You are on page 1of 14

Document Status: Approved Livelink ID: 29819300 Version: 2

Department: HE-IM-OE-RDE-A&D/SW1 Tel: ++49 89 6207-3243


Fax: ++49 89 6207-7371
Health Care Name SE: W. Haug

Distribution List:
AGFA Mortsel phone AGFA Munich phone AGFA Peissenberg phone
H. Blockx 8207 C. Stenzel 3022 T. Schelle 241
O. Hujer 3243 J. Geiger 254
M. Hölscher 4782
J. Hansemann 3169
V. Voigtländer 3772
M. Probst 3891
F. Frei 3114
E. Widemann 3622

Software Release Note


for ADC Compact Plus / CR75.0 / CR85-X

Version Date of Build AG-Sachnummer


ACP_5103 22-Jan-2010 F8.5146.0828.0

To be signed in Livelink by:


Dedicated for
Preference Test Init. Market Test  Cust.-Release
PL PL PL
W. Haug
SIV SIV SIV
M. Probst
Verif. Verif. Verif.
J. Hansemann
Abbreviation: Service Service
PE = Project Leader Development Ch. Stenzel
SIV = System Integration and Verification PM PMM
Verif = Leader Software Verification
PMM = Product Marketing Manager Ch. Uytterhoeven
Service = Global Support Organisation VEM
VEM = Manager Value Engineering / Maintenance
F. Frei
Identification of Deliveries:
Progr.-Name_Version Sachnummer Type (Number of Disk(s))
CR75.0/CR85.0 Firmware / ACP_5103 F8.5146. 0977.0 Harddisk Formatter (1 Disk)
CR75.0/CR85.0 Firmware / ACP_5103 F8.5146. 0978.0 Portex Installer (2 Disks)
CR75.0/CR85.0 Firmware / ACP_5103 F8.5146. 0979.0 Release Disks (3 Disks)
CR75.0/CR85.0 Firmware / ACP_5103 F8.5146. 0980.0 Production Tools (1 Disk)
CR75.0/CR85.0 Firmware / ACP_5103 F8.5146. 0981.0 Languages (1 Disk)

Identification of Archival:
Version Label Place of Archives
ACP_5103 Development: //sun005/home/disg/projects/adc/release/acp_5103/
Distribution:
\\de.local\dfs\de\muc\R&D\06_Software\SW_Releases\5146_ADC_Compact_Plus\
ACP_5103

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 1
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2

Table of Contents
1 REFERENCES ............................................................................................................................3
2 SCOPE........................................................................................................................................4
3 HARDWARE REQUIREMENTS ..................................................................................................4
4 RECOMMENDED WORKSTATION SOFTWARE ........................................................................4
5 INSTALLATION ON ARIEL / OBERON.......................................................................................5
5.1 Release Diskettes Set........................................................................................................................ 5
5.2 Generation of the Release Diskettes Set .......................................................................................... 5
5.3 Installation Procedure for the Application Software ...................................................................... 6
5.3.1 Installation on a Virgin Hard-Disk.............................................................................................. 6
5.3.2 Upgrade of version ACP_5007................................................................................................... 7
5.3.3 Upgrade of versions less or equal ACP_4007.............................................................................. 7
5.3.4 Installation of the CPF Configuration File................................................................................... 8
5.4 Installation of the Production Tools ................................................................................................ 8
6 INSTALLATION ON REVIVE.......................................................................................................9
6.1 Preparation of USB Stick ................................................................................................................ 9
6.2 Installation of Portex and the Application Software....................................................................... 9
6.2.1 Installation on a Virgin Hard-Disk.............................................................................................. 9
6.2.2 Upgrade from Previous Versions ................................................................................................ 9
6.2.3 Upgrade from Previous Versions, Using the LCD Terminal ...................................................... 10
6.2.4 Installation of the CPF Configuration File................................................................................. 10
6.3 Installation of the Production Tools .............................................................................................. 10
7 RELEASE INFO ........................................................................................................................11
7.1 What is new in ACP_5103 compared to ACP_5102?.................................................................... 11
7.1.1 Changes in Customer Functionality .......................................................................................... 11
7.1.2 Changes in Service Functionality.............................................................................................. 11
7.1.3 Changes in Production Tools .................................................................................................... 11
7.1.4 Resolved Problems................................................................................................................... 11
7.2 Non conformities to Specifications ................................................................................................ 11
7.2.1 Customer Functionality ............................................................................................................ 11
7.2.2 Service Functionality ............................................................................................................... 11
7.2.3 Production Tools...................................................................................................................... 11
7.2.4 Data flow specification............................................................................................................. 11
7.3 Known Bugs................................................................................................................................... 12
7.4 Restrictions .................................................................................................................................... 12
8 IMPORTANT REMARKS...........................................................................................................12
9 ABBREVIATIONS .....................................................................................................................12
10 TESTS AND RELATED DOCUMENTS......................................................................................12

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 2
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2

1 References
Document Rev. Document Reference
Consolidated Requirements and 5 Livelink ID: 19316934
Testing_CR85-X
CR 85-X User Interface Requirements 2 Livelink ID: 20221482
CR85 User Interface Requirements 4 Livelink ID: 18981228
Service Level
Data Flow Specification 1 Livelink ID: 21212575
CR MD4.2 Extremity for CR25.0-CR75.0 1.0 I:\de\muc\amos\g-
archiv\FT\Projekte\CompRad\A_Extremities\Sp
and QS 3.0 ecs\CR MD4.2 Extremity for CR25.0-CR75.0
and QS 3.0.pdf
Software Requirement Specification - CR 1 Livelink ID: 23260391
MD4.2 Extremity
Release Qualification Plan ACP_51xx - 2 Livelink ID: 25915902
Version 1.0
Software Requirements ACP 41xx 2 Livelink ID: 20242082
DIOV Requirement List logHV formula 5 Livelink ID: 10784288
CDD CR85 and CR35 with USB 3 Livelink ID: 24430672
Software Requirements ACP_50xx 3 Livelink ID: 24457695

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 3
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2

2 Scope
The version ACP_5103 supports the following device types. The behaviour of the device adapts
dependent on the type code:

 Typ 5146/100/200, Compact Plus,K2025JA CR MP 3510 READER


 Type 5146/101/201, CR75.0, K2025JA CR MP3510 READER W/50µ
 5148/100, CR85.0

3 Hardware Requirements
 Compact PCI-Rack
 200 MHz Ariel Board (A329646.x). or
Oberon Board (A341076.>3)
 EEPROM Mon 4.26 (GECDCPU114A)
 MSB, Master Scan-Board, Revision: 0
with:
– Scan IO-Board Boot EPLD, Version: 0
– Scan IO-Board serial PROM, Version: 0
 Floppy (no SCSI)
 SCSI Hard-Disk
Or
 Revive-Rack
 Revive Board (F8.4851.1510.1).
 EIDE Hard-Disk (F7.0486.1428.0).
 USB Stick (F7.0486.1429.0).

 Stepper motor boards GS IOB-SINSTEP (F8.9499.8100.0 … F8.9499.8100.6) and


GS IOB-5STEP(F8.9499.8120.2 … F8.9499.8130.0).
Firmware(Download) for IOB-SINSTEP: SM5_1113.hex
Firmware(Download) for IOB-5STEP: SM5_1113.hex

4 Recommended Workstation software

At least version or higher Application supported


QS 3.5.372 SU1 GenRad
QS 3.5.372M Mammo
NX 1.0.3203 GenRad

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 4
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2

5 Installation on Ariel / Oberon

5.1 Release Diskettes Set


The complete Release Disk Set for the ADC Compact Plus / CR75.0 / CR 85-X Firmware of Release acp_5103
consists of 8 floppy disks.

Arcticle Code / "Sachnummer" Name Disk number/total Volume label


(Number of Disks)
F8.5146.0977.0 "Harddisk Formatter / Portex 1" 1/8 --
(1 Diskette)
F8.5146.0978.0 "Portex" 2/8 --
(2 Diskettes) 3/8 --
F8.5146.0979.0 "CR75/85 Application" 4/8 ACP5103_1_3
(3 Diskettes) 5/8 ACP5103_2_3
6/8 ACP5103_3_3
F8.5146.0981.0 "CR75/85 Language Disk" 7/8 --
(1 Diskette) (Holds the language files.)
F8.5146.09780.0 "CR75/85 Production Tools" 8/8 --
(1 Diskette) (This disk is not distributed to field.)

(Labelling rule for the application disks: ACPNNNN_X_Y were NNNN is the Release Number and X =
1 ... 3 defines the disk number and Y = 3 is the total number of application disks.)

5.2 Generation of the Release Diskettes Set

You can create a complete Floppy set (8 Diskettes) as follows:

 Take 8 formatted diskettes (HDD, 1.44MB, DOS); non-empty disks will be erased.
 Open a shell (cmd, command).
 Change to the distribution directory ... \5146_ADC_Compact_Plus\acp_5103 mentioned on the
cover page.
 Start "floppy.bat" (without parameters) for information about the "floppy" tool.
 Start "floppy.bat c" and follow the program advices, sticking your labels (as presented above) to
the diskettes.

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 5
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2

5.3 Installation Procedure for the Application Software

5.3.1 Installation on a Virgin Hard-Disk

This is necessary on new devices in production or if you exchanged the hard disk. In this case the
Portex operating system must be installed before the application software can be installed.
When installing an ADC Compact Plus, replace CR75.0 with ADC Compact Plus where necessary.

 Take the Harddisk formatter floppy of ACP_5103 and insert it into the disk drive of the Digitizer.
 Make sure the Digitizer is disconnected from the network (since a default IP-Address & hostname
will be configured)
 Connect a PC with terminal program or a serial terminal to service RS232 port (9600 baud,
Xon/Xoff, no parity, 8 bit, 1 stop bit).
 Press the Reset on CPU-Board or turn Power Off/On of the Digitizer.
 Follow the instructions given on the terminal or as follows:
 Confirm partitioning/formatting of the hard disk on your terminal with ‘y’.
 Wait until you get the final message on the terminal, or wait until control LED of floppy disk drive is
off.
 Insert the floppy Portex Installer 2/2 ACP_5103 into the disk drive of the CR75.0 Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
 Wait until you get the final message on the terminal, or wait until control LED of floppy disk drive is
off.
 Remove the floppy.
ATTENTION: do not insert any other floppy!
 Press the Reset on CPU-Board or turn Power Off/On the CR75.0 Digitizer.
 Wait until you get the final message on the terminal or wait approximately 2 minutes.

 Insert the first floppy of the ACP_5103 Release Disks (labelled ACP5103_1_3) into the disk drive
of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
 Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.

 Insert the second floppy of the ACP_5103 Release Disks (labelled ACP5103_2_3) into the disk
drive of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
 Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.

 Insert the third floppy of the ACP_5103 Release Disks (labelled ACP5103_3_3) into the disk
drive of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On of the Digitizer.
 Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.

 Remove the floppy, press the Reset on CPU-Board or turn Power Off/On of the Digitizer. and wait
approx. 2 minutes (system extracts SW release)
 Press the Reset on CPU-Board or turn Power Off/On of Digitizer and wait until system is up.

Note:
 If you exchanged the hard disk, restore the original device specific parameters (using the Backup
Floppy) by following the guidelines of the Service Menu Topic Install from floppy, Machine specific
data.
Take care to use the right Backup Disk for the Digitizer in the backup restore procedure. (A
backup disk with the original Machine Specific Data is available for any machine from production

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 6
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2
in case of loss or fault).
CAUTION: Without restoring the device specific parameters, no operation is possible!
 If the previous software was less than ACP_1201, then close the machine doors and scan or
erase one cassette (for calibration of PM monitoring, see MUCft01014). This cassette needs not to
be exposed.

5.3.2 Upgrade of version ACP_5007

 Insert the first floppy of the ACP_5103 Release Disks (labelled ACP5103_1_3) into the disk drive
of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
 Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.

 Insert the second floppy of the ACP_5103 Release Disks (labelled ACP5103_2_3) into the disk
drive of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
 Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.

 Insert the third floppy of the ACP_5103 Release Disks (labelled ACP5103_3_3) into the disk
drive of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On of the Digitizer.
 Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.

 Remove the floppy, press the Reset on CPU-Board or turn Power Off/On of the Digitizer. and wait
approx. 2 minutes (system extracts SW release)
 Press the Reset on CPU-Board or turn Power Off/On of Digitizer and wait until system is up.

5.3.3 Upgrade of versions less or equal ACP_4007

 Insert the floppy Portex Installer 1/2 into the disk drive of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On.
 Wait until you get the final message on a Terminal connected to the service RS232 port, or wait
until control LED of floppy disk drive is off.

 Insert the floppy Portex Installer 2/2 into the disk drive of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On.
 Wait until you get the final message on a Terminal connected to the service RS232 port, or wait
until control LED of floppy disk drive is off.
 Remove the floppy.
ATTENTION: do not insert any other floppy!
 Press the Reset on CPU-Board or turn Power Off/On the Digitizer.
 Wait until you get the final message on the terminal.

 Insert the first floppy of the ACP_5103 Release Disks (labelled ACP5103_1_3) into the disk drive
of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
 Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 7
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2

 Insert the second floppy of the ACP_5103 Release Disks (labelled ACP5103_2_3) into the disk
drive of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On of the CR75.0 Digitizer.
 Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.

 Insert the third floppy of the ACP_5103 Release Disks (labelled ACP5103_3_3) into the disk
drive of the Digitizer.
 Press the Reset on CPU-Board or turn Power Off/On of the Digitizer.
 Wait until you get the final message on the terminal or wait until control LED of floppy disk drive is
off.

 Remove the floppy, press the Reset on CPU-Board or turn Power Off/On of the Digitizer. and wait
approx. 2 minutes (system extracts SW release)
 Press the Reset on CPU-Board or turn Power Off/On of Digitizer and wait until system is up.
 If the previous software was less than ACP_1201, then close the machine doors and scan or
erase one cassette (for calibration of PM monitoring, see MUCft01014). This cassette needs not to
be exposed.
 Go to the service menu on the LCD screen, select Install data and Languages and follow the
instructions to install the language files. This is important, because old language files are no
longer valid. Switch off and on the digitizer to activate the new language files.
 Refresh your backup.

5.3.4 Installation of the CPF Configuration File

 Make sure the Digitizer is disconnected from the network (never connect the Digitizer to the
network before configuring it with a valid CPF-file!)
 Generate the CPF-file with the QS, the VIPS or the CCM
 Parse the CPF-File using the guidelines of the Service Menu Topic Install from floppy, CPF-File
follow the instructions displayed on the user interface.
 Reboot the system.
 Reconnect Digitizer to the network
 If the previous software was less than ACP_1201, then close the machine doors and scan or
erase one cassette (for calibration of PM monitoring, see MUCft01014). This cassette needs not to
be exposed.
 Refresh your backup by following the guidelines of Service Menu Topic Save on floppy, Machine
specific data

5.4 Installation of the Production Tools

The production tools are installed for production purposes only and removed again before the device
is delivered to the customer. The production tools will never be installed and must not be used in
the field!

 Make sure to use the correct Release Disk for the desired release in the installation procedure.
 Insert the floppy Production Tools into the disk drive of the Digitizer.
 Connect the Service PC to the service plug (serial) and start a terminal program (9600 baud,
Xon/Xoff, no parity, 8 bit, 1 stop bit).
 Type in: A:INSTALL.
 Wait until the final message appears and remove the disk. No reboot is necessary to use the
production tools.

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 8
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2

6 Installation on Revive

6.1 Preparation of USB Stick

Plug the USB stick into a PC and copy the complete software release directory structure to there. This
can also be done by unzipping the release zip file to the USB stick.
Verify the directory structure on the USB stick. It should look as follows. The USB stick may contain
software versions and a back-up. In case it contains a back-up, the USB stick is additionally named
<TypeNumber><SerialNumber>BAC (this is necessary for restore). Older backup folder names have
the extension BCO.

(Root) ─┬─ acp_5103 ─┬─ portex_installer_1


│ ├─ portex_installer_2
│ ├─ harddisk_formatter
│ ├─ disk1
│ ├─ disk2
│ ├─ disk3
│ ├─ production_tools (optional for production)
│ └─ languages
└─ ACP_<SerialNumber>.BAC

6.2 Installation of Portex and the Application Software

6.2.1 Installation on a Virgin Hard-Disk

 Disconnect the digitizer from the network. This is necessary since a default IP-Address and
hostname is configured.
 Connect the PC to the RS 232 serial port of Revive and start a terminal program with baud rate
115200.
 Using the Explorer on a PC, go to the folder acp_5103 on the USB stick and double-click on the
file Prepare_Formatting.cmd. (This has to be done every time, the software is being installed,
because the created startup.com is deleted after execution)
 Plug the USB stick into Revive and switch off and on the digitizer.
 When asked about formatting the hard disk on the terminal, answer with yes.
 Wait until the digitizer has started.
 Install the backup from the USB stick, using the service menu on the digitizer LCD terminal.
 Switch off and on the digitizer.
 Connect the digitizer to the network.

6.2.2 Upgrade from Previous Versions

 Using the Explorer on a PC, go to the folder acp_5103 on the USB stick and double-click on the
file Prepare_Installation.cmd. (This has to be done every time, the software is being installed,
because the created startup.com is deleted after execution)
 Plug the USB stick into Revive and switch off and on the digitizer.
 Wait until the digitizer has started.

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 9
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2
6.2.3 Upgrade from Previous Versions, Using the LCD Terminal

This does not install Portex.


Make sure, that the USB stick contains only one software version.
At the digitizer LCD terminal, select Service  Install data  Software and follow the instructions.

6.2.4 Installation of the CPF Configuration File

See "Installation on Ariel / Oberon".

6.3 Installation of the Production Tools

The production tools are installed for production purposes only and removed again before the device
is delivered to the customer. The production tools will never be installed and must not be used in
the field!

 Connect the PC to the RS 232 serial port of Revive and start a terminal program with baud rate
115200.
 Plug the USB stick into Revive.
 On the terminal, call a:/acp_5103/InstProd.com

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 10
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2

7 Release Info

7.1 What is new in ACP_5103 compared to ACP_5102?

7.1.1 Changes in Customer Functionality

None.

7.1.2 Changes in Service Functionality

None.

7.1.3 Changes in Production Tools

See MPM00003644.

7.1.4 Resolved Problems

MPM00003644 Software change for Hamamatsu photomultiplier


The modification avoids problems during sensitivity adjustment at production time.
No change in customer environment.

7.2 Non conformities to Specifications

7.2.1 Customer Functionality

None.

7.2.2 Service Functionality

None.

7.2.3 Production Tools

None.

7.2.4 Data flow specification

None.

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 11
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2

7.3 Known Bugs

MPM00001728 SALh values of CR85/35 tend to be around 1750 SAL (too low).
MPM00001956 Saugerarm Eingaberoboter streift an Stange Kassettenteil.
MPM00002035 TagReader HALT recovery may omit retry.
(The functionality itself is not affected.)
MPM00002090 Info counters deleted in case tag reader has no serial number.
(This error may only occur at production site.)
MPM00002279 Slight deviation from specified user interface behaviour at shading calibration.
No influence on functionality.
MPM00002309 The correct function of the fan of the Compact PCI or Revive rack cannot be checked
directly. This should be done indirectly by checking the CPU temperature.
MPM00002594 (HQ_0902060001) Solve error 20507 (laser power too low).
MPM00002597 (GSC081030149) Locked Condition 20507 every 24 hours.
Duplicate to MPM00002594.
MPM00002626 New PORTEX release for correct USB frequency using already onboard 48MHz
oscillator.
MPM00003155 Triangles in Mammo images.

7.4 Restrictions

8 Important Remarks
A new Portex Version (2006.04 for Oberon / 2008.04 for Revive) accompanies the releases acp_5007
and acp_5103. In case of an upgrade from a version < acp_5007, an update of Portex is mandatory.

In case of upgrade from a version < acp_1007 to acp_5103, a new BOL/BOS Calibration is necessary
because of changed IP stop position before start of scan.

In field test devices with Software Releases ACP_20xx (Mammo Fieldtest), mammo_factor was
manually edited to a different value. This value "survives" a software update. In order to get the correct
value also at these devices, mammo_factor has to be edited manually to the default 1.19. Then the
entry in scnxxx00.u00 disappears and the device has the same status as a new series machine.

9 Abbreviations
IP Image Plate
PMT Photo Multiplier
QS Quality System (image processing station for digitizers)

10 Tests and Related Documents


See release qualification plan as mentioned in chapter “References” and affiliated release qualification
report in LiveLink.

DATUM NAME LL ID: 29819300 TYP-Nr.


BEARB. 12.05.10 Haug Software Release Note 5146/100/200
GEPR. Comp.+ / CR75.0 / CR 85-X 5146/101/ 201
NORM ACP_5103 5148/100

41101 AGFA-GEVAERT BLATT


FOTOTECHNIK F8.5146.0828.0 12
Ä-Z Ä-Nr. DAT. NAME MÜNCHEN ERSATZ FÜR: AUS 12
Document Status: Approved Livelink ID: 29819300 Version: 2

Details as of PDF Creation Date


Document Metadata
Title: Release Note ACP_5103
Livelink ID: 29819300
Version#: 2
Version Date: 2010/05/12 03:06 PM CET
Status: Approved on 2010/05/17 09:41 AM CET
Owner: Werner Haug (aghwe)
Created By: Werner Haug (aghwe)
Created Date: 2010/04/09 10:50 AM CET
PDF Creation Date: 2010/05/17 10:43 AM CET

This document was approved by:

Signatures:

1. Manfred Probst (agobs) on 2010/05/12 04:13 PM CET


2. Christian Stenzel (agszn) on 2010/05/14 08:49 AM CET
3. Christine Uytterhoeven (amegt) on 2010/05/17 09:36 AM CET
4. Jörg Hansemann (agjhe) on 2010/05/17 08:52 AM CET
5. Werner Haug (aghwe) on 2010/05/12 03:10 PM CET
6. Florian Frei (agmou) on 2010/05/14 09:13 AM CET

Detailed Approver History:

 Approval Workflow started on 2010/05/12 03:09 PM CET


 Approval task originally assigned to and completed by Florian Frei (agmou)
on 2010/05/14 09:13 AM CET
 Approval task originally assigned to and completed by Christine
Uytterhoeven (amegt) on 2010/05/17 09:36 AM CET
 Approval task originally assigned to and completed by Christian Stenzel
(agszn) on 2010/05/14 08:49 AM CET
 Approval task originally assigned to and completed by Jörg Hansemann
(agjhe) on 2010/05/17 08:52 AM CET
 Approval task originally assigned to and completed by Manfred Probst
(agobs) on 2010/05/12 04:13 PM CET
 Approval task originally assigned to and completed by Werner Haug (aghwe)
on 2010/05/12 03:10 PM CET
Document Status: Approved Livelink ID: 29819300 Version: 2

Version & Status History


Version# Date Created Status
2 2010/05/12 03:06 PM CET Approved - 2010/05/17
1 2010/04/09 10:50 AM CET

You might also like