You are on page 1of 4

Identification of DC corona generating SF6

decomposition gases adsorbed on CNT gas sensor


using FTIR Spectroscopy
Yul Martin, Zhenyu Li, Takuya Tsutsumi, Junya Kiminobu Imasaka
Suehiro Dept. of Electrical Engineering
Dept. of Electrical and Electronic Engineering Kyushu Sangyo University
Kyushu University Fukuoka, Japan
Fukuoka, Japan
suehiro@ees.kyushu-u.ac.jp
Shinya Ohtsuka
Dept. of Electrical Eng. and Electronics
Kyushu Institute of Technology
Fukuoka, Japan

Abstract—It is important to detect decomposition of insulating gas emission (light) had been developed [2]. Moreover, the gas
SF6 caused by partial discharges for gas-insulated switchgear (GIS). sampling method detecting decomposition gas products, has
In the previous result, the author developed a novel partial discharges been also regarded as effective technique. Up to now,
sensing using carbon nanotube (CNT) gas sensor. In recently, an decomposition products were successfully detecting using a
application of high voltage DC for long distance electric power
detection tube, a gas chromatography, a mass spectrometer
transmission has been often used in various countries as the
interconnection system and overload demand. In this study we and an ion mobility spectrometer [6]. The detection tube is
applied the CNT gas sensor to detect partial discharges generated by handy and can detect main decomposition products such as
high voltage DC. There was polarity profile for DC compared AC sulfur dioxide (SO2) or hydrogen fluoride (HF) at ppm-levels.
discharges. It was found, response of the sensor in the case of However, the detection tube suffers from poor accuracy and is
negative discharges was much higher than positive discharges under not suitable for automated remote monitoring. On the other
the same experimental condition. And the decomposed gases for each hand, the gas chromatography and the spectrometers with high
case were identified by FT-IR spectroscopy. The result indicated sensitivity are costly and suitable for laboratory-based analysis
absorption peaks correspond to SF4 and SOF2 of positive discharges rather than on-line monitoring. It is essential to develop a new
was much higher than that of negative discharges.
type of gas sensor, which is easy to use on-site and sensitive
enough to detect slight PD in SF6.
Keywords-DC discharges;partial discharges;SF6decomposition
gases;gas insulated switchgear; carbon nanotube; gas sensor It is well known that SF6 decomposition gases under
electrical discharge and overheating are consists of SF4, F2,
I. INTRODUCTION SOF2, SO2F2, SOF4, SO2, and HF as a result produced of SF6
decomposition and the reactions with water vapor and oxygen
From1960, a high voltage direct current (HVDC) has been under discharges condition [3]. To date, several kinds of gas
recognized to be used for long distance electric power sensor has been proposed to detect SF6 decomposition gases
transmission caused have had advantageous to support the [3-4]. For example, ion-conductive solid electrolyte, which
growth of renewable energy and integration into electricity could cause an electrochemical reaction with HF to generate
grids with the higher capacity and efficiency [1]. Gas insulated ion current, has been successfully used as a SF6 decomposition
Switchgear (GIS) for HVDC transmission filled with sulfur gas sensor [3]. However, it is be used to detect only arc
hexafluoride (SF6) as insulation and arc-quenching medium, discharge and local overheating in GIS.
are widely used because of high reliability and small size.
However, the partial discharges (PD) caused by small In the previous study [6], the authors proposed a new
protrusion, void, or impurities decompose SF6 resulting detection method of PD generated in SF6 gas using a gas
deterioration of insulation. A sensitive PD measurement is sensor composed of carbon nanotubes (CNTs). Recently,
necessary for evaluation of the insulation condition. Early interest in CNTs has been rapidly growing in various scientific
detection of PD can provide a way to predict and avoid the and engineering fields. Especially, CNT-based gas sensors [6]
fatal failure. have received considerable attention because of their
outstanding properties such as faster response,higher
For the PD measuring of physical effect as well as ultra sensitivity, lower operating temperature and a wider variety of
high frequency (UHF), acoustic emission (sound), optical

1
978-1-4244-6890-4/10$26.00 ©2010 IEEE 663 TENCON 2010
detectable gas. The authors have proposed a new fabrication In order to analyze SF6 decomposed products generated
method of a gas sensor composed of multi-wall carbon HVDC partial discharges adsorbed on CNT of the gas sensor
nanotubes (MWCNTs) using dielectrophoresis (DEP) [7]. and KBr card comprising CNT was placed same position of
the CNT gas sensor in tank. HVDC was applied to needle to
In this study, it was investigated that the sensor response to generated PD for 4,8 or 18 hours. After the PD exposure, the
decomposition gas SF6 generated HVDC PD and component Kbr card was used for taking FT-IR spectra. Within 30
of the gas by FT-IR spectroscopy. The FT-IR spectra of SF6 minutes after the PD exposure. FT-IR analysis was done using
decomposition gas were taken from a KBr card comprising FTIR-620 (Jasco Corporation, Japan) within 30 minutes after
CNT. The card was exposed by the positive and the negative the PD exposure. All our experimental were carried out at
partial discharges. At that time, several components of the atmosphere pressure (0.1 MPa SF6) and ambient temperature.
decomposed gas were adsorbed on CNT in Kbr card. We
considered the adsorbed gas causes the sensor response. FT-IR The decomposition of SF6 by positive and negative corona
absorbance peaks were observed around 735 cm-1 and 480 discharges has been investigated using FT-IR spectroscopy.
cm—1. Based on this result, SF4 and SOF2, which were Before the FT-IR analysis of PD decomposition, the FT-IR
attributed responding the sensor after PD generated. In our spectra measured using sample placed in ambient air or SF6
case, absorbance peaks of positive were larger than that of gas for hours in order to obtain background sognal.
negative. Additionally, it was discussed the sensor response of
negative partial discharges was higher than that of positive..
III. EXPERIMENTAL RESULT
Figure 2 shows FTIR spectra at various measurement with
II. EXPERIMENTAL METHOD KBr card only and Kbr Card including CNT gas sensor at
The objective of the experiment was to investigate negative and positive polarity applied voltage U = + 15 kV or
adsorbed components of the SF6 decomposition gases using -15 kV for 4 hours. This result suggests that KBr card only did
FT-IR spectroscopy and comparing with the measurement by not detection by FT-IR, which that needed uses CNT as
passive detector tubes. The schematic diagram of experimental adsorbed gas molecules. Remarkable absorbed peak were
setup for PD detection is shown in Fig.1 [9]. The experimental appeared around of 735 cm-1and 480 cm-1. The absorbance at
setup fundamentally consisted of an high voltage DC power 735 cm-1 and 480 cm-1 was not observed without the PD and
source, a model stainless steel cylinder tank (50 cm inner increased with PD duration (Fig.4 and 5). Comparing the
diameter, 90 cm long), an electrode system (point-to-plane absorbance peaks between negative and positive PD polarities,
electrode system made of stainless steel with 50 µm point the peaks of absorbance of the positive PD much larger were
radius and 10 mm gap length), CNT sensors, and an than that negative.
impedance measuring system (a lock-in amplifier controlled
0.2
by a PC). The electrode system was vertically placed in the
tank, which was filled with fresh SF6 gas at atmospheric Negative PD
pressure before each test run. The CNT sensors were placed in
the stainless steel cylinder tank and in the same plane as the 0.15
plane electrode. The fabrication method of CNT gas sensor
has been described in detail in the previous paper [8]. The
distance x of the sensor from electrode was 10 cm along the Positive PD
Absorbance

0.1
horizontal axis of the stainless steel cylinder tank.The
impedance change of the sensor response was continuously
measured at room temperature and 100 kHz. DC high voltage KBr card with CNT
at +15 kV and -15 kV was applied to the point electrode to 0.05
generate corona discharge (simulated PD). The PD signals
were detected using a shunt resistance of 50Ω inserted KBr card only
between the plane electrodes to ground. 0
KBr card
KBr CNT
Positive
Negative
-0.05
1600 1200 800 400

Wavenumber (cm-1)

Fig.2 FTIR spectra of SWCNT with the KBr card without


CNT and with CNT gas sensor.
Fig.1 Experimental setup of PD exposure of CNT gas sensor

2
664
Figure 3 summarizes previously reported FT-IR spectra of appeared at 735 cm-1 and 480 cm-1. According to some
SF6 and typical SF6 decomposition gas species, which can be literatures, this absorbance peaks correspond to SF4 or SOF2
adsorbed on CNT gas sensor [7]. These data were used as that of decomposition by product as shows at Fig.3 [11] This
reference for identifying the FT-IR spectra obtained in this result also the same with our previous study using high voltage
study. Figure 4 depict that the FT-IR spectra obtained under AC [10].
negative PD generated in SF6 for 4,8 and 18 hour, Comparing the absorbance peaks between the negative and
respectively. Remarkable absorbed peaks were appeared the positive PD polarities. The peaks of positive were higher
around of 735 cm-1 and 480 cm-1. The absorbance at 735 cm-1 than that of the negative The absorbance peaks values of the
and 480 cm-1 was not observed without the PD duration. negative were smaller as about 0.15 and 0.1 and 480 cm-1
rather than that of the positive. Both of the value of absorbance
peaks increasing as gas molecule can adsorbed the
decomposition gases on CNT gas sensor surface depend on
discharges energy at duration time PD exposure. In this case
positive DC polarity faster reaches peak compared with the
negative DC polarity. However, we compare with the result of
response of CNT gas sensor [8], which the author had been
research for detecting SF6 decomposition generated by HVDC.
As long as that the data of negative DC polarity much higher
than positive in the same experimental condition.

PDC-18h
PDC-8h
2.5 PDC-4h
No PD

Figure 3 Summary of adsorption bands of main SF6


decomposition by product and interference gas. 2
Absorbance

0.6 1.5

NDC 18h
NDC 8h
0.5 NDC 4h 1
No PD

0.5
0.4
Absorbance

0
0.3
900 800 700 600 500 400

Wavenumber (cm-1)
0.2

Fig.5 FTIR spectra of CNT gas sensor exposed to PD


0.1
generated by HVDC under positive voltage polarity.

0 IV. DISCUSSION
900 800 700 600 500 400
A. Identification and determines of SF6 decomposition product
Wavenumber (cm-1) by FTIR.
When SF6 is in environment of electrical discharges,
including arc or glow, spark, and partial discharges (corona
Figure 4 FT-IR spectra of CNT gas sensor exposed to PD discharges) will dissociate into lower order sulphur fluorides,
generated by HVDC under negative voltage polarity. sulphur and fluorine. It is known that major primary
In this result, absorbance peak after occurred the SF6 decomposition products such as SF5,SF4, and SF2 are produced
decomposition gases increased with the PD application time. It in corona discharges by electron impact as follows [5] :
means amounts of decomposition product tend to increase with SFx + e-  SFx-1 + F + e-1 (6≥x≥1) (1)
increasing time of exposure. Figure 5 shows the FT-IR spectra
obtained under PD generated in SF6 for the duration time with The stain less steel chamber of high voltage system equipment
positive voltage polarities. The absorbance peaks tends is usually not a contamination-free environment. Some of the

3
665
dissociation of SF6 will react with contaminant such as water higher level as well as 11 ppm level but SO2 more reactive
vapour (H2O) or oxygen (O2) and with the metal surfaces and gases.
electrodes of the discharges chamber to form stable gaseous
Considering the both result, the FTIR spectroscopy and
molecular species including gaseous sulfur oxyfluoride, SOF2,
SO2F2, and SOF4; sulfur oxide, SO2; and hydrofluoric acid,HF; passive detector tube test, it was suggested that SF4 and SOF2
were possible decomposition component to respond for CNT
as well as solid product such as metal fluoride [10]. For
example, SF4 react with H2O and undergo hydrolysis as : gas sensor response. However, it is necessary to investigate
deeply for describing adsorption of molecule gas on CNT gas
SF4 + H2O  SOF2 + 2 HF (2) sensor for the positive higher than the negative one. It was
identified that SF4 and SOF2 were adsorbed on CNT by FT-IR
SOF2 + H2O  SO2 + 2 HF (3)
spectroscopy.
The final decomposition products, SO2 and HF, have been
widely employed as consequence, a reduction in SF6 V. CONCLUSION
concentration takes place and the insulating efficiency of the
system is degraded. Eventually, the SF6 must be renewed, if We have performed a series of experiment to explore SF6
failure of insulation is to be prevented. decomposition gases in HVDC partial discharges. FT-IR
spectra showed that SF4 and SOF2 gas molecules were
In this study, however, we did not observe absorption of adsorbed on CNT gas sensor under exposing the PD in SF6.
spectra SO2 and HF due to background absorption of H2O,
shown in the Fig.3. But it did not mean that SO2 and HF
molecules were not adsorbed on CNT. On the other hand, ACKNOWLEDGMENT
adsorption of SF4 or SOF2 (or both) was highly expected from This result was partially supported by TEPCO Research
Fig.4 and 5. The reason of why difficult to detect SF4 or SOF2 Foundation.
separately is that hydrolysis of SF4 into SOF2 is so fast.
Therefore, it has been considered and quantified together as REFERENCES
(SF4+ SOF2) in the past [10]. Although theirs separation are
impossible at this point, the FTIR spectroscopy revealed that
[1] W.Breuer,D.Povl, D.Retzman, and E. Teltsch,” Trends for future HVDC
SF4 or SOF2 (or both of them) were adsorbed on CNT. application,” CEPSI conference 2006. The 16th Conference of the
B. Response of CNT gas sensor to HF and SO2 Electric Power Supply Industry
[2] S.Tominaga, H.Kuwahara, K.Hiroka, T. Yoshioka, SF6 gas- analysis
Although adsorption of SF4 or SOF2 on surface of CNTs technique and its application for evaluation of internal condition in SF6
was realized, it has not been clarified yet if they were gas equipment, IEEE Trans.Power Appar.Syst PAS-100 (1981) 4196-
responsible for PD response of CNT gas sensor [11]. In order 4206
to discuss this point, the CNT gas sensor responses to HF and [3] B.Belmany,J.Casanova,AM Casanova, “ SF6 decomposition under
power arc-chemical aspect”,IEEE Tran.Elect.Insul., vol.26,pp.1177-
SO2 at ppm level were investigated. The experimental set up 1182, 1991..
and procedure were almost similar to experimental before but [4] T.Minagawa,M.Kawada,S.Yamauchi, MKamei,Nishida, “Development
there were a little different. In this experiment we applied of SF6 gas sensor,” Surf. Coat Technol., vol. 169/170, pp. 643–645.
voltage 20 kV for generating PD in SF6 (0.1 Mpa) and setting 1991
gap length at 10 mm, PD exposure time about 18 to 20 hours. [5] R.Irawan,G.B.Scelsi and G.A.Woolsey “Optical fiber sensing of SF6
After the PD stopped, a part of SF6 decomposition gases (25 degradation in high voltage switchgear,” J.Nonlin. Opt Phys., vol
mL) was taken gas –tight syringe and the gas inject ed into a 10,pp.181-195,2001..
small chamber (200µL in volume) for four times during each [6] J.Suehiro,G.Zhou, and M.Hara, “Detection of partial discharges in SF6
gas using a carbon nanotube based gas sensor,” J. Sensors and Actuators
10 minutes. Before injected SF6 decomposition gases within B,105, 2005, pp. 164-169.
small chamber, we must flowing N2 gas as carrier for 5 [7] H.M.Heise, R.Kurte, P.Fisher,D.Klockow, and P.R.Janissek, “Gas
minutes. After that we injected SF6 decomposition gases within analysis by infrared spectroscopy as a tool for electrical fault diagnostics
chamber for four times during each 10 minutes. The last step of in SF6 insulated equipment ”, Fressenius J.Anal.Chem, vol.358,pp.793-
this experiment that flowing of N2 gases again during 30 799,1997.
minutes. The result was found conductance change under [8] J.Suehiro,G.Zhou, and M.Hara,” Fabrication of a carbon nanotubes-
positive and negative DC polarity. based gas sensor using dielectrophoresis and its application for ammonia
detection by impedancy spectroscopy”,J.Phys.D:
In this case, an application of CNT gas sensor for off-line Appl.Phys.,vol.36,pp.L109-L114,2003.
diagnosis (detection was carried out 18 hours later after PD [9] Z Li,Y Martin,T.Tsutsumi,K.Imasaka,J.Suehiro,S.Ohtsuka,” Responses
stopped) using syringe and gas detector tube was realized. It of carbon nanotube gas sensor to partial discharges generated by DCHV
of different polarities in SF6,” Proc. of National Convention of IEEJ
was found that the CNT gas sensor higher response to the (2010).
negative polarity than that positive. The series of experiment to
[10] W.ding,R.Hayashi,K.Ochi,J.Suehiro,K.Imasaka,M.Hara,N.Sano,E.Naga
detect HF and SO2 using detector tube (pump type) at ppm o,and T.Minagawa,” Analysis of PD-generated SF6 decomposition gases
level with indicate the color change for both of negative and adsorbed on carbon nanotubes. IEEE Transac. On DEI, vol
positive DC polarity. Target of this experiment to detect 13,No.6,2006.
decomposition of gases: HF and SO2 in the decomposition gas. [11] Y.Martin, Z.Li,T.Tsutsumi,K.Imasaka,J.Suehiro,S.Ohtsuka,” Analysis of
Passive detector tube was detected HF at 1 to 2 ppm level and SF6 decomposition gases absorbed on carbon nanotubes under DCHV
partial discharges”, Proc. of National Convention of IEEJ (2010).
SO2 at over 40 ppm for both polarities. This result suggested it
was necessary long period of discharges time to detect HF to

4
666

You might also like