Order No. Tag No. Job Name FIT-603 Agency End User Rgst. No. Plant Name Quantity 1 Model&Suf.Code Reduced bore type select Detecter Inner dia. :E10 150mm 6in Fluid Name Hydrogen Kind of Fluid :C20,F10 Gas Maximum Flow Rate :C45,B10,F40 30000 Nm3/h Minimum Flow Rate 5000 Nm3/h Scale Condi. :C32,C35,F19,F22 0degC 1atm 0% Operating Temperature :C31,D21 90.0 degC Maximum Temperature 116.0 degC Operating Pressure 400.0 psi gauge Operating Pressure(abs) :C34,F21 414.7 psi abs Minimum Operating Pressure 40.00 psi gauge Deviation Factor :C36,F23 1.000 Density at Scale Condition 0.00561 lb/cf Density at Oper. Condition :C26,D26,F27 0.11907 lb/cf Viscosity 0.012 cP Total Rate :B45 0.00000 Nm3/pulse Pulse Rate :B21 0.00000 Nm3/pulse Low Cut Value :D10 4684 Nm3/h Flange Piping Size : 200mm 8in Reduced Detector Size Selection Turn down Pres. drop
R2 19.7 1.23 psi
R1 19.7 0.23 psi
0 20 40 60 80 100% Max.=30000 Nm3/h
Measurable Zone Accuracy Zone (*) High Velocity Zone Min. Flow Rate Accuracy dividing point(35m/s) (out of accuracy) (* Refer to GS01F06A00-01E) Selected Detecter Inner dia. R2 R1 Result Good Good Reason of NG Reynolds Number at Max. Flow 847143 572493 Measurable Maximum Flowrate(Nm3/h) 42240.353 92491.281 Measurable Minimum Flowrate(Nm3/h) 2138.935 4683.504 Flow Velocity at Maximum Flowrate(m/s) 56.818 25.948 Measurable Minimum Flow Velocity(m/s) 4.051 4.051 Min. Flow for Specified Accuracy(1.0%)(Nm3/h) 2138.935 4683.504 Predicted Error --- --- at Low Reynolds Number(of reading) CHECKED.BY