Professional Documents
Culture Documents
KLa measurement
Objectives
1. To determine KLa of a fermentation system by dynamic gassing out techniques
depends upon the monitoring of the increase in dissolve oxygen in agitation and
aeration range.
2. To monitor the increase in dissolved oxygen over an adequate range, it is
necessary to fast decrease the O2 level to a low value. Two methods can be
employed to achieve this lowering of the dissolved oxygen concentration; non-
fermentative and fermentative.
3. To study the effect of medium viscosity on KLa value.
Apparatus
Bioreactor including pO2 probe.
Stopwatch
Chemicals
NaCl
Antifoam
Distilled water
3. For slope adjustment, the air supply is activated and the stirring speed is adjusted at
the operating value. The medium should be optimally gassed (max. flow rate intended
for the process) and mixed. At a stable display of the measured value we can calibrate
this as 100 pO2".
4. After calibration, the gas supply rate required for the start up of the intended
fermentation process can be adjusted on the rotameter of the control unit. Note that the
rotameter is calibrated according to standard conditions (temperature 20°C, with air at 2
barabs). If it is important to maintain precise operating air flow-rates for further
calculations, this makes it necessary to recalculate the indicated flowrate with a
correction factor.
The calibration of the p02 -electrode is made in the culture vessel after autoclaving and
under the conditions of fermentation.
Non-fermentative method
In this technique, the oxygen concentration of the solution is lowered by gassing the
liquid out with nitrogen gas, so that the solution is "scrubbed" free of oxygen. Aeration is
then initiated at a constant sir flow rate and the increase in dissolved oxygen tension
(DOT) is monitored using dissolved oxygen electrode. The profile of DOT during
deaeration and aeration is shown in Figure 1. Increase in DOT during aeration can be
expressed by Eq. 1;
Figure 1: Dynamic gassing out for the determination of KLa values. Aeration was
terminated at point A and recommenced at point B.
1. Set the agitation speed of 500 rpm and 1.0L/min. Purge the nitrogen gas until reach
0% DO. Determine K L a of stirred tank reactor at different air flow rate (0.5, 1.0, 1.5,
2.0 and 2.5 L/min). For this experiment, set the agitation speed at 500 rpm.
2. Determine the effect of increasing agitation speed (200, 400, 600, 800 and 1000
rpm) on K L of a 2 L stirred tank fermenter. For this experiment, set the air flow rate at
1 L/min.
3. In experiment 1 and 2, the fermenter is filled with 1.5 L of distilled water.
4. Investigate the effect of salt (NaCI) and antifoam addition to distilled water on K L a.
In this experiment, add 1.5 g of NaCI to 1.5 L distilled water in a fermenter.
Determined the K L at a 500 rpm and air flow rate of 1 L/min. Then, add 5 mL of
antifoam in a salt solution and determine KLa at the same agitation speed and air
flow rate.
1. Plot a graph to show the effect of air flow rate and agitation speed on K La. Also
discuss the effect of the addition of salt and antifoam on KLa.
2. Compare the KLa value determined using different rpm and air flow rate.
3. Discuss the possible cause of error in determination of KLa by using this dynamic
gassing out technique.
40 100
60 120
80 130
100 140
120 150
130 160
140 170
150 180
160 190
170 200
180 210
190 220
200 230
210 240
220 250
230 260
240 270
250 280
260 290
270 300
280
60 20
time CL (% ∆CL/∆t CL(averag Ln(CE – 40
80
(s) saturation) e) CL )
100 0 60
12020 80
13040 100
14060 120
15080 130
160100 140
170120 150
180130 160
190140 170
200150 180
210160 190
220170 200
230180 210
240190 220
250200 230
260210 240
270220 250
280230 260
290240 270
300250 280
260 290
270 300
280 Air flow rate: L/min
Volume liquid: L
290
Note: CE = 100% saturation
300
150 140
160
150
160
170
180
170
190 180
200 190
210 200
220 210
230 220
240 230
250 240
260
250
270 260
280 270
290 280
300 290
300