Professional Documents
Culture Documents
HRA 2
Service Manual
Table of Contents
1 LASER POWER CHECK ................................................................................................................................. 4
1.1 LASER SOURCES ............................................................................................................................................ 4
1.2 MEASUREMENT OF LASER POWER WITH OPHIR POWERMETER ....................................................................... 4
1.3 TSE-02 POWER SENSOR FOR ADJUSTMENT PROCEDURES ............................................................................... 7
1.4 BEAMPATH OF BLUE LASER, AOM ................................................................................................................ 8
1.5 LASER TO FIBER COUPLER ............................................................................................................................. 9
1.6 REALIGNMENT OF SAPPHIRE FIBER COUPLING ............................................................................................ 10
12 APPENDIX.................................................................................................................................................... 59
12.1 HRA 2 PROGRAMMABLE COMPONENTS ...................................................................................................... 59
HRA 2 Service Manual
Art.Nr. 96 040-002 Page4 of 59 TC01_HRA2 Service Manual.doc
HRA2 Laser Laser type Wavelength [nm] Output power at objective (av. over 10 sec.)
(colour) @ 30° HR 100% @ 30° HS 100%
FA-Laser Optically pumped 488 nm 260 – 300 µW 240 – 280 µW
Redfree Refl.-Laser solid state laser (blue)
(Coherent)
ICGA-Laser Laser diode 785 nm 2.35 – 2.65 mW 2.2 – 2.5 mW
(infrared)
IR Refl.-Laser Laser diode 820 nm < 150 µW < 150 µW
(infrared) (typ. 50 –65 µW) (typ. 45 – 60 µW)
On/off switch
(cannot be seen in
this photograph)
Operations
Buttons for
operation
Menu button
Detector head
with HRA2
adaptor
HRA 2 Service Manual
Art.Nr. 96 040-002 Page6 of 59 TC01_HRA2 Service Manual.doc
Detector head
Choice of laser with HRA 2
Scale
to be monitored adaptor
This TSE sensor is only an adjustment tool but not a powermeter. During installation the laser powers have
to be checked with a calibrated Ophir power meter!
The adjustment is used in the adjustment process after reinsertion of the fiber into the 4 axis manipulator during the
setup of the system. In this case the use of the TSE-02 is advantageous, as the display is slow enough to average the
power fluctuations due to the laser modulation.
Procedure:
Place detector head on the objective
Switch the TSE-02 knob to FA.
Operate the HRA 2 at 30°, HS in either FA or RF mode.
Adjust the Kinematics four axis manipulator according to the short or detailed adjustment instructions (v.i.)
so that the scale reading reaches its maximum. (As a rule of thumb for a correctly installed fiber the reading
should be in between 4.5 and 6 scale units.)
HRA 2 Service Manual
Art.Nr. 96 040-002 Page8 of 59 TC01_HRA2 Service Manual.doc
When the modulation is disabled (e.g. by disconnecting the RF-power from the AOM drive, all laser power is
transmitting the AOM crystal (acousto-optical modulator) without being deflected. This beam is called 0. order
beam.
However, when the modulation is activated, the blue laser beam is split in several beams. The beam next to the 0.
order beam (left side) is the 1. rder beam and is coupled into the single mode fiber.
retaining pin
B1 B2
A1 A2
Retaining pin locked
due to allen screw
Push this spring button 4 x locking nuts for
down for inserting fiber alignment screws
coupler or prealignment
tool
HRA 2 Service Manual
Art.Nr. 96 040-002 Page10 of 59 TC01_HRA2 Service Manual.doc
1. For inserting the fiber: Release the tension screw D. 1. For inserting the fiber: Push down spring button E,
Insert the fibre input coupler. Make sure, that the while inserting the fiber input coupler. Make sure, that
polarization key E fits into the keyway F on the fiber the polarization key E fits into the keyway F on the fiber
carrier. Increase tension on screw D.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page11 of 59 TC01_HRA2 Service Manual.doc
2. During alignment procedure laser power can either be 2. During alignment procedure laser power can either be
monitored using adjustment aid TSE-02 (to be preferred) monitored using adjustment aid TSE-02 (to be preferred)
or by periodic averaging over a 10s time scale (takes or by periodic averaging over a 10s time scale (takes
much more time). Attach the sensor to the HRA2 much more time). Attach the sensor to the HRA2
objective lens. objective lens.
3. Tilt alignment: Optimise A1 & B1 for optimum fibre 3. Tilt alignment: Optimise A1 & B1 for optimum fibre
output. It is important, that the tension screws C &D are output.
always in the correct range. Alignment is not possible if
these screws are tightened too much or; if they do not
generate a preload tension (too loose).
4. Horizontal translation: (only required if step 3 does 4. Horizontal translation: (only required if step 3 does
not result in sufficient laser power). not result in sufficient laser power).
Although the tilt of the input optic is optimised, the
Although the tilt of the input optic is optimised, the
alignment still can be improved by translating the input
alignment still can be improved by translating the input
optic in horizontal or vertical direction. For horizontal
optic in horizontal or vertical direction. For horizontal
translation adjust the screws A1 and A2 stepwise in the
translation adjust the screws A1 and A2 stepwise in the
same direction, i.e. adjust A1 a quarter of a turn counter
same direction, i.e. adjust A1 a quarter of a turn counter
clockwise. The output power will go down completely.
clockwise. The output power will go down completely.
Then adjust A2 in same direction also about a quarter of
Then adjust A2 in same direction also about a quarter of
a turn and maximize output power. If the power is higher
a turn and maximize output power. If the power is higher
than in the beginning, repeat the procedure in the same
than in the beginning, repeat the procedure in the same
direction. If the power decreased repeat the procedure in
direction. If the power decreased repeat the procedure in
the other direction i.e. clockwise. If the power does not
the other direction i.e. clockwise. If the power does not
change significantly, the optimum has been found
change significantly, the optimum has been found
5. Vertical Translation: (only required if step 3 and 4 5. Vertical Translation: (only required if step 3 and 4
do not result in sufficient laser power). do not result in sufficient laser power).
Repeat the same procedure (4.) for vertical translation by Repeat the same procedure (4.) for vertical translation by
manipulating the screws B1 and B2. manipulating the screws B1 and B2.
6. Carefully secure all the alignment screwsA1..B2 by 6. Carefully secure all the alignment screwsA1..B2 by
tightening the locking nuts J. The locking nuts should not tightening the locking nuts. The locking nuts should not
be tightened too much, it should be still possible to be tightened too much, it should be still possible to
manipulate the alignment screws in a small range. manipulate the alignment screws in a small range.
8. As the fine tuning of the adjustment depends on 8. New versions are in general less sensitive to
temperature inside the housing, a final adjustment of temperature changes inside the housing. Nevertheless we
horizontal and vertical tilt (screws A1 and B1) in warm recommend to close the cover and recheck the laser
condition is required. Therefore close the cover and wait power after about 30 minutes. If necessary, reopen the
for about 30 minutes. Then reopen the housing and housing and optimise A1 and B1 again.
optimise A1 and B1 again.
Please note: In case no output power at all can be measured after inserting the fiber coupler, the optimisation
procedure can not be accomplished. In this case a special prealignment tool is required for the coarse alignment. In
doubt, please contact Heidelberg engineering for technical support.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page12 of 59 TC01_HRA2 Service Manual.doc
• Place the scale target on the scan field aperture tubus. Make sure, that the scale is oriented
horizontally resp. vertically.
• Switch to Live mode, IR Refl. and make sure, that the filter is set to angiography mode. The IR
laser power has to be set to typically 4%, otherwise the image is overexposed. This can be done either
by switching the Touch Panel into Service mode (red background) or by opening one of the start-up
windows in the HeScript software.
• As the scan field aperture is exactly positioned in the intermediate plane, the scale itself can be
positioned only 1 mm in front of the intermediate plane. Thus the image is not sharp. Therefore set the
Z-module to the position, where the best image of the scale is obtained. For example, if the home
position of the Z-module is at 240 steps, you should see the sharpest image of the scale at about 230
steps. This can be done either in the service panel, which is opened by pressing Ctrl & Alt & Shift &D
within the HRA 2 acquisition module or by opening the Z-module procedure in the HeScript startup
menu.
• Measure the scan field size. It is convenient for the measurements to activate to the extended
service live image mode in the service panel, where the scan field is limited on top, left and right side
by green lines. The specified values are listed in the following table.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page13 of 59 TC01_HRA2 Service Manual.doc
30° x 30° 166.5 +/- 1.5 scale units 166.5 +/- 1.5 scale units
20° x 20° No check necessary 110.5 +/- 1 scale units
15° x 15° No check necessary 83 +/- 1 scale units
Saved image
@ 30° x 30°
HRA 2 Service Manual
Art.Nr. 96 040-002 Page14 of 59 TC01_HRA2 Service Manual.doc
Please note:
a.) The scan field in horizontal direction does not need to be checked as the scan amplitude for 20° and 15° is
identical to the scan amplitude at 30°. The 20° and 15° images are just cropped in X-direction.
b.) At 20° and 15° there is some difference between horizontal size (X-direction) of live mode images and
saved images. The reason for this is, that the sinus correction is not done in live mode, but only when the
images are stored. Of course the size of the saved images correspond to the specified image sizes of 20°
resp. 15°.
• Move the camera as far as possible towards positive diopters (max. distance between
objective arm and camera body) in order to have access to the 3 allen screws! Remove
the three screws (wrench size 3.0)!
• Remove the refraction knob (allen set screw, wrench size 1.5)
• Remove the two allen screws (wrench size 2.0) on the left side (patients view). These
screws ensure thermal contact of the final amplifier stages of the scanner drivers to the
housing. Thermal contact is very important to remove the heat from inside the camera,
i.e. the camera housing acts as heat sink for the whole camera.
• Remove the allen screw on the backside of the camera (wrench size 3.0)!
• Push the housing carefully backwards until you can remove it completely. Disconnect
the ground cable inside the housing.
When closing the camera housing just proceed the opposite way. Extremely care has to be taken
when remounting the housing on the optic block, to make sure, that no cable is squeezed.
Especially the channel (see images) needs to be free of cabl
The channel between optic block and mounting plate needs to be free of cables, when remounting
the housing cover.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page18 of 59 TC01_HRA2 Service Manual.doc
+3V3
+5V
-12V
GND
+12V
1=+5V 1=L1 1=+5V 1=X-Scanner 1=IR MOD 1=ICG MOD 1=Y-Drive 7=AGC
2=ID 0 2=L2 2=Obj 0 2=GND 2=GND 2=GND 2=Y-Drive 8=GND
3=ID 1 3=L3 3=Obj 1 3=X-PD- 3=IR VCC 3=ICG VCC 3=GND 9=Y-PD-
4=ID 2 4=L4 4=GND 4=X-PD+ 4=IR Ref 4=ICG Ref 4=GND 10=Y-PD+
5=GND 5=Home 5= 5= 5=IR Adj 5=ICG Adj 5=+12V
6=+5V 6=GND 6=+5V 6= 6= 6= 6=GND
7=ID 3 7=Z-Pos
8=+5V 8=GND
HRA 2 Service Manual
Art.Nr. 96 040-002 Page19 of 59 TC01_HRA2 Service Manual.doc
fixation
mirror
FA-laser
IR-Laser
ICG-Laser
pinhole
90° deflection
mirror
For the adjustment of the turning mirrors again access to the optical intermediate plane is required. Therefore the
telescope needs to be disconnected as described in the previous section. A homogeneous, fluorescent target is
required for the alignment procedure of pinhole focussing and laser sources. Position the target on the scan field
aperture and move the stepper of Z-module in order to obtain sharp images on the screen. Open the service panel
with the hot key Ctrl & Alt & Shift & D for access to the stepper motor. The adjustment is done in order to achieve
a brightest possible image. However, at the same time the brightness distribution should be homogeneous and
circular symmetric. When activating in the service panel the extended service image mode a histogram of the
brightness distribution and a mean value of the gray values is displayed.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page20 of 59 TC01_HRA2 Service Manual.doc
Select “hra2.dev” from the list of devices shown in the startup dialog, check the language setting and click on “OK”.
The first screen (“Scripts”) allows to execute scripts that control the HRA 2. This is a function that is especially
useful for further diagnostics in a service case.
Select the respective script in listbox (1). The script content will be displayed in field (6). Select the “run” button (2)
to start the script. The text of button (2) will then change to “stop” and another hit of this button will halt script
execution.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page25 of 59 TC01_HRA2 Service Manual.doc
The currently executed line is indicated in field (6) by a green bar. Possible log output is dumped to field (5). Button
(4) lets you choose another script directory and button (3) lets you delete scripts.
When using an up to date HRA 2 service program CD, you can check, if a firmware update is required. Start the
service program as described above and click on the “Startup” menu. Then select procedure “Firmware-, Flash- and
Board Versions” (see section “Collecting Version Information” below):
Old versions, that are recommended to update are shown in red colour. Acceptable versions are shown in blue
colour, current versions (up to date according to the service CD) are shown in green.
To update the HRA 2 internal firmware, you will need the following parts:
1. The HE-ISPD-JP1 in-circuit programming device for the parallel port and a the appropriate connection
cable to the HRA 2 power supply. A driver is provided on the CD ROM. This driver must be installed
before the HE-ISPD-JP1 device can be used.
2. A Firmware update file with file extension *.HFU
3. The HRA service program (CD)
Connect the HE-ISPD-JP1 device to the parallel (printer) port of the PC and connect the yellow plug to the
appropriate socket at the HRA 2 power supply back side. In case, it is hard to insert the plug into the socket, use a
19mm wrench to slightly open the screw nut.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page26 of 59 TC01_HRA2 Service Manual.doc
Start the service program as described above and click on the “Startup” menu. The following dialog will appear:
Please note, that you must have administrative user privileges on Windows to be able to install the driver. After the
driver installation, the computer must be rebooted.
After reboot, restart the service program and select the procedure “Firmware upload” again.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page27 of 59 TC01_HRA2 Service Manual.doc
Click on the “Browse…” button (1) in section “Camera/Power Supply”, change to directory “firmware-HRA2”,
select the firmware file “HRA2-Firmware-x.y.z.hfu” (where x.y.z represents the firmware version) for upload and
click on “Open”:
HRA 2 Service Manual
Art.Nr. 96 040-002 Page28 of 59 TC01_HRA2 Service Manual.doc
The complete path and file name of the firmware update file will be shown in the “Firmware Upload” dialog:
NOTE: Close all other running applications before upload. Do not stop the program or disconnect the cable
before the “Firmware Upload” dialog closes automatically. Any interrupt in the programming procedure
may cause serious damage to the electronic circuits of the HRA 2!
Click on the “Start CAM/PWS firmware upload …” button (2). A new dialog will be shown in red colour:
After the upload procedure has been finished, select procedure “Firmware-, Flash- and Board Versions” (see section
“Collecting Version Information” below) to verify the upload process.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page29 of 59 TC01_HRA2 Service Manual.doc
Double-click on the procedure “Firmware upload”. The following dialog will appear:
Click on the “Browse…” button (1) in section “Touch Panel”, change to directory “firmware-HRA2”, select the
firmware file “TSP-Firmware-x.y.z.hfu” (where x.y.z represents the firmware version) for upload and click on
“Open”:
HRA 2 Service Manual
Art.Nr. 96 040-002 Page30 of 59 TC01_HRA2 Service Manual.doc
Click in “FW Info Details” (1) to open the IModule version dialog. The following dialog will be shown:
Check bootloader version (entry “BL Version”) and application version (entry “APP Version”) and compare with
the version provided with the latest HRA 2 service CD.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page32 of 59 TC01_HRA2 Service Manual.doc
If required, the IModule bootloader or application firmware can be updated from within the “IModule
configuration” dialog. Click on “Browse…”(2) and a file selection box will be opened:
Change to folder “firmware-hra2” and select file “IModuleApp-x.y.hfu” for the application firmware or
“IModuleL2Bl-x.y.hfu” for the bootloader firmware (where x.y represents the firmware file version) and click on
“Open”.
Select “Write Firmware:” (3). The following dialog will be shown.
The serial numbers of the camera and power supply are shown in frame “EEPROM” (1).
To download the flash contents, click on “Download flash content to file” (button 2), select a file name, and start the
download procedure.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page34 of 59 TC01_HRA2 Service Manual.doc
The actual versions of the different HRA 2 components will be shown. Click on the “Print” button (1), if a version
report is required. The report will be sent to the standard printer automatically.
Old versions, that are recommended to update are shown in red colour (see sections “Firmware Update” above on
how to update firmware versions). Acceptable versions are shown in blue colour, current versions (up to date
according to the service CD) are shown in green.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page35 of 59 TC01_HRA2 Service Manual.doc
7.9 Z-Module
Start the service program as described above and click on the “Startup” menu.
Double-click on the procedure “Adjusting scanner amplitudes”. The following dialog will appear:
The z-stepper control (1) allows to move the internal z-module step by step. The refractive power corresponding to
the current z-module position is displayed in diopters. Note that this module directly accesses the z-module while
bypassing upper firmware layers. Therefore it is necessary to perform a z-module reset before resuming to
normal system operation if the stepper position was changed manually.
The second control (2) allows a simple stepper random walk test. The first entry is the number of stepper
movements to be performed, the second entry determines the maximum number of steps of a single stepper move.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page38 of 59 TC01_HRA2 Service Manual.doc
The dialog shows the parameters of the refraction correction system. Entry (1) shows the current refraction setting,
entry (2) the change of refractive power per mm camera head movement, and entry (3) shows the current zero
position relative to the position encoder zero position.
Note: Selecting this screen resets the z-module. Therefore any “myopic lens” setting is reset to zero diopters.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page39 of 59 TC01_HRA2 Service Manual.doc
Description:
1. Halts y scanner by selecting a zero-ramp and x-scanner. The x-scanner can only be halted if the camera
service jumper is set. Do never look into the beam when the scanners are disabled!
2. Allows selection of laser mode and laser powers.
3. Shows the IR laser lookup table (translation table between power in percent and internal potentiometer
setting).
4. Shows the ICGA laser lookup table (see 3.)
5. Blue laser potentiometer setting when the laser is switched off (see below).
6. Time after that the laser potentiometer is set to 100% after blue laser selection (see below)..
7. Time after that the AOM lets the laser through after blue laser selection (see below)..
HRA 2 Service Manual
Art.Nr. 96 040-002 Page41 of 59 TC01_HRA2 Service Manual.doc
100%
75%
50%
0%
Laser switched off Laser switched on
Time
100%
75%
50%
25%
0%
Laser switched off Laser switched on Time
HRA 2 Service Manual
Art.Nr. 96 040-002 Page42 of 59 TC01_HRA2 Service Manual.doc
Security Configurations
Start the service program as described above and click on the “Startup” menu.
Double-click on the procedure “Adjusting scanner amplitudes”. The following dialog will appear:
Line (1) allows to set the blue laser power to 100 resp. 125%. Check the box to update the current monitor voltages
at a regular interval.
Line (2) and (3) show the current thresholds of the internal diode lasers for IR and ICGA. Check the respective box
to update the laser currents at a regular interval.
Start the service program as described above and click on the “Debug” menu. connect e.g. the camera RS232
debugging connector to e.g. COM3, select “COM3” on the radio buttons in the dialog below and click on “Online”.
All debugging messages will be logged into the text area of the dialog. This text can be deleted using the “Clear”
button or saved to a file using the “Save” button.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page44 of 59 TC01_HRA2 Service Manual.doc
As soon as you enable the “Service mode ON” checkmark, the system will enter the service mode and will show up
the following behaviour:
1. A detected “Scanner-Interlock” or “Laser-Interlock” will be ignored by the HRA 2 firmware, but it will not
be ignored be the blue laser in the power supply. In this case, the laser light will go off, but the system does
not go to its interlock state.
2. A detected “MCU-Interlock” will be ignored by the HRA 2 firmware. In this case, the blue laser will not be
affected and stays on.
3. Any interlock detected will be shown as warning message only (yellow colour) in the PC acquisition
module status bar.
4. Image acquisition is disabled as long as service mode is on.
5. The background of the acquisition module dialog is shown in pink
6. The LCD display on the Power Supply will periodically show the message “SERVICE MODE, DO NOT
USE THE SYSTEM!”
7. Power supply and camera are periodically sending debug messages (see “Service Panel” in chapter “PC
Acquisition module” for more information on debug messages).
VERY IMPORTANT:
NEVER EXPOSE A PATIENT TO THE LASER BEAM WHILE “SERVICE MODE” IS ACTIVE
BECAUSE OF THE REDUCED SECURITY LEVEL!
HRA 2 Service Manual
Art.Nr. 96 040-002 Page45 of 59 TC01_HRA2 Service Manual.doc
VERY IMPORTANT:
NEVER EXPOSE THE PATIENT TO THE LASER BEAM WHILE “SERVICE MODE” IS ACTIVE!
The behaviour of the acquisition module differs in the following aspects from its normal operation:
1. Image acquisition is disabled.
2. Interlocks errors are shown in Yellow colour and are shown only temporarily in the status bar
HRA 2 Service Manual
Art.Nr. 96 040-002 Page46 of 59 TC01_HRA2 Service Manual.doc
To open up the service panel press the following keys simultaneously: Ctrl & Alt & Shift % D. The following
dialog will appear on the screen:
10. Allows to select the HRA 2 devices which shall be debugged. Only messages from the specified devices
will be appended to the debug message screen (8). Click on “Start” to start debugging for the specified
devices.
11. The button “Reset Memory” allows to remove acquired images from memory BEFORE saving. All images
not saved to disk will be permanently lost!
HRA 2 Service Manual
Art.Nr. 96 040-002 Page48 of 59 TC01_HRA2 Service Manual.doc
In normal operation the power supply LCD shows the following values: Blue Laser Error Bits
In the case of an safety interlock, the last interlock numbers are displayed. There are three coarse interlock Types:
1. Scanner Interlock x- or y-scanner malfunction.
2. Laser Interlock Monitor diode in camera detects too much (blue) laser intensity.
3. MCU Interlock The firmware has detected an erroneous or inconsistent internal state. This
Interlock type is only displayed when whether 1) nor 2) applies.
Note that the blue laser supply voltage is switched off (“-” instead of “L”) while the HECOM connection persists.
The left number is always the power supply internal interlock cause, the right number shows the last incoming
interlock cause number. A complete list of interlock numbers is found in the appendix.
If the system is in service mode, a warning is shown in a regular time interval (see above).
Do not use the system with patients when this warning appears.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page50 of 59 TC01_HRA2 Service Manual.doc
1.
press
once
Go to freeze mode. Press the bottom left corner once and then the top right corner thrice. A correct push is
acknowledged by a “beep” sound. An incorrect sequence must be started anew.
If the sequence was right, a new screen with a warning appears, press “Continue” to reach the service screens.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page51 of 59 TC01_HRA2 Service Manual.doc
10.2 Overview
The left column (1) always shows values related to the selection in column (2). The values are updated
approximately every 5 seconds; during update the touch screen does not react to input. Column (3) provides special
functions likewise dependent on the selection in column (2).
Values in colum (1) on the “Overwiew” service screen are (from top to bottom):
a. Camera head board temperature
b. Blue laser base plate temperature
c. Blue laser diode temperature
d. Blue laser controller board temperature
e. Power supply board temperature
f. Measured +5 V supply voltage (measured inside power supply unit)
g. Measured +12 V supply voltage (measured inside power supply unit)
h. Measured -12 V supply voltage (measured inside power supply unit)
i. Detected eye side: OD (right eye) or OS (left eye)
j. ID of the currently selected laser mode
k. ID of the currently selected scan angle
This screen shows the serial numbers and firmware/software versions of all connected HECOM components.
Additionally the EEPROM content versions and PLD firmware versions of camera and power supply and the
camera flash content version are displayed.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page53 of 59 TC01_HRA2 Service Manual.doc
The left column shows the touch screen calibration values. These values should be identical to the ones in hra2.ini
once the acquisition module is started.
10.4 Camera
Values in column (1) on the “Camera” service screen are (from top to bottom):
• Camera head board temperature
• Measured +5 V supply voltage (measured inside camera head)
• Measured +12 V supply voltage (measured inside camera head)
• Measured -12 V supply voltage (measured inside camera head)
• Laser monitor (ON) voltage
• Laser monitor (ON) safety threshold
• Y scanner amplitude voltage
• Y scanner amplitude safety threshold
• X scanner amplitude voltage
• X scanner amplitude safety threshold
Displayed values in column (1) on the “PowSup” service screen are (from top to bottom):
a. Blue laser base plate temperature
b. Blue laser diode temperature
c. Blue laser controller board temperature
d. Power supply board temperature
e. Measured +5 V supply voltage (measured inside power supply unit)
f. Measured +12 V supply voltage (measured inside power supply unit)
g. Measured -12 V supply voltage (measured inside power supply unit)
HRA 2 Service Manual
Art.Nr. 96 040-002 Page56 of 59 TC01_HRA2 Service Manual.doc
11 Safety Interlocks
General errors/warnings are reported via HECOM and shown in the PC status bar. There are 4 severity levels:
• Information
• Warning A system parameter is near its limits
• Error An irregular system state is detected
--> System may go to freeze mode
• Severe Error A safety relevant error is detected
--> Safety interlock
Safety relevant errors are additionally reported via the interlock lines. When a security interlock is reached, the
system cannot be used any more before it is switched off and on again.
A Safety Interlock has the following consequences:
Camera
• The current laser mode can’t be changed any more. In particular it is no longer possible to switch to freeze
mode
• The internal fixation target can’t be changed any more
• The fixation targets in all four corners are blinking
• All lasers are switched off
Note: If an Interlock occurs during a live mode, the Scanners will not be switched off (the system state is safer
with the scanners moving)!
Power supply
• The blue laser supply voltage will be switched off (The “L” on the LCD display changes to “-”)
• One of three interlock types will be displayed:
o LASER INTERLOCK Blue laser power was too high either when the laser was modulated or
modulated off
o SCANNER INTERLOCK One of the scanners does not work properly
o MCU INTERLOCK One micro controller has detected an irregular internal state. This type
is overruled by laser or scanner interlock.
• The last internal (left side) and external interlock number will be displayed
Touch Panel
• The freeze screen is shown. A red on/off button and a textual warning indicate the interlock state. The
panel is locked.
PC Acquisition Module
• A list of interlock numbers is shown. Note that consecutive faults may be included.
HRA 2 Service Manual
Art.Nr. 96 040-002 Page57 of 59 TC01_HRA2 Service Manual.doc
5 MCU Camera Measured current of IR laser was too high CAM 1.0.0.0
6 MCU Camera PLD error bits were not set after FREEZE mode was entered CAM 1.0.3.11
7 MCU Camera Laser mode in PLD differed from lasermode in MCU CAM 1.1.0.0
20 MCU Power Supply After switching on the scanner supply voltage the
scanner interlock line does not get inactive (after a time)
21 MCU Power Supply Serial communication with Sapphire laser not working
22 MCU Power Supply Sapphire laser reports that it is still switched on although it was switched off
23 MCU Power Supply +12 V supply voltage missing
24 MCU Power Supply -12 V supply voltage missing
25 SCANNER Power Supply Negative edge on scanner interlock line detected
26 SCANNER Power Supply Scanner interlock line is active (low)
(if not error 25)
27 LASER Power Supply Negative edge on laser interlock line detected
28 LASER Power Supply Laser interlock line is active (low)
(if not error 27)
29 MCU Power Supply Blue laser does not react to active scanner interlock (i.e. service jumper on laser controller is set). Note: PWS 1.0.3.14
This Error is detected on startup but reported on the first attempt to change to live.
30 MCU Power Supply Blue laser does not react to active MCU interlock (i.e. service jumper on laser controller is set). Note: PWS 1.0.3.14
This Error is detected on startup but reported on the first attempt to change to live.
31 MCU Power Supply Blue Laser is in Error state unexpectedly. Note: This Error may be detected on startup but reported on PWS 1.0.3.14
the first attempt to change to live.
65 LASER Camera Too much light on monitor diode when laser is modulated off
66 LASER Camera Too much light on monitor diode when laser is modulated on
67 LASER Camera Error 65 + 66
68 MCU Camera PLL is not locked in
69 LASER Camera Error 68 + 65
70 LASER Camera Error 68 + 66
71 LASER Camera Error 68 + 65 + 66
72 SCANNER Camera Amplitude of y-scanners ist too low
73 LASER + SCANNER Camera Error 72 + 65
74 LASER + SCANNER Camera Error 72 + 66
75 LASER + SCANNER Camera Error 72 + 65 + 66
76 SCANNER Camera Error 72 + 68
77 LASER + SCANNER Camera Error 76 + 65
78 LASER + SCANNER Camera Error 76 + 66
79 LASER + SCANNER Camera Error 76 + 65 + 66
80 SCANNER Camera Amplitude of x-scanners ist too low
81 LASER + SCANNER Camera Error 80 + 65
82 LASER + SCANNER Camera Error 80 + 66
83 LASER + SCANNER Camera Error 80 + 65 + 66
84 SCANNER Camera Error 80 + 68
85 LASER + SCANNER Camera Error 84 + 65
86 LASER + SCANNER Camera Error 84 + 66
87 LASER + SCANNER Camera Error 84 + 65 + 66
88 SCANNER Camera Error 80 + 72
89 LASER + SCANNER Camera Error 88 + 65
90 LASER + SCANNER Camera Error 88 + 66
91 LASER + SCANNER Camera Error 88 + 65 + 66
92 SCANNER Camera Error 80 + 72 + 68
93 LASER + SCANNER Camera Error 92 + 65
94 LASER + SCANNER Camera Error 92 + 66
95 LASER + SCANNER Camera Error 92 + 65 + 66
96 MCU Both Connection to critical system component lost CAM 1.0.3.0,
PWS 1.0.4.0
HRA 2 Service Manual
Art.Nr. 96 040-002 Page58 of 59 TC01_HRA2 Service Manual.doc
ERROR_LIVE_FREEZE
Broadcast
CMD_INTERLOCK
Broadcast
y-scanner amplitude Monitor signal too high Laser mode
below threshold when modulation is on •SYS_STATE_INTERLOCK •SYS_STATE_INTERLOCK
inconsistent •Change to live mode disabled
•Laser mode change disabled
Scanner interlock Monitor diode interlock
line active (low) line active (low)
&
CMD_INTERLOCK
Broadcast
J3
Sapphire Interlock State * Overruled by scanner or laser interlock
HRA 2 Service Manual
Art.Nr. 96 040-002 Page59 of 59 TC01_HRA2 Service Manual.doc
12 Appendix
PC Camera
Eye Explorer Micro controller program Gate Array (PLD)
- Firmware version
EEPROM
HRA2
HECOM -Serial / #
Acquisition Module -PLD firmware version / #
Dinterbus
HECOM
HECOM