Professional Documents
Culture Documents
à
ñ
à
×
à
J
?
!
"
J
#
3
©
""!
$$
J$
c
à
3
?
?
c
×
In PTC thermistors the resistance increases with temperature. They are
manufactured using barium titanate and are not normally used in temperature
measurement. These devices are selected when a significant change in the
resistance is required at a specific temperature or current. Examples of
applications include temperature sensing, switching, protection of windings in
electric motors and transformers.
Semiconductor devices, consisting of either diodes or integrated circuit transistors, have
only been commonly used in industrial applications for a few years, but they were first
invented several decades ago. They have the advantage of being relatively
inexpensive,
but one difficulty that affects their use is the need to provide an external power supply
c
to the sensor.
Thermography, or thermal imaging, involves scanning an infrared radiation detector
across an object. The information gathered is then processed and an output in the form
of the temperature distribution across the object is produced. Temperature
measurement
over the range from _20°C up to C1500°C is possible. Elements of the system are
shown in Figure 14.14.
The radiation detector uses the same principles of operation as a radiation pyrometer
in inferring the temperature of the point that the instrument is focused on from
a measurement of the incoming infrared radiation. However, instead of providing a
c
measurement of the temperature of a single point at the focal point of the instrument,
the detector is scanned across a body or scene, and thus provides information about
temperature distributions. Because of the scanning mode of operation of the instrument,
radiation detectors with a very fast response are required, and only photoconductive
or photovoltaic sensors are suitable. These are sensitive to the portion of the infrared
spectrum between wavelengths of 2 ȝm and 14 ȝm.
Simpler versions of thermal imaging instruments consist of hand-held viewers that
are pointed at the object of interest. The output from an array of infrared detectors is
directed onto a matrix of red light-emitting diodes assembled behind a glass screen, and
the output display thus consists of different intensities of red on a black background,
with the different intensities corresponding to different temperatures. Measurement
resolution is high, with temperature differences as small as 0.1°C being detectable.
Such instruments are used in a wide variety of applications such as monitoring product
flows through pipework, detecting insulation faults, and detecting hot spots in furnace
linings, electrical transformers, machines, bearings etc. The number of applications is
extended still further if the instrument is carried in a helicopter, where uses include
scanning electrical transmission lines for faults, searching for lost or injured people
and detecting the source and spread pattern of forest fires.
More complex thermal imaging systems comprise a tripod-mounted detector
connected to a desktop computer and display system. Multi-colour displays are
commonly used in such systems, where up to 16 different colours represent different
bands of temperature across the measured range. The heat distribution across the
measured body or scene is thus displayed graphically as a contoured set of coloured
bands representing the different temperature levels. Such colour-thermography systems
find many applications such as inspecting electronic circuit boards and monitoring
production processes. There are also medical applications in body scanning.
c
The principle of acoustic thermometry was discovered as long ago as 1873 and uses
the fact that the velocity of sound through a gas varies with temperature according to
the equation:
vD
_
ÞRT/M _14.10_
where v is the sound velocity, T is the gas temperature, M is the molecular weight of
the gas and both R and Þ are constants. Until very recently, it had only been used for
measuring cryogenic (very low) temperatures, but it is now also used for measuring
higher temperatures and can potentially measure right up to 20 000°C. However, typical
inaccuracy is š5%, and the devices are expensive (typically £6000 or $10 000). The
various versions of acoustic thermometer that are available differ according to the
technique used for generating sound and measuring its velocity in the gas. If ultrasonic
Temperature-indicating devices known as Seger cones or pyrometric cones are
commonly used in the ceramics industry. They consist of a fused oxide and glass
material that is formed into a cone shape. The tip of the cone softens and bends over
when a particular temperature is reached. Cones are available that indicate
temperatures
over the range from 600°C to C2000°C.
Butter end
3
3emiconductor-based temperature sensors are used in an increasing proportion
of temperature measurement applications, mainly due to the need to
monitor integrated circuit temperatures in many electronics applications.
3ensors based on simple transistor circuits can be readily incorporated as part
of an integrated circuit to provide on-board diagnostic or control capability.
The typical temperature range for these devices is ±55°C to 150°C. The
temperature range for some smart integrated circuits can be extended to
185°C.
The majority of semiconductor junction sensors use a diode-connected
bipolar transistor. If the base of the transistor is shorted to the collector (see
Figure 6.22) then a constant current flowing in the remaining p±n junction (base
c
_r
0F =
_ ln _ F
_ (6.44)
3
The forward voltage drop across a p n junction increases with decreasing
temperature. For some semiconductors, the relationship between voltage and
temperature is almost linear; in silicon this occurs between 400 K and 25K
c
with a corresponding sensitivity of approximately 2.5 mV/K. The two most
commonly used semiconductors for thermometry are GaAs and 3i. The
typical voltage sensitivities for these devices are illustrated in Figure 6.24. For
silicon at temperatures below about 25 K, when the forward voltage
approaches 1.1V, the characteristic function for the voltage temperature
relationship changes. As can be seen in Figure 6.24, 3i diodes give a lower
output than GaAs diodes but have better stability and are cheaper and easily
interchangeable. Generally rectifying diodes are used and these can be potted
in a small container. Commercial versions are available, for example the
1.25 mm in diameter and 0.75 mm long DT-420 device, from Lake 3hore
Cryotronics. Zener diodes have also been used to indicate temperature
(3zmyrka-Grzebyk and Lipinski, 1995).
The virtues of diode thermometers are their low price, a simple voltage±
temperature relationship, a relatively large temperature range (1 to 400 K), no
need for a reference bath or junction, relatively high sensitivity, an uncertainty
lower than ±50 mK (Krause and Dodrill, 1986) and simplicity of operation
with a constant current source and a digital voltmeter. Measurements down to
1K do, however, require careful calibration. Errors can occur if the supply
current is not a true DC but has an AC component due to, say, noise induced
in the circuit from improper shielding, electrical grounds or ground loops. To
minimize noise effects, all the instrumentation should be electrically shielded
and proper grounding techniques used (Morrison, 1998). The diode current
supply should have a single ground, generally at the voltmeter, which then
requires a floating current source. A current between 1_A and 100_A can be
used but generally needs to exceed 10_A in order to overcome noise
problems. However, this can cause problems associated with self-heating at
very low temperatures.
w
c
Infrared thermometers can be classified in a number of groups based on the
principle of operation:
_ spectral band thermometers
_ total radiation thermometers
_ ratio thermometers
_ multiwaveband thermometers
_ special-purpose thermometers and methods
_ thermal imagers.
w
Thermography or thermal imaging involves determining the spatial distribution
of thermal energy emitted from the surface of an object. This information
can be manipulated to provide qualitative and quantitative data of the
distribution of temperature on a surface. In their usual form they comprise an
optical system, a detector, processing electronics and a display. A typical
handheld thermal imager is shown in Figure 9.27 and images resulting from
the use of such devices in Figures 9.28 and 9.29. Thermal imagers do not
require any form of illumination in order to operate and this makes them
highly attractive to military and surveillance users. However, these devices
tend to optimized to produce an image rather than quantitative information on
the distribution of temperatures.
It is possible to make use of a single detector with some form of scanner to
transmit the radiation signal from specific regions of the optical system to
enable a two-dimensional image of the temperature distribution to be built up,
c
c
(Figure 9.30). This principle can be extended by the use of a linear detector
array as illustrated in Figure 9.31. An alternative is to use an array of detectors
and this form is known as a staring array (Figure 9.32). It is usually necessary
to cool single and linear array detectors because of the level of incident
c
energy. This can be achieved in a number of ways ranging from introducing
liquid nitrogen into a dewar incorporated into the device to the use of a
miniature 3terling engine.
The optimum waveband for a thermal imager, as for most other infrared
thermometers, is dictated by the wavelength distribution of the emitted
radiation, the transmission characteristics of the atmospheric environment
between the imager and the target and by the characteristics of the available
detector technology. The optical windows for air shown in Figure 9.15
between 3 and 5_m, with a notch at 4.2_m due to CO2 absorption, and
between 7.5 and 14_m make these a common choice for the selection of
detectors. The band 3±5_m is commonly referred to as medium-wave
infrared (MWIR) and that between 7.5 and 14_m as long-wave infrared
(LWIR). The emissivity of most naturally occurring objects and organic paints
is high in the long-wave infrared but is lower and more variable in the
medium-wave infrared. Metallic surfaces tend to have lower emissivity in
either band. As such, the use of a thermal image to provide quantitative
information for the temperature distribution, particularly of a surface
c
3
Rayleigh spectra can be obtained using continuous wave and pulsed lasers to
c
excite the flow. The principal components for scattering based measurements
consist of an optical system and spectrometer to observe the gas sample and
a pulsed or continuous wave laser to excite the particles in the gas. A typical
system is illustrated in Figure 10.5. In Rayleigh scattering the collected signal
will typically be a factor of 109 smaller than the pump signal from the laser,
making it susceptible to corruption from other processes such as Mie
scattering, optical effects and background radiation. Also, in order to analyse
the spectra it is usually necessary to know the individual concentrations of the
species in the flow. As a result this technique must be undertaken very
carefully. The range for Rayleigh scattering is from approximately 293 K to
9000 K (e.g. see Farmer and Haddad, 1988). Applications have included
plasmas (Murphy and Farmer, 1992; Bentley, 1996), combustor flames (Barat
, 1991), sooting flames (Hoffman , 1996), and supersonic flows
(Miles and Lempert, 1990).