Professional Documents
Culture Documents
Service Guide
Rapidlab 1200 Series Service Guide
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its
affiliates to provide such services. Assemblers and other persons who are not employed by or otherwise
directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
SMN 10488397
REF 02101783
Rev. B
This Service Guide, and the software described in it, are protected by copyright laws. No part of this
Service Guide or the described software may be copied, reproduced, translated or reduced to any
electronic medium or machine-readable form without the prior written consent of Siemens Healthcare
Diagnostics except that you may make one copy of the program CD solely for back-up purposes.
Rapidlab, RapidQC, AutomaticQC, Rapidlink, AQC, Complete, and Rapidsystems are registered
trademarks of Siemens Healthcare Diagnostics. All other products mentioned are trademarks of their
respective owners.
Electronics ..............................................................................................................................29
Replacing the Drive Bay Assembly....................................................................................29
Replacing the Polychromator Tray.....................................................................................30
Replacing the Power Supply Tray......................................................................................30
Replacing the Fan..............................................................................................................30
Replacing the Rear Interface Board...................................................................................31
Replacing the Main Board Assembly .................................................................................31
Front Wall ...............................................................................................................................31
Replacing the Cartridge Interface Assembly......................................................................31
Replacing the Pump Motors/Roller Cages.........................................................................32
Replacing the Wash RFID Board.......................................................................................32
Replacing the Reagent RFID Board ..................................................................................33
Replacing the Valve Actuator.............................................................................................33
Replacing the Valve Drive Motor .......................................................................................40
Replacing the Waste Sensor Assembly.............................................................................40
Measurement Module.............................................................................................................40
Replacing the Measurement Module .................................................................................40
Replacing the Measurement Module Door ........................................................................42
Replacing the Contact Assembly .......................................................................................42
Replacing the Pressure Sensor .........................................................................................43
CO-ox Module ........................................................................................................................43
Replacing the CO-ox Module.............................................................................................43
Replacing the CO-ox Pump Motor .....................................................................................43
Replacing FD3 ...................................................................................................................44
Display and Printer .................................................................................................................44
Removing the Display Assembly .......................................................................................44
Electronic Zero Procedure......................................................................................................44
Parts Required ...................................................................................................................45
Procedure ..........................................................................................................................45
CAUTION: Please observe the warning and hazard statements appearing throughout
the Service Guide. If the instrument is used in a manner not specified by Siemens, the
protection provided by the equipment may be impaired.
Warranty Period
The limited warranty period generally commences upon installation of the original instrument
at the customer’s location and extends for a period of one year thereafter, unless otherwise
specifically agreed to by and between Siemens (or its authorized distributors) and customer
in a writing signed by duly authorized representatives of both parties (sales representatives
are generally not authorized representatives of Siemens for these purposes).
Service calls
The customer may obtain service for instruments during normal business hours by
contacting the nearest Siemens technical support provider or authorized distributor.
Customers, with some exceptions, may also request service to be delivered or an exchange
to be initiated outside normal business hours, including evenings, weekend days, or
nationally observed holidays, by contacting the nearest Siemens location or authorized
distributor. Service performed outside normal hours is subject to a surcharge unless the
customer has in place a service product option that provides service at the time/s requested.
Replacement of parts
In performing service, Siemens or its authorized distributors will provide appropriate parts to
repair the instrument, or will arrange for the exchange of the instrument or affected parts, at
no charge with the exception of certain parts or subassemblies that are considered
Customer Replaceable Items. Customer replaceable items include, but are not limited to, the
following items: lamps, electrodes or sensors (which are covered by a separate warranty),
reagents, calibrators, controls, paper, and pens. Consult the appropriate system operator’s
guide for a complete list of customer replaceable items for any specific model of instrument.
Siemens reserves the right to change the design or construction of specific models of
instruments at any time without incurring any obligation to make such changes available to
individual customers or instruments. If Siemens notifies customers of a change that improves
the performance or reliability of their instrument, and requests to retrofit that instrument, the
customer must agree to allow Siemens or an authorized distributor, at Siemens’ expense, to
retrofit components or make design changes, which will not adversely affect the instrument’s
performance characteristics.
OSHA requirements
When service is required at a customer location, the customer must provide the Siemens
representative with adequate facilities that comply with the regulations of the Secretary of
Labor under the Occupational Safety and Health Act (OSHA) of 1970, as amended.
Software License
No title or ownership of software is transferred to the customer. The software component of
this Siemens system and any of its modules are merely licensed to the customer for its own
use on such system.
Any software (including documentation) provided for the system contains proprietary
information constituting valuable trade secrets and is protected by federal copyright law.
The software may not be disclosed, in whole or in part, to third parties or duplicated in any
form or medium except as necessary for program execution and archival storage.
Prior to change of ownership, Siemens must be contacted by the original customer to
establish terms for the transfer of the software license.
To prevent electrical shock or damage to the system, remove power from the
system, as described in this procedure.
1. Touch Status > Shutdown.
Caution: Cartridges installed in the system remain stable for six hours without
power. To maintain cartridge stability, do not remove power from the system for more than
six hours if a cartridge is installed.
2. When the Shutdown screen is displayed, turn off system power and unplug the power
cord.
3. Disinfect the exterior surfaces of the system. Refer to Chapter 5 of the Rapidlab 1200
Series Operator's Guide.
4. Move the system to its new location.
5. Plug in the power cord and turn on the power switch.
To prevent electrical shock or damage to the system, remove power from the system
as described in this procedure.
Biohazard: Refer to Appendix A of the Rapidlab 1200 Series Operator's Guide for
recommended precautions when working with biohazardous materials.
Clean the sample path with bleach as described in Chapter 5 of the Rapidlab 1200 Series
Operator's Guide RL1200_OPERATORS_GUIDE_06621382rs.pdf#page=133.
Note: These procedures should done with the Reagent and Wash doors open. Also, if a sink
is available, remove the manifold and flush it over the sink.
1. Flush the tubing for the RCx reagent:
a. Connect a piece of tubing to a syringe.
b. Connect the syringe tubing to the connector for the RCx reagent.
Inject water at 1.
e. Flush with a minimum of 12 mL air.
f. Collect the water in an absorbent swab or equivalent.
4. Flush the tubing for the measurement module:
a. Remove tubing from the pinch valve.
b. Disconnect the tubing from the left side of the pressure detector.
c. Connect the syringe tubing to the pressure detector.
This alternative procedure requires a keyboard and a WindowsXP driver (on CDROM or
floppy) from Microsoft or from the printer manufacturer.
Printer manufacturers frequently include drivers for different printers, and for different
operating systems, on the same CDROM. You will need to search the CDROM to locate the
WindowsXP driver for the printer being installed.
1. To begin, enable the system by going into diagnostics using the password of the day or
the service password.
Note: Failure to do this step will suppress the task manager and prevent access to the
Cmd window.
2. Access the Windows Run dialog or any Windows Cmd window; to open a Cmd window
press <Ctrl+Alt+Del>, access Task Manager, then access menus File (Alt-F), New Task
(Run...), and enter cmd.exe as the task to run. Note: you may have to manually move
the Windows Task Manager to access the Cmd.exe screen, which may be overlain.
5. Select Next, then deselect Automatically detect and install my Plug and Play
printer. Be sure that Local printer attached to this computer is selected. Select
Next.
7. Within Windows Explorer, switch to the CD or floppy drive that contains the driver
software. There may be folders with names such as "DRIVERS", "SETUP", "INSTALL",
10. Before closing all Windows boxes, you may need to go to Printing Preferences in the
printer menu to choose the page size. By default the format is LETTER, but many sites
use the A4 format.
Once the USB printer driver is installed on the Rapidlab 1200 system, you may continue on
to Selecting Printer Options in the Rapidlab 1200 system. The Setup -> Printer Options
screens allow you to perform the following functions:
enable or disable the internal and external printer
select the number of copies for patient and QC sample reports
select printer settings for each type of report
The Setup -> Printer Options screen displays options based on the selected report. Select
several different reports to view the available options.
To set up printer options, perform the following steps:
1. At the Setup screen, select Printer and Devices -> Printer Options.
2. At the Printer Options screen, select the report type from the scrollable list. The system
displays only those options that are available for that report.
3. Select the printer that you want to print this report: Internal Printer or External Printer.
To print this report on both printers, complete these steps for one printer, then select the
other printer.
Main Assembly
Electronics
Caution: Do not interrupt power during this sequence. A power interruption can
significantly affect the calibration and impact the accuracy of your pO2 results
1. At the Ready screen, select the instrument icon. The Status screen opens.
2. Select Diagnostics.
3. Enter the service password or password of the day.
4. Select Sensors.
5. Select Calibrate.
6. Wipe the contact area of the EZ blank sensor lightly with dry gauze, or a clean dry cotton
cloth.
7. When prompted, remove the pO2 sensor.
8. Insert the EZ blank sensor.
pO2 Calibration
1. Close the door. You must close the door to continue the calibration.
2. When prompted to confirm that sensors were replaced or removed, select Yes.
3. Select Continue. The calibration takes approximately 15 seconds.
4. When prompted, reinstall the pO2 sensor.
5. Close the door.
Front Wall
Do not reverse the motors when removing or replacing them or the system will not
function correctly.
Tools Required
Procedure
A. Preparation
1. In main menu select Diagnostics.
2. Enter password: 1 2 3 4 5
3. Select Cartridges.
4. Select Eject Both Cartridges.
5. Remove both cartridges.
Note: Verify that the Alignment Tool reads zero using Gauge Plate before checking
alignment.
Measurement Module
Caution: Do not interrupt power during this sequence. This significantly affects the
calibration and impacts the accuracy of your pO2 results
1. At the Ready screen, select the instrument icon. The Status screen opens.
2. Select Diagnostics.
3. Enter the service password or password of the day.
4. Select Sensors.
5. Select Calibrate.
6. Wipe the contact area of the EZ blank sensor lightly with dry gauze, or a clean dry cotton
cloth.
7. When prompted, remove the pO2 sensor.
8. Insert the EZ blank sensor.
pO2 Calibration
1. Close the door. You must close the door to continue the calibration.
2. When prompted to confirm that sensors were replaced or removed, select Yes.
3. Select Continue. The calibration takes approximately 15 seconds.
4. When prompted, reinstall the pO2 sensor.
5. Close the door.
6. When prompted to confirm that sensors were replaced or removed, select Yes.
7. Select Continue. The system may require you to repeat this calibration. This ensures
that 2 successive calibrations are within accepted tolerance. To repeat the calibration,
return to step 5 under Electronic Zero Procedure.
8. Select Return 2 times.
The system automatically calibrates when this procedure is completed. When the calibration
finishes you may continue with normal operation.
Caution: Do not interrupt power during this sequence. This significantly affects the
calibration and impacts the accuracy of your pO2 results
1. At the Ready screen, select the instrument icon. The Status screen opens.
2. Select Diagnostics.
3. Enter the service password or password of the day.
4. Select Sensors.
5. Select Calibrate.
6. Wipe the contact area of the EZ blank sensor lightly with dry gauze, or a clean dry cotton
cloth.
7. When prompted, remove the pO2 sensor.
8. Insert the EZ blank sensor.
pO2 Calibration
1. Close the door. You must close the door to continue the calibration.
2. When prompted to confirm that sensors were replaced or removed, select Yes.
3. Select Continue. The calibration takes approximately 15 seconds.
4. When prompted, reinstall the pO2 sensor.
5. Close the door.
6. When prompted to confirm that sensors were replaced or removed, select Yes.
CO-ox Module
Caution: Do not touch the end of the bundle. Be sure the end is kept free of dirt.
5. Disconnect all other cables.
6. Remove the four screws securing the CO-ox Module in place.
7. Slide the module out from the front of the instrument.
8. Replace the module by following these steps in reverse order.
Replacing FD3
1. Shut Off system power and unplug the power cord.
2. Disconnect the FD3 tubing from the Measurement Module and from the nipple on the
CO-ox optics assembly.
3. Loosen the two captive screws behind the front partition, and rotate the front of the
instrument forward.
4. From the rear, squeeze the release tab on the the FD3 connector and disconnect it from
FD3. Do not tug on the wires.
5. FD3 is snap fitted in place. Wedge a flat-bladed screwdriver between the well wall and
one of the tabs on FD3 to unseat it while pulling from the front.
6. Replace FD3 by following these steps in reverse order.
Parts Required
Electronic Zero (EZ) Blank Spare 05353899
Service Password or Password Of The Day. Call your local technical support provider to
obtain this information.
Procedure
Caution: Do not interrupt power during this sequence. A power interruption can
significantly affect the calibration and impact the accuracy of your pO2 results
1. At the Ready screen, select the instrument icon. The Status screen opens.
2. Select Diagnostics.
3. Enter the service password or password of the day.
4. Select Sensors.
5. Select Calibrate.
6. Wipe the contact area of the EZ blank sensor lightly with dry gauze, or a clean dry cotton
cloth.
7. When prompted, remove the pO2 sensor.
8. Insert the EZ blank sensor.
pO2 Calibration
1. Close the door. You must close the door to continue the calibration.
2. When prompted to confirm that sensors were replaced or removed, select Yes.
3. Select Continue. The calibration takes approximately 15 seconds.
4. When prompted, reinstall the pO2 sensor.
5. Close the door.
6. When prompted to confirm that sensors were replaced or removed, select Yes.
7. Select Continue. The system may require you to repeat this calibration. This ensures
that 2 successive calibrations are within accepted tolerance. To repeat the calibration,
return to step 5 under Electronic Zero Procedure.
8. Select Return 2 times.
The system automatically calibrates when this procedure is completed. When the calibration
finishes you may continue with normal operation.
Section 4: Troubleshooting
Overview
The Rapidlab 1200 diagnostic system is comprised of four levels, each designed to provide
easy to understand, useful information regarding system functions and components. Each
level provides a different type of information and evokes a different user response, based on
the type of information, and when, how, and where it is provided. The following subsections
describe each of the four levels listed below:
Level 1: Informative
Level 2: Predictive
Level 3: Error identification
Level 4: Corrective
The goal of the informative level is to provide users with information that is continually
available to assess current system performance. Inherent in this level are performance
monitoring and data management capabilities required to collect and report the information.
Although this information is always available, it is not provided automatically. Users must
request the information. Examples of the type of “informative” information that can be
retrieved by the user to check system functions are:
Status logs, including lists of error messages and warnings
System logs that provide statistics and cycle counts
The predictive level allows users to resolve conditions before actual errors occur, decreasing
possible downtime. The system provides warnings and other predictive information to assist
users in anticipating future error conditions, and makes specific recommendations about the
action(s) required. Inherent in this level are performance monitoring and data management
capabilities that collect and report information. This level relies on trends and historical data,
not single events.
The system provides “predictive” information automatically to the user. Examples of
predictive information include:
Reagent Cartridge - Time remaining before replacement required
Wash Cartridge - Time remaining before replacement required
Waste Bottle – Reports when container almost full and when it is full
The error identification level provides feedback to the user that the system has detected an
error that is affecting system performance. The information explains the error condition and
the required corrective action. Inherent in this level are performance monitoring, and data
management capabilities required to detect faults.
The system provides “error identification” information automatically two ways: a text
message for the user, and a numeric code (qualifier) for service personnel, provided either in
the status Event Log or on the Recall screen. Examples of error identification diagnostics
include:
tHb out of calibration
No fluid detected
Systems errors
Error messages are classified and handled according to the severity of the fault, and the
required user action. These messages are ordered according to the date and time of their
occurrence, with the most recent appearing first.
The goal of the corrective level is to assist users in correcting faults as quickly and easily as
possible, in order to decrease system downtime. This level provides troubleshooting support:
information that provides the user with explanations and problem resolutions. Inherent in this
level is the knowledge base required to provide intuitive and effective help to isolate and
correct faults.
Troubleshooting support information includes the following:
Individual screen instructions - provided as part of the user interface and included on
each screen, where applicable, to explain the actions the user must take. These
messages may be found in the 1200 Series UI Specifications for Analysis Sequence,
Status, Data Recall and Setup. Content and conditions that display error message boxes
are not within the scope of this chapter.
Glossary
Acronyms
Exception messages, in the form of D-codes and operational exceptions, are displayed in the
Data Recall section where an historical log is kept for 250 errors. Unresolved or open
exceptions are displayed in a Status event list that is found in the Status section.
Exceptions and system messages are displayed in chronological order, with the most
recently generated message posted at the top of the list.
Error Clearing
An exception message remains on the Status Events Log until the condition that led to its
generation is no longer in effect.
In the D-codes and Operational Exceptions Sections, the How Cleared item defines the
method of clearing the exception.
Error Dependency
An effort will be made to suppress an error if another existing error may have directly or
indirectly caused it. For example, MOST D_codes will be suppressed by the presence of
D73 CO-ox Sample Chamber Position Error. By suppressing these dependent errors,
diagnostics is properly directed to the relevant failure.
Only a D73 will stop sample analysis. All other D-Codes will only prevent reporting of tHb
results.
By definition, instrument in the Not_Ready state is an instrument that will not accept sample
analysis requests. A Not_Ready instrument will have the Status Screen displayed and,
depending on additional information, an error message box displayed. MIT operations may
be accessible when the system is Not_Ready. The display of the Primary screen, however,
indicates that the unit is ready, as determined by the sensor icon status.
Clot Removal
Unlike other measured parameters, the calibration state of the CO-oximeter (tHb parameter)
is affected by many, non-fatal electronic and optical errors. In order to clarify the behavior of
the tHb parameter, the following table, which has the following columns, has been
generated:
1. Error – the Dcode/Operational Exception (along with its internal RTP/UIP #define from
the code) for all errors that could impact the tHb parameter calibration state.
2. tHb icon state (\, X, or “None”) – the symbol used with the tHb parameter on the
Analysis screen to indicate the calibration state of the parameter.
“\” means that the parameter is out of calibration, but the system is likely to be able to
correct the issue.
“X” means that the parameter is out of calibration, and the system is unlikely to be
able to correct the issue without user intervention.
“None” means that the parameter should not be out of calibration for this error.
3. Back In Cal with Zero – the error can be cleared (and tHb brought back in calibration)
simply by running the CO-ox Zero sequence and passing the tests.
4. Back In Cal with Slope – the error can be cleared (and tHb brought back in calibration)
by running the CO-ox Zero and Slope (Full tHb calibration) sequence and passing the
tests.
5. Permanently Dead? – the error cannot be cleared without rebooting the instrument.
6. Needs To Track What Calibrations Were Missed – the error is such that we need to
track what calibrations have occurred since the error was detected, and run the most
thorough calibration that was “missed” while the error was present, prior to returning tHb
to "in calibration."
Chart
D2 \ Yes
CX_THB_SLP_DRIFT_ERR
D3 \ Yes
CX_THB_SLOPE_ERR
D70:2 \ Yes
CX_ZERO_DRIFT_ERR
D70:3 \ Yes
CX_DETECTOR_OOR_ERR
D70:4 \ Yes Yes
CX_DARK_NOISE_ERR
D70:7 \ Yes Yes
CX_DARK_CURRENT_OOR_ERR
D70:9 X Yes
CX_ALIGN_PROM_READ_ERR
D70:11 \ Yes Yes
CX_NEON_LAMP_FAILURE_ERR
D-Codes
A fault detection event is an event that is detectable by the system and for which the system
has a defined response. Fault detection events are identified by D-codes, which describe a
possible instrument component failure. D-codes require the user to follow a certain
troubleshooting procedure, which may include the execution of one or more MITs (Manually
Initiated Tests).
D-codes are described as follows:
Qualifiers Lists the different qualifiers that further define the
D-code.
Description Defines the D-code and describes how it is
generated.
When Detected Describes the instrument operations that result in
the generation of the D-code.
Current Sample If the D-code is generated during sample analysis,
this item describes the outcome of the assay.
System Response Describes the behavior of the system after the D-
code is generated. In response to the D-code, the
system places itself in a troubleshooting mode. In
some cases, the instrument is set to the
Not_Ready state, which means that the system
displays the Status screen.
allows operations in “Setup” and “Recall,” but
no sample analysis will be permitted.
How Cleared Describes how the D-code may be cleared.
Error Message The message that appears in the Events Log to
inform the user of the error.
MIT (Manually Initiated Names the MITs used to diagnose the D-code
Test) failure.
Fault Injection Method Describes how to force the particular D-code to
occur.
D5 No Endpoint: <qualifier>
D6 Noise Error:<qualifier>
Qualifiers None
Description The 200Cal bag PO2 differs significantly from the
assigned value based on measurement of room air
pO2 and barometric pressure. The calculation
algorithm and limits are defined in the Rapidlab
Next Generation Calibration Process Specification
Doc. No. 129306.
A PO2 Reagent Error is set only if the pO2 sensor
has none of the following D-codes: D2, D3, or D4.
Note that it is assumed by this diagnostic that the
barometer reading has been calibrated by the user.
When Detected D8 is tripped during a full calibration (note that the
diagnostic actually runs immediately prior to a full
calibration). A D8 will be set whenever the
200Cal_PO2_Bias is >200Cal_PO2_Bias threshold.
The calibration that sets the D8 can be either a
scheduled calibration, or a repeat calibration.
D8 can also be detected on 1-point or 2-point
calibrations following R-Cart replacement since
V2.0 software.
Current Sample Not applicable. Occurs during calibration.
System Response During Calibration:
1. Calibration report contains spaces for pO2.
2. Posts D8 message.
3. Prints D8 code message with calibration results.
4. If LIS turned on, sends appropriate LIS
message.
5. Sets pO2 icon of the failed channel to
“Permananetly Out of Cal.”
6. Repeats calibration (see Calibration Process
Doc. No. 129306).
Next Sample Channels that are Not In Cal will not be measured.
Qualifiers None
Description After the sample was detected at FD1, and the
sample moved into the measurement block, the
sample was not detected at FD2 within the time
allowed, FD1 still detects sample, and no vacuum
error is detected. This may indicate that the fluidic
system has malfunctioned, or that a fluid detector
has malfunctioned or was not calibrated.
Qualifiers 1 = FD1
2 = FD2
3 = FD3
Description The 180 or 90 signal from indicated fluid detector is
out of calibration range. See Rapidlab Next
Generation Calibration process Specification Doc.
No. 129306 for ranges.
When Detected At power-up initialization (PUI), during fluid detector
calibration after cartridge replacement, and once
per day at the beginning of a Full Calibration.
Current Sample Not applicable
System Response 1. Posts D19 message.
2. Sets instrument to Not_Ready.
3. Goes to Status screen and displays the
troubleshooting message.
Next Sample System cannot analyze samples in Not_Ready
state.
How Cleared power-up initialization (PUI)
Error Message ID_FLUID_DETECTOR_DIAG “D19 Fluid Detector
Error:"
MIT None
Fault Injection Method Disconnect the fluid detector leads.
Qualifiers None
Description The pressure sensor error trips if the pressure
signal A/D reading is outside the range: ±1 volt.
When Detected Continuously (read once a second.)
System Response 1. Posts D20 message and enters error into status
log.
2. The gas sensors will be put out of cal.
3. Goes to Status screen and displays
troubleshooting message.
Next Sample System is not available for sample analysis until
error is cleared.
How Cleared Signal is monitored and will clear automatically
when the value is within acceptable range. A 2-
point calibration must be completed successfully
before the gas sensors come back into calibration.
Error Message ID_PRESSIGN_DIAG “D20 Pressure Signal Out Of
Range”
MIT None
Qualifiers None
Description The instrument software encountered a fatal
condition due to a UIP-RTP communication failure
from which it can not recover.
When Detected May occur any time, during or between operations.
Current Sample Is discarded if the error is detected before
measurement has been completed.
Is accepted if the error is detected after
measurement is completed on Results Screen.
System Response 1. Posts D21 message.
2. If running a user initiated sample, informs the
user that the sample has been aborted.
3. Terminates current operation, if any.
4. If LIS turned on, sends appropriate LIS
message.
5. Sets instrument to Not_Ready.
6. Goes directly to Status screen.
7. Displays fatal error message.
8. In case of RTP failure, the error condition may
be encoded using RTP LEDs and sent to the
UIP for logging by OM. The RTP will be unable
to set LEDs in response to some failures.
Qualifiers None
Description The barometric pressure error trips if any of the
following conditions is true:
The calibrated or uncalibrated pAtm is outside
the range:
523 to 800 mmHg.
The barometric pressure sensor offset is
outside the range:
–30.0 to +30.0 millivolts.
When Detected 1. During power-up initialization
2. Before a calibration or sample
3. During the user-initiated Barometric Pressure
Calibration
4. By the Temperature/pAtm MIT
During Barometric Pressure calibration, the user
interface software rejects any entered Barometric
Pressure value that is outside the range 523 to 800
mmHg. D22 also trips if the entered value is inside
the range, but is still high (or low) enough to
generate an out-of-range slope.
Current Sample Maintained
System Response 1. Posts D22 message and enters error into
status log. The gas sensors will be put out of
cal if Barometirc Pressure is being used for
other than calculated parameters such as are
used for D8 or D23:1.
2. Prints D22 error message on sample reports if
calculated parameters that depend on
barometric pressure are turned on in Setup.
Calculated parameters that depend on BP will
not be reported.
Fault Injection Method Replace probe tubing with clear tygon tubing and
observe cal preslugs. After the 2nd preslug,
disconnect the tubing as follows:
For qualifier 3, 4, 5, disconnect exit tubing
from Measurement Module before CAL
reagent gets to FD1.
For qualifier 11, 12, 13, use forcep to clamp
the exit tube immediately after the CAL
reagent gets to FD1, and before it gets to
FD2.
Fault Injection Method Replace probe tubing with clear tygon tubing and
observe cal preslugs. During the wash following the
calibrations, disconnect the exit tubing from
Measurement Module.
D23:14
Qualifiers 1. Level 1
2. Level 2
3. Level 3
4. Not Used
5. Not Used
Description An error was detected while delivering AQC. This
could result from any of the following conditions:
1. Out of AQC reagent.
2. Obstruction in the fluidic path of the Reagent
Cartridge.
3. Disconnected AQC connector or AQC reagent
bag tubing.
4. Bubbles or short sample in AQC reagent during
positioning to FD2.
5. AQC reagent not detected by FD1.
6. Air leak in fluidic path of AQC cartridge.
Note: AQC hardware currently prevents the test
for an Air leak in the fluidic path of the AQC
cartridge from being implemented.
These errors occur when a valve fails to move to the correct position.
Qualifiers 1. AQC Cartridge movement failure
2. AQC Cartridge fails to eject
Error Message ID_AQC_VALVE_DIAG - "D32 AQC Cartridge
Valve Error:"
D32:1
D32:2
Qualifiers None
Description Reagent Cartridge valve fails to move to certain
position because of a valve driver failure.
When Detected 1. During the valve move that occurs during almost
all instrument operations.
2. During the Valve Test MIT, if the valve fails to
travel to the appropriate position.
3. On power up.
Note: Reagent Cartridge valve errors are not
detected during the ejection phase of reagent
cartridge replacement or when running diagnostics.
Qualifiers None
Description Cartridge Loader fails to move to either engaged or
retracted position because of a loader drive failure.
The failure can be a drive motor malfunction or a
position sensor malfunction.
D37:2, D37:3
D39 Obstruction
Qualifiers None
Description The system is unable to properly move and position
fluids from the sample device through the
measurement module past FD2. It is detected by
excessive vacuum in the measurement module.
When Detected All sampling (Patient, QC, Clean and Condition).
Current Sample Discarded
System Response 1. Posts D-code.
2. If LIS turned on, sends appropriate LIS
message.
3. Displays a message stating that an obstruction
was detected while aspirating the sample.
4. Sets the pCO2 and pO2 sensors out of
callibration and calls for an uninterruptible 2-
point calibration, due as soon as possible.
5. System automatically initiates clot removal
procedure. See clot removal procedure.
The system must run a 2-point calibration before it
becomes available for sample analysis.
Next Sample Sampling is allowed once the D-code is resolved
either by a backflush or cartridge replacement.
How Cleared Clot removal procedure; successful wash and 2-
point calibration
Error Message ID_CLOTPRESENT_DIAG "D39 Obstruction"
MIT None
Fault Injection Method In the Analyze screen, press the Analyze button.
Insert a syringe but clamp off the measurement
module exit tubing.
Qualifiers None
Description This D-code error applies only to 126X models only.
The active to inactive glucose signal ratio
(calculation algorithm per Rapidlab Next Generation
Calibration process Specification Doc. No. 129306)
is calculated and compared to a valid range after all
analytes have endpointed during the 0 Cal segment
of a 2-point calibration. If the Active/Inactive ratio
falls outside the valid range, D50 is posted.
When Detected During all 2-point cal operations. The test is skipped
if the Glucose sensor was turned off in Setup.
Current Sample Not applicable. Occurs during 2-point calibration.
Qualifiers None
Description This D-code error applies only to 126X models
only.
The active to inactive lactate signal ratio
(calculation algorithm per Rapidlab Next
Generation Calibration Process Specification Doc.
No. 129306) is calculated and compared to a valid
range after all analytes have endpointed during the
0 Cal segment of a 2-point calibration. If the
Active/Inactive ratio falls outside the valid range,
D51 is posted.
When Detected During all 2-point cal operations. The test is
skipped if the Lactate sensor was turned off in
setup.
Current Sample Not applicable. Occurs during 2-point calibration.
Qualifiers None
Description This D-code error applies only to 126X models with
Type 0 R-Cart installed.
The number of exposures of the Lactate sensor to
ZeroCal reagent has exceeded a threshold over a
defined period (calculation algorithm per Rapidlab
Next Generation Calibration Process Specification”
Doc. No. 129306).
When Detected Following all ZeroCal reagent loads (including
calibrations, MIT Fluidics or R-Cart Prime
operations).
Current Sample Not applicable. Occurs during non-sampling
operations.
D70:2
MIT None
Fault Injection Method Introduce fluid while calibrating so CO-ox will fail on
first attempt. Error should be posted and printed on
Full Calibration Report. Then cleared on the second
attempt to cal.
D70:3
Next Sample If other parameters are still “in cal,” next sample is
analyzed for those other parameters.
How Cleared Successful Zero
MIT None
Fault Injection Method Introduce fluid while calibrating so CO-ox will fail on
first attempt. Error should be posted and printed on
Full Calibration Report. Then cleared on the second
attempt to cal.
D70:7
D70:11
D70:12
Qualifiers None
Description After the user inserted a device and pressed the
Analyze key, no CO-ox sample was detected at
Fluid Detector 3 within the maximum time allowed.
This may indicate a clot, fluidic system malfunction,
fluid detector malfunction, or too small sample.
When Detected During CO-ox Sample
Current Sample CO-ox results are discarded. Blood gas, pH,
electrolytes and metabolites data are retained.
Qualifiers None
Description D73 indicates a failure of the Sample chamber to
reach an open or closed position. Specific
descriptions of tests are covered in the BIT
sections.
When Detected Power-up initialization (PUI) and each Zero, Full
calibration, Wash and Sample analysis, CO-ox
sample chamber MIT.
Current Sample CO-ox results are discarded. Blood gas, pH,
electrolytes and metabolites data are retained.
System Response 1. Posts the D73 message.
2. Sets tHb icon to CO-ox Not_Ready state.
3. Sends COOXERROR message to LIS.
Next Sample System changes to Blood Gas Only mode. CO-ox
sample measurements will NOT be allowed until the
D73 is removed.
How Cleared System shutdown and restart
Successful full calibration
CO-ox Sample Chamber MIT
Error Message ID_CX_SAMP_CHAMBER_DIAG "D73 COox
Chamber Position Error"
MIT CO-ox Sample Chamber Test
Fault Injection Method Remove the drive coupling.
Qualifiers None
Description D75 indicates a failure of the CO-ox Halogen Lamp.
Specific descriptions of tests are covered in the BIT
sections.
When Detected Power-up initialization (PUI) and each Zero, all
calibrations, Wash and Sample analysis.
Current Sample CO-ox results are discarded. Blood gas, pH,
electrolytes and metabolites data are retained.
System Response 1. Posts the D75 message.
2. Sets tHb icon to CO-ox Not_Ready state.
3. Posts lamp failure icon on the Analysis Home
screen.
Next Sample System changes to Blood Gas Only mode. CO-ox
sample measurements will NOT be allowed until the
D75 is resolved.
How Cleared System shutdown and restart
System shutdown, lamp replacement, and system
restart
Lamp calibration
If tHb is out of calibration, a tHb only, or Full 2-point
cal, must be run to bring it back into calibration.
Error Message ID_CX_LAMP_FAILURE_DIAG "D75 Lamp Failure"
MIT CO-ox Lamp On/Off or Lamp Cal
Fault Injection Method Disconnect the bulb from back of system.
System Messages
Expand "System Messages" at left to see an alphabetical listing of messages.
Qualifiers None
When Detected During the OCT performed during the wash that
follows each sample measurement, or at endpoint if
sampling device is still present.
Current Sample Continued
System Response Qualifier 1:
1. Posts error message.
2. Repeats the wash once.
3. If the OCT fails during the repeat wash, this
error is cleared and D70:2 on page 101 is
posted to the events log.
Qualifier 2: Operator Error
The wash is not repeated until the sampling device
is removed.
Next Sample No effect on next sample.
How Cleared Successful OCT in the repeat.
Error Message ID_CX_OCT_DIAG "Additional Wash Required:"
Fault Injection Method Same as D70:2.
Qualifiers None
Qualifiers None
Qualifiers None
Qualifiers None
Bubbles in Sample
Qualifiers None
Qualifiers None
Cal Overdue
Qualifiers None
Qualifiers None
Fault Injection Method Remove the CO-ox cartridge when the system is
not in the middle of a CO-ox sample chamber
movement.
Qualifiers None
Qualifiers None
Qualifiers None
Qualifiers None
Insufficient Sample
Qualifiers None
Qualifiers None
Next Sample Not accepted until the error is cleared. Once the
Measurement Module Door has been opened, any
pending scheduled AQC samples will be delayed
until either the “Calibration Yes/No” prompt has
been answered, or a Full 2-point Calibration takes
place.
How Cleared ID_UNSOLICITEDDOOROPEN_DIAG
"Measurement Module Door Opened"
Error Message Automatically, as soon as the measurement
module door is closed.
Fault Injection Method Open the measurement module door after
measurement has started.
No Paper in Printer
Qualifiers None
Description The roll printer Paper Out sensor did not detect
paper because the printer is out of paper. This error
occurs only when the roll printer is turned on
through the Printing Options screen.
When Detected During printing when paper runs out.
Current Sample Maintained. Results are not printed, or only partially
printed.
System Response 1. Posts exception to status log.
2. Disables printing.
3. If LIS turned on, sends appropriate LIS
message.
4. Continues current operation.
5. Returns to Analysis home screen.
Next Sample Future samples are accepted for analysis, but
results will not be printed.
How Cleared Install paper.
Error Message ID_PAPEROUT_DIAG "No Paper in Printer"
Fault Injection Method Install a short roll of paper, with only a few inches of
paper left.
Qualifiers None
QC Material Expired
Qualifiers None
QC Failed: <qualifier>
QC Missed: <qualifier>
Qualifiers 1, 2, 3
Description Grace period has expired and RQC or AQC has not
been run.
When Detected At end of grace period
Current Sample Not affected
System Response 1. Places parameters for the level missed Out of
QC.
2. Posts error to Status log.
3. If LIS turned on, sends appropriate LIS
message.
Next Sample Samples cannot be accepted until QC has been run
or overidden.
How Cleared When QC missed is overridden.
Error Message ID_QC_MISSED_DIAG “Scheduled QC Not Done: "
Fault Injection Method Schedule AQC or RQC and prevent QC samples
from running until the grace period expires.
QC Overridden: <qualifier>
Qualifiers None
Qualifiers None
Qualifiers None
Qualifiers None
Qualifiers None
Qualifiers None
SulfHb >1.5%
Qualifiers None
Temp Warning
Qualifiers None
Qualifiers None
Qualifiers None
When Detected After door is closed, the cartridge is loaded, but the
EEPROM is not detected.
Current Sample No effect
System Response 1. Goes to Status screen.
2. Posts exception message.
3. Sets instrument to Not_Ready state.
4. Selects Replace and Wash Cartridge buttons.
5. Displays error message box with generic
replace Wash cartridge message.
6. If LIS turned on, sends appropriate LIS
message.
7. Waits for user to proceed with Wash cartridge
replacement.
Next Sample Fluidic operations are not accepted in this state.
How Cleared Automatically, when a valid cartridge is installed
and the system detects the cartridge EEPROM.
Error Message ID_NOWASHWASTE_DIAG "No W Cartridge"
Fault Injection Method Illegally removes the Wash Cartridge and closes
the door.
Installs a Wash Cartridge without an EEPROM.
Qualifiers None
Qualifiers None
Fault Injection Method Partially fill waste bottle (at least 250 mL) with a
solution and place on the stand.
Waste Full
Qualifiers None
Qualifiers None
CO-ox Tests
The Dark Current/Dark Noise BIT verifies that the dark current and total electrical offset of
the Measurement System are below a maximum level so they will not interfere with
measurement accuracy and precision. It also verifies that the noise level of the measurement
electronics is within expected levels.
Description This test is run at power-up initialization (PUI) and
before each Zero, Full calibration and Sample
measurement.
For a detailed description of the test refer to Doc. #
123387 Chapter 11, Section 11.5.1.
Component Under Test Photo diodes and preamp circuit
D-codes D70:4 - (CO-ox Dark Noise Error)
D70:7 - (CO-ox Dark Current Out Of Range Error)
The Alignment PROM test verifies that alignment data were read correctly.
Description This test is only run at power-up initialization (PUI).
Checksum of alignment parameters should match a
pre-determined value.
Component Under Test Alignment PROM
D-codes D70:9 - (CO-ox Align PROMREAD Error)
The Lamp On Test verifies that the lamp is illuminated and that its light output is within
tolerance. The CO-ox lamp calibration requirements are in the Rapidlab 1200 CO-oximeter
Sample and Calibration Processing Specifications, Doc. # 124158.
Description This test is performed during every Sample, Slope or
Zero measurement and whenever the Lamp MIT is
run. It assumes a previous successful Lamp
Calibration and Voltage Test.
First the lamp is turned ON and the lamp control
error signal (LMP_ERROR_SIG) is measured. This
signal should change by no more than 10 mV from
the value measured during lamp calibration.
If, after 2.5 seconds, the lamp error signal differs
from the calibration value by more than 10 mV, a
D75 error is reported.
Component Under Test Lamp, Lamp control electronics
D-codes D75 - Lamp Failure
D76 - COox Electronics Error: 8 (Lamp Control
Error)
The Zero Calibration Drift Test determines whether the Zero baseline has drifted enough to
cause serious errors in the measurement results.
This test checks for the presence of the CO-ox Sample Chamber.
Description 1. This test continuously (every second) monitors
the cell flag status except during sequences and
while there is no D73 slide cell error, since this
may indicate a flag/hardware problem.
2. Cartridge removal is suspected when both Open
and Close position flags are FALSE. Debounces
the flag status (three times in a row) so that an
instantaneous change does not cause an event.
The system immediately sends a message to the
UIP that Coox cartridge removal was detected.
This test checks for the presence of the CO-ox Sample Chamber.
Description This test is run at power-up initialization (PUI) and
before each Zero and Slope calibration.
For a detailed description of the test refer to Doc. #
123387 Chapter 11, Section 11.5.6 (first test).
Component Under Test Interface electronics between polychromator board
and main board
D-codes D76:2 - COox Electronics Error (Polychromator
Board -Main Board Interface Error)
This test verifies the integrity of the data path between the Polychromator Board and the
Main Board.
Description This test is run at power-up initialization (PUI), and
during every Zero and Slope calibration and
Sample.
Whenever a new Integration Time is loaded into the
polychromator board, the new value is compared to
the old value.
Component Under Test Interface electronics between polychromator board
and main board.
D-codes D76:9 - COox Electronics Error (Polychromator
Board -Main Board Interface Error)
This test measures the offset of the ADC on the Polychromator Board and verifies that it is
within an acceptable range.
Description This test is run at power-up initialization (PUI) and
before each Zero and Slope calibration.
The ground input to the ADC on the polychromator
board is selected. A GO pulse is issued to the
polychromator board and the resulting 256 values
are averaged. If the average value is outside the
range 7F00H to 80FFH, an error is reported.
Component Under Test ADC
D-codes D76:3 - COox Electronics Error (ADC/DAC Error)
ADC Calibration
The ADC is calibrated upon power-up initialization (PUI) and before every Zero calibration. It
is also possible to test the ADC to make sure that it is calibrating correctly.
Description When the ADC is calibrated, a timer is used to verify
that the calibration was completed successfully in
less than 100 ms. The error condition is asserted if
the calibration is not finished within 100 ms.
Component Under Test ADC
D-codes D76:3 - COox Electronics Error: 3 (ADC/DAC Error)
This test uses the ADC to read back the output of the DAC. This test will verify operation of
the DAC, DAC reference, ADC, and ADC reference.
Description This test is run at power-up initialization (PUI) and
before each Zero and Slope calibration.
The DAC input to the ADC is selected. The value
03D7H is written to the DAC. A GO pulse is issued
and the resulting 256 values are collected and
averaged. The average value should be within the
range 3C79H to 3E6DH. Then the value 0800H is
written to the DAC. A GO pulse is issued and the
resulting 256 values collected and averaged. The
average value should be within the range 7F06H to
80FAH. Finally, the value 0C28H is written to the
DAC. A GO pulse is issued and the resulting 256
values are collected and averaged. The average
value should be within the range C192H to C386H.
If any of these conditions are not met, an error is
reported.
Component Under Test ADC
D-codes D76:3 - COox Electronics Error (ADC/DAC Error)
This test is to verify that the gain resistors in the readout circuit on the Polychromator Board
are correct.
Description This test is run only at power-up initialization (PUI).
The DAC is loaded with the value 016FH. A dark
measurement is then initiated. The DAC is then
loaded with 05D0H, and another dark measurement
is done. The difference between the two
measurements, in ADC counts, should be 32,768
counts ±200 counts. An error is reported if the
difference between the two measurements is
greater than 32,768 counts ±200 counts.
Component Under Test Gain resistors on the Polychromator Board
D-codes D76:4 - COox Electronics Error (Readout Circuit
Gain Error)
This test verifies that the ground noise on the Polychromator Board is acceptable.
Description This test is run at power-up initialization (PUI) and
before each Zero and Slope calibration.
Ground is selected as the ADC input and a GO
pulse issued. The resulting 256 values are collected
and the SD computed. If the SD is greater than 3.30
ADC counts, an error is reported.
Component Under Test ADC circuit
D-codes D76:5 - COox Electronics Error (Noisy Ground)
Verifies that the lamp control DAC operates over its full range.
Description This test is run at start-up and any time a lamp
control error is detected.
It is a test of the integral non-linearity of the DAC.
The DAC is alternately set to 0, mid, and full scale;
and the output is read at each level. At each level
the value must be within 100 mV (after subtracting
the offset of the Main Analog-to-Digital Converter):
Programs a 0 value into the DAC and reads 0
±100 mV.
Programs 2047 into the DAC and reads 2500
±100 mV.
Programs 4095 into the DAC and reads 5000
±100 mV.
Component Under Test MAIN BOARD Lamp DAC circuit
D-codes D76:6 - COox Electronics Error (Lamp or Heater
DAC Error)
The lamp calibration is intended to adjust the 1200 Series CO-ox main lamp to between
4190.0 mV and 4200.0 mV.
Description The complete calibration is done during power-up
initialization (PUI) and when the Lamp Calibration
MIT is run.
Initial Lamp ON Checks
Turns the lamp off, waits 5 seconds for the lamp to
cool, and measures the lamp voltage
(LAMP_V_MEAS signal). If the lamp voltage
exceeds 500 mV, sets the D75 error and quits
(does not proceed to next step).
With the lamp driver circuit disabled (LMP_ON = 0),
writes a decimal 100 into the Lamp DAC. Turns on
the lamp (LMP_ON = 1) and waits (up to 2500 ms)
for the lamp_error to come within limits (0.0V ±
20mV). Reads the lamp voltage.
Incremental Adjustment Of Lamp Voltage
If no error was encountered up to this point, the
lamp is incrementally adjusted in a loop as
described below. In order to accommodate system-
to-system variability and still reach the desired lamp
voltage quickly, adjustments are made in
increments that start large and decrease during the
process. The following table should be used.
Stage DAC Increment Minimum Lamp Voltage
0 100 3700
1 50 4000
2 25 4100
3 12 4152
4 5 4180
5 1 4190
This test monitors the magnitude of shifts in the polychromator alignment. It also serves as a
calibration of the polychromator alignment wavelength-to-pixel mapping.
Description During each calibration, the neon lamp is turned on
and 50 polychromator measurements are performed
and summed. The centroid (position in pixel units) of
the filtered neon spectral line is computed and
compared to the position that was measured during
the factory alignment of the polychromator. The
difference between the new measured position and
the position measured at the factory is called the
wavelength shift. If the wavelength shift exceeds
±1.5 pixels, a D70:12 error is posted. If there is not
enough light signal to perform the computation, a
D70:11 error is posted. If the computation is
successful, the shift of the neon spectral line is used
as an adjustment to all subsequent measurements
until the next Wavelength Shift Test.
Component Under Test Polychromator alignment, neon lamp
Motor Tests
This test monitors the movement of the valves to insure that the correct position is reached.
Description This test is performed every time the reagent valve
is moved.
Sequences commands valve movement and waits
for in-position event to be set. If the in-position
event is not set within a timeout period, the
sequence sets the error. In order for the in-position
event to be set, the move must arrive at the
requested “slot” within limits on the total steps taken
and steps from last “slot.” If the move fails the
system tries to find its “home slot,” then retries the
move once again. Special processing is used to
reduce the effects of mechanical backlash in the
valve system.
Component Under Test Main / AQC valve, position detectors, motors,
cables, and main board.
D-codes D33:1 - Main valve error (many places)
D32:2 - AQC valve error (many places)
This test monitors the movement of the CO-ox sample chamber, cartridge door, and loader
motor movement to insure that the correct position is reached.
Description This test is performed whenever the loader or ejector
motors are moved.
The sequence is to command motor movement and
wait for the “in position” event to be set. If the “in
position” event (open or closed) is not set within a
timeout period, the the error is set. In order for the
“in position” event to be set the motor must arrive at
the requested position within limits on the total steps
taken. For the CO-ox sample chamber and ejector
(cartridge door) motor, after the object position is
reached the motor must move off the object position
to relax the sample chamber, or return to the home
position.
Component Under Test CO-ox Sample Chamber / Ejector Motor and door /
Loader Motor and “cassette mechanism,” position
detectors, cables, and main board.
D-codes D73 - CO-ox Chamber Position Error (many places)
D36 - Loader Motor Error
D37:1 - Ejector Motor Movement error
D37:2 - Reagent Door Failed to Open
D37:3 - Wash Door Failed to Open
This test checks that the Micro Stepper Motor Controller Responds to the reset command.
Description This test is performed whenever the Sample, Wash,
or CO-ox pump is turned on.
The system initializes the microcontroller that
controls the sample, wash, and CO-ox pumps, and
reads its status hardware register. The error is set if
the microcontroller is not ready after delay.
Component Under Test Microcontroller, mainboard
D-codes D35:10a - Control Board Error, micro stepper
control not responding
This test checks that the Micro Stepper Motor Controller responds to commands.
Description This test is performed whenever the Sample, Wash,
or CO-ox pump is turned on.
Commands can only be written to the
microcontroller when it is in the ready state. Each
command requires four writes to the
microcontroller’s control register. If the
microcontroller does not return to the ready state
after each write, this error is set. Two failures are
required to generate this error.
Component Under Test Microcontroller, mainboard
D-codes D35:10b - Control Board Error, micro stepper
controller timed out
This tests for current being drawn by a pump’s motor when it is turned on.
Description This test is performed during wash and pump MITs
(currently not being tested).
Turn the pump on. Wait for signal to become stable
(100 ms). Read hardware motor current status
register. The status register should read “1” with a
motor on and “0” with all motors off. Ensure that
only one motor is running while the test is being
conducted.
Component Under Test Sample Pump, Wash Pump, or CO-ox Pump
D-codes D35:13a and b - Pump motor error
Fluidic Tests
This test calibrates the fluid detectors when there is air in front of the detectors.
Description Air Test
This test is run at power-up initialization (PUI) and
every third Full Point Cal.
Sequence code moves the main valve to the vent
position and turns on the sample and CO-ox pumps
to aspirate air into the CO-ox and Electrode
measurement modules. The analog values read
from the fluid detectors are verified to be within
calibration tolerances.
Component Under Test FD1 / FD2 / FD3, DAS Subsystem, Sample path
D-codes D19:1 - FD1 Out of Range
D19:2 - FD2 Out Of Range
D19:3 - FD3 Out Of Range
This test confirms that fluid is reaching Fluid Detector 1, which is located at the input of the
main measurement module.
Description This test is run at every calibration, sample test,
and AQC.
Turns Sample / Wash Pump on with valve set to
aspirate fluid position. Waits until fluid is detected at
FD1, or sets error after timeout.
Component Under Test Main Board, Pump Tubing, Sample / Wash Pump
and their motors, Sensor installation, Fluid Detector
1, Mux board, Wash cartridge, Reagent cartridge,
No Sample device, and main measurement module.
Failure could also be caused by clot or partial
obstruction.
This test confirms that fluid is reaching fluid detector 2, which is located at the output of the
main measurement module.
Description This test is run at every calibration, sample test,
and AQC.
Turns Sample / Wash Pump on with fluid at FD1.
Waits until fluid is detected at FD2, or sets error
after timeout.
Component Under Test Main Board, Pump Tubing, Sample / Wash Pump
and their motors, Sensor installation, Fluid
Detector 2, Mux board, and main measurement
module.
Failure could also be caused by clot or partial
obstruction.
This test confirms that fluid is reaching fluid detector 3, which is located at the input of the
CO-ox module.
Description This test is run at every calibration, sample test,
and AQC (12X5 Models only).
Turns CO-ox Pump on with valve set to aspirate
position. Waits until fluid is detected at FD3, or sets
error after timeout.
Component Under Test Main Board, Pump Tubing, CO-ox Pump and its
motor, CO-ox Sample chamber correct installation,
Fluid Detector 3, Wash cartridge, Reagent
cartridge, and No Sample device.
Failure could also be caused by clot or partial
obstruction.
D-codes D71 - No Sample Detected at FD3
D78 – No Cal Reagent detected at FD3
This test detects samples that are too short to cover the sensor module.
Description This operational error is detected during patient
sample measurement.
Fluid detector readings are monitored every 10 ms.
An insufficient sample event is set by the
measurement code when 20 consecutive FD1
readings are determined to be air prior to the
sample arriving at FD2.
Component Under Test Sample
Description The System will run the Air Leak Test when it is
initiated by the MIT.
The Air Leak Test seals off the normal vent and
reagent ports and runs the sample pump to create a
positive over-pressure condition. Over-pressure is
used because it does not have the detrimental
effects that creating a vacuum does (specifically on
the Gas Sensors).
Stage 1
1. Close the pinch valve.
2. If the analyzer has CO-ox hardware (RL 1245 or
RL 1265), move the CO-ox slide cell to the
closed position.
3. Run the sample pump backwards (utilizing the
one-way valve in the manifold) to create a
pressure of 1.5 psi. Shut the sample pump off
when > 1.5 (but less than 2.0 psi) has been
achieved, or when 15 seconds has elapsed,
whichever occurs first.
4. If 15 seconds elapsed without achieving at
leasst 1.5 psi of pressure, or if the pressure
exceeded 2.0 psi, the Air Leak Test Diagnostic
sets the D23:6 D-Code, opens the pinch valve,
and exits.
5. Monitor the pressure decay for 14 seconds (this
time allows for a more sensitive pressure
monitoring test to run). The Air Leak Test
Diagnostic will set the D23:6 D-Code, open the
pinch valve, and exit if any of the following three
tests fail:
The measured pressure drops below 0.25
psi (during pressure monitoring).
The measured pressure drops 1.0 mV in
any given second (during pressure
monitoring).
The 13-second Pressure Monitoring Test
fails. The 13-second Pressure Monitoring
Test computes the average pressure from
seconds 1, 2, and 3; then computes the
average pressure from seconds 11, 12, and
13. The two averages are then compared. If
the pressure has dropped > 0.75 mV, the
13-Second Test fails.
6. If 14 seconds elapses and none of the above
tests fail, the Air Leak Test Diagnostic opens the
pinch valve and advances to stage 2.
Stage 2
1. Move the main valve to the RCx position.
2. Run the sample pump backwards (utilizing the
one-way valve in the manifold) to create a
pressure of 1.5 psi. Stop the sample pump
when > 1.5 psi (but <2.0 psi) has been achieved
or when 15 seconds has elapsed, whichever
occurs first.
3. If 15 seconds elapsed without achieving at least
1.5 psi of pressure, or if the pressure exceeded
2.0 psi, the Air Leak Test Diagnostic sets the
D23:6 D code and exits.
4. Monitor the pressure decay for 14 seconds. The
Air Leak Test Diagnostic will set the D23:6 D
code and exit if any of the following conditions
are true:
The measured pressure drops below 0.25
psi (during pressure monitoring).
The measured pressure drops 1.0 mV in
any given second (during pressure
monitoring).
The 13-second Pressure Monitoring Test
fails.
5. If 14 seconds elapses and none of the above
tests fail, the Air Leak Test Diagnostic advances
to stage 3.
Stage 3
1. Move the main valve to the vent position.
2. If the analyzer does not have CO-ox hardware
(RL 1240 or RL 1260), the leak test has passed,
so exit the test.
3. If the analyzer has CO-ox hardware (RL 1245 or
RL 1265), move the CO-ox slide cell to the open
position.
4. Move the main to the RCx Position.
5. Run the sample pump backwards (utilizing the
one-way valve in the manifold) to create an
over-pressure of 1.5 psi. Stop the sample pump
after > 1.5 but less than 2.0 psi has been
achieved, or when 15 seconds has elapsed,
whichever occurs first.
6. If 15 seconds elapsed without achieving at least
1.5 psi of pressure, or if the pressure exceeded
2.0 psi, the Air Leak Test Diagnostic sets the
D23:6 D-Code and exits.
7. The Air Leak Test Diagnostic will set the D23:6
D-Code, open the pinch valve, and exit if any of
the following conditions are true:
The measured pressure drops below 0.25
psi (during pressure monitoring).
The measured pressure drops 1.0 mV in
any given second (during pressure
monitoring).
The 13-second Pressure Monitoring Test
fails.
8. If a the Leak test passes, clear any leak errors
and exit the Diagnostic successfully.
Component Under Test System fluidic path
D-codes D34:2 - Sensor Offset out of range
D34:3 - Sensor Slope out of range
This test calibrates the Waste Subsystem and sets errors if the calibration fails.
Description The waste sensor requires a 2-point calibration on
installation and on request. The process involves
removal of the waste bottle and determining the
offset with zero weight (offset) and then filling the
waste bottle (350 mL) and computing the slope.
This calibration is designed to eliminate differences
in waste sensors.
Component Under Test Waste Subsystem, DAS Subsystem
D-codes D34:2 - Sensor Offset out of range
D34:3 - Sensor Slope out of range
This test corrects the Waste Subsystem output voltage and sets errors if the calibration fails.
Description This test is performed during each calibration.
The waste sensor contains an internal calibration
signal and a voltage input port. The goal of a one-
point calibration is to provide a voltage to the input
port such that the calibration signal output is
reduced to zero within tolerance.
Component Under Test Waste Subsystem, DAS Subsystem
D-codes D34:1 - Sensor DAC Internal Test voltage out of
range
This test monitors the waste sensor to detect if the waste bottle is missing, almost full, or full.
Description This test is performed during each calibration.
The analog voltage of the waste sensor is
compared against values established during waste
calibration. Depending on the value read, the test
will set error flags to indicate if the waste bottle is
missing, almost full, or full.
Component Under Test Waste subsystem
Heater Tests
This test monitors the temperatures of the heaters to ensure that they are above minimum
acceptable values.
Description The heater failure test checks if the temperature is
above minimum when the thermal system is in a
stable state (time, control state = main control, and
sequence not running). Limits are different for each
heater.
Component Under Test Block Heater / Pre-Heater / Door Heater / and CO-
ox Heater, Main board, thermistors
D-codes D38:4 - Block Heater failure
D38:7 - Pre-Heater failure
D77:2 - CO-ox Heater failure
D38:9 - Door Heater failure
D38:12 - If error persists
This test monitors the temperatures of the heaters to ensure that they are below maximum
acceptable values.
Description The heater out of range test checks that that
temperature is below maximum when the thermal
system is in a stable state (time, control state =
main control, and sequence not running) test. Limits
are different for each heater.
Component Under Test Block Heater / Pre-Heater / Door Heater / and CO-
ox Heater, Main board, thermistors
This test monitors the heater DAC (Digitial Analog Converts) for activity.
Description The Heater DAC Activity test is run at PUI, when
the thermal system is turned on, and when another
thermal error is detected.
The heaters are turned on and the voltage output
by the DACs is read and tested against tolerances.
Testing is done at three levels:
1. DAC Min counts
2. DAC Mid counts
3. DAC Max counts
Component Under Test Pre-heater / Block heater / CO-ox heater / Door
heater, main board, DAS subsystem
This test monitors the sample path temperature at the end of the sample measurement.
Description This test is run at the end of a patient and QC
sample measurement. It is not run after an
Automatic QC sample.
Depending on what state the system is in, selects
the range to be tested against. Compares the
sample path temperature at the end of
measurement to:
Out of range high limit (Error)
Out of range low limit (sample error)
Component Under Test Sample (Operation Exception)
Expand "Analog Subsystem (DAS) Tests" at left to see the list of tests.
This test monitors the DAS calibration values for meeting required criteria.
Description The DAS Calibration Tests are performed at PUI
and before every calibration. (Currently not being
activated by SEQ.)
Check that the calibration values are within
tolerance:
1. DAS CAL OFFSET
2. DAS CAL GAIN
Component Under Test DAS Subsystem
This test checks the DAC reference voltage for a within tolerance condition.
Description The DAS DAC Reference Test is performed at
power-up initialization (PUI) and before every
calibration. (Currently not being activated by SEQ.)
Reads the 5 volt reference channel and verifies that
it is within tolerance.
Component Under Test DAS Subsystem
This test checks the operation of the DAC by setting it to several values and reading the
voltage back through the A/D.
This test checks the common mode rejection of the A/D by reading a known voltage from the
DAC and then reading the other leg of the signal.
Description The DAS Common Mode Rejection Test is
performed at power-up initialization (PUI) and before
every calibration. (Currently not being activated by
SEQ.)
Adds the DAC high signal to the DAC low signal.
The sum should be close to zero. Tests that it is
within tolerance.
Component Under Test DAS Subsystem
This test monitors the IRQ5 signal and ensures that it responds correctly.
Description The Read A/D Tests are performed continuously.
Delay until IRQ5 signal is cleared, or error. Start
conversion and ensure that IRQ5 signal gets set, or
error.
Component Under Test DAS Subsystem
This test checks to see if the DAS bus and DAS interrupt are working.
Description The DAS Bus and Interrupt Test is performed
continuously.
1. Attempts to clear the A/D Interrupt Line.
D35:11 b, c, d, f
2. Attempts to set the A/D Interrupt Line.
D35:11 a, e
Component Under Test DAS Subsystem
This test checks the ability of the DAS system to accurately modify the premux gains in the
system.
Description The DAS Check Premux Gain Test is performed at
power-up initialization (PUI) and before every
calibration. (Currently not being activated by SEQ.)
There are two programmable gain amplifiers
(PGA1 & PGA2) in the DAS subsystem. Each of
these amplifiers may be set to a gain of 1, 2, 4, 8,
or 16. This test programs the two amplifiers to give
a gain of 16 in the following combinations, and
ensures that all resulting values are within
tolerance:
PGA1 = 01 & PGA2 = 16
PGA1 = 02 & PGA2 = 08
PGA1 = 04 & PGA2 = 04
PGA1 = 08 & PGA2 = 02
PGA1 = 16 & PGA2 = 01
Component Under Test Programmable gain amplifiers, DAS Subsystem
This test checks the performance of the offset capability of the system.
Description The DAS Check Premux Offset Test is performed at
power-up initialization (PUI) and before every
calibration. (Currently not being activated by SEQ.)
The same Analog source signal (PREMUX_1) is
read by the A/D converter at the following gains at a
constant offset (250 mV), and the resulting values
compared to tolerances:
Gain of 1
Gain of 2
Gain of 4
Gain of 8
Gain of 16
If any of the “LED Code” failures occur, the real-time processor (RTP) determines that it is
fatally broken and may not be able to communicate with the user interface processor (UIP).
Under these conditions the RTP sets the LED display to a light pattern, attempts to send a
message for OM logging, and does not respond further. The RTP may not be “well enough”
for the logging message to reach OM but a good first step in determining the cause of a D21
is to look at the OM log.
There are other reasons for the RTP to cause a D21 error including:
The RTP may also fail to respond if a noise spike has caused the RTP to reset, LEDs
would be all off (Code 00).
The RTP CPU (68332) could also detect an interrupt that the system is not set up to
respond to, in which case the LED code will be the vector receiving the interrupt. (This is
new, unproven code ported from the Boot PROMs.)
The RTP could also become lost in which case the error code could still be changing or
be meaningless.
The user is made aware of any of these failures by the UIP setting a D21 error when the
RTP does not respond as expected.
The Boot PROMs on previous programs also set LED codes (normally around 0x40 and
above).
LED Codes
The following table defines the displayed LED Code for conditions under which the RTP
does a controlled crash. An “X” in the “How Set” column indicates tat the LED code is not
currently being used.
Hex Code Meaning How Set
80 SPURIOUS_INTERRUPT_0 A spurious level 0 interrupt was detected
84 SPURIOUS_INTERRUPT_4 X
91 OS_TCREATE_ERR_NOSTK X
92 OS_TCREATE_ERR_TINYSTK X
94 OS_TCREATE_ERR_OBJDEL X
95 OS_TCREATE_ERR_OBJID X
96 OS_TCREATE_ERR_ACTIVE X
97 OS_TCREATE_ERR_ILLRSC X
98 OS_TCREATE_UNKNOWN X
99 pNA_INIT_ERROR X
9C BAGS_ARE_FULL X
AB T_IDENT_ERR_OBJNF X
AC HARDWARE_READ_FAILURE X
B0 LAC_GAIN_LESS_ZERO X
B1 BAD_ARRAY_INDEX X
B2 MEAS_SOFTWARE_ERROR X
B3 PUT_DATA_PCCMD_ERROR1 X
B4 PUT_DATA_PCCMD_ERROR2 X
BB SOCK_ENFILE X
BC SOCK_ENOBUFS X
BE SOCK_UNKNOWN X
BF SETSOCK_EBADS X
C0 SETSOCK_EINVALID X
C1 SETSOCK_ENOPROTOOPT X
C2 SETSOCK_ENOBUFS X
C3 SETSOCK_UNKNOWN X
C4 BIND_EBADS X
C5 BIND_EINVALID X
C6 BIND_EADDRINUSE X
C7 BIND_EADDRNOTAVAIL X
C8 BIND_UNKNOWN X
C9 LISTEN_EINVALID X
CA LISTEN_EOPNOTSUPP X
CB LISTEN_UNKNOWN X
DD MBPEND_FAILED X
DE OSLOCK_FAILED X
DF MBPOST_FAILED X
Central to the 68332 exception processing sequence is a vector table that occupies 1024
bytes of memory. This table is located at addresses 0x000000 to 0x0003ff. The table is
organized into 256 vectors. Each vector is a 32-bit address that will be loaded into the
program counter as part of the exception processing sequence.
The code located at this vector address will be executed if the exception occurs. During
initialization the software places the address of vector_wrapper() ISR into all unused entries
of the exception table. The vector_wrapper() ISR decodes from which address it was called
and sets the LEDs to this value.
The following table shows what Hex Code will be encoded on the LEDs if an unhandled
interrupt through that vector is detected. Where the “Exception” column entry is in
parentheses (), it indicates that that vector is used by the system for the purpose indicated.
05 Divide By 0 29 Trap 9
17 Unassigned 08 3B Unassigned 09
1D Level 5 Interrupt To To
20 Trap 0 To To
23 Trap 3 To To
25 Trap 5
Expand "User Interface Processor (UIP) BITs" at left to see the list of tests.
This test monitors system power resources, using the Advanced Configuration & Power
Interface (ACPI) IC.
Description The ACPI chip is configured and periodically polled
to see if the power supplies and monitored
temperatures are within tolerance. The power
supplies must have a tolerance of ±5% within the
resolution of the chip, and temperature range from
0 – 70 °C
Component Under Test Power supplies and internal thermal conditions
This test checks the system's ability to eject the AQC cartridge.
Description The AQC Cartridge's EEPROM is monitored and, if
the user interface processor (UIP) can still read the
EEPROM after the cartridge was to be ejected, this
error is set.
Component Under Test AQC stepper motor and AQC cartridge
CommunicationsTest
This test checks the connection to an external device using the RS332 or Ethernet port.
Description Whenever a message (other than the LIS
acknowledge message) is sent to an external
device, an acknowledgement is expected from the
external device within a timeout period (8 seconds).
If a message is not acknowledged within 8 seconds,
the message is sent a second time, and the system
again waits 8 seconds for acknowledgement. If still
no acknowledgement is received, the link is
considered failed, the D60:1 error is posted in the
event log, and the “Link Down” icon is displayed in
the UI. The system will continue to attempt to send
new messages, but will not resend queued
messages (patient, QC, and calibration data) until
the link has been re-established.
In addition, for networked systems, during initial
configuration of the communication link, the system
will wait 30 seconds for the remote system
(Rapidlink) to initiate a connection. If a connection
request is not received within 30 seconds, the
system repeats its initial configuration work and
again waits 30 seconds for the remote system to
initiate a connection. If a connection request is not
received within 30 seconds of the second attempt,
the link is considered failed, the D60:1 message is
posted in the event log, and the “Link Down” icon is
displayed in the UI.
Component Under Test Serial port, Ethernet port, External Data
Management System, Cables
D-codes D60:1 - LIS3 or LIS4 error
Download Errors
This test confirms that the ACEX logic IC and RTP are configured on startup.
Description The startup process must load a JAM file into the
ACEX logic IC to configure this chip. The RTP
program must also be loaded using the BDM port
on the 68332. Failure to accomplish these functions
causes these errors.
Component Under Test ACEX IC, RTP processor and memory, parallel
ports
Cal Overdue
No Paper in Printer
QC Material Expired
QC Missed
Miscellaneous Tests
This test monitors the occurrence of spurious interrupts detected by the 68332 hardware.
Description The 68332 microprocessor will occasionally (MTBF
greater than 30 days) detect an invalid interrupt.
Normally this means nothing. Logic is in place to
record spurious interrupts as to what interrupt level
it occurred at; and, if it was a level 4 interrupt (most
likely) to check the FIFO for data. If more than one
spurious interrupt occurs in an operation, the RTP
will initiate a controlled crash with error code
displayed in the LEDs.
Component Under Test System interrupt structor
D-codes D20
Description The waste subsystem has one input and two output
signals. Besides the sensor reading output, there is
an electronic error output. An input signal is
provided to calibrate the electronic error in the
subsystem. The calibration input is varied to reduce
the electronic error output. If this error cannot be
reduced below a minimum limit, an error code is
set.
Component Under Test Waste subsystem electronics
D-codes D23:6
AQC Valve
Description The AQC Valve MIT tests the integrity of the AQC
cartridge valve.
CO-ox Pump
Description This MIT allows the user to eject the AQC cartridge.
Component Under Test AutomaticQC cartridge, AutomaticQC cartridge
interface assembly, motor and motor mount,
stepper motor and valve drive assembly.
Test Criteria Tests whether or not the AQC cartridge can be
ejected. If the AQC cartridge fails to eject, a D32
error will be set.
D-codes D32:2 AQC Valve Error
Test Procedure 1. Select Cartridges from the Diagnostics screen.
2. User should open the AQC slide. This will
enable the Eject AQC Cartridge button.
3. Select Eject AQC Cartridge from the
Cartridges screen.
Note: The AQC cartridge is no longer valid after it is
ejected from the system.
A user may choose to run this MIT to observe the pumps while running the Sample Pump
MITs. The user needs to eject both cartridges to replace the reagent manifold, which is a
required yearly maintenance task.
Description This MIT allows the user to eject the Reagent
cartridge and the wash cartridge.
Eject R Cartridge
A user may chose to run this MIT to observe the pumps while running the Sample Pump
MITS.
Description This MIT allows the user to eject the Reagent
cartridge.
Component Under Test Loader motor, optical position detector
Eject W Cartridge
A user may chose to run this MIT to obseve the pumps while running the Sample Pump
MITS.
Component Under Test W Cart, valve drive assembly, all fluidic path
connections (including the CO-ox module if the
instrument is a 1245 or 1264), FD1, FD2, FD3
Test Criteria If excess vacuum is detected, the sequence is
aborted.
D-codes D23: 5
D23: 13
D23:14
D78 CO-ox Chamber: No RCX at FD3 Error
Test Procedure Touch the Load RCX button in the Fluidic
Functions screen within Diagnostics, and verify a
consistent flow rate of fluid.
Results Setting and clearing of D-codes as appropriate.
Lamp Cal
The CO-ox Calibration Lamp test verifies the proper operation of the CO-ox Tungsten
Halogen lamp and its associated power driver and control circuitry. The CO-ox Lamp
Calibration test may be used to perform a lamp calibration.
Description The “Lamp Cal” test turns the lamp on and initiates
a full CO-ox lamp calibration with embedded
diagnostics. A “Lamp Cal” will test the lamp and its
associated control circuitry, as well as automatically
determine the optimum control settings for the
lamp.
Component Under Test CO-ox Tungsten Halogen lamp and its associated
control electronics.
Test Criteria If one of the D-codes listed below is tripped, the
message “FAILED” is printed; otherwise, “PASSED”
is printed. Upon exiting this test menu the lamp
must always be turned OFF
D-codes D75 CO-ox Lamp Failure
D76:8 CO-ox Electronics error
Test Procedure Select the Lamp Cal button on the CO-ox screen.
All buttons are disabled and the instrument
performs the Lamp Calibration and Voltage Test. At
the end of the test, the lamp is turned OFF by the
system, the test status (passed/failed) is printed,
and the tHb icon is set to the Not-In-Cal state,
regardless of test outcome.
The tHb icon comes back into calibration the next
time a successful CO-ox zero is run.
Results Test results (PASSED or FAILED) are printed at the
completion of the CO-ox Lamp Test.
The CO-ox On/Off Lamp test verifies the proper operation of the CO-ox Tungsten Halogen
lamp, and its associated power driver and control circuitry. The CO-ox lamp test may be
used to directly control the lamp while problems with the illumination system are being
diagnosed.
Description The Lamp On/Off Test provides the user direct
control to turn the lamp ON and OFF using the most
recent lamp calibration settings to control the
intensity of the lamp.
Component Under Test CO-ox Tungsten Halogen lamp and its associated
control electronics
Test Criteria The user observes that the halogen lamp can be
turned on or off.
D-codes D75 CO-ox Lamp Failure
D76:8 CO-ox Electronics Error
Test Procedure Select CO-ox on the Diagnostics screen.
Last Cal
Leak Test
Measurement
Pinch Valve
Printer
R Valve
The Sample Chamber Test provides the user with the ability to move the sample chamber
(CO-ox optical sample chamber/cuvette) to the CLOSED and then OPEN position.
Description This MIT provides the user with the ability to test
the CO-ox sliding sample chamber and its
associated mechanical hardware and circuitry.
During run time, the sample chamber will open and
close to perform the most optimum wash of the CO-
ox optical chamber.
Component Under Test Sample chamber assembly, sample chamber
drive/interface and its associated electronics
Test Criteria An error flag indicating a sample chamber position
error shall be detected if the sample chamber fails
to move to the targeted position during this MIT.
Sample Pump
Temperature/pAtm
The Voltage and Temperatures test will read the UIP temperature and voltage values from
the ACPI chip, convert them to their proper resolutions, and then generate a report of the
values.
If any of the values are outside of their acceptable ranges, the appropriate D35:4x errors will
be set.
Description The test issues read/write commands to the I2C
bus in order to interact with the ACPI chip. This is
an electronic hardware component that monitors
typical computer voltages and temperatures.
Component Under Test Power supplies for +2.5 volts, 24 volts (heater), 3.1
volts, +5 volts, +12 volts, –12 volts, external
temperature sensor, and the internal chip
temperature
Test Criteria The values from the ACPI chip registers are read in
and converted into expected precision values based
on equations provided in the ACPI Chip Data Sheet
documentation provided by the vendor. If the limits
for any value are exceeded, the appropriate D-code
will be set. In all cases, a printout will be generated.
Wash Pump
Waste Detector
Test Criteria If the weights of an empty and full waste bottle are
within range, a D34 will be cleared. If either weight
is out of range, a D34 will be set.
D-codes D34 Waste Sensor Calibration Error
Test Procedure 1. Select Waste Detector on the Diagnostics
screen.
2. Follow instructions presented on the screen.
Results Errors are set or cleared appropriately. Future
reports of waste bottle percent full will be more
accurate.
The Wavelength Calibration Test provides the user with the ability to test the Wavelength
Calibration of the CO-oximeter.
Description The Wavelength Calibration Test checks neon lamp
operation, the sample cell’s optical path,
polychromator measurement functions, and
polychromator alignment.
Component Under Test Lamp control electronics, neon lamp, optical fiber
bundles, sample cell, and polychromator
Test Criteria The sample chamber must be in the closed
position, and the lamp turned on. Measures the
Neon spectrum, calculates the Neon line centroid
and compares it with alignment centroid of same
line stored in the polychromator to calculate
polychromator shift.
QC Troubleshooting
QC Temperature Correction
It has been reported that pCO2 and pO2 results are inversely affected by ambient
temperature. If your laboratory routinely operates in the 15 to 20 °C or 24 to 28 °C range, or
experiences seasonal variations in temperature, you can adjust the ranges to account for
these temperature effects as shown below.
For example, if your laboratory typically operates at 24 to 26 °C and the Level 2 pO2 range
in the Expected Values Chart is 92.0 to 108.0 mmHg, you can adjust the range by
subtracting 3.49 from the higher and lower limits as follows:
Adjusted range = (92.0 – 108.0 mmHg) -3.49 = 88.5 – 104.5 mmHg
Parameter/Level (mmHg)
15 to 17 °C 18 to 20 °C 21 to 23 °C 24 to 26 °C 27 to 28 °C
pCO2 level 1 2.32 1.16 0.00 –1.16 –2.32
pO2 level 1 11.75 5.88 0.00 –5.88 –11.75
pO2 level 2 6.97 3.49 0.00 –3.49 –6.97
Parameter/Level (kPa)
15 to 17 °C 18 to 20 °C 21 to 23 °C 24 to 26 °C 27 to 28 °C
pCO2 level 1 0.31 0.15 0.00 –0.15 –0.31
pO2 level 1 1.57 0.78 0.00 –0.78 –1.57
pO2 level 2 0.93 0.46 0.00 –0.46 –0.93
Part Numbers
Description Part Number BAN
Rapidlab 1240/1245 electrode test 02759061 02759061
simulator
Rapidlab 1260/1265 electrode test 02757409 02757409
simulator
Background Information
The Rapidlab 1200 electrode test simulator is a tool to help troubleshoot performance
problems. By providing simulated sensor output signals that the Rapidlab 1200 reads, the
electrode test simulator helps determine if the problem is associated with the sensors on the
system or with the interface hardware.
Note: For a Rapidlab 1240/1245, only the first 3 lines are valid; for a Rapidlab 1260/1265, all
11 lines are valid.
If a value falls outside the expected range, there is probably a fault with the contact
assembly or MUX PCB.
Sample Handling
See Customer Bulletin 18021, Handling and Storing Samples.
Note: Some spare parts are listed under their next higher assembly. Measurement Module,
for example, contains all subassemblies pertaining to it.
Bubbler
CO-ox Lamp
DC Power Harness
Door Spring
Caution:
Observe precautions for
handling electrostatic-
sensitive devices. Use an
antistatic mat and wrist
strap when handling PC
boards.
Perform Electronic Zero
Procedure
Hard Drive 40 GB 2.5 10327182 07899201 07899201
in.
Fan, Cooling
Filter Material
Grommet Kit
Illumination Housing
Measurement Module
Assembly 1265
Main Heater 1240 10327246 07923625 07923625
Illumination Assembly
1240/1245 10315496 01571719 01571719
Illumination Assembly
1260/1265 10315477 01566294 01566294
Mux Cable
Optical Switch
Peripump Motor
PM Kit
SMN BAN Item Number Additional Information
10312939 00192358 00192358
Polychromator Assembly
Power Cord, UK
SMN BAN Item Number Additional Information
10323838 06139440 00171416X
Pressure Sensor
Reagent Door
Top Door
Wash Door
Waste Assembly
Note: If the touch screen calibration dialog does not appear, restart the system again
and repeat steps 1 and 2.
CAUTION: Touch only the button that appears on the screen, and touch this
button only once each time it appears. Touching any other part of the screen or touching
the button more than once causes the screen to become inoperable. If the touch screen
becomes inoperable, repeat steps 1 through 4.
3. Touch the Touch Screen Calibration button.
The button disappears for 2 seconds, and then reappears in the lower right-hand area of
the screen.
4. Touch the Touch Screen Calibration button when it reappears.
Materials required
Circuit Descriptions
Introduction
The Rapidlab 1200 series systems electronic system consists of the following PC boards:
IMain/ETX Module UIP (User Interface Processor)
Display
Dual-opto
Preamplifier
Mux
Drive Interface
Rear Interface
RFID
IMain PC Board
The ETX UIP processor provides the ISA address, data, control, and power signals.
An RS232C port (COM1) routes data through the IMain board to the display interface
for communications to the touch screen.
A second RS232C port (COM2) routes LIS data through the IMain board to the rear
interconnect board.
A third RS232C port (COM3) routes barcode data from the IMain board to the Rear
Interconnect interface connector for communications to the external barcode reader.
The flat panel interface routes the active LCD display signals from the IMain board
assembly’s UIP processor to the display module interface.
The Utility interface provides UIP reset and speaker functions to IMain board.
The diskette interface provides diskette signals to the drive Interface from the IMain
board.
The IDE drive interface provides IDE drive signals to the drive interface from the
IMain board.
The drive interface provides IDE and diskette signals to the drive module from the
IMain board and provides the Drive Module with protected power to operate the
diskette, hard drive, and CD-ROM drive.
The Display/Printer board Interface provides power and ground signals for the
thermal printer, LCD display, and LCD backlight. It also provides signals from the UIP
COM1 interface, the UIP LPT2 interface, and the UIP flat panel interface.
The rear interconnect board interface provides COM2, COM3 and the USB bus
signals to the Rear Interconnect board from the Main board.
UIP I2C Communications circuitry provides a bi-directional serial bus controller.
Allows the user interface processor (UIP) and the RTP 68332 processor to
communicate with each other.
Bidirectional 2 kilobyte FIFO
Provides Status flags for each FIFO memory
Provides a reset function for the FIFO memory.
The 68332 processor circuitry generates address, data, and control signals for the
RTP bus and the reset circuitry to restart the system.
RAM circuitry provides volatile memory for the storage of 2.0 MB of data memory.
Provides signals to, and processes output signals from, the Multiplexer board.
Processes inputs from the three thermistors, the measurement block, the control
multiplexer outputs, the fluid detector outputs, and the system reference and DAC
signals.
Steers the heater DAC outputs, the heater DAC reference, the microstepper circuit
reference voltages, and the lamp voltage, current and error voltage to the multiplexer.
Provides DAC circuitry with channels for the CO-ox lamp, the heater control, the fluid
detector, stepping motor current reference circuits and for creating a bridge reference
voltage for the pressure transducer.
Provides ADC circuitry process multiplexer circuit signals.
Compares fluid detector to reference sources.
Control Circuitry
CO-oximeter Circuitry
The lamp servo drive circuit drives the halogen lamp and transmits the light through
the optical fibers and controls the on/off cycle.
The calibration lamp circuit drives the neon lamp.
The Polychromator board interface allows the Main board to control and
communicate with the Polychromator board assembly.
Drives the luer valve motor, the wash motor, and the sample motor.
Communicates with the RFID memory device in the reagent cartridge.
Drives the cartridge loader motor and the cartridge ejector motor.
Communicates with the RFID memory device in the wash cartridge.
The System Status circuit monitors the power supplies, the stepper motor current. It
also Monitors the RTP processor and controls the reset state.
The Power System Input and Protection circuit protects the power supplies from
overvoltage.
The System Input Interface connects the system power supply to the IMain board
and supplies power to a cooling fan near the power supply.
Display PC Board
The Display PC board converts signals from the IMain board to operate the following
components:
Thermal printer
Touch screen
Liquid crystal display (LCD)
Speaker
In addition, the display board provides the following electrical functions:
Circuitry to provide diagnostic capability
Conversion of ASCII data from the host to actual printed reports
Detection, location, measurement, and reporting of inputs from the touch screen
Generation of the voltage necessary to turn the LCD backlight on and off
Dual-opto PC Board
The Dual-opto PC boards provide the luer position mechanism a way to determine where the
valve actuator is located for the measurement and AutomaticQC valves. Two optical
interrupter switches accomplish this task.
Multiplexer PC Board
The mux board provides all requirements for measurement module functions. The Premux
board selects amplified electrode signals from the Preamplifier (Preamp) module, adds a
programmable offset and gain, and sends this signal to the Data Acquisition System (DAS)
for digitizing. The mux board provides the following functions:
Diagnostic circuitry
Stable bias voltage for the pO2 sensor
Regulated power supply for the Preamp module
The Preamp Board amplifies signals from the sensors before they are processed by the rest
of the system. The Preamp PC module provides the following functions:
Diagnostic circuitry
Bias connection for the pO2 sensor
Over voltage and current protection on all inputs
High impedance differential amplifier for the pCO2 sensor
Low bias current transimpedance amplifier for the pO2 sensor
Signal connection to the sample ground and temperature thermistor
High impedance differential amplifiers for the pH, K+, Ca++, Cl–, and Na+ sensors
Low bias current transimpedance output amplifiers and bias amplifier for Glucose
sensor
Low bias current transimpedance output amplifiers and bias amplifier for Lactate
sensor
The ETX board is a VIA 1000 MHz microprocessor with the following components:
Via PN 133T chipset, providing support for
Ì L2 cache
Ì DRAM
Ì CPU-to-PCI bridge
Ì IDE controller with primary channel routed to 2.5 inch hard drive interface;
secondary channel routed to IDE CD ROM devices
Ì Interrupt controller
Ì Timer/counters
Ì DMA controller
Ì Real-time clock (RTC)
Miscellaneous Functions
Peripheral Interfaces
Serial Ports
Parallel Port
RFID Coupler
RFID Tags are installed in every Reagent and Wash cartridge to store reagent assigned
values and cartridge identification. The RFID board acts as a transponder to read/write to the
RFID Tags.
Neon Assembly
The neon board assembly contains a neon gas discharge lamp, lenses, a narrow bandpass
interference filter, and a driver board.
The output light (monochromatic) from this assembly is used as a reference to check the
wavelength measurement accuracy of the polychromator. This check occurs at the same
time and frequency as the automatic tHb zero. Any optical or mechanical misalignment in the
polychromator, which may cause wavelength measurement inaccuracies, are corrected
during the zero calibration.
Functional Descriptions
Introduction
Rapidlab 1200 Series systems are intended for laboratory testing of blood gases,
electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous,
and capillary whole blood samples.
System Parameters
Fluid Flow
After aspiration, the sample moves through the sample tubing to the measurement module
where it is analyzed by the sensors. Waste moves from the measurement module to the
waste bottle through the waste tubing.
On the 1245 and 1265 systems, the CO-ox sample moves through the CO-ox sample tubing
to the sample chamber for analysis, and exits as waste through the pump tubing and reagent
manifold tubing.
Reagent/Wash Assembly
The reagent/wash manifold consists of reagent cartridge, wash cartridge, tubing, and check
valves.
The reagent cartridge is loaded into the reagent door, and the wash cartridge is loaded into
the wash door.
Reagent and Wash Doors
The reagent and wash doors have the following components:
Reagent cartridge door assembly
Reagent door
Reagent door hinge
Door latch
Wash cartridge door assembly
Wash door
Wash door hinge
Door latch wash
The reagent and wash cartridges contain foil bags of reagents and the mechanical
components to interface the cartridge to the system.
Cartridges are self-contained and non-serviceable. The use life of a cartridge is independent
of sample volume. The system prompts you when to replace the cartridge.
When a cartridge is installed and locked into place, pierce pins puncture fittings in each bag
and create a fluid path. The pierce pins, which are part of the interface assembly, create a
seal between the pierce probes and the fitting on each bag.
The platen area has four tubes. The two tubes on the right serve as the flow paths for foam
wash, and connect the reagent cartridge to the wash cartridge. The two tubes on the left
serve as the flow paths for KCl, sample, and reagents.
Reagent Cartridge
The reagent cartridge contains Zero and 200 reagents described in the table below.
Electrolytes, pH, glucose, and gases in the reagents are NIST traceable.
Reagent Ingredients Volume
Sample Entry
The sample entry components are part of the reagent cartridge, and perform the following
functions:
Provide an insertion point for syringes and capillary tubes
Hold syringes, capillary tubes, or aspiration adapters during sample aspiration
The sample entry port components consist of a luer and capillary seal. The capillary seal has
two passages that provide pathways for the wash and waste fluids.
The sample port attaches to the sliding valve assembly by means of two snaps. During
aspiration from a capillary tube, the capillary seal provides a seal around the sample probe.
The sample port is attached to the sliding valve, which moves to different positions to
perform the following functions:
Aspirates samples from capillary tubes and syringes
Aspirates the wash fluid
Aspirates QC samples
Selects and aspirates reagents
The sliding valve has nine positions, each of which allows fluid, air, or sample to be pumped
through a thin, stainless-steel probe in the center of the valve. The probe remains stationary
while the valve moves back and forth around the probe to change the hydraulic path of the
system.
The following table shows the valve positions and their functions:
Position Name Function
SYRINGE_POS moves the valve to place the probe tip into the syringe, which
creates the fluid path for sample aspiration
Syringe Aspiration aspirates syringe or QC samples
Vent provides an opening to advance syringe samples while the device
is in place
Capillary Aspiration aspirates samples from capillary tubes; uses the front hole in the
probe
Vent AQC controls the aspiration of an external QC sample; uses the front
hole of the probe
Zero Reagent Load delivers the zero calibration reagent to the measurement module;
uses the rear hole in the probe
200 Reagent Load delivers the 200 calibration reagent to the measurement module;
uses the rear hole in the probe
Wash Reagent Load delivers steady wash to the measurement module; uses the rear
hole in the probe
RCx Reagent Load delivers the RCx calibration reagent to the measurement module;
uses the rear hole in the probe
Luer Wash empties the luer area from a port in the bottom of the sample entry;
uses the front hole of the probe
The sample pump compresses the tubing on the cartridge to generate flow. Peristaltic action,
created by the pinching of tubing segments in two places by adjacent rollers, pulls fluid
through the tubing.
Reagent Manifold
The fluid connectors on the reagent manifold engage with the fluid connectors on the
cartridges to create the fluid path for reagents, samples, and waste.
Measurement Module
The measurement module contains sensors, fluid detectors, and mechanical components
that interface the sensors to the system.
Thermal Block
The thermal block maintains heat stability, and provides proper alignment and fit of the
sensors to provide an optimal flow path to minimize fluid residue.
Preamplifier Assembly
The preamplifier assembly contains the sensor contacts and the preamplifier electronics.
Sensor contacts are electrically isolated from one another to provide acceptable contact-to-
contact impedance. The preamplifier is thermally controlled to minimize electronic drift due to
ambient temperature fluctuations.
Fluid Detectors
Two optical sensors are used as fluid detectors. Fluid detector 1 (FD1) is located in front of
the first sensor. Fluid detector 2 (FD2) is located after the last sensor. The system uses
information from the sensors to control the movement and positioning of fluids during the
system sequences.
The main block heater controls the temperature for the sensors and the preamplifier board.
The control temperature of the heater is determined by the sample ground/temperature
sensor. The main block, preheater, and heater use Fuzzy Logic to control the temperature.
The system temperatures are shown in the table below.
Device Range
Sample ground/temperature sensor 37.0 ±0.15 °C
Preheater 37.6 ±0.30 °C
Door 36.5 ±0.50 °C
Main block heater Adjusts dynamically to adapt to
the environment and maintains
the sample temperature
Preheater
The preheater, controlled at 37.6 ±0.3 °C, heats fluid entering the measurement module prior
to positioning within the sample block. The fluid temperature must be 37.0 ±0.15 °C before
measurement.
The door contains a clear lens and a heater. A door-open detector is located in the
measurement module. A clear lens magnifies the sample path for greater visibility. The door
heater limits the heat loss from the block and minimizes the effects of room temperature
fluctuations.
Illumination Assembly
The sensors are lit with a series of white LEDs located on the illumination board assembly.
Sensors
Electrodes used for measurement in the RapidLab 1200 systems are called sensors The
measurement technology is based on electrochemical, biochemical, and optical physics. The
sensors detect analytes present in the sample and form the sample path.The sensors have
the following characteristics: a molecular or ion-specific recognition mechanism, a transducer
mechanism, and a signal processor system.
The molecular recognition mechanism gives a sensor its identity. Each sensor is designed to
selectively measure the activity of a specific substance. Although many elements in a
sample may interact with a sensor, the sensor is highly selective for one substance over
others. The common recognition mechanism used in many RapidLab 1200 sensors is a
membrane designed to be selective for a specific substance.
The transducer mechanism converts the potential generated by the molecular recognition
mechanism to an electrical signal. In the RapidLab 1200 systems, this is accomplished
through potentiometry or amperometry. Potentiometry is the technology that measures the
difference in potential (voltage) between two electrodes (the molecular recognition
mechanism) in a solution without applied current. Amperometry is a technique that involves
applying voltage between an anode and a cathode, and then measuring the current
generated as the measured analyte is reduced at the cathode.
The signal processing system conditions the electronic signal from the sensor, through
electronic smoothing and noise filtering. The electronic signals are then converted into a
concentration expressed in recognizable units of measurement.
Electrochemistry involves the measurement of current or voltage occurring in an
electrochemical cell. The cell consists of two or more electrodes that interact with a chemical
in solution and are connected to an electrical system. Many of the sensors, like the pH
sensor, are designed with ion-selective electrode (ISE) technology. An ISE contains a
specially designed membrane that provides sensor selectivity. Selectivity is the ability of the
sensor to interact with a specific ion in solution. The membrane separates an inner,
reference element, which is immersed in a fixed electrolyte solution, from the sample.
During analysis, a membrane potential develops as a result of the interaction of the analyte
(ion) at the membrane. The membrane potential is related to the amount of substance being
measured in the sample. The half-cell potential in the sensor consists of the inner reference
element potential plus the membrane potential.
Sample Ground/Temperature Sensor
The sample ground/temperature sensor provides electrical grounding for stable sensor
readings and also measures the sample temperature. The sample ground/temperature
sensor is divided into two parts: the ground junction and the thermistor (temperature sensor).
These two parts are housed in the same block.
The ground junction is formed by a spring loaded contact and a sample path tube. The
ground references the fluid in the sample path tube to the analog ground, which eliminates a
possible source of variation in the sensor output signal.
A permeable cellulose membrane separates the KCl solution from the sample and provides
the ionic conduction between the KCl solution and the sample. The membrane completes
the conductive path to the sample from the fixed half-cell potential that is required for the
measurement.
The Ag wire conducts the half-cell potential of the reference sensor to the measurement
device where it is compared to the potential of the measuring sensor. The potential
difference measured reflects the concentration of analyte in the sample. Although the
reference sensor provides a constant potential from sample to sample, the potential
difference measured between sensors varies with each sample.
The potassium sensor contains a silver/ silver chloride wire surrounded by an electrolyte
solution that has a fixed concentration of potassium ions. The membrane consists of the
ionophore valinomycin immobilized in a plasticized PVC (polyvinyl chloride) matrix. This
membrane separates the electrolyte solution from the sample. Valinomycin is a neutral ion
carrier that is highly selective for potassium ions over other clinically encountered cations.
The chloride sensor contains a silver/silver chloride wire surrounded by an electrolyte
solution that has a fixed concentration of chloride ions. The membrane is a derivitized
quaternary ammonium compound that is immobilized in a polymer matrix. This membrane
acts as an ion exchanger with a high selectivity for chloride ions over other ions present in
the sample, and separates the electrolyte solution from the sample.
The calcium sensor contains a silver and silver chloride wire surrounded by an electrolyte
solution that has a fixed concentration of calcium ions. The membrane, which consists of an
ionophore imbedded in a polyvinyl chloride membrane, separates the electrolyte solution
from the sample. The ionophore is a compound that is highly selective for calcium ions over
other ions.
Waste Assembly
After analysis is completed, the waste assembly accepts waste reagents and samples
pumped from the fluidic components.
The waste sensor is covered by a membrane. The weight of the waste bottle pushes the
membrane down and drives a load cell to indicate the volume of fluid in the waste bottle.
Display Assembly
The display tilts on a horizontal axis and locks at the chosen tilt. The display support holds
the display and the pawl provides stability through the stepped mechanism when the display
is set to the varying tilt positions.
The display board contains all the circuitry necessary to drive and control the LCD, the
printer, and the touch screen. There are no service settings or jumpers on the display board.
The display board communicates with the main board.
The display module contains the LCD, the touch screen, the printer, the display board, and
display support. In addition the display module contains roll of printer paper and the paper
cover.
The LCD is a 10.4-inch display with active matrix color LCD that provides 65536 colors. The
touch screen is overlaid on the LCD. The touch screen consists of two electrically resistive
sheets that are superimposed over the viewing area of the display. The sheets are separated
from each other by a small air gap. When the screen is touched, the two sheets are
electrically shorted. The resolution of the touch screen is determined by control circuitry.
The printer provides reports for samples, calibrations, and diagnostics. The printer consists
of a platen, a motor and gear mechanism, a thermal printhead, and two sensors. One sensor
detects when the printhead is up and the other sensor detects when the printer is out of
paper. The printer is controlled by the display board. The printer cable carries the data
signals, motor control signals, the paper-empty sensor signal, and the printhead-up sensor
signal.
Interface Wall
Front Wall
Pumps
The pumps move the reagents, samples, and waste through the system. The sample pump
moves samples and reagents through the system. The wash pump moves waste to the
waste bottle.
Each pump consists of a motor, pump housing, motor mount, spring plate, and springs. The
sample pump generates flow by compressing the tubing against the fixed platen of the
cartridge. As the individual rollers rotate through the platen area, a segment of the tubing is
pinched in two places by adjacent rollers. The fluid is pushed through the tubing by the
peristaltic action of the moving rollers.
Cartridge Interface
The cartridge interface frame attaches the reagent cartridge to the system and consists of
the following components:
Interface frame
Location pins
Pierce pins
Valve actuator
Thermal cover
Camshaft
Roller cage
Dual-opto PCB assembly
Valve drive motor assembly
Interface frame slide assembly
The interface frame holds all devices and parts necessary to interface the system with the
reagent cartridge.
Locator pins position the cartridge on the cartridge interface frame. Three bag piercing pins
are below the locator pins. The pierce pins force the pierce probes to puncture the reagent
bags and thereby create the fluid path. The pierce probes engage with the fitting on the
reagent bags to make a seal.
Four cable clamps secure the cables that extend from the connector block, the stepper
motor, and the optical detector board to the main board.
The valve actuator moves the sliding valve. The luer mount extension on the sliding valve
connects with the valve actuator when the reagent cartridge locks on the cartridge interface
module. During cartridge initialization, the actuator moves back and forth until it automatically
engages the luer mount extension. A stepper motor, which has nine positions and left and
right limit stops, controls the actuator.
The bidirectional valve actuator motor and the position detectors, which are part of the
optical detector board, move and position the valve. Two sensors in conjunction with the
position detectors determine valve position.
The roller cage mounts on the motor shaft. The roller cage is designed to manually push on
and pull off of the motor shaft for easy cleaning of the platen. The extended cap on the
sample pump roller cage allows the user to grasp the pump and turn it during manual
operation.
AutomaticQC Interface
The AutomaticQC interface holds all devices and parts necessary to interface the system
with the AutomaticQC cartridge. The AutomaticQC interface consists of the following
assemblies:
AQC frame assembly
AQC latch assembly
Removal of an access panel provides access to the latch assembly where the AutomaticQC
cartridge connects to the system. A sliding metal latch plate in the latch assembly engages
four standoffs on the AutomaticQC cartridge to secure the cartridge to the system. A valve
actuator that extends through the latch assembly moves the sliding valve on the
AutomaticQC cartridge. The actuator moves until it engages a vertical slot in the sliding
valve. The bidirectional stepper motor and position detectors move the valve along the probe
to select the required fluid path for flow of QC material from the AutomaticQC cartridge to the
reagent cartridge. The stepper motor has five stop positions.
With the AutomaticQC cartridge engaged to the system, the cartridge connector located on
the front of the AutomaticQC cartridge connects the AutomaticQC cartridge to the reagent
cartridge. Fluidic connectivity between the two cartridges is provided by a single 0.5 mm ID
tube that passes from the base of the sliding valve in the AutomaticQC cartridge through the
cartridge connector and into the reagent cartridge. The cartridge connector must be closed
for the system to operate.
Automatic QC Cartridge
The AutomaticQC cartridge supplies sufficient QC material to perform at least one sample
analysis of each level, three times per day for the life of the cartridge . Analysis of the
AutomaticQC levels is done at pre-programmed intervals, which are defined as part of
system setup.
The AutomaticQC cartridge contains three levels of quality control material uniquely
formulated to provide verification of performance at several points in the clinical range of the
Rapidlab 1200 system.
Level Volume Contents
1 75 mL Buffered bicarbonate solution with Na+, K+, Ca++, Cl-, carbon
dioxide, oxygen, nitrogen, dye, glucose, surfactant, and
preservative
2 115 mL Buffered bicarbonate solution with Na+, K+, Ca++, Cl-, carbon
dioxide, oxygen, nitrogen, dye, glucose, surfactant, and
preservative
3 155 mL Buffered bicarbonate solution with Na+, K+, Ca++,
-
Cl , carbon dioxide, oxygen, nitrogen, dye, glucose, surfactant,
and preservative
The AutomaticQC cartridge is mounted on the right-hand side of the system where a support
bracket secures the cartridge in place. An access panel must be removed before the
cartridge is installed. Four standoffs attach the cartridge to the system.
Fluidic Components
The AutomaticQC cartridge consists of the following components:
Valve assembly
QC material
Tubing
Cartridge lever
Pierce pins
Cartridge connector
The valve assembly consists of a sliding valve, valve seal, valve lubricant, and AutomaticQC
probe.
The pierce probes engage fittings on the reagent bags to provide a seal and to puncture the
bags to create fluid paths for the QC material to flow from the AutomaticQC cartridge to the
reagent cartridge. A total of three pierce probes service the five bags. The AutomaticQC
cartridge is shipped with the lever in its open position.
The cartridge connector is mounted on the front of the AutomaticQC cartridge. During
installation of the cartridge, the connector is moved horizontally and connects with the
reagent cartridge. The connector provides the sample path from the AutomaticQC cartridge
to the reagent cartridge. The cartridge connector slide must be fully engaged for the
AutomaticQC cartridge to function.
CO-oximeter
The RapidLab 1245 and 1265 systems contain a CO-oximeter (CO-ox) that measures the
concentration of total hemoglobin and hemoglobin fractions in whole blood samples.
The CO-ox assembly analyzes samples for the following parameters:
tHb
FO2Hb
FCOHb
The drive wheel is attached to the drive shaft of the motor, and engages a mechanism in the
measurement cartridge that moves the sample chamber to the open and closed positions.
The fibers at the end of the fiber bundle form a line that defines the input slit to the
polychromator. The first mirror collimates light from the slit so that nearly parallel light falls on
the diffraction grating. The diffraction grating disperses the light coming from the collimating
mirror into its component wavelengths. The second mirror images the light from the grating
onto the photodiode detector array where the various wavelength intensities are measured.
The diode array board converts light into an electrical signal. The polychromator board
converts the signal from the diode array board from an analog to a digital value for further
processing.
The pump generates flow by compressing the tubing against a fixed platen. As the individual
rollers rotate through the platen area, a segment of the tubing is pinched off in two places by
two adjacent rollers. The fluid in the segment of tubing is pushed ahead through the tubing
by the peristaltic action of the moving rollers. The length of tubing is minimized on the CO-ox
pump to keep the fluid path volume to a minimum.
Proper operation of the pump requires that the gap between the fixed platen and the rollers
be between 1.47 mm (0.058 inches) and 1.91 mm (0.075 inches). Flow rates, tubing life, and
motor life can be adversely affected by gaps outside of this range.
The CO-ox pump is designed to provide a flow rate of 5 µL/second (±10%) during slow
aspirate speed. This ensures that the 50 µL portion of the sample is moved through the
system at a speed that does not cause performance errors, such as incomplete hemolysis or
excess separation. There is no automatic pump speed calibration. Pump speed is fixed at 3
revolutions per minute (rpms), which is equivalent to 0.05 revolutions per second (rps).
The housing contains optical filters that limit the lamp's output light (spectrum) to the visible
range, and lenses that collect and focus light into the fiber bundle.
The output intensity of the lamp is controlled by a closed-loop optical feedback system. This
is accomplished by routing some of the output fibers from the fiber bundle, which is
connected to the illumination housing assembly, to a photo-feedback sensor located on the
IMain PCB board. This sensor continuously monitors lamp output and generates a signal
that regulates power to the lamp.