AR-310A/AR-330A/AR-360A
ARK-510A/ARK-530A/ARK-560A
SERVICE MANUAL
September 2014
Pages in total: 218
XAR17ARDA001G/E
XAR17ARDA001G/E
Table of Contents
1. INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2. SAFETY PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3. SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
3 . 1 Classifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3 . 2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4. TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
5. SUBTROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . .19
5 . 1 Initial Screen is not Displayed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5 . 1 . 1 Power supply voltage check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5 . 1 . 2 Backlight of color LCD does not light up . . . . . . . . . . . . . . . . . . . . . . 21
5 . 2 Error Message Appears . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5 . 2 . 1 ERR001 (EEPROM error) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5 . 2 . 2 ERR002 (Clock error). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5 . 2 . 3 PD ERROR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5. 2. 3. 1 PD LED (30601-EA55) does not light up. . . . . . . . . . . . . . . 23
5 . 2 . 4 ERR011 (Time-out error in characters [OUT]) . . . . . . . . . . . . . . . . . . 24
5 . 2 . 5 ERR012 (Time-out error for receiving data [OUT]) . . . . . . . . . . . . . . 25
5 . 2 . 6 ERR013 (Receiving error [OUT]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 7 ERR014 (Receiving code error [OUT]). . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 8 ERR015 (Reply time-out [OUT]). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 9 ERR016 (DTR-DSR error [OUT]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 10 ERR017 (Data error [OUT]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 11 ERR018 (Command error [OUT]). . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 12 NO DAT (Data empty [OUT]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 13 ERR021 (Time-out error in characters [IN]) . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 14 ERR022 (Time-out error for receiving data [IN]) . . . . . . . . . . . . . . . . 25
5 . 2 . 15 ERR023 (Receiving error [IN]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 16 ERR024 (Receiving code error [IN]). . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 17 ERR025 (Reply time-out [IN]). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5 . 2 . 18 ERR026 (DTR-DSR error [IN]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5 . 2 . 19 ERR027 (Data error [IN]) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5 . 2 . 20 ERR028 (Command error [IN]). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5 . 2 . 21 ERR031 (Up-and-down tracking error). . . . . . . . . . . . . . . . . . . . . . . . 26
5 . 2 . 22 ERR032 (Left-to-right tracking error) . . . . . . . . . . . . . . . . . . . . . . . . . 27
5 . 2 . 23 ERR033 (Back-and-forth tracking error) . . . . . . . . . . . . . . . . . . . . . . 28
5 . 2 . 24 ERR034 (Up/down chinrest error) . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5 . 2 . 25 NO PAPER. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3
XAR17ARDA001G/E
4
XAR17ARDA001G/E
5
XAR17ARDA001G/E
8. ADJUSTMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103
8 . 1 Entering Adjustment Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
8 . 1 . 1 MENU mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
8 . 1 . 2 CALIB MODE SELECT (for AR). . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
8 . 1 . 3 ALIGNMENT CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
8 . 1 . 4 Measurement mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
8 . 1 . 5 Parameter settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
8 . 1 . 6 CALIB MODE SELECT (for KM) . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
8 . 1 . 7 EEPROM ERROR DISP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
8 . 2 Jig Attachment Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
8 . 2 . 1 Service optical adjustment jig (30670-5300) . . . . . . . . . . . . . . . . . . 107
8. 2. 1. 1 Joint jig (30670-1100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
8. 2. 1. 2 Reticle illumination jig (30670-1200) . . . . . . . . . . . . . . . . . 109
8 . 2 . 2 0-D model eye 2 (32177-0618) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
8 . 2 . 3 Dioptric tester (99104-G020) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
8 . 2 . 4 Simple calibration model eye (32107-6200). . . . . . . . . . . . . . . . . . . 111
8 . 2 . 5 Chinrest attachment jig (32107-1100) . . . . . . . . . . . . . . . . . . . . . . . 111
8 . 2 . 6 22-model eye jig (32171-0600) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
6
XAR17ARDA001G/E
7
XAR17ARDA001G/E
9. SUPPLEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .177
9 . 1 Appearance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
9 . 1 . 1 Front view. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
9 . 1 . 2 Rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
9 . 2 Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
9 . 3 Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
9 . 3 . 1 Complete wiring diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
9. 3. 1. 1 30602-E000 (AR-330A and ARK-530A) . . . . . . . . . . . . . . 185
9. 3. 1. 2 30601-E000 (AR-310A and ARK-510A) . . . . . . . . . . . . . . 186
9. 3. 1. 3 30603-E000 (AR-360A) and 30703-E000 (ARK-560A) . . . 187
9 . 3 . 2 Partial writing diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
9 . 4 Error Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202
9 . 5 Parameter List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
9 . 5 . 1 AR-310A/AR-330A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
9 . 5 . 2 AR-360A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206
9 . 5 . 3 ARK-510A/ARK-530A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 208
9 . 5 . 4 ARK-560A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210
9 . 6 Jig List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 212
9 . 7 Adjustment Item List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213
9 . 7 . 1 ASSYs, boards, and adjustment items. . . . . . . . . . . . . . . . . . . . . . . 213
9 . 7 . 2 Electrical components and adjustment items . . . . . . . . . . . . . . . . . . 214
9 . 7 . 3 Mechanical components and adjustment items . . . . . . . . . . . . . . . . 215
9 . 8 Screw List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216
8
XAR17ARDA001G/E
1 INTRODUCTION
This Service Manual is for the NIDEK AUTO REFRACTOMETER and AUTO REF/KERATOME-
TER models: AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A.
For correct service, thorough understanding of the contents in this manual is required prior to the
service.
Refer to the Operator’s Manual and Parts List for the AR-310A, AR-330A, AR-360A, ARK-510A,
ARK-530A, and ARK-560A.
In case the device cannot be repaired according to the procedures described in this manual,
please report the serial number of the device and details of the symptom or symptoms.
Refer to the Service Manual Annex for important changes.
2 SAFETY PRECAUTIONS
2 . General precautions
1 ) Repairs by the Service Manual or Service Manual Annex must be performed by per-
sonnel who have undergone training at NIDEK.
2 ) Conduct repairs according to the procedures described in the Service Manual or Ser-
vice Manual Annex.
3 ) Never wipe the covers using an organic solvent such as paint thinner.
Doing so may ruin the surface and impair the appearance of the device.
4 ) When conducting repairs, turn off the power switch and disconnect the power cord
from the wall outlet unless the power needs to be on.
5 ) A part emitting invisible laser beam (Class 3B laser) is used inside the device. (How-
ever, the device is classified as a Class 1 laser product indicating that the part does
not emit hazardous irradiation to the eye.)
4 . Maintenance cautions
1 ) Never look directly at the beam from the phototransmitter ASSY (invisible laser: Class
3B laser) or block it with any part of the body.
2 ) Take antistatic precautions prior to repair.
3 ) Take proper care against electric shock during repair.
4 ) Tighten or loosen screws with proper tools.
5 ) Never drop parts or screws inside the device.
6 ) Prepare storage cases so as not to lose the removed screws or parts.
5 . Adjustment precautions
1 ) Confirm that the following conditions are met before installation.
a . A place which is level and stable without vibration and shock
b . No exposure to direct sunlight or ultraviolet rays
c . A place with minimal external light sources
d . No exposure to water
e . A place where temperature and humidity meet the specifications for use
f . Dust-free and smoke-free
2 ) Never use the adjustment jigs for purposes other than those intended.
3 SPECIFICATIONS
3.1 Classifications
1 . Type of protection against electrical shock: Class I
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A are classi-
fied as Class I devices.
A Class I device is one in which, in addition to basic insulation, protection against
electric shock is provided by connecting the device to a grounding source to prevent
accessible metal parts from becoming electrically charged, should the basic insulation
fail.
2 . Degree of protection against electrical shock: Type B applied part
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A are classi-
fied as devices with a Type B applied part.
A Type B Applied Part provides a particular degree of protection against electric
shock, particularly regarding the following:
- allowable leakage currents
- reliability of the protective ground connection (if applicable)
3 . Conformity to electromagnetic compatibility standard
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A conform
to the EMC standard of IEC60601-1-2: 2007.
4 . Degree of protection against harmful ingress of water: IPX0
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A provide
no protection against ingress of water with harmful effects.
Avoid splashing water or other liquids on or near these devices.
5 . Degree of safety for application in the presence of a flammable anesthetic mixture with air,
or with oxygen or nitrous oxide.
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A are not
suitable for use in the presence of a flammable anesthetic mixture with air, or with
oxygen or nitrous oxide.
Do not operate the device near flammable type materials.
6 . Method(s) of sterilization or disinfection recommended by the manufacturer
The forehead rest and chinrest can be cleaned with a cloth dampened with rubbing
alcohol as necessary.
7 . Mode of operation
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A are classi-
fied as continuous operating devices.
8 . Installation and use
The AR-310A, AR-330A, AR-360A, ARK-510A, ARK-530A, and ARK-560A are classi-
fied as stationary devices.
3.2 Specifications
1 . Objective measurement of refractive error (AR measurement)
Spherical power (S) -30.00 to +25.00 D (VD = 12 mm)
0.01/0.12/0.25 D increments
Cylindrical power (C) -12.00 to +12.00 D
0.01/0.12/0.25 D
Cylinder axis (A) 0 to 180°
1°/5° increments
Vertex distance 0/10.5/12/13.75/15/16.5 mm
Minimum pupil diameter ø2 mm
measurable
Relaxation of Auto fogging system
accommodating eye
Chart Scenery chart
2 . PD measurement
Measurement range 30 to 85 mm
(Near PD = 28 to 80 mm, Near working distance = 40
cm)
1 mm increments
3 . Measurement of corneal curvature (KM measurement) (ARK-510A, ARK-530A, and ARK-
560A)
Corneal curvature radius 5.00 to 13.00 mm
(R1, R2, AVE) 0.01 mm increments
Corneal refractive power 25.96 to 67.50 D (n = 1.3375)
(R1, R2, AVE) 0.01/0.12/0.25 D increments
n = 1.3375/1.336/1.332
Corneal cylindrical power -12.00 to +12.00 D
(CYL) 0.01/0.12/0.25 D increments
Cylinder axis (A) 0 to 180°
1°/5°increments
KM measurable area ø3.3 mm (for mire ring/corneal curvature radius of 7.7
mm)
ø6.0 mm (for four points/corneal curvature radius of
7.7 mm)
Size of mire ring ø2 mm and ø3.3 mm on the cornea (for corneal curva-
ture of 7.7 mm)
Chart Scenery chart
4 . Sagittal radius measurement
Measurable angle Within 25° from center (SUP., INF, TEMP, and NAS.)
Sagittal curvature 5.00 to 13.00 mm
0.01 mm increments
Eccentricity (E) -4.10 to 2.05
5 . CS measurement (ARK-510A, ARK-530A, and ARK-560A)
Measurement range 10.0 to 14.0 mm
0.1 mm increments
6 . PS measurement (ARK-510A, ARK-530A, and ARK-560A)
Measurement range 1.0 to 10.0 mm
0.1 mm increments
11 . Other functions
Auto shot The measurement automatically starts when the eye
is properly aligned.
Alignment/observation 5.7-inch color LCD panel
method
Printer Thermal line printer with auto cutter
Width 58 mm
Interface connectors RS-232C: Two ports (IN/OUT: two-phase connectable
system)
USB: One port
LAN: One port
4 TROUBLESHOOTING
No
Is the initial screen displayed? 5.1 Initial Screen is not Displayed (p19)
Yes
Yes
Does an error message appear? 5.2 Error Message Appears (p22)
No
No
Can the function buttons be operated? 5.3 Function Buttons are not Operational (p40)
Yes
Does rotating the joystick move the measuring unit No 5.5 Rotating Joystick does not Move Measuring
up and down? Unit Up or Down (p42)
Yes
No
Does the chinrest move up and down smoothly? 5.6 Chinrest does not Move Up or Down (p44)
Yes
No
Can printing be performed? 5.9 Printing cannot be Performed (p60)
Yes
5 SUBTROUBLESHOOTING
Yes
Does the backlight of the color LCD (30601-E005) No 5.1.2 Backlight of color LCD does not light up
light up? (p21)
Yes
End
Yes
Is the voltage between the P001 1st and 3rd pins No Replace the primary power supply unit (30601-
on the primary power supply unit (30601-EA01) the
EA01).
same as the input voltage (see 9.3 [p185])?
Yes
Are the voltages between the J51 4th and 1st, 5th
and 2nd, and 6th and 3rd pins on the switching No Replace the Switching power supply (80602-
power supply (80602-00102) DC +15 V (see 9.3 00102) (see 7.3.20 [p96]).
[p185])?
Yes
Are there any breaks in the power supply output Yes Replace the power supply output cable (30601-
cable (30601-EA51) (see 9.3 [p185])? EA51).
No
Are there any breaks in the BA02 power supply Yes Replace the BA02 power supply cable (30601-
cable (30601-EA11) (see 9.3 [p185])? EA11).
No
End
No
No
No
Yes
End
Yes
Is the problem resolved? End
No
No
Does ERR002 occur? End
Yes
5.2.3 PD ERROR
Yes
Is the PD phototransmitter soiled? Clean the PD phototransmitter.
No
No
Is the problem resolved? End
Yes
End
End
End
5.2.25 NO PAPER
Replace the printing paper. (Refer to the operator's manual.)
Yes
Yes
Is the problem resolved? End
No
Are there any breaks in the printer power supply Yes Replace the printer power supply cable (30601-
cable (30601-EA31) (see 9.3 [p185])? EA31).
No
Yes
Is the problem resolved? End
No
Yes
Is the problem resolved? End
No
No
Are there any breaks in the sensor motor (30601- Yes Replace the Sensor motor (30601-E035) (see
EA35) (see 9.3 [p185])? 7.3.19 [p95]).
No
Yes
Is the problem resolved? End
No
Are there any breaks in the CL sensor (30703- Yes Replace the CL sensor (30703-EA38) (see 7.3.9
EA38) (see 9.3 [p185])? [p87]).
No
Yes
Is the problem resolved? End
No
Yes
Is the problem resolved? End
No
End
Yes
Does ERR502 appear? Replace the card.
No
End
Yes
Is the problem resolved? End
No
No
Does ERR602 occur? Replace the USB device.
Yes
Yes
Is the problem resolved? End
No
Yes
No
Is the network setting correct? Correct the network setting (see 8.16 [p174]).
Yes
Yes
Is the problem resolved? End
No
Yes
No
Is the network setting correct? Correct the network setting (see 8.16 [p174]).
Yes
Yes
Are there any breaks in the LAN cable? Replace the LAN cable.
No
Yes
Is the problem resolved? End
No
No
Is write access to the shared folder enabled? Enable write access to the shared folder.
Yes
Yes
No
Is the network setting correct? Correct the network setting (see 8.16 [p174]).
Yes
Yes
Are there any breaks in the LAN cable? Replace the LAN cable.
No
No
Is the destination PC connected by LAN? Connect the destination PC by LAN.
Yes
Is the network setting of the destination PC No Correct the network setting of the destination PC
correct? (see 8.16 [p174]).
Yes
Is the password entered in the password field of the No Enter the correct password in the password field of
network setting correct? the network setting (see 8.16 [p174]).
Yes
Is the user account setting of the destination PC No Correct the user account setting of the destination
correct? PC.
Yes
Does a shared folder with the same name as the No Create in the destination PC a shared folder with
one set in the device exist in the destination PC? the same name as the one set in the device.
Yes
Yes
Yes
Are there any breaks in the LAN cable? Replace the LAN cable.
No
No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).
Yes
End
5.4 Measuring Unit does not Move Smoothly along Base Unit
Clean the sliding plate (32105-M102).
Yes
Is the problem resolved? End
No
Does the measuring unit move smoothly to the No Replace the Sliding plate (32105-M102) (see 7.4.3
right, left, forward, and backward? [p99]).
Yes
0.8 V or more
Yes
Are there any breaks in the BA02 power supply Yes Replace the BA02 power supply cable (30601-
cable (30601-EA11) (see 9.3 [p185])? EA11).
No
No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).
Yes
End
Is there any breaks in the chinrest (30601-EA53) Yes Replace the Chinrest (30601-EA53) (see 7.3.12
(see 9.3 [p185])? [p89]).
No
No
No
Is the mire ring displayed? 5.7.2 Mire ring is not displayed (p49)
Yes
No
Is the focus LED (30601-EA42) displayed? 5.7.3 Focus LED is not displayed (p49)
Yes
No
Is the chart visible? 5.7.4 Chart is not visible (p50)
Yes
Does the measurement start when the start button No 5.7.6 Measurement does not start even when start
is pressed in manual mode? button is pressed (p54)
Yes
No
Is the PD measurement data normal? 5.7.9 PD measurement data is not normal (p57)
Yes
No
Is the KM measurement data normal? 5.7.10 KM measurement data is not normal (p58)
Yes
No
Can the KM periphery be measured? 5.7.11 KM periphery cannot be measured (p58)
Yes
No
Is the CS measurement data normal? 5.7.12 CS measurement data is not normal (p59)
Yes
No
Is the PS measurement data normal? 5.7.13 PS measurement data is not normal (p59)
Yes
End
No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).
Yes
End
End
Are there any breaks in the focus LED (30601- Yes Replace the Focus LED(30601-EA42) (see 7.3.10
EA42) (see 9.3 [p185])? [p88]).
No
End
Are there any breaks in the chart LED (30601- Yes Replace the Chart LED (30601-EA76) (see 7.3.14
EA76) (see 9.3 [p185])? [p91]).
No
No
Is auto tracking icon 3D displayed? Display auto tracking icon 3D.
Yes
No
Is the auto tracking icon displayed? Display the auto tracking icon.
Yes
No
Is auto tracking icon 3D displayed? Display auto tracking icon 3D.
Yes
5.7.6 Measurement does not start even when start button is pressed
Activate manual mode.
No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).
Yes
End
5.7.7 Measurement does not start even when alignment and focus are achieved
Yes
Does an error appear? 5.2 Error Message Appears (p22)
No
Yes
Does the measurement start automatically? End
No
Does the phototransmitter ASSY (30601-3310) No Replace the Phototransmitter ASSY (30601-3310)
light up? (see 7.1.3 [p70]).
Yes
No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).
Yes
End
Does the phototransmitter ASSY (30601-3310) No Replace the Phototransmitter ASSY (30601-3310)
light up? (see 7.1.3 [p70]).
Yes
Yes
End
No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73]).
Yes
End
No
Is the problem resolved? Replace the Main board (see 7.2.1 [p73])..
Yes
End
No
Yes
Is the problem resolved? End
No
Yes
Is the problem resolved? End
No
Yes
Is the problem resolved? End
No
6 REMOVAL PROCEDURE
2%
2%
2%
2%
5$
5$
3&
7 REPLACEMENT PROCEDURE
7.1 ASSYs
7.1.1 Joystick ASSY (30601-2600)
Replacement part: 30601-2600
1 . Remove the rear body cover ASSY (30601-7000) (see 6.6 [p65]).
2 . Remove the GND cable of the joystick.
3 . Unscrew BS3 × 6 (n = 4) to replace the
joystick ASSY (30601-2600).
4 . Reassemble the parts in the reverse order.
%6
$5
%6
7.2 Boards
7.2.1 Main board
Replacement part: 30601-BA01P (for the AR-310A)
30602-BA01P (for the AR-330A)
30603-BA01P (for the AR-360A)
30701-BA01P (for the ARK-510A)
30702-BA01P (for the ARK-530A)
30703-BA01P (for the ARK-560A)
1 . Display the PARAMETER SETTING screen (see 8.1.5 [p105]).
2 . Press the print button to print out the parameter settings of the device.
3 . Remove the tracking board (30601-BA02) (see 7.2.2 [p74]).
4 . Remove the rear body cover ASSY (30601-7000) (see 6.6 [p65]).
5 . Unscrew AS3 × 6 (n = 4) to remove the
guard (30601-M225). $#
6 . Disconnect all connectors on the main
board (30601-BA01).
7 . Unscrew BS3 × 8 (n = 2) to remove the
#5
main board (30601-BA01) together with
the double locking spacers (80745-01103
[n = 4]).
8 . Remove the double locking spacers $5
(80475-01103 [n = 4]).
9 . Replace the main board (30601-BA01).
10. Reassemble the parts in the reverse order.
11 . After the replacement, turning on power to
the device displays the following errors.
1 ) The AR-310A, AR-330A, AR-360A: ERR065
2 ) The ARK-510A, ARK-530A, ARK-560A: ERR066
* Do not turn off power to the device until the error is displayed.
* If the model of the replaced main board differs from that of the device, ERR062 to
ERR064 occur, and the device does not start up.
12. Turn the power off, then turn it on again.
13. Enter MENU mode, and select 9. PARAM-
2#4#/'6'45'66+0)
ETER SETTING (see 8.1.1 [p103]).
14. Press the up or down button to select the ,#2#0
(14'+)0
desired destination language. 810*1((
15. Press the set button, and confirm that OK 1%7.75
+0%
is displayed at the bottom of the screen. %*+0#
16. Press the exit button twice.
17. Turn the power off, and turn it on again.
18. Display the PARAMETER SETTING
screen (see 8.1.5 [p105]). 1-
19. Set the parameters based on the parame-
ter settings of the device that was printed
out earlier.
$#
6 . After the replacement, turning on power to the device displays the following errors.
1 ) The AR-310A, AR-330A, AR-360A: ERR065
2 ) The ARK-510A, ARK-530A, ARK-560A: ERR066
* Do not turn off power to the device until the error is displayed.
* If the model of the replaced main board differs from that of the device, ERR062 to
ERR064 occur, and the device does not start up.
7 . Turn the power off, and turn it on again.
8 . Rewrite the EEPROM data (see 8.15 [p173]).
$5
$#
2%
$#
26
$#
$5
26
7.2.10 PD boards
7.2.10.1 BA08 PD CCD board (32105-BA08)
Replacement part: 32105-BA08
1 . Remove the measuring unit front cover ASSY (30601-6300) (see 6.3 [p64]).
2 . Disconnect all connectors on the BA08 PD
CCD board (32105-BA08).
3 . Unscrew BS3 × 6 (n = 2) to replace the
BA08 PD CCD board (32105-BA08).
* The BA08 PD CCD board (32105-
BA08) must be used in combination
with the driver board (30601-BA03),
artwork number PC5433 (see 7.2.3
[p75]).
4 . Reassemble the parts in the reverse order.
5 . Perform PD CALIBRATION (see 8.9
[p163]). $#
$5
2%
'#
$5
'# $5
$5
'#
$5
'#
$5
'#
$5
D motor (30601-EA28).
5 . Reassemble the parts in the reverse order.
#5
6 . Adjust the position of the bracket (30601-
M211) by lightly pressing the belt (32156- **
M098) so that the belt bends about 3 - 4
mm.
7 . Adjust the tension of the belt (32156-
M098) so that the tension after the
replacement becomes almost the same as
before.
8 . Perform AUTO TRC ADJUST (see 8.11 [p168]).
'#
%-
%-
'#
%-
%-
5$
$5
4 . Unscrew SB3 × 10 (n = 3) to remove the
chinrest cover (32105-M041).
5$
5 . Unscrew PC3 × 6.
6 . Unscrew HH3 × 5 (n = 3), and move the
bearing (32105-M048) upward. 2%
7 . Unscrew HH3 × 5 and CK 2 × 5 (n = 2) to
remove the motor brace (32105-M080).
8 . Unscrew SB3 × 6 (n = 2) and PC2 × 5 (n =
2) to replace the chinrest (30601-EA53).
9 . Reassemble the parts in the reverse order.
5$
**
5 . Use the steps below to adjust the LED intensity when the previous chart LED (30601-
EA76A) of the devices with serial numbers earlier than those shown in the following table is
replaced with the new one (30601-EA76B).
1 ) Serial numbers of post-design change devices
Countries other than
Model USA China
USA and China
AR-310A 120001— 150621— 136114—
AR-330A 220001— 250005— 230659—
AR-360A 820001— 850001— 830312—
ARK-510A 320342— 350157— 334477—
ARK-530A 421330— 450004— 431698—
ARK-560A 920348— 950001— 930622—
3 ) Enter MENU mode, and select 7. LED ADJUST MODE (see 8.1.1 [p103]).
0
'#
26
$5
$5
26
26
5$
**
6%
6%
8 ADJUSTMENT
. 4
. 4
. 4
. 4
24'5556#46$76610
4
.
5
5
%
%
##
4
-
5
4
%
4
#
4
-
5
4
%
4
#
' $#
, , ,
#4&,$#':'$1#4&
,
,
,
'
,
,
$#
6 . Connect the connector of the EA39 extension cable (32557-EA11) to J1 of the mire ASSY.
7 . Connect the connector of the EA39 extension cable (32557-EA11) to J9 on the driver board
(30601-BA03).
,
,
,
,
,
, ,
'#
,
$# ,
'#
6Q
, 6Q
'#
6Q
, 6Q$#
,
(%
)
button.
6 . Turn the turret plate (30670-M071) of the
chart test jig (30670-2300) to select the
model eye convex lens (32527-G001).
7 . Perform alignment of the model eye con- )
)
vex lens (32527-G001) so that the focus-
ing indicator displays the optimum
condition.
8 . Fasten the measuring unit with the locking
lever.
9 . Fasten the unit while the focusing indica- )
tor is in the optimum condition.
* Maintain the alignment of the convex ) Marking-off line
lens during adjustment. ⟌ᦠ߈✢
Reticle
ࣞࢳࢡࣝ༑Ꮠ⥺
4 ) Use the steps below to remove the reticle illumination jig (30670-1200) from the
devices with serial numbers earlier than those shown in the following table.
USA Countries other than USA
(S/N) (S/N)
AR-310A 120001 134438
AR-330A 220001 230544
AR-360A 820001 830108
ARK-510A 320182 333207
ARK-530A 420905 431269
ARK-560A 920153 930260
**
5 ) Use the steps below to remove the reticle illumination jig (30670-1200) from the
devices with serial numbers shown in the following table and later.
USA Countries other than USA
(S/N) (S/N)
AR-310A 120001 134438
AR-330A 220001 230544
AR-360A 820001 830108
ARK-510A 320182 333207
ARK-530A 420905 431269
ARK-560A 920153 930260
++
5 . Adjust the position of the observation CCD camera as follows:
1 ) Attach 0-D model eye 2 (32177-0618) to the joint jig (30670-1100) (see 8.2.2 [p110]).
0
2 ) Press the next button to display %%&215+6+10
1$5'4
the 1-2 CCD POSITION (OBSER.)
screen. 470
3 ) Press the function buttons to display
RUN and LED. .'& 0
* Adjust the brightness of the SLD
light with the up or down
button so that adjustment can 56#46595#8'
be performed easily.
AR SLD light
#45.& శ 0-D
&model
ᮨဳ⌒ eye
2 SLD
5.&light brightness
శߩࠆߐ
5 ) Loosen SB3 × 6 (n = 2) to adjust the
position of the camera ASSY
(30601-3340) so that the SLD light is
aligned to the centering box ( ) of
the observation CCD camera.
5$
%%&215+6+10
1$5'4
470
.'& 0
56#46595#8'
AR SLD light
#45.& శ
6 ) After the adjustment is complete, remove 0-D model eye 2 (32177-0618) (see 8.2.2
[p110]).
6 . Adjust the inclination of the observation CCD camera as follows:
1 ) Attach the reticle illumination jig (30670-1200) (see 8.2.1.2 [p109]).
0
2 ) Press the next button to display *14+<106#.&+4'%6+10
the 1-3 HORIZONAL DIRECTION
screen. 470
* Press the function buttons to
display RUN and LED, and .+0' 0
adjust the brightness of the reti-
cle illumination jig with the up
or down button so that
adjustment can be performed
easily.
3 ) Press the function buttons to display
RUN and LINE.
8 ) After the adjustment is complete, press the exit button and turn off power to the
device.
7 . Reassemble the removed units and covers.
8 . Perform ALIGNMENT CALIBRATION (see 8.6 [p129]).
9 . Perform KM calibration (see 8.8 [p156]).
* Applies only to the ARK-510A, ARK-530A, and ARK-560A.
10. Perform AUTO TRC ADJUST (see 8.11 [p168]).
11 . Perform CS/PS calibration (see 8.10 [p167]).
* Applies only to the ARK-510A, ARK-530A, and ARK-560A.
$4#-' 0
AR SLD
#45.& light
శ
6 . Loosen SB3 × 6 (n = 2).
5$
AR SLDశ
#45.& light
0
8 . Press the next button to display the 5- %%&215+6+10
1$5'4
2 CCD POSITION (OBSER.) screen.
9 . Press the function buttons to display RUN 470
and LED.
10. Confirm that the centering box ( ) of the .'& 0
observation CCD camera and the SLD
light are in alignment.
11 . After the adjustment is complete, press
56#46595#8'
the exit button and turn off power to the
device.
12. Reassemble the removed units and cov- AR#45.&
SLD light
శ
ers.
13. Perform AR CALIBRATION (see 8.7 [p130]).
$4#-' 0
AR SLD light
#45.& శ
#45.&
AR SLD light శ
9 . After the adjustment is complete, press the exit button and turn off power to the device.
10. Reassemble the removed units and covers.
11 . Perform CHART CALIBRATION (see 8.7.5 [p135]).
%
%
4 . Follow the steps below to check the operating noise of the motor when the subjective cylin-
der ASSY is operated.
1 ) CL1 side of the subjective cylinder
ASSY.
6
15
6 . Move the subjective focus base ASSY
Measuring
(30703-3300) back and forth so that the window
chart can be seen through the measuring
window.
Uniform
8 . Follow the steps below to adjust the optical axis of the chart.
1 ) Move the focus base ASSY (30601-
3300) to the position where the chart
can be seen through the dioptric tes-
ter.
2 ) Confirm that the chart does not shift
to the right or left at any position of -
12 D, 0 D, or +12 D.
$5
3 ) If the chart shifts to the right or left,
loosen BS3 × 6 (n = 2) to adjust the
position of the chart mirror ASSY
(30601-3270).
4 ) Follow the steps below so that the center of the dioptric tester target grid is inside the
inner circle of the clock dial chart (the balloon of the scenery chart) at any position of -
12 D, 0 D, and +12 D.
a . Loosen HH3 × 3 to turn chart
lens holder 2 (30601-M317) of
the chart lens 2 ASSY.
**
**
9 . After the adjustment is complete, press the exit button and turn off power to the device.
10. Reassemble the removed units and covers.
11 . Perform chart calibration (see 8.7.5 [p135]).
%#.+$4#6+10(+0+5*
%#.+$4#6+10(+0+5*
8.7 AR CALIBRATION
8.7.1 AR CALIBRATION
1 . Attach the 22-model eye jig (see 8.2.6 [p112]).
2 . Display the CALIB MODE SELECT
screen (for AR), and select AR (see 8.1.2 #4%#.+$4#6+10
[p104]). /1&'.';'07/$'4
%7561/
3 . Select CUSTOM with the up or down
or left button.
2 ) Perform alignment of the model eye,
and press the start button to start
the AR CALIBRATION. 24'5556#46
3 ) Change the displayed diopter with
6 . After AR CALIBRATION is complete, press the exit button and turn off power to the
device.
4
.
5
5
%
%
##
2 ) For the ARK-510A, ARK-530A, and
ARK-560A, the screen as shown to ᵏ %;. 㧸
the right appears.
4
-
5
4
%
4
#
8 . Press the auto button to select 3D auto tracking and auto shot .
8.7.2 AR AXIS
1 . Attach the cylinder axis adjustment jig (32177-1500) (see 8.2.7 [p112]).
2 . Display the CALIB MODE SELECT screen (for AR), and select AR AXIS (see 8.1.2 [p104]).
3 . Align the marking-off line to the three hori-
zontal lines of the cylinder axis adjustment #4#:+5%#.+$4#6+10
jig.
&')
4 . Set the same numeric value as the nomi-
nal value for 180º that is indicated on the
base of the cylinder axis adjustment jig
with the up or down or button.
24'5556#46
12. After the adjustment is complete, press the exit button and turn off power to the device.
13. If the measured values are not within the specifications, perform AXIS OFFSET (see 8.7.3
[p133]).
5#8'
9 . After the adjustment is complete, press the exit button twice and turn off power to the
device.
5#8'
5 . When the measured value for 0.00 D is -4.00 D, changing the numeric value results in as
below.
1 ) When the numeric value is changed to +0.50 D, the measured value becomes -3.50 D.
2 ) When the numeric value is changed to -0.50 D, the measured value becomes -4.50 D.
6 . After the adjustment is complete, press the exit button twice and turn off power to the
device.
2 ) Press the down button, and follow the steps below to adjust the +12.12 D posi-
tion.
a . Turn the turret plate to select
%*#46%*'%-
the +12.00 D lens (33510-
G033).
&
3 ) Press the down button, and follow the steps below to adjust the -12.00 D position.
a . Turn the turret plate to select
%*#46%*'%-
the -12.00 D lens (33510-
G018). &
7 . Press the exit button twice and turn off the power switch.
8 . Remove the chinrest attachment jig (32107-1100) and chart test jig (32145-2100 or 30670-
2300).
button.
5#8'
2 ) Press the down button, and follow the steps below to adjust the +12.12 D posi-
tion.
a . Turn the turret plate to select
%*#46%#.+$4#6+10
the +12.00 D lens (33510-
G033).
&
3 ) Press the down button, and follow the steps below to adjust the -12.00 D position.
a . Turn the turret plate to select
%*#46%#.+$4#6+10
the -12.00 D lens (33510-
G018). &
2 ) Press the down button, and follow the steps belows to check the +12.12 D posi-
tion.
a . Turn the turret plate to select %*#46%*'%-
the +12.00 D lens (33510-
G033).
&
3 ) Press the down button, and follow the steps below to check the -12.00 D position.
a . Turn the turret plate to select
%*#46%*'%-
the -12.00 D lens (33510-
G018). &
8.7.5.2 AR-360A/ARK-560A
8.7.5.2.1 Chart test jig
* If the adjustment jig is not available, return the device to the factory without performing
adjustment.
1 . Attach the chart test jig (30670-2300) (see 8.2.9 [p114]).
2 . Display the CALIB MODE SELECT screen (for AR), and select CHART (see 8.1.2 [p104]).
3 . Perform STAR CALIB as follows:
* Pressing the page down button five times returns the subjective cylinder
ASSY to the original position.
5 ) If the chart is not focused at every angle, repeat adjustment starting from Step 2).
0
6 ) Press the NEXT button to display the THETA2 CALIB screen.
5 ) Perform adjustment with the left- or right+ button so that the whole clock dial
chart can be seen in uniform clarity.
6 ) Perform Steps 2) to 5) twice.
7 ) Press the SAVE button to save the position.
0
* Pressing the SAVE button displays the NEXT button.
0
8 ) Press the NEXT button to display the THETA1 CHECK screen.
6 . Perform THETA1 CHECK as follows:
1 ) Turn the turret plate to select a position without any lens.
1 ) Select STAR CHECK with the up or down button, then press the execute
button.
2 ) Follow the steps below to check the 0.00 D position.
a . Turn the turret plate to select a position without any lens.
b . Perform adjustment with the 56#4%*'%-
left- or right+ button so &
that the whole clock dial chart
is focused.
c . Press the start button, and
confirm that the displayed
numeric value is within 0.00
±0.25 D.
3 ) Press the down button, and follow the steps below to check the +12.12 D posi-
tion.
a . Turn the turret plate to select 56#4%*'%-
the +12.00 D lens (33510-
G033).
&
b . Perform adjustment with the
4 ) Press the down button, and follow the steps below to check the -12.00 D position.
a . Turn the turret plate to select 56#4%*'%-
the -12.00 D lens (33510-
G018). &
b . Perform adjustment with the
0
left- or right+ button so
that the whole clock dial chart
is focused.
c . Press the start button, and
confirm that the displayed
numeric value is -12.00 ±0.25
D.
0
12. Press the next button to display the BALOON CALIB screen.
5#8'
2 ) Press the down button, and follow the steps below to adjust the +12.12 D posi-
tion.
a . Turn the turret plate to select the +12.00 D lens (33510-G033).
5#8'
3 ) Press the down button, and follow the steps below to adjust the -12.00 D position.
a . Turn the turret plate to select the -12.00 D lens (33510-G018).
0
4 ) Press the next button to display the BALOON CHECK screen.
2 ) Press the down button, and follow the steps below to check the +12.12 D posi-
tion.
a . Turn the turret plate to select the +12.00 D lens (33510-G033).
3 ) Press the down button, and follow the steps below to check the -12.00 D position.
a . Turn the turret plate to select the -12.00 D lens (33510-G018).
5#8'
4 ) Press the down button, and follow the steps below to adjust the +12.12 D posi-
tion.
a . Turn the turret plate to select the +12.00 D lens (33510-G033).
5#8'
5 ) Press the down button, and follow the steps below to adjust the -12.00 D position.
a . Turn the turret plate to select the -12.00 D lens (33510-G018).
5 ) Perform adjustment with the left- or right+ button so that the whole clock dial
chart can be seen in uniform clarity.
6 ) Perform Steps 2) to 5) again.
7 ) Press the SAVE button to save the position.
0
* Pressing the SAVE button displays the next button.
0
8 ) Press the NEXT button to display the THETA1 CHECK screen.
6 . Perform THETA1 CHECK as follows:
1 ) Turn the turret plate to select the position without any lens.
0
5 ) Press the next button to display the BALOON CALIB screen.
12. Perform BALOON CALIB as follows:
1 ) Follow the steps below to adjust the 0.00 D position.
a . Turn the turret plate to select a position without any lens.
2 ) Press the down button, and follow the steps below to adjust the +12.12D position.
a . Turn the turret plate to select the +12.00 D lens (33510-G033).
3 ) Press the down button, and follow the steps below to adjust the -12.00 D position.
a . Turn the turret plate to select the -12.00 D lens (33510-G018).
0
4 ) Press the next button to display the BALOON CHECK screen.
13. Perform BALOON CHECK as follows:
1 ) Follow the steps below to check the 0.00 D position.
a . Turn the turret plate to select a position without any lens.
3 ) Press the down button, and follow the steps below to check the -12.00 D position.
a . Turn the turret plate to select the -12.00 D (33510-G018) position.
8.8 KM CALIBRATION
* Applies only to the ARK-510A, ARK-530A, and ARK-560A.
8.8.1 KM CALIB
1 . Attach the KM calibration jig (30670-0300) (see 8.2.11 [p116]).
2 . Display the CALIB MODE SELECT screen (for KM), and select KM CALIB (see 8.1.6
[p106]).
* When the software version is earlier than Ver. 2.03, press the reset button.
* When the software version is Ver. 2.03 or later, the reset button is not displayed.
3 . Perform KM CALIBRATION as follows:
1 ) Align the relevant steel ball of the
KM calibration jig (30670-0300) in -/%#.+$4#6+10
the order of the following numeric OO
values so that the focusing indicator
displays the optimum condition.
a . 7.94 mm
b . 5.95 mm
c . 9.13 mm %#.+$4#6+10(+0+5*
2 ) Press the start button to perform KM
CALIBRATION.
* Perform calibration of all three
steel balls.
3 ) Press the SAVE button to save the setting.
7 . If the center measurement values are not within the specifications, perform correction as
follows:
* When the measurement values for each steel ball of the KM calibration jig (30670-
0300) differ by the same amount, change all correction factors by the same numeric
value.
* When the measurement values for only some steel balls of the KM calibration jig
(30670-0300) differ by the same amount, change the correction factors close to the
radiuses of such steel balls.
1 ) Select the steel ball to be changed
-/#&,756
with the up or down button. #&,756)#+0
2 ) Change the numeric value with the 4
4
4
right or left button. 4
* Perform correction by taking the
median value of specifications
as a guide.
* Change the numeric value by
0.0010 (as a guide) for each 5#8' -/#:/+0&+((
4
correction of 0.01 mm. 4
3 ) Press the start button, and confirm
the measurement values.
4 ) Press the SAVE button to save the setting.
8 . Repeat Steps 5 to 6 until the center measurement values for each steel ball of the KM cali-
bration jig are within the specifications.
9 . After the adjustment is complete, press the exit button to return to the KM ADJUST
screen.
10. Perform ADJUST PHI6 GAIN (see 8.8.2.2 [p159]).
7 . If the peripheral measurement values are not within the specifications, perform correction
as follows:
* When the measurement values for each steel ball of the KM calibration jig (30670-
0300) differ by the same amount, change all correction factors by the same numeric
value.
* When the measurement values for only some steel balls of the KM calibration jig
(30670-0300) differ by the same amount, change the correction factors close to the
radiuses of such steel balls.
1 ) Select the steel ball to be changed
-/#&,756
with the up or down button. #&,756)#+0
#&,7562*+)#+0
2 ) Change the numeric value with the 4
4
right or left button. 4
4
* Perform correction by taking the
median value of specifications
as a guide.
* Change the numeric value by
-/#:/+0&+((
5#8' 2/#:/+0&+((
0.0010 (as a guide) for each 4
correction of 0.01 mm. 4
3 ) Press the start button, and confirm
the measurement values.
4 ) Press the SAVE button to save the setting.
8 . Repeat Steps 5 to 6 until the peripheral measurement values for each steel ball of the KM
calibration jig are within the specifications.
8 . If the center KM measurement values are not within the specifications, perform the correc-
tion as follows:
11 . Press the exit button and turn off the power switch.
8.9 PD CALIBRATION
8.9.1 CENTER POSITION
1 . Attach the PD checking jig (32545-1002) (see 8.2.13 [p116]).
* A scale can be used as substitute for the PD checking jig (32545-1002).
2 . Enter MENU mode, and select 15. PD ADJUST MODE (see 8.1.1 [p103]).
3 . Select CENTER POSITION with the up
2&#&,756/1&'
or down button, then press the %'06'4215+6+10
execute button. )#+01((5'6
%#.+$4#6+10
Flap
⓭
Groove
┄ߺ
㪌㪇
[p164]). 215+6+10
5#8' 1((5'6
8.9.2 PD CALIBRATION
1 . Attach the PD checking jig (32545-1002) (see 8.2.13 [p116]).
* A scale can be used as substitute for the PD checking jig (32545-1002).
2 . Enter MENU mode, and select 15. PD ADJUST MODE (see 8.1.1 [p103]).
)#+0
㪉㪇
1((5'6
)#+0
㪇
1((5'6
)#+0
㪐㪇
1((5'6
)#+0
㪈㪇
1((5'6
button.
*$,1
6$9(
㻝㻜
2))6(7
㪐㪇
1((5'6
6*4'5
56#4659%#2674'
5#8'
button.
3 ) Align the left cursor to a position 12
mm away from the right cursor with
1 ) Align the right cursor to the graduation of the scale with the right or left but-
ton.
3 ) Align the left cursor to a position 5 mm away from the right cursor with the right or
left button.
8 . Press the SAVE button to save the setting.
9 . Measure the CS value at 12 mm and PS value at 5 mm, and confirm that the measured val-
ues are within the following the specifications.
1 ) CS: 12 mm ±0.2 mm
2 ) PS: 5 mm ±0.2 mm
10. If the CS or PS measurement values are not within the specifications, recalibrate the items
that are not within the specifications.
or down button
a . The first row: upward
b . The second row: downward
5#8'
c . The third row: left-and-right
direction
d . The fourth row: backward-
and-forward direction
2 ) Change the numeric value with the right , right + , left , or left - button.
* The right and left buttons increase or decrease the numeric value by 1
51(694+6'/1&'
4'#&+0) 94+6'
51(694+6'/1&'
or left button.
. +2#&&4'5
b . Change the numerical value
with the right button, left
button, or joystick.
d . Press the delete button to delete a character in the entry field. Then move
the cursor in the entry field one character to the left.
2 ) Select YES, NO, or LOW CONF with the ight button, left button, or joystick.
a . YES: Sends the measurement ring image.
b . NO: Does not send the measurement ring image.
c . LOW CONF: Sends the measurement ring image when the confidence index is
7 or less.
13. Check the network connection.
1 ) Press the TEST button.
2 ) The following message is displayed for about two seconds along with a beep.
a . When the LAN cable is connected prop-
erly, “CONNECTION OK” is displayed.
%100'%6+101-
b . If the LAN cable is not connected prop-
erly, “ERR771 NO NETWORK CABLE” 㧱㧾㧾㧣㧣㧝
is displayed. 㧺㧻ޓ㧺㧱㨀9㧻㧾㧷ޓ㧯㧭㧮㧸㧱
9 SUPPLEMENT
9.1 Appearance
9.1.1 Front view
Function buttons
LCD
Memory indicator
Printer cover
Start button
Joystick
Locking lever
Power switch
Forehead rest
Eye level
marker
Measuring window Chinrest
up/down
button
Chinrest
PD window
9.2 Labels
Front and side views
USA
China ––––––
Canada ––––––
Rear view
72120(75(6$16&217$&7
72120(752$62)),2
121&217$&772120(7(5
7210(752'(12&217$&72
0$
&/$66/('352'8&7
웢 웳월월웑웢
$8725()5$&720(7(5 For technical assistance please call:
,1387 9윟+]9$ 㻝㻙㻤㻜㻜㻙㻤㻣㻠㻙㻡㻞㻣㻠
6(512 1111 ;;;;
0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0'
Destination For countries other than USA, Canada, and China USA
&/$66/('352'8&7
웢 웳월웑웢
$8725()5$&720(7(5
,1387 9윟+]9$
AR-310A
6(512 1111 ;;;;
0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0' 30601-M803-C
&/$66/('352'8&7
웢 웳월월웑웢
$8725()5$&720(7(5
,1387 9윟+]9$
AR-330A
6(512 1111 ;;;;
0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0' 30602-M803-C
&/$66/('352'8&7
웢 웳월웑웢
$8725()5$&720(7(5
,1387 9윟+]9$
AR-360A ––––––
6(512 1111 ;;;;
0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0&
ID
label
&/$66/('352'8&7
㸿 㹐㹉㸳㸯㸮㸿
$8725().(5$720(7(5
,1387 9㹼+]9$
ARK-510A
6(512 1111 ;;;;
0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0' 30701-M803-C
&/$66/('352'8&7
㸿 㹐㹉㸳㸱㸮㸿
$8725().(5$720(7(5
,1387 9㹼+]9$
ARK-530A
6(512 1111 ;;;;
0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0' 30702-M803-C
&/$66/('352'8&7
㸿 㹐㹉㸳㸴㸮㸿
$8725().(5$720(7(5
INPUT 9㹼+]9$
ARK-560A
SER.NO. 1111 ;;;;
0$(+$0$
+,52,6+,&+2
*$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0&
72120(75(6$16&217$&7
72120(752$62)),2
Multilingual label 121&217$&772120(7(5 ––––––
7210(752'(12&217$&72
0$
Destination For countries other than USA, Canada, and China USA
MarcoFreeDial ––––––
㻝㻙㻤㻜㻜㻙㻤㻣㻠㻙㻡㻞㻣㻠
㸿 㹐㸱㸯㸮㸿
㸿㹓㹒㹍 㹐㹃㹄㹐㸿㹁㹒㹍㹋㹃㹒㹃㹐
ൟো $5$
ѻકৡ⿄˖ܼ㞾ࡼ⬉㛥偠ܝҾ
,1387 9㹼+]9$
䕧ܹ⬉⑤˖$&9+]9$
䖤㸠ᮍᓣ˖䖲㓁䖤㸠 1R1111
0$(+$0$ ⊼ݠᷛޚ㓪ো˖<=%-$3
+,52,6+,&+2
*$0$*25,
ए⭫఼Ẅ⊼ݠ䆕к㓪ো˖亳㥃ⲥẄ䖯ᄫো
$,&+,-$3$1 ⫳ѻӕϮ˖ Ⱬⶲᢼ嚦
0$'(,1-$3$1
0$'(,1-$3$1 0$ ഄഔ˖⠅ⶹও㪆䚵Ꮦᣒ⬎ࠡ⌰⬾ഄ 0$
㸿 㹐㸱㸱㸮㸿
㸿㹓㹒㹍 㹐㹃㹄㹐㸿㹁㹒㹍㹋㹃㹒㹃㹐
රਧ $5$
ӗ૱〠˖ޘ㠚ࣘ⭥㝁傼ݹԚ
,1387 9㹼+]9$
䗃⭥ޕⓀ˖$&9+]9$
䘀㹼ᯩᔿ˖䘎㔝䘀㹼 1R1111
0$(+$0$ ⌘ḷ߶㕆ਧ˖<=%-$3
+,52,6+,&+2
*$0$*25,
५⯇ಘỠ⌘䇱Җ㕆ਧ˖ഭ伏㦟ⴁỠ䘋ᆇㅜਧ
$,&+,-$3$1 ⭏ӗԱъ˖ Ⱬⶲᢼ嚦
0$'(,1-$3$1
0$'(,1-$3$1 0$ ൠ൰˖⡡⸕৯㫢䜑ᐲ⸣⭪ࡽ⎌⮚ൠ 0$
㸿 㹐㸱㸴㸮㸿 &/$66/('352'8&7
ID
label
㸿 㹐㹉㸳㸯㸮㸿
㸿㹓㹒㹍 㹐㹃㹄㹉㹃㹐㸿㹒㹍㹋㹃㹒㹃㹐
ൟো $5.$
ѻકৡ⿄˖ܼ㞾ࡼ⬉㛥偠ܝҾ
,1387 9㹼+]9$
䕧ܹ⬉⑤˖$&9+]9$
䖤㸠ᮍᓣ˖䖲㓁䖤㸠 1R1111
0$(+$0$ ⊼ݠᷛޚ㓪ো˖<=%-$3
+,52,6+,&+2
*$0$*25,
ए⭫఼Ẅ⊼ݠ䆕к㓪ো˖亳㥃ⲥẄ䖯ᄫো
$,&+,-$3$1 ⫳ѻӕϮ˖ Ⱬⶲᢼ嚦
0$'(,1-$3$1
0$'(,1-$3$1 0$ ഄഔ˖⠅ⶹও㪆䚵Ꮦᣒ⬎ࠡ⌰⬾ഄ 0$
㸿 㹐㹉㸳㸱㸮㸿
㸿㹓㹒㹍 㹐㹃㹄㹉㹃㹐㸿㹒㹍㹋㹃㹒㹃㹐
ൟো $5.$
ѻકৡ⿄˖ܼ㞾ࡼ⬉㛥偠ܝҾ
,1387 9㹼+]9$
䕧ܹ⬉⑤˖$&9+]9$
䖤㸠ᮍᓣ˖䖲㓁䖤㸠 1R1111
0$(+$0$ ⊼ݠᷛޚ㓪ো˖<=%-$3
+,52,6+,&+2
*$0$*25,
ए⭫఼Ẅ⊼ݠ䆕к㓪ো˖亳㥃ⲥẄ䖯ᄫো
$,&+,-$3$1 ⫳ѻӕϮ˖ Ⱬⶲᢼ嚦
0$'(,1-$3$1
0$'(,1-$3$1 0$ ഄഔ˖⠅ⶹও㪆䚵Ꮦᣒ⬎ࠡ⌰⬾ഄ 0$
㸿 㹐㹉㸳㸴㸮㸿 &/$66/('352'8&7
㸿㹓㹒㹍 㹐㹃㹄㹉㹃㹐㸿㹒㹍㹋㹃㹒㹃㹐
㸿 㹐㹉㸳㸴㸮㸿
$8725().(5$720(7(5
,1387 9㹼+]9$
INPUT 9㹼+]9$
ARK-560A 6(512 1111 ;;;;
SER.NO. 1111 ;;;;
0$(+$0$
+,52,6+,&+2 0$(+$0$
*$0$*25, +,52,6+,&+2
$,&+,-$3$1 *$0$*25,
$,&+,-$3$1
0$'(,1-$3$1 0$ 0$'(,1-$3$1 0&
Underside view
For USA, Canada, and China
30601-EA11
30601-EA12
30601-EA13
30601-EA14
30601-EA23
30601-EA24
30601-EA25
30601-EA28
30601-EA30
30601-EA31
30601-EA32
30601-EA34
30601-EA36
30601-EA39
30601-EA42
30601-EA44
30601-EA51
30601-EA52
30601-EA53
(Motor is included.)
30601-EA54
30601-EA55
30601-EA63
30601-EA64
30601-EA65
30601-EA72
30601-EA75
30601-EA76
30601-EA80
30601-EA81
30601-EA82
30601-EA83
30601-EA93
type
(With motor)
(With motor)
(With motor)
*For the details of ERR061 to ERR068 caused by the combination of the main board and driver
board, see the table below.
Main board (30601-BA01)
-
Unwritten AR-310A AR-330A AR-360A ARK-510A ARK-530A ARK-560A
Unwritten ERR061 ERR065 ERR065 ERR065 ERR066 ERR066 ERR066
AR-310A ERR067 - ERR062 ERR062 ERR064 ERR064 ERR064
9.5.2 AR-360A
Parameter item No. Item USA Countries other than China
USA and China
1 STEP 0.25 D 0.25 D 0.25 D
2 VERTEX D. 13.75 mm 12.00 mm 12.00 mm
3 AXIS STEP 1° 1° 1°
AR 4 MEAS MODE Con. Con. Con.
5 AI MODE YES YES YES
6 AR CONTINUE 3 3 3
7 AR THUMBNAIL YES YES YES
31 PRINT MANUAL MANUAL MANUAL
32 ECONO. PRINT NO NO NO
33 PRINT&CLEAR NO NO NO
34 PRINT DENCITY MIDDLE MIDDLE MIDDLE
PRINT1 35 PATIENT NO. YES YES YES
36 SET PATIENT NO. 0001 0001 0001
37 NAME PRINT YES YES YES
38 DATE FORMAT M/D/Y M/D/Y M/D/Y
39 PRINT COMMENT YES YES YES
41 AR PRINT ALL ALL ALL
43 CONF. INDEX YES YES YES
44 ERROR DATA YES YES YES
PRINT2
45 CAT MARK NO NO NO
46 ERROR PRINT NO NO NO
47 PRINT FORMAT OBJ->SBJ OBJ->SBJ OBJ->SBJ
51 SE PRINT NO NO NO
52 EYE PRINT NO NO NO
53 TL PRINT NO NO NO
PRINT3
54 CL PRINT NO NO NO
55 NEAR PD PRINT NO NO NO
56 WORKING D. 16-inch 40 cm 40 cm
61 WINDOW CHECK NO NO NO
62 TRACKING SW TRC/ASHOT TRC/ASHOT TRC/ASHOT
63 AUTO PD YES YES YES
64 SLEEP 5 min 5 min 5 min
FUNCTION 65 BEEP Low Low Low
66 LCD BRIGHTNESS NORMAL NORMAL NORMAL
67 ICON OFF NO NO NO
68 ALIGN QUALITY FINE FINE FINE
69 MAN FOCUS DISP YES YES YES
71 I/F MODE NIDEK NIDEK NIDEK
72 I/F FORMAT SHORT SHORT SHORT
73 BAUD-RATE 9600 9600 9600
COMMUNICATION
74 BIT LENGTH 8 8 8
75 CR CODE NO NO NO
76 LM DATA PRINT NO NO NO
9.5.3 ARK-510A/ARK-530A
Parameter item No. Item USA Countries other than China
USA and China
1 STEP 0.25 D 0.25 D 0.25 D
2 VERTEX D. 13.75 mm 12.00 mm 12.00 mm
3 AXIS STEP 1° 1° 1°
AR 4 MEAS MODE Con. Con. Con.
5 AI MODE YES YES YES
6 AR CONTINUE 3 3 3
7 AR THUMBNAIL YES YES YES
11 KM UNIT D mm mm
12 KM DISPLAY R1, R2 R1, R2 R1, R2
KM 13 REF. INDEX 1.3375 1.3375 1.3375
14 KM CONTINUE 3 3 3
15 PERIPHERAL NO NO NO
21 COMPARE SW MANUAL MANUAL MANUAL
22 COMPARE(AR) SE SE SE
COMPARE
23 ADD SW YES YES YES
24 ADD SELECT 1.75 D 1.75 D 1.75 D
31 PRINT MANUAL MANUAL MANUAL
32 ECONO. PRINT NO NO NO
33 PRINT&CLEAR NO NO NO
34 PRINT DENCITY MIDDLE MIDDLE MIDDLE
PRINT1 35 PATIENT NO. YES YES YES
36 SET PATIENT NO. 0001 0001 0001
37 NAME PRINT YES YES YES
38 DATE FORMAT M/D/Y M/D/Y M/D/Y
39 PRINT COMMENT YES YES YES
41 AR PRINT ALL ALL ALL
43 CONF. INDEX YES YES YES
44 ERROR DATA YES YES YES
PRINT2
45 CAT MARK NO NO NO
46 ERROR PRINT NO NO NO
47 PRINT FORMAT R->L R->L R->L
51 SE PRINT NO NO NO
52 EYE PRINT NO NO NO
53 TL PRINT NO NO NO
PRINT3
54 CL PRINT NO NO NO
55 NEAR PD PRINT NO NO NO
56 WORKING D. 16-inch 40 cm 40 cm
61 WINDOW CHECK NO NO NO
62 TRACKING SW TRC/ASHOT TRC/ASHOT TRC/ASHOT
63 AUTO PD YES YES YES
64 SLEEP 5 min 5 min 5 min
FUNCTION 65 BEEP Low Low Low
66 LCD BRIGHTNESS NORMAL NORMAL NORMAL
67 ICON OFF NO NO NO
68 ALIGN QUALITY NORMAL NORMAL NORMAL
69 MAN FOCUS DISP YES YES YES
9.5.4 ARK-560A
Parameter item No. Item USA Countries other than China
USA and China
1 STEP 0.25 D 0.25 D 0.25 D
2 VERTEX D. 13.75 mm 12.00 mm 12.00 mm
3 AXIS STEP 1° 1° 1°
AR 4 MEAS MODE Con. Con. Con.
5 AI MODE YES YES YES
6 AR CONTINUE 3 3 3
7 AR THUMBNAIL YES YES YES
11 KM UNIT D mm mm
12 KM DISPLAY R1, R2 R1, R2 R1, R2
KM 13 REF. INDEX 1.3375 1.3375 1.3375
14 KM CONTINUE 3 3 3
15 PERIPHERAL NO NO NO
21 COMPARE SW MANUAL MANUAL MANUAL
22 COMPARE(AR) SE SE SE
COMPARE
23 ADD SW YES YES YES
24 ADD SELECT 1.75 D 1.75 D 1.75 D
31 PRINT MANUAL MANUAL MANUAL
32 ECONO. PRINT NO NO NO
33 PRINT&CLEAR NO NO NO
34 PRINT DENCITY MIDDLE MIDDLE MIDDLE
PRINT1 35 PATIENT NO. YES YES YES
36 SET PATIENT NO. 0001 0001 0001
37 NAME PRINT YES YES YES
38 DATE FORMAT M/D/Y M/D/Y M/D/Y
39 PRINT COMMENT YES YES YES
41 AR PRINT ALL ALL ALL
43 CONF. INDEX YES YES YES
44 ERROR DATA YES YES YES
PRINT2
45 CAT MARK NO NO NO
46 ERROR PRINT NO NO NO
47 PRINT FORMAT AR->KM AR->KM AR->KM
51 SE PRINT NO NO NO
52 EYE PRINT NO NO NO
53 TL PRINT NO NO NO
PRINT3
54 CL PRINT NO NO NO
55 NEAR PD PRINT NO NO NO
56 WORKING D. 16-inch 40 cm 40 cm
61 WINDOW CHECK NO NO NO
62 TRACKING SW TRC/ASHOT TRC/ASHOT TRC/ASHOT
63 AUTO PD YES YES YES
64 SLEEP 5 min 5 min 5 min
FUNCTION 65 BEEP Low Low Low
66 LCD BRIGHTNESS NORMAL NORMAL NORMAL
67 ICON OFF NO NO NO
68 ALIGN QUALITY NORMAL NORMAL NORMAL
69 MAN FOCUS DISP YES YES YES
o
o
o
o
o
7.1.2 Measuring ASSY (30601-9E00) (p70)
o
o
o
7.1.3 Phototransmitter ASSY (30601-3310) (p70)
o
o
7.1.4 Mire ASSY (30601-4000) (p71)
7.1.5 LCD ASSY(30601-6100) (p71)
213
o
7.1.6 Subjective cylinder ASSY (30703-3700) (p72)
o
7.1.7 Chart disk ASSY (30703-3800) (p72)
o
o
7.2.3 Driver board (30601-BA03) (p75)
7.2.4 Base board (30601-BA05) (p76)
7.2.5 LCD board (30601-BA06) (p77)
7.2.6 Left panel SW board (30601-BA07) (p77)
7.2.7 U/D SW board (30601-BA09) (p78)
7.2.8 Sensor carry board (30601-BA14) (p78)
7.2.9 Right panel SW board (30601-BA15) (p79)
o
(p129)
(p122)
(p121)
XAR17ARDA001G/E
Replacement part
o
7.3.4 R/L motor (30602-EA26) (p84)
o
7.3.5 F/B motor (30602-EA27) (p85)
o
7.3.6 U/D motor (30601-EA28) (p85)
214
o
7.3.7 Origin sensor (30601-EA36) (p86)
o
7.3.8 CL motor (30703-EA37) (p86)
o
7.3.9 CL sensor (30703-EA38) (p87)
o
o
7.3.10 Focus LED(30601-EA42) (p88)
o
o
7.3.11 KM6 LED (30601-EA44) (p88)
7.3.12 Chinrest (30601-EA53) (p89)
o
7.3.13 PD LED (30601-EA55) (p90)
9.7.2 Electrical components and adjustment items
o
7.3.17 Color LCD (30601-E005) (p94)
7.3.18 Brushless motor (30601-E011) (p94)
o
ࢿࢪ㢮グྕ⾲6&5(:6/,67