Temperature Gradient Plate MFFT 10 / MFFT 20

ISO 2115 – ASTM D 2354


The new designed temperature gradient plate
for measuring the Minimum Film Forming Temperature , MFFT

The minimum film forming The Coesfeld MFFT has been A constant flow acc. to the
temperature (MFFT) is the developed to give the user a standards can be set with a
minimum temperature at which comfortable possibility to built in flow meter.
synthetic latex, emulsions, measure the MFFT according
polymers or adhesives will the practice described in the
coalesce when laid on a ASTM D 2354 and ISO 2115.
substrate as a thin film.
The ideal way to determine The temperature gradient plate
MFFT is to simultaneously dry consists of a high precision
the emulsion under test over a chromium measuring plate, with
range of temperatures to see equispaced temperature
which is the lowest at which sensors beneath the surface.
film-forming has occurred. This The flat hinged perspex cover
is best achieved in practice by For measuring the temperature, provides thermal and
using the Coesfeld MFFT the MFFT 20 is equipped with atmospheric insulation whilst
gradient plate. 20 Pt-100 temperature sensors. allowing constant visual
(MFFT 10: 10 Pt-100 sensors). inspection of an experiment.
Three steps of the film-forming The microprocessor controller
guaranteed an optimum
temperature control.
1. Evaporation of water
The temperature set points for
the MFFT 20 can vary in a wide
2. Packing and deformation of particles range from – 30° to +250°C
with a maximum possible
gradient on the surface of 100°.
3. Diffusion of water out of film and (MFFT 10: -5° to +80°C with a
coalescence of particles. maximum gradient of 20°).
Option
A film of the latex under test is Purge gas is dried with an Cooling devices with different
cast along the stage and left to integrated membrane dryer and ranges of capacity adapted to
dry. A visual inspection of the flows over the platen. the customer requirements are
dry film is then made to available.
determine the position along Different kinds of film casters
the length of the film where it completed the components of
changes from the coalesced to the temperature gradient plate.
the non-coalesced state. A non-
coalesced film shows whitening
and/or cracking. The lowest
temperature at which the film is
coalesced is reported as the
MFFT.

Coesfeld GmbH & Co. KG Telefon: +49 231 912 980 0 E-Mail: mail@coesfeld.com
Tronjestraße 8 * 44319 Dortmund Telefax: +49 231 179 885 Internet: www.coesfeld.com
art.-no.: 50-130
type: Temperature Gradient Plate
MFFT 20
minimum film forming temperature MFFT

acc. to current DIN ISO 2115 – DIN 53 366 – ASTM D 2354 – ASTM D 1465 –
standards: ISO 2115 – ISO/DIS 4622

description: chromium-plated measuring plate:


measuring length: 500 mm
measuring width: 180 mm
temperature display: -30…+250°C
temperature gradient: max. 100 K (depends on the external cryostat)
electronic temperature controller with digital temperature display and
temperature preselection as well as excess temperature protection
20 built-in temperature sensors
 accuracy: 0,1°C
 display of nominal value and actual value of the temperature of
the hot and cold side
 scroll function to request the temperature measuring stations
 manual selection of the single temperature measuring stations
 integrated membrane drying unit with digital flowmeter
 hood made of PMMA, for temperatures up to +80°C
 option: hood made of stainless steel with cover made of safety
glass for temperatures up to +250°C

necessary The necessary cooling unit for operating does not belong to the scope of
accessories: delivery of the MFFT 20. It must be ordered separately.

The working temperature range and the maximum temperature gradient


depend on the effective cooling power of the connected cryostat.

We recommend the standard cryostat type RP845

Working temperature range and temperature gradient MFFT 20 with


connected cryostat RP845

Cryostats for other working temperature ranges and gradients upon request

Coesfeld GmbH & Co. KG Telefon: +49 231 912 980 0 E-Mail: mail@coesfeld.com
Tronjestraße 8 * 44319 Dortmund Telefax: +49 231 179 885 Internet: www.coesfeld.com
technical data: dimensions (L x W x H): 800 x 350 x 320 mm
(with closed cover)

weight: 50 kg

power supply: 230 V, 50 Hz, 1,5 kW

MFFT 20 with digital flow meter

Film caster for 6 parallel


tests

Coesfeld GmbH & Co. KG Telefon: +49 231 912 980 0 E-Mail: mail@coesfeld.com
Tronjestraße 8 * 44319 Dortmund Telefax: +49 231 179 885 Internet: www.coesfeld.com
Membrane dryer for compressed air or nitrogen

Digital flow meter 0 – 10 l/min

Temperature controller

Coesfeld GmbH & Co. KG Telefon: +49 231 912 980 0 E-Mail: mail@coesfeld.com
Tronjestraße 8 * 44319 Dortmund Telefax: +49 231 179 885 Internet: www.coesfeld.com
Overview scetch MFFT with dropping border and grooves

Dropping border

Dropping border

Groove 0,3 mm depth

Coesfeld GmbH & Co. KG Telefon: +49 231 912 980 0 E-Mail: mail@coesfeld.com
Tronjestraße 8 * 44319 Dortmund Telefax: +49 231 179 885 Internet: www.coesfeld.com