Professional Documents
Culture Documents
a b
Positively charged detector
Electron beam
Gun chamber V0
A
Cascade electrons
Anode aperture
Projection apertures
+
Upper pressure
limiting aperture Gaseous atom
V
Vacuum manifold
Lower pressure
limiting aperture
and integrated Earthed sample
detector
Specimen stage
The gas is not simply a passive participant in the scope of this article, and will only be touched upon
Pressure Pressure
range zone imaging process, there simply to open up the range of briefly in what follows.
10–7 torr Gun chamber
samples studied, it plays a key role in signal detection1,2. Recent research in the ESEM field is of two types.
At its simplest,it can be seen that as electrons (secondary There is an increasing range of applications in the field
10–6 torr Upper column
or backscattered) are emitted, they also have to travel of materials that use the full power of the instrument —
10–4 torr EC2 through the gas. The secondary electrons in particular, these include a significant amount of work on ceramics
10–1 torr EC1 because of their low energies, have a high collision cross- such as cement, which use the ESEM’s ability to image
section with the gas molecules, and ionization of the uncoated insulators. There are also interesting
10 torr Specimen
molecules has a significant probability of occurring. developments in the study of natural fibres, which are
Each such ionizing collision leads to the generation of not only insulators but also may be partially hydrated.
an additional daughter electron, so that a cascade Finally there are (fewer) examples of using the
Figure 1 Schematic of ESEM. amplification occurs, shown schematically in Fig. 1b. instrument to look at aqueous dispersions, or actual
a,The different pressure zones All the electrons produced are drawn towards the water droplets, as well as biological samples maintained
in the ESEM column.The gun positively biased detector: this is not the standard in their native state. This aspect of biological studies is
can be maintained at high Everhardt–Thornley detector used in CSEM, which the least well developed because the number of workers
vacuum, while a system of could not work under these pressures, but a specially in biological departments who are entering the field is
differential pumping- and designed detector appropriate for these conditions. still small. For any of these applications there is the
pressure-limiting apertures The original ESEM (which is a trademark of FEI/Philips) possibility of carrying out dynamic, in situ experiments,
allows the chamber to be detects charge flow in the gas, but as other in which the state of the sample is changed whilst
maintained at a pressure of a manufacturers enter the market (selling so-called continuing to image3,4. Such changes could be in
few torr (1 torr = 133 Pa). variable pressure instruments, which typically are not temperature5, state of hydration6, or even by carrying
b,The cascade amplification yet able to operate at such high pressures for technical out a chemical reaction7,8.
process.This process occurs reasons, limiting their ability to maintain hydrated The second developmental area is in what might be
within the chamber, where the samples in their fully hydrated state), a variety of termed ESEM physics. Few groups are doing this kind of
gas molecules are ionized by different signals are being used for detection. This article work, but it is vital to the successful interpretation of
the electrons emitted from the will, for simplicity, concentrate simply on the electronic images. Experiments in this area deal with more
sample. Each ionizing collision detection in the ESEM. quantitative analysis of what happens in the gas cascade,
gives rise to a daughter As well as the daughter electrons produced in the exploring the effect of charge deposition on electron
electron, which, like the original ionizing collisions, positive ions are also produced. yield in different types of experiments, and
electron, is accelerated towards These drift down towards the sample, and hence serve consideration of contrast and detection mechanisms.
the positively charged detector, to compensate charge build-up at the surface of
and further collisions can occur insulators. It is for this reason that insulators do not WET SAMPLE IMAGING
en route to the detector. need to be coated to prevent charging artefacts and
The positive ions drift back consequent loss of image quality. However, it should be As indicated above, the ESEM can operate with a
towards the sample surface. noted that the precise role of ions in affecting the whole significant gas pressure around the sample.In principlethis
imaging process is still incompletely understood and is means that hydrated samples can be kept fully hydrated,
an active field of study; this complexity is beyond the but in practice achieving this takes some thought.
Pressue (torr)
area to be imaged10, the state of hydration can be
maintained. It is also important during the whole
Condense Evaporate
imaging process to be aware of the state of the sample 10
relative to the chamber pressure. Figure 2 shows the
form of the saturated vapour pressure curve for water
as a function of temperature. If one tries to work at 5 Gas
room temperature, it can be seen that the water vapour
pressure required to maintain a saturated vapour in
the chamber is impossibly high. Imaging under these 0
conditions is currently not practical. To achieve 0 5 10 15 20 25
saturation, it is desirable to drop the sample
Temperature (°C)
temperature to just above freezing, and then only a
modest gas pressure is needed to stabilize water.
Usually this is done by the use of a Peltier-chip-
controlled cooling stage, making it possible to image are still obvious — there is no need for the standard Figure 2 Saturated vapour
samples containing water. type of sample preparation including drying, possibly pressure of water as a function
Figure 3a shows a colloidal dispersion of a fixing, and coating required for CSEM. This removes of temperature.The useful range
poly(methyl methacrylate) latex, containing spherical many of the potential pitfalls of artefacts being of working conditions for the
particles ~250 nm across. In Fig. 3a, the individual introduced during preparation. And such work is ESEM are shown; it can be seen
particles near the sample surface can clearly be seen as beginning to throw up examples of where tried and that the state of the sample can
dark against the light background of the continuous tested routes of sample preparation are now being be changed by small alterations
water phase.Watching such a sample shows the particles shown to have introduced artefacts28,31–33. in temperature.
undergoing constant Brownian motion. During this Will ESEM become the microscopy of choice for
process, collisions between particles can occur and, biologists? At this time the answer to this is far from
depending on the interparticle potential, sticking may clear. The ability to image cells and other biological
occur. This process can be studied in more detail by assemblies including whole organisms without
deliberately setting instrument conditions to favour dehydration or coating — let alone staining or fixing —
aggregation: as Fig. 2 shows, it is a simple matter to cross is clearly very attractive. But as yet it is not known
the saturated vapour pressure curve in either direction whether living cells can be viewed but not killed by the
by slightly increasing or decreasing the temperature. electron beam. Clearly, the lower the magnification used
To study aggregation, dehydrating conditions can be set the lower the electron dose and therefore the greater the
up to pull surface water off the sample. Figure 3b shows
a sample of the same latex but now dispersed in a salt-
containing solution after a short period of dehydration:
in this case a fractal aggregate has started to form. a b
Armed with this knowledge of how to control the
thermodynamic state of the sample, a whole range of 5 µm
new, dynamic experiments become possible (other
kinds of dynamic experiments will be discussed later).
In essence, the sample chamber becomes a
microprocessing unit. Significant work has been done
in this area using ESEM on cement11–14 and related
inorganic materials7,15, as well as colloidal aggregation
leading to film formation, in which the packing of
particles can be interrogated at intermediate steps as the
film dries (Fig. 4)16–18.Another type of experiment has
involved the growth of Langmuir films at an air–water
interface19. In this case, the structure of the films grown 2 µm
under conditions of different concentrations can be
examined to see how the packing alters.As a further
example, Fig. 5 shows how contact angles can be
determined from condensation of water onto a Figure 3 ESEM images of colloidal dispersions under different conditions. a,A colloidal dispersion of
cellulose fibre20. 250 nm poly(methyl methacrylate) latex particles dispersed in water.The particles undergo Brownian
Not all samples of course involve free water; many motion and may collide with each other. b,A dispersion of the same particles now dispersed in a 0.078 M
are simply ‘damp’. Such types of samples include magnesium sulphate solution. In this solution, the interparticle potential is altered such that any collisions
natural biopolymers (as fibres21–24 or gels25, lead to a reasonably high sticking probability, and a fractal structure develops as aggregation proceeds.
biofilms26,27 and many truly biological samples28–30). The exact nature of the aggregate is strongly dependent on the salt concentration; in the absence of any
Nevertheless, the advantages of working in an ESEM salt the sticking probability is low, and colloidal crystals rather than fractals form.