You are on page 1of 16

®

URODYN 1000 System


6WDWLRQDU\8URIORZ5HFRUGHU

User Guide

$
®
Urodyn 1000 User Guide

Copyright © 2000 Medtronic Functional Diagnostics A/S. All rights reserved.

The contents of this manual are the property of Medtronic Functional Diagnostics A/S. Any repro-
duction in whole or in part is strictly prohibited.

At the time of printing / transfer to the CD-ROM, this manual correctly described the device and its
functions. However, as modifications may have been carried out since the production of this man-
ual, the system package may contain one or more addenda to the manual. This manual including
any such addenda must be thoroughly read, before using the device.

2 Contents
®
Urodyn 1000 User Guide

Contents
Contents ................................................................................................................... 3

Safety Information................................................................................................... 5
Product Description..................................................................................................... 5
Intended Use............................................................................................................... 5
Symbols .................................................................................................................... 6
Setup Keys.................................................................................................................. 6
Parts List ..................................................................................................................... 6
Routine use of system .............................................................................................. 7
Prepare Equipment ..................................................................................................... 7
Operation modes ..................................................................................................................... 7
Start up the System.................................................................................................................. 7
Device Verification ...................................................................................................... 7
Standard Settings ....................................................................................................... 7
Change Settings ......................................................................................................... 8
Description of factory settings ..................................................................................... 8
Print Recording ........................................................................................................... 8
Report ......................................................................................................................... 8
Troubleshooting....................................................................................................... 9
Out of Paper............................................................................................................................ 9
Secure Printer Arm ................................................................................................................. 9
No Sensor Connected.............................................................................................................. 9
Calibration Failed .................................................................................................................. 9
Printer too hot......................................................................................................................... 9
Technical failure ..................................................................................................................... 9
Device description.................................................................................................. 10
System ...................................................................................................................... 10
Legend...................................................................................................................... 10
Maintenance .......................................................................................................... 11
Cleaning.................................................................................................................... 11
Calibration ............................................................................................................. 12
Device Calibration..................................................................................................... 12
Technical Data....................................................................................................... 13
Model .................................................................................................................................... 13
Number of Channels ............................................................................................................. 13
Filters.................................................................................................................................... 13
Power Supply ........................................................................................................................ 13
Languages:............................................................................................................................ 13
Operating Conditions............................................................................................................ 13
Accuracy ............................................................................................................................... 13
Measures and weights .............................................................................................. 13
Contents 3
®
Urodyn 1000 User Guide

Sensor..................................................................................................................... 14
Sensor type:........................................................................................................................... 14
Recording method ................................................................................................................. 14
Parameters ............................................................................................................................ 14
Classification ......................................................................................................... 15
Authorized Service Centers................................................................................................... 16

4 Contents
®
Urodyn 1000 User Guide

Safety Information Intended Use


The Urodyn® 1000 is intended to acquire and
This product is CE marked in conformity with
analyze data and to provide post acquisition
the requirements of the Medical Device Direc-
analysis of voiding and voiding disorders.
tive 93/42/EEC.

Adhere to the following recommendations for WARNING Only professional users should use
safe operation of the device: this device.
 Never use the device near X-rays, mobile
telephones, CB radios or other forms of radio Medtronic Functional Diagnostics A/S is only
communication, and/or electromagnetic responsible for the reliability and performance
fields. These may affect the performance of of the device if the following are strictly ob-
the device. served:
 To avoid hazards, only use sensors that have
 Authorized personnel (see back of guide for
been designed for use with the system (see
authorized service centers) carry out all re-
Sensor section).
pairs and modifications.
 Do not attempt to open the case.
 The device must be used in accordance with
 Do not immerse the device in water or any the instructions given in this section.
other liquid.
 The device must be used only for its intended
 Do not change printer paper during an inves- use.
tigation.
If the above points are not strictly observed, the
 Do not allow patients to connect a sensor warranty will be considered invalid.
themselves.
 Do not connect the sensor to anything other WARNING Danger of electrical ignition. The
than the appropriate input on the device. device is not intended for use with anesthetic
 Never attempt to connect a sensor to any gases mixed with air, oxygen or nitrous oxide.
electrical outlet apart from the one specified
on the device (see Device description).
 You should observe the specific safety stan- CAUTION In the United States, Federal Law
dards and regulations of your own country, restricts this device to sale by or on the order of
regarding electrical connections to the main a physician or licensed practitioner.
supply in hospitals and wet environments.
 Note: To disconnect the mains from the de- WARNING The instructions in this manual are
vice, remove the plug from the power socket. meant for recommendation only. They are not
 Do not keep non-compliant equipment in the intended to replace or negate hospital patient
patient environment. If in doubt contact your care policy.
authorized service center for further infor-
mation.

Product Description
The Urodyn® 1000 is a stationary medical de-
vice used to record physiological data, such as
those associated with voiding disorders.

Safety Information 5
®
Urodyn 1000 User Guide

Symbols Parts List


The package should contain the following
Explanations of the symbols found on the de-
items:
vice:
1 Urodyn® 1000
1 Power cord
 Attention, see accompanying documents. 3 Paper rolls
1 Registration Card
1 Urodyn® 1000 User Guide (This manual)

 Body Floating safety rating


1 Changing the language of Urodyn® 1000
1 Cleaning Instructions
1 Wall Mounting Instruction

! Input

Standby/Manual start/stop

$ CE mark with notified body ID

Setup Keys

C Cancel, returns you to the previous loca-


tion without saving any changes

Down, moves the selection bar to the next


item down

Up, moves the selection bar to the next


item up

Enter button, executes the current selec-


tion

6 Symbols
®
Urodyn 1000 User Guide

Routine use of system


This is done as follows:
Prepare Equipment 1. When prompted by the device, or during the
installation, pour the contents of the calibra-
Operation modes tion bottle into the Flow Transducer. This
The Urodyn® 1000 has three different operation provides a known volume (e.g. 400ml), pour
modes: at 10-40 ml/second.
2. Wait for the test report to be printed.
1. Power save: a number of functions are
automatically deactivated by the system in 3. Validate the device result against the poured
order to save power. The mode is changed to volume. If the difference between the result
standby either by pressing the Standby but- obtained and the actual volume is greater
ton or automatically by the built-in On pe- than the volume accuracy figure (see the
riod function, if selected. Technical Data section), then the device
should be calibrated, as described in the
2. Standby: all functions are enabled and the
Calibration section.
device is ready to record. Recording starts
when the Standby button is pressed or when 4. Empty the beaker /pot.
a flow is detected.
3. Recording: data is being recorded. NOTE The device will prompt you to verify at
regular intervals (weekly), to ensure optimal
Start up the System performance. This interval can be changed in
the Setup Menu under Verification.
Connect the system, as described in Device de-
scription. The device will start up in power save
or, if within the designated On period, in
standby mode. Standard Settings
If the device starts in power save mode, to begin
The device is set up to use factory settings at
using it press the Standby button.
delivery. This can be changed in the Setup
When this screen is displayed, the system is menu, using Modify existing.
ready for use.
Use Factory settings will clear any customer
settings and return the device to its factory set-
tings.

Recording starts when the Standby button is


pressed or when a flow is detected.

Device Verification

The device is factory calibrated, but to ensure


optimal performance it is recommended that it
should be verified, before the system is used.
Routine use of system 7
®
Urodyn 1000 User Guide

 Printer settings: the user can choose between


Change Settings printing 0, 1 or 2 additional copies of the re-
To use the factory settings, provided with the cording. The user can also switch on or off
device, no changes are necessary. If you wish to an option to enter patient information before
modify the factory settings, select Setup from the recording starts.
the main menu then Modify existing from the  Sensitivity: allows the selection of the sensi-
Setup screen. The configurable items are dis- tivity (ml/s) for the recording. Values avail-
played in a list with the optional settings under- able are: 50, 25 and 10 ml/s.
neath. Use Up and Down keys to scroll between  On period: allows the user to set the start and
items and Enter to select the preferred setting. stop times for the on period.
Press Cancel to return to the previous item.
 No flow time: this is the period of time, after
a recording, that the device continues to rec-
Description of factory ord data after the flow has fallen to zero.
Fixed values from 10 seconds to 2 minutes
settings are available. The default value is 20 sec-
onds.
NOTE In the following items the system's de-
fault settings (where applicable) are underlined.
Print Recording
The default is to print during each study. To
 Time format/Date format: are determined by obtain an additional copy of the study, press the
the system's language setting. When this is Start/Standby button before the time limit is
English (U.S.), the system will use the 12- exceeded.
hour (a.m./p.m.) and Month/Day/Year for- The device will maintain up to 20 studies for
mats. If the language is set to English (U.K.), safekeeping and for printing out copies later.
the system will use the 24-hour and Patients can be found in the Previous Patients
Day/Month/Year formats. For all other lan- menu. Using the arrows (Up and Down keys),
guages the system will use the 24-hour and select the patient, then press Enter to print.
Day/Month/Year formats.
 Set time: the system's internal clock can be Note when the memory is full the oldest data
set up. Select Time then Change will be erased, one entry at a time.
Hours/Change Minutes  Use the Up/Down
controls to increase/decrease the values dis-
played.
 Set date:the system's internal date can be
Report
setup. Select Date then Change
Months/Change years Use the Up/Down IMPORTANT: Exercise diligence in the inter-
controls to increase/decrease the values dis- pretation of results from this system. It is always
played. advisable to check the program analysis results
 Language settings: the user can select the against the signal tracing when a major conse-
relevant local language. For a complete list quence such as surgery occurs as a result of an
of available languages, see Technical Data. investigation.
The default language is English (U.S.).
 Verification interval: can be set to prompt the
user to verify the accuracy of the devices.
The interval between verifications can be: a
week, a month, six months or a year.
 Filter settings: the user can choose between
the Standard or Thompson settings.

8 Troubleshooting
®
Urodyn 1000 User Guide

Troubleshooting
The device has an error detection system that, in there is a build-up of residue. If there is, then
the event of a failure, will suggest an action or clean the device and try to calibrate again.
display an error code. If an error code is shown
always note it down, as this will help your
service representative to solve the problem. WARNING If you choose to ignore this mes-
sage by pressing ENTER standard calibration
values will be used instead of the real ones.
Out of Paper This may produce incorrect results!

Printer too hot


If this screen is displayed: The printer motor has
overheated.
Load paper into the device. Instructions on how
to do this can be found on the inside of the de- This can happen if there is a high turnover of
vice lid. patients, where all the studies are printed out.
To avoid damaging the device, allow it to cool
Secure Printer Arm down before reuse.

Open the device's lid and secure the printer arm.


Technical failure
If this screen is displayed:
No Sensor Connected
&RQQHFW VHQVRU

Note down the error code and call your service


Ensure that the Flow Transducer is connected to representative.
the device.

Calibration Failed

If this screen appears, there may be a calibra-


tion/ sensor problem. Ensure that all connec-
tions are secure. If the screen is still displayed,
check whether the sensor is obstructed or if

Troubleshooting 9
®
Urodyn 1000 User Guide

Device description
System Legend
 1. Urodyn® 1000

 2. Display
3. Power on indicator
 4. Standby, start/stop switch
  5. IrDA Port
6. Lid

 7. Lid catch
 8. Set up keys


 9. Printer
 10. Secure screw to wall bracket

11. Paper location

 12. Printer lever
13. Paper output
 14. Power input connector

15. Power supply cable
16. Electrical outlet
 17. Flow Transducer input
 18. Flow Transducer connector
 19. Flow Transducer
20. Flow Transducer stand

10 Device description
®
Urodyn 1000 User Guide

Maintenance
On a weekly basis, test the device for proper  Do not immerse the device in any liquid.
function and inspect cables and sensor for cuts  To clean accessories, follow the cleaning in-
and other signs of damage. If in doubt, replace structions as described in their individual
the relevant parts. The device should be cleaned User Guides, provided with those accesso-
routinely, as described below. No additional ries.
maintenance is required.

Cleaning
The Urodyn® 1000 should be cleaned once a
month. Clean the device as follows:

 Turn off the device by unplugging it from the


power socket.
 Wipe the device with a damp cloth and mild
detergent.

Maintenance 11
®
Urodyn 1000 User Guide

Calibration
Device Calibration
The device is factory calibrated, but to ensure
optimal performance it is recommended that it
should be calibrated regularly.

This is done as follows:

1. Using the control keys, select Test from the


main menu.
2. Select Calibration
3. The device should be calibrated using a
known volume of 400ml, measured using the
calibration bottle. When prompted by the de-
vice, pour the contents of the calibration bot-
tle into the Flow Transducer (pour at 10-40
ml/second).
4. Wait for the test report to be printed.
5. Empty the beaker /pot.

12 Calibration
®
Urodyn 1000 User Guide

Technical Data
Model 5VQTCIG%QPFKVKQPU
®
Urodyn 1000 6TCPURQTVCVKQP
4GSWKTGOGPVU
Number of Channels Temperature -40°C to +70°C
(-40°F to 158°F)
1 channel
Humidity 10% - 100%rh
(including condensation)
Filters Atmospheric pressure
Standard (0-1 Hz –3dB) 700hPa - 1060hPa
Thompson (0-0.25 Hz –3dB)
Operating Conditions
Power Supply Temperature +10°C to +40°C
Power Supply: Mains (50°F to 104°F)
Input: 110-240V, 50-60Hz Humidity 30% - 75%rh
Power consumption: max. 100 VA Atmospheric pressure
Type of Protection: Safety Class I 700hPa - 1060hPa

Languages: Accuracy
English (U.S.), English (UK), Danish, Dutch, Flow Transducer
French, German, Greek, Italian, Portuguese, Flow: +/-3% (relative)
Spanish, Swedish and Finnish. +/- 0.2 ml/s (absolute)
Volume: +/-3% (relative)
+/- 1.0 ml (absolute)

Note: Accuracy is dependent on calibration.

Measures and weights

Width Depth Height Weight

Metric 320 mm 210 mm 70 mm Maximum 1.7kg


U.S. 12.5 ” 8.0 ” 3.0 ” Maximum 3.8lbs.

All values are approximate.

Technical Data 13
®
Urodyn 1000 User Guide

Sensor
The device supports a Flow Transducer sensor. Parameters
The device has 2 channels (internally) with dif-
The device, together with the sensor supports
ferent filtering (Standard and Thompson). The
the following parameter:
amplifier is controlled by the device’s software.
 Flow.

Sensor type:
 Flow Transducer: REF 9034K010.

Recording method
 Continuous on demand or automatic detec-
tion of flow.
 Sampling rate: 16Hz.
 Resolution: 12 bit A/D.
 Measuring range: 0-1000 ml @ 0-50 ml/s.

14 Sensor
®
Urodyn 1000 User Guide

Classification
Type of protection against electric shock:  Not intended for direct cardiac application.
 Class I.
Degree of protection against ingress of water:
Degree of protection against electric shock:  IP 20.
 Type BF (Body Floating): Applied part pro-
vides a specific degree of protection against Degree of safety in presence of inflammable
electric shock, particularly in regard to the anesthetics:
amount of current leakage that is allowed.  The device is not intended for use with an-
esthetic gases mixed with air, oxygen or ni-
 The applied part is mechanically isolated trous oxide. Danger of electrical ignition.
(floating).
Mode of operation:
 Continuous or intermittent operation.

Classification 15
Authorized Service Centers

Denmark USA
Medtronic Functional Diagnostics A/S Medtronic Functional Diagnostics, Inc.
Tonsbakken 16-18 3850 Victoria Street North MS v215
DK-2740 Skovlunde Shoreview, MN 55126-2978
Denmark USA
Tel. +45 44 57 90 00 Tel. +1-612-514-1700
Fax +45 44 57 90 10 Fax +1-612-514-1710
E-mail uro.mfd@medtronic.com Toll-free in the USA: 1-800-227-3191
Http://www.mfd.medtronic.com

Manufactured by:

Medtronic Functional Diagnostics A/S


Tonsbakken 16-18
DK-2740 Skovlunde
Denmark
Tel. +45 44 57 90 00
Fax +45 44 57 90 10
E-mail: uro.mfd@medtronic.com
http://www.mfd.medtronic.com

Printed in Denmark, March, 2000 Reg. No 9034M1041 

:KHQ/LIH'HSHQGVRQ0HGLFDO7HFKQRORJ\

You might also like