Professional Documents
Culture Documents
V e n t i l a t o r Sy s t em
Service M anual
4-070496-00 Rev A
August, 2003
Definit ions
This m anual uses three special i ndicators to convey i nform ation of a specifi c nature.
They include:
Warning
Indicates a condition that can endanger the patient or the ventilator operator.
Caut ion
Indicates a condition that can damage the equipment.
NOTE:
Indicates points of particular interest that make operation of the ventilator more efficient or
convenient.
i 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual Rev. A (08/ 03)
Preface
Puritan Ben nett does recom m end that users of its products that require cleaning and
sterili zation /di si nfection consider the National Standards and Recommended Practices for
Sterilization publi sh ed by th e Associati on for the Advan cem ent of M edi cal
Instrum entati on (AAM I), as well as th e foll owi ng Cen ter for Di sease Con trol (CDC)
publi cati ons: Guidelines for Maintenance of In-use Respiratory Therapy Equipment and
Guidelines for Prevention of Nosocomial Pneumonia.
• Pati en ts on li fe-support equipm ent should be appropri ately m oni tored by competen t
m edical personnel an d sui tabl e m oni torin g devices.
• The 840 Ventil ator System i s not in tended to be a com prehensive m on itori ng device an d
does n ot acti vate alarm s for all types of dan gerous con diti on s for pati ents on li fe-support
equipm en t.
• For a thorough un derstan din g of ventil ator operati on s, be sure to th orough ly read the 840
Ventilator System Operator's and Technical Reference Manual before attem pti ng to use the
system .
• Before activating any part of th e venti lator, be sure to check the equipm ent for proper
operati on and, i f appropri ate, run SST (Short Sel f Test).
• Do n ot use sh arp obj ects to m ake sel ections on the graphical user interface (GUI) display
or keyboard.
• Federal law (U.S.) restricts the sale of thi s device to, or by th e order of, an y ph ysi cian.
• Check the ven tilator peri odical ly as outlin ed in this m anual; do n ot use i f defective.
Im m ediatel y replace parts th at are broken , missi ng, obviously worn, distorted, or
contam in ated.
• An altern ative source of ven tilation shoul d al ways be availabl e when usin g th e 840
Ventil ator System .
• The 840 Venti lator System is a m em ber of the 800 Series™ fam ily of products. An y
accessory whose m odel n um ber i s 80x (for exampl e, th e 802 Backup Power Source or 806
Com pressor Un it) operates with al l 800 Series ven ti lators. An accessory whose m odel
num ber is 84x operates onl y wi th a m odel 840 Ven ti lator System .
2003 – 10
For venti lators whose seri al num bers begi n wi th 3510, the year of m anufacture is in dicated by
the seri al num ber’s 5th and 6th digits. In the exam ple below, a ventil ator with seri al num ber
3510021223 was the 1223rd uni t buil t i n 2002.
3510 02 1223
Sequential number
Year of manufacture
iii 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual Rev. A (08/ 03)
SECTION
C ONTENTS
1 General information
1.1 How to use this manual ............................................................................................... 1-1
1.2 General product description ........................................................................................ 1-1
1.3 Configuration information ...........................................................................................1-2
1.4 Accessories.................................................................................................................. 1-2
1.5 Specifications .............................................................................................................. 1-3
1.6 Compliance and approvals .......................................................................................... 1-7
1.7 Technical information..................................................................................................1-8
1.8 Range, resolution, accuracy, and new patient/ default settings .................................. 1-10
1.9 Tools, equipment, and service materials ....................................................................1-19
1.10 Periodic maintenance .............................................................................................. 1-22
1.11 Service kits ..............................................................................................................1-24
1.12 Controls and indicators ......................................................................................... 1-24
1.13 Onscreen symbols and abbreviations....................................................................... 1-41
1.14 Ventilator serial numbers and software version ........................................................ 1-41
1.15 Service philosophy...................................................................................................1-41
2 Theory of operation
2.1 Major ventilator subassemblies....................................................................................2-1
2.1.1 Breath delivery unit (BDU) ...................................................................................2-2
2.1.2 Graphic user interface (GUI) ................................................................................2-2
2.1.3 806 Compressor Unit .......................................................................................... 2-3
2.1.4 802 Backup Power Source (BPS) .......................................................................... 2-4
2.1.5 Cart ..................................................................................................................... 2-4
2.1.6 Patient system ..................................................................................................... 2-5
2.2 Operational overview .................................................................................................. 2-6
2.3 Pneumatic system ....................................................................................................... 2-9
2.3.1 Inspiratory module ............................................................................................ 2-14
2.3.1.1 Gas supply conditioning subsystem .........................................................2-15
2.3.1.2 Flow control subsystem ...........................................................................2-20
2.3.1.3 Safety valve and inspiration monitoring subsystem .................................. 2-23
2.3.1.4 Inspiratory module operation .................................................................. 2-28
2.3.2 Patient System ................................................................................................... 2-31
2.3.2.1 Patient system components.....................................................................2-32
2.3.2.2 Patient system operation ......................................................................... 2-33
2.3.3 Exhalation module............................................................................................. 2-33
2.3.3.1 Exhalation module components ..............................................................2-34
2.3.3.2 Exhalation module operation ................................................................... 2-35
840 Ventilator System Service Manual 4-070496-00 Rev. A (08/ 03) vii
Cont ent s
2.3.4 806 Compressor Unit ........................................................................................ 2-38
2.3.4.1 806 Compressor unit components.......................................................... 2-38
2.3.4.2 806 Compressor unit operation............................................................... 2-41
2.4.1 Summary of electrical components ................................................................... 2-42
2.4.2 Overview of electrical system operation ............................................................ 2-47
2.4.3 ac distribution components............................................................................... 2-50
2.4.3.1 Power cord ............................................................................................. 2-50
2.4.3.2 ac panel .................................................................................................. 2-50
2.4.3.3 Power switch........................................................................................... 2-53
2.4.4 Power supply .................................................................................................... 2-54
2.4.5 BPS................................................................................................................... 2-56
2.4.6 Card Cage ........................................................................................................ 2-58
2.4.6.1 Motherboard PCB................................................................................... 2-60
2.4.6.2 BDU CPU PCB......................................................................................... 2-63
2.4.6.3 Analog interface (AI) PCB........................................................................ 2-67
2.4.7 Data key subsystem .......................................................................................... 2-69
2.4.8 GUI................................................................................................................... 2-69
2.4.8.1 GUI CPU PCB ......................................................................................... 2-69
2.4.8.2 Touch frame PCB .................................................................................... 2-75
2.4.8.3 Keyboard assembly with knob ................................................................. 2-76
2.4.8.4 GUI LED PCB........................................................................................... 2-78
2.4.8.5Backlight inverter PCB and LCD lamps ........................................................ 2-79
2.4.8.6 GUI alarm assembly ................................................................................ 2-80
2.4.9 BDU LED PCB ................................................................................................... 2-81
2.4.10 Inspiratory electronics PCB.............................................................................. 2-81
2.4.11 Exhalation transducer PCB .............................................................................. 2-83
2.4.11.1BD (continuous-tone) alarm assembly ....................................................... 2-85
2.4.12 806 Compressor unit ...................................................................................... 2-85
2.4.12.1806 compressor unit ac power distribution components and motor.......... 2-86
2.4.12.2 806 compressor PCB............................................................................... 2-86
2.4.12.3806 Compressor unit operation ................................................................ 2-90
2.5 Breath delivery .......................................................................................................... 2-92
2.5.1 Inspiration......................................................................................................... 2-92
2.5.1.1Pressure triggering ...................................................................................... 2-94
2.5.1.2 Flow triggering ....................................................................................... 2-94
2.5.1.3 Time-cycling method .............................................................................. 2-94
2.5.1.4 Operator triggering................................................................................. 2-94
2.5.2 Exhalation ......................................................................................................... 2-94
2.5.2.1Time-cycling method .................................................................................. 2-95
2.5.2.2 End-inspiratory flow method ................................................................... 2-95
2.5.2.3 Airway pressure method.......................................................................... 2-95
2.5.2.4 Time limit................................................................................................ 2-95
2.5.2.5 High circuit pressure limit........................................................................ 2-95
2.5.2.6 High ventilator pressure limit .................................................................. 2-95
viii 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Cont ent s
2.6 Other hardware operations .......................................................................................2-97
2.6.1 Gas supply and control...................................................................................... 2-97
2.6.2 Data monitoring................................................................................................ 2-97
2.6.3 Pressure transducer autozero ............................................................................. 2-98
2.6.4 Power monitoring and power fail handling......................................................2-100
2.6.4.1 Loss of power source .............................................................................2-100
2.6.4.2 Supply voltage monitoring .................................................................... 2-101
2.7.1 Safety valve open (SVO) state.......................................................................... 2-101
2.7.2 Occlusion handling ......................................................................................... 2-103
2.7.2.1 When the ventilator declares an occlusion .............................................2-103
2.7.2.2 Occlusion detection and handling ......................................................... 2-103
3 Self t est s
3.1 Introduction ................................................................................................................3-1
3.2 How to enter Service Mode ......................................................................................... 3-1
3.3 Self tests and background checks ................................................................................ 3-1
3.3.1 POST................................................................................................................... 3-1
3.3.2 SST...................................................................................................................... 3-1
3.3.3 EST ..................................................................................................................... 3-2
3.3.4 Background checks.............................................................................................. 3-2
3.3.5 When self tests are run ........................................................................................ 3-2
3.4 Power on self test (POST) ............................................................................................3-3
3.4.1 Safety ................................................................................................................. 3-7
3.4.2 POST characteristics ............................................................................................3-7
3.4.3 POST following power interruptions.................................................................... 3-8
3.4.4 POST user interface ............................................................................................. 3-8
3.4.5 Structure of POST................................................................................................ 3-9
3.5 SST (short self test) ....................................................................................................3-11
3.5.1 When to run ......................................................................................................3-11
3.5.2 Hardware requirements..................................................................................... 3-12
3.5.3 Running SST......................................................................................................3-13
3.6 EST (extended self test) ............................................................................................. 3-17
3.6.1 Description........................................................................................................3-17
3.6.2 When to run ......................................................................................................3-17
3.6.3 Hardware requirements..................................................................................... 3-18
3.6.4 Running EST...................................................................................................... 3-18
4 Service mode
4.1 Accessing service mode ............................................................................................... 4-1
4.2 Service mode functions ............................................................................................... 4-3
4.2.1 SST RESULT: Displaying SST results......................................................................4-5
4.2.2 DIAG LOG: Displaying error and status logs......................................................... 4-5
4.2.3 Alarm log: Displaying the alarm history ............................................................... 4-7
4.2.4 VENT CONFIG: Displaying software revisions and serial numbers ........................4-7
4.2.5 OPERATION TIME: Displaying ventilator and compressor elapsed hours.............. 4-7
4.2.6 TEST SUMMARY: Displaying results from EST and SST ......................................... 4-7
4.2.7 EST: Extended self test ......................................................................................... 4-7
4.2.8 DATE/ TIME: Adjusting date and time................................................................... 4-7
4.2.9 EXIT: Exiting service mode................................................................................... 4-7
5 Performance verification
5.1 Tools, test equipment, and service materials ............................................................... 5-1
5.2 When to run ............................................................................................................... 5-1
5.3 Preliminary ventilator cleaning and inspection ............................................................ 5-4
5.4 Preliminary ventilator setup ......................................................................................... 5-5
5.5 Preliminary calibrations and tests ................................................................................ 5-5
5.6 Performance verification guidelines............................................................................. 5-7
5.7 Performance verification tests...................................................................................... 5-7
5.7.1 Electrical safety test ............................................................................................ 5-8
5.7.2 Ground isolation check ....................................................................................... 5-8
5.7.3 Extended self test (EST) ....................................................................................... 5-9
5.7.4 Regulator setting verification............................................................................... 5-9
5.7.5 Serial loopback test (10.4-inch GUI only) .......................................................... 5-10
5.7.6 Performance verification using PTS 2000 Performance
Test System and BreathLab 840 VTS software ................................................... 5-10
5.7.7 Manual ventilator check using equipment other than
PTS 2000 Performance Test System ................................................................... 5-10
6 Diagnostic codes
6.1 Introduction................................................................................................................ 6-1
6.2 Reference Tables ......................................................................................................... 6-1
6.3 Troubleshooting.......................................................................................................... 6-2
6.4 POST fault handling ................................................................................................... 6-2
6.5 Diagnostic CPU LED arrays.......................................................................................... 6-3
6.6 Diagnostic codes......................................................................................................... 6-4
6.6.1 How to interpret diagnostic codes ...................................................................... 6-4
6.7 Organization of diagnostic codes table ....................................................................... 6-5
6.8 System Diagnostic Log and BDU POST analog devices test ....................................... 6-54
7 Alarm handling
7.1 Alarm classifications..................................................................................................... 7-1
7.2 Responding to alarms..................................................................................................7-2
xii 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Cont ent s
8.15.5 Power supply assembly.................................................................................... 8-48
8.15.5.1 Removing power supply assembly ........................................................... 8-48
8.15.5.2 Installing power supply assembly............................................................. 8-48
8.15.6 Power switch (S1)............................................................................................ 8-49
8.15.6.1 Removing power switch (S1) ...................................................................8-49
8.15.6.2 Installing power switch (S1).....................................................................8-49
8.15.7 Humidifier receptacle (100 – 120 V models only) ............................................ 8-50
8.15.7.1 Removing humidifier receptacle............................................................... 8-50
8.15.7.2 Installing humidifier receptacle ................................................................ 8-50
8.15.7.3 ac panel................................................................................................... 8-51
8.15.8 Inspiratory module ..........................................................................................8-52
8.15.8.1 Inspiratory module modifications............................................................. 8-52
8.15.8.2 Additional noise suppression (ferrite cores) .............................................. 8-52
8.15.8.3 Inspiratory module O-rings......................................................................8-53
8.15.8.4 Air inlet filter (F2)..................................................................................... 8-53
8.15.8.5 Fascia panel ............................................................................................. 8-54
8.15.8.6 Oxygen and air pressure switches (PS1 and PS2) .....................................8-56
8.15.8.7 PSOL cartridge ........................................................................................8-57
8.15.8.8 Removing inspiratory module..................................................................8-58
8.15.8.9 Leak testing inspiratory module ............................................................... 8-58
8.15.8.10Installing inspiratory module ................................................................... 8-59
8.15.8.11Oxygen sensor (OS) ................................................................................8-59
8.15.8.12Inspiratory check valve (CV3) ..................................................................8-62
8.15.8.13Right-side plate ....................................................................................... 8-62
8.15.8.14Inspiratory electronics PCB ...................................................................... 8-64
8.15.8.15Left-side plate..........................................................................................8-64
8.15.8.16PSOL manifold ........................................................................................ 8-65
8.15.8.17Oxygen and air flow sensors (Q1 and Q2) ............................................... 8-67
8.15.8.18Safety valve ............................................................................................. 8-68
8.15.8.19Check valve assembly, regulator assembly, and flow sensor manifold ......8-70
8.15.8.20Oxygen and air regulators (REG1 and REG2) ........................................... 8-72
8.15.8.21Inspiratory pressure transducer autozero solenoid (SOL1) ........................8-73
8.15.8.22Inspiratory floor assembly ........................................................................ 8-73
8.15.9 Exhalation module........................................................................................... 8-74
8.15.9.1 Exhalation collector vial (ECV) and expiratory filter (F9) ........................... 8-74
8.15.9.2 Removing exhalation module cover ......................................................... 8-74
8.15.9.3 Removing exhalation module .................................................................. 8-76
8.15.9.4 Installing exhalation module....................................................................8-76
8.15.9.5 Exhalation valve (EV) ............................................................................... 8-77
8.15.9.6 Exhalation flow sensor (Q3) ..................................................................... 8-80
8.15.9.7 Exhalation transducer PCB....................................................................... 8-81
8.15.9.8 Expiratory pressure transducer autozero solenoid (SOL2) ......................... 8-82
8.15.9.9 Exhalation heater (EXH HTR) and check valve (CV5) ................................ 8-83
8.15.10 BDU housing ................................................................................................. 8-84
8.15.10.1Removing BDU housing........................................................................... 8-84
8.15.10.2Installing BDU housing ............................................................................ 8-86
8.15.10.3M otherboard PCB ...................................................................................8-86
8.15.10.4BDU alarm assembly................................................................................ 8-87
840 Ventilator System Service Manual 4-070496-00 Rev. A (08/ 03) xiii
Cont ent s
8.15.11 Power indicator............................................................................................. 8-88
8.15.11.1Removing power indicator ...................................................................... 8-88
8.15.11.2Installing power indicator........................................................................ 8-89
8.15.11.3Alarm blindmate cable............................................................................ 8-89
8.15.11.4Inspiratory blindmate cable..................................................................... 8-90
8.15.11.5dc power supply blindmate cable ........................................................... 8-91
8.15.11.6ac power supply blindmate harness ........................................................ 8-92
8.15.12 Release handle .............................................................................................. 8-94
8.15.12.1Removing release handle ........................................................................ 8-94
8.15.12.2Installing release handle.......................................................................... 8-94
8.16 806 compressor unit ............................................................................................... 8-94
8.17 Servicing the 806 compressor ................................................................................. 8-95
8.17.1 Compressor inlet filter..................................................................................... 8-95
8.17.2 Removing and installing the compressor inlet filter ......................................... 8-95
8.17.3 Removing compressor from cart ..................................................................... 8-95
8.17.3.1 Disconnecting compressor from BDU...................................................... 8-95
8.17.3.2 Removing compressor from cart.............................................................. 8-96
8.17.4 Removing top cover........................................................................................ 8-96
8.17.5 Installing top cover ......................................................................................... 8-97
8.17.6 Cooling fans.................................................................................................... 8-97
8.17.6.1 Removing a cooling fan........................................................................... 8-97
8.17.6.2 Installing a fan......................................................................................... 8-98
8.17.7 Removing the back panel................................................................................ 8-98
8.17.8 Reinstalling the back panel.............................................................................. 8-98
8.17.9 Replacing an accumulator fitting and O-ring ................................................... 8-99
8.17.10 Pneumatic hoses........................................................................................... 8-99
8.17.11 Plenum assembly ........................................................................................ 8-100
8.17.11.1Removing the plenum assembly............................................................ 8-101
8.17.12 Heat exchanger (HE) ................................................................................... 8-102
8.17.12.1Replacing the heat exchanger ............................................................... 8-102
8.17.13 Removing and replacing the ac power cord ................................................ 8-102
8.17.14 Removing and replacing the data cable ...................................................... 8-102
8.17.15 Compressor PCB......................................................................................... 8-103
8.17.15.1Removing and replacing the compressor PCB....................................... 8-103
8.17.16 Air dryer (dryer) and solenoid valve assembly.............................................. 8-103
8.17.16.1Removing and reinstalling the air dryer
and solenoid valve assembly and replacing filters8-104
8.17.17 Compressor assembly ................................................................................. 8-105
8.17.17.1Removing the compressor assembly...................................................... 8-105
8.17.17.2Replacing the coalescing filter element ................................................. 8-106
8.17.18 Replacing the compressor panels................................................................ 8-107
8.17.19 Reinstalling the plenum assembly................................................................ 8-108
8.17.20 Reconnecting electrical cables.................................................................... 8-108
8.17.21 Replacing the Tinnerman clips .................................................................... 8-108
8.17.22 Reinstalling the back panel/accumulator assembly ...................................... 8-108
8.17.23 Replacing the main inlet filter and reinstalling the top................................. 8-109
8.17.24 Reinstalling the compressor module............................................................ 8-109
8.17.25 Running performance verification test ......................................................... 8-109
xiv 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Cont ent s
8.18 Backup power source (BPS) ................................................................................... 8-110
8.18.1 Removing BPS............................................................................................... 8-110
8.18.2 Installing BPS................................................................................................. 8-111
8.18.3 Battery pack ..................................................................................................8-111
8.18.3.1 Removing battery pack..........................................................................8-111
8.18.3.2 Installing battery pack ...........................................................................8-112
8.18.4 BPS PCB ........................................................................................................ 8-113
8.18.4.1 Removing BPS PCB................................................................................ 8-113
8.18.4.2 Installing BPS PCB.................................................................................. 8-113
8.19 Cart .......................................................................................................................8-113
8.19.1 Casters ..........................................................................................................8-113
8.19.1.1 Removing casters...................................................................................8-113
8.19.1.2 Installing casters....................................................................................8-114
8.19.2 Removing/ installing GUI mount ....................................................................8-114
8.19.3 Removing/ installing flex arm inserts ..............................................................8-115
9 Part s list
9.1 How to use this parts list.............................................................................................. 9-1
9.2 840 Ventilator System patient system and accessories ................................................ 9-3
9.2.1 840 Ventilator System NeoMode patient system and accessories ........................ 9-6
9.3 Flex arm assembly ..................................................................................................... 9-8
9.3.1 Oxygen hose assemblies ................................................................................... 9-10
9.3.2 Air hose assemblies ........................................................................................... 9-12
9.3.3 Power cords ..................................................................................................... 9-14
9.4 Ventilator major assemblies ......................................................................................9-16
9.4.1 Label kits .......................................................................................................... 9-18
9.4.2 10.4-inch GUI (graphic user interface) assembly ...............................................9-26
9.4.3 10.4-inch GUI (graphic user interface) handle .................................................. 9-28
9.4.4 10.4-inch graphic user interface (GUI) rotor assembly ...................................... 9-30
9.4.5 10.4-inch GUI keyboards and speaker assembly ................................................9-32
9.4.6 9.4-inch graphic user interface (GUI) ................................................................ 9-35
9.4.6.19.4-inch GUI front housing assembly ........................................................... 9-38
9.4.6.29.4-inch GUI rear housing assembly ............................................................ 9-40
9.4.7 Breath delivery unit (BDU) ................................................................................ 9-42
9.4.7.1Exhalation module....................................................................................... 9-46
9.4.7.2Inspiratory module ...................................................................................... 9-49
9.4.7.3BDU cover ...................................................................................................9-58
9.4.7.4BDU chassis assembly ..................................................................................9-60
9.4.7.5BDU chassis kit ............................................................................................9-62
9.4.8 806 compressor unit ........................................................................................9-64
9.4.8.1806 compressor base assembly.................................................................... 9-66
9.4.8.2806 compressor unit enclosure assembly ..................................................... 9-68
9.4.8.3806 compressor unit plenum assembly........................................................9-70
9.4.9 Backup power source (BPS) .............................................................................. 9-72
9.4.10 Cart assembly ................................................................................................. 9-74
xvi 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
SECTION
FIGURES
840 Ventilatory System Service Manual 4-070496-00 Rev. A (08/ 03) xvii
Figures
Figure 2-32. Exhalation module components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-36
Figure 2-33. 806 compressor on cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-38
Figure 2-34. 806 Compressor Pneumatic diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-38
Figure 2-35. 806 components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-39
Figure 2-36. Heat exchanger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-40
Figure 2-37. 806 water trap assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-40
Figure 2-38. Air dryer assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-40
Figure 2-39. 806 back panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-41
Figure 2-40. 806 cooling fans . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-41
Figure 2-41. Electrical system block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-49
Figure 2-42. ac panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-51
Figure 2-43. 840 Ventilator System interconnect diagram – ac panel . . . . . . . . . . . . . . . . . . . . . . . . . 2-52
Figure 2-44. Power switch (S1) and indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-53
Figure 2-45. Power supply assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-54
Figure 2-46. 840 Ventilator System interconnect diagram – Power distribution . . . . . . . . . . . . . . . . . 2-55
Figure 2-47. BPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-56
Figure 2-48. BPS battery pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-57
Figure 2-49. BPS PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-57
Figure 2-50. Card cage with all PCBs installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-58
Figure 2-51. 840 Ventilator System interconnect diagram – Card cage . . . . . . . . . . . . . . . . . . . . . . . . 2-59
Figure 2-52. Motherboard PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-60
Figure 2-53. Motherboard PCB in place. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-61
Figure 2-54. Motherboard PCB block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-62
Figure 2-55. BD CPU PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-64
Figure 2-56. AI PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-67
Figure 2-57. Data key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-69
Figure 2-58. 10.4” GUI CPU PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-70
Figure 2-59. 9.4” GUI CPU PCB and backlight inverter PCB in place . . . . . . . . . . . . . . . . . . . . . . . . . . 2-70
Figure 2-60. 840 Ventilator System interconnect diagram – GUI 10.4-inch LCD panels . . . . . . . . . . . . 2-72
Figure 2-61. 840 Ventilator System interconnect diagram – GUI 9.4-inch LCD panels . . . . . . . . . . . . . 2-74
Figure 2-62. Touch Frame PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-75
Figure 2-63. Keyboard assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-77
Figure 2-64. GUI LED PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-78
Figure 2-65. 10.4” GUI LCD panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-79
Figure 2-66. 9.4-inch LCD panels and backlight tubes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-80
Figure 2-67. GUI alarm assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-80
Figure 2-68. BDU LED PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-81
Figure 2-69. Inspiratory electronics PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-81
Figure 2-70. 840 Ventilator System interconnect diagram – Inspiratory module . . . . . . . . . . . . . . . . . 2-82
Figure 2-71. Exhalation transducer PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-83
Figure 2-72. 840 Ventilator System interconnect diagram – Exhalation module . . . . . . . . . . . . . . . . . 2-84
Figure 2-73. BD alarm assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-85
Figure 2-74. 806 compressor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-85
Figure 2-75. 806 Compressor fans. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-86
Figure 2-76. 806 compressor PCBA installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-86
Figure 2-77. 806 compressor PCB block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-88
Figure 2-78. 840 Ventilator System interconnect diagram – Compressor unit . . . . . . . . . . . . . . . . . . . 2-89
Figure 2-79. Compressor operational sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-91
Figure 2-80. Compressor unit start-up sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-92
xviii 4-070496-00 Rev. A (08/ 03) 840 Ventilatory System Service Manual
Figures
Figure 2-81. Inspiration gas flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-93
Figure 2-82. Exhalation gas flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-96
Figure 2-83. Pressure transducer autozero mode gas flow diagram . . . . . . . . . . . . . . . . . . . . . . . . . . 2-99
Figure 2-84. Power loss sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-100
Figure 2-85. Safety valve open diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-102
Figure 2-86. Pressure release, patient circuit occluded diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-104
Figure 3-1. Patient circuit setup for SST. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Figure 3-2. EST setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Figure 3-3. EST screens during testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
840 Ventilatory System Service Manual 4-070496-00 Rev. A (08/ 03) xix
Figures
Figure 8-30. LCD panel and backlight panel assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-37
Figure 8-31. Replacing the 9.4-inch GUI cooling vent filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-38
Figure 8-32. 9.4-inch rotor housing assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-39
Figure 8-33. BDU. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-40
Figure 8-34. BDU connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-43
Figure 8-35. Sure-Lock™ retainer and power cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-44
Figure 8-36. BDU release handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-44
Figure 8-37. BD CPU PCB and AI PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-45
Figure 8-38. BD card cage and PCBs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-46
Figure 8-39. BDU I/ O panel connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-47
Figure 8-40. Replacing power supply assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-48
Figure 8-41. Replacing power switch (S1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-49
Figure 8-42. Replacing humidifier receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-50
Figure 8-43. ac panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-51
Figure 8-44. Adding ferrites to air and oxygen transducer harnesses . . . . . . . . . . . . . . . . . . . . . . . . . . 8-53
Figure 8-45. Replacing F2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-54
Figure 8-46. Removing inspiratory module fascia panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-54
Figure 8-47. Oxygen filter assembly and pressure switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-55
Figure 8-48. Pressure switches. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-56
Figure 8-49. Replacing PSOL cartridge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-57
Figure 8-50. Removing inspiratory module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-58
Figure 8-51. Opening oxygen sensor access port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-59
Figure 8-52. Detail of oxygen sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-60
Figure 8-53. Replacing oxygen sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-61
Figure 8-54. Inspiratory check valve (CV3) assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-62
Figure 8-55. Removing inspiratory module right-side plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-63
Figure 8-56. Replacing inspiratory module left-side plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-65
Figure 8-57. PSOL manifold ready for removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-66
Figure 8-58. Replacing oxygen and air flow sensors (Q1 and Q2) and safety valve. . . . . . . . . . . . . . . . 8-67
Figure 8-59. Oxygen and air flow sensors (Q1 and Q2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-68
Figure 8-60. Safety valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-69
Figure 8-61. Disassembling flow sensor manifold/regulator assembly/check valve assembly . . . . . . . . 8-71
Figure 8-62. Removing REG1 and REG2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-72
Figure 8-63. Removing exhalation module cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-74
Figure 8-64. Exhalation module disassembled . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-75
Figure 8-65. Replacing exhalation module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-76
Figure 8-66. Exhalation sample port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-77
Figure 8-67. Replacing exhalation module components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-78
Figure 8-68. Exhalation flow sensor (Q3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-81
Figure 8-69. Replacing exhalation transducer PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-82
Figure 8-70. Replacing exhalation heater (EXH HTR) and check valve (CV5) . . . . . . . . . . . . . . . . . . . . 8-83
Figure 8-71. Removing BDU housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-85
Figure 8-72. Removing motherboard PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-87
Figure 8-73. BDU alarm and motherboard PCB cable connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-88
Figure 8-74. Replacing power indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-89
Figure 8-75. Cable connections to ac panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-90
Figure 8-76. Replacing inspiratory blindmate cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-91
Figure 8-77. Replacing dc power supply blindmate cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-92
Figure 8-78. Replacing ac power supply blindmate harness. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-93
Figure 9-1. 840 Ventilator System patient system and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Figure 9-2. NeoMode patient circuit and accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Figure 9-3. Flex arm assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
Figure 9-4. Oxygen hose assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11
Figure 9-5. Air hose assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
Figure 9-6. Power cords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-15
Figure 9-7. Ventilator major assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-17
Figure 9-8. Label kits (Sheet 1 of 2). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-23
Figure 9-9. 10.4-inch GUI assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-27
Figure 9-10. 10.4-inch GUI handle assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-29
Figure 9-11. 10.4-inch GUI rotor assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-31
Figure 9-12. 10.4-inch GUI keyboard and speaker assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-33
Figure 9-13. 9.4-inch GUI assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-37
Figure 9-14. 9.4-inch GUI front housing assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-39
Figure 9-15. 9.4-inch GUI rear housing assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-41
Figure 9-16. Breath delivery unit (BDU) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-45
Figure 9-17. Exhalation module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-47
Figure 9-18. Inspiratory module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-51
Figure 9-19. Proportional solenoid (PSOL) valve assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-53
Figure 9-20. Inspiratory module floor assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-57
Figure 9-21. BDU cover assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-59
Figure 9-22. BDU chassis assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-61
Figure 9-23. BDU chassis kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-63
Figure 9-24. 806 compressor unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-65
Figure 9-25. 806 compressor base assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-67
Figure 9-26. Compressor unit enclosure assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-69
Figure 9-27. Compressor plenum assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-71
Figure 9-28. Backup power source (BPS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-73
Figure 9-29. Cart assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-75
840 Ventilatory System Service Manual 4-070496-00 Rev. A (08/ 03) xxi
Figures
xxii 4-070496-00 Rev. A (08/ 03) 840 Ventilatory System Service Manual
SECTION
TABLES
840 Ventilatory System Service Manual 4-070496-00 Rev. A (08/ 03) xxiii
Tables
Table 4-9. Cal Info Duplication errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Table 4-10. Serial number setup errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Table 4-11. Datakey update error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Table 4-12. Serial Loopback Test error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
xxiv 4-070496-00 Rev. A (08/ 03) 840 Ventilatory System Service M anual
SECTION
S EC T I O N
General information 1 1
This section provides i ntroductory in form ati on on the Puri tan Ben nett 840 Ven til ator
System. In cluded are a descripti on of th e venti lator, i ncl uding specifi cati ons, requi red tools
and test equipm ent, schedul e of m ai ntenance, and control s and in dicators.
840 Ventilator System Service Manual 4-070496-00 Rev. A (08/ 03) 1-1
1 General information
The BPS (Backup Power Supply) provi des dc power to th e BDU in the event th at ac power is
l ost. A n ew, full y charged BPS all ows the ventil ator to function (wi thout compressor or
h umi di fi er) for at l east 30 m in utes; thus, th e BPS can power th e venti lator for transport
purposes wi th in the respi ratory care faci li ty. The BPS sh ould al ways be conn ected to the
venti lator duri ng operation.
The optional com pressor uni t provides com pressed ai r to the BDU, and can be used in place
of wall or bottl ed air for norm al operati on. The compressor uni t is powered by an d
com m un icates wi th the BDU.
NOTE:
An external air source is required when performing service mode calibrations and
performance verification testing.
1.4 Accessories
The followi ng accessories are eith er required or can be used with th e venti lator.
Pa tien t c ircu it : A variety of reusable Puritan Bennett patient ci rcui ts, adult an d pedi atric,
with and wi thout w ater traps, an d wi th and without h eated wi re, i s avail abl e. Consult the
“ Part num bers” appendi x of th e 840 Ventilator System Operator’s and Technical ReferenceManual
for pati ent circui t orderi ng in form ation.
H u m id ifica t ion d ev ic e: Th e 840 Venti lator System supports th e use of an optional
h um idi fication devi ce, in cludin g a heated humi difier, h eat and moisture exchan ger (HM E),
or heated wire. A m oun ti ng bracket i s avail abl e for the Fi sher & Paykel M 480/ M 730
Hum i difier. An ac socket for the humi di fier is avail abl e on 100 and 120 V ventil ators.
Rem o t e n u rse ca ll u n it: An anal og output con nector permi ts connection to a rem ote nurse
call uni t. (Puri tan Ben nett does not supply nurse’s call un its or cabl es.) Refer to th e 840
Ventilator System Operator’s and Technical ReferenceManual for nurse cal l specifi cations.
1-2 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
1.5 Specificat ions
Dimensions BDU: 330 mm high x 457 mm wide x 254 mm deep (13 in. high x 18 in. wide x 10 in. deep)
GUI: 460 mm high x 394 mm wide x 170 mm deep (18.1 in. high x 15.5 in. wide x 6.7 in. deep)
BPS: 83 mm high x 244 mm wide x 254 mm deep (3.25 in. high x 9.6 in. wide x 10 in. deep)
Cart: 998 mm high x 582 mm wide x 602 mm deep (39.3 in. high x 22.9 in. wide x 23.7 in. deep)
804 Compressor: 417 mm high x 458 mm wide x 362 mm deep (16.4 in. high x 18 in. wide x 14.25
in. deep)
806 Compressor: 425 mm high x 458 mm wide x 362 mm deep (17 in. high x 18 in. wide x 14.25
in. deep)
Oxygen and air inlet Pressure: 241 to 690 kPa (35 to 100 psi)
supplies
Warning
Due to excessive restriction of the Air Liquide, SIS, and Dräger hose assemblies, reduced
ventilator performance levels may result when oxygen or air supply pressures < 50 psi
(345 kPa) are employed.
Oxygen sensor life The oxygen sensor should be replaced two years after date of manufacture, or as often as necessary.
Actual sensor life depends on operating environment; operation at higher temperature or O2% levels
will shorten sensor life.
Gas mixing system Range of flow from the mixing system: Can be set to 150 L/ min standard temperature and pressure,
dry (STPD). Additional flow is available (up to 80 L/ min for pediatric patients whose IBW ≤ 24 kg,
and up to 200 L/ min for adults whose IBW > 24 kg) for compliance compensation.
Leakage from one gas system to another: Meets standard EN 60601-2-12.
Operating pressure range: 35 to 100 psi (241 to 690 kPa)
Air/ oxygen regulator bleed: Up to 3 L/ min
NOTE:
Above values obtained using the following ventilator settings at 22 °C ambient temperature: mode, A/ C; mandatory
type, PC; IBW, 85 kg; fTOT, 20/ min; PSUPP, 30 cmH2O; TI, 1 s; Rise Time Percent (was Flow Acceleration), 50%; O2%,
50%; PMEAN, 50 cmH2O; PSENS, 3 cmH2O. Input power specifications are for ventilators with Fisher & Paykel M R730
humidifiers. (Humidifier connection only available on 100 – 120 V ventilators.)
Warning
In the event of a defective earth conductor, connecting equipment to the auxiliary mains
socket outlet(s) (that is, the humidifier or compressor connections) may increase patient
leakage current to values that exceed the allowable limits.
1-4 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Table 1-1: Ventilat or specificat ions (cont inued)
NOTE:
BPS battery life specifications are approximate. To ensure maximum battery life, maintain full
charge and minimize the number of complete discharges.
Communications Rem ot e alarm (nurse’s call) port (Figure 1-1). Allows medium- and high-urgency alarm conditions
capabilities to be annunciated at locations away from the ventilator (for example, when the ventilator is in an
isolation room). The ventilator signals an alarm using a normally open or a normally closed signal.
The ventilator asserts a remote alarm when there is an active medium- or high-urgency alarm
condition, unless the alarm silence function is active. The remote alarm port is a 4-pin female
connector. Allowable current is 500 mA at 30 V dc (maximum).
4 1
3 2
8-00020
Pin Signal
2 Relay common
4 Not connected
Communication RS-232 (serial) port (Figure 1-2). A 9-pin male connector configured as data terminal equipment
capabilities (cont) (DTE). Allowable current is 0.2 A at 10 V dc (maximum).
1 2 3 4 5
8-00019
6 7 8 9
Figure 1-2. 840 Vent ilat or System RS-232 serial port pinout
Pin Signal
1 Not connected
2 Receive data (RxD)
3 Transmit data (TxD
4 Data terminal ready (DTR), terminated high
5 Ground (GND)
6 Not connected
7 Request to send (RTS)
8 Clear to send (CTS)
9 Not connected
NOTE:
The connection of accessories or equipment to the ventilator’s RS-232 and remote alarm ports
requires electrical separation of conductive earth (ground) connections between different items of
equipment of a system. The system cable assembly shielding can interconnect the ventilator and
other equipment, increasing the risk of excessive enclosure leakage current from external
equipment. For safe connection and prevent a double earth ground, the shield conductor must have
separation of 4 mm between the equipment and ventilator conductors, as shown below:
Minimum shield
To 840 Ventilator RS-232 and Cable conductive separation
remote alarm input connectors shielding 4 mm
1-6 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
1.6 Compliance and approvals
The 840 Ventil ator System was developed in accordance with perti nent FDA guidances and
North Am erican an d EN 46001/ISO 13485 i nternati onal standards (Table 1-2). Th e
m anufacturin g faci li ty for this product i s EN 46001 certifi ed.
The ven til ator’s IEC 60601-1/EN 60601-1 classi ficati on i s Protecti on cl ass I, Type B, internal ly
powered, IPX 1 drip-proof equi pm en t, continuous operation.
St andards/ cert ificat ions Configurat ions Cert ificat ion agency
Nor t h Am erica
Eur opean
Approved to the type test requirements of 220 – 240 V, 50 Hz TÜV Product Service
Annex III of the Medical Device Directive. 220 – 240 V, 60 Hz
EN 60601-1
EN 60601-1-1
EN 60601-1-2
EN 794-1
EN 60601-1-4 --
NOTE:
When pressure units are set to hPa, pressure delivery and spirometry are subject to an additional 2% error.
M aximum working pressure 100 cmH2O (102 hPa), ensured by high pressure limit
90 cmH2O (pressure-based ventilation)
Volume measurements:
Type: Hot film anemometer.
Sensing position: Exhalation compartment
M easurements: Exhaled tidal volume (range: 0 to 6,000 mL); total minute volume
(range: 0 to 99.9 L)
Results of ventilator patient Inspiratory pressure drop from inlet of open safety valve to outlet port without
circuit testing (using circuits inspiratory filter:
identified for use with 840 At 5 standard liters per minute (SL/ min): 0.06 cmH2O
Ventilator) At 30 SL/ min: 0.28 cmH2O
At 60 SL/ min: 0.95 cmH2O
Inspiratory pressure drop across inspiratory filter:
At 5 SL/ min: 0.17 cmH2O
At 30 SL/ min: 0.56 cmH2O
At 60 SL/ min: 1.37 cmH2O
Inspiratory pressure drop from inlet of open safety valve with inspiratory filter:
At 5 SL/ min: 0.17 cmH2O
At 30 SL/ min: 0.84 cmH2O
At 60 SL/ min: 2.32 cmH2O
Pressure drop across 1.68 m (5.5 ft) inspiratory or expiratory limb with water trap, to
patient wye:
Neonatal patient circuit N/ A (no water trap)
Pediatric patient circuit at 30 SL/ min: 0.73 cmH2O
Adult patient circuit at 60 SL/ min: 1.05 cmH2O
1-8 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Table 1-3: Technical inform at ion (continued)
Results of ventilator patient Pressure drop across 1.22 m (4 ft) inspiratory or expiratory limb without water trap, to
circuit testing (using circuits patient wye:
identified for use with 840 Neonatal patient circuit at 5 SL/ min: 0.45 cmH2O (inspiratory limb)
Ventilator) (continued)
Neonatal patient circuit at 5 SL/ min: 0.40 cmH2O (expiratory limb)
Pediatric patient circuit at 30 SL/ min: 0.56 cmH2O
Adult patient circuit at 60 SL/ min: 0.70 cmH2O
Pressure drop across Fisher & Paykel humidifier and lead-in tube:
Neonatal patient circuit at 5 SL/ min: 0.14 cmH2O
Pediatric patient circuit at 30 SL/ min: 0.28 cmH2O
Adult patient circuit at 60 SL/ min: 0.93 cmH2O
Internal volume:
Inspiratory pneumatics: 50 mL ±5 mL
Expiratory pneumatics: 1000 mL ±25 mL (including expiratory filter and collector vial)
The 840 Ventilator automatically adjusts for volume losses due to gas compressibility
(that is, automatic compliance compensation), subject to a maximum delivered volume
of 2500 mL.
NOTE:
• Patient circuit testing specifications are with the ventilator powered off, and are based on the
recommended configurations shown in the 840 Ventilator System Operator’s and Technical Reference
Manual.
• To ensure that compliance compensation functions correctly, the user must run SST with the circuit
configured as intended for use on the patient.
Bacteria filter efficiency 99.97% for nominal particle size of 0.3 µm (micron) at 100 L/ min
1.8 Range, resolut ion, accuracy, and new pat ient / default set t ings
Ranges, resol utions, accuracies, and new pati en t defaults for ven til ator setti ngs, alarm
settin gs, and patient data are listed in Tabl e 1-4.
Table 1-4: Ventilat or range, resolut ion, accuracy, new pat ient/ default s
Set t ing Range, resolut ion, accuracy, new pat ient / default
1-10 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Table 1-4: Ventilator range, resolut ion, accuracy, new pat ient / default s (continued)
Set t ing Range, resolut ion, accuracy, new pat ient / default
Apnea peak inspiratory Range: 1.0 to 30 L/ min when patient circuit is neonate
flow (VMAX) 3.0 to 60 L/ min when patient circuit is pediatric
3.0 to 150 L/ min when patient circuit is adult
Resolution: 0.1 L/ min for peak flows from 3 to 20 L/ min
1 L/ min for peak flows above 20 L/ min
Accuracy: ± (0.5 + 10% of setting) L/ min of the flow command input to the flow controller,
at the end of each control interval, after the first 100 milliseconds of inspiration.
New patient: Maximum of 1.0 or (0.87 x IBW) L/ min with NEONATAL patient circuit
M aximum of 3.0 or (0.572 x IBW) L/ min with PEDIATRIC patient circuit
M aximum of 3.0 or (0.435 x IBW) L/ min with ADULT patient circuit
Constant during rate change Range: Inspiratory time, I:E ratio, or expiratory time
Resolution: Not applicable
Accuracy: Not applicable
New patient: Inspiratory time
Table 1-4: Vent ilator range, resolut ion, accuracy, new pat ient/ defaults (cont inued)
Set t ing Range, resolut ion, accuracy, new pat ient / default
Flow sensitivity (VSENS) Range: 0.1 to 10.0 L/ min with NEONATAL patient circuit
0.2 to 20.0 L/ min with PEDIATRIC or ADULT patient circuit
Resolution: 0.1 L/ min
Accuracy: Not applicable
New patient: 1.0 L/ min with NEONATAL patient circuit
2.0 L/ min with PEDIATRIC patient circuit
3.0 L/ min with ADULT patient circuit
Humidification type Range: HME, non-heated expiratory tube, or heated expiratory tube
Resolution: Not applicable
Accuracy: Not applicable
Default: non-heated expiratory tube
Ideal body weight (IBW) Range: 0.5 kg (1.1 lb) to 7 kg (15.4 lb) with NEONATAL patient circuit
3.5 kg (7.7 lb) to 35 kg (77 lb) with PEDIATRIC patient circuit
7.0 kg (15.4 lb) to 150 kg (330 lb) with ADULT patient circuit
Resolution: 0.1 kg for 0.5 kg to 3.5 kg
0.5 kg for 4.0 to 9.5 kg
1 kg for 10 to 50 kg
5 kg for 50 to 100 kg
10 for 100 to 150 kg
Accuracy: Not applicable
New patient: 3.0 kg with NEONATAL patient circuit
15 kg with PEDIATRIC patient circuit
50 kg with ADULT patient circuit
1-12 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Table 1-4: Ventilator range, resolut ion, accuracy, new pat ient / default s (continued)
Set t ing Range, resolut ion, accuracy, new pat ient / default
Patient circuit type Range: Neonatal (if Neo-mode option is active), Pediatric, or Adult
Resolution: Not applicable
Accuracy: Not applicable
Peak inspiratory flow (VMAX) Range: 1.0 to 30 L/ min with NEONATAL patient circuit
3.0 to 60 L/ min with PEDIATRIC patient circuit
3 to 150 L/ min with ADULT patient circuit
Resolution: 0.1 L/ min for flows of 3 to 20 L/ min; 1 L/ min for flows above 20 L/ min
Accuracy: ± (0.5 + 10% of setting) L/ min of the flow command input to the flow controller,
at the end of each control interval, after the first 100 milliseconds of inspiration.
Table 1-4: Vent ilator range, resolut ion, accuracy, new pat ient/ defaults (cont inued)
Set t ing Range, resolut ion, accuracy, new pat ient / default
Safety ventilation Settings are identical to new patient values, except: mode = A/ C, mandatory type = PC,
respiratory rate = 16/ min, inspiratory time = 1 s, inspiratory pressure = 10 cmH2O, PEEP =
3 cmH2O, trigger type = pressure, O2% = 100% (21% if O2 not available), patient circuit
type = last set value or ADULT if none available, humidification type = last set value or
NON-HEATED EXP TUBE if none available, humidifier volume = last set value or 480 mL if
none available.
Alarm settings in safety ventilation: high circuit pressure = 20 cmH2O, high exhaled minute
volume = OFF, high exhaled tidal volume = OFF, high respiratory rate = OFF, low exhaled
mandatory tidal volume = OFF, low exhaled minute volume = 0.05 L, low exhaled
spontaneous tidal volume = OFF
1-14 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Table 1-4: Ventilator range, resolut ion, accuracy, new pat ient / default s (continued)
Set t ing Range, resolut ion, accuracy, new pat ient / default
Setting limits for volume Tidal volume: 25 mL ≤ VT ≤ 2500 mL; 1.16 mL/ kg ≤ VT ≤ 45.7 mL/ kg (default 7.25 mL/ kg)
control (VC) mandatory Inspiratory time: 0.2 s ≤ TI ≤ 8 s
breaths Expiratory time: 0.2 s ≤ TE ≤ 59.8 s
I:E ratio: 1:299 ≤ I:E ≤ 1:4.00
Flow (at 1/ min ≤ f ≤ 100/ min):
3 L/ min ≤ V ≤ 60 L/ min for IBW ≤ 24 kg
150 L/ min ≤ V for IBW > 24 kg
M inute volume (using square flow pattern, I:E = 1:1, and f ≥ 30/ min):
30 L/ min ≤ VE for IBW < 24 kg
30 L/ min < VE < 75 L/ min for IBW 24 to 54 kg
M aximum VE = 75 L/ min for IBW 55 to 150 kg
TI is a function of VT, flow pattern, TPL, and VMAX
TE is a function of VT, flow pattern, TPL, and f
I:E is the result of TI and TE
Any combination of settings for VT, VMAX, TPL, f, and flow pattern that violates these
boundaries is rejected. Refer to the Technical Reference section of the 840 Ventilator System
Operator’s and Technical Reference Manual for more details.
Table 1-4: Vent ilator range, resolut ion, accuracy, new pat ient/ defaults (cont inued)
Set t ing Range, resolut ion, accuracy, new pat ient / default
Setting limits for pressure Inspiratory pressure: PI = 5 to 90 cmH2O; PI + PEEP ≤ 90 cmH2O; PI + PEEP
control (PC) mandatory + 2 cmH2O ≤ 1PMEAN
breaths Inspiratory time: 0.2 s ≤ TI ≤ 8 s
Expiratory time: 0.2 s ≤ TE ≤ 59.8 s
I:E ratio: 1:299 ≤ I:E ≤ 1:4.00
Respiratory rate: 1/ min ≤ f ≤ 100/ min
High circuit pressure limit: 7 cmH2O ≤ ↑ PMEAN ≤ 100 cmH2O
TI is a function of f (for I:E or TE constant during rate change) and TE. TE is a function of f
(for I:E or TI constant during rate change) and TI. I:E is a function of f (for TI or TE constant
during rate change), TI, and TE.
Any combination of settings for PI, PEEP, 1PMEAN, f, TI, I:E, or TE that violates these
boundaries is rejected. Refer to the Technical Reference section of the 840 Ventilator System
Operator’s and Technical Reference Manual for more details.
Setting limits when selected Support pressure: PSUPP = 0 to 70 cmH2O; PSUPP + PEEP ≤ 90 cmH2O
support type is pressure PEEP: PEEP = 0 to 45 cmH2O; PEEP + 7 cmH2O ≤ 1PMEAN
support (PS) High circuit pressure limit: PSUPP + PEEP + 2 cmH2O ≤ 1PMEAN
Any combination of settings for PSUPP, PEEP, or 1PMEAN that violates the above boundaries
is rejected. Refer to the Technical Reference section of part of the 840 Ventilator System
Operator’s and Technical Reference Manual for more details.
1-16 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Table 1-4: Ventilator range, resolut ion, accuracy, new pat ient / default s (continued)
Set t ing Range, resolut ion, accuracy, new pat ient / default
Low exhaled minute volume Range: < high exhaled minute volume limit
limit (4VE TOT) OFF (immediately following the soft limit of 0.01 L/ min) to 10 L/ min with
NEONATAL patient circuit
0.05 L/ min ≤ 4VE TOT ≤ 30 L/ min with PEDIATRIC patient circuit
0.05 L/ min ≤ 4VE TOT ≤ 60 L/ min with ADULT patient circuit
Resolution: 0.005 L for 0.01 to 0.50 L; 0.05 L for 0.50 to 5.0 L; 0.50 L for 5.0 to 60.0 L
New patient: maximum of 0.01 or ((20 x 7.25 x IBW x 0.70/ 1000) - 0.05) with NEONATAL
patient circuit
((14 x 7.25 x IBW x 0.70/ 1000) - 0.05) with PEDIATRIC patient circuit
((10 x 7.25 x IBW x 0.70/ 1000) - 0.05) with ADULT patient circuit
End expiratory pressure New symbol for end expiratory pressure (monitored) (was PEEND)
(PEEP) Range: -20.0 to 130 cmH2O
Resolution: 0.1 cmH2O for -20.0 to 9.9 cmH2O; 1.0 cmH2O for 10 to 130 cmH2O
Accuracy: ± (2 + 4% of reading) cmH2O (relative to pressure measured at the exhalation
side of the patient wye when the end-expiratory flow is less than 60 lpm)
Table 1-4: Vent ilator range, resolut ion, accuracy, new pat ient/ defaults (cont inued)
Set t ing Range, resolut ion, accuracy, new pat ient / default
M ean airway pressure New symbol for mean airway pressure (was PCIRC).
(PMEAN) Range: -20.0 to 120 cmH2O
Resolution: 0.1 cmH2O for -20.0 to 9.9 cmH2O; 1 cmH2O for 10 to 120 cmH2O
Accuracy: ± (3 + 4% of reading) cmH2O relative to pressure measured at the exhalation
side of the patient wye.
1-18 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
1.9 Tools, equipment , and service mat erials
The tools, equi pm ent, and service m ateri al s listed in Tabl e 1-5 are used to servi ce th e 840
Ventil ator System . Refer to Secti on 5 for a l ist of required tool s, equipm ent, and servi ce
m aterials specific to perform an ce veri fi cati on.
Digital multimeter (DMM) accurate to Fluke Model 87 or equivalent Performance verification, general
3 decimal places, with test leads troubleshooting
Table 1-5: Tools, equipment , and service mat erials (cont inued)
Electrical safety analyzer capable of Dale M odel 600, Dale Technology Inc., Performance verification
measuring ground resistance and P.O. Box 196, 401 Claremont Ave.,
leakage current Thornwood, NY 10594 USA,
800.544.3253
http:/ / www.daletech.com
or equivalent
Flex tube, 21.0 in. (53.4 cm) (gold 4-018506-00 EST, performance verification
standard tube) (2 required)
1-20 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Table 1-5: Tools, equipment , and service materials (continued)
Pneumatic calibration analyzer or Puritan Bennett PTS2000 Performance Performance verification, regulator
equivalent devices capable of Test System (4-074686-00) (includes adjustment. Barometer used for
measuring oxygen percent, flow, BTPS accessory kit with serial cable) atmospheric pressure transducer
volume, pressure, and barometric calibration.
pressure. Oxygen analyzer connector
tee. Required accuracies: NOTE:
• Flow: 2.75% of reading ±0.05 slpm If you use a device other than
the PTS 2000 Performance Test
• Volume: 2% of reading or ±1 digit System, refer to the “ Manual
• Low pressure (-150 to +150 ventilator check” in Section 5 of
cmH2O): 0.75% of reading ±0.04 this manual.
cmH2O
• High pressure (0 to 150 psig): 1.0%
of reading ±0.1 psi
• Oxygen percentage: ±2% oxygen
• Barometric pressure: Range: 10 to
16 psia; resolution: 0.0 to 1 psia;
operating temperature: 10 to
40 °C; measurement accuracy:
±0.75 of reading; response: ≤ 100
ms.
Stoppers, no. 1, 5, and 5.5 Local supplier SST, EST, performance verification
(with center bore and luer fitting:
4-071856-00)
Stopper, no. 5.5 with center bore and 4-076467-00 Performance verification
Fitting, luer, bulkhead 4-012470-00
Table 1-5: Tools, equipment , and service mat erials (cont inued)
Table 1-6 lists th e peri odic m ain tenance activities requi red for the 840 Ven tilator System . See
the Venti lator Inform ati on screen for total h ours of operation for th e venti lator and
com pressor. For detai ls on patient system m ai ntenan ce, refer to the 840 Ventilator System
Operator’s and Technical Reference Manual.
1-22 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Several times a day or as Patient circuit: inspiratory and expiratory Check for water build-up, empty, and clean
required by your limbs as necessary.
institution’s policy
Inspiratory and expiratory bacteria filters Inspect and check resistance across
inspiratory and expiratory filters before every
use, after 15 days of continuous use in
expiratory limb, or if you suspect excess
resistance. SST checks the resistance of the
expiratory filter.
Collector vial, water traps, and drain bag Check and empty as needed.
Every 250 hours (or more Compressor inlet filter Either vacuum filter or wash filter in a warm
often, if required) detergent solution, rinse, and dry well.
Replace filter when it shows signs of wear.
Varies: Every year or 100 Reusable expiratory or expiratory bacteria Replace. Sterilize between patients and
autoclave cycles. Consult filters circuit changes, or according to your
product Directions for institution’s policy. Sterilize before
Use. nondestructive disposal.
Every 10,000 hours Various parts Use appropriate preventive maintenance kit
Every 15,000 hours (see Table 1-7).
NOTE:
If any part found in a preventive
maintenance kit requires replacement
before the recommended interval
elapses, consider installing the entire
kit anyway.
10,000 hours GUI and BDU 4-079056-00 Filter, air inlet (F2) with O-ring
Filter, oxygen (F1)
Filter, oxygen inlet (F3) with O-ring
Installation instructions
Label, preventive maintenance
Lamps, fluorescent backlight
(for 9.4“ color LCD displays)
Spring
1-24 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
16
15
12
14
13
8-00189
1 2 3 4 5 6 7 8 9 10 11
Figure 1-3. M onochrome GUI front view (show ing all keys)
Index
Labeling Funct ion
(Figure 1-3)
1 Screen lock key. When the yellow light on the screen lock key is lit, touching
the screen or off-screen controls (including the knob and ACCEPT key) has
no effect until you press the screen lock key again. New alarms (or when an
alarm’s urgency level escalates) automatically unlock the screen and controls.
8-00435
The screen lock allows you to clean the touch screen and prevents
inadvertent changes to settings and displays.
or
8-10001
Index
Labeling Funct ion
(Figure 1-3)
8-10002
8-10003
(US version only)
4 Alarm volume key. Allows you to adjust the alarm volume when you hold
down this key while turning the knob. You cannot turn off alarm volume.
8-00407
or
8-10004
5 Alarm silence key. Turns off alarm sound for 2 minutes. The yellow light on
the alarm silence key lights during the silence period, and turns off if you
press the alarm reset key or the 2-minute interval times out. A new, high-
urgency alarm cancels the silence. Alarms that can be silenced (lockable)
include all Patient-Data Alarms and Circuit Disconnect (see Table 7-2 for
8-00402 details).
Each time you press the alarm silence key, the silence period resets to 2
minutes. Each time you press the alarm silence key (whether or not there is
an active alarm), the keypress is recorded in the alarm log.
1-26 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Table 1-8: GUI front view (cont inued)
Index
Labeling Funct ion
(Figure 1-3)
8-00441
8-00411
or
8-10005
US version only
8 Delivers 100% oxygen (if available) for 2 minutes and calibrates the oxygen
sensor. The green light on this key lights to indicate that 100% O2 delivery is
active. Pressing this key again restarts the 2-minute delivery interval. You may
cancel the 2-minute interval by touching the CANCEL button on the GUI
touch screen.
8-00401
Oxygen sensor calibration can be tested using a procedure in the 840
Ventilator System Operator’s and Technical Reference Manual.
10 Allows you to measure auto-PEEP. Hold this key down until the maneuver
begins (at the next ventilator-initiated inspiration), and release when the
measurement is stable, up to 20 s. The measured values for intrinsic and total
PEEP are displayed at the end of the expiratory pause. Displays and freezes
the most recently selected graphics, allowing you to see when expiratory
8-00419 pressure stabilizes.
EXP PAUSE is not functional in SPONT. It has no effect during the inspiratory
phase of a breath. The pause is canceled and inspiration begins if you release
the EXP PAUSEkey, the patient triggers an inspiration, an alarm occurs, or the
maximum 20-s pause interval elapses. If flow triggering is active, backup
pressure sensitivity is used to detect patient effort. Only one expiratory pause
is allowed during a breath. Expiratory pause requests are ignored in apnea
ventilation, safety ventilation, occlusion status cycling (OSC), and idle mode.
During an expiratory pause, occlusion alarm detection is suspended. During
expiratory pause, the apnea interval is extended by the amount of time the
pause is active. In SIMV, the cycle during which the pause becomes active
(and the next scheduled VIM will occur) is extended by the amount of time
the pause is active. For purposes of I:E ratio calculation, expiratory pause is
considered part of the exhalation phase.
Index
Labeling Funct ion
(Figure 1-3)
8-00415
8-00406
14 Adjusts the value of a setting. A button that is highlighted means that the
knob is linked to that setting. Where applicable, turning the knob clockwise
increases the value, and turning the knob counterclockwise decreases the
value.
8-00433
15 System operation
indicators
8-00448
non-US version
or
8-10007
US version
1-28 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Table 1-8: GUI front view (cont inued)
Index
Labeling Funct ion
(Figure 1-3)
8-00447
non-US version
or
8-10007
US version
15 Red safety valve open (SVO) indicator. Illuminates when the ventilator has
(cont) entered its safe state and opened its safety valve to allow the patient to
breathe unassisted from room air.
8-00459
non-US version
or
8-10008
US version
Index
Labeling Funct ion
(Figure 1-3)
Green BPS ready indicator. The ventilator senses that the BPS is installed,
operational, and has at least 2 minutes of estimated run time.
8-00460
or
BATTERY
READY
US version only
On BPS power indicator. When yellow bar to the right of a lit BPS ready
indicator (battery symbol) is lit, ventilator is operating on BPS, and ac power
is insufficient to support ventilator operation. During BPS operation, power
to the compressor unit and the humidifier outlet (if available) is off.
8-00457
or
BATTERY
ON
US version only
Green compressor ready indicator. The compressor logic cable and air supply
hose are connected to the ventilator. The compressor is up to operating
pressure but not supplying gas to the ventilator. The compressor motor turns
on intermittently to keep the compressor chamber pressurized.
8-00461
or
COM PRESSOR
READY
US version only
or
COM PRESSOR
ON
US version only
1-30 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
2 3
Index
Labeling Funct ion
(Figure 1-4)
2 RS-232 10.4-inch GUI only: Two serial ports with 9-pin male connector configured as
data terminal equipment (DTE).
NOTE:
Allowable current is 0.2 A at 10 V dc (maximum).
2 3
5
4
1
7
8-00197
Index
Labeling Function
(Figure 1-5)
2 Ventilator operation
indicators
8-00447
non-US version
or
8-10007
US version
1-32 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Table 1-10: BDU front view (cont inued)
Index
Labeling Funct ion
(Figure 1-5)
2 Red safety valve open (SVO) indicator. (See Table 1-8, item 15.)
(cont)
8-00459
non-US version
or
8-10008
US version
8-00454
non-US version
8-10009
US version
Index
Labeling Function
(Figure 1-5)
2 Red loss of GUI indicator. The ventilator has detected a malfunction that
(cont) prevents the GUI from reliably displaying or receiving information.
8-00455
non-US version
or
8-10009
US version
non-US version
or
8-10006
US version
4 (Humidifier outlet, shown North-American style electrical receptacle for humidifier. Receptacle
covered) available in 100 – 120 V ventilators only. Ventilator supports a humidifier
rated for up to 2.3 A (270 VA) with a maximum leakage current of 50 µA.
7 (Collector vial drain port) Collector vial drain port. Use to attach drainage bag.
1-34 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
2
1
8-00196
Index
Labeling Funct ion
(Figure 1-6)
1 TEST TEST (service) button. Enables service mode. When you turn on the
ventilator and press this button after the first beep from the BDU, the
ventilator is placed into service mode (for example, to run EST).
2 PTS 2000 Puritan Bennett PTS 2000 Performance Test System connection.
Index
Labeling Function
(Figure 1-6)
8-00461
or
Compressor
US version
8-00427
or
Display
(GUI)
US version
1-36 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
2 3
1
5 8-00161
Index
Labeling Funct ion
(Figure 1-7)
8-00430 NOTE:
A humidifier connection is only available on 100 – 120 V
or ventilators.
Compressor &
humidifier
circuit
breaker
US version
or
Vent ilat or
circuit
breaker
US version
or
ac input
8-00405
US version
Index
Labeling Function
(Figure 1-7)
or
Compressor out let :
5.6 A max
US version
2
1
8-01298
Index
Labeling Funct ion
(Figure 1-8)
1 (High-pressure air fitting) DISS male, DISS female, NIST, Air Liquide, or SIS fitting
2 (High-pressure oxygen DISS male, DISS female, NIST, Air Liquide, or SIS fitting
fitting)
1-38 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
Index
Labeling Funct ion
(Figure 1-8)
1 (Warning label) Warns user of hazards associated with the operation of the 840 ventilator and
GUI
2 (Serial number label) Unique assigned number. M ust be the same as the GUI serial number stored
on the data key.
8-00199
Index
Labeling Function
(Figure 1-10)
8-00456
or
8-00462
US version
1-40 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
General information 1
1.13 Onscreen symbols and abbreviat ions
Consult th e 840 Ventilator System Operator’s and Technical Reference Manual to i nterpret these.
1-42 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
SECTION
S EC T I O N
Theory of operation 2 2
This section detai ls th e operation al th eory of the 840 Ven tilator System an d con tain s the
fol lowin g i nform ati on :
• descri ption of m aj or ventilator assem bli es
• overview of venti lator operati on
• descri ption of the pneum ati c system
• descri ption of the electri cal system , i ncl udi ng prin ted ci rcui t boards (PCBs)
• detail ed description of the in teractions am ong all ven tilator com pon ents during breath
deli very and under certain other con ditions
• oth er h ardware operations
• em ergen cy m odes of operation
8-01203
840 Ventilator System Service Manual 4-070496-00 Rev. A (08/ 03) 2-1
2 Theory of operation
8-00187
2-2 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
8-01159
8-8-01132
8-00199
2.1.5 Cart
The opti on al cart, sh own i n Figure 2-6, m ounts system com ponents an d accessories,
i ncl udi ng th e compressor unit. It also provi des m obili ty for th e ven ti lator. Brakes on the front
casters prevent th e cart from rolli ng an d turn ing.
8-01119
2-4 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.1.6 Patient system
The patien t system (adul t, reusable versi on shown i n Fi gure 2-7) i ncl udes reusabl e or single-
patient use tubi ng plus an i nspiratory fil ter th at prevents gas-borne parti cles from exi tin g the
ventil ator. A hum i difi er or heat and m oi sture exch an ger (HM E), water traps, and other
accessori es can be added.
8-00276
NeoM ode, a software option, requires the use of the n eonatal patient system . In addition to
the pati ent tubi ng an d i nspiratory fi lter, the neonatal patien t system i ncl udes a di sposabl e
expi ratory fi lter, a coll ector vial , an d a special m ounti ng pl ate, as shown in Fi gure 2-8.
(From patient)
Neonate
patient circuit
mounting plate
Humidifier
8-01140
Graphic user
interface (GUI)
Exhalation Inspiratory
module: module:
Active exhalation valve PSOLs
Pressure transducer Safety valve
Flow sensor Oxygen sensor
Pressure transducers
Flow sensors
Air Oxygen
regulator regulator
Air
supply
Expiratory
filter Oxygen
supply
(Expiratory (Inspiratory
Collector limb) limb) Inspiratory
vial filter
8-00001
Patient
Humidification
circuit
device
2-6 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
The GUI logs an event in th e di agnostic log an d declares a ventil ator inoperative condition if:
• An y raw si gnal data from th ree BDU transmi ssions withi n 24 hours is corrupted.
• The GUI does n ot recei ve data from the BDU wi thin the time requi red.
• The GUI determ in es th at raw data is vali d, but settings or alarm li mi ts are not being
handl ed properl y.
The ventil ator uses fl ow or pressure tri ggeri ng to recogn ize patien t effort. W hen pressure
triggering (P-TRIG) is selected, the ventil ator moni tors pressure i n th e patient circui t. As th e
patient draws gas from the circuit and the airway pressure drops by at least the value sel ected
for pressure sensiti vity (VSENS), the ventil ator triggers a breath .
W hen flow triggering (V-TRIG) i s selected, the BDU m ai ntai ns a constant fl ow of gas th rough
the pati ent circuit (called base flow) during the later part of exhalati on . The venti lator
i ndirectly m easures pati ent flow (assum in g mi ni mal leaks) by m on itori ng the di fferen ce
between th e in spi ratory an d expiratory flow sensor measurem ents. If th e patient i s not
i nspiring, any difference between th e del ivered and the ex hal ed flow i s due to sen sor
i naccuracy or leaks in th e patient system . To com pensate for leaks i n the pati ent system , the
operator can i ncrease th e flow sen sitivi ty, which i deal ly equal s desired flow sen siti vity + l eak
fl ow.
As th e patient inspires from th e base flow, th e ventil ator m easures less exhaled fl ow whi le
del ivered fl ow rem ains con stant. As th e patient con ti nues to i nspire, th e di fference between
the two flows, m easured by the in spiratory and expi ratory tran sducers, i ncreases.
The ventil ator decl ares an i nspi ration when th e fl ow i nspired by the patient (that is, the
difference between th e m easured fl ows) is equal to or greater th an the operator-selected value
for fl ow sensi tivi ty (VSENS).
As a backup m ethod of tri ggering inspirati on , a pressure sen siti vity of 1 cm H 2O i s also in
effect. Thi s setti ng is the m ost sen sitive settin g that i s stil l large enough to avoid autocycl ing,
yet wil l trigger with acceptable patien t effort.
Air and oxygen from cyli nders, wall suppli es, or com pressor (ai r only) enter the ventilator
through h oses and fitti ngs (the fittings are avail abl e i n several versions). On ce insi de th e
ven til ator, ai r and oxygen are regulated to pressures appropriate for th e venti lator, th en
m ixed accordin g to the operator-sel ected O2 %.
The ventil ator delivers th e m i xed air an d oxygen through the inspiratory module, and out to
the pati ent. Th e oxygen concentrati on of the deli vered gas is m on itored here, using a
gal vanic oxygen sensor. The gal vani c sensor gen erates a voltage proportional to th e oxygen
concen trati on . The ventil ator al arm s i f the m oni tored oxygen concentrati on i s m ore than
seven percent above or below the O2 % setti ng, or bel ow 18%. A safety valve rel ieves pati ent
pressure if necessary (for exam ple, if the patient ci rcui t i s ki nked or occluded). Th e operator
selects the hum i difi cation type, wh ich th e venti lator uses to correct for gas tem perature and
h um idity.
Ventil ator in spi ratory pn eum atics consist of two parall el ci rcuits: one for oxygen an d on e for
air. The prim ary el em en ts of th e in spi ratory pneum atics are two proportional solenoid valves
(PSOLs), whi ch control the flow of gas deli vered to the patient. Air and oxygen flow sensors,
along wi th pressure signals from the inspiratory and exhalati on m odules, provide feedback
that is used by the BDU CPU to con trol the PSOLs. As a result, th e ven ti lator suppl ies m ixed
breath in g gas to the pati en t accordin g to operator-set variables. Th e m i xed air an d oxygen
passes through th e pati en t circui t external to th e ven tilator.
The patient circuit includes th e com pon ents external to the ventil ator th at route gas between
the ventil ator and the pati ent. These com pon ents incl ude the inspiratory filter (which protects
agai nst contam i nati on between the patient and venti lator), a hum idi fication devi ce, the
i nspi ratory an d expi ratory l im bs of the pati ent circuit (th e tubi ng through which the gas
travel s), a collector vial (whi ch protects the expiratory system from bulk m oi sture i n the
exhaled gas), an d an expiratory filter (whi ch l im its the bacteria in th e patient’s exhal ed gas
from escapin g to room air or contam i nati ng the ventil ator).
The ven ti lator activel y control s the exhal ation valve, wh ich i s accurately posi tioned by
software through out i nspiration and exhal ati on, and all ows the ven tilator to deliver
aggressive breath s whi le m in im i zi ng pressure overshoots, con troll ing PEEP, an d reli evi ng
excess pressures. The exhalation system m on itors the gas leaving the pati ent circuit for
spirom etry.
Throughout the respi ratory cycle, pressure tran sducers mon itor inspiratory, expiratory, and
atm ospheric pressures. Th e tem perature of th e exh aled gas i s heated to a tem perature above
i ts dew point to prevent con densation in the exh alation com partment.
Power to operate the ven ti lator com es from ac m ains (wall) power or th e BPS. The in tegral
power suppl y is desi gned to protect again st excessi ve vol tages, tem peratures, or current
draws. A power cord retai ner preven ts accidental disconn ection . A pow er swi tch cover
protects agai nst spil ls and accidental ac power-off.
The ven til ator i ncl udes th e BPS that suppl ies dc power to the ven til ator i f ac power is l ost. A
fully ch arged BPS operatin g under n om inal am bient con diti on s can power th e venti lator for
at least 30 m i nutes. Th e BPS rech arges duri ng operation from ac power. Th e GUI an d BPS
i ndi cate when th e ventil ator i s operati ng on the BPS. Th e BPSdoes n ot power th e compressor
unit or the hum idifi er pow er outl et.
In som e situati ons, the ven ti lator decl ares a ventilator inoperativecondi ti on. A hardware fai lure
or critical software error th at could com promi se safe ventil ation triggers thi s con dition.
W h en th is occurs, th e ven tilator i noperative in dicator lights an d the ven ti lator enters th e
safety valve open (SVO) state. To correct a ven tilator inoperative condi tion, th e ven tilator m ust
be turned off, then powered on again; at power on, a quali fied service tech nician must run
extended self test (EST). The ven tilator m ust pass EST before norm al ven til ation can resum e.
In the safety val ve open (SVO) state, the safety val ve allows th e patient to breathe room ai r
unassisted. The ventil ator rem ai ns i n the SVO state unti l the con dition that caused the SVO
state is corrected or, if the ven tilator decl ared a venti lator in operative con diti on , the power
on self test (POST) verifies that power l evel s to the venti lator are acceptable an d that th e
m aj or electron ics system s are functioni ng correctl y.
If the ven til ator enters the SVO state, the safety val ve open i ndi cator ligh ts and a hi gh-
urgen cy alarm sounds. The venti lator enters the SVO state if a h ardware or softw are fail ure
occurs th at coul d comprom ise safe ventil ation, both air and oxygen suppli es are lost, or an
occlusion is detected. In case of a m alfun ction that prevents software from openi ng the safety
valve, th ere is also an an alog circui t that opens th e safety val ve i f system pressure exceeds 100
to 120 cm H 2O.
2-8 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.3 Pneumat ic syst em
This subsection descri bes th e 840 Ven til ator pneum atics, as follows:
• Inspiratory m odule: Con tain s the fol lowin g pneum atic subsystem s:
– Gas supply con diti oning subsystem
– Fl ow control subsystem
– Safety val ve/ inspiration m onitorin g subsystem
• Pati ent system
• Exhalation m odule
• Compressor uni t (optional)
Fi gure 2-10 and Figure 2-11 are diagram s of the pneumatic system . Table 2-1 l ists th e
compon ents withi n the pn eum ati c system .
Exhalat ion
m odule
Air Oxygen
Exhaled
supply supply
gas
Elect ronic
Com pressor Backup
modules
module pow er
(hardware
source
and software)
8-00141
C NC
Patient Wye
SOL2
PE EXH
HTR NO
WT WT
CV5
F9
H P Oxygen, Wall
Q3 Reg Oxygen
H P Air, Wall/Comp
Humidifier
ECV Reg Air
Mixed Gas F8
Atmosphere
Vacuum
EV To Patient
Water
Vent PSOL1
PS1 Q1
F7
F3
REG1
TP1 CV3
SV
F5
F1 OS
CV2
PS2
NO
Vent
Q2 SOL1
CV4 PSOL2
F2 REG2
F6
NC C
TP2
WT1 F4
PA PI
RV
M/C
HE PC
Accumulator
R1/F11
F13
SOL3
NC
F10
WT2 NO C
F12
Dryer
HB
8-01145
Figure 2-11. Pneumatic syst em diagram
2-10 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Reference
Component Descript ion
designat or
Inspirat or y module
Fitting, inlet Connects external oxygen and air sources to ventilator via hoses.
Fittings include diameter index system standard (DISS) (male or
--
female), noninterchangeable screw thread (NIST), Air Liquide, and
Sleeved Index System (SIS).
Manifold, flow sensor Holds bases of flow sensors Q1 and Q2, and houses pressure valves
TP1 and TP2 and pneumatic noise filters F6 and F7.
Manifold, PSOL/ SV Houses proportional solenoid valves (PSOL1 AND PSOL2) and safety
--
valve (SV).
Orifice, inspiratory pressure Bypasses inspiratory check valve to relieve pressure when an occlusion
-- relief is present in exhalation circuit while safety valve is open. Minimizes
rebreathing of exhaled gas during SVO.
Check valve, air/ compressor CV2 opens to admit external compressed air and CV4 closes to isolate
CV2 compressor unit. When CV4 opens to allow compressor-supplied
CV4 compressed air, CV2 closes to prevent compressed air (compressor
source) from venting out the air (external source) inlet fitting.
Check valve, inspiratory Opens to supply inspiratory gas and restricts exhalation flow in reverse
CV3
direction.
F3 Filter, inlet, oxygen/ air Filters matter greater than 0.3 µm (micron).
F2
F5 Filter, screen, oxygen/ air impact Filters large debris from REG1 and REG2. These filters are part of the
F4 regulator assemblies (one in each regulator).
Filter, pneumatic noise, oxygen/ Conditions gas flow by eliminating swirling of gas induced by elbows
F7
air and restrictions. These filters are part of the flow sensor manifold (two
F6
in each manifold).
Sensor, oxygen (percentage) Measures partial pressure of oxygen in inspired gas. Range is 21 to
OS
100% O2.
Pressure transducer, inspiratory Measures pressure (psig) at outlet manifold. Located on inspiratory
PI
electronics PCB.
PS1 Pressure switch, oxygen/ air Opens when pressure is less than 20.0 psig nominal.
PS2 Closes when pressure is greater than 31.5 psig nominal.
PSOL1 Proportional solenoid valve, 0 to 200 L/ min BTPS output (intermittent) or 0 to 180 L/ min BTPS
PSOL2 oxygen/ air output (steady state).
Q1 Sensor, flow, oxygen/ air Measures oxygen or air flow before PSOL.
Q2
REG1 Regulator, oxygen/ air Reduces input supply pressure (35 to 100 psig, flow up to 200 L/ min
REG2 BTPS) to output pressure (9 psig minimum to 12 psig maximum).
Reference
Component Descript ion
designat or
TP1 Pressure valve, oxygen/ air Allows measurement of REG1 and REG2 output.
TP2
WT1 Water trap, air Houses air inlet filter (F2) and includes a manual drain.
Collector vial, exhalation Collects water (up to 250 mL with the full line at 200 mL) resulting
ECV from condensation in patient circuit. Neonatal collector vial collects up
to 25 mL water (with the full line at 15 mL) .
Filter, inspiratory (main flow) Filters matter greater than 0.3 µm (micron) (nominal) at 100 L/ min
F8
flow.
Filter, expiratory Filters matter greater than 0.3 µm (micron) (nominal) at 100 L/ min
F9
flow.
WT Trap, water Collects excessive water. Present only on certain patient circuits.
-- Mounting plate (neonatal only) Allows neonatal expiratory filter to be installed in place of adult filter.
Check valve, exhalation Opens during exhalation to let exhaled gas into exhalation system.
CV5
Prevents rebreathing when safety valve is open.
Exhalation valve Electronically controlled, electrically operated valve that opens during
exhalation (as required to maintain positive end expiratory pressure
EV
(PEEP)/ continuous positive airway pressure (CPAP). Closed during
inspiration.
PE Pressure transducer, expiratory M easures pressure (psig) at a port on exhalation transducer PCB.
2-12 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Table 2-1: Pneumat ic component descriptions (continued)
Reference
Component Descript ion
designat or
Accumulat or Accumulator Four 1-L cylinders store compressed air generated by the compressor.
Air dryer Removes water vapor from the compressed air to lower the dew point
Dryer
below ambient temperature.
F10 Filter 0.3 micron filter that filters gas going to the ventilator.
F12 Filter, intake silencer Filters and silences the compressor intake.
Housing base Collects water emptied from the water trap where it evaporates using
HB
heat from the compressor motor and air flow from the fans.
HE Heat exchanger Cools the compressed air allowing water vapor to condense.
Motor/ compressor assembly Supplies compressed air to the ventilator. Includes starting capacitor,
M/ C
shock mounts, inlet filter, and relief valve.
PC Compressor pressure transducer Differential pressure transducer that measures accumulator pressure.
Water trap Collects condensate as the compressed air cools in the heat
WT2 exchanger. Automatically drains collected water to the housing base
where it evaporates.
8-01158
8-01293
2-14 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
To Patient
Vent
PSOL1
PS1 Q1
F7
F3
REG1
TP1 CV3
SV
F5
F1 OS
OXYGEN
SUPPLY
CV2
PS2
NO
Vent
Q2 SOL1
CV4 PSOL2
F2 REG2
F6
NC C
TP2
AIR
SUPPLY WT1 F4
PA PI
8-01294
Oxygen
inlet filter
(F3)
Check valve
assembly
(CV2/ CV4) Water trap (WT1)
Air inlet
filter (F2)
8-00277
Vent
PS1
F7
F3
REG1
TP1
F5
F1
OXYGEN CV2
SUPPLY PS2
Vent
CV4
F2 REG2
F6
TP2
AIR
SUPPLY WT1 F4
8-01287
Figure 2-16. Gas supply condit ioning subsyst em gas flow diagram
2-16 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
The air side of the gas supply conditionin g subsystem i ncl udes the fol lowin g com ponents:
• The a ir h ose a n d in let fit t in g connect an external high-pressure air source to th e
venti lator. Inlet fi tti ngs are avail able i n DISS m ale, DISS fem al e, NIST m ale, Ai r Liquide,
an d SIS m ale fittin gs.
• The a ir pressu re sw it ch (PS2) sen ses a loss of supply pressure. Thi s 12 V, norm all y open
swi tch closes when gas pressure reaches 31.5 psig n om i nal and opens when gas pressure
reach es 20.0 psig n om i nal .
• The a ir in let filt er (F2) protects th e pneumatic system s from debris by trappin g parti cles
down to 0.3 µm (m icron).
• The coalescin g a ir w a t er t ra p (W T1 ) removes con densati on from the hi gh-pressure ai r.
Th is conden sati on can be caused by pressure drops betw een the w all outlet an d water
trap. W T1 al so houses F2. It has a screw-type drai n valve.
• The a ir a n d com pressor ch eck va lves (CV2 a n d CV4 ), h oused i n the check valve
m ani fold, are h igh-pressure un idi recti on al valves th at prevent ventil ator backfl ow into
the air system . CV2 open s and CV4 closes wh en wall air pressure exceeds com pressor ai r
pressure (or n o com pressor is in stal led). CV4 open s and CV2 closes wh en com pressor air
exceeds wall air pressure (or w all ai r is not connected). W hen wal l air pressure is equal to
com pressor air pressure, both check val ves are open .
• The a ir regu la t or (REG2) is a sprin g-loaded, venti ng li ne regul ator that m ain tain s 9 to 12
psi g output th roughout a 35 to 100 psig suppl y range, from less th an 1 L/ mi n to 200 L/
m in. Its n om in al output i s 10.5 psi g at 2 L/ m in . REG2 vents in side the BDU. Th e a ir
im pa ct screen filt er (F4), h oused in REG2, removes debri s enteri ng REG2.
• The a ir pressu re va lve (TP2) checks the oxygen regulator settin g (REG2).
• The a ir pn eu m a t ic n oise filt er (F6), housed in the flow sensor m an ifold, con diti on s gas
flow by eli mi natin g swi rl ing of gas in duced by elbows an d restri ctions.
8-00222
Oxygen
inlet filter
(F3)
Oxygen
impact
filter
(F1)
8-00278
8-00278
2-18 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Air regulator
(REG2)
Oxygen regulator
(REG1)
8-00280
CV4 Seat
CV2 Seat Compressor
check valve
(CV4)
Oxygen seat
8-01231
On th e air side, th e air in let fil ter (F2) protects the pneum atic system from debris and W T1
protects the pn eum atic system from water condensation. Th e air transfer tube directs th e air
to the ch eck val ve m anifold, which contai ns th e air ch eck valve (CV2) and the com pressor
check valve (CV4). Th ese check val ves direct air from th e highest pressure to the next part of
the system .
NOTE:
Use an external (optional) water trap if there is condensation in the piping system.
On the oxygen si de, two fil ters protect the pneum ati c system from debris. The first fi lter, the
oxygen i m pact fil ter (F1), traps particles down to 65 µm (m icrons) an d th e subsequent fil ter,
the oxygen inl et fi lter (F3), traps particl es down to 0.3 µm (m icron ). Th e oxygen transfer tube
di rects oxygen to the ch eck val ve m ani fold. There is no ch eck val ve for oxygen .
On both th e oxygen and air sides, fi ltered gas passes th rough screen fil ters F5 an d F4, wh ich
are part of regul ators REG1 and REG2. The regulators maintain a stable nom inal output of
10.5 psi g. Pn eum atic noise filters F7 and F6, part of the fl ow sensor m anifol d, condition gas
fl ow by eli m inating swi rl ing of gas i nduced by elbows and restricti ons.
Pressure valves TP1 an d TP2 let you check the regul ator setti ngs. Th is i s don e as part of the
perform ance verifi cati on (see Secti on 5).
PSOL1
PSOL2
Q1
Q2
8-00281
2-20 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
PSOL1
Q1
OXYGEN
SUPPLY
SV
Q2
PSOL2
AIR SUPPLY
8-01286
8-01184
Oxygen and air flow sensors (Q1 and Q2) wit h ferrit es inst alled
8-00282
Oxygen and air proport ional solenoid valves (PSOL1 and PSOL2)
2-22 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
W hil e the ventil ator is in the PSOLs cl osed m ode, the followi ng are true:
• The flow requiremen t is determ ined by BDU software (lookup table or calculate). Required
curren t i s then cal cul ated and i s added to li ft-off curren t.
• Current i s sent to the PSOL. A reasonabl eness ch eck is performed on this current.
• Flow i s checked at flow sensors Q1 an d Q2.
• Target flow and fl ow at th e sensor are com pared (error determ in ed). The curren t needed to
reposition the PSOL i s calcul ated and sen t to th e appli cabl e PSOL.
• The PSOL current i s checked for reason abl eness.
W hil e the ventil ator is in the PSOLs opened m ode, the foll owi ng are true:
• Flow i s checked at flow sensors Q1 an d Q2.
• Target flow and fl ow at th e sensor is com pared (error determ ined). Current needed to
reposition the PSOL i s calcul ated and sen t to th e appli cabl e PSOL.
• The PSOL current i s checked for reason abl eness.
Target gas flow i s derived differentl y for each breath type (volum e-l im ited m an datory,
pressure-li mi ted m andatory, spontaneous, and spontan eous with pressure support). The flow
control algorith m com pensates for agin g com ponen ts. It also prevents oversh oot.
Inspiratory check
Absolute pressure valve (CV3) housing
transducer (PA)
Inspiratory
pressure
transducer
(PI)
Oxygen Safety
sensor valve (SV)
(OS)
Inspiratory
pressure
transducer
autozero
solenoid
(SOL1)
8-00152
To Patient
PSOL1
CV3
SV
OS
NO
SOL1
PSOL2
NC C
PA PI
8-01291
Figure 2-23. Inspirat ion monit oring subsyst em gas flow diagram
2.3.1.3.1 Safet y valve and inspirat ion monit oring subsyst em component s
The safety val ve section in cludes th e foll ow ing com ponen ts, shown in Figure 2-23:
• The sa fet y va lve (SV) con si sts of a +12 V solen oid actuator th at i s norm al ly energi zed
(closed) wh ile th e venti lator operates. The closed valve prevents gas from escaping to the
room and thus al lows th e venti lator to del iver gas through the outl et m an ifold to the
patient.
SV has two im portan t safety functi on s. First, if outlet pressure exceeds the m axi mum
operati ng pressure or if pressure m easured at the expiratory pressure transducer (PE)
reaches 100 cm H 2O, th e valve passively rel ieves the excess pressure. The actual SV reli ef
pressure m ay vary dependi ng on the pati ent fl ow rate. Second, i f the control system
detects a catastroph ic fai lure, the valve i s open ed an d the pati ent can freely breath e room
air.
Under som e ci rcum stances, i ncl udi ng when a venti lator in operative condi tion i s declared,
SV is de-energi zed, open ing the patien t ci rcuit to atm osphere an d venting pressure.
Typi cally, i f patient pressure exceeds 100 cm H 2O (m easured at PE), th e BD CPU
com m an ds SV to de-energize. If pressure at SV exceeds 115 cm H 2O nomi nal , SV is
m echan ical ly forced open.
• The ou t let m a n ifold h ouses the oxygen sensor (OS), th e in spi ratory ch eck val ve (CV3),
and the pressure rel ief ori fice. (See Figure 2-24.)
• The in spira t ory ch eck va lve (CV3) opens to supply inspiratory gas. It also restricts
exhal ation flow in the reverse direction wh en SV is open .
The inspirati on moni tori ng section in cl udes the fol lowing com ponents:
• The oxygen sen sor (OS) m on itors the oxygen concentrati on of delivered gas. This gal van ic
sensor gen erates a voltage proporti onal to the oxygen con cen tration . The ventil ator
alarm s if the m onitored oxygen concen trati on i s n ot wi thi n ±7% of the oxygen
percentage settin g.
2-24 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
• Two soli d-state di fferential pressure transducers m oni tor th e pressure of th e in spi ratory
an d exh aled gases. The in spir a t or y pressu re t ra n sd u cer (PI), located on the in spiratory
electronics PCB at the inspiratory lim b, mon itors the pressure of the inspiratory gas. The
expira t ory pressu re t ra n sd u cer (PE), located in the exh al ation com partm en t, m on itors
the pressure of the exh al ed gas.
• The in spir a t ory pressu re t ra n sd u cer a u t ozero solen oid (SOL1) is a +6 V, three-way
solen oi d. It i s energized periodi call y to set a zero referen ce or offset voltage for the
inspi ratory pressure transducer (PI). During the autozero procedure, SOL1 conn ects PI to
atm osphere.
• The a b solu t e pressu re t ra n sd u cer (PA), l ocated on th e inspi ratory electroni cs PCB,
m easures atmospheri c pressure (psia). One port is open ed to the atm osphere an d one port
i s con nected to a sealed vacuum ch am ber.
8-00269
Seat
Check valve
Leaf
housing
Oxygen
sensor
Oxygen
sensor
port
8-01122
Figure 2-24. Safety valve and inspirat ion monitoring subsyst em component s
8-00284
PA
PI
8-00285
Absolut e and inspirat ory pressure t ransducers (PA and PI) on inspirat ory elect ronics PCB
8-00283
Figure 2-24: Safet y valve and inspirat ion monit oring subsystem component s (continued)
2-26 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.3.1.3.2 Safet y valve sect ion operat ion
Durin g n orm al operation, the safety valve is energi zed closed so th at gas is deli vered to the
patient.
Sh ould the operating pressure of the patient system exceed 100 cm H 2O (m easured at PE),
compon ents in the safety val ve subsystem work together to relieve excess pressure.
If pressure at the safety valve reach es nomi nal 115 cm H 2O, th e valve is forced open.
To Patient
PSOL1
CV3
SV
OS
NO
SOL1
PSOL2
NC C
8-01131
PA PI
In case of a catastroph ic failure, such as a power fai lure, SV is de-energized. This acti on
perm its the patien t to breathe room air. CV3, whi ch i s bui lt i nto the outlet m an ifold,
preven ts rebreath in g from room air by en surin g th at exh alati on occurs onl y th rough the
exh alati on valve.
15 PSOL/ SV m an ifol d
4
16 Air proporti on al solenoid
8-01283
valve (PSOL2)
17 Safety val ve (SV)
Figure 2-26. Air flow diagram
18 Outl et m an ifold
19 Inspiratory ch eck val ve (CV3)
20 Pressure rel ief orifice
21 Oxygen sensor (OS)
22 Inspiratory outl et
23 Inspi ratory pressure tran sducer (PI)
24 Absolute pressure transducer (PA)
25 Inspi ratory pressure tran sducer autozero solenoi d (SOL1)
26 Inspiratory el ectronics PCB
Figure 2-26. Air flow diagram
2-28 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Gas in. Wal l air en ters the ven ti lator at the air i nlet fitti ng. The ventil ator requires wall air
pressure m ain tained between 35 and 100 psi g wi th i nterm ittent fl ow up to 200 L/ mi n. If wal l
air fal ls below compressor pressure, an optional com pressor uni t suppli es air to the ventil ator
at the ch eck val ve m anifold.
Pressure monitoring. Wal l ai r is monitored by PS2 (norm al ly open). As air pressure increases
above 31.5 psig nom inal, PS2 closes, indicating sufficient wall air pressure to operate the
ven til ator.
Filtration. Th e ai r in let fil ter (F2) traps particl es down to 0.3 µm (m icron), protecting th e
pneum ati c system from debri s. Coalescing-type water trap W T1 coll ects condensation.
Wall or compressor air. Extern all y suppli ed air con tinues through a transfer tube to th e check
valve m ani fold, wh ich contain s the air ch eck val ve (CV2) an d th e com pressor ch eck val ve
(CV4). Depen din g on wal l air pressure, either wal l air or com pressor output is suppli ed. If wall
air pressure fal ls below 26 psig, com pressor air i s supplied to REG2.
Regulation. From th e check valve m an ifold, air enters REG2 vi a the air i mpact screen filter
(F4). F4 traps parti cl es. REG2 reduces air pressure from 35 – 100 psig to 9 – 12 psi g (10.5 psi g
n omi nal). It ven ts 2 L/m i n insi de th e BDU.
Turbulence smoothing. Low-pressure ai r enters th e flow sensor m ani fol d, whi ch houses the ai r
pressure valve (TP2) and the air pneum atic noise fil ter (F6). TP2 i s used to check REG2
settings. F6 conditi on s gas flow by eli m inating swi rl ing of gas i nduced by elbows and
restri cti on s.
Air and oxygen mixing. Air con ti nues on to the air flow sensor (Q2), wh ere air flow i s m easured.
Software calcul ates air flow at Q2 and target flow. Software-controll ed current is appli ed to
PSOL2, open ing i t to al low a predetermi ned ai r flow through PSOL2. The PSOL/SV m ani fol d
h ouses proportional solenoid valves PSOL1 and PSOL2, and th e safety valve (SV). M ixed gas
i s deli vered to the patien t ci rcui t through the inspiratory outl et.
Safety valve. Should the operatin g pressure of the patient circui t exceed 100 cm H 2O (m easured
at PE), the safety valve opens. The outlet m anifold houses the oxygen sen sor (OS), th e
i nspiratory check val ve (CV3), and th e pressure relief orifice. CV3 prevents rebreath in g from
the room by en suri ng that exhal ation occurs onl y th rough th e exhal ati on valve.
Inspiration monitoring. Th e oxygen concen trati on of the deli vered gas is m easured by the
oxygen sensor (OS). The ven til ator alarm s i f th e m onitored oxygen concentration is not
wi thin ±7% of the oxygen percentage setting. Pressure m easurem ents are made by two
differenti al pressure tran sducers.
The inspiratory pressure transducer autozero solenoid (SOL1) is energized periodicall y to set a
zero referen ce or offset voltage for the i nspiration pressure transducer.
Gas in. Oxygen en ters the ven ti lator at the oxygen in let fitti ng. The ven til ator requires
oxygen pressure m ai ntain ed between 35 an d 100 psi g wi th interm i tten t flow up to
200 L/m i n.
Pressure monitoring. Oxygen is m onitored by PS1 (normally open ). As oxygen pressure
i ncreases above 31.5 psi g nom inal, PS1 cl oses, indicatin g sufficien t oxygen pressure to
operate the ventil ator.
Filtration. The oxygen im pact fil ter (F1) traps particles down to 65 µm (m icrons). Th e oxygen
i nlet fi lter (F3) traps parti cles down to 0.3 µm . These fil ters act togeth er, protecti ng the
pn eum atic system from debri s. Oxygen contin ues through a transfer tube, the check valve
m an ifold, and on to REG1. PSOL1 provides prim ary, and REG1, secondary, cross-
contami nati on protection. Thus, no oxygen check valve i s requi red i n the check valve
m an ifold.
Regulation. Oxygen enters REG1 through the oxygen im pact screen fi lter (F5). Th is fi lter traps
l arge parti cles to prevent accel eration an d to reduce parti cle tem perature. REG1 reduces
oxygen pressure from 35 – 100 psi g to 9 – 12 psig (10.5 psi g nom inal). It vents at a m axim um
of 2 L/ m in (outside th e BDU to prevent oxygen bui ldup).
2-30 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Turbulencesmoothing. Low-pressure oxygen enters th e flow sensor m ani fol d, wh ich h ouses the
oxygen pressure val ve (TP1) and the oxygen pneum ati c noise fi lter (F7). TP1 checks REG1
settings. F7 conditi on s gas flow by eli m inating swi rl ing of gas i nduced by elbows and
restri cti on s.
Oxygen and air mixing. Oxygen contin ues on to the oxygen flow sensor (Q1), where oxygen
fl ow is m easured. Software calculates oxygen flow at Q1 and target fl ow. Software-control led
current is appli ed to PSOL1, openi ng it and allowi ng a predeterm ined oxygen flow through
PSOL1. The PSOL/ SV m an ifold houses proporti onal solenoi d valves PSOL1 an d PSOL2 and
safety val ve SV. M i xed gas is deli vered to the pati ent ci rcuit through th e i nspi ratory outlet.
Safety valve. If the operati ng pressure of th e patient ci rcuit ex ceeds 100 cm H 2O (m easured at
PE), SV opens. Th e outlet m anifold houses th e galvan ic oxygen sensor (OS), the inspiratory
check val ve (CV3), and the pressure rel ief orifice. CV3 prevents rebreath in g from the room by
directi ng exhaled gas to th e exhalati on valve.
Inspiration monitoring. The oxygen concen trati on of th e delivered gas is m easured by OS. The
ven til ator alarm s i f the m on itored oxygen concentration i s n ot wi thin 7 percentage points of
the O 2 percentage settin g. Pressure m easurem en ts are made by two soli d-state differen tial
pressure transducers.
SOL1 i s energized peri odi cally to set a zero reference or offset voltage for the in spi ration
pressure transducer.
Inspiratory
filter
(F8)
Humidification
device
Water trap
(2 PL)
Wye
8-00276
Figure 2-28. Patient syst em (minus exhalat ion collect or vial and expiratory filt er)
PATIENT WYE
WT WT
F9
ECV F8
TO PATIENT
8-01290
Caut ion
To protect the exhalation assembly, always use a Puritan Bennett expiratory filter or a
recommended DAR expiratory filter.
2-32 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.3.2.2 Pat ient syst em operat ion
A predeterm i ned oxygen/ ai r gas m ixture flows from the ven tilator, through F8, and the
opti onal hum idi fication devi ce. W T col lects excessive m oisture (due to conden sation ), and
i nspiration gas is deli vered to the patient through the wye. Exh aled gas flows th rough the
wye, to th e ECV, an d through F9.
8-01175
C NC
SOL2
PE
EXH
HTR NO
CV5
F9
Q3
FROM
PATIENT
ECV
EV
8-01295
2-34 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
• The exh a la t ion va lve (EV) cl oses durin g i nspiration to
prevent del ivered gas from ven ti ng to the atm osphere.
It opens during exh alati on to m aintain th e operator-
sel ected PEEP/ CPAP. Th is active exhalation valve i s
acti vel y controll ed at al l tim es el ectron ically. Th e
exhal ation valve is activel y coordi nated w ith th e
PSOLs to all ow very fast rise ti m e and to m i nim ize
pressure spikes that woul d otherwise occur.
Using an active exh alati on val ve al so m i ni m izes
pressure rises during in spi rati on wh en a patien t
cough s; th e valve is open ed and excess pressure
vented.
• The expira t or y pressu re t ra n sd u cer a u t ozero solen oid
(SOL2) is a +12 V, th ree-way sol enoid that is opened
periodical ly to set a zero referen ce or offset voltage for
the expi ratory pressure tran sducer (PE). Both the
inspi rati on and expiratory pressure tran sducers are
zeroed togeth er. Duri ng the autozero procedure, both
solen oi ds are energi zed so th at am bi en t pressure i s
suppl ied to both ports of both transducers. An average
offset pressure vol tage for each pressure tran sducer is 8-00286
calcul ated and tested, an d stored in data m em ory. Exhalat ion valve
• The expira t or y p ressu re t ra n sdu cer (PE) (on th e
exhal ation transducer PCB) m easures th e exhalati on system pressure. Th e ventilator uses
exhal ation pressure readings in breath del ivery calcul ations. The tran sducer is autozeroed
vi a SOL2.
• A pressu re port all ows pressure m easurements i n th e exhalati on m odule.
8-01176
8-00236
2-36 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
8-00287
8-00240
8-00239
8-01257
RV
M/C
HE PC
Accumulator
R1/F11
F13
SOL3
NC
F10
WT2 NO C
F12
Dryer
HB
8-01133
2-38 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
• An em ergency relief va lve (RV) protects the com pressor system agai nst overpressurizati on
by open ing at 36 psi g.
• Cool ing fa n s blow air across a h ea t exch a n ger (H E) wh ich rem oves h eat from the
com pressed ai r. The air flow from the fans al so hel ps to cool the m otor/ com pressor.
• A w a t er t ra p (W T2) col lects the conden sation form ed as th e com pressed air travels
through the heat exchan ger. Th e water trap automatical ly dispenses its contents in to th e
h ou sin g ba se (H B) wh ere i t evaporates.
• A membrane a ir d ryer l owers the dew poi nt below am bi ent tem perature by rem ovi ng
resi dual water vapor from the com pressed ai r prior to bei ng stored i n th e accumulator. An
attached silen cer (F1 3) reduces n oi se from the air drying process.
• The com pressor u n loa d in g solen oid (SOL3) opens at com pressor start-up to reduce the
l oad on th e com pressor. As a secondary function, excess fl ow is ven ted through SOL3 to
the air dryer durin g periods of low dem and to assi st in rem oving water vapor from the air
dryer.
• Com pressed air, sufficient to m eet the ventil ator’s ful l fl ow requi rem ents, is stored in the
a ccu m u la t or.
• The 0.3 m icron filt er (F10 ) screens particl es from com pressed air fl ow ing to the venti lator.
• The com pressor pressu re t r a n sd u cer (PC) m easures pressure in the accumulator an d th e
pressure si gn al i s used to control the operati on of th e solen oi d, com pressor m otor, and
fan s. Air goin g to the pressure tran sducer passes through a rest rict or (R1 ) to reduce
pressure pul sati on s and a filt er (F11 ) to prevent contam in ants from enteri ng the
transducer.
Pressure
relief valve
To heat exchanger
Silencer M otor/compressor
Cooling fans
0.3 micron
filter
8-01264
Water
trap
Filt er
element
8-01262
To accumulator
To pressure
Unloading transducer (PC)
solenoid (SOL3)
To solenoid Filter (F10)
From heat
exchanger
8-01266 Air dryer Silencer
2-40 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Accumulator
Air dryer
assembly
M otor/
Silencer compressor
8-01270
8-01256
Duri ng the start-up sequen ce, a solenoid valve (SOL3) opens 1 second prior to the tim e the
com pressor m otor starts, and rem ai ns open for 0.5 seconds afterwards, reducing the load on
the com pressor. A tran sducer on th e PCBA m easures the accum ulator pressure, an d the signal
i s used to control the compressor operation . If pressure in th e accum ul ator drops below 13
psig, the com pressor shuts down and the ventil ator delivers 100% O 2 to the pati ent.
In stand-by mode, th e com pressor sh uts off w hen th e pressure reaches 27 psi g i n the
accum ulator, an d turns on w hen th e pressure drops to 22.5 psig. (This pressure drop can
occur because of sm all leaks i n the com pressor system or because of cool in g of com pressed air
i n the accumulator.)
The com pressor PCBA contain s an hour m eter that records the elapsed run tim e of the
com pressor. This i nform ati on is stored in the EEPROM on the PCBA, as i s th e com pressor’s
serial number. If a replacem ent PCBA is required, th e EEPROM from th e existing PCBA can be
rem oved an d i nstalled on th e n ew board, preservin g th e com pressor elapsed tim e and seri al
n um ber.
For more in form ation regardin g the 806 com pressor operati on , see Section 2.4.12 and
Figure 2-76.
NOTE:
The 10.4-inch GUI features color LCD panels, updated GUI CPU, touch frame, and
backlight inverter PCBs, and three serial communications ports. There is a single
serial communications port on the older 9.4-inch monochrome GUI (no longer
available) and the 9.4-inch color GUI color LCD panels, and one serial
communications port.
2-42 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
For a di scussi on of el ectropn eum ati c parts, or el ectrical parts used in cl ose conjunction with
pneum ati c parts, see Secti on 2.3. For wi ring detai l s, see Fi gure 2-41, Figure 2-43, Figure 2-46,
Fi gure 2-51, Figure 2-61, Figure 2-70, Fi gure 2-72, and Figure 2-78. Table 2-2 descri bes the
el ectronic parts.
Reference
Component Descript ion
Designat or
CB1 Main circuit breaker Provides over-current protection to the power supply.
CB2 Auxiliary circuit breaker Provides over-current protection to the humidifier and compressor.
J2 Potential equalization Provides attachment point to equalize electrical potential between the
connector ventilator and other equipment (such as electrical safety analyzer).
J3 Humidifier receptacle Located at the front of the ventilator, receives ac power from power
relay K1 and provides ac power to the humidifier.
J4 Compressor receptacle Receives ac power from power relay K1 and provides ac power to the
compressor via the compressor ac power cord.
K1 Power relay Switches power to the humidifier and compressor receptacles when ac
power is present and ventilator power switch is ON.
-- Power supply Converts ac and dc (from BPS) inputs to supply dc power to ventilator
electronics. Outputs include GUI +5V, GUI +12V, VH +5V, VH +12V, VH
+15V, VH -15V, BPS +36V, and relay +12V.
-- Back-up power source (BPS) With new, fully charged batteries, provides at least 30 minutes of back-
up power in the event of an ac line failure or temporary disconnect.
Output signals from the BPS PCB include charging*, charged*, BPS
model, IBATT, VBATT, and E-BP.
NOTE:
When the ventilator is running on battery power, the
compressor and humidifier are non-operational.
Reference
Component Descript ion
Designat or
-- Motherboard PCB Provides the main electrical interconnect between the breath delivery
CPU PCB, analog interface PCB, breath delivery LED PCB, exhalation
module, inspiratory module, power supply, power switch, and BD
alarm.
-- Breath delivery (BD) CPU PCB Contains the electronics and software that control all breath delivery
functions in the 840 ventilator. Communicates with the GUI CPU to
respond to operator inputs and display ventilation parameters.
-- Analog interface (AI) PCB Provides the interface for all analog signals in the ventilator. The AI and
BD CPU PCBs together provide the main intelligence and drive for the
mechanical devices and electronic sensors used in ventilation.
-- BD LED PCB Contains visual alarm indicators and LED drivers to annunciate safety
valve open, loss of GUI, and ventilator inoperative conditions. All
indicators are powered by +12V.
S1 Power switch Switches power to the power relay (K1), and power supply, enabling
its secondary outputs used by ventilator electronics.
-- BD audio alarm A continuous-tone alarm sounds if one or more of the following events
occurs: a ventilator inoperative condition, loss of communication
between the BD and GUI CPU PCBs, detection of a GUI alarm fault,
power switch failure, or low voltage condition detected by the +5V
sentry on the AI PCB.
-- Data key Required for proper ventilator operation, the data key stores BDU and
GUI serial numbers, hours of ventilator operation, and enables
ventilator software options.
-- Exhalation transducer PCB Produces a voltage signal proportional to pressure in the patient circuit
exhalation limb using the exhalation pressure transducer (PE). The
PCB uses a +10V reference.
PE Exhalation transducer M easures the pressure difference in the patient circuit exhalation limb
relative to atmospheric air pressure.
EXH HTR Exhalation heater Heats the patient’s exhaled gas, preventing condensation in the
exhalation module. The 16 W exhalation heater uses +12V.
Q3 Exhalation flow sensor Measures exhalation flow for use in determining net flow to the
patient and for spirometry. Flow sensor drivers use +12V and +5V.
Thermistor supply voltage is +12V.
SOL2 Exhalation pressure transducer This +12V, 3-way solenoid periodically exposes the exhalation pressure
autozero solenoid transducer to atmospheric pressure for setting a zero-reference offset
voltage.
2-44 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Table 2-2: Elect ronic Component Descriptions (cont inued)
Reference
Component Descript ion
Designat or
-- Inspiratory electronics PCB Produces voltage signals proportional to the pressure in the patient
circuit inspiratory limb and atmospheric pressure using the inspiratory
pressure transducer (PI) and absolute pressure transducer (PA),
respectively. Also provides an interface for the oxygen sensor. The PCB
uses a +10V reference.
PA Absolute pressure transducer Measures atmospheric pressure for use in volume delivery calculations.
PI Inspiratory pressure transducer Measures pressure in the patient circuit inspiratory limb relative to
atmospheric pressure.
SV Safety valve This +12V valve opens when de-energized (during POST, at pressures
> 100 cm H2O, during ventilator inoperative conditions, or when both
source gases are lost). Energized closed all other times.
PSOL2 Air PSOL A +12V proportional solenoid valve. Regulates the flow of air in the
inspiratory circuit by adjusting flow proportionally to the supplied
current which is under feed-back control from the Q2 flow sensor
signal and the inspiratory pressure transducers.
PSOL1 Oxygen PSOL A +12V proportional solenoid valve. Regulates the flow of oxygen in
the inspiratory circuit by adjusting flow proportionally to the supplied
current which is under feed-back control from the Q1 flow sensor
signal and the inspiratory pressure transducers.
SOL1 Inspiratory pressure transducer This +12V, 3-way solenoid periodically exposes the inspiratory pressure
autozero solenoid transducer to atmospheric pressure for setting a zero-reference offset
voltage.
Q1 Oxygen flow sensor Measures inspiratory oxygen flow. M easurements are used to control
PSOL1. Flow sensor drivers use +12V and +5V. Thermistor supply
voltage is +12V.
Q2 Air flow sensor Measures inspiratory air flow. M easurements are used to control
PSOL2. Flow sensor drivers use +12V and +5V. Thermistor supply
voltage is +12V.
PS1 Oxygen pressure switch Senses inlet oxygen pressure and opens if pressure drops below 20.0
psig nominal. Closes when pressure is greater than 31.5 psig nominal.
PS2 Air pressure switch Senses inlet air pressure and opens if pressure drops below 20.0 psig
nominal. Closes when pressure is greater than 31.5 psig nominal.
-- Oxygen sensor Senses inlet oxygen pressure and opens if pressure drops below 20.0
psig nominal. Closes when pressure is greater than 31.5 psig nominal.
Reference
Component Descript ion
Designat or
-- Graphical user interface (GUI) Provides microprocessor control and monitoring of the display and
CPU PCB user interface system. Communicates with the BD CPU PCB via an
Ethernet connection.
Backlight inverter PCB Converts +12V input into high-voltage ac that powers the fluorescent
lamps that backlight the LCD display panels. Pulse width modulation
from the GUI CPU board determines the magnitude of the ac voltage
which determines the brightness of the displays. Two backlight
inverter PCBs individually control the upper and lower LCD panels on
the 10.4-inch GUI. On the 9.4-inch GUI, one PCB performs the
backlight functions for both displays.
Touch frame PCB Contains the electronics and firmware to detect inputs to the GUI
touchscreen by means of locating the logical X/ Y coordinate position
of an operator's touch.
LCD displays Two variable voltage, 9.4-inch (older version) or 10.4-inch (current
version) color LCD flat panels display ventilator settings and patient
information.
GUI LED PCB Contains LED drivers and light bars that illuminate visual alarm and
status indicators on the GUI LED panel.
Keyboard/ knob PCB A membrane keyboard assembly containing the multiposition rotary
encoder knob is used to change selected ventilator settings, accept or
clear settings changes, provide alarm silence, reset, and volume adjust
functions, actuate screen lock and information functions, and deliver
manual inspiration, inspiratory and expiratory pause, and 100% O2.
RS-232 connectors Three (10.4-inch GUI) or one (9.4-inch GUI) RS-232 connectors
provide for serial communications with external devices such as
printers or ventilator monitors.
Remote alarm connector Enables/disables an external alarm device (such as a nurse’s call). Relay
contacts are closed during normal ventilation, and open during an
alarm condition or when unit is off.
VGA controllers Plug-in modules (on some versions of the GUI CPU board) that
provide a video interface between the GUI CPU PCB and the LCD flat
panels. The VGA controllers are integrated into the current version of
the GUI CPU PCB.
GUI alarm assembly The ventilator’s primary alarm, emits multiple frequency alarm sounds
under control of the GUI CPU PCB.
-- Compressor PCB Controls the functions of the compressor motor, cooling fans, and
unloading solenoid.
SOL3 Unloading solenoid A +12V, 3-way solenoid controlled by logic signals that reduces back
pressure on the compressor during start-up and periods of low
demand. Also vents excess accumulator pressure to assist in
transporting water vapor from the air dryer.
2-46 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Table 2-2: Elect ronic Component Descriptions (cont inued)
Reference
Component Descript ion
Designat or
PC Compressor pressure transducer Located on the compressor PCB, measures accumulator pressure and
outputs are used by compressor control logic.
2-48 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
EXH M ODULE
HUMIDIFIER
OUTLET
EXH PRESS
XDUCER
A/ Z SOL
(SOL2) AC PANEL 100-120 V
POWER RELAY
COM PRESSOR UNIT J4 ONLY
COM PRESSOR PCB K1
EXH HEATER J1 CB2
(EXH HTR) J6 BLU
PC RELAY
BRN
J1
UNLOADING
EXH FLOW SOLENOID
SENSOR (SOL3) CB1 E4
(Q3) TH1 AC BLU
J5 FILTER BRN
PCB E3
EXH VALVE J2 J3
(EV) EQUIPOTENTIALITY
CONNECTOR J4
COMP
EXH PCB
M OTOR
TRANSDUCER
FAN 1 FAN 2
PE STARTING
CAPACITOR
elecsys dgm
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 2-49
2 Theory of operation
2.4.3.2 ac panel
The ac panel (Figure 2-42 and Fi gure 2-43), a si ngle fi el d-repl aceabl e un it (FRU), i ncl udes
com ponen ts that appl y ac to the ventil ator head, compressor unit, and optional hum idifier.
The ac panel includes ac receptacl e J1, circui t breakers CB1 and CB2, ac fil ter PCB, power relay
K1, com pressor receptacle J4, and potenti al equali zation con nector J2. A hum i difi er
receptacle (J3), which is conn ected to but is not a part of the ac pan el, is on the front of the
venti lator.
• The a c recept a cle (J1) receives facil ity ac power via the power cord.
Warning
Ensure that the power cord retainer bracket is properly installed and secures the power cord
to the ac receptacle (J1).
• The ac panel h ouses two push-to-reset type circui t breakers. Th e 5 A m a in circu it brea ker
(CB1) li mi ts curren t to the power supply. The 10 A (100 – 120 V) or 5 A (220 – 240 V)
a u xilia ry circu it brea ker (CB2) li m its curren t to the com pressor and hum idi fier ci rcuits.
• The a c filt er PCB reduces ac lin e noi se to and from the ventil ator.
• The power rela y (K1 ) swi tches power to h um idifier receptacle J3 and com pressor
receptacle J4 when th e power switch (S1) is on. The rel ay is powered from a dedicated
+12 V power suppl y output, whi ch i s active when the power switch i s on .
• The com pressor recept a cle (J4) recei ves ac power vi a K1 an d provi des the ac power outlet
for the com pressor power cord.
• The h u m id ifier recept a cle (J3) receives ac power via K1 an d provides an ac power outlet
for the hum i difi er power cord.
• The pot en t ia l equ a liza t ion con n ect or (J2 ) i nterfaces th e equipm en t an d th e potential
equali zation bus bar.
2-50 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
ac receptacle
(J1)
M ain
circuit
breaker
(CB1)
8-00288
Power
relay
(K1)
Fuse
Auxiliary (compressor/
humidifier circuit breaker (CB2) 8-00246
acac power
Auxiliary mains
receptacle
circuit breaker
EXTERIOR OF
AC PANEL
Potential
Compressor/ humidifier equalization
circuit breaker point
Compressor
receptacle
GRN/YEL
BRN
J1
C1 BLU 4
2
L N
G CB1
BLU
INTERIOR OF E2 E1 BRN
AC PANEL J2
C2 1 3
K1 GRN/YEL
2 4
E4 E3 BRN
CB2
J4
L N 1 3
L G
BRN BLU
GRN/YEL
BRN
BRN
BRN
BLU
BLU BLU
1 3 BLU
4 6 BLU BRN
7 9 BRN
A B
LEGEND:
CB1 Power supply circuit breaker
CB2 Auxiliary mains circuit breaker
J1 Mains inlet
BRN
J2 Potential equalization point BRN
BLU
J4 Compressor outlet To power supply
K1 Power relay RED BLK compartment
8-00125
AC PANEL M ODULE
2-52 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.4.3.3 Power swit ch
The logic-dri ven rocker-type power swi tch (S1), sh own i n Fi gure 2-44, en ables the power
supply secondary outputs that are used by ven ti lator el ectronics. W h en the power switch is
off, the power rel ay is open , preven ting ac distribution to the humi difi er and com pressor
receptacl es J3 and J4. Regardl ess of the power switch positi on , ac power i s supplied di rectl y to
the power suppl y to perm it charging of the BPS. Secondary circuits that power ventil ator
el ectroni cs are n ot active wh en the power switch is off.
A cover protects th e power sw itch an d prevents i t from acciden tall y bei ng turned off. An LED
i ndicator beside the power swi tch lights to indi cate th at power i s avai labl e to the ventilator.
Power
switch
(S1)
LED
Indicator
8-00289
8-00169
2-54 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Comms
outlet/ hmd
cable assy
Humidifier
outlet (J1) dc blindmate
(100 – 120 V only)
INSPIRATORY
M ODULE Connects to Inspiratory
inspiratory blindmate
COM PARTM ENT
module To ac
panel
POWER SUPPLY
To backup
power supply
(BPS)
8-00128
Figure 2-46. 840 Vent ilator Syst em interconnect diagram – Pow er dist ribut ion
2.4.5 BPS
The BPS, sh own in Fi gure 2-47, provi des a dc power source durin g a bri ef ac li ne fai lure or
tem porary disconnect. A ful ly charged BPS suppli es at least 30 m i nutes of power backup to
the ventil ator. Th e BPS does not power the opti onal com pressor or hum i difier; these operate
from ac li ne power.
NOTE:
It is assumed that the ventilator is powered from an outlet connected to the hospital’s
emergency generator ac backup system.
W h en th e ven tilator power switch (S1) is on, the BPS provi des a +19 to +30 V supply voltage
to the power supply. The ventil ator software m oni tors thi s voltage to determ in e wh eth er the
battery i s present and the battery status. Th e softw are triggers a m edium-urgen cy alarm if th e
battery capacity falls below 2 m inutes run tim e.
The BPSelectroni cs charge th e battery pack as necessary w henever ac is present and suffi ci ent
(> 80 V), even i f the power switch is off. An i ndi cator di splays the batteri es’ ch arging status:
yel low m eans the batteries are chargin g, and green m eans th e batteri es are are approxim ately
75% charged rel ative to n om in al. Good batteri es can be ch arged withi n 8 h ours. Th e BPS has
overcurrent (overcharging) protection.
NOTE:
Battery packs, BPS assemblies, or units with BPS’s that are unused or in storage for over 6
months, or have been depleted by continuous usage require a minimum of 8 hours to
recharge the battery pack.
The BPS contai ns a battery pack (Fi gure 2-48) (w hich i ncl udes two +12 V, lead-aci d batteri es
an d a 15 A fuse) an d a PCB (Fi gure 2-49).
8-00199
2-56 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Fuse
8-00273
8-00272
M otherboard PCB
8-00156
2-58 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
Power
BD alarm switch (SI)
To ac
panel
J7 J8
J6
J4
J2
P2 P3
CPU PCB
BREATH DELIVERY
UNIT (BDU)
TEST
switch
(SW2)
P1
LED
indicator
GUI-to-BDU
cable assy
P9 P10
AI PCB
PTS 2000
connection
9-pin J1
J6
P5
8-00124
AI PCB connectors
BD CPU PCB
connectors
8-00290
2-60 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
M otherboard PCB
8-00157
P1 J7
BD +5 V, BD +12 V
J8
PWRSW ON
J11
TO BDU BDU LEDS
LED PCB
J9
PWRSW POSITIONS
FROM PWRSW
TO ALARM
PWR RELAY
COMP CTRL/
TO PWR RELAY
J10
DATA, ADDRESS,
(AC PANEL)
CTRL, STATUS
STATUS
J13
+15 V, -15 V
TO/ FROM
AI PCB
EXH MOD PWR & SENS
TO/ FROM
EXHALATION
MODULE
BD ALARM
+12 V EXH AZ
J12
INSP MOD PWR & SENS
TO/ FROM
INSPIRATORY
INSP MODULE CTRL
MODULE
8-00300
2-62 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.4.6.2 BDU CPU PCB
The BD CPU PCB (Fi gure 2-55), in conj un cti on with the AI PCB, provi des mi croprocessor
control of all breath deli very functi ons for th e 840 Ventil ator System . It al so com municates
wi th the GUI CPU PCB for displ ay and control i nformation from the operator. All an alog
sign al s to and from the sensors and actuators of th e system are controll ed by software
run nin g in the BD CPU.
The PCB resi des wi thin the card cage, next to the AI PCB. Th ese two boards are closel y
coupl ed and provide the m ai n intell igen ce and drive for all the mechanical devices an d
el ectroni c sen sors used in ven tilati on.
The BD CPU PCB pl ugs i nto the m otherboard usin g two standard DIN 96-pi n con nectors. The
first con nector (P2) is used for power. The second con nector (P3) is an En han ced Eurocard
DIN 96 connector wi th additi on al con necti on s for power an d ground.
The BD CPU PCB con trols an d m oni tors the status of ven ti lator power, th e BPS, th e
compressor, the AI PCB, and the opti on PCBs. The BD CPU PCB includes these circuits, wh ich
perform th e in dicated functions:
• The CPU a n d con t rol circu it s i ncl ude a M otorola 68040 mi croprocessor (wh ich in cludes
an onboard m ath coprocessor), operating at 24 M Hz, al ong with supportin g devices to
perform ven tilator control functi ons. In addi ti on to executin g instructi ons, th ese
function s i ncl ude passi ng data back an d forth between m em ory an d I/O devices;
gen erati ng address signal s used to access m em ory locations; generatin g read/ write,
tim in g, and other control si gn als; and processi ng in terrupts from the various system
devi ces.
• The PCB’s m em ory i ncl udes 2 M B of fl ash m em ory (soldered; n o sockets) th at con tain s the
operational software for the BD. Thi s software is down loadabl e vi a a PC. A boot mem ory
PROM contains the ventil ator’s in iti al izati on and POST code and provides a basi c
com m un ications program to perm it downloadi ng of system software. Downloaded
software is not written to the boot memory PROM ; it cannot be written to i n the fiel d.
4 M B of dyn am ic random access memory (DRAM ) i s used for on goin g cal cul ations and
data storage and as a m essage an d com m and buffer for the Eth ernet control ler.
• 16 KB of NOVRAM circu it (nonvolati le RAM ) stores essenti al, persisten t variables an d
confi gurati on i nform ati on; current breath m ode setti ngs (for obtaini ng defaul t
param eters at power on); POST and exten ded self test (EST) fault inform ati on; the reset
stack poi nter; and oth ers (see Table 2-4). Th e NOVRAM devices are soldered in place (no
sockets). Alth ough i nform ati on i n these devi ces is lost wh en the BD CPU PCB is repl aced,
m uch of this data is updated durin g POST and EST. Critical data speci fic to a particular
un it (i ncl udin g system serial n um bers) is stored on a data key devi ce whi ch stays with the
unit.
Each NOVRAM devi ce con si sts of an internal fl ash m em ory device (whi ch has the
advantages of bei ng abl e to retain data without th e appli cati on of power, an d wh ich also
can be easil y written to) and a static RAM device. Upon power-up, the data stored in fl ash
mem ory i s retri eved and stored in stati c RAM . Duri ng norm al operation, the NOVRAM
behaves like stati c RAM . Upon detectin g l ow suppl y voltage, the NOVRAM stores data
from static RAM i nto i ntern al flash m emory for nonvol atile storage. This operation takes
2.5 m s. The BD CPU PCB provi des an i solated, decoupled power suppl y (V CAP),
indepen dent of the +5 V supply, to the NOVRAM to en sure th e power-down ram p i s long
en ough to store the stati c RAM con ten ts.
TEST button
• The rea l-t im e clock tracks the ven tilator operational tim e and sh ort-durati on power l osses
(up to 60 m i nutes). W hen the ven til ator i s operati ng, th e real -ti m e cl ock is powered by
V CC. W hen power is lost or rem oved, a bui lt-in battery powers the cl ock. Thi s battery has
a 10-year l ife expectancy.
• The w a t ch d og circu it m oni tors th e BD CPU for safe operation. W h en a tim e-out occurs,
indicati ng a l ack of bus acti vity, a system reset occurs. Probl ems with the CPU, clock, or
software can cause a watchdog tim e-out.
• The GUI-BD U in t erfa ce provides for Ethernet comm un ication between both BD an d GUI
CPU PCBs. This control ler sends an in terrupt to th e BD CPU to inform it of a
com m un ication from the GUI CPU. Com m an ds an d m essages to be tran sferred (vi a direct
m em ory access, DM A) are stored in DRAM . The GUI-BDU i nterface uses isolation
tran sform ers an d an isolated power supply, but i s connected to standard system ground.
There i s no isolation from any external devi ce that m ay be attached.
• An RS-2 32C ch a n n el i s avail abl e for developm en tal use onl y.
• A d ia gn ost ic LED a rra y (with supporting circuitry) indicates the status of the BD CPU
PCB. Duri ng POST, they i ndi cate the curren t test step. A n in th LED di spl ays th e
supervisory m ode status of th e CPU.
2-64 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
• The pow er fa il d et ect ion circu it causes an NM I to be generated if the power suppl y
gen erates a power fail si gnal . The power supply generates a power fail si gnal i f the output
of the bul k suppl y i s less than 19.30 V.
• The d igit a l I/O in t er fa ce circu it provi des th ese outputs:
– A signal (redundant) to declare a ven ti lator i noperative con diti on and to l ight an LED.
– A signal to place the BDU hardware into a safety val ve open (SVO) state an d to li gh t an
LED.
– BD LED PCB con trol signals, i ncl uding loss of user i nterface (LOUI) LED, safety valve
open (SVO) LED, and ven til ator inoperative (VENT INOP) LED.
– Com pressor control sign als (representin g on, standby, an d off). The on com m and runs
the compressor contin uously wh en the ventil ator is operating on compressed air (that
i s, when the wall air suppl y h as dropped bel ow 14 psi g). Standby causes the
compressor to cycle to main tain accum ulator pressure; wall ai r pressure drops to
between 14 and 26 psi g, and the com pressor suppl em en ts th e wall air pressure. The off
comm and overrides the com pressor autom atic cycl ing control (to m aintain
accum ulator pressure) to en sure the com pressor is off for self test purposes.
– Power m ode si gnal suspen ds battery charging to all ow a true readin g of battery
voltage.
– Low ac power si gnal (ac < 80% of n omi nal) to com pressor.
– Si gn als to pl ace the venti lator in to the service m ode (i n conj unction with a
m om entary push-button swi tch) and to di sable the two POST 10-s tim ers (n ecessary to
all ow test m ode).
Th e di gi tal I/O interface provi des th ese i nputs:
– Com pressor status signal s, in cl udi ng: com pressor present; com pressor com partmen t
overtem perature; ac i nput to com pressor low; compressor accum ul ator pressure
adequate for ven tilation; com pressor i n stan dby m ode, com pressor off, or com pressor
on .
– BPS status si gnal s including: battery bei ng charged, battery in use, and battery fully
charged.
– Servi ce m ode enabled
• The AI PCB in t er fa ce provides for com m uni cation between the BD CPU an d th e AI PCB.
• The BDU LED displa y d river s in terface the BDU LED PCB.
• The ser vice m ode sw it ch is a push-button on the board edge used to activate thi s mode.
The BD CPU PCB generates an NM I un der any of these con dition s:
• Ethernet parity error detected
• Power fail si gn al
• A/ D con verter system error
Background alarm Status of all DEVICE ALERT alarms BD alarm statuses are stored in
status BD NOVRAM. GUI alarm
statuses are stored in GUI
NOVRAM.
Diagnostic logs This includes the System Diagnostic Log, System Information BD events are stored in BD
Log, and EST/ SST Diagnostic Log (which are actually stored in NOVRAM . GUI events are
two separate locations -- one for EST, one for SST). stored in GUI NOVRAM .
Alarm history log History of alarm events for the current/ previous patient GUI NOVRAM
NOTE:
The alarm history log is automatically cleared upon
successful completion of a new patient setup. (A new
patient setup is required to start normal ventilation after
exiting the service mode.)
EST and SST result Result and status information for each EST and SST test BD information is stored in BD
information NOVRAM . GUI information is
stored in GUI NOVRAM .
POST test status Status of each of the POST/ kernel tests BD statuses are stored in BD
NOVRAM . GUI statuses are
stored in GUI NOVRAM .
2-66 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.4.6.3 Analog int erface (AI) PCB
The AI PCB (Figure 2-56) provides an in terface between the ven tilator’s m i croprocessor
circuits an d i ts an alog system s vi a an IP bus. Th ese analog system s in cl ude the com pressor, dc
el ectroni cs, val ves, BD (con tinuous-tone) audio alarm , and tran sducers. The com pressor
control and dc status sign als and dc power pass th rough the AI PCB. The PCB i ncludes these
circuits, wh ich perform th e i ndi cated functions:
• The d igit a l in t erfa ce circu it provides for com m un ication between the BD CPU PCB an d
the analog ci rcui ts on the AI PCB.
• The pressu re a n d flow sen sor filt ers m i nim i ze pn eum atic n oi se i n th ese com pon ents’
readi ngs.
• The +10 V referen ce produces a reference voltage for use by anal og-to-digi tal and digi tal-
to-an alog converters and pressure transducers (in spi ratory electron ics PCB and exhalati on
transducer PCB).
• The d a t a key in t er fa ce provides clock and data l atch es to handl e the flow of data between
the BD CPU an d data key.
• The t h erm ist or a m plifiers am pl ify i nput sign als from the m an ifold heater therm istor and
the motor m agnet th erm i stor, both located in th e exhalati on compartm ent.
• The va lve con t rol a n d d rive circu it provides drive si gnals for th e PSOLs, EV, an d th e EV
stabili zer devi ce.
PTS 2000
connector
Data key
connector
Compressor
connector
AI LEDs
8-00267
• The BD a la rm circu it acti vates th e BD al arm . The alarm i s activated under any of these
con di ti ons:
– A venti lator in operati ve condi ti on is declared
– The BD CPU detects a l oss of comm unicati on wi th the GUI
– A GUI alarm faul t i s detected
– The power swi tch fail s
– The +5 V sen try on th e AI PCB detects a low voltage condition
Under m ost con diti on s, the +5 V powers th e al arm. If the +5 V sentry circuit on the AI
PCB detects a power fai lure, a power fail capaci tor powers the alarm for a m i nim um of 2
m in.
• The PSOL pow er d iscon n ect circu it rem oves power from the PSOL drivers wh en the
ventil ator is in the safety val ve open state.
• The exh a la t ion va lve circu it drives the ex hal ation valve m otor. Pressure readi ngs provide
feedback to th e software that determ ines the am ount of valve dampening requi red. The
circui t includes pow er di scon nect ci rcuitry, wh ich rem oves power from the exhal ation
valve dri ver when th e ven tilator is i n the safety valve open state.
• The sa fet y va lve circu it control s an d drives the safety val ve. Full power to the safety val ve
pul ls the valve cl osed. After a sh ort in terval , power i s reduced to keep the val ve closed.
The safety valve circui t ensures full power i s applied in th e even t of a loss of +12 V
(required to reopen the val ve).
• The volt a ge sen t ry circu it creates si gnals that are proportional to voltages used in the
system . Th ese signal s are used in software m oni toring of th ese vol tages: +10 V referen ce
and +5, +12, +15, an d -15 V supply vol tages.
In addition, th e sentry circuit generates a global reset i f +5 V i s out of ran ge. A rel ay, wh ich
i s part of the sentry ci rcuit, en sures th at a power fail reset occurs during vol tage tran sients.
• The BPS m od el sign a l b u ffer provides a signal THAT yi elds BPSm odel type i nform ati on in
an an alog form. The software reads th is sign al to determ in e whether the BPS i s currentl y
con nected to the ventil ator.
• Other bu ffers provide condi tioni ng for vari ous ven tilator sign als.
• The com p ressor in t erfa ce circu it , i n conj un ction with the com pressor PCB, controls and
m on itors th e com pressor operation.
2-68 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.4.7 Dat a key subsyst em
NOTE:
Call your Puritan Bennett representative if the data key requires replacement due
to loss or failure.
The 840 Ven til ator System uses a data key (Figure 2-57) to record data speci fic to a particular
ven til ator uni t. The data key provides a way to retain data when PCBs or the battery are
rem oved from the ven til ator. Th e data key pl ugs into a receptacle on the venti lator. Data is
read from or written to flash m emory on the data key. Thi s serial data is tran sferred to th e BD
CPU th rough th e data key i nterface. The data key must always be installed in the ventilator for
proper operation.
The data key stores this data:
• Seri al numbers of th e GUI and BDU.
• Hours of ventilator operation .
2.4.8 GUI
The GUI (Fi gure 2-60), wh ich detach es from the ven til ator h ead for servici ng, is an enclosure
that houses th e GUI CPU PCB, keyboard, GUI LED PCB, backlight i nverter PCBs, knob, two
10.4-inch or 9.4-i nch color LCD flat panels, a touch fram e PCB, and an al arm speaker.
GUI CPU
PCB
Backlight
inverter PCB
8-00166
Figure 2-59. 9.4” GUI CPU PCB and backlight inverter PCB in place
2-70 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
The GUI CPU PCB i ncludes th ese circuits, whi ch perform the in dicated functions:
• The CPU a n d con t rol circu it includes a M otorola 68040 m icroprocessor (whi ch i ncl udes
an onboard m ath coprocessor), operating at 24 M Hz, al ong with supportin g devices to
perform ven tilator control functi ons. In addi ti on to executin g instructi ons, th ese
function s i ncl ude passi ng data back an d forth between m em ory an d I/O devices;
gen erati ng address signal s used to access m em ory locations; generatin g read/ write,
tim in g, and other control si gn als; and processi ng in terrupts from the various system
devi ces.
• The PCB’s m em or y incl udes 4 M B of flash m em ory (soldered; n o sockets) that contain s the
operational software for the GUI. A boot PROM contai ns ventil ator in iti alization and
POST code and provides a basic com m un ications program to perm it dow nloading of
system software. Downl oaded software is not written to th e boot m em ory PROM ; the
PROM cann ot be written to i n the fi eld. 4 M B of DRAM i s used for on goin g cal cul ati ons
an d data storage and as a m essage an d com m and buffer for th e Eth ernet con troll er.
• 64 KB of NOVRAM (nonvolati le RAM ) stores essential, persisten t variabl es and
confi gurati on i nform ati on; current breath m ode setti ngs (for obtaini ng defaul t
param eters at power on); POST and EST fault i nform ati on; the reset stack pointer; and
oth ers (see Table 2-3). The NOVRAM devices are soldered in place (no sockets). Al though
inform ation in these devices is l ost when the GUI CPU PCB is repl aced, m uch of this data
is updated duri ng POST and EST. Critical data specifi c to a parti cular uni t (i ncl udi ng
system serial n um ber) i s stored on a data key devi ce, wh ich stays with th e uni t.
Each NOVRAM device consists of an i nternal flash mem ory devi ce (wh ich has th e
advan tages of bei ng abl e to retai n data wi thout the application of power, and whi ch also
can be easily written to) an d a stati c RAM devi ce. Duri ng power on, data stored in the
flash m em ory i s retrieved an d stored in static RAM . Durin g n orm al operation, the
NOVRAM behaves li ke static RAM . Upon detectin g l ow supply vol tage, th e NOVRAM
stores data from static RAM i nto i ntern al flash m em ory for nonvolati le storage. This
operation takes 10 m s. Th e GUI CPU PCB provi des an isol ated, decoupled power supply
(V CAP), i ndepen dent of the +5 V suppl y, to the NOVRAM to ensure th e power-down ramp
is long en ough to store the static RAM con tents.
• The rea l-t im e clock tracks the venti lator’s operational time and short-duration power
losses (up to 60 m in utes). W hen th e ven tilator i s operatin g, th e real-ti me clock i s powered
by V CC. W hen power is lost or rem oved, a bui lt-in battery powers the cl ock. Th is battery
has a 10-year l ife expectan cy. Th e real -tim e clock can be rem oved from i ts socket for field
replacem ent.
• +5 a n d +12 V m on it ors signal the mi croprocessor (vi a a n onm askable i nterrupt) wh en
+5 V i s too h igh or +12 V is out of ran ge. Thi s ci rcuit also generates a power fail reset i f the
+5 V output drops below +4.8 V.
• The w a t ch d og circu it moni tors the CPU for safe operation. W hen a ti me-out occurs,
indi catin g a lack of bus activity, a system reset occurs. Probl em s wi th the CPU, cl ock, or
software can cause a watchdog tim e-out.
• The GUI-BD U con t roller provides for Ethernet com m un icati on between both BD an d GUI
CPU PCBs. Th e control ler sen ds an in terrupt to th e GUI CPU to i nform i t of a
com m un ication from th e BD CPU. Com m an ds and messages to be transferred (via DM A)
are stored i n DRAM . The GUI-BDU i nterface uses i solati on transformers an d an isol ated
power supply, but i s conn ected to stan dard system ground. There i s n o i solati on from any
external devi ce that m ay be attached to it.
• Three RS-232 C ch a n n els provide output for digi tal com m un ications in terface and
external comm uni cations for servi ce mode. They are electri cal ly isol ated for safety. The
9.4-in ch GUI has onl y one RS-232 channel , al so electrical ly isol ated.
Backlight PCB
Upper LCD
Backlight cable panel
4-076689-SP Flex circuit,
video
Backlight PCB
Lower LCD
panel
4-076689-SP
GUI LED PCB
J11
Flex circuit,
RS-232
J19 J20
To BDU
RS-232
Remote alarm
J2 Keyboard
cable assy
Rear of keyboard
Touch frame
cable assy
GUI M ODULE GUI10_4wiring
Figure 2-60. 840 Ventilator System int erconnect diagram – GUI 10.4-inch LCD panels
2-72 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
• D ia gn ost ic LED a rra y uses ei gh t discrete LEDs, in con juncti on wi th supporti ng ci rcuitry,
to indi cate the status of the GUI. Duri ng POST, th e array in dicates the current test step. A
nin th LED displays the supervisory m ode status of the CPU, and a 10th LED in dicates
power on.
• The VGA vid eo con t rol circu it , com pri sin g two VGA LCD controll er ICs w ith 1 M B video
RAM , i nterfaces the two 640 x 480 x 256K col or 10.4-i nch LCD panel s. Older GUI versi on s
with 9.4-in ch, 4K color or m on ochrome (16 greyscale levels) displays contai nin g th e older
GUI CPU board contai n plug-i n IP in terface VGA LCD controll ers that in terface the LCD
panels.
• An in fra red t ou ch sca n n er in t er fa ce, which includes its own m icrocontroller, mon itors
the touch screen.
• The LED PCB in t erfa ce lets th e m icroprocessor control the di screte LEDs on th e GUI LED
PCB.
• The LCD ba ckligh t con t rol circu it con trols the brigh tness of the two fluorescent lam ps
that li ght th e fl at panel s based on software in put. It generates a pulse width m odulated
si gn al to th e backli gh t in verter PCB(s) (two PCBs on 10.4-in ch GUIs; one PCB on 9.4-inch
GUIs), wh ich controls the brightn ess of both LCD panels. There is n o user-adj ustable
bri ghtness con trol on color displays.
• The rem ot e a la rm rela y in t er fa ce enabl es/ disabl es an extern al al arm device (such as a
nurse’s call ). The rel ay contacts are closed when the un it i s i n a norm al ventil ation state
an d open ed wh en th e un it is off or in an al arm state.
• The sou n d gen era t or in t erfa ce produces the venti lator’s al arm sounds. Con sisti ng of a
m icrocon troller an d audio sign al processing hardware, th e interface produces ISO-
standard alarm sounds and provides keyboard entry audible feedback.
• The rot a r y en coder/d ecod er circu it determi nes the direction and am ount of knob
m ovem ent. Th e opti cal encoder i ncorporates an em i tter secti on , tw o codewheels, and a
detector secti on . Each codewheel has a pattern photographi call y pl ated on it. As th e knob
sh aft revolves, th e codewh eel s rotate with respect to th e em itter and ph otodetector
secti ons, causi ng th e l ight beam to be in terrupted by the pattern of spaces an d bars on th e
codewheels. Th e detectors are posi tioned such that a light period on one ph otodetector
corresponds to a dark period on the other photodetector. Usi ng the photodetector
outputs, the decoder can determ i ne th e knob posi tion.
• The keyboa rd in t erfa ce reads the keyswitches and controls the lighting of the LEDs on the
keyboard.
The GUI CPU PCB generates an NM I un der any of these condi tions:
• GUI +5 V i s hi gh
• GUI +12 V i s out of ran ge
• Ethernet parity error detected
• SAAS m icrocon troller fai lure
Backlight
control
cable assy
Lower LCD
panel
GUI LED PCB
Cable assy
P1
J4 J5
J3
J7 J8 Color display
J9 cable assy
Color display
J16 cable assy
Keyboard
J5
cable assy
Figure 2-61. 840 Ventilator Syst em int erconnect diagram – GUI 9.4-inch LCD panels
2-74 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.4.8.2 Touch frame PCB
The touch fram e PCB i s responsible for detectin g operator inputs on the GUI screen. Different
touch fram e designs are used in the 10.4-in ch GUI and the 9.4-i nch GUI.
On the 10.4-in ch GUI, th e touch fram e PCB comm unicates with the GUI CPU PCB usi ng a
serial in terface, an d operates from a sin gle 5V supply, drawi ng no more th an 100m A. Infrared
(IR) tran sm ittin g LEDs are m oun ted on all sides of the PCB and IR detectors, at whi ch
i nvisible IR light beam s are aim ed, are l ocated at strategic poin ts aroun d th e PCB. Thi s
arran gem ent of LEDs and detectors define an in tersecti ng pattern of li gh t beams th at is used
to determ i ne th e l ocation of an input on th e GUI screen. See Figure 2-62.
Photosensitive
diodes
Light-emitting
diodes
A m icro-con troll er located on th e touch fram e PCB scans the touchscreen for operator inputs
by swi tchi ng on th e LEDs one at a ti me in a pseudo random pattern , and then sam pl ing the
output results from the logi cally associ ated IR detectors. A com plete scan of al l the LEDs
occurs 39 ti m es per secon d. W h en the GUI screen is touch ed, som e of the li gh t beam s don ’t
reach th ei r associated IR detectors, and as a result, there is no output from those detectors and
they are consi dered "blocked." Th e m icrocon troller ch ecks pairs of LEDs and IR detectors for
bl ocked beam s, uses a mathem atical coordi nate con version routin e to l ocate the touch in th e
i ntersectin g area of l ight beam s, and sends the information to the GUI CPU board for
processin g.
To i m prove touch fram e rel iabil ity, each LED output i s received by two IR detectors. The
bl ocked beam s are detected si m ul tan eously in di fferent intersecti ng areas provi din g
redun dancy whi ch all ows conti nued operation of th e touch fram e in the event an LED or IR
detector burns out. Durin g n orm al conditions, wh en all LEDs and IR detectors are
functi onin g properly, th e redun dant detectors are used to provi de a m ore accurate output by
averagin g the locations of th e blocked beam s.
On th e 9.4-inch GUI, the touch fram e com m uni cates wi th th e GUI CPU PCB vi a a Carrol l-
Touch propri-*+9I CPU PCB sequenti ally pul ses the LEDs, creating an invisibl e grid of IR l ight
beam s just in fron t of th e di spl ay surface. The phototransistors sense these l ight beam s.
W h en th e screen is touch ed, the light beam s from certain LEDs are obstructed and do not
reach their correspon din g ph ototransi stors. The interface constantl y m onitors the presen ce,
or in thi s case the absence, of th e IR light beam s recei ved by the phototransi stors and can
thus determ ine th e X - and Y-coordi nates of the touch acti vati on.
On both th e 10.4-in ch and 9.4-inch GUIs, th e touch fram e i s attached to th e face of the
di spl ay, wh ere i t is con cealed behi nd an IR-transparent bezel (a type of en cl osure that is
opaque to the eye, but th at al lows IR l ight to pass through).
The m ultiposition knob assem bl y, w hich is part of th e keyboard assem bly, perm i ts ventil ator
setting selection s or chan ges. Kn ob encoder/ decoder circui try on the GUI CPU PCB
determ in es th e di recti on and positi on of th e shaft based on en coder outputs.
2-76 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
8-00294
2-78 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.4.8.5 Backlight invert er PCB and LCD lamps
The 10.4-inch GUI uses two backl ight i nverter PCBs to convert +12V to a h igh-voltage ac
l evel that indivi dually powers the fluorescen t lam p tubes in the upper and lower di spl ays.
These lam ps backl igh t th e LCD flat panels. Indi vidual pulse-width m odul ated si gnals from
the GUI CPU PCB determ in e the m agn itude of the voltage, which in turn determi nes the
brigh tness of each LCD pan el. (There is n o user-accessi bl e brightness or contrast adj ustm en t
avail abl e wi th col or LCD panels.) Th e l am ps on th e 10.4-in ch LCD panels do not requi re
replacem ent.
The 9.4-i nch GUI uses on e backl ight i nverter PCB to perform th e sam e fun cti on described
above, and con trols th e backli ght for both LCD panels si m ultaneousl y. Th e LCD l amps are
part of the 10,000-h our preventive m ai nten ance ki t on th e 9.4-in ch GUI.
LCD panels
8--00184
Tube
element
8-00185
8-00295
2-80 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.4.9 BDU LED PCB
The BDU LED PCB (vent head
LED PCB), sh own in Figure 2-68,
contain s three LED-based visual
i ndi cators an d LED drivers.
These com ponen ts operate
under control of th e BD CPU
PCB. The in dicators (safety valve
open , ventil ator inoperative, and
l oss of GUI) backl igh t the
ven til ator head status panel . Al l
backlights are powered from
+12 V.
8 00118
Figure 2-68. BDU LED PCB
Figure 2-68. BDU LED PCB
PA
PI
8-00285
Oxygen
sensor
4-072214-00
Inspiratory
pressure transducer/
autozero solenoid
(SOL1)
4-072215-SP
Blindmate connector
P2 44 pin
Air flow
sensor (Q2)
4-072211-SP Oxygen flow
sensor (Q1)
4-072211-SP
Oxygen flow
sensor
Air flow
cable assy
sensor
cable assy 4-071322-00
4-071321-00
Press switch
harness assy
4-071326-00
INSPIRATORY M ODULE
8-00123
Figure 2-70. 840 Ventilat or System int erconnect diagram – Inspirat ory module
2-82 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.4.11 Exhalation t ransducer PCB
The ex hal ation transducer PCB (Figure 2-71 an d Figure 2-72), which i s housed i n the
exh alation m odule, produces a voltage si gn al represen ti ng the pati ent pressure in th e
exh alati on circuit. Th is sign al is routed to the AI PCB and used by software. The expiratory
pressure transducer (PE), which is on th is PCB, senses th e pressure di fferen ce in the
exh alation circui t rel ative to am bi ent air pressure. The PCB uses a +10 V reference.
8-00296
Expiratory
pressure transducer/
autozero solenoid (SOL2)
Exhalation
heater assembly
Exhalation
flow sensor (Q3)
To
Exhalation motherboard
module Exhalation I/ O
cable assy cable assy
P1 J2
J13
Expiratory
Exhalation pressure transducer
valve (PE)
Exhalation
transducer PCB
8-00126
EXHALATION M ODULE
Figure 2-72. 840 Vent ilat or System int erconnect diagram – Exhalation module
2-84 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.4.11.1 BD (cont inuous-t one) alarm assembly
The BD (con tin uous-tone) alarm assem bl y i s sh own i n Fi gure 2-73. It i s activated under the
conditi on s described i n Section 2.4.6.2.
Under m ost conditions, the +5 V powers the al arm. If the +5 V sentry circui t on the AI PCB
detects a power fai lure, a power fail capacitor powers the al arm for a m i nim um of 2 m inutes.
BD alarm
assembly
8-00173
8-01257
2.4.12.1 806 compressor unit ac power dist ribut ion component s and mot or
M ain s power i s appl ied to th e com pressor receptacle J4 through power rel ay K1 wh en the
power suppl y i s on .
The ac power to the com pressor un it i s provi ded by a field-replaceabl e power cord plugged
i nto receptacle J4. Th e ac-powered fans (Fi gure 2-75), an d m otor receive power vi a th e
com pressor PCB and are protected from overtem perature con diti on s by a th erm ostat (TH1).
The fans operate whenever th e com pressor is on . Ci rcui t breaker, CB2, protects the
com pressor from over-current condi tions.
The m otor compressor uses a 25 µF (100 V uni ts), 15 µF (120 V un its), or a 12.5 µF
(220 – 240 V, 50/60 Hz units) starting capacitor.
8-01256
Data
cable
Pressure Pow er
transducer cord
hose
Solenoid
Compressor cable
cable
Fan
cables
Ground
8-01272
2-86 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
The PCB includes th ese ci rcuits, wh ich perform the in dicated fun ctions:
• The fla sh m em ory circu it s include a flash m emory devi ce, an el apsed ti me counter, and
ci rcui ts to perm it seri al data tran sfer between the AI PCB and the com pressor PCB. Th e
coun ter moni tors (in seconds) com pressor m otor operation. The counter value is
periodical ly read by th e ven tilator, w hich th en stores th e total el apsed ti m e (in hours) i n
com pressor flash m em ory. During a power-on reset, the ventilator l oads the current
com pressor fl ash m em ory value i nto the coun ter. The flash m em ory device al so contai ns
other com pressor-speci fic i nform ation, including th e com pressor i dentifi cation (ID).
• The m ot or st a rt -u p circu it i ncl udes a relay, ti mers, and l ogi c that togeth er orch estrate th e
m otor start-up sequen ce. Th is sequence i nvol ves energizi ng SOL3, which vents
com pressor output, and then energizin g th e relay to apply power to the com pressor.
(Venting com pressor output m om entaril y reduces the l oad on the com pressor.)
• The com pressor pressu re t r a n sd u cer (PC) circu it m onitors accum ul ator pressure vi a
several com parators in th e ci rcuit. W hen a com parator detects an accumulator pressure
equal to or greater than predefi ned trip poin ts, i t si gnals the com pressor control l ogic.
• The com pressor con t rol logic controls the m otor, fans, and SOL3. It also provi des
com pressor status signals. Secti on 2.4.12.3 describes how th is l ogic controls com pressor
com ponen t operation. This logic sh uts off th e com pressor wh en ac i s inadequate or the
therm ostat on the PCB detects overtem perature.
J1
12 V COMPS
UNLOADING
SOLENOID SOLENOID
CKT (SOL3)
J2
FAN1
J6
FROM
ac PANEL J3
FAN2
J4
MOTOR
COMP
COMPRESSOR
CONTROL
HOUSING
CKTS
TO/ FROM
AI PANEL
EEPROM
CKTS
COMP OVERTEMP
+5 V COMP
THERMOSTAT
+12 V COMP
806compbd
2-88 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
To AI PCB
M otor mounting
plate
BDU compressor
dc cable
4-074223-00
Compressor
motor
Top view of compressor
Compressor base
plate
GRN/ YEL
Compressor PCB
4-075802-SP
Unloading
solenoid
J5 J6
(SOL3)
J4
J3 J2
J1
Hose to BDU
To ac panel
806conn1
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 2-89
2 Theory of operation
2-90 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
27 psig
Accumulator
pressure 22.5 psig
Compressor
motor
27 psig
Accumulator
pressure
22.5 psig
SOL3
Compressor
motor
Breath
interval 8-00195
1 3
SOL3
(Dump compressor
head pressure)
1.0 s 0.5 s
M otor relay
8-00191
NOTE:
• Inspiratory and expiratory flows and tidal volumes in the ventilator are compliance-
compensated and corrected to body temperature and pressure, saturated (BTPS).
• Exhaled volumes delivered/ measured by the ventilator meet specified accuracies
when conventional humidification, heated wire systems, and heat-moisture
exchangers (HM Es) are used and SST is successfully completed.
2.5.1 Inspiration
Duri ng in spiration (Fi gure 2-81), the exh al ation valve (EV) i s en ergized cl osed and gas flows
to the patien t.
An in spiration i s triggered when any of the fol lowin g happen s:
• The venti lator senses patient in spi ratory effort (pressure or fl ow triggeri ng).
• The venti lator’s breath tim in g dictates.
• The operator presses the M ANUAL INSP key.
• The venti lator’s softw are oth erw ise dictates (such as apn ea tim e-out).
NOTE:
The ventilator does not autocycle when pressure sensitivity is greater than 1
cmH2O or when flow sensitivity is greater than 1 L/ min for pediatric patients or
1.5 L/ min for adult patients.
2-92 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
C NC Patient Wye
SOL2
PE EXH
HTR NO
WT WT
CV5
F9
H P Oxygen, Wall
Q3 Reg Oxygen
H P Air, Wall/Comp
Humidifier
ECV Reg Air
Mixed Gas F8
Atmosphere
Vacuum
EV To Patient
Water
Vent
PSOL1
PS1 Q1
F7
F3
REG1
TP1 CV3
SV
F5
F1 OS
CV2
PS2
NO
Vent
Q2 SOL1
CV4 PSOL2
F2 REG2
F6
NC C
TP2
WT1 F4
PA PI
RV
M/C
HE PC
Accumulator
R1/F11
F13
SOL3
NC
F10
WT2 NO C
F12
Dryer
HB
i8-01288
2-94 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.5.2.1 Time-cycling met hod
Durin g pressure- and volum e-based m andatory breaths, the ti me-cyclin g m eth od operates.
This m ethod uses a specifi ed i nspiratory tim e to term inate i nspiration and transi tion to
exh alation. The ven tilator term inates inspiration based on the set or com puted val ue for
i nspiratory tim e.
C NC Patient Wye
SOL2
PE
EXH
HTR NO
WT WT
CV5
F9
H P Oxygen, Wall
Q3 Reg Oxygen
H P Air, Wall/Comp
Humidifier
ECV Reg Air
Mixed Gas F8
Atmosphere
Vacuum
EV To Patient
Water
Vent
PSOL1
PS1 Q1
F7
F3
REG1
TP1 CV3
SV
F5
F1 OS
CV2
PS2
NO
Vent
Q2 SOL1
CV4 PSOL2
F2 REG2
F6
NC C
TP2
WT1 F4
PA PI
RV
M/C
HE PC
Accumulator
R1/F11
F13
SOL3
NC
F10
WT2 NO C
F12
Dryer
HB
8-01285
2-96 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.6 Ot her hardw are operat ions
Breath type Indicates the type and phase of the Type: Control, assist, or
breath being delivered. spontaneous
Phase: Inspiration or
exhalation
Delivered O2% (O2%) Percentage of oxygen in the gas Oxygen sensor (OS) 0 to 103%
delivered to the patient. The high and
low O2% alarms are set internally and
are based on the set O2% value.
End expiratory pressure Pressure at end of previous breath’s Based on expiratory and -20.0 to 100 cmH2O
(PE END) expiratory phase. Reflects level of any inspiratory pressure
active lung PEEP. transducer (PE and PI)
readings
End inspiratory pressure Pressure at end of current breath’s Based on expiratory and -20.0 to 130 cmH2O
(PI END) inspiratory phase. If plateau is active, it inspiratory pressure
reflects level of end-plateau pressure. transducer (PE and PI)
readings
Exhaled minute volume Calculated total of exhaled volumes for Net flow, based on 0.00 to 99.9 L
.
(V E TOT) mandatory and spontaneous breaths exhalation (Q3), oxygen
for previous 1-minute interval. It is (Q1), and air (Q2) flow
compliance- and BTPS-compensated. sensor readings
Exhaled tidal volume Exhaled volume for previous Net flow, based on 0 to 6000 mL
(VTE) mandatory or spontaneous breath. It is exhalation (Q3), oxygen
compliance- and BTPS-compensated. (Q1), and air (Q2) flow
sensor readings
NOTE:
A significant change to the O2% setting can cause the VTE (exhaled tidal volume) to be
transiently displayed as lower or higher than the actual exhaled volume. This is a result of
initial spirometry calculations and does not reflect actual volume exhaled by the patient.
M ean airway pressure Average airway pressure over entire Based on expiratory and -20.0 to 120 cmH2O
(PMEAN) breath cycle of previous breath, inspiratory pressure
regardless of type. transducer (PE and PI)
readings
M aximum circuit Indicates the maximum pressure Based on expiratory and -20.0 to 130 cmH2O
pressure during the previous breath, relative to inspiratory pressure
(PPEAK) the patient wye, including the transducer (PE and PI)
inspiratory and expiratory phases. readings
Updated at the beginning of the next
inspiration.
Spontaneous minute Displays a calculated total of the Net flow, based on 0.00 to 99.9 L
volume (VE SPONT) volumes exhaled by the patient for exhalation (Q3), oxygen
spontaneous breaths for the previous (Q1), and air (Q2) flow
1-minute interval. Values for sensor readings
mandatory breaths during this period
are not included. The displayed value
is compliance- and BTPS-
compensated. Updated at the
beginning of the next inspiration.
Total respiratory rate Displays a calculated value of the 0 to 200 breaths/ min
(fTOT) number of mandatory and
spontaneous breaths delivered to the
patient for the previous 1-minute
interval. Updated at the beginning of
the next inspiration.
2-98 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
C NC Patient Wye
SOL2
PE
EXH
HTR NO
WT WT
CV5
F9
H P Oxygen, Wall
Q3 Reg Oxygen
H P Air, Wall/Comp
Humidifier
EVC Reg Air
Mixed Gas F8
Atmosphere
Vacuum
EV To Patient
Water
Vent
PSOL1
PS1 Q1
F7
F3
REG1
TP1 CV3
SV
F5
F1 OS
CV2
PS2
NO
Vent
Q2 SOL1
CV4 PSOL2
F2 REG2
F6
NC C
TP2
WT1 F4
PA PI
RV
M/C
HE PC
Accumulator
R1/F11
F13
SOL3
NC
F10
WT2 NO C
F12
Dryer
HB
8-01142
Figure 2-83. Pressure t ransducer aut ozero mode gas flow diagram
ac input < 80 V ac
level
31.50 V dc
Bulk 24 V dc
supply
24 V dc nominal 19.30 V dc
BPS or external
battery output
NMI
(generated by
power fail)
2-100 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.6.4.2 Supply volt age monit oring
The venti lator ch ecks for and respon ds to out-of-tolerance supply voltages, as foll ows:
Where
Volt age Source Not es
monit ored
GUI CPU PCB If voltage is low, this circuit issues a PCB reset. If
voltage is high, an NMI is generated.
+10 V reference AI PCB AI PCB Converted into a scaled voltage for reading by
software during background and other checks.
C NC Patient Wye
SOL2
PE
EXH
HTR NO
WT WT
CV5
F9
H P Oxygen, Wall
Q3 Reg Oxygen
H P Air, Wall/Comp
Humidifier
EVC Reg Air
Mixed Gas F8
Atmosphere
Vacuum
EV To Patient
Water
Vent
PSOL1
PS1 Q1
F7
F3
REG1
TP1 CV3
SV
F5
F1 OS
CV2
PS2
NO
Vent
Q2 SOL1
CV4 PSOL2
F2 REG2
F6
NC C
TP2
WT1 F4
PA PI
RV
M/C
HE PC
Accumulator
R1/F11
F13
SOL3
NC
F10
WT2 NO C
F12
Dryer
HB
8-01143
2-102 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Theory of operation 2
2.7.2 Occlusion handling
The ventil ator detects severe pati ent circui t occl usions to protect the patien t agai nst excessive
airway pressures over exten ded peri ods of ti m e. The venti lator is also designed to detect
patient circui t di scon nects, because they can cause the patien t to receive li ttl e or no gas from
the venti lator, and require i m mediate cli nical attenti on . Figure 2-86 depi cts the ven tilator
pneum ati cs during occlusion handl ing.
NOTE:
During a severe occlusion, apnea detection, expiratory pause, manual inspirations, and
maneuvers are suspended, and the ↑ PMEAN (high airway pressure) alarm limit is disabled.
The GUI does allow you to change ventilator settings.
C NC Patient Wye
SOL2
PE
EXH
HTR NO
WT WT
CV5
F9
H P Oxygen, Wall
Q3 Reg Oxygen
H P Air, Wall/Comp
Humidifier
EVC Reg Air
Mixed Gas F8
Atmosphere
Vacuum
EV To Patient
Water
Vent
PSOL1
PS1 Q1
F7
F3
REG1
TP1 CV3
SV
F5
F1 OS
CV2
PS2
NO
Vent
Q2 SOL1
CV4 PSOL2
F2 REG2
F6
NC C
TP2
WT1 F4
PA PI
RV
M/C
HE PC
Accumulator
R1/F11
F13
SOL3
NC
F10
WT2 NO C
F12
Dryer
HB
8-01144
2-104 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
SECTION
S EC T I O N
Self tests 3 3
3.1 Int roduct ion
This section describes the 840 Ven ti lator System buil t-in sel f tests, i ncl uding h ow to run
them. For a li sting of ventil ator diagnosti c codes an d messages generated duri ng these self
tests, refer to Section 6.
3.3.1 POST
POST tests th e i ntegrity of the ven tilator’s electroni cs with out operator i ntervention. It i s
executed when the ventil ator is powered up, before i t enters servi ce mode, or i f th e ventil ator
detects selected faul t conditions. Breath deli very cannot start until th e venti lator com pletes
POST with no m ajor errors.
3.3.2 SST
The SST is a short (about 3 mi nutes) and si m ple sequen ce of tests that verifies proper
operation of breath del ivery hardware (in cluding pressure and fl ow sensors), ch ecks the
patient circui t (i ncl udi ng tubin g, hum idi fi cation device, and fi lters) for leaks, and m easures
the ci rcui t com pl ian ce and resistan ce. SST also checks the exhal ation fil ter resistance. SST i s a
user-ini tiated ch eck prim aril y in tended for use by the operator. The ven ti lator does n ot begi n
SST i f it sen ses that a patien t is conn ected.
840 Ventilator System Service Manual 4-070496-00 Rev. A (08/ 03) 3-1
3 Self tests
3.3.3 EST
EST is a user-ini ti ated sel f test, intended to be run by the service tech ni cian, that verifies the
i ntegri ty of th e ventilator’s subsystems usin g operator parti cipation. EST ch ecks the
pn eum ati cs, m em ory, safety system , front panel con trols, i ndi cators, di gital an d anal og
electronics, power suppli es, tran sducers and opti on s. EST requi res a “ gold stan dard” test
ci rcuit, avail abl e from Puritan -Ben nett. All required software support to perform an EST i s
residen t on th e ven ti lator. EST testing, excludin g tests of opti on al equipm en t, such as the
com pressor, takes about 15 min utes to com pl ete.
SST A short sequence of tests intended to be run • Before patient is connected to ventilator
primarily by the operator. Checks pneumatics • After patient circuit or humidifier is changed
and electronics. SST also characterizes system • When ventilator is serviced
leaks and system/ tubing compliance to • Every 15 days
compensate during breath delivery. POST is run
as part of SST, both at the start and end.
1. POST will run only if there is no BPSinstalled, or if the BPS battery is discharged. Otherwise, during a temporary loss of ac power, the vent
will automatically switch to BPS power, and POST will not run.
3-2 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
Table 3-1: Self t ests (cont inued)
NOTE:
• SST is not run as part of EST. To
determine patient circuit resistance and
compliance, run SST.
• The performance verification,
described in Section 5 of this manual, is
a more thorough test of the ventilator
to verify specifications are met.
Ongoing Includes these checks: memory, power supply, Automatically, continually during ventilation
background checks dc voltages, pressure transducers, analog input
validity, and more. These checks do not disrupt
normal operation.
ac monitor voltage X
BPS model X X
Pneum at ics
System leak X
3-4 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
Table 3-2: Components test ed by self t ests (cont inued)
Elect ronics
Keyboard X partially X
Knob X
Ethernet controller X X
Safet y syst em
NOTE:
A fault identified in POST indicates that the ventilator or an associated component
is defective. A defective ventilator or associated component should be repaired
before the ventilator is returned to service, unless it can be determined with
certainty that the defect cannot create a hazard for the patient, or add to the risks
which may arise from other hazards.
There are tw o versions of POST: l on g POST, whi ch is run when the user cycl es power, and
short POST, wh ich is run fol lowin g an un intended h ardware reset.
NOTE:
The graphic user interface (GUI) indicates a POST is in progress but does not
indicate a long or short POST.
The graphic user interface (GUI) and the breath delivery unit (BDU) subsystems each has its own
POST that tests the major hardware electronics systems. POST does not check the ventilator’s
pneumatics, options, or accessories that are not directly related to ventilation. POST is designed to
detect major problems before proceeding to normal ventilation, and to provide a confidence
check before a patient is connected to the ventilator.
POST routin es are ordered so that each routi ne requires successively m ore operati onal
h ardw are than the last. Thi s sequence all ow s POST to system ati cal ly excl ude electronic
com pon en ts as causes of system m alfunctions.
3-6 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
3.4.1 Safety
The ventil ator does not provide ventil atory support to the pati ent duri ng POST. Th e
ven til ator alarm s if POST l asts lon ger than 10 secon ds or i f an unexpected fault i s detected.
POST i s designed to m i nim i ze the delay until n orm al venti lati on begi ns and to provi de
i mm ediate notification in case a fault is detected.
Durin g POST the ven tilator proportional sol en oid valves (PSOLs) are closed and the
exh alati on valve an d safety valve are open to al low the patien t to breathe room air, and the
ven til ator displays a m essage that POST i s in progress.
Once POST i s complete, ventil ator startup (foll owi ng power-up) or norm al ven ti lati on begin s,
unless service m ode i s requested or the ventilator detects any of the fol lowin g:
• An un corrected m ajor system fault.
• An un corrected m aj or POST fault.
• An un corrected sh ort self test (SST) fail ure or n on -overri dden SST al ert.
• An un corrected exten ded self test (EST) fai lure or non-overridden EST al ert.
• The ven ti lator i s turned on for the first ti me fol lowin g a software down load, but has not
yet successful ly com pleted one of th e fol lowing: exhal ation valve cali bration, flow sen sor
calibration, atm ospheri c pressure transducer cal ibrati on , SST, or EST.
• An un com pl eted system in itiali zation .
NOTE:
When a compressor is installed and wall air is not present, there may be a short
interval following a successful POST before the compressor achieves operational
pressures. If so, the ventilator annunciates a NO AIRSUPPLY alarm, which resets as
soon as the compressor charges the system to operational pressure.
To ensure that th ere is an alarm i f th e central processin g uni t (CPU) fails, audi o, visual , an d
rem ote alarm s are norm all y on, and turn off once system in iti ali zation (that is, the process
that occurs betw een POST com pl eti on and the start of ventil ation) is com pl eted and
com m un ication is establ ished.
An alarm turns on i f POST l asts m ore than 10 seconds or if POST restarts three ti mes without
com pl eti on. Th e 10-secon d ti m er is a redundant check i n case POST fail s to al arm upon
detectin g a faul t. The ch eck for three restarts can detect a con ti nuous l oop, and prevents
POST from run nin g for m ore th an 10 seconds.
NOTE:
Puritan Bennett recommends that a BPS is always installed on the ventilator. If
there is a loss of ac power with a charged BPS installed, the ventilator will switch
to the dc source and will not run POST upon restoration of ac power.
3-8 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
3.4.5 St ruct ure of POST
W hen POST i s started, two “ POSTs” actually run si m ultaneously: the graph ic user in terface
(GUI) an d the breath del ivery un it (BDU) POSTs. These two POSTs run i ndependen tl y of each
other.
The fol lowin g phases compri se th e GUI POST and BDU POST:
Ph a se 1 (k ern el) - Tests the m i crocon troll ers and associated circui try. For each POST, thi s
secti on of code resi des in a sin gl e PROM al ong with kernel operatin g system and
download util ity. Th is part of POST is not down loadabl e. It is i dentical for both th e GUI
an d BDU POSTs, wi th the exception of the roll in g th un der test, wh ich i s confi ned to th e
BDU CPU.
Ph a se 2 - Tests com ponen ts outside of the kern el. Thi s portion of POST differs for the GUI
an d BDU subsystem s.
Ph a se 3 - Consi sts of tests com m on to both CPUs, and those th at depen d on the
venti lator operating system to i niti ali ze som e h ardware com ponents before th e test.
Table 3-3 com pares the BDU and GUI POST sequen ces, i ndi cates the tests perform ed an d
shows which indicators are active durin g each phase.
Sympt oms
How t o respond
What you see and hear What is happening
• POST continues to end. • Previous run of POST failed • Run EST/ service ventilator.
• Ventilator inoperative alarm and was not corrected,
(audible and visible). although POST passed
• Ventilation disallowed. subsequent run.
• Ventilator inoperative
condition declared.
• POST continues to end. • M inor fault detected. • Check System Diagnostic Log
• Ventilation proceeds. for any associated error codes.
• A DEVICE ALERT is • Rerun POST/ EST.
annunciated.
GUI CPU POST • Initializes processor • Tests GUI on-board • Tests GUI CPU • Updates NOVRAM
• Tests power, GUI electronics, GUI 5V memory manage- • Initializes hardware
CPU, time of day sentry, GUI 12V ment and floating
clock, kernel sentry, POST point units, down-
DRAM/ NOVRAM, NOVRAM, bus load OS boot,
boot PROM , timer, DRAM , application boot
interrupt timer, SAAS, Ethernet
watchdog timer, controller,
flash memory unexpected reset
checksum, umpire test
EEPROM checksum
No communication occurs between GUI and • GUI and BDU CPUs • Ventilator begins
BDU CPU boards during Phases 1 and 2 establish flowing air at
communication 10 L/ min to check
for patient
connection
BDU CPU POST • Initializes processor • Tests BDU on- • Tests BDU CPU • Initializes hardware
• Tests power, BDU board electronics, memory manage-
CPU, time of day VH 5V, VH 12V, ment and floating
clock, kernel POST NOVRAM, point units, down-
DRAM/ NOVRAM, bus timer, DRAM , load OS boot,
boot PROM , vent INOP, application boot
interrupt timer, Ethernet controller,
watchdog timer, service mode
flash memory switch, unexpected
checksum, reset umpire test,
EEPROM • AI PCB test
checksum, rolling • Analog devices
thunder test test:
PS1/ PS2
Q1/ Q2/ Q3
PSOL1/ PSOL2
PA/ PI/ PE
O2 sensor
safety valve
exhalation valve
exh. valve heater
• ADC and DAC tests
• ac voltage test
Visual and • GUI LEDs: SVO LED • GUI LEDs: all LEDs • GUI LEDs: all OFF • GUI: start-up
audible indicat ors ON ON except green screen displays if
• BDU LEDs: SVO • BDU LEDs: all LEDs “ normal” indicator POST passes
LED ON, loss of UI ON • BDU LEDs: all OFF • BDU: all LEDs off if
LED ON • BDU continuous • GUI audible alarm POST passes
• BDU continuous tone alarm OFF turned on then off • No audible alarms/
tone alarm ON (2 beeps if OK) LEDs active1
• BDU continuous
tone alarm turned
on and off twice
AI PCB LEDs Red LED ON Red LED ON Green LED ON Green LED flashing
1. This is true only if both air and O 2 supplies are connected and have adequate pressure, and the patient circuit is not connected, and no
failures or fault conditions are detected by ongoing diagnostics (including POST 10 second timer).
3-10 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
3.5 SST (short self t est )
Warning
• Always disconnect the ventilator from the patient before running SST. Running SST while
the ventilator is connected to the patient can injure the patient
• When running SST, make sure the patient circuit is configured exactly as it will be used on
the patient (for example, with same accessories). If accessories are changed or added to
the patient circuit after SST has completed, you must run SST again with the new
configuration installed to calculate for compliance and check for leaks.
• An ALERT identified in SST indicates that the ventilator or an associated component is
defective. A defective ventilator or associated component should be repaired before the
ventilator is returned to service, unless it can be determined with certainty that the defect
cannot create a hazard for the patient, or add to the risks which may arise from other
hazards.
NOTE:
• If the ventilator has not reached operating temperature from recent usage, allow it to
warm up for at least 10 minutes in service mode before running SST, to ensure accurate
testing.
• Puritan Bennett recommends that you run SST every 15 days, between patients, after a
major service or repair (refer to Table 5-2 on page 6), and when you change the patient
circuit. Puritan Bennett recognizes that the protocol for running SST varies widely among
health care institutions. It is not possible for Puritan Bennett to specify or require specific
practices that will meet all needs, or to be responsible for the effectiveness of those
practices.
•
Table 3-6 l ists the tests that com prise SST. For m ore detail s about these tests and associ ated
diagnosti c codes, see Secti on 6 of th is m anual. For a theory of operation of SST, consult th e
840 Ventilator System Operator’s and Technical Reference Manual.
NOTE:
To ensure that compliance compensation
functions correctly, the user must run SST
with the circuit configured as intended for
use on the patient.
NOTE:
Due to the excessive restriction of Air
Liquide, Dräger, and SIS air/ oxygen hose
assemblies, certain SST tests may fail when
supply pressures< 50 psi (345 kPa) are used
in conjunction with these hoses.
NOTE:
Due to the excessive restriction of Air
Liquide, Dräger, and SIS air/ oxygen hose
assemblies, certain SST tests may fail when
supply pressures< 50 psi (345 kPa) are used
in conjunction with these hoses.
1. If necessary, the compressor can be used as the only gas source. However, only the gas pathway, flow sensor, and PSOL
of the single gas (air or O2) will be verified during SST.
3-12 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
3.5.3 Running SST
1 If attached, rem ove gold stan dard circuit (test circui t desi gn ed for use with EST) from
venti lator. Instal l th e circuit to be used on patien t. (Fi gure 3-1).
Do n ot attach a test l ung or a humi di fi er to the circuit at thi s ti m e.
NOTE:
To ensure accurate circuit resistance measurement, check that the circuit is not
obstructed and is properly connected to the ventilator.
2 En sure that pati ent i s n ot con nected to circuit and that patient w ye i s unblocked.
3 Turn on power switch to start norm al ventil ation.
4 Verify that the ventil ator is operati ng on full ac power before runni ng SST. If ful l ac
power is n ot present, test fai lures may result.
5 At Ventilator Startup screen (lower GUI screen ), touch SST, then press TEST button (on
si de of ven ti lator) wi thi n 5 seconds. Wai ting l on ger than 5 secon ds can cels th e SST
prom pt.
Tubing configured
for use on patient
6 At SST Setup screen (l ower GUI screen), select patient circui t an d h um idification types,
then press ACCEPT.
Warning
Incorrectly specifying the patient circuit type or humidifier during SST setup, or changing
either type after you have run SST can affect the accuracy of compliance calculation and the
delivered and measured exhaled tidal volumes. You must rerun SST after changing the
patient circuit type, installing a new patient circuit, or after adding or changing accessories.
NOTE:
During ventilation, the humidification type can be changed by touching the
OTHER SCREENS button, then the More Settings button. However, the preferred
operation in this case is to rerun SST after making such a change.
7 The venti lator autom ati cal l y i niti ates th e test sequence (Table 3-6).
In st a lla t ion of h u m id ifier: Th e SST Fl ow Sensor test wi ll pause durin g executi on and
prom pt for i nstallation of humi difi er. If appl icabl e, fil l the hum idi fier with water and
connect as shown in Fi gure 3-1. (Th e hum idi fi er m ust be fi lled with water to ensure
proper com pliance com pensation.) Press ACCEPT to con ti nue SST.
The SST Expi ratory Fi lter, Circuit Resi stan ce, and Compli ance Cal ibrati on tests requi re
your i ntervention , and wil l wait indefi nitely for your response. Respond when
prom pted; then press ACCEPT. Otherwise you don’t need to do anythi ng until a test
result is ALERT or FAILURE, or SST is com plete.
8 As each test is perform ed, the SST Status screen shows test results (see Table 3-7). SST
results can be viewed in servi ce m ode in th e SST Results subscreen . You can also review
resul tin g error i nform ati on th rough the EST/SST Diagnostic Log after SST is compl eted
(refer to Section 6).
9 You can touch EXIT SST during SST to halt testin g. You can touch EXIT SST again to
resum e testin g, or press ACCEPT to restart the ventilator (if SST h as not detected an
ALERT or FAILURE).
Warning
• To ensure ventilation that correctly compensates for circuit resistance and compliance, do
not exit SST and begin normal ventilation until the ent ire SST has been successfully
completed with the circuit to be used on the patient installed.
• To ensure reliable SST results, do not repeat an individual test with a different patient
circuit if the test result is FAILURE or ALERT. If you suspect a defective patient circuit,
restart SST from the beginning with a different patient circuit.
NOTE:
• If SST is interrupted, you can restart normal ventilation as long as no errors have occurred
that would otherwise prevent it.
• Tests that result in an alert or failure and are successfully rerun are not written into the
EST/ SST Diagnostic Log if they are successfully rerun during the same SST test sequence.
10 W h en all of the tests in SST are com plete, th e SST Status screen di spl ays all in divi dual test
results an d SST outcome. Table 3-8 sum m arizes overal l SST outcom es and how to
proceed in each case.
11 To begi n norm al ven til ati on (if SST has not detected an ALERT or FAILURE), touch EX IT
SST, th en press ACCEPT. The ventil ator rerun s POST, th en displays th e Ventilator Startup
screen.
3-14 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
SST Flow Sensor Test 1. Verifies that at least one gas is con- When prompted, do the following:
nected. (If only one gas is connected • M ake sure inspiratory filter is installed.
then only that gas pathway (flow sen- • Connect patient circuit with inspiratory
sor and PSOL) is verified during SST.) filter, but without the humidifier.
2. Cross-checks air and oxygen flow sen- • Block wye.
sors against exhalation flow sensor at • When later prompted, and If applicable,
four different gas flows. install the humidifier with jar filled with
3. Verifies that proportional solenoid water.
(PSOL) command current is in range
for each test flow.
NOTE:
To ensure proper compliance
compensation when using a
humidifier, make sure the jar is full of
water.
Circuit Pressure Test 1. Verifies that system is running on ac. FAILURE if not passed (cannot be
2. Exercises air and oxygen PSOLs. overridden).
3. Checks inspiration and exhalation
autozero solenoids.
4. Cross-checks inspiration and exhala-
tion pressure transducers at various
pressures.
Circuit leak Displays drop in circuit pressure in 10 s. Overriding an ALERT could cause improper
(Determines ability of circuit to hold compliance compensation, inaccurate tidal
pressure.) volume delivery, or autocycling.
FAILURE reported if test detects excessive
leak.
Expiratory filter Displays pressure drop across expiratory Connect and disconnect tubing when
filter. prompted.
Overriding an ALERT could cause inaccurate
patient pressure estimation.
FAILURE if test detects exhalation
compartment occlusion, expiratory filter
occlusion or damage, or you did not follow
prompts to detach and reattach tubing
correctly.
Circuit Resistance Displays pressure drop across inspiratory Unblock wye when prompted.
and expiratory limbs, including effect of all Overriding an ALERT could cause inaccurate
devices on each limb (filters, humidifier, patient pressure estimation.
water traps). FAILURE if test detects excessive or low limb
resistance, or if you did not follow the
prompt to unblock the wye.
Compliance calibration Displays patient circuit compliance. Block and unblock wye when prompted.
Overriding an ALERT could cause improper
compliance compensation or inaccurate
tidal volume delivery. FAILURE if test detects
out of range compliance.
ALERT Test results not ideal, but not Repeat t est by touching REPEAT.
critical. SST halts. Skip t o next t est by touching NEXT.
Repeat SST from t he beginning by touching RESTART SST.
Exit SST in order to service ventilator by touching EXIT SST.
FAILURE A critical problem has been Repeat failed t est by touching REPEAT.
detected, and SST cannot complete Repeat SST from t he beginning by touching RESTART SST.
until the ventilator passes the failed Exit SST in order to service ventilator by touching EXIT SST.
test.
If t he SST
It means: Do t his:
out come is:
ALERT One or more faults were detected. Repeat SST from t he beginning by touching RESTART SST.
If it can be determined with Override ALERT, as allowed by your institution’s protocol, by
certainty that this cannot create a touching OVERRIDE, followed by EXIT SST.
hazard for the patient, or add to
the risks which may arise from
other hazards, the user can choose
to override the ALERT status and
authorize ventilation.
FAILURE One or more critical faults were Repeat SST from t he beginning by pressing RESTART SST.
detected. A ventilator inoperative Exit SST in order to service ventilator or review error codes by
condition is declared. pressing EXIT SST.
3-16 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
3.6 EST (ext ended self t est )
Warning
• Always disconnect the ventilator from the patient before running EST. Running EST while
the ventilator is connected to the patient can injure the patient.
• A fault identified in EST indicates that the ventilator or an associated component is
defective. A defective ventilator or associated component should be repaired before the
ventilator is returned to service, unless it can be determined with certainty that the defect
cannot create a hazard for the patient, or add to the risks which may arise from other
hazards.
NOTE:
• If the ventilator has not reached operating temperature from recent usage, allow it to
warm up for at least 10 minutes in service mode before running EST to ensure accurate
testing.
• Be aware that each time EST is performed, the patient alarm log is cleared.
Compressed air source, 241 to 690 kPa (35 to 100 Local supplier
psi) or compressor option
NOTE:
Due to excessive restriction of Air Liquide,
Dräger, and SISair/oxygen hose assemblies,
certain EST tests may fail when supply
pressures < 50 psi (345 kPa) are used in
conjunction with these hoses.
Oxygen source, 241 to 690 kPa (35 to 100 psi) Local supplier
NOTE:
Due to excessive restriction of Air Liquide,
Dräger, and SISair/oxygen hose assemblies,
certain EST tests may fail when supply
pressures < 50 psi (345 kPa) are used in
conjunction with these hoses.
3-18 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
Gold
standard
tube
8-01211
NOTE:
• If EST is interrupted, you can restart normal ventilation as long as no errors have occurred
that would otherwise prevent it. You must press the EXIT button on the GUI screen then
the ACCEPT key to exit EST properly.
• Tests that result in an alert or failure and are successfully rerun during the same test
sequence are not written into the EST/ SST Diagnostic Log.
• It may be useful to complete EST even with errors, because information on multiple errors
can facilitate troubleshooting.
7 W h en all of the tests i n EST are com plete, the Extended Self Test screen displays al l
i ndi vidual test results and EST outcom e. Refer to Tables Tabl e 3-10, Table 3-11, and
Tabl e 3-12 for addi tional inform ation.
18 Jul
18 Jul 03
03
NOTE:
These screens are examples only. Your unit may vary slightly in appearance.
3-20 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
Circuit Pressure Test Checks that the system is running on ac power, prompts • Leaks or occlusions in the
user to set up the vent for EST, and exercises the PSOLs. expiratory filter or elsewhere will
Calibrates inspiratory and exhalation transducers to cause this test to fail.
atmospheric pressure. • Connect ac power if ventilator is
running on battery power and ac
Cross-checks inspiration and exhalation pressure is not connected.
transducers at various pressures.
• M ake sure green BPS ready
Checks inspiration and exhalation autozero solenoids. indicator is lit on GUI LED panel.
• M ake sure air and oxygen are
connected.
• Remove inspiratory filter and
connect gold standard circuit as
directed.
Flow Sensors Cross Controla los sensores de flujo de aire y oxígeno contra • Leaks or occlusions in the
Check Test el sensor de flujo de exhalación en cinco flujos de gas expiratory filter or elsewhere will
diferentes. Verifica que la corriente de comando del cause this test to fail.
solenoide proporcional (PSOL) esté dentro del rango • Low source gas pressures may also
para cada flujo de prueba. Realiza aire / O 2 Calibración cause this test to fail.
de corriente de despegue PSOL. • Exhalation flow sensor has
separate air and O2 calibration
tables. During EST, the system
uses the calibration table(s)
appropriate for the source gas(es)
used during EST.
Gas Supply/ SV Test Tests air and oxygen pressure switch gross functionality, • Leaks or occlusions will cause this
tests PSOL forward leak, tests safety valve cracking test to fail.
pressure, and tests flow sensor zero offsets. • Block and unblock To patient port
as directed.
• Disconnect and reconnect air and
oxygen as directed.
SM Leak Test Checks for system leaks by pressurizing the system and • Leaks or occlusions in the
verifying that pressure drops by no more than 5 cmH2O expiratory filter or elsewhere will
in 10 s. cause this test to fail.
GUI Keyboard Test Verifies that GUI keys are functioning by prompting the Press keys within 15 s when
operator to press a key. prompted.
GUI Knob Test Verifies GUI knob rotation direction by prompting the Turn GUI knob as prompted, within
operator to rotate the knob in one direction, then the 15 s.
other.
GUI Lamp Test Verifies that GUI LEDs are functioning by prompting the Observe GUI indicators and press
operator to acknowledge that the LEDs are turned on. ACCEPT (if on) or CLEAR (if not on).
BDU Lamp Test Verifies that BDU lights are functioning. Observe BDU indicators and press
ACCEPT (if on) or CLEAR (if not on).
NOTE:
The ventilator inoperative and safety valve open
LEDs on the GUI are hard-wired to the BDU and
are considered BDU LEDs.
GUI Audio Test Verifies that GUI audio is functioning by prompting the Listen for GUI audio. Press ACCEPT (if
operator to listen for the alarm sound. on) or CLEAR (if not on).
GUI Nurse Call Verifies that nurse’s call relay is functioning. If not testing nurse’s call, press
CLEAR. If testing nurse’s call, press
ACCEPT, then verify nurse’s call
operation as prompted.
BDU Audio Test Verifies that BDU audio is functioning. Also verifies that Listen for BDU audio alarm.
power fail capacitor can operate loss-of-power alarm for
at least 2 minutes.
PSOL Loopback Test Checks PSOL loopback and command currents at three
different points by comparing the difference between
the two currents with the maximum allowable
difference.
Safety System Test Tests safety valve and inspiratory check valve operation
by checking for occlusions when the safety valve is open,
checking safety valve loopback current, and checking
inspiratory check valve reverse flow.
Exp Valve Loopback Test Checks exhalation valve loopback and command current
at three different points by comparing the difference
between the two currents with the maximum allowable
difference.
Exp Valve Pressure Verifies that exhalation valve seal and poppet friction are
Accuracy Test acceptable.
Exp Valve Test Verifies that current versus pressure values in flash
memory correspond with actual installed exhalation
valve.
Exp Heater Test Verifies that gas flow temperature increases and
decreases in response to commanding the heater on and
off.
Compressor Test Tests compressor operation. This test checks the timer, • If wall air is connected, disconnect
compressor pressure switch, and compressor operating wall air as directed.
states (run, standby, and disabled). • If ac is not connected, connect ac
power as directed.
Compressor Load Test Tests compressor operation. This test checks the
compressor’s ability to maintain sufficient air pressure
during worst-case breath delivery (200 lpm).
Compressor Leak Test Verifies that a fully charged compressor (if installed) does • Disconnect and connect wall air as
not leak enough to cause the compressor to turn on directed.
within 1 minute while in standby mode with wall air • Test takes 1 minutes to complete.
connected.
Analog Data Display Reads and displays analog data channels in engineering O2 sensor data is displayed in DAC
units. counts.
GUI Serial Port Test Verifies that GUI serial port can send and receive data.
Battery Test With ac disconnected, tests BPS under discharging and Test takes 5 minutes to complete.
charging conditions.
3-22 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Self tests 3
If t he t est result
It means: Do this:
is:
ALERT Test results not ideal, but not Repeat t est by touching REPEAT.
critical. EST halts. Skip t o next t est by touching NEXT.
Repeat EST from the beginning by touching RESTART EST.
Exit EST in order to service ventilator or review error codes by
touching EXIT EST.
FAILURE A critical problem has been Repita la prueba fallida tocando REPEAT.Skip para la
detected, and EST cannot siguiente prueba tocando NEXT.Repita EST desde el
complete until the ventilator passes principio tocando RESTART EST. Salga de EST para reparar
the failed test.
el ventilador o revisar los códigos de error por
tocando EXITEST.
If t he EST
It means: Do t his:
out come is:
ALERT One or more faults were detected. Repeat EST from t he beginning by touching RESTART EST.
Override t he alert in order to service ventilator or review error
codes by touching OVERRIDE, followed by EXIT EST.
NOTE:
Although you can override an alert in EST, normal
ventilation is not allowed until the problem is corrected
and all EST tests passed.
FAILURE One or more critical faults were Repeat EST from t he beginning by touching RESTART EST.
detected. A ventilator inoperative Exit EST in order to service ventilator or review error codes by
condition is declared. touching EXIT EST.
3-24 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
SECTION
S EC T I O N
Service mode 4 4
The 840 Ven til ator System operates in two m odes: pati ent ventil ation and service m ode. The
service m ode i s i ntended for use by a trai ned service techn ici an to aid in ven til ator testing
and troublesh ootin g an d to perform system cal ibrations.
This section detai ls th e service m ode, includin g prom pts an d expected system responses.
Warning
Do not enter the service mode while a patient or other person is connected to the ventilator.
To do so may cause patient injury or death. Normal ventilator operation is not accessible
while in the service mode.
NOTE:
During service mode startup, certain functions may automatically execute to
program data into BDU and/ or GUI NOVRAM . Refer to Section 4.2.16 for details.
While in the Service Mode, normal ventilation is not allowed.
840 Ventilator System Service Manual 4-070496-00 Rev. A (08/ 03) 4-1
4 Service mode
4-2 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service mode 4
4.2 Service mode funct ions
Table 4-1 sum m arizes the ven tilator’s servi ce m ode functions. Fi gure 4-2 shows h ow to
n avi gate amon g th e functi on s.
It is im portant to al ways run th e fi rst four EST tests before perform ing any ventil ator sen sor
cali brati ons. Al l system l eaks m ust be el im inated before proceedi ng wi th any cal ibrati on s.
Upper screen
SST RESULT Displays the latest SST test results, including test name, date and time
each test was last run, individual test results, and overall test outcome.
NOTE:
When reviewing SST results, be aware that not all of the results
for each SST test may be from the same date and time run.
DIAG LOG Displays the contents of System Diagnostic Log, System Information
Log, and EST/ SST Diagnostic Log.
Alarm log Displays the history of most recent ventilator alarms for the previous
patient.
NOTE:
Running EST or selecting New Patient will erase this log.
VENT CONFIG Displays software revisions and serial numbers. Software revisions
displayed include: GUI and BDU boot PROM and application software
as well as GUI audible alarm subsystem (SAAS). Serial numbers
displayed are for BDU, GUI, and compressor unit (if installed); these
serial numbers are read from the data key and compressor (if
installed).
TEST SUM M ARY Displays the latest EST and SST test results, including date and time
each was last run and overall test outcome.
Low er screen
EST Displays the details of the last EST run. Runs EST to thoroughly test the
operational integrity of the ventilator, both electronics and
pneumatics.
UPPER SCREEN
Other
EST DATE/ TIM E EXIT Screens
At mospheric Serial
Service External Exp Valve Vent Inop Flow Sensor Pressure Datakey Loopback
M ode Setup Test Control Calibration Test Calibration Transducer Update Test
Calibration (10.4 GUI only)
8-00210
LOWER SCREEN
4-4 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service mode 4
4.2.1 SST RESULT: Displaying SST result s
This functi on lets you read the l atest SST test resul ts, includin g: test nam e, date an d ti m e each
test w as last run, i ndi vidual test results, an d overal l test outcom e.
NOTE:
SST lets you exit and resume patient ventilation if required. Thus, the results shown are not
necessarily all from the same SST run. If a partial SST run is performed, the results of the first few
tests might be more recent than the results of later tests.
NOTE:
System Diagnostic Information and EST/ SST Diagnostic logs can only be erased by
using the 840 VTS Breath Labs Software. The alarm log is automatically erased
each time a new patient is setup is selected and when EST is run.
This functi on lets you read the contents of th e venti lator’s error and status logs. The l ogs are
as fol lows:
• The Syst em D ia gn ost ic Log stores data on unexpected conditions detected duri ng POST,
backgroun d checks. This in cl udes alerts an d fai lures.
• The Syst em In form a t io n Lo g (Fi gure 4-3) stores data on venti lator even ts, in cludin g
detail s on GUI/ BDU com m un icati on .
• The EST/ SST D ia gn o stic Lo g (Figure 4-4) stores data on overall test outcomes and even ts
durin g SST an d EST.
03
03
03
03
03
03
03
03
03
03
03
03
03
4-6 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service mode 4
4.2.3 Alarm log: Displaying t he alarm history
This functi on l ets you read the history of th e m ost recen t venti lator al arm events. Thi s
i ncludes such i nform ati on as autoreset, user reset, alarm si lence, and chan ges in urgency. Th e
alarm log is autom ati cal ly cl eared when a new patient setup i s com pl eted. W hen you exit th e
service m ode to start ventil ation, a new pati ent setup is required.
See Section 7 to in terpret the events in the alarm l og.
4.2.4 VENT CONFIG: Displaying soft ware revisions and serial numbers
This functi on di spl ays software revisions an d seri al num bers. Software revisions di spl ayed
i nclude: GUI an d BDU boot PROM an d applicati on software as well as GUI audible al arm
subsystem (SAAS). Serial num bers di splayed are for BDU, GUI, and com pressor unit (if
i nstalled); these serial numbers are read from the data key.
4.2.5 OPERATION TIM E: Displaying vent ilat or and compressor elapsed hours
This functi on displ ays accum ul ated venti lator and com pressor operati onal tim e, i n hours.
This h el ps determ i ne wh en preventive m ai ntenance is needed.
4.2.6 TEST SUM M ARY: Displaying results from EST and SST
This functi on displays the latest EST and SST overall test outcom es, includin g date and tim e
each was l ast run.
NOTE:
The performance verification, described in Section 5 of this manual, is a more
thorough test of the ventilator to verify it meets specifications.
Serial Loopback Test (10.4-inch Tests the serial communication ports on the 10.4-inch GUI.
GUI only)
Dat akey Updat e Transfers the operational hours from the ventilator to the new
datakey. For use after software options have been installed.
Service M ode Set up Lets you adjust nominal line voltage lower limit, baud rate for RS-232
port, and lets you choose the unit for display of pressures (cmH2O or
hPa).
Ext ernal Test Cont rol Lets you test the ventilator using a computer, 840 VTS Breath Labs
Software, and PTS 2000 tester, to run performance verification. Also
used to download new application software to the 840.
Exp Valve Calibrat ion Performs calibration. For use after a new exhalation valve, AI PCB, or
BDU CPU PCB is installed, or as part of the performance verification.
Vent Inop Test Exercises ventilator inoperative circuitry during five test phases.
Flow Sensor Calibrat ion Performs calibration. For use after a flow sensor replacement or as part
of the performance verification.
Nominal Line Volt age Nominal line voltage lower limit. This setting is used to detect low
power alarm conditions. It has no direct effect on power supply
function. Improper line voltage settings will cause a Device Alert,
preventing normal operation.
Ext ernal Cont rol Baud Rat e For GUI RS-232 port. The baud rate takes effect when you restart the
ventilator in the service mode.
4.2.12 Ext ernal Test Cont rol: Performing remot e vent ilat or t esting
This functi on l ets you test th e ven tilator usin g a computer to conduct the ventilator
perform ance verifi cati on . Refer to Secti on 5 for detail s. W hen the Extern al Text Control m ode
i s active, press the Other Screen s button or power th e venti lator off then on to exit the
function .
4-8 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service mode 4
4.2.13 Exp Valve Calibration
NOTE:
• If the ventilator has not reached operating temperature from recent usage, allow it to
warm up for at least 10 minutes in service mode before performing this calibration, to
ensure a successful calibration.
• Ensure that the ventilator has passed the Leak Test in EST prior to performing this
calibration.
This functi on builds the exhal ation valve cali bration table. This table l ists exhal ati on valve
DAC comm ands versus exh alati on pressure. The exh alation val ve cal ibrati on requires a gol d
standard test ci rcuit and l asts approxi mately 3 m in utes.
If th e cali bration fail s, try rerunn in g i t. If it conti nues to fail , refer to Table 4-4 to troublesh oot
any errors. Run EST to assist i n troubl eshooting.
NOTE:
A system leak is a common source of failure, particularly expiratory filter leaks.
Make sure the expiratory filter collector vial is secured tightly, the drain port is
capped, and the filter is mounted securely to the ventilator. M ake sure the test
circuit is securely attached.
Bad exp valve loopback current Failure Exhalation valve and/ or analog interface PCB
may be out of specification. Try cleaning the
exhalation valve if inspection warrants and re-
run the calibration.
Both insp and exp autozero failed Failure Analog Interface PCB may be out of
specification.
Calibration failed: current limit exceeded Failure Exhalation valve and/ or analog interface PCB
may be out of specification.
Calibration failed: gain resolution Failure Exhalation valve and/ or analog interface PCB
may be out of specification.
Calibration failed: Pressure build time- Failure Verify that gold standard circuit is still
out. connected and expiratory filter is locked in
place. Verify that seal is placed over collector
vial drain and vial is securely fastened to
expiratory filter.
Exhalation pressure transducer autozero Failure Exhalation pressure transducer DAC count at
failed 0 cmH2O is out of range.
Exhalation valve temperature out of Failure Exhalation valve temperature out of range.
range
Flow sensor cross-check: unable to Failure Flow controller is not able to establish a 5 L/ min
establish flow test flow via the air flow sensor.
Flow sensor cross-check failed Failure Exhalation air flow is out of range with respect
to the 5 L/ min inspiration-controlled test flow.
Verify gold circuit is connected. Check for
expiratory filter leaks.
Flow sensor info invalid Failure Flow sensor calibration table checksum error
occurred.
Pressure sensor cross-check alert: Alert Pressure cross-check alert occurred at one or
approaching spec limit more of five test pressures, no cross-check
failures occurred.
Pressure sensor cross-check failed Failure Pressure transducer cross-check failed at one or
more of these five test pressures (5, 25, 50, 75,
and 100 cmH2O).
Unable to establish flow Failure Flow controller is not able to establish a 5 L/ min
test flow via the exhalation flow sensor.
Unable to program flash Failure Unable to program calibration table into BDU
NOVRAM. Check BDU +12 V supply and BDU
CPU PCB.
4-10 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service mode 4
4.2.14 Vent Inop Test
This functi on exerci ses th e ven tilator i noperative circuitry.
Vent Inop A ten second test PSOLs/ exhalation valve/ safety valve inactive.
Vent Inop B ten second test PSOLs/ exhalation valve/ safety valve inactive.
If a test ph ase fai ls (“ Ven t Inop Test fai led” message i n lower screen ), poweri ng the venti lator
dow n and back up into servi ce m ode wi ll in voke the n orm al service m ode. Troubl eshoot as
i ndicated in Table 4-6. You m ust repeat the Vent In op Test and pass it wi th no fail ures before
ven tilation is al lowed.
GUI vent inop test phase BDU alarm (alone) Check BDU alarm or AI PCB.
NOTE:
The BD CPU PCB controls the
ventilator inoperative and safety
valve open LEDs on both the
BDU and GUI CPU PCBs,
sending the two LED signals to
the GUI via the BDU-GUI cable.
BDU alarm and ventilator inoperative Check BDU-GUI cable, BD CPU PCB, or
and safety valve open LEDs GUI CPU PCB.
PSOLs/ exhalation valve/ safety valve Check BD CPU PCB or AI PCB. If alarm
inactive and/ or LED failure occurred also, check
BDU-GUI cable, GUI CPU.
NOTE:
• If the ventilator has not reached operating temperature from recent usage, allow it to
warm up for at least 10 minutes in service mode before performing this calibration, to
ensure a successful calibration.
• It is very important that the ventilator passes the leak test during EST prior to running a
flow sensor calibration.
• A system leak is a common source of failure, particularly expiratory filter leaks. Make sure
the expiratory filter collector vial is secured tightly, the drain port is capped, and the filter
is mounted securely to the ventilator. Make sure the test circuit is securely attached.
• Use both air and oxygen gas sources to run Flow Sensor calibration.
This fun cti on buil ds a tabl e of exh alati on flow sensor (Q3) offsets. Th e flow sensor cali bration
requires a gold stan dard test circuit.
If th e cali bration fai ls, make sure that the test ci rcui t is attach ed and th at th e expiratory fi lter
i s securel y connected and free of l eaks, th en try rerun ni ng it. If the cali bration conti nues to
fai l, refer to Table 4-7 to troubleshoot an y errors.
4-12 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service mode 4
4.2.15 Running Flow Sensor Calibration
1 Enter Service M ode.
2 En sure both air an d oxygen are con n ected to the ventil ator.
3 Con nect the gol d stan dard circuit between the To Patient an d From Patien t ports.
4 En sure that there are n o leaks or occl usions by run ning EST and ensure th at th e
venti lator passes the EST Leak Test. Once thi s test has passed, exi t EST.
5 Sel ect the Other Screens button .
6 Sel ect Flow Sen sor Cal ibrati on .
7 Press the Start button and press Accept.
The ventil ator autom atical ly starts the cali bration. W hen th e cali brati on i s com pl ete, press
the Oth er Screen s button again to return to the Other Screen s m enu.
Air offset out of range Failure 1. Replace air flow sensor (Q2).
2. Replace exhalation flow sensor (Q3).
3. Replace AI PCB.
Cannot achieve minimum air flow Failure System cannot sustain a minimum 60 L/ min flow. Check
air supply pressure and REG2.
Cannot achieve minimum O2 flow Failure System cannot sustain a minimum 60 L/ min flow. Check
oxygen supply pressure and REG1.
No air connected Failure Wall air pressure switch (PS2) detects no air connected.
Verify air is connected.
NOTE:
When calibrating the atmospheric pressure transducer, make sure that the
pressure in the room is stable and not subject to pressure changes caused by door
closures, heavy foot traffic, or the startup of an air conditioning system. Any such
environmental changes will result in an incorrect adjustment being made to the
ventilator.
This functi on cali brates the atmospheric pressure tran sducer usi ng an external barom eter.
Refer to Table 1-5 for barom eter accuracy speci ficati ons.
Allow the barom etri c pressure readi ng to stabil ize, then in put th e pressure (in m m Hg)
m easured by th is external barometer, as prom pted.
(1 m m Hg = 1.333 m bars = 133.33 Pa = 0.019 psia.)
If th e cali bration fai ls, try rerun nin g it, m akin g sure you correctly in put th e barom etri c
pressure. If the cal ibrati on contin ues to fail , try rerun nin g EST, then repl aci ng th e i nspiratory
pressure tran sducer autozero solenoi d (SOL1) an d th en the inspiratory el ectron ics PCB.
Unable to read air flow sensor Error reading air flow sensor (Q2) serial EEPROM
data. Check Q2 and AI PCB.
Unable to read O2 flow sensor Error reading oxygen flow sensor (Q1) serial
EEPROM data. Check Q1 and AI PCB.
Unable to read the expiratory flow sensor Error reading exhalation flow sensor (Q3) serial
EEPROM data. Check Q3 and AI PCB.
Unable to receive flash data GUI did not receive all calibration data from BDU
within 10 s. Check BDU/ GUI cable, BD CPU PCB, or
GUI CPU PCB.
4-14 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service mode 4
4.2.17 Serial num ber setup
This functi on copies the seri al num bers on the data key i nto NOVRAM . This copy operation
i s perform ed on ly if BDU or GUI NOVRAM con tain s a default seri al n um ber (for exam pl e,
after th e BDU an d/ or GUI CPU PCB i s replaced or after n ew softw are is dow nloaded).
NOTE:
• You are given the option to “ Proceed without setting serial number.” You may want to
use this option for troubleshooting purposes. For example, if a new BDU or GUI CPU PCB
is inserted temporarily for testing, you may find it useful to maintain the default serial
number on that PCB; otherwise, by setting the serial number the test CPU PCB will
become “ committed” to the ventilator used to initialize it. If you choose not to initialize
the serial numbers, be aware that ventilation is not allowed until serial numbers are
copied from the data key.
• It is possible to have one CPU with a valid serial number in NOVRAM but the other CPU
containing the default serial number. For example, when the BDU only is downloaded
with new software, the BDU serial number in NOVRAM is set to the default, but the GUI
serial number in NOVRAM is unaffected.
• Do not use the same data key to initialize the serial numbers in more than one ventilator.
Data key is not installed If a data key is installed but not being detected, try
inserting the key into a known good ventilator whose
serial numbers have already been initialized. Then
power the ventilator up into service mode and verify
that the ventilator sees an invalid data key (that is, the
message “ Serial number(s) doesn’t match the data key”
is displayed). If the data key is still not detected, replace
it; otherwise, replace the AI PCB. Alternatively, try using
a known good data key in the suspect ventilator and
verify that it is detected.
NOTE:
If no data key is inserted, you can proceed with
service mode but ventilation is not allowed.
Serial number(s) doesn’t match the data key Press ACCEPT to confirm and proceed with service
mode.
NOTE:
If an invalid data key is inserted, you can proceed
with service mode but ventilation is not allowed.
NOTE:
If the datakey is updated before calibrations have been run, the operational hours
will be lost or become corrupted.
4-16 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service mode 4
4-18 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
SECTION
S EC T I O N
Performance verification 5 5
Conduct th e perform ance verifi cation to ch eck the perform an ce of th e 840 Ventil ator
System. Field performance veri fication testing i ncl udes electri cal safety tests, the extended
self test (EST), plus speci fic perform ance veri fication tests, using the Puritan Bennett PTS2000
Perform ance Test System with 840 VTSsoftware. A m anual ventilator check can be perform ed
to check gen eral ven til ator perform an ce.
NOTE:
Performance verification procedures do not apply to ventilator accessories. Refer
to the specific accessory operator’s or service manual for performance verification
information. Malfunctioning accessories may affect some ventilator functions and
may result in false test results.
840 Ventilator System Service Manual 4-070496-00 Rev. A (08/ 03) 5-1
5 Performance verification
Table 5-1: Tools, equipment , and service mat erials required for
performance verificat ion
Pentium 166 Mhz computer, equipped, at a minimum, with the Local supplier X
following:
• 32 M B RAM (64 M B for NT systems)
• 800x600 screen display resolution with 256 colors
• Two available serial ports
• Windows® 95, 98, 2000, XP or NT 4.0 operating system
• Mouse or equivalent pointer device
• 2x or higher CD-ROM drive
• 1 Gigabyte available hard disk space
• One available TCP/ IP network port with BNC connector (if software
download desired)
EtherLink III 3C589C LAN PC card for 10BASE-T or later revision, or Software
equivalent card Download
only
EtherLink III 3C589C Combo Cable for EtherLink III PC Card or Software
equivalent Download
only
840 Software Download Cable (BNC to BNC) 4-075731 or Local Supplier Software
Download
only
Digital multimeter (DMM ) accurate to 3 decimal places, with test Fluke Model 87 or equivalent
leads
* Th ese parts m ay be purch ased in dividually or by orderin g an 840 VTS Accessories Kit, P/N 4-076599-00.
5-2 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Performance verification 5
Table 5-1: Tools, equipment, and service materials required for
performance verificat ion (cont inued)
Electrical safety analyzer capable of measuring ground resistance and Dale Model 600, Dale
leakage current Technology Inc., P.O. Box 196,
401 Claremont Ave.,
Thornwood, NY 10594 USA,
800.544.3253
http:/ / www.daletech.com/
or equivalent
Filter, inspiratory, Re/ Flex (for use on PTS 2000 port) 4-074600-00 X
Gold standard tube [flex tube, 21.0 in. (53.4 cm)] (quantity = 2) *4-018506-00 X
Patient circuit, adult (reusable, without water traps or heated wire) *N-4401008 X
(quantity = 2)
Pneumatic calibration analyzer or equivalent devices capable of Puritan Bennett PTS 2000 X
measuring oxygen percent, flow, BTPS volume, pressure, and Performance Test System (also requires
barometric pressure. Oxygen analyzer connector tee. Required (4-074686-00) (includes 840 VTS
accuracies: accessory kit with serial cable) software)
• Flow: 2.75% of reading ±0.05 slpm for performance verification; or
• Volume: 2% of reading or ±1 digit equivalent device for manual
• Low pressure (-150 to +150 cmH2O): 0.75% of reading ±0.04 ventilator check
cmH2O
• High pressure (0 to 150 psig): 1.0% of reading ±0.1 psi
• Oxygen percentage: ±2% oxygen
• Barometric pressure: Range: 10 to 16 psia; resolution: 0.0 to 1 psia;
operating temperature: 10 to 40 °C; measurement accuracy: ±0.75
of reading; response: ≤ 100 ms
Serial Card, Socket I/ O Ruggedized or equivalent (Provides second Socket Communications Corp.
serial port) Newark, CA
http:/ / www.socketcom.com/
* Th ese parts m ay be purchased in dividually or by orderin g an 840 VTS Accessories Kit, P/ N 4-076599-00.
Table 5-1: Tools, equipment , and service mat erials required for
performance verificat ion (cont inued)
Static-dissipative field service kit (includes wrist strap, static dissipative 4-018149-00
mat, and earth (ground) cord)
Tubing, silicone, 1/ 8 in. ID x 1/ 4 in. OD, 4.75 in. (12.1 cm) *4-008578-00 X
Vacuum cleaner, ESD-safe, with 0.2 µm filter (rated for photocopiers Local supplier
and laser printers)
* Th ese parts m ay be purch ased in dividually or by orderin g an 840 VTS Accessories Kit, P/N 4-076599-00.
Warning
To prevent disease transmission, use personal protective equipment when handling
contaminated bacterial filters or other patient accessories. Refer to the 840 Ventilator System
Operator’s and Technical Reference Manual for instructions on sterilizing patient system parts.
Caut ion
Do not soak any portion of the ventilator in solvent, alcohol, or any other cleaning agent.
Soaking ventilator components may damage the ventilator.
5-4 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Performance verification 5
5.4 Preliminary vent ilat or set up
NOTE:
A humidification device is not necessary for performance verification, but is
required for completion of SST.
Performance
Flow sensor
calibration
calibration
calibration
Download
Soft ware
cycle1
Type of service
Ventilator installation 1 -- 2 3 4 5 6 7 8 92
10,000-hour preventive 1 -- 2 3 4 5 6 7 8 9
maintenance (ventilator or
compressor)
Oxygen sensor 2 1
5-6 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Performance verification 5
5.6 Performance verificat ion guidelines
NOTE:
To ensure that the ventilator performs within specifications, Puritan Bennett
recommends that you use the Puritan Bennett PTS2000 Performance Test System
in conjunction with the BreathLab 840 Verification Test Software. If this
equipment is not available, you can use an alternative test that checks some
parameters but does not guarantee the integrity of the entire system. Refer to
Table 5-3 to determine the approximate test limits if you use equipment other
than the PTS 2000 Performance Test System.
To en sure systematic performance verification and l ogical fault diagn osis, perform tests in the
order gi ven. Foll ow th ese general guideli nes when runni ng the perform ance verification:
Warning
Follow accepted safety procedures for electrical equipment when making connections,
adjustments, or repairs.
Warning
Do not enter the service mode while a patient or other person is connected to the ventilator.
To do so may cause patient injury or death. While in the service mode, normal ventilator
operation is not accessible.
NOTE:
The procedures that follow do not verify the performance of accessories. Verify the
performance of accessories using the appropriate procedures in the applicable
operator’s or service manual.
Warning
If the ventilator fails an electrical safety test, do not proceed to the next electrical safety test
until the problem is corrected and the ventilator is retested.
Caut ion
To prevent possible equipment damage, do not attempt to perform HI-POT testing on the
ventilator. The ventilator design incorporates high-voltage protective devices that may be
damaged if HI-POT testing is performed on the entire system. The ventilator is HI-POT-
tested during manufacturing, using processes that do not damage the protective devices.
NOTE:
• After servicing the ventilator, perform an electrical safety test before putting the
unit back into operation.
• Before performing the electrical safety test, make sure the compressor unit (if
installed) and all accessories are connected and operational.
5-8 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Performance verification 5
5.7.3 Ext ended self test (EST)
NOTE:
• A test (gold standard) patient circuit and a no. 1 stopper are required to perform EST.
• The nurse’s call circuit is tested during EST and may be bypassed by selecting the CLEAR
button.
• If the ventilator has not reached operating temperature from recent usage, allow it to
warm up for at least 10 minutes in service mode before running EST, to ensure accurate
testing.
NOTE:
During testing, it is sometimes necessary for the user to perform test-related tasks,
such as disconnecting a gas supply or blocking the To Patient port. For each such
step, messages are displayed in the prompt area.
3 Verify that EST com pletes al l test with a PASS result. Perform ance verificati on i s not
com pl ete if any EST test results in an ALERT, OVERRIDE, or FAILURE. If a test fails, refer
to Secti on 8 for repair i nform ation .
Caution
To prevent damage to ESD-sensitive components, always follow ESD guidelines when
performing this procedure.
NOTE:
Regulator performance must be verified each time performance verification or a
manual ventilator check is performed. Use the appropriate type of regulator
calibration hose (see Table 5-1).
4 Rem ove h ose from test port, an d verify that anal yzer does n ot register pressure. Replace
test port cover.
5 Repeat steps 3 an d 4 for oxygen si de.
NOTE:
If regulators are out of specification, refer to Section 8 for service and repair of the
inspiratory module.
5.7.6 Per formance verification using PTS2000 Performance Test Syst em and
Breat hLab 840 VTS soft ware
1 Assem ble test equipm en t indi cated in Tabl e 5-1. (Software in stal lati on i nstructi ons are
provi ded with BreathLab 840 VTS Software.)
2 Turn on com puter, an d bri ng up th e VTS program .
3 Turn venti lator on , and bri ng up in service m ode.
4 Usin g venti lator test program , press START TEST to begin test appli cation.
5 Fol low prom pts di splayed on computer screen to com plete perform an ce veri ficati on .
Access th e 840 VTS Software Hel p Screen for addi tional in form ation.
5.7.7 M anual ventilat or check using equipment ot her t han PTS 2000
Perform ance Test System
NOTE:
• To ensure proper ventilator performance, Puritan Bennett recommends that you use the
Puritan Bennett PTS 2000 Performance Test System along with the BreathLab 840 VTS
software.
• If you are using a pneumatic analyzer other than the PTS 2000 Performance Test System,
you must also have an oxygen analyzer available.
• The manual ventilator check (i.e., without PTS2000 PTS and 840 VTS software) does not
satisfy the manufacturer’sperformance testing requirements. The manual ventilator check
can be used as a general indicator of ventilator condition.
1 Before beginni ng your ventil ator check, perform these al arm tests:
a. Verify that rem ote al arm (n urse’s cal l) relay is fun ctional by in voki ng an alarm an d
veri fying that your nurse’s call system regi sters appropriate alarm .
b. Verify that the BDU annunciates the “ l oss of GUI” alarm when GUI is disconnected.
c. Verify that low-, m edi um -, and h igh -urgen cy alarm s sound by creating an alarm ed
event for each of these. Refer to Operator’s and Technical Reference Manual for
assi stan ce.
2 M easure ven tilator parameters, and verify that they m eet specifications i n Tabl e 5-3.
5-10 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Performance verification 5
NOTE:
To accurately measure the parameters listed in Table 5-3, add the tolerance of
your measurement device to the target value of the ventilator. The test results you
obtain are approximate and may not guarantee the accuracy of the ventilator.
5-12 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
SECTION
S EC T I O N
Diagnostic codes 6 6
6.1 Int roduct ion
This section con tain s in form ation to assi st i n th e i nterpretation of di agn ostics codes,
i ndi cated in the venti lator’s error an d status l ogs, as well as di agnostic codes reported by the
diagnosti c LED arrays on the BD CPU PCB and the GUI CPU PCB.
The ventil ator’s di agnostic an d status l ogs also contai n oth er information that can be useful
wh en troubl eshooti ng the ventil ator. The logs are:
• Syst em Dia gn ostic Lo g: Stores data on un expected condi tions detected during POST,
backgroun d checks, an d cal ibrati on s. Th is i ncl udes al erts an d failures.
• Syst em In fo rm a t io n Lo g: Stores data on ventil ator events, in cl udi ng detai ls on
GUI/BD com m un i cati on .
• EST/ SST D ia gn o stic Lo g: Stores data on un ex pected condi tions detected duri ng SST and
EST, al ong with overall resul ts at com pletion of SST an d EST.
You access th ese l ogs through the service m ode, as described in Section 4.2.2.
Table Page
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-1
6 Diagnostic codes
NOTE:
The alphanumeric diagnostic code for each of the self tests is not identified in this
manual. This information is not necessary for field service repairs.
If th e system detects a faul t duri ng POST (or a background check), each of the four pairs of
LEDs i n th e correspondi ng LED array wil l li ght, in sequen ce, creati ng a flashi ng effect. Th e
sam e LED array th en di spl ays the di agnostic code corresponding to th e detected faul t. Thi s
LED pattern of display—fl ash ing LED pai rs fol lowed by a diagn osti c code—cycles
conti nuousl y an d is a visual i ndi cator of a system fault. Use the tables i n this section to
convert the diagnostic code LED pattern to an al phanum eric di agnostic code.
Additi on all y, under a fault condi tion, the processor of the corresponding CPU PCB attem pts
to l og the fault i nform ati on , with a ti m e-stamp, in its nonvolati le ran dom access m em ory
(NOVRAM ). Upon compl etion of its POST, each processor reports i ts test results to th e GUI
processor for di spl ay an d transfer i nto the System Diagnosti cs Log.
POST fail ures are cl assified as m i nor or m aj or faults:
Min or PO ST fa u lt i s a fault that does not affect venti lati on or pati ent safety ch ecks. A m i nor
fault does n ot i nterrupt the regular POST sequence. The GUI di splays POST fault in form ation
an d logs it in to NOVRAM of the corresponding CPU PCB. Th e System Diagn ostic Log di spl ays
an alert. Normal venti lati on wi ll begin if POST detects a m in or faul t.
Ma jo r POST fa u lt is a fault that affects venti lati on or pati ent safety checks. A m aj or faul t
i nterrupts th e regular sequen ce of POST and causes a VENT INOP condi tion. Fault
i nform ati on is sen t to the GUI (i f possible) an d to the LED array on the correspondi ng CPU
PCB (BD or GUI). Th e safety valve an d exh alati on val ve rem ai n open to allow the pati en t to
6-2 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
breath e room ai r. Th e software does n ot perm it normal venti lation whi le a m aj or faul t exists.
To clear a m aj or fault, both CPU processors m ust com pl ete thei r respective POST, wi th no
m ajor POST faul ts. If onl y a mi nor fault exists, n orm al ven tilati on wi ll begin .
Status LED
diagnostic code LEDs
(read from top
to bottom)
8-00156
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-3
6 Diagnostic codes
Error identifier
Diagnostic type
Reportability classification
D or H BDU problem that resulted in soft reset during operation (as opposed
to during POST)
6-4 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-1: Values of diagnost ic code digit s
b B Background check
C Communications test
E EST
N Nonmaskable interrupt
S SST
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-5
6 Diagnostic codes
• The fifth col um n provides addi tional in form ation and/ or identi fies possi bl e causes.
• The sixth colum n suggests how to correct th e con diti on . Th ese actions are sequen ced to
correct th e m ost probable m al functi on or to present the m ost efficien t corrective action
first. The proposed fi xes li sted, however, m ay not al ways correct th e parti cul ar probl em .
In addition to the actions suggested in the tables i n th is section, keep the fol lowing i n mi nd
when troubleshooti ng the ventil ator:
• Verify secure connections of cabl es and ven ti lator m odul es.
• If possi bl e, run full EST, bypassi ng an y failures or faults that may occur, to further
diagn ose a problem .
• Replace BD or GUI CPU PCB only after al l other rem edies have been attem pted. Contact
Puritan-Ben nett Techni cal Support for further assistance.
• Check for system l eaks, particularly at the expi ratory fil ter. M ake sure the expi ratory fil ter
collector vial is secured ti gh tl y, th e drain port is capped, and the fil ter i s m ounted securel y
to the venti lator. M ake sure th e test circui t is securely attached.
• Before runnin g SST or EST, ensure th at th e un it h as been warm ed up for at l east 10
m inutes i n am bient temperature. Fail ure to warm up th e venti lator m ay resul t in fal se
flow sen sor or pressure transducer fai lures.
W h en troubl eshooting SST or EST, al ways use a kn own good patien t ci rcuit or gol d standard
ci rcuit and filter(s).
6-6 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
AE0110 Unable to EST Flow Alert Flow controller unable 1. Make sure oxygen supply is
establish O2 sensors cross to establish and control connected and unrestricted.
flow check Test oxygen flow at 2. Verify oxygen regulator pres-
120 L/ min. sure is set between 9-12 psi.
3. Switch PSOLs to see if failure
transfers to the other gas side.
If yes, return the air PSOL to its
original position and replace
the oxygen PSOL.
4. Run the leak test in EST to
check for leaks/ occlusions.
5. Perform a flow sensor calibra-
tion.
6. Replace Q1.
7. Replace the Inspiratory Elec-
tronics PCB.
8. Replace the AI PCB.
AE0111 Unable to EST Flow Alert Flow controller unable 1. Make sure air supply is con-
establish air flow sensors cross to establish and control nected and unrestricted.
check Test air flow at 120 L/ min. 2. Verify air regulator pressure is
set between 9-12 psi.
3. Switch PSOL's to see if failure
transfers to the other gas side.
If yes, return Oxygen PSOL to
its original position and replace
the Air PSOL.
4. Run the leak test in EST to
check for leaks/ occlusions.
5. Perform a flow sensor calibra-
tion.
6. Replace Q2.
7. Replace the Inspiratory Elec-
tronics PCB.
8. Replace the AI PCB.
AE0213 Air PSOL leak EST Gas Alert Air PSOL (PSOL2) 1. Check for leaks around the Q2
Supply/ SV Test forward leak was O-rings.
detected via excessive 2. Make sure air supply is con-
pressure buildup nected and unrestricted.
( > 50 cmH2O but 3. Reseat Air PSOL.
< 100 cmH2O) in 4. Replace Air PSOL.
blocked inspiratory 5. Replace AI PCB.
module.
AE0214 O2 PSOL leak EST Gas Alert Oxygen PSOL (PSOL1) 1. Check for leaks around the Q1
Supply/ SV Test forward leak was O-rings.
detected via excessive 2. Make sure oxygen supply is
pressure buildup connected and unrestricted.
( > 50 cmH2O but 3. Reseat oxygen PSOL.
< 100 cmH2O) in 4. Replace oxygen PSOL.
blocked inspiratory 5. Replace AI PCB.
module.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-7
6 Diagnostic codes
AE0306 Test circuit not EST Leak Test Alert Pressure not detected 1. M ake sure test circuit is prop-
connected on expiratory side erly connected.
2. Replace expiratory bacteria
filter.
3. Check for leaks around the Q3
flow sensor.
4. Check/ replace the exhalation
valve.
5. Replace the expiratory pressure
transducer PCB.
AE0601 GUI High Alarm EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
LED fails. Test indicate LED not on. cable between the GUI LED
PCB and the GUI CPU PCB.
2. Replace the GUI LED PCB.
AE0602 GUI M edium EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
Alarm LED fails. Test indicate LED not on. cable between the GUI LED
PCB and the GUI CPU PCB.
2. Replace the GUI LED PCB.
AE0603 GUI Low Alarm EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
LED fails. Test indicate LED not on. cable between the GUI LED
PCB and the GUI CPU PCB.
2. Replace the GUI LED PCB.
AE0604 GUI Normal LED EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
fails. Test indicate LED not on. cable between the GUI LED
PCB and the GUI CPU PCB.
2. Replace the GUI LED PCB.
AE0605 GUI Batt Backup EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
LED fails. Test indicate LED not on. cable between the GUI LED
PCB and the GUI CPU PCB
2. Replace the GUI LED PCB.
AE0606 GUI On Batt Pwr EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
LED fails. Test indicate LED not on. cable between the GUI LED
PCB and the GUI CPU PCB.
2. Replace the GUI LED PCB.
AE0607 GUI Compressor EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
Ready LED fails. Test indicate LED not on. cable between the GUI LED
PCB and the GUI CPU PCB.
2. Replace the GUI LED PCB.
AE0608 GUI Compressor EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
Operating LED Test indicate LED not on. cable between the GUI LED
fails. PCB and the GUI CPU PCB.
2. Replace the GUI LED PCB.
AE0609 GUI 100% O2 EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
LED fails. Test indicate LED not on. cable between the GUI LED
PCB and the GUI CPU PCB.
2. Replace keyboard.
AE0610 GUI Alarm EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
Silence LED fails. Test indicate LED not on. cable between the GUI LED
PCB and the GUI CPU PCB.
2. Replace the GUI LED PCB.
6-8 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
AE0611 GUI Screen Lock EST GUI Lamp Alert CLEAR key pressed to 1. Check/ replace interconnect
LED fails. Test indicate LED not on. cable between the GUI LED
PCB and the GUI CPU PCB.
2. Replace the GUI LED PCB.
AE0702 Bad Vent inop EST BD Lamp Alert CLEAR key pressed to Replace BD LED PCB.
LED Test indicate one or both
ventilator inoperative
LEDs not on.
AE0703 Bad SVO LED EST BD Lamp Alert CLEAR key pressed to Replace BDU LED PCB.
Test indicate one or both
SVO LEDs not on.
AE0704 Bad Loss of GUI EST BD Lamp Alert CLEAR key pressed to Replace BDU LED PCB.
LED Test indicate loss of GUI LED
not on.
AE1001 Air PSOL EST PSOL Alert Air PSOL (PSOL2) 1. Verify that the air supply is
loopback Loopback Test loopback current out of good.
current OOR range of drive current. 2. Verify air regulator pressure set
to between 9 and 12 psi.
3. Switch PSOLs to see if failure
transfers to the other gas side.
If yes, return oxygen PSOL
(PSOL1) to its original position
and replace PSOL2.
4. Switch Q1 and Q2, run a flow
sensor calibration and rerun
test. If the problem transfers to
the other gas side, return Q1 to
its original position and replace
Q2.
5. Replace the AI PCB.
6. Replace the Inspiratory
Electronics PCB.
AE1002 O2 PSOL EST PSOL Alert Oxygen PSOL (PSOL1) 1. Verify that the oxygen supply is
loopback Loopback Test loopback current out of good.
current OOR range of drive current. 2. Verify oxygen regulator pres-
sure set to between 9 and 12
psi.
3. Switch PSOLs to see if failure
transfers to the other gas side.
If yes, return the air PSOL to its
original position and replace
PSOL1.
4. Switch Q1 and Q2, run a flow
sensor calibration and rerun
test. If the problem transfers to
the other gas side, return Q2 to
its original position and replace
Q1.
5. Replace the AI PCB.
6. Replace the Inspiratory
Electronics PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-9
6 Diagnostic codes
AE1104 Insp check valve EST Safety Alert It took too little time to 1. M ake sure test circuit is con-
test failed System Test relieve excess pressure nected.
through open safety 2. M ake sure CV3 is not installed
valve, indicating backward.
inspiratory check valve 3. Replace CV3.
(CV3) may be damaged
or incorrectly mounted.
AE1201 Exp valve EST Exp Valve Alert Exhalation valve 1. Verify that the system has no
loopback Loopback Test loopback current is out leaks or occlusions by running
current OOR of range of drive the leak test in EST.
current. 2. Clean exhalation valve
diaphragm.
3. Calibrate the exhalation valve.
4. Clean/ replace the exhalation
valve.
5. Replace the AI PCB.
AE1305 Seal test failed EST Exp Valve Alert Seal test ∆ pressure is 1. Verify that the system has no
Seal Test above alert level but leaks or occlusions by running
below failure level. the leak test in EST.
2. Clean the exhalation valve.
3. Calibrate exhalation valve.
4. Replace the exhalation valve.
5. Replace the AI PCB.
AE1600 Compressor EST Status Ventilator did not sense No action required.
Test - Not Compressor a compressor attached
installed Test and skipped test.
AE1601 Wall air pressure EST Alert Wall air pressure switch 1. M ake sure air supply is discon-
detected Compressor (PS2) detected air after nected.
Test user was prompted to 2. Disconnect PS2 and rerun test.
disconnect air. If test passes, replace PS2.
AE1602 ac power not EST Alert System is still running 1. Plug in ventilator power cord
connected Compressor on battery power after and check the cord connec-
Test prompting user to tion at the ventilator.
connect ac power. 2. Check ac.
Compressor can only 3. Disconnect BPS to isolate prob-
run on ac (facility) lem.
power. 4. Replace power supply.
AE1603 Compressor EST Alert Compressor pressure 1. Run compressor leak test to
pressure not Compressor transducer (PC) check for leaks.
detected Test indicates that a. Use leak detector to check
compressor air is not for leaks at the accumulator
present, although fittings and other tubing
compressor motor is connections.
on. b. Verify no leaks at CV2 within
the ventilator by plugging
the air inlet fitting.
2. Replace compressor PCB.
3. Contact Puritan-Bennett Tech-
nical Support.
6-10 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
AE1604 Run mode time EST Alert Compressor timer is not Listen for motor. If motor is on,
OOR Compressor running while replace compressor PCB.
Test compressor motor is Otherwise, replace compressor.
on.
AE1607 Disabled mode EST Alert Compressor timer is Replace compressor PCB.
time OOR Compressor running while
Test compressor motor is
off.
AE1608 Unable to test EST Alert Compressor standby 1. Troubleshoot code AE1604 or
standby mode Compressor mode check cannot be AE1607, if present.
Test run, because of inability 2. Replace compressor PCB.
to verify timer
functionality during
compressor run and
disabled tests.
AE1609 Standby mode EST Alert Compressor motor is Replace compressor PCB.
time OOR Compressor still running. During
Test standby mode test
phase, compressor
motor should
eventually turn off.
AE1610 Unable to EST Alert Compressor load test Troubleshoot code AE1603 or
perform Compressor cannot be run, because AE1606.
compressor load Test of inability to verify PC
test during compressor run
and disabled tests.
AE1611 Compressor EST Alert Compressor unable to 1. Run compressor leak test to
load test failed Compressor maintain minimum check for compressor leak.
Test pressure under worst- a. Use leak detector to check
case breath delivery for leaks at the accumulator
waveform. fittings and other tubing
connections.
b. Verify no leaks at CV2 within
the ventilator by plugging
the air inlet fitting.
2. Replace compressor.
3. Replace compressor PCB.
AE1700 Compressor EST Status Ventilator did not sense No action required.
Leak Test - Not Compressor a compressor attached
installed Leak Test and skipped test.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-11
6 Diagnostic codes
AE1702 Wall air pressure EST Alert Wall air pressure switch 1. M ake sure air supply is discon-
detected Compressor (PS2) detected air nected.
Leak Test presence after user was 2. Disconnect PS2 and rerun test.
prompted to If test passes, replace PS2.
disconnect air.
AE1703 Wall air pressure EST Alert Wall air pressure switch 1. M ake sure air supply is con-
not detected Compressor (PS2) did not detect air nected.
Leak Test presence after user was 2. Run Gas Supply/ SV Test.
prompted to connect
air.
AE1901 GUI touch: Error EST GUI Touch Alert Touch screen error 1. Clean touch screen, removing
Test occurred (e.g., blocked any obstructions.
beam or low-level 2. Replace touch frame PCB.
error), buffer cannot be 3. Replace GUI CPU PCB.
read, or report is
invalid.
AE2001 Bad GUI serial EST GUI Serial Alert While in loopback Replace GUI CPU PCB.
port Port Test mode, failed to verify
received message was
identical to transmitted
message.
AE2101 Battery not EST Battery Alert BPSnot fully charged at 1. Allow BPS to fully charge, then
charged Test start of test. repeat test.
2. Replace BPS PCB.
3. Replace battery pack.
4. Replace BD CPU PCB.
AE2102 Battery not EST Battery Alert BPS not discharging 1. Verify ac power is disconnected
discharging Test after ac power was when prompted.
disconnected. 2. Replace BPS PCB.
3. Replace battery pack.
4. Replace BD CPU PCB.
6-12 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
AE2103 Bad Backup EST Battery Alert While BPS was 1. Ensure that the battery pack is
Power Supply Test discharging, BPS fully charged by checking that
voltage dropped below the green LED on the BPS is lit
accepted level or prior to initiating an EST. If the
dropped too quickly. amber LED is lit, allow the unit
to charge the batteries prior to
rerunning EST.
2. Replace the battery pack.
3. Replace the BPS PCB.
4. Replace the AI PCB.
5. Replace the power supply.
AE2104 Battery not EST Battery Alert BPS not charging after 1. Verify that ac power is recon-
charging Test ac power was nected when prompted and
reconnected. that ac is good.
2. Replace BPS PCB.
3. Replace the battery pack.
4. Replace the power supply.
5. Replace the BD CPU.
6. Replace the AI PCB.
AE2300 GUI Nurse Call EST Test Status User pressed CLEAR to No action required.
Test - Not indicate nurse’s call
installed device not installed.
Test was skipped.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-13
6 Diagnostic codes
AS0010 Unable to SST flow sensor Alert Flow controller unable 1. M ake sure oxygen supply is
establish O2 test to establish and control connected and unrestricted.
flow oxygen flow at 2. Verify that the oxygen supply
120 L/ min. meets minimum pressure
requirements.
3. Verify that the patient circuit
system has no leaks or occlu-
sions. If not sure, run the first
four tests in EST to get to the
leak test. If the leak test passes
in EST, exit out of EST. Correct
the leak or occlusion in the
patient circuit used in SST and
rerun or try another patient cir-
cuit.
4. Check the oxygen inlet filter
assembly.
5. Verify that the oxygen regula-
tor pressure is set between 9-
12 psi.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the air side, return Q2 back to
its original position and replace
Q1.
8. Switch PSOLs to see if the fail-
ure transfers to the air side. If
yes, return PSOL2 back to its
original position and replace
PSOL1.
6-14 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
AS0011 Unable to SST flow sensor Alert 1. Flow controller 1. Make sure air supply is con-
establish air flow test unable to establish nected and unrestricted.
and control air flow 2. Verify that the air supply meets
at 120 L/ min. minimum pressure require-
ments.
3. Verify that the patient circuit
system has no leaks or occlu-
sions. If not sure, run the first
four tests in EST to get to the
leak test. If the leak test passes
in EST, exit out of EST. Correct
the leak or occlusion in the
patient circuit used in SST and
rerun or try another patient cir-
cuit.
4. Check the air inlet filter (F2).
5. Verify that the air regulator
pressure is set between 9-12
psi.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the oxygen side, return Q1
back to its original position and
replace Q2.
8. Switch PSOLs to see if the fail-
ure transfers to the oxygen
side. If yes, return PSOL1 back
to its original position and
replace PSOL2.
AS0012 O2 pressure not SST flow sensor Alert Oxygen pressure not Make sure oxygen supply is
detected test detected via PS1. Only connected.
air available for SST.
AS0013 Wall air pressure SST flow sensor Alert Air pressure not Make sure air supply is connected.
not detected test detected via PS2 or PC.
Only oxygen available
for SST.
AS0202 Excessive leak SST Circuit leak Alert Pressure drops to 85 1. Make sure patient circuit is
cmH2O in 10 s, connected and is not leaking.
then in 10 s more by 2. Check exhalation valve opera-
≥ 10 cmH2O. tion.
3. Verify connections of bacteria
filters and humidifier are
secure.
AS0305 Occluded SST Expiratory Alert Pressure drop across Consider replacing expiratory
expiratory filter filter filter > 2 cmH2O, but filter.
< 3 cmH2O.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-15
6 Diagnostic codes
AS0311 Low expiratory SST Expiratory Alert Pressure drop 1. Repeat test, following direc-
filter ∆P filter across expiratory filter tions more closely.
< 0.4 cmH2O, but 2. Replace filter.
> 0.1 cmH2O.
AS0403 Occluded SST Circuit Alert Inspiratory limb Check for occluded patient tubing.
inspiratory limb Resistance pressure > 8.5 cmH2O
(adult) or 5.5 cmH2O
(pediatric), but
< 12.5 cmH2O (adult)
or 7.5 cmH2O
(pediatric), indicating
occlusion.
AS0406 Occluded SST Circuit Alert Expiratory limb Check for occluded patient tubing.
exhalation limb Resistance pressure > 8.5 cmH2O
(adult) or 5.5 cmH2O
(pediatric) but > 12.5
cmH2O (adult) or
7.5 cmH2O (pediatric),
indicating occlusion.
AS0407 Unable to reach SST Circuit Alert During characterization Check for kinked or occluded
min peak flow Resistance of total circuit patient tubing.
resistance over a range
of flows, peak flow
< 80 L/ min but > 60 L/
min.
AS0411 Unable to reach SST Circuit Alert During characterization Check for kinked or occluded
min peak flow Resistance of total circuit patient tubing.
resistance over a range
of flows, peak flow was
less than alert
threshold.
AS0413 Insp limb SST Circuit Alert Inspiratory limb 1. M ake sure inspiratory filter is
resistance low Resistance pressure < 0.6 cmH2O installed.
(adult) or 5.5 cmH2O 2. Repeat test, following direc-
(pediatric), but tions more closely.
> 0.2 cmH2O. 3. Replace inspiratory filter.
4. Replace patient circuit.
AS0414 Exp limb SST Circuit Alert Expiratory limb Replace patient circuit.
resistance low Resistance pressure < 0.6 cmH2O
(adult) or 0.5 cmH2O
(pediatric), but
> 0.2 cmH2O.
AS0505 Excessive SST Alert High compliance 1. M ake sure correct patient tub-
compliance Compliance >6 mL/ cmH2O (adult) ing type was specified in SST.
calibration or 4.5 mL/ cmH2O 2. Replace patient circuit.
(pediatric), but
< 12 mL/ cmH2O (adult)
or 9 mL/ cmH2O
(pediatric).
6-16 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
AS0507 Compliance low SST Alert High compliance falls 1. Make sure correct patient tub-
Compliance below 1.56 mL/ cmH2O ing type was specified in SST.
calibration (adult) or 1.34 mL/ 2. Replace patient circuit with a
cmH2O (pediatric) but known good circuit and filter
not lower than set.
1.05 mL/ cmH2O. 3. Run EST to check the operation
of PI and PE during the Circuit
Pressure Test. Replace applica-
ble pressure transducer.
4. Run Atmospheric Pressure
Transducer calibration.
AS0509 Excessive SST Alert Low compliance > 1. Make sure correct patient tub-
compliance Compliance 6 mL/ cmH2O (adult) or ing type was specified in SST.
calibration 4.5 mL/ cmH2O 2. Replace patient circuit with a
(pediatric), but < known good circuit and filter
12 mL/ cmH2O (adult) set.
or 9 mL/ cmH2O 3. Run EST to check the operation
(pediatric). of PI and PE during the Circuit
Pressure Test. Replace applica-
ble pressure transducer.
4. Run Atmospheric Pressure
Transducer calibration.
AS0511 Compliance low SST Alert Low compliance falls 1. Make sure correct patient tub-
Compliance below 1.56 mL/ cmH2O ing type was specified in SST.
calibration (adult) or 1.34 mL/ 2. Replace patient circuit with a
cmH2O (pediatric), but known good circuit and filter
not below set.
1.05 mL/ cmH2O. 3. Run EST to check the operation
of PI and PE during the Circuit
Pressure Test. Replace applica-
ble pressure transducer.
4. Run Atmospheric Pressure
Transducer calibration.
Dxxxxx Assertion Background Failure that System generated a Replace the BDCPU PCB.
Checks (BD) results in a reset to correct a
POST or a boundary check or
reset possible data
corruption of control
variables.
DT0002 Bus error / Background Failure that Hardware trap from an Replace the BD CPU PCB.
Access fault Checks (BD) results in a access fault due to a
POST or a bad memory chip, bad
reset control line, or a
hardware timing issue.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-17
6 Diagnostic codes
NOTE:
For all "FE" codes (failures reported during EST), diagnose the problem and perform the repair.
Run a complete EST to reset the EST test failure in memory. If there is a failed EST test result in
memory, normal operation is prevented until the failed EST test is rerun and passes.
FE0003 Failed to reach EST Circuit Failure Unable to build 1. M ake sure that test circuit is
test pressure Pressure Test pressure (air) to installed and air supply or com-
10 cmH2O at pressor is available and good.
5 L/ min within time-out 2. Verify that the air regulator
period. pressure is set between 9-12
psi.
3. Verify that the patient circuit
system has no leaks or occlu-
sions by running the first four
tests in EST to get to the leak
test.
FE0004 Cross-check EST Circuit Failure Inspiratory/ expiratory 1. Verify that the system has no
failed Pressure Test pressure transducer leaks or occlusions. If not sure,
readings at 10 cmH2O run the first four tests in EST to
test pressure are too far run the leak test
apart. 2. If the leak test passes in EST,
exit out of EST. Review the
data for the Circuit Pressure
Test to determine which pres-
sure transducer is out of range:
PI or PE. Replace applicable
pressure transducer.
3. Replace AI PCB.
FE0005 Bad insp EST Circuit Failure Inspiratory pressure 1. Replace SOL1.
autozero Pressure Test reading (taken after 2. Replace Inspiratory Electronics
solenoid inspiratory pressure PCB.
transducer autozero 3. Replace AI PCB.
solenoid (SOL1)
actuated) out of range
(-0.60 to 0.60 cmH2O).
FE0006 Bad exp EST Circuit Failure Expiratory pressure 1. Replace SOL2.
autozero sol Pressure Test reading (taken after 2. Replace exhalation transducer
expiratory pressure PCB.
transducer autozero 3. Replace AI PCB.
solenoid (SOL2)
actuated) out of range
(-0.60 to 0.60 cmH2O).
6-18 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
FE0007 Cross-check EST Circuit Failure Inspiratory/ expiratory 1. Replace the expiratory bacteria
failed Pressure Test pressure transducer filter.
measurements at 50 or 2. Verify that the system has no
100 cmH2O test leaks or occlusions. If not sure,
pressure are too far run the first four tests in EST to
apart. run the leak test.
3. If the leak test passes in EST,
exit out of EST. Review the data
for the Circuit Pressure Test to
determine which pressure
transducer is out of range: PI or
PE. Replace applicable pres-
sure transducer.
4. Replace AI PCB.
FE0008 Failed to reach EST Circuit Failure Unable to build 1. Make sure air supply or com-
test pressure Pressure Test pressure (air) to 50 or pressor is available.
100 cmH2O at 5 L/ min 2. Check for system leak at the
within time-out period. expiratory bacteria filter or O2
sensor. If not sure, run the first
four tests in EST to run the leak
test.
FE0010 ac power not EST Circuit Failure System is still running 1. Plug in ventilator power cord.
connected Pressure Test on battery power after 2. Disconnect BPS to isolate prob-
prompting user to lem.
connect ac power. EST 3. Replace power supply.
can only run on ac
(facility) power.
FE0101 O2 flow sensor EST Flow Failure Oxygen flow sensor 1. Make sure oxygen supply is
cross check sensors cross (Q1) cross-check failed. connected and unrestricted.
failed check Test 2. Verify that the oxygen supply is
good.
3. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
4. Run a flow sensor calibration.6.
Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the air side, return Q2 back to
its original position and replace
Q1.
5. Switch PSOLs to see if the fail-
ure transfers to the air side. If
yes, return PSOL2 back to its
original position and replace
PSOL1.
6. Replace exhalation flow sensor
(Q3)
7. Replace AI PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-19
6 Diagnostic codes
FE0102 O2 PSOLcurrent EST Flow Failure Oxygen PSOL (PSOL1) 1. M ake sure oxygen supply is
out of range sensors cross current is out of range connected and unrestricted.
check Test with respect to flow 2. Verify that the oxygen supply is
sensor (Q1). good.
3. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
4. Switch PSOLs to see if the fail-
ure transfers to the air side. If
yes, return PSOL2 back to its
original position and replace
PSOL1.
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the air side, return Q2 back to
its original position and replace
Q1.
FE0103 Air flow sensor EST Flow Failure Inspiratory module air 1. M ake sure air supply is con-
cross check sensors cross flow sensor (Q2) cross- nected and unrestricted.
failed check Test check failed. 2. Verify that the air supply is
good.
3. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
4. Run a flow sensor calibration.
5. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the oxygen side, return Q1
back to its original position and
replace Q2.
6. Switch PSOLs to see if the fail-
ure transfers to the air side. If
yes, return PSOL back to its
original position and replace
PSOL2
7. Replace exhalation flow sensor
(Q3).
8. Replace AI PCB.
6-20 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
FE0104 Air PSOL current EST Flow Failure Air PSOL (PSOL2) 1. Make sure air supply is con-
out of range sensors cross current is out of range nected and unrestricted.
check Test with respect to air flow 2. Verify that the air supply is
sensor (Q2). good.
3. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
4. Switch PSOLs to see if the fail-
ure transfers to the air side. If
yes, return PSOL back to its
original position and replace
PSOL2
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the oxygen side, return Q1
back to its original position and
replace Q2.
FE0106 Unable to EST Flow Failure Flow controller unable 1. Make sure oxygen supply is
establish O2 sensors cross to establish and control connected.
flow check Test oxygen flow at 60, 5, 2. Check regulated oxygen pres-
and 1 L/ min. sure.
3. Replace PSOL1 or Q1.
4. Replace AI PCB.
FE0107 Unable to EST Flow Failure Flow controller unable 1. Make sure air supply is con-
establish air flow sensors cross to establish and control nected.
check Test air flow at 60, 5, and 1 2. Check regulated air pressure.
L/ min. 3. Perform flow sensor calibra-
tion.
4. Replace PSOL2 or Q2.
5. Replace AI PCB.
FE0108 O2 zero flow EST Flow Failure Inspiratory flow > 1. Verify no leaks at the Q1
check failed sensors cross 0.153 L/ min with O-rings.
check Test oxygen PSOL (PSOL1) 2. Run EST Gas Supply Test to
commanded to 0 check for PSOL1 leak. Remove
(closed). and then reseat PSOL1
3. Replace PSOL1.
4. Perform a flow sensor calibra-
tion.
5. Replace Q1.
FE0109 Air zero flow EST Flow Failure Inspiratory flow > 0.153 1. Verify no leaks at the Q2
check failed sensors cross L/ min with air PSOL O-rings.
check Test (PSOL2) commanded 2. Run EST Gas Supply Test to
to 0 (closed). check for PSOL2 leak. Remove
and then reseat PSOL2
3. Replace PSOL2.
4. Perform a flow sensor calibra-
tion.
5. Replace Q2.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-21
6 Diagnostic codes
FE0204 Wall air pressure EST Gas Failure Wall air not detected 1. M ake sure air supply is con-
not detected Supply/ SV Test initially and wall air still nected.
not detected after user 2. Check PS2: Remove the wiring
was prompted to connectors from PS2 and
connect wall air. jumper the connectors to each
other. If the system now rec-
ognizes the air supply, replace
PS2.
FE0205 O2 pressure not EST Gas Failure Oxygen pressure not 1. M ake sure oxygen supply is
detected Supply/ SV Test detected initially and connected.
oxygen pressure still 2. Check PS1: Remove the wiring
not detected after user connectors from PS1 and
was prompted to jumper the connectors to each
connect oxygen. other. If the system now rec-
ognizes the air supply, replace
PS1.
FE0206 O2 pressure EST Gas Failure Oxygen pressure switch 1. M ake sure oxygen supply is dis-
detected (O2 Supply/ SV Test detected presence of connected.
not oxygen after user was 2. Replace Inspiratory Electronics
disconnected) prompted to PCB.
disconnect oxygen. 3. Replace PS1.
FE0207 Air PSOL leak EST Gas Failure Air PSOL (PSOL2) 1. Check for leaks around the Q2
Supply/ SV Test forward leak was O-rings.
detected via excessive 2. Verify that the safety valve
pressure buildup (> 100 relieves pressures above 100
cmH2O) in blocked cmH2O. Listen for the pressure
inspiratory module. relief while watching the
numeric digital display of sys-
tem pressure in the upper
screen. If the safety valve does
not crack open, replace the
safety valve.
3. Replace PSOL2.
FE0208 Wall air pressure EST Gas Failure Wall air pressure switch 1. M ake sure air supply is discon-
detected Supply/ SV Test (PS2) detected wall air nected.
after user was 2. Replace Inspiratory Electronics
prompted to PCB.
disconnect it. 3. Replace PS2.
FE0209 O2 PSOL leak EST Gas Failure Oxygen PSOL (PSOL1) 1. Check for leaks around the Q1
Supply/ SV Test forward leak was O-rings.
detected via excessive 2. Verify that the safety valve
pressure buildup (> 100 relieves pressures above 100
cmH2O) in blocked cmH2O. Listen for the pressure
inspiratory module. relief while watching the
numeric digital display of sys-
tem pressure in the upper
screen. If the safety valve does
not crack open, replace the
safety valve.
3. Replace PSOL1.
6-22 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
FE0210 SV pressure EST Gas Failure Safety valve cracking 1. Make sure To patient port is
relief failed Supply/ SV Test pressure and/ or peak blocked.
steady-state pressure is 2. Check for leaks at the O2 sen-
out of range. sor, SOL1 or PI.
3. Replace the safety valve.
4. Replace PI.
FE0211 O2 pressure not EST Gas Failure PS1 did not detect 1. Make sure oxygen supply is
detected Supply/ SV Test oxygen after user was connected.
prompted to connect it. 2. Replace Inspiratory Electronics
PCB.
FE0212 Compressor EST Gas Failure Compressor pressure Replace compressor PCB.
pressure Supply/ SV Test transducer (PC)
detected detected that
compressor was
pressurized.
FE0215 Air zero flow EST Gas Failure During zero-flow check, 1. Make sure gas supplies are dis-
check failed Supply/ SV Test air flow sensor (Q2) connected.
reads > 0.05 L/ min. 2. Make sure compressor is not
running.
3. Verify no leaks at the Q2
O-rings.
4. Check for PSOL2 leak.
5. Remove and then reseat
PSOL2.
6. Perform a flow sensor calibra-
tion.
7. Replace PSOL2.
8. Replace Q2.
FE0216 O2 zero flow EST Gas Failure During zero-flow check, 1. Make sure gas supplies are dis-
check failed Supply/ SV Test oxygen flow sensor connected.
(Q1) reads > 0.05 L/ 2. Make sure compressor is not
min. running.
3. Verify no leaks at the Q2
O-rings.
4. Check for PSOL2 leak.
5. Remove and then reseat
PSOL2.
6. Perform a flow sensor calibra-
tion.
7. Replace PSOL2.
8. Replace Q2.
FE0217 Exp zero flow EST Gas Failure During zero-flow check, 1. Make sure gas supplies are dis-
check failed Supply/ SV Test exhalation flow sensor connected.
(Q3) reads > 0.1 L/ min. 2. Make sure compressor is not
running.
3. Verify no leaks at the Q3
O-rings.
4. Perform a flow sensor calibra-
tion.
5. Replace Q3.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-23
6 Diagnostic codes
FE0301 Excessive leak EST Leak Test Failure System pressure 1. M ake sure test circuit is con-
dropped below failure nected and is not leaking.
pressure level. 2. Check for system leaks or
occlusions especially at the O2
sensor and expiratory filter.
3. Verify secure connection of
expiratory filter.
4. Check exhalation valve opera-
tion.
FE0305 Unable to EST Leak Test Failure System cannot attain 1. M ake sure test circuit is con-
establish leak test starting nected and is not leaking.
pressure pressure using oxygen 2. Check exhalation valve opera-
or air within time-out tion, and verify secure connec-
period. tion of expiratory filter.
FE0401 Accept key fails. EST GUI Failure Wrong key pressed or 1. Repeat test.
Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0402 Clear key fails. EST GUI Failure Wrong key pressed or 1. Repeat test.
Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0403 Insp. Pause key EST GUI Failure Wrong key pressed or 1. Repeat test.
fails. Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0404 Exp. Pause key EST GUI Failure Wrong key pressed or 1. Repeat test.
fails. Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0405 Man Insp fails. EST GUI Failure Wrong key pressed or 1. Repeat test.
Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0406 100% O2/ CAL EST GUI Failure Wrong key pressed or 1. Repeat test.
key fails. Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0407 Info key fails. EST GUI Failure Wrong key pressed or 1. Repeat test.
Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0408 Alarm Reset key EST GUI Failure Wrong key pressed or 1. Repeat test.
fails. Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0409 Alarm Silence EST GUI Failure Wrong key pressed or 1. Repeat test.
key fails. Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0410 Alarm Volume EST GUI Failure Wrong key pressed or 1. Repeat test.
key fails. Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0411 Screen EST GUI Failure Wrong key pressed or 1. Repeat test.
brightness fails. Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
6-24 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
FE0412 Screen contrast EST GUI Failure Wrong key pressed or 1. Repeat test.
key fails. Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0413 Screen lock key EST GUI Failure Wrong key pressed or 1. Repeat test.
fails. Keyboard Test key not pressed within 2. Replace keyboard.
15 s.
FE0501 Bad knob EST GUI Knob Failure Knob was not turned in 1. Repeat test.
Test direction as prompted 2. Replace keyboard.
within 15 s.
FE0801 SAAS EST GUI Audio Failure CLEAR key pressed to 1. Verify the GUI alarm cable con-
(Safety Audible Test indicate GUI audio nection to the GUI CPU PCB.
Alarm System) diagnostic failed. 2. Replace GUI alarm assembly.
test failed 3. Replace GUI CPU PCB.
FE0901 Bad alarm cable EST BDU Audio Failure Alarm cable voltage is 1. Make sure BD alarm cable is
Test out of range (< 3.5 or connected.
> 5.05 V). 2. Replace BD alarm.
3. Replace AI PCB.
4. Replace BD alarm cable.
FE0902 Bad power fail EST BDU Audio Failure Power failure capacitor 1. Make sure BD alarm cable is
cap Test initial voltage is out of connected.
range (< 4.5 or > 5.05 2. Replace BD alarm.
V). 3. Replace AI PCB.
4. Replace BD alarm cable.
5. Replace Motherboard PCB.
FE0903 Bad power fail EST BDU Audio Failure Power failure capacitor 1. Make sure BD alarm cable is
cap Test final voltage is out of connected.
range or RC constant 2. Replace BD alarm.
< 60 s. 3. Replace AI PCB.
FE0904 Bad BD audio EST BDU Audio Failure CLEAR key pressed to 1. Make sure BD alarm cable is
Test indicate user did not connected.
hear alarm, although 2. Replace BD alarm.
alarm was active. 3. Replace AI PCB.
4. Replace BD alarm cable.
FE1101 Safety valve EST Safety Failure Excessive safety valve Replace safety valve.
occluded System Test back pressure when
safety valve is open.
FE1102 Bad safety valve EST Safety Failure Safety valve loopback 1. Replace safety valve.
driver or System Test current is out of range 2. Replace AI PCB.
loopback during one or more of 3. Replace Inspiratory Electronics
timed test points. PCB.
FE1103 Insp check valve EST Safety Failure It took too long to 1. Make sure proper test circuit is
test failed System Test relieve excess pressure used.
through open safety 2. Make sure CV3 is not installed
valve, indicating backward.
inspiratory check valve 3. Replace CV3.
(CV3) is occluded or 4. Replace Safety Valve (SV).
test circuit is too large.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-25
6 Diagnostic codes
FE1105 Unable to EST Safety Failure Flow controller unable 1. M ake sure air supply is con-
establish flow System Test to establish and control nected and unrestricted.
gas flow at 60 L/ min. 2. Check air regulator pressure.
3. Run flow sensor calibration.
4. Switch PSOLs to see if test
passes. If it does, return PSOL1
to its original position and
replace PSOL2.
5. Replace Q2.
FE1301 Seal test failed EST Exp Valve Failure Seal test ∆ pressure is 1. Clean exhalation valve.
Seal Test above failure level. 2. Calibrate exhalation valve.
3. Run flow sensor calibration.
4. Replace exhalation valve.
5. Replace AI PCB.
6. Replace PE.
FE1302 Exp valve temp EST Exp Valve Failure Exhalation valve 1. Ensure the unit has been
OOR Seal Test magnet temperature warmed up at ambient temper-
out of range (10 to ature for at least 10 minutes.
100oC). 2. Calibrate the exhalation valve
(EV).
3. Replace exhalation valve (EV).
4. Replace AI PCB.
FE1303 Unable to EST Exp Valve Failure Flow controller unable 1. M ake sure proper test circuit it
establish exp Seal Test to establish and control used and that there are no
flow air flow measured by leaks or occlusions.
exhalation flow sensor 2. M ake sure air supply is still con-
(Q3). nected.
3. Run flow sensor calibration.
4. Replace Q3.
5. Replace Q2.
6. Replace the exhalation valve.
FE1304 Exp valve not EST Exp Valve Failure Exhalation valve table 1. Check for leaks or occlusions.
calibrated Seal Test checksum isnot valid or Correct and then run the exha-
last calibration lation valve calibration.
performed was not 2. Replace the exhalation valve.
completed successfully. 3. Replace the AI PCB.
4. Replace Q3.
5. Replace Q2.
FE1401 Bad calibration EST Exp Valve Failure Measured system 1. Calibrate exhalation valve.
Test pressure at one or more 2. Replace exhalation valve.
test points is out of 3. Replace AI PCB.
range. 4. Replace the Exhalation Pressure
Transducer PCB.
FE1402 Exp valve not EST Exp Valve Failure Exhalation valve table 1. Check for leaks or occlusions.
calibrated Test checksum isnot valid or Correct and then run the exha-
last calibration lation valve calibration.
performed was not 2. Replace the exhalation valve.
completed successfully. 3. Replace the AI PCB.
4. Replace Q3.
5. Replace Q2.
6-26 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
FE1403 Unable to EST Exp Valve Failure Flow controller unable 1. Check for leaks or occlusions.
establish exp Test to establish and control Correct and then run the exha-
flow air flow measured by lation valve calibration.
exhalation flow sensor 2. Replace the exhalation valve.
(Q3). 3. Replace the AI PCB.
4. Replace Q3.
5. Replace Q2.
FE1501 Unable to EST Exp Heater Failure Flow controller unable 1. Make sure air supply is con-
establish air flow Test to establish and control nected and unrestricted.
60 L/ min air flow. 2. Check air regulator pressure.
3. Run flow sensor calibration.
4. Switch PSOLs to see if test
passes. If it does, return PSOL1
to its original position and
replace PSOL2.
5. Replace Q2.
FE1502 Bad exp heater EST Exp Heater Failure Exhalation heater 1. Replace exhalation heater.
Test temperature did not 2. Replace AI PCB.
rise sufficiently when 3. Replace interconnect cable
heater was turned on. between exhalation compart-
ment and motherboard.
FE1503 Bad exp heater EST Exp Heater Failure Heater temperature did 1. Replace exhalation heater.
Test not drop sufficiently 2. Replace AI PCB.
during an interval after
heater was turned off.
FE2201 Low exp ∆P EV Velocity Failure Expiratory ∆ P too low, 1. Check for leaks or occlusions.
Transducer Test indicating velocity Correct and then run the exha-
transducer not lation valve calibration.
responding properly. 2. Run flow sensor calibration.
3. Replace exhalation valve.
4. Replace Q3.
5. Replace Q2.
FE2301 Nurse call stuck EST GUI Nurse Failure CLEAR key pressed to 1. Make sure test equipment is
on Call Test indicate nurse’s call connected properly to remote
(remote alarm) relay is alarm port.
stuck on when it should 2. Replace GUI CPU PCB.
be off.
FE2302 Nurse call stuck EST GUI Nurse Failure CLEAR key pressed to 1. Make sure test equipment is
off Call Test indicate nurse’s call connected properly to remote
(remote alarm) relay is alarm port.
stuck off when it should 2. Replace GUI CPU PCB.
be on.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-27
6 Diagnostic codes
NOTE:
For all "FS" codes (failures reported during SST), if the corrective action indicated is to run an EST
to further diagnose the problem, perform the repair after the EST, verify the fix in EST, and then
rerun SST in order to enter normal ventilator operation. If there is a failed SST test result in
memory, normal operation until is prevented until the failed SST test is rerun and passes.
FS0001 O2 flow sensor SST Flow Failure Oxygen flow sensor 1. Correct the leak or occlusion in
cross check Sensor Test (Q1) cross-check failed. the SST patient circuit and
failed rerun or try known good
patient circuit and filter set.
2. Verify that the ventilator sys-
tem has no leaks or occlusions
by running the first four tests in
EST to run the leak test.
a. If the leak test fails in EST,
troubleshoot the expiratory
filter or ventilator system for
leaks or occlusions.
b. If the leak test passes in EST,
troubleshoot the SST patient
circuit.
3. Run a flow sensor calibration.
4. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the air side, return Q2 back to
its original position and replace
Q1.
5. Replace Q3
6. Switch PSOLs to see if the fail-
ure transfers to the air side. If
yes, return PSOL2 back to its
original position and replace
PSOL1.
FS0002 O2 PSOLcurrent SST Flow Failure Oxygen PSOL (PSOL1) 1. M ake sure oxygen supply is
out of range Sensor Test command current is connected and unrestricted.
out of range with 2. Verify that the oxygen supply is
respect to oxygen flow good.
sensor (Q1). 3. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
4. Run a flow sensor calibration.
5. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the air side, return Q2 back to
its original position and replace
Q1.
6. Replace Q3.
7. Switch PSOLs to see if the fail-
ure transfers to the air side. If
yes, return PSOL2 back to its
original position and replace
PSOL1.
6-28 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
FS0003 Air flow sensor SST Flow Failure Air flow sensor (Q2) 1. Correct the leak or occlusion in
cross check Sensor Test cross-check failed. the SST patient circuit and
failed rerun or try known good
patient circuit and filter set.
2. Verify that the ventilator sys-
tem has no leaks or occlusions
by running the first four tests in
EST to run the leak test.
3. If the leak test fails in EST, trou-
bleshoot the expiratory filter or
ventilator system for leaks or
occlusions.
4. If the leak test passes in EST,
troubleshoot the SST patient
circuit.
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the oxygen side, return Q1
back to its original position and
replace Q2.
7. Replace Q3.
8. Switch PSOLs to see if the fail-
ure transfers to the oxygen
side. If yes, return PSOL1 back
to its original position and
replace PSOL2.
FS0004 Air PSOL current SST Flow Failure Air PSOL (PSOL2) 1. Make sure air supply is con-
out of range Sensor Test command current is nected and unrestricted.
out of range with 2. Verify that the air supply is
respect to air flow good.
sensor (Q2). 3. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
4. Run a flow sensor calibration.
5. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the oxygen side, return Q1
back to its original position and
replace Q2.
6. Replace Q3.
7. Switch PSOLs to see if the fail-
ure transfers to the oxygen
side. If yes, return PSOL1 back
to its original position and
replace PSOL2.
FS0005 Gas not SST Flow Failure No gas is available Connect gas supply.
connected Sensor Test (neither PS1, PS2, nor
PC detects pressure).
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-29
6 Diagnostic codes
FS0006 Unable to SST Flow Failure Flow controller unable 1. M ake sure oxygen supply is
establish O2 Sensor Test to establish and control connected and unrestricted.
flow oxygen flow at 60, 5, 2. Verify that the oxygen supply is
and 1 L/ min. good.
3. Check regulated oxygen pres-
sure.
4. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the air side, return Q2 back to
its original position and replace
Q1.
7. Replace exhalation flow sensor
(Q3).
8. Switch PSOLs to see if the fail-
ure transfers to the air side. If
yes, return PSOL2 back to its
original position and replace
PSOL1.
FS0007 Unable to SST Flow Failure Flow controller unable 1. M ake sure air supply is con-
establish air flow Sensor Test to establish and control nected unrestricted.
air flow at 60, 5, and 2. Verify that the air supply is ade-
1 L/ min. quate.
3. Check regulated air pressure.
4. Verify that the system has no
leaks or occlusions. If not sure,
run the first four tests in EST to
run the leak test.
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the
flow sensor calibration and
EST. If the failure transfers to
the oxygen side, return Q1
back to its original position and
replace Q2.
7. Replace exhalation flow sensor
(Q3).
8. Switch PSOLs to see if the fail-
ure transfers to the oxygen
side. If yes, return PSOL1 back
to its original position and
replace PSOL2.
6-30 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
FS0008 O2 zero flow SST Flow Failure Inspiratory flow > 0.153 1. Verify no leaks at the Q1
check failed Sensor Test L/ min with oxygen O-rings.
PSOL (PSOL1) 2. Run EST Gas Supply Test to
commanded to 0 check for PSOL1 leak.
(closed). 3. Remove and then reseat
PSOL1.
4. Replace PSOL1.
5. Perform a flow sensor calibra-
tion.
6. Replace Q1.
FS0009 Air zero flow SST Flow Failure Inspiratory flow > 0.153 1. Verify no leaks at the Q2
check failed Sensor Test L/ min with air PSOL O-rings.
(PSOL2) commanded 2. Run EST Gas Supply Test to
to 0 (closed). check for PSOL2 leak.
3. Remove and then reseat
PSOL2.
4. Replace PSOL2.
5. Perform a flow sensor calibra-
tion.
6. Replace Q2.
FS0101 Inspiratory SST circuit Failure Inspiratory pressure 1. Replace Inspiratory Electronics
autozero out of pressure test transducer ADC count PCB
range at 0 cmH2O is out of 2. Replace AI PCB.
range.
FS0102 Expiratory SST circuit Failure Expiratory pressure 1. Replace exhalation transducer
autozero out of pressure test transducer ADC count PCB.
range at 0 cmH2O is out of 2. Replace AI PCB.
range. 3. Replace SOL1.
FS0103 Failed to reach SST circuit Failure Unable to build 1. Verify that the patient circuit
test pressure pressure test pressure (air) to 10 system has no leaks or occlu-
cmH2O at sions or use a known good
5 L/ min within time-out patient circuit and filter set.
period. 2. Verify that the air supply or
compressor is available and
adequate.
3. Verify that the ventilator sys-
tem has no leaks or occlu-
sions. If not sure, run the first
four tests in EST to run the leak
test.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-31
6 Diagnostic codes
FS0104 Cross-check SST circuit Failure Inspiratory/ expiratory 1. Verify that the patient circuit
failed pressure test pressure transducer system has no leaks or occlu-
readings at 10 cmH20 sions or use a known good
test pressure are too far patient circuit and filter set.
apart. 2. Verify that the ventilator sys-
tem has no leaks or occlu-
sions. If not sure, run the first
four tests in EST to run the leak
test.
3. If the leak test passes in EST,
exit out of EST. Review the
data for the Circuit Pressure
Test to determine which pres-
sure transducer is out of range:
PI or PE. Replace applicable
pressure transducer.
4. Replace AI PCB.
FS0105 Bad insp SST circuit Failure Inspiratory pressure 1. Replace SOL1.
autozero pressure test reading taken after 2. Replace Inspiratory Electronics
solenoid actuating inspiratory PCB.
pressure transducer 3. Replace AI PCB.
autozero solenoid
(SOL1) is out of range
(-0.60 to 0.60 cmH2O).
FS0106 Bad exp SST circuit Failure Expiratory pressure 1. Replace SOL2.
autozero sol pressure test reading taken after 2. Replace exhalation transducer
actuating expiratory PCB.
transducer autozero 3. Replace AI PCB.
solenoid (SOL2) is out
of range (-0.60 to
0.60 cmH2O).
FS0107 Cross-check SST circuit Failure Inspiratory/ expiratory 1. Replace the expiratory bacteria
failed pressure test pressure transducer filter.
readings at 50 or 100 2. Verify that the system has no
cmH2O test pressure leaks or occlusions. If not sure,
are too far apart. run the first four tests in EST to
run the leak test.
3. If the leak test passes in EST,
exit out of EST. Review the
data for the EST Circuit Pres-
sure Test to determine which
pressure transducer is out of
range: PI or PE. Replace appli-
cable pressure transducer.
4. Replace AI PCB.
FS0108 Failed to reach SST circuit Failure Unable to build 1. M ake sure air supply or com-
test pressure pressure test pressure (air) to 50 or pressor is available.
100 cmH2O at 2. Check for system leak.
5 L/ min within time-out
period.
6-32 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
FS0201 Excessive leak SST Circuit leak Failure System pressure 1. Make sure patient circuit is
dropped below failure connected and is not leaking.
pressure level. 2. Check exhalation valve opera-
tion, and verify secure connec-
tions of bacteria filters, water
traps, and humidifier.
FS0205 Unable to SST Circuit leak Failure Unable to attain leak 1. Make sure patient circuit is
establish test starting pressure connected and is not leaking.
pressure within time-out period 2. Check exhalation valve opera-
using oxygen or air. tion, and verify secure connec-
tions of bacteria filters, water
traps, and humidifier.
FS0301 Unable to SST Expiratory Failure Flow controller unable Make sure gas supply is still
establish flow filter to establish and control connected.
gas flow.
FS0306 Patient circuit SST Expiratory Failure A test flow was used to Make sure patient circuit is
not filter verify that circuit was disconnected at expiratory filter.
disconnected disconnected, and an
expiratory flow was
detected.
FS0307 Patient circuit SST Expiratory Failure A test flow was used to Make sure patient circuit is
not filter verify that circuit was connected at expiratory filter.
reconnected reconnected, and an
expiratory flow was not
detected.
FS0310 Low expiratory SST Expiratory Failure Pressure drop across 1. Repeat test, following direc-
filter ∆P filter expiratory filter < 0.1 tions more closely.
cmH2O. 2. Replace filter.
FS0401 Unable to SST Circuit Failure Flow controller unable Make sure gas supply is connected.
establish flow Resistance to establish and control
gas flow.
FS0402 Occluded SST Circuit Failure Inspiratory limb Check for occluded patient tubing.
inspiratory limb Resistance pressure > 12.5 cmH2O
(adult) or 7.5 cmH2O
(pediatric), indicating
occlusion.
FS0404 Unable to reach SST Circuit Failure During characterization Check for kinked or occluded
min peak flow Resistance of total circuit patient tubing.
resistance over a range
of flows, peak flow was
less than failure
threshold.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-33
6 Diagnostic codes
FS0405 Occluded SST Circuit Failure Expiratory limb Check for occluded patient tubing.
exhalation limb Resistance pressure > 12.5 cmH2O
(adult) or 7.5 cmH2O
(pediatric), indicating
occlusion.
FS0408 Insp limb SST Circuit Failure Inspiratory limb 1. M ake sure inspiratory filter is
resistance low Resistance pressure < 0.2 cmH2O. installed.
2. Repeat test, following direc-
tions more closely.
3. Replace inspiratory filter.
4. Replace patient circuit.
FS0409 Exp limb SST Circuit Failure Expiratory limb Replace patient circuit.
resistance low Resistance pressure < 0.2 cmH2O.
FS0410 Unable to reach SST Circuit Failure During characterization Check for kinked or occluded
min peak flow Resistance of total circuit patient tubing.
resistance over a range
of flows, peak flow was
less than failure
threshold.
FS0412 Wye not SST Circuit Failure Unable to pressurize 1. Block wye when prompted.
blocked Resistance system within time-out 2. Check for circuit leaks.
period.
FS0501 Unable to SST Failure System cannot Check patient circuit for leaks.
pressurize Compliance pressurize to one of the Replace as necessary.
circuit calibration test pressure points.
FS0502 Unable to SST Failure System cannot Check patient circuit for leaks.
pressurize Compliance pressurize to one of the Replace as necessary.
circuit calibration test pressure points.
FS0503 Compliance SST Failure Out-of-range pressure 1. Repeat test, making sure circuit
calculation Compliance ratio calculation. is undisturbed during test.
failure calibration 2. Repeat SST to retest pressure
transducers.
3. Run EST and verify all pneu-
matics tests pass.
FS0504 Excessive SST Failure High compliance > 1. M ake sure correct patient tub-
compliance Compliance 12 mL/ cmH2O (adult) ing type was specified in SST.
calibration or 9 mL/ cmH2O 2. Replace patient circuit.
(pediatric).
FS0506 Compliance low SST Failure High compliance falls 1. M ake sure correct patient tub-
Compliance below 1.05 mL/ cmH2O ing type was specified in SST.
calibration 2. Replace patient circuit.
FS0508 Excessive SST Failure Low compliance > 1. M ake sure correct patient tub-
compliance Compliance 12 mL/ cmH2O (adult) ing type was specified in SST.
calibration or 9 mL/ cmH2O 2. Replace patient circuit.
(pediatric).
FS0510 Compliance low SST Failure Low compliance falls 1. M ake sure correct patient tub-
Compliance below 1.05 mL/ cmH2O ing type was specified in SST.
calibration 2. Replace patient circuit.
6-34 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
HN0129 NM I: DRAM Other (BDU) BDU reset 1. Check GUI/ BD cable connec-
Parity Error tions.
2. Replace power supply.
3. Replace BD CPU PCB.
HN0130 NM I: Ethernet Other (BDU) BDU reset 1. Check GUI/ BD cable connec-
Parity Error tions.
2. Replace power supply.
3. Replace BD CPU PCB.
KB0001 Bad safety valve Background Vent inop Measured voltage on 1. Replace safety valve.
switched side check (BDU) safety valve switched 2. Replace AI PCB.
side indicates valve is 3. Replace Inspiratory Electronics
not in expected state. PCB.
4. Replace interconnect cable
between the mother board and
the Inspiratory Electronics PCB.
KB0002 Bad expiratory Background Vent inop Exhalation flow sensor 1. If possible, run EST and use the
flow check (BDU) (Q3) reading out of Flow Sensor Cross Check test
range to determine if Q3 is failing.
Replace Q3 if applicable.
2. Perform flow sensor calibra-
tion.
3. Replace AI PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-35
6 Diagnostic codes
KB0003 Bad O2 PSOL Background Vent inop Oxygen PSOL (PSOL1) 1. If possible, run EST to deter-
current check (BDU) current out of range mine if PSOL1 is failing. Switch
PSOLs to see if the failure trans-
fers to the other side. Replace
PSOL1 if applicable.
2. Replace AI PCB.
3. Replace Inspiratory Electronics
PCB.
4. Replace interconnect cable
between the motherboard and
the Inspiratory Electronics PCB.
KB0004 Bad air PSOL Background Vent inop Air PSOL (PSOL2) 1. If possible, run EST to deter-
current check (BDU) current out of range mine if PSOL2 is failing. Switch
PSOLs to see if the failure trans-
fers to the other side. Replace
PSOL2 if applicable.
2. Replace AI PCB.
3. Replace Inspiratory Electronics
PCB.
4. Replace interconnect cable
between the motherboard and
the Inspiratory Electronics PCB.
KB0005 Bad exp motor Background Vent inop Current to exhalation 1. Perform exhalation valve (EV)
current check (BDU) valve motor out of calibration.
range 2. Replace EV.
3. Replace AI PCB.
KB0007 Bad exp Background Vent inop PE reading out of range 1. Replace exhalation transducer
pressure check (BDU) PCB.
2. Replace AI PCB.
KB0008 Bad insp Background Vent inop PI reading out of range 1. Replace Inspiratory Electronics
pressure check (BDU) PCB.
2. Replace AI PCB.
KB0009 Air flow out of Background Vent inop Q2 reading too high 1. Perform flow sensor calibra-
range HIGH check (BDU) tion.
2. Replace Q2.
3. Replace AI PCB.
KB0012 O2 flow out of Background Vent inop Q1 reading too high 1. Perform flow sensor calibra-
range HIGH check (BDU) tion.
2. Replace Q1.
3. Replace AI PCB.
KB0016 Bad BD 10V Background Vent inop BDU +10 V supply out Replace AI PCB.
supply check (BDU) of range
KB0017 Bad BD 12 V Background Vent inop BDU +12 V supply out 1. Replace AI PCB.
supply check (BDU) of range 2. Replace power supply or other
power input components.
KB0018 Bad BD 15 V Background Vent inop BDU +15 V supply out 1. Replace AI PCB.
supply check (BDU) of range 2. Replace power supply or other
power input components.
6-36 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
KB0019 Bad BD -15 V Background Vent inop BDU -15 V supply out 1. Replace AI PCB.
check (BDU) of range 2. Replace power supply or other
power input components.
KB0022 Bad BD 5 V Background Vent inop BDU +5 V supply out of 1. Replace AI PCB.
supply check (BDU) range 2. Replace power supply or other
power input components.
KB0023 O2 PSOL stuck Background Vent inop Oxygen PSOL (PSOL1) 3. If possible, run EST to deter-
check (BDU) command current out mine if PSOL1 is failing. Switch
of range PSOLs to see if the failure trans-
fers to the other side. Replace
PSOL1 if applicable.
4. Replace AI PCB.
5. Replace Inspiratory Electronics
PCB.
6. Replace interconnect cable
between the motherboard and
the Inspiratory Electronics PCB.
KB0024 Air PSOL stuck Background Vent inop Air PSOL (PSOL2) 1. If possible, run EST to deter-
check (BDU) command current out mine if PSOL2 is failing. Switch
of range PSOLs to see if the failure trans-
fers to the other side. Replace
PSOL2 if applicable.
2. Replace AI PCB.
3. Replace Inspiratory Electronics
PCB.
4. Replace interconnect cable
between the motherboard and
the Inspiratory Electronics PCB.
KB0025 Air PSOL stuck Background Vent inop Air PSOL (PSOL2) 1. If possible, run EST to deter-
open check (BDU) command current mine if PSOL2 is failing. Switch
reading indicates PSOL PSOLs to see if the failure trans-
stuck open fers to the other side. Replace
PSOL2 if applicable.
2. Replace AI PCB.
3. Replace Inspiratory Electronics
PCB.
4. Replace interconnect cable
between the motherboard and
the Inspiratory Electronics PCB.
KB0026 O2 PSOL stuck Background Vent inop PSOL1 command 1. If possible, run EST to deter-
open check (BDU) current reading mine if PSOL1 is failing. Switch
indicates PSOL stuck PSOLs to see if the failure trans-
open fers to the other side. Replace
PSOL1 if applicable.
2. Replace AI PCB.
3. Replace Inspiratory Electronics
PCB.
4. Replace interconnect cable
between the motherboard and
the Inspiratory Electronics PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-37
6 Diagnostic codes
KB0030 Bad safety valve Background Vent inop Current to safety valve 1. Replace safety valve.
current check (BDU) out of range 2. Replace AI PCB.
3. Replace Inspiratory Electronics
PCB.
4. Replace interconnect cable
between the motherboard and
the Inspiratory Electronics PCB.
KB0031 Insp pressure Background Vent inop Inspiratory pressure 1. Replace Inspiratory Electronics
stuck check (BDU) transducer (PI) reading PCB.
indicates transducer is 2. Replace AI PCB.
stuck
KB0032 Exp pressure Background Vent inop Expiratory pressure 1. Replace exhalation transducer
stuck check (BDU) transducer (PE) reading PCB.
indicates transducer is 2. Replace AI PCB.
stuck
KB0033 Insp pressure Background Vent inop Inspiratory pressure 1. Replace inspiratory pressure
autozero offset check (BDU) transducer (PI) reading transducer autozero solenoid
failed following autozero out (SOL1).
of range 2. Replace Inspiratory Electronics
PCB.
3. Replace AI PCB.
KB0034 Exp pressure Background Vent inop Expiratory pressure 1. Replace expiratory pressure
autozero offset check (BDU) transducer (PE) reading transducer autozero solenoid
failed following autozero out (SOL2).
of range 2. Replace exhalation transducer
PCB.
3. Replace AI PCB.
KB0044 BD NOVRAM Background Vent inop BD NOVRAM failure Replace BD CPU PCB.
checksum error check (BDU)
KB0053 BD EEPROM Background Vent inop BD EEPROM failure Replace BD CPU PCB.
checksum error check (BDU)
KB0064 LV Ref out of Background Vent inop Low-voltage reference Replace AI PCB.
range check (BDU) out of range
KB0065 SV current out Background Vent inop Safety valve current out 1. Replace safety valve.
of range check (BDU) of range 2. Replace Inspiratory Electronics
PCB.
3. Replace AI PCB.
4. Replace interconnect cable
between the motherboard and
the Inspiratory Electronics PCB.
6-38 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
KB0079 BK vent inop Background Vent inop Ventilator inoperative 1. Run the VENT INOP test.
occurred check (BDU) signal read by BDU is a. If the test passes, run EST to
active. This could mean reset the Device Alert. Run
GUI has declared a the unit for 48 hours and
ventilator inoperative then put back into service if
condition. the code cannot be dupli-
cated.
b. If the test fails, troubleshoot
per Table 6-5, VENT INOP
Test troubleshooting.
KP0002 Integer Unit POST (BDU) Vent inop Replace BD CPU PCB.
Test
KP0003 DRAM Refresh POST (BDU) Vent inop Replace BD CPU PCB.
Timer Test
KP0004 Kernel DRAM POST (BDU) Vent inop Replace BD CPU PCB.
Test
KP0005 Boot EPROM POST (BDU) Vent inop Replace BD CPU PCB.
Checksum Test
KP0006 POST Phase 2 POST (BDU) Vent inop Replace BD CPU PCB.
Initialization
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-39
6 Diagnostic codes
KP0009 Rolling Thunder POST (BDU) Vent inop POST has started (but 1. Replace AI PCB.
Test not run to completion) 2. Replace BD CPU PCB.
three or more times, for 3. Replace motherboard.
reasons unrelated to
power.
KP0011 Time of Day POST (BDU) Vent inop Replace BD CPU PCB.
Clock Test
KP0012 Timer Test POST (BDU) Vent inop Timer failed Replace BD CPU PCB.
comparison test with
real-time clock
(external timing
reference)
or
Timer failed to generate
interrupt
KP0013 Watchdog POST (BDU) Vent inop Watchdog timer did Replace BD CPU PCB.
Timer Test not time out as
expected
KP0015 FLASH Memory POST (BDU) Vent inop Replace BD CPU PCB.
Checksum Test
KP0082 Bus Timer Test POST (BDU) Vent inop Activity has ceased on Replace BD CPU PCB.
local bus
or
Bus time-out circuit did
not generate an
interrupt as expected
KP0083 NMI Source POST (BDU) Vent inop Replace BD CPU PCB.
Register Test
KP0084 POST DRAM POST (BDU) Vent inop Replace BD CPU PCB.
Test
6-40 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
KP0087 Unexpected POST (BDU) Vent inop POST has been invoked 1. Check associated errors in the
Reset Umpire three times in 24 System Diagnostic and System
Test operational hours due Information log to determine
to unexpected resets. why POST was invoked three
times within 24 hours.
If the codes indicate that the BD
CPU generated the resets,
replace the BD CPU (Other
codes may be present
indicating that the GUI lost
communications with the BD).
2. Run a complete EST to check
for analog device failures.
KP0089 Floating Point POST (BDU) Vent inop Replace BD CPU PCB.
Unit Test
KP0090 DRAM Parity POST (BDU) Vent inop Replace BD CPU PCB.
Circuit Test
KP0113 Analog Interface POST (BDU) Vent inop 1. Replace BD CPU PCB.
PCB Test 2. Replace AI PCB.
3. Replace motherboard.
KP0114 ADC Test POST (BDU) Vent inop Analog-to-digital 1. Replace AI PCB.
converter data not 2. Replace motherboard.
available to CPU as
expected
KP0115 DAC Test POST (BDU) Vent inop Digital-to-analog 1. Replace AI PCB.
converter data not as 2. Replace motherboard.
expected
KP0116 Analog Devices POST (BDU) Vent inop Critical analog data out 1. Refer to Section 6.8 to identify
Test of range which analog device is failing.
2. Run EST to get more informa-
tion on the analog device fail-
ure.
3. Replace the AI PCB.
4. Replace the motherboard.
5. Replace the interconnect cable
between the Inspiratory Elec-
tronics PCB and the mother-
board.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-41
6 Diagnostic codes
KP0120 Safe State POST (BDU) Vent inop Places PSOLs and 1. Look up associated codes
System Test exhalation valve into reported in Diagnostic Log
safe state and verifies and/ or the BD CPU LED array
that they cannot be to get more information about
commanded. other possible system, soft-
ware, or hardware faults.
Troubleshoot those codes.
2. Run a Ground Isolation Test to
troubleshoot shorts.
3. Try another inspiratory mod-
ule and see if the code can be
reset by running the VENT
INOP, EST and POST tests.
a. If the unit passes with a
known good inspiratory
module, replace the Inspira-
tory Electronics PCB.
b. If the unit fails, replace the
inspiratory blind mate inter-
connect cable.
4. Replace the Inspiratory Elec-
tronics PCB.
5. Replace the AI PCB.
6. Replace the BD CPU PCB.
KP0128 Download POST (BDU) Vent inop Replace BD CPU PCB.
operating
system (OS)
Boot
KP0130 PB-MON Boot POST (BDU) Vent inop Replace BD CPU PCB.
LB0006 Bad exhalation Background Alert Exhalation valve coil 1. Replace exhalation valve.
valve coil temp. check (BDU) temperature out of 2. Replace AI PCB.
range 3. Replace interconnect harness
between the exhalation com-
partment connector and the
motherboard.
LB0010 Air flow out of Background Alert Air flow sensor (Q2) 1. Perform flow sensor calibra-
range LOW check (BDU) reading too low tion.
2. Replace Q2.
3. Replace AI PCB.
6-42 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
LB0011 Bad air flow Background Alert Air flow sensor (Q2) 1. Replace Q2.
temperature check (BDU) thermistor reading out 2. Replace AI PCB.
of range
LB0013 O2 flow out of Background Alert Oxygen flow sensor 1. Ensure that the ventilator has
range LOW check (BDU) (Q1) reading too low been warmed up for at least 10
minutes in service mode at
ambient temperature.
2. Run a flow sensor calibration.
3. Replace Q1.4.Replace the AI
PCB.
LB0014 Bad O2 flow Background Alert Oxygen flow sensor 1. Replace Q1.
temp. check (BDU) (Q1) thermistor reading 2. Replace AI PCB.
out of range
LB0015 Bad expiratory Background Alert Expiratory flow sensor 1. Replace Q3.
flow temp check (BDU) (Q3) thermistor reading 2. Replace AI PCB.
out of range
LB0020 Bad GUI 12V Background Alert GUI +12 V supply out of 1. Check cabling from BD CPU
supply check (BDU) range PCB to GUI CPU PCB.
2. Replace AI PCB.
3. Replace power supply.
4. Check whether or not the cable
from the GUI to the BDU was
reconnected to the unit while
the ventilator power was on. If
yes, run EST to reset the
DEVICE ALERT.
LB0021 Bad GUI 5V Background Alert GUI +5 V supply out of 1. Check cabling from BD CPU
supply check (BDU) range. PCB to GUI CPU PCB.
2. Replace AI PCB.
3. Replace power supply.
4. Check whether or not the cable
from the GUI to the BDU was
reconnected to the unit while
the ventilator power was on. If
yes, run EST to reset the
DEVICE ALERT.
LB0028 Bad O2 sensor Background Alert Oxygen sensor 1. Calibrate oxygen sensor.
OOR check (BDU) (percentage) reading 2. Replace oxygen sensor.
out of range
LB0035 Bad power fail Background Alert Power fail capacitor 1. Replace AI PCB.
capacitor check (BDU) voltage out of range 2. Replace power supply.
voltage
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-43
6 Diagnostic codes
LB0036 Alarm cable Background Alert Alarm cable voltage out 1. Check cabling to main alarm.
error check (BDU) of range 2. Replace BD alarm.
3. Replace AI PCB.
LB0043 ac switch stuck Background Alert Power switch is in off 1. Replace power switch.
check (BDU) state but power fail 2. Replace AI PCB.
NMI did not occur
LB0045 BD Time of Day Background Alert Real-time clock failure Replace BD CPU PCB.
failed check (BDU)
LB0048 Bad backup Background Alert BPS voltage out of 1. Replace BPS battery pack.
power supply check (BDU) range 2. Replace AI PCB.
voltage 3. Replace BPS PCB.
LB0049 Bad backup Background Alert BPS current out of 1. Replace BPS battery pack.
power supply check (BDU) range 2. Replace AI PCB.
current 3. Replace BPS PCB.
LB0050 Bad backup Background Alert Model information Replace BPS PCB.
power supply check (BDU) incorrect
model
LB0051 Bad exp heater Background Alert Exhalation heater 1. Replace exhalation heater.
check (BDU) thermistor reading out 2. Replace AI PCB.
of range 3. Replace interconnect cable
between the exhalation com-
partment and the mother-
board.
LB0057 Compressor bad Background Alert Data in compressor Replace compressor PCB.
eprom data check (BDU) EPROM not as expected
LB0058 Loss of GUI Background Alert 1. This code indicates that the BD
communication check (BDU) logged a loss of communica-
tions with the GUI. Check for
codes generated by the GUI
indicating a GUI fault.
2. Check cabling from BD CPU
PCB to the GUI CPU PCB.
3. Replace GUI CPU PCB or BD
CPU PCB.
6-44 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-45
6 Diagnostic codes
LC1xxx Varies Communi- Alert 1. Ensure that the green LED is lit
cations error on the front of the BPS.
(BDU) a. If the green LED is not lit,
allow the BPSto charge until
the green LED lights.
b. If the green LED does not
light (within 6-8 hours or
less), replace the battery
pack.
c. Replace the BPS PCB
2. Once the green LED is lit, run
EST to test the BPS and use EST
diagnostics to determine if fur-
ther corrective action is
needed.
LP0008 Kernel POST (BD) Alert Noncritical NOVRAM Replace the BD CPU PCB.
NOVRAM Test data failure
LP0087 Unexpected POST (BDU) Alert POST was invoked due 1. Check all logs to see associated
Reset Umpire to a watchdog time-out errors occurring around the
Test (unexpected reset) same time and date as this
code.
2. Run EST to get more informa-
tion.
LP0088 POST NOVRAM POST (BDU) Alert Noncritical NOVRAM 1. Check all logs to see associated
Test data failure errors occurring around the
same time and date as this
code.
2. Run EST to get more informa-
tion.
LP0112 Ventilator INOP POST (BDU) Alert Ventilator inoperative 1. Replace the interconnect cable
Test signal not in expected between the Inspiratory Elec-
state tronics PCB and the mother-
board.
2. Run VENT INOP Test.
3. Replace BD CPU PCB.
LP0114 ADC Test POST (BD) Alert Replace AI PCB.
6-46 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
LP0116 Analog Devices POST (BDU) Alert Noncritical analog data 1. Refer to Table 6-3: "Address
Test out of range codes for BDU POST analog
devices test errors".
2. Replace AI PCB.
LP0121 Service Switch POST (BDU) Alert Service switch appears 1. Turn off ventilator. Release
Stuck Test to be closed. (Only switch.
requests to enter 2. Check for mechanical obstruc-
service mode made tion to switch.
after this test are 3. Replace BD CPU PCB.
honored by system.)
LP0122 ac Voltage Test POST (BDU) Alert ac monitor voltage not 1. Make sure ventilator is plugged
within set nominal line into ac source.
voltage ±20% 2. Make sure nominal line voltage
is set correctly.
U04004 Assertion Failure Background Alert Can occur if there is an 1. Check all cable connections.
check (GUI) intermittent connection 2. Perform Ground Isolation Test.
to any digital I/ O line. 3. Replace GUI CPU PCB.
U05013 Assertion Failure Background Alert Can occur if there is an 1. Check all cable connections.
check (GUI) intermittent connection 2. Perform Ground Isolation Test.
to any digital I/ O line. 3. Replace GUI CPU PCB.
U08002 Assertion Failure Background Alert Occurs when the GUI is 1. Check for wet expiratory or
check (BDU) overloaded with invalid inspiratory filter.
data from the serial 2. Check all cable connections:
port. Caused by the GUI, Serial, Internal cables.
host device transmit- 3. Review the ventilator logs,
ting bad data that is including patient logs, observe
generating over-run, any occurrences of rapidly
parity, or framing repeating messages. This pat-
errors. The GUI gener- tern may contribute to a buffer
ates a soft reset or overflow situation that the ven-
assertion to clear the tilator will attempt to clear by
condition and resyn- running POST.
chronize with the BDU. 4. Perform Ground Isolation Test.
5. Replace GUI CPU PCB.
U08012 Assertion Failure Background Alert The GUI generates a 1. Check all cable connections.
check (GUI) soft reset or assertion to 2. Perform Ground Isolation Test.
clear the condition and 3. Replace GUI CPU PCB.
re synchronize with the
BDU.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-47
6 Diagnostic codes
U14019 Assertion Failure Background Alert The GUI generates a 1. Check all cable connections.
check (GUI) soft reset or assertion to 2. Perform Ground Isolation Test.
clear the condition and 3. Replace GUI CPU PCB.
resynthesized with the
BDU.
U14021 Assertion Failure Background Alert If the time stamps 1. Check all cable connections.
check (GUI) between the GUI and 2. Perform Ground Isolation Test.
the BD are out of range, 3. Replace GUI CPU PCB.
the GUI generates a soft
reset or assertion to
clear the condition
and re synchronize with
the BDU.
UT0002 Assertion Failure Background Alert A reset of the GUI has NOTE:
check (GUI) occurred as a result of a Refer to Section 6.10.1 for
hardware-trap further information
detection. regarding for this error
code.
VN0129 NMI: DRAM Other (GUI) GUI reset 1. Check GUI/ BD cable connec-
Parity Error tions.
2. Replace power supply.
3. Replace GUI CPU PCB.
VN0130 NMI: Ethernet Other (GUI) GUI reset 1. Check GUI/ BD cable connec-
Parity Error tions.
2. Replace power supply.
3. Replace GUI CPU PCB.
VN0132 NMI: +5 V dc Other (GUI) GUI reset 1. Check GUI/ BD cable connec-
Overvoltage tions.
2. Replace power supply.
3. Replace GUI CPU PCB.
VN0136 NMI: +12 V dc Other (GUI) GUI reset 1. Check GUI/ BD cable connec-
Overvoltage tions.
2. Replace power supply.
3. Replace GUI CPU PCB.
VN0144 NMI: +12 V dc Other (GUI) GUI reset 1. Check GUI/ BD cable connec-
Undervoltage tions.
2. Replace power supply.
3. Replace GUI CPU PCB.
XB0066 Monitor alarms Background Vent inop Call Technical Support or your
fail checks (GUI) Puritan Bennett representative.
XB0067 Monitor apnea Background Vent inop Call Technical Support or your
alarm fails checks (GUI) Puritan Bennett representative.
XB0068 Monitor apnea Background Vent inop Call Technical Support or your
int fails checks (GUI) Puritan Bennett representative.
XB0069 Monitor hip fails Background Vent inop Call Technical Support or your
checks (GUI) Puritan Bennett representative.
6-48 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
XB0070 Monitor insp Background Vent inop Call Technical Support or your
time fails checks (GUI) Puritan Bennett representative.
XB0071 Monitor no data Background Vent inop Call Technical Support or your
checks (GUI) Puritan Bennett representative.
XB0072 Monitor data Background Vent inop Call Technical Support or your
corrupted checks (GUI) Puritan Bennett representative.
XP0001 Processor POST (GUI) GUI inop Replace GUI CPU PCB.
Initialization
XP0002 Integer Unit POST (GUI) GUI inop Replace GUI CPU PCB.
Test
XP0003 DRAM Refresh POST (GUI) GUI inop Replace GUI CPU PCB.
Timer Test
XP0004 Kernel DRAM POST (GUI) GUI inop Replace GUI CPU PCB.
Test
XP0005 Boot EPROM POST (GUI) GUI inop Replace GUI CPU PCB.
Checksum Test
XP0006 POST Phase 2 POST (GUI) GUI inop Replace GUI CPU PCB.
Initialization
XP0007 Addressing POST (GUI) GUI inop Replace GUI CPU PCB.
Mode Test
XP0008 Kernel POST (GUI) GUI inop Replace GUI CPU PCB.
NOVRAM Test
XP0010 Interrupt POST (GUI) GUI inop Replace GUI CPU PCB.
Controller Test
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-49
6 Diagnostic codes
XP0011 Time of Day POST (GUI) GUI inop Replace GUI CPU PCB.
Clock Test
XP0012 Timer Test POST (GUI) GUI inop Replace GUI CPU PCB.
XP0013 Watchdog POST (GUI) GUI reset Watchdog timer did 1. M ake sure ventilator is plugged
Timer Test not time out as in to wall power and BPS is
expected fully charged.
2. Replace GUI CPU PCB.
XP0015 FLASH Memory POST (GUI) GUI inop Replace GUI CPU PCB.
Checksum Test
XP0081 Memory POST (GUI) GUI inop Replace GUI CPU PCB.
Management
Unit Test
XP0082 Bus Timer Test POST (GUI) GUI inop Replace GUI CPU PCB.
XP0083 NM I Source POST (GUI) GUI inop Replace GUI CPU PCB.
Register Test
XP0084 POST DRAM POST (GUI) GUI inop Replace GUI CPU PCB.
Test
XP0087 Unexpected POST (GUI) GUI inop POST has been invoked Check associated errors in log. Call
Reset Umpire three times in 24 Technical Support or your Puritan
Test operational hours due Bennett representative.
to unexpected resets.
XP0089 Floating Point POST (GUI) GUI inop Replace GUI CPU PCB.
Unit Test
XP0090 DRAM Parity POST (GUI) GUI inop Replace GUI CPU PCB.
Circuit Test
XP0097 SAAS Self-Test POST (GUI) GUI inop Replace GUI CPU PCB.
Start
6-50 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
XP0098 SAAS Self-Test POST (GUI) GUI inop Replace GUI CPU PCB.
End
XP0128 Download OS POST (GUI) GUI inop Replace GUI CPU PCB.
Boot
XP0129 Application OS POST (GUI) GUI inop Replace GUI CPU PCB.
Boot
XP0130 PB-M ON Boot POST (GUI) GUI inop Replace GUI CPU PCB.
XP0131 Application POST (GUI) GUI inop Replace GUI CPU PCB.
Boot
ZB0046 GUI Time of Day Background Alert Replace GUI CPU PCB.
failed checks (GUI)
ZB0047 GUI NOVRAM Background Alert Noncritical NOVRAM 1. Replace GUI CPU PCB.
checksum error checks (GUI) data failure 2. Call Technical Support or your
Puritan
Bennett representative.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-51
6 Diagnostic codes
ZB0086 Init Loss of BD Background Alert A status message See associated error codes.
communication checks (GUI) indicating that
communication with
BD was lost.
ZB0090 Touch Screen Background Alert 1. Remove visible items from GUI
blocked checks (GUI) screen.
2. Clean GUI screen.
3. Replace touchscreen PCB.
ZB0091 Touch Screen Background Status Status message
resumed checks (GUI) indicates that a touch
screen blocked error is
cleared.
ZC0xxx Varies Communi- Alert 1. Ensure that the green LED is lit
cations error on the front of the BPS.
(BDU) a. If the green LED is not lit,
allow the BPSto charge until
the green LED lights.
b. If the green LED does not
light (within 6-8 hours or
less), replace the battery
pack.
c. Replace the BPS PCB
2. Once the green LED is lit, run
EST to test the BPS and use EST
diagnostics to determine if fur-
ther corrective action is
needed.
6-52 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-2: 840 Vent ilat or diagnostic codes (cont inued)
ZC2000 DCI parity error Communica- Alert Check parity of host device.
tions error
(GUI)
ZC2001 DCI input buffer Communica- Alert 1. Check cabling between ventila-
overflow error tions error tor and host.
(GUI) 2. Check baud rate of host device.
ZC2002 DCI command Communica- Alert Undefined command 1. Check host software for com-
error tions error received patibility with ventilator.
(GUI) 2. Check cabling between ventila-
tor and host.
ZC2003 DCI unknown Communica- Alert 1. Check host software for com-
error tions error patibility with ventilator.
(GUI) 2. Check cabling between ventila-
tor and host.
ZP0008 Kernel POST (GUI) Alert Noncritical NOVRAM Call Technical Support or your
NOVRAM Test data failure Puritan Bennett representative.
ZP0085 Ethernet Self- POST (GUI) Alert Spec says failure Replace GUI CPU PCB.
Test Start
ZP0086 Ethernet Self- POST (GUI) Alert Spec says failure Replace GUI CPU PCB.
Test End
ZP0087 Unexpected POST (GUI) Alert POST was invoked due Check associated errors in log. No
Reset Umpire to an unexpected reset. action required unless other errors
Test are present.
ZP0088 POST NOVRAM POST (GUI) Alert Noncritical NOVRAM Replace GUI CPUPCB.
Test data failure
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-53
6 Diagnostic codes
6.8 Syst em Diagnost ic Log and BDU POST analog devices t est
Anal og device errors reported durin g th e BDU POST anal og devices test are recorded in th e
System Di agnostic Log. These errors are reported as di agn ostic codes KP0116 an d LP0116.
In form ation in the NOTES col um n of the l og iden ti fies the analog ch ann el on whi ch the
error occurred.
Using an exam ple of a BDU POST analog devi ce error, as i t m i gh t be reported in the System
Di agnosti c Log:
The di agnosti c codes, KP0116 (BDU) an d LP0116 (GUI), are m aj or faults (VENT INOP) th at
can occur durin g the Anal og Devices Test (Ph ase 2 POST).
If th ere are m ultiple KP0116 an d/ or LP0116 diagnosti c codes, ch eck the PC channel s and
determ in e i f they are all th e sam e. If the chann el s are al l th e sam e, repl ace the associated
an al og device or the AI PCB. If the chan nels are differen t, but ori gi nate from the sam e
m odul e, check and/ or replace the associated m odul e, in terconnect cabl e or AI PCB.
Using the fifth an d sixth digits of error address reported in the System Diagnosti c Log an d
Table 6-3, th e an alog channel on w hich th e error occurred can be iden tified.
Table 6-3: Address codes for BDU POST analog devices test errors
If t he 5t h and 6t h
digit s of t he error Paramet er M easured Replace This:
address are:
6-54 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-3: Address codes for BDU POST analog devices t est errors
If t he 5t h and 6t h
digit s of t he error Paramet er M easured Replace This:
address are:
12 O2 Sensor O2 Sensor
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-55
6 Diagnostic codes
6-56 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-4: BDU and GUI—Phase 1 (kernal) POST diagnost ic codes (cont inued)
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-57
6 Diagnostic codes
BDU
BDU Test BDU Correct ive
BDU Test Fault
Descript ion Diag. Code Act ion
Type
ac Voltage Test Comparesac monitor voltage M inor 1. Check the diagnostic code’s
to specified range corrective actions in Table 6-2.
2. Make sure ventilator is plugged
into ac source.
LP0122 3. Make sure nominal line volt-
age is set correctly.
KP0114
Analog Devices Test Performs a range check on Major 1. Check the diagnostic code’s
each of the 32 analog or corrective actions in Table 6-2.
channels M inor 2. Refer to Section 6.8 for addi-
tional troubleshooting informa-
KP0116 or tion regarding this fault.
LP0116 3. Replace affected component.
4. Replace AI PCB.
5. Call Technical Support or your
Puritan Bennett representative.
Analog Interface Test Tests analog interface Major 1. Check the diagnostic code’s
circuitry corrective actions in Table 6-2.
2. Replace BD CPU PCB.
3. Replace AI PCB.
KP0113 4. Replace mother board.
Bus Timer Test * M onitors activity on the local Major 1. Check the diagnostic code’s
bus corrective actions in Table 6-2.
2. Replace BD CPU PCB.
KP0082
KP0115
DRAM M emory Test * Tests DRAM by writing and Major 1. Check the diagnostic code’s
verifying data patterns to corrective actions in Table 6-2.
memory 2. Replace BD CPU PCB.
KP0084
DRAM Parity Error Checks the ability to detect a Major 1. Check the diagnostic code’s
Test * parity error upon reading corrective actions in Table 6-2.
defective parity from memory 2. Replace BD CPU PCB.
KP0090
*
Tested in Short POST
6-58 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-5: BDU only—Phase 2 POST diagnostic codes (continued)
BDU
BDU Test BDU Correct ive
BDU Test Fault
Descript ion Diag. Code Act ion
Type
LP0086
Ethernet Self-Test Internal routine verifies Major 1. Check the diagnostic code’s
Start functionality of Ethernet corrective actions in Table 6-2.
controller 2. Replace BD CPU PCB.
KP0085
NMI Register Test * Verifies NMI source register is Major 1. Check the diagnostic code’s
in reset state (no bits set) corrective actions in Table 6-2.
2. Replace BD CPU PCB.
KP0083
Phase 2 Initialization* Start non-kernel portion of Major 1. Check the diagnostic code’s
POST corrective actions in Table 6-2.
2. Replace BD CPU PCB.
KP0006
Phase 2 NOVRAM Compares NOVRAM data to Minor 1. Check the diagnostic code’s
Test * valid states and ranges corrective actions in Table 6-2.
2. Call Technical Support or your
Puritan Bennett representative.
LP0088
Safe State Test Places PSOLs and exhalation Major 1. Check the diagnostic code’s
valve into safe state and corrective actions in Table 6-2.
verifies that they cannot be 2. Replace AI PCB.
commanded 3. Replace BD CPU PCB.
KP0120
Serial Device Test Writes to serial EEPROM to Major 1. Check the diagnostic code’s
verify accessibility corrective actions in Table 6-2.
2. Replace AI PCB.
3. Replace BD CPU PCB.
KP0117
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-59
6 Diagnostic codes
BDU
BDU Test BDU Correct ive
BDU Test Fault
Descript ion Diag. Code Act ion
Type
Service Switch Test Checks that the service mode M inor 1. Check the diagnostic code’s
Start * switch is not stuck, and corrective actions in Table 6-2.
prevents entry into Service 2. Turn off ventilator. Release
M ode from a power fail switch.
condition LP0121 3. Check for mechanical obstruc-
tion to switch.
4. Replace BD CPU PCB.
Unexpected Reset Checks for unexpected resets Major 1. Check the diagnostic code’s
Umpire Test * of the CPU. Three resets or corrective actions in Table 6-2.
(strikes) in 24 hours will cause M inor 2. Check associated errors in the
a VENT INOP condition. System Diagnostic and System
KP0087 or Information log to determine
LP0087 why POST was invoked three
times within 24 hours.
a. If the codes indicate that the
BD CPU generated the
resets, replace the BD CPU
(Other codes may be
present indicating that the
GUI lost communications
with the BD).
b. If the codes indicated that
the GUI CPU generated the
resets, replace the GUI CPU
(Other codes may be
present indicating that the
BD lost communications
with the GUI.
3. Run a complete EST to check
for analog device failures.
VENT INOP Test Checks to see if the ventilator Major 1. Check the diagnostic code’s
is in the VENT INOP state corrective actions in Table 6-2.
2. Check cabling from BD CPU
PCB to GUI CPU PCB.
KP0112 3. Replace BD CPU PCB.
4. Call Technical Support or your
Puritan Bennett representative.
*
Tested in Short POST
6-60 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
DRAM Parity Error Checks the ability to Major 1. Check the diag-
Test detect a parity error nostic code’s cor-
upon reading rective actions in
defective parity from Table 6-2.
memory XP0090 XP0090 2. Replace GUI CPU
PCB.
NMI Register Test Verifies NMI source Major 1. Check the diag-
register is in reset nostic code’s cor-
state (no bits set) rective actions in
Table 6-2.
XP0083 XP0083 2. Replace GUI CPU
PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-61
6 Diagnostic codes
6-62 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-63
6 Diagnostic codes
Table 6-8: Interrupt errors and test failures – POST self t est s
6-64 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-8: Interrupt errors and test failures – POST self test s (cont inued)
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-65
6 Diagnostic codes
Code and
Test descript ion and failure modes Correct ive act ions
message
Failure: No gas is available (neither PS1, PS2, nor FS0005 Connect gas supply.
PC detects pressure). Gas not
connected
Failure: Only air is available (only PS2 or PC AS0012 Make sure oxygen supply is connected.
detects pressure). O2 pressure not
detected
Failure: Only oxygen is available (only PS1 detects AS0013 Make sure air supply is connected.
pressure). Wall air pressure
not detected
Failure: Flow controller unable to establish and AS0010 1. M ake sure oxygen supply is connected and
control oxygen flow at 120 L/ min. Unable to unrestricted.
establish O2 flow 2. Verify that the oxygen supply is good.
3. Verify that the patient circuit system has no
leaks or occlusions. If not sure, run the first
four tests in EST to get to the leak test. If the
leak test passes in EST, exit out of EST. Correct
the leak or occlusion in the patient circuit used
in SST and rerun or try another patient circuit.
4. Verify that the oxygen regulator pressure is set
between 9-12 psi.
5. Switch PSOLs to see if the failure transfers to
the air side. If yes, return PSOL2 back to its
original position and replace PSOL1.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the flow sensor cali-
bration and EST. If the failure transfers to the
air side, return Q2 back to its original position
and replace Q1.
6-66 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-9: SST diagnost ic codes (cont inued)
Code and
Test descript ion and failure modes Correct ive act ions
message
Failure: Flow controller unable to establish and FS0006 1. Make sure oxygen supply is connected and
control oxygen flow at 60, 5, and 1 L/ min. Unable to unrestricted.
establish O2 flow 2. Verify that the oxygen supply is good.
3. Check regulated oxygen pressure.
4. Verify that the system has no leaks or occlu-
sions. If not sure, run the first four tests in EST
to run the leak test.
5. Switch PSOLs to see if the failure transfers to
the air side. If yes, return PSOL2 back to its
original position and replace PSOL1.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the flow sensor cali-
bration and EST. If the failure transfers to the
air side, return Q2 back to its original position
and replace Q1.
8. Replace Q3.
Failure: Q1 and Q3 readings are too far apart. FS0001 1. Check for leak.
O2 flow sensor 2. Verify that safety valve is closed.
cross check failed 3. Perform flow sensor calibration.
4. Replace Q1.
5. Replace Q3.
6. Replace AI PCB.
Failure: PSOL1 command current is out of range. FS0002 1. Check for leak.
O2 PSOL current 2. Verify safety valve is closed.
out of range 3. Check regulated oxygen pressure.
4. Replace PSOL1.
5. Replace AI PCB.
Failure: Oxygen flow > 0.153 L/ min. FS0008 1. Verify no leaks at the Q1
O2 zero flow O-rings.
check failed 2. Run EST Gas Supply Test to check for PSOL1
leak.
3. Remove and then reseat PSOL1.
4. Replace PSOL1.
5. Perform a flow sensor calibration.
6. Replace Q1.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-67
6 Diagnostic codes
Code and
Test descript ion and failure modes Correct ive act ions
message
Failure: Flow controller unable to establish and AS0011 1. M ake sure air supply is connected and unre-
control air flow at 120 L/ min. Unable to stricted.
establish air flow 2. Verify that the air supply is good.
3. Verify that the patient circuit system has no
leaks or occlusions. If not sure, run the first
four tests in EST to get to the leak test. If the
leak test passes in EST, exit out of EST. Correct
the leak or occlusion in the patient circuit used
in SST and rerun or try another patient circuit.
4. Verify that the air regulator pressure is set
between 9-12 psi.
5. Switch PSOLs to see if the failure transfers to
the oxygen side. If yes, return PSOL1 back to
its original position and replace PSOL2.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the flow sensor cali-
bration and EST. If the failure transfers to the
oxygen side, return Q1 back to its original
position and replace Q2.
Failure: Flow controller unable to establish and FS0007 1. Make sure air supply is connected unrestricted.
control air flow at 60, 5, and 1 L/ min. Unable to 2. Verify that the air supply is adequate.
establish air flow 3. Check regulated air pressure.
4. Verify that the system has no leaks or occlu-
sions. If not sure, run the first four tests in EST
to run the leak test.
5. Switch PSOLs to see if the failure transfers to
the oxygen side. If yes, return PSOL1 back to
its original position and replace PSOL2.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the flow sensor cali-
bration and EST. If the failure transfers to the
oxygen side, return Q1 back to its original
position and replace Q2.
8. Replace Q3.
6-68 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-9: SST diagnost ic codes (cont inued)
Code and
Test descript ion and failure modes Correct ive act ions
message
Failure: Q2 and Q3 readings are too far apart. FS0003 1. Correct the leak or occlusion in the SST patient
Air flow sensor circuit and rerun or try known good patient
cross-check circuit and filter set.
failed 2. Verify that the ventilator system has no leaks or
occlusions by running the first four tests in EST
to run the leak test.
3. If the leak test fails in EST, troubleshoot the
expiratory filter or ventilator system for leaks or
occlusions.
4. If the leak test passes in EST, troubleshoot the
SST patient circuit.
5. Switch PSOLs to see if the failure transfers to
the oxygen side. If yes, return PSOL1 back to
its original position and replace PSOL2.
6. Run a flow sensor calibration.
7. Switch Q1 and Q2. Rerun the flow sensor cali-
bration and EST. If the failure transfers to the
oxygen side, return Q1 back to its original
position and replace Q2.
8. Replace Q3.
Failure: PSOL2 command current out of range. FS0004 1. Make sure air supply is connected and unre-
Air PSOL current stricted.
out of range 2. Verify that the air supply is good.
3. Verify that the system has no leaks or occlu-
sions. If not sure, run the first four tests in EST
to run the leak test.
4. Switch PSOLs to see if the failure transfers to
the oxygen side. If yes, return PSOL1 back to
its original position and replace PSOL2.
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the flow sensor cali-
bration and EST. If the failure transfers to the
oxygen side, return Q1 back to its original
position and replace Q2.
7. Replace Q3.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-69
6 Diagnostic codes
Code and
Test descript ion and failure modes Correct ive act ions
message
Failure: Air flow > 0.153 L/ min. FS0009 1. Verify no leaks at the Q2
Air zero flow O-rings.
check failed 2. Run EST Gas Supply Test to check for PSOL2
leak.
3. Remove and then reseat PSOL2.
4. Replace PSOL2.
5. Perform a flow sensor calibration.
6. Replace Q2.
Failure: Inspiratory pressure transducer ADC FS0101 1. Replace Inspiratory Electronics PCB.
count at 0 cmH2O is out of range. Inspiratory 2. Replace AI PCB.
autozero out of
range
Failure: Expiratory pressure transducer ADC count FS0102 1. Replace exhalation transducer PCB.
at 0 cmH2O is out of range. Expiratory 2. Replace AI PCB.
autozero out of 3. Replace SOL1.
range
Failure: Unable to build pressure to 10 cmH2O at FS0103 1. Verify that the patient circuit system has no
5 L/ min within time-out period. Failed to reach leaks or occlusions or use a known good
test pressure patient circuit and filter set.
2. Verify that the air supply or compressor is
available and adequate.
3. Verify that the ventilator system has no leaks or
occlusions. If not sure, run the first four tests
in EST to run the leak test.
6-70 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-9: SST diagnost ic codes (cont inued)
Code and
Test descript ion and failure modes Correct ive act ions
message
Failure: Inspiratory/ expiratory pressure transducer FS0104 1. Verify that the patient circuit system has no
readings are too far apart. Cross-check leaks or occlusions or use a known good
failed patient circuit and filter set.
2. Verify that the ventilator system has no leaks or
occlusions. If not sure, run the first four tests
in EST to run the leak test.
3. If the leak test passes in EST, exit out of EST.
Review the data for the Circuit Pressure Test to
determine which pressure transducer is out of
range: PI or PE. Replace applicable pressure
transducer.
4. Replace AI PCB.
Failure: Unable to build pressure to 50 or FS0108 1. Make sure gas source is available.
100 cmH2O at 5 L/ min within time-out period. Failed to reach 2. Check for system leak.
test pressure
Failure: Inspiratory/ expiratory pressure transducer FS0107 1. Replace the expiratory bacteria filter.
readings are too far apart. Cross-check 2. Verify that the system has no leaks or occlu-
failed sions. If not sure, run the first four tests in EST
to run the leak test.
3. If the leak test passes in EST, exit out of EST.
Review the data for the EST Circuit Pressure
Test to determine which pressure transducer is
out of range: PI or PE. Replace applicable
pressure transducer.
4. Replace AI PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-71
6 Diagnostic codes
Code and
Test descript ion and failure modes Correct ive act ions
message
Circuit leak
Failure: Unable to attain leak test starting pressure FS0205 1. M ake sure test circuit is connected and is not
within time-out period. Unable to leaking.
establish 2. Check exhalation valve operation, and verify
pressure secure connection of expiratory filter.
Failure: System pressure dropped below FS0201 1. M ake sure test circuit is connected.
85 cmH2O. Excessive leak 2. Check for system leaks.
3. Check exhalation valve operation.
4. Verify secure connection of expiratory filter.
Failure: Pressure drops to 85 cmH2O in 10 s, then FS0201 1. M ake sure patient circuit is connected.
in 10 s more by ≥ 30 cmH2O. Excessive leak 2. Check for system leaks.
3. Check exhalation valve operation.
4. Verify secure connections of bacteria filters and
humidifier.
Failure: Pressure drops to 85 cmH2O in 10 s, then AS0202 1. M ake sure patient circuit is connected.
in 10 s more by ≥ 10 cmH2O. Excessive leak 2. Check for system leaks.
3. Check exhalation valve operation.
4. Verify secure connections of bacteria filters and
humidifier.
Failure: Inspiratory/ expiratory pressure transducer AS0206 Make sure test circuit is connected properly.
readings are too far apart. Test circuit not
connected
Expirat or y f ilt er
Failure: Unable to deliver stable gas flow. FS0301 Make sure gas supply is connected.
Unable to
establish flow
6-72 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-9: SST diagnost ic codes (cont inued)
Code and
Test descript ion and failure modes Correct ive act ions
message
Failure: Expiratory flow detected, indicating that FS0306 Make sure patient circuit is disconnected at
circuit not disconnected. Patient circuit expiratory filter.
not disconnected
Failure: Unable to deliver stable gas flow. FS0301 Make sure gas supply is connected.
Unable to
establish flow
Failure: Expiratory flow reading indicates that FS0307 Make sure patient circuit is connected at
circuit not reconnected. Patient circuit expiratory filter.
not reconnected
Failure: Pressure drop across filter > 2 cmH2O, but AS0305 Consider replacing expiratory filter.
< 3 cmH2O. Occluded
expiratory filter
Failure: Pressure drop across filter > 3 cmH2O. FS0304 Replace expiratory filter.
Occluded
expiratory filter
Failure: Pressure drop across expiratory filter AS0311 1. Repeat test, following directions more closely.
< 0.4 cmH2O, but > 0.1 cmH2O. Low expiratory 2. Replace filter.
filter ∆P
Failure: Pressure drop across expiratory filter FS0310 1. Repeat test, following directions more closely.
< 0.1 cmH2O. Low expiratory 2. Replace filter.
filter ∆P
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-73
6 Diagnostic codes
Code and
Test descript ion and failure modes Correct ive act ions
message
Failure: Unable to pressurize system within time- FS0412 1. Block wye when prompted.
out period. Wye not blocked 2. Check for circuit leaks.
Failure: Peak flow < alert threshold but > failure AS0411 Check for kinked or occluded patient tubing.
threshold. Unable to reach
min peak flow
Failure: Peak flow < failure threshold. FS0410 Check for kinked or occluded patient tubing.
Unable to reach
min peak flow
Failure: Flow controller unable to establish and FS0401 Make sure gas supply is connected.
control gas flow. Unable to
establish flow
Failure: Inspiratory limb pressure > 8.5 cmH2O AS0403 Check for occluded patient tubing.
(adult) or 5.5 cmH2O (pediatric), but Occluded
< 12.5 cmH2O (adult) or 7.5 cmH2O (pediatric), inspiratory limb
indicating occlusion.
Failure: Inspiratory limb pressure > 12.5 cmH2O FS0402 Check for occluded patient tubing.
(adult) or 7.5 cmH2O (pediatric), indicating Occluded
occlusion. inspiratory limb
Failure: Inspiratory limb pressure < 0.6 cmH2O AS0413 1. M ake sure inspiratory filter is installed.
(adult) or 5.5 cmH2O (pediatric), but > Insp limb 2. Repeat test, following directions more closely.
0.2 cmH2O. resistance low 3. Replace inspiratory filter.
4. Replace patient circuit
Failure: Inspiratory limb pressure < 0.2 cmH2O. FS0408 1. M ake sure inspiratory filter is installed.
Insp limb 2. Repeat test, following directions more closely.
resistance low 3. Replace inspiratory filter.
4. Replace patient circuit
6-74 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-9: SST diagnost ic codes (cont inued)
Code and
Test descript ion and failure modes Correct ive act ions
message
Failure: Peak flow < 80 L/ min but > 60 L/ min. AS0407 Check for kinked or occluded patient tubing.
Unable to reach
min peak flow
Failure: Peak flow < 60 L/ min. FS0404 Check for kinked or occluded patient tubing.
Unable to reach
min peak flow
Failure: Expiratory limb pressure > 8.5 cmH2O AS0406 Check for occluded patient tubing.
(adult) or 5.5 cmH2O (pediatric) but Occluded
> 12.5 cmH2O (adult) or 7.5 cmH2O (pediatric), exhalation limb
indicating occlusion.
Failure: Expiratory limb pressure > 12.5 cmH2O FS0405 Check for occluded patient tubing.
(adult) or 7.5 cmH2O (pediatric), indicating Occluded
occlusion. exhalation limb
Failure: Expiratory limb pressure < 0.6 cmH2O AS0414 Replace patient circuit
(adult) or 0.5 cmH2O (pediatric), but > Exp limb
0.2 cmH2O. resistance low
Failure: Expiratory limb pressure < 0.2 cmH2O. FS0409 Replace patient circuit
Exp limb
resistance low
Failure: System cannot pressurize to one of the FS0501 Check patient circuit for leaks. Replace as
test pressure points. Unable to necessary.
pressurize circuit
Failure: System cannot pressurize to one of the FS0502 Check patient circuit for leaks. Replace as
test pressure points. Unable to necessary.
pressurize circuit
Failure: Out-of-range pressure ratio calculation. FS0503 1. Repeat test, making sure circuit is undisturbed
Compliance during test.
calculation 2. Repeat SST to retest pressure transducers.
failure 3. Run EST and verify all pneumatics tests pass.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-75
6 Diagnostic codes
Code and
Test descript ion and failure modes Correct ive act ions
message
Failure: High compliance > 6 mL/ cmH2O (adult) AS0505 1. M ake sure correct patient tubing type was
or 4.5 mL/ cmH2O (pediatric), but Excessive specified in SST.
< 12 mL/ cmH2O (adult) or 9 mL/ cmH2O compliance 2. Replace patient circuit.
(pediatric).
Failure: High compliance > 12 mL/ cmH2O (adult) FS0504 1. M ake sure correct patient tubing type was
or 9 mL/ cmH2O (pediatric). Excessive specified in SST.
compliance 2. Replace patient circuit.
Failure: High compliance falls below AS0507 1. M ake sure correct patient tubing type was
1.56 mL/ cmH2O (adult) or 1.34 mL/ cmH2O Compliance low specified in SST.
(pediatric) but not lower than 1.05 mL/ cmH2O. 2. Replace patient circuit with a known good cir-
cuit and filter set.
3. Run EST to check the operation of PI and PE
during the Circuit Pressure Test. Replace
applicable pressure transducer.
4. Run Atmospheric Pressure Transducer calibra-
tion.
Failure: High compliance falls below FS0506 1. M ake sure correct patient tubing type was
1.05 mL/ cmH2O. Compliance low specified in SST.
2. Replace patient circuit.
Failure: Low compliance > 6 mL/ cmH2O (adult) or AS0509 1. M ake sure correct patient tubing type was
4.5 mL/ cmH2O (pediatric), but < 12 mL/ cmH2O Excessive specified in SST.
(adult) or 9 mL/ cmH2O (pediatric). compliance 2. Replace patient circuit with a known good cir-
cuit and filter set.
3. Run EST to check the operation of PI and PE
during the Circuit Pressure Test. Replace
applicable pressure transducer.
4. Run Atmospheric Pressure Transducer calibra-
tion.
Failure: Low compliance > 12 mL/ cmH2O (adult) FS0508 1. M ake sure correct patient tubing type was
or 9 mL/ cmH2O (pediatric). Excessive specified in SST.
compliance 2. Replace patient circuit.
Failure: Low compliance falls below AS0511 1. M ake sure correct patient tubing type was
1.56 mL/ cmH2O (adult) or 1.34 mL/ cmH2O Compliance low specified in SST.
(pediatric), but not below 1.05 mL/ cmH2O. 2. Replace patient circuit with a known good cir-
cuit and filter set.
3. Run EST to check the operation of PI and PE
during the Circuit Pressure Test. Replace
applicable pressure transducer.
4. Run Atmospheric Pressure Transducer calibra-
tion.
Failure: Low compliance falls below FS0510 1. M ake sure correct patient tubing type was
1.05 mL/ cmH2O. Compliance low specified in SST.
2. Replace patient circuit.
6-76 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence
Code and
Descript ion Display Correct ive act ions
message
Failure: System is still running on battery FE0010 1. Plug in ventilator power cord.
power. ac power not 2. Disconnect BPS to isolate
connected problem.
3. Replace power supply.
Failure: Unable to build pressure (air) to FE0003 Inspiratory 1. Make sure that test circuit is
10 cmH2O at 5 L/ min within time-out period. Failed to reach and expiratory installed and air supply or com-
test pressure pressure (only pressor is available and good.
if test fails) 2. Verify that the air regulator pres-
sure is set between 9-12 psi.
3. Verify that the patient circuit sys-
tem has no leaks or occlusions by
running the first four tests in EST
to get to the leak test.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-77
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: Inspiratory/ expiratory pressure FE0004 1. Verify that the system has no leaks
transducer readings are too far apart. Cross-check or occlusions. If not sure, run the
failed first four tests in EST to run the
leak test
2. If the leak test passes in EST, exit
out of EST. Review the data for
the Circuit Pressure Test to deter-
mine which pressure transducer is
out of range: PI or PE. Replace
applicable pressure transducer.
3. Replace AI PCB.
Failure: Unable to build pressure (air) to 50 or FE0008 Inspiratory 1. Make sure air supply or compres-
100 cmH2O at 5 L/ min within time-out period. Failed to reach and expiratory sor is available.
test pressure pressure (only 2. Check for system leak at the expi-
if test fails) ratory bacteria filter or O2 sen-
sor. If not sure, run the first four
tests in EST to run the leak test.
6-78 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: Inspiratory/ expiratory pressure FE0007 1. Replace the expiratory bacteria fil-
transducer readings are too far apart. Cross-check ter.
failed 2. Verify that the system has no leaks
or occlusions. If not sure, run the
first four tests in EST to run the
leak test.
3. If the leak test passes in EST, exit
out of EST. Review the data for the
Circuit Pressure Test to determine
which pressure transducer is out
of range: PI or PE. Replace appli-
cable pressure transducer.
4. Replace AI PCB.
Failure: Flow controller unable to establish and AE0110 1. Make sure oxygen supply is con-
control oxygen flow at 120 L/ min. Unable to nected and unrestricted.
establish O2 2. Verify oxygen regulator pressure is
flow set between 9-12 psi.
3. Switch PSOLs to see if failure
transfers to the other gas side. If
yes, return the air PSOL to its orig-
inal position and replace the oxy-
gen PSOL.
4. Run the leak test in EST to check
for leaks/ occlusions.
5. Perform a flow sensor calibration.
6. Replace Q1.
7. Replace the Inspiratory Electronics
PCB.
8. Replace the AI PCB.
Failure: Flow controller unable to establish and FE0106 1. Make sure oxygen supply is con-
control oxygen flow at 60, 5, and 1 L/ min. Unable to nected.
establish O2 2. Check regulated oxygen pressure.
flow 3. Replace PSOL1 or oxygen flow
sensor (Q1).
4. Replace AI PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-79
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: Q1 and Q3 readings are too far apart. FE0101 1. Make sure oxygen supply is con-
O2 flow sensor nected and unrestricted.
cross check 2. Verify that the oxygen supply is
failed good.
3. Verify that the system has no leaks
or occlusions. If not sure, run the
first four tests in EST to run the
leak test.
4. Switch PSOLs to see if the failure
transfers to the air side. If yes,
return PSOL2 back to its original
position and replace PSOL1.
5. Run a flow sensor calibration.6.
Switch Q1 and Q2. Rerun the
flow sensor calibration and EST. If
the failure transfers to the air side,
return Q2 back to its original posi-
tion and replace Q1.
Failure: PSOL1 command current is out of FE0102 1. Make sure oxygen supply is con-
range. O2 PSOL nected and unrestricted.
current out of 2. Verify that the oxygen supply is
range good.
3. Verify that the system has no leaks
or occlusions. If not sure, run the
first four tests in EST to run the
leak test.
4. Switch PSOLs to see if the failure
transfers to the air side. If yes,
return PSOL2 back to its original
position and replace PSOL1.
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the
flow sensor calibration and EST. If
the failure transfers to the air side,
return Q2 back to its original posi-
tion and replace Q1.
Failure: Oxygen flow > 0.153 L/ min. FE0108 1. Verify no leaks at the Q1 O-rings.
O2 zero flow 2. Run EST Gas Supply Test to check
check failed for PSOL1 leak. Remove and then
reseat PSOL1
3. Replace PSOL1.
4. Perform a flow sensor calibration.
5. Replace Q1.
6-80 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: Flow controller unable to establish and AE0111 1. Make sure air supply is connected
control air flow at 120 L/ min. Unable to and unrestricted.
establish air 2. Verify air regulator pressure is set
flow between 9-12 psi.
3. Switch PSOL's to see if failure
transfers to the other gas side. If
yes, return Oxygen PSOL to its
original position and replace the
Air PSOL.
4. Run the leak test in EST to check
for leaks/ occlusions.
5. Perform a flow sensor calibration.
6. Replace Q2.
7. Replace the Inspiratory Electronics
PCB.
8. Replace the AI PCB.
Failure: Flow controller unable to establish and FE0107 1. Make sure air supply is connected.
control air flow at 60, 5, and 1 L/ min. Unable to 2. Check regulated air pressure.
establish air 3. Perform flow sensor calibration.
flow 4. Replace PSOL2 or air flow sensor
(Q2).
5. Replace AI PCB.
Failure: Q2 and Q3 readings are too far apart. FE0103 1. Make sure air supply is connected
Air flow sensor and unrestricted.
cross check 2. Verify that the air supply is good.
failed 3. Verify that the system has no leaks
or occlusions. If not sure, run the
first four tests in EST to run the
leak test.
4. Switch PSOLs to see if the failure
transfers to the air side. If yes,
return PSOL back to its original
position and replace PSOL2
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the
flow sensor calibration and EST. If
the failure transfers to the oxygen
side, return Q1 back to its original
position and replace Q2.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-81
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: PSOL2 command current out of range. FE0104 1. Make sure air supply is connected
Air PSOL and unrestricted.
current out of 2. Verify that the air supply is good.
range 3. Verify that the system has no leaks
or occlusions. If not sure, run the
first four tests in EST to run the
leak test.
4. Switch PSOLs to see if the failure
transfers to the air side. If yes,
return PSOL back to its original
position and replace PSOL2
5. Run a flow sensor calibration.
6. Switch Q1 and Q2. Rerun the
flow sensor calibration and EST. If
the failure transfers to the oxygen
side, return Q1 back to its original
position and replace Q2.
Failure: Air flow > 0.153 L/ min. FE0109 1. Verify no leaks at the Q2
Air zero flow O-rings.
check failed 2. Run EST Gas Supply Test to check
for PSOL2 leak. Remove and then
reseat PSOL2
3. Replace PSOL2.
4. Perform a flow sensor calibration.
5. Replace Q2.
Failure: ACCEPT key was pressed but PS2 still FE0204 1. Make sure air supply is connected.
open. Wall air 2. Check PS2: Remove the wiring
pressure not connectors from PS2 and jumper
detected the connectors to each other. If
the system now recognizes the air
supply, replace PS2.
6-82 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: ACCEPT key was pressed but PS1 still FE0205 1. Make sure oxygen supply is con-
open. O2 pressure nected.
not detected 2. Check PS1: Remove the wiring
connectors from PS1 and jumper
the connectors to each other. If
the system now recognizes the air
supply, replace PS1.
Failure: Stable flow could not be established FE0210 1. Make sure To patient port is
within allotted time. SV pressure blocked.
relief failed 2. Check for leaks at the O2 sensor,
SOL1 or PI.
3. Replace the safety valve.
4. Replace PI.
Failure: Cracking pressure out of range. FE0210 1. Make sure To patient port is
SV pressure blocked.
relief failed 2. Check for leaks at the O2 sensor,
SOL1 or PI.
3. Replace the safety valve.
4. Replace PI.
Failure: Neither PSOL could deliver 100 L/ min FE0210 1. Make sure To patient port is
within prescribed count. SV pressure blocked.
relief failed 2. Check for leaks at the O2 sensor,
SOL1 or PI.
3. Replace the safety valve.
4. Replace PI.
Failure: Peak pressure is too high. FE0210 1. Make sure To patient port is
SV pressure blocked.
relief failed 2. Check for leaks at the O2 sensor,
SOL1 or PI.
3. Replace the safety valve.
4. Replace PI.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-83
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: PS1 closed instead of open. FE0206 1. Make sure oxygen supply is dis-
O2 pressure connected.
detected 2. Replace Inspiratory Electronics
(O2 not PCB.
disconnected) 3. Replace PS1.
Failure: Air PSOL (PSOL2) forward leak was AE0213 1. Check for leaks around the Q2
detected via excessive pressure buildup Air PSOL leak O-rings.
(> 50 cmH2O but < 100 cmH2O) in blocked 2. Make sure air supply is connected
inspiratory module. and unrestricted.
3. Reseat Air PSOL.
4. Replace Air PSOL.
5. Replace AI PCB.
Failure: Air PSOL (PSOL2) forward leak was FE0207 1. Check for leaks around the Q2 O-
detected via excessive pressure buildup Air PSOL leak rings.
(> 100 cmH2O) in blocked inspiratory module. 2. Verify that the safety valve relieves
pressures above 100 cmH2O. Lis-
ten for the pressure relief while
watching the numeric digital dis-
play of system pressure in the
upper screen. If the safety valve
does not crack open, replace the
safety valve.
3. Replace PSOL2.
Failure: PS2 closed instead of open. FE0208 1. Make sure air supply is discon-
Wall air nected.
pressure 2. Replace Inspiratory Electronics
detected PCB.
3. Replace PS2.
6-84 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: During zero-flow check, Q2 reads FE0215 1. Make sure gas supplies are discon-
> 0.05 L/ min. Air zero flow nected.
check failed 2. Make sure compressor is not run-
ning.
3. Verify no leaks at the Q2
O-rings.
4. Check for PSOL2 leak.
5. Remove and then reseat PSOL2.
6. Perform a flow sensor calibration.
7. Replace PSOL2.
8. Replace Q2.
Failure: During zero-flow check, Q1 reads FE0216 1. Make sure gas supplies are discon-
> 0.05 L/ min. O2 zero flow nected.
check failed 2. Make sure compressor is not run-
ning.
3. Verify no leaks at the Q2
O-rings.
4. Check for PSOL2 leak.
5. Remove and then reseat PSOL2.
6. Perform a flow sensor calibration.
7. Replace PSOL2.
8. Replace Q2.
Failure: During zero-flow check, Q3 reads > FE0217 1. Make sure gas supplies are discon-
0.1 L/ min. Exp zero flow nected.
check failed 2. Make sure compressor is not run-
ning.
3. Verify no leaks at the Q3
O-rings.
4. Perform a flow sensor calibration.
5. Replace Q3.
Failure: PS1 open instead of closed. FE0211 1. Make sure oxygen supply is con-
O2 pressure nected.
not detected 2. Replace Inspiratory Electronics
PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-85
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: Oxygen PSOL (PSOL1) forward leak AE0214 1. Check for leaks around the Q1 O-
was detected via excessive pressure buildup O2 PSOL leak rings.
(> 50 cmH2O but < 100 cmH2O) in blocked 2. Make sure oxygen supply is con-
inspiratory module. nected and unrestricted.
3. Reseat oxygen PSOL.
4. Replace oxygen PSOL.
5. Replace AI PCB.
Failure: Oxygen PSOL (PSOL1) forward leak FE0209 1. Check for leaks around the Q1 O-
was detected via excessive pressure buildup O2 PSOL leak rings.
(> 100 cmH2O) in blocked inspiratory module. 2. Verify that the safety valve relieves
pressures above 100 cmH2O. Lis-
ten for the pressure relief while
watching the numeric digital dis-
play of system pressure in the
upper screen. If the safety valve
does not crack open, replace the
safety valve.
3. Replace PSOL1.
Leak Test
Failure: System cannot attain leak test starting FE0305 1. Make sure test circuit is con-
pressure using oxygen or air within time-out Unable to nected.
period. establish 2. Check for system leaks.
pressure 3. Check exhalation valve operation.
4. Verify secure connection of expira-
tory filter.
Failure: System pressure dropped below FE0301 1. Make sure test circuit is con-
85 cmH2O. Excessive leak nected.
2. Check for system leaks.
3. Check exhalation valve operation.
4. Verify secure connection of expira-
tory filter.
6-86 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: Pressure not detected on expiratory AE0306 Pressure drop 1. Make sure test circuit is properly
side (inspiratory/ expiratory pressure transducer Test circuit in 10 s connected.
readings are too far apart). not connected 2. Replace expiratory bacteria
filter.
3. Check for leaks around the Q3
flow sensor.
4. Check/ replace the exhalation
valve.
5. Replace the expiratory pressure
transducer PCB.
Failure: Pressure drops to 85 cmH2O in 10 s, FE0301 1. Make sure test circuit is connected
then in 10 s more by ≥ 5 cmH2O. Excessive leak and is not leaking.
2. Check for system leaks or occlu-
sions especially at the O2 sensor
and expiratory filter.
3. Verify secure connection of expira-
tory filter.
4. Check exhalation valve operation.
Failure: Wrong key pressed or key not pressed FE0401 1. Repeat test.
within 15 s. Accept key 2. Replace keyboard.
fails.
Failure: Wrong key pressed or key not pressed FE0402 1. Repeat test.
within 15 s. Clear key fails. 2. Replace keyboard.
Failure: Wrong key pressed or key not pressed FE0403 1. Repeat test.
within 15 s. Insp. Pause 2. Replace keyboard.
key fails.
Failure: Wrong key pressed or key not pressed FE0404 1. Repeat test.
within 15 s. Exp. Pause key 2. Replace keyboard.
fails.
Failure: Wrong key pressed or key not pressed FE0405 1. Repeat test.
within 15 s. Man Insp fails. 2. Replace keyboard.
Failure: Wrong key pressed or key not pressed FE0406 1. Repeat test.
within 15 s. 100% O2 / 2. Replace keyboard.
CAL key fails.
Failure: Wrong key pressed or key not pressed FE0407 1. Repeat test.
within 15 s. Info key fails. 2. Replace keyboard.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-87
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: Wrong key pressed or key not pressed FE0408 1. Repeat test.
within 15 s. Alarm Reset 2. Replace keyboard.
key fails.
Failure: Wrong key pressed or key not pressed FE0409 1. Repeat test.
within 15 s. Alarm Silence 2. Replace keyboard.
key fails.
Failure: Wrong key pressed or key not pressed FE0410 1. Repeat test.
within 15 s. Alarm Volume 2. Replace keyboard.
key fails.
Failure: Wrong key pressed or key not pressed FE0411 1. Repeat test.
within 15 s. Screen 2. Replace keyboard.
brightness
fails.
Failure: Wrong key pressed or key not pressed FE0412 1. Repeat test.
within 15 s. Screen 2. Replace keyboard.
contrast key
fails.
Failure: Wrong key pressed or key not pressed FE0413 1. Repeat test.
within 15 s. Screen lock 2. Replace keyboard.
key fails.
Failure: CLEAR key pressed to indicate LED not AE0601 1. Check/ replace interconnect cable
on. GUI High between the GUI LED PCB and the
Alarm LED GUI CPU PCB.
fails. 2. Replace the GUI LED PCB.
Failure: CLEAR key pressed to indicate LED not AE0602 1. Check/ replace interconnect cable
on. GUI Medium between the GUI LED PCB and the
Alarm LED GUI CPU PCB.
fails. 2. Replace the GUI LED PCB.
Failure: CLEAR key pressed to indicate LED not AE0603 1. Check/ replace interconnect cable
on. GUI Low between the GUI LED PCB and the
Alarm LED GUI CPU PCB.
fails. 2. Replace the GUI LED PCB.
Failure: CLEAR key pressed to indicate LED not AE0604 1. Check/ replace interconnect cable
on. GUI Normal between the GUI LED PCB and the
LED fails. GUI CPU PCB.
2. Replace the GUI LED PCB.
6-88 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: CLEAR key pressed to indicate LED not AE0605 1. Check/ replace interconnect cable
on. GUI Batt between the GUI LED PCB and the
Backup LED GUI CPU PCB.
fails. 2. Replace the GUI LED PCB.
Failure: CLEAR key pressed to indicate LED not AE0606 1. Check/ replace interconnect cable
on. GUI On Batt between the GUI LED PCB and the
Pwr LED fails. GUI CPU PCB.
2. Replace the GUI LED PCB.
Failure: CLEAR key pressed to indicate LED not AE0607 1. Check/ replace interconnect cable
on. GUI between the GUI LED PCB and the
Compressor GUI CPU PCB.
Ready LED 2. Replace the GUI LED PCB.
fails.
Failure: CLEAR key pressed to indicate LED not AE0608 1. Check/ replace interconnect cable
on. GUI between the GUI LED PCB and the
Compressor GUI CPU PCB.
Operating 2. Replace the GUI LED PCB.
LED fails.
Failure: CLEAR key pressed to indicate LED not AE0609 1. Check/ replace interconnect cable
on. GUI 100% O2 between the GUI LED PCB and the
LED fails. GUI CPU PCB.
2. Replace keyboard.
Failure: CLEAR key pressed to indicate LED not AE0610 1. Check/ replace interconnect cable
on. GUI Alarm between the GUI LED PCB and the
Silence LED GUI CPU PCB.
fails. 2. Replace keyboard.
Failure: CLEAR key pressed to indicate LED not AE0611 1. Check/ replace interconnect cable
on. GUI Screen between the GUI LED PCB and the
Lock LED fails. GUI CPU PCB.
2. Replace keyboard.
BD Lam p Test
Failure: CLEAR key pressed to indicate one or AE0702 Replace BDU LED PCB.
both ventilator inoperative LEDs not on. Bad Vent inop
LED
Failure: CLEAR key pressed to indicate one or AE0703 Replace BDU LED PCB.
both SVO LEDs not on. Bad SVO LED
Failure: CLEAR key pressed to indicate LED not AE0704 Replace BDU LED PCB.
on. Bad Loss of
GUI LED
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-89
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: CLEAR key pressed to indicate GUI FE0801 1. Verify the GUI alarm cable connec-
audio diagnostic failed. SAAS tion to the GUI CPU PCB.
(Safety Audible 2. Replace GUI alarm assembly.
Alarm System) 3. Replace GUI CPU PCB.
test failed
Failure: CLEAR key pressed to indicate relay is FE2301 1. Make sure test equipment is con-
stuck on when it should be off. Nurse call nected properly to remote alarm
stuck on port.
2. Replace GUI CPU PCB.
Failure: CLEAR key pressed to indicate relay is FE2302 1. Make sure test equipment is con-
stuck off when it should be on. Nurse call nected properly to remote alarm
stuck off port.
2. Replace GUI CPU PCB.
BD Aud io Test
Failure: Voltage out of range FE0901 1. Make sure BD alarm cable is con-
(< 3.5 or > 5.05 V). Bad alarm nected.
cable 2. Replace BD alarm.
3. Replace AI PCB.
4. Replace BD alarm cable.
Failure: Voltage out of range FE0902 1. Make sure BD alarm cable is con-
(< 4.5 or > 5.05 V). Bad power fail nected.
cap 2. Replace BD alarm.
3. Replace AI PCB.
4. Replace BD alarm cable.
5. Replace M otherboard PCB.
6-90 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: CLEAR key pressed to indicate user did FE0904 1. Make sure BD alarm cable is con-
not hear alarm, although alarm was active. Bad BD audio nected.
2. Replace BD alarm.
3. Replace AI PCB.
4. Replace BD alarm cable.
Failure: Final capacitor voltage higher than FE0903 1. Make sure BD alarm cable is con-
initial voltage. Bad power fail nected.
cap 2. Replace BD alarm.
3. Replace AI PCB.
Failure: PSOL2 loopback current out of range AE1001 1. Verify that the air supply is good.
(OOR) of drive current. Air PSOL 2. Verify air regulator pressure set to
loopback between 9 and 12 psi.
current OOR 3. Switch PSOLs to see if failure
transfers to the other gas side. If
yes, return oxygen PSOL (PSOL1)
to its original position and replace
PSOL2.
4. Switch Q1 and Q2, run a flow
sensor calibration and rerun test.
If the problem transfers to the
other gas side, return Q1 to its
original position and replace Q2.
5. Replace the AI PCB.
6. Replace the Inspiratory
Electronics PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-91
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: PSOL1 loopback current out of range AE1002 1. Verify that the oxygen supply is
(OOR) of drive current. O2 PSOL good.
loopback 2. Verify oxygen regulator pressure
current OOR set to between 9 and 12 psi.
3. Switch PSOLs to see if failure
transfers to the other gas side. If
yes, return the air PSOL to its orig-
inal position and replace PSOL1.
4. Switch Q1 and Q2, run a flow sen-
sor calibration and rerun test. If
the problem transfers to the other
gas side, return Q2 to its original
position and replace Q1.
5. Replace the AI PCB.
6. Replace the Inspiratory
Electronics PCB.
Failure: Flow controller unable to establish and FE1105 1. Make sure air supply is connected
control gas flow at 60 L/ min. Unable to and unrestricted.
establish flow 2. Check air regulator pressure.
3. Run flow sensor calibration.
4. Switch PSOLs to see if test passes.
If it does, return PSOL1 to its origi-
nal position and replace PSOL2.
5. Replace Q2.
Failure: Excessive safety valve back pressure FE1101 Replace safety valve.
when safety valve is open. Safety valve
occluded
Failure: Safety valve loopback current out of FE1102 1. Replace safety valve.
range. Bad safety 2. Replace AI PCB.
valve driver or 3. Replace Inspiratory Electronics
loopback PCB.
6-92 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: It took too long to relieve excess FE1103 1. Make sure proper test circuit is
pressure through open safety valve, indicating Insp check used.
inspiratory check valve (CV3) is occluded or test valve test 2. Make sure CV3 is not installed
circuit is too large. failed backward.
3. Replace CV3.
4. Replace Safety Valve (SV).
Failure: It took too little time to relieve excess AE1104 1. Make sure test circuit is con-
pressure through open safety valve, indicating Insp check nected.
inspiratory check valve (CV3) may be damaged valve test 2. Make sure CV3 is not installed
or incorrectly mounted. failed backward.
3. Replace CV3.
Failure: Exhalation valve loopback current is AE1201 1. Verify that the system has no leaks
out of range (OOR) of drive current. Exp valve or occlusions by running the leak
loopback test in EST.
current OOR 2. Clean exhalation valve
diaphragm.
3. Calibrate the exhalation valve.
4. Clean/ replace the exhalation
valve.
5. Replace the AI PCB.
Failure: Exhalation valve table checksum invalid FE1304 1. Check for leaks or occlusions. Cor-
or last calibration performed was not Exp valve not rect and then run the exhalation
completed successfully. calibrated valve calibration.
2. Replace the exhalation valve.
3. Replace the AI PCB.
4. Replace Q3.
5. Replace Q2.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-93
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: Exhalation valve magnet temperature FE1302 1. Ensure the unit has been warmed
out of range (OOR) (10 to 100 °C). Exp valve up at ambient temperature for at
temp OOR least 10 minutes.
2. Calibrate the exhalation valve
(EV).
3. Replace exhalation valve (EV).
4. Replace AI PCB.
Failure: Flow controller unable to establish and FE1303 1. Make sure proper test circuit it
control air flow measured by exhalation flow Unable to used and that there are no leaks or
sensor (Q3). establish exp occlusions.
flow 2. Make sure air supply is still con-
nected.
3. Run flow sensor calibration.
4. Replace Q3.
5. Replace Q2.
6. Replace the exhalation valve.
Failure: Seal test ∆ pressure is above failure FE1301 1. Clean exhalation valve.
level. Seal test failed 2. Calibrate exhalation valve.
3. Run flow sensor calibration.
4. Replace exhalation valve.
5. Replace AI PCB.
6. Replace PE.
Failure: Seal test ∆ pressure is above alert level AE1305 1. Verify that the system has no leaks
but below failure level. Seal test failed or occlusions by running the leak
test in EST.
2. Clean the exhalation valve.
3. Calibrate exhalation valve.
4. Replace the exhalation valve.
5. Replace the AI PCB.
6-94 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: Exhalation valve table checksum invalid FE1402 1. Check for leaks or occlusions. Cor-
or last calibration performed was not Exp valve not rect and then run the exhalation
completed successfully. calibrated valve calibration.
2. Replace the exhalation valve.
3. Replace the AI PCB.
4. Replace Q3.
5. Replace Q2.
Failure: Flow controller unable to establish and FE1403 1. Make sure air supply is connected.
control air flow measured by exhalation flow Unable to 2. Make sure proper test circuit is
sensor (Q3). establish exp used.
flow
Failure: Expiratory ∆ P too low, indicating FE2201 1. Calibrate exhalation valve and
velocity transducer not responding properly. Low exp ∆P retest.
2. Replace exhalation valve.
Failure: Flow controller unable to establish and FE1501 Make sure air supply is connected.
control 60 L/ min flow. Unable to
establish air
flow
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-95
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: Heater temperature did not rise FE1502 1. Replace exhalation heater.
sufficiently during an interval after heater was Bad exp 2. Replace AI PCB.
turned on. heater
Failure: Heater temperature did not drop FE1503 1. Replace exhalation heater.
sufficiently during an interval after heater was Bad exp 2. Replace AI PCB.
turned off. heater
Failure: System is still running on battery power AE1602 1. Plug in ventilator power cord and
(i.e., when polled, either BPS charged or BPS ac power not check the cord connection at the
charging line is active). Compressor can only connected ventilator.
run on ac (facility) power. 2. Check ac.
3. Disconnect BPS to isolate prob-
lem.
4. Replace power supply.
Failure: PS2 is closed when it should be open. AE1601 1. Make sure air supply is discon-
Wall air nected.
pressure 2. Disconnect PS2 and rerun test. If
detected test passes, replace PS2.
6-96 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: Calculated run time not within range, AE1604 Listen for motor. If motor is on,
perhaps because compressor timer is not Run mode replace compressor PCB. Otherwise,
running while compressor motor is on. time OOR replace compressor.
Failure: PC indicates that compressor air is not AE1603 1. If compressor system is an 804
present, although compressor motor is on. Compressor model, replace with the 806
pressure not model.
detected 2. Run compressor leak test to check
for leaks.
a. Use leak detector to check for
leaks at the accumulator fittings
and other tubing connections.
b. Verify no leaks at CV2 within
the ventilator by plugging the
air inlet fitting.
3. Replace compressor PCB.
Failure: Calculated disabled time not within AE1607 Replace compressor PCB.
range, perhaps because compressor timer is Disabled
running while compressor motor is off. mode time
OOR
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-97
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: Calculated standby mode time not AE1609 Replace compressor PCB.
within range, perhaps because compressor Standby mode
motor is still running. During standby mode time OOR
test phase, compressor motor should eventually
turn off.
Failure: Compressor load test cannot be run, AE1610 Troubleshoot code AE1603 or
because of inability to verify PC functionality. Unable to AE1606.
perform
compressor
load test
Failure: PC reading indicates compressor AE1611 1. Run compressor leak test to check
unable to maintain minimum pressure under Compressor for compressor leak.
worst-case breath delivery waveform. load test failed a. Use leak detector to check for
leaks at the accumulator fittings
and other tubing connections.
b. Verify no leaks at CV2 within
the ventilator by plugging the
air inlet fitting.
2. Replace compressor.
3. Replace compressor PCB.
Failure: Compressor leak test cannot be run, AE1701 1. Troubleshoot accompanying code
because of inability to verify timer functionality. Unable to AE1604 or AE1607.
perform 2. Replace compressor PCB.
compressor
leak test
6-98 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: PS2 is closed, but it should be open. AE1702 1. Make sure air supply is discon-
Compressor control is restored. Wall air nected.
pressure 2. Disconnect PS2 and rerun test. If
detected test passes, replace PS2.
Failure: PS2 is open, but it should be closed. AE1703 1. Make sure air supply is connected.
Compressor control is restored. Wall air 2. Run Gas Supply/ SV Test.
pressure not
detected
Failure: Touchscreen error occurred (e.g., AE1901 1. Clean touchscreen, removing any
blocked beam or low-level error), buffer cannot GUI touch: obstructions.
be read, or report is invalid. Error 2. Replace touchframe PCB.
3. Replace GUI CPU PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-99
6 Diagnostic codes
Code and
Descript ion Display Correct ive act ions
message
Failure: Received message not identical to AE2001 Replace GUI CPU PCB.
transmitted message. Bad GUI serial
port
Bat t er y Test
Failure: BPS not fully charged. AE2101 1. Allow BPS to fully charge, then
Battery not repeat test.
charged 2. Replace BPS PCB.
3. Replace battery pack.
4. Replace BD CPU PCB.
Failure: BPS voltage dropped below accepted AE2103 1. Ensure that the battery pack is
level or dropped too quickly. Bad Backup fully charged by checking that the
Power Supply green LED on the BPS is lit prior to
initiating an EST. If the amber LED
is lit, allow the unit to charge the
batteries prior to rerunning EST.
2. Replace the battery pack.
3. Replace the BPS PCB.
4. Replace the AI PCB.
5. Replace the power supply.
6-100 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
Diagnostic codes 6
Table 6-10: EST t est sequence (cont inued)
Code and
Descript ion Display Correct ive act ions
message
Failure: BPS not charging after ac power was AE2104 1. Verify that ac power is recon-
reconnected. Battery not nected when prompted and that
charging ac is good.
2. Replace BPS PCB.
3. Replace the battery pack.
4. Replace the power supply.
5. Replace the BD CPU.
6. Replace the AI PCB.
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 6-101
6 Diagnostic codes
The NOTEScol um n i n the log contains the UT0002 Faul t Address (FA). Usin g the FA, you can
determ i ne i f there is a problem with an VGA control ler, and iden tify wh ether it i s th e
upper LCD PCB or the lower LCD PCB. If you cann ot locate a faul t code in Table 6-1 that
m atch es th e three di gi ts of the reported fault code then replace the GUI CPU PCB.
Table 6-11: UT0002 Fault Addresses for LCD invert er PCB errors
The exam ple gi ven in dicates th at the upper VGA Controll er PCB shoul d be repl aced.
6-102 4-070496-00 Rev A (08/ 03) 840 Ventilator System Service Manual
SECTION
S EC T I O N
Alarm handling 7 7
This section describes how to respond to ven tilator al arm s. It al so l ists, in al phabetic order,
m essages displ ayed by the venti lator wh en it detects al arm conditions.
Table 7-1: How the vent ilator responds to alarm condit ions
High Requires immediate Red high-urgency ! ! ! Sequence of 5 tones that If a high-urgency alarm
attention to ensure indicator flashes rapidly. repeats twice, pauses, goes away
patient safety Alarm message flashes at then repeats again spontaneously
top of upper screen. (autoresets), its indicator
remains lit (not flashing)
until you press the alarm
reset key.
840 Ventilator System Service Manual 4-070496-00 Rev. A (08/ 03) 7-1
7 Alarm handling
}
The two highest-
3VTE MAND 2 of last 4 mand breat hs ≤ set limit. Check for leaks, changes i n
patient's R& C. priority active
alarm messages
11:20 25 Jun 2000 are displayed here.
The remedy
message suggests
how to resolve the
alarm condition.
8-01197
3 Correct th e condi tion, if n ecessary, referri ng to Table 7-2. You can ch eck th e alarm log
(Fi gure 7-2). You m ay wan t to ch eck th e conten ts of th e System Di agnostic Log
(accessi bl e through the service men u, Section 4) for related di agnostic codes. The
troublesh ootin g procedures li sted are sequen ced to correct th e m ost probable
m alfunction or to present the m ost efficient corrective action first. The proposed fi xes
listed, h owever, m ay n ot al ways correct the particul ar problem.
NOTE:
• You can change alarm settings even when alarms are active.
• Completing a new patient setup or running EST clears the alarm log.
• When more than one alarm is active and their alarm messages vary in their degree of
seriousness, you should assume that the most serious message is applicable.
4 If necessary, cl ear the m essage or reset the al arm by pressing the al arm reset key.
For specifics about al arm operation, see th e 840 Ventilator System Operator’s and Technical
Reference Manual.
7-2 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Alarm handling 7
8-01198
"Lockabl e" al arm s incl ude all PATIENT DATA an d CIRCUIT DISCONNECT alarm s an d can be
silenced for two mi nutes by pressing the al arm si lence key. A n ew, h igh -urgency alarm
cancel s the si len ce. Each ti m e you press th e alarm silence key, th e sil ence peri od resets to two
m in utes. Each ti me you press the alarm si lence key (whether or n ot there is an active alarm ),
the keypress is recorded i n the alarm log.
Analysis Remedy
Base message Descript ion Correct ive act ion
message message
APNEA The set apnea interval has Check patient and settings.
(lockable) elapsed without the ventilator,
patient, or operator triggering a
breath. Resets when patient
initiates 2 consecutive breaths.
Analysis Remedy
Base message Descript ion Correct ive act ion
message message
CIRCUIT No Check patient/ Ventilator has recovered from 1. Check patient circuit con-
DISCONNECT ventilation. ventilator unintended power loss, detects nections.
(lockable) status. circuit disconnect, and switches 2. Run EST.
to idle mode; upper screen
displays elapsed time without
ventilator support. Resets when
ventilator senses reconnection.
DEVICE ALERT A background check or POST has Check System Diagnostic Log
detected a problem. Resets for any associated error codes.
when ventilator passes EST or Rerun POST/ EST.
POST.
7-4 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Alarm handling 7
Table 7-2: Alarm messages (cont inued)
Analysis Remedy
Base message Descript ion Correct ive act ion
message message
Last spont
breath ≥ set
1PPEAK limit
Last 3 spont
breaths ≥ set
1PPEAK limit
Last 4 or more
spont breaths
≥ set 1PPEAK
limit
Analysis Remedy
Base message Descript ion Correct ive act ion
message message
1VTI SPONT Delivered inspiratory volume ≥ 1. For TC: Check for leaks,
(lockable) high inspiratory volume limit tube type/ I.D. setting.
(breath type = TC or VS) 2. For VS: Check patient and
settings.
LOSS OF Power switch is on, but there is Turn power switch off to reset
POWER insufficient ac and BPS power. alarm. Restore ac power.
There may not be a visual
indicator for this alarm, but an
independent audio alarm sounds
for at least 120 s.
LOW AC Ventilator Power Mains (ac) power has dropped 1. Check integrity of ac
POWER currently not interrupt below 80% of nominal for 1 s. power source.
affected. possible. Ventilator continues operation as 2. If ventilator software was
close to settings as possible. recently downloaded or if
Resets when there is no low ac ventilator was moved
power signal for 1 s. between locations with dif-
ferent voltages, check Ser-
vice Mode Setup
NOTE: configuration.
Compressor operation 3. Check connections of ac
may be disrupted during power distribution compo-
low ac events. nents.
LOW BATTERY Operational Replace or Resets when BPS has more than 1. Check integrity of ac
time < 2 allow approximately 2 min of power source.
minutes. recharge. operational time remaining. 2. Check connections of ac
power distribution compo-
nents.
7-6 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Alarm handling 7
Table 7-2: Alarm messages (cont inued)
Analysis Remedy
Base message Descript ion Correct ive act ion
message message
3O2% M easured Check patient, The O2% measured during any 1. Calibrate oxygen sensor.
O2% < set gas sources, phase of a breath cycle is 7% 2. Run EST.
O2%. O2 analyzer & (12% during the first hour of 3. Replace oxygen sensor.
ventilator. operation) or more below the 4. Inspect oxygen PSOL
O2% setting for at least 30 s, or (PSOL1).
below 18%. (These percentages
increase by 5% for 4 minutes
following an increase in the O2%
setting.) Alarm updated at 1-s
intervals.
3VTE MAND 2 of last 4 Check for Exhaled mandatory tidal volume Run SST.
(lockable) mand. breaths leaks, changes ≤ set limit. Alarm updated
≤ set limit. in patient‘s whenever exhaled mandatory
RSTAT & CSTAT. tidal volume is recalculated.
3V E TOT .
V E TOT ≤ set Check patient Total minute volume ≤ set limit. Run SST.
(lockable) limit for ≤ 20s. & settings. Alarm updated whenever
exhaled minute volume is
.
V E TOT ≤ set Check patient recalculated.
limit for > 20s. & settings.
.
V E TOT ≤ set Check patient
limit for > 40s. & settings.
Analysis Remedy
Base message Descript ion Correct ive act ion
message message
NO AIR SUPPLY Ventilation Check air Ventilator delivers100% oxygen. 1. Check air inlet filter.
continues as source. Resets if air supply connected. 2. Check pressure switch con-
set. Only O2 nections.
available. 3. Run EST.
Compressor Check air Ventilator delivers100% oxygen. 1. Check air inlet filter.
inoperative. source. Resets if air supply connected. 2. Check pressure switch con-
Ventilation nections.
continues as 3. Run EST.
set. Only O2
available.
Ventilation Check patient Ventilator delivers 100% oxygen 1. Check air inlet filter.
continues as & air source. instead of set O2%. Resets if air 2. Check pressure switch con-
set except supply connected. nections.
O2% = 100 3. Run EST.
Compressor Check patient Ventilator delivers 100% oxygen 1. Check air inlet filter.
inoperative. & air source. instead of set O2%. Resets if air 2. Check pressure switch con-
Ventilation supply connected. nections.
continues as 3. Run EST.
set, except
O2% = 100.
NO O2 SUPPLY Ventilation Check patient Resets if oxygen supply 1. Check oxygen inlet filter.
continues as and O2 connected. 2. Check pressure switch con-
set. Only air source. nections.
available. 3. Run EST.
Ventilation Check patient Ventilator delivers 21% oxygen 1. Check oxygen inlet filter.
continues as & O2 source instead of set O2%. Resets if 2. Check pressure switch con-
set, except oxygen supply connected. nections.
O2% = 21. 3. Run EST.
7-8 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Alarm handling 7
Table 7-2: Alarm messages (cont inued)
Analysis Remedy
Base message Descript ion Correct ive act ion
message message
SCREEN BLOCK Possible Remove Possible blocked beam (GUI) or 1. Remove obstruction.
blocked beam obstruction or touch screen fault. 2. Service ventilator.
or touch service
screen block. ventilator.
SEVERE Little/ no Check patient. Ventilator enters occlusion status Run EST.
OCCLUSION ventilation. Provide cycling (OSC) and upper screen
alternate displays elapsed time without
ventilation. ventilator support.
Clear
occlusions;
drain circuit.
VOLUME NOT Last 2 mand Check patient Ventilator cannot deliver target Check patient and settings.
DELIVERED breaths, and setting for volume.
(This alarm pressure > 2PPEAK.
applies to VC+ max allowable
and VS level
breaths.)
Last 10 or
more mand
breaths,
pressure >
max allowable
level
7-10 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
SECTION
S EC T I O N
840 Ventilator System Service Manual 4-070496-00 Rev. A (08/ 03) 8-1
8 Service and repair
8-2 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.5 Elect rical cables and pneumat ic t ubing
• To ensure proper reassem bl y, note or label wi re and tube posi tions before disconn ectin g
parts.
• To avoi d shreddin g a sil icone tube when rem ovi ng from a fi tting, gently pull the tube
whi le turni ng. Do n ot attem pt to rem ove si li cone tubing from barbed fitti ngs i n
inspi ratory modul e. Removal of tubing from non-barbed fitti ngs onl y i s permi ssi bl e.
• M ake sure al l tubes, and harnesses or cabl es, are installed usi ng tie wraps, as specified.
M ake sure wiring does not i nterfere w ith, an d cannot be dam aged by, hi nged or m ovi ng
venti lator parts.
Warning
The failure to replace damaged warning, caution, and identification labels may result in
personal injury or equipment/property damage.
8.10 Test ing, calibrat ion, and ot her post -service procedures
After you com pl ete the ventil ator repair, do th e foll ow ing before pl acing th e ventil ator on a
patient:
• Visual ly verify th at all pn eum ati c and electrical parts are properly con nected an d th at all
parts are properly i nstalled. Then, wi th a l ight tug, verify th at connecti ons are secure an d
that parts are securely attached. Li sten for any un characteristic soun ds (pn eum atic
leaki ng, vi brations, gri ndi ng, squeakin g, or others).
• Perform th e addi ti on al service, testin g, and cal ibrati on activities li sted i n Tabl e 8-1. The
num bers in the colum ns i ndi cate the sequence in whi ch to perform these activi ties.
• Keep a m ain ten an ce l og of all repai rs. M ake sure servi ce records and other docum en tation
are com pleted.
8-4 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
Table 8-1: Test ing and calibration requirement s
Performance
Flow sensor
calibrat ion
calibrat ion
calibrat ion
Download
Soft ware
cycle1
Type of service
Ventilator installation 1 -- 2 3 4 5 6 7 8 92
10,000-hour preventive 1 -- 2 3 4 5 6 7 8 9
maintenance (ventilator or
compressor)
Oxygen sensor 2 1
Warning
Connectors and tubes with the proprietary Bennett barbed cuff fittings are intended for use
only with like fittings. They are not interchangeable with ISO-standard cone and socket
fittings. A leak-tight connection cannot be ensured if these two fitting types are combined.
Adapters may be used to connect Bennett barbed cuff fittings to ISO-standard cone and
socket fittings.
There are two confi gurations of the GUI in th e field: the older 9.4-in ch GUI an d the newer
10.4-inch GUI. The table below li sts the m ajor differences between the two versions.
Communication Ports Single null port Single null port and two serial ports
Print Screen function? Yes; S/ W revision H (US only or Yes; S/ W revision H (US only or
equivalent) and higher equivalent) and higher
The GUI, shown in Fi gure 8-1, is a detachable m odule that can be m ounted on the opti onal
cart or shelf. W hen the ventilator is m ounted on a sh el f, an optional cabl e all ows you to place
the GUI up to 10 ft from the BDU.
8-6 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
Touch screen
interface
Rotary
encoder
knob
Keyboard 8-00189
GUI rear
GUI replacement part GUI mount ed GUI det ached housing
removed
Alarm assembly X X
Touchframe PCB X X
Keyboard assembly X
The service and repair procedures that fol low are organi zed i nto procedures for th e
10.4-i nch GUI, fol lowed by procedures for th e 9.4-inch GUI.
8-01299 8-01297
M ounting Cable
M oun tin g guide
bracket
bracket
Cable
guide
GUI
interface Release
Release knobkn ob
cable GUI
in terface
con n ect ion
8-01328 8-00182
8-8 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.13.2 Removing the 10.4-inch GUI cable assembly
1 Using a Phi lli ps screwdriver, rem ove the screw securi ng the outer h andl e. (Refer to
Figure 8-4.)
2 Rem ove th e top handl e and set aside.
3 Using a sm al l flat blade screwdri ver, unscrew the capti ve screws securi ng the GUI cable.
Gently rem ove the GUI cable and set i t aside.
Top handle
GUI
interface cable
Bottom
Diagnostic handle
LED array
8-01218
Caution
Use care not to overtighten captive screws.
2 Align the tabs on the top h andl e i nto the slots in the bottom h andle. Press h andl e into
pl ace.
3 In sert screw and gentl y ti gh ten un ti l snug.
4 Gen tly separate the rear panel from the front housing and set aside. If necessary, a sm all
flat-bladed screwdriver m ay be used to separate the GUI rear panel from the fron t.
Caut ion
If you use a small flat-blade screwdriver to separate the rear and front housings, take care
not to damage the GUI front and rear panels.
8-01223
8-01324
Figure 8-6. 10.4-inch GUI rear warning label and serial number tag
8-10 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.13.6 Rem oving t he 10.4-inch CPU shield
1 Using a Phi lli ps screw driver, rem ove the 11 screws th at secure the CPU shi eld to the
LCD/GUI bracket (Figure 8-7).
NOTE:
Remove only those screws that go through the shield and not those located in a
cutout.
CPU shield
GUI LED
cable PCB
RS-232
shield
Keyboard
PCB cable
Alarm
speaker cable
(not shown)
Touchframe 8-01217
PCB cable
Figure 8-7. Removing the 10.4-inch CPU shield and GUI PCB
Caution
To maintain an isolated ground system, ensure the nylon washer is placed between the
inverter PCB and the standoff on the GUI CPU PCB when mounting the backlight inverter
PCBs on the GUI CPU PCB.
4 For each of th e backli ght in verter PCBs, di sconn ect th e h arn ess that connects the LCD to
the backli ght in verter an d the cable from th e backl ight i nverter to the GUI CPU PCB
(Figure 8-8).
5 Rem ove each of the PCBs an d place i nto a con ducti ve bag.
backlight inverter/
CPU harness
LCD/backlight Backlight
driver harness driver PCB
(one of two)
Standoffs (2)
Caut ion
Handle the flex circuit cable with extreme care to prevent damage to the flex circuit
connectors.
RS-232 shield
Screw-lock screws
8-12 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
4 Using a Phill ips screwdriver, rem ove th e 10 screws securi ng th e PCB to th e GUI housing.
5 Gentl y ti lt the GUI CPU PCB to th e right and disconn ect the LCD pan el fl ex cables (2)
from the board (J2, J11).
6 Rem ove th e GUI CPU PCB an d place i nto a con ducti ve bag. Set the PCB asi de.
NOTE:
If installing a new 10.4-inch CPU PCB, ensure the touch panel jumpers are set to
the 10.4-inch touchframe style, “ MKG TOUCH.” The five jumpers are located on
the lower left side of the PCB (Figure 8-10).
Jumpers
LCD
shield
EM I Harness from
gaskets backlight
inverter to
LCD panel
LCD
mounting
screws CPU
(4 ea.) shield
8-01325
8-14 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.13.10 10.4-inch LCD panels
Caution
• The LCD panels are easily soiled or damaged. Use caution not to touch the surface of the
LCD panel or the inside of the GUI window as fingerprints will result.
• Do not allow the LCD panels to be exposed to the ambient environment any longer than
necessary as dirt and debris will be deposited on the LCD panels. Should this occur, use
an ESD-protected air source to gently blow away any dust or debris.
Caution
The touchframe PCB should not be flexed or bent as damage will occur. Use caution when
handling and avoid touching emitters or receptors mounted on the board.
GUI LED
PCB
GUI LED
cable
Touchframe
GUI
PCB
alarm
8-01157
Touchframe Keyboard
cable cable
Figure 8-12. 10.4-inch GUI t ouchframe PCB and GUI LED PCB locat ions
8-1220
8-16 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.13.14 10.4-inch GUI alarm assembly
The GUI alarm i s located beh in d the GUI fron t pan el and can on ly be accessed by rem oving
the rear GUI h ousi ng.
Caution
When handling the GUI alarm assembly, be careful not to damage the speaker element.
Speaker
harness
Speaker
element
GUI front
housing
8-01221
Caut ion
Do not touch the inside (coated side) of the GUI window. If necessary, use finger cots or
protective gloves. The window is treated for glare resistance and touching it creates
smudges, which cannot be removed without damaging the window.
8-18 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.13.17 10.4-inch rot or housing
The rotor h ousi ng is easi ly replaced without disassembli ng the GUI h ousi ng.
Fascia
panel
GUI rear
panel
Rotor
Rotation
spring
Brake shoes
8-01337
Caut ion
Be careful not to touch the inside (coated side) of the window. If necessary, use finger cots
or protective gloves. The window is treated for glare resistance and touching it creates
smudges which cannot be removed without damaging the window.
Bezel
Gasket
Screwdriver
placement
Window
Window
gasket
LCD
panels
8-00176
8-00104
8-20 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.14.2 9.4-inch GUI window
Caution
• Be careful not to touch the inside (coated side) of the window. If necessary, use finger cots
or protective gloves. The window is treated for glare resistance and touching it creates
smudges which cannot be removed without damaging the window.
• Avoid cleaning the coated side of the window. If necessary, gently blow off dust or debris
using compressed air.
The window serves as the GUI touch screen; it i s held i n place by th e bezel .
Caution
Be careful not to touch the inside (coated side) of the window. If necessary, use finger cots
or protective gloves. The window is treated for glare resistance and touching it creates
smudges which cannot be removed without damaging the window.
NOTE:
When installing the window, be sure the coated-side (with the painted black strip)
is facing to the inside of the unit.
GUI tilt
positions
8-01339
NOTE: The top of the bezel is narrower than the bottom. When replacing the bezel,
position the narrower end at the top.
Caut ion
Carefully remove the keyboard assembly to avoid damaging the GUI front housing.
2 Slide a flat tool under one edge of keyboard l am in ate and pull up to break keyboard's
adhesive bond to fron t h ousi ng.
3 Once th e adhesi ve has been broken , disconn ect ribbon cable from keyboard.
4 Con tin ue to pull keyboard assem bly away from front housin g. Rem ove en tire assem bl y
includi ng keyboard lami nate, keyboard PCB, an d rotary encoder.
8-22 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
5 Gently rem ove keyboard assem bly and place i n conducti ve bag or on ESD-protected m at.
6 Using al coh ol, clean an y rem ai ni ng adhesi ve from front housin g, being careful not to let
debri s fall i nto the GUI housin g.
NOTE: Do not attempt to reuse the 9.4-inch keyboard assembly after removal.
Keyboard
assembly
Keyboard
laminate
8-00189
8-00116
fig8-19c.tif
8-00262
Caut ion
The keyboard assembly is held in place by a quick bonding adhesive that surrounds the
inside perimeter of the keyboard laminate. Once the adhesive is exposed, it immediately
bonds to anything it touches. Therefore, use extreme care when removing the protective
strip concealing the adhesive, and while aligning the keyboard laminate to the GUI's front
panel.
1 Al ign keyboard over housing and connect keyboard PCB ribbon cable (dark stri pe goes to
pin 1 of GUI CPU PCB J20) (Figure 8-18). Verify that cable is positi vel y l ocked into pl ace.
2 On in si de of keyboard lamin ate, careful ly peel back protective strip th at con ceal s
adhesive.
3 Takin g care not to touch keyboard assem bly to GUI fron t housin g, al ign keyboard
lam i nate (starting with side contai nin g rotary encoder) with keyboard i nset on fron t
housing.
4 Carefully press keyboard i n pl ace by appl yin g pressure to perim eter of lami nate; th en,
check keyboard perim eter to ensure i t i s securely bonded to GUI h ousin g.
M ounting Cable
Cable
M oun tin g guide
guide
bracket
bracket
GUI
interface
connection
Release
Release knobkn ob
GUI
in terface
con n ect ion
8-01328 8-00182
8-24 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.14.6 9.4-inch GUI rear housing
Caution
To prevent damage to ESD-sensitive components, always follow ESD guidelines when
disassembling the GUI.
Top handle
Interface cable
Bottom
handle
Captive screw
(2 PL)
Interface cable
connector
8-00321
8-00108
Figure 8-20. Removing 9.4-inch GUI handle assembly and int erface cable
5 Loosen th e two captive screws that retai n RS-232 port cover; rem ove cover.
6 Unplug nurse’s call port cover.
7 Rem ove 19 screws from GUI rear h ousi ng (Figure 8-21) an d set aside.
8 Gently separate GUI rear and fron t housi ng. If necessary, use a sm all fl at-bladed
screwdriver to separate the GUI rear pan el from the front.
Caution
If you do use a small flat-blade screwdriver to separate the GUI rear and front housings, use
care not to damage the GUI front or rear housing.
1-1/2-in. screw
(13 PL)
3/ 4-in. screw
3/8-in. screw
(5 PL)
8-00109
NOTE:
The screws that secure the GUI rear housing to the front housing come in three
lengths. Refer to Figure 8-21 for correct lengths.
1 With GUI front housi ng face down, ali gn rear and fron t housi ngs.
2 Feed rem ote alarm cover through rear housing.
3 Check perim eter of housin g to m ake sure cables are not being pi nched.
4 Gen tly press rear housin g in to front h ousi ng un ti l two enclosures full y seat.
5 Install GUI rear housin g (Secti on 8.14.6.1).
Caut ion
Ensure cables are not pinched between the front and rear housings. Otherwise, damage to
the cables may result.
6 Install and h and-tighten 19 screws; then , usin g a screwdriver, ti gh ten until sn ug. D o n o t
o vert igh ten .
7 Insert the bottom han dle into h andle seati ng on rear housing (Fi gure 8-20). Install and
han d-tighten two screws; then , using a screwdriver, tighten until snug. Do n o t
o vert igh ten .
8 Connect fem al e end of GUI in terface cable to GUI CPU PCB con nector located in si de
bottom han dle. Tighten two i nterface cable captive screws to GUI CPU PCB un ti l snug.
Do n o t o vert igh t en .
9 Position the i nterface cable into th e handl e sl ots.
10 Sn ap top han dle in to place. In stal l screw an d ti gh ten un til snug. Do n o t ov ert igh t en .
11 Install GUI (Section 8.14.5).
8-26 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.14.7 9.4-inch GUI alarm assembly
The GUI alarm assem bl y is located behi nd the GUI front panel.
J16
J6
J9
J7
Alarm
assembly
Backlight
inverter
PCB
J3
J2
8-00166
Alarm
assembly
8-00110
8.14.8 9.4-inch GUI backlight invert er PCB and GUI LED PCB
The backli ght i nverter PCB, whi ch is part of th e LCD backl igh t control circui t, con trols th e
bri ghtn ess of both LCD panels.
The GUI LED PCB i nterface lets th e GUI m i croprocessor con trol th e di screte diagnosti c LEDs
on the LED PCB.
8-28 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.14.8.2 Inst alling 9.4-inch GUI backlight invert er PCB
1 If appl icable, in stal l backli ght in verter PCB/ GUI CPU PCB ribbon cable to backli ght
i nverter PCB con nector J1.
2 With ribbon cable facing GUI CPU PCB an d com ponents facin g up, ali gn backlight
i nverter PCB to four stan doffs (Figure 8-24).
3 M anuall y in stall and tigh ten four screws; then, usi ng a screwdriver, tigh ten until snug.
Do n o t o v ert igh ten .
4 Connect backli ght inverter PCB/ GUI CPU PCB ribbon cable to GUI CPU PCB con nector
J7 (Fi gure 8-22).
5 Con nect two LCD panel cabl es to backl igh t inverter PCB conn ectors J2 and J3.
6 In stall GUI rear h ousi ng (Secti on 8.14.6.2).
J2
J3
Backlight inverter
PCB
Standoff
(4 PL)
GUI LED
PCB
8-00111
Figure 8-24. Replacing 9.4-inch backlight invert er and GUI LED PCBs
CPU
Screw (13 PL) protective
shield
J20
J19
Keyboard
cable
Touch
frame
cable
8-00112
8.14.10 9.4-inch video cont roller and VGA LCD cont roller PCBs
Later version s of th e 9.4-in ch GUIs use a CPU PCB that is com mon to both 9.4-inch and
10.4-inch GUIs. This l ater PCB design incl udes two bui lt-in VGA control lers th at i nterface
with the LCD panels. If replaci ng an older GUI CPU PCB in a 9.4-in ch GUI with the newer
PCB, discard the VGA controll ers.
8-30 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.14.10.1 Removing 9.4-inch VGA LCD cont roller PCBs (older CPU PCB)
1 Rem ove GUI rear housing (Section 8.14.6).
2 Rem ove CPU EM I shield (Secti on 8.14.9.1).
3 Grasp sides of desired VGA LCD controller PCB and pul l up to di sengage pi ns from GUI
CPU PCB con nectors (Figure 8-26). Place i n conducti ve bag or on ESD-protected m at.
VGA LCD
controller
PCBs
J5
J4
J15
J14
8-00113
8.14.10.2 Inst alling 9.4-inch VGA LCD cont roller PCBs (older CPU PCB)
NOTE:
The two plug-in IP interface VGA LCD controller PCBs are identical and can be
mounted at either GUI CPU PCB location (J14/ J15 or J4/J5). The pin sets at each
end of the video controller PCBs are different, preventing the video controller
PCBs from being mounted backward on the GUI CPU PCB.
1 Ori en t pins on video control ler PCB so they ali gn wi th correspondin g connectors on GUI
CPU PCB (Fi gure 8-26).
2 Gently press vi deo control ler PCB un ti l ful ly seated.
3 Install GUI EM I sh ield (Section 8.14.9.2).
4 In stall GUI rear h ousi ng (Secti on 8.14.6.2).
Caut ion
When removing the GUI CPU PCB, be careful not to damage surface components. Do not
flex t he PCB.
Jumpers
1 Al ign GUI CPU PCB to el even (11) screwholes in mountin g bracket (Figure 8-22). Install
and hand-ti ghten 11 screws. Th en use a screwdri ver to tigh ten the screws un ti l snug. D o
n o t o ver tigh ten .
2 Connect the backl igh t extender cable assem bl ies from the backli ght inverter PCB
conn ectors J2 and J3.
8-32 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
3 Con nect these cables to GUI CPU PCB:
• LCD h arn esses (2) to J6 and J16
• Backl igh t i nverter PCB ri bbon cable to J7
• GUI LED PCB ri bbon cabl e to J3
• GUI al arm harness to J9
4 Align CPU EM I sh ield to 13 screwholes on CPU bracket (Figure 8-25). Install an d h and-
tighten 13 screws. Then , use a screwdri ver to tighten th e screws unti l snug. Do n o t
o vert igh t en .
5 Con nect these cables to bottom of GUI (Figure 8-22):
• Touchfram e PCB ribbon cable to GUI CPU PCB connector J19
• Keyboard ribbon cabl e to GUI CPU PCB conn ector J20
6 In stall GUI rear h ousi ng (Secti on 8.14.6.2).
Caution
• Handle the touchframe PCB with care: do not flex the assembly, damage any of the
surface components, or touch emitters or receivers.
• Take care when disconnecting the LCD harnesses; do not pull on wires. These harnesses
are delicate.
Touchframe PCB
LCD
Backlight
adapter
extender cable
cable
assemblies
LCD
panels
LCD
harnesses
LCD
adapter
cable
CPU bracket
assembly
8-00114
LCD
harnesses
Figure 8-28. Replacing t he 9.4-inch t ouchframe PCB and LCD panel assembly
8-34 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
Short
cable
Long
Foam cable
side
Standoff
Plastic spacer
8-00364
Caut ion
• Avoid touching the front of the LCD panels. Smudges are difficult to remove.
• Use care not to scratch the front of LCD panels.
NOTE:
LCD panels are not interchangeable due to backlight wire lengths. Do not reverse
displays. Complete the backlight replacement for one LCD panel first, then
complete the second backlight replacement.
7 Gen tly li ft the LCD panel from the bracket an d place the panel face down on an
ESD-protected m at.
8 Rem ove the LCD in terface PCB (wi th LCD harnesses attach ed) from th e rear of th e LCD
panel and set asi de (Fi gure 8-30).
9 Rem ove the backli gh t pan el (Figure 8-30):
a. Using a sm al l flat-bl aded screwdriver, strai ghten the smal l m etal tabs around the
perim eter of th e LCD panel.
b. Carefull y rem ove the backl ight panel .
c. Di scard i f replaci ng the panel .
8-36 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
LCD
interface
PCB
Detail:
Bottom of LCD
Backlight interface PCB
panel
Reattach to
backlight cable
extender assembly
(not shown)
Bend and
straighten metal
tabs and secure
backlight panel
LCD panel
8-00374
Caut ion
To prevent damage to filter-retaining studs, do not attempt to remove the pushnuts that
hold the foam filters in place.
Caut ion
Ensure cables are not pinched between the front and rear housings as you reassemble the
GUI unit. Otherwise, damage to the cables may result.
Pushnut
Foam
filter
Foam
filter Pushnut
8-00263
8-00117
8-38 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.14.15 9.4-inch rot or housing
The blue rotor h ousi ng supports the GUI. W hen th e GUI's rel ease knob is sli d to th e right, th e
rotor housin g opens and the m odule is released.
Fascia
panel
GUI rear
panel
Blue rotor
housing
Spring
8-00077
Brake
shoes
8-0131 9 8-01298
The BDU is a removabl e uni t that can be m oun ted to the opti on al cart or shel f (using th e
shelf-m oun t kit).
M any BDU parts can be repl aced with out rem oving the uni t from i ts m ounti ng platform .
Table 8-4 in dicates th e l evel of di sassem bl y required for al l BDU fi eld-replaceabl e uni ts
(FRUs).
As you con duct th e foll owi ng repl acement and in stall ati on procedures for the BDU, you m ay
fi nd it useful to refer to Tabl e 2-2, and Fi gure 2-11 in Secti on 2 of th is manual.
Caut ion
To prevent damage to ESD-sensitive components, always follow ESD guidelines when
servicing the BDU.
8-40 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
M odules removed
BDU
replacement part Pow er Inspirat ory Exhalat ion BDU Card cage
BDU BPS
supply module m odule housing PCBs
AI PCB X
BD CPU PCB X
Power supply X
assembly
Power indicator X X X X X X X
Pressure switches
(PS1/ PS2)
PSOL cartridges X
(PSOL1/ PSOL2)
Inspiratory X
electronics PCB
Oxygen sensor X
Safety valve X
Inspiratory check X
valve (CV3)
Regulators (REG1/ X
REG2)
Inspiratory X
pressure
transducer
autozero solenoid
(SOL1)
Exhalation valve
(EV)
Exhalation flow
sensor (Q3)
Exhalation X
transducer PCB
Filter seal
Expiratory X
pressure
transducer
autozero solenoid
(SOL2)
Exhalation heater X
Exhalation check X
valve (CV5)
Sample port X
M otherboard PCB X X X X X X X
BDU alarm X X X X X X
assembly
Alarm blindmate X X X X X X X
cable
Inspiratory X X X X X X X
blindmate cable
dc power supply X X X X X X X
blindmate cable
ac power supply X X X X X X X
blindmate cable
ac panel X
Humidifier X
receptacle
8-42 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.1 Removing BDU
Warning
To prevent personal injury, take appropriate measures when removing the BDU from its
mounting platform as the unit weighs 18.2 kg (40.1 lb) when fully loaded.
Caution
Use extreme care when moving the BDU. Do not jar or drop the unit.
2 Pull out BDU release handle (Fi gure 8-36) to di sengage BDU locking bracket; whi le
hol din g handle out, sli de BDU off bracket whil e resti ng it on edge of m ounting platform .
Release h andle.
3 Carefull y lift BDU off platform from rear of venti lator, and pl ace on a fi rm , clean surface.
Air Oxygen
Compressor hose hose
hose
GUI cable
Compressor
power cord
BDU
release
handle
8-00264 8-00265
8-44 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.4 Analog int erface (AI) PCB and breat h delivery (BD) CPU PCB
The AI and the BD CPU PCBs (Fi gure 8-37) reside i n the card cage of the BDU.
Data
key
Lookin g i nto th e card cage, the AI PCB sits in th e far right sl ot (toward the fron t of the
ven til ator), and the BD CPU PCB sits in second sl ot from the ri gh t (behind the AI PCB). Refer
to Fi gure 8-37.
Caution
The data key is matched to each ventilator. Ventilator is inoperable without the correct data
key installed. Do not lose the data key.
Card
cage
BD CPU
PCB
2 PL
AI PCB
8 PL 8-00120
I/ O panel
8-46 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
Test button
PTS 2000
connector
Data key
GUI
connector data connector
Caution
Grounding of the BDU can be compromised is the I/ O panel is incorrectly installed.
5 Using a 3/16-i n. nutdri ver, in stal l two latch in g post screws that secure panel to PTS2000
conn ector, and tigh ten until snug. Do n o t o v er tigh t en .
6 Insert data key, i nstall data key cover, and con nect cables to panel.
Caut ion
To prevent electrical shock hazard, always unplug the power cord from facility power and
disconnect the BPS cable from the power supply before servicing the power supply.
8-00169
8-48 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.6 Pow er swit ch (S1)
Warning
To prevent electrical shock hazard, always unplug the power cord from facility power and
disconnect the BPS cable from the power supply assembly before servicing ac components.
8-00154
8-01161 8-00154
Caut ion
Route the humidifier cables so they do not obstruct the ac power supply harness and
blindmate bracket, located at the bottom rear of the power supply compartment.
NOTE:
Consult this manual or the humidifier vendor for the appropriate bracket model
when mounting a humidifier to the ventilator.
8-50 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.7.3 ac panel
8.15.7.3.1 Removing ac panel
1 Rem ove i nspi ratory m odule (Section 8.15.8.8). Th is is n ecessary to provide access to th e
panel .
2 If con nected, disconnect venti lator an d compressor (i f in stal led) power cords from ac
panel , and BPS cabl e from BDU.
3 Rem ove six screw s that secure panel (Figure 8-43) to BDU chassi s. (Two screws also secure
BPS con nector cover to pan el .)
4 Taki ng care n ot to disturb electrical conn ections on i nside of ac panel , gently separate
panel from chassi s.
5 To provide better access to com pon ent con necti on s, cut ti e wrap that secures wi res to
wal l of ac power di stri buti on com partment.
6 Using n eedlenose pl iers, disconnect th e wires from the ac panel com pon ents:
• red wire from termi nal 10 of relay
• bl ack wire from term inal 12 of relay
• bl ue wi re from top l eft-han d termi nal of ac fil ter PCB
• brown wire from top righ t-hand term inal of ac fil ter PCB
• brown wire from m i ddle left-h and term i nal of compressor ac receptacl e
• bl ue wi re from m iddle ri ght-h and term in al of compressor ac receptacl e
7 Using 5/ 16-in . nutdriver, di sconn ect ground wires from ground stud of ac panel .
8 Rem ove ac pan el.
Warning
Replacing key inspiratory module components requires the use of leak detector fluid to
ensure a good gas seal. Use extreme caution when using leak detector in the vicinity of
electronics. Thoroughly dry all components following use of leak detector.
8-52 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
Caution
When replacing damaged O-rings, carefully remove and install to prevent damage to the
manifold ports.
Check for damaged O-rings as you disassemble th e in spi ratory m odul e. If O-rings are
dam aged, pry them off of the ports. Lubricate replacem en t O-rings wi th Krytox grease, th en
gently stretch O-ri ng over base of th e port an d positi on i n place.
Allow replaced O-rin gs to settle i n place; then verify their i ntegrity.
NOTE:
Replace the air inlet filter (F2) and O-ring every 10,000 hours. They are part of the
10,000-hour preventive maintenance kit.
8.15.8.4.1 Removing F2
1 M ake sure ai r source is di sconn ected from venti lator.
2 Drai n any con densate from ai r inlet fil ter bowl by unscrew ing captive drai n plug at
bottom (Figure 8-45). Ti gh ten pl ug.
3 Unscrew bowl.
4 Unscrew fil ter.
Bowl
Top
Groove
Filter
Air filter
assembly Drain
(F2) plug
O-ring 8-00212
8-01118
Fascia
panel
8-00213
8-54 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.8.5.2 Inst alling fascia panel
1 Insert l ip, at top of fascia panel, under ri m of BDU pl asti c h ousi ng and gently press fascia
panel in place (Fi gure 8-46).
2 Install two screws th at secure fascia pan el to BDU ch assis and tighten unti l snug. Do n o t
o vert igh t en .
NOTE:
Replace the oxygen impact filter (F1), oxygen inlet filter (F3) with a new O-ring
and spring every 10,000 hours. (These items are part of the 10,000-hour
preventive maintenance kit.)
Oxygen
filter
assembly
8-00214
8.15.8.6 Oxygen and air pressure swit ches (PS1 and PS2)
8.15.8.6.1 Removing PS1 and PS2
1 Rem ove fascia pan el (Section 8.15.8.5.1).
2 Rem ove oxygen fi lter assem bly (Secti on 8.15.8.5.3).
3 Di sconn ect harness from term inals at top of appli cabl e swi tch (Figure 8-47).
4 Usin g 9/16-in . open-en d wren ch , loosen pressure swi tch (Figure 8-48). Rem ove.
8-00222
8-56 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.8.7 PSOL cart ridge
8.15.8.7.1 Removing PSOL cart ridge
1 Rem ove fascia panel (Secti on 8.15.8.5.1).
2 Pull in spiratory m odule out to access PSOLs:
a. Remove two screws th at retain i nspi ratory m odul e.
b. Fi rm ly grasp bottom of i nspiratory module, by ai r and oxygen i nlet fi tti ngs. Pull
module out of BDU chassi s unti l nyl on l ine atop PSOL m ani fold is visi bl e (Figure
8-49).
3 At top of PSOL m ani fol d, in sert blade of sm all, flat-bl aded screwdri ver into etched notch
adj acen t to nyl on l ine. Sl ide bl ade un der nyl on l in e an d pull l in e up, creating a loop.
Grasp loop and pull li ne com pl etel y out.
4 Di sconnect appropriate PSOL harness from in spi ratory electroni cs PCB. PSOL at your left
i s air PSOL (PSOL2); PSOL at your right i s ox ygen PSOL (PSOL1).
Caution
Place the PSOL cartridge on a clean work surface. Do not drop the assembly.
5 Carefull y pul l cartri dge out of PSOL m ani fol d and set on a clean work surface. Veri fy th at
al l four O-rin gs are present and i n good condi ti on.
Nylon Nylon
line line
8-00223 8-00224
Oxygen
PSOL
(PSOL1)
Air PSOL
(PSOL2)
8-00225
Inspiratory
8-00226 module 8-00163
8-58 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.8.10 Inst alling inspirat ory module
1 Leak-test in spiratory m odule (Secti on 8.15.8.9).
1 Firm ly grasp bottom of i nspi ratory m odul e, by air an d oxygen in let fittings an d al ign
i nspi ratory m odul e wi th sl ots in inspiratory com partmen t (Figure 8-50).
2 Carefull y slide m odul e i nto i nspi ratory com partm ent and to back of BDU chassi s.
3 In stal l two screws that secure inspi ratory modul e to BDU chassis and tighten un ti l snug.
Do n o t o v ert igh ten .
4 In stall fasci a pan el (Section 8.15.8.5.2).
8-01329
NOTE:
The access cover is permanently attached to the instrument by an retaining strap.
4 Press wi re retai ner tab away from the connector to release th e sen sor cable connector.
an d gentl y rem ove the conn ector from the oxygen sensor.
Access port
cover
Cover
retaining
strap
Sensor Check Valve
harness Housing
Sensor Oxygen
harness sensor 8-01332
connector
Caut ion
Ensure that the oxygen sensor O-ring is properly seated on the sensor before installing in the
ventilator. Failure to properly seat the O-ring may result in leaks.
7 Insert the th readed end of th e oxygen sensor i nto the cavi ty on th e check valve h ousi ng
and fin ger-tighten the sensor w ith out usin g excessi ve force. As th e sensor is in stal led,
ensure that it i s not cross-threaded.
Caut ion
To prevent cracking of the sensor body, do not overtighten as you screw it into place on the
check valve housing.
8 Connect the sensor harn ess connector to the oxygen sensor wi th th e ri dge on the cabl e
connector orien ted tow ards th e wire retainer tab on the oxygen sensor.
9 Repl ace the access port cover by fi rst sl idi ng i ts top flap i nto the openin g on the top of
the BDU housi ng.
10 Usin g both th um bs, seat the port cover in th e port by press the two outside corn ers of
the lower fl ap at the j uncture of th e housing edge, fittin g the in to the h ousi ng openi ng.
11 Contin ue using both th um bs and firm ly press th e lower flap i nto pl ace, workin g aroun d
the flap from the outside corners to the bottom center to seal th e access cover. Be sure
that th e cover properly seals th e h ousin g openin g.
12 Cali brate oxygen sensor by pressin g 100% O 2/CAL 2 m i n key. Veri fy that the calibrati on
passes.
13 Run an SST to check the system before placi ng the venti lator i n service.
8-60 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.8.11.2 Removing oxygen sensor (no access port )
If your BDU unit does not have an access port on the top right edge of the h ousi ng, use the
fol lowin g i nstructi on s to replace the oxygen sensor.
1 Rem ove i nspiratory m odul e (Section 8.15.8.8).
2 Di sconnect oxygen sensor h arn ess from the sensor (Figure 8-53). Be sure to press the wi re
retai ner tab i nside the recess of th e oxygen sen sor to rel ease the connector.
3 Unscrew oxygen sen sor from PSOL m anifol d.
Oxygen
sensor
8-00153
Caution
Ensure that the oxygen sensor O-ring is properly seated on the sensor before installing in the
ventilator. Failure to properly seat the O-ring may result in leaks.
2 In sert th e threaded end of th e oxygen sensor i nto th e cavity on th e ch eck valve housing
an d finger-ti gh ten the sensor with out using excessi ve force. As the sensor is install ed,
ensure that it is not cross-th readed.
Caution
To prevent cracking of the sensor body, do not overtighten as you screw it into place on the
check valve housing.
NOTE:
If replacing an earlier version of the inspiratory check valve assembly (silver) with a
newer version (black), ensure the proper fit of the inspiratory filter. Otherwise, the
patient circuit may leak.
8-62 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.8.13.1 Removing right -side plat e
1 Rem ove i nspiratory m odul e (Section 8.15.8.8).
2 Di sconnect th ese from in spiratory electron ics PCB:
• Ai r PSOL (PSOL2) harness from P6
• Ai r fl ow sensor (Q2) harness from P4
• Pressure swi tch harness from P7
• Oxygen fl ow sensor (Q1) harness from P3
• Oxygen PSOL (PSOL1) harn ess from P5
• Oxygen sensor harn ess from P1
• Safety val ve h arn ess from P9 (adjacent to P1)
• Inspi ratory pressure tran sducer autozero solenoid (SOL1) harness from P10
• Tube from i nspiratory pressure transducer (PI) port
3 Ori en t in spi ratory m odule with ri ght-si de pl ate on top.
4 Rem ove six screws th at secure righ t-si de pl ate to ch assis (Figure 8-55):
• Th ree flat-head screws on flat side of pl ate
• One screw at fron t of pl ate
• Two screws at rear of pl ate
5 Rem ove righ t-side pl ate, wi th in spiratory electronics PCB attached, an d place on ESD-
protected m at.
8-00227
8-64 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
3 screws on
left side of
left-side plate
1 screw at front
of left-side plate
2 screws at rear
of left-side plate
Th is p h o t o w ill
b e rep la ced w it h n ew sh ot .
8-00152
NOTE:
If seal is damaged, refer to Section 9 for replacement adhesive tape.
PSOL
manifold Th is p h o t o w ill b e rep la ced .
Remove
(4 PL)
8-00172
NOTE:
The PSOL manifold slides between the left- and right-side plates. If the fit is too
tight, loosen the flat-head screw on the left side plate, allowing the plate to be
lifted slightly to provide sufficient clearance for the manifold.
1 Ali gn flow sensor tubes on PSOL m ani fol d to vents on air sen sor m ani fol d and gentl y
press m anifold until tubes ful ly seat in to vents.
2 Install four flat-head screws that secure PSOL m ani fol d to right- an d l eft-side pl ates
(Fi gure 8-57). Tighten screws. Do n o t o v er tigh t en .
3 If thi rd flat-head screw on left-side pl ate was l oosened to in stal l PSOL m an ifold, tigh ten
unti l sn ug.
4 Connect tube attach ed to top of SOL1 to beneath PSOL man ifold body.
8-66 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
5 M ake these con nection s to inspiratory el ectroni cs PCB:
• Safety val ve harn ess to P9 (th is conn ector i s located between P1 and P10)
• Oxygen sensor harn ess to P1
• Ai r PSOL (PSOL2) harness to P6
• Ai r fl ow sensor (Q2) harness to P4
• Pressure swi tch (PS1 and PS2) h arness to P7
• Oxygen fl ow sensor (Q1) harness to P3
• Oxygen PSOL (PSOL1) harn ess to P5
6 Install i nspi ratory m odul e (Section 8.15.8.10).
PSOL
manifold
Safety
valve
Air flow
sensor
(Q2)
Oxygen
flow
sensor
(Q1)
8-00177
Figure 8-58. Replacing oxygen and air flow sensors (Q1 and Q2) and safet y valve
O-ring
O-ring
8-01184
Figure 8-59. Oxygen and air flow sensors (Q1 and Q2)
8-68 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8-00269
Caution
To prevent component damage when installing the safety valve, do not apply leak detector
fluid.
8.15.8.19 Check valve assembly, regulat or assembly, and flow sensor manifold
8.15.8.19.1 Removing check valve assembly, regulat or assembly, and flow sensor
manifold
1 Rem ove ri ght-si de plate (Section 8.15.8.13.1) and left-side plate (Section 8.15.8.15).
2 Rem ove PSOL m ani fol d (Secti on 8.15.8.16.1).
3 Rem ove four in set screws that secure ch eck valve assem bly an d fl ow sensor m an ifold to
inspiratory fl oor (Fi gure 8-61).
4 Disconn ect oxygen vent tube from inspiratory floor.
5 Pull in g i n direction of check valve assem bly, rem ove fl ow sensor m anifold/regulator
assem bly/ check valve assembly from plate.
6 Pull fl ow sensor m anifold/regulator assem bl y/ check valve assem bly apart i nto three
pieces as shown.
8.15.8.19.2 Inst alling check valve assembly, regulat or assembly, and flow sensor
manifold
1 M ake sure four O-ri ngs on ch eck val ve assem bl y (Figure 8-61) are presen t an d properly
positi on ed (two O-rings on air and ox ygen ports and two O-rings in tran sfer tube port
cavities).
2 Al ign air an d oxygen ports (on ch eck val ve assem bl y) to ai r and oxygen ports on
regulator assem bly.
3 Gently press ch eck val ve assem bl y again st regul ator assem bly until it reach es stop post.
NOTE:
It may be necessary to flex the inspiratory floor plate down in order to seat the
check valve assembly to the regulators.
4 Install two screws that secure check val ve assem bl y to fl oor. Tigh ten screws un til snug.
Do n o t o vert igh t en .
5 Install PSOL m ani fol d (Secti on 8.15.8.16.2).
6 Install ri gh t-si de pl ate (Section 8.15.8.13.2) and left-side plate (Section 8.15.8.15.2).
8-70 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
Regulator
assembly
Check
valve
assembly
Flow
sensor
manifold
Inspiratory
plate 8-00229
Inspiratory
floor plate
8-00230
Regulator
Check assembly
valve
assembly Flow
sensor
manifold
8-00231
Figure 8-61. Disassembling flow sensor manifold/ regulat or assembly/ check valve assembly
REG1 REG2
REG2
REG1 SOL1
fig8-51a.tif
Regulator
plate
SOL1
8-00232 8-00233
NOTE:
REG1 and REG2 are not interchangeable and must be installed in their appropriate
locations. REG1 has an output port, which distinguishes it from REG2.
1 Insert regul ator in appropri ate cutout i n regulator plate, and ali gn m ounti ng screw to
plate (Figure 8-62). Usi ng 7/16-in. socket or open -end wren ch, i nstall retaini ng nut and
tigh ten until snug.
2 If you are instal li ng REG1, con nect vent tube to REG1 output port, at bottom of
inspiratory fl oor pl ate.
3 Instal l regulator assem bly with pl ate (Secti on 8.15.8.19.2).
8-72 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.8.21 Inspirat ory pressure t ransducer aut ozero solenoid (SOL1)
Caution
To prevent damage to tubing, do not attempt to remove silicone tubing from solenoid
barbed fittings. Solenoid FRUs have tubing pre installed.
8.15.9.1 Exhalat ion collect or vial (ECV) and expirat ory filt er (F9)
1 Push the bl ue latch up to release coll ector vial/ expi ratory fi lter.
2 Rem ove assem bly, and di sassem ble as required. For com pl ete detail s on usin g the
collector vial and expi ratory filter, consul t th e 840 Operator’s and Technical Reference
Manual.
NOTE:
To ensure that all patient circuit connections are leak-tight, perform a circuit leak
test by running SST every time you install the filter on a ventilator. Consult the 840
Operator’s and Technical Reference Manual.
Remove
screw
8-00160
8-01171 8-00160
8-01318
8-74 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
Exhalation heater
Heater
harness
Fascia panel
Exhalation
Expiratory pressure check valve
transducer autozero (CV5)
solenoid (SOL2)
Ferrites
Exhalation
flow sensor
(Q3)
Exhalation
valve (EV) Exhalation port
tube
Exhalation
module
harness
Exhalation
transducer PCB
Exhalation
module
harness
Exhalation I/ O cable
Exhalation connector
module
8-01172
8-76 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.9.5 Exhalat ion valve (EV)
8.15.9.5.1 Exhalat ion met abolic sample port
The 800 Seri es ven til ators were ini tial ly equipped wi th a m etaboli c sample port, located on
the fron t fasci a panel of the exhal ation m odul e (Figure 8-66). This sam pl e port h as si nce been
el im in ated and a di fferen t styl e exhalation val ve has been im plem ented.
This n ew exhal ati on valve can be used on origin al 840 ventilators with the metaboli c sam ple
port, but a m in or m odifi cati on to th e exhal ati on m odule is requi red. W hen replaci ng an
exh alation valve wi th a m etabol ic port, you must first elim i nate th e m etabol ic port from the
unit, as descri bed in the fol lowing i nstruction s.
Fascia
panel
Exhalation
sample
port
5 Rem ove and di scard the sam pl e port tubing and the Luer fittin g com ponents.
6 Cover the sam pl e port open ing in the fasci a panel with the l abel provi ded wi th the new
exhalation valve FRU.
7 Remove exhal ation valve from the ex hal ation m odule. (Section 8.15.9.5.3)
8 Install new exh alation valve. (Section 8.15.9.5.4)
9 Install exhal ation m odule in BDU. (Secti on 8.15.9.4).
Caut ion
• To prevent equipment damage, do not attempt to disassemble the exhalation valve. It is
factory-tested as an assembly.
• When removing the exhalation valve, take care not to damage the flow sensor that sits
atop the exhalation valve. Do not drop or jar the valve.
Expiratory Exhalation
pressure heater
transducer assembly
autozero
solenoid (SOL2)
Exhalation
flow sensor
(Q3)
Exhalation
valve (EV)
Exhaust
port and tube Exhalation
module
harness
Exhalation
transducer PCB
8-01163
8-78 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.9.5.4 Inst alling exhalat ion valve (EV)
Caution
• When installing the exhalation valve, take care not to pinch or damage the O-ring at the
bottom of the flow sensor tube.
• Note that improperly installing the exhaust port can result in oxygen accumulation in the
exhalation compartment.
NOTE:
If replacing the exhalation valve with the newer version, note the absence of the
metabolic port. Refer to Section 8.15.9.5.1 for details regarding the elimination of
the metabolic port.
1 Con nect exhal ation m odul e harn ess to con nector on exhal ati on val ve (Fi gure 8-67).
2 With exhalati on valve port facin g out, careful ly m aneuver exhal ation val ve so it sl ides
up onto bottom of flow sensor tube.
3 Con nect sam pl e port tube to top of val ve.
4 Al ign four exh alation valve screwholes to exhal ati on m odule side pan el .
5 W h ile holding exhalati on valve in pl ace, m anually in sert and tigh ten four screws to hold
valve in pl ace; th en usi ng screwdri ver, ti gh ten screws until sn ug. D o n o t o vert igh t en .
6 In stal l exhaust port by sliding grooved portion of exhaust port onto plastic guide; th en
sl ip top of exh aust port over exhalati on port. M ake sure port is firm ly i n place.
7 Con nect exhal ation m odule harness to exh alation I/ O cable con nector.
8 Install exhal ation m odule cover (Section 8.15.9.2.1).
4 Insert the cotton swabs between th e poppet seal and the m etalli c seat of the exhalation
valve. Wi th your in dex finger, l igh tly touch the poppet to preven t rotati on whi le
cleanin g.
Caut ion
Do not attempt to remove the seal from the poppet for cleaning. Otherwise, permanent
damage to the seal may result.
Caut ion
Care should be taken to prevent dislodging the seal from the popet while blowing.
Caut ion
When disconnecting Q3, be careful not to jar or drop it.
8-80 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8-00236
Caution
Ensure that no damage to the O-ring occurs during installation of Q3.
Exhalation
transducer
PCB
8-01169
Caut ion
To prevent damage to tubing, do not attempt to remove silicone tubing from solenoid
barbed fittings. Solenoid FRUs have tubing pre installed.
8-82 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.15.9.8.2 Inst alling SOL2
1 Ori ent SOL2 so th at con nector on side faces downward (Figure 8-67). Install SOL2 to
exhalati on m odule plate using two screws. D o n o t o v er t igh t en .
2 Con nect exhal ation m odul e harn ess to SOL2.
3 Con nect tube from top of SOL2 to conn ector n ear top of exh alation modul e. Conn ect
tube from side of SOL2 to transducer on exhalati on transducer PCB.
4 Install exhal ation m odule (Section 8.15.9.4).
8.15.9.9 Exhalat ion heat er (EXH HTR) and check valve (CV5)
Caution
The exhalation heater may be hot. Be careful when removing it.
8.15.9.9.1 Removing exhalat ion heat er (EXH HTR) and check valve (CV5)
1 Rem ove exh alati on m odule (Section 8.15.9.3).
2 Di sconnect exhal ati on heater (Fi gure 8-70) cabl e from exhalation m odul e harn ess.
3 Loosen th ree screws that secure heater to module.
4 Gently l ift heater to separate i t from flow sen sor tube. Place heater on flat surface.
5 Gently rem ove ch eck val ve from ori fice on exhalati on m odul e.
Exhalation
heater
Exhalation heater
8-00239
Figure 8-70. Replacing exhalat ion heater (EXH HTR) and check valve (CV5)
8.15.9.9.2 Inst alling exhalat ion heat er (EXH HTR) and check valve (CV5)
1 Inspect seal surroun din g orifice that supports exhal ation check valve for cracks or
damage. Replace CV5 assem bl y if required (seal is part of assem bly).
2 M ake sure leaf is lyi ng fl at, then pl ace check valve (Fi gure 8-70) (l eaf side up) i nto groove
of seal surroun din g ori fice.
3 M ake sure O-ri ng at top of fl ow sensor tube is in pl ace.
4 Al ign h eater over fl ow sensor tube and gentl y press heater assembly onto tube wh il e
ali gni ng three screw locati ons to exhal ati on m odule.
5 Ti ghten three screws unti l sn ug. Do n o t o vert igh t en .
6 Con nect h eater cable to exh al ation m odule harness.
7 Install exhal ation m odule (Secti on 8.15.9.4).
8-84 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
Power
switch
Power
indicator
8-00154
Gasket ac panel
Exhalation
I/ O cable
connector
8-00241 8-00242
Warning
To reduce the risk of fire hazard due to oxygen enrichment in the power supply
compartment, make sure the gasket forms a complete seal between the exhalation module
and power supply compartment.
4 From in si de exhalati on com partm en t, install two screws and standoffs th at secure
exhal ation I/O cabl e conn ector to chassi s. M ake sure l atch ing block “ barbs” are facing
out.
5 Reach i nside power supply com partm ent and conn ect h arn ess to power i ndi cator and
power swi tch.
6 Con nect BDU LED PCB cable to PCB.
7 Install exhal ation m odule (Secti on 8.15.9.4).
8 Install in spiratory m odule (Secti on 8.15.8.10).
9 Install power supply assem bly (Section 8.15.5.2).
10 Install BDU (Section 8.15.3).
Caut ion
Be careful not to damage the upper and lower card guides.
3 Usin g 1/8-in. (3-mm ) flat-bladed screwdri ver, loosen two capti ve screws on each of four
bl indm ate cabl es con nected to m oth erboard PCB (Figure 8-73). Disconn ect cabl es.
4 Usin g 3/16-i n. n utdriver, rem ove ei ght stan doffs th at secure motherboard PCB
connectors to ch assis.
5 Rem ove seven screws that secure m otherboard PCB to back pan el of card cage (Fi gure
8-72). Gently remove m otherboard PCB th rough card cage open in g.
6 Place PCB in a con ductive bag or on ESD-protected m at.
8-86 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
Screw
(7 PL)
Card
guide
8-00243
8-00244 8-00245
Motherboard
BDU alarm
assembly
Alarm
blindmate
cable
Standoff
(8 PL)
Inspiratory
blindmate
cable
dc power supply
blindmate cable
Exhalation
I/ O cable
8-00173
Figure 8-73. BDU alarm and mot herboard PCB cable connections
Warning
To prevent electrical shock hazard, always unplug the power cord from facility power and
disconnect the BPS cable from the power supply assembly before servicing ac components.
8-88 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
Power indicator
8-00154
Auxiliary
(compressor/ humidifier)
circuit breaker (CB2)
8-00246
8-90 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
To inspiratory
module
Inspiratory
blindmate
cable
To motherboard PCB
8-00179 8-00154
Warning
To reduce the risk of fire hazard due to oxygen enrichment in the power supply
compartment, make sure the gasket forms a complete seal between the inspiratory module
and power supply compartment.
4 Pl ace bracket over ch assi s studs and secure wi th three locknuts. Usi ng 11/32-i n.
n utdriver or socket wrench, tighten until snug (Figure 8-76). D o n o t o v er tigh t en .
5 In stall BDU housi ng (Section 8.15.10.2).
To
power
supply
dc power
supply
blindmate
cable
To
motherboard PCB
8-00180 8-00154
8-92 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
To Bracket
power
indicator ac power
supply
To ac
blindmate
receptacle
harness
To power
supply
To ac panel
ground stud 8-00178 8-00154
Warning
To reduce the risk of fire hazard due to oxygen enrichment in the power supply
compartment, make sure the gasket forms a complete seal between the inspiratory module
and power supply compartment.
Caution
The obstruction of the harness within the bracket could prevent proper alignment of power
supply contacts.
8 Using two sm all ti e wraps, rebun dle wires on ei ther side of roun d grom m et. Cut ti e wrap
ends fl ush with the ti e wrap body.
9 Install four chassi s-retaini ng screws at rear of ac panel.
10 In stal l BDU h ousi ng (Section 8.15.10.2).
NOTE:
The 806 compressor is the replacement for the 804 compressor assembly.
The 806 compressor un it i s a detachabl e m odul e consistin g of th e com pressor, com pressor
PCB, accum ul ator, and heat exchanger. Th e compressor uni t is m ountable on ly when used
with the opti onal cart.
M ost routi ne repairs and preventive m ai nten ance to th e com pressor uni t requi re th at th e
com pressor be removed from th e cart. This section descri bes how to perform service an d
repai rs on the 806 com pressor, as well as perform an ce veri fi cati on.
To ensure th at reassem bly is correctly done, label al l w ires and tube positi on s before you
begin discon necti ng parts. Take precaution s to prevent dirt an d oth er particl es from enterin g
the interi or of th e com pressor. Always properl y dispose of the ori gin al parts that have been
replaced by new parts.
Repl acem en t parts for th e origi nal com pressor, model 804, are no longer avai labl e. Contact
Puritan-Ben nett Techni cal Support for additional i nform ati on .
Be su re to r ea d a ll a p p lica b le in str u ct io n s com p let ely b efore d isa ssem b ly.
8-94 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.17 Servicing t he 806 compressor
NOTE:
To disconnect a pneumatic hose held in place by a compressible clip, use flat-
nosed pliers to compress the nubs on the clamp. Slide the clamp up the hose 1 to
2 inches.
Accessory
drawer
Compressor
handle
Compressor
Casters module
(facing out) (slide forward)
8-01335
Warning
To prevent personal injury and equipment damage, have someone assist you when lifting
the heavy ventilator and its components.
8-01259
8-96 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.17.5 Inst alling t op cover
1 Pl ace th e m ai n in let fil ter and support baffl e i nto position at top of front panel.
2 Pl ace th e top cover assembly on th e compressor.
3 Tighten th e four (4) screws (two on each si de) to secure the cover to the side panel s.
Caution
When removing the fan hardware:
• Be careful not to drop the screws or washers into the plenum assembly and onto the heat
exchanger.
• Be careful not to misplace the star washers as they provide ground to the fans.
3 Facin g th e front of the com pressor, rem ove the four (4) Phil li ps screws and star w ashers
that secure the fan to th e top of th e pl enum assembly.
4 Di sconnect th e electrical cable from the fan.
5 Rem ove th e fan .
8-01256
Caut ion
Verify the flow direction of the fan prior to installing the fan in place. The flow should be
directed into the compressor compartment.
1 Based on fl ow direction, properl y orien t th e replacem en t fan and place over the hole in
the pl enum assem bl y.
2 Connect the fan cord to the fan and place the fan on the fan-support panel .
3 Reinstall the four (4) screws and star washers.
4 Secure th e fan cable to the fan’s m oun ti ng screw with a cabl e ti e.
5 Reinstall the baffle and in take fi lter.
6 Reinstall top cover an d secure screw s.
Fitting
and Accumulator
O-ring (4) intake port 8-01270
8-98 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.17.9 Replacing an accumulator fitt ing and O-ring
Each of the four accumulators has a fitti ng with an O-ri ng that conn ects the accum ul ator to
the accumulator intake port.
1 Rem ove back panel (Secti on 8.17.7).
2 W h ile supportin g th e back pan el, disconn ect th e hose from the accumulator i ntake port.
3 Using a nut driver, rem ove th e four (4) nuts that secure the bottl es to th e rear panel.
4 Rem ove th e accum ul ator assem bly from th e back pan el .
5 Rem ove the compressible hose clam ps and disconnect the hoses from either si de of the
accum ulator.
6 Unscrew the accumulator fittin g and rem ove from th e bottle.
7 Replace th e accum ulator fittin g, applyi ng a sm al l am ount of Krytox grease to the O-ri ng
before reinstalli ng on th e accum ul ator.
8 Resecure the hoses to the accum ulator wi th the com pressible hose clam ps.
9 Rein stal l the accum ulator assem bly and place the accum ulator cover over the four (4)
studs.
10 Secure the accum ulator assembly to the back panel usi ng four (4) nuts. Hand tighten.
NOTE:
• To prevent the shredding of the silicone hoses when disconnecting from barbed fittings,
gently pull while turning the hose.
• Use a flat-nosed pliers to compress the nubs on compressible clamp.
Solenoid filter
outlet port
8-01271
Figure 8-83. Disconnect ing filter out let port from main supply hose
6 Di sconn ect the hose at the heat exchanger i nlet conn ecti on that runs from the
compressor outl et (Figure 8-84).
Disconnect
inlet connection
here
8-01334
7 Di sconn ect the hose at the water trap outl et that runs to the air dryer inl et port.
8 Loosen the two (2) Ph il lips screws that secure the i ntern al pl enum panel to th e left an d
ri ght panels.
9 Loosen the two (2) Ph il lips screws that secure the external pl enum pan el th e l eft and
ri ght panels.
8-100 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.17.11.1 Removing t he plenum assembly
1 Rem ove top cover (Section 8.17.4).
2 Rem ove rear panel assem bl y (Section 8.17.7).
3 Di sconnect th e electri cal cabl es from the com pressor PCB as foll ows (Fi gure 8-85):
• solenoid cabl e at J1
• two (2) fan cables at J2 an d J3.
• ac in put cord at J6
• m otor cabl e at J5
4 Release th e data cable by sl idi ng the l ock bracket up an d careful ly disconnecti ng the
cabl e.
5 Use a 11/32" n ut dri ver to rem ove the nut securin g the ground cable to th e grounding
l ug. (Locate the ground con nection on the pl en um base, adj acent to the com pressor
PCB.)
6 Di sconnect th e groun din g cable an d reinstall the nut for recon necting th e groun d l ater.
7 Di sconnect th e pn eum ati c h oses (Section 8.17.10.0.1).
Pressure transducer
hose
Fan connectors
M otor
cable 8-01272
8 Facin g th e rear of the com pressor, place your thumbs under the edge of the plen um
assem bl y top pan el and li ft th e panel free from the si de panel .
9 Sli gh tl y l ift (or prop open) the pl enum assem bly enough to gain access to th e h eat
exch anger ports.
10 Di sconnect th e hose at th e heat exch an ger outlet port that runs to the w ater trap i nlet.
11 Taki ng care not to dam age the heat exch an ger fin s, carefull y l ift th e plenum assem bly
from the top of th e com pressor an d set assem bl y aside.
8-102 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.17.15 Compressor PCB
The rem oval of th e com pressor PCB requires the partial or com plete rem ove of the pl enum to
gai n n eeded access.
NOTE:
If you are installing a new PCB, it is necessary to transfer the compressor serial
number and operational hours information by removing the EEPROM from the
old board and transferring it to the new board.
Caution
Verify the polarity of the EEPROM before installing on the new PCB: When transferring the
EEPROM, ensure that the notch on the EEPROM matches the notch on the 806 PCB before
installing.
9 Rein stal l the new board by first ali gn ing the board with th e si de panel an d pl aci ng the
bottom of the board i nto the board retain er.
10 Reattach the cable con necti on s to th e PCB as foll ows:
• solenoid cabl e at J1
• two (2) fan cables at J2 an d J3.
• ac in put cord at J6
• m otor cabl e at J5
11 Attach the data cable to th e PCB and sli de i ts lock i nto place.
8.17.16.1 Removing and reinst alling t he air dryer and solenoid valve assembly and
replacing filt ers
1 Rem ove th e top cover assem bly an d set aside (Section 8.17.4).
2 Rem ove th e rear panel and set aside (Section 8.17.7).
3 Cut the cable ti e securi ng the air dryer assem bly to the clam p (Fi gure 8-86).
Solenoid
outlet
Air
dryer filter
hose Air dryer
muffler/filter
Solenoid
purge outlet
hose
Cable tie
8-01263
8-104 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
19 Recon nect th e short h ose from th e ai r dryer to the solen oi d purge outl et fittin g.
Use a flat-nose pli ers to sli de th e hose clam p over the fitti ng to hold the h ose i n place.
20 Rein stal l the ai r dryer and solenoi d val ve assem bly in the compressor.
21 Secure th e assem bl y wi th a cable ti e (Fi gure 8-86) by sli din g th e cable tie under th e
bracket and aroun d the air dryer assem bl y.
22 Tighten cable tie and cut off excess.
To accumulator
To pressure
Unloading transducer (PC)
Solenoid
solenoid valve
To solenoid (SOL3) Filter (F10)
assembly
From heat
Air dryer exchanger
assembly
Air dryer Filter/muffler 8-01266
Warning
To prevent personal injury and equipment damage, have someone assist you when lifting
the heavy compressor assembly.
7 Rein stal l the com pressor assem bl y by ori enti ng the com pressor assembly wi th the
com pressor i ntake fi lter on the left.
8 Carefull y l ower th e com pressor in to the com partm ent. Sl ightl y ti lt th e com pressor
forward to posi ti on i n pl ace.
9 Usin g a m agn etic Phi lli ps screwdriver, rei nstall th e four (4) ori gin al screws that secure
the com pressor assem bl y to th e base of the uni t.
Water
8-01260
Coalescing
filter
element
Water
trap
5 Usin g a m agn eti c Phi lli ps screwdriver, rem ove the two (2) screws that secure the water
trap bracket to the base of the m odule. Set the screws aside.
6 Place the water trap assem bly on a work surface.
8-106 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
7 Unscrew the ri ng th at secures th e water trap bowl to the fil ter head.
8 Rem ove th e bowl and ring and set aside.
9 Unscrew the fi lter elem ent retainer and set aside.
10 Verify that the O-ri ng i s in place aroun d th e filter h ead.
11 In stall the new coalesci ng filter elem ent.
12 Replace th e ori gin al fil ter elem ent retai ner.
13 Rein stal l the water trap bowl and hold it i n place by rein stal li ng the original rin g.
14 Using a m agnetic Phi lli ps screwdriver, rem oun t th e water trap assem bl y bracket to th e
base of the compressor usin g th e ori gi nal two (2) screws.
15 Reinstall hose on outl et si de of coalescin g filter.
16 Rein stal l ai r dryer (Section 8.17.16.1).
17 Rein stal l the back panel assem bl y (Secti on 8.17.8) an d the pl enum assem bl y (Section
8.17.11.1).
NOTE:
If replacing a side panel, use the spare piece of foam in the foam kit to configure
the panel to match the panel removed. Install the Tinnerman clips to the rear of
the panel.
NOTE:
The solenoid electrical cable also connects to the PCB at J1. Refer to reinstallation
instructions for the air dryer and solenoid valve assemblies in (Section 8.17.16).
8-108 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
3 W h ile resti ng the bottom of th e rear pan el on th e base of the uni t.
4 Take th e h ose that connects th e solen oi d outlet to the accum ulator in let port and route
thi s h ose beh in d th e m ai n supply hose.
5 W h ile supporti ng the rear panel i n place, reconnect the h ose runn in g from the m ain
supply tee to the accum ul ator inl et port.
6 Reposi tion the hose cl am p over th e i nlet port to h ol d th e hose in pl ace.
7 Taki ng care n ot to crim p h oses or cabl es, press th e back pan el in to pl ace and secure to th e
unit usi ng the original si x (6) Phi lli ps screws.
8 Once the rear panel/ accumulator assembly is secured i n place, fi ni sh tigh ten in g th e four
(4) Phi lli ps screws th e secure the plenum assem bl y to th e si de panel s.
Warning
To prevent personal injury and equipment damage, have someone assist you when lifting
the heavy ventilator and its components.
1 Rein stal l the com pressor onto the cart and secure usi ng the captive hardware on th e cart.
2 Reattach and secure the com pressor data cabl e.
3 Reattach the compressor power cord.
4 Secure th e power cord usi ng the power cord retainer. Secure the retai ner with two (2)
screws.
5 Reattach the compressor supply tube.
GUI mount
BPS
Rear caster
(2 PL)
Front caster
(2 PL)
8-01208
8-110 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.18.2 Inst alling BPS
1 From fron t of cart, sl ide brackets at top of BPS onto rail s, then push BPS to back (Fi gure
8-90). Secure BPS to rai ls wi th captive screws.
2 Open protective cover on BDU con nector, and conn ect BPS cabl e to BDU (Figure 8-91).
Tighten capti ve screws.
Back of BPS
BPS
cable
8-00038
Caution
To prevent damage to ESD-sensitive components, always follow ESD guidelines when
disassembling the BPS.
2 PL
Fascia
panel
4 PL
Front
panel
J1 Floor
assembly
Battery
J3 pack
Chassis
Nut
(4 PL)
BPS
PCB
Enclosure
8 PL
8-00134
8-112 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.18.4 BPS PCB
8.19 Cart
8-114 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Service and repair 8
8.19.3 Rem oving/ inst alling flex arm inserts
On older style carts, rem ove and i nstall th e flex arm inserts as fol lows:
1 Rem ove flex arm or any item secured i n flex arm in sert (socket).
2 Remove flat-h ead screw from cart di rectl y beneath in sert in cart. Rem ove insert.
8-01402
3 Pl ace i nsert i nto cavi ty in cart, ali gni ng screwhol es i n cart and insert.
4 From beneath the cavity of the cart, i nsert fl at-head screw i nto insert and tighten unti l
sn ug. Verify insert is tigh t an d does n ot rotate.
On newer. ready to assem bl e (RTA) carts, rem ove an d i nstal the flex are i nserts as fol lows:
1 Sli de locki ng ri ng off the GUI m oun t.
2 Install new insert.
3 Rein stal l locki ng ring.
8-116 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
SECTION
S EC T I O N
Parts list 9 9
9.1 How t o use t his part s list
Th is sectio n begin s by sh o win g th e en tire ven tilator system , in clu d in g accesso ries.
Su bseq uen t figu res sh o w ven tilato r sub assem b lies an d th eir com pon en t p arts.
Th is sectio n uses th e followin g co n ven tio n s:
• Two d ash es (-) in th e Part no. colu m n in d icate th at th e p art is n o t o rd erab le; it is sh o wn for
referen ce on ly.
• Bullets (• ) in th e descriptio n colum n in d icate levels o f in d en tation . For exam p le, a p art
with a n o n -b u lleted d escriptio n in clu d es all th e su bseq uen t sin gle-bu lleted p arts. A p art
with a sin gle-bulleted descrip tion in clu d es all th e su bseq u en t do u ble-bulleted p arts, an d
so on .
• Ab breviation s an d acron ym s used in th is section are listed in Tab le 9-1.
840 Ventilator System Service Manual 4-070496-00 Rev. A (08/ 03) 9-1
9 Parts list
A/ R As Required
BD Breath Delivery
EX LK External Lock
FIL Fillister-head
FH Flat-head
HEX Hexagonal
I/ O Input/ output
ID Inside Diameter
IN LK Internal Lock
OD Outside Diameter
PAN Panhead
P/ N Part Number
SOC Socket
SR Split-ring
9-2 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
9.2 840 Vent ilat or Syst em pat ient syst em and accessories
It em no.
Part no. Quant it y Description
(Figure 9-1)
4-075659-00 • M ounting kit, Fisher & Paykel 480/ 730 humidifier, ventilator-
powered (120 V), cart-mount, Spanish/ Portuguese
20 -- 1 • • Plate with cutout, humidifier, English/ French
• • Plate with cutout, humidifier, Spanish/ Portuguese
It em no.
Part no. Quant it y Descript ion
(Figure 9-1)
9-4 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
25 15
(4 PL)
24
3
4
23
20 (4 PL) 2
Fisher & Paykel 480/ 730 humidifier,
ext ernally-powered (220-240 V) 6
8
22 (3 PL)
18 9
10
19
7
5
21
17
12
13
8-00190
840 Ventilator System Service Manual 4-070496-00 Rev A (08/ 03) 9-5
9 Parts list
9.2.1 840 Vent ilat or Syst em NeoM ode pat ient syst em and accessories
It em no.
Part no. Quant it y Descript ion
(Figure 9-2)
9-6 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
2
3
8-00137
It em no.
Part no. Quant it y Descript ion
(Figure 9-3)
9-8 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
6 (3 PL)
5 (3 PL)
3 (3 PL)
8
2 (6 PL)
1 (3 PL)
10
7
11
12
13
8-01229
It em no.
Part no. Quant it y Descript ion
(Figure 9-4)
1 4-001474-00 1 Hose assembly, oxygen, DISS female x DISS female, clear, 3-m (10-ft) (for
USA) (See Section 9.2 for NHA.)
2 4-074710-00 1 Hose assembly, oxygen, DISS female x DISS male, 3-m (10-ft) (for
Canada) (See Section 9.2 for NHA.)
3 4-074697-00 1 Hose assembly, oxygen, Air Liquide female x Air Liquide female, 3-m (10-
ft) (for France) (See Section 9.2 for NHA.)
4 4-074711-00 1 Hose assembly, oxygen, SISfemale x SIS female, 3-m (10-ft) (for Australia,
New Zealand) (See Section 9.2 for NHA.)
5 4-074698-00 1 Hose assembly, oxygen, NIST female x BOC male, 3-m (10-ft) (for United
Kingdom, Ireland) (See Section 9.2 for NHA.)
6 4-074715-00 1 Hose assembly, oxygen, DISS female x Dräger male, 3-m (10-ft) (for
Germany) (See Section 9.2 for NHA.)
7 4-074700-00 1 Hose assembly, oxygen, NIST female x NIST female, 3-m (10-ft) (for
Netherlands) (See Section 9.2 for NHA.)
8 4-074702-00 1 Hose assembly, oxygen, DISSfemale x DISSfemale, green, 3-m (10-ft) (for
Israel, Japan, Saudi Arabia) (See Section 9.2 for NHA.)
9 4-074705-00 1 Hose assembly, oxygen, DISSfemale x DISSfemale, white, 3-m (10-ft) (for
Egypt, India, Italy, Kuwait, Poland, Portugal, South Africa) (See Section 9.2
for NHA.)
10 4-074708-00 1 Hose assembly, oxygen, DISS female x DISS female, blue, 3-m (10-ft) (for
Switzerland) (See Section 9.2 for NHA.)
9-10 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
1 (clear)
8 (green)
9 (white)
10 (blue)
6 8-00073
It em no.
Part no. Quant it y Descript ion
(Figure 9-5)
1 4-006541-00 1 Hose assembly, air, DISS female x DISSfemale, clear, 3-m (10-ft) (for USA)
(See Section 9.2 for NHA.)
2 4-074709-00 1 Hose assembly, air, DISSfemale x DISSmale, 3-m (10-ft) (for Canada) (See
Section 9.2 for NHA.)
3 4-074696-00 1 Hose assembly, air, Air Liquide female x Air Liquide female, 3-m (10-ft) (for
France) (See Section 9.2 for NHA.)
4 4-074712-00 1 Hose assembly, air, SIS female x SIS female, 3-m (10-ft) (for Australia, New
Zealand) (See Section 9.2 for NHA.)
5 4-074713-00 1 Hose assembly, air, NIST female x BOC male, 3-m (10-ft) (for United
Kingdom, Ireland) (See Section 9.2 for NHA.)
6 4-074714-00 1 Hose assembly, air, DISS female x Dräger male, 3-m (10-ft) (for Germany)
(See Section 9.2 for NHA.)
7 4-074701-00 1 Hose assembly, air, NIST female x NIST female, 3-m (10-ft) (for
Netherlands) (See Section 9.2 for NHA.)
8 4-074703-00 1 Hose assembly, air, DISS female x DISS female, black, 3-m (10-ft) (for
Israel, Japan, Kuwait, Poland, Portugal), 3-m (10-ft) (See Section 9.2 for
NHA.)
9 4-074706-00 1 Hose assembly, air, DISS female x DISS female, white/ black, 3-m (10-ft)
(for Egypt, India, Italy) (See Section 9.2 for NHA.)
10 4-074707-00 1 Hose assembly, air, DISS female x DISS female, yellow, 3-m (10-ft) (for
Switzerland) (See Section 9.2 for NHA.)
-- -- 1 Common module for the following: DISS male, DISS female, Air Liquide,
NIST, CIG (See Section 9.4.7.2 for details.)
9-12 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
1 (clear)
8 (black)
9 (white/black)
10 (yellow)
6 11 (white)
8-00074
It em no.
Part no. Quant it y Descript ion
(Figure 9-6)
1 4-078107-00 1 Power cord, , latching, for North America (See Section 9.2 for
NHA.)
9-14 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
1 2
3 4
5 6
7 8
8-00070
9
It em no.
Part no. Quant it y Descript ion
(Figure 9-7)
1. Contact your Puritan Bennett representative to obtain 806 compressor unit ordering information.
9-16 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
3
4
5
8-01204
It em no.
Part no. Quant it y Descript ion
(Figure 9-8)
-- 4-075143-00 1 Label kit, ventilator, 100 V, 50/ 60 Hz, Japanese (See Section 9.4 for
NHA.)
4-075413-00 Label kit, ventilator, 120 V, 60 Hz, English, US (See Section 9.4 for
NHA.)
4-071559-00 Label kit, ventilator, 120 V, 60 Hz, English/ French (See Section 9.4
for NHA.)
4-075099-00 Label kit, ventilator, 120 V, 60 Hz, Spanish/ Portuguese (See Section
9.4 for NHA.)
4-071590-00 Label kit, ventilator, 220-240 V, 50 Hz, English/ French (See Section
9.4 for NHA.)
4-075436-00 Label kit, ventilator, 220-240 V, 60 Hz, English (See Section 9.4 for
NHA.)
4-075059-00 Label kit, ventilator, 220-240 V, 50 Hz, German (See Section 9.4 for
NHA.)
4-075400-00 Label kit, ventilator, 220-240 V, 50 Hz, Portuguese (See Section 9.4
for NHA.)
4-075072-00 Label kit, ventilator, 220-240 V, 50 Hz, Spanish/ Italian (See Section
9.4 for NHA.)
9-18 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
Label kits parts list (cont inued)
It em no.
Part no. Quant it y Description
(Figure 9-8)
It em no.
Part no. Quant it y Descript ion
(Figure 9-8)
9-20 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
Label kits parts list (cont inued)
It em no.
Part no. Quant it y Description
(Figure 9-8)
12 4-075101-00 1 • Label kit, 806 Compressor, 100 V, 50/ 60 Hz, Japanese (Not
shown)
4-074015-00 • Label kit, 806 Compressor, 120 V, 60 Hz, English/ French (Not
shown)
It em no.
Part no. Quant it y Descript ion
(Figure 9-8)
9-22 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
20
19
1 6 5
2
21
18
13
8-01181 8-0002
8-00298
14
8-00029
8-00030
8-01182 8-01183
10
11
8-00033 8-00034
9-24 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual
Parts list 9
9-26 4-070496-00 Rev. A (08/ 03) 840 Ventilator System Service Manual