2. Snap the MGC-S Calibration Cap (provided with the detector) into place
over the sensor ports.
(If the detector is not already activated (blank screen), activate it by pressing the
power button continuously, until detector switched on)
3. Connect them with the hose with inner diameter 4 or 6 mm.
5. Once the display shows SHOW STATUS press power button once again
6. When display shows L.BUMP Press power button once again
8. Once all of the sensors have been tested, the detector will go into
alarm. Turn off the regulator and remove the MGC-S Calibration Cap.
The detector will return to normal operation after a short period of
time.
10. To check that the Bump Test is recorded follow the steps 4 and 5
You must see the indication L.BUMP and the present date.
IMPORTANT:
The whole procedure from the step 7 to step 8 takes less than
10-15 seconds