You are on page 1of 5

COULOMETRIC TITRATION OF ASCORBIC ACID

WITH ELECTROGENERATED IODINE


Aguilar, J., Antazo, F., Lumayag, L., Naguit, P., Ragos, E.
Department of Chemistry, College of Science
University of Santo Tomas

ABSTRACT
Coulometry is an electroanalytical method used to measure the quantity of electricity
used in either to transform the substance in a redox reaction or to produce a reactant in an
analytical reaction and is based by on the Faraday’s Law. The experiment aims to calculate the
amount of ascorbic acid using coulometric titration with electrogenerated iodine. A mixture of
NaNO3, KI, starch indicator, and the aliquot of the ascorbic sample with a red carbon electrode
in the solution attached and a crucible also containing the NaNO3 and KI solution but with the
black carbon electrode inside the solution was titrated together until it the solution reached a
purple color. The time on which the reaction is completed is recorded and is marked as the end
point of the titration. Based on the time recorded, the amount of ascorbic acid was calculated.
Results show that an average mol of Vitamin C or ascorbic acid was computed to be 1.918x10 -5
mol with a percentage error of 35.17%. Possible sources of error are personal errors and
instrumental error.

INTRODUCTION F is the Faraday number (96.500 coulombs


= 26.8 A hours) [1].
Coulometry is the universal term
referring to an electroanalytical method Iodine is very useful in titration methods,
which consists in the measurement of the such as Iodometric and Iodimetric titration.
quantity of electricity used either to These methods are used to determine the
transform the substance to be determined in concentration of vitamin-C, sodium
a redox reaction or to produce a reactant of thiosulphate, amount of copper in super
an analytical reaction. Thus the species to conductor etc [3]. Vitamin C is also called
be determined reacts either at one of the ascorbic acid and antioxidant. It is titrated in
electrodes or with a reagent produced by a solution with electrogenerated iodine
electrolysis. The second method is also (Equation 1) [1]. The endpoint of the titration
referred to as coulometric titration [1]. is indicated by purple or blue coloration due
This method is based on Faraday’s to the reaction of excess iodine with a
second law: starch indicator. Ascorbic easily oxidize to
dehydroascorbic acid. Thus in the
iodometric titration of ascorbic acid, the
ascorbic acid is quantitatively oxidized by
where m is the amount of substance iodine (Figure 1) [1] [2] [3].
transformed in the electrolysis, M is the
molecular weight of substance, n is the
2 I-→ I2 + 2e- (Equation 1)
number of electrons involved in the
electrode reaction, Q is the amount of
electricity used in the reaction in coulombs,
crucible while the red carbon electrode was
in the solution of the beaker. It is important
that the stirring is begun before switching on
Figure 1. The reaction mechanism for the the apparatus. After observing a color
oxidation of iodine change, the time at which the solution turns
completely purple was noted marking as the
A basic requirement in coulometric endpoint of the titration. The duration of the
analysis consists in selecting an electrode titration was determined. Multiple trials are
reaction which takes place 100% current highly recommended for more accurate and
efficiency as the amount of electricity is precise results.
employed to calculate the quantity of the
substance to be determined. Current RESULTS AND DISCUSSION
efficiency can be affected by interfering
reactions, e.g., the solvent or the supporting Coulometric titration is a method in
electrolyte can react at the electrode, the which the titrating agent is generated in a
electrode material can react with some solution electrochemically in the titration
components of the solution, or a secondary cell. Constant current is applied and
reaction can take place involving the monitored as a function of time. This kind of
product of electrolysis [1]. titration uses a current system to accurately
quantify the concentration of the species of
By a filtering crucible covered with interest.
glass frit, the cathode is isolated from the
Table 1. Time elapsed when the violet color
test solution to maintain 100% generating shows each trials
efficiency. Using Faraday’s law, the amount TRIAL TIME (min)
of ascorbic acid was calculated [1]. 1 6:55
2 9:44
METHODOLOGY 3 6:29
4 7:33
A. Materials and reagents
In performing the experiment, 0.5g Vitamin Table 2. Mean value of the time elapsed
MEAN 7:40 min or 460 s
C was crushed and diluted to 100 mL then
filtered before titrating. Apart from this, a The iodine is generated at the
mixture of 0.1 M NaNO3/ 0.1 M KI and 0.5 anode. A titration cell is consisted of two
% starch indicator was freshly prepared. parts: an anodic and cathodic compartment.
As the iodine is added during
B. Coulometric titration of the sample titration, the ascorbic acid is oxidized to
A mixture of 75 mL of 0.1 M NaNO3/ 0.1 M dehydroascorbic acid, while the iodine is
KI, 5mL starch indicator, and 0.5 mL aliquot reduced to iodide ions.
of the ascorbic acid sample was placed in a 𝑎𝑠𝑐𝑜𝑟𝑏𝑖𝑐 𝑎𝑐𝑖𝑑 + 𝐼2 → 2 𝐼 − + 𝑑𝑒ℎ𝑦𝑑𝑟𝑜𝑎𝑠𝑐𝑜𝑟𝑏𝑖𝑐 𝑎𝑐𝑖𝑑
beaker. Inside the beaker, a magnetic stirrer
was used and above it is a filtering crucible Due to this reaction, the iodine
that also contains 10 mL of 0.1 M NaNO3/ formed is immediately reduced to iodide as
0.1 M KI mixture. The Black carbon long as there is any ascorbic acid present.
electrode was inserted into the filtering Once all the ascorbic acid has been
oxidised, the excess iodine is free to react current measurement, (4) error in the
with the starch indicator, forming the blue- measurement of time, and (5) titration error
black starch-iodine complex. This is marks due to the difference between the
the endpoint of the titration. The time is equivalence point and the end point.
measured until the blue-violet color spreads Situations wherein the indicator error is the
throughout the solution. limiting factor would prove difficult to
Electrodes have large surface areas correct. However, proper observation is
so the current density is low. advised.
The values were tested using the Q-
test due to the outlier, 9:44, as seen on CONCLUSION
Table 1. This is to test if 9:44 is to be
accepted or rejected. The formula: Coulometric titration is a method in
which the titrating agent is produced in a
𝑋𝑎 − 𝑋𝑏 solution through electrolysis. It was
𝑄= successfully used to determine the mole
𝑅
concentration of vitamin c in the solution.
where; The data, due to producing an outlier,
Xa = is the suspected outlier underwent the Q-test. After calculations, the
Xb = value closest to Xa outlier, 9:44 min, was accepted which
R = Range of the values (highest to lowest) means that she will not be removed from
After the computation, the Q was the equation all throughout the experiment.
accepted which provides the leeway to The true value and the experimental value
consider it with the other values. of the mole concentration of vitamin c was
The number of moles of the vitamin computed to be able to get the percentage
c used was calculated. This was used as error. Based on the time recorded, the
the true value. The experimental values amount of ascorbic acid was calculated.
were calculated using the time lapsed in Results show that an average mol of
seconds. Using these values, the Vitamin C or ascorbic acid was computed to
percentage error was determined. The be 1.918x10-5 mol with a percentage error of
formula for finding the percentage error is, 35.17%. After resulting to 35.17% error,
several factors could be behind such a large
𝐸𝑉 − 𝑇𝑉 percentage error. All of the objectives for
% 𝑒𝑟𝑟𝑜𝑟 = 𝑥100
𝑇𝑉 this experiment were met, however, strict
surveillance is suggested for the the repeat
of this experiment to avoid errors.
The resulting percentage error was
35.17%. There are several factors that
REFERENCES
could have led to such a high percentage
error.
[1] Pungor, E. (1995). A practical guide to
instrumental analysis. Boca Raton: CRC
There are different factors to
Press.
consider in coulometric titration. It could be
(1) variation in the current during
[2] Iodometric Titration. (n.d.). Retrieved
electrolysis, (2) departure of the process
March 15, 2018, from
from 100% current efficiency, (3) error in the
https://chemistry.tutorvista.com/analytical-
chemistry/iodometric-titration.html

[3] Skoog, D. A., West, D. M., Holler, F. J.,


& Crouch, S. R. (2014). Fundamentals of
analytical chemistry. Belmont, CA:
Brooks/Cole, Cengage Learning.
APPENDIX Conclusion: 0.6719 < 0.829, outlier is
accepted
TIME (CONVERTED TO SECONDS)

60 𝑠𝑒𝑐 TRUE VALUE OF MOL VIT C


1. 6: 55 𝑚𝑖𝑛 ( 1 𝑚𝑖𝑛 ) = 360 + 5
= 415 𝑠𝑒𝑐𝑜𝑛𝑑𝑠 1 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶 0.5
1. (0.5 𝑔)( 176.12 𝑔 )(100)
𝑚𝑜𝑙

60 𝑠𝑒𝑐
= 1.419𝑥10−5 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶
2. 9: 44 𝑚𝑖𝑛 (
1 𝑚𝑖𝑛
) = 540 + 44 =
548 𝑠𝑒𝑐𝑜𝑛𝑑𝑠 EXPERIMENTAL VALUE OF MOL VIT C

2. a.
1 𝑚𝑜𝑙 𝐼 1 𝑚𝑜𝑙 𝐼 1 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶
60 𝑠𝑒𝑐
3. 6: 29 𝑚𝑖𝑛 ( 1 𝑚𝑖𝑛 ) = 360 + 29 = (389 𝑠)(0.00804)(96,487 2𝑐)(2 𝑚𝑜𝑙 𝑒2− )( 1 𝑚𝑜𝑙 𝐼 )
2

389 𝑠𝑒𝑐𝑜𝑛𝑑𝑠 = 1.621𝑥10−5 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶

2. b.
1 𝑚𝑜𝑙 𝐼 1 𝑚𝑜𝑙 𝐼 1 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶
60 𝑠𝑒𝑐
4. 7: 33 𝑚𝑖𝑛 ( 1 𝑚𝑖𝑛 ) = 420 + 33 = (215 𝑠)(0.00804)(96,487 2𝑐)(2 𝑚𝑜𝑙 𝑒2− )( 1 𝑚𝑜𝑙 𝐼 )
2

453 𝑠𝑒𝑐𝑜𝑛𝑑𝑠 = 1.729𝑥10−5 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶

2. c.
1 𝑚𝑜𝑙 𝐼 1 𝑚𝑜𝑙 𝐼 1 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶
MEAN TIME (453 𝑠)(0.00804)(96,487 2𝑐)(2 𝑚𝑜𝑙 𝑒2− )( 1 𝑚𝑜𝑙 𝐼 )
2
= 1.887𝑥10−5 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶
6: 55 + 9: 44 + 6: 29 + 7: 33
= 7: 40 𝑚𝑖𝑛𝑠
4 2. d.
1 𝑚𝑜𝑙 𝐼2 1 𝑚𝑜𝑙 𝐼2 1 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶
(584 𝑠)(0.00804)( )( − )( )
96,487 𝑐 2 𝑚𝑜𝑙 𝑒 1 𝑚𝑜𝑙 𝐼2
415 + 548 + 389 + 453 = 2.433𝑥10−5 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶
= 460 𝑠𝑒𝑐𝑜𝑛𝑑𝑠
4
AVERAGE MOL VIT C (EXPERIMENT)
Q-TEST AT 95% CONFIDENCE
INTERVAL 1.621𝑥10−5 + 1.729𝑥10−5 + 1.887𝑥10−5 + 2.433
4
TIME, in increasing order
389 seconds
415 seconds = 1.918x10-5 mol Vit C
453 seconds
PERCENTAGE ERROR
584 seconds
1.918𝑥10−5 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶 − 1.419𝑥10−5 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶
584 − 453 𝑥 100
𝑄= = 0.6718 1.419𝑥10−5 𝑚𝑜𝑙 𝑉𝑖𝑡 𝐶
584 − 389
% 𝑒𝑟𝑟𝑜𝑟 = 35.17%
Number of
Computed
Observatio Q95
Q
ns
4 0.829 0.6719

You might also like