You are on page 1of 11

Designation: E 1012 – 05

Standard Practice for


Verification of Test Frame and Specimen Alignment Under
Tensile and Compressive Axial Force Application1
This standard is issued under the fixed designation E 1012; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope 3.1.2 notched section—the section perpendicular to the


1.1 Included in this practice are methods covering the longitudinal axis of symmetry of the specimen where the
cross-sectional area is intentionally at a minimum value in

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
determination of the amount of bending that occurs during the
application of tensile and compressive forces to notched and order to serve as a stress raiser.
unnotched test specimens in the elastic range and to plastic 3.1.3 nominal percent bending in notched specimens—the
strains less than 0.002. These methods are particularly appli- percent bending in a hypothetical (unnotched) specimen of
cable to the force application rates normally used for tension uniform cross section—equal to the minimum cross section of
testing, creep testing, and uniaxial fatigue testing. the notched specimen, the eccentricity of the applied force in
the hypothetical, and the notched specimens being the same.
2. Referenced Documents (See 11.1.5.) (This definition is not intended to define strain at
2.1 ASTM Standards: 2 the root of the notch.)
E 6 Terminology Relating to Methods of Mechanical Test- 3.1.4 reduced section—the specimen length between the
ing fillets.
E 8 Test Methods for Tension Testing of Metallic Materials 3.2 Definitions of Terms Specific to This Standard:
E 83 Practice for Verification and Classification of Exten- 3.2.1 alignment—the condition of a testing machine and
someter System fixturing (including the test specimen) which can introduce
E 251 Test Methods for Performance Characteristics of bending moments into a specimen during the application of
Metallic Bonded Resistance Strain Gages tensile or compressive forces.
E 466 Practice for Conducting Force Controlled Constant 3.2.1.1 Discussion—This is the overall state of alignment
Amplitude Axial Fatigue Tests of Metallic Materials comprising machine and specimen components.
E 1237 Guide for Installing Bonded Resistance Strain 3.2.2 apparatus—the components of the machine and fix-
Gages turing to be used for testing. This includes all components that
will be used repeatedly for multiple tests.
3. Terminology 3.2.2.1 Discussion—While the strain gaged specimen is not
3.1 Definitions of Terms Common to Mechanical Testing: used for subsequent specimen testing it is included as part of
3.1.1 For definitions of terms used in this practice that are the apparatus.
common to mechanical testing of materials, see Terminology 3.2.3 axial strain—the average of the longitudinal strains
E 6. measured at the surface on opposite sides of the longitudinal
axis of symmetry of the specimen by multiple strain-sensing
devices located at the same longitudinal position as the reduced
1
section.
This practice is under the jurisdiction of ASTM Committee E28 on Mechanical
Testing and is the direct responsibility of Subcommittee E28.04 on Uniaxial Testing. 3.2.3.1 Discussion—This definition is only applicable to
Current edition approved June 1, 2005. Published July 2005. Originally approved this standard. The term is used in other contexts elsewhere in
in 1989. Last previous edition approved in 1999 as E 1012 – 99.
2
mechanical testing.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
3.2.4 bending strain—the difference between the strain at
Standards volume information, refer to the standard’s Document Summary page on the surface and the axial strain (see Fig. 1). In general, the
the ASTM website. bending strain varies from point to point around and along the

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

1
E 1012 – 05

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
NOTE 1—A bending strain, 6B, is superimposed on the axial strain, a, for low-axial strain (or stress) in (a) and high-axial strain (or stress) in (b). For
the same bending strain 6B, a high-percent bending is indicated in (a) and a low-percent bending is indicated in (b).
FIG. 1 Schematic Representations of Bending Strains (or Stresses) That May Accompany Uniaxial Loading

reduced section of the specimen. Bending strain is calculated require the application of tensile or compressive forces under
as shown in Section 11. conditions where alignment is important. The objective is to
3.2.5 eccentricity—the distance between the line of action implement the use of common terminology and methods for
of the applied force and the axis of symmetry of the specimen verification of alignment of test machines, associated fixtures
in a plane perpendicular to the longitudinal axis of the and test specimens.
specimen. 4.2 Unless otherwise specified, axiality requirements and
3.2.6 machine alignment—the condition of the testing ma- verifications should be optional when testing is performed for
chine and all rigid parts of the load train which can introduce acceptance of materials for minimum strength and ductility
bending moments into a specimen during subsequent force requirements. This is because any effects especially from
application. excessive bending, would be expected to reduce strength and
3.2.7 maximum bending strain—the largest value of bend- ductility properties and give conservative results. There may be
ing strain at the position along the length of the reduced section no benefit from improved axiality when testing high ductility
of a straight unnotched specimen at which bending is mea- materials to determine conformance with minimum properties.
sured. (For notched specimens, see 4.9.) Whether or not to improve axiality should be a matter of
3.2.8 percent bending—the bending strain times 100 di- negotiation between the material producer and the user.
vided by the axial strain.
5. Verification of Alignment
3.2.9 rated force—a force at which the alignment is being
measured. 5.1 For ease of reference in other practices, test methods,
3.2.10 specimen alignment—the condition of the test speci- and product specifications, the most commonly used methods
men including the non-rigid parts of the fixturing and the for verifying alignment are listed in Section 6.
positioning of the specimen within the grips which can 5.2 A numerical requirement for alignment should specify
introduce bending moments into the specimen during subse- the force, specimen dimensions, and temperature at which the
quent force application. measurement is to be made. An alternate method employed
when strain levels are of particular importance may be used as
4. Significance and Use described in Practice E 466. When this method is used, the
4.1 It has been shown that bending stresses that inadvert- numerical requirement should specify the strain levels, speci-
ently occur due to misalignment between the applied force and men dimensions and temperature at which the measurement is
the specimen axes during the application of tensile and to be made.
compressive forces can affect the test results. In recognition of 5.2.1 The force at which the bending strain is specified may
this effect, some test methods include a statement limiting the be stated in terms of a yield strength or other nominal specimen
misalignment that is permitted. The purpose of this practice is stress.
to provide a reference for test methods and practices that NOTE 1—For a misaligned load train, the percent bending usually

2
E 1012 – 05
decreases with increasing applied force. (See Curves A, B, and C in Fig. testing without removing the specimen from the testing ma-
2.) However, in some severe instances, percent bending may increase with chine or making any other adjustments that would affect
increasing applied force. (See Curve D in Fig. 2.) alignment during the period between verification and testing.
5.3 Alignment requirements and results can refer to either These type of verifications provide the best measure of the true
an overall test machine capability or to a specific test. This bending strain in a specific test specimen.
distinction should be noted in the results.
NOTE 3—Maintaining a small force on the specimen between verifica-
5.3.1 Verifications of overall test machine capability should tion and testing may be necessary to retain alignment on test machines
be made using a specimen and apparatus made to a similar with non rigid fixturing.
design and of similar materials as those that will be used during
testing, except that any specimen notches may be eliminated. 6. Methods of Verification of Alignment
The same specimen may be used for successive verifications. 6.1 Use this method for verification of machine alignment
The materials and design should be such that only elastic and for measurement of specimen alignment on a particular test
strains occur at the rated force. In cases where the expected test or at specified test conditions.
specimen material is not yet known, use good engineering
6.1.1 Machine Alignment—This part of the method de-
judgement to select a specimen made of a commonly used
scribes the initial alignment of the rigid parts of the fixturing.

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
material for verification.
Machine alignment is initially established when first installing
NOTE 2—To avoid damage to the verification specimen, the sum of the a test machine and when setting up a particular type of rigid
axial strain and the maximum bending strain should not exceed the elastic fixturing configuration on a testing machine. While it may not
limit. change appreciably over time, catastrophic failures in the load
5.3.2 Verifications of specific specimens that are to become train (fixturing or test specimen) or wear may establish the
test specimens following the alignment procedure shall be need to measure and readjust the machine alignment. The
made on the specimen to be tested just prior to or during the machine alignment should be performed any time a change in

NOTE 1—Curve A: Machine 1, threaded grip ends (11)


NOTE 2—Curve B: Machine 2, buttonhead grip ends (11)
NOTE 3—Curve C: Machine 3, grips with universal couplings (7)
NOTE 4—Curve D: schematic representation of a possible response from a concentrically misaligned load train (16)
FIG. 2 Effects of Applied Force on Percent Bending for Different Testing Machines and Gripping Methods

3
E 1012 – 05
the rigid fixturing is required. Machine alignment is often circumference of a cylindrical specimen have been effectively
viewed as a “coarse” alignment. used for this purpose. Displacement measurement devices need
6.1.2 Specimen Alignment—This part of the method de- to have sufficient resolution to detect very small differences in
scribes the positioning and subsequent alignment of the speci- displacements around the specimen. If this method is used
men and all the non-rigid fixturing in the load train. It requires these displacements must be converted to strain before apply-
the use of either a strain gaged specimen of specific geometry ing the bending calculations. Strain should be calculated using
or a mechanical alignment fixture that uses other types of an effective gage length as described in ASTM E 21.
displacement gages to measure the strain applied to the NOTE 5—When multiple extensometers are used, the strain may be
specimen. The strain-gaged specimen is discussed in Section 8. determined by arithmetically averaging outputs. Electrical outputs are
The mechanical alignment fixture is described in Section 7. A thought to be more accurate and reproducible than mechanical outputs.
description of the type of alignment measuring configuration
7.4 Additional Machine and Fixturing Considerations:
(that is, strain gaged specimen or mechanical alignment fix-
7.4.1 Poorly made components and multiple interfaces in a
ture) should be included in the report. Strain gaged specimens
load train can cause major difficulty in attempting to align a test
usually provide better resolution of strain readings, particularly
system. All components in the load train should be machined
at low levels, than do alignment fixtures so they are more
within modern machine shop practices with attention paid to

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
commonly used for this method of measurement. Specimen
perpendicularity, concentricity, flatness and surface finish. The
alignment is often viewed as a “fine” alignment.
number of components should be kept to a minimum.
7. Apparatus 7.4.2 Situations can arise where acceptable alignment can-
not be achieved for a given machine, fixturing and specimen. In
7.1 When multiple strain sensors are used as in 6.1.2,
these cases, redesign and fabrication of any of the components
specimen size limitations may dictate the use of electrical
may be needed to achieve acceptable alignment.
resistance strain gages rather than extensometers or alignment
fixtures employing mechanical linkages. Strain sensors, such as 8. Test Specimen
mechanical, optical, or electrical extensometers, as well as wire
8.1 This practice refers to cylindrical specimens, thick
resistance or foil strain gages, can provide useful displacement
rectangular specimens, and thin rectangular specimens. The
data. The sensitivity of displacement measurement required by
actual specimen geometry is dictated by the test standard to be
an applicable standard or specification depends on the amount
used. These specimens are usually hourglass shaped with a
of bending permitted.
reduced gage section, although other specimens such as those
7.2 For verification using an alignment fixture as in 6.1.2, a
used for compression testing are acceptable.
single extensometer of the nonaveraging type may be used by
8.2 This practice is valid for metallic and nonmetallic test
rotating it to various positions around the perimeter during
specimens.
successive force applications and repeating the measurements
8.3 Quality of machining of test specimens is critical.
as described in 10.5. In general, repeated force applications to
Important features include straightness, concentricity, flatness,
strain levels approaching yielding are not good laboratory
and surface finish. In particular, specimens used for compres-
practice because they may affect the subsequently measured
sion testing may be of the type that uses two parallel plates to
results by deforming or fatiguing the specimen.
apply compression to the ends of the specimen. In these cases,
NOTE 4—Repositioning the extensometer around the specimen does not the parallelism of the specimen ends is extremely important as
usually give highly precise and reproducible results, but nevertheless is a described in ASTM Method E 9.
technique which is useful for detecting large amounts of bending. 8.4 The design of a strain gaged specimen should follow the
7.3 Mechanical fixturing for measurement of strain on a same guidelines as design of standard test specimens. For static
specimen can be an effective way to measure and allow for in (tensile, compressive and creep) testing, specimens conforming
situ adjustments to improve alignment on a test specimen. to Test Methods E 8 are appropriate. For fatigue testing
Fixtures that attach to the specimen shoulders and measure applications, specimens conforming to ASTM E 606 are ap-
displacements at four equally spaced positions around the propriate. The strain-gaged specimen should be as close

NOTE 1—w equals width of specimen.


NOTE 2—d equals distance from edge of specimen to centerline of strain sensor.
FIG. 3 Locations of Strain Sensors on Specimens of Rectangular Cross Section (Numbers Indicate Positions of Strain Sensors)

4
E 1012 – 05
dimensionally to the expected test specimens as possible so different product forms are described in the Test Specimen section of Test
that the same grips and fixturing to be used during testing will Methods E 8.
be used during alignment. The material used for the strain- 9. Calibration and Standardization
gaged specimen should be as close as possible to expected test
specimen materials. If the expected test material is not known, 9.1 All conditioning electronics and data acquisition devices
it is acceptable to use a specimen of a common material that used for the determination of testing system alignment shall be
has similar elastic properties to expected test materials. The calibrated where applicable. The calibration results shall be
alignment specimen should be carefully inspected and the traceable to the National Institute of Standards and Technology
dimensions recorded prior to application of the strain gages. (NIST) or another recognized National Metrology Institute.
9.1.1 Calibration of strain gaged specimens is very difficult
8.5 Strain Gages should be selected that have known stan-
and is not required by this standard. However, great care
dardized performance characteristics as described in Test
should be taken in the manufacture of strain gage specimens
Methods E 251. Strain gage manufacturers provide detailed
used for the determination of alignment. With the exception of
information about the strain gages available. Gages with gage
cases where the gaged specimen is bent, the sources of
lengths of approximately 10 % of the reduced section of the
measurement error due to individual gage misalignment and
specimen or less should be selected. The gages should be as
differences in gage sensitivity can be minimized by acquiring

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
small as practical to avoid any strain averaging effects with
rotational and repeatability data runs.
adjacent gages. Temperature compensated gages that are all of
9.2 Extensometers should be verified in accordance with
the same type and from the same batch should be used.
Practice E 83. Typically extensometers that meet the ASTM
8.6 Strain gages should be installed according to procedures classification B-2 are adequate for most determinations of
outlined in Guide E 1237. A commonly accepted method is to alignment.
make precision shallow longitudinal marks by scribing where 9.3 Strain gages should conform to the requirements of Test
the strain gages are to be applied. The gages are then applied Methods E 251.
with the scribe marks as the longitudinal axis. This can also be
used to mark the transverse axis. This method has the added 10. Procedure
benefit that the gage placements can be inspected at any time
10.1 Temperature variations during the verification test
after the installation. Note that surface preparation is often
should be within the limits specified in the methods or practices
required for mounting strain gages that can have an influence
which require the alignment verification.
on subsequent mechanical properties. For this reason, the strain
10.2 Machine Alignment—This section describes the initial
gaged specimen should not be expected to supply standard
alignment of the rigid parts of the fixturing. Machine alignment
mechanical properties as a normal test specimen would.
is usually established when setting up a particular type of rigid
8.7 Strain gages are to be arranged in at least two sets of fixturing configuration on a testing machine. While it often
four with each set mounted on one of two strain measurement does not change appreciably over time, shock from cata-
planes. For cylindrical specimens, the gages are equally spaced strophic failure in the load train (within the fixturing or test
at 90 degrees to one another around the circumference of the specimen) or wear may establish the need to measure and
specimen. For thick rectangular specimens (that is, those with readjust the machine alignment. Before continuing with sub-
width to thickness ratios of less than three), gages are to be sequent fine alignment activities, one should always be sure
mounted in the center of each of the four faces. For thin that the machine alignment is acceptable.
rectangular specimens (that is, those with width to thickness 10.2.1 Inspect all tooling for the ability of the force bearing
ratios of three or larger), the gages are to be mounted on the surfaces to properly mate with one another and with the
two larger faces in pairs of back to back sensors that are alignment specimen and subsequent test specimens. This
equidistant from the specimen center line. Strain gage place- includes but is not limited to concentricity, perpendicularity
ment is shown in Fig. 3 for rectangular specimens. The sets of and parallelism measurements. Other measurements may be
gages are to be spaced at a distance of 0.753 the reduced needed for specific types of grips and tooling. Re-machine
section length with each set positioned equidistant from the specific parts of the fixturing if necessary.
longitudinal center of the specimen. 10.2.2 Assemble the rigid portion of the fixturing, and
NOTE 6—While the maximum bending strain is usually best measured inspect the position of the tooling on one end of the specimen
using gages placed near the ends of the reduced section, a third set of attachment point with respect to the position of the tooling on
gages located at the geometric center of the reduced specimen may also be the other end of the opposite specimen attachment point. This
used. is often done with a dial indicator setup that allows the user to
NOTE 7—For thick rectangular specimens, the differences in adjacent establish both linear (concentric or parallel) and angular
dimensions of the gage section can lead to differences in the sensitivities differences between the centerlines of the tooling on each end
of gages on these surfaces. This in turn can lead to difficulties in making
of the specimen attachment points. Fig. 4 illustrates linear
adjustments to bring a test setup into good alignment.
(concentric and parallel) and angular differences between the
NOTE 8—Arrays of three gages at 120 degree spacing on cylindrical
specimens are acceptable if there is a compelling reason to use this tooling on the two ends of the rigid portion of the test machine.
configuration. Caution should be taken with this configuration as adjust- Special alignment fixtures may also be employed. Specific
ments to the test machine and fixturing become more complicated and less tolerances are beyond the scope of this standard, but should
intuitive. In addition, it is more difficult to detect a malfunctioning gage. adequate alignment be unachievable, misalignment of these
NOTE 9—Geometry and dimensions of test specimens taken from components may be the reason. Test machines that allow the

5
E 1012 – 05

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
FIG. 4 Illustration of Test Machine Frame and Fixturing (A) Properly Aligned (B) With Concentric Misalignment
and (C) with Angular Misalignment

user to adjust the position of the normally fixed crosshead including the specimen should fit smoothly together with no
should be set up in the position that will be used during testing. extra play. Re-machine specific parts of the fixturing if neces-
Movement of the normally fixed crosshead during testing can sary.
affect alignment results. If moving the normally fixed cross- 10.3.2 Mark the position of any portion of the fixturing that
head during routine testing (that is, between specimens) is will be moved (that is, unthreaded or otherwise repositioned)
needed, the inspection should be performed several times to during the course of normal testing relative to the fixed portion
assure that movement can be made and the crosshead reposi- of the fixturing. This is to assure that the components can be
tioned to the same location without appreciably affecting put together the same way each time.
alignment. 10.3.3 Inspect cylindrical specimens for concentricity and
10.2.3 Adjust the position of the tooling on one end of the perpendicularity between gage section and loading surfaces.
specimen attachment point with respect to the position of the Inspect flat specimens for parallelism and perpendicularity
tooling on the other end of the opposite specimen attachment between the gage section and the gripping surfaces. This is
point to minimize the perpendicularity and the concentricity most often done using a machinist microscope or an optical
(cylindrical specimens) and parallelism (flat specimens). This comparitor. Other measurements may be needed for specimens
may require loosening the attachment hardware of the fixturing of unusual configuration. Document results.
of one end, tapping or shimming it into position and retight- 10.3.4 If a strain gaged specimen is to be used, select a
ening it. Alignment adjustment fixturing is commercially suitable specimen based on inspection results and apply strain
available to facilitate this often-tedious process. gages as described in Section 8. Since this can be a time
10.3 Specimen Alignment—This section describes the posi- consuming and expensive process it is best to have this step
tioning and subsequent alignment of the specimen and all the planned out well in advance of needing the strain gaged
non-rigid fixturing in the load train. It requires the use of either specimen.
a strain gaged specimen of specific geometry or a specialized 10.3.5 Install the strain-gaged specimen or the specimen
alignment fixture that uses other types of displacement gages to and the alignment fixture into the assembly. Zero the strain
measure the strain applied to the specimen. The strain-gaged readings with no force applied to the specimen. It is best to do
specimen is discussed in Section 8. The specialized alignment this with the specimen unattached from one of the sets of grips.
fixture is described in Section 7. The act of gripping a specimen on both ends can be enough to
10.3.1 Inspect any tooling not already inspected as in 10.2.1 introduce unwanted bending.
(the non-rigid parts of the assembly). Establish the position of 10.3.6 Attach the specimen to the remaining grip.
the specimen for fixturing setups with non-rigid members by 10.3.7 Apply a small force to make sure all sensors are
assembling the inspected parts of the load train. Connections, reading properly and then remove the force (see Note 3).

6
E 1012 – 05
10.3.8 Plan the force application cycle such that the ma- 10.3.12 Calculate the percent bending either for the entire
chine applies the maximum force expected during routine force cycle or for the discrete force intervals using the formulas
testing, unless this force induces plasticity in the test specimen. given in Section 11.
If the maximum expected test force is enough to induce 10.3.13 If the calculated percent bending does not meet
plasticity in the test specimen, apply only enough force to well requirements from the test specification, adjustments will need
describe the alignment within the elastic region of the speci- to be made. The commercially available alignment adjustment
men. The actual force level in these cases should be agreed fixturing facilitates this process.
upon with the customer and well documented. This may be a 10.3.14 Small adjustments can have a significant effect on
tensile force, a compressive force, or both. The force may be the measurements. Adjustments are typically made at 90-
applied either manually or automatically. degree intervals around the specimen.
10.3.9 Apply the force cycle while recording strain mea- 10.3.15 Specimens bent in the shape of an “S” (see Fig. 5)
surements continuously if the recording equipment has the require adjustments to be made to the concentricity (for
capability. If this is not possible, at least ten discrete points cylindrical specimens) or perpendicularity (for flat specimens)
between zero force and the maximum force should be re- of the fixturing.
corded. These should be evenly spaced through the force cycle. 10.3.16 Specimens bent in the shape of a “C” (see Fig. 5)

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
If both tension and compression are to be used, record data in require adjustments to the angularity of the fixturing.
a similar manner for both. At a minimum, record the strain at NOTE 10—A combination of the concentricity and angularity adjust-
zero applied force when the specimen is attached to the grips ments are often required to achieve good alignment.
on both ends and at the maximum force conditions (tensile or
compression) expected during upcoming testing. If mechanical 10.3.17 When adjustments are completed, perform force
or hydraulically actuated grips that lock the specimen in place application cycle again as in 10.3.5, record strain information
are used, record the strain at zero applied force both before and as in 10.3.6 through 10.3.11 and perform calculations as in
after the locking mechanisms have been engaged. This shows 10.3.12. Reassess alignment quality as in 10.3.13 and readjust
the influence of the locking mechanism on the bending of the again as necessary.
specimen. 10.4 Specific Test Specimen Alignment—The highest level
of alignment is achieved by direct alignment of a specific test
10.3.10 The force application cycle shall be repeated sev-
specimen. This can be performed by carefully applying strain
eral times with the specimen in the same position to assure
gages to the specimen or by using a high precision alignment
repeatability. Repeatability must be within acceptable limits
fixture that uses other types of displacement gages to measure
related to the testing standard’s requirement for total bending.
the strain applied to the specimen. In either case, the presence
For example, if the overall allowable bending strain as pre-
of the strain measuring devices can have some impact on the
scribed by a testing standard is 62 %, an acceptable demon-
true specimen behavior. Since this level of alignment requires
strated repeatability might be 60.25 % depending on other
gages to be put on each individual test specimen, it can be very
error contributing factors.
costly and is often not necessary if the previous steps show
10.3.11 Since the bending measurements are to be made acceptable alignment. The need for this level of alignment
under the same conditions as those seen in a test, it is important should be agreed upon with the customer before attempting it.
to simulate the various ways of installing the specimen and the
non-rigid portion of the fixturing. To simulate this, the align- 11. Calculation and Interpretation of Results
ment specimen should be removed and repositioned in the
grips. Installing the specimen in the same orientation as it 11.1 Results of interest usually include axial strain, local
previously was installed will provide information on repeat- bending strains, maximum bending strain, and percent bend-
ability of the specimen. Installing the specimen in another ing.
orientation (that is, rotating it or inverting it) will further 11.1.1 Cylindrical Specimens, Three Strain Sensors—For
characterize the alignment of the fixturing. Strain gaged three strain gages or extensometers, equally spaced around the
specimens always have some eccentricity, though preparation circumference of a specimen of circular-cross section in a place
as described in Section 8 will minimize this. Alignment perpendicular to and at the center of the gage length, see the
specimens can get damaged or bent over time and use. Careful following equations:
handling and storage will minimize this. Still, measurements axial strain, a 5 ~e1 1 e2 1 e3!/3 (1)
should be made such that eccentricities to the alignment where:
specimen can be characterized and separated from the fixturing e1, e2, and e3 = measured strains at the three locations, and
contribution to the overall alignment. This can be accom- where e1 $ e2 $ e3.
plished by measuring strains under the force cycle described in
b1 5 e1 2 a (2)
10.3.8 in the original orientation, 180 degrees and again back
in the original orientation. In addition, rotating the specimen b 2 5 e2 2 a
and measuring the alignment at 90, 180 and 270 degrees b3 5 e3 2 a
relative to the original position will simulate all possible
specimen installation positions. Inverting the specimen and where:
b = bending strain.
repeating these measurements will add to the confidence level
of the overall alignment. u 5 tan21@~2/=3!~b2/b1 1 1/2!# (3)

7
E 1012 – 05

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
FIG. 5 Illustration of Strain Gaged Specimen (A) Properly Aligned, (B) With Concentric Misalignment and (C) With Angular Misalignment

where: 11.1.3.1 For thick specimens of rectangular cross section


u = direction of maximum bending and is measured from with strain sensors placed as shown in Fig. 3a, see the
the highest-reading strain sensor toward the next following equation:
highest-reading strain sensor. Finally, axial strain, a 5 ~e1 1 e2 1 e3 1 e4!/4 (10)
B 5 b1/cos u (4) where e1 and e3 are measured strains at the center of the
where: specimen thickness on opposite faces, and e2 and e4 are
B = maximum bending strain. corresponding values for the wide faces.
11.1.3.2 The local bending strains b1, b2, b3, b4 are calcu-
PB 5 ~B/a! 3 100 (5)
lated by the equations in 11.1.2.
where: 11.1.3.3 The maximum bending strain, B, is calculated from
PB = percent bending. the following equation:
11.1.2 Cylindrical Specimens, Four Strain Sensors—For B 5 | b1 2 b3 | /2 1 | b2 2 b4 | /2 (11)
four strains gages or extensometers, equally spaced around the
circumference of specimens of circular cross section, see the 11.1.3.4 Percent bending, PB, is calculated as follows:
following equations: PB 5 ~B/a! 3 100 (12)
axial strain, a 5 ~e1 1 e2 1 e3 1 e4!/4 (6) 11.1.4 Thin Rectangular Specimens, Four Strain Sensors:
where e1, e2, e3, and e4 are the measured strains at the four 11.1.4.1 For thin specimens of rectangular cross section
locations and the subscript indicates the order around the with strain sensors placed as shown in Fig. 3b, see the
specimen. following equation:
local bending strain, b1 5 e1 2 a (7) axial strain, a 5 ~e5 1 e6 1 e7 1 e8!/4 (13)
b 2 5 e2 2 a 11.1.4.2 Equivalent strains at the center of the four faces, if
b 3 5 e3 2 a strain sensor placement were possible as shown in Fig. 3a, are
b 4 5 e4 2 a given by:
e1 5 a 2 [a 2 ~e5 1 e8!/2]@w/~w 2 2d!# (14)
and maximum bending strain,
e3 5 a 2 [a 2 ~e6 1 e7!/2]@w/~w 2 2d!#
B 5 1/2=~b1 2 b3! 2 1 ~b2 2 b4! 2 (8)
e2 5 ~e5 1 e6!/2
and
e4 5 ~e7 1 e8!/2
PB 5 ~B/a! 3 100 (9)
where, as shown in Fig. 3b:
11.1.3 Thick Rectangular Specimens, Four Strain Sensors:

8
E 1012 – 05

w = width of the broad face, and 12.1.3 Rated maximum force used in verification indicating
d = distance from edge of specimen to position of strain tension or compression. If tension and compression are used,
sensor. include rated maximum for both,
11.1.4.3 The maximum bending strain B, and the percent 12.1.4 Description of specimen (material and dimensions),
bending, PB, are calculated from the equations in 11.1.3.3 and 12.1.5 Description of strain measuring equipment, includ-
11.1.3.4. ing:
11.1.4.4 The equations for the rectangular cross section, 12.1.5.1 Type of strain measuring device (strain gages,
given in 11.1.3, are used to complete the calculation. extensometer, etc.),
11.1.5 For tests on notched specimens of circular cross 12.1.5.2 Precision and sensitivity of the strain measuring
section, the nominal percent bending at the root of the notch is system,
obtained by calculating the percent bending in the reduced 12.1.5.3 Location (s) of the sensors on the specimen,
section as described in 11.1.1 or 11.1.2 and multiplying the 12.1.5.4 Method of attaching the strain sensor(s) to the
result by the ratio of the diameter of the reduced section to the specimen, and
diameter at the root of the notch. 12.1.6 Description of fixturing, including method of grip-
11.1.6 For tests on notched specimens of rectangular cross ping, dimensions and types of couplings and joints, length of

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
section with the notch root axis in the thickness direction, the load train, and any other pertinent details.
nominal percent bending is calculated as follows:
13. Precision and Bias
@b1~h/h8! 1 b2#
a 3 100 (15) 13.1 The precision of the measurement of specimen align-
ment under applied tensile forces varies with such test condi-
where: tions as temperature, stress, configuration of load train, and
h = the distance between the notched sides adjacent to the material. At present, the available data are not of a type that
notch, permits meaningful analysis of the precision of the measure-
h8 = the distance between notch roots, and b1, b2 are ment. It is the intention of Committee E28 to obtain the
defined in 11.1.2. necessary data from an interlaboratory test program based on
11.1.6.1 Similarly, when the notches are in the width face, this practice.
the nominal percent bending is calculated as follows: 13.2 The bias of the measurement of specimen alignment
b1 1 [b2~h/h8!# under application of tensile and/or compressive forces varies
a 3 100 (16) with such test conditions as temperature, stress, quality of
machining of test specimens, fixturing and material. Since the
12. Report bending strains used to measure alignment are determined from
12.1 Report the following information: ratios of strain measurements from three or more strain
12.1.1 Values of bending strain or percent bending, the sensors, the absolute accuracy of the strain sensor calibration is
corresponding forces (or strains) for these values, and method not important (see 9.1). No direct measure of bias is available,
used, because the identical test conditions cannot be duplicated
12.1.2 Test temperature, during a calibration run and an actual test.

APPENDIX

(Nonmandatory Information)

X1. SOURCES AND EFFECTS OF SPECIMEN MISALIGNMENT UNDER APPLIED AXIAL TENSILE
OR COMPRESSIVE FORCES

X1.1 Source of Misalignment Under Applied Axial bottom grip centerline, poor conformance of specimen center-
Tensile Forces line to top and bottom grip centerlines, and asymmetric
X1.1.1 The usual procedure in a uniaxial tension test is to machining of the test specimen itself. A combination of these
apply a tensile force to a specimen through grips attached to a three sources of misalignment always operates in any test
test machine with suitable fixturing and then correlate the strain under tensile forces. The occurrence of misalignment is recog-
response of the specimen, as measured with an appropriate nized in a wide range of Mechanical Testing and Fatigue and
strain measuring transducer, with the applied stress. In the case Fracture activities dealing with a variety of materials.
of ideal alignment, the top and bottom grip centerlines are X1.1.2 The characteristic elastic strain gradients resulting
precisely in line with one another and with the centerlines of from misalignment are such that the extreme elastic strains
other components of the loading train. Moreover, they are occur at the surface. These gradients can significantly influence
precisely in line with the specimen centerline. Finally, the the results of a tension test, especially results at strains less
specimen is symmetric about its centerline. Departures from than 0.002 where significant plastic strain and accompanying
the ideal situation are caused by poor alignment of the top and strain hardening have not yet contributed to evening out the

9
E 1012 – 05
gradients. Therefore, it is important to recognize the effects of analyzed using a series of displacement gages and character-
misalignment on the stresses and strains measured in studies of izing the displacements encountered between the compressive
the fracture strength of materials in a brittle state, stress-rupture load bearing components. This can be a complex entity to
life, creep, notched-tensile specimens, fatigue, plastic micros- accurately measure as the surfaces may deflect in any of the
train, alloy strengthening, and surface-sensitive strength. three orthogonal directions, and in a non-linear fashion (12).
X1.1.3 The objective of any effort to improve alignment is X1.2.3 Acceptable alignment in both tension and compres-
to bring the centerlines of all pertinent fixturing into precise sion can be difficult to achieve. Adjustments using alignment
alignment. Logically, the first piece of hardware on which to enhancing fixturing often have the opposite effect in tension
focus attention is the testing machine itself. Testing machines than they do in compression. For this reason, a compromise
as-received from manufacturers may have deviations between between the quality of alignment in tension and the quality of
top and bottom grip centerline positions of 0.03 to 3.18 mm alignment in compression may be needed.
(0.001 to 0.125 in.) or more. Moreover, further misalignment
may develop as applied forces cause machine frame deflection X1.3 Effects of Misalignment on Test Results
or as nonaxial crosshead separation occurs. In the worst case,
deviations in this range have been reported to lead to eccen- X1.3.1 Bending stresses associated with misalignment be-
tween the fixturing and the specimen axes have been shown to

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
tricities resulting in a 50 to 100 % difference between extreme
surface bending strains and average strain. affect the results of tension, compression and combined
X1.1.4 Another important factor in the alignment process is tension-compression tests (13-19). In routine tension tests of
the tolerances specified for the machining of fixturing and test most engineering materials, bending stresses will be insignifi-
specimens. In ordinary machine shop practice, tolerances cant if sufficient plastic flow occurs during the test to eliminate
usually range +0.05 to +0.25 mm (+0.002 to +0.010 in.). These the bending stresses. However, when testing under conditions
tolerances may not be tight enough and may contribute to poor where plastic flow is limited by inherent brittleness of the test
alignment when the components of a loading train are as- specimen material, or by need for measurements near the
sembled. In the worst case, these tolerances have been reported elastic limit, or when plasticity is confined to a small volume
to lead to eccentricities resulting in a 50 to 100 % difference (specimens with stress concentration such as notches), small
between extreme surface bending strains and average strain. misalignment may give rise to variable bending stresses which
X1.1.5 There are two further considerations for the devel- have noticeable effects on the test results. For example,
opment of good alignment. One deals with the type of fixturing Morrison (8) noted that the yield stress of carefully machined
used. Some of these include threaded-vs-nonthreaded joints, mild steel specimens tested in torsion exhibited a 61 %
spherical seats and universal joints with low friction, cross variation from the mean, whereas the yield stresses of the same
flexures, fluid couplings, and other couplings which tend not to steel specimens tested in tension exhibited a 65 % variation.
transmit a bending stress. The other relates to specimen design, Morrison concluded that the larger variation in tensile yield
such as length and length-to-diameter ratio. The approach to stresses resulted from misalignment rather than from micro-
promoting good alignment has been discussed in several papers structural variations, and he stated that “with the ordinary
(1-11).3 standard of accuracy in cutting the screwed ends of the
specimens, the slackness in the thread was quite sufficient to
X1.2 Sources of Misalignment under Applied Axial allow the specimen to take up and retain under load an
Compressive Forces eccentricity in the shackles which would account for the
variation in results.”
X1.2.1 Misalignment in compression takes on similar char-
acteristics to misalignment in tension, however different as- X1.3.2 Schmieder et al (9, 10) found that bending ranged
pects of the test machine, fixturing and test specimen can cause from 5 to 27 % and depended on specimen attachment to the
it. Compressive force application to a specimen usually makes fixturing, prior force application, and type of testing machine.
use of an entirely different set of mating surfaces than tensile These authors concluded that “most of the nonaxiality of
force application. Force is applied to threads on the opposite loading appears to be due to loose threads or machining
side, grip surfaces can change, crossheads must be locked from imperfections in the couplings.” Jones and Brown (11) dem-
opposite sides, and actuators must be forced from opposite onstrated that, at fixed stress, simply rotating a load-train
sides to that of tensile force application. For this reason, component through 360° about the longitudinal axis changed
alignment in tension is often completely different from align- the percentage of bending by a factor of more than 5, from 8 to
ment in compression. 43 %. In an experiment with other equipment, Jones and
X1.2.2 Machine Lateral Stiffness—An additional com- Brown (11) found that bending could be varied between about
pounding problem in compression is the ability of the test 2 and 14 %, depending on the relative rotational positions of
machine to maintain its rigidity during compressive force the specimen and of the top and bottom grips. Hence, a fourth
application. If extreme difficulties are encountered in achieving item which influences bending might be added to the three
adequate specimen alignment in compression, it may be cited by Schmieder et al, namely, the rotational registry of the
because of poor lateral stiffness of the test machine. This can be components of the fixturing.
X1.3.3 Robinson (13) reported a 40 to 60 % decrease in the
uniaxial tension–tension fatigue life of steel bolts when the
3
The boldface numbers in parentheses refer to the list of references at the end of bending microstrain increased by a factor of two. Jones et al
this standard. (14) demonstrated a continuous decrease (ranging from 80 to

10
E 1012 – 05
90 %) of notch-rupture life of a chromium-molybdenum- quantifying uncertainties in low cycle fatigue testing. This
vanadium steel, at 414 MPa 538°C (60 ksi 1000°F), as work illustrates the difficulty sometimes seen in achieving
eccentricity increased from a negligible value to 2.5 mm (0.1 proper alignment and the effect it has on test results. In the
in.) Christ (15) showed that results of plastic microstrain study, laboratories are categorized by the classification of
studies and other pre-yield studies are ambiguous unless effects alignment they were able to achieve. The test results are shown
of misalignment on the average microstrain are recognized. in relation to this classification. The laboratories with the best
Attention was directed to this point by McVetty (16) as early as alignment had the highest fatigue lives. It is clear from this data
1928, but it has been frequently overlooked since then. that poor alignment causes artificially low lives in low cycle
X1.3.4 Kandil (17) demonstrated the effects of misalign- fatigue testing. This work led to a VAMAS procedure that
ment on fatigue life results in an interlaboratory study aimed at includes the classification system for quality of alignment (18).

REFERENCES

(1) Christ, B. W., and Swanson, S. R., “Alignment Problems in the Tensile Creep Machines,” Elevated Temperature Testing Problem Areas,

FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO BHARAT FORGE LIMITED - MUNDHWA, PUNE
Test,” Journal of Testing and Evaluation, Vol 4, No. 6, November ASTM STP 488, ASTM, 1971, p. 43.
1976, pp. 405–417. (11) Jones, M. H., and Brown, W. F., “An Axial Loading Creep Machine,”
(2) Wu, H. C., and Rummler, D. R., “Analysis of Misalignment in the ASTM Bulletin, ASTM No. 211, January 1956, p. 53.
Tension Test,” Transactions: Series H, Journal of Engineering Mate- (12) Lohr, R. D. “Materials Testing Machine Design—Reducing Speci-
rials and Technology, Vol 101, n.1, ASME January 1979, pp, 68–74. men Bending through Improved Alignment and Lateral Stiffness,”
(3) Holmes, A. M. C., “Continuous Servo-Controlled Alignment of Notes from The Effects of Misalignment on Uniaxial Testing, a
Specimens in Materials Testing,” Experimental Mechanics, Vol 15, workshop sponsored by ASTM Committee E28, Norfolk, VA, Nov 4,
No. 9, September 1975, pp. 358–364. 1998.
(4) Webb, J. N., “A System For the Axial Loading in Creep Specimens,” (13) Robinson, D., “Misalignment Detector for Axial Loading Fatigue
Structures Dept., Royal Aircraft Establishment, Farnborough, England. Machines,” Technical Note 480, National Bureau of Standards,
Her Majesty’s Stationary Office, London, 1977. Washington, DC, 1972.
(5) Jones, M. H., Bubsey, R. T., Succop, G., and Brown, W. F., Jr., “Axial
(14) Jones, M. H., Shannon, J. L., Jr., and Brown, W. D., Jr., “Influence of
Alignment Fixture For Tension Tests of Threaded Specimens,” Journal
Notch Preparation and Eccentricity of Loading on Notch Rupture
of Testing and Evaluation, Vol 2, September 1974, p. 378.
Life,” Proceedings, ASTM, Vol 57, 1957, p. 833.
(6) Jones, M. H., and Brown, W. F., Jr., “Note on Performance of Tapered
(15) Christ, B. W., “Effects of Misalignment on the Pre-Macro Yield
Grip Tensile Loading Devices,” Journal of Testing and Evaluation, Vol
Region of the Uniaxial Stress–Strain Curve,” Metallurgical Transac-
3, No. 3, 1975, pp. 179–181.
tions, AIME, Vol 4, No. 8, 1973, pp. 1961–1965.
(7) Penny, R. K., Ellison, E. G., and Webster, G. A., “Specimen Alignment
and Strain Measurement in Axial Creep Tests,” Materials Research (16) McVetty, P. G., “Testing of Materials at Elevated Temperatures,”
and Standards, Vol 6, No. 2, 1966, pp. 76–84. Proceedings, ASTM, Vol 28, 1928, p. 60.
(8) Morrison, J. L. M., “The Yield of Mild Steel with Particular Reference (17) “Private Communication to B. W. Christ from H. S. Starrett,”
to the Effect of Size of Specimen,” Journal and Proceedings, The Southern Research Institute, Birmingham, AL, April 1979.
Institution of Mechanical Engineers, London, Vol 140, No. 3, 1940, p. (18) Kandil, F. A., “Recent Intercomparisons on Low Cycle Fatigue and
193–223. Alignment Measurements,” VAMAS Report No. 41, ISSN 1016-2186,
(9) Schmieder, A. K., “Measuring the Apparatus Contribution to Bending Feb 2003.
in Tension Specimens,” Elevated Temperature Testing Problem Areas, (19) Kandil, F. A., “A Procedure for the Measurement of Machine
ASTM STP 488, ASTM, 1971, p. 15. Alignment in Axial Testing,” VAMAS Report No. 42, ISSN 1016-
(10) Schmieder, A. K., and Henry, A. T., “Axiality Measurements on Fifty 2186, Feb 2003.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

11

You might also like