Professional Documents
Culture Documents
White Paper
If the surface area of the liquid is increased, more molecules are present
at the surface, and work must be done for this to occur. Therefore, the
surface has excess Gibbs energy relative to the interior of the liquid.
Surface tension is the excess energy per unit area (force per unit length;
SI unit is Nm-1). Typically mN/m (which is equivalent to dynes/ cm) is used
as the unit for surface and interfacial tension.
Surface tension values for some pure liquids are shown in (Table 1).
It should be noticed that surface tension of water is high compared to
most of the organic liquids. Only glycerol has surface tension close to
that of water. Both water and glycerol are highly polar liquids containing
hydrogen bonds which explain the high surface tensions.
One of the main factors affecting surface tension of pure liquids is the 70
temperature. Experimentally it has been found that the surface tension
65
decreases almost linearly with temperature (see surface tension of water
as a function of temperature in figure 2). When temperature increases, 60
the molecular thermal activity increases causing a decrease in cohesive
55
interaction. This causes decrease in surface tension. 0 20 40 60 80 100
Temperature (ºC)
Another factor affecting the surface tension is pressure. In most appli-
cations the pressure does not play a role, but needs to be considered [Figure 2] Surface tension of water as a function
when processes that happen under high pressures are studied. These of temperature (Data from3).
include for example enhanced oil recovery (EOR) and supercritical fluids.
In general, dissolution of the gas in liquid increases with the increasing
pressure, decreasing the surface tension. Interfacial tension between CO2 60
40
35
Force tensiometry
Force
Force
be known.
water
F= ρp glwt+2γ(w+t)cosθ-ρι gtwh
where the first term is due to the mass of the plate, second is the
wetting force and third is due to buoyancy. As the Wilhelmy plate is
assumed to be wetted completely with the liquid the term cos θ will
go to one and the equation can be rewritten as Equation 2.
ΔF = 2γ(t+w) ⇒ γ = ΔF , if w ≫ t
2w
Wilhelmy plate measurements can also be done in so called static or
continuous mode where Wilhelmy plate remains in contact with liquid
during the entire measurement cycle.
Platinum rod
Both of the above mentioned approaches require a relatively large
amount of the liquid (typically above 10 milliliters) to be used to ensure
complete wetting of the probe. It is possible to use a sample vessel
with a smaller diameter to decrease the total volume required. However,
there is a limit. When the edges of the sample vessel and the probe are
too close to each other, the balance can be affected by the meniscus
that forms between the liquid and the edge of the vessel. To avoid this
problem, the user must allow enough room between the probe and
the side wall of the sample container, ~ 2 mm or so. In any of these
techniques, the accuracy of the measurement is affected by the accuracy
with which one measures the geometry of the probe. The measurement
accuracy of the platinum rod is likely to be less than the du Noüy ring
or the Wilhelmy plate which translates to larger errors in the surface /
interfacial tension results. Therefore the platinum rod approach is only
recommended when the volume of the sample is limited.
The method and calculations are based on the same principle as the
Wilhelmy plate-method.
γ = ΔρgR0
β
where γ is the surface tension, Δρ is the density difference between
fluids, g is the gravitational constant, R0 is the drop radius of curvature
[Figure 7] Pendant drop method based on optical drop at the apex and β is the shape factor. β can be defined through the
shape analysis allows using very small sample volumes Young-Laplace equation expressed as 3 dimensionless first order
(~5 to 20 microliters). equations as shown in Figure 7.
Comparison of methods
Force and optical tensiometers are excellent tools for surface and
interfacial tension measurements. They provide fast and simple measure-
ment methods for wide range of applications. As with any measurement
methods, all the above discussed techniques have their advantages and
disadvantages9.
With all the force tensiometry techniques the geometry of the probe
is especially important. Wilhelmy plate and platinum rod are relatively
robust but du Noüy rings are very delicate and deform easily. For correct
surface and interfacial tension measurements it is crucial that the plane
of the ring remain parallel to the interface. With the proper handling of
the du Noüy ring, deformation of the ring can be avoided.
Conclusions
In this white paper, we have given a short introduction to
surface and interfacial tension as well as reviewed the most
common measurement techniques. We have also discussed
the advantages and limitations related to each measurement
type. The instrument should be chosen according to your
application. For more information on the selection of the
best method for your application please refer to white
paper titled “How to select the best measurement method
for your application?”.
About us
Biolin Scientific is a leading Nordic instrumentation company with roots
in Sweden, Denmark and Finland. Our customers include companies
working with pharmaceuticals, energy, chemicals, and advanced materials,
as well as academic and governmental research institutes. Our precision
instruments help discover better drugs faster, develop better solutions
for energy and materials, and perform research at the frontiers of science
and technology.
Contact us at info@biolinscientific.com