You are on page 1of 2

Accessed from 10.6.1.

1 by JNJconsumer on Thu May 11 14:38:36 EDT 2017

2602 Adenosine / Official Monographs USP 40

ASSAY Acceptance criteria


• PROCEDURE Any individual impurity: NMT 1.0%
Mobile phase: Dissolve 2.0 g of monobasic potassium Total impurities: NMT 1.5%
phosphate in 800 mL of water. Add 5 mL of 1.0 M tet-
rabutylammonium dihydrogen phosphate, dilute with SPECIFIC TESTS
water to 980 mL, and mix. Add 20 mL of acetonitrile. • PH 〈791〉: 4.5–7.5
System suitability solution: 0.03 mg/mL each of USP • PARTICULATE MATTER IN INJECTIONS 〈788〉: It meets the re-
Adenosine RS and inosine dissolved in warm water (50° quirements for small-volume injections.
to 55°), and diluted with water • BACTERIAL ENDOTOXINS TEST 〈85〉: When the product is
Standard solution: 0.03 mg/mL of USP Adenosine RS used for rapid intravenous injection, it contains NMT
dissolved in warm water (50° to 55°), and diluted with 11.62 USP Endotoxin Units/mg of adenosine. When the
water to volume. Before addition of the warm water, if product is used for continuous peripheral intravenous in-
sodium chloride is present in the Injection, add 0.01 mL fusion, it contains NMT 5.95 USP Endotoxin Units/mg of
of a solution of sodium chloride (0.9 in 100) per mL of adenosine.
the anticipated final volume of the Standard solution. • OTHER REQUIREMENTS: It meets the requirements under
Sample solution: Nominally 0.03 mg/mL of adenosine, Injections and Implanted Drug Products 〈1〉.
from a suitable volume of Injection in water
Chromatographic system ADDITIONAL REQUIREMENTS
(See Chromatography 〈621〉, System Suitability.) • PACKAGING AND STORAGE: Preserve in tight, single-dose
Mode: LC containers, preferably of Type I glass, and store at con-
Detector: UV 254 nm trolled room temperature.
Column: 3.9-mm × 30-cm; packing L1 • USP REFERENCE STANDARDS 〈11〉
Flow rate: 2.5 mL/min USP Adenosine RS
Injection volume: 10 µL USP Endotoxin RS
Run time: 2.5 times the retention time of adenosine
System suitability
Samples: System suitability solution and Standard
solution
.

[NOTE—The relative retention times of inosine and Medical Air


adenosine are 0.43 and 1.0, respectively.]
Suitability requirements DEFINITION
Resolution: NLT 6.0 between adenosine and inosine, Medical Air is a natural or synthetic mixture of gases consist-
System suitability solution ing largely of nitrogen and oxygen. It contains NLT
Tailing factor: NMT 2.0 for the adenosine peak, Sys- 19.5% and NMT 23.5%, by volume, of oxygen (O2).
tem suitability solution
Relative standard deviation: NMT 1.5%, Standard IDENTIFICATION
solution • A. The paramagnetic signal exhibited by the Sample gas
Analysis in the Assay confirms the presence of oxygen.
USP Monographs

Samples: Standard solution and Sample solution • B. The Sample gas in the Assay meets the assay Accep-
Calculate the percentage of the labeled amount of tance criteria.
adenosine (C10H13N5O4) in the portion of Injection ASSAY
taken: • PROCEDURE
Result = (rU/rS) × (CS/CU) × 100 The certified standards called for in the following test are
listed in Reagents, Indicators, and Solutions.
rU = peak response from the Sample solution Zero gas: Nitrogen certified standard
rS = peak response from the Standard solution Span gas: 21% Oxygen certified standard. [NOTE—See
CS = concentration of USP Adenosine RS in the Reagents, Indicators, and Solutions.]
Standard solution (mg/mL) Sample gas: Medical Air
CU = nominal concentration of adenosine in the Mode: Paramagnetic oxygen measurement (see Medical
Sample solution (mg/mL) Gases Assay 〈415〉)
Acceptance criteria: 90.0%–110.0% Analysis: Determine the concentration of oxygen in
percentage by volume of Medical Air using a suitable
IMPURITIES paramagnetic analyzer.
• ORGANIC IMPURITIES Acceptance criteria: 19.5%–23.5% of oxygen by
Mobile phase, System suitability solution, Standard volume
solution, Chromatographic system, and System suit-
ability: Proceed as directed in the Assay. IMPURITIES
Sample solution: Nominally 0.3 mg/mL of adenosine See Impurities Testing in Medical Gases Assay 〈413〉. The de-
from a volume of Injection, in water tector tubes called for in the following tests are listed in
Analysis Reagents, Indicators, and Solutions.
Sample: Sample solution If the label indicates that Medical Air is a synthetic mixture
Calculate the percentage of each impurity in the vol- of oxygen and nitrogen, and where oxygen complies to
ume of Injection taken: Oxygen USP and Nitrogen complies to Nitrogen NF, then
the Impurities tests are not required.
Result = (rU/rT) × 100 • LIMIT OF CARBON DIOXIDE
Sample: Detector tube manufacturer’s recommended
rU = peak response for each impurity volume ±5% of Medical Air
rT = sum of the responses of all of the peaks Analysis: Pass the Sample through a carbon dioxide de-
tector tube at the rate specified for the tube by the
detector tube manufacturer.
Acceptance criteria: NMT 500 ppm
• LIMIT OF CARBON MONOXIDE
Sample: Detector tube manufacturer’s recommended
volume ±5% of Medical Air

Official from May 1, 2017


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 10.6.1.1 by JNJconsumer on Thu May 11 14:38:36 EDT 2017

USP 40 Official Monographs / Alanine 2603

Analysis: Pass the Sample through a carbon monoxide VB = Titrant volume consumed by the Blank (mL)
detector tube at the rate specified for the tube by the N = actual normality of the Titrant (mEq/mL)
detector tube manufacturer. F = equivalency factor, 89.09 mg/mEq
Acceptance criteria: NMT 10 ppm W = Sample weight (mg)
• LIMIT OF SULFUR DIOXIDE Acceptance criteria: 98.5%–101.5% on the dried basis
Sample: Detector tube manufacturer’s recommended
volume ±5% of Medical Air IMPURITIES
Analysis: Pass the Sample through a sulfur dioxide de- • RESIDUE ON IGNITION 〈281〉: NMT 0.15%
tector tube at the rate specified for the tube by the • CHLORIDE AND SULFATE, Chloride 〈221〉
detector tube manufacturer. Standard solution: 0.70 mL of 0.020 N hydrochloric
Acceptance criteria: NMT 5 ppm acid
• LIMIT OF NITRIC OXIDE AND NITROGEN DIOXIDE Sample: 1.0 g of Alanine
Sample: Detector tube manufacturer’s recommended Acceptance criteria: NMT 0.05%
volume ±5% of Medical Air • CHLORIDE AND SULFATE, Sulfate 〈221〉
Analysis: Pass the Sample through a nitric oxide–ni- Standard solution: 0.30 mL of 0.020 N sulfuric acid
trogen dioxide detector tube at the rate specified for Sample: 1.0 g of Alanine
the tube by the detector tube manufacturer. Acceptance criteria: NMT 0.03%
Acceptance criteria: NMT 2.5 ppm • IRON 〈241〉: NMT 30 ppm
• LIMIT OF WATER AND OIL
Analysis: Support one container in an inverted position Delete the following:
(with the valve at the bottom) for 5 min. Cautiously
open the valve slightly, maintaining the container in an •• HEAVY METALS, Method I 〈231〉: NMT 15 ppm• (Official 1-
inverted position. Vent the gas with a barely audible
.

Jan-2018)
flow against a stainless steel mirror for a few seconds. • RELATED COMPOUNDS
Acceptance criteria: No liquid is discernible on the Mobile phase: 0.008 N sulfuric acid solution
mirror. System suitability solution: A mixture of 0.05 mg/mL
ADDITIONAL REQUIREMENTS of USP Fumaric Acid RS, 0.05 mg/mL of USP Maleic
• PACKAGING AND STORAGE: Preserve in pressurized contain- Acid RS, and 3 mg/mL of USP Malic Acid RS in water
ers. Container connections shall be appropriate for air. Maleic acid standard solution: 0.05 mg/mL of USP
Adaptors shall not be used to connect containers to pa- Maleic Acid RS in water
tient use supply system piping or equipment. Malic acid standard solution: 3 mg/mL of USP Malic
• LABELING: Label states if Medical Air is a synthetic mix- Acid RS in water
ture of Oxygen USP and Nitrogen NF. Where it is piped Fumaric acid standard solution: 0.05 mg/mL of USP
directly from the collecting tank to the patient point of Fumaric Acid RS in water
use, label each outlet “Medical Air”. Sample solution: 100 mg/mL of Alanine in water
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)

USP Monographs
Mode: LC
Detector: UV 214 nm
Column: 7.8-mm × 30-cm; 9-µm packing L17
.

Alanine Column temperature: 30°


Flow rate: 0.6 mL/min
Injection volume: 10 µL
System suitability
Sample: System suitability solution
[NOTE—See Table 1 for the relative retention times.]
Suitability requirements
C3H7NO2 89.09 Resolution: NLT 1.5 between maleic acid and malic
L-Alanine [56-41-7].
acid
DEFINITION Relative standard deviation: NMT 5.0% for each of
Alanine contains NLT 98.5% and NMT 101.5% of L-alanine fumaric acid, maleic acid, and malic acid
(C3H7NO2), calculated on the dried basis. Analysis
Samples: Standard solutions and Sample solution
IDENTIFICATION Calculate the percentage of each specified acid in the
• A. INFRARED ABSORPTION 〈197K〉 portion of Alanine taken:
ASSAY Result = (rU/rS) × (CS/CU) × 100
• PROCEDURE
Sample: 80 mg of Alanine rU = peak response of maleic acid, malic acid, or
Titrimetric system fumaric acid from the Sample solution
(See Titrimetry 〈541〉.) rS = peak response of maleic acid, malic acid, or
Mode: Direct titration fumaric acid from the corresponding
Titrant: 0.1 N perchloric acid VS Standard solution
Endpoint detection: Potentiometric CS = concentration of USP Maleic Acid RS, USP
Blank: 3 mL of formic acid in 50 mL of glacial acetic Malic Acid RS, or USP Fumaric Acid RS in the
acid corresponding Standard solution (mg/mL)
Analysis: Dissolve the Sample in a mixture of 3 mL of CU = concentration of Alanine in the Sample solution
formic acid and 50 mL of glacial acetic acid, and titrate (mg/mL)
with Titrant. Calculate the percentage of any unspecified impurity in
Calculate the percentage of L-alanine (C3H7NO2) in the the portion of Alanine taken:
portion of Alanine taken:
Result = (rU/rS) × (CS/CU) × 100
Result = [(VS − VB) × N × F × 100]/W
rU = peak response of any unspecified impurity
VS = Titrant volume consumed by the Sample (mL) from the Sample solution

Official from May 1, 2017


Copyright (c) 2017 The United States Pharmacopeial Convention. All rights reserved.

You might also like