Greiner - Guidelines for automatic analyzers – Mindray-BS 480 Edition 2016-10-21

Application: HDL -C
DIRECT METHOD

Preparation: The reagent is ready to use.


Storage: Refrigerate at 2-8°C
Stability: Please see universal package insert sheet

Test Information Judgment Criteria


Test Name HDL-G Absorbance - -
No. ** Incre. Test -
Full Name HDL Decre. Test -
Std. No. ** Lin. Range 1 200
Reaction Parameters Lin. Limit -
Reac. Type Endpoint Subs. Limit -
Pri. Wave. 605nm Reference
Sec. Wave. 700 nm Serum Low High
Direction Increase Male 35 80
Rgt. Blank 48 49 Female 35 80
Reac. Time 80 82 CALIBRATION
Result Setup Rule Two-point linear
Decimal 0 Replicate 2
Unit mg/dl Factor -
Slope 1
Inter. 0 ** User defined

Reagent Volume
R1 (µL) 180
R2 (µL) 60
Sample Volume
Standard (µL) 2.3 - -
Increased (µL) 4.6 - -
Decreased (µL) 2.0 - -