Flow Calibration with Adjustment


20847277-2644915
¸
53207330 FCP-7.2
Purchase order number Calibration rig

CL-844427-50 / Endress+Hauser Flowtec AG 981.748 l/s ( 100%)


Order Nº/Manufacturer Calibrated full scale

53P5H-ER1B1BA0BAAG Service interface


Order code Calibrated output

PROMAG 53 P 20" 1.2926


Transmitter/Sensor Calibration factor

EA072419000 2
Serial Nº Zero point

8400-FIT-85101A 22.5 °C
Tag Nº Water temperature

Flow Flow Duration V target V meas. o.r.* Measured error % o.r.


[%] [l/s] [s] [l] [l] [%] Tolerance limit: ±0.2% o.r.* ± Zero stability

40.2 394.4 70.1 27657.0 27656.0 0.00 1.5

40.2 394.4 70.1 27644.3 27645.5 0.00


1
100.3 984.3 70.1 69036.9 69117.7 0.12
- - - - - - 0.5

- - - - - - 0
- - - - - -
- - - - - - -0.5

- - - - - - -1
- - - - - -
- - - - - - -1.5

*o.r.: of rate 0 10 20 30 40 50 60 70 80 90 100 [%]


Flow

For detailed data concerning output specifications of the unit under test, see Technical Information (TI), chapter Performance characteristics.
Traceability to the national standard for all test instruments used for the calibration is guaranteed.
Endress+Hauser Flowtec operates ISO/IEC 17025 accredited calibration facilities in Reinach (CH), Cernay (FR), Greenwood (USA),
Aurangabad (IN) and Suzhou (CN).