AGFA Mortsel phone AGFA Munich phone AGFA Peissenberg phone
H. Blockx 8207 C. Stenzel 3022 T. Schelle 241
O. Hujer 3243 J. Geiger 254
M. Hölscher 4782
J. Hansemann 3169
V. Voigtländer 3772
M. Probst 3891
F. Frei 3114
E. Widemann 3622
Identification of Archival:
Version Label Place of Archives
ACP_5103 Development: //sun005/home/disg/projects/adc/release/acp_5103/
Distribution:
\\de.local\dfs\de\muc\R&D\06_Software\SW_Releases\5146_ADC_Compact_Plus\
ACP_5103
Table of Contents
1 REFERENCES ............................................................................................................................3
2 SCOPE........................................................................................................................................4
3 HARDWARE REQUIREMENTS ..................................................................................................4
4 RECOMMENDED WORKSTATION SOFTWARE ........................................................................4
5 INSTALLATION ON ARIEL / OBERON.......................................................................................5
5.1 Release Diskettes Set........................................................................................................................ 5
5.2 Generation of the Release Diskettes Set .......................................................................................... 5
5.3 Installation Procedure for the Application Software ...................................................................... 6
5.3.1 Installation on a Virgin Hard-Disk.............................................................................................. 6
5.3.2 Upgrade of version ACP_5007................................................................................................... 7
5.3.3 Upgrade of versions less or equal ACP_4007.............................................................................. 7
5.3.4 Installation of the CPF Configuration File................................................................................... 8
5.4 Installation of the Production Tools ................................................................................................ 8
6 INSTALLATION ON REVIVE.......................................................................................................9
6.1 Preparation of USB Stick ................................................................................................................ 9
6.2 Installation of Portex and the Application Software....................................................................... 9
6.2.1 Installation on a Virgin Hard-Disk.............................................................................................. 9
6.2.2 Upgrade from Previous Versions ................................................................................................ 9
6.2.3 Upgrade from Previous Versions, Using the LCD Terminal ...................................................... 10
6.2.4 Installation of the CPF Configuration File................................................................................. 10
6.3 Installation of the Production Tools .............................................................................................. 10
7 RELEASE INFO ........................................................................................................................11
7.1 What is new in ACP_5103 compared to ACP_5102?.................................................................... 11
7.1.1 Changes in Customer Functionality .......................................................................................... 11
7.1.2 Changes in Service Functionality.............................................................................................. 11
7.1.3 Changes in Production Tools .................................................................................................... 11
7.1.4 Resolved Problems................................................................................................................... 11
7.2 Non conformities to Specifications ................................................................................................ 11
7.2.1 Customer Functionality ............................................................................................................ 11
7.2.2 Service Functionality ............................................................................................................... 11
7.2.3 Production Tools...................................................................................................................... 11
7.2.4 Data flow specification............................................................................................................. 11
7.3 Known Bugs................................................................................................................................... 12
7.4 Restrictions .................................................................................................................................... 12
8 IMPORTANT REMARKS...........................................................................................................12
9 ABBREVIATIONS .....................................................................................................................12
10 TESTS AND RELATED DOCUMENTS......................................................................................12
1 References
Document Rev. Document Reference
Consolidated Requirements and 5 Livelink ID: 19316934
Testing_CR85-X
CR 85-X User Interface Requirements 2 Livelink ID: 20221482
CR85 User Interface Requirements 4 Livelink ID: 18981228
Service Level
Data Flow Specification 1 Livelink ID: 21212575
CR MD4.2 Extremity for CR25.0-CR75.0 1.0 I:\de\muc\amos\g-
archiv\FT\Projekte\CompRad\A_Extremities\Sp
and QS 3.0 ecs\CR MD4.2 Extremity for CR25.0-CR75.0
and QS 3.0.pdf
Software Requirement Specification - CR 1 Livelink ID: 23260391
MD4.2 Extremity
Release Qualification Plan ACP_51xx - 2 Livelink ID: 25915902
Version 1.0
Software Requirements ACP 41xx 2 Livelink ID: 20242082
DIOV Requirement List logHV formula 5 Livelink ID: 10784288
CDD CR85 and CR35 with USB 3 Livelink ID: 24430672
Software Requirements ACP_50xx 3 Livelink ID: 24457695
2 Scope
The version ACP_5103 supports the following device types. The behaviour of the device adapts
dependent on the type code:
3 Hardware Requirements
Compact PCI-Rack
200 MHz Ariel Board (A329646.x). or
Oberon Board (A341076.>3)
EEPROM Mon 4.26 (GECDCPU114A)
MSB, Master Scan-Board, Revision: 0
with:
– Scan IO-Board Boot EPLD, Version: 0
– Scan IO-Board serial PROM, Version: 0
Floppy (no SCSI)
SCSI Hard-Disk
Or
Revive-Rack
Revive Board (F8.4851.1510.1).
EIDE Hard-Disk (F7.0486.1428.0).
USB Stick (F7.0486.1429.0).
(Labelling rule for the application disks: ACPNNNN_X_Y were NNNN is the Release Number and X =
1 ... 3 defines the disk number and Y = 3 is the total number of application disks.)
Take 8 formatted diskettes (HDD, 1.44MB, DOS); non-empty disks will be erased.
Open a shell (cmd, command).
Change to the distribution directory ... \5146_ADC_Compact_Plus\acp_5103 mentioned on the
cover page.
Start "floppy.bat" (without parameters) for information about the "floppy" tool.
Start "floppy.bat c" and follow the program advices, sticking your labels (as presented above) to
the diskettes.
This is necessary on new devices in production or if you exchanged the hard disk. In this case the
Portex operating system must be installed before the application software can be installed.
When installing an ADC Compact Plus, replace CR75.0 with ADC Compact Plus where necessary.
Take the Harddisk formatter floppy of ACP_5103 and insert it into the disk drive of the Digitizer.
Make sure the Digitizer is disconnected from the network (since a default IP-Address & hostname
will be configured)
Connect a PC with terminal program or a serial terminal to service RS232 port (9600 baud,
Xon/Xoff, no parity, 8 bit, 1 stop bit).
Press the Reset on CPU-Board or turn Power Off/On of the Digitizer.
Follow the instructions given on the terminal or as follows:
Confirm partitioning/formatting of the hard disk on your terminal with ‘y’.
Wait until you get the final message on the terminal, or wait until control LED of floppy disk drive is
off.
Insert the floppy Portex Installer 2/2 ACP_5103 into the disk drive of the CR75.0 Digitizer.
Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
Wait until you get the final message on the terminal, or wait until control LED of floppy disk drive is
off.
Remove the floppy.
ATTENTION: do not insert any other floppy!
Press the Reset on CPU-Board or turn Power Off/On the CR75.0 Digitizer.
Wait until you get the final message on the terminal or wait approximately 2 minutes.
Insert the first floppy of the ACP_5103 Release Disks (labelled ACP5103_1_3) into the disk drive
of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.
Insert the second floppy of the ACP_5103 Release Disks (labelled ACP5103_2_3) into the disk
drive of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.
Insert the third floppy of the ACP_5103 Release Disks (labelled ACP5103_3_3) into the disk
drive of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On of the Digitizer.
Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.
Remove the floppy, press the Reset on CPU-Board or turn Power Off/On of the Digitizer. and wait
approx. 2 minutes (system extracts SW release)
Press the Reset on CPU-Board or turn Power Off/On of Digitizer and wait until system is up.
Note:
If you exchanged the hard disk, restore the original device specific parameters (using the Backup
Floppy) by following the guidelines of the Service Menu Topic Install from floppy, Machine specific
data.
Take care to use the right Backup Disk for the Digitizer in the backup restore procedure. (A
backup disk with the original Machine Specific Data is available for any machine from production
Insert the first floppy of the ACP_5103 Release Disks (labelled ACP5103_1_3) into the disk drive
of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.
Insert the second floppy of the ACP_5103 Release Disks (labelled ACP5103_2_3) into the disk
drive of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.
Insert the third floppy of the ACP_5103 Release Disks (labelled ACP5103_3_3) into the disk
drive of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On of the Digitizer.
Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.
Remove the floppy, press the Reset on CPU-Board or turn Power Off/On of the Digitizer. and wait
approx. 2 minutes (system extracts SW release)
Press the Reset on CPU-Board or turn Power Off/On of Digitizer and wait until system is up.
Insert the floppy Portex Installer 1/2 into the disk drive of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On.
Wait until you get the final message on a Terminal connected to the service RS232 port, or wait
until control LED of floppy disk drive is off.
Insert the floppy Portex Installer 2/2 into the disk drive of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On.
Wait until you get the final message on a Terminal connected to the service RS232 port, or wait
until control LED of floppy disk drive is off.
Remove the floppy.
ATTENTION: do not insert any other floppy!
Press the Reset on CPU-Board or turn Power Off/On the Digitizer.
Wait until you get the final message on the terminal.
Insert the first floppy of the ACP_5103 Release Disks (labelled ACP5103_1_3) into the disk drive
of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.
Insert the second floppy of the ACP_5103 Release Disks (labelled ACP5103_2_3) into the disk
drive of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.
Insert the third floppy of the ACP_5103 Release Disks (labelled ACP5103_3_3) into the disk
drive of the Digitizer.
Press the Reset on CPU-Board or turn Power Off/On of the Digitizer.
Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.
Remove the floppy, press the Reset on CPU-Board or turn Power Off/On of the Digitizer. and wait
approx. 2 minutes (system extracts SW release)
Press the Reset on CPU-Board or turn Power Off/On of Digitizer and wait until system is up.
If the previous software was less than ACP_1201, then close the machine doors and scan or
erase one cassette (for calibration of PM monitoring, see MUCft01014). This cassette needs not to
be exposed.
Go to the service menu on the LCD screen, select Install data and Languages and follow the
instructions to install the language files. This is important, because old language files are no
longer valid. Switch off and on the digitizer to activate the new language files.
Refresh your backup.
Make sure the Digitizer is disconnected from the network (never connect the Digitizer to the
network before configuring it with a valid CPF-file!)
Generate the CPF-file with the QS, the VIPS or the CCM
Parse the CPF-File using the guidelines of the Service Menu Topic Install from floppy, CPF-File
follow the instructions displayed on the user interface.
Reboot the system.
Reconnect Digitizer to the network
If the previous software was less than ACP_1201, then close the machine doors and scan or
erase one cassette (for calibration of PM monitoring, see MUCft01014). This cassette needs not to
be exposed.
Refresh your backup by following the guidelines of Service Menu Topic Save on floppy, Machine
specific data
The production tools are installed for production purposes only and removed again before the device
is delivered to the customer. The production tools will never be installed and must not be used in
the field!
Make sure to use the correct Release Disk for the desired release in the installation procedure.
Insert the floppy Production Tools into the disk drive of the Digitizer.
Connect the Service PC to the service plug (serial) and start a terminal program (9600 baud,
Xon/Xoff, no parity, 8 bit, 1 stop bit).
Type in: A:INSTALL.
Wait until the final message appears and remove the disk. No reboot is necessary to use the
production tools.
6 Installation on Revive
Plug the USB stick into a PC and copy the complete software release directory structure to there. This
can also be done by unzipping the release zip file to the USB stick.
Verify the directory structure on the USB stick. It should look as follows. The USB stick may contain
software versions and a back-up. In case it contains a back-up, the USB stick is additionally named
<TypeNumber><SerialNumber>BAC (this is necessary for restore). Older backup folder names have
the extension BCO.
Disconnect the digitizer from the network. This is necessary since a default IP-Address and
hostname is configured.
Connect the PC to the RS 232 serial port of Revive and start a terminal program with baud rate
115200.
Using the Explorer on a PC, go to the folder acp_5103 on the USB stick and double-click on the
file Prepare_Formatting.cmd. (This has to be done every time, the software is being installed,
because the created startup.com is deleted after execution)
Plug the USB stick into Revive and switch off and on the digitizer.
When asked about formatting the hard disk on the terminal, answer with yes.
Wait until the digitizer has started.
Install the backup from the USB stick, using the service menu on the digitizer LCD terminal.
Switch off and on the digitizer.
Connect the digitizer to the network.
Using the Explorer on a PC, go to the folder acp_5103 on the USB stick and double-click on the
file Prepare_Installation.cmd. (This has to be done every time, the software is being installed,
because the created startup.com is deleted after execution)
Plug the USB stick into Revive and switch off and on the digitizer.
Wait until the digitizer has started.
The production tools are installed for production purposes only and removed again before the device
is delivered to the customer. The production tools will never be installed and must not be used in
the field!
Connect the PC to the RS 232 serial port of Revive and start a terminal program with baud rate
115200.
Plug the USB stick into Revive.
On the terminal, call a:/acp_5103/InstProd.com
7 Release Info
None.
None.
See MPM00003644.
None.
None.
None.
None.
MPM00001728 SALh values of CR85/35 tend to be around 1750 SAL (too low).
MPM00001956 Saugerarm Eingaberoboter streift an Stange Kassettenteil.
MPM00002035 TagReader HALT recovery may omit retry.
(The functionality itself is not affected.)
MPM00002090 Info counters deleted in case tag reader has no serial number.
(This error may only occur at production site.)
MPM00002279 Slight deviation from specified user interface behaviour at shading calibration.
No influence on functionality.
MPM00002309 The correct function of the fan of the Compact PCI or Revive rack cannot be checked
directly. This should be done indirectly by checking the CPU temperature.
MPM00002594 (HQ_0902060001) Solve error 20507 (laser power too low).
MPM00002597 (GSC081030149) Locked Condition 20507 every 24 hours.
Duplicate to MPM00002594.
MPM00002626 New PORTEX release for correct USB frequency using already onboard 48MHz
oscillator.
MPM00003155 Triangles in Mammo images.
7.4 Restrictions
8 Important Remarks
A new Portex Version (2006.04 for Oberon / 2008.04 for Revive) accompanies the releases acp_5007
and acp_5103. In case of an upgrade from a version < acp_5007, an update of Portex is mandatory.
In case of upgrade from a version < acp_1007 to acp_5103, a new BOL/BOS Calibration is necessary
because of changed IP stop position before start of scan.
In field test devices with Software Releases ACP_20xx (Mammo Fieldtest), mammo_factor was
manually edited to a different value. This value "survives" a software update. In order to get the correct
value also at these devices, mammo_factor has to be edited manually to the default 1.19. Then the
entry in scnxxx00.u00 disappears and the device has the same status as a new series machine.
9 Abbreviations
IP Image Plate
PMT Photo Multiplier
QS Quality System (image processing station for digitizers)
Signatures: