You are on page 1of 40

3/20/2014

Modern Rules
for an Old Practice:
Sensor Placement for GxP Temperature Mapping

Paul Daniel
Senior Regulatory Compliance Expert

Page 2 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

1
3/20/2014

Recent History

 Health Canada
– GUI-0069 – Guidelines for Temperature Control of
Drug Products during Storage and Transportation

 ISPE
– Good Practice Guide: Cold Chain Management

Page 3 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Recent History

 PDA
– Technical Report 58 – Risk Management for
Temperature Controlled Distribution

 CDSCO (India)
– Guidelines on Good Distribution Practice for Biological
Products

Page 4 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

2
3/20/2014

Recent History

 CFDA (China)
– Good Supply Practices for Pharmaceutical Products

 EMA (Europe)
– (2013/C 68/01) Good Distribution Practice of Medicinal
Products for Human Use

 USP 36
– Chapter <1079> Good Storage and Distribution
Practices for Drug Products

Page 5 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Goals
 Review Mapping Regulations.

 Discuss 5 Rules for Sensor Placement.

 Review Current Guidance on Sensor Placement.

Page 6 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

3
3/20/2014

Regulations and Mapping

 The regulations don’t tell us how to map, or even to map.

 Regulatory Summary
1. Determine if environmental parameters affect product quality.
2. Do stability testing to determine appropriate storage specifications.
3. Ensure that storage spaces are controlled to meet specifications.

 Mapping and monitoring is industry GMP response to #3.

 Therefore, regulations aren’t very helpful for sensor placement!

Page 7 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Resources for Sensor Placement

 Tradition and Precedence


 What has been done at your facility before?

 Equipment
 What equipment do you already have?

 Science
 What makes sense based on scientific principles?
 Guidance

Page 8 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

4
3/20/2014

5 Rules
Map the Extremes.

Map in 3 Dimensions.

For Large Spaces, Map Storage Only.

Identify and Address Variables.

If it’s worth mapping, its worth monitoring.

Page 9 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Rule 1: Map the Extremes.


 Geometric
 Place sensors at each end of the storage area.

 Conditions
 Apply validation principle of worst cases.

Page 10 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

5
3/20/2014

Parts of a Cube

Corner – 3 planes

Edge – 2 planes

Side – 1 plane

Space – 0 planes

Page 11 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Parts of a Cube

Corner – 3 planes

Rule 1: Map the Extremes

Space – 0 planes

Page 12 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

6
3/20/2014

Corollary 1A: If ≤ 2m3, use 9 + 1.


TOP

SIDE

FRONT

Page 13 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

9 spots = worst case?

 Air Flow:
 Corners have least air circulation.
 Center has most air circulation.

 Heat Exchange:
 Corners have the most exposure (3 planes)
 Center has the least exposure (0 planes)

Page 14 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

7
3/20/2014

How many? Where? < 20m3

TOP

FRONT SIDE

Page 15 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

How many? Where? < 20m3


TOP

SIDE

FRONT

Page 16 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

8
3/20/2014

How many? Where? < 20m3


TOP

SIDE

FRONT

Page 17 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

How many? Where? < 20m3


TOP

SIDE

FRONT

Page 18 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

9
3/20/2014

TOP
How many? Where? < 20m3

SIDE

FRONT

Page 19 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Corollary 1B: If ≤ 20m3, use 15 +1.

TOP

SIDE

FRONT

Page 20 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

10
3/20/2014

Guidance describes 9 +1 and 15 +1

 ISPE Good Practice Guide: Cold Chain Management (2011)

Page 21 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Rule 2: Map in 3 Dimensions.

3 planes.

Left to right.

Page 22 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

11
3/20/2014

Rule 2: Map in 3 Dimensions.

3 planes.

Top to bottom.

Page 23 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Rule 2: Map in 3 Dimensions.

3 planes.

Front to back.

Page 24 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

12
3/20/2014

Corollary 2A: If ≥ 20m3, use Stacks of 3.

High

Middle

Low

Page 25 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 26 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

13
3/20/2014

Stacks of 3, in 3 dimensions.

Page 27 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 28 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

14
3/20/2014

Stacks of 3, in 3 dimensions.

Page 29 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 30 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

15
3/20/2014

Corollary 2A: If ≥ 20m3, use Stacks of 3.

High

Middle

Low

Page 31 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Corollary 2B: Remove sensors if possible.

 Remove “extra” sensors.


 15 +1 diagram.
 With “Stacks of 3, in 3-D”,
it would have 27 sensors.

Page 32 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

16
3/20/2014

Stacks of 3, in 3 dimensions.

Page 33 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 34 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

17
3/20/2014

More published support....

“Completing a three-dimensional temperature profile should


be achieved by measuring points at not less than three
dimensional planes in each direction/axis—top-to-bottom,
left-to-right, front-to-back…”

USP 36 Chapter <1079> Good Storage and Distribution


Practices for Drug Products (2013)

Page 35 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Rule 3: If ≥ 20m3, map storage only.


 As the space gets larger, it is not necessary to map hallways
and access areas.

 Map only the shelves and areas


where product is actually stored.

 Requires controls to prevent


storage in wrong areas.
 SOPs
 Training
 Signs

Page 36 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

18
3/20/2014

More published support....

“Completing a three-dimensional temperature profile should


be achieved by measuring points at not less than three
dimensional planes in each direction/axis—top-to-bottom,
left-to-right, front-to-back, where product will be present.”

USP 36 Chapter <1079> Good Storage and Distribution


Practices for Drug Products (2013)

Page 37 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Rule Summary (so far…)


 Rule 1 – Map the Extremes
 Corollary 1A – If ≤ 2m3, use 9+1.
 Corollary 1B – If ≤ 20m3, use 15+1

Page 38 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

19
3/20/2014

Rule Summary (so far…)


 Rule 1 – Map the Extremes
 Corollary 1A – If ≤ 2m3, use 9+1.
 Corollary 1B – If ≤ 20m3, use 15+1

 Rule 2 - Map in 3 Dimensions


 Corollary 2A – If ≤ 20m3, use Stacks of 3
 Corollary 2B – Remove Sensors if Possible

Page 39 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Rule Summary (so far…)


 Rule 1 – Map the Extremes
 Corollary 1A – If ≤ 2m3, use 9+1.
 Corollary 1B – If ≤ 20m3, use 15+1

 Rule 2 - Map in 3 Dimensions


 Corollary 2A – If ≤ 20m3, use Stacks of 3
 Corollary 2B – Remove Sensors if Possible

 Rule 3 - If ≥ 20m3, map storage only.

Page 40 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

20
3/20/2014

It’s just guidance. It’s a place to start.

 “…additional points may be needed depending on airflow


sources/characteristics, shelving (storage locations), external
temperature sources, and previous experience with similar
units and their thermal behavior.”
– ISPE Good Practice Guide: Cold Chain Management (2011)

Page 41 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

More from USP 36 Chapter <1079>...


 The following factors should be considered during
temperature mapping of storage locations:

 Size of the space  Outside temperature variability


 HVAC Locations  Weekly workflow variation
 Sun-facing walls  Equipment movement
 Low ceilings or roofs  Storage patterns of product
 Geography  Equipment capabilities
 Airflow  SOPs.

Page 42 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

21
3/20/2014

Rule 4: Identify Variables


Volume HVAC Vents and Returns
Temperature ∆ Air Circulation
Height Control Sensors
Exterior Walls Energy Sources
Construction Materials Racks and Shelving
Doors and Windows Traffic Patterns
Lighting Human Factors
Gradients

Page 43 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Volume

Page 44 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

22
3/20/2014

Temperature ∆

Page 45 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Height

Page 46 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

23
3/20/2014

Exterior Walls

Page 47 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Construction Materials

Page 48 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

24
3/20/2014

Doors and Windows

Page 49 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Lighting

Page 50 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

25
3/20/2014

Gradients

Page 51 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

HVAC Vents and Returns

Page 52 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

26
3/20/2014

Air Circulation

Page 53 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Control Sensors

Page 54 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

27
3/20/2014

Energy Sources

Page 55 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Racks, Shelves, and Product

Page 56 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

28
3/20/2014

Traffic Patterns

Page 57 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Human Factors

Page 58 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

29
3/20/2014

Rule 4: Identify Variables

 Place sensors near


representative variables!

 Adjust ideal “grid” to intersect


variables.

Page 59 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

SIDE

TOP

FRONT (South Wall)

Page 60 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

30
3/20/2014

SIDE

TOP

FRONT (South Wall)

Page 61 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

SIDE

TOP

FRONT (South Wall)

Page 62 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

31
3/20/2014

SIDE

TOP

FRONT (South Wall)

Page 63 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

SIDE

TOP

FRONT (South Wall)

Page 64 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

32
3/20/2014

Distance Between Points

Temperature

Distance (X)

Page 65 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Distance Between Points


Temperature

Distance (X)

Page 66 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

33
3/20/2014

Distance Between Points

Temperature

Distance (1/2X)

Page 67 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Distance Between Points


Temperature

Distance (1/2X)

Page 68 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

34
3/20/2014

Distance Between Points

Temperature

Distance (1/4X)

Page 69 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Distance Between Points


Temperature

Distance (1/8X)

Page 70 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

35
3/20/2014

Distance Between Points

Temperature

Distance (1/16X)

Page 71 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

SIDE

TOP

FRONT (North Wall)

Page 72 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

36
3/20/2014

Rule 5: If it is worth mapping…


It is worth monitoring.

 Find the hot and cold spots and monitor them.

 Select the right monitoring solution.

 Qualify the monitoring system.

Page 73 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Monitoring....
“Temperature monitoring… documents the temperatures
during the qualification and operation. It allows conclusions on
whether the product is stored and shipped within the approved
temperature range. Therefore, an appropriate temperature
monitoring strategy should be selected and qualified.”

PDA Technical Report No. 58: Risk Management for


Temperature Controlled Distribution (2012)

Page 74 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

37
3/20/2014

Rule Summary
 Rule 1 – Map the Extremes
 Corollary 1A – If ≤ 2m3, use 9+1.
 Corollary 1B – If ≤ 20m3, use 15+1

 Rule 2 - Map in 3 Dimensions


 Corollary 2A – If ≤ 20m3, use Stacks of 3
 Corollary 2B – Remove Sensors if Possible

 Rule 3 - If ≥ 20m3, map storage only.

 Rule 4 – Identify Variables

 Rule 5 – If it is worth Mapping, it is worth Monitoring

Page 75 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

What about product temperature?

 Its always best to measure air temperature.


 Worst case scenario provides best challenge.

“It is normal practice to use air temperature as the reference


source for the stored [product]…” and is ‘the worst
case representation of the stored product
temperatures.”
-ISPE Good Practice Guide:
Cold Chain Management (2011)

Page 76 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

38
3/20/2014

What about mapping humidity?


 The same concepts apply, but concerned with sources of moisture.

 Use the same number of sensors unless you have an expert who
understands humidity theory.

“Where relative humidity is a critical factor, then humidity sensors could be


located in the same locations – or fewer locations used with the impact of the
temperature considered for other locations – as absolute
humidity will be very similar throughout the space.”
-ISPE Good Practice Guide:
Cold Chain Management (2011)

Page 77 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

Summary
 Reviewed Mapping Regulations.
 Explored 5 Rules for Sensor Placement for Mapping Storage Areas.
 Reviewed Mapping Guidance.

Page 78 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

39
3/20/2014

Thanks for Attending!

Paul Daniel
Senior Regulatory Compliance Expert
Vaisala, Inc.

Email: paul.daniel@vaisala.com

Page 79 / Sensor Placement for GxP Temperature Mapping – MSC 2014 / ©Vaisala

40