Professional Documents
Culture Documents
J..... ,o,Applied
9
Institut far Technische Chemie der Universitiit Hannover, Callinstrage 3, D-3000 Hannover 1,
Federal Republic of Germany
1. Introduction
In the foregoing paper (part I) (Buchholz and Schiigerl, 1979) photographic and elec-
trical conductivity methods were discussed. In the present paper two further methods
are considered, the electrooptical and light scattering methods. The latter was not in-
vestigated by the authors. However, based on the literature data, it will also be dis-
cussed and compared with other methods.
2. Electrooptical Method
The electrooptical probe was developed by Todtenhaupt (1971) and Pilhofer (Pilhofer
and Miller, 1972) (Pilhofer, 1974) as well as by Jekat (1975) and improved by Brentrup
et al. (1977). Figure 1 shows the principle of this method. A negatively pressurized
capillary tube, which is equipped with a funnel inlet draws off the two phase system.
In the capillary tube, the bubble is deformed into a cylindrical plug which passes two
light gates. By passing the gas/liquid interface through the gate, the intensity of the
light at the detector is at first increased, then decreased and increased again due to
focusing deflection and focusing of the light beam. At gates 1 and 2 the plug causes
near square-wave signals with the durations At 1 and ~t 2. The time lag between the two
signals is ZXtl2. The bubble volume is given, neglecting the curvatures of the plug at its
two ends, by Eq. (1):
V B = h A D ZXt1/At12 (1)
where h is a calibration constant and A D the cross section area of the capillary. Since
Eq. (1), used for the evaluation of the measurements made by the electrooptical
0171--1741/79/0006/0315/~ 01.80
316 R. Buchholz and K. Schiigerl
f,-,
Ate2
T
(b)
(a)
~;Z. b
Fig. 1. a. Electrooptical sensor (Pilhofer and Miiller, 1972). a = capillary; b = inlet funnel; c = light
source; d = light detector; e = bubble; f = liquid, distance of the gates 1 and 2; b = 132 mm.
course of light beam with liquid in the capillary; - - - course of light beam with bubble in
the capillary, b. Signals at the gates 1 and 2
method, and Eq. (2b) of part I (Buchholz and Schiigerl, 1979), used for the evalua-
tion of the measurements carried out b y the electrical conductivity method, are
analogue, the same electronics (Fig. 3 in part I) are applied for both of these probes.
The application of this probe can be limited b y the following shortcomings: at high
removal rates the bubbles are dispersed in the inlet funnel and/or in the capillary tube;
at low removal rates the bubbles can be dispersed if t h e y are much larger than the inlet
funnel; at too low removal rates and in a coalescence promoting media, the bubbles
can coalesce in the inlet funnel; b y using a capillary with diameter larger than the
bubble diameter, no signal processing is possible; only by an isokinetic removal rate
is segregation of the bubbles with different diameters possible. However, isokinetic
removal rate promotes coalescence; at too high bubble concentration a separation of
the signals caused by different bubbles is not possible.
For bubbles with narrow size distributions, most of these difficulties can be
avoided b y using a suitable capillary diameter and removal rate. In case of bubbles with
broad size distributions, no satisfactory solution is possible, if one uses only a single
capillary diameter: in a capillary with a medium diameter, the small bubbles cannot
be detected and the large bubbles are dispersed. To evaluate the size distribution of
bubbles with broad size distribution, the use of capillaries with different diameters
is necessary.
By using focusing lenses to improve resolution of the light beam (Brentrup, 1977),
b y applying capillary tubes with an inner surface which was made hydrophobic to
diminish the hazard to bubble dispersion, by reducing the shear stress in the capillary,
and, to improve the signal quality, by eliminating the liquid film which adheres to the
capillary surface causing noisy base lines (Buchholz, 1978), the quality of this meas-
uring technique was significantly improved. The hydrophobic surface made it possible
to use high removal rates without dispersing the bubbles in the capillary.
Bubble C o l u m n Bioreactors II 317
120~
Fig. 2. Combination of five optical sensors (dip-sticks)
(Calderbank and Pereira, 1977). k = fiber optics; l = flange; m =
spacer; n = stainless steel hypodermic tubing; o = electrode tip;
p = stainless steel tube ; q = glass electrode. Longitudinal distance
from the tip of sensor i to the tip of sensor 2, 3 or 4: a = 2.93 mm;
transverse distance from the tip of sensor 1 to the tip of sensor 2,
3 or 4: b = 1.03 mm; transverse distance from the tip of sensor 1
to the tip of sensor 5: c = 4.95 mm; length of the sensor tip: d =
1 4 . 2 1 m m ; e = 8 c m , f = 2 4 c m , g = 5 cm, b=O.079cm, i=
b; c-
0.398 cm (i
318 R. Buchholz and K. Schiigerl
tOO
0.80
060
040
0.20
0,00
I I
1.00
I
JY
200 3.00 &O0 5.00 6,00
• p
700
d(mm]
Fig. 3. Distribution of bubble diameter estimated by the electrooptical sensor. Bubble column 14
cm in diameter, 267 cm high. Porous plate distribution, 50 #m mean pore diameter. Distilled
water. Superficial gas velocity 1.07 cm s-1. Distance from the gas distributor 132 cm. d = 3.2 cm,
d s = 3.45 cm
Critical Points: The light b e a m across t h e capillary must be exactly adjusted; the
Schmitt trigger threshold voltage must be set for b o t h channels to 40% of m a x i m u m
voltage; the sensor capillary diameter must be selected carefully according to the bub-
ble size distribution. F o r narrow b u b b l e size distributions, a single capillary is adequate.
The capillary diameter should be a b o u t half of the m e a n b u b b l e diameter. The use of
too small a sensor capillary diameter causes b u b b l e redistribution in the inlet funnel.
By t h e application of too large a capillary diameter, signals caused b y b u b b l e s with
diameters smaller t h a n the sensor capillary c a n n o t be separated; for wide b u b b l e size
distributions, two or more sensor capillaries m u s t be used. The construction o f the
overall b u b b l e size distribution b y means of particular ones, evaluated b y the differ-
ent sensor capillaries, is fairly difficult; the inside surface of the inlet f u n n e l and the
sensor capillary should be made hydrophobic. This avoids dispersion of the bubbles
inside the capillary at high sucking velocities due to the adhesion of the liquid o n the
capillary surface, yielding high shear velocities at the wall.
n_~
}-n/Ad
~,~
o,6
r -
iii/
o.2
I 2 3 4 5 6 7 8
wSG=2fl3 cm/s d(mm)
n___
o,a
(2,6 /"
.iI
/I
0,~ ,./ /
/ /
0 1 2 3 4 5 6 7
w = 1,~Tcralz d(rora,~
Fig. 4. Comparison of the bubble size distributions measured at the wall by photographic, electri-
cal conductivity - and electrooptical methods. Distilled water, h = 132 cm, WSL = 1.8 cm/s, I. At
low superficial gas velocities, wSG. - - electrooptical method; - - - photographic method;
. . . . electrical conductivity method
produced using a capillary (aerator) of definite diameter. The bubble sizes were esti-
mated by measuring the number and total volume of many hundred bubbles. The bub-
ble size was altered by varying the capillary (aerator) diameter. The measurements
showed that for the 1.5 mm (inside diameter) sensor capillary, which was used for
electrooptical measurements in distilled water, Eq. (1) did not hold for bubbles less
than 2.5 mm in diameter. Below this diameter, no linear relation prevails. All bubbles
between 2.0 and 2.4 m m in diameter were counted in one channel, and those less
than 2.0 mm were hardly detected. A comparison of bubble size distributions which
were measured at the wall by electrooptical and photographic methods (Figs. 4 and
5), indicate that the small bubbles were missing in the bubble size distributions
evaluated by the electroopticaI method with a 1.5 mm diameter sensor capillary. Fur-
thermore, the larger bubbles appeared less frequently in the distribution evaluated by
photographic measurements. Therefore, d and d s evaluated by the electrooptical
method were higher than those measured by the photographic method (Fig. 6).
Furthermore, at high superficial gas velocities (e.g., at wSG = 5.33 cm s-1, Fig. 5)
large bubbles are dispersed in the funnel and sensor capillary producing small bubbles
320 R. B u c h h o l z a n d K. Schfigerl
n
~n/Z~d
0,8
O.6
0`,
Oj
0 1 2 3 ~ 5 6 7 8
w~33cm/s d(mm)
n
T nlxd
II i ~"
0,2 . /9
1 Z 3 4 5 6 7 8
wS6=3,2Ocm/s d(mm)
Fig. 5. Comparison of the bubble size distributions measured at the wall by photographic, electri-
cal conductivity - and electrooptical methods. Distilled water, b = 132 cm; WSL = 1.8 cm/s. II. At
high superficial gas velocities, WSG. For symbols see Figure 4
a ~s
9 9 electrooptico~ method
o A photogrophical method
WsL= 1,8 c m / s
porous plate 50,urn
distiUed water
Highly ;election
~queous viscous Fermenta- vith regard
aaodel model ti on Bubble size o t he
dethod ned ia media media distribution ?osition ,ubble size ~emarks
All of them can be applied to aqueous solutions. The electrical conductivity sensor
in unsuitable for organic liquids. Except for electrooptical sensors, sensors can be
applied to highly viscous media. Because of the non-spherical shape of large bubbles,
the double sensor-sensors d o not yield enough information to reliably estimate bubble
volumes and shapes. This also applies for pure liquids or solutions with antifoam agents.
Four or five sensor-sensors are necessary when large bubbles are present.
Except for the flash photography, the methods are not restricted to a given column
position. Difficulties arise due to the different sensitivity of the methods for bubbles
with unlike sizes. Flash photography can handle the broadest bubble size distribution,
however, it suppresses large bubbles.
The electrical conductivity and/or glass fiber sensor suppresses bubbles with d <
0.6 mm and/or d 1 <~ mm. Electrooptical sensors can only handle bubbles with narrow
size distributions. If the diameter d k of the sensor capillary, is much larger than d,
then the bubbles will not be detected. When d is much larger than dK, the bubbles
will be dispersed and a false bubble size distribution will result. Thus, dK must be care-
fully adjusted to d in case of narrow distributions. For broad distributions, two or
more sensor capillaries are needed.
Electrical conductivity probes and glass fiber sensors may be sterilized if they are
appropriately constructed. In the authors laboratory, the position dependence of bub-
ble size distributions during fermentation is measured by four electrical conductivity
sensors simultaneously. The sterilization problem of the electrooptical sensors is a dif-
ficult one, because the fermentation medium must be removed from the reaction and
is usually fed back into it. Hence, the entire sampling by-pass system must be sterilized.
No simple method is known, by which it is possible to reliably measure a very broad
bubble size distribution in bubble columns.
Acknowledgements. The authors gratefully acknowledge the financial support given by the Ministry
of Research and Technology in Bonn, Federal Republic of Germany.
Symbols
References
Brentrup, L., Onken, U., Irrgang, W., Wagner, F. (1977). "Bioreaktoren", H. Keune
and R. Scheunemann, eds., DFVLR p. 137
Buchhoiz, R. (1979). Dissertation, Technical University Hannover
Buchholz, R., Schiigerl, K. (1979). European J. Appl. Microbiol. Biotechnol. 6,
301-313
Burgess, J.M., Calderbank, P.H. (1975). Chem. Eng. Sci. 30,743
Calderbank, P.H. (1958). Trans. Inst. Chem. Engrs. 36, 443; lbid (1959). 37,173
Calderbank, P.H. (1960). Int. Symp. Destillation, Brighton (Inst. Chem. Eng.) p. 51
Calderbank, P.H. (1967. Mass transfer in fermentation equipment. In: Biochemistry
and biological engineering science, N. Blakeborough, ed. vol. 1, p. 101. New York:
Academic Press
Calderbank, P.H., Pereira, J. (1977). Chem. Eng. Sci. 32, 1427
Delhaye, J.M. (1969). Proc. 11th Nat. ASME/AIChE Heat Transfer Conf. on Two
Phase Flow Instrumentations. Minneapolis, Minn. p. 58
Geake, J.E., Smaller, C. (1975). The Chemical Engineer, p. 301 May
J ekat, H. (1975). Dissertation, Technical University Munich
Pilhofer, T., Miller, H.D. (1972). Chem. Ing. Techn. 44, 295
Pilhofer, T. (1974). Chem. Ing. Techn. 46,913 (MS 149/74)
Sauter, J. (1928). Forsch. Arb. Geb. Ing. 2-8,312
Todtenhaupt, E.K. (1971). Chem. Ing. Techn. 43,337
Trice, V.G., Roger, W.A., Rushton, J.H. (1956). AIChE-Journal 2, 205
Vermeulen, T., Williams, G.M., Langlois, G.E. (1955). Chem. Eng. Progress 51, 85