You are on page 1of 6

PCCP

View Article Online


PAPER View Journal | View Issue

A pacemaker powered by an implantable biofuel cell


Cite this: Phys. Chem. Chem. Phys., 2013,
operating under conditions mimicking the human
15, 6278
blood circulatory system – battery not included
Mark Southcott,wa Kevin MacVittie,wb Jan Halámek,*b Lenka Halámková,bc
William D. Jemison,*a Robert Lobeld and Evgeny Katz*b

Biocatalytic electrodes made of buckypaper were modified with PQQ-dependent glucose dehydrogenase
on the anode and with laccase on the cathode and were assembled in a flow biofuel cell filled with
serum solution mimicking the human blood circulatory system. The biofuel cell generated an open
circuitry voltage, Voc, of ca. 470 mV and a short circuitry current, Isc, of ca. 5 mA (a current density of
0.83 mA cm2). The power generated by the implantable biofuel cell was used to activate a pacemaker
connected to the cell via a charge pump and a DC–DC converter interface circuit to adjust the voltage
produced by the biofuel cell to the value required by the pacemaker. The voltage–current dependencies
were analyzed for the biofuel cell connected to an Ohmic load and to the electronic loads composed of
the interface circuit, or the power converter, and the pacemaker to study their operation. The correct
Received 1st March 2013, pacemaker operation was confirmed using a medical device – an implantable loop recorder. Sustainable
Accepted 6th March 2013 operation of the pacemaker was achieved with the system closely mimicking human physiological
DOI: 10.1039/c3cp50929j conditions using a single biofuel cell. This first demonstration of the pacemaker activated by the
physiologically produced electrical energy shows promise for future electronic implantable medical
www.rsc.org/pccp devices powered by electricity harvested from the human body.

1. Introduction were designed and investigated for in vivo electrical power


generation. In addition to obvious challenges typical for biofuel
Implantable devices harvesting energy from biological sources cells, such as improving efficiency and operational stability,5,6
and based on electrochemical, mechanical and other energy problems resulting from the operation in the biological environ-
transducers are currently receiving lots of attention.1–4 Among ment were studied and at least partially resolved.20–25
biofuel cells5,6 harvesting electrical power from biological The energy collected from the body can be utilized to activate
sources and based on the biocatalytic activity of enzymes,7–12 various microelectronic devices. While some microelectronic
microbial cells13–16 and biological organelles,17 micro-sized devices can work over a fairly broad range of electrical operating
implantable biofuel cells18,19 operating in vivo in a living body conditions others, such as a pacemaker, require precise voltage
have attracted significant attention in recent research. Biofuel levels and voltage regulation to operate correctly. Thus certain
cells implanted in the retroperitoneal space of a freely moving classes of electronic devices powered by implantable energy
rat,20 in the blood vessel in a rabbit ear,21 in the abdomen of an sources will require careful attention to not only energy and
insect,22 or in the hemolymph of snails,23 clams24 and lobsters25 power considerations, but also to voltage scaling and voltage
regulation. This requires appropriate interfacing between
a
Department of Electrical and Computer Engineering, Clarkson University,
the energy harvesting device and the microelectronic device.
Potsdam, NY 13699, USA. E-mail: wjemison@clarkson.edu Therefore, another problem appeared when the biofuel cells
b
Department of Chemistry and Biomolecular Science, Clarkson University, Potsdam, were aimed to power electronic devices: achieving the appro-
NY 13699, USA. E-mail: jhalamek@clarkson.edu, ekatz@clarkson.edu priate voltage necessary. The voltage produced by the biofuel
c
Department of Biology, Clarkson University, Potsdam, NY 13699, USA
d
cells is thermodynamically limited by the redox potentials of the
Fletcher Allen Cardiology, University of Vermont College of Medicine,
Medical Center Campus, McClure 1, 111 Colchester Avenue, Burlington,
chemical species (enzyme redox centers or mediators in the case
VT 05401-1473, USA of direct26 or mediated electron transfer,27 respectively). Usually
† M. Southcott and K. MacVittie contributed equally to this work. the open circuit voltage generated by biofuel cells based on the

6278 Phys. Chem. Chem. Phys., 2013, 15, 6278--6283 This journal is c the Owner Societies 2013
View Article Online

Paper PCCP

oxidation of glucose and reduction of oxygen hardly exceeds described elsewhere.39 1-Pyrenebutanoic acid succinimidyl ester
0.5 V, being decreased upon consuming current from the cells;5–8 (PBSE) was purchased from AnaSpec Inc. Serum from a human
at best the voltage was measured as high as 0.78 V while operating male (type AB), 3-(N-morpholino)propanesulfonic acid (MOPS-
under non-physiological conditions.28 However, the operational buffer) and other standard chemicals were purchased from
voltage of modern microelectronics is usually several volts, based Sigma-Aldrich and used as supplied without any further purifica-
on standard batteries as electrical power supplies.29 The mis- tion. Water used in all of the experiments was ultra pure
match between the biofuel cell output voltage and the electronic (18.2 MO cm) from a NANOpure Diamond (Barnstead) source.
devices input voltage was compensated by assembling biofuel
cells in series electrically, thus increasing the total output 2.2 Electrode preparation
voltage,24,30–32 and by collecting produced electrical energy Buckypaper composed of compressed multi-walled carbon
in capacitors/charge pumps for the burst release in short nanotubes (Buckeye Composites; NanoTechLabs, Yadkinville,
pulses.33–36 The first approach, while being successful in NC) was used as the electrode material. Electrodes were washed
assembling microbial fuel cells in stacks,30–32 has very limited with isopropyl alcohol with moderate shaking for 15 min at
applicability for the implantable devices since the serial connection room temperature prior to their modification. The electrodes
between the opposite poles of individual fuel cells implanted in a were incubated with PBSE, 10 mM, in ethanol with moderate
body has a finite resistance associated with the biological tissue shaking for 1 h at room temperature, subsequently rinsed with
resulting in internal power dissipation and reduced stack voltage.25 ethanol to remove excess PBSE and then with MOPS-buffer
In order to solve the problem, single biofuel cells were implanted in (50 mM, pH 7.0) to remove ethanol. The biocatalytic anodes
different animals or in separated flow devices mimicking blood were prepared by immobilization of PQQ-GDH: the PBSE-
circulation systems and then connected in series.25 This achieved functionalized electrodes were incubated for 1 hour in the
the desired voltage increase, but had very little hope for future solution of PQQ-GDH (2.4 mg mL1) in MOPS-buffer (50 mM,
practical applicability for implantable devices. The second pH 7.0) containing Na2SO4 (100 mM) and CaCl2 (1 mM). The
approach based on electronic interface devices such as charge biocatalytic cathodes were prepared by immobilization of
pumps and other forms of DC–DC conversion has already been laccase: the PBSE-functionalized electrodes were incubated
applied for activating a wireless transmitting electronic device, for 1 hour in the solution of laccase (1.5 mg mL1) in potassium
however using non-implantable enzyme-based36 or microbial phosphate buffer (10 mM, pH 7.0). The immobilization reactions
biofuel cells.33 Application of an interface device to increase proceeded at room temperature with moderate shaking. The
the voltage is rather well-known,37 however it should be enzyme-modified electrodes were stored (4 1C) in the same buffer
remembered that the voltage increase is achieved at the until used in the biofuel cell. Characterization of the enzyme-
expense of the current consumed by the charge pump, thus modified electrodes (cyclic voltammetry, enzyme content, etc.) was
putting additional demand on the current output of the biofuel described in detail elsewhere.23,24
cell. Implantable micro-size electrical energy generators
connected to an electrical interface can be effectively used for 2.3 Biofuel flow cell design and measurements
activating micro-electronic devices operating in the short-
pulses regime, while the time between pulses is used for the The flow cell with the dimensions shown in the sketch, Fig. 1,
accumulation of energy.38 However, a single implantable bio- included the buckypaper electrodes modified with PQQ-GDH
fuel cell connected to a charge pump has never been used for and laccase in the anode and the cathode, respectively, and the
the continuous powering of an electronic device. The current inlet/outlet of the cell were connected to plastic tubes of
produced by the biofuel cell will be particularly important for 0.5 mm internal diameter. A human serum solution with the
the steady-state operation of the load-device rather than short glucose concentration of 6.4 mM was pumped (MINIPULS 3,
pulses operation. The present paper demonstrates for the first Gilson) at the volumetric rate of 58.9 mL min1, mimicking
time interfacing between an implantable biofuel cell operating blood flow in a human capillary of 0.008 mm at the linear rate
under conditions mimicking human physiology and a continuously of 1 mm s1 characteristic of a resting person.40 Note that the
operating pacemaker, while the mismatch between the output volumetric flow rate was scaled up to generate the same linear
voltage of the biofuel cell and the input voltage of the pacemaker rate in the tube of a bigger diameter. The polarization curve of
was compensated by a charge pump and a DC–DC converter the biofuel cell was obtained by measuring voltage–current
interface circuit.

2. Experimental
2.1 Chemicals
PQQ-dependent glucose dehydrogenase (PQQ-GDH; E.C. 1.1.5.2,
from microorganisms – not specified by the company) was
purchased from Toyobo Co., Japan, and used as supplied. Laccase
(E.C.1.10.3.2, from Trametes versicolor) was obtained from Sigma-
Aldrich and used in experiments after the purification procedure Fig. 1 Sketch of the biofuel flow cell.

This journal is c the Owner Societies 2013 Phys. Chem. Chem. Phys., 2013, 15, 6278--6283 6279
View Article Online

PCCP Paper

by the pacemaker. The circuit was constructed using a standard


printed circuit board.37

3. Result and discussion


The biocatalytic electrodes based on the buckypaper conducting
support (composed of compressed multi-walled carbon nano-
tubes, CNTs)23–25,39,41–43 with the geometrical area of 6 cm2 were
integrated into a flow cell filled with human serum solution
moving at the flow rate of 58.9 mL min1 with the help of a
peristaltic pump. The human serum solution was spiked with
glucose to the concentration of 6.4 mM to mimic human
physiological conditions. The glucose oxidizing anode was
modified with PQQ-dependent glucose dehydrogenase (PQQ-
Fig. 2 The open pacemaker prior to its wiring to the external power source. GDH) and the oxygen reducing cathode was modified with
laccase. The enzymes were attached to the CNTs with the help
of the bifunctional reagent 1-pyrenebutanoic acid succinimidyl
values on the external load variable resistance using a multi- ester (PBSE) following the published procedure.23–25,39 PQQ-
meter (Meterman 37XR). GDH and laccase enzymes are known for the direct electron
transfer with the electrode support composed of CNTs.44,45
2.4 Pacemaker Characterization of the enzyme-modified electrodes (cyclic voltam-
The pacemaker (Affinity DR 5330L, St. Jude Medical) was metry, enzyme content, etc.) was described in detail elsewhere.23–25
an explanted device that had been removed and discarded It should be noted that the electrode size (electrochemically active
following battery depletion (as per the standard pacemaker area) was increased compared to the previous experiments23–25 to
replacement protocol). The metal shell of the pacemaker was provide enough current for the operation of the interface circuit
cut and the device was opened, allowing the removal of and the connected pacemaker.
the internal battery and the connection of the leads of the The biofuel cell operating under flow conditions was connected
electronic chip to the external biofuel cell, Fig. 2. to a variable load resistance and polarization was measured
demonstrating the open circuitry voltage, Voc, ca. 470 mV and short
circuitry current, Isc, ca. 5 mA, Fig. 4 curve a. It should be noted that
2.5 Charge pump and DC–DC converter design the biofuel cell polarization function was measured with the Ohmic
A charge pump and a DC–DC converter were used to amplify resistance and the derived voltage and current are both functions of
the voltage produced by the biofuel cell to 3 volts, a level the used resistance. Conversely, the electronic devices – the charge
required by the pacemaker. Fig. 3 shows the schematic diagram pump and the connected pacemaker – are not simple Ohmic loads
of the circuitry. The two primary components are a Seiko S-882z and they require specific current and voltage for their correct
charge pump and a Seiko S-8353 DC–DC converter. The charge operation. Measuring the application current demand in order
pump can take an input voltage, VIN, as low as 0.3 volts, and to predict if a given device will operate with the biofuel cell is an
generate an output of 2 volts. The DC–DC converter further important step needed for interfacing of biofuel cells and
conditions this signal to the desired output level, VOUT, of electronic devices which are powered by them. The current
3 volts. Thus the combination of the charge pump and the and voltage demanded by an electronic device are normally
DC–DC converter can take a nominal biofuel cell input voltage fixed at some point, and a change in the input voltage will cause
range of 0.3–0.5 V and generate the fixed 3 volt output required the device to work improperly or damage it. The pacemaker

Fig. 3 Power converter circuit schematic. Abbreviations used in the circuit shown are standard for electronic circuit annotation.37

6280 Phys. Chem. Chem. Phys., 2013, 15, 6278--6283 This journal is c the Owner Societies 2013
View Article Online

Paper PCCP

interface circuit (Vbfc and Vcp, respectively), Icp(Vbfc) is the


interface circuit current draw as a function of Vbfc, Iapp is
the application device (pacemaker) current draw as seen by
the biofuel cell, and k is the interface circuit’s efficiency defined
as the ratio of the power output and the power input. The Icp
values for the variable voltage applied to the power converter
were measured experimentally (see the text above and Fig. 4,
curve b), while the second term in eqn (1) is an extension of the
ideal transformer current equation,46 with the charge pump
efficiency taken into account. Note that the Vcp value is fixed at
3 V, which is the voltage required by the pacemaker and the Iapp
value should be known for the device being used, in this case
the pacemaker has a peak current draw of ca. 100 mA. These
values are shown in the Fig. 4 inset, and are specific to the
devices used in this study. The efficiency of the power converter
loaded with a pacemaker equivalent resistance was calculated
Fig. 4 Voltage–current dependences of the studied assembly: the polarization curve to be ca. 60%, based on the measured voltage and current
of the biofuel cell measured on the variable resistances – voltage and current
inputs and outputs of the charge pump. The V–I plot derived
produced by the cell as the function of the Ohmic resistance load (a), measured
current consumed by the charge pump–DC–DC interface circuit upon application of
from eqn (1) shows the calculated application current demand
the variable voltage from the external power source demonstrating that the interface curve, Fig. 4, curve c, compared with the measured biofuel cell
circuit can supply constant current over a wide input voltage range (b), calculated polarization curve, Fig. 4, curve a. Because the application
(eqn (1)) current consumed by the charge pump–DC–DC interface circuit when a current demand curve lies below the polarization curve we
variable voltage is applied and the pacemaker is connected as the load (c). Green
know that the biofuel cell should be able to power our applica-
point on the polarization curve and the outcoming arrows show the experimental
operational voltage–current when the biofuel cell is loaded with the charge pump
tion (the pacemaker). Theoretically, upon connecting to the
connected to the pacemaker. The diagram at the top shows the fixed operational pacemaker through the charge pump, the biofuel cell voltage
voltage–current parameters in the system. will drop from Voc to the voltage where the biofuel cell polariza-
tion curve and the load V–I curve first intersect, where it will stay
during operation of the device. The calculated intersection point
used in the present study (Affinity DR 5330L, St. Jude Medical) was roughly estimated at ca. 400 mV with a current draw of
operates at 3 V and 100 mA peak current draw, measured upon ca. 1.2 mA. The experimentally determined steady-state operat-
powering the pacemaker from a power supply with a variable output ing point was ca. 300 mV and 2 mA. The difference between the
(Tenma 72-8700) and registering the pacemaker-generated signals calculated and experimental operating points is most likely due
on an oscilloscope (Agilent DSO-3000).25 Deviating too much from to the biofuel cell output varying over the 45 minutes operation
that input voltage will cause the pacemaker to stop operating, so to of the charge pump and the pacemaker.
ensure the pacemaker is supplied with the proper voltage, a charge The biofuel cell mimicking an implantable device was
pump and a DC–DC converter interface circuit were used to amplify connected to the charge pump and the DC–DC converter inter-
the output voltage of the biofuel cell to match the expected input face circuit, which was further connected to a pacemaker medically
voltage of the pacemaker. The process of boosting the voltage to the used for regulating the heart’s rhythm by administering electrical
level required by the pacemaker comes at a cost of drawing more charges to the myocardial tissue,47 Fig. 5A. In order to analyze
current from the fuel cell, since power is conserved (ignoring circuit
efficiency). Two factors must be considered in the analysis of the
operation of the electrical interface, first is the interface current
demand curve without a load (i.e. without the connected pace-
maker). Using a laboratory power supply we varied the voltage
supplied to our charge pump and DC–DC converter interface circuit
and measured the current drawn by the charge pump in order to
obtain a curve of current as a function of applied voltage, Fig. 4,
curve b. The second factor is the efficiency of the charge pump and
DC–DC conversion. The biofuel cell current required for the pace-
maker application was calculated according to eqn (1): Fig. 5 (A) Experimental setup which includes (from left to right): (a) a sensor
  device for the Medtronic CareLink Programmer, Model 2090, (b) Medtronic
Vcp
 Iapp Reveal XT, Model 9529, an implantable loop recorder (ILR), (c) Affinity DR
 Vbfc 5330L, St. Jude Medical, a pacemaker, (d) the charge pump–DC–DC interface
Ibfc Vbfc ; Vcp ¼ Icp ðVbfc Þ þ (1)
k circuit, (e) the biofuel flow cell with the inlet/outlet connected to a peristaltic
pump (not shown in the scheme). (B) Registered pulses generated by the
where Ibfc(Vbfc,Vcp) is the biofuel cell current draw as a function pacemaker when it is powered by the standard battery. (C) Registered pulses
of the voltage of the biofuel cell and the output voltage of the generated by the pacemaker when it is powered by the biofuel cell.

This journal is c the Owner Societies 2013 Phys. Chem. Chem. Phys., 2013, 15, 6278--6283 6281
View Article Online

PCCP Paper

Fig. 6 Left: typical shape of the electrical pulse generated by a pacemaker (adapted from ref. 52 with permission). Right: an oscilloscope measuring the electrical
pulses produced by the pacemaker activated using the biofuel cell through the charge pump.

the pacemaker performance, the pacemaker output leads were human–machine hybrids.54 The present study is a step on the
connected to the implantable loop recorder (ILR) – a subcuta- long path to the design of bioelectronic self-powered ‘‘cyborgs’’
neous electrocardiographic (ECG) monitoring device. This which can autonomously operate using power from biological
device is medically used for diagnosis and monitoring of the sources. In the next step of this research, which is underway in
heart in patients with recurrent episodes of palpitations or the lab, we will achieve long term operation of the pacemaker
syncope, for long-term monitoring in patients at risk for or with powered by the biofuel cell resulting in the device steady-state
documented atrial fibrillation (AF), and for risk stratification in operation without a battery, but harvesting electrical power
patients who have sustained a myocardial infarction (MI) and from living biological sources.
as well as for those who have certain genetic disorders.48–50 In
the present setup it was used as a medically useful analyzer of Acknowledgements
the electrical pulses produced by the pacemaker receiving the
power from the biofuel cell. The loop recorder output was wire- This research was supported by the National Science Foundation
lessly read using the sensor device of the Medtronic CareLink (Award CBET-1066397) and by the Semiconductor Research
Programmer, Model 2090, typically used for the programming of Corporation (Award 2008-RJ-1839G). The authors acknowledge
pacemakers and loop recorders after implantation and for the several helpful suggestions from Professor Todd Wey of
remote control of their operation.51 The electrical pulses produced Lafayette College.
by the pacemaker and registered by the ILR are very similar when
the pacemaker is powered by a standard battery (WG9438),
References
Fig. 5B, and by an implantable biofuel cell, Fig. 5C, thus
confirming the correct pacemaker operation while receiving 1 U. Schröder, Angew. Chem., Int. Ed., 2012, 51, 7370–7372.
power from the biofuel cell through the charge pump and the 2 C. Himes, E. Carlson, R. J. Ricchiuti, B. P. Otis and
DC–DC converter interface circuit. In addition to the registration B. A. Parviz, IEEE Trans. NanoTechnol., 2010, 9, 2–5.
of the pacemaker-generated pulses with the loop recorder 3 T. Starner and J. A. Paradiso, in Low Power Electronics Design,
described above, we analyzed the electrical pulse shape using ed. C. Piquet, CRC Press, 2004, 1–35.
an oscilloscope similarly to our previous work.25 The electrical 4 J. M. Donelan, Q. Li, V. Naing, J. A. Hoffer, D. J. Weber and
output cable of the pacemaker was connected to an oscilloscope A. D. Kuo, Science, 2008, 319, 807–810.
(Agilent DSO-3000) for recording the generated electrical pulses. 5 F. Davis and S. P. J. Higson, Biosens. Bioelectron., 2007, 22,
The obtained pulse shape corresponded to the standard shape 1224–1235.
of pulses typical for pacemakers,52,53 Fig. 6. The biofuel cell 6 E. Katz, A. N. Shipway and I. Willner, in Handbook of Fuel
produced electrical power which was sufficient for the pace- Cells - Fundamentals, Technology, Applications, ed. W. Vielstich,
maker operation for at least 5 hours (longer stability test was not H. Gasteiger and A. Lamm, Wiley, 2003, vol. 1, Part 4, ch. 21,
performed in this preliminary study). pp. 355–381.
7 J. A. Cracknell, K. A. Vincent and F. A. Armstrong, Chem.
4. Conclusion Rev., 2008, 108, 2439–2461.
8 M. J. Moehlenbrock and S. D. Minteer, Chem. Soc. Rev.,
In conclusion we note that the activation of the pacemaker with 2008, 37, 1188–1196.
the single biofuel cell closely mimicking human physiological 9 N. Mano, F. Mao and A. Heller, ChemBioChem, 2004, 5,
conditions was achieved for the first time. The present result 1703–1705.
promises in the future various medical electronic implants 10 M. J. Moehlenbrock, T. K. Toby, A. Waheed and
powered by implanted biofuel cells and resulting in bionic S. D. Minteer, J. Am. Chem. Soc., 2010, 132, 6288–6289.

6282 Phys. Chem. Chem. Phys., 2013, 15, 6278--6283 This journal is c the Owner Societies 2013
View Article Online

Paper PCCP

11 R. A. Rincon, C. Lau, H. R. Luckarift, K. E. Garcia, E. Adkins, 35 T. Hanashi, T. Yamazaki, W. Tsugawa, S. Ferri,


G. R. Johnson and P. Atanassov, Biosens. Bioelectron., 2011, D. Nakayama, M. Tomiyama, K. Ikebukuro and K. Sode,
27, 132–136. Biosens. Bioelectron., 2009, 24, 1837–1842.
12 E. H. Yu and K. Scott, Energies, 2010, 3, 23–42. 36 T. Hanashi, T. Yamazaki, W. Tsugawa, K. Ikebukuro and
13 F. Scholz and U. Schröder, Nat. Biotechnol., 2003, 21, 1151–1152. K. Sode, J. Diabetes Sci. Technol., 2011, 5, 1030–1035.
14 M. H. Osman, A. A. Shah and F. C. Walsh, Biosens. Bioelectron., 37 F. Pan and T. Samaddar, Charge Pump Circuit Design,
2010, 26, 953–963. McGraw-Hill Professional, New York, 2006.
15 B. E. Logan, Microbial Fuel Cells, John Wiley & Sons, Inc., 38 P. P. Mercier, A. C. Lysaght, S. Bandyopadhyay,
Hoboken, NJ, 2008. A. P. Chandrakasan and K. M. Stankovic, Nat. Biotechnol.,
16 J. J. Fornero, M. Rosenbaum and L. T. Angenent, Electro- 2012, 30, 1240–1243.
analysis, 2010, 22, 832–843. 39 G. Strack, H. R. Luckarift, R. Nichols, K. Cozart, E. Katz and
17 D. Bhatnagar, S. A. Xu, C. Fischer, R. L. Arechederra and G. R. Johnson, Chem. Commun., 2011, 47, 7662–7664.
S. D. Minteer, Phys. Chem. Chem. Phys., 2011, 13, 86–92. 40 P. Vennemann, R. Lindken and J. Westerweel, Exp. Fluids,
18 A. Heller, Phys. Chem. Chem. Phys., 2004, 6, 209–216. 2007, 42, 495–511.
19 S. C. Barton, J. Gallaway and P. Atanassov, Chem. Rev., 2004, 41 C. W. N. Villarrubia, R. A. Rinćon, V. K. Radhakrishnan,
104, 4867–4886. V. Davis and P. Atanassov, ACS Appl. Mater. Interfaces, 2011,
20 P. Cinquin, C. Gondran, F. Giroud, S. Mazabrard, A. Pellissier, 3, 2402–2409.
F. Boucher, J.-P. Alcaraz, K. Gorgy, F. Lenouvel, S. Mathé, 42 A. Zebda, C. Gondran, A. Le Goff, M. Holzinger, P. Cinquin
P. Porcu and S. Cosnier, PLoS ONE, 2010, 5, e10476. and S. Cosnier, Nat. Commun., 2011, 2, art. #370.
21 T. Miyake, K. Haneda, N. Nagai, Y. Yatagawa, H. Onami, 43 L. Hussei, S. Rubenwolf, F. von Stetten, G. Urban,
S. Yoshino, T. Abe and M. Nishizawa, Energy Environ. Sci., R. Zengerle, M. Krueger and S. Kerzenmacher, Biosens.
2011, 4, 5008–5012. Bioelectron., 2011, 26, 4133–4138.
22 M. Rasmussen, R. E. Ritzmann, I. Lee, A. J. Pollack and 44 D. Ivnitski, P. Atanassov and C. Apblett, Electroanalysis,
D. Scherson, J. Am. Chem. Soc., 2012, 134, 1458–1460. 2007, 19, 1562–1568.
23 L. Halámková, J. Halámek, V. Bocharova, A. Szczupak, 45 N. S. Parimi, Y. Umasankar, P. Atanassov and R. P. Ramasamy,
L. Alfonta and E. Katz, J. Am. Chem. Soc., 2012, 134, 5040–5043. ACS Catal., 2012, 2, 38–44.
24 A. Szczupak, J. Halámek, L. Halámková, V. Bocharova, 46 R. C. Dorf and J. A. Svoboda, Introduction to Electric Circuits,
L. Alfonta and E. Katz, Energy Environ. Sci., 2012, 5, 7th ed. Wiley, Hoboken, ch. 11, pp. 531–535.
8891–8895. 47 K. A. Ellenbogen, G. N. Kay, C.-P. Lau and B. L. Wilkoff,
25 K. MacVittie, J. Halámek, L. Halámková, M. Southcott, Clinical Cardiac Pacing, Defibrillation, and Resynchronization
W. D. Jemison, R. Lobel and E. Katz, Energy Environ. Sci., Therapy, 4th edn, Elsevier, Philadelphia, PA, 2011.
2013, 6, 81–86. 48 M. Brignole, R. Sutton, C. Menozzi, R. Garcia-Civera,
26 A. L. Ghindilis, P. Atanasov and E. Wilkins, Electroanalysis, A. Moya, W. Wieling, D. Andresen, D. G. Benditt and
1997, 9, 661–674. P. Vardas, Eur. Heart J., 2006, 27, 1085–1092.
27 I. Willner and E. Katz, Angew. Chem., Int. Ed., 2000, 39, 49 K. Seidl, M. Rameken, S. Breunung, J. Senges, W. Jung,
1180–1218. D. Andresen, A. van Toor, A. D. Krahn and G. J. Klein,
28 N. Mano, F. Mao, W. Shin, T. Chen and A. Heller, Chem. Europace, 2000, 2, 256–262.
Commun., 2003, 518–519. 50 A. D. Krahn, G. J. Klein, A. C. Skanes and R. Yee, Pace-Pacing
29 B. Razavi, Fundamentals of Microelectronics, Wiley, 2013. Clin. Electrophysiol., 2004, 27, 657–664.
30 I. Ieropoulos, J. Greenman and C. Melhuish, Bioelectro- 51 F. Kusumoto and N. Goldschlager, Clin. Cardiol., 2010, 33,
chemistry, 2010, 78, 44–50. 10–17.
31 I. Ieropoulos, J. Greenman and C. Melhuish, Int. J. Energy 52 S. S. Barold, R. X. Stroobandt and A. F. Sinnaeve, Cardiac
Res., 2008, 32, 1228–1240. Pacemakers and Resynchronization Step-by-Step, Wiley-Blackwell,
32 P. Aelterman, K. Rabaey, H. T. Pham, N. Boon and Chichester, UK, 2010, p. 23.
W. Verstraete, Environ. Sci. Technol., 2006, 40, 3388–3394. 53 Manufacturer Specifications - Affinity DR 5330, St. Jude
33 A. Meehan, H. Gao and Z. Lewandowski, IEEE Trans. Power Medical: http://www.medwow.com/med/pacemaker/st-jude-
Electronics, 2011, 26, 176–181. medical/affinity-dr-5330/7305.model-spec.
34 I. Ieropoulos, J. Greenman, C. Melhuish and I. Horsfield, 54 The Bionic Human: Health Promotion for People with
Proc. of the Alife XII Conference, Odense, Denmark, 2010, Implanted Prosthetic Devices, ed. F. E. Johnson, K. S. Virgo,
pp. 733–740. Humana Press, Totowa, NJ, 2006.

This journal is c the Owner Societies 2013 Phys. Chem. Chem. Phys., 2013, 15, 6278--6283 6283

You might also like