, refer to your Cary operating
instructions.

Connections

25-pin D-range connector on flying lead

Hose connections

2 barbs to fit 10 mm ID tubing if performing temperature controlled measurements

Weight

Packed: 9.0 kg (19.8 lb)


Unpacked: 4.4 kg (9.7 lb)

Dimensions (W x D x H)

Packed: 900 x 460 x 310 mm (35.5 x 18.1 x 12.2 in)


Unpacked: 510 x 125 x 110 mm (20.1 x 4.9 x 4.3 in)

Getting started
Unpacking

When you receive your 18 Cell Holder, unpack it and check that you have received all
of the necessary parts, referring to the packing list below:

Packing List Part number


18 Cell Holder accessory 00 100727 00
Hexagonal wrench
Specification sheet 85 101651 00

Caution
Inspect all parts for damage in transit. Any damage should be reported immediately.
Please refer to the accessory by model and part number.

Installation

You will need:

• The Accessory Cable kit or the Dissolution Accessory Cable kit (depending on
your analytical requirements)
• Hexagonal wrench, supplied with the 18 Cell Holder.
Refer to Spare parts for ordering information.

To install the 18 Cell Holder:

1. Remove the sample compartment front panel and lid from the instrument (if
fitted).

2. Clear the sample compartment of any sample holders.

3. Install the appropriate accessory cable in the sample compartment according to its
instructions (see Accessory Cable kit or Dissolution Accessory Cable kit).

4. Place the 18 Cell Holder in the sample compartment, aligning the locating holes
over the mounting pins in the floor of the sample compartment.

5. Slide the cell block forward and use the hexagonal wrench supplied to tighten the
securing screw.

6. Slide the cell block back into position ensuring that the cable from the 18 Cell
Holder is clear of the cell block. Connect the cable from the 18 Cell Holder to the 25
pin connector on the sample compartment wall.

Note
You should only slide the cell block manually when the accessory is not
powered i.e. plugged into the accessory cable.

The 18 Cell Holder is now installed and ready for use.

You should now install the Extended Sample Compartment.

Using the Temperature Probe

Refer to the Temperature Probe accessory instructions for details on fitting and using
the temperature probe with the 18 Cell Holder.

Maintenance
The outside of the 18 Cell Holder should be kept clean by wiping with a dampened
(with either water or alcohol) soft cloth. Do not use abrasive cleaning agents. If the
inside of the cuvette positions require cleaning, wrap a dampened soft cloth around a
metal rod or spatula and gently wipe the inside surfaces.

Also ensure that the printed circuit board is kept dry and free from dust.